GeoPrecision ML5WR10 Wireless Datalogger Family User Manual log hc2
GeoPrecision GmbH Wireless Datalogger Family log hc2
Contents
- 1. Manual__log_hc2_v2
 - 2. Manual__log_pt1000_v2
 
Manual__log_hc2_v2

LOG-HC2  Wireless  Temperature/Humidity  Data  Logger  for
Rotronic AG Hydroclip HC2
Handling
The logger consists  of  the  electronics (housing model dependent: Stainless  Steel  or  Plastic,
20x160  mm)  and  the  HydroClip2  HC2  sensor  (very  high  precision  sensor,  integrated  or
separate  cable)  from  Rotronic  AG.  The  sensor  head  can  be  replaced,  different  variants  are
available. 
Please note: The HC2 sensor is not included, because
different versions exist.
Important:  The  sensor  is  protected  against  dust  and
can  withstand  condensed  water,  but  should  not  be
exposed to liquid water.
If  the  enclosure  must  be  opened  (e.g.  for  battery
replacement), please check  the  o-ring  (sealing) very carefully!  Use  lubricant  (to  protect  the
o-ring) before reassembling.
We recommend to lubricate the enclosure additionally after reassembling and before using in
the  field  (e.g.  by  Spray  Vaseline).  Please  ensure,  that  no  force  affects  the  connector
sensor<->logger in the field. 
About the range: Optimal range (line of sight in open environment) could be up to 300 meters
for  loggers  with  separate  antennas.  If the  sensor  cable  (if  applicable)  is  used  as  antenna,  a
range  of  100  meters  is  realistic,  for  plastic  housings  with  internal  antenna,  30  meters  are
realistic.  In  the  2nd  case  the  first  20  cm  of  the  sensor  cable  are  mostly  important  for  the
transmission. On request, special beam antennas with very high gain are available from us.
Please note: The devices are available with several different radio frequencies, depending on
the local regulations of the country (e.g.: 433 MHz in Europe, parts of Africa and Asia, 905
MHz in USA and Canada, ...). It is in the responsibility of the user, not to use devices that are
not allowed in the country of usage.
Technical Shortform  LOG-HC2  Wireless Temperature/Humidity Data Logger 15.07.2013
Don't expose
the sensor to
liquid water.

Technical Data
Logger: Temperature range: -40°C to +85°C.  Radio  transmission  frequency:  a)  EC,
parts of Africa and Asia): 433.92 MHz (harmonised frequency for license free
operation  within  the  EC,  Switzerland,  Norway,  Iceland.  (for  other  countries
please  check  the  local  regulations.  b)    USA/Canada:  905  MHz.  Effective
emitted energy < 5mW
Memory: 512-2048kB (non volatile) Flash memory. Up to 100.000-400.000 measures: 1
measure  typically  uses  5.5  Bytes  per  value  (Software  Version  1.1,  will  be
reduced  to  2-3  Bytes  in  a  following  revision).  Each  HK-record  (time  stamp
and optional HK-data) typically requires another 6 to 9 bytes. Since HK-values
are recorded only after a selectable number of measures, (factory default is 6). 
Example:  24  measures,  2  values  per  day  require  24*5.5*2  =    264  Bytes.
Additionally  4  HK-records  require  additional  4  *  6..9  =  36  Bytes.  Sum:  ca.
300 Bytes per day. Conclusion: 2048kB will be good for (theoretically) ca. 20
years without clearing the memory. The memory has a duty cycle of  >100 000
clearing cycles.
HK-data: Internally  the  battery  voltage  and  internal  temperature  can  be  recorded.  The
internal temperature is only calibrated on  demand. Calibrated accuracy is +/-
1°C  in  the  range  -20°C  to  +40°C.  The  battery  voltage  is  measured  with  a
resolution of 1 mV. 
Sensor: Any  type  of  HygroClip2  (replaceable  and  calibrated)  from  ROTRONIC  AG
(www.rotronic.de). The sensor range is 0-100% rH and -50°C to +100°C. For
the  accuracies  and  extended  temperature  ranges contact  ROTRONIC  (please
note: the logger itself is only designed for temperatures from -40°C to +60°C
(+85°C  for  short  times)).  Optionally  the  sensor  can  be  connected  in  4-wire
technology. Sensor cables are possible from 20 cm to 1 meter. 
Battery: SB-AA11 from www.vitzrocell.com :
3.6 Volt Lithium (Li-SoCl2), 2400mAh. 
Max. Pulse load: <20 mA
Size: AA with solder terminal
Low self discharge rate (less than 1% after 1 year of storage at +20°C)
Non flammable
Non restricted for transport
The battery can be replaced by the user. Similar types of battery are available
from all major battery production companies. 
About the battery voltage: The voltage is around 3.4-3.6 V at 20°C, but „drops
down“  to  3.1-3.3°C  at  -40°C.    The  battery  voltage  does  not  significantly
reflect the remaining power of the battery, the „temperature drops“ show this
better, as described below.
Technical Shortform  LOG-HC2  Wireless Temperature/Humidity Data Logger 15.07.2013

Calculation the battery capacity
4 main factors impact the battery capacity/lifetime:
 Constant load: about 10µA for the current version Mode „Sleeping“ with
   periodically checking the radio. So 1 mAh is good for ca. 100 hours of „Sleeping“
 Pulse load: about 10mA for 1.7 seconds per measure, so 1 mAh is good for
    ca. 210 measures
 Self discharge: about 10% after 10 years
 Temperature cycles: difficult to predict, could be up to 50% (worst case) for
   harsh environments like high mountain sites.
As a rough estimation: Theoretically 2400 mAh are good for  >25  years of  „Sleeping“ or >
500000 measures (this is almost 60 years for 1 measure per hour), if no self discharge is
assumed. Practically we normally calculate only  with 1/3 of the capacity (the rest is for
spare):  For 1 measure per hour each year requires in total 130 mAh (42 mAh for the
measures and 88mAh for „Sleeping“), with  the rest of 2400 mAh / 3 = 800 mAh still more
than 6 years of operation should be possible. Hence we would recommend to replace the
battery with these settings  after 5 years or later, if „temperature drops“ for cold phases rises
significantly.
Compliance informations (only for USA/Canada):
a) User Information according to FCC 15.21:
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s
authority to operate the equipment. 
b) Part 15 Statement according FCC 15.19/RSS Gen Issue 3 Sect. 7.1.3
This device complies with Part 15 of the FCC Rules and with Industry Canada license-exempt RSS standard(s).
Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2)
this device must accept any interference received, including interference that may cause undesired operation.
Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de
licence. L'exploitation est autorisée aux deux conditions suivantes : (1) l'appareil ne doit pas produire de
brouillage, et (2) l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique subi, même si le
brouillage est susceptible d'en compromettre le fonctionnement.
LOG-HC2:   FCC-ID: 2AAREHC2R10 / IC Certification No IC: 11303A-HC2R10
(Label assembled over battery screw cap)
Technical Shortform  LOG-HC2  Wireless Temperature/Humidity Data Logger 15.07.2013
Model: LOG-HC2
 FCC-ID: 2AAREHC2R10
IC: 11303A-HC2R10
1cm
2.54cm
. 
Model: LOG-HC2 
FCC-ID: 2AAREML5WR10 
IC: 11303A-ML5WR10
FCC-ID: 2AAREML5WR10 / IC Certification No IC: 11303A-ML5WR10