Leland Legacy Sample Pump Cat. No. 100 3002 Spec Sheet Publication 1620 PDF

User Manual: 1620

Open the PDF directly: View PDF PDF.
Page Count: 2

SKC Inc. 724-941-9701 SKC-West 714-992-2780 SKC Gulf Coast 281-859-8050 SKC South 434-352-7149
www.skcinc.com
High flows — 5 to 15 L/min
Provides the high flows and long run times of a vacuum-style pump in
a compact, portable, battery-operated sampler within specified back
pressure range
Designed for use with impactors, spore traps, 37-mm filters,
low-volume PUF tubes, and other low-pressure applications
Long 24-hour runs on one charge with Li-Ion battery
Suitable for unattended ambient air sampling
Low noise, size, and weight
Ideal for indoor air studies or unattended ambient air sampling
Longer runs + Higher flows = Enhanced sensitivity for
measuring low concentrations
Patented* highly accurate isothermal flow control system
Flexible programming options
Manual three-button programmability using the large built-in keypad
PC programmability with DataTrac Software for Leland Legacy
Create complete running sequences, download sampling history,
and generate exposure reports for ISO 9000 or quality programs
Large easy-to-read LCD
Displays battery status indicator, flow fault,
run time data, and sampling parameters
CalChek automatic calibration feature
Provides hands-free direct communication to a Defender calibrator
Adjustable datalogging interval from 3 seconds to 8 hours
with DataTrac Software
Rugged and convenient case design
Tough rubber overmolding protects the pump and
provides a sure grip
Anti-static thermoplastic material
Inlet port with removable protective cover
Easy-access computer interface and battery charging jack under
protective cover
Performance with Sioutas Impactor verified by EPA-ETV
Featured in deployable sampler systems
CE marked
* U.S. Patent No. 5,892,160
Leland Legacy is not recommended for high back pressure applications such as asbestos
clearance sampling.
Certi cations
performance
verified
Leland Legacy Sample Pump
Flows: 5 to 15 L/min
High Flows, Long Runs Enhance Measurement Sensitivity!
8 inches
3.9 inches Leland Legacy® is the perfect partner for the Sioutas Impactor
Charging port
Secure belt clip
Data port
Protective inlet lter
Protective cover
Bright status LED
Top view
Publication 1620 Rev 1702
SKC Inc. 724-941-9701 SKC-West 714-992-2780 SKC Gulf Coast 281-859-8050 SKC South 434-352-7149
www.skcinc.com
Quick View
Recommended Accessories
Chargers
Battery Charging Adapter
Cat. No. 223-248
CalChek Communication Cable
Cat. No. 210-502
Noise-reducing Case
TSI 4146 Calibration Kit
Cat. No. 740-4146
DataTrac Software for Leland Legacy
Cat. No. 877-92
Low-volume PUF Tube Holder
Cat. No. 224-29P
Filter Holders
Tubing
Flow Rate (ml/min)
5 to 15 (L/min)
Weight in Ounces (grams)
36 (1000)
Compensation Range (inches water)
Up to 12 at 10 L
Built-in Timer/Clock
Clock
Constant Flow
Yes
Programmable
Yes
PC-compatible
Yes
Multi-tube Sampling
N/A
Flow Fault Feature
Yes
RFI/EMI Shielded
Yes
MSHA-approved Models Available
N/A
ATEX Models Available
N/A
CE Marked
Yes
Corrects for Changes in
Atmospheric Pressure
Yes
Corrects for Changes in
Temperature
Yes
Battery Type
Li-Ion (7.4 V, 12 Ah, 88.8 Wh)
Battery Check
Yes
Tubing
Requires 3/8-inch ID tubing
** Leland Legacy pumps contain Li-Ion batteries and may be subject to special shipping regulations.
Pump and Kits Cat. No.
Leland Legacy Pump** with Li-Ion battery pack and screwdriver set, requires charger 100-3002
Starter Kit includes pump, single charger, Tygon tubing (3 feet, 1/4-inch ID), and collar clip
with cable tie 100-240 V 100-3002-S
Single Pump Kit**# includes Leland Legacy pump and single charger, in a Pelican carry case
5-pack Pump Kit**# includes 5 Leland Legacy pumps, Take Charge 5 Multi-charger, and
DataTrac Software, in a deluxe Pelican case
100-240 V
100-240 V
100-3002K
100-3002K5
Accessories
TSI 4146 Calibration Kit, ow measurement from 0.01 to 20 L/min, includes calibrator, soft-side
case, mounting lugs, 1/4-inch ID tubing, battery pack, 6 AA batteries, inlet lter, dampening module,
NIST certi cate, and manual 740-4146
Chargers
Single
Take Charge 5 Multi-charger
100-240 V
100-240 V
223-241
223-441
CalChek Communication Cable, required for CalChek calibration 210-502
Replacement Parts
Replacement Battery Pack,** Li-Ion
Replacing batteries with non-approved battery packs voids any warranty. P75692
Replacement Filter/O-ring Set, 5 lters and 1 O-ring P40021B
Replacement Inlet Filters, pk/50 P40021A
Leland Legacy Sample Pump
Performance Pro le
Flow Rate Accuracy ± 5% of set-point after calibration
Timing Accuracy 1 min/mo at 25 C
Atmospheric Pressure Accuracy ± 3% in Hg
Typical Run Time Battery: 24 hrs at 9 L/min with Sioutas Impactor (approx. 13 in water back pressure)
24 hrs at 5 L/min with low-volume PUF tube
24 hrs at 10 L/min with IMPACT Sampler in DPS System
24 hrs at 8 L/min with 8 L/min Respirable PPI
Using charger with AC power: Extended run times
Charge Time
(varies with battery capacity
and level of discharge)
15 hrs with approved charger
Temperature Range Operating: 32 to 104 F (0 to 40 C)
Storage: -4 to 95 F (-20 to 35 C)
Charging: 32 to 113 F (0 to 45 C)
Altitude The pump can apply correction to volumetric ow during sampling for weather-related or
altitude variations from the atmospheric pressure established at calibration up to at least
7500 feet above and 5000 feet below sea level.
Timer Display Range 1 to 99999 min (69 days); if run time exceeds 69 days, timer display rolls over
Time Display Time of day in hours and minutes (12 or 24-hr clock) with AM and PM indicators
Volume Display Continually updated based on corrected ow rate multiplied by sampling time
When volume exceeds 99,999 liters, the pump will continue to run normally but an O_FLO
Error will appear on the LCD.
Noise Level 62.5 dBA - pump without case
52 dBA - pump housed in noise-reducing case (optional accessory)
Flow Fault After 15 sec, pump goes into Hold, retains accumulated run time and historical data, and
displays fault icon. Auto-restart is attempted every 20 sec up to 10 times (adjustable with
DataTrac Software)
Flow Control An internal isothermal ow sensor measures ow directly and continuously. Sensor readings
are used in a ow monitoring algorithm to maintain calibrated volumetric ow. In addition, built-
in atmospheric temperature and pressure sensors provide readings to correct volumetric ow
for these parameters when they vary from point of calibration.
Weight 36 oz (1 kg)
Measured 3 feet (1 meter) from pump operating at 10 L/min and 12 inches water back pressure
Ordering Information
Leland Legacy
requires 3/8-inch ID tubing.
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which provides SKC’s sole liability and the buyer’s
exclusive remedy. To view the complete SKC Limited Warranty and Return Policy, go to http://www.skcinc.com/warranty.asp.

Navigation menu