Endress and Hauser and Co FMR6XEL Level Probing Radar User Manual Micropilot FMR60

Endress and Hauser GmbH and Co Level Probing Radar Micropilot FMR60

Contents

User Manual 1

Download: Endress and Hauser and Co FMR6XEL Level Probing Radar User Manual Micropilot FMR60
Mirror Download [FCC.gov]Endress and Hauser and Co FMR6XEL Level Probing Radar User Manual Micropilot FMR60
Document ID3933672
Application IDiPt0hsfc15W1ZAqsSUm6dQ==
Document DescriptionUser Manual 1
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize162.52kB (2031526 bits)
Date Submitted2018-07-22 00:00:00
Date Available2018-07-25 00:00:00
Creation Date2018-03-19 15:09:27
Producing SoftwareAntenna House PDF Output Library 6.2.735 (Windows (x64))
Document Lastmod2018-03-19 15:11:05
Document TitleMicropilot FMR60
Document CreatorAH XSL Formatter V6.2 MR13 for Windows (x64) : 6.2.15.22885 (2015/12/14 12:50JST)
Document Author: 00

DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
TI01302F/00/EN/04.18
71394723
Products
Solutions
Services
Technical Information
Micropilot FMR60
Free space radar
Level measurement in liquids
Application
• Continuous, non-contact level measurement of liquids,
pastes and sludges
• PTFE drip-off antenna
• Maximum measuring range: 50 m (164 ft)
• Temperature: –40 to +130 °C (–40 to +266 °F)
• Pressure: –1 to +16 bar (–14.5 to +232 psi)
• Accuracy: ± 1 mm (0.04 in)
• Linearity protocol (3-point, 5-point)
Your benefits
• Innovative drip-off antenna made of PTFE
• Reliable measurement thanks to improved focusing and small beam angle,
particularly in vessels with many internal fittings
• Safety by design - ensures highest safety
• Easy, guided commissioning with intuitive user interface
• Bluetooth® wireless technology for commissioning, operation and maintenance via
free iOS / Android app SmartBlue
• Maximum reliability thanks to multi-echo tracking
• HistoROM configuration memory makes for easier commissioning,
maintenance and diagnostics
• SIL2 as per IEC 61508, SIL3 for homogeneous or diverse redundancy
• Easy proof testing for SIL and WHG
• RFID TAG - easy identification of measuring points for improved data access
• Heartbeat Technology
Table of contents
Important document information . . . . . . . . . . . . . . . 4
Safety symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Electrical symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Symbols for certain types of information . . . . . . . . . . . . . .
Symbols in graphics . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Symbols at the device . . . . . . . . . . . . . . . . . . . . . . . . . . .
Terms and abbreviations . . . . . . . . . . . . . . . . . . . . .
Registered trademarks . . . . . . . . . . . . . . . . . . . . . . . 7
Function and system design . . . . . . . . . . . . . . . . . . . 8
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Operating frequency . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Transmission power . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Linearization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Protocol-specific data . . . . . . . . . . . . . . . . . . . . . . . . . .
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Terminal assignment . . . . . . . . . . . . . . . . . . . . . . . . . .
Device plug connectors . . . . . . . . . . . . . . . . . . . . . . . . .
Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . .
Current consumption . . . . . . . . . . . . . . . . . . . . . . . . . .
Power supply failure . . . . . . . . . . . . . . . . . . . . . . . . . .
Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . .
Cable entries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Overvoltage protection . . . . . . . . . . . . . . . . . . . . . . . . .
11
11
12
12
12
12
14
14
18
19
20
20
20
20
21
21
21
Performance characteristics . . . . . . . . . . . . . . . . . . 22
Reference operating conditions . . . . . . . . . . . . . . . . . . .
Reference accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measured value resolution . . . . . . . . . . . . . . . . . . . . . .
Response time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Influence of ambient temperature . . . . . . . . . . . . . . . . .
Influence of gas phase . . . . . . . . . . . . . . . . . . . . . . . . .
22
22
22
22
23
23
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Installation conditions . . . . . . . . . . . . . . . . . . . . . . . . . 24
Installation: Drip-off antenna PTFE 50 mm / 2" . . . . . . . . . 28
Container with heat insulation . . . . . . . . . . . . . . . . . . . . 29
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Ambient temperature range . . . . . . . . . . . . . . . . . . . . .
Ambient temperature limits . . . . . . . . . . . . . . . . . . . . .
Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . .
30
30
31
V. 4, Rev. 2, 19-03-2018
Climate class . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Altitude according to IEC61010-1 Ed.3 . . . . . . . . . . . . . .
Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . .
Vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . .
Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . .
31
31
31
31
31
Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Process temperature, process pressure . . . . . . . . . . . . . . . 32
Dielectric constant . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Mechanical construction . . . . . . . . . . . . . . . . . . . .
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Materials: GT19 housing (plastic) . . . . . . . . . . . . . . . . . .
Materials: GT20 housing (die-cast aluminum, powdercoated) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Materials: antenna and process connection . . . . . . . . . . .
Materials: Weather protection cover . . . . . . . . . . . . . . . .
34
34
37
38
39
40
41
Operability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Operating concept . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Local operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operation with remote display and operating module
FHX50 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operation via Bluetooth® wireless technology . . . . . . . . . .
Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . .
SupplyCare inventory management software . . . . . . . . . .
43
44
44
46
Certificates and approvals . . . . . . . . . . . . . . . . . . .
49
CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
RoHS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
RCM-Tick marking . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Ex approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Dual seal according to ANSI/ISA 12.27.01 . . . . . . . . . . . .
Functional safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
WHG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Pressure equipment with allowable pressure
≤ 200 bar (2 900 psi) . . . . . . . . . . . . . . . . . . . . . . . . . .
Radio standard EN 302729-1/2 . . . . . . . . . . . . . . . . . . .
Radio standard EN 302372-1/2 . . . . . . . . . . . . . . . . . . .
FCC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Industry Canada . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CRN approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Test, certificate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Hard-copy product documentation . . . . . . . . . . . . . . . . .
Other standards and guidelines . . . . . . . . . . . . . . . . . . .
42
43
49
49
49
49
49
49
49
49
50
51
51
51
51
53
53
54
Ordering information . . . . . . . . . . . . . . . . . . . . . . . 55
Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . .
3-point linearity protocol . . . . . . . . . . . . . . . . . . . . . . .
5-point linearity protocol . . . . . . . . . . . . . . . . . . . . . . .
Customer-specific configuration . . . . . . . . . . . . . . . . . . .
Tagging (TAG) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
55
56
57
58
58
58
Application Packages . . . . . . . . . . . . . . . . . . . . . . . 59
Heartbeat Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . .
59
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Heartbeat Verification . . . . . . . . . . . . . . . . . . . . . . . . .
Heartbeat Monitoring . . . . . . . . . . . . . . . . . . . . . . . . .
60
61
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Device-specific accessories . . . . . . . . . . . . . . . . . . . . . .
Communication-specific accessories . . . . . . . . . . . . . . . .
Service-specific accessories . . . . . . . . . . . . . . . . . . . . . .
System components . . . . . . . . . . . . . . . . . . . . . . . . . . .
62
66
67
67
Supplementary documentation . . . . . . . . . . . . . . .
68
Standard documentation . . . . . . . . . . . . . . . . . . . . . . . . 68
Safety Instructions (XA) . . . . . . . . . . . . . . . . . . . . . . . . 68
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
Important document information
Safety symbols
Symbol
Meaning
DANGER
DANGER!
This symbol alerts you to a dangerous situation. Failure to avoid this situation will
result in serious or fatal injury.
WARNING
WARNING!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can
result in serious or fatal injury.
CAUTION
CAUTION!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can
result in minor or medium injury.
NOTICE
Electrical symbols
Symbol
NOTE!
This symbol contains information on procedures and other facts which do not result in
personal injury.
Meaning
Direct current
Alternating current
Direct current and alternating current
Ground connection
A grounded terminal which, as far as the operator is concerned, is grounded via a
grounding system.
Protective Earth (PE)
A terminal which must be connected to ground prior to establishing any other
connections.
The ground terminals are situated inside and outside the device:
• Inner ground terminal: Connects the protectiv earth to the mains supply.
• Outer ground terminal: Connects the device to the plant grounding system.
Symbols for
certain types of information
Symbol
Meaning
Permitted
Procedures, processes or actions that are permitted.
Preferred
Procedures, processes or actions that are preferred.
Forbidden
Procedures, processes or actions that are forbidden.
Tip
Indicates additional information.
Reference to documentation.
Reference to page.
Reference to graphic.
Visual inspection.
Symbols in graphics
Symbol
Meaning
1, 2, 3 ...
Item numbers
1. , 2. , 3. …
Series of steps
A, B, C, ...
Views
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Symbols at the device
Symbol
Meaning
A-A, B-B, C-C, ...
Sections
Hazardous area
Indicates a hazardous area.
Safe area (non-hazardous area)
Indicates the non-hazardous area.
Symbol
Meaning
Safety instructions
Observe the safety instructions contained in the associated Operating Instructions.
Temperature resistance of the connection cables
Specifies the minimum value of the temperature resistance of the connection cables.
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
Terms and abbreviations
Term/abbreviation Explanation
BA
Document type "Operating Instructions"
KA
Document type "Brief Operating Instructions"
TI
Document type "Technical Information"
SD
Document type "Special Documentation"
XA
Document type "Safety Instructions"
PN
Nominal pressure
MWP
Maximum Working Pressure
The MWP can also be found on the nameplate.
ToF
Time of Flight
FieldCare
Scalable software tool for device configuration and integrated plant asset management
solutions
DeviceCare
Universal configuration software for Endress+Hauser HART, PROFIBUS,
FOUNDATION Fieldbus and Ethernet field devices
DTM
Device Type Manager
DD
Device Description for HART communication protocol
DC
Relative dielectric constant εr
Operating tool
The term "operating tool" is used in place of the following operating software:
FieldCare / DeviceCare, for operation via HART communication and PC
BD
Blocking Distance; no signals are analyzed within the BD.
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Registered trademarks
HART®
Registered trademark of the HART Communication Foundation, Austin, USA
Bluetooth®
The Bluetooth® word mark and logos are registered trademarks owned by the Bluetooth SIG, Inc. and
any use of such marks by Endress+Hauser is under license. Other trademarks and trade names are
those of their respective owners.
Apple®
Apple, the Apple logo, iPhone, and iPod touch are trademarks of Apple Inc., registered in the U.S.and
other countries. App Store is a service mark of Apple Inc.
Android®
Android, Google Play and the Google Play logo are trademarks of Google Inc.
KALREZ®, VITON®
Registered trademark of DuPont Performance Elastomers L.L.C., Wilmington, USA
TEFLON®
Registered trademark of E.I. DuPont de Nemours & Co., Wilmington, USA
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
Function and system design
Measuring principle
The Micropilot is a "downward-looking" measuring system, operating based on the frequency
modulated continuous wave method (FMCW). The antenna emits an electromagnetic wave at a
continuously varying frequency. This wave is reflected by the product and received again by the
antenna.
A0032017
1
FMCW principle: transmission and reflection of the continuous wave
Reference point of measurement
Distance between reference point and product surface
The frequency of this wave is modulated in the form of a sawtooth signal between two limit
frequencies f1 and f2:
f2
∆f
∆t
f1
A0023771
2
FMCW principle: result of frequency modulation
Transmitted signal
Received signal
This results in the following difference frequency at any time between the transmitted signal and the
received signal:
Δf = k Δt
where Δt is the run time and k is the specified increase in frequency modulation.
Δt is given by the distance D between the reference point R and the product surface:
D = (c Δt) / 2
where c is the speed of propagation of the wave.
In summary, D can be calculated from the measured difference frequency Δf. D is then used to
determine the content of the tank or silo.
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Input
Measured variable
The measured variable is the distance between the reference point and the product surface. The level
is calculated based on "E", the empty distance entered. Optionally, the level can be converted to other
variables (volume, mass) by linearization (32 value pairs).
Measuring range
Maximum measuring range
Device
Antenna 1)
Maximum measuring range
FMR60
GA:
Drip-off, PTFE 50 mm / 2"
50 m (164 ft)
1)
Feature 070 in the product structure
Usable measuring range
The usable measuring range depends on the antenna size, the medium's reflective properties, the
installation position and any possible interference reflections.
The following tables describe the groups of media as well as the achievable measuring range as a
function of application and media group. If the dielectric constant of a medium is unknown, it is
recommended to assume media group B to ensure a reliable measurement.
Media groups
Media group
DC (εr)
Example
A0
1.2 to 1.4
Butane, liquid nitrogen, liquefied hydrogen
1.4 to 1.9
non-conducting liquids, e.g. liquefied gas 1)
1.9 to 4
non-conducting liquids, e.g. benzene, oil, toluene, …
4 to 10
e.g. concentrated acids, organic solvents, esters, aniline, alcohol, acetone, …
> 10
conducting liquids, e.g. aqueous solutions, dilute acids and alkalis
1)
NH3 falls into medium group A.
For dielectric constants (DC values) of many media commonly used in various industries refer
to:
• the Endress+Hauser DC manual (CP01076F)
• the Endress+Hauser "DC Values App" (available for Android and iOS)
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
FMR60
Storage tank
Buffer tank
Process tank with agitator
A0034565
A0034567
A0034566
Calm product surface (e.g. intermittent
filling, filling from bottom, immersion
tubes)
A0
Moving surfaces (e.g. continuous filling,
from above, mixing jets)
A0
A0
Turbulent surface (e.g. filling from above, agitators, baffles)
(7)
(13)
(13)
(23)
(23)
(23)
12
(39)
13
(43)
23
(75)
15
(49)
25
(82)
28
(92)
40
(131) 50
(164)
A0034584
44
(144)
A0034579
A0034574
Measuring range [m (ft)]
Operating frequency
Approx. 80 GHz
Up to 8 devices can be installed in the same tank without interfering with each other.
Transmission power
10
• Peak power: 6.3 mW
• Average output power: 63 µW
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Output
Output signal
HART
Signal coding
FSK ±0.5 mA over current signal
Data transmission rate
1 200 Bit/s
Galvanic isolation
Yes
Bluetooth® wireless technology
Device version
Ordering feature 610 "Accessory mounted", option NF "Bluetooth"
Operation / configuration
By the SmartBlue app.
Range under reference conditions > 10 m (33 ft)
Encryption
Encrypted communication and password encryption prevent incorrect
operation by unauthorized persons.
Switch output
For HART devices, the switch output is available as an option. See product structure, feature 20:
"Power Supply, Output", option B: "2-wire; 4-20mA HART, switch output"
Switch output
Function
Open collector switching output
Switching behavior
Binary (conductive or non-conductive), switches when the programmable switch
point is reached
Failure mode
non-conductive
Electrical connection values U = 16 to 35 VDC, I = 0 to 40 mA
Internal resistance
RI < 880 Ω
The voltage drop at this internal resistance has to be taken into account on
planning the configuration. For example, the resulting voltage at a connected
relay must be sufficient to switch the relay.
Insulation voltage
floating, Insulation voltage 1 350 VDC to power supply aund 500 VAC to ground
Switch point
freely programmable, separately for switch-on and switch-off point
Switching delay
freely programmable from 0 to 100 s, separately for switch-on and switch-off
point
Number of switching cycles corresponds to the measuring cycle
Signal source
device variables
•
•
•
•
•
•
Level linearized
Distance
Terminal voltage
Electronic temperature
Relative echo amplitude
Diagnostic values, Advanced diagnostics
Number of switching cycles unlimited
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
11
Signal on alarm
Depending on the interface, failure information is displayed as follows:
• Current output (for HART devices)
– Failsafe mode selectable (in accordance with NAMUR Recommendation NE 43):
Minimum alarm: 3.6 mA
Maximum alarm (= factory setting): 22 mA
– Failsafe mode with user-selectable value: 3.59 to 22.5 mA
• Local display
– Status signal (in accordance with NAMUR Recommendation NE 107)
– Plain text display
• Operating tool via HART communication or service interface (CDI)
– Status signal (in accordance with NAMUR Recommendation NE 107)
– Plain text display
Linearization
The linearization function of the device allows the conversion of the measured value into any unit of
length or volume. Linearization tables for calculating the volume in cylindrical tanks are preprogrammed. Other linearization tables of up to 32 value pairs can be entered manually or semiautomatically.
Galvanic isolation
All circuits for the outputs are galvanically isolated from each other.
Protocol-specific data
HART
Manufacturer ID
17 (0x11)
Device type ID
0x112B
HART specification
7.0
Device description files (DTM, DD)
Information and files under:
• www.endress.com
• www.fieldcommgroup.org
HART load
min. 250 Ω
HART device variables
The measured values can be freely assigned to the device variables.
Measured values for PV (primary variable)
• Level linearized
• Distance
• Electronic temperature
• Relative echo amplitude
• Area of incoupling
• Analog output adv. diagnostics 1
• Analog output adv. diagnostics 2
Measured values for SV, TV, FV (second, third and fourth variable)
• Level linearized
• Distance
• Electronic temperature
• Terminal voltage
• Relative echo amplitude
• Absolute echo amplitude
• Area of incoupling
• Analog output adv. diagnostics 1
• Analog output adv. diagnostics 2
Supported functions
• Burst mode
• Additional transmitter status
Wireless HART data
12
Minimum start-up voltage
16 V
Start-up current
3.6 mA
Start-up time
65 s
Minimum operating voltage
14.0 V
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Endress+Hauser
Multidrop current
4.0 mA
Set-up time
15 s
V. 4, Rev. 2, 19-03-2018
13
Power supply
Terminal assignment
2-wire: 4-20mA HART
2- w
4-2 ire le
el
4-2
0m
ope
10
Sp
711 re pa
08x rt
xx
mm
4...
HA 20mA
RT
HA mA
[21 RT
–
–
4...20 mA
1-c
han
[16
nel
ove
1 4-2 2
rvo
ltag
0m
pro
tec
tion
–
–
4...20 mA
A0011294
3
14
Terminal assignment 2-wire; 4-20mA HART
Without integrated overvoltage protection
With integrated overvoltage protection
Active barrier with power supply (e.g. RN221N): Observe terminal voltage
HART communication resistor (≥ 250 Ω): Observe maximum load
Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
Analog display device: Observe maximum load
Cable screen; observe cable specification
4-20mA HART (passive): Terminals 1 and 2
Overvoltage protection module
Terminal for potential equalization line
Cable entry
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
2-wire: 4-20mA HART, switch output
10
2- w
4-2 ire
HA mA
[02 RT
/03
PF
ope
++
31 -42
3+
Sp
711 re pa
08x rt
xx
1 2
10
4-
mm
4...
HA 20 mA
RT
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
–
≥250 Ω 4...20 mA
2 3
3+
3 4-2 4
11
0m
A/
2-c
han
4[17
–
nel
ove
rvo
lt
1 4-2 2
FIE 0mA
LD
BU
pro
age
tec
tion
≥250 Ω 4...20 mA
2 3
A0013759
4
10
11
Endress+Hauser
Terminal assignment 2-wire; 4-20mA HART, switch output
Without integrated overvoltage protection
With integrated overvoltage protection
Active barrier with power supply (e.g. RN221N): Observe terminal voltage
HART communication resistor (≥ 250 Ω): Observe maximum load
Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
Analog display device: Observe maximum load
Cable screen; observe cable specification
4-20mA HART (passive): Terminals 1 and 2
Switch output (open collector): Terminals 3 and 4
Terminal for potential equalization line
Cable entry for 4-20mA HART line
Cable entry for switch output line
Overvoltage protection module
V. 4, Rev. 2, 19-03-2018
15
2-wire: 4-20mA HART, 4-20mA
14
13
12
–
1 #
0m
o pe
Sp
711 re pa
08x rt
xx
1 2
10
mm
4 ...
3 4
11
el
4-2
20m
2- w
4-2 ire le
HA mA
[04 RT
/05
4 ...
HA 20mA
RT
4...20 mA
–
2 #
4...20 mA
5 6
–
1 #
3 4
4...20 mA
4-2
0m
10
2-c
han
[17
–
2 #
nel
A/
1 -
4FIE 20mA
ove
rvo
LD
ltag
ep
ro
tec
tion
4...20 mA
5 6
A0013923
5
10
11
12
13
14
Terminal assignment 2-wire, 4-20 mA HART, 4...20mA
Without integrated overvoltage protection
With integrated overvoltage protection
Connection current output 2
Connection current output 1
Supply voltage for current output 1 (e.g. RN221N); Observe terminal voltage
Cable screen; observe cable specification
HART communication resistor (≥ 250 Ω): Observe maximum load
Connection for Commubox FXA195 or FieldXpert SFX350/SFX370 (via VIATOR Bluetooth modem)
Analog display device ; observe maximum load
Analog display device ; observe maximum load
Supply voltage for current output 2 (e.g. RN221N); Obeserve terminal voltage
Overvoltage protection module
Current output 2: Terminals 3 and 4
Terminal for the potential equalization line
Cable entry for current output 1
Cable entry for current output 2
This version is also suited for single-channel operation. In this case, current output 1 (terminals
1 and 2) must be used.
16
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Connection examples for the switch output
For HART devices, the switch output is available as an option. See product structure, feature 20:
"Power Supply, Output", option B: "2-wire; 4-20mA HART, switch output"
3+
Ri
3+
Ri
4A0015909
6
4-
Connection of a relay
A0015910
Suitable relays (examples):
• Solid-state relay: Phoenix Contact OV-24DC/480AC/5 with mounting
rail connector UMK-1 OM-R/AMS
• Electromechanical relay: Phoenix Contact PLC-RSC-12DC/21
7
Connection of a digital input
Pull-up resistor
Digital input
For optimum interference immunity we recommend to connect an external resistor (internal
resistance of the relay or Pull-up resistor) of < 1 000 Ω.
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
17
Device plug connectors
For the versions with fieldbus plug connector (M12 or 7/8"), the signal line can be connected
without opening the housing.
Pin assignment of the M12 plug connector
A0011175
Pin
Meaning
Signal +
not connected
Signal -
Ground
Pin
Meaning
Signal -
Signal +
Not connected
Screen
Pin assignment of the 7/8" plug connector
A0011176
18
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Supply voltage
An external power supply is necessary.
Various power supply units can be ordered as an accessory from Endress+Hauser.
2-wire, 4-20mA HART, passive
"Power
supply,
output" 1)
"Approval" 2)
Terminal
voltage U at
device
Maximum load R,
depending on the supply voltage
U0 of the power supply unit
A:
2-wire;
4-20mA
HART
• Nonhazardous
• Ex nA
• Ex ic
• CSA GP
14 to 35 V 3)
R [W]
Ex ia / IS
14 to 30 V 3)
• Ex d(ia) /
XP
• Ex ic(ia)
• Ex nA(ia)
• Ex ta / DIP
14 to 35 V 3) 4)
500
10
20
14
30
25
35
U0 [V]
A0031745
Ex ia + Ex
d(ia) / IS + XP
1)
2)
3)
4)
Feature 020 in the product structure
Feature 010 in the product structure
If the Bluetooth modem is used, the minimum supply voltage increases by 2 V.
At ambient temperatures TTa ≤ –20 °C, a terminal voltage U ≥ 16 V is required to start the device with the
min. error current (3.6 mA).
"Power
supply,
output" 1)
"Approval" 2)
Terminal
voltage U at
device
Maximum load R,
depending on the supply voltage
U0 of the power supply unit
B:
2-wire;
4-20 mA
HART,
switch
outpu
• Nonhazardous
• Ex nA
• Ex nA(ia)
• Ex ic
• Ex ic(ia)
• Ex d(ia) / XP
• Ex ta / DIP
• CSA GP
16 to 35 V 3)
R [W]
• Ex ia / IS
• Ex ia + Ex
d(ia) / IS +
XP
16 to 30 V 3)
1)
2)
3)
Endress+Hauser
14 to 30 V 3)
500
10
16
20
27
30
35
U0 [V]
A0031746
Feature 020 in the product structure
Feature 010 in the product structure
If the Bluetooth modem is used, the minimum supply voltage increases by 2 V.
V. 4, Rev. 2, 19-03-2018
19
"Power
supply,
output" 1)
"Approval" 2) Terminal
voltage U at
device
C:
2-wire;
4-20mA
HART,
4-20mA
All
16 to 30 V 3)
Maximum load R,
depending on the supply voltage
U0 of the power supply unit
R [W]
500
10
16
20
27
30
35
U0 [V]
A0031746
1)
2)
3)
Power consumption
Feature 020 in the product structure
Feature 010 in the product structure
If the Bluetooth modem is used, the minimum supply voltage increases by 2 V.
Integrated polarity reversal protection
Yes
Permitted residual ripple with f = 0 to 100 Hz
USS < 1 V
Permitted residual ripple with f = 100 to 10 000 Hz
USS < 10 mV
"Power supply; Output" 1)
Power consumption
A: 2-wire; 4-20mA HART
< 0.9 W
B: 2-wire; 4-20mA HART, switch output
< 0.9 W
C: 2-wire; 4-20mA HART, 4-20mA
< 2 x 0.7 W
1)
Feature 020 of the product structure
Current consumption
HART
Nominal current
3.6 to 22 mA the start-up current for multidrop mode can be parametrized
(is set to 3.6 mA on delivery)
Breakdown signal (NAMUR NE43) adjustable: 3.59 to 22.5 mA
Power supply failure
• Configuration is retained in the HistoROM (EEPROM).
• Error messages (incl. value of operated hours counter) are stored.
Potential equalization
No special measures for potential equalization are required.
If the device is designed for hazardous areas, observe the information in the documentation
"Safety Instructions" (XA).
20
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Cable entries
Connection of power supply and signal line
To be selected in feature 050 "Electrical connection"
• Gland M20; Material dependent on the approval:
– For Non-Ex, ATEX, IECEx, NEPSI Ex ia/ic:
Plastics M20x1.5 for cable ⌀5 to 10 mm (0.2 to 0.39 in)
– For Dust-Ex, FM IS, CSA IS, CSA GP, Ex nA:
Metal M20x1.5 for cable ⌀7 to 10 mm (0.28 to 0.39 in) 1)
– For Ex d:
No gland available
• Thread
– ½" NPT
– G ½"
– M20 × 1.5
• Plug M12 / Plug 7/8"
Only available for Non-Ex, Ex ic, Ex ia
Connection of remote display FHX50
Dependent on feature 030: "Display, Operation":
• "Prepared for display FHX50 + M12 connection":
M12 socket
• "Prepared for display FHX50 + custom connection":
M16 cable gland
Cable specification
• Devices without integrated overvoltage protection
Pluggable spring-force terminals for wire cross-sections 0.5 to 2.5 mm2 (20 to 14 AWG)
• Devices with integrated overvoltage protection
Screw terminals for wire cross-sections 0.2 to 2.5 mm2 (24 to 14 AWG)
• For ambient temperature TU≥60 °C (140 °F): use cable for temperature TU +20 K.
HART
• A normal device cable suffices if only the analog signal is used.
• A shielded cable is recommended if using the HART protocol. Observe grounding concept of the
plant.
Overvoltage protection
If the measuring device is used for level measurement in flammable liquids which requires the use of
overvoltage protection according to DIN EN 60079-14, standard for
test procedures 60060-1 (10 kA, pulse 8/20 μs), an overvoltage protection module has to be
installed.
Integrated overvoltage protection module
An integrated overvoltage protection module is available for 2-wire HART devices.
Product structure: Feature 610 "Accessory mounted", option NA "Overvoltage protection".
Technical data
Resistance per channel
2 × 0.5 Ω max.
Threshold DC voltage
400 to 700 V
Threshold impulse voltage
< 800 V
Capacitance at 1 MHz
< 1.5 pF
Nominal arrest impulse voltage (8/20 μs)
10 kA
External overvoltage protection module
HAW562 or HAW569 from Endress+Hauser are suited as external overvoltage protection.
1)
The material of the gland is dependent on the housing type; GT19 (plastic housing) and GT20 (aluminum housing): nickel-coated brass (CuZn).
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
21
Performance characteristics
Reference operating
conditions
•
•
•
•
•
Temperature = +24 °C (+75 °F) ±5 °C (±9 °F)
Pressure = 960 mbar abs. (14 psia) ±100 mbar (±1.45 psi)
Humidity = 60 % ±15 %
Reflector: metal plate with diameter ≥ 1 m (40 in)
No major interference reflections inside the signal beam
Reference accuracy
Typical data under reference operating conditions: DIN EN IEC 61298-2 / DIN EN IEC 60770-1;
percentage values in relation to the span.
Output:
digital
analog 1)
Accuracy (Sum of nonlinearity, nonrepeatability
and hysteresis) 2)
Measuring distance up to 0.8 m (2.62 ft): max.
±4 mm (±0.16 in)
±0.03 %
Measuring distance > 0.8 m (2.62 ft): ±1 mm (±0.04 in)
±0.02 %
Non-repeatability
1)
2)
3)
3)
≤ 1 mm (0.04 in)
Add error of the analogous value to the digital value.
If the reference conditions are not met, the offset/zero point arising from the mounting conditions may be
up to ±4 mm (0.16 in). This additional offset/zero point can be compensated for by entering a correction
(parameter "level correction") during commissioning.
The non-repeatability is already considered in the accuracy.
Differing values in near-range applications
∆ [mm (in)]
4 (0.16)
1 (0.04)
-1 (-0.04)
-4 (-0.16)
0.8 (2.62)
D [m (!)]
A0032636
8
Δ
Measured value resolution
Maximum measured error in near-range applications
Maximum measured error
Reference point of the distance measurement
Distance from reference point of antenna
Dead band according to DIN EN IEC 61298-2 / DIN EN IEC 60770-1:
• Digital: 1 mm
• Analog: 1 µA
Response time
2)
22
The response time can be configured. The following step response times (in accordance with
DIN EN IEC 61298-2 / DIN EN IEC 60770-1) 2) are when damping is switched off:
According to DIN EN IEC 61298-2 / DIN EN IEC 60770-1, the step response time is the time that elapses after an abrupt change in the input
signal until the change in the output signal has adopted 90% of the steady-state value for the first time.
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Sampling rate
≥1.3 s–1 at U ≥ 24 V
Response time
< 3.6 s
Influence of ambient
temperature
The measurements are performed according to DIN EN IEC 61298-3 / DIN EN IEC 60770-1
• Digital (HART): average TK = 2 mm/10 K
• Analog (current output):
– Zero point (4 mA): average TK = 0.02 %/10 K
– Span (20 mA): average TK = 0.05 %/10 K
Influence of gas phase
High pressure reduces the speed of propagation of the measuring signals in the gas/vapor above the
medium. This effect depends on the type of gas phase and its temperature. This results in a
systematic measured error that increases with increasing distance between the reference point of
the measurement (flange) and the surface of the product. The following table shows this measured
error for a few typical gases/vapors (with regard to the distance, a positive value means that a too
large distance is measured):
Gas phase
Air/nitrogen
Hydrogen
Water (saturated steam)
Temperature
Pressure
°C
°F
1 bar (14,5 psi)
10 bar (145 psi)
25 bar (362 psi)
20
68
0.00 %
0.22 %
0.58 %
200
392
–0.01 %
0.13 %
0.36 %
400
752
–0.02 %
0.08 %
0.29 %
20
68
–0.01 %
0.10 %
0.25 %
200
392
–0.02 %
0.05 %
0.17 %
400
752
–0.02 %
0.03 %
0.11 %
100
212
0.02 %
180
356
2.1 %
263
505,4
4.15 %
310
590
364
687
With a known, constant pressure, it is possible to compensate for this measured error with a
linearization, for example.
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
23
Installation
Installation conditions
Mounting position
A0016882
• Recommended distance A wall - nozzle outer edge: ~1/6 of the container diameter. However, the
device must not under any circumstances be mounted closer than 15 cm (5.91 in) to the tank wall.
• Not in the center (2) as interference can cause signal loss.
• Not above the filling curtain (3).
• The use of a weather protection cover (1) is recommended to protect the transmitter from direct
sunlight or rain.
24
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Internal container fittings
α
A0031777
Avoid the location of internal fittings (limit switches, temperature sensors, struts, vacuum rings,
heating coils, baffles etc.) inside the signal beam. Take into account the beam angle →  26.
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
25
Avoiding interference echoes
α
A0031813
Metal deflection plates installed at an angle to scatter the radar signals help prevent interference
echoes.
Optimization options
• Antenna size
The larger the antenna the smaller the beam angle α,resulting in fewer interference echoes
→  26.
• Mapping
Measurement can be optimized by electronically suppressing interference echoes.
Beam angle
α
W= 2 . D . tan_
α
A0031824
9
Relationship between beam angle α, distance D and beamwidth diameter W
The beam angle is defined as the angle α where the energy density of the radar waves reaches half
the value of the maximum energy density (3dB width). Microwaves are also emitted outside the
signal beam and can be reflected off interfering installations.
26
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
FMR60
A0032080
Antenna
1)
1)
Drip-off,
PTFE 50 mm / 2"
Beam angle α
6°
Distance (D)
Beamwidth diameter W
5 m (16 ft)
0.52 m (1.70 ft)
10 m (33 ft)
1.05 m (3.44 ft)
15 m (49 ft)
1.57 m (5.15 ft)
20 m (66 ft)
2.10 m (6.89 ft)
25 m (82 ft)
2.62 m (8.60 ft)
30 m (98 ft)
3.14 m (10.30 ft)
35 m (115 ft)
3.67 m (12.04 ft)
40 m (131 ft)
4.19 m (13.75 ft)
45 m (148 ft)
4.72 m (15.49 ft)
50 m (164 ft)
5.24 m (17.19 ft)
Feature 070 in the product structure
Measurement through a ball valve
A0034564
• Measurements can be performed through an open full bore ball valve without any problems.
• At the transitions, no gap may be left exceeding 1 mm (0.04 in).
• Diameter of opening of ball valve must always be equivalent to pipe diameter; avoid edges and
constrictions.
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
27
Measuring through plastic tank ceilings or dielectric windows
• Dielectric constant of the medium: εr ≥ 10
• The distance from the lower edge of the antenna to the tank ceiling or window should be about
100 mm (4 in).
• If possible, avoid mounting locations where condensation or build-up might occur.
• In case of outdoor mounting, the space between antenna and vessel has to be protected from the
elements.
• Do not mount any potential reflectors (e.g. pipes) outside the tank in the signal beam.
Suitable thickness of the tank ceiling or dielectric window
Material
PE
PTFE
DK (= εr)
2.3
2.1
Optimum thickness
1)
Installation: Drip-off
antenna PTFE 50 mm / 2"
1.25 mm
(0.049 in) 1)
1.3 mm (0.051)
1)
PP
Perspex
2.3
3.1
1.25 mm
(0.049 in) 1)
1.07 mm
(0.042 in) 1)
or an integer multiple of this value; however, take into account that the microwave transparency is
significantly reduced when increasing the thickness of the window.
Aligning the antenna axis
Align the antenna vertically to the product surface.
Attention:
The maximum reach of the antenna can be reduced if it is not installed perpendicular to the
product.
Radial alignment of the antenna
Based on the directional characteristic, radial alignment of the antenna is not necessary.
Information concerning nozzles
Hmax
The maximum nozzle length Hmax depends on the nozzle diameter D:
øD
A0032209
28
Inner nozzle diameterD
Maximum nozzle height Hmax 1)
50 to 80 mm (2 to 3.2 in)
750 mm (30 in)
80 to 100 mm (3.2 to 4 in)
1 150 mm (46 in)
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Inner nozzle diameterD
Maximum nozzle height Hmax 1)
100 to 150 mm (4 to 6 in)
1 450 mm (58 in)
≥ 150 mm (6 in)
2 200 mm (88 in)
1)
In case of longer nozzles, a reduction of the measuring performance is to be expected.
Note the following if the antenna does not project out of the nozzle:
• The end of the nozzle must be smooth and free from burrs. The edge of the nozzle should be
rounded if possible.
• Mapping must be performed.
• Please contact Endress+Hauser for applications with nozzles that are higher than indicated in
the table.
Information concerning threaded connections
• When screwing in, turn by the hex bolt only.
• Tool: open-ended wrench 55 mm
• Maximum permissible torque: 50 Nm (36 lbf ft)
Container with heat
insulation
A0032207
If process temperatures are high, the device should be included in the usual container insulation
system (2) to prevent the electronics from heating as a result of thermal radiation or convection. The
insulation should not be higher than the neck of the device (1).
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
29
Environment
Ambient temperature range
Measuring device
–40 to +80 °C (–40 to +176 °F)
Local display
–20 to +70 °C (–4 to +158 °F), the readability of the display may be impaired at
temperatures outside the temperature range.
Outdoor operation in strong sunlight:
• Mount the device in the shade.
• Avoid direct sunlight, particularly in warm climatic regions.
• Use a weather protection cover (see accessories).
The following diagrams only consider functional aspects. Additional restrictions may apply for
certified device versions. See the separate Safety Instructions for more information →  68.
Ambient temperature limits
In the event of temperature (Tp) at the process connection, the permitted ambient temperature (Ta)
is reduced as indicated in the following diagram (temperature derating) in the table header.
FMR60
Ta
P1
Seal 1)
A3:
FKM Viton GLT, –40 to 80 °C (–40 to 176 °F)
P2
P3
Tp
P5
P4
Temperature specifications: °C (°F)
Housing type 2)
A0032024
P1
P2
P3
Tp
Ta
Tp
Ta
A:
GT19 two-chamber
Plastic PBT
–40
(–40)
60
(140)
60
(140)
60
(140)
C:
GT20 two-chamber
Aluminum, coated
1)
2)
P4
Tp
Ta
Tp
P5
Ta
Tp
Ta
80
54
80
(176) (129.2) (176)
–40
(–40)
–40
(–40)
–40
(–40)
–40
76
76
76
80
76
80
(–40) (168.8) (168.8) (168.8) (176) (168.8) (176)
–40
(–40)
–40
(–40)
–40
(–40)
Feature 090 in the product structure
Feature 040 in the product structure
FMR60
Ta
P1
Seal 1)
A4:
FKM Viton GLT,
–40 to 130 °C (–40 to 266 °F)
P2
P2+
P3
Tp
P5
P4
Temperature specifications: °C (°F)
Housing type
2)
P1
Tp
P2
Ta
Tp
P2+
Ta
Tp
Ta
P3
Tp
P4
Ta
Tp
P5
Ta
Tp
Ta
A:
GT19 two-chamber
Plastic PBT
–40
60
60
60 113 45
130
22
130 –40 –40 –40
(–40) (140) (140) (140) (235.4)(113) (266) (71.6) (266) (–40) (–40) (–40)
C:
GT20 two-chamber
Aluminum, coated
–40
76
76
76 110 71
130
60
130 –40 –40 –40
(–40) (168.8)(168.8)(168.8) (230) (159.8)(266) (140) (266) (–40) (–40) (–40)
1)
2)
30
A0032025
Feature 090 in the product structure
Feature 040 in the product structure
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Storage temperature
–40 to +80 °C (–40 to +176 °F)
Climate class
DIN EN 60068-2-38 (test Z/AD)
Altitude according to
IEC61010-1 Ed.3
• Generally up to 2 000 m (6 600 ft) above MSL.
• Above 2 000 m (6 600 ft) if the following conditions are met:
– Ordering feature 020 "Power supply; Output" = A, B, C, E or G (2-wire versions)
– Supply voltage U < 35 V
– Supply voltage of overvoltage category 1
Degree of protection
• With closed housing tested according to:
– IP68, NEMA6P (24 h at 1.83 m under water surface)
– For plastic housing with transparent cover (display module): IP68 (24 h at 1.00 m under water
surface) 3)
– IP66, NEMA4X
• With open housing: IP20, NEMA1
• Display module: IP22, NEMA2
Degree of protection IP68 NEMA6P applies for M12 PROFIBUS PA plugs only when the
PROFIBUS cable is plugged in and is also rated IP68 NEMA6P.
Vibration resistance
DIN EN 60068-2-64 IEC 60068-2-64 at 5 to 2 000 Hz: 1.5 (m/s2)2/Hz
Electromagnetic
compatibility (EMC)
Electromagnetic compatibility in accordance with all of the relevant requirements outlined in the
EN 61326 series and NAMUR Recommendation EMC (NE 21). For details, please refer to the
Declaration of Conformity 4).
A normal device cable suffices if only the analog signal should be used. Use a shielded cable for
digital communication (HART/ PA/ FF).
Maximum measured error during EMC testing: <0.5 % of the span. By way of derogation, for devices
with a plastic housing and see-through lid (integrated display SD02 or SD03) the measured error can
be up to 2 % of the span in the event of strong electromagnetic radiation in the 1 to 2 GHz frequency
range.
3)
4)
This restriction is valid if the following options of the product structure have been selected at the same time: 030 ("Display, Operation") = C
("SD02") or E ("SD03"); 040 ("Housing") = A ("GT19").
Available for download at www.de.endress.com.
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
31
Process
Process temperature, process
pressure
The pressure ranges indicated can be reduced by the choice of process connection. The nominal
pressure (PN) indicated on the nameplate refers to a reference temperature of 20 °C, and of
100 °F for ASME flanges. Observe pressure-temperature dependency.
Please refer to the following standards for the pressure values permitted at higher
temperatures:
• EN 1092-1: 2001 Tab. 18
In terms of their stability-temperature property, the materials 1.4435 and 1.4404 are
grouped in EN 1092-1 table 18 under 13E0. The chemical composition of the two materials
can be identical.
• ASME B 16.5a – 1998 Tab. 2-2.2 F316
• ASME B 16.5a – 1998 Tab. 2.3.8 N10276
• JIS B 2220
FMR60 threaded process connection
[bar] ([psi]) p
16 (232)
0 (0)
-1 (-14.5)
-40
(-40)
(+32)
+130
(+266)
Tp
[°C]
([°F])
A0032614
 10
FMR60: Permitted range for process temperature and process pressure
Feature 90, seal: A4, FKM Viton GLT
FMR60 threaded process connection
Feature 100 "Process
connection"
Feature 90
"Seal"
Process temperature
range
• GGJ:
Thread ISO228 G1-1/2
• RGJ:
Thread ANSI MNPT1-1/2
A4:
FKM Viton GLT
–40 to +130 °C
(–40 to +266 °F)
1)
32
Process pressure range
prel =
–1 to 16 bar
(–14.5 to 232 psi) 1)
The pressure range may be further restricted in the event of a CRN approval
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
FMR60 flange process connection
[bar] ([psi]) p
3 (43.5)
0 (0)
-1 (-14.5)
-40
(-40)
(+32)
+80
(+176)
Tp
+130 [°C]
(+266) ([°F])
A0032615
 11
FMR60: Permitted range for process temperature and process pressure
Process connection: flange PP
Process connection: flange 316L
FMR60 flange process connection
Feature 100 "Process connection"
Process temperature range
• XJG:
UNI flange 3"/DN80/80A, PP
• XKG:
UNI flange 4"/DN100/100A, PP
• XLG:
UNI flange 6"/DN150/150A, PP
–40 to +80 °C (–40 to +176 °F)
• XJJ:
UNI flange 3"/DN80/80A, 316L
• XKJ:
UNI flange4"/DN100/100A, 316L
• XLJ:
UNI flange 6"/DN150/150A, 316L
–40 to +130 °C (–40 to +266 °F)
1)
Dielectric constant
Process pressure range
prel = –1 to 3 bar (–14.5 to 43.5 psi)
pabs < 4 bar (58 psi) 1)
The pressure range may be further restricted in the event of a CRN approval
For liquids
εr ≥ 1.9
Please contact Endress+Hauser for applications with lower dielectric constants than indicated.
For dielectric constants (DC values) of many media commonly used in various industries refer
to:
• the Endress+Hauser DC manual (CP01076F)
• the Endress+Hauser "DC Values App" (available for Android and iOS)
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
33
Mechanical construction
Dimensions
Dimensions of the electronics housing
R110
99 (3.9)*
90 (3.54)
ø10
ø106 (4.17)
113.5 (4.47)
144 (5.67)
6 (4
.17
78 (3.07)
A0011346
 12
Housing GT19 (Plastics PBT); Dimensions in mm (in)
*for devices with integrated overvoltage protection.
R110
98 (3.86)*
90 (3.54)
141.5 (5.57)
145.5 (5.73)
61 (2.4)
ø112 (4.41)
ø108.5 (4.27)
ø10
3.5
(4.
07)
78 (3.07)
A0020751
 13
Housing GT20 (Alu coated); Dimensions in mm (in)
*for devices with integrated overvoltage protection.
34
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
FMR60: Antenna and process connection
78 (3.07)
205 (8.07)
107 (4.21)
127 (5.00)
22 (0.87)
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
ø42.5 (1.67)
ø42.5 (1.67)
A0031560
 14
Endress+Hauser
Dimensions: mm (in)
Process connection: G1-1/2" or MNPT1-1/2" thread
Process connection: UNI flange 3"/DN80/80A to 6"/DN150/150A
Reference point of the measurement
Bottom edge of housing
V. 4, Rev. 2, 19-03-2018
35
UNI flanges for FMR60
G 1 1/2
A-A
45°
20 (0.79)
A0034611
 15
Dimensions: mm (in)
Feature 100: process connection
Suitable for
• XJG:
UNI flange 3"/DN80/80A, PP
• XJJ: UNI flange 3"/DN80/80A, 316L
• 3" 150lbs
• DN80 PN16
• 10K 80A
150 mm (5.9 in)
160 mm (6.3 in)
200 mm (7.9 in)
19 mm (0.75 in)
• XKG:
UNI flange 4"/DN100/100A, PP
• XKJ: UNI flange 4"/DN100/100A, 316L
• 4" 150lbs
• DN100 PN16
• 10K 100A
175 mm (6.9 in)
190.5 mm (7.5 in)
228.6 mm (9 in)
19 mm (0.75 in)
• XLG:
UNI flange 6"/DN150/150A, PP
• XLJ:
UNI flange 6"/DN150/150A, 316L
• 6" 150lbs
• DN150 PN16
• 10K 150A
240 mm (9.4 in)
241.3 mm (9.5 in)
285 mm (11.2 in)
23 mm (0.9 in)
36
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Weight
Housing
Part
Weight
GT19 housing - plastic
Approx. 1.2 kg (2.7 lb)
GT20 housing - aluminum
Approx. 1.9 kg (4.2 lb)
Antenna and process connection
Device
Antenna 1)
Weight of antenna /
process connection
FMR60
GA:
Drip-off, PTFE DN50
Max. 2 kg (4.41 lb) + flange weight 2)
1)
2)
Endress+Hauser
Order code 070
For flange weights (316/316L) see Technical Information TI00426F.
V. 4, Rev. 2, 19-03-2018
37
Materials: GT19 housing
(plastic)
2.1
2.2
5.2
5.1
A0013788
No. Part
Material
PBT
2.1 Cover of the electronics compartment
•
•
•
•
2.2 Cover of the terminal compartment
• Cover: PBT
• Cover seal: EPDM
• Thread-coating: Graphite-based lubricant varnish
• Screw: A4-70
• Clamp: 316L (1.4404)
Lock at the housing neck
Cover glass: PC
Cover frame: PBT-PC
Cover seal: EPDM
Thread-coating: Graphite-based lubricant varnish
5.1 Dummy plug, cable gland, adapter or plug
(depending on the device version)
• Dummy plug, depending on the device version:
– PE
– PBT-GF
• Cable gland, depending on the device version:
– Nickel-plated brass (CuZn)
– PA
• Adapter: 316L (1.4404/1.4435)
• Seal: EPDM
• M12 plug: Nickel-plated brass 1)
• 7/8" plug: 316 (1.4401) 2)
5.2 Dummy plug, cable gland or adapter (depending on
the device version)
• Dummy plug, depending on the device version:
– PE
– PBT-GF
– Nickel-plated steel
• Cable gland, depending on the device version:
– Nickel-plated brass (CuZn)
– PA
• Adapter: 316L (1.4404/1.4435)
• Seal: EPDM
Dummy plug or M12 socket (depending on the
device version)
• Dummy plug: Nickel-plated brass (CuZn)
• M12 socket: Nickel-plated GD-Zn
Pressure relief stopper
Nickel-plated brass (CuZn)
Ground terminal
•
•
•
•
Adhesive nameplate
Plastic
1)
2)
38
Housing
Screw: A2
Spring washer: A4
Clamp: 304 (1.4301)
Holder: 304 (1.4301)
For the version with M12 plug the sealing material is Viton.
For the version with 7/8" plug, the sealing material is NBR.
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Materials: GT20 housing
(die-cast aluminum, powdercoated)
2.1
2.2
5.2
5.1
A0036037
Endress+Hauser
Nr. Part
Material
• Housing: AlSi10Mg(<0,1% Cu)
• Coating: Polyester
Housing, RAL 5012 (blue)
2.1 Cover of the electronics compartment; RAL 7035
(gray)
•
•
•
•
•
Cover: AlSi10Mg(<0,1% Cu)
Window: Glass
Cover seal: NBR
Seal of the window: NBR
Thread-coating: Graphite-based lubricant varnish
2.2 Cover of the terminal compartment; RAL 7035
(gray)
• Cover: AlSi10Mg(<0,1% Cu)
• Cover seal: NBR
• Thread-coating: Graphite-based lubricant varnish
Cover lock
• Screw: A4
• Clamp: 316L (1.4404)
Lock at the housing neck
• Screw: A4-70
• Clamp: 316L (1.4404)
5.1 Dummy plug, cable gland, adapter or plug
(depending on the device version)
• Dummy plug, depending on the device version:
– PE
– PBT-GF
• Cable gland, depending on the device version:
– Nickel-plated brass (CuZn)
– PA
• Adapter: 316L (1.4404/1.4435)
• Seal: EPDM
• M12 plug: Nickel-plated brass 1)
• 7/8" plug: 316 (1.4401) 2)
5.2 Dummy plug, cable gland or adapter (depending on
the device version)
• Dummy plug, depending on the device version:
– PE
– PBT-GF
– Nickel-plated steel
• Cable gland, depending on the device version:
– Nickel-plated brass (CuZn)
– PA
• Adapter: 316L (1.4404/1.4435)
• Seal: EPDM
Dummy plug or M12 socket (depending on the
device version)
• Dummy plug : Nickel-plated brass (CuZn)
• M12 socket: Nickel-plated GD-Zn
Pressure relief stopper
Nickel-plated brass (CuZn)
V. 4, Rev. 2, 19-03-2018
39
Nr. Part
Material
Ground terminal
•
•
•
•
Adhesive nameplate
Plastic
1)
2)
Materials: antenna and
process connection
Screw: A2
Spring washer: A2
Clamp: 304 (1.4301)
Holder: 304 (1.4301)
For the version with M12 plug the sealing material is Viton.
For the version with 7/8" plug, the sealing material is NBR.
FMR60
A0031812
40
No.
Component part
Material
Antenna
Antenna: PTFE
Antenna seal: Viton (FKM)
Antenna adapter
316L / 1.4404
Housing adapter
316L / 1.4404
Housing seal
EPDM
Process connection
316L / 1.4404
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Materials: Weather
protection cover
A0015473
Endress+Hauser
No
Part: Material
Protection cover: 316L (1.4404)
Molded rubber part (4x): EPDM
Clamping screw: 316L (1.4404) + carbon fibre
Bracket: 316L (1.4404)
Ground terminal
• Screw: A4
• Spring washer: A4
• Clamp: 316L (1.4404)
• Holder: 316L (1.4404)
• Washer: A4
• Cheese head screw: A4-70
V. 4, Rev. 2, 19-03-2018
41
Operability
Operating concept
Operator-oriented menu structure for user-specific tasks
• Commissioning
• Operation
• Diagnostics
• Expert level
Operating languages
• English
• Deutsch
• Français
• Español
• Italiano
• Nederlands
• Portuguesa
• Polski
• русский язык (Russian)
• Svenska
• Türkçe
• 中文 (Chinese)
• 日本語 (Japanese)
• 한국어 (Korean)
• Bahasa Indonesia
• tiếng Việt (Vietnamese)
• čeština (Czech)
Feature 500 of the product structure determines which of these languages is preset on delivery.
Quick and save commissioning
• Interactive wizard with graphical interface for easy commissioning via FieldCare/DeviceCare
• Menu guidance with brief explanations of the individual parameter functions
• Standardized operation at the device and in the operating tools
Integrated data storage device (HistoROM)
• Enables transfer of configuration when changing electronic modules
• Records up to 100 event messages in the device
• Records up to 1000 measured values in the device
• Saves the signal curve on commissioning which can later be used as a reference.
Efficient diagnostics increase measurement reliability
• Remedy information is integrated in plain text
• Diverse simulation options and line recorder functions
Integrated Bluetooth module (option for HART devices)
• Easy and fast setup via SmartBlue (app)
• No additional tools or adapters required
• Signal curve via SmartBlue (app)
• Encrypted single point-to-point data transmission (Fraunhofer-Institut, third party, tested)
andpassword-protected communication via Bluetooth® wireless technology
42
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Local operation
Operation
with
Pushbuttons
Touch Control
Order code for
"Display;
Operation"
Option C "SD02"
Option E "SD03"
A0036312
Display
elements
4-line display
A0036313
4-line display
white background lighting; switches to red in
event of device error
Format for displaying measured variables and status variables can be individually configured
Permitted ambient temperature for the display: –20 to +70 °C (–4 to +158 °F)
The readability of the display may be impaired at temperatures outside the temperature
range.
Operating
elements
local operation with 3 push buttons (, , ) external operation via touch control; 3
optical keys: , , 
Operating elements also accessible in various hazardous areas
Additional
functionality
Data backup function
The device configuration can be saved in the display module.
Data comparison function
The device configuration saved in the display module can be compared to the current device
configuration.
Data transfer function
The transmitter configuration can be transmitted to another device using the display module.
Operation with remote
display and operating
module FHX50
A0036314
 16
Endress+Hauser
FHX50 operating options
Display and operating module SD03, optical keys; can be operated through the glass of the cover
Display and operating module SD02, push buttons; cover must be removed
V. 4, Rev. 2, 19-03-2018
43
Operation via Bluetooth®
wireless technology
A0034939
 17
Remote operation
Possibilities for remote operation via Bluetooth® wireless technology
Transmitter power supply unit
Smartphone / tablet with SmartBlue (app)
Transmitter with Bluetooth® wireless technology
Via HART protocol
A0036169
 18
44
Options for remote operation via HART protocol
PLC (programmable logic controller)
Transmitter power supply unit, e.g. RN221N (with communication resistor)
Connection for Commubox FXA191, FXA195 and Field Communicator 375, 475
Field Communicator 475
Computer with operating tool (e.g. DeviceCare/FieldCare , AMS Device Manager, SIMATIC PDM)
Commubox FXA191 (RS232) or FXA195 (USB)
Field Xpert SFX350/SFX370
VIATOR Bluetooth modem with connecting cable
Transmitter
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DeviceCare/FieldCare via service interface (CDI)
A0032466
 19
Endress+Hauser
DeviceCare/FieldCare via service interface (CDI)
Service interface (CDI) of the instrument (= Endress+Hauser Common Data Interface)
Commubox FXA291
Computer with DeviceCare/FieldCare operating tool
V. 4, Rev. 2, 19-03-2018
45
SupplyCare inventory
management software
SupplyCare is a web-based operating program for coordinating the flow of material and information
along the supply chain. SupplyCare provides a comprehensive overview of the levels of
geographically distributed tanks and silos, for instance, providing complete transparency over the
current inventory situation, regardless of time and location.
Based on the measuring and transmission technology installed onsite, the current inventory data are
collected and sent to SupplyCare. Critical levels are clearly indicated and calculated forecasts provide
additional security for material requirements planning.
The main functions of SupplyCare:
Inventory visualization
SupplyCare determines the inventory levels in tanks and silos at regular intervals. It displays current
and historical inventory data and calculated forecasts of future demand. The overview page can be
configured to suit the user's preferences.
Master data management
With SupplyCare you can create and manage the master data for locations, companies, tanks,
products and users, as well as user authorization.
Report Configurator
The Report Configurator can be used to create personalized reports quickly and easily. The reports
can be saved in a variety of formats, such as Excel, PDF, CSV and XML. The reports can be
transmitted in many ways, such as by http, ftp or e-mail.
Event management
Events, such as when levels drop below the safety stock level or plan points, are indicated by the
software. In addition, SupplyCare can also notify pre-defined users by e-mail.
Alarms
If technical problems occur, e.g. connection issues, alarms are triggered and alarm e-mails are sent to
the System Administrator and the Local System Administrator.
Delivery planning
The integrated delivery planning function automatically generates an order proposal if a pre-set
minimum inventory level is undershot. Scheduled deliveries and disposals are monitored
continuously by SupplyCare. SupplyCare notifies the user if scheduled deliveries and disposals are not
going to be met as planned.
Analysis
In the Analysis module, the most important indicators for the inflow and outflow of the individual
tanks are calculated and displayed as data and charts. Key indicators of material management are
automatically calculated and form the basis for optimizing the delivery and storage process.
Geographical visualization
All the tanks and the tank inventories are represented graphically on a map (based on Google Maps).
The tanks and inventory situations can be filtered by tank group, product, supplier or location.
Multi-language support
The multi-language user interface supports 9 languages, thereby enabling global collaboration on a
single platform. The language and settings are recognized automatically using the browser settings.
SupplyCare Enterprise
SupplyCare Enterprise runs by default as a service under Microsoft Windows on an application server
in an Apache Tomcat environment. The operators and administrators operate the application via a
Web browser from their workstations.
46
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
10
A0034288
 20
10
Endress+Hauser
Example of inventory management platform with SupplyCare Enterprise SCE30B
SupplyCare Enterprise (via Web browser)
SupplyCare Enterprise installation
SupplyCare Enterprise on mobile devices (via Web browser)
Ethernet/WLAN/UMTS
Fieldgate FXA42
Power supply 24 V DC
Modbus TCP via Ethernet as server/client
Converter from Modbus to HART Multidrop
HART Multidrop
4 x 4 to 20 mA analog input (2-wire/4-wire)
V. 4, Rev. 2, 19-03-2018
47
Cloud-based application: SupplyCare Hosting
SupplyCare Hosting is offered as a hosting service (software as a service). Here, the software is
installed within the Endress+Hauser IT infrastructure and made available to the user in the Endress
+Hauser portal.
WWW
10
A0034289
 21
10
Example of inventory management platform with SupplyCare Hosting SCH30
SupplyCare Hosting installation in Endress+Hauser data center
PC workstation with Internet connection
Warehouse locations with Internet connection via 2G/3G with FXA42 or FXA30
Warehouse locations with Internet connection with FXA42
Fieldgate FXA42
Power supply 24 V DC
Modbus TCP via Ethernet as server/client
Converter from Modbus to HART Multidrop
HART Multidrop
4 x 4 to 20 mA analog input (2-wire/4-wire)
With SupplyCare Hosting, users do not need to make the initial software purchase or install and run
the IT infrastructure needed. Endress+Hauser constantly update SupplyCare Hosting and enhance
the capability of the software in conjunction with the customer. The hosted version of SupplyCare is
thus always up-to-date and can be customized to meet different customer requirements. Other
services are also offered in addition to the IT infrastructure and the software that is installed in a
secure, redundant Endress+Hauser data center. These services include defined availability of the
global Endress+Hauser Service and Support Organization and defined response times in a service
event.
48
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Certificates and approvals
Currently available certificates and approvals can be called up via the product configurator.
CE mark
The measuring system meets the legal requirements of the applicable EC guidelines. These are listed
in the corresponding EC Declaration of Conformity together with the standards applied.
Endress+Hauser confirms successful testing of the device by affixing to it the CE mark.
RoHS
The measuring system complies with the substance restrictions of the Restriction on Hazardous
Substances Directive 2011/65/EU (RoHS 2).
RCM-Tick marking
The supplied product or measuring system meets the ACMA (Australian Communications and Media
Authority) requirements for network integrity, interoperability, performance characteristics as well
as health and safety regulations. Here, especially the regulatory arrangements for electromagnetic
compatibility are met. The products are labelled with the RCM- Tick marking on the name plate.
A0029561
Ex approval
•
•
•
•
•
•
•
•
•
ATEX
IECEx
CSA
FM
NEPSI
KC
INMETRO
TIIS 5)
EAC
Additional safety instructions must be followed for use in hazardous areas. Please refer to the
separate "Safety Instructions" (XA) document included in the delivery. Reference to the applicable XA
can be found on the nameplate.
For details on the available certificates and associated XAs: →  68
Dual seal according to
ANSI/ISA 12.27.01
The devices have been designed according to ANSI/ISA 12.27.01 as dual seal devices, allowing the
user to waive the use and save the cost of installing external secondary process seals in the conduit
as required by the process sealing sections of ANSI/NFPA 70 (NEC) and CSA 22.1 (CEC) These
instruments comply with the North-American installation practice and provide a very safe and costsaving installation for pressurized applications with hazardous fluids.
Further information can be found in the Safety Instructions (XA) of the relevant devices.
Functional safety
Use for level monitoring (MIN, MAX, range) up to SIL 3 (homogeneous or diverse redundancy),
independently evaluated by TÜV Rheinland in accordance with IEC 61508, refer to the "Functional
Safety Manual" for information.
WHG
WHG approval: No. requested
Pressure equipment with
allowable pressure
≤ 200 bar (2 900 psi)
Pressure instruments with a flange and threaded boss that do not have a pressurized housing do not
fall within the scope of the Pressure Equipment Directive, irrespective of the maximum allowable
pressure.
Reasons:
According to Article 2, point 5 of EU Directive 2014/68/EU, pressure accessories are defined as
"devices with an operational function and having pressure-bearing housings".
5)
in preparation
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
49
If a pressure instrument does not have a pressure-bearing housing (no identifiable pressure chamber
of its own), there is no pressure accessory present within the meaning of the Directive.
Radio standard
EN 302729-1/2
The devices comply with the Level Probing Radar (LPR) radio standard EN 302729-1/2. The devices
are approved for unrestricted use inside and outside closed containers in countries of the EU and the
EFTA . that have already implemented this standard.
The following countries are those that have currently implemented the directive:
Belgium, Bulgaria, Germany, Denmark, Estonia, France, Greece, UK, Ireland, Iceland, Italy,
Liechtenstein, Lithuania, Latvia, Malta, The Netherlands, Norway, Austria, Poland, Portugal,
Romania, Sweden, Switzerland, Slovakia, Spain, Czech Republic and Cyprus.
Implementation is still underway in all of the countries not listed.
Please note the following for operation of the devices outside of closed vessels:
1.
Installation must be carried out by properly trained, expert staff.
2.
The device antenna must be installed in a fixed location pointing vertically downwards.
3.
The installation site must be located at a distance of 4 km from the astronomy stations listed
below or otherwise approval must be provided by the relevant authority. If the device is
installed at a distance of 4 to 40 km from one of the listed stations, it must not be installed at a
height of more than 15 m (49 ft) above the ground.
Astronomy stations
Country
Name of the station
Latitude
Longitude
Germany
Effelsberg
50°31'32" North
06°53'00" East
Finland
Metsähovi
60°13'04" North
24°23'37" East
Tuorla
60°24'56" North
24°26'31" East
Plateau de Bure
44°38'01" North
05°54'26" East
Floirac
44°50'10" North
00°31'37" West
Cambridge
52°09'59" North
00°02'20" East
Damhall
53°09'22" North
02°32'03" West
Jodrell Bank
53°14'10" North
02°18'26" West
Knockin
52°47'24" North
02°59'45" West
Pickmere
53°17'18" North
02°26'38" West
Medicina
44°31'14" North
11°38'49" East
Noto
36°52'34" North
14°59'21" East
Sardinia
39°29'50" North
09°14'40" East
Poland
Fort Skala Krakow
50°03'18" North
19°49'36" East
Russia
Dmitrov
56°26'00" North
37°27'00" East
Kalyazin
57°13'22" North
37°54'01" East
Pushchino
54°49'00" North
37°40'00" East
Zelenchukskaya
43°49'53" North
41°35'32" East
Sweden
Onsala
57°23'45" North
11°55'35" East
Switzerland
Bleien
47°20'26" North
08°06'44" East
Spain
Yebes
40°31'27" North
03°05'22" West
Robledo
40°25'38" North
04°14'57" West
Penc
47°47'22" North
19°16'53" East
France
Great Britain
Italy
Hungary
As a general rule, the requirements outlined in EN 302729-1/2 must be observed.
50
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Radio standard
EN 302372-1/2
The devices comply with the Tanks Level Probing Radar (TLPR) radio standard EN 302372-1/2 and
are approved for use in closed containers. For installation, points a to f in Annex B of EN 302372-1
must be taken into consideration.
FCC
This device complies with Part 15 of the FCC rules. Operation is subject to the following two
conditions: (1) This device may not cause harmful interference, and (2) this device must accept any
interference received, including interference that may cause undesired operation.
[Any] changes or modifications not expressly approved by the party responsible for compliance could
void the user's authority to operate the equipment.
The devices are compliant with the FCC Code of Federal Regulations, CFR 47, Part 15, Sections
15.205, 15.207, 15.209.
In addition, the devices are compliant with Section 15.256 . For these LPR (Level Probe Radar)
applications the devices must be professionally installed in a downward operating position. In
addition, the devices are not allowed to be mounted in a zone of 4 km around RAS stations and
within a radius of 40 km around RAS stations the maxium operation height of devices is 15 m (49 ft)
above ground.
Industry Canada
Canada CNR-Gen Section 7.1.3
This device complies with Industry Canada licence-exempt RSS standard(s). Operation is subject to
the following two conditions: (1) This device may not interference, and (2) this device must accept
any interference, including interference that may cause undesired operation of the device.
Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts
de licence. L'exploitation est autorisée aux deux conditions suivantes : (1) l'appareil ne doit pas
produire de brouillage, et (2) l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique
subi, même si le brouillage est susceptible d'en compromettre le fonctionnement.
[Any] changes or modifications not expressly approved by the party responsible for compliance could
void the user's authority to operate the equipment.
• The installation of the LPR/TLPR device shall be done by trained installers, in strict compliance
with the manufacturer’s instructions.
• The use of this device is on a “no-interference, no-protection” basis. That is, the user shall accept
operations of high-powered radar in the same frequency band which may interfere with or
damage this device. However, devices found to interfere with primary licensing operations will be
required to be removed at the user’s expense.
• This device shall be installed and operated in a completely enclosed container to prevent RF
emissions, which can otherwise interfere with aeronautical navigation.
• The installer/user of this device shall ensure that it is at least 10 km from the Dominion
Astrophysical Radio Observatory (DRAO) near Penticton, British Columbia. The coordinates of the
DRAO are latitude 49°19ʹ15ʹʹ N and longitude 119°37ʹ12ʹʹ W. For devices not meeting this 10 km
separation (e.g., those in the Okanagan Valley, British Columbia,) the installer/user must
coordinate with, and obtain the written concurrence of, the Director of the DRAO before the
equipment can be installed or operated. The Director of the DRAO may be contacted at
250-497-2300 (tel.) or 250-497-2355 (fax). (Alternatively, the Manager, Regulatory Standards
Industry Canada, may be contacted.)
The model FMR60L is a submodel of the FMR60 that fulfills the requirements for use as LPR
(Level Probe Radar).
CRN approval
Some device versions have CRN approval. Devices are CRN approved if the following two conditions
are met:
• The device has a CSA approval (product structure: feature 010 "Approval")
• The device has a CRN-approved process connection as per the following table:
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
51
Feature 100 in the product structure
RGJ
Process connection
Thread ANSI MNPT1-1/2, 316L
• Process connections that do not have CRN approval are not listed in this table.
• Please refer to the product structure to discover which process connections are available for a
specific device type.
• CRN-approved devices bear the registration number (tbd.) on the nameplate.
An additional pressure restriction applies for the device versions listed in the table below if they
are CRN-approved. The pressure range indicated in the "Process" section →  32 continues to
apply for device versions not listed in the table below (in preparation).
52
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Test, certificate
Feature 580 "Test, certificate" Description
JA
3.1 Material certificate, wetted metal parts, EN10204-3.1 inspection
certificate
Test reports, declarations and inspection certificates are available in electronic format in the
W@M Device Viewer:
Enter the serial number from nameplate (www.endress.com/deviceviewer)
This concerns the options for the following order codes:
• 550 "Calibration"
• 580 "Test, certificate"
Hard-copy product
documentation
Endress+Hauser
Hard-copy versions of the test reports, declarations and inspection certificates can also be order via
order code 570 "Service", option I7 "Hard-copy product documentation". The documents are then
supplied with the product.
V. 4, Rev. 2, 19-03-2018
53
Other standards and
guidelines
54
• EN 60529
Degrees of protection provided by enclosures (IP code)
• EN 61010-1
Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use
• IEC/EN 61326
"Emission in accordance with Class A requirements". Electromagnetic compatibility (EMC
requirements).
• NAMUR NE 21
Electromagnetic compatibility (EMC) of industrial process and laboratory control equipment
• NAMUR NE 43
Standardization of the signal level for the breakdown information of digital transmitters with
analog output signal.
• NAMUR NE 53
Software of field devices and signal-processing devices with digital electronics
• NAMUR NE 107
Status classification as per NE107
• NAMUR NE 131
Requirements for field devices for standard applications.
• IEC61508
Functional safety of safety-related electric/electronic/programmable electronic systems
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Ordering information
Ordering information
Detailed ordering information is available from the following sources:
• In the Product Configurator on the Endress+Hauser website: www.endress.com -> Click "Corporate"
-> Select your country -> Click "Products" -> Select the product using the filters and search field ->
Open product page -> The "Configure" button to the right of the product image opens the Product
Configurator.
• From your Endress+Hauser Sales Center: www.addresses.endress.com
Product Configurator - the tool for individual product configuration
• Up-to-the-minute configuration data
• Depending on the device: Direct input of measuring point-specific information such as
measuring range or operating language
• Automatic verification of exclusion criteria
• Automatic creation of the order code and its breakdown in PDF or Excel output format
• Ability to order directly in the Endress+Hauser Online Shop
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
55
The following points must be considered if option F3 (3-point linearity protocol) was selected in
feature 550 ("Calibration").
3-point linearity protocol
The 3 points of the linearity protocol are defined as follows:
500 (19.7)
6000 (236)
A0032642
 22
Points of the 3-point linearity protocol; engineering unit: mm (in)
Distance from the reference point R to the first measuring point
Reference point of measurement
First measuring point
Second measuring point (in the middle between the first and third measuring point)
Third measuring point
Measuring point
Position
1st measuring point
• At a distance A from the reference point
• A = length of antenna + 500 mm (19.7 in)
• Minimum distance: Amin = 1 000 mm (39.4 in)
2nd measuring point
In the middle between the 1st and 3rd measuring point
3rd measuring point
6 000 mm (236 in) below the reference point R
The position of the measuring points can vary by ±1 cm (±0.04 in).
The linearity check is performed under reference operating conditions.
56
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
5-point linearity protocol
The following points must be considered if option F4 (5-point linearity protocol) was selected in
feature 550 ("Calibration").
The 5 points of the linearity protocol are evenly distributed over the measuring range (0% - 100%).
Empty calibration (E) and Full calibration (F) must be specified in order to define the measuring
range 6).
The following restrictions must be considered when selecting E and F:
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
A0032643
Minimum distance between reference point R and
100% mark
Minimum span
Maximum value for "Empty
calibration"
A ≥ length of antenna + 200 mm (8 in)
Minimum value: 400 mm (16 in)
F ≥400 mm (16 in) E ≤24 m (79 ft)
The linearity check is performed under reference operating conditions.
The selected values for Empty calibration and Full calibration are only used to create the
linearity protocol. Afterwards, the values are reset to the default values specific for the
antenna. If values other than the default values are required, they must be ordered as a
customized parameterization →  58.
6)
If (E) and (F) are not specified, antenna-dependent default values will be used instead.
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
57
Customer-specific
configuration
Tagging (TAG)
If the option IJ: "Customized parameterization HART", IK "Customized parameterization PA" or IL
"Customized parameterization FF" has been selected in feature 570 "Service", presettings that differ
from the default settings can be selected for the following parameters:
Parameters
Communication
protocol
Picklist / range of values
Setup → Unit of length
• HART
• PA
• FF
•
•
•
•
Setup → Empty calibration
• HART
• PA
• FF
max. 70 m (230 ft)
Setup → Full calibration
• HART
• PA
• FF
max. <70 m (230 ft)
Setup → Extended setup → Curr. output 1 or 2 → Damping
HART
0 to 999.9 s
Setup → Extended setup → Curr. output 1 or 2 → Failure
mode
HART
• Min
• Max
• Last valid value
Expert → Comm. → HART config. → Burst mode
HART
• Off
• On
Ordering feature
895: Marking
Option
Z1: Tagging (TAG), see additional spec.
Position of the measuring
point marking
To be selected in the additional specifications:
• Tag plate Stainless Steel
• Self-adhesive paper label
• Supplied label/plate
• RFID TAG
• RFID TAG + Tag plate Stainless Steel
• RFID TAG + Self-adhesive paper label
• RFID TAG + Supplied label/plate
Definition of the measuring
point designation
To be defined in the additional specifications:
3 lines containing up to 18 characters each
in
ft
mm
The measuring point designation appears on the selected label and/or the
RFID TAG.
Designation in the Electronic The first 32 characters of the measuring point designation
Name Plate (ENP)
Designation on the display
module
Services
7)
58
The first 12 characters of the measuring point designation
The following services can be selected via the product structure in the Product Configurator 7):
• PWIS-free (PWIS = paint-wetting impairment substances)
• Customized parameterization HART →  58
• Customized parameterization PA →  58
• Customized parameterization FF →  58
• W/o tooling DVD (FieldCare)
• Hard-copy product documentation
Feature 570 in the product structure
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Application Packages
Heartbeat Diagnostics
Availability
Available in all device versions.
Function
• Continuous self-monitoring of the device.
• Diagnostic messages output to
– the local display.
– an asset management system (e.g. FieldCare/DeviceCare).
– an automation system (e.g. PLC).
Advantages
• Device condition information is available immediately and processed in time.
• The status signals are classified in accordance with VDI/VDE 2650 and NAMUR recommendation
NE 107 and contain information about the cause of the error and remedial action.
Detailed description
See Operating Instructions of the device (→  68); chapter "Diagnostics and trouble shooting"
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
59
Heartbeat Verification
Availability
Available for the following options of feature 540 "Application Package":
• EH: Heartbeat Verification + Monitoring
• EJ: Heartbeat Verification
Device functionality checked on demand
•
•
•
•
Verification of the correct functioning of the measuring device within specifications.
The verification result provides information about the condition of the device: Passed or Failed.
The results are documented in a verification report.
The automatically generated report supports the obligation to demonstrate compliance with
internal and external regulations, laws and standards.
• Verification is possible without interrupting the process.
Advantages
• No onsite presence is required to use the function.
• The DTM 8) triggers verification in the device and interprets the results. No specific knowledge is
required on the part of the user.
• The verification report can be used to prove quality measures to a third party.
• Heartbeat Verification can replace other maintenance tasks (e.g. periodic check) or extend the
test intervals.
SIL/WHG-locked devices 9)
• The Heartbeat Verification module contains a wizard for the proof test which must be performed
at appropriate intervals for the following applications:
– SIL (IEC61508/IEC61511)
– WHG (German Water Resources Act)
• To perform a proof test, the device must be locked (SIL/WHG locking).
• The wizard can be used via FieldCare, DeviceCare or a DTM-based process control system.
In the case of SIL-locked and WHG-locked devices, it is not possible to perform verification
without additional measures (e.g. by-passing of the output current) because the output current
must be simulated (Increased safety mode) or the level must be approached manually (Expert
mode) during subsequent re-locking (SIL/WHG locking).
Detailed description
SD01870F
8)
9)
60
DTM: Device Type Manager; controls device operation via DeviceCare, FieldCare or a DTM-based process control system.
Only relevant for devices with SIL or WHG approval: order code 590 ("Additional approval"), option LA ("SIL") or LC ("WHG").
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Heartbeat Monitoring
Availability
Available for the following options of feature 540 "Application Package":
EH: Heartbeat Verification + Monitoring
Function
• In addition to the verification parameters, the corresponding parameter values are also logged.
• Existing measured variables, such as the echo amplitude, are used in the Foam detection and
Build-up detection wizards.
"Foam detection" wizard
• The Heartbeat Monitoring module contains the Foam detection wizard.
• This wizard is used to configure automatic foam detection, which detects foam on the product
surface on the basis of the reduced signal amplitude. Foam detection can be linked to a switch
output in order to control a sprinkler system, for example, which dissolves the foam.
• This wizard can be used via FieldCare, DeviceCare or a DTM-based process control system.
"Build-up detection" wizard
• The Heartbeat Monitoring module contains the Build-up detection wizard.
• The wizard is used to configure automatic buildup detection, which detects the buildup of deposits
on the antenna on the basis of the increased area of the coupling signal. Buildup detection can be
linked to a switch output in order to control a compressed air system, for example, to clean the
antenna.
• This wizard can be used via FieldCare, DeviceCare or a DTM-based process control system.
Advantages
• Early detection of changes (trends) to ensure plant availability and product quality.
• Use of information for the proactive planning of measures (e.g. cleaning/maintenance).
• Identification of undesirable process conditions as the basis to optimizing the facility and the
processes.
• Automated control of measures to remove foam or buildup.
Detailed description
SD01870F
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
61
Accessories
Device-specific accessories
Accessory
Weather protection cover
Description
222 (8.74)
Weather protection cover
A0015466
54
(2
.13
35°
156
38
65°
(6.1
4)
°
(1.
275 (10.8)
164 (6.46)
55
5)
300 (11.8)
255.1 (10)
A0015472
 23
Weather protection cover; Dimensions: mm (in)
The weather protection cover can be ordered together with the device (product structure, feature 620 "Accessory
 Enclosed",
option PB "Weather Protection Cover").
Alternatively, it can be separately ordered as an accessory; order code 71162242.
62
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Mounting bracket adjustable
Accessories
Description
Mounting bracket
adjustable
99.5 (3.9)
260 (10.2)
48(1.9)
65 (2.6)
122 (4.8)
9(0.4)
ø9(0.4)
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
12(0.5)
35(1.4)
100 (3.9)
16 (0.6)
A0032295
Material: 304 (1.4301)
Suitable for housing 1):
– A: GT19 dual compartment, plastic PBT
– C: GT20 dual compartment, Alu, coated
Suitable for antenna 2):
GA: Drip-off, PTFE DN50
Suitable for process connection 3):
– GGJ: thread ISO228 G1-1/2, 316L
– RGJ: thread ANSI MNPT1-1/2, 316L
Order number: 71336522
There is no conductive connection between the mounting bracket and the transmitter housing.
 •• Risk
of electrostatic charge.
• Integrate the mounting bracket in the local potential equalization system.
• Only to be fastened at stable materials (e.g. metal, brick, concrete) using suitable fastening material (to be
supplied by customer).
1)
2)
3)
Feature 040 in the product structure
Feature 070 in the product structure
Feature 100 in the product structure
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
63
Remote display FHX50
Accessories
Description
Remote display FHX50
1.
2.
3.
4.
A0019128
• Material:
– Plastic PBT
– 316L/1.4404
– Aluminum
• Degree of protection: IP68 / NEMA 6P and IP66 / NEMA 4x
• Suitable for display modules:
– SD02 (push buttons)
– SD03 (touch control)
• Connecting cable:
– Cable supplied with device up to 30 m (98 ft)
– Standard cable supplied by customer up to 60 m (196 ft)
• Ambient temperature range: –40 to 80 °C (–40 to 176 °F)
• Ambient temperature range (option): –50 to 80 °C (–58 to 176 °F) 1)
display should be used, order the device version "Prepared for display FHX50" (feature 030, version
 • IfL ortheM).remote
For the FHX50, you must select option A: "Prepared for display FHX50" under feature 050 "Measuring
device version".
• If the device version "Prepared for display FHX50" was not originally ordered and a FHX50 display is to be
retrofitted, you must select version B "Not prepared for display FHX50" under feature 050: "Measuring device
version" when ordering the FHX50. In this case, a retrofit kit for the device is supplied with the FHX50. The kit
can be used to prepare the device so that the FHX50 can be used.
Use of the FHX50 may be restricted for transmitters with an approval. A device can only be retrofitted with the
 FHX50
if the option L or M ("Prepared for FHX50") is listed under Basic specifications, item 4 "Display, operation" in
the Safety Instructions (XA) for the device.
Also pay attention to the Safety Instructions (XA) of the FHX50.
is not possible on transmitters with:
 Retrofitting
• An approval for use in areas with flammable dust (dust ignition-proof approval)
• Ex nA type of protection
 For details, see document SD01007F.
1)
64
This range is valid if option JN "Ambient temperature transmitter –50 °C (–58 °F)" has been selected in ordering feature 580 "Test, Certificate". If
the temperature is permanently below –40 °C (–40 °F), failure rates may be increased.
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Overvoltage protection
Accessory
Description
Overvoltage protection for
2-wire-devices
OVP10 (1 channel)
OVP20 (2 channel)
A0021734
Technical data
• Resistance per channel: 2 * 0.5 Ωmax
• Threshold DC voltage: 400 to 700 V
• Threshold impulse voltage: < 800 V
• Capacitance at 1 MHz: < 1.5 pF
• Nominal arrest impulse voltage (8/20 μs): 10 kA
• Suited for wire cross-sections: 0.2 to 2.5 mm2 (24 to 14 AWG)
with device
 Ordering
The overvoltage protection module is preferably ordered with the device. See product structure, feature 610
"Accessory mounted", option NA "Overvoltage protection". Separate ordering of the module is only necessary if a
device is to retrofitted with the overvoltage protection.
code for retrofitting
 Order
• For 1-channel devices (feature 020, option A)
OVP10: 71128617
• For 2-channel devices (feature 020, option B, C, E or G)
OVP20 : 71128619
Hosuing lid for retrofitting
In order to keep the necessary safety distances, the housing lid needs to be replaced if the device is retrofitted with
the overvoltage protection. Depending on the housing type, the order code of the suitable lid is as follows:
• GT18 housing: Lid 71185516
• GT19 housing: Lid 71185518
• GT20 housing: Lid 71185516
for retrofitting
 Restrictions
Depending on the approval of the transmitter the usage of the OVP module may be restricted. A device may only be
retrofitted with an OVP module if the option NA (overvoltage protection) is quoted unter Optional Specifications in
the Safety Instructions (XA) pertaining to the device.
 For details refer to SD01090F.
Gas-tight feedthrough
Accessories
Description
Gas-tight feedthrough
Chemically inert glass feedthrough; prevents gases from entering the electronics housing
To order with the device: product structure, feature 610 "Accessory mounted", option NC "Gas-tight feedthrough"
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
65
Communication-specific
accessories
Accessory
Description
Commubox FXA195
HART
For intrinsically safe HART communication with FieldCare via the USB interface.
Accessory
Description
Commubox FXA291
Connects Endress+Hauser field devices with CDI interface (= Endress+Hauser
Common Data Interface) to the USB interface of a computer.
Order code: 51516983
 For details refer to Technical Information TI00404F
 For details refer to Technical Information TI00405C
Accessory
Description
HART Loop Converter
HMX50
Evaluates the dynamic HART variables and converts them to analog current signals
or limit values.
Order code: 71063562
details refer to Technical Information TI00429F and Operating
 For
Instructions BA00371F
Accessory
Description
WirelessHART Adapter
SWA70
Connects field devices to a WirelessHART network.
The WirelessHART adapter can be mounted directly at a HART device and is easly
integrated into an existing HART network. It ensures safe data transmission and
can be operated in parallel with other wireless networks.
 For details refer to Operating Instructions BA00061S
Accessories
Description
Connect Sensor Fully integrated, battery-powered gateway for simple applications with SupplyCare Hosting.
FXA30/FXA30B Up to 4 field devices with 4 to 20 mA communication (FXA30/FXA30B), serial Modbus
(FXA30B) or HART (FXA30B) can be connected. With its robust design and ability to run for
years on the battery, it is ideal for remote monitoring in isolated locations. Version with LTE
(USA, Canada and Mexico only) or 3G mobile transmission for worldwide communication.
For details, see "Technical Information" TI01356S and Operating Instructions
 BA01710S.
Accessories
Description
Fieldgate
FXA42
Fieldgates enable communication between connected 4 to 20 mA, Modbus RS485 and
Modbus TCP devices and SupplyCare Hosting or SupplyCare Enterprise. The signals are
transmitted either via Ethernet TCP/IP, WLAN or mobile communications (UMTS). Advanced
automation capabilities are available, such as an integrated Web-PLC, OpenVPN and other
functions.
For details, see "Technical Information" TI01297S and Operating Instructions
 BA01778S.
Accessories
Description
SupplyCare
Enterprise
SCE30B
Inventory management software that visualizes levels, volumes, masses, temperatures,
pressures, densities or other tank parameters. The parameters are recorded and transmitted
by means of gateways of the type Fieldgate FXA42.
This Web-based software is installed on a local server and can also be visualized and
operated with mobile terminals such as a smartphone or tablet.
For details, see "Technical Information" TI01228S and Operating Instructions
 BA00055S
66
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Accessories
Description
SupplyCare
Hosting SCH30
Inventory management software that visualizes levels, volumes, masses, temperatures,
pressures, densities or other tank parameters. The parameters are recorded and transmitted
by means of gateways of the type Fieldgate FXA42, FXA30 and FXA30B.
SupplyCare Hosting is offered as a hosting service (Software as a Service, SaaS). In the
Endress+Hauser portal, the user is provided with the data over the Internet.
For details, see "Technical Information" TI01229S and Operating Instructions
 BA00050S.
Accessory
Description
Field Xpert SFX350
Field Xpert SFX350 is a mobile computer for commissioning and maintenance. It
enables efficient device configuration and diagnostics for HART and FOUNDATION
fieldbus devices in the non-Ex area.
 For details, see Operating Instructions BA01202S
Accessory
Description
Field Xpert SFX370
Field Xpert SFX370 is a mobile computer for commissioning and maintenance. It
enables efficient device configuration and diagnostics for HART and FOUNDATION
fieldbus devices in the non-Ex area and the Ex area.
 For details, see Operating Instructions BA01202S
Service-specific accessories
Accessory
Description
DeviceCare SFE100
Configuration tool for HART, PROFIBUS and FOUNDATION Fieldbus devices
 Technical Information TI01134S
is available for download at
 • DeviceCare
www.software-products.endress.com. The download requires a
registration in the Endress+Hauser software portal.
• Alternatively, a DeviceCare DVD can be ordered with the device. Product
structure: Feature 570 "Service", Option IV "Tooling DVD (DeviceCare
Setup)".
FieldCare SFE500
FDT-based Plant Asset Management tool.
Helps to configure and maintain all field devices of your plant. By supplying status
information it also supports the diagnosis of the devices.
 Technical Information TI00028S
System components
Accessory
Description
Graphic Data Manager
Memograph M
The graphic data manager Memograph M provides information on all the relevant
process variables. Measured values are recorded correctly, limit values are
monitored and measuring points analyzed. The data are stored in the 256 MB
internal memory and also on an SD card or USB stick.
details refer to Technical Information TI00133R and Operating
 For
Instructions BA00247R
RN221N
Active barrier with power supply for safe separation of 4 to 20 mA current circuits.
Provides bi-directional HART transmission.
details refer to Technical Information TI00073R and Operating
 For
Instructions BA00202R
RNS221
Transmitter supply for 2-wire sensors or transmitters exclusively for non-Ex areas.
Provides bi-directional communication using the HART communication sockets.
details refer to Technical Information TI00081R and Operating
 For
Instructions KA00110R
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
67
Supplementary documentation
For an overview of the scope of the associated Technical Documentation, refer to the following:
• The W@M Device Viewer: enter the serial number from the nameplate
(www.endress.com/deviceviewer)
• The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the
2-D matrix code (QR code) on the nameplate.
The following document types are available:
In the Download Area of the Endress+Hauser Internet site: www.endress.com → Downloads
Standard documentation
Micropilot FMR60
Correlation of documentations to the device:
Device
Power supply, Communication
output 1)
Document type
Document code
FMR60
A, B, C, K, L
Operating Instructions
BA01618F
Brief Operating Instructions
KA01251F
Description of Device
Parameters
GP01101F
1)
Safety Instructions (XA)
HART
Feature 020 in the product structure
Depending on the approval, the following Safety Instructions (XA) are supplied with the device. They
are an integral part of the Operating Instructions.
The nameplate indicates the Safety Instructions (XA) that are relevant to the device.
Feature 010
Approval
Feature 020 "Power Supply; Output"
68
1)
B 2)
C 3)
BA
ATEX II 1G Ex ia IIC T6 Ga
XA01549F
XA01549F
XA01549F
BB
ATEX II 1/2G Ex ia IIC T6 Ga/Gb
XA01549F
XA01549F
XA01549F
BC
ATEX II 1/2G Ex ia/db [ia Ga] IIC T6 Ga/Gb
XA01552F
XA01552F
XA01552F
BG
ATEX II 3G Ex ec IIC T6 Gc
XA01551F
XA01551F
XA01551F
BH
ATEX II 3G Ex ic IIC T6 Gc
XA01551F
XA01551F
XA01551F
B2
ATEX II 1/2G Ex ia IIC T6 Ga/Gb, 1/2D Ex ia IIIC T85°C Da/Db
XA01555F
XA01555F
XA01555F
B3
ATEX II 1/2G Ex ia/db [ia Ga] IIC T6, Ga/Gb 1/2D Ex ta/tb IIIC T85°C Da/Db
XA01556F
XA01556F
XA01556F
B4
ATEX II 1/2G Ex ia IIC T6 Ga/Gb, Ex ia/db [ia Ga] IIC T6 Ga/Gb
XA01553F
XA01553F
XA01553F
CB
CSA IS Cl.I Div.1 Gr.A-D
XA01612F
XA01612F
XA01612F
CC
CSA XP Cl.I Div.1 Gr.A-D [Ex ia]
XA01613F
XA01613F
XA01613F
C2
CSA IS Cl.I,II,III Div.1 Gr.A-G, Ex ia, NI Cl.1 Div.2 [Ex ia]
XA01612F
XA01612F
XA01612F
C3
CSA XP Cl.I,II,III Div.1 Gr.A-G, Zn0/1, NI Cl.I Div.2 [Ex ia]
XA01613F
XA01613F
XA01613F
FA
FM IS Cl.I Div.1 Gr.A-D
XA01615F
XA01615F
XA01615F
FB
FM IS Cl.I,II,III Div.1 Gr.A-G, AEx ia, NI Cl.1 Div.2
XA01615F
XA01615F
XA01615F
FC
FM XP-IS Cl.I Div.1 Gr.A-D, AIS Cl.I Div.1 Gr.A-D
XA01616F
XA01616F
XA01616F
FD
FM XP-IS Cl.I Div.1 Gr.A-D, Zn0/1, DIP-IS Cl.II,III Div.1 Gr.E-G, NI Cl.I Div.2
XA01616F
XA01616F
XA01616F
GA
EAC 0Ex ia IIC T6...T3 Ga X
XA01617F
XA01617F
XA01617F
GB
EAC Ga/Gb Ex ia IIC T6...T3 X
XA01617F
XA01617F
XA01617F
GC
EAC Ga/Gb Ex ia/db [ia Ga] IIC T6...T3 X
XA01618F
XA01618F
XA01618F
IA
IEC Ex ia IIC T6 Ga
XA01549F
XA01549F
XA01549F
IB
IEC Ex ia IIC T6 Ga/Gb
XA01549F
XA01549F
XA01549F
V. 4, Rev. 2, 19-03-2018
Endress+Hauser
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
Micropilot FMR60
Feature 010
Approval
Feature 020 "Power Supply; Output"
1)
B 2)
C 3)
IC
IEC Ex ia/db [ia Ga] IIC T6 Ga/Gb
XA01552F
XA01552F
XA01552F
IG
IEC Ex ec IIC T6 Gc
XA01551F
XA01551F
XA01551F
IH
IEC Ex ic IIC T6 Gc
XA01551F
XA01551F
XA01551F
I2
IEC Ex ia IIC T6 Ga/Gb, Ex ia IIIC T85°C Da/Db
XA01555F
XA01555F
XA01555F
I3
IEC Ex ia/db [ia Ga] IIC T6 Ga/Gb, Ex ta/tb IIIC T85°C Da/Db
XA01556F
XA01556F
XA01556F
I4
IEC Ex ia IIC T6 Ga/Gb, Ex ia/db [ia Ga] IIC T6 Ga/Gb
XA01553F
XA01553F
4)
XA01631F
XA01553F
4)
XA01631F 4)
JA
JPN Ex ia IIC T6 Ga
XA01631F
JB
JPN Ex ia IIC T6 Ga/Gb
XA01631F 4)
XA01631F 4)
XA01631F 4)
JC
JPN Ex d [ia] IIC T6 Ga/Gb
XA01632F 4)
XA01632F 4)
XA01632F 4)
JG
JPN Ex nA IIC T6 Gc
XA01725F 4)
XA01725F 4)
XA01725F 4)
JH
JPN Ex ic IIC T6 Gc
XA01725F 4)
XA01725F 4)
XA01725F 4)
J2
JPN Ex ia IIC T6 Ga/Gb, JPN Ex ia IIIC T85°C Da/Db
XA01728F 4)
XA01728F 4)
XA01728F 4)
J3
JPN Ex d [ia] IIC T6 Ga/Gb, JPN Ex ta/tb IIIC T85°C Da/Db
XA01729F 4)
XA01729F 4)
XA01729F 4)
J4
JPN Ex ia IIC T6 Ga/Gb, JPN Ex d [ia] IIC T6 Ga/Gb
XA01726F 4)
XA01726F 4)
XA01726F 4)
KA
KC Ex ia IIC T6 Ga
XA01623F 4)
XA01623F 4)
XA01623F 4)
KB
KC Ex ia IIC T6 Ga/Gb
XA01623F 4)
XA01623F 4)
XA01623F 4)
KC
KC Ex ia/db [ia Ga] IIC T6 Ga/Gb
XA01624F 4)
XA01624F 4)
XA01624F 4)
MA
INMETRO Ex ia IIC T6 Ga
XA01620F
XA01620F
XA01620F
MB
INMETRO Ex ia IIC T6 Ga/Gb
XA01620F
XA01620F
XA01620F
MC
INMETRO Ex ia/db [ia Ga] IIC T6 Ga/Gb
XA01622F
XA01622F
XA01622F
MG
INMETRO Ex ec IIC T6 Gc
XA01621F
XA01621F
XA01621F
MH
INMETRO Ex ic IIC T6 Gc
XA01621F
XA01621F
XA01621F
NA
NEPSI Ex ia IIC T6 Ga
XA01625F
XA01625F
XA01625F
NB
NEPSI Ex ia IIC T6 Ga/Gb
XA01625F
XA01625F
XA01625F
NC
NEPSI Ex ia/d [ia Ga] IIC T6 Ga/Gb
XA01627F
XA01627F
XA01627F
NG
NEPSI Ex nA IIC T6 Gc
XA01626F
XA01626F
XA01626F
NH
NEPSI Ex ic IIC T6 Gc
XA01626F
XA01626F
XA01626F
N2
NEPSI Ex ia IIC T6 Ga/Gb, NEPSI Ex iaD 20/21 T85
XA01629F
XA01629F
XA01629F
N3
NEPSI Ex ia/d [ia Ga] IIC T6 Ga/Gb, NEPSI Ex tD A20/A21 IP6X T85°C
XA01630F
XA01630F
XA01630F
8A
FM/CSA IS+XP-IS Cl.I,II,III Div.1 Gr.A-G, AIS Cl.I,II,III Div.1 Gr.A-G
XA01612F
XA01615F
XA01616F
XA01612F
XA01615F
XA01616F
XA01612F
XA01615F
XA01616F
* 4)
1)
2)
3)
4)
2-wire; 4-20mA HART
2-wire; 4-20mA HART, switch output
2-wire; 4-20mA HART, 4-20mA
in preparation
Endress+Hauser
V. 4, Rev. 2, 19-03-2018
69
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
71394723
www.addresses.endress.com
DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT DRAFT
*71394723*

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.7
Linearized                      : Yes
Author                          : 00
Create Date                     : 2018:03:19 15:09:27+01:00
Keywords                        : Technical, Information;, Micropilot, FMR60;, Free, space, radar
Modify Date                     : 2018:03:19 15:11:05+01:00
Subject                         : Free space radar
Language                        : EN
XMP Toolkit                     : Adobe XMP Core 5.2-c001 63.139439, 2010/09/27-13:37:26
Format                          : application/pdf
Creator                         : 00
Description                     : Free space radar
Title                           : Micropilot FMR60
Creator Tool                    : AH XSL Formatter V6.2 MR13 for Windows (x64) : 6.2.15.22885 (2015/12/14 12:50JST)
Metadata Date                   : 2018:03:19 15:11:05+01:00
Producer                        : Antenna House PDF Output Library 6.2.735 (Windows (x64))
Trapped                         : False
Document ID                     : uuid:166b1183-cee4-4a59-b39a-322ad8f7e261
Instance ID                     : uuid:37b458cf-df11-4253-aaec-40007a99c7b0
Page Mode                       : UseOutlines
Page Count                      : 72
EXIF Metadata provided by EXIF.tools
FCC ID Filing: LCGFMR6XEL

Navigation menu