AVITA BPW-1718R Blood Pressure Monitor User Manual

AVITA Corporation Blood Pressure Monitor

User Manual

Download: AVITA BPW-1718R Blood Pressure Monitor User Manual
Mirror Download [FCC.gov]AVITA BPW-1718R Blood Pressure Monitor User Manual
Document ID3639131
Application IDDf01btZFg3qDxQH+aO+eGQ==
Document DescriptionUser Manual
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize45.02kB (562773 bits)
Date Submitted2017-11-15 00:00:00
Date Available2017-11-15 00:00:00
Creation Date2017-11-07 11:51:35
Producing SoftwareAcrobat Distiller 9.5.2 (Windows)
Document Lastmod2017-11-07 11:51:35
Document TitleMicrosoft Word - BP18R Manual 72-B18RMN-AV1_V2_171102.doc
Document CreatorPScript5.dll Version 5.2.2
Document Author: docscaner

Contents
Wrist Type Blood Pressure Monitor
Model: BPM18R
Introduction ...................................................... 1
Important Information Before Use ................... 2
Blood Pressure Monitor Features .................... 3
Description of LCD Display.............................. 4
Battery Installation ........................................... 5
Setting the Date and Time ............................... 6
Applying Your Cuff........................................... 7
Taking a blood pressure measurement ........... 8
Irregular Heartbeat Detector ............................ 9
World Health Organization (WHO) ................ 10
Memory Function ........................................... 11
Troubleshooting ............................................. 12
Error Codes.................................................... 12
Operating the Bluetooth function ................... 13
Care and Maintenance .................................. 14
Technical Specification .................................. 14
FCC Statement .............................................. 15
EMC Tables ................................................... 16
Explanation of Symbols ................................. 17
INSTRUCTION MANUAL
Please read this instruction manual carefully
before operating this unit.
Introduction
BPM18R Series Blood Pressure Monitor applies the
Oscillometric method to measure human systolic / diastolic
blood pressure and heart rate. All values are shown on the
LCD monitor and using the Bluetooth technology (BLE4.0)
or USB to transfer the related information of your blood
pressure measurement. This device is designed only for
adults.
CAUTION:
No modification of this equipment is allowed.
Keep your equipment out of the reach of children, pets
and pest.
The Patient is an intended operator.
Risk of explosion if battery is replaced by an incorrect
type.
Dispose of used batteries according to the instructions.
Important Information Before Use
When using this product, please be sure to follow all the
notes listed below. Any action against these notices may
cause injury or affect the accuracy.
1. WARNING – Do not modify this equipment without
authorization of the manufacturer.
2. Blood pressure measurements should only be interpreted
by a physician or a trained health care professional who
is familiar with your medical history.
Through regular use of this device and recording of your
measurements, you can keep your physician informed of
the changes in your blood pressure.
3. Perform your measurement in a quiet place. You should
be seated in a relaxed position.
4. Avoid smoking, eating, taking medication, alcohol
consumption or physical activity 30 minutes prior to
taking a reading. If you are exhibiting signs of stress,
avoid taking your measurement until the feeling
subsides.
5. Rest 15 minutes prior to taking a reading.
6. Remove any constrictive clothing or jewelry that may
interfere with the cuff placement.
7. Keep the monitor stable during measurement to achieve
an accurate reading. Remain still; do not talk during the
measurement.
8. Record your daily blood pressure and pulse readings on a
chart.
9. Take your readings at the same time, each day or as
recommended by your physician to get an accurate
indication of change in your true blood pressure.
10.Wait a minimum of 2~3 minutes between readings to
allow for the blood vessels to return to normal. The wait
time may vary depending on your individual
physiological characteristics.
11.Although such cases are rare, for those with an
extremely weak pulse or irregular pulse, errors may
result which prevent proper measurement. If abnormal
variations are noticed, consult with your physician or
trained healthcare professional.
12.This device is intended for adult use. While taking a
measurement, you can stop the inflation or deflation
process of the cuff at any time by pressing the POWER
button.
13.It is recommended that the accuracy should be checked
by the manufacturer every two years.
14.If you would require any assistance during setting up,
using or maintaining, please contact the manufacturer.
Under any circumstances, the measurement taking
result is ONLY for reference.
Any further medical action should follow the advice of
doctors.
Blood Pressure Monitor Features
Description of LCD Display
Battery Installation
Setting the Date and Time
It is necessary to set the date and time for the unit every
Low battery warning:
It is necessary to replace the batteries
when the Low Battery symbol “
”
appears on the display, or when the
display does not turn on after the
POWER button is pressed.
time batteries are initially installed or replaced.
Replacing the Battery:
2. Press the “MEMORY” button to advance the display to
1. Slide the battery cover on the bottom of the monitor.
2. Insert or replace 2 x 1.5 V AAA batteries into the battery
compartment, ensuring
to match the indicated polarity symbols. Always use new
batteries.
3. Replace the battery cover.
1. After replacing batteries, the “Year” will begin to flash
on the display.
the desired “Year”, press“POWER” button to confirm the
“Year”.
3. Next, the “Month” will blink. Repeat steps 2 to set the
“Month” and “Date”, then “hour”, then “Minute”.
4. After setting the minute, the unit will automatically exit
out or press power button the date/time setting mode and
briefly show the word OFF before shutting down.
Battery-operated
1. Always use NEW batteries; never mix old and new
batteries.
2. Do not use rechargeable batteries; rechargeable batteries
can vary greatly in quality and strength.
Use of rechargeable batteries may affect the performance of
this device.
3. Please properly dispose of the batteries away from small
children and heat.
4. It is recommended to remove the batteries if the unit will
not be used for an extended period of time.
5. Batteries must be disposed of in accordance with local
environmental and institutional policies.
6. Remove the battery during extended storage.
Applying Your Cuff
It is important to avoid smoking,
Taking a blood pressure measurement
Please read the preceding portions of this manual prior to
taking your first reading.
consumption or physical activity 30
1. Rest your elbow on a solid surface with your palm facing
upward. Elevate your wrist so that the cuff is at the same
level as your heart. Relax your left hand.
minutes prior to taking a reading. If
2. Press the POWER button. This will turn the power on.
eating, taking medication, alcohol
for any reason you are unable to or
should not use your left wrist,
please modify the instructions for cuff application to your
right wrist.
Your physician can help you identify which wrist is best for
3. After the self-test, the values for
the last reading will appear
on the display.
4. The blood pressure monitor start
to measure.
5. The ‘Easy Positioning System’ helps to ensure
the arm is in the correct position. Simply move
your arm up and down until the wrist stays right height.
6. The measurement starts automatically once
the correct position has been obtained. Do not
move or speak while the measurement is in
progress.
you to take measurements from.
Applying Your Cuff
1. Remove any constrictive clothing or jewelry that may
interfere with cuff placement.
2. Be seated at a table or desk with your feet flat on the
floor.
3. Hold your wrist in front of you with your palm facing
upward.
4. Apply the preformed cuff to your wrist so that the digital
display face is positioned on the inside area of your wrist
facing you.
5. Adjust the cuff 0.5~1 cm from the
edge of the head of the ulna bone.
6. The cuff should fit comfortably, yet
snugly around your wrist.
7. The cuff will automatically begin to inflate, with
the display showing the increasing pressure in
the cuff. As the pressure increases, an arrow
pointing up will appear on the display.
8. When the inflation has reached optimum
level, the display will begin to show the
decreasing pressure; the screen will display an
arrow pointing down while you feel the pressure
of the cuff decrease.
9. To detect the heartbeat, the heartbeat symbol will appear
and continuous flashes on the LCD display.
10. Your blood pressure measurement and pulse will display
simultaneously on the screen.
11. The Hypertension Indicator will indicate
your reading range on the display separately.
12.The reading will automatically be stored in
memory.
13. Select POWER to turn the unit off and
conserve energy and battery life. The unit will
automatically shut-off approximately 2
minutes.
Blood pressure naturally varies from
one wrist to the other; therefore, measure your blood
pressure on the same wrist to ensure comparability of the
two readings.
Irregular Heartbeat Detector
World Health Organization (WHO)
Your digital blood pressure monitor
This unit features our unique Hypertension Indicator. The
features an Irregular Heartbeat Detector.
World Health Organization has established globally accepted
Irregular Heartbeats may influence the
standards for the assessment of high or low blood pressure
results of the measurement. If the monitor
readings. The below chart should be considered only as
detects the Irregular Heartbeats during
a guideline, always consult with your physician or health care
” will appear on the display
measurement, the symbol “
professional to interpret your individual results.
with the measurement values. You can take another
measurement to make sure the values are not influenced by
moving during measurement or Irregular Heartbeat.
2. If this symbol appears frequently, we recommend you
consult your physician.
3. This device does not replace a cardiac examination, but
serves to detect pulse irregularities at an early stage.
10
Memory Function
Troubleshooting
Recalling Measurements in Memory:
1. Press and release the “MEMORY” button. The unit
will first display the average of
the last 3 stored measurements.
2. Continue to press the “MEMORY” button to
successively view the next previously stored
measurements.
Measurements will
appear on
the display from most
current to oldest; the
memory number will appear on the lower right corner.
3. All results for a given measurement will display,
Error Codes
including measurement results, pulse rate,
Hypertension Indicator, and date/time stamp.
4. When the number of readings exceeds 90, the oldest
data will be replaced with the new record.
5. Press the Power button to turn the monitor OFF at
any time during review of the stored measurements.
Clearing Measurements from Memory:
From power display off, press and hold down
the “MEMORY” button until the display shows
CLr. This indicates that all measurements have
been erased.
11
12
Data Transferring
Bluetooth function:
What You Need
An Android device with Android version 4.3 or above and
hardware support for Bluetooth 4.0.
An iOS device with iOS version 5 or above and hardware
support for Bluetooth 4.0. All devices Apple released
since the iPhone 4S (including the 4S) do, the older ones
don’t.
Note:
Please refer to the instruction manual of your smart
phone for how to activate the Bluetooth function.
Set Up Process
1. Download an app which supports Bluetooth 4.0 from
the iTune App Store or Google Play.
2. Enable Bluetooth on your mobile device.
3. Open the app and activate the function.
4. Turn on the BPM18R.
5. The app should automatically detect your BPM18R.
Tap the BPM18R that is showed in the device list.
6. Your BPM18R is now successfully connected to
your mobile device. Every measure reading will be
transfer to your mobile device automatically.
You can send results in memory zone as following step
1.Press Memory recall button of blood pressure monitor
to enter memory mode. The symbol AVG and M will
appear on the display and the number of memory index is
“00”.
2.Turn on the bluetooth function of device to search the
blood pressure monitor. Then the blood pressure monitor
will be searched and connected the bluetooth device.
3.After the connection is completed, you can press
Care and Maintenance
1. Clean the device and wrist cuff carefully only with a
slightly moistened cloth.
2. Do not immerse the device in water. It may cause
damage if water enters.
3. Do not use any gas, strong detergent, or solvents to
clean the device, including the wrist cuff.
4. Use a soft, dry cloth to clean your monitor.
5. Do not store in direct sunlight, dust or excessive
humidity.
6. Avoid extreme temperatures.
7. Do not disassemble the monitor or cuff.
8. Remove batteries if the monitor will not be used for
an extensive period of time.
Technical Specification
Measuring range :
• Calibration Accuracy :
• Operating environment :
10oC~40oC (50oF~104oF)
with relative humidity up to 85% (non condensing)
• Storage/ Transportation environment :
-20oC~+50oC (-4oF~+122oF)
with relative humidity up to 85% (non condensing)
• Power Source :
• Weight :
• Dimensions :
• Cuff circumference (M Size) : approx. 12.5~21 cm
2 x 1.5 V AAA batteries
approx. 106g (exclude batteries)
approx. 58mm x 88.5mm x 28.3mm (W x H x D)
memory recall button to display the memory result on
LCD. Each result showed on LCD, the result will be
transmitted by bluetooth immediately.
USB function:
Connect the blood pressure monitor to your PC using the
USB Cable.
appears on the display. Begin the data
transfer in the PC software. During the data transfer, an
animation is shown on the display.
• A successful data transfer is displayed
as in figure 1.
If the data transfer is unsuccessful, an error
message appears as in figure 2.
In this case, interrupt the PC connection and
start the data transfer again.
After 30 seconds of not being in use or if communication
with the PC is interrupted, the blood pressure monitor
switches itself off automatically.
13
14
FCC Statement
Note: This equipment has been tested and found to
comply with the limits for a Class B digital device,
pursuant to Part 15 of the FCC Rules.
These limits are designed to provide reasonable
protection against harmful interference in a residential
installation.
This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful
interference to radio communications.
However, there is no guarantee that interference will
not occur in a particular installation. If this equipment
does cause harmful interference to radio or television
reception, which can be determined
by turning the equipment off and on, the user is
encouraged to try to correct the interference by one of
the following measures:
1. Reorient or relocate the receiving antenna.
2. Increase the separation between the equipment and
receiver.
3. Connect the equipment into an outlet on a circuit
different from that to which the receiver is
connected.
4. Consult the dealer or an experienced adio/TV
technician for help.
5. FCC Caution: To assure continued compliance,
(example – use only shielded interface cables when
connecting to computer or peripheral devices). Any
changes or modifications not expressly approved by
the party responsible for compliance could void the
user’s authority to operate this equipment.
6. This device complies with Part 15 of the FCC Rules.
Operation is subject to the following two conditions:
(1) this device may not cause harmful interference,
and (2) this device must accept any interference
received, including interference that may cause
undesired operation.
7. This device complies with FCC radiation exposure
limits set forth for an uncontrolled environment. End
users must follow the specific operating instructions
for satisfying RF exposure compliance. This
transmitter must not be colocated or operating in
conjunction with any other antenna or transmitter.
EMC Tables
AViTA BPM18R is intended for use in the electromagnetic environment
specified below. The customer or the user of AViTA BPM18R must
make sure that it is used in such an environment.
Guidance and manufacturer’s declaration - Electromagnetic emissions
Professional healthcare
HOME HEALTHCARE
Phenomenon
facility environment a)
ENVIRONMENT a)
Conducted and a)
CISPR 11
radiated RF
Group 1 Class B
EMISSIONS
Harmonic
Not applicable
distortion
Voltage
Not applicable
fluctuations and
flickering
a) The equipment is suitable for use in Home Health Environments and
Professional Health Care Environments limited to patient rooms and
respiratory treatment facilities in hospital or clinics. The more restrictive
acceptance limits of Group 1 Class B (CISPR 11) have been considered and
applied. The equipment is suitable for use in the mentioned environments
when directly connected to the Public Mains Network.
b) The test is not applicable in this environment unless the ME EQUIPMENT
and ME SYSTEM used will be connected to the PUBLIC MAINS NETWORK
and the power input is otherwise within the scope of the Basic EMC
standard.
Guidance and manufacturer’s declaration - Electromagnetic immunity Enclosure port
Phenomenon
Basic EMC
Immunity test levels
standard or test Professional HOME HEALTHCARE
method
healthcare ENVIRONMENT
facility
environment
ELECTROSTATIC IEC 61000-4-2 ± 8kV contact
DISCHARGE
± 2 kV, ±4kV ±, ±8 kV, ±15 kV air
Radiated RF EM IEC 61000-4-3 a)
10 V/m b) 80MHz - 2.7
fields
GHz 80% AM at 1kHz
Proximity fields
IEC 61000-4-3 COMPLIANT
from RF wireless
NOTE: Further information about
communications
distances to be maintained between
equipment
portable and mobile RF
communications equipment
(transmitters) and the AViTA BPM18R
can be requested from AViTA using the
contact information provided in this
manual. However, it is advisable to
keep the electromechanical aerosol
equipment at an adequate distance of,
at least, 0.5 m from mobile phones or
other RF communications transmitters
to minimise possible interference.
RATED power
IEC 61000-4-8 30 A/m c)
frequency
50 Hz or 60 Hz
magnetic fields.
a) The equipment is suitable for use in Home Health Environments and
Professional Health Care Environments limited to patient rooms and
respiratory treatment facilities in hospital or clinics. The more restrictive
IMMUNITY acceptance limits have been considered and applied.
b) Before modulation is applied.
c) This test level assumes a minimum distance of at least 15 cm between
the ME EQUIPMENT or ME SYSTEM and sources of power frequency
magnetic fields.
15
16
Explanation of Symbols
Follow instructions for use
Disposal information: Should you wish to
dispose of the article, do so in accordance
with current regulations. Details are available
from your local authority
“ON/OFF” (push-push)
Type of protection of applied part against
electric shock, body floating
This product meets the basic safety and
essential performance requirements
indicated in the IP22 conditioning test
(protection against solid foreign objects of
12.5mm ∅ and greater and against
vertically falling water drops when enclosure
tiled up to 15°)
Temperature limits
Manufacturer’s name and address
SN YYMWWWXXXXX
SN: Product Serial Number
YY: year, M:month, WWW: working sheet,
XXXXX: serial no.
17
18
72-B18RMN-AV1

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : Yes
Author                          : docscaner
Create Date                     : 2017:11:07 11:51:35+08:00
Modify Date                     : 2017:11:07 11:51:35+08:00
XMP Toolkit                     : Adobe XMP Core 4.2.1-c043 52.372728, 2009/01/18-15:08:04
Format                          : application/pdf
Title                           : Microsoft Word - BP18R Manual 72-B18RMN-AV1_V2_171102.doc
Creator                         : docscaner
Creator Tool                    : PScript5.dll Version 5.2.2
Producer                        : Acrobat Distiller 9.5.2 (Windows)
Document ID                     : uuid:4f4466a9-f171-4fd6-9040-04990e73c982
Instance ID                     : uuid:293d48db-7a7c-444d-81b4-368a59b48204
Page Count                      : 10
EXIF Metadata provided by EXIF.tools
FCC ID Filing: UV3BPW-1718R

Navigation menu