Transtek Medical Electronics TMB1014BT BLOOD PRESSURE MONITOR User Manual
Guangdong Transtek Medical Electronics Co.,Ltd BLOOD PRESSURE MONITOR
Contents
- 1. User Manual
- 2. Users Manual
User Manual
Version:1.0 User Manual Blood Pressure Monitor TMB-1014-BT Wrist Type Zhongshan Transtek Electronics Co., Ltd Jinâan Road, Minzhong, Zhongshan 528441, Guangdong, China Thank you very much for selecting TRANSTEK Blood Pressure Monitor TMB-1014-BT. To use the monitor correctly and safely, please read the manual thoroughly. Please well keep this manual in order to reference in future. CATALOGUE INTRODUCTION 1 Safety information LCD display signal Monitor components BEFORE YOU START 4 The Choice of power supply Installing and replacing the batteries Setting Date, time and measurement unit MEASUREMENT 7 Tie the cuff Strat a measurement THE OPERATION OF RECALLING RECORDS 9 Recall the records Delete the records INFORMATION FOR USER 11 Tips for operation Maintenances ABOUT BLOOD PRESSURE 13 What are systolic pressure and diastolic pressure? What is the standard blood pressure classification? Why does my blood pressure fluctuate throughout the day? Why the blood pressure I get from the hospital is different from home? If the result is the same if measuring on the right wrist? TROUBLESHOOTING 15 SPECIFICATIONS 16 CONTACT INFORMATION 17 Thank you for selecting TRANSTEK wrist blood pressure Monitor (TMB-1014-BT). The monitor features blood pressure measurement, pulse rate measurement and auto-save the result. The design provides you with many years of reliable service. Reading taken by the TMB-1014-BT are equivalent to those obtained by a trained observer using the cuff and stethoscope auscultation method. This manual contains important safety and care information, and provides step by step instruction for using the product. Read the manual thoroughly before using the product. INTRODUCTION Features: Systolic blood pressure Diastolic blood pressure Pulse rate Historic record of up to 60 measurements Safety information The below signs might be in the user manual, labeling or other component. They are the requirement of standard and using. Caution:Consult accompanying documents Type B applied part CE Mark: conforms to essential requirements of the Medical Device Directive 93/42/EEC. DISPOSAL: Do not dispose this product as unsorted municipal waste. Collection of such waste separately for special treatment is necessary. Manufacturer Specifies serial number Direct current EC REP Authorized Representative in the European Community CAUTION This device is intended for adult use only. This device is intended for no-invasive measuring and monitoring of arterial blood pressure. It is not intended for use on extremities other than the wrist or for functions other than obtaining a blood pressure measurement. Do not confuse self-monitoring with self-diagnosis. This unit allows you to monitor your blood pressure.Do not begin or end medical treatment based solely physician for treatment advice. If you are taking medication,consult your physician to determine the most appropriate time to measure your blood pressure. Never change a prescribed medication without consulting your physician. This unit is not suitable for continuous monitoring during medical emergencies or operations. If the cuff pressure exceeds 40 kPa (300 mmHg),the unit will automatically deflate. Should the cuff not deflate when pressures exceeds 40 kPa (300 mmHg),detach the cuff from the wrist and press the START/STOP button to stop inflation. To avoid measurement errors, carefully read this manual before using the product. The equipment is not AP/APG equipment and not suitable for use in the presence of a flammable anesthetic mixture with air of with oxygen or nitrous oxide. The operator shall not touch the battery and the patient simultaneously. The user must check that the equipment functions safely and see that it is in proper working condition before being used. The manufacturer does not require such preventive inspections by other persons. The Max. temperature of apply part can be achieved is 42.5â under the environmental temperature of 40â. Manufacture will make available on request circuit diagrams, component part list, etc. INTRODUCTION INTRODUCTION LCD display signal Monitor components SYSTOLIC MEM/UP BUTTON DIASTOLIC GRADE PULSE RATE LCD DISPLAY 6 < 0 % 2 / ' ( 6 & 5 ,3 7 ,2 1 START/OFF BUTTON TIME SET/DOWN BUTTON ( ; 3 /$ 1 $ 7 ,2 1 Systolic blood pressure High pressure result Diastolic blood pressure Low pressure result Pulse Pulse/minute Low battery Batteries are low and need to be replaced mmHg Measurement unit the blood pressure Grade The grade of the blood pressure For instructions,refer to Page 13 Arrhythmia Arrhythmia is detected Bluetooth Bluetooth is in an open position Error Blood pressure monitor has detected error User The current selected user Recalling The erecords will be showed Time Currently time CUFF (Type B applied part) BATTERY COMPARTMENT BEFORE YOU START BEFORE YOU START The installment and replacement of battery Setting Date, time and measurement unit Slide off the battery cover. Install the batteries by matching the correct polarity, as shown below. Always use the correct battery type (2 alkaline LR03AAA-size). Replace the cover. It is important to set the clock before using your blood pressure monitor, so that a time stamp can be assigned to each record that is stored in the memory. In sure the âlockâ key is glided to the unlock mode.(year :2010â2050; time:24 H)) 1.Pressing âSETâ for 3 seconds to enter the mode for [Hour] setting. MEM SET 2.Press the âMEMâ to change the [Hour]. MEM Replace the batteries whenever the below happen The shows SET The display dims The display does not light up CAUTION 3.When you get the right [Hour], press âSETâ to set down and turn to next step automatically. MEM Remove batteries if the device is not likely to be used for some time. The old battery is harmful to the environment, so please disposal with other daily trash. Remove the old battery from the device and follow your local recycling guidelines. SET BEFORE YOU START BEFORE YOU START Bluetooth phonesâ test software 4.Repeat the 2 and 3 to set the [Minute]. 1.Open the Bluetooth 4.0 special test MEM mobile phones after electrify blood pressure monitor. 2.After operating Bluetooth Test program (It is about 8 m from mobile to blood pressure monitor), the âdiscovered deviceâ column will remind âsearching for deviceâ... SET 5.Repeat the 2 and 3 to set the [YEAR], [MONTH] and [DAY]. MEM MEM 3 .Long press the "START" button after power off the blood pressure monitor, itâll show moving "oâ. SET SET 4.At this moment, the âdiscovered deviceâ column of mobile will display devices list. 7.After the unit is set, the right picture will show, then it turn off automatically. 5.Choose detected device from the list. 7KHUHLV³¼´RQWKHULJKW BEFORE YOU START 6.When the blood pressure monitor show â[ ]â, it means matching successfully; When it displayâE1â, it indicates matching unsuccessfully. 7.If matching successfully, it will show the lately measuring record on the âreceive dataâ column. 8.Choose âBloodpressure Recordâ to view the history record. MEASUREMENT Positioning the wrist cuff 1 .Remove all accessories (watch, bracelet, etc) from your left wrist. If your physician has diagnosed you with poor circulation in your left wrist, use your right wrist. 2.Roll or push up your sleeve to expose the skin. 3.Apply the cuff to your left wrist with your palm facing up. 4. Position the edge of the cuff about 1-1.5cm. 5. Fasten the wrist cuff around your wrist, leaving no extra room between the cuff and your skin. If the cuff is too loose, the measurement will not be accurate. Resting for 5 minutes before measuring. Wait at least 3 minutes between measurements. This allows your blood circulation to recover. For a meaningful comparison, try to measure under similar conditions. For example, take daily measurements at approximately the same time, on the same wrist, or as directed by a physician. 9.Click âbackâ to return main interface of testing software and go on testing next blood pressure monitor. MEASUREMENT MEASUREMENT Start the Measurement Please press the SET to choose the user 1 or user 2 when the blood pressure monitor turn off 1.Press the START/STOP, it will finish the whole measurement automatically. Adjust the zero automatically. Inflating and measuring automatically. Display and save the results automatically. START STOP 2.A f te r me as u rin gďź i tâ ll d is pl ay t he mo v in g âoâ a nd s en d dat a at th e s ame t ime. 3.If d ata i s s en t s u cc es s fu ll y, go b ac k to d is p la y th e i nte rfa c e o f me as uri ng re su lt s. If the data failed to sendďźgo back to display the interface of measuring results after display âE9â. 4.P re s s the S TA RT/ S TOP to power off, otherwise it will turn off automatically within 1 minute. START STOP THE OPERATION OF RECALLING RECORDS THE OPERATION OF RECALLING RECORDS Recalling the records 1.Press the âMEMâ to show the last 3 average of the record. Deleting a measurement record from memory MEM If you did not get the correct measurement, you can delete all results by following below steps. 1.Hold pressing âMEMâ for 3 MEM seconds, when the memory is SET shown. SET 2. Press the âMEMâ or âSETâ to get the record you want. MEM UP 2.When flash âDelete allâ, press the âSETâ to delete the memory. MEM SET DOWN SET CAUTION The most recent record (1) is shown first. Each new measurement is assigned to the first (1) record. All other records are pushed back one digit (e.g., 2 becomes 3, and so on), and the last record (60) is dropped from the list. 3.If you donât want to delete the records, press âS TA R T/ S TOPâ to escape. 4.If there is no record. the right display will show. INFORMATION INFORMATION FOR USER Tips for measurement Maintenance It can cause incorrectness if the measurement are taken in the following circumstances. Wait at least 1 hour after dinner or drinking Wait at least 20 minutes after taking a bath In a very cold enviroment Immediate measurement after tea, coffee, smoking In order to get the best performance, please follow the below instructions. Put in a dry place and avoid the sunshine Avoid touching water, clean it with a dry cloth in case. Avoid the intense shaking and collision Avoid the dusty and unstabletemperature environment Using the wet clothing to remove the dirt Avoid washing the cuff When talking or moving your fingers When you want to discharge urine Clean the cuff with a soft dry cloth. Do not use any abrasive or volatile cleaners. In order to reduce the pollution of the environment, please return to the manufacture when the product is close to the end of life. Instructions for correct replacement of interchangeable or detachable parts specified by MANUFACTURER as replaceable by SERVICE PERSONNEL. Please calibrate the blood pressure monitor in specific institute once every two years to ensure the precise measurementăIn order to get the best performance, please follow the below instructions for storage. ABOUT BLOOD PRESSURE ABOUT BLOOD PRESSURE What are systolic pressure and diastolic pressure? When ventricles contract and pump blood out of the heart, blood pressure reaches its maximum value, the highest pressure in the cycle is known as systolic pressure. When the heart relaxes between heartbeats, the lowest blood pressure is diastolic pressure. Systolic Diastolic blood discharging artery blood entering vein press relax What is the standard blood pressure classification? Why the blood pressure I get from the hospital is different from home? yellow light green light red light SYS 120 120~129 130~139 140~159 160~179 !180 DIA 80 80~84 85~89 90~99 100~109 !110 CAUTION Only a physician can tell you your normal blood pressure range and the point at which you are at risk. Consult your physician to obtain these values. If the measurements taken with these products fall outside the range, consulty. 1. Individual blood pressure varies every in one day, it also affected by the way you tie your cuff and the your measurement position, so please take the measurement at the same condition. 2.The varies of the pressure is greater if the person take medicine. 3.Waiting at least 4-5 minutes for another measurement. Below illustrates the blood pressure classification mode by World Health Organization (WHO) and International Society of Hypertension(ISH) in 1999. %ORRG /HYHO 3UHVVXUH PP+J Why my blood pressure is varies even in one day? The blood pressure is different even during 24 hour because of the weather,emotion, exercise etc, specially the âwhite coatâ in hospital which makes the results are higher than the ones at home. The attention need to pay when you measure you blood pressure at home: If the cuff is tied properly. If the cuff is too tight or too loose. If the cuff is tied on the wrist. If you feel anxious pressured. You had better take deep breath 2-3 times before beginning. Advice:adjust yourself for 4-5 minutes until you calm down. If the result is the same if measuring on the right wrist? It is ok for both wrists, but there will be some different results for different person, so suggest you measure the same wrist every time. SPECIFICATIONS TROUBLESHOOTING This section includes a list of error messages and frequently asked questions for problems you may encounter with your blood pressure monitor. If the products not operating as you think it should, check here before arranging for servicing. PROBLEM SYMPTOM No power Low batteries Error massage Display is dim or will not light up. REMEDY Batteries are exhausted. Replace with new batteries Batteries are inserted incorrectly. Show on Batteries are low. the display Display moder Digital LCD V.A.36x41mm Measurement mode Measurement range correctly Replace with new batteries Data communication has failed Make sure that phoneâs Bluetooth is on or within the distance range E 2 shows The cuff is very tight Refasten the cuff and then measure again. E 3 shows The pressure of the cuff is excess. Relax for a moment and then measure again. E 9 shows Product has not been activated. Reactivated E 10 or E 11 shows movement can affect the The monitor detected measurement.Relax for a motion while measuring. moment and then measure again. E 20 shows The measurement Loosen the clothing on process does not detect the arm and then the pulse signal. measure again. E 21 shows Measure incorrectly. A calibration error occurred. 2*AAA alkaline batteries Insert the batteries E 1 shows EExx,shows on the display. CHECK THIS Power supply Relax for a moment and then measure again. Retake the measurement. If the problem persists, contact the retailer or our customer service department for further assistance.Refer to the warranty for contact information and return instructions. Accuracy Oscillographic testing mode Pressure: 0kpa-40kpaďź0mmHg-300mmHgďź pulse value:(40-199)times/minute Pressure: 15Ä-25ÄwithinÂą0.4kpa(3mmHg) 10Ä-40Ä(out of 15Ä-25Ä) vithinÂą0.8kpa(6mmHg) pulse value:Âą5% Normal working condition Temperature:5Ä to 40Ä5HODWLYHKXPLGLW\ Atmospheric pressure: 86kPa to 106kPa Storage & transportation condition Temperature:-20Ä to 60Ä Relative humidity:10% to 93% Measurement perimeter of the wrist About 13.5cm-19.5cm Weight External dimensions Attachment Mode of operation Degree of protection Approx.120g(Excluding the dry cells) Approx.80Ă65Ă22mm 2*AAA alkaline batteries,user manual Continuous operation Type B applied part Protection against ingress of water IPXO Software version V01 Device classification Internally Powered ME Equipment WARNING: No modification of this equipment is allowed. CONTACT INFORMATION & R Q WDFW,Q IR UP DWLR Q For more information about our products, please visit www.transtek.cn.you can get customer service, usual problems and customer download, transtek will serve you anytime. Manufactured by: ZHONGSHAN TRANSTEK ELECTRONICES CO.,LTD Conpany: ZHONGSHAN TRANSTEK ELECTRONICES CO.,LTD Address: Jinâan Road, Minzhong, Zhongshan,528441,Guangdong, China Authorized European Representative: Conpany: MDSS - Medical Device Safety Service GmbH Address: Schiffgraben 41, 30175 Hannover, Germany Complied European Standards list Risk management EN 980: 2008 User manual EN 1041: 2008 General requirements EN 60601-1:1990+A1+A2+A13 for safety EN 1060-1:2001/A1:2002 sphygmomanometers EN 1060-3:1997/A1:2005 General requirements for all EQUIPMENT and SYSTEMS Guidance and manufactureâs declaration â electromagnetic emission The TMB-1014-BT is intended for use in the electromagnetic environment specified below. The customer of the user of the TMB-1014-BT should assure that it is used in such and environment. Emission test Compliance Electromagnetic environment â guidance RF emissions CISPR 11 Group 1 The TMB-1014-BT uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. RF emission CISPR 11 Class B Harmonic emissions IEC 61000-3-2 Not applicable Voltage fluctuations/ flicker emissions IEC 61000-3-3 Not applicable EN/ISO 14971:2007 Labelling Non-invasive Table 1 Guidance and manufactureâs declaration â electromagnetic emissions- EN 1060-4: 2004 The TMB-1014-BT is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. Electromagnetic EN 60601-1-2:2001+A1Ë2006 compatibility 1) Caution: This unit has been thoroughly tested and inspected to assure proper performance and operation! 2) Caution: this machine should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, this machine should be observed to verify normal operation in the configuration in which it will be used 3) This product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided, and this unit can be affected by portable and mobile RF communications equipment. 4) The guidelines for avoiding or identifying and resolving adverse electromagnetic effects on other equipment that may result from operation of the ME EQUIPMENT or ME SYSTEM. 5) Each frequency or frequency band of reception; the preferred frequency or frequency band, if applicable, and the bandwidth of the receiving section of the ME EQUIPMENT or ME SYSTEM in those bands; 6) Warning: the TMB-1014-BT may be interfered with by other equipment, even if that other equipment complies with CISPR EMISSION requirements. 7) Each frequency or frequency band of transmission, the type and frequency characteristics of the modulation and the EFFECTIVE RADIATED POWER. Table 2 Guidance and manufactureâs declaration â electromagnetic immunity â for all ME EQUIPMENT and ME SYSTEMS Table 4 Guidance and manufactureâs declaration â electromagnetic immunity â for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING Guidance and manufactureâs declaration â electromagnetic immunity Guidance and manufactureâs declaration â electromagnetic immunity The TMB-1014-BT is intended for use in the electromagnetic environment specified below. The customer or the user of TMB-1014-BT should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electrostatic discharge (ESD) IEC 61000-4-2 Âą6 kV contact Âą8 kV air Electrical fast transient/burst IEC 61000-4-4 Âą2 kV for power supply lines Surge IEC 61000-4-5 Âą1 kV line(s) to line(s) Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 <5% UT (>95% dip in UT) for 0.5 cycle <5% UT (>95% dip in UT) for 0.5 cycle 40% UT (60% dip in UT) for 5 cycles 40% UT) (60% dip in UT) for 5 cycles 70% UT (30% dip in UT) for 25 cycles 70% UT (30% dip in UT) for 25 cycles <5% UT (>95% dip in UT) for 5 sec Power frequency 3A/m (50Hz) magnetic field IEC 61000-4-8 NOTE Electromagnetic environment - guidance Floors should be wood, concrete or ceramic tile. If floor are covered with synthetic material, the relative humidity should be at least 30%. Mains power quality should Âą2kV for power supply lines be that of a typical commercial or hospital environment. Mains power quality should Âą1 kV be that of a typical differential mode commercial or hospital environment. The TMB-1014-BT is intended for use in the electromagnetic environment specified below. The customer or the user should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Âą6 kV contact Âą8 kV air <5% UT (>95% dip in UT) for 5 sec 3A/m Mains power quality should be that of a typical commercial or hospital environment. If the user of the TMB-1014-BT requires continued operation during power mains interruptions, it is recommended that the TMB-1014-BT be powered from an uninterruptible power supply or a battery. Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. UT is the a.c. mains voltage prior to application of the test level. Conducted RF 3 Vrms IEC 61000-4-6 150 kHz to 80 MHz 3 Vrms Radiated RF 3 V/m 80 MHz IEC 61000-4-3 to 2.5 GHz 3 V/m Electromagnetic environment guidance Portable and mobile RF communications equipment should be used no closer to any part of the YS-6100, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Portable and mobile RF communications equipment should be used no closer to any part of the including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance d = 1.167 d = 1.167 80 MHz to 800 MHz d = 2.333 800 MHz to 2.5 GHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range. Interference may occur in the vicinity of equipment marked with the following symbol: NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the ELE007839V1 is used exceeds the applicable RF compliance level above, the ELE007839V1 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the ELE007839V1. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. Table 6 Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM â for ME EQUIPMENT or ME SYSTEM that are not LIFE-SUPPORTING Recommended separation distances between portable and mobile RF communications equipment and the ELE007839V1 Fitness Equipment. The TMB-1014-BT is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the TMB-1014BT can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the TMB-1014-BT as recommended below, according to the maximum output power of the communications equipment. Rated maximum output Separation distance according to frequency of transmitter power of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2,5 GHz d = 2.333 d = 1.167 d = 1.167 0,01 0.167 0.167 0.233 0,1 0.369 0.369 0.738 10 1.167 1.167 2.333 3.690 3.690 7.388 100 11.67 11.67 23.330 For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. This device complies with Part 15 and Part 18 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. Caution: The user is cautioned that changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 and Part 18 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: âReorient or relocate the receiving antenna. âIncrease the separation between the equipment and receiver. âConnect the equipment into an outlet on a circuit different from that to which the receiver is connected. âConsult the dealer or an experienced radio/TV technician for help. This equipment complies with FCC and IC radiation exposure limits set forth for an uncontrolled environment. End user must follow the specific operating instructions for satisfying RF exposure compliance. This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter.
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.3 Linearized : No Encryption : Standard V2.3 (128-bit) User Access : Print, Print high-res Page Count : 15 Create Date : 2012:10:10 16:20:24+08:00 Title : User Manual.pdf Producer : GPL Ghostscript 8.64 Author : Millie_LiWM Creator : PScript5.dll Version 5.2 Modify Date : 2012:10:10 16:20:27+08:00EXIF Metadata provided by EXIF.tools