8520 DustTrak Aerosol Monitor Operation And Service Manual TSI Dust Trak

User Manual: Manual

Open the PDF directly: View PDF PDF.
Page Count: 89

Model 8520
DUSTTRAKTM Aerosol Monitor
TRAKPROTM Data Analysis Software enclosed
Operation and Service Manual
1980198, Revision R
June 2006
Exposure Monitoring
Model 8520
DUSTTRAKTM Aerosol Monitor
Operation and Service Manual
1980198, Revision R
June 2006
SHIP/MAIL TO:
TSI Incorporated
500 Cardigan Road
Shoreview, MN 55126-3996
USA
U.S.
Technical Support:
(800) 874-2811/(651) 490-2811
Fax:
(651) 490-3824
E-mail address:
answers@tsi.com
Website:
http://www.tsi.com
INTERNATIONAL
Technical Support:
(001 651) 490-2811
Fax:
(001 651) 490-3824
Copyright ©
TSI Incorporated / 1994–2006 / All rights reserved.
Address
TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
Fax No.
(651) 490-3824
Limitation of Warranty and Liability (effective July 2000)
Seller warrants the goods sold hereunder, under normal use and service as described in the
operator's manual, shall be free from defects in workmanship and material for twenty-four (24)
months, or the length of time specified in the operator's manual, from the date of shipment to the
customer. This warranty period is inclusive of any statutory warranty. This limited warranty is
subject to the following exclusions:
a. Hot-wire or hot-film sensors used with research anemometers, and certain other components
when indicated in specifications, are warranted for 90 days from the date of shipment.
b. Parts repaired or replaced as a result of repair services are warranted to be free from defects
in workmanship and material, under normal use, for 90 days from the date of shipment.
c. Seller does not provide any warranty on finished goods manufactured by others or on any
fuses, batteries or other consumable materials. Only the original manufacturer's warranty
applies.
d. Unless specifically authorized in a separate writing by Seller, Seller makes no warranty with
respect to, and shall have no liability in connection with, goods which are incorporated into
other products or equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS stated
herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR
PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE.
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR
BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES,
INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED
ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL
BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE
PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE
GOODS. IN NO EVENT SHALL SELLER BE LIABLE FOR ANY SPECIAL,
CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE
RESPONSIBLE FOR INSTALLATION, DISMANTLING OR REINSTALLATION COSTS
OR CHARGES. No Action, regardless of form, may be brought against Seller more than 12
months after a cause of action has accrued. The goods returned under warranty to Seller's
factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND
LIABILITY, which contains the complete and exclusive limited warranty of Seller. This
LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms
waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are to our
customers, our service policy is designed to give prompt attention to any problems. If any
malfunction is discovered, please contact your nearest sales office or representative, or call TSI
at (800) 874-2811 (USA) or (001 651) 490-2811 (International).
i
CONTENTS
SAFETY INFORMATION .................................................................... V
1 UNPACKING AND PARTS IDENTIFICATION ............................ 1
Unpacking the DUSTTRAK Aerosol Monitor.................................... 1
Parts Identification for the DUSTTRAK Aerosol Monitor.................. 2
2 SETTING-UP................................................................................. 5
Supplying Power to the DUSTTRAK Aerosol Monitor ..................... 5
Installing the Batteries............................................................. 5
Using the AC Adapter ............................................................. 5
Instrument Setup ........................................................................... 6
Setting-up TRAKPRO Data Analysis Software ......................... 6
Connecting the DUSTTRAK Aerosol Monitor to the
Computer............................................................................. 6
Set-Up the Communications Port ........................................... 6
Setting the Real-Time Clock................................................... 7
Programming the Date/Time Using TRAKPRO Software ......... 8
Manually Setting the Real-Time Clock ................................... 8
Connecting the Optional Portable Printer...................................... 8
Connecting/Wiring the Analog/Alarm Output Connector............... 9
3 OPERATION ............................................................................... 11
Overview...................................................................................... 11
DUSTTRAK Aerosol Monitor Keypad Functions............................ 11
ON/OFF Key................................................................................ 11
SAMPLE Key............................................................................... 12
TIME CONSTANT Key................................................................ 12
Programming the Available Time Constants............................... 13
STATISTICS Key......................................................................... 13
PRINT Key................................................................................... 14
SAMPLING MODE Key............................................................... 14
Survey Mode......................................................................... 14
LOG Modes........................................................................... 15
CLEAR MEMORY Key................................................................ 15
LOGGING INTERVAL Key.......................................................... 15
Programming the Logging Intervals using TRAKPRO Software ... 16
Ï and Ð Keys ............................................................................. 17
CALIBRATE Key ......................................................................... 17
Display/Keypad Lockout Switch .................................................. 17
Remote Sampling........................................................................ 17
Respirable Mass Sampling with the Cyclone.............................. 18
Using the Cyclone ....................................................................... 18
Using the 1.0 or 2.5 µm Inlet Conditioners.................................. 19
Using the 10 µm Nozzle .............................................................. 20
Analog Output.............................................................................. 21
ii
Alarm Output................................................................................22
Alarm Output Specifications..................................................23
Programming the Alarm Setpoint..........................................24
Upgrading the DUSTTRAK Aerosol Monitor to Add Analog and
Alarm Functions .......................................................................24
Identifying an Instrument with Analog/Alarm Functions........24
Identifying an Instrument that may be Upgraded to Add
Analog/Alarm Functions.....................................................25
Identifying an Instrument That May Not Be Upgraded..........25
Programming Advanced Modes: LOG 2 and LOG 3...................26
Sample Protocol for LOG 2 and LOG 3 Modes...........................29
Things You Should Know About Taking Pre-Programmed
Samples....................................................................................30
Memory Considerations...............................................................31
Custom Calibrations ....................................................................31
Determining the Calibration Factor for a Specific Aerosol .......... 32
Setting Custom Calibration Factor from Keypad.........................33
Setting Custom Calibration Factors using TRAKPRO Software....34
Converting Stored Data to Calibrated Data................................. 35
4 MAINTENANCE..........................................................................37
Maintenance Schedule................................................................37
Zero Checking/Re-Zeroing ..........................................................38
Cleaning the 1.0 µm and 2.5 µm Inlet Conditioners.................... 39
Cleaning the 10 µm Inlet and Sample Tube................................40
Replacing the Internal Filters.......................................................41
Setting the Flow Rate ..................................................................43
Cleaning the Cyclone ..................................................................44
Storage Precautions....................................................................44
5 TROUBLESHOOTING................................................................45
A SPECIFICATIONS.......................................................................49
B MODEL 8520-1 DUSTTRAK ENVIRONMENTAL ENCLOSURE
OPERATION AND MAINTENANCE MANUAL..........................51
Unpacking the DUSTTRAK Environmental Enclosure...................51
Spare Parts..................................................................................51
Parts Identification: Environmental Enclosure.............................52
SETTING-UP...............................................................................54
Installing the Aerosol or Respirable Aerosol Inlet .......................54
Install Water Trap Bottle..............................................................56
Correct Placement of Velcro Straps ............................................57
Install DUSTTRAK Monitor into Environmental Enclosure.............57
Connecting Tubing ......................................................................59
Supplying Power to the DUSTTRAK Aerosol Monitor.................... 59
Using the External Battery Pack ...........................................60
Install External Battery Pack .................................................60
iii
Mounting to a Surveyor Tripod.................................................... 61
Downloading a Data File ............................................................. 62
Real-Time Monitoring.................................................................. 62
Setting-up TRAKPROTM Data Analysis Software .......................... 62
Transporting the Environmental Enclosure................................. 63
Wiring the Analog/Alarm Cable in the Enclosure ........................ 63
OPERATION ............................................................................... 64
Overview...................................................................................... 64
How to Properly Orient the Environmental Enclosure................. 64
Changing and Re-charging the Battery Packs............................ 64
Receiving Data from the DUSTTRAK Aerosol Monitor.................. 64
Zeroing the DUSTTRAK Aerosol Monitor ...................................... 65
Setting the Flow Rate.................................................................. 65
Locking the Environmental Enclosure......................................... 65
Checklist for Sampling with the Environmental Enclosure.......... 66
MAINTENANCE.......................................................................... 67
Daily Maintenance Checks.......................................................... 67
Cleaning the Aerosol Inlet ........................................................... 67
Cleaning the Respirable Aerosol Inlet......................................... 67
When to Change the Battery....................................................... 67
Replacing the Battery Pack......................................................... 67
Recharging the Battery Packs..................................................... 68
Battery Pack Life ......................................................................... 68
Emptying the Water Trap ............................................................ 68
Storage Precautions.................................................................... 69
TROUBLESHOOTING THE ENVIRONMENTAL
ENCLOSURE.............................................................................. 70
SPECIFICATIONS: ENVIRONMENTAL ENCLOSURE............. 73
INDEX ................................................................................................ 75
AVAILABLE APPLICATION NOTES
DUSTTRAK, Theory of Operation...........................................ITI-036
Frequently Asked Questions: DUSTTRAK Aerosol
Monitor ..............................................................................ITI-039
Serial Interface Commands and Connections:
Model 8520 .......................................................................ITI-044
DUSTTRAK Analog Output .....................................................ITI-073
DUSTTRAK Alarm Output.......................................................ITI-074
To obtain any of the listed Application Notes, contact TSI at:
U.S. (800) 874-2811/(651) 490-2811, Fax: (651) 490-3824
International (001 651) 490-2811, Fax: (001 651) 490-3824
iv
These Application Notes can also be found under TSI’s web site:
http://www.tsi.com
v
Safety Information
When operated according to the manufacturer’s instruction, this device is a
Class I laser product as defined by U.S. Department of Health and Human
Services standards under the Radiation Control for Health and Safety Act of
1968. A certification and identification label like the one shown below is
affixed to each instrument.
May, 2006
THIS LASER PRODUCT COMFORMS
TO 21 CFR 1040.10 BUREAU OF
RADIOLOGICAL HEALTH AND
SAFETY REGULATIONS.
TSI INCORPORATED
500 CARDIGAN ROAD
SHOREVIEW, MN 55126 USA
There are no user-serviceable parts inside this instrument. Performing
services other than those described in this manual may result in exposure to
harmful (invisible) laser radiation. A warning label like the one shown
below is affixed to the internal laser device.
DANGER: INVISIBLE LASER
RADIATION WHEN OPEN. AVOID
DIRECT EXPOSURE TO BEAM.
WARNING: NO USER SERVICEABLE
PARTS INSIDE. REFER SERVICING
TO QUALIFIED PERSONNEL.
WARNING
!
The DUSTTRAK monitor Alarm Output function should not
be used to detect hazardous conditions or to provide an
alarm for protecting human life, health, or safety.
vi
1
Chapter 1
Unpacking and Parts Identification
Carefully unpack the Model 8520 DUSTTRAKTM Aerosol Monitor from the
shipping container. Use the tables and illustrations below to make certain
that there are no missing components. Contact TSI immediately if anything
is missing or damaged.
Unpacking the DUSTTRAK Aerosol Monitor
Compare all the components you received with those listed in the table
below. If any parts are missing, contact TSI.
Quantit
y
Item Descri
p
tion Pa
r
t/Model
1 Model 8520 Aerosol Monitor 8520
1 Carrying Case 800686
4 C-size Alkaline Batteries 1208018
1 AC adapter
115 V, NEMA-1-15
230 V, Eur., CEE 7/16
230 V, Great Britain
240 V, Australian
2613033
2613078
800169
2613105
1 Data Analysis Software CD-ROM 800700
1 Zero Filter 800663
1 Computer Cable, 9-pin to RJ-45 800563
1 25-Pin to 9-Pin Serial Cable Adapter 962003
1 Model 8520 DUSTTRAK Aerosol Monitor
Operation and Service Manual 1980198
1 Calibration Certificate
1 10 mm Nylon Dorr-Oliver Cyclone 800665
1 Flowmeter, 0–2.5 L/min 800669
4 feet Plastic Tubing, 1/4 in I.D. 800668
1 Small Screwdriver 800661
1 Sample Nozzle Removal Tool 800662
8 Spare Internal Filter Element 1602188
1 Wrist Strap 2913028
1 Shoulder Strap 2913120
1 Sample Nozzle Cleaning Brush 800699
10 Cotton Swab
1 Inlet Nozzle, 2.5 µm 1508112
1 Inlet Nozzle, 1.0 µm 1508113
1 Impactor Plate 1508114
1 Tube, Grease 801163
1 Analog Output/Alarm:connector, with cable 801652
Chapter 1
2
Parts Identification for the DUSTTRAK Aerosol Monitor
Figures 1–1 and 1–2 identify the parts of the Model 8520 DUSTTRAK
Aerosol Monitor. Become familiar with these components before
proceeding.
Figure 1–1: Top and Front of the DUSTTRAK Monitor
1. Display 7. External Power Socket
2. Keypad 8. Sample Inlet Nozzle and Port
3. Battery Cover 9. Flow Adjustment Screw
4. Battery Cover
Thumb Screw 10. Analog Output/Alarm
Connector
5. Data Port 11. Cyclone Holder Clip
6. Display/Keypad
Lockout Switch 12. Exhaust Port
2
1
3
4
5 67
8 9
11 12
10
Unpacking and Parts Identification 3
Figure 1–2: DUSTTRAK Aerosol Monitor Accessories
1. Zero Filter 9. AC Adapter
2. Flowmeter 10. Cyclone
3. Wrist Strap 11. 25-Pin to 9-Pin Adapter
4. 2.5 µm Inlet Nozzle 12. Shoulder Strap
5. 1.0 µm Inlet Nozzle 13. Sample Tube Cleaning Brush
6. Impactor Plate 14. Sample Tube Removal Tool
7. Grease 15. Internal Filter Elements
8. Computer Cable 16. Analog/Alarm Cable (not
shown)
1
2
3
8
9
10 11 12 13 14
15
4567
Chapter 1
4
5
Chapter 2
Setting-Up
Supplying Power to the DUSTTRAK Aerosol Monitor
The Model 8520 DUSTTRAKTM Aerosol Monitor must be powered in one of
two ways: four size C batteries or the supplied AC adapter.
Installing the Batteries
Remove the battery cover and slide the battery holder out of the battery
compartment and insert four size C batteries.
Hold the wires out of the way and slide the battery holder back into the
battery compartment. Tuck the wires in and then re-install the battery
cover (see Figure 2–1).
Figure 2–1: Tuck the Wires In and Close the Cover
Using the AC Adapter
The AC adapter allows you to power the DUSTTRAK monitor from an
AC wall outlet. When using the AC adapter, the batteries (if installed)
will be bypassed. The AC adapter will not charge the batteries.
The DUSTTRAK monitor has an internal, non-user accessible battery that
is used to keep logged data intact when power is turned off. Changing
the C-size batteries or disconnecting the AC adapter will not cause data
to be lost. This battery will last for years. TSI will install a new battery,
if necessary, when the unit is returned to the factory for service.
Chapter 2
6
Instrument Setup
The DUSTTRAK monitor comes with special software called TRAKPROTM
Data Analysis Software, which is designed to provide you with maximum
flexibility and power when using the DUSTTRAK monitor. The following
sections describe how to install the software and set up the computer.
Setting-up TRAKPRO Data Analysis Software
TRAKPRO software contains a very comprehensive Help Function. This
utility provides all the necessary information to guide you in all aspects
of software operation.
Connecting the DUSTTRAK Aerosol Monitor to the Computer
Each DUSTTRAK monitor comes equipped with an RS-232 cable and a
25-pin to 9-pin serial cable adapter. One end of the cable is a 25-pin D
subminiature connector labeled COMPUTER; the other end is an RJ-45
modular connector that connects with the logging instrument. Serial port
connectors always have pins (male) on the computer side.
1. Locate an available serial port on your computer: COM1, COM2,
COM3, or COM4.
2. If the port has a 25-pin connector, you do not need the adapter. If
the port has a 9-pin connector, plug the 25-pin end of the adapter
into the RS-232 cable.
3. Connect the RS-232 cable to the available serial port on your
computer.
4. Connect the RJ-45 connector to the DUSTTRAK monitor.
Set-Up the Communications Port
To communicate with the DUSTTRAK monitor, the software must be
configured for the proper COM port. The TRAKPRO software can be
manually set to operate on a specific COM port, or it can automatically
find a DUSTTRAK monitor that is attached to any COM port. To set up
the COM port, do the following:
1. Turn on the DUSTTRAK monitor and start the TRAKPRO software.
2. Select Communications from the Instrument Setup menu. The
following dialog is displayed:
Setting-Up 7
3. Select the following:
Serial
Port Select the name of the serial port to which the
logging device is connected: COM1, COM2,
COM3, or COM4.
Baud
Rate Select the baud rate for the port. Higher baud rates
are recommended to transfer data at a faster rate.
Select a lower baud rate only if you are having
trouble communicating at a higher rate.
4. Select Test to verify that you have set up the communications port
properly. The system displays an informative message indicating
whether it was able to establish communications.
5. As an alternate, you can select Find Port, to have the TRAKPRO
software search the available COM ports, looking for an attached
DUSTTRAK monitor.
6. Select OK to accept the setup or Cancel to discard the changes.
Note: Some computers do not communicate reliably at baud rates
above 9600.
Setting the Real-Time Clock
The DUSTTRAK monitor has an internal real-time clock that keeps track
of the time of day (The format is HH:MM where HH is the hour in 24-
hour format and the MM is minutes) and the date. It is very important
for the DUSTTRAK monitor to have the time and date correctly set;
otherwise, date and time stamping of recorded data and calibrations will
not be correct.
There are two ways to set the time and date. The first is to use the
supplied TRAKPROTM Data Analysis Software.
Chapter 2
8
Programming the Date/Time Using TRAKPRO Software
To program the DUSTTRAK monitor date and time:
1. Make sure the DUSTTRAK monitor is connected to the computer
and turned on.
2. Select Parameters, then Clock from the Instrument Setup menu.
The TRAKPRO software retrieves the current date and time settings
from the DUSTTRAK monitor and displays them in the following
dialog:
3. The system date and time (from the computer) can be transferred to
the DUSTTRAK monitor using the “arrows” keys. Alternately, the
date and time can be manually entered into the dialog box.
4. Select Send to reprogram the DUSTTRAK monitor.
Manually Setting the Real-Time Clock
To set the time and date with the keypad, you must press and hold the
SAMPLE key while the DUSTTRAK monitor displays the time of day
during its power-up. Release when the DUSTTRAK monitor “beeps.”
You will have an opportunity to view and/or change the hours, minutes,
year, month, and day of month in sequence. Use the up and down arrow
keys (ÏÐ) to change a setting. Use the SAMPLE key to store each
setting and advance to the next one.
Connecting the Optional Portable Printer
To connect the portable printer to the DUSTTRAK monitor, locate Printer
Interface Cable and connect the 9-pin end labeled “PRINTER” to the printer
and the other end to the data port on the DUSTTRAK. monitor Always turn the
DUSTTRAK monitor on BEFORE the printer. If the printer prints question
marks (??????), asterisks (******), or random characters, reset it by turning
05/09/2006 05/09/2006
Setting-Up 9
it off and then on again. If necessary, refer to the Portable Printer Operation
and Service Manual.
Connecting/Wiring the Analog/Alarm Output Connector
The Model 8520 DUSTTRAK Aerosol Monitor is capable of providing an
analog output voltage signal that is proportional to the displayed
concentration. It also contains a switch closure that is tied to an alarm value
(see Chapter 3, “Operation,” for complete specifications and operational
information for these features). The DUSTTRAK monitor is supplied with an
output cable. The cable contains a 4-pin, mini-DIN connector. See
Figure 2-2 below.
1. Analog Ground (-) 3. Alarm Positive (+)
2. Analog Output (+) 4. Alarm Ground (-)
Figure 2–2: Analog/Alarm Connector Pin-Outs
Since every application of this function will have unique requirements, the
customer is responsible for making connections to their own equipment. The
output cable contains a label, showing the wiring diagram/pin-outs.
Please see TSI Application Notes ITI-073 and ITI-074 for complete
information and examples of wiring and using the Analog/Alarm Outputs.
These Application Notes are available at the TSI website,
http://www.tsi.com.
1
23
4
Chapter 2
10
11
Chapter 3
Operation
Overview
The Model 8520 DUSTTRAKTM Aerosol Monitor has four modes of
operation, Survey, LOG 1, LOG 2, and LOG 3. When the DUSTTRAK
monitor is first turned on it will be in Survey mode, which is used to display
real-time readings and to determine statistics such as average, minimum, and
maximum readings. LOG 1 mode is used to record individual data points for
later analysis using a fixed protocol. LOG 2 and LOG 3 modes have a user-
defined protocol, set up using TRAKPROTM Data Analysis Software.
TRAKPRO software is used for analysis of data taken in any of the three LOG
modes, but cannot be used on samples taken in Survey mode.
DUSTTRAK Aerosol Monitor Keypad Functions
When pressing the keys on the front panel, the DUSTTRAK monitor beeps to
confirm the function. If you press a key and the DUSTTRAK monitor does not
beep, the DUSTTRAK monitor does not allow that function during the
selected sampling mode.
ON/OFF Key
Press the ON/OFF key to power the DUSTTRAK monitor. The DUSTTRAK
monitor immediately begins an internal self-check, while illuminating all the
display digits and symbols. If a problem is detected, the display shows the
message “SERVICE” along with a number to indicate that the DUSTTRAK
monitor requires servicing. Refer to the Chapter 5: “Troubleshooting,” for
information regarding service numbers. If the “SERVICE” message appears,
the DUSTTRAK monitor pauses until any key is pressed.
When the DUSTTRAK monitor completes its internal self-check, it displays
the approximate percentage of battery life remaining. The DUSTTRAK
monitor displays the battery symbol + – when the battery voltage becomes
very low. After the battery symbol appears, the DUSTTRAK monitor runs for
approximately 60 minutes before displaying the message “LO” (for a few
seconds) and then automatically turning off. This feature is accurate for
alkaline batteries only.
Chapter 3
12
Note: The percentage life remaining will not be accurate for NiCd
batteries. The battery symbol appears when battery voltage
becomes low, but the DUSTTRAK monitor runs considerably less
than 60 minutes before displaying the message “LO” and turning
off.
CO
2
SAMPLE
CALIBRATE ZERO RECORDING
HR:MN:SEC TEST ID LOG 123
CO2 SAMPLE
ELAPSED TIME
CONSTANT
AVERAGE
MIN MA
X
SERVICE
% MEMORY
PPM
°F°C
%RH
m
g
/m3
Figure 3–1: DUSTTRAK Display with All Elements Shown
After displaying the percentage of battery life remaining, the current time set
on the internal real-time clock is displayed. When the self-check is complete,
the DUSTTRAK monitor will be in Survey mode.
SAMPLE Key
Press the SAMPLE key to start/stop data sampling. The word “SAMPLE
appears in the upper right corner of the display while the DUSTTRAK monitor
is taking a sample. When sampling is stopped, the DUSTTRAK monitor
automatically scrolls through statistics for the sample that just ended.
TIME CONSTANT Key
Momentarily press and release the TIME CONSTANT key to view the
current time-constant. To change the time-constant, press and hold the key
down. The available time-constant choices will scroll on the display. When
the desired value is displayed, immediately release the key.
The time-constant is actually an averaging period. The DUSTTRAK monitor
display is always updated every second. However, the reading displayed is
the average reading over the last time-constant period. For example, if the
current time-constant is set to 10-seconds, the display shows readings
averaged over the previous 10 seconds, updated every second. This is also
called a 10-second “moving average.”
As configured at the factory, the available time-constant values are 1, 5, 10,
15, and 30 seconds. The internal list of time-constant values can be altered
using TRAKPRO Data Analysis Software supplied with the DUSTTRAK
monitor. The range of time constants allowed is 1–60 seconds. See the
following instructions.
Operation 13
Programming the Available Time Constants
To program the list of time constants using TRAKPRO Data Analysis
Software:
1. Make sure the DUSTTRAK monitor is connected to the computer and
turned on.
2. Select Parameters, then Time Constants from the Instrument Setup
menu. TRAKPRO software retrieves the current time constant settings
from the DUSTTRAK monitor and displays them in the following dialog:
3. Enter a value for each of the five available time constants. (The range is
limited to 1-60 seconds.)
4. Select Send.
The DUSTTRAK monitor is reprogrammed to offer the time constants you
have specified.
STATISTICS Key
Use the STATISTICS key to sequentially view the average, minimum, and
maximum readings as well as the elapsed time of the most recently sampled
data. If one of the LOG modes is active, a test identification number will be
displayed also. Press the STATISTICS key once to display the average
reading, again to display the minimum reading, again to display the
maximum reading, and again to display the elapsed time for that sample (and
again for the test ID if in LOG mode). If you press the STATISTICS key a
fifth time (sixth time if in LOG mode), the DUSTTRAK monitor switches
back into the currently selected measuring mode. You must sequence
through all four statistic displays (i.e., press the STATISTICS key five
Chapter 3
14
times, six if in LOG mode) before the DUSTTRAK monitor goes back into the
measuring mode.
PRINT Key
Use the PRINT key to print information on the optional Portable Printer.
The information printed will be different depending on what the DUSTTRAK
monitor is currently doing.
When the DUSTTRAK monitor is displaying real-time readings, pressing the
PRINT key causes the reading to be printed along with the time and date.
Each time the PRINT key is pressed, one reading will print. The reading
printed reflects the current time-constant; therefore, it is the same as the
displayed reading.
When the DUSTTRAK monitor is displaying any statistic, pressing the PRINT
key causes the current statistics to print.
When the DUSTTRAK monitor is in one of the LOG modes and is idle
(“%MEMORY” is displayed), pressing the PRINT key causes the logging
setup for the current LOG mode to print.
During the power-up sequence, if you press and hold the PRINT key after
the time is displayed, a printout will be generated showing certain system
information (printer must be connected and turned on).
SAMPLING MODE Key
The SAMPLING MODE key allows you to select among the four sampling
modes: Survey, LOG 1, LOG 2, and LOG 3. Each time you press the
SAMPLING MODE key, the DUSTTRAK monitor sequences to the next
mode. When the DUSTTRAK monitor is in Survey mode, the current
measurement will be shown on the display. When one of the LOG modes is
selected, the LOG mode number, i.e. LOG 1, LOG 2, or LOG 3 will be
displayed along with the percentage of free memory available.
Survey Mode
When the DUSTTRAK monitor is first turned on, it will always be in
Survey mode. Survey mode allows you to make real-time measurements
of aerosol concentration, and to obtain statistics for that sample. The
statistics include the average, minimum, and maximum values, as well
as the elapsed time for that sample. Individual data points are not
recorded (this can be done in any of the LOG modes). Each new sample
taken in Survey mode clears the previous sample data from memory.
Data taken in Survey mode remains in memory until another sample is
made, or until samples are taken in one of the LOG modes. Turning the
Operation 15
DUSTTRAK monitor off will not erase data. Use the SAMPLE key to
start and stop a sample.
LOG Modes
There are three LOG modes: LOG 1, LOG 2, and LOG 3. When one of
these modes is selected using the SAMPLING MODE key, the LOG
mode number will be listed at the top of the display. The LOG modes
allow you to record aerosol concentration data points for later retrieval
and analysis using the software provided with the instrument.
Use the SAMPLE key to start and stop recording. The word
“RECORDING” appears at the top of the display when recording is in
progress.
The frequency that data is recorded can be set for LOG 1 mode with the
LOGGING INTERVAL key. Use TRAKPRO software for LOG 2 and
LOG 3 modes.
Data recorded using one of the LOG modes can only be erased by using
the CLEAR MEMORY key. Turning the DUSTTRAK monitor off will
not erase data. Recording another sample with one of the LOG modes
without clearing memory first will cause new data to be added to the
existing data (using a new test ID).
CLEAR MEMORY Key
Use the CLEAR MEMORY key to erase all data. The CLEAR MEMORY
key will not respond unless the DUSTTRAK monitor is first put into one of
the three LOG modes by using the SAMPLING MODE key.
Note: Before clearing memory, you should first download the data to a
computer through the TRAKPRO software. There is only one block of
memory in the DUSTTRAK monitor. Clearing the memory for one
LOG mode clears memory for all LOG modes.
To clear memory, press and hold the CLEAR MEMORY key until the
countdown reaches zero, then release quickly. This prevents accidental
erasure of data. Releasing the key too soon or too late prevents memory
from being cleared.
LOGGING INTERVAL Key
Use the LOGGING INTERVAL key to view or set the frequency/averaging
period for recording data in LOG 1 mode or to view the current interval
setting in LOG 2 and LOG 3 mode. Use TRAKPRO software to set the
logging interval for LOG 2 and LOG 3 mode (see following section). The
LOGGING INTERVAL key will not respond unless the DUSTTRAK monitor
is first put into one of the three logging modes. Press the LOGGING
Chapter 3
16
INTERVAL key momentarily to view the current logging interval. Press and
hold the LOGGING INTERVAL key to sequence through the available
choices and release the key when the desired interval is on the display (LOG
1 mode only).
The logging interval is both a frequency and an averaging period. For
example, when the logging interval is set to 30-minutes, readings will be
recorded at 30-minute intervals. Each reading will be the average value
measured over a 30-minute interval.
As shipped from the factory, the available logging intervals for LOG 1 mode
are 1 second, 1 minute, 5 minutes, 15 minutes, and 30 minutes. Use the
TRAKPRO Data Analysis Software to alter these values.
Programming the Logging Intervals using TRAKPRO Software
To program the list of logging intervals available for LOG 1 mode:
1. Make sure the DUSTTRAK monitor is connected to the computer and
turned on.
2. Select Parameters, then Logging Intervals from the Instrument
Setup menu. TRAKPRO software retrieves the current logging intervals
from the DUSTTRAK monitor and displays them in the following dialog:
3. Enter a value for each of the five available logging intervals (the range
is from 1 second to 59 minutes and 59 seconds).
4. Select Send.
The DUSTTRAK monitor is reprogrammed to offer the logging intervals you
have specified.
Operation 17
Ï and Ð Keys
The two arrow keys are used to adjust readings when calibrating the
DUSTTRAK monitor and for adjusting the time and date for the internal real-
time clock.
CALIBRATE Key
Use the CALIBRATE key to put the DUSTTRAK monitor into calibration
mode. See Chapter 4, “Maintenance” for details on how to calibrate.
Display/Keypad Lockout Switch
Recording data over extended time periods often requires leaving the
DUSTTRAK monitor unattended. To reduce the risk of having an
unauthorized person either intentionally or inadvertently interrupt the
measurements, you can lock the display and keypad.
The lockout switch is on the backside of the DUSTTRAK monitor between the
data port and the external power socket. It is a small slide switch and is
recessed so that a pointed instrument must be used to move it.
With the switch in the right side position (nearest the power socket), all
keypad and display functions will work normally. With the switch in the left
side position (closest to the data port), the display and keypad will go into
lockout mode. There are two ways to use the lockout switch. You can lock
the keypad after recording starts, or you can put the switch into the lockout
position prior to when recording starts. If you select the second method, you
will be able to operate all functions normally until the SAMPLE key is
pressed in one of the LOG modes. At that time the keypad automatically
locks. When the display and keypad are locked, the display shows the words
RECORDING LOG X Ð where “X” is the current LOG mode number.
Remote Sampling
The DUSTTRAK Aerosol Monitor can be used for remote sampling by
attaching a length of tubing to the inlet. For example, it can be set up for a
test and hidden away where only the sample tubing extends into the area to
be monitored. A 4-foot length of tubing is included with the DUSTTRAK
monitor for remote sampling purposes.
Note: Using a sampling tube longer than 4 feet is not recommended
because particle transport losses in the tube may adversely affect
the measurement. Also, do not use Teflon or silicone tubing
because they can cause significant transport losses. Tygon® tubing
is a good choice, as is metal tubing if it is grounded to dissipate
static charges.
®Tygon is a registered trademark of Norton company.
Chapter 3
18
Respirable Mass Sampling with the Cyclone
The 10-mm Nylon Dorr-Oliver Cyclone included with your aerosol monitor
can be used to discriminate between the respirable fraction and other
portions of the ambient aerosol. Four micrometers (µm) is internationally
accepted as the 50 percent cut-off size for respirable aerosol mass. Particles
larger than 4 µm impact onto the surfaces of the upper respiratory tract and
cannot reach the lungs. The cyclone accessory provided with the DUSTTRAK
monitor is designed to provide a cut-off at 4 µm. This is specified as a 50
percent cut-off at 4 µm.
The cyclone works by forcing the particle-laden air sample to swirl inside
the cyclone body. Larger (higher mass) particles cannot follow the air stream
and become trapped, while smaller particles stay in the air stream and pass
through. When using the cyclone, you can assume that all particles smaller
than the cut-off size pass through and all larger particles become trapped in
the grit pot.
The cut-off size for any cyclone is dependent on flow rate. It is very
important that the sample flow rate through the DUSTTRAK monitor be
set at 1.7 liters per minute (L/min). If some other flow rate is set, the
cut-off size will not be at 4 µm.
Using the Cyclone
To use the cyclone, be sure that you have the black inlet nozzle in place and
that the blue impactor plate is removed from the inlet. Then you must adjust
the sample flow rate to 1.7 L/min. See Chapter 4: “Maintenance,” for details.
After the flow rate is properly set, attach it to the DUSTTRAK monitor inlet port
using the section of plastic tubing supplied with the cyclone. Slide the cyclone
body into the holder provided on the back of the DUSTTRAK monitor. See
Figure 3–2.
Operation 19
Figure 3–2: Connecting the Cyclone
1. DUSTTRAK Sample Inlet 3. Cyclone Holder
2. Cyclone
Using the 1.0 or 2.5 µm Inlet Conditioners
The two blue inlet nozzles appear similar to the standard black inlet nozzle
but are engraved with their respective cut size. The blue nozzles should
always be used with the blue impactor plate. The black nozzle has a cut size
of 10 μm and should never be used with the impactor plate.
Install either the 1.0 or 2.5 μm inlet conditioner as follows.
1. Turn the DUSTTRAK monitor off.
2. Remove the black inlet nozzle.
3. Evenly coat the center plateau of the blue impactor plate with a thin
layer of the supplied grease (Figure 3–3). Avoid getting the grease on
other parts of the impactor plate. If you do, simply wipe off the extra.
Notes: Do not use grease other than that supplied with this kit. This grease
has been chosen for its low outgassing and good wicking properties
to hold a large mass of particles.
The grease on the impactor plate increases the amount of time that
you can sample before particle bounce begins to affect the impactor
performance. Particle bounce results from particles not sticking to
the impactor plate and falling back into the aerosol stream.
4. Insert the blue impactor plate into the DUSTTRAK monitor.
5. Select the blue nozzle with the cut size you want and make sure there is
an O-ring inside.
1
2
3
Chapter 3
20
6. Thread the nozzle on the inlet.
7. Turn on the DUSTTRAK monitor and verify that the flow rate is at 1.7
L/min. You might need to make a small adjustment when you switch
nozzles.
The DUSTTRAK monitor is calibrated to the respirable fraction of ISO
12103-1, A1 (formerly called ultrafine Arizona test dust or SAE ultrafine).
To increase the accuracy while using the inlet conditioners, you will need to
calibrate in the aerosol you wish to test. Please refer to the manual for
calibration details.
Figure 3–3: Installing the 1.0 or 2.5 µm Inlet Conditioners
Because of the much higher percentage of particles being removed from the
stream when using the smaller cut-size inlets, you will need to clean the inlet
more frequently than for the standard 10 μm nozzle (black.) See
maintenance schedule in Chapter 4 for the inlet cleaning frequency.
Using the 10 µm Nozzle
1. Remove the blue nozzle and blue impactor plate.
Note: The blue impactor plate must be removed from the inlet when
using the black nozzle. Performance of the 10 micrometer nozzle
will be severely degraded if the impactor plate is not removed.
2. Thread the black nozzle onto the DUSTTRAK monitor inlet.
3. Verify that the flow rate is at 1.7 L/min.
4. The DUSTTRAK monitor now measures aerosol with an upper particle
size limit of 10 μm.
Operation 21
Analog Output
Newer models of the Model 8520 DUSTTRAK Aerosol Monitor contain an
analog output feature. This means the instrument is capable of providing an
analog voltage signal that is proportional to mass concentration.
Note: Before using this feature, it is important to read this operational
section. For further information, please see Application Note
ITI-073, which describes the Analog Output programming and
operation. This Application Note is available at the TSI website,
http://www.tsi.com.
Because the DUSTTRAK monitor measures aerosol over such a wide dynamic
range, it is necessary to select the appropriate range before using the analog
output feature. The instrument may be programmed to one of four options,
using the TRAKPRO software. When shipped from the factory, the instrument
is programmed to the lowest range. To select the analog output range, do the
following:
1. Turn on the DUSTTRAK monitor and start the TRAKPRO software
(version 3.2 or higher).
2. Select Parameters, Analog Output, from the Instrument Setup menu.
The following dialog is displayed:
3. Select one of the four ranges, and press Send.
Please note the following considerations when using the analog output
feature.
The analog output function is always “On.” You do not need to activate
this function.
The interpretation of the analog output voltage is directly tied to the
chosen range. For example, at the lowest range, each “0.001 mg/m3” of
Chapter 3
22
aerosol is equivalent to “0.050 VDC.” If the instrument display shows
“0.018 mg/m3,” the analog output voltage would be “18 x 0.050 VDC”
or “0.9 VDC” (see Application Note ITI-073, for more specifications
and examples).
Select the lowest range consistent with your desired application. For
example, if you typically measure aerosol mass concentrations around
0.025 mg/m3, you should use the first range. This will provide you with
the best resolution (strongest signal) corresponding to the mass
concentration.
If your application involves high-mass concentrations (wood dust, etc.)
you may want to choose the second or third range.
If the DUSTTRAK monitor measures mass concentrations that are greater
than the selected range, the analog voltage will “rail” at 5.0 volts
(maximum output; no change in signal beyond that point).
All of the ranges are referenced back to “zero” mg/m3.
Alarm Output
Newer models of the DUSTTRAK Aerosol Monitor contain an alarm output
feature. This means the instrument is capable of providing an alarm signal
when the measured mass concentration exceeds a certain threshold.
Note: Before using this feature, it is important to read this operational
section. For further information, please see Application Note
ITI-074, which describes the Alarm Output programming and
operation. This Application Note is available at the TSI website,
http://www.tsi.com .
Operation 23
Alarm Output Specifications
Please note the following considerations when using the analog output
feature.
WARNING
!
The DUSTTRAK monitor Alarm Output function should not
be used to detect hazardous conditions or to provide an
alarm for protecting human life, health or safety.
The alarm output function default status is “Off.” You must activate
this function and set the alarm level, before using the alarm output
(see section below, on programming the Alarm Output).
You must supply your own powered alarm system. The DUSTTRAK
monitor only contains a switch that closes (activates) the alarm,
when the detected levels exceed the threshold.
The user alarm must be DC-powered only! The alarm system must
not exceed a maximum supply voltage rating of 15 VDC or a
current draw of 1 amp.
Caution
!
The alarm switch must not be wired to AC power! Failure
to properly install the user alarm could damage the
DUSTTRAK instrument and/or void the instrument
warranty! Please read and follow all instructions before
wiring or operating the user alarm.
The alarm switch contained within the DUSTTRAK monitor has an
electrical polarity. It must be wired with the supply voltage
connected to the positive pole. If the user alarm is wired with
reverse polarity, it will not function properly.
The alarm threshold must be set between the values of 0.002 and
100 mg/m3.
The alarm switch will turn on the instant the setpoint is reached. It
will remain “latched” for a minimum of 5 seconds.
While the alarm is triggered, the DUSTTRAK monitor’s internal
beeper will sound once per second, and the display will alternately
exhibit the characters “AL” to indicate an alarm condition.
The alarm switch will turn off when the measured concentration
falls 5% below the setpoint (dead-band).
The alarm output function operates totally independently of the
analog output.
Chapter 3
24
Programming the Alarm Setpoint
To program the alarm output function, do the following:
1. Turn on the DUSTTRAK monitor and start the TRAKPRO software
(version 3.2 or higher).
2. Select Parameters, Alarm Setpoint, from the Instrument Setup
menu. The following dialog is displayed:
3. Turn “On” the alarm function by checking the “Alarm Enabled”
check box. Enter a value within the acceptable range. In this
example, the DUSTTRAK monitor is programmed to a setting of
“0.100 mg/m3.” Press the Send button, to program the instrument.
Upgrading the DUSTTRAK Aerosol Monitor to Add Analog and
Alarm Functions
The newest version of the DUSTTRAK Aerosol Monitor contains the analog
output and alarm functions. Older instruments may be hardware compatible
with these functions and may only require a factory upgrade to install these
features. Instruments manufactured before August 1999 are not capable of
being upgraded to add these features. These three scenarios are summarized
below.
Identifying an Instrument with Analog/Alarm Functions
If your instrument contains an Analog/Alarm Output connector on the
back of the instrument, it is fully functional for both analog and alarm
output functions.
Operation 25
Identifying an Instrument that may be Upgraded to Add
Analog/Alarm Functions
If your DUSTTRAK Aerosol Monitor was manufactured after August
1999 (serial number 21960 or later) and does not contain the
analog/alarm output connector, it may be upgraded at the factory to add
this feature. To verify this status, simply attach the instrument to the
TRAKPRO software and attempt to program the Analog or Alarm
Setpoint functions. You should receive the following message:
This instrument may be returned to the factory, for a Cleaning,
Calibration and Upgrade (there is a charge for this service).
Identifying an Instrument That May Not Be Upgraded
If your DUSTTRAK Aerosol Monitor was manufactured prior to August
1999 (serial number 21959 or earlier), it is not hardware compatible
with these functions. It is not possible to upgrade these instruments to
add analog/alarm functions. To verify this status, simply attach the
instrument to the TRAKPRO software and attempt to program the Analog
or Alarm Setpoint functions. You should receive the following message:
Chapter 3
26
Programming Advanced Modes: LOG 2 and LOG 3
Use LOG 2 or LOG 3 modes for unattended recording and setting user
protocols. With LOG 2 and LOG 3 modes you can set the start date, start
time, test length, logging interval, number of tests and the time delay
between tests. All or selected parameters can be set.
To program a protocol for LOG 2 or LOG 3 mode:
1. Make sure the DUSTTRAK monitor is connected to the computer and
turned on.
2. Select Logging Setup from the Instrument Setup menu. The TRAKPRO
software retrieves the current settings for LOG 2 and LOG 3 modes
from the DUSTTRAK monitor and displays them in the following dialog:
05/09/2006 05/10/2006
Operation 27
The following table summarizes the information displayed in the
DUSTTRAK Logging Protocols dialog box:
Serial Number Displays the serial number of the logging
instrument.
Number of tests
logged Displays the number of tests currently logged
and stored in the logging instrument.
Available
Memory (%) Displays the percent of available memory in
the logging instrument.
LOG 2 and LOG 3 Mode Protocols
Channels Displays the channels selected for sampling in
LOG 2 and LOG 3 modes.
Start Date Displays the start date for LOG 2 and LOG 3
modes.
Start Time Displays the start time for LOG 2 and LOG 3
modes.
Log interval Displays the log interval for LOG 2 and LOG
3 modes.
Test length Displays the test length for LOG 2 and LOG 3
modes.
Number of tests Displays the number of tests for LOG 2 and
LOG 3 modes.
Time between
tests Displays the time between tests for LOG 2
and LOG 3 modes.
Percent
memory
required.
Displays the percent of logger memory
required to perform a LOG 2 or a LOG 3
mode sample. To store the results of a LOG 2
or LOG 3 mode sample, the Available
Memory must be equal to or greater than the
Percent memory required.
Chapter 3
28
3. Enter the following for LOG 2 and LOG 3 modes:
Channels Select the channels for which you want to log
data. In the case of the DUSTTRAK monitor,
there is only one channel to select: Aerosol.
Start Date
Start Time Enter the date and time to begin the sample:
If you enter a blank for a start date, the
sample begins whenever the specified start
time occurs.
If you enter a blank for the start time, both
start date and start time are ignored, and
the sample begins when the operator
manually starts the sample.
Log interval Enter the log interval to use for the test.
Test length Enter the length for the sample:
If you enter a value, the instrument
automatically turns off when the last test is
complete.
If you enter a blank, the operator must
manually stop the sample.
Number of tests Enter the number of tests to perform.
Time between
tests If you have specified more than one Number
of tests, enter the time between tests. If you
enter 0 or blank, the next test is started
immediately after the last test is complete.
While you are entering values for LOG 2 and LOG 3 modes, the
Percent Memory Required is dynamically updated to show the amount
of logger memory required to take the programmed sample. If the
protocol you have defined requires more than 100% of memory, you
can decrease the amount of memory required by manipulating the
following protocol parameters:
Increase the logging interval.
Decrease the length for the test.
Decrease the number of tests.
The settings for each LOG mode must not require more than 100% of
the logger memory.
Note: If the Percent memory required is greater than the Available
memory, the logging instrument automatically stops the test
when memory is full.
4. When you have finished defining the parameters for LOG 2 and LOG 3
modes, select Send.
Operation 29
5. You can now disconnect the DUSTTRAK monitor and cable from the
computer.
Sample Protocol for LOG 2 and LOG 3 Modes
The following steps describe how to program a sample protocol for LOG 2
or LOG 3 mode.
The sample protocol for LOG 2 is set to take unattended aerosol readings for
one day, 11/02/2004. The logging sample begins at 8:00 a.m. and continues
for eight hours.
The sample protocol for LOG 3 is set up to take unattended aerosol readings
for two days, beginning on 11/02/2004. The logging sample begins at 8:00
a.m. and continues for eight hours. The instrument is off for 16 hours, and
then repeats the eight hours test on the following day.
The following graphic gives the appearance of the dialog box displayed in
the TRAKPRO software, with these particular logging parameters.
To program this logging example, do the following:
1. Make sure the DUSTTRAK monitor is connected to the computer and
turned on.
05/09/2006 05/10/2006
Chapter 3
30
2. Select Logging Setup from the Instrument Setup menu. The TRAKPRO
software retrieves the current settings for LOG 2 and LOG 3 modes
from the DUSTTRAK monitor and displays them in the previous dialog.
3. Enter the following for LOG 2 and LOG 3:
Setting LOG 2 LOG 3
Channels Aerosol Aerosol
Start Date 05/09/2006 05/10/2006
Start Time 08:00 08:00
Log interval 01:00 05:00
Test length 00:08:00 00:08:00
Number of tests 1 2
Time between
tests 00:00:00 00:16:00
4. Select Send. The logging instrument is programmed for the mode 2 and
mode 3 protocols.
5. Note that the LOG 2 test requires 1% of the available memory and
LOG 3 requires less than 1% of the memory. A total of 100% of the
memory is available for use.
6. You can now disconnect your DUSTTRAK monitor from the computer.
Refer to other sections of this Operation and Service Manual for details
on making measurements using LOG 2 and LOG 3 modes.
After programming the DUSTTRAK monitor with the TRAKPRO Data
Analysis Software, take the DUSTTRAK monitor to the desired location and
turn it on. Put it into LOG 2 or LOG 3 mode (whichever you programmed)
using the SAMPLING MODE key. Press the SAMPLE key to initiate the
program.
If you have set a start time and/or date, the display toggles between the next
“TEST ID” and the “ELAPSED TIME 0” message. If the test start time is
greater than one minute away, the DUSTTRAK monitor shuts off until one
minute before the test start time. This indicates that the DUSTTRAK monitor
is waiting until the programmed start time and dates occur. To prevent
tampering, lockout the display and keypad at this time.
Things You Should Know About Taking Pre-Programmed
Samples
If you press the SAMPLE key during programmed operation, the
program terminates (unless the keypad is locked).
Operation 31
If the programmed start time/date has already passed, pressing the
SAMPLE key has no effect. The program will never execute.
It is always best to enter both a start time and start date! However,
setting the start time but no start date causes the DUSTTRAK monitor to
start at the specified time regardless of the date. The instrument may not
shut down prior to beginning logging.
If no start time is set, the DUSTTRAK monitor waits for you to press the
SAMPLE key and then starts sampling immediately.
If no test length is set, the DUSTTRAK monitor takes samples continuously
until the SAMPLE key is pressed to stop sampling or until memory is full.
When a pre-programmed test ends, the DUSTTRAK monitor automatically
shuts off.
Memory Considerations
The DUSTTRAK monitor has a great deal of memory and you will not
normally have to be concerned with running out. The DUSTTRAK monitor
can store more than 31,000 data points. Therefore 1% of memory is about
310 data points.
This should be considered when selecting a logging interval. Shorter logging
intervals use memory more quickly than longer intervals.
The logging interval and the available memory determine the maximum
possible duration of a data logging session. The equation below can be used
to determine any memory, recording time or logging interval restraints.
[]
[][ ]
Interval Logging315
Time Elapsed
Memory% ×
(Note: Elapsed Time and Logging Interval are in units of minutes.)
Custom Calibrations
In most situations, the DUSTTRAK monitor with its built-in data logger can
provide very good information on how the concentration of an aerosol
changes for different processes over time. Factory calibration to the
respirable fraction of standard ISO 12103-1, A1 test dust (formerly Arizona
Test Dust) allows comparisons between measurements where the source or
type of dust is predominately the same. Because optical mass measurements
are dependent upon particle size and material properties, there may be times
in which a custom calibration would improve your accuracy for a specific
aerosol.
Chapter 3
32
The DUSTTRAK monitor has several features to aid in obtaining good
accuracy for a particular aerosol.
The DUSTTRAK monitor’s custom calibration factor can be changed
through the instrument keypad. All future readings from the DUSTTRAK
monitor will correspond to a specific aerosol until the calibration factor
is changed back to the factory setting.
Custom calibration factors can be stored in a table and downloaded to
the DUSTTRAK monitor using the TRAKPRO Data Analysis Software. All
future measurements will correspond to a specific aerosol until the
calibration factor is changed back to the factory setting.
A single set of logged data can be converted to data calibrated to a
specific aerosol with the use of the TRAKPRO software. This conversion
can be done by knowing either the true mass concentration for the
logged data or the calibration factor for the aerosol. Future
measurements will continue to be read and logged with the original
calibration factor (normally 1.0) and will not be converted
automatically.
All of these options require that you determine a true mass concentration
(e.g., gravimetric analysis) for the aerosol you want to measure. The true
mass concentration is used to calculate the custom calibration factor for that
aerosol. Once you have a custom calibration factor, you can reuse it each
time you make measurements in the same or similar aerosol environment.
Determining the Calibration Factor for a Specific Aerosol
The DUSTTRAK monitor is factory calibrated to the respirable fraction of
standard ISO 12103-1, A1 test dust. The DUSTTRAK monitor can be easily
calibrated to any arbitrary aerosol by adjusting the custom calibration factor.
The DUSTTRAK monitor’s custom calibration factor is assigned the value of
1.00 for the standard ISO test dust. This procedure describes how to
determine the calibration factor for a specific aerosol. Using a value of 1.00
will always revert back to the factory calibration.
To determine a new calibration factor you need some way of accurately
measuring the concentration of aerosol, hereafter referred to as the reference
instrument. A gravimetric analysis is often the best choice, though it is
limited to nonvolatile aerosols.
To make an accurate calibration you must simultaneously measure the
aerosol concentration with the DUSTTRAK monitor and your reference
instrument.
1. Zero the DUSTTRAK monitor as described in the Maintenance chapter.
Operation 33
2. Put the instrument in Log Mode 1.
3. Set the logging interval by pressing and holding the LOGGING
INTERVAL button down. The default choices for the logging intervals
can be reprogrammed using TRAKPRO software. One minute (i.e.,
“01:00”) is often a good choice.
4. Locate the DUSTTRAK monitor and the reference sampler together so
that they are measuring from the same area.
5. Start sampling aerosol with both instruments at the same time.
Note: Greater accuracy will be obtained with longer samples. The
time you permit for sampling often depends on the reference
instrument and characteristics of the measured aerosol. It may
take some time to collect sufficient aerosol onto a filter cassette
for accurate gravimetric analysis. Refer to instructions of your
reference instrument for sampling times.
6. Stop sampling with both instruments at the same time.
7. Record the DUSTTRAK monitor average concentration. Press the
STATISTICS key to view the average for the last test. If you wish to
determine a custom calibration factor for this test only, you can instead
receive the data into the TRAKPRO software.
8. Determine the mass concentration in mg/m3 from your reference
instrument. For gravimetric sampling this means having the gravimetric
sample weighed.
9. Compute the new calibration constant, NewCal, using the following
formula:
CurrentCal
ionConcentratDustTrak
ionConcentrat Reference
NewCal
=
The calibration factor, OldCal, can be obtained from the DUSTTRAK
monitor. In Survey mode, press and hold the CALIBRATE key down
until the display counts down to zero. Release immediately when zero is
reached and “CALIBRATE ZERO” appears on the display. If not, try
again until it does. Press CALIBRATE again and the current calibration
factor will be displayed with flashing up and down arrows. Record this
value to make your calculation. Press CALIBRATE again to exit
calibration mode.
Setting Custom Calibration Factor from Keypad
1. In Survey mode, press and hold the CALIBRATE key down until the
DUSTTRAK monitor counts down to zero. Release the button
Chapter 3
34
immediately when zero is reached and “CALIBRATE ZERO” appears
on the display. If not, try again until it does.
2. Press CALIBRATE again and the current calibration factor will be
displayed with flashing up and down arrows.
3. Use the up and down arrows on the keypad to adjust the calibration
factor. To revert back to factory calibration, set to 1.00.
4. Press SAMPLE to store the value and exit calibration mode, or press
CALIBRATE to exit without saving the new value.
Setting Custom Calibration Factors using TRAKPRO Software
The TRAKPRO software allows you to store several custom calibration
factors along with a brief title for the aerosol or environment the calibration
was taken. A calibration factor can be downloaded to the DUSTTRAK
monitor when you want to make measurements of that aerosol or
environment. All future measurements will be adjusted by the new custom
calibration factor until you change it to some other value. To set this new
custom calibration factor using TRAKPRO software, do the following:
1. Connect your DUSTTRAK monitor to the computer and start the
TRAKPRO software.
2. Select DUSTTRAK Model 8520 from the Options menu. Or, select
DUSTTRAK from list of available instruments on the Drop-Down menu
on the menu bar.
3. Select Calibration in the Instrument Setup menu. The following
dialog box is displayed.
05/09/2006
05/09/2006
05/09/2006
Operation 35
4. Click the Set DUSTTRAK Calibration button. The following dialog box
is displayed.
5. Enter a new Calibration Factor along with a new Name and press
Add. This adds the new Calibration Factor to the list of available
factors. In the example above, the “Wood Dust” factor, with a value of
1.45, was added to the list.
Note: This step does not automatically download the new value to
the DUSTTRAK monitor. This must be done in a separate step.
6. Highlight the desired Defined Calibration factor (either the new one or
a previous factor) by clicking on it, and press Send. This downloads the
new Calibration Factor to the DUSTTRAK monitor. You can enter up to
25 different custom calibrations.
7. The DUSTTRAK monitor response for all subsequent measurements will
be multiplied by the new factor. To reset the DUSTTRAK monitor to
factory calibration, simply send the factory setting of 1.
Converting Stored Data to Calibrated Data
A single set of logged data can be converted to data calibrated to a specific
aerosol with the use of the TRAKPRO software. This conversion can be done
by knowing either the true mass concentration for the logged data or the
calibration factor for the aerosol. To perform this conversion, follow these
steps:
Note: This process converts existing data files using a new calibration
factor. It does not affect the current reading of the DUSTTRAK
monitor. Future measurements will continue to be read and logged
with the original calibration factor (normally 1.0) and will not be
converted automatically.
1. Connect your DUSTTRAK monitor to the computer and start the
TRAKPRO software.
Chapter 3
36
2. Select DustTrak Model 8520 from the Options menu under Software
Configure. Or, select DustTrak from list of available instruments on
the drop-down menu on the menu bar.
3. Open the file that contains the data to be converted. Highlight a single
test within that data file.
4. Select Define DustTrak Calibration from the Options menu. The
following dialog box is displayed.
5. Enter the New Mass Concentration value (in mg/m3). The New
Calibration Factor is calculated and automatically updated. Press OK.
6. When you click OK, a new data set of the same name will be generated
calibrated to the mass concentration that you entered. This data set will
be appended to the same file. The original data remains.
7. If you want to adjust the data to a previously determined custom
calibration factor without performing a gravimetric calibration, enter the
new calibration factor. The New Mass Concentration automatically
updates.
05/09/2006,
37
Chapter 4
Maintenance
The DUSTTRAK monitor can be maintained and calibrated in the field using
the instructions below. Even so, TSI recommends that you return your
DUSTTRAK monitor to the factory for annual calibration. For a reasonable
fee, we will quickly clean and calibrate the unit and return it to you in “as
new” working condition, along with a Certificate of Calibration. This
“annual checkup” helps ensure that the DUSTTRAK monitor is always in
good operating condition.
The DUSTTRAK monitor is factory calibrated to the standard ISO 12103-1,
A1 test dust (formerly Arizona Test Dust). The calibration data is stored
internally and cannot be accessed by you. This standard test dust is used
because of its wide particle size distribution. This makes the internal
calibration representative of an average of most types of ambient aerosol you
might encounter.
For many applications, the factory calibration provides accurate
measurements. There may be times when more accurate mass concentration
data is needed because the aerosol being sampled is significantly different
than the standard ISO test dust. To accommodate such a measurement, the
DUSTTRAK monitor has been designed to allow for field calibration.
Note: There are no user-serviceable parts inside this instrument. TSI
recommends that you return the DUSTTRAK monitor to the factory
for any required maintenance or service not described in this
manual.
Maintenance Schedule
Your Model 8520 DUSTTRAK Aerosol Monitor requires maintenance on a
regular basis. Table 4–1 lists the factory recommended maintenance
schedule.
Some maintenance items are required each time the DUSTTRAK monitor is
used or on an annual basis. Other items are scheduled according to how
much aerosol is drawn through the instrument. For example, cleaning the
inlet sample tube is recommended after 350 hours of sampling a 1 mg/m3
concentration of aerosol. This recommendation should be pro-rated
according to how the instrument is used. 350 hours at 1 mg/m3 is the same
amount of aerosol as 700 hours at 0.5 mg/m3 or 175 hours at 2 mg/m3, etc.
Chapter 4
38
When using the 1.0 or 2.5 μm inlet nozzles you need to clean the inlet more
frequently than for the standard 10 μm nozzle (black) because of the much
higher percentage of particles being removed from the stream. Table 4–1
represents estimates of how long you can run before cleaning is necessary.
The actual time you can run between cleanings varies with the size
distribution of particles you are measuring.
Table 4–1. Recommended Maintenance Schedule
Item Frequency
Perform zero check Daily (or before long tests)
Clean 1.0 μm nozzle 5 hr. at 1 mg/m3*
Clean 2.5 μm nozzle 30 hr. at 1 mg/m3*
Clean 10 μm nozzle, inlet, and sample tube 350 hr. at 1 mg/m3*
Replace internal filters 700 hr. at 1 mg/m3*
Clean cyclone Before each use
Return to factory for cleaning and
calibration Annually
*Pro-rated, see discussion above.
The DUSTTRAK monitor keeps track of the accumulated amount of aerosol
drawn through it since its last cleaning. When sample tube cleaning or
internal filter replacement is due, the display shows the message
“SERVICE 4” or “SERVICE 5,” respectively, during power-up. Press any
key to bypass the message, but be sure to perform the maintenance
procedures at your earliest convenience.
TSI recommends that you perform a zero check each day the DUSTTRAK
monitor is used, before running any extended tests, and after the instrument
experiences a significant environmental change. Examples of significant
environmental changes would be ambient temperature changes that exceed
15 °F (8 °C) or moving from locations with high aerosol concentrations to
low concentrations.
Zero Checking/Re-Zeroing
1. Put the DUSTTRAK monitor in Survey mode.
2. Put zero filter on aerosol sample inlet. See Figure 4–1.
3. Set the time-constant to 10 seconds. Press and hold the TIME
CONSTANT key until “10” is displayed, then release.
4. Wait 10–60 seconds for displayed values to settle to zero.
5. If the displayed value is between -0.001 and +0.001 mg/m3, the
DUSTTRAK monitor does not need adjustment. If the displayed value
exceeds this limit, follow steps 7 to 9 below to re-zero the instrument.
Maintenance 39
Note: Negative mass concentration readings are an indication that
the DUSTTRAK monitor needs to be re-zeroed. The negative
reading of –0.001 in step 5 above is the only time when a
negative reading is acceptable.
Figure 4–1: Zero Checking the DUSTTRAK Aerosol Monitor
6. Press and hold the CALIBRATE key and wait for the displayed
countdown to reach 0, then immediately release the key. The message
“CALIBRATE ZERO” is displayed… if not, try again.
7. Press the SAMPLE key and wait for the 60-second countdown. When
the countdown is completed, the current calibration constant will be
displayed.
8. Press the CALIBRATE key again to return to survey mode. The re-
zeroing process is now completed.
Cleaning the 1.0 µm and 2.5 µm Inlet Conditioners
1. Turn the DUSTTRAK monitor off.
2. Remove the inlet nozzle and impactor plate.
3. Wipe the old grease and other residue off from the impactor plate and
nozzle. Once the nozzle and impactor plate are removed from the
DUSTTRAK monitor, compressed air can be used to blow them off. Do
not blow directly into the DUSTTRAK inlet. You can use isopropanol
or soap and water to clean the parts, but be sure they are dry before
reassembly.
4. Inspect the inlet for residue and clean as necessary.
5. Reapply grease to the impactor plate.
6. Reassemble the inlet conditioners.
Chapter 4
40
Cleaning the 10 µm Inlet and Sample Tube
The aerosol inlet needs to be cleaned regularly. The inlet nozzle should be
cleaned based on the schedule in Table 4–1 above or especially when
changing from very high concentrations of aerosols to low concentrations.
The “SERVICE 4” message will be displayed during power-up when nozzle
cleaning is due. The DUSTTRAK monitor pauses when any of the service
messages come on. Press any key to bypass the message and continue. The
“SERVICE 4” message is a one-time indicator to remind you that inlet port
cleaning is due. This message will automatically reset.
1. Move to a relatively clean environment. (Aerosol concentration less
than 0.1 mg/m3.)
2. Turn the DUSTTRAK monitor off.
Figure 4–2: Exploded View of Inlet Assembly
3. Remove the sample inlet nozzle and O-ring. The O-ring will usually
stay inside the nozzle.
4. Use the sample tube removal tool to unscrew the tube from the inside of
the inlet port. The cross-pattern on one end of the tool mates with the
end of the sample tube.
5. Use the hook end of the tool to pull the sample tube out of the inlet port.
6. Clean the sample tube and inlet port. A cotton swab can be used to clean
the inside of the inlet port, cap, O-ring, and ends of the tube. The swabs
can be dampened with water or a light solvent (e.g., isopropanol).
Note: Do not use compressed air to clean the inlet area. Compressed
air forces particulates into the optics chamber and
contaminates the optics.
Maintenance 41
The inside of the sample tube can be cleaned using the sample tube
cleaning brush provided, along with a light solvent. Dry the tube by
blowing it out with compressed air, or let it air-dry thoroughly. Be
careful not to blow particles into the DUSTTRAK monitor inlet port.
7. Insert the sample tube back into the port and screw it in using the
sample tube removal tool.
8. Replace the inlet nozzle and O-ring.
9. Perform a zero check and re-zero if necessary.
10. Turn the DUSTTRAK monitor off and turn it on again while pressing and
holding the CALIBRATE key. This action resets the internal circuit that
keeps track of the total aerosol quantity that has gone through the
instrument. The inlet port cleaning procedure is now completed.
Replacing the Internal Filters
The internal filters should be replaced based on the schedule in Table 4–1 or
when the “SERVICE 5” is displayed during power-up. This “service
reminder” will appear each time at power-up, until you reset this message.
To reset the “SERVICE 5” reminder, do the following:
1. Turn the instrument off.
2. Press and hold the CALIBRATE key while turning on the instrument.
3. The service message is now reset.
If you change the filters before the “SERVICE 5” message appears, you
should reset this reminder by using the same technique.
The operation of the built-in total mass tracking circuit is dependent upon
the instrument calibration and the instrument flow rate. The assumed flow
rate is 1.7 L/min.
To replace the internal filters, follow the procedure below. Refer to
Figure 4-3.
1. Move to a relatively clean environment (aerosol concentration less than
0.1 mg/m3).
2. Turn the DUSTTRAK Aerosol Monitor off.
Chapter 4
42
Figure 4–3: Internal Filter Replacement
3. Turn the aerosol monitor over on its keypad so the filter wells are facing
up.
4. Use the small flat-blade screwdriver to unscrew the retaining screws at
the center of each filter well. The screws are “captive” and will stay in
the caps.
5. Grip the screw heads between your thumb and forefinger. Lift to pull
the filter well caps off. There is an O-ring under each cap that may lift
out with the cap.
6. Pull out the internal filter elements and discard them.
7. Clean the inside of the filter wells if necessary with a mild solvent like
isopropanol and a cotton swab. Do not use compressed air.
8. Replace the internal filter elements with TSI part no. 800666.
Note: Only the recommended high efficiency particulate air (HEPA)
filters should be used. The DUSTTRAK monitor should never be
run without filters in place. Doing so will contaminate the
optics and probably require you to send the unit back to the
factory for cleaning.
9. Replace the O-rings and caps.
10. Use the small screwdriver to tighten the hold-down screws until “finger-
tight.” Do not over-tighten.
Maintenance 43
11. Zero the mass tracking circuit to turn the service message off. This is
done by pressing and holding the CALIBRATE key during power-up.
The procedure is now completed.
Setting the Flow Rate
The sample flow rate through the DUSTTRAK monitor is set at the factory at
1.7 L/min. The flow rate can be easily adjusted if necessary by following the
procedure below. You need an instrument that precisely measures flow rate
for use as a reference such as the flowmeter supplied with the DUSTTRAK
monitor.
Note: The mass concentration readings are not affected by flow rate
changes because the volume of sampled air “seen” by the
DUSTTRAK detector is fixed. Flow rate is critical when using the
cyclone or other aerosol separation devices.
1. Clean the inlet port as described in this chapter. If necessary, replace the
internal filters as described in this chapter.
2. Connect the flowmeter to the DUSTTRAK monitor sample port using the
tubing provided.
3. With the DUSTTRAK monitor running, use the supplied screwdriver to
turn the flow adjustment screw located on the back of the DUSTTRAK
monitor (see Figure 4–4). Observe the reading on the flowmeter
(reading is taken at the center of the ball) and adjust the flow until the
desired reading is obtained. The factory setting is 1.7 L/min.
4. Disconnect the flowmeter. The procedure is now completed.
Figure 4–4: Adjusting the Flow Rate
Chapter 4
44
Cleaning the Cyclone
The 10-mm Nylon Dorr-Oliver Cyclone should be cleaned prior to each use.
In most cases, simply cleaning the grit pot will be all that is needed. Inspect
the inside of the cyclone body regularly and clean it if necessary.
1. Detach the cyclone from the DUSTTRAK monitor.
2. Unscrew the grit pot from the bottom of the cyclone and pull the top off
the cyclone by pulling on the plastic tubing.
3. Hold the open end of the grit pot down and tap it on a hard surface to
dislodge particles. Repeat with the cyclone body.
Note: If dirt is visible inside either the grit pot or the cyclone body, it
may be necessary to blow compressed air into the cyclone
parts and/or to clean them with soap and water. A mild solvent
like isopropanol may also be used. Make certain that the
cyclone is perfectly dry before using it.
4. Re-assemble the cyclone. Refer to Figure 4–5 below. The cyclone
cleaning procedure is now completed.
Figure 4–5: Exploded View of 10 mm Nylon Dorr-Oliver Cyclone
Storage Precautions
When storing the DUSTTRAK monitor for more than 30 days, you should
remove the batteries. This prevents damage due to battery leakage.
This instrument must be stored in a location where the temperature remains
between –20 and 60°C (–4 and 140°F).
45
Chapter 5
Troubleshooting
The table below lists the symptoms, possible causes, and recommended
solutions for common problems encountered with the DUSTTRAK monitor.
Symptom Possible Cause Corrective Action
Erratic zero
reading. Leak.
Dirty inlet port and/or
sample tube.
Sample nozzle missing.
Internal filter(s) not
installed properly
(leaking).
Check connections for leaks.
Replace zero filter.
Tighten inlet cap.
Clean inlet port. Clean or
replace tubing.
Remove inlet port cap and
check sample nozzle. Replace
if missing.
Inspect internal filter wells to
make certain the filters and o-
rings are seated properly.
Replace internal filters if
necessary.
No display. Unit not switched on.
Low or dead batteries.
Batteries installed
backwards.
Dirty battery contacts.
Switch unit on.
Replace the batteries or plug in
the AC adapter.
Observe polarity indication on
battery holder diagram.
Clean the battery contacts.
No keypad
response. Display/Lockout switch
in lockout position. Slide lockout switch to normal
position (to the right).
Incorrect
function
displayed.
Two keys have been
pressed at the same time. Press only one key at a time.
Chapter 5
46
Symptom Possible Cause Corrective Action
Battery
symbol is
displayed.
(constant or
blinking)
Low battery charge.
Incorrect AC adapter.
Low AC line Voltage
Dirty battery contacts.
Replace batteries or use AC
adapter.
Replace with the correct AC
adapter.
Correct the AC line voltage or
use batteries.
Clean the battery contacts.
Analog
output does
not work
Cable/connector not
correctly installed.
Output wired with reverse
polarity.
Make sure cable connector is
fully seated.
Make sure analog out (+) and
analog ground (-) are wired
correctly to data-logger.
Analog
output is not
in proportion
to display
Analog output range in
DUSTTRAK monitor may
be set incorrectly.
Data logger scaling factor
may be set incorrectly.
Program analog output to
desired range:
0 – 0.100 mg/m3
0 – 1.00 mg/m3
0 – 10.0 mg/m3
0 - 100 mg/m3
Scaling factor must be
consistent with DUSTTRAK
monitor.
Alarm
output does
not work.
Alarm does
not turn on
correctly.
Alarm function not turned
on.
Alarm setting incorrect.
Cable/connector not
correctly installed.
Alarm output wired with
reverse polarity.
Alarm function is “off” by
default. It must be programmed
“on” (using TRAKPRO
software).
Program alarm setting using
TRAKPRO software.
Make sure cable connector is
fully seated.
Alarm wires are polarized.
Voltage input must be wired to
alarm input (+).
Maintenance 47
Symptom Possible Cause Corrective Action
Display
reads
“MEMORY
0%.”
Memory is full. Download memory to a PC if
desired, then clear memory.
SERVICE
and “1”
displayed.
Memory has been cleared
due to temporary loss of
power from internal
backup battery.
Factory service required if
condition persists. Press any
key to bypass.
SERVICE
and “2”
displayed.
Calibration memory error.
Internal calibration data
corrupted.
Factory service required.
SERVICE
and “3”
displayed.
Backup battery low. Factory service required. Press
any key to bypass.
SERVICE
and “4”
displayed.
The sample inlet nozzle is
due for cleaning. Clean the nozzle and sample
inlet port. Press any key to
bypass.
SERVICE
and “5”
displayed.
Internal filters are due for
replacement. Replace internal filters and
reset service reminder. See
“Replacing the Internal Filters”
for more information.
SERVICE
and “6”
displayed.
Flow obstruction on
previous power up.
Internal pump failing,
indicated by inability to
adjust flow rate to full
range.
Remove obstruction if still
present. Press any key to
bypass.
If Service 6, recurs on each
power up even after bypassing,
factory service may be
required.
Chapter 5
48
Symptom Possible Cause Corrective Action
SERVICE
and “7”
displayed.
High humidity or laser
failure. Typically occurs if the unit is
turned ON either after sampling
under high humidity conditions
or when instruments are stored
at a different temperature than
that of the sampling area.
Bypass this error by pressing
“Calibrate” key. Zero the
DUSTTRAK by following the
instructions given in Chapter 4,
Maintenance,” under section
Zero Checking/Re-Zeroing”.
If the unit cannot be zeroed, let
the DUSTTRAK equilibrate to
ambient conditions and try
zeroing the DUSTTRAK again. If
the Service 7 error still persists,
it might be due to laser failure
and the unit needs to be
returned to TSI.
49
Appendix A
Specifications
Specifications are subject to change without notice.
Sensitivity:
Sensor type:..................... 90° light scattering, laser diode
Range: ............................. 0.001–100 mg/m3 (calibrated to respirable
fraction of standard ISO 12103-1, A1 test
dust)
Resolution: ...................... ±0.1 % of reading or ±0.001 mg/m3,
whichever is greater
Zero stability:.................... ±0.001 mg/m3 over 24 hours using
10-second time-constant
Particle size range: .......... 0.1–approximate 10 µm (Upper limit is
dependent on flow rate)
Flow rate range:............... 1.4–2.4 L/min.
Temperature coefficient: .. +0.001 mg/m3 per °C (for variations from
temperature at which DUSTTRAK monitor
was zeroed)
Instrument Temperature Range:
Operating range: ............. 0 °C to 50 °C (32 °F to 120 °F)
Storage range: ................ -20 °C to 60 °C (-4 °F to 140 °F)
Time Constant:
Range:.............................. Adjustable from 1 to 60 seconds
Data Logging:
Data points:...................... >31,000 (21 days logging every minute)
Logging interval:............... Adjustable from 1 second to 1 hour
Analog Output (selected instruments only):
Selectable ranges: ........... 0 – 0.100 mg/m3
0 – 1.0 mg/m3
0 – 10.0 mg/m3
0 – 100 mg/m3
Voltage output.................. 0 – 5 VDC
Output impedance............ 0.01 ohms
Maximum current ............. 15 mA
Appendix A
50
Alarm Output (selected instruments only):
Range............................... 0.002 to 100 mg/m3
Maximum voltage input .... 15 VDC
Maximum current ............. 1 amp
Physical:
External dimensions:........ 221 mm x 150 mm x 87 mm (8.7 in x
5.9 in x 3.4 in)
Weight (with batteries): ... 1.5 kg (3.3 lb.)
Display:............................. 4 digit LCD, 15 mm (0.6 in) digit height
Factory Maintenance:
Factory clean/calibrate:.... Annually
User Maintenance:
User calibration: ............... As needed
Clean nozzle: ................... After every 350 hours at 1 mg/m3
(e.g., 350 hours at 1 mg/m3, 35 hours at
10 mg/m3 or 3.5 hours at 100 mg/m3)
Change internal filters: ..... After every 700 hours at 1 mg/m3
(e.g., 700 hours at 1 mg/m3, 70 hours at
10 mg/m3 or 7 hours at 100 mg/m3)
Serial Interface:
Type: ................................ RS-232
BAUD rate: ....................... 1200
Data Bits:.......................... 8
Stop Bits:.......................... 1
Data Format: .................... ASCII
Power Requirements:
External Adapter: ............. 6 VDC, 300 ma
or
Batteries: .......................... 4 C-size Alkaline
Battery run-time:............... 16 hours (typical)
EMI/RF Immunity:
1. Complies with Emissions Directive Standard: EN50081-1:1992
2. Complies with Immunity Directive Standard: EN50082-1:1992
51
Appendix B
Model 8520-1 DUSTTRAK Environmental
Enclosure Operation and Maintenance Manual
Unpacking the DUSTTRAK Environmental Enclosure
Carefully unpack the Model 8520-1 DUSTTRAK Environmental Enclosure
from the shipping container. Use the tables and illustrations below to make
certain that there are no missing components. Contact TSI immediately if
anything is missing or damaged.
Table B–1: Model 8520-1 DUSTTRAK Environmental Enclosure
Quantity Item Description Part/Model
1 Model 8520-1 DUSTTRAK
Environmental Enclosure Case 8520-1
1 7.5 V Universal Input Battery Charger 1208033
2 6 V, 10 A-Hr Lead/Acid Battery 801563
1 Aerosol Inlet Assembly 801565
1 Water Trap Bottle 2002032
2 Velcro Strap 2913125
2 Tygon Plastic Tubing, 1/4 in ID 3001252
1 DB9 Connector Dust Caps 1304132
1 Transportation Inlet Port Plug 2602013
1 Tripod Mount Screw Plug 5010005
1 Model 8520 DUSTTRAK Aerosol
Monitor Operation and Service Manual 1980198
Spare Parts
The following kit contains replacement parts that may be needed if the
standard parts are lost or destroyed. It contains O-rings, storage caps (DB9
and inlet) and replacement tubing.
Quantity Item Description Part/Model
1 Environmental Enclosure Spare parts 801566
Appendix B
52
Parts Identification: Environmental Enclosure
Figures B–1and B–2 identify the parts of the Model 8520-1 DUSTTRAK
Environmental Enclosure. Please become familiar with these components
before proceeding.
Figure B–1: Inside and Outside Views of
Environmental Enclosure
1. Inlet Assembly (in storage) 5. Feed Thru DB9 Adapter
2. DUSTTRAK/Battery Bracket 6. Inlet Assembly (sampling)
3. Water Trap Bottle 7. Inlet Ring
4. Inlet Mounting Brackets
4
1
3
2
7
5
6
Setting-Up 53
Figure B–2: DUSTTRAK Environmental Enclosure Accessories
1. 6 V-10 Ah Battery
Pack (X2)
2. 7.5 V Universal
Input/Dual Step Charger
3. Aerosol Sampling Inlet
4. Respirable Sampling
Inlet (optional)
5. Water Trap Bottle
6. Velcro Straps
7. DB9 Dust Cap
8. Transport Inlet Plug
9. Tripod Mount Screw Plug
10. Tubing
11. Manual (not shown)
Appendix B
54
Setting-Up
The setup of the Environmental Enclosure is an important part in allowing
reliable and accurate sampling of aerosols in a wide range of conditions. TSI
cannot ensure accurate measurements if any of the components are set up
incorrectly. Damage to the enclosure or its components can result if these
procedures are not followed.
WARNING
!
The DUSTTRAK monitor is not rated for intrinsic safety. The
DUSTTRAK monitor, with the Environmental Enclosure, must
never be operated under conditions where there is a risk of fire
or explosion.
Caution
!
The enclosure is designed to be water resistant to rain or spray.
It is not designed to be waterproof when immersed. Setting it in
a pool of water will result in flooding the inner compartment with
water. This will severely damage both your DUSTTRAK monitor
and battery pack. Do not set the Environmental Enclosure in
water!
Installing the Aerosol or Respirable Aerosol Inlet
There are two inlet assemblies, either of which may be stored inside the lid
of the Environmental Enclosure, allowing ambient aerosols to be sampled
efficiently in a wide range of conditions. Proper installation into its sampling
position ensures that the DUSTTRAK monitor measures an accurate ambient
aerosol mass concentration. Follow these instructions to install either of the
aerosol inlets:
Hold the inlet assembly by the knurled section on the tube and remove it
from the holding bracket by unscrewing it (see Figure B-3).
Caution
!
Always hold the inlet assembly by the knurled portion of the
tube. Do not twist on the inlet assembly cap.
Carefully slide the inlet through the upper mounting bracket and remove
the inlet from the box.
Remove the Inlet Transport Plug from the top of the box.
Setting-Up 55
Figure B–3: Unscrew Inlet from Inner Bracket
Thread the inlet assembly into upper inlet ring found on the top of the
Environmental Enclosure (See Figure B-4).
Figure B–4: Thread the Inlet into the Inlet Ring
Notes: Make sure that a thin film of vacuum grease is coating the O-ring
on the inlet tube to ensure a good seal before installing it.
Do not use the Respirable Aerosol Inlet in the rain. As it will
aspirate rainwater into the cyclone body.
Appendix B
56
Install Water Trap Bottle
The translucent bottle that attaches to the bottom of the inner inlet is used to
collect any water that is drawn into the sampled flow. This prevents water
from reaching the DUSTTRAK monitor and damaging it. The bottle is
installed on the Environmental Enclosure when it is shipped. If the bottle is
removed for cleaning, follow these precautions when re-installing:
Before screwing the bottle into inner inlet ring, be sure that the O-ring
found in the bottom of the threaded hole is greased. This ensures an air-
tight seal.
When threading the bottle into the inlet be careful not to damage its
threads. Figure B-5 demonstrates how to install the water trap correctly.
Make sure the bottle has been tightened securely.
Figure B–5: Installing the Water Trap Bottle
Setting-Up 57
Correct Placement of Velcro Straps
The battery and the DUSTTRAK monitor are held in place by Velcro straps.
One end of each strap is secured into place. Figure B-6 shows the correct
placement of each strap (before the DUSTTRAK monitor and battery pack are
installed).
Figure B–6: Placement of Velcro Straps
Install DUSTTRAK Monitor into Environmental Enclosure
To correctly install the DUSTTRAK monitor follow the instructions below:
Slip the free end of the Velcro strap through the slot on the DUSTTRAK
monitor portion of the main bracket (see Figure B-7).
Figure B–7: Slip Velcro Strap Through Slot in Bracket
Appendix B
58
Slide the DUSTTRAK monitor under the Velcro and into the bracket with
the DUSTTRAK monitor inlet towards the center of the case (see
Figure B-8).
Figure B–8: Sliding DUSTTRAK Monitor into the Bracket
Tighten the Velcro Strap around the DUSTTRAK monitor (see
Figure B-9).
Figure B–9: Tighten the Velcro Strap
Setting-Up 59
The DUSTTRAK monitor is now safely secured. When access is needed to the
keypad covered by the Velcro strap, loosen it to expose them. Be sure to
reattach the strap when done using the keypad.
Connecting Tubing
A length of tubing is provided to transport the aerosol from the external inlet
to the DUSTTRAK inlet. Before attempting to sample, make sure that this
tubing is securely fastened to the barbs from both of the inlets. When closing
the enclosure, make sure that this tube does not get kinked. The best way to
prevent this is to pull the tube towards the front of the box as you close it
(see Figure B-10).
Figure B–10: Keeping the Tube Unkinked
Supplying Power to the DUSTTRAK Aerosol Monitor
The Model 8520 DUSTTRAK Aerosol Monitor may be powered in one of two
ways when used with the 8520-1 Environmental Enclosure:
1. Using the standard internal battery pack (consisting of 4, C size
batteries)
2. Using the supplied external battery pack.
The internal battery pack will supply power for approximately 16 hours of
operation (may be less under cold conditions). This is adequate for single
shift (8-hour) monitoring, but will not work for 24-hour operation.
Appendix B
60
Caution
!
TSI does not recommend using the standard AC adapter when
operating outside. This device is not rated for external
exposure.
Using the External Battery Pack
The battery packs supplied with the Environmental Enclosure allow for
extended use of the DUSTTRAK monitor. These battery packs enable the
DUSTTRAK monitor to operate for at least 24 hours (even in cold
conditions). The battery packs should be rotated and properly charged
after each use, to ensure a prolonged lifetime (see Maintenance section
for further information). Make sure that the internal C size batteries are
installed correctly within the DUSTTRAK monitor to make it possible to
continue sampling when changing the external battery packs.
Install External Battery Pack
To correctly install the Battery Pack follow the instructions below:
Slip the free end of the Velcro strap through the slot on the battery
pack portion of the main bracket (See Figure B-11).
Figure B–11: Slip Velcro Strap Through Slot.
Slide the battery pack into the correct space in the sheet metal
bracket such that the leads are oriented up, and the PC board is in
the upper left hand quarter.
Tighten the Velcro strap around the battery pack (See Figure B-12).
Setting-Up 61
Figure B–12: Tighten Velcro Around Battery.
Connect the external battery pack to the DUSTTRAK monitor by
inserting the power plug protruding from the battery to the power
jack in the top of the DUSTTRAK monitor.
Mounting to a Surveyor Tripod
The Environmental
Enclosure may be
mounted to a standard
surveyor tripod equipped
with a 58”-11 threaded
stud. Mounting the
Enclosure on a tripod is
useful when doing
outdoor site monitoring.
Figure B-13 shows the
Environmental Enclosure
mounted on a surveyor
tripod.
Figure B–13: Enclosure
Mounted on a Tripod.
Caution
!
Do not place the Environmental Enclosure in direct contact
with the ground. The bottom of the Environmental Enclosure
contains an exhaust port that may allow water to enter the
enclosure, resulting in damage to the DUSTTRAK monitor.
Appendix B
62
Downloading a Data File
To download a file under normal data-logging operations, you would do the
following:
Open the Environmental Enclosure.
Discontinue sampling.
Attach the supplied RJ45 to the RS-232 cable from the DUSTTRAK
monitor to a notebook computer.
Download the data file using TRAKPRO software (refer to Chapter 2 of
this manual for further details).
Real-Time Monitoring
The Environmental Enclosure comes equipped with an external DB9
connector. This allows the DUSTTRAK monitor to be connected to a
computer for continuous real-time monitoring. To set up the Environmental
Enclosure for real-time monitoring follow these instructions:
Connect the DUSTTRAK monitor to the DB9 connector, which is
mounted on the inside of the enclosure, using the Computer Interface
Cable provided with the DUSTTRAK monitor.
Attach a standard 9-pin serial cable (not supplied) from the DB9
connector on the outside of the enclosure to a computer serial port. The
computer can be configured to query the DUSTTRAK monitor when in a
terminal emulator mode. See Application Note ITI-044 for more
information on these commands.
Setting-up TRAKPROTM Data Analysis Software
Information on setting-up and using TRAKPRO software can be found in
Chapter 2 of this manual.
Setting-Up 63
Transporting the Environmental Enclosure
When transporting the Environmental Enclosure it is important to store its
components correctly. The following steps should be taken to ensure that no
damage is done to the enclosure or its components during transportation:
Attention
!
The Environmental Enclosure is designed to protect equipment
during stationary sampling only.
Do not transport or ship equipment inside the Environmental
Enclosure.
Unscrew the aerosol inlet from the top of the case, and install it into the
inlet-mounting bracket inside the lid of the enclosure.
Insert the Inlet Transport Plug into the opening in the inlet ring in the
top of the box.
Empty any water from the water trap.
Remove DUSTTRAK monitor and battery pack.
If mounted on a tripod, take the enclosure off the tripod and screw in the
tripod mount plug.
Wiring the Analog/Alarm Cable in the Enclosure
You may desire to install the Analog/Alarm cable to the DUSTTRAK monitor,
when operated inside the Environmental Enclosure. The serial connector,
installed in the side of the enclosure, may be used for this purpose.
Note: If additional penetrations are made into the wall of the enclosure,
you should take precautions to install a sealing (water-tight)
connector around the cable.
64 Appendix B
Operation
Overview
The DUSTTRAK Environmental Enclosure can be used in conjunction with
the DUSTTRAK monitor for many different applications. Its primary use is in
outdoor applications to give the DUSTTRAK monitor protection from the
elements and the ability to sample efficiently in different wind speeds. The
enclosure and extended-life battery may also be advantageous in indoor
industrial applications, to provide additional security and protection to the
DUSTTRAK monitor.
How to Properly Orient the Environmental Enclosure
The Environmental Enclosure should be set up in a location where it can
sample the particles of interest. It should be placed “out in the open,” away
from obstructions which may affect wind currents. It should not, for
example, be placed at the corner of a building, which would cause swirling
wind currents and result in poor particle sampling.
The Environmental Enclosure should be used in wind conditions with speeds
of 22 mph or less to obtain the most accurate readings. An increase in wind
speed over 22 mph can decrease the sampling efficiency of the inlet to under
the efficiency specified by PM-10 standards. If wind gusts of over 22 mph
are present, the data collected is still valid, but be aware that the readings
will be slightly lower than the actual mass concentration of aerosol present.
See Application Note ITI-060 for more information on the wind speed
dependency of the external inlet’s sampling characteristics.
Changing and Re-charging the Battery Packs
The battery pack within the Environmental Enclosure should be changed
every 24 hours of use during the daily maintenance check described in the
Maintenance Section. A voltage cutoff switch will cut the power to the
DUSTTRAK monitor when the battery voltage reaches 5.1 V. This ensures
that the battery does not become deeply discharged. Preventing this
condition will prolong the life of the battery.
Receiving Data from the DUSTTRAK Aerosol Monitor
The DUSTTRAK monitor records data in either the manual sampling mode
(LOG 1) or the programmable modes (LOG 2, LOG 3). See the DUSTTRAK
User Manual for more information on setting up the logging modes. The
data can be downloaded by opening the Environmental Enclosure lid,
attaching the supplied RJ45/RS-232 cable from the DUSTTRAK monitor to a
computer COM port, and using the RECEIVE DATA command within the
TRAKPRO software.
Operation 65
If desired, the DUSTTRAK RJ45 cable may be attached directly to the inside
of the DB9 connector located on the side of the Environmental Enclosure. A
serial cable (not supplied) may then be attached from the outside of the DB9
connector to the computer COM port.
Note: In order to download a data file the DUSTTRAK monitor cannot be
in a sampling mode.
The primary use for the external DB9 connector is to facilitate real-time data
monitoring. An external computer or data logger can be directly attached to
the DUSTTRAK monitor, while still keeping the instrument inside its
weatherproof enclosure. The DUSTTRAK monitor can be queried through a
set of ASCII commands to provide periodic information.
Zeroing the DUSTTRAK Aerosol Monitor
The DUSTTRAK monitor should always be zeroed before beginning a
sample. See the instructions in Chapter 3 of this manual for more
information. If possible, the instrument should be zeroed under stable
ambient temperature conditions (since variations in temperature will have a
small impact upon the DUSTTRAK monitor readings). If this is not possible
(for example, 24-hour outdoor sampling, with wide temperature swings) the
instrument should be kept closed inside the Environmental Enclosure. The
thermal mass of the Enclosure will dampen out the affect of temperature
swings, radiant heat, etc.
Note: See the DUSTTRAK monitor Specifications in Appendix A for more
information on the temperature dependence.
In nearly all cases, the temperature impact upon the DUSTTRAK monitor
reading is very small in comparison to the ambient particle concentration.
Setting the Flow Rate
The aerosol inlet on the Environmental Enclosure was designed to operate at
a 1.7 liters/minute flow rate. Before each sample, the flow rate should be
verified and/or adjusted as described in the DUSTTRAK monitor Maintenance
section (Chapter 4).
Locking the Environmental Enclosure
To avoid theft of the DUSTTRAK monitor, when left inside the Environmental
Enclosure, a padlock should be used to lock the box. There are two holes on
the front of the box that allow for a standard padlock to be attached. A
padlock is not supplied with the 8520-1.
Appendix B
66
Checklist for Sampling with the Environmental Enclosure
Before beginning a sample check to see that all of the following conditions
are satisfied:
The flow rate is set to 1.7 liters/minute.
All components are properly installed into the enclosure as described in
the Setup section.
The DUSTTRAK monitor has been zeroed at the temperature at which it
will be sampling (if possible).
The Environmental Enclosure has been put in a place clear of any
obstructions that will affect the flow around the enclosure (putting it on
a tripod is optional).
The Enclosure is not resting directly on the ground (no standing water).
The DUSTTRAK monitor has been set to the appropriate logging mode.
The Enclosure is locked shut to prevent theft or vandalism to
instrument.
Plans have been made to check for maintenance and data collection
every 24 hours after initial set up.
After the completion of monitoring, follow the procedures for transporting
the Environmental Enclosure in the Setting-Up section of this appendix. This
prevents the instrument from becoming damaged.
Maintenance 67
Maintenance
Daily Maintenance Checks
The DUSTTRAK monitor should be checked a minimum of once a day to
change the battery pack and make sure the instrument is operating properly.
A visual inspection of the instrument and case can ensure that the water trap
is empty, the sampling inlet is unobstructed and the tubing is not kinked.
Cleaning the Aerosol Inlet
Under normal operating conditions, the external aerosol inlet will remain
clean and unobstructed. Under very dirty conditions; however, the inlet may
become obstructed with vegetative debris or insects. To ensure the inside of
the aerosol inlet is clean, periodically blow compressed air through the
bottom of the inlet.
Caution
!
Do not disassemble the Environmental Enclosure aerosol inlet
for any reason. It is not designed for field service and
reassembly.
Cleaning the Respirable Aerosol Inlet
Follow the same procedure found in Chapter 4 of this manual in the section
entitled “Cleaning the Cyclone.”
When to Change the Battery
The battery packs are designed to provide power to the DUSTTRAK monitor
for a minimum of 24 hours, even under cold ambient conditions. To ensure
uninterrupted operation, the packs should be replaced daily. If a battery pack
has been forgotten and left connected to a running DUSTTRAK monitor, the
extended-life battery will automatically quit delivering current to the
DUSTTRAK monitor after its voltage drops below 5.1. At this point the
internal, C-cell battery pack automatically switches over and begins
powering the DUSTTRAK monitor with no loss of data.
Note: An internal battery backs up the DUSTTRAK monitor memory. The
logged data will not be lost even if both battery packs lose power.
Replacing the Battery Pack
When changing the battery pack, you do not have to interrupt the sampling
process as long as fresh batteries are installed in the DUSTTRAK monitor’s
internal battery compartment. Replace the discharged external battery with
the newly charged one, taking care to secure the fully charged battery with
the Velcro strap. When closing the case make sure that the tubing does not
become kinked. Pull the tubing in front of the battery bracket while closing
Appendix B
68
the case to prevent this from occurring. To prevent long term damage and
loss of capacity recharge the battery promptly after removing it.
Recharging the Battery Packs
The following steps should be followed to correctly recharge the battery
packs:
Disconnect the battery pack from the DUSTTRAK monitor and remove it
from the Environmental Enclosure.
Take it to a protected area where it can be charged undisturbed for 12 to
15 hours.
Check on the bottom of the charger to see that the red voltage switch is
set appropriately to either 115V or 230V.
Supply power to the charger. The green LED lights on the front of the
charger.
Insert the power plug protruding from the battery pack into the output
jack attached to the charger. The charger will now quick-charge the
battery. This is indicated by an illuminated yellow LED.
Charging is complete when the yellow LED shuts off. A trickle charge
will maintain the full charge. A fully discharged battery will require up
to 15 hours to fully recharge.
Note: Do Not Store A Battery Pack That Has Not Been Fully Charged.
Storing a discharged battery for any length of time will negatively
affect the battery life.
Battery Pack Life
Lead acid batteries have a finite life of about 300 charging cycles. This
means that the two battery packs last around two years without need of
replacement. This assumes that the batteries are being used to run a
DUSTTRAK monitor every day, and that they are properly cared for and
rotated daily. To help track battery lifetime, an In-service Date sticker has
been attached to the front of the battery pack to inform you when each
battery pack was shipped from TSI. Replacement battery packs may be
ordered from TSI. See the previous section of this appendix on “Parts
Identification” for more information.
Emptying the Water Trap
If any water or moisture has accumulated in the water trap, be sure to
unscrew the water trap from the internal inlet and empty it. Take care when
removing and reinstalling the bottle to prevent damage to the threads. The
bottle must be securely hand tightened to seal against the O-ring to prevent
leaking.
Maintenance 69
Storage Precautions
This instrument must be stored in a location where the temperature remains
between –20 and 60°C (–4 and 140°F).
Appendix B
70
Troubleshooting the Environmental Enclosure
The table below lists the symptoms, possible causes and recommended
solutions for common problems encountered with the DUSTTRAK
Environmental Enclosure.
Symptom Possible Cause Corrective Action
DUSTTRAK
monitor does
not turn on.
Uncharged battery.
External battery pack not
plugged to DUSTTRAK
monitor.
Bad battery pack (no
longer able to be
recharged).
Connect to charger and wait
until the yellow LED turns off.
Plug battery power plug into
the power jack in DUSTTRAK
monitor.
Replace with new battery pack.
Readings are
unusually
low.
Zero was not checked on
DUSTTRAK monitor.
Plugged external inlet.
Sampling in wind speeds
over 22 mph.
Enclosure located near an
object that obstructs the
flow.
Leak in the inlet.
Re-zero DUSTTRAK monitor at
desired sampling conditions.
Remove inlet from the
Enclosure; blow out debris with
compressed air. Rinse with
clean air, if needed. DO NOT
DISASSEMBLE COVER ON
INLET ASSEMBLY.
Take off water trap bottle and
clean all exposed surfaces.
DUSTTRAK monitor will under-
sample slightly in high wind
speeds. No correction possible.
Move the Environmental
Enclosure into a more open
area.
Make sure the upper inlet
assembly and the water trap are
screwed in tightly.
Make sure tubing is secured on
the enclosure barb and the
DUSTTRAK monitor.
Troubleshooting the Environmental Enclosure 71
Symptom Possible Cause Corrective Action
Readings are
unusually
low (cont.).
Leak in the inlet.
(cont.) Make sure inlet assembly
O-ring is in place.
Re-grease the O-rings sealing
the inlet assembly and the
water trap bottle.
DUSTTRAK
monitor
Display
reads
“MEMORY
0%”.
Memory is full. Download memory to a PC if
desired then clear memory.
73
Specifications: Environmental Enclosure
Specifications are subject to change without notice.
Sampling Conditions:
Operating Temperature: . 0 °C to 50 °C (32 °F to 120 °F)
Wind Speed: .................... 0 m/s to 10 m/s (0 mph to 22.5 mph)
Aerosol Concentration
Range: ............................. 0.001–100 mg/m3 (calibrated to respirable
fraction of standard ISO 12103-1, A1 test
dust)
Storage Temperature:...... -20 °C to 60 °C (-4 °F to 140
Zero stability:.................... ±0.001 mg/m3 over 24 hours using
10-second time-constant
Particle size range: .......... 0.1–approximate 10 µm (Upper limit is
dependent on flow rate)
Flow rate: ........................ 1.7 L/min.
Temperature coefficient: .. +0.001 mg/m3 per °C (for variations from
temperature at which DUSTTRAK monitor
was zeroed)
Physical:
External dimensions:........ 406 mm x 330 mm x 191 mm
(16 in. x 13 in. x 7.5 in.)
Weight (with battery and
DUSTTRAK monitor): ......... 7.2 kg (15.8 lb.)
USER MAINTENANCE:
Maintenance Check/
Data Collection:................ Daily
Clean inlet: ....................... Weekly or daily if concentrations of over
30 mg/cm3 are measured.
Re-grease O-rings: .......... As needed
Power Requirements:
External Battery Pack: ..... 6 VDC, 10 Ah
Battery run-time: .............. 24–60 hours (typical)
75
Index
A
AC adapter, 3, 5
accessories, 3
adapter, 1
adjusting flow rate, 43
advanced modes, 26
aero filter, 3
aerosol monitor, 1
accessories, 3
components, 2
connecting to computer, 6
keypad functions, 11
maintenance, 37
parts identification, 2
setting up, 5
specifications, 49
troubleshooting, 45
alarm output, 22
specifications, 23
alarm output function default, 23
alarm setpoint
programming, 24
screen, 24
alarm switch, 23
alarm threshold, 23
alkaline batteries, 1
analog output, 21
analog output range screen, 21
analog output/alarm connector, 1
analog output/alarm thumb screw, 2
analog/alarm cable, 3
analog/alarm connector pin-outs, 9
analog/alarm output connector
connecting, 9
arrow keys, 17
available memory, 27
available time constants
programming, 13
B
battery cover, 2
battery installation, 5
baud rate, 7
C
calibrate key, 17
calibrated data
converted stored data, 35
calibration certificate, 1
calibration factor
for specific aerosol, 32
setting from keypad, 33
carrying case, 1
channels, 27, 28
Class I, v
cleaning
cyclone, 44
inlet and sample tube, 40
inlet conditioner, 39
clear memory key, 15
communications port
setting up, 6
components, 2
computer cable, 1, 3
connecting analog/alarm output
connector, 9
connecting portable printer, 8
converting stored data to calibrated
data, 35
cotton swab, 1
custom calibrations, 31
cyclone, 3, 18, 44
cleaning, 44
respirable mass sampling, 18
cyclone holder clip, 2
D
data file
environmental enclosure, 62
data port, 2
determining calibration factor for
specific aerosol, 32
display, 2
display/keypad lockout switch, 17
connector, 2
Index
76
E
environmental enclosure, 51
aerosol inlet, cleaning, 67
aerosol monitor, 57
aerosol or respirable aerosol
inlet, 54
analog/alarm cable, 63
battery pack, 64
battery pack, recharging, 68
battery pack, replacing, 67
battery, charging, 67
checklist for sampling, 66
data file, 62
external battery pack, 60
flow rate, setting, 65
internal battery pack, 60
locking, 65
maintenance, 67
operation, 64
orienting, 64
parts identification, 52
power to aerosol monitor, 59
real-time monitoring, 62
receiving data, 64
respirable aerosol inlet,
cleaning, 67
setting up, 54
spare parts, 51
specifications, 73
storage precautions, 69
surveyor tripod, 61
transporting, 63
troubleshooting, 70
tubing, 59
unpacking, 51
velcro straps, 57
water traip, emptying, 68
water trap bottle, 56
zeroing, 65
exhaust port, 2
external power socket, 2
F
flow adjustment screw, 2
flow rate
adjusting, 43
setting, 43
flowmeter, 3
G–H
grease, 3
I–J
impactor plate, 1, 3
inlet assembly, 40
inlet conditioner, 19
cleaning, 39
inlet nozzle, 1, 3
inlet tube
cleaning, 40
installing batteries, 5
instrument setup, 6
internal filter
replacing, 41
internal filter element, 1
internal filter elements, 3
K
keypad, 2
functions, 11
L
log interval, 27, 28
log mode, 15
logging interval key, 15
logging intervals
programming, 16
M
maintenance, 37
environmental enclosure, 67
schedule, 37
manually setting real-time clock, 8
memory considerations, 31
Model 8520. (see aerosol monitor)
N
newcal, 33
nozzle
10 µm, 20
number of tests, 27, 28
number of tests logged, 27
Nylon Dorr-Oliver cyclone. See also
cyclone
Index 77
O
oldcal, 33
on/off key, 11
operation, 11
environmental enclosure, 64
options menu, 34
P–Q
packing list, 1
parts identification, 2
percent memory required, 27
plastic tubing, 1
portable printer, 8
pre-programmed samples, 30
print key, 14
programming
advanced modes, 26
alarm setpoint, 24
available time constants, 13
date/time, 8
R
real-time clock
manually setting, 8
setting, 7
real-time monitoring
environmental enclosure, 62
remote sampling, 17
replacing internal filter, 41
respirable mass sampling with
cyclone, 18
re-zeroing, 38
S
safety information, v
sample inlet nozzle, 2
sample key, 12
sample nozzle cleaning brush, 1
sample nozzle removal tool, 1
sample protocol, 29
sample tube
cleaning, 40
sample tube cleaning brush, 3
sample tube removal tool, 3
sampling mode key, 14
log mode, 15
survey mode, 14
sensitivity, 49
serial number, 27
serial port, 7
service policy, 2
setting custom calibration factor
keypad, 33
setting custom calibration factors
TRAKPRO software, 34
setting flow rate, 43
setting real-time clock, 7
setting up, 5
communications port, 6
environmental enclosure, 54
instrument, 6
TRAKPRO software, 6
shoulder strap, 1, 3
specifications, 49
alarm output, 23
environmental enclosure, 73
start date, 27, 28
start time, 27, 28
statistics key, 13
storage precautions, 44
survey mode, 14
T
test length, 27, 28
time between tests, 27, 28
time constant key, 12
TRAKPRO software
programming date/time, 8
programming logging intervals, 16
setting custom calibration
factors, 34
setting up, 6
troubleshooting, 45
environmental enclosure, 70
U–V
unpacking, 1
environmental enclosure, 51
upgrading, 24
using cyclone, 18
using inlet conditioners, 19
Index
78
W–X–Y
warning, v
warranty, 2
wrist strap, 1, 3
Z
zero checking, 38
zero filter, 1
TSI Incorporated
500 Cardigan Road, Shoreview, MN 55126 U.S.A.
Telephone: 1-800-874-2811 (toll-free) or 651-490-2811, Fax: 651-490-3824,
E-mail: answers@tsi.com
P/N 1980198 Rev. R Copyright © 2006 by TSI Incorporated All rights reserved