PRODUCT SPECIFICATIONS

Vanquish Charged Aerosol Detectors
The collective power of chromatography
LC that takes your productivity to new heights

Vanquish platform benefits
� Precision and reproducibility to meet every application demand
� High detector sensitivity and low baseline noise
� Reduced maintenance, easier setup with Thermo ScientificTM ViperTM fingertight fittings
� Seamless integrated solutions for comprehensive analyte detection
Keywords Vanquish Horizon, Vanquish Flex, Vanquish Core, Vanquish Duo, HPLC, UHPLC, Vanquish Charged Aerosol Detector F, Vanquish Charged Aerosol Detector H, Inverse Gradient

Discover what you're missing Thermo ScientificTM VanquishTM Charged Aerosol Detectors change the way you view every sample and open new opportunities for discovery and routine analysis. They deliver the flexibility and performance required for analytical R&D and the simplicity and reproducibility needed for manufacturing QA/QC.
� Estimate relative amounts even without standards by utilizing consistent detector response for all non-volatile compounds independent of chemical structure.
� Excellent application flexibility is achieved with adjustable evaporation temperature for improved analyte response.
� The FocusJet concentric nebulizer ensures compatibility with analytical HPLC, UHPLC and microflow LC without any hardware changes.
� The Vanquish Charged Aerosol Detector H offers versatility for research and method development while the Vanquish Charged Aerosol Detector F provides a robust feature set for everyday analysis.
� Regardless of configuration, Vanquish Charged Aerosol Detectors provide a highly integrated solution with optimized fluidic connections and singlepoint intelligent control through Thermo ScientificTM ChromeleonTM CDS (chromatography data system) software.

Operating mode Nebulization Mobile phase flow rate Evaporation temperature Maximum data collection Dynamic range Filter time constant Inert gas supply Gas consumption Internal gas pressure GLP features PC connections Analog signal output Safety features Wetted parts Biocompatible Power requirements Environmental conditions Dimensions (h � w � d) Weight

Product specification

Vanquish Charged Aerosol Detector H

Vanquish Charged Aerosol Detector F

Charged aerosol detection

FocusJet concentric flow design

0.01�2.0 mL/min

0.2�2.0 mL/min

Settable range: Ambient +5 to +100 �C

Select: 35, 50, 70 �C

200 Hz

100 Hz

Up to 4 orders of magnitude

4th order low-pass Bessel digital with time constant based user selections: 0.1, 0.2, 0.5, 1.0, 2.0, 3.6, 5.0, 10.0 s

Nitrogen (recommended) or compressed air Inlet Pressure: 482�551 kPa (4.8�5.5 bar, 70�80 psi)

Typically 4 L/min

Electronically controlled pressure regulation system

Predictive performance functions: gas filter change interval, service monitoring period, system parameters logged in the Thermo ScientificTM Chromeleon CDS Audit Trail

USB 2.0; 3-port-HUB to connect further Vanquish modules

0�1 V DC with full scale output range: 1 pA to 500 pA in 1-2-5 sequence (installable option)

Power-up diagnostics, inlet gas over-pressure relief, detection of high evaporation temperature, pump flow shutdown, fluid leak detection and safe leak handling

Stainless steel (type 316), Nitronic� 60 stainless steel, PEEKTM, Simriz�, fused silica

No

100/240 VAC, 50/60 Hz, max. 150 W/255 VA

Operation: 5�35 �C; 20�80% RH (non condensing), max. 2000 m above sea-level, Storage: -20�45 �C max. 60% RH (non condensing)

192 � 420 � 620 mm (7.6 � 16.5 � 24.4 in.)

17.5 kg (38.6 lbs.)

Ordering information
Description
Vanquish Charged Aerosol Detector H Vanquish Charged Aerosol Detector F Accessories Corona Nitrogen 1010 Generator Installation kit included (compressed air inlet source required: 760 kPa, 110 psi) Corona Air Compressor 230 V Oil-free air compressor for pairing with 6295.0200 above. Installation kit included. Corona Air Compressor 110 V Oil-free air compressor for pairing with 6295.0200 above. Installation kit included.
Find out more at thermofisher.com/CAD
� 2015-2021 Thermo Fisher Scientific Inc. All rights reserved. Nitronic is a registered trademark of AK Steel Corp. Simriz is a registered trademark of Freudenberg-NOK. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. PS71516-EN 0321M

Part number VH-D20-A VF-D20-A
6295.0200 6295.0300 6295.0350