Agilent Technologies 5185 5991 Users Manual 4&15mL Concentrators Instructions 1

5185-5991 50748603-c0c4-4b76-9d3e-a225439d04ae

5185-5991 to the manual 50748603-c0c4-4b76-9d3e-a225439d04ae

2015-02-02

: Agilent-Technologies Agilent-Technologies-5185-5991-Users-Manual-413454 agilent-technologies-5185-5991-users-manual-413454 agilent-technologies pdf

Open the PDF directly: View PDF PDF.
Page Count: 2

DownloadAgilent-Technologies Agilent-Technologies-5185-5991-Users-Manual 4&15mL Concentrators Instructions 1
Open PDF In BrowserView PDF
5185-5991 5 KDa MWCO 4 mL
5188-5201 30 KDa MWCO 4 mL
5188-5202 50 KDa MWCO 4 mL

5188-2798 5 KDa MWCO 15 mL
5188-2799 30 KDa MWCO 15 mL
5188-2800 50 KDa MWCO 15 mL

Agilent Spin Concentrators for Proteins
Instructions – 4 mL Concentrators

5,000 Molecular Weight Cut-Off (MWCO)
4 mL Spin Concentrator, Part No. 5185-5991

Equipment required
4 mL Spin Concentrators, pack of 25
Centrifuge
Rotor type
Swing bucket or Fixed angle
Minimum rotor angle
25°
Rotor cavity
To fit 15 mL (17 mm) conical bottom tubes
Minimum effective speed
2,000 g
Recommended speed
5,000 g
Maximum speed
7,500 g
Concentrate recovery
Pipette type
Fixed or variable volume
Recommended tip
Thin gel loader type
Technical specifications
Concentrator sample volume capacity
Swing bucket rotor
1-4 mL
Fixed angle rotor
1-4 mL
Dimensions
Total Length
122 mm
Width
17 mm
Active Membrane Area
2.0 cm2
Hold-up volume of membrane
<10 µL
Dead stop volume
20 µL
pH range
pH 1 – 9
Materials of construction
Body
Polycarbonate
Filtrate vessel
Polypropylene
Concentrator cap
Polycarbonate
Membrane
Polyethersulfone

Concentrating Proteins with 4 mL Concentrators:
1 Fill concentrator with sample within the working volume
capacity range noted in the table above (ensure that the hinged
lid is fully seated).
2 Insert assembled concentrator into centrifuge (when fixed
angle rotors are used, angle the concentrators so that the
printed windows face upwards/outwards).
3 Centrifuge at recommended speeds, taking care not to exceed
the maximum g force indicated in the table above (20 minutes
at 7,500 g is recommended for 30´ concentration of 4 mL at
10°C).
4 Once the desired concentration is achieved, remove assembly
and recover sample from the bottom of the concentrate pocket
with a pipette.

Desalting/Buffer Exchange
1 Concentrate sample to desired level or at least
5´.
2 Empty filtrate container.
3 Refill concentrator with an appropriate buffer.
4 Concentrate the sample again and repeat the
process until the concentration of the
contaminating species is sufficiently reduced.
Typically three wash cycles will remove 99% of
initial salt content.
See Notes on next page.

Instructions – 15 mL Concentrators
Equipment required
Centrifuge
Rotor type
Minimum rotor angle
Rotor cavity

30,000 Molecular Weight Cut-Off (MWCO)
15 mL Spin Concentrator, Part No. 5188-2799

15 mL Spin Concentrators, pack of 10
Swing bucket or Fixed angle
25°
To fit 30 x 114 mm conical bottom tubes or
29 x 104 mm round bottom tubes (50mL)
1,000 g
2,000 g
3,000 g

Minimum effective speed
Recommended speed
Maximum speed
Concentrate recovery
Pipette type
Fixed or variable volume
Recommended tip
Thin gel loader type
Technical specifications
Concentrator sample volume capacity
Swing bucket rotor
2-15 mL
Fixed angle rotor
2-8 mL
Dimensions
Total Length
76 mm
Width
25.5 mm
Active Membrane Area
4 cm2
Hold-up volume of membrane
<20 µL
Dead stop volume
50 µL
pH range
pH 1 – 9
Materials of construction
Body
Polycarbonate
Filtrate vessel
Polypropylene
Concentrator cap
Polycarbonate
Membrane
Polyethersulfone

Concentrating Proteins with 15 mL Concentrators:
1 Fill concentrator with sample within the working volume
capacity range noted in the table above, taking care not to
touch the membrane surface, and then close the hinged lid
(ensure that the lid is fully seated).
2 Place concentrator into 50 mL centrifuge tube (filtrate
collection vessel – not supplied, user must provide).
3 Insert assembled concentrator into centrifuge (when fixed
angle rotors are used, angle the concentrators so that the
printed windows face upwards/outwards). Spin at
recommended speeds, taking care not to exceed the maximum
g force indicated in the table above (20-40 minutes is
common).
4 Once the desired concentration is achieved, remove assembly
and recover sample from the bottom of the concentrate pocket
with a pipette.
Desalting/Buffer Exchange
See 4 mL Spin Concentrator instructions on the reverse side.
For more information or technical assistance,
please call toll free: 1-800-227-9770 or visit
our web site at: www.agilent.com/chem/bioreagents

Notes:
· Spin concentrators are single use disposable
items. Re-use is not recommended.
· For research use only.
· Select the most appropriate membrane cut-off for
your sample. For maximum recovery, select a
MWCO at least 50% smaller than the molecular
size of the species of interest.
· These concentrators are designed for use with
biological fluids or aqueous solutions only, not
for organic solvents or strong acids or bases.
· Flow rates are affected by MWCO, porosity,
sample size, concentration, viscosity, centrifuge
force, and temperature. Adjust centrifuge force
and time accordingly for your samples.
· Do not autoclave. Sterilize with a 70% ethanol
solution.

Agilent Technologies, Inc.
2850 Centerville Road
Wilmington, DE 19808-1610 USA

05/2004A



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.3
Linearized                      : Yes
Create Date                     : 2004:05:12 08:42:39-04:00
Modify Date                     : 2004:05:12 08:42:39-04:00
Creator                         : AdobePS5.dll Version 5.2
Page Count                      : 2
Creation Date                   : 2004:05:12 12:42:39Z
Mod Date                        : 2004:05:12 12:42:39Z
Producer                        : Acrobat Distiller 5.0.5 (Windows)
Metadata Date                   : 2004:05:12 12:42:39Z
Title                           : Microsoft Word - 4&15mL concentrators instructions 1.doc
EXIF Metadata provided by EXIF.tools

Navigation menu