754

HART Mode

Users Guide

July 2011 Rev. 1, 8/12
© 2011-2012 Fluke Corporation. All rights reserved.
All product names are trademarks of their respective companies.

LIMITED WARRANTY AND LIMITATION OF LIABILITY
This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. This
warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or
abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke’s behalf.
To obtain service during the warranty period, contact your nearest Fluke authorized service center to obtain return
authorization information, then send the product to that Service Center with a description of the problem.
THIS WARRANTY IS YOUR ONLY REMEDY. NO OTHER WARRANTIES, SUCH AS FITNESS FOR A PARTICULAR
PURPOSE, ARE EXPRESSED OR IMPLIED. FLUKE IS NOT LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR
CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries
do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of
liability may not apply to you.
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
11/99

Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands

Table of Contents

Title

Page

Introduction ....................................................................................................................
How to Contact Fluke .....................................................................................................
HART Channel Selection ...............................................................................................
Connect to a HART Transmitter .....................................................................................
Supported vs. Generic Transmitters ...............................................................................
Communication Operations ............................................................................................
See Process Variables ..............................................................................................
Setup Operations.......................................................................................................
Basic .....................................................................................................................
Sensor ..................................................................................................................
Device Identification..............................................................................................
HART Output ........................................................................................................
HART Information .................................................................................................
Service Operations ....................................................................................................
Abort Softkey .............................................................................................................

i

1
1
2
3
6
7
9
9
9
10
10
11
11
11
12

754
Users Guide
Analog Mode and HART Mode Interaction ....................................................................
HART Calibration ...........................................................................................................
HART Mode Menus for Adjustment ...............................................................................
Calibrate a Supported HART Transmitter ......................................................................
Loop Test ..................................................................................................................
Output Trim ...............................................................................................................
Sensor Trim ..............................................................................................................
Duplicate a Transmitters Basic Information ...................................................................

ii

12
12
12
14
15
16
16
18

List of Tables

Table
1.

Title

Page

Supported vs. Non-Supported Instruments ........................................................................... 6

iii

754
Users Guide

iv

List of Figures

Figure
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.

Title

Page

HART Channel Selection.......................................................................................................
Active Device Screen ............................................................................................................
HART Transmitter Connection ..............................................................................................
Supported Device Screen......................................................................................................
754 HART Mode Menu Tree .................................................................................................
Device Variables Screen .......................................................................................................
Basic Setup Screen ...............................................................................................................
Sensor Setup Screen ............................................................................................................
Device Identification Screen ..................................................................................................
HART Output Screen ............................................................................................................
HART Information Screen .....................................................................................................
Block Diagram of a HART Transmitter ..................................................................................
Calibrator Mode Select Screen..............................................................................................
Loop Test Screen ..................................................................................................................
Output Trim Screen ...............................................................................................................

v

2
4
5
7
8
9
9
10
10
11
11
13
14
15
16

754
Users Guide
16.
17.
18.

Sensor Trim Screen .............................................................................................................. 16
Sensor Trim Zero Screen...................................................................................................... 17
Basic Screen......................................................................................................................... 18

vi

How to Contact Fluke

Introduction
 Warning
To prevent possible electrical shock, fire, or
personal injury, read the Safety Information
in the 753/754 Users Manual before you use
the 754 Documenting Process Calibrator.
Adjustment is necessary for analog transmitters during
calibration. With HART (Highway-Addressable Remote
Transducer) transmitters, adjustments are by remote
command. For these adjustments, use of a
communication tool and a Calibrator is necessary. The
754 (Product or Calibrator) gives communication and
calibration functions in one tool.
This guide tells you how to use the HART communication
mode. HART mode is a procedure for the Product to
communicate over its serial HART interface to a HART
transmitter. See the 753/754 Users Manual for safety
information, analog mode use instructions, specifications,
and more general data. All of the Product functions shown
in the 753/754 Users Manual are available and can be
used with supported HART transmitters. The only
transmitters addressed in this manual are HART
transmitters.

To contact Fluke, call one of the following telephone
numbers:
•

Technical Support USA: 1-800-44-FLUKE
(1-800-443-5853)

•

Calibration/Repair USA: 1-888-99-FLUKE
(1-888-993-5853)

•

Canada: 1-800-36-FLUKE (1-800-363-5853)

•

Europe: +31 402-675-200

•

Japan: +81-3-6714-3114

•

Singapore: +65-6799-5566

•

Anywhere in the world: +1-425-446-5500

Or, visit Fluke's website at www.fluke.com.
To register your product, visit http://register.fluke.com.
To view, print, or download the latest manual supplement,
visit http://us.fluke.com/usen/support/manuals.

Note
Some specific HART transmitter commands are
only available with supported devices.

1

754
Users Guide

HART Channel Selection
For HART communication, use the HART connector on
the side of the Product or the mA jack on the front of the
Product. The procedure you use depends on the
application and preference.
Use Setup mode to choose the correct HART channel:
1.

Push s. The first setup screen shows. See
Figure 1.

2.

Push  or  to select HART Channel.

3.

Push .

4.

Push  or  to select HART Port or mA Jack.

5.

Push .

gnx19s.bmp

The selection is kept in Product memory. The Product
uses this selection until it is changed in setup mode.
Depending on the selection, HART Port or mA Jack
shows at the top of the display when you push r.
Note
The figures in this guide show the mA Jack in
use for HART communication.

2

Figure 1. HART Channel Selection

HART Mode
Connect to a HART Transmitter

Connect to a HART Transmitter

•

To connect and start communication with a HART
transmitter:
Note
If only a communication connection to a
transmitter that is powered in a loop is
necessary, attach the HART alligator clips to the
loop power terminals on the transmitter and push
r.
1.

Connect the Product mA jacks to the loop power
terminals of the transmitter. See Figure 3.

2.

Connect the HART interface cable into the HART
jack, and then connect the alligator clips to the
terminals in step 1. There is no incorrect polarity. If
the HART signal is directed through the mA jack in
setup mode, this connection is not required.
Push r.
If the transmitter is not powered by a loop power
supply, push the Loop Power softkey to start loop
power.

3.
4.

•
•

Notes
The Product supplies loop power
through an internal series resistance of
250 Ω.
If the Product shows a measurement of
0 mA, make sure you have not
reversed the current leads.
If an external loop power supply is
used, a resistance of between 230 Ω
and 270 Ω connected in series with the
external loop supply and the transmitter
is necessary.

5.

The Product tries to connect to Poll Address 0 (single
transmitter per loop). If no connection is made, push
the Poll softkey to examine Poll Addresses 1 through
15 (multidrop).

6.

When the Product establishes communication with
the transmitter, the Active Device screen shows. See
Figure 2. In a multidrop configuration, you must
choose a transmitter from a list, and push .

3

754
Users Guide
The Active Device screen supplies this data for all
transmitters:
•
•
•
•
•
•
•

gnx06s.bmp

Figure 2. Active Device Screen

4

Poll address (if not 0)
Model number and Tag ID
PV (Primary Variable)
PVAO (digital representation of the Analog Output)
PV LRV (PV Lower Range Value)
PV URV (PV Upper Range Value)
Softkeys for accessing HART operation menus

HART Mode
Connect to a HART Transmitter

754

DOCUMENTI

NG PROCE

SS CALIBR

ATOR

+

–

gks61.eps

Figure 3. HART Transmitter Connection

5

754
Users Guide

Supported vs. Generic Transmitters
The Product communicates with most HART transmitters. “Supported Transmitters” are transmitters in which the Product is
programmed to use their device-dependent commands and software version. “Generic transmitters” are transmitters that are
not programmed in the Product. Table 1 shows the operations that are available for supported vs. generic transmitters in
single-point and multidrop configurations.
Table 1. Supported vs. Generic Transmitters (Devices)
Supported Transmitter
Menu
Top Level

Active Device screen
Basic (read/write, cloning capability)
Sensor

Setup

(Temperature) Sensor (read/write)
Device Identification (read/write)
HART Output (read/write)
HART Information (read only)
Loop Test

Service

Pressure Zero Trim
Output Trim
Sensor Trim

Process

6

Generic Transmitter

Operation

Detailed process information

Single Point

Multidrop

Single Point

Multidrop

•
•
•

•
•
•

•
•

•
•

•
•
•
•
•
•
•
•
•

•
•
•
•
Not Available

•
Not Available

•
•

Read only

•
•
•
•
•
•

Read only

•
•
•
Not Available

•
Not Available

Not Available

Not Available

•

•

HART Mode
Communication Operations

1.

2.

Note
Sensor trim is supplied for supported
transmitters, with some exceptions, as identified
in the list of supported transmitters that you can
see on the display.
To show a list of supported transmitters and software
versions:
•

If the Product is connected to a HART
transmitter, push the Abort softkey, and then
More Choices. Continue to Step 2.

•

If the Product is not connected to a HART
transmitter, push r and then More Choices.
Continue to Step 2.

Push the Device Revs softkey. The screen in Figure
4 is shown.

gnx07s.bmp

Figure 4. Supported Device Screen
3.

Push  or  to highlight the applicable
manufacturer, and push . A list of model
numbers is shown.

4.

Push  or  to highlight the model number, and
push . A list of software versions is shown.

Communication Operations
Figure 5 shows the HART Mode menu tree. Menu
selections are determined by the type of transmitter that
you use and the configuration connection type (current
loop or multidrop).

7

754
Users Guide

PROCESS

· View process variables
· View variable map
· Re-map process

Basic

variables

· Tag
· PV units
· LRV, URV
· Damping
· Transfer function

(Dual sensor
temperature devices)

Sensor

· Keypad input
· Adjust URV, LRV to applied
values

Sensor 1

· Sensor serial number
· Sensor lower and upper limits
· Sensor minimum span
Temperature devices only:

Device
Identification

SETUP

· Software Version
· Final assembly number
· Tag
· Message
· Date
· Description

HART
Output

SERVICE
Output trim
Sensor trim

HART
Information

Sensor 2

· Change Sensor Type
· Change Sensor Connections
(Dual sensor temperature
devices)

· Dual Sensor Config

Loop test
Pressure zero trim

· Change Sensor Type
· Change Sensor Connections

· Manufacturer
· Model
· Device HART ID
· Software revision
· Hardware revision
· Number of preambles

(Dual sensor temperature
devices)
· Write protect
· Alarm state
· HART poll address
· HART burst mode
· HART burst command

ABORT

gnx03f.eps

Figure 5. 754 HART Mode Menu Tree

8

HART Mode
Communication Operations
See Process Variables

Setup Operations

From the Active Device screen:

The Setup softkey gives access to these five setup
functions:

1.

Push the Process softkey to see more device
variables and their continuously updated values.

2.

Select View Process Variables.

3.

Push .

To see more data, push the Next Page softkey. See
Figure 6.

•
•
•
•
•

Basic
Sensor
Device Identification
HART Output
HART Information

Basic
Use the Basic setup screen to replicate a transmitter as
described at the end of this manual.
From the Active Device screen, push the Setup and
Basic softkeys. See Figure 7.

gnx08s.bmp

Figure 6. Device Variables Screen

gnx09s.bmp

Figure 7. Basic Setup Screen

9

754
Users Guide
Sensor

Device Identification

This is where you can see data about the sensor in the
transmitter, that includes serial number, limits, and span.
The limits shown are the absolute limits for the sensor.
(The Upper Range Value (URV) and Lower Range Value
(LRV) are different, and are viewable and programmable
through the Basic Setup screen.)

You can program the Tag, Message, Date, and Descriptor
registers in the transmitter when you use this screen.
From the Active Device screen, push the Setup and
Device Identification softkeys to see data about the
transmitter. See Figure 9.

From the Active Device screen, push the Setup and
Sensor softkeys to access the Sensor Setup screen. See
Figure 8.
You can also program the Sensor Type and Sensor
Connection registers in a supported temperature
transmitter using this screen. Screen examples do not
show Temperature device menu items.

gnx11s.bmp

Figure 9. Device Identification Screen

gnx10s.bmp

Figure 8. Sensor Setup Screen

10

HART Mode
Communication Operations
HART Output
From the HART Output screen you can change the Poll
Address (0 = single transmitter, any other address =
multidrop), and control burst mode.

From the Active Device screen push the Setup and
HART Information softkeys to access the read-only
HART Information screen. See Figure 11.

From the Active Device screen, push the Setup and
HART Output softkeys to access the read/write HART
Output screen. See Figure 10.

gnx13s.bmp

Figure 11. HART Information Screen

Service Operations
gnx12s.bmp

Figure 10. HART Output Screen

HART Information
This screen shows more complete data about the
transmitter model, hardware and software revision
numbers, and how many preambles it sends.

The Service softkey gives you access to Loop Test,
Pressure Zero Trim (where applicable), Output Trim, and
Sensor Trim operations. For generic transmitters, only
Loop Test, Output Trim, and Pressure Zero Trim are
available (see Table 1.) The trim (adjustment) operations
are later in this manual.
Note
Loop Test and Output Trim are not available if
the transmitter is in multidrop mode.

11

754
Users Guide
Abort Softkey

HART Calibration

The Abort softkey stops the communication operation in
process, and gives control to the previous screen. From
the Active Device screen, Abort calls up the browser, in
which you can see the list of connected transmitters.

An analog transmitter has one step of electrical
conversion from a measured physical parameter to
a 4-20 mA current loop output. A HART transmitter
has the three blocks. See Figure 12.

Analog Mode and HART Mode
Interaction

It can be necessary to examine and adjust the Input
block, the Output block, or both. For example, if your
application requires the Primary Variable (PV) to be
correct when read by a host computer, you must calibrate
the Input stage.

Analog mode is usual Product operation, as given in the
753/754 Users Manual. HART Communication mode
starts when r is pushed.
Push r to change between HART and analog modes, or
push  to change to analog mode from HART mode.
This lets the transmitter automatically set up analog mode
for applicable measure and source functions, if
necessary.
For supported transmitters, the change to analog mode
goes to the MEASURE/SOURCE screen. This makes it
easy to continue with an “as found” calibration.

If your application requires that the 4-20 mA current
output value accurately show what the Input block is
measuring, you must calibrate the Input and Output block.
Transmitters in multidrop systems, in which more than
one is wired in parallel, do not use their Output blocks.
Their analog outputs are all held at an idling level of 4 mA
no matter what the Input block measures.

HART Mode Menus for Adjustment

For generic transmitters, the change to analog mode
makes you choose the MEASURE or SOURCE screen,
from which you select the applicable function.

Sensor Trim in HART mode is when you adjust the Input
block. Output Trim is when you adjust the Output block.
These adjustments are made from the Service menu.

For supported or generic transmitters, when you push r
to return to HART mode, the Active Device screen is
shown. The HART serial communication connection stays
active as you change between HART and analog modes.

For pressure transmitters, Pressure Zero Trim is an
additional adjustment. This adjustment is the same as
when you set the lower sensor point at zero. All three
operations are energized from the HART mode Service
menu.

12

HART Mode
HART Mode Menus for Adjustment
754
754

Sensor trim
Pressure zero trim

Physical
variable

Input

DOCUMENTING PROCESS CALIBRATOR

PV

PVAO

Computation

Loop test
Output trim

Output

4-20 mA
loop

HART transmitter
gnx02f.eps

Figure 12. Block Diagram of a HART Transmitter

13

754
Users Guide

Calibrate a Supported HART Transmitter
An “as found” and “as left” transmitter calibration has an
easier, more automated procedure for HART transmitters
than for analog transmitters. The calibration procedure is
the same as given in the 753/754 Users Manual. How to
set up the calibration template, and how you adjust the
transmitter are different.
Note
If you start a calibration procedure from an
installed task, do not push Task until the Product
is connected to the HART transmitter and
communication in analog mode is established.
The subsequent procedure assumes that you know how
to use a 754 Series Documenting Process Calibrator to
calibrate analog transmitters, and are not running an
installed Task.
1.

Make the applicable measure, source, and HART
interface connections between the Product and the
transmitter.

2.

Push r to establish communication.

3.

Push r to change to analog mode.

4.

You are shown a set of selections for analog mode.
Use  and  to select one of the measure/source
choices. This is where you select measure mA
(analog output), or measure PV, which does not
involve the Output block of the transmitter. See
Figure 13.

14

gnx14s.bmp

Figure 13. Calibrator Mode Select Screen
5.

Push .

6.

Push the As Found softkey.

7.

Push  or  followed by  to select an
instrument calibration procedure. You can see that
the calibration template is installed with applicable
data. You can make changes to items if necessary.
The Error% value must be entered before you
continue.

8.

Continue with calibration as given in the 753/754
Users Manual. When calibration is complete, do
step 9.

9.

Push the Adjust softkey to go back to HART mode
and launch the Service menu for the transmitter.

HART Mode
Calibrate a Supported HART Transmitter
10. Do an Output Trim, and/or a Sensor Trim. These are
Service operations. See “Output Trim” and “Service
Trim”. When you are done with the trim procedures,
push the Done softkey.

To do a Loop Test:
1.

From the Active Device screen, push the Service
softkey, followed by Loop Test. See Figure 14.

11. Push the As Left softkey and do the calibration
procedure again to make sure that the transmitter
passes.

Loop Test
The Loop Test feature sends a command to the
transmitter to set its output block to a specified value. Use
this to examine the calibration of the output block, or to
verify a correct indication on an external loop reading
device.

gnx15s.bmp

Figure 14. Loop Test Screen
2.

Push a softkey to command the transmitter to set its
analog output to the related value, or manually record
a value with the numeric buttons.

The transmitter acknowledges the selected PVAO (digital
representation of the Output stage) in the middle of the
screen. The Product shows the measured value at the top
of the screen so you can use Loop Test as a fast
procedure to check the calibration of the transmitter’s
Output block.

15

754
Users Guide
Output Trim

Sensor Trim

An Output Trim adjusts the transmitter’s Output block.
You can do an Output Trim on generic and supported
transmitters.

You can only do a Sensor Trim on a supported
transmitter. A Sensor Trim adjusts the Input block of a
HART transmitter. If your application does not use the
Output block of the transmitter, Sensor Trim is the only
adjustment necessary.

To do an Output Trim:
1.

From the Active Device screen, push the Service
softkey and then Output Trim. See Figure 15.

A Sensor Trim can involve one or more trim points. This
is controlled by the transmitter’s software.
To do a Sensor Trim:
1.

From the Active Device screen, push the Service
softkey, and then Sensor Trim. See Figure 16.

gnx16s.bmp

Figure 15. Output Trim Screen
2.

Push Fetch to put the mA value you measure with
the Product into the dialog box.

3.

Push Send to trim the low analog output point.

4.

To trim the high analog output point, do steps 2 and
3 again. Follow the prompts on the display.

gnx17s.bmp

16

Figure 16. Sensor Trim Screen

HART Mode
Calibrate a Supported HART Transmitter
2.

Push  or  to select the Sensor Trim operation,
and push .

3.

4.

5.

Follow the instructions on the display. For pressure,
there is a message to connect a pressure module to
the Product, and a message to push  to zero the
pressure module.

Record the necessary trim value the Product must
produce, and push . In the case of a pressure
transmitter (as above), push Fetch to put the value
that is measured by the Fluke pressure module into
the dialog box.

6.

Push Trim.

Push Continue. See Figure 17.

7.

If asked, do steps 2 and 3 again to trim the remaining
points. Follow the display prompts.

Note
Select trim points at the LRV or URV values. To
review these values, push the Show LRV/URV
softkey.
If you receive an error that shows excessive correction for
the selected trim point, it is possible to repair the problem
with a series of sensor trims. Start the sensor trims in
small increments from the previous trim point to the
necessary trim point.

gnx 18s.bmp

Figure 17. Sensor Trim Zero Screen

17

754
Users Guide

Duplicate a Transmitters Basic
Information

4.

Push the Store Page softkey to put the settings into
Product memory (not the transmitter).

The Product lets you duplicate the Basic Setup data from
one transmitter to a different transmitter. You can copy
generic and supported transmitters.

5.

Disconnect the Product from the transmitter and
connect it to the transmitter that you will configure.

6.

Push the Abort softkey to start communication.

To use the Product to duplicate a transmitter:

7.

Push the Setup softkey, and then select Basic to go
back to the Basic Setup screen.

8.

Push the Recall Page softkey to refresh the Product
display with the duplicated parameter settings. The
settings are not transmitted to the transmitter at this
time.

9.

Push  or  and  to select parameters
individually, change them, or leave them as
duplicated, and push the Send softkey to transmit
them to the transmitter.

1.

Connect the Product to the transmitter you want to
duplicate. Only a communication connection is
necessary.

2.

From the Active Device screen, push the Setup
softkey.

3.

Push  or  keys to select Basic from the list on
the screen and push . See Figure 18.

gnx09s.bmp

Figure 18. Basic Screen

18



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : Yes
Author                          : 351116
Create Date                     : 2012:07:24 13:01:28+05:30
Modify Date                     : 2012:07:24 08:17:21-07:00
Has XFA                         : No
XMP Toolkit                     : Adobe XMP Core 5.2-c001 63.139439, 2010/09/27-13:37:26
Format                          : application/pdf
Creator                         : 351116
Title                           : Microsoft Word - 0master.doc
Creator Tool                    : PScript5.dll Version 5.2.2
Metadata Date                   : 2012:07:24 08:17:21-07:00
Producer                        : Acrobat Elements 10.0.0 (Windows)
Document ID                     : uuid:7d0cca12-d903-4d97-83ef-b0011edef53e
Instance ID                     : uuid:fe44e10b-3382-46ad-8135-ce1c24a5b1f0
Page Mode                       : UseOutlines
Page Count                      : 26
EXIF Metadata provided by EXIF.tools

Navigation menu