WHITESTAR SIGNATURE PRO System 8  • Error Messages Troubleshooting and Diagnostics RX Only –NGP Z370500 Rev. B 1116 8-38     513 IV Pole communications error. Probable cause: Communication error. Corrective action: 1.   Prime the system.  2.   If the error still does not clear or if the error continues to occur, document the error message and contact AMO for technical service.  514 IV Pole error.  Probable cause: Communication error.  Corrective action:  1.   Prime the system.  2.   If the error still does not clear or if the error continues to occur, document the error message and contact AMO for technical service.  515 IV Pole calibration error. Probable cause: Communication error. Corrective action: 1.   Run Self Test.  2.   If the error still does not clear or if the error continues to occur, document the error message and contact AMO for technical service.  516 IV Pole communication error. Probable cause: Communication error. Corrective action: 1.   Prime the system.  2.   If the error still does not clear or if the error continues to occur, document the error message and contact AMO for technical service.
WHITESTAR SIGNATURE PRO System 8  • Error Messages Troubleshooting and Diagnostics RX Only –NGP Z370500 Rev. B 1116 8-39     517 IV Pole is jammed.  Probable cause:  •  IV Pole motor failure. •  There is an obstruction that does not allow the IV pole to move.  Corrective action:  1.   Check the IV pole for obstructions.  2.   Prime the system.  3.   If the error still does not clear or if the error continues to occur, document the error message and contact AMO for technical service.  601 Tune excessive vacuum error. Probable cause: There is too much vacuum. Corrective action: 1.   Check tubing connections.  2.   Check the handpiece. Replace the tip if needed.  3.   Retune the handpiece.  4.   If the error still does not clear, replace the handpiece with a new handpiece, and tune the handpiece.  5.   If the error still does not clear or if the error continues to occur, document the error message and contact AMO for technical service.  605 Tune no handpiece error. Probable cause: Hardware failure. Corrective action: 1.   Attach handpiece to the system.  2.   Retune the handpiece.  3.   If the error still does not clear, replace the handpiece with a new handpiece, and tune the handpiece.  4.   If the error still does not clear or if the error continues to occur, document the error message and contact AMO for technical service.
RX Only –NGP  Z370500 Rev. B 1116 9-1   9  WARRANTY AND MAINTENANCE  Warranty Statement
WHITESTAR SIGNATURE PRO System 9  • Warranty and Maintenance RX Only –NGP  Z370500 Rev. B 1116 9-2     Warranty Statement Abbott Medical Optics Inc. warrants for a period of one year (12 months) from the date of installation of the WHITESTAR SIGNATURE PRO System console, foot pedal, wireless remote control, the programmable power pole, and the phaco handpiece to be free from defects in materials and workmanship when properly installed, maintained, and used for the intended purpose. In no event shall the date of installation be deemed to occur later than six months from the date of delivery of the WHITESTAR SIGNATURE PRO System by Abbott Medical Optics Inc. to the common carrier.  You are required to provide AMO with prompt written notice of any defect or defects and permit AMO to have access to the WHITESTAR SIGNATURE PRO System within a reasonable time after notification of any defect. In the event that you do not provide AMO with prompt notification of a defect or permit access to the WHITESTAR SIGNATURE PRO System within a reasonable time after notification, AMO is hereby released from all liability with reference to such defect and subsequent damage, if any, resulting from the inability of AMO to correct such defect.  MISUSE AND MISHANDLING ARE NOT COVERED UNDER WARRANTY. AMO’s sole obligation is to repair or replace, at AMO’s option, the defective part(s), which shall be returned freight prepaid to AMO by you and then returned to you freight collect once repaired or replaced.  The irrigation/aspiration handpiece set, diathermy forceps and diathermy cord are warranted for ninety (90) days. This warranty applies only to the original purchaser/user of the device and may not be transferred. This warrant will be void if and only so long as the equipment is used in any country other than the country to which it was originally shipped by AMO.  THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING WITHOUT LIMITATION THE WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL AMO BE LIABLE TO THE PURCHASER/USER FOR CONSEQUENTIAL, SPECIAL, INCIDENTAL, OR PUNITIVE DAMAGES, LOSS OF PROFITS OR LOSS OF USE. The liability of AMO for breach of this warranty shall, in any case, be limited to the total amount paid by you for the WHITESTAR SIGNATURE PRO System furnished by AMO and any work performed by AMO in connection therewith. Extended Warranty  Extended warranty contracts (excluding the phaco handpiece) are available. Contact AMO for information on the availability of an extended warranty contract. Note: If outside the United States, contact your local AMO office or representative for warranty information.
WHITESTAR SIGNATURE PRO System 9  • Warranty and Maintenance RX Only –NGP  Z370500 Rev. B 1116 9-3     Maintenance  User maintenance of the WHITESTAR SIGNATURE PRO System is only for those adjustments and corrective actions in the Error Messages and Troubleshooting, and Diagnostics sections of this manual. There are no user serviceable components within the console and you must not attempt to access the internal components. Any attempt to do so will void the warranty. AMO recommends, at least annually, routine or periodic maintenance of the WHITESTAR SIGNATURE PRO System by an AMO representative. AMO recommends the measurement of PE resistance and leakage current according to IEC 601-1 every two years.  If a problem continues following setup, check-out and troubleshoot as per the procedures in this manual, then contact AMO for corrective action (1-877-AMO-4-LIFE in USA 1-877-266-4543 or 1-800-511-0911 for outside the USA).
RX Only – NGP Z370500 Rev. B 1116 10-1   10 SPECIFICATIONS  Physical Specifications  Environmental Specifications  Electrical Specifications  Phacoemulsification Specifications  Irrigation and Aspiration Specifications  Vitrectomy Specifications  Diathermy Specifications  Diathermy Power Graphs  Diathermy Power versus Load Impedance  Phaco Power Graphs  Phaco Tip Velocity   Bottle Height Pressure
WHITESTAR SIGNATURE PRO System 10  • Specifications RX Only – NGP Z370500 Rev. B 1116 10-2     English / U.S. Metric: WHITESTAR SIGNATURE PRO System Console    Console dimensions (with display) Width 24 inches 61 cm  Depth 24 inches 61 cm  Height 59 inches 150 cm Power cord length  20 feet 6 meters Electrical enclosure current leakage  IEC 60601-1 compliance UL 60601-1  Weight (including IV pole)  185 pounds 84 kg Foot Pedal    Single Linear Foot Pedal    Dimensions Width 12 inches 31 cm  Length 10.5 inches 27 cm  Height 5.5 inches 14 cm Weight  10.0 pounds 4.5 kg Advanced Control Pedal    Dimensions Width 10.5 inches 27 cm  Length 14 inches 36 cm  Height 5.5 inches 14 cm Weight  15 pounds 7 kg Cord Length  11.8 feet 36 cm Advance Linear Pedal    Dimensions Width 9.82 inches 25 cm  Length 15.57 inches 39.6 cm  Height 4.88 inches 12.4 cm Weight  13 pounds 5 kg Cord Length  12 feet 366 cm Programmable IV Pole    Maximum travel  41 inches 106 cm Velocity  2.4 inches/sec 6 cm/sec. Maximum lift weight  2.43 pounds 1.1 kg    Physical Specifications Table 10.1 – System Physical Specifications
WHITESTAR SIGNATURE PRO System 10  • Specifications RX Only – NGP Z370500 Rev. B 1116 10-3    Phaco Power 1%-5% in 1% increments,   5-100%, in 5% increments Peristaltic Pump Vacuum Level 0–650 mmHg in 5 mmHg increments Pump Flow Rate 0-60 cc/minute Venturi Pump Vacuum Level 0-600 mmHg in 5 mmHg increments      English / U.S. Metric: Wireless Remote Control     Width 5 inches 13 cm  Length 5 inches 13 cm  Height 1.5 inches 4 cm  Weight 2.0 pounds .9 kg   Environmental Specifications Table 10.2 – Environmental Specifications  Environmental Specifications Operating Temperature 10 to 40°C Humidity Up to 95% RH, non-condensing  Ambient Pressure See Vitrectomy Specifications for additional information. Storage/Transportation Conditions Storage Temperature -20 to 60°C Humidity Up to 95% RH, non-condensing   Electrical Specifications Table 10.3 – Electrical Specifications   Voltage Frequency Rated Power Fuse Rating Enclosure Current Leakage System 100-240 Vac 50-60 Hz 750 VA 6.3A, 250V, Bussman GDA <500 µA   Phacoemulsification Specifications Table 10.4 – Phacoemulsification Specifications
WHITESTAR SIGNATURE PRO System 10  • Specifications RX Only – NGP Z370500 Rev. B 1116 10-4   Peristaltic Pump Vacuum Level 0–650 mmHg in 5 mmHg increments Pump Flow 0-60 cc/minute Venturi Pump Vacuum Level 0-600 mmHg in 5 mmHg increments  Peristaltic Pump Vacuum Level 0–650 mmHg in 5 mmHg increments  Pump Flow Rate 10-60 cc/minute, in 1 cc increments Venturi Pump Vacuum Level 0-600 mmHg in 5 mmHg increments Atmospheric Pressure/Altitude Density Altitude (DA) conditions are 2,500 cpm from sea level to 3,000 feet (914.4 m) elevation and shall be de-rated by 100cpm per 1,000ft (304.8 m) additional elevation up to a maximum of 10,000 ft above sea level. Indicated Cut Rate Actual Cut Rate Upper Limit (cpm) Actual Cut Rate Lower Limit (cpm) 2500 3000 2000    Irrigation and Aspiration Specifications Table 10.5 – Irrigation and Aspiration Specifications        Vitrectomy Specifications Table 10.6 – Vitrectomy Specifications
WHITESTAR SIGNATURE PRO System 10  • Specifications RX Only – NGP Z370500 Rev. B 1116 10-5   Power Adjustment 5% to 100%, in 5% increments Diathermy Power (maximum power into rated load) 8.5 Watts into 350 Ohms 386 kHz Diathermy Type Bipolar Rated Peak Voltage 160 Volts Connector of diathermy forcep/ probe Ø.072 RECEPTACLE (2)               All measurements are in inches.    Diathermy Specifications Table 10.7 – Diathermy Specifications
WHITESTAR SIGNATURE PRO System 10  • Specifications RX Only – NGP Z370500 Rev. B 1116 10-6     Diathermy Power Graphs  Figure 10.1 – Diathermy Output Power (Typical) 10 Ohm                         Y Axis = Watts  X Axis = Settings  Figure 10.2 – Diathermy Output Power (Typical) 50 Ohm                          Y Axis = Watts  X Axis = Settings
WHITESTAR SIGNATURE PRO System 10  • Specifications RX Only – NGP Z370500 Rev. B 1116 10-7     Figure 10.3 – Diathermy Output Power (Typical) 200 Ohm                          Y Axis = Watts  X Axis = Settings   Figure 10.4 – Diathermy Output Power (Typical) 350 Ohm                          Y Axis = Watts  X Axis = Settings
WHITESTAR SIGNATURE PRO System 10  • Specifications RX Only – NGP Z370500 Rev. B 1116 10-8     Figure 10.5 – Diathermy Output Power (Typical) 500 Ohm                         Y Axis = Watts  X Axis = Settings   Figure 10.6 – Diathermy Output Power (Typical) 1000 Ohm                          Y Axis = Watts  X Axis = Settings
WHITESTAR SIGNATURE PRO System 10  • Specifications RX Only – NGP Z370500 Rev. B 1116 10-9     Diathermy Power versus Load Impedance Figure 10.7 – Diathermy Output Power (50% Setting)                         Y Axis = Watts  X Axis = Impedance (Ohms)   Figure 10.8 – Diathermy Output Power (100% Setting)                             Y Axis = Watts  X Axis = Impedance (Ohms)
WHITESTAR SIGNATURE PRO System 10  • Specifications RX Only – NGP Z370500 Rev. B 1116 10-10     Phaco Power Graphs Figure 10.9 – Longitudinal Phaco Power                       Y Axis = Output Power (Watts)  X Axis = GUI Settings %   Figure 10.10 – ELLIPS FX Phaco Power Linear Mode                             Y Axis = Power (Watts)  X Axis = GUI Settings %
WHITESTAR SIGNATURE PRO System 10  • Specifications RX Only – NGP Z370500 Rev. B 1116 10-11     Phaco Tip Velocity Nominal Velocity for the ELLIPS FX handpiece and the WHITESTAR handpiece at 100% power, continuous, with the OPOR3020L phaco tip is shown in the following table.   Handpiece Nominal Longitudinal Velocity (m/s) ELLIPS FX (690880 (A)) 8.85 WHITESTAR (690697 (A)) 11.13   Bottle Height Pressure The WHITESTAR SIGNATURE PRO System has a built in IV pole to generate positive pressure using balanced salt solution to provide positive irrigation fluid. Measurement points:  •  All measurements are taken with a pressure meter at the same height as the pressure sensor in the WHITESTAR SIGNATURE PRO System. •  This position is defined as 41 inches or 101.6 cm from the floor.   1. Center of Drip Chamber 2. Center of pressure sensor is 41 inches or 104.2 cm from the floor
WHITESTAR SIGNATURE PRO System 10  • Specifications RX Only – NGP Z370500 Rev. B 1116 10-12     An IV bottle height reported as 0 cm includes an offset of 7.9 inches or 20 cm to make sure that a positive pressure is always provided.   Table 10.8 – Expected Pressure as a function of Bottle Height  Bottle Height cm Theoretical Pressure mmHg Corrected Pressure At Patient Eye Level mmHg 0 0.00 14.80 20 14.80 29.60 40 29.60 44.40 60 44.40 59.20 80 59.20 74.00 100 74.00 88.80 105 77.70 92.50
RX Only – NGP Z370500 Rev. B 1116 11-1   11   ACCESSORIES AND PARTS REORDERING       Accessories and Parts List
WHITESTAR SIGNATURE PRO System 11  • Accessories and Parts Reordering RX Only – NGP Z370500 Rev. B 1116 11-2     Table 11.1 – WHITESTAR SIGNATURE PRO System Part Number   Name Description Part number WHITESTAR SIGNATURE PRO Console An ophthalmic microsurgical system. NGP680301  List of Accessories with Part Numbers The following table is a list of accessories that you can use to reorder parts for your WHITESTAR SIGNATURE PRO System. Your AMO representative can advise you of recommended inventory levels based your facility's volume of phacoemulsification procedures. You can order any of the items listed below through your AMO representative or directly through AMO customer service. Your AMO representative can confirm the availability of these items in your area.  Table 11.2 – Accessories and Parts List   System Accessories Part Number Description Phaco Handpiece 690697 WHITESTAR Phaco Handpiece, Coaxial 690880 ELLIPS FX Phaco Handpiece OPOR0019L 19 Gauge LAMINAR Flow Phaco Tip, 0° Round Tip OPOR1519L 19 Gauge LAMINAR Flow Phaco Tip, 15° Round Tip OPOR3019L 19 Gauge LAMINAR Flow Phaco Tip, 30° Round Tip OPOR4519L 19 Gauge LAMINAR Flow Phaco Tip, 45° Round Tip OPOR0020L 20 Gauge LAMINAR Flow Phaco Tip, 0° Round Tip OPOR1520L 20 Gauge LAMINAR Flow Phaco Tip, 15° Round Tip OPOR3020L 20 Gauge LAMINAR Flow Phaco Tip, 30° Round Tip OPOR4520L 20 Gauge LAMINAR Flow Phaco Tip, 45° Round Tip OPOCR3019L Disposable LAMINAR Flow Phaco Tip, 19 Gauge, 30° Curved OPOCR3020L Disposable LAMINAR Flow Phaco Tip, 20 Gauge, 30° Curved OPOS19L 19 Gauge LAMINAR Flow Irrigation Sleeve and Test Chamber OPOS20L 20 Gauge LAMINAR Flow Infusion Sleeve and Test Chamber OPOMTWL LAMINAR Flow Tip Wrench I/A Handpiece OPOIA20CRV SOLO I/A Handpiece, 20 Ga., Curved Silicone Sleeve Tip OPOIA2045D SOLO I/A Handpiece, 20 Ga., 45° Silicone Sleeve Tip OPOIA20STR SOLO I/A Handpiece, 20 Ga., Straight Silicone Sleeve Tip OPOIA20KIT SOLO I/A Handpiece Kit OM055002 PHACOFIT Titanium Handle OM05510110 PHACOFIT I/A Tip Curved 0.5 mm Port
WHITESTAR SIGNATURE PRO System 11  • Accessories and Parts Reordering RX Only – NGP Z370500 Rev. B 1116 11-3     System Accessories Part Number Description  OM05510112 PHACOFIT O-Ring's I/A Tip (5PK) OM05510113 PHACOFIT Titanium, Multi-Tip Set OM05510114 PHACOFIT Cleaning/Flushing Kit OM05510115 PHACOFIT Tip, Luer Adapter (Aspiration) OM05510116 PHACOFIT Tip, 45 Degree Silicone (No Sleeve) OM0551011 PHACOFIT I/A Tip Straight 0.3 MM Port OM0551012 PHACOFIT I/A Tip Curved 0.3 mm Port OM0551013 PHACOFIT I/A Tip 45 Degree 0.3 mm Port OM0551014 PHACOFIT I/A Tip 90 Degree 0.3 mm Port OM0551015 PHACOFIT I/A Tip Binkhorst 0.3 mm Port OM0551016 PHACOFIT I/A Tip Straight SI 0.3 mm Port OM0551017 PHACOFIT I/A Luer Adapter, F/F OM0551018 PHACOFIT I/A Tip Curved, SI, 0.3 mm Port OM0551019 PHACOFIT I/A Tip Straight 0.5 MM Port OM0551020J PHACOFIT Athlete I/A Tip Straight Type OM0551021J PHACOFIT Athlete I/A Tip 45 Degree Type DUA02020 DUET STARTER KIT, 20GA DU02100 DUET ASPIRATION HANDLE DU02200 DUET IRRIGATION HANDLE DU02301 DUET 20G CURVED IRRIGATOR CLOSED END DU02302 DUET 20G CURVED ASPIRATION 0.3MM PORT CAP POL DU02303 DUET 19G FINE IRRIGATING CHOPPER DU02305 DUET 20G OLSON II IRRIGATING CHOPPER DU02325 DUET MST Aguirre Duet Tip DU02335 DUET MST Fine/Nagahara Tip Diathermy Handpiece SCP680600 Diathermy Power Cable K121085 Bipolar Forceps, Curved Iris 0.5 mm Tip K147000 18 Gauge Straight Diathermy Pencil, reusable Vitrectomy Cutter NGP0020 20 gauge Vitreous Cutter NGP0023 23 gauge Vitreous Cutter NGP0025 25 gauge Vitreous Cutter
WHITESTAR SIGNATURE PRO System 11  • Accessories and Parts Reordering RX Only – NGP Z370500 Rev. B 1116 11-4     System Accessories Part Number Description FUSION Pack OPO70 FUSION Pack (disposable, sterile) OPO71 FUSION Dual Pump Pack (disposable, sterile) Wireless Remote Control NGP680137 WHITESTAR SIGNATURE PRO System Remote Control 0100-1834 AC Charger, SIGNATURE Wireless Remote Foot Pedal NGP680701 WHITESTAR SIGNATURE System Single Linear Foot Pedal NGP680702 WHITESTAR SIGNATURE System Advanced Control Pedal NGP680703 WHITESTAR SIGNATURE PRO System Advanced Linear Foot Pedal 0100-1835 AC Charger, SIGNATURE Wireless Foot Pedal 0100-0055 CABLE ASSY, WHITESTAR SERIES FOOT PEDAL (Single Linear Foot Pedal and Advanced Control Pedal) 0100-6303 Cable Assy, Advanced Linear Foot Pedal to Console, Signature 0100-1836 LI-BATTERY, SIGNATURE PRO Advanced Linear Pedal Miscellaneous 2410-0049-L Power cord, hospital grade, 20 ft., 6.1 meters OM05510114 I/A Cleaning Kit 0100-0750 WHITESTAR SIGNATURE System Dust Cover MSR309 Sterilization Tray Operator’s Manual NGP Z370494 WHITESTAR SIGNATURE PRO System Operator’s Manual (Bulgarian) NGP Z370495 WHITESTAR SIGNATURE PRO System Operator’s Manual (Chinese) NGP Z370496 WHITESTAR SIGNATURE PRO System Operator’s Manual (Czech) NGP Z370497 WHITESTAR SIGNATURE PRO System Operator’s Manual (Croatian) NGP Z370498 WHITESTAR SIGNATURE PRO System Operator’s Manual (Danish) NGP Z370499 WHITESTAR SIGNATURE PRO System Operator’s Manual (Dutch) NGP Z370500 WHITESTAR SIGNATURE PRO System Operator’s Manual (English) NGP Z370501 WHITESTAR SIGNATURE PRO System Operator’s Manual (Estonian) NGP Z370502 WHITESTAR SIGNATURE PRO System Operator’s Manual (Finnish) NGP Z370503 WHITESTAR SIGNATURE PRO System Operator’s Manual (French)
WHITESTAR SIGNATURE PRO System 11  • Accessories and Parts Reordering RX Only – NGP Z370500 Rev. B 1116 11-5     System Accessories Part Number Description  NGP Z370504 WHITESTAR SIGNATURE PRO System Operator’s Manual (German) NGP Z370505 WHITESTAR SIGNATURE PRO System Operator’s Manual (Greek) NGP Z370506 WHITESTAR SIGNATURE PRO System Operator’s Manual (Hungarian) NGP Z370507 WHITESTAR SIGNATURE PRO System Operator’s Manual (Italian) NGP Z370508 WHITESTAR SIGNATURE PRO System Operator’s Manual (Japanese) NGP Z370509 WHITESTAR SIGNATURE PRO System Operator’s Manual (Korean) NGP Z370510 WHITESTAR SIGNATURE PRO System Operator’s Manual (Latvian) NGP Z370511 WHITESTAR SIGNATURE PRO System Operator’s Manual (Lithuanian) NGP Z370512 WHITESTAR SIGNATURE PRO System Operator’s Manual (Norwegian) NGP Z370513 WHITESTAR SIGNATURE PRO System Operator’s Manual (Polish) NGP Z370514 WHITESTAR SIGNATURE PRO System Operator’s Manual (Portuguese) NGP Z370515 WHITESTAR SIGNATURE PRO System Operator’s Manual (Portuguese-Brazilian) NGP Z370516 WHITESTAR SIGNATURE PRO System Operator’s Manual (Romanian) NGP Z370517 WHITESTAR SIGNATURE PRO System Operator’s Manual (Russian) NGP Z370518 WHITESTAR SIGNATURE PRO System Operator’s Manual (Serbian) NGP Z370519 WHITESTAR SIGNATURE PRO System Operator’s Manual (Slovakian) NGP Z370520 WHITESTAR SIGNATURE PRO System Operator’s Manual (Slovenian) NGP Z370521 WHITESTAR SIGNATURE PRO System Operator’s Manual (Spanish) NGP Z370522 WHITESTAR SIGNATURE PRO System Operator’s Manual (Swedish) NGP Z370523 WHITESTAR SIGNATURE PRO System Operator’s Manual (Turkish)
WHITESTAR SIGNATURE PRO System 11  • Accessories and Parts Reordering RX Only – NGP Z370500 Rev. B 1116 11-6     System Accessories Part Number Description  NGP Z370526 WHITESTAR SIGNATURE PRO System Operator’s Manual (Traditional Chinese) NGP Z370524 WHITESTAR SIGNATURE PRO System Operator’s Manual CD, Multilingual
RX Only – NGP Z370500 Rev. B 1116 12-1   12 INDEX    A Accessories 11-1 Add a Surgeon 4-3 Advanced Control Pedal 1-4, 1-18, 2-5, 10-2 Calibration 4-23 Connections 4-17 Default Settings 4-16 Setup 4-16 Switch Assignment 4-17 Testing 4-23 Threshold Configuration 4-22 Advanced Linear Pedal 1-4, 2-5 Battery Replacement 8-4 Connections 4-25 Handle 2-7 Setup 4-25 Switch Assignment 4-26 Test 4-32 Threshold Configuration 4-30 Anterior chamber too deep 8-9 Anterior Segment Surgery 3-2 Aspiration Specifications 10-4 Aspiration Rate Occlusion 5-24 Assign Order Program 5-29 Surgeon 4-35  B Back-light button 8-7 Backup All 6-21 Balance Irrigation/Aspiration 3-14 Batteries Remote control 2-9 Battery Charging 4-16, 4-25, 8-6 Replacement 8-4 Begin Tune 3-13 Blocked Aspiration port 5-26 BLUETOOTH 8-7 Bottle Height IA 5-11 Phaco 5-8 Vitrectomy 5-12 Bypass 3-11, 3-15  C Calibrate Foot Pedal Advanced Control Pedal 4-23 Single Linear 4-14 Touch Screen 6-11 CASE Mode 5-26 CASE ONE Touch 1-7, 5-27 CASE Vacuum 5-27 Chamber partially collapses 8-11 Changing Mode Parameters 3-16 Charging Battery 4-16, 4-25, 8-6 Checkout System 3-17 Cleaning Touch Screen 7-2 Cleaning/Care 7-2 Common Problems 8-2 Connecting Handpieces 1-9
WHITESTAR SIGNATURE PRO System 12  • Index RX Only – NGP Z370500 Rev. B 1116 12-2    Connections Advanced Control Pedal 4-17 Front Panel 3-2 Rear Panel 3-3, 4-8 Console Dimensions 10-2 Continuous Burst 5-13 Phaco Power 5-15 Continuous Irrigation 1-5, 3-11 Cup Fill 3-10 Continuous Phaco Power 5-14 Copy a Program 5-29 Create Surgeon 5-2 Cup Fill 3-10 Custom Duty Cycles 5-16 Cut Mode Vitrectomy 5-12 Cut Rate Vitrectomy 5-12 Cutter Vitrectomy 3-7   D Database Delete 6-25 Import/Export 6-19 Restore 6-23, 6-25 Date Set 6-16 Default Settings Diathermy 5-12 Foot Pedal Advanced Control Pedal 4-16 Single Linear 4-7 Irrigation/Aspiration 5-11 Phaco 5-8 Vitrectomy 5-12 Delete Database 6-25 Program 5-2, 5-28 Delete a Backup Database 6-25 Diagnostics 8-1 Diathermy 1-7, 1-10 Default Settings 5-12 Forceps 3-6 Handpiece 3-6, 7-2 Output Power 10-9 Diathermy Handpiece Part Numbers 11-3 Dimensions 10-2 Dual Pump 1-5 Duty Cycles 5-15   E Edit Program 5-3 Surgeon 4-4 Effective Phaco Time 3-21 Electrical Specifications 10-3 ELLIPS Handpiece 1-4, 3-4, 3-5 End Case 3-21 Environmental Specifications 10-3 EPT 3-21 Export 6-19  F Failed to pair 8-8 Feedback Foot Pedal 4-7 Fluidics 1-3 Fluidics Mode Phaco 5-8
WHITESTAR SIGNATURE PRO System 12  • Index RX Only – NGP Z370500 Rev. B 1116 12-3    Foot Pedal 1-4 Advanced Control Pedal 1-18 Calibrate 4-23 Connections 4-17 Default Settings 4-16 Setup 4-16 Switch Assignment 4-17 Testing 4-23 Threshold Configuration 4-22 Advanced Linear Pedal 2-5 Configuration 4-30 Connections 4-25 Setup 4-25 Test 4-32 Feedback 4-7 Modes Vitrectomy 5-6 Part Number 11-4 Pitch 2-7 Pitch Regions 4-7 Positions 2-7, 5-16 Single Linear 1-4, 2-5 Calibration 4-14 Default Settings 4-7 Setup 4-6 Switch Assignment 4-9 Testing 4-14 Threshold Configuration 4-12 Specifications 10-2 Forceps 9-2 Diathermy 3-6 Front Panel 3-2 Fuse Replacement 8-2 FUSION 5-26 FUSION Fluidics 5-20, 5-22, 5-26 FUSION Tubing Pack 2-3, 11-4   G Graphic User Interface 1-5 Graphs Phaco Power 10-10 GUI 1-5 H Handle Advanced Linear Pedal 2-7 Handpiece Cleaning 7-2 Connecting 1-9 IA 3-6 Phaco 3-5 Setup Diathermy 3-6 I/A 3-6 Phaco 3-4 Vitrectomy 3-7 Ultrasonic 1-9 Warranty 9-2 Help 8-14 High Power Limit 5-19 High Power Pulse Phaco Power 5-14   I IA Handpiece 3-6 IAC 5-6 ICA 5-6 ICE Pulse Shaping WHITESTAR Technology 5-19 Import 6-20 Import/Export Database 6-19 Inspection Instructions 2-2 Install Tubing Cassette 3-9 Irrigation Changing 1-8 Continuous 3-11 Flow does not stop 8-9 Inadequate 1-9 Irrigation/Aspiration 1-7 Balance 3-14 Default Settings 5-11
WHITESTAR SIGNATURE PRO System 12  • Index RX Only – NGP Z370500 Rev. B 1116 12-4     IV Pole Does not work 8-8 Programmable 1-10, 2-8 Setup 3-3 Testing 6-3   K Kick 5-19  L LAMINAR Flow 3-6 Language Selection 6-14 Locked Program 5-28 Long Pulse Phaco Power 5-14 Power 5-13 Low batteries 8-8 Low Power Limit 5-19 Low Power Pulse Phaco Power 5-14 Lower Threshold 5-27   M Maintenance 9-3 Max Aspiration IA 5-11 Vitrectomy 5-12 Max Power Diathermy 5-12 Max Vacuum IA 5-11 Vitrectomy 5-12 Maximum Height IV Pole 6-2 Maximum Vacuum 5-20 Mayo Tray Drape 2-3 Miscellaneous Part Numbers 11-4 Mode Programming 1-6 Modes Changing Parameters 3-16 Monitor Drape 2-3 Multiple Burst Phaco Power 5-15 Power 5-13 Multiple Mode Programming 1-6   N No aspiration 8-10 No diathermy or poor diathermy 8-13 No phacoemulsification 8-12 No sound when using diathermy 8-13 No vitrectomy cutting 8-13   O Occluded Aspiration Rate 5-21 Occlusion Aspiration Rate 5-24 Vacuum Threshold 5-24 Venting 5-26 Occlusion Mode Phaco Settings 5-21 One Touch CASE 5-27 Operating Modes 1-6 Output Power Diathermy 10-9  P Pairing Remote Control 6-10 Wired 6-7 Wireless 6-5 Wireless Remote Control 8-7 Panel Front 3-2 Rear 3-3 Parameter Settings WHITESTAR Technology 5-16 Parameters Programming 1-5 Parts 11-1
WHITESTAR SIGNATURE PRO System 12  • Index RX Only – NGP Z370500 Rev. B 1116 12-5     Passive Reflux 5-28 Pencil Probe Diathermy 3-6 Percent Kick High End of Range 5-19 Low End of Range 5-19 Peristaltic Pump 10-4 Phaco Default Settings 5-8 Handpiece 3-4 Phaco Handpiece 1-4, 1-10, 3-4, 3-5, 7-2 Part Numbers 11-2 Phaco Power Continuous 5-14 Graph 10-10 Phacoemulsification 1-2, 1-6, 1-9, 3-4, 8-12 Physical Specifications 10-2 Pitch 2-7 Pitch Regions Foot Pedal 4-7 Poor aspiration 8-10 Poor phacoemulsification 8-12 Positions Foot Pedal 2-7 Power Continuous Burst 5-13 Cord Length 10-2 Failure 1-9 Long Pulse 5-13 Multiple Burst 5-13 Short Pulse 5-13 Single Burst 5-13 Switch 8-2 Power Delivery Diathermy 5-12 pps 5-14 Precautions Safety 1-8 Prime Bypass 3-11 Prime/Tune 1-5, 3-10, 3-11 Priming Vitrectomy 3-15 Program Assign Order 5-29 Delete 5-2 Lock 5-28 Locked 5-28 Select 3-8 Program Name Edit 5-3 Programmable IV Pole 1-6, 1-10, 2-8 Setup 3-3 Specifications 10-2 Testing 6-3 Pulse Off Time 5-16 On time 5-16 Pulses per second 5-14 Rate 5-16 Pulse Shaping 5-18 Pump Ramp Threshold IA 5-11   R Rear Panel 3-3 Reduced/insufficient irrigation flow 8-9 Reflux 5-28 Passive 5-28 Remote Control Pairing 6-10 Wireless 2-8 Remote Control, Wireless Setup 6-5 Reordering 11-1 Replacing Batteries 1-14 Fuse 8-2 Reset Timers 3-14, 3-18 Restore Database 6-25 Restore All 6-24 Restore Database 6-23 Restriction of Hazardous Substances 1-22 RoHS 1-22
WHITESTAR SIGNATURE PRO System 12  • Index RX Only – NGP Z370500 Rev. B 1116 12-6    S Screen, Touch Calibrate 6-11 Selecting a Language 6-14 Self Test 3-8 Service Interval 6-18 set vacuum units 1-7 Settings Occlusion Mode 5-21, 5-25 Post Laser Cataract 5-10 Setup Advanced Control Pedal 4-16 Surgical Media Center (SMC) 6-15 System 3-2 Wireless Remote Control 6-5 Short Pulse Phaco Power 5-14 Power 5-13 Shutdown Sequence 3-20 System 3-20 Side vitrectomy 5-6 Single Burst Phaco Power 5-15 Power 5-13 Single Linear Foot Pedal Calibration 4-14 Default Settings 4-7 Setup 4-6 Switch Assignment 4-9 Testing 4-14 Threshold Configuration 4-12 Sleep Mode 8-7 SMC (Surgical Media Center) 2-9 Software Versions View 6-12 SOLO Handpiece 3-6 Sounds Set and test 4-33 Specifications Electrical 10-3 Environmental 10-3 Foot Pedal 10-2 Fuse 8-2 IA 10-4 IV Pole 10-2 Phacoemulsification 10-3 Physical 10-2 Vitrectomy 10-4 Startup 3-8 Status 8-14 Sterilization 3-20 Storage Conditions 10-3 Surgeon Add 4-3, 5-2 Assign Order 4-35 Copy 5-29 Surgical Media Center (SMC) 2-9 Setup 6-15 Switch Assignment Advanced Control Pedal 4-17 Advanced Linear Pedal 4-26 Single Linear Foot Pedal 4-9 Symbol Definitions 1-15 System Checkout 3-17 Components 2-2 Console 1-5 Description 1-2 Disposal 1-22 Does not turn on 8-7 Self Test 3-8 Shutdown 3-20 System Date and Time 6-16 System Self Test 6-13   T Technical Service 8-2 Test Advanced Control Pedal 4-23 Advanced Linear Pedal 4-32 Foot Pedal Single Linear 4-14 System 6-13 Wireless Remote Control 6-10 Test Chamber 2-3, 11-2
WHITESTAR SIGNATURE PRO System 12  • Index RX Only – NGP Z370500 Rev. B 1116 12-7     Threshold Vacuum 5-24 Threshold Configuration Advanced Control Pedal 4-22 Advanced Linear Pedal 4-30 Single Linear Foot Pedal 4-12 Time Set 6-16 Timers Reset 3-14 Touch Screen Calibrate 6-11 Cleaning 7-2 Does not respond 8-8 Using 3-16 Tubing Cassette 2-3 Install 3-9 Tune 3-11 Begin 3-13  U Ultrasonic (U/S) Time 3-21 Ultrasonic Handpiece 1-9 Units of Measure 6-17 Units of measure Set units of measure for vacuum 6-17 Unpair 8-7 Up Time Threshold 5-27 Upper Threshold 5-27 Using too much fluid 8-10 Touch Screen 3-16 Venturi pump 5-27 View Software Versions 6-12 Vitrectomy 1-7, 1-10, 3-7 Advanced Control Pedal Modes 4-17 Cutter 1-10 Default Settings 5-12 Foot Pedal Modes 4-7 Handpiece 3-7 Prime 3-15   W Wake-up 2-5 Warnings 1-8, 1-11 WHITESTAR ICE Technology 1-3 WHITESTAR Technology 1-2, 5-15 Wired Pairing 6-7 Wireless Pairing 6-5 Wireless Remote Control 1-4, 1-11, 2-8 Does not pair 8-7 Part Number 11-4 Wireless Remote Test 6-10 Wireless Setup Foot Pedal 6-5 Remote Control 6-5   Y Yaw 2-7   V Vacuum monitoring 5-20 pressure 5-26 Set units of measure 6-17 Vacuum units 1-7 Venting Occlusion 5-26
WHITESTAR SIGNATURE PRO System 12  • Index RX Only – NGP Z370500 Rev. B 1116 12-8

Navigation menu