Advanced operation(key-operator mode)This chapter gives an overview of functions for the advanced user:!Overview of key-operator functions!Quality Control!Preventive maintenance schedule!Cleaning the exterior!Cleaning the print head!Troubleshooting checklistChapter 3
46 2831A EN 20041201Advanced operation (key-operator mode)DRYSTAR 5302 USER MANUALOverview of key-operator functionsThe key-operator menus make it possible to use the Drystar 5302 advancedfunctions.For general information on the key functions of the Drystar 5302, refer to ‘Theuser interface’ on page 20.OverviewThe Drystar 5302 features the following functions on the main menu level ofthe key-operator mode:These functions are described in detail in the Drystar 5302 Reference manual.Menu item FunctionPage (Ref. Man.)Show settings To consult the current settings of the printer.50Change settings To change the current settings of the printer.56Print imageTo print one of the Drystar 5302 test images.To load and print images from an external CF-card.86Save configuration To make a backup of the printer settings.90Restore configuration To restore the backup of the printer settings.92Calibration To calibrate the printer.97Service actions To view error, repair and maintenance data.105Quality Control To perform the QC procedure.(User Manual)47 Installation To install or update the Drystar 5302 software.118Refer to the indicated page of the Drystar 5302 Reference manual for an explanation of the function and the appropriate procedures.
472831A EN 20041201 Advanced operation (key-operator mode)DRYSTAR 5302 USER MANUALQuality ControlIn order to establish and maintain consistent image quality, a regularevaluation of image quality is advised.The Drystar 5302 contains an automatic QC feature that has been designed tocomply with the grayscale reproduction constancy test, according to theinternational standard IEC 1223-2-4.Local Regulations may require other procedures.The Drystar 5302 QC procedure consists of two main steps:•Before initial use, establishing a number of reference values that will be used for further follow-up and verifying initial image quality. •After establishing these values, performing regular daily, weekly and annual quality tests.The results of these tests are recorded on Quality Control Charts.The QC image (Refer to ‘QC test image’ on page 52) has several additionalfields where the QC data can be filled in. This image should be filed as part ofthe QC procedure.For more information, please refer to ‘Quality Control Charts’ on page 75.
48 2831A EN 20041201Advanced operation (key-operator mode)DRYSTAR 5302 USER MANUALEstablishing the reference values and verifying image qualityAfter installation of a new Drystar 5302 and before initial use you mustestablish Quality Control aim values. These values will be used as the baseline for comparison when daily Quality Control is done. These values shouldbe determined again after major service, repair or software update.The following Quality Control aim values must be determined:•The daily operating density levels. Refer to ‘Establishing the daily operating reference density levels’ on page 49.•Drystar 5302 image geometry. Refer to ‘Establishing the image geometry reference values’ on page 52.Once Quality Control aim values are established you must evaluate theSpatial Resolution, the Artifact Levels and the Low Contrast Visibility todetermine if the image quality is acceptable. Refer to ‘Verifying AcceptableSpatial Resolution, Artifact Levels and Low Contrast Visibility’ on page 54.The Quality Control aim values, the Spatial Resolution and Artifact Levels andthe Image Geometry values are all recorded on the Quality Control charts.Refer to ‘Quality Control Charts’ on page 75.On these charts, the following test conditions are also recorded:•The type and serial number of the Drystar 5302.•The type and emulsion number of the film used to determine the reference values.•The time (day, month, year) that the values were established.Before you can establish the daily operating levels, the Drystar 5302 must be switched on for at least 15 minutes and it must be calibrated as well. Refer to ‘Switching on the Drystar 5302’ on page 30 and ‘Performing the calibration procedures’ on page 97 of the Drystar 5302 Reference manual.
492831A EN 20041201 Advanced operation (key-operator mode)DRYSTAR 5302 USER MANUALEstablishing the daily operating reference density levelsThis procedure enables you to establish the base line values for:•Low density•Mid density•High densityTo establish the daily operating levels, proceed as follows:1Press the Key-operator key   to enter the Key-operator mode.2Press the down key seven times, followed by the ok key to select ‘QC’.The ‘QC’ screen appears:3Press the ok key to continue.The Drystar 5302 will automatically print the QC Test image.4After the image is printed, the system will display the optical density values:The densitometer of the Drystar 5302 is calibrated at installation. Authorized service personnel should recalibrate the densitometer annually or after major service or repair.1 Show settings2 Change settings3 Print image4 Save configuration5 Restore configuration6 Calibration7 Installation8 QC quit ok  selectKey-operatormain menuQCStart printing test imageto perform:daily / weeklyand yearly control quit okKey-operatorQCQCInternal Density readingsLow density :0.26Mid density :1.35High density :1.89(copy on control chart) okKey-operatorQC
50 2831A EN 20041201Advanced operation (key-operator mode)DRYSTAR 5302 USER MANUALThe displayed values represent the following steps on the test film:•Low density: the density of the Low density step.Target: 0.4.•Mid density: the density value of the Mid density step.Target: 1.2.•High density: the density value of the High density step.Target: 2.0.5Record the density levels on the Drystar 5302 Chart 1 (‘Determination of the operating levels’). Refer to ‘Quality Control Charts’ on page 75.6Press the ok key. The following screen is displayed:If the mid density value does not meet or exceeds the recommended values, the reason must be found and the problem solved before any further clinical films can be printed.Refer to ‘Preventive maintenance schedule’ on page 59 and ‘Maintaining image quality and resolving image quality problems’ on page 173  of the Drystar 5302 Reference manual, or call your local Agfa service organization.QCProceed with the QC proce-dure as stated in the User man-ual  okKey-operatorQC
512831A EN 20041201 Advanced operation (key-operator mode)DRYSTAR 5302 USER MANUAL7Press the ok key to return to the main menu.8Repeat 1 steps through 7 once a day for five consecutive days, as indicated on the Drystar 5302 Chart 1.9Calculate the average value of the densities from the five images. These values represent operating levels, or aim values, for each density.10 Record the respective aim (average) values as the ‘Operating levels’ on the Drystar 5302 Charts 2a and 2b (‘daily Drystar 5302 control chart’). Refer to ‘Quality Control Charts’ on page 75.The calculated ‘Operating levels’ should be as following:11 These charts will be used for the daily quality test. For more information, refer to ‘Performing the daily QC test’ on page 55.Operating Level Value(according IEC 1223-2-4 or beter)Low density  0.4 ± 0.05Mid density  1.2 ± 0.15High density  2.0 ± 0.2
52 2831A EN 20041201Advanced operation (key-operator mode)DRYSTAR 5302 USER MANUALEstablishing the image geometry reference values1Print the QC test image or use the previously printed test image.You should obtain an image looking like this (without the dimensions A and B):QC test image 2To determine the reference values for geometry, measure the dimensions A and B of the geometric square on the test image.Make sure to measure distance A from the left edge of the left line to the right edge of the right line and distance B from the upper edge of the upper line to the lower edge of the lower line.We strongly recommend using a 30 cm (12-inch) machinist scale with 0.5 mm divisions (1/64 inch).Drystar 5302Quality Control Test ImageDate: __________Time: __________Initials: ________       Weekly Spatial Resolution TestGeometry Test Daily density testMax DHi DMid DLo DBase + FogDD Density Difference(Hi D - Lo D)Sets of dots visible?Yes NoDimension ADimension B1 2
532831A EN 20041201 Advanced operation (key-operator mode)DRYSTAR 5302 USER MANUAL3Record these values as reference dimensions A ref and B ref on the Drystar 5302 Chart 4 (‘Drystar 5302 Geometric Consistency Control Chart’). Refer to ‘Quality Control Charts’ on page 75.These charts will be used for the annual quality test. For more information, refer to ‘Performing the Annual QC tests’ on page 58.4Save this film for future reference.
54 2831A EN 20041201Advanced operation (key-operator mode)DRYSTAR 5302 USER MANUALVerifying Acceptable Spatial Resolution, Artifact Levels and Low ContrastVisibility1Print the QC Test image or use the previously printed QC Test image used to establish the daily operating density levels.2Visually check the QC test image for artifacts: no significant disturbing artifacts should be visible.3Check the spatial resolution in each of the three ovals. Within each oval there are three groups, each having five dots. All five dots of each group must be visible with a magnifying glass. The smallest cluster of 5 dots are only visible if the viewing conditions are good.4Record these values at the top of the Drystar 5302 Chart 3 (Drystar 5302 Artifacts and Spatial Resolution Control Chart). Refer to ‘Quality Control Charts’ on page 75.5Check the Low Contrast Visibility at both the high (100 / 95%) and low end (0 / 5%) of the density scale. You should be able to see the circle in the square (refer to item 1 on the ‘QC test image’ on page 52) and the upper circle (refer to item 2 on the ‘QC test image’ on page 52).6These charts will be used for the weekly quality test. For more information, refer to ‘Performing the Weekly QC tests’ on page 57.Good viewing conditions are important for the correct interpretation of both diagnostic and test images. Make sure that the lightbox intensity (luminance) is between 2000 and 4000 cd/m² (4500 and 6500 °K). Use a magnifying glass and use shutters to collimate. Make sure the ambient light is low.In case of significant artifacts or insufficient spatial resolution, the reason must be found and the problem solved before any further clinical films can be printed.Refer to ‘Preventive maintenance schedule’ on page 59 and ‘Maintaining image quality and resolving image quality problems’ on page 173  of the Drystar 5302 Reference manual, or call your local Agfa service organization.
552831A EN 20041201 Advanced operation (key-operator mode)DRYSTAR 5302 USER MANUALPerforming quality control (QC) testsThe following procedures must be performed daily, weekly or annually asindicated.The reason for performing quality control tests is to determine if anysignificant image quality variation or deterioration has occurred which mayrequire corrective action. Comparing the results of the tests with the referencevalues previously established does this.This procedure allows the operator to take the necessary preventive actionsbefore any image quality loss can take place.Performing the daily QC test1Turn on the Drystar 5302 and wait at least for 15 minutes. Refer to ‘Switching on the Drystar 5302’ on page 30.2Press the Key-operator   key to enter the Key-operator mode.3Press the down key seven times, followed by the ok key to select ‘QC’.The ‘QC’ screen appears:4Press the ok key to continue.The Drystar 5302 will automatically print the QC Test image.This test must be performed every day before any clinical film can be processed.1 Show settings2 Change settings3 Print image4 Save configuration5 Restore configuration6 Calibration7 Installation8 QC quit ok  selectKey-operatormain menuQCStart printing test image to perform:daily / weeklyand yearly control quit okKey-operatorQC
56 2831A EN 20041201Advanced operation (key-operator mode)DRYSTAR 5302 USER MANUALAfter the image is printed, the system will display the optical density values:5Record the density values on the Drystar 5302 Charts 2A and 2B (Drystar 5302 Daily Density Control Chart’). Also record the date and time of the test on the charts and on the QC test images. Refer to ‘Quality Control Charts’ on page 75.6Press the ok key. The following screen is displayed:7Press the ok key to return to the main menu.In case the measure results are not within the aim values, the reason for the unacceptable density variations must be identified and resolved before any further clinical films can be processed. This may include repeating the film calibration procedure.For possible causes of non-compliance and the respective actions, refer to ‘Preventive maintenance schedule’ on page 59 and ‘Maintaining image quality and resolving image quality problems’ on page 173 of the Drystar 5302 Reference manual.QCInternal Density readingsLow density :0.19Mid density :1.25High density: 1.78(copy on control chart) okKey-operatorQCQCProceed with the QC proce-dure as stated in the User man-ual okKey-operatorQC
572831A EN 20041201 Advanced operation (key-operator mode)DRYSTAR 5302 USER MANUALPerforming the Weekly QC testsSpatial Resolution, Artifact Test and Low Contrast VisibilityTo identify artifacts and verify spatial resolution you must perform thefollowing test weekly or as needed for troubleshooting image qualityproblems.1Check the QC test image visually for artifacts: no significant disturbing artifacts should be visible.2Check the spatial resolution.The test film also shows three squares which each contains an oval. These 3 ovals contain 3 groups, each having 5 dots. All five dots of each group must be visible with a magnifying glass. The smallest cluster of 5 dots are only visible if the viewing conditions are good.3Check the Low Contrast Visibility at both the high (100 / 95%) and low end (0 / 5%) of the density scale. You should be able to see the circle in the square (refer to item 1 on the ‘QC test image’ on page 52) and the upper circle (refer to item 2 on the ‘QC test image’ on page 52).4Record these values on the Drystar 5302 Chart 3 (Drystar 5302 Artifacts and Spatial Resolution Control Chart).Good viewing conditions are important for the correct interpretation of both diagnostic and test images. Make sure that the lightbox intensity (luminance) is between 2000 and 4000 cd/m² (4500 and 6500 °K). Use a magnifying glass and use shutters to collimate. Make sure the ambient light is low.In case of significant artifacts, insufficient spatial resolution or failure of any other recommended QC tests, the cause of the problem must be identified, and corrective action must be taken before the Drystar 5302 can be used for any further clinical imaging.Refer to‘Preventive maintenance schedule’ on page 59 and ‘Maintaining image quality and resolving image quality problems’ on page 173 of the Drystar 5302 Reference manual, or call your local Agfa service organization for assistance.
58 2831A EN 20041201Advanced operation (key-operator mode)DRYSTAR 5302 USER MANUALPerforming the Annual QC tests Geometric Consistency Test To be able to notice fluctuations in image size and aspect ratio, you mustperform this procedure once a year.1First, perform the daily test.2Use the QC test image of the weekly test and measure the dimensions A and B of the geometric square. Refer to ‘Establishing the image geometry reference values’ on page 52.3Record these values as measured dimensions A and B on Chart 4 (‘Drystar 5302 Geometric Consistency Control Chart’).4Compare the measured A and B dimensions with the reference dimension values, A ref and B ref on the Drystar 5302 Chart 4 (‘Drystar 5302 Geometric Consistency Control Chart’).The differences between measured dimensions of A and B and the reference values A ref and B ref should be less than or equal to 1.0%.5Check for image distortion.6Calculate the aspect ratio by dividing A by B.The result must be 1 +/- 0.01Make sure to measure distance A from the left edge of the left line to the right edge of the right line and distance B from the upper edge of the upper line to the lower edge of the lower line.We strongly recommend using a 30 cm (12-inch) machinist scale with 0.5 mm divisions (1/64 inch).If the image size or distortion values are outside of limits, contact Agfa service to resolve the problem.
592831A EN 20041201 Advanced operation (key-operator mode)DRYSTAR 5302 USER MANUALPreventive maintenance scheduleThe Drystar 5302 is designed for trouble-free service. Maintenance andcleaning involve only some minor user tasks. Refer to the following pages forthe appropriate cleaning procedure.Safety guidelines•Do not lubricate the printer.•Do not attempt to disassemble the printer.•Do not touch the resistor line of the print head.•Always switch off the Drystar 5302 and disconnect the power cord from the outlet before carrying out any maintenance work inside the printer.Interval What to do? PageAd hoc ‘Cleaning the exterior’ 60When image quality tends to degrade. An appropriate warning message is displayed.‘Cleaning the print head’ 61To prevent damage to the printer while performing maintenance, observe the following safety precautions:Film jam removal or cleaning the printer head can be done without switching the power off. Nevertheless, care should be taken and the ‘Safety precautions’ on page 9 should be respected.
60 2831A EN 20041201Advanced operation (key-operator mode)DRYSTAR 5302 USER MANUALCleaning the exterior1Switch off the Drystar 5302 by following the procedure as described in ‘Switching off the Drystar 5302’ on page 32.2Remove the power plug from the socket.3Wipe the exterior of the printer with a clean, soft, damp cloth.Use a mild soap or detergent if required but never use an ammonia–based cleaner. Be careful not to get any liquid in the power cord port.4Plug in the printer and switch it on by following the procedure as described in ‘Switching on the Drystar 5302’ on page 30.
612831A EN 20041201 Advanced operation (key-operator mode)DRYSTAR 5302 USER MANUALCleaning the print headTo clean the print head:1Press the Key-operator key to enter the key-operator mode.2On the key-operator main menu, press the Down key five times, followed by the Confirm key to select ‘Calibration’.3On the Calibration menu, press the Down key, followed by the Confirm key to select ‘Therm. Head clean.’.4The ‘Thermal head cleaning’ screen will give step by step instructions on what to do:Print head cleaning must be done when image quality problems occur.1 Show settings2 Change settings3 Print image4 Save configuration5 Restore config.6 Calibration7 Service actions8 Quality control9 InstallationKOCACALIBRATION1 Film Calibration2 Therm. Head clean.THERMAL HEADCLEANINGOpen top coverCA
62 2831A EN 20041201Advanced operation (key-operator mode)DRYSTAR 5302 USER MANUAL5Open the top cover.6As soon as the top cover is opened, the ‘Thermal head cleaning’ screen continues giving the following instructions:7Open the hold-down bracket.THERMAL HEADCLEANINGClean thermal headClose top coverCA
632831A EN 20041201 Advanced operation (key-operator mode)DRYSTAR 5302 USER MANUAL8Open the print head unit.9Locate and check on sight the print head resistor line.The print head unit can be warm.Be careful not to touch the print head resistor line with your fingers.
64 2831A EN 20041201Advanced operation (key-operator mode)DRYSTAR 5302 USER MANUAL10 Clean the print head resistor line.Gently pass over the resistor line a few times with a lint free cloth, slightly moistened with Isopropyl alcohol or Ethanol. Do this only in one direction, i.e. from left to right, without lifting the cloth.11 Close the print head unit, the hold-down bracket and finally the top cover.After you have cleaned the print head resistor line and you have closed the top cover, the printer will restart automatically.Do not apply any pressure on the print head because this pressure may cause damage on the interconnections underneath the print head.If residue dust is present as part of the cleaning procedure it will disappear after a few prints.
652831A EN 20041201 Advanced operation (key-operator mode)DRYSTAR 5302 USER MANUALTroubleshooting checklistThe table below lists some general problems which can occur when workingwith the Drystar 5302.•The Drystar 5302 does not print.•The quality of the printed images is bad (printing remains possible).Refer to the appropriate pages of the Drystar 5302 Reference manual.Action Refer toPage (Ref. Man.)Check the Drystar 5302 ‘Checking the status indicator LED’ 157‘Checking the connections’ 158‘Checking the print queue’ 160Remove a jammed film‘Film input tray jams’ 162‘Film transport jams (clearing from the top)’ 165‘Unauthorized opening of the printer’ 167Resolve error messages ‘Checking error messages’ 159Handle CF-card errors ‘Checking CF-card error messages’ 159Resolve film identificationproblems‘The Film Identification tag is not readable’ 168Action Refer toPage (Ref. Man.)Resolve film quality problems ‘Persistent white dots or lines appear in the transport direction’ 175Resolve warning messages ‘Maintenance messages’ 177Have electrical or mechanical defects repaired by skilled personnel only!
66 2831A EN 20041201Advanced operation (key-operator mode)DRYSTAR 5302 USER MANUAL
Equipment informationsheetAppendix A
68 2831A EN 20041201Equipment information sheetDRYSTAR 5302 USER MANUALSpecificationsProduct descriptionType of product  PrinterCommercial name Drystar 5302Original seller/manufacturer Agfa-Gevaert N.V.LabellingTÜV-, cULus-Certification Mark, CE-markingCCC MarkDimensionsDimensions (approx. values in cm)•Unpacked: widthtbd, lengthtbd, height tbd• Packed: width tbd, length tbd, height tbdWeight • Unpacked: approx. 90 kg• Packed: approx. 120 kgRAM memory 512 MbMass storage media (internal/external) Compact Flash Type IIElectrical connectionOperating voltage 100-127 V; 220-240 V ACNo external mains fusesMains frequency 50/60 Hz
692831A EN 20041201 Equipment information sheetDRYSTAR 5302 USER MANUALNetwork connectivityEthernet / connectors RJ45 twisted pair for 10/100Base-TX; Serial RS232 connectionNetwork protocols (TCP/IP services) FTP, Telnet, HTTP, SMTPImage formats DICOM (Default)TIFFPostscript Not availablePower consumption - heat dissipationDuring operation 250 W - 900 kJ/hIn standby 70 W - 252 kJ/hPeak power (absolute max. rating) 530 W - 1908 kJ/hProtection againstElectrical shocks Class 1 (grounded)Ingress of water IPXØEnvironmental conditions (operation)Room temperature Between +15°C and +30°CRelative humidity Between 20% and 75%Note: Films may not become wet!Atmospheric pressure 70 kPa - 106 kPaEnvironmental storage conditionsClimate conditions for storage are in accordance with EN60721-3-1-class 1K4.Room temperature Between -25°C and 55°C (storage) Relative humidity Between 10% and 100%Absolute humidity Between 0.1 g/m3 and 35 g/m3Rate of change of temperature 1°C/minAtmospheric pressure 70 kPa - 106 kPa
70 2831A EN 20041201Equipment information sheetDRYSTAR 5302 USER MANUALEnvironmental transport conditionsClimate conditions for transport are in accordance with EN60721-3-2-class 2K4.Temperature Between -40°C and 70°C (transport)Relative humidity not combined with rapid temperature changes 95% at +45°CNoise emission (method of measurement in accordance with DIN 45635 part 19)During operation Max. 64 dBAIn standby Max. 54 dBAConsumablesDrystar DT 2B and Drystar DT 2C 8x10”, 10x12”, 11x14”, 14x14” and 14x17”film sizesPrint technologyDirect thermal printingReliabilityEstimated product life(if regularly serviced and maintained according to Agfa instructions)> 5 years and > 125,000 filmsService interventions Max. 2 interventions / 3 yearsEarthquake (standard) Meets the CA requirements
712831A EN 20041201 Equipment information sheetDRYSTAR 5302 USER MANUALImaging Array - Diagnostic areaFilm size 8x10”8” dimensions in pixels8” dimensions in mm10” dimensions in pixels10” dimensions in mmDiagnostic area 2375.76 188.64 3072.24 243.94Film size 10x12”10” dimensions in pixels10” dimensions in mm12” dimensions in pixels12” dimensions in mmDiagnostic area 3072.24 243.94 3652.84 290.04Film size 11x14”11” dimensions in pixels11” dimensions in mm14” dimensions in pixels14” dimensions in mmDiagnostic area 3348.06 265.84 4358.13 346.04Film size 14x14”14” dimensions in pixels14” dimensions in mm14” dimensions in pixels14” dimensions in mmDiagnostic area 4358.13 346.04 4302.72 341.64Film size 14x17”14” dimensions in pixels14” dimensions in mm17” dimensions in pixels17” dimensions in mmDiagnostic area 4358.13 346.04 5232.19 415.44
72 2831A EN 20041201Equipment information sheetDRYSTAR 5302 USER MANUALViewing the System info area on a filmOn the top right corner of each film, a “System info” area will be printed.This info can only be read using a magnifying glass.The System info area contains info about:•Printer: (serial number, densitometer info, film counts, software version, etc.),•Controller (image source, date, time, etc.).For more detailed information, refer to the Drystar 5302 Servicedocumentation.FilmDiagnostic areaClear borderBlack borderSystem info areaTransport direction
732831A EN 20041201 Equipment information sheetDRYSTAR 5302 USER MANUALOptions and accessoriesMobile / Earthquake provisionsHardwareThe OPTIONAL mobile/earthquake installation kit allows you to use theDrystar 5302 in a van, or to use it in unstable environment.It contains the necessary equipment to fix the printer onto a table and hasprovisions for easy service access.The mobile/earthquake installation kit is delivered with the necessarymounting instructions.SoftwareNo additional software for mobile/earthquake use is required.ABC codeABC code: tbd
74 2831A EN 20041201Equipment information sheetDRYSTAR 5302 USER MANUALConnectivityConnectivity with Agfa equipment•Connected via VIPS or CR QS•ADC Compact•ADC Compact Plus•ADC Solo•CR 25.0•CR 75.0•ADR Thorax•Impax•MG3000•Paxport•MULTIFLEXFor more information, contact your Agfa representative.Connectivity with non-Agfa equipmentDrystar 5302 is a Dicom printer and can therefore be connected to allmodalities supporting Dicom. Although, to ensure optimal operation andimage quality, Agfa has made the effort to test and release the Drystar 5302with most of modalities on the market. For the complete list or if you want tocheck on a specific modality, contact your Agfa representative.
Quality Control ChartsAppendix B
76 2831A EN 20041201Quality Control ChartsDRYSTAR 5302 USER MANUALDrystar 5302: Determination of Operating Levels Chart 1Imager Type: __________ Serial #: _________________ Date ____________________Film Type: ____________ Emulsion #: ______________Densitometer Internal: _________________ (default selection)External: Type ____________ Serial #: _________________Day 1 Day 2 Day 3 Day 4 Day 5MonthDayInitialsLow DensityAverage of 5 Values = operating (aim) level “Low Density”Mid Density  Average of 5 Values = operating (aim) level "Mid Density"High DensityAverage of 5 Values = operating (aim) level “High Density”Step 1: Print QC Test images on five consecutive days. Record the optical densities measurements in the tables below. After five days, average the values to determine the operating (aim) levels for each of the parameters. Step 2: Copy the operating (aim) levels to Charts 2A/B ('Daily Density Control Chart')
772831A EN 20041201 Quality Control ChartsDRYSTAR 5302 USER MANUALDrystar 5302 Daily Density   Control Chart  Imager Type: __________ Serial #: _____________ Film Type:____________ Emul #:___________ Densitometer  Internal: ____________________  (default selection)  External: Type _______________  Serial #:________________ Date:Initia ls:Low DensityMid DensityUpper Control limit =+0.10Low Density AimLower Control Limit =-0.10Upper Control limit =+0.20Mid DensityAim(High - Low)Lower Control Limit =-0.20Chart 2A
78 2831A EN 20041201Quality Control ChartsDRYSTAR 5302 USER MANUALDrystar 5302 Daily Density   Control Chart  Imager Type: __________ Serial #: _____________ Film Type:____________ Emul #:___________        High Density Upper Control limit =+0.20 Lower Control Limit =-0.20High Density AimChart 2BDensitometer Internal: ____________________  (default selection)External: Type _______________  Serial #:________________
792831A EN 20041201 Quality Control ChartsDRYSTAR 5302 USER MANUALChart 3Drystar 5302 Artifacts and Spatial ResolutionControl Chart Test Frequency:     Weekly Drystar 5302 Serial # _____________Initial Reference Test DateInitial Reference ArtifactsInitial Reference Dot Visibility Initial Reference Low Contrast Month Day Artifacts Visibility of all DotsMonth Day Artifacts Visibility of all DotsMonth Day Artifacts Visibility of all DotsMonth Day Artifacts Visibility of all DotsMonth Day Artifacts Visibility of all DotsLow Contrast VisibilityLow Contrast VisibilityLow Contrast VisibilityLow Contrast VisibilityLow Contrast Visibility
80 2831A EN 20041201Quality Control ChartsDRYSTAR 5302 USER MANUALChart 4Drystar 5302 Geometric ConsistencyControl ChartTest Frequency: Annually or as required Drystar 5302 Serial # ______________Reference DimensionsDate:Measured DimensionsDate:Consistency Aspect Ratio Aref A: A/Aref A/BBref B: B/BrefReference DimensionsDate:Measured DimensionsDate:Consistency Aspect Ratio Aref A: A/Aref A/BBref B: B/Bref
812831A EN 20041201 Quality Control ChartsDRYSTAR 5302 USER MANUAL