3D IMAGING and SIMULATIONS CRSCANNER2 Computed Radiography Scanner User Manual
3D IMAGING & SIMULATIONS CORP. Computed Radiography Scanner
User manual
User Manual
Dental Computed Radiography Reader
Doc No. : TM-801-EN
Rev 0.1.2 Nov. 2012
Part No. : CR-FPM-11-001-EN
3DISC, FireCR, Quantor and the 3D Cube are trademarks of 3D Imaging & Simulations Corp,
South Korea, and its affiliates. All other trademarks are held by their respective owners and
are used in an editorial fashion with no intention of infringement. The data in this publication
are for illustration purposes only and do not necessarily represent standards or specifications,
which must be met by 3D Imaging & Simulations Corp. All information contained herein is
intended for guidance purposes only, and characteristics of the products and services described
in this publication can be changed at any time without notice. Products and services may not
be available in your local area. Please contact your local sales representative for availability
information. 3D Imaging & Simulations Corp. strives to provide as accurate information as
possible, but shall not be responsible for any typographical error.
© Copyright 2010 3D Imaging & Simulations Corp, all rights reserved, printed, and published
in South Korea by 3D Imaging & Simulations Corp.
User Manual
FireCR Dental TM-801-EN
2
3DISC Americas
22560 Glenn Dr, Suite 116
Sterling, VA 20164 USA
Tel : 1-703-430-6080
E-mail : info@3DISCimaging.com
Contact
815, Tamnip-Dong, Yuseong-Gu,
Daejeon, Korea
Tel : 82-42-931-2100
Fax : 82-42-931-2299
Website : www.3DISCimaging.com
E-mail : info@3DISCimaging.com
3DISC Europe
Gydevang, 39-41, 3450 Alleroed, Denmark
Tel : 45-88-276-650
E-mail : info@3DISCimaging.com
User Manual
FireCR Dental TM-801-EN
3
The device complies with Part 15 of the FCC Rules. Operation is subject to the condition
that this device does not cause harmful interference.
Note: This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide
reasonable protection against harmful interference when the equipment is operated in a
commercial environment. This equipment generates, uses, and can radiate radio frequency
energy and, if not installed and used in accordance with the instruction manual, may cause
harmful interference to radio communications. Operation of this equipment in a residential
area is likely to cause harmful interference in which case the user will be required to correct
the interference at his own expense. Modifications not expressly approved by the manufacturer
could void the user's authority to operated the equipment under FCC rules.
User Manual
FireCR Dental TM-801-EN
4
Warnings and Used Symbols
To ensure the safety of patients, staff and other persons, any changes to software and hardware
delivered by 3D Imaging & Simulations Corp. may only be made with prior written permission
from 3D Imaging & Simulations Corp.
Please read the respective manuals of the connected software, such as acquisition and
diagnostic software, before starting to use the FireCR Dental system.
The following symbols will be used throughout this manual:
D
ANGER
The functionality of the system can be destroyed in the case of
incorrect use.
If unauthorized changes have been made to delivered system and
accessories, the warranty by 3D Imaging & Simulations Corp.
becomes void. 3D Imaging & Simulations Corp. will not accept
any responsibility or liability for the improper functioning of the
product in such a case.
W
ARNING
The functionality of the system can be limited in the case of
incorrect use. Hints that require special attention.
N
OTE
Notes represent information that is important to know but which do
not affect the functionality of the system.
D
ANGER
This equipment is indoor use only and all the communication
wirings are limited to inside of the building.
User Manual
FireCR Dental TM-801-EN
5
General Safety Guidelines
All the safety and operating instructions should be read carefully before this device is operated.
This device has been designed and tested to meet strict safety requirements applicable to
medical equipment, and has been supplied in a safe condition. To ensure personnel and
patient safety, the device shall be operated and serviced in compliance with all procedures,
warnings and precautions during all phases of operation and service of this device. Failure to
comply to with safety guidelines may result in injury to service personnel, operator, or patient.
3D Imaging & Simulations Corp. assumes no liability for failure to comply.
If this device is not used as specified, the protection provided by the device could be impaired.
This device must be used in a normal condition only.
Installation, service and operation of this device should only be undertaken by qualified trained
personnel. The operator should study instructions and precautions carefully before starting to
use the device listed here and throughout the manual.
There are no user serviceable parts inside this device. The device should only be opened and
serviced by qualified service personnel. Failure to heed this warning may result in injury to
service personnel or damage to equipment, and void any and all warranties. If there is a
service problem, please contact 3D Imaging & Simulations Corp. or authorized dealer.
Do not spill liquids on the device, and never operate the device in a wet environment.
Keep the device from radiators and heat sources.
Use the device only with accessories supplied with this device.
This device is intended to be grounded. Plug power cord into properly grounded electrical
outlets. This cord is equipped with three-prong plugs to help ensure proper grounding.
This device contains static sensitive components. Proper static handling procedures and
equipment must be used when servicing this device.
User Manual
FireCR Dental TM-801-EN
6
Do not look inside of the device.
If any of the following conditions occur, unplug the device from the electrical outlet and contact
authorized service personnel.
The power cord or power adapter is damaged.
An object has fallen into the device.
The device has been exposed to water.
The device has been dropped or damaged.
The device does not operate correctly when the operating instructions are followed.
Federal law restricts this device to sale by or on the order or a physician
Intended Use
This device is a Dental Computed Radiography System and intended for use in producing digital
X-Ray images for dental radiography purposes. It comprises of reader, reusable imaging plate
and workstation software. It scans X-Ray exposed imaging plate and produces X-Ray image
in digital form. Then, digital image is transferred to workstation for further processing and
routing. This device is intended to be operated in a radiological environment by qualified staff.
User Manual
FireCR Dental TM-801-EN
7
Index of contents
Chapter 1. Introduction .................................................................... 9
Chapter 2. Unpacking .................................................................... 10
2.1. Inspection for Damage ................................................................... 10
2.2. Identify the Components ................................................................. 10
Chapter 3. Setting Up .................................................................... 12
3.1. Positioning ...................................................................................... 12
3.2. Identify Important Features ............................................................. 13
3.2.1. Reader Connecting Part ........................................................... 13
3.2.2. Touch Display Panel ................................................................. 14
3.3. Computer Requirements ................................................................ 18
3.3.1. Recommended Requirement .................................................... 18
3.3.2. Minimum Requirement ............................................................. 18
3.4. Installation of Acquisition and Diagnostic Software......................... 19
3.5. Connect the Cable and Power Cord ............................................... 19
3.5.1. Connect USB Interface Cable................................................... 19
3.5.2. Connect Ethernet Cable ........................................................... 20
3.5.3. Connect the Power Cord .......................................................... 21
3.5.4. Installation Report ..................................................................... 22
Chapter 4. Operating ..................................................................... 23
4.1. System Specifications .................................................................... 23
4.2. Operation Conditions ...................................................................... 24
4.3. Operating Instructions .................................................................... 25
4.3.1. Turn on the Reader ................................................................... 25
4.3.2. Turn on the Computer ............................................................... 25
4.3.3. Imaging Plate Placement and Removal .................................... 25
4.3.4. Getting a Scanned Image ......................................................... 26
4.3.5. Circuit Functions ....................................................................... 26
User Manual
FireCR Dental TM-801-EN
8
Chapter 5. Symbols ....................................................................... 28
5.1. Manufacturer’s Declaration ............................................................ 29
5.1. - Electromagnetic Emission ........................................................... 29
5.2. Manufacturer’s Declaration - Electromagnetic Immunity ................ 30
5.3. Guidance and Manufacturer ’s Declaration – Electromagnetic
Immunity ..................................................................................................... 33
5.4. Laser Safety Statement .................................................................. 34
Chapter 6. Warranty and Repair Service ....................................... 35
6.1. Standard Warranty.......................................................................... 35
6.2. Repair Service ................................................................................ 35
6.3. Out of Warranty Repair Service ...................................................... 35
6.4. Shipping ......................................................................................... 36
Chapter 7. Technical Assistance .................................................... 37
User Manual
FireCR Dental TM-801-EN
9
Chapter 1. Introduction
FireCR Dental Reader
• Rapidly view high-quality digital images
• Easy-to-use system increases patient throughput
• Full-featured, affordable dental solution
Experience the Benefits of Digital Imaging
The full-featured Fire CR Dental Reader from 3D Imaging & Simulations Corp. rapidly and
affordably delivers high-quality digital images for busy dental practices.
Compact & Affordable
The FireCR Dental Reader is compact and affordable, helping to increase patient throughput
and improve the overall productivity of your practice. With its small footprint, the reader fits
seamlessly into even the most space-challenged dental offices and exam rooms.
Elegant Design & Streamlined Operation
The FireCR Dental Reader’s elegant design belies a powerful yet easy-to-use system that gets
the job done day in and day out.
The reader is DICOM 3.0 compatible with existing systems and uses a full range of low-cost,
reusable bitewing and intraoral imaging plates that are easier and faster to position than
intraoral digital sensors.
Unique Touch Screen Panel
The FireCR Dental Reader features a unique color touch-screen LCD panel for seamless
device operation. With no physical push buttons on the unit and an intuitive user interface,
entering operator instructions has never been easier.
Built-in Erase & UV Sterilization
A built-in erase function eliminates the need to purchase an additional device to erase imaging
plates prior to reuse, improving efficiency and reducing costs. And, an integrated UV disinfection
system sterilizes imaging plates that may be contaminated during handling, ensuring a hygienic
workflow and protecting the safety of patients and staff.
Unmatched Flexibility to Help Your Practice Grow
In this increasingly competitive environment, show patients that your practice offers the latest in
digital imaging technology for improved patient results. Whether you’re looking for a central
reader that can distribute images throughout your facility, or an affordable exam-room based
solution, the next-generation FireCR Dental Reader is the natural choice for practices looking
for a full-featured system that delivers high-quality digital images—along with the flexibility that
helps them grow.
User Manual
FireCR Dental TM-801-EN
10
Chapter 2. Unpacking
2.1. Inspection for Damage
FireCR Dental Reader is shipped in a custom designed container to protect the reader from
external shock. Before unpacking the reader, inspect the shipping container for damage. In
case the container is damaged, notify the shipper immediately.
2.2. Identify the Components
Open the shipping container and identify each of these components.
Part No. Item
CR-FP-11-001 FireCR Dental Reader
CR-FPA-01-002 Power Adapter
CR-FPA-02-001 USB 2.0 Interface Cable
CR-FPA-02-002 RJ45 CAT.6 UTP Cable 2M (direct type)
CR-FPA-03-00X Power Cord
CR-FPM-11-001 FireCR Dental User Manual
CR-PKM-11-004 IP Storage Case
CR-FP-12-001 Size 0 Imaging Plate (22mm x 31mm)
CR-FP-12-002 Size 1 Imaging Plate (24mm x 40mm)
CR-FP-12-003 Size 2 Imaging Plate (31mm x 41mm)
CR-FP-12-004 Size 3 Imaging Plate (27mm x 54mm)
CR-FP-12-005 Size 4c Imaging Plate (48mm x 54mm)
CR-FPA-15-001 Size 0 IP Hygienic Bag
CR-FPA-15-002 Size 1 IP Hygienic Bag
CR-FPA-15-003 Size 2 IP Hygienic Bag
CR-FPA-15-004 Size 3 IP Hygienic Bag
CR-FPA-15-005 Size 4c IP Hygienic Bag
W
ARNING
If the FireCR Dental needs to be returned to manufacturer or one
of its representatives, the reader must be repacked in the original
container with all accessories.
User Manual
FireCR Dental TM-801-EN
11
W
ARNING
Use of Power Cord;
Type SJT or SVT, min. 18AWG, 3-Conductor, VW-1 125V, min 10A
(or 250V, 10A). Max 3.0m long; One end with Hospital Grade
Type, NEMA 5-15P for 125V or NEMA 6-15P for 250V. Other end
with appliance coupler. “CAUTION Grounding reliability can only
be achieved when the equipment is connected to an equipment
receptacle marked “Hospital Only” or “Hospital Grade”.
For connection to supply not in USA, make sure the power cord is
the correct type that is required in your area.
W
ARNING
Improper disposal of this product may result in environmental
contamination. When disposing of this equipment, contact 3D
Imaging & Simulations Corp.’s representative or related organs
of government. Do not dispose of any part of this equipment
without consulting a 3D Imaging & Simulations Corp.’s
representative first.
3D Imaging & Simulations Corp. does not assume any
responsibility for damage resulting from disposal of this equipment
without consulting 3D imaging & Simulations Corp.
N
OTE
AC/DC Adapter
Manufacturer : Bridge Power corp.
Model : BPM050X24XXX
This adapter is passed by IEC60601-1.
W
ARNING
Use the device passed IEC60950-1 or IEC60601-1 for the product
connected via USB port.
User Manual
FireCR Dental TM-801-EN
12
Chapter 3. Setting Up
3.1. Positioning
The reader must be placed on a rigid and flat desk or tabletop with at least 5 cm (2 inch) free
space on both sides, 10 cm (4 inch) on rear side and 15 cm (6 inch) on front side for imaging
plate insertion. Its space requirements are shown below.
Figure 1. Space Requirements (Top View)
D
ANGER
Never place the reader on the floor.
Install in a location that is level and stable. Installation in an
unsuitable location can cause accident, or deterioration in image
quality.
Free space of 10 cm (4 inch)
on backside, at least, to
allow the power switch,
power cord and interface
cable to be reached by hand
at all times.
Free space of 15 cm (6 inch)
on front side, at least, for
imaging plate insertion and
removal.
Front Side Backside
W
ARNING
Unsuitable Installation Sites
Locations with excessive humidity or dust
Locations subject to high temperature
Locations subject to shaking or vibration
Locations exposed to considerable electrical or magnetic
noise, or other forms of electromagnetic energy
Locations with poor heat radiation
User Manual
FireCR Dental TM-801-EN
13
3.2. Identify Important Features
Look over the reader and features shown in this section. User will need to know where these
features are when user operates the reader in later chapters.
3.2.1. Reader Connecting Part
Figure 2. Reader Connecting Part
W
ARNING
Sliding the reader may result in internal damage or misalignment
of the optics.
External vibration or shock during scanning may affect image
quality. The reader must be placed on rigid, flat and reinforced
desk or tabletop.
W
ARNING
Do not place anything on top of the reader.
W
ARNING
This equipment may be interfered or interfere by electromagnetic
or other interference.
Assure a distance of minimum 1.0 m with neighboring equipment.
USB 2.0
Power Inlet
Power Switch
100Mbps Ethernet port
User Manual
FireCR Dental TM-801-EN
14
3.2.2. Touch Display Panel
Screen displays reader’s status and control of the reader can be done by touch screen.
Figure 3. Touch Display Panel
display Status Remark
Booting screen
When the reader is turned on,
booting screen is displayed during
system initialization.
Sleeping
Power saving mode – IP tray
moves back into the reader and
door will be closed.
Touch display panel
User Manual
FireCR Dental TM-801-EN
15
Disconnection USB / Ethernet is disconnected.
No software Software is not running on PC.
Updating Firmware
Firmware is being updated.
Ready for scan,
but no IP
The reader is ready for scan, but
IP is not placed on the tray.
Ready for Scan IP size is recognized and the
reader is ready.
User Manual
FireCR Dental TM-801-EN
16
Scanning Scanning is in progress.
Manual Erasing Erasing image from IP manually.
Preview Preview of scanned image.
Setting mode Shows information and settings of
the reader.
Scanning
resolution setting Choosing scanning resolution.
User Manual
FireCR Dental TM-801-EN
17
Auto sleep setting Choosing time duration until sleep
mode.
Network setting Manual input for network related
settings.
DHCP setting Choosing DHCP mode for
automatic network settings.
IP Address setting Type in IP address manually.
Subnet mask
setting Type in subnet mask manually.
User Manual
FireCR Dental TM-801-EN
18
Default gateway
Setting
Type in gateway address
manually.
3.3. Computer Requirements
3.3.1. Recommended Requirement
Operation System Microsoft Windows 7 (32 bit or 64 bit)
CPU Core Duo / Core2 Processor
Memory RAM 4GB or more
Hard Disk 300GB Free Hard Disk Space
Network 1Gbps Ethernet
Video 32 bit Color Display
Video Resolution 1280 x 1024
3.3.2. Minimum Requirement
Operation System Microsoft Windows 7 (32 bit or 64 bit)
CPU Core Duo / Core2 Processor
Memory RAM 2GB or more
Hard Disk 80GB Free Hard Disk Space
Network 1Gbps Ethernet
Video 32 bit Color Display
Video Resolution 1280 x 900
User Manual
FireCR Dental TM-801-EN
19
3.4. Installation of Acquisition and Diagnostic Software
Refer to Acquisition and Diagnostic Software manual.
3.5. Connect the Cable and Power Cord
3.5.1. Connect USB Interface Cable
The reader interfaces with computer via USB2.0 cable.
1. Use the USB cable inside the shipping container.
2. Connect the cable to the reader’s USB2.0 port, located on the rear of the reader.
3. Connect the other end of the cable to the USB2.0 port on the computer.
Figure 4. USB Connection
D
ANGER
This equipment is indoor use only and all the communication
wirings are limited to inside of the building.
W
ARNING
Do not pull out the USB cable during scanning.
User Manual
FireCR Dental TM-801-EN
20
3.5.2. Connect Ethernet Cable
The reader interfaces with computer via Ethernet cable (RJ45 CAT.6 UTP).
1. Connect the cable to the reader’s Ethernet port, located on the rear of the reader.
2. Connect the other end of the cable to the Ethernet port of the Ethernet-hub.
3. If you want to connect to PC directly, use crossed type cable instead of normal Ethernet
cable.
Figure 5. Ethernet Connection
D
ANGER
This equipment is indoor use only and all the communication
wirings are limited to inside of the building.
W
ARNING
Do not pull out the Ethernet cable during scanning.
User Manual
FireCR Dental TM-801-EN
21
3.5.3. Connect the Power Cord
1. Connect the power cord to the reader, located on the rear side.
2. Connect the other end of the cord to a grounded power outlet.
Figure 6. Power Connection
D
ANGER
This equipment must only be connected to supply mains with
protective earth. Use only a three-wire cord that has grounding.
This is a safety feature. If you are unable to insert the plug into
the outlet, contact your electrician to replace your obsolete ou
tlet.
Do not defeat the safety purpose of the grounding-type plug.
D
ANGER
Do not use with any electrical power supply that does not meet the
ratings displayed on the power adapter. Use with any other
power adapter may lead to fire or electrocution.
D
ANGER
Do not use any power adapter and power cord other than the
power adapter and power cord included with the system. Doing
so may lead to fire, electrical shock, or electrocution.
User Manual
FireCR Dental TM-801-EN
22
3.5.4. Installation Report
After installation of the reader, fill in Installation Report from (Appendix I) and send to 3D
Imaging & Simulations Corp. service department by fax or e-mail.
Fax : +82-42-931-2299
E-mail : support@3DISCimaging.com
W
ARNING
Socket-outlet shall be installed near the device and shall be easily
accessible.
Do not place the device where difficult to access to appliance inlet.
Do not unplug the power cord or turn the power switch off during
scanning.
User Manual
FireCR Dental TM-801-EN
23
Chapter 4. Operating
4.1. System Specifications
Sampling Pixel Pitch SD 64um
HD 35um
Pixel Matrix (Size 0)
SD 343 x 484
HD 628 x 885
Pixel Matrix (Size 1)
SD 375 x 625
HD 685 x 1143
Pixel Matrix (Size 2)
SD 484 x 640
HD 886 x 1171
Pixel Matrix (Size 3)
SD 421 x 843
HD 771 x 1542
Pixel Matrix (Size 4c) SD 750 x 843
HD 1370 x 1542
Accepted Imaging Plate Size 0, 1, 2, 3, 4c
Gray Scale Resolution 16 bit
Eraser Embedded
UV Sterilizer Embedded
Computer Interface USB 2.0 / 100Mbps Ethernet
Dimensions 265 (H) x 120 (W) x 318 (D) mm
10.4 (H) x 4.7 (W) x 12.5 (D) inch
Weight 4.7 kg
10.4 lbs
Power Requirement 100 ~ 240V / 50 ~ 60Hz
Image File Format DICOM 3.0, TIFF, BMP, JPEG
* Specifications subject to change without notice.
** Specific results may vary since operating conditions fluctuate.
User Manual
FireCR Dental TM-801-EN
24
4.2. Operation Conditions
Indoor use only
Operating Temperature 15°C ~ 30°C (59°F ~ 86°F)
Temperature Gradient 0.5°C / Min
Relative Humidity 15% ~ 95% (non-condensing)
Storage Temperature - 10°C ~ 50°C (14°F ~ 122°F)
Storage Humidity 15% ~ 95% (non-condensing)
Storage Atmospheric Pressure 500 ~ 1,060 hPa
Transportation Temperature - 10°C ~ 50°C (14°F ~ 122°F)
Transportation Humidity 15% ~ 95% (non-condensing)
Transportation Atmospheric Pressure 500 ~ 1,060 hPa
Installation Category II
Pollution Degree 2
Ingress of Liquids IPX0
Altitude Up to 2,000m
Protective Class Class 1
Equipment Maintenance No user maintenance is required and no user
service is allowed. Please contact technical
support if there is a problem.
Cleaning Do not try to clean inside of the reader.
Wipe outside of the reader for dust removing
with soft and dry cloth.
WARNING
There are no user serviceable parts inside the reader. The
reader should only be opened and serviced by qualified service
personnel. Failure to heed this warning may result in injury to
service personnel or damage to equipment, and void any and all
warranties. If there is a service problem, please contact 3D
Imaging & Simulations Corp. or authorized dealer.
User Manual
FireCR Dental TM-801-EN
25
4.3. Operating Instructions
4.3.1. Turn on the Reader
Turn on the reader. Power switch is on the rear side of the reader.
4.3.2. Turn on the Computer
Turn on the computer. Acquisition and Diagnostic Software must be installed before operate
the reader.
4.3.3. Imaging Plate Placement and Removal
Place the imaging plate on the tray to start scan. The imaging plate can be removed when
scanning and erasing are completed. Gently pull up the imaging plate.
Figure 7. Imaging Plate Placement and Removal
D
ANGER
This device uses laser. Avoid looking inside of the reader.
W
ARNING
Do not place the imaging plate in wrong direction or upside down
when it is being placed on the tray.
Tray
User Manual
FireCR Dental TM-801-EN
26
In order to scan or erase the IP, locate the IP on the tray correctly and push the tray into the
reader fully until interlock holds the tray.
4.3.4. Getting a Scanned Image
To getting a scanned image, refer to Acquisition and Diagnostic Software manual.
4.3.5. Circuit Functions
W
ARNING
Locate the IP in correct position.
W
ARNING
Do not reuse hygienic bag.
User Manual
FireCR Dental TM-801-EN
28
Chapter 5. Symbols
Symbol Description
.Manufacturer
Date of Manufacture
Equipment Power ON
Equipment Power OFF
Warning, Consult Accompany Documents
General mandatory action manual
General prohibition indication
User Manual Reference
Directive on Waste Electrical and Electronic Equipment
Authorised Representative in the European
Community
Keep Dry
User Manual
FireCR Dental TM-801-EN
29
Fragile
Handle with care
This side up
Non-ionizing electromagnetic radiation
FCC ID :
X68CRSCANNER2 FCC Mark
Medical Equipment
WITH RESPECT TO ELECTRIC SHOCK
FIRE, AND MECHANICAL HAZARDS ONLY
IN ACCORDANCE WITH UL60601-1 / CAN / CSA CSS.2 No. 601.1
3SE3
CE Mark
5.1. Manufacturer’s Declaration- Electromagnetic Emission
The FireCR Dental system
is intended for use in the electromagnetic environment specified
below. The customer or the user of FireCR Dental system
should assure that it is used in such
an environment
Emission test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1 The
FireCR
Dental system uses RF energy only
for its internal function. Therefore. Its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment
RF emissions
CISPR 11
Class B The Model FireCR Dental is suitable for use in
all establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonics emission
IEC 61000-3-2
A
Voltage fluctuation
IEC 61000-3-3
Complies
User Manual
FireCR Dental TM-801-EN
30
5.2. Manufacturer’s Declaration - Electromagnetic Immunity
The FireCR Dental system
is intended for use in the electromagnetic environment specified
below. The customer or the user of FireCR Dental system
should assure that it is used in such
an environment
Immunity test
IEC 60601
Test level Compliance level
Electromagnetic
Environment -guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV Contact
8 kV Air
6 kV Contact
8 kV Air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%
Electrical fast
Transient / burst
IEC 61000-4-4
2kV for power supply lines
1kV for input/output lines
2kV for power supply lines
1kV for input/output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
1 kV differential mode
2 kV common mode
1 kV differential mode
2 kV common mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Power frequency
(50/60Hz)
Magnetic field
IEC 61000-4-8
3.0 A/m 3.0 A/m Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Voltage dips, short
Interruptions and
Voltage variations
on power supply
input lines
IEC 61000-4-11
<5% Uт (>95% dip in Uт)
for 0.5cycle
40% Uт (60% dip in Uт )
for 5 cycle
70% Uт (30% dip in Uт)
for 25 cycle
<5% Uт (<95% dip in Uт )
for 5 s
<5% Uт (>95% dip in Uт)
for 0.5cycle
40% Uт (60% dip in Uт )
for 5 cycle
70% Uт (30% dip in Uт)
for 25 cycle
<5% Uт (<95% dip in Uт )
for 5 s
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the BSVD-1000 system
requires continued operation
during power mains
interruptions, it is
recommended that the FireCR
Dental system be powered
from an uninterruptible power
supply or a battery
User Manual
FireCR Dental TM-801-EN
31
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
150 kHz to 80 MHz
Portable and mobile RF
communications equipment
should be used no closer to
any part of the FireCR Dental
system, including cables, than
the recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
Radiated RF
IEC 61000-4-3
3 V/m
80.0 MHz to 2.5 GHz
3 V/m
80.0 MHz to 2.5 GHz
Recommended separation
distance
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as deter-mined by
an electromagnetic site survey,
(a) Should be less than the
compliance level in each
frequency range (b).
Interference may occur in the
vicinity of
equipment marked with the
following symbol:
N
ote 1)
Uт is the A.C. mains voltage prior to application of the test level.
Note 2) At 80 MHz and 800 MHz, the higher frequency range applies.
Note 3) These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
User Manual
FireCR Dental TM-801-EN
32
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
EUT is used exceeds the applicable RF compliance level above, the EUT should be observed to
verifynormal operation. If abnormal performance is observed, additional measures may be necessary, such
as re-orienting or relocating the EUT.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [V1] V / m.
Recommended Separation Distances Between Portable and Mobile RF Communications
Equipment and the FireCR Dental system.
The FireCR Dental
system is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The user of the FireCR Dental
system can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the FireCR Dental system as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output
power (W) of transmitter
Separation distance (m) according to frequency of transmitter
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5
GHz
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.70 3.70 7.37
100 11.70 11.70 23.30
For transmitters rated at a maximum output power not listed above, the recommended separation
distance (d) in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
N
ote
1:
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects, and people.
Immunity and Compliance Level
Immunity test IEC 60601 Test Level Actual Immunity Level Compliance Level
Conducted RF
IEC 61000-4-6
3 Vrms, 150 kHz to 80
MHz
3 Vrms, 150 kHz to 80
MHz
3 Vrms, 150 kHz to 80
MHz
Radiated RF
IEC 61000-4-3
3 V/m, 80 MHz to 2.5
GHz
3 V/m, 80 MHz to 2.5
GHz
3 V/m, 80 MHz to 2.5
GHz
User Manual
FireCR Dental TM-801-EN
33
5.3. Guidance and Manufacturer ’s Declaration –
Electromagnetic Immunity
The FireCR Dental system
is intended for use in the electromagnetic environment specified
below. The customer or the user of FireCR Dental system
should assure that it is used in such an
environment
Immunity test
IEC 60601
Test level Compliance level
Electromagnetic
environment
-
guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80MHz
3 Vrms
150 kHz to 80 MHz
FireCR Dental system must be used
only in a shielded location with a
minimum RF shielding effectiveness
and, for each cable that enters the
shielded location with a minimum RF
shielding effectiveness and, for each
cable that enters the shielded location
Radiated RF
IEC 61000-4-3
3 V/m
80.0 MHz to 2.5GHz
3 V/m
80.0 MHz to 2.5GHz
Field strengths outside the shielded
location from fixed RF transmitters, as
determined by an electromagnetic site
survey, should be less than 3V/m.a
Interference may occur in the vicinity of
equipment marked with the following
symbol:
N
ote
1)
These guidelines may not apply
in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
Note 2)
It is essential that the actual shielding effectiveness and filter attenuation of the shielded
location be verified to assure that they meet the minimum specification.
a
-
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field
strength outside the shielded location in which the EUT is used exceeds 3V/m, th
e EUT should be
observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as relocating
the EUT or using a shielded location with a higher RF shielding effectiveness and filter attenuation.
User Manual
FireCR Dental TM-801-EN
34
5.4. Laser Safety Statement
The Computed Radiography Reader is Certified in the U.S. to Conform to the Requirements of
DHHS 21 CFR, chapter 1 Subchapter J for Class I(1) Laser Products, and Elsewhere is
Certified as a Class I(1) Laser Product Conforming to the Requirements of IEC 60825-1 :
2007. Class I(1) Laser Products are not Considered to be Hazardous. The Laser System and
Computed Radiography Reader are Designed so there is never any Human Access to Laser
Radiation above a Class I(1) level during normal Operation, user Maintenance or Prescribed
Service Condition.
• Wavelength : 658 nm (Typ.)
• Beam Divergence
- Paraller : 9.5 degrees (-2.5/+2.5)
- Perpendicular : 17 degrees (-3/+3)
• Maximum Power of Energy Output : 80 mW (CW)
W
ARNING
Never operate or service the product with the protective cover removed
from Laser/Reader assembly.
The reflected beam, although invisible, can damage your eyes.
When using this product, these basic safety precautions should always be
followed to reduce risk of fire, electric shock and personal injury
CAUTION
Use of controls or adjustments or performance of procedures other than
those specified herein may result in hazardous radiation exposure
User Manual
FireCR Dental TM-801-EN
35
Chapter 6. Warranty and Repair Service
6.1. Standard Warranty
3D Imaging & Simulations Corp. warrants its non-consumable hardware products to be free
from defects in materials and workmanship. The warranty covers the cost of parts and labor to
repair the product. Please keep the shipping container for future use. Products returned to
the factory for repair should be properly packaged. To obtain warranty service, follow the
procedure described in the Repair Service section. Failure to do so will cause long delays and
additional expense to the customer.
The warranty is valid when the product is used for its intended purpose and does not cover
products which have been modified without written permission from 3D Imaging & Simulations
Corp., or which have been damaged by abuse, accident or connection to incompatible
equipment.
This warranty is in lieu of all other warranties, expressed or implied.
6.2. Repair Service
The company reserves the right to cease providing repair maintenance, parts and technical
support for its non-consumable hardware products five years after a product is discontinued.
Technical support for old versions of software products will cease 12 months after they are
upgraded or discontinued.
6.3. Out of Warranty Repair Service
Out of warranty repair service is available in selected geographical locations. Contact the
supplier for current terms and rates.
User Manual
FireCR Dental TM-801-EN
36
6.4. Shipping
The FireCR Dental Reader is a solidly built system designed to survive shipping around the
world. However, in order to avoid damage during shipping, the FireCR Dental Reader must
be properly packaged.
In general, the best way to package the FireCR Dental Reader is in the original factory
container. If this is no longer available, we recommend that user carefully wraps the FireCR
Dental Reader in at least 75 mm (3 inch) of foam or bubble pack sheeting. The wrapped
device should then be placed in a sturdy cardboard carton. Mark the outside of the box with
word FRAGILE and an arrow showing which way is up.
We do not recommend using loose foam pellets to protect the FireCR Dental Reader. If the
carton is dropped by the shipper, there is a good chance that the device will shift within the
loose pellet packing and be damaged.
If user needs to ship the FireCR Dental Reader to another location, or back to the factory, and
user does not have a means to adequately package it, user can order additional shipping
container. This may seem an expense user would like to avoid, but it is inexpensive compared
to the cost of repairing an instrument that has sustained shipping damage.
It is user’s responsibility to package the system properly before shipping. If the packaging is
inadequate, and the system is damaged during shipping, the shipper will not honor user’s claim
for compensation.
User Manual
FireCR Dental TM-801-EN
37
Chapter 7. Technical Assistance
If user has any questions about installing or using the device, contact your 3D Imaging &
Simulations Corp representative or your local dealer.
3D Imaging & Simulations Corp.
815, Tamnip-Dong, Yuseong-Gu, Daejeon, Korea
Tel : 82-42-931-2100 Fax : 82-42-931-2299
www.3DISCimaging.com
User Manual
FireCR Dental TM-801-EN
38
Appendix I
Installation Report
Please complete this report at the time of installation and submit the
completed form signed by customer to:
Fax : +82-42-931-2299
E-mail : support@3DISCimaging.com
Date of Installation :
Customer Information
Hospital / Institute
Name
Address
Tel
Fax
E-mail
Installer Information
Company
Name
Address
Tel
Fax
E-mail
System Information
Model FireCR Dental Reader
System S/N
Installer’s Signature: Date:
Customer’s Signature: Date: