Alcon Research 212-1 Constellation Vision System User Manual NGVS Book 1 indb

Alcon Research Ltd. Constellation Vision System NGVS Book 1 indb

User Manual part 4

2.76 8065751025
Constellation® Vision System
is constant regardless of the vacuum level. In Linear the limit varies depending on
the vacuum level.
Phaco Pulse Off Time Up/Down Buttons - Sets the Phaco pulse off time.
Phaco Pulse Off Time Progress Bar and Numeric Value - Shows the current Phaco
pulse off time.
Phaco Pulse Off Time Limit Control Drop Down List - Determines whether
Figure 2-58 Surgery Screen: Phaco Mode - Pulsed Submode
Aspiration Panel
Vacuum Panel
Configuration Button
the pulse off time pulse type limit is decreasing, fixed, or increasing. In Fixed
the limit is constant regardless of the vacuum level. In Linear the limit varies
depending on the vacuum level.
Phaco Pulse Configuration Button - Displays the configuration dialog. Simplifies
large value changes by providing slider controls.
Phaco Mode: PULSED SUBMODE
In the Phaco Custom submode fixed U/S power (Power %) can be adjusted for Rate
8065751025 2.77
Constellation® Vision System
(pulses per second-pps) and Time On (%) U/S power pulses. The pulse is controlled
with the footpedal from detent two to full treadle depression. The actual percent of
time U/S power is activated is reflected in the Actual window.
The controls for each panel in Pulsed submode have the following functions:
Aspiration Panel:
Same as Burst submode.
Vacuum Panel:
Same as Burst submode.
Ultrasound Panel:
Power End Up/Down Buttons - Sets the power at full treadle.
Power Progress Bar and Numerical Value - Shows the current power value.
Phaco Power Configuration Button - Displays the configuration dialog.
Phaco Pulse Rate Up/Down Buttons - Set the Phaco pulse rate in pulses per
second (pps).
The following controls are available only in the Advanced view:
Smart Pulse - Indicator for smart pulse. Becomes visible when power is greater
than 0 and pulse on time is less than 20ms.
Power Limit Control Type Drop List Determines whether the power limit is
fixed or linear. In Fixed, the limit is constant regardless of the vacuum level.
In Linear, the limit varies depending on the vacuum level.
Phaco Pulse Time On Up/Down Buttons - Sets the Phaco time on percentage.
Phaco Pulse Configuration Button - Displays the configuration popup.
Phaco Mode: CONTINUOUS SUBMODE
In the Phaco Continuous submode provides continuous fixed U/S power
(Power %). Phaco power is controlled with the footpedal from detent two to full
treadle depression. The actual percent of time U/S power is activated is reflected in
the Actual window.
The controls for each panel in Continuous submode have the following functions:
Aspiration Panel:
Same as Burst submode.
Vacuum Panel:
Same as Burst submode.
2.78 8065751025
Constellation® Vision System
Figure 2-59 Surgery Screen: Phaco Mode - Continuous Submode
Aspiration Panel
Vacuum Panel
Configuration Button
8065751025 2.79
Constellation® Vision System
Ultrasound Panel:
Phaco Power End Up/Down Buttons - Sets the power at full treadle.
Phaco Power Progress Bar and Numerical Value - Shows the current power value.
Ultrasound Configuration Button - Displays the configuration popup.
The following controls are available only in the Advanced view:
Smart Pulse - Indicator for smart pulse. Becomes visible when power is greater
than 0 and pulse on time is less than 20ms.
Phaco Power Limit Control Drop DownList - Determines whether the power limit
Figure 2-60 Surgery Screen: Fragmentation Mode - Fixed Submode
Aspiration Panel
Vacuum Panel
Configuration Button
is fixed or linear. In Fixed, the limit is constant regardless of the vacuum level. In
Linear, the limit varies depending on the vacuum level.
Fragmentation Mode
Fragmentation mode is enabled when a fragmentation handpiece is connected to the
system and the Fragmentation Surgery step is pressed. Three submodes are available, each
with its own default U/S power and vacuum limits. U/S power is enabled and disabled
using the
footswitch or by pressing the power setting on the touch screen (except in
momentary mode). Micro and proportional reflux are available via the footswitch.
U/S Pulse controls are available in all submodes.
Fragmentation Mode: FIXED SUBMODE
In Fixed submode, the vacuum level is controlled linearly from 0 to the programmed
value whereas the ultrasound power level is fixed. The controls for each panel in
Fixed submode have the following functions:
Vacuum Panel:
2.80 8065751025
Constellation® Vision System
Vacuum End Up/Down Buttons - Sets the vacuum level at full treadle.
Vacuum Progress Bar and Numerical Value - Displays the current vacuum level
Vacuum Configuration Button - Displays the configuration popup. Simplifies
large value changes by providing slider controls. Also provides panel specific
configuration capabilities such as Drainbag Change.
The following controls are available only in the Advanced view:
Flow Limit On/Off Button - Turns flow limit on or off. When the maximum rate is
reached, the vacuum level can no longer be increased. If the Flow Limit is turned
off, the vacuum level is limited only by the Start and End values.
Flow Limit Up/Down Buttons - Sets the maximum flow rate.
Flow Limit Progress Bar and Numerical Value - Displays the current flow rate.
Flow / Vacuum Toggle - Switches between Flow Mode and Vacuum Mode.
Flow Limit Configuration Button - Displays the configuration popup.
Reflux Label - Read Only display of the current reflux configuration together with
its current value.
Reflux Configuration Button - Displays the Reflux configuration popup.
Aspiration Panel:
Flow End Up/Down Buttons - Sets the flow level at full treadle.
Flow Progress Bar and Numerical Value - Displays the current flow level.
Aspiration Configuration Button - Displays the configuration popup.
Vacuum Limit On/Off Button - Turns vacuum limit on or off. When the maximum
rate is reached, the flow can no longer be increased. If the Vacuum Limit is turned
off, the flow is limited only by the Start and End values.
The following controls are available only in the Advanced view:
Vacuum Limit Up/Down Buttons - Sets the maximum vacuum value.
Vacuum Limit Progress Bar and Numerical Value - Displays the current vacuum
level.
Vacuum Limit Configuration Button - Displays the configuration popup.
Flow / Vacuum Toggle - Switches between Flow Mode and Vacuum Mode.
Ultrasound Panel:
Power On/Off Button - Turns power on or off.
8065751025 2.81
Constellation® Vision System
Power End Up/Down Buttons - Sets the power at full treadle. The Start value
tracks with the End value.
Power Progress Bar and Numerical Value - Displays the current power value.
Ultrasound Configuration Button - Displays the configuration popup.
The following controls are available only in the Advanced view:
Pulse Rate On/Off Button - Turns the Pulse Rate on or off.
Pulse Rate Up/Down Buttons - Sets the pulse rate.
Pulse Time On Up/Down Buttons - Sets the time on percentage.
Pulse Configuration Button - Displays the configuration popup.
Fragmentation Mode: LINEAR SUBMODE
Linear submode is very similar to the Fixed submode with the exception that the
treadle start values for both vacuum and ultrasound power are fixed at 0.
Figure 2-61 Surgery Screen: Fragmentation Mode - Linear Submode
Aspiration Panel
Vacuum Panel
Configuration Button
2.82 8065751025
Constellation® Vision System
Vacuum Panel:
Same as Fixed submode.
Flow Panel:
Same as Fixed submode.
Ultrasound Panel:
Same as Fixed submode except that the Start value is always zero instead of tracking
with the End value.
Fragmentation Mode: MOMENTARY SUBMODE
Figure 2-62 Surgery Screen: Fragmentation Mode - Momentary Submode
Aspiration Panel
Vacuum Panel
Configuration Button
8065751025 2.83
Constellation® Vision System
Momentary submode is similar to the Fixed Sub Mode with the exception that
ultrasound is only active when the appropriate footswitch button is pressed.
Vacuum Panel:
Same as Fixed submode.
Flow Panel:
Figure 2-63 Surgery Screen: Irrigation/Aspiration Mode
Same as Fixed submode.
Ultrasound Panel:
Same as Fixed submode except that there is no On/Off button to turn U/S power Off.
Irrigation/Aspiration (IA) Mode
In IA mode, the Surgery screen contains panels for controlling an IA handpiece.
The vacuum pressure is always controlled linearly and vacuum cannot be turned
off. When the system detects an occlusion, the word "Occlusion" is displayed on
the extraction panel (in either vacuum mode or flow mode) under the primary limit
control’s label and an alarm sounds.
In “Standard” display mode, only the Flow mode controls are available. The controls
for each panel in IA mode have the following functions:
Vacuum Panel:
Vacuum End Up/Down Buttons - Sets the vacuum level at full treadle.
Vacuum Progress Bar and Numerical Value - Displays the current vacuum level
Vacuum Configuration Button - Displays the configuration popup. Simplifies
large value changes by providing slider controls. Also provides panel specific
configuration capabilities such as Drainbag Change.
2.84 8065751025
Constellation® Vision System
The following controls are available only in the Advanced view:
Flow Limit On/Off Button - Turns flow limit on or off. When the maximum rate is
reached, the vacuum level can no longer be increased. If the Flow Limit is turned
off, the vacuum level is limited only by the Start and End values.
Flow Limit Up/Down Buttons - Sets the maximum flow rate.
Flow Limit Progress Bar and Numerical Value - Displays the current flow rate.
Flow / Vacuum Toggle - Switches between Flow Mode and Vacuum Mode.
Flow Limit Configuration Button - Displays the configuration popup.
Flow Limit Control Drop Down List - Determines whether the flow limit is fixed
or linear. In Fixed, the limit is constant regardless of the flow level. In Linear, the
limit varies depending on the flow level.
Reflux Label - Read Only display of the current reflux configuration together with
its current value.
Reflux Configuration Button - Displays the Reflux configuration popup.
Aspiration Panel:
Flow End Up/Down Buttons - Sets the flow level at full treadle.
Flow Progress Bar and Numerical Value - Displays the current flow level.
Aspiration Configuration Button - Displays the configuration popup.
Vacuum Limit On/Off Button - Turns vacuum limit on or off. When the maximum
8065751025 2.85
Constellation® Vision System
rate is reached, the flow can no longer be increased. If the Vacuum Limit is turned
off, the flow is limited only by the Start and End values.
The following controls are available only in the Advanced view:
Vacuum Limit Up/Down Buttons - Sets the maximum vacuum value.
Figure 2-64 Surgery Screen: Extrusion Mode
Vacuum Limit Progress Bar and Numerical Value - Displays the current vacuum
level.
Vacuum Limit Configuration Button - Displays the configuration popup.
Flow / Vacuum Toggle - Switches between Flow Mode and Vacuum Mode.
Extrusion Mode
In Extrusion Mode the Surgery screen contains the Vacuum panel necessary for
controlling an extrusion handpiece. The vacuum pressure is always controlled linearly
and vacuum cannot be turned off.
In “Standard” display mode, only the Vacuum mode controls are available. The
controls for each panel in Extrusion mode have the following functions:
Vacuum Panel:
Vacuum End Up/Down Buttons - Sets the vacuum level at full treadle. The Start
value is always zero.
Vacuum Progress Bar and Numerical Value - Displays the current vacuum level
Vacuum Configuration Button - Displays the configuration popup. Simplifies
large value changes by providing slider controls. Also provides panel specific
configuration capabilities such as Drainbag Change.
The following controls are available only in the Advanced view:
Flow Limit On/Off Button - Turns flow limit on or off. When the maximum rate is
reached, the vacuum level can no longer be increased. If the Flow Limit is turned
2.86 8065751025
Constellation® Vision System
off, the vacuum level is limited only by the Start and End values.
Flow Limit Up/Down Buttons - Sets the maximum flow rate.
Flow Limit Progress Bar and Numerical Value - Displays the current flow rate.
Flow / Vacuum Toggle - Switches between Flow Mode and Vacuum Mode.
Flow Limit Configuration Button - Displays the configuration popup.
Flow Limit Control Drop Down List - Determines whether the flow limit is fixed
or linear. In Fixed, the limit is constant regardless of the flow level. In Linear, the
limit varies depending on the flow level.
Reflux Label - Read Only display of the current reflux configuration together with
its current value.
Reflux Configuration Button - Displays the Reflux configuration popup.
Aspiration Panel:
Flow End Up/Down Buttons - Sets the flow level at full treadle.
Flow Progress Bar and Numerical Value - Displays the current flow level.
Flow Configuration Button - Displays the configuration popup.
The following controls are available only in the Advanced view:
Vacuum Limit On/Off Button - Turns vacuum limit on or off. When the maximum
rate is reached, the flow can no longer be increased. If the Vacuum Limit is turned
off, the flow is limited only by the Start and End values.
Vacuum Limit Up/Down Buttons - Sets the maximum vacuum value.
Vacuum Limit Progress Bar and Numerical Value - Displays the current vacuum
level.
Vacuum Limit Configuration Button - Displays the configuration popup.
8065751025 2.87
Constellation® Vision System
Flow / Vacuum Toggle - Switches between Flow Mode and Vacuum Mode.
Reflux Label - Read Only display of the current reflux configuration together with
its current value.
Reflux Configuration Button - Displays the Reflux configuration popup.
Cutting Panel:
Cut Rate End Up/Down Buttons - Sets the cut rate.
Cut Rate Progress Bar and Numerical Value - Displays the current cut rate.
Cutting Configuration Button - Displays the configuration dialog.
Laser Mode
In Laser Mode the Surgery screen contains panels for controlling a laser probe as well
as an aspirating handpiece (which could be the laser probe). The laser function has
Figure 2-65 Surgery Screen: Laser Mode
two modes: Ready and Firing.
Laser Modes:
Ready Mode - Laser is ready to fire.
Standby – Laser cannot be fired.
Treatment Mode:
Repeat – Generates laser shots continuously when the footswitch treadle is down
Single Shot – When the footswitch treadle is pressed down, the laser fires a single
shot.
Continuous – The laser keeps firing as long as the footswitch treadle is down.
2.88 8065751025
Constellation® Vision System
In “Standard” display mode, only the Vacuum mode controls are available. The
controls for each panel in Laser mode have the following functions:
Vacuum Panel:
Vacuum End Up/Down Buttons - Sets the vacuum level at full treadle. The Start
value is always zero.
Vacuum Progress Bar and Numerical Value - Displays the current vacuum level
Vacuum Configuration Button - Displays the configuration popup. Simplifies
large value changes by providing slider controls. Also provides panel specific
configuration capabilities such as Drainbag Change.
The following controls are available only in the Advanced view:
Flow Limit On/Off Button - Turns flow limit on or off. When the maximum rate is
reached, the vacuum level can no longer be increased. If the Flow Limit is turned
off, the vacuum level is limited only by the Start and End values.
Flow Limit Up/Down Buttons - Sets the maximum flow rate.
Flow Limit Progress Bar and Numerical Value - Displays the current flow rate.
Flow / Vacuum Toggle - Switches between Flow Mode and Vacuum Mode.
Flow Limit Configuration Button - Displays the configuration popup.
Flow Limit Control Drop Down List - Determines whether the flow limit is fixed
or linear. In Fixed, the limit is constant regardless of the flow level. In Linear, the
limit varies depending on the flow level.
Reflux Label - Read Only display of the current reflux configuration together with
its current value.
Reflux Configuration Button - Displays the Reflux configuration popup.
Aspiration Panel:
Flow End Up/Down Buttons - Sets the flow level at full treadle.
Flow Progress Bar and Numerical Value - Displays the current flow level.
Flow Configuration Button - Displays the configuration popup.
The following controls are available only in the Advanced view:
Vacuum Limit On/Off Button - Turns vacuum limit on or off. When the maximum
rate is reached, the flow can no longer be increased. If the Vacuum Limit is turned
off, the flow is limited only by the Start and End values.
Vacuum Limit Up/Down Buttons - Sets the maximum vacuum value.
Vacuum Limit Progress Bar and Numerical Value - Displays the current vacuum
level.
Vacuum Limit Configuration Button - Displays the configuration popup.
Flow / Vacuum Toggle - Switches between Flow Mode and Vacuum Mode.
Reflux Label - Read Only display of the current reflux configuration together with
its current value.
Reflux Configuration Button - Displays the Reflux configuration popup.
Laser Panel:
Laser Mode - Selects laser Standby or Ready mode.
Port Selection - Selects the port used for treatment.
Treatment Mode - Selects the Treatment mode: Repeat, Single Shot, or
Continuous
Power Up/Down Buttons - Adjusts the laser power level.
8065751025 2.89
Constellation® Vision System
Exposure Time Up/Down Buttons - Adjust the laser exposure time. Only available
in Repeat or Single Shot modes.
Interpulse Time Up/Down Buttons - Adjusts the laser interpulse time. Only
available in Repeat mode.
Aiming Beam Up/Down Buttons - Displays the current relative intensity of the
aiming beam.
LIO Illumination Intensity Up/Down Buttons - Adjusts the LIO illumination
intensity.
Shot Count Label - Displays the current shot count. Also allows for the shot
count to be reset.
Total Energy Label - Displays the total energy. Also allows for the total energy to
be reset.
Laser Configuration Button - Displays the configuration dialog. Simplifies large
value changes by providing slider controls.
NOTE: In tethered mode, all front panel controls except for the emergency
switch are disabled.
Forceps Mode
The Forceps Pressure Panel controls the pressure for opening and closing pneumatic
forceps. Figure 2-66 shows the Forceps surgery screen. The controls for each panel in
Forceps mode have the following functions:
Forceps Panel:
Pressure Start Up/Down Buttons - Sets the pressure at treadle start.
Pressure End Up/Down Buttons - Sets the pressure at full treadle.
Pressure Progress Bar and Numerical Value - Displays the current pressure value.
Forceps Configuration Button - Displays the configuration popup.
Figure 2-66 Surgery Screen: Forceps Mode
2.90 8065751025
Constellation® Vision System
Figure 2-67 Surgery Screen: Scissors Mode - MultiCut
Calibrate On/Off Button - Enters Calibration Mode allowing the user select and
set the pressure at Start and Close.
- When Start is selected, the Start pressure is applied to the port and the Start Up/
Down buttons are active on the Pressure control to allow adjustment of the Start
pressure.
- When Close is selected, the Close pressure is applied to the port and the Close
Up/Down buttons are active on the Pressure control to allow adjustment of the
Close pressure.
- Pressing the "Save Calibration Values" button saves these values.
Scissors Mode
The Scissors mode provides cutting capability using pneumatically powered
scissors controlled by the footswitch. The pneumatically powered scissors uses a
microscissors snapped onto the Alcon/Grieshaber pneumatic handpiece, which is
connected via tubing to a front panel pneumatic port. Two submodes are available:
MultiCut and Prop (Proportional). In Scissors Mode the surgery screen contains
panels for controlling scissors and bimanual forceps.
Scissors Mode: MULTICUT SUBMODE
In the MultiCut submode, pressing the footpedal activates the scissors at a Cut Rate
proportional to the footpedal position up to the preset End limit, set by pressing the
up/down
buttons next to the End limit readout. Single cuts are possible by adjusting
the cut rate to 1 cpm and momentarily pressing the footpedal.
The controls for each panel in Scissors MultiCut submode have the following
functions:
Cutting Panel:
Cut Rate End Up/Down Buttons - Sets the cut rate.
8065751025 2.91
Constellation® Vision System
Cut Rate Progress Bar and Numerical Value - Displays the current cut rate.
Cutting Configuration Button - Displays the configuration popup.
Forceps Panel:
Bimanual On/Off Button - Turning Bimanual mode on allows the control of the
pneumatic forceps while the system is in Scissors mode. All controls function the
same as when the system is in Forceps mode (see Forceps mode for a description).
Figure 2-68 Surgery Screen: Scissors Mode - Proportional
Scissors Mode: PROP SUBMODE
The Proportional submode provides proportional control of the opening and closing
of the scissor blades, dependent upon the amount of footpedal depression. The
amount of scissors closure is indicated in the Pressure panel. When the footpedal is
fully depressed, the Close Limit is reached, and scissors should be fully closed.
Pressing the On/Off button in the Calibration panel displays the Calibration panel that
allows the user to adjust the scissors Start and Close Pressures.
The controls for each panel in Scissors Proportional submode have the following
functions:
Cutting Panel:
Pressure Start Up/Down Buttons - Sets the pressure at treadle start.
Pressure End Up/Down Buttons - Sets the pressure at full treadle.
Pressure Progress Bar and Numerical Value - Displays the current cut rate.
Cutting Configuration Button - Displays the configuration popup.
Calibrate On/Off Button - Enters Calibration Mode allowing the user select and
2.92 8065751025
Constellation® Vision System
set the pressure at Start and Close.
- When Start is selected, the Start pressure is applied to the port and the Start Up/
Down buttons are active on the Pressure control to allow adjustment of the Start
pressure.
- When Close is selected, the Close pressure is applied to the port and the Close
Up/Down buttons are active on the Pressure control to allow adjustment of the
Close pressure.
- Pressing the "Save Calibration Values" button saves these values.
Forceps Panel:
Bimanual On/Off Button - Turning Bimanual mode on allows the control of the
pneumatic forceps while the system is in Scissors mode. All controls except for
the footswitch function the same as when the system is in Forceps mode (see
Forceps mode for a description).
Viscous Fluid Control (VFC) Mode
The VFC (Viscous Fluid Control) mode provides pressure at the front panel VFC
connector for fluid injection (i.e., silicone oil), or vacuum for extraction, through the
VFC tubing set to a syringe. Using vacuum, the VFC mode also provides a means of
extruding fluid through the VFC consumables.
CAUTION
Always use Alcon-supplied Viscous Fluid Control Kits and follow all Directions
for Use. Do not use VFC without the plunger/stopper supplied with the kit. Do
not aspirate fluids directly into the console; this will cause damage to the console,
increase risk of electrical shock, and void all warranties.
Viscous Fluid Control (VFC) Mode: Extract
Submode
Figure 2-69 Surgery Screen: VFC Mode - Extract Submode
8065751025 2.93
Constellation® Vision System
Vacuum for fluid extraction is provided to the syringe proportional to footpedal
depression. Max Limit vacuum is reached at full depression. The controls on the
Extraction panel have the following functions:
Vacuum End Up/Down Buttons - Sets the vacuum level at full treadle. Start value
is always zero.
Vacuum Progress Bar and Numerical Value - Displays the current vacuum level.
Extraction Configuration Button - Displays the configuration that simplifies large
value changes by providing slider controls.
Viscous Fluid Control (VFC) Mode:
Inject Submode
In this submode injection pressure to the syringe is provided proportional to the
footpedal position up to the Max Limit setting. Vacuum is available in Bimanual mode
by pressing the Bimanual On/Off button.
WARNINGS!
Double check the cannula connected to the syringe for a tight connection. It must
not be allowed to come loose.
Adjust the Max Limit air pressure in accordance with the viscosity of fluids to be
injected.
Vacuum Panel controls:
Bimanual-Vacuum On/Off Button - Turns vacuum on or off. When Bimanual
mode is on, the control displayed is a dual value control.
Vacuum End Up/Down Buttons - Sets the vacuum level at full treadle. The Start
value is always zero.
Vacuum Progress Bar and Numerical Value - Displays the current vacuum level.
Figure 2-70 Surgery Screen: VFC Mode - Injection Submode
2.94 8065751025
Constellation® Vision System
END CASE
The End Case Screen, selected by pressing the End Case tab, displays case related
metrics as shown in Figure 2-71. It is comprised of a Summary panel area and a
tabbed interface area for displaying the various metrics associated with the case. The
screen also provides various controls for renaming the case, printing the End Case
and Laser forms, and customizing the form’s content and layout via the Setup Form
popup.
Figure 2-71 End Case Screen: Anterior Tab
Viewing of the various metrics associated with the case is accomplished by selecting
the associated tabs - Anterior, Posterior, Laser, or Consumables. Pressing the Rename
button above the tabs, displays a keyboard which enables the user to rename the case
from the default name (date of surgery).
The Print and Print Laser Form buttons are used to obtain a print-out of the End
Case Form and Laser Form respectively. The Laser Form is comprised of the End
Case Form’s first header and the Laser metrics table only. Modification of the End
Case Form’s layout and content is accomplished using the Setup Form button. When
pressed, the End Case Form Setup popup is displayed.
End Case Summary Panel
The Summary panel area lists those procedure steps that were used during the case
along with their cumulative times. For a step to be included in the summary list, work
must have been performed using the step. In most cases, work is performed when the
treadle is depressed. Therefore, steps that are defined in the procedure but not actually
used to perform work, are not be listed. This could occur, for example, if the surgeon
used the 1st step of a procedure, selected the 2nd step but did not depress the treadle,
and then proceeded ahead to the 3rd step to perform work. In this scenario, steps 1
and 3 would be included in the summary list, but step 2 would not be included in the
summary since no work was performed with it.
8065751025 2.95
Constellation® Vision System
In computing a step’s cumulative time, it is necessary for work to be done while
the step is selected in order for the current elapsed time to be added to the step’s
cumulative time. Therefore, a user may access a step and interact with the various
touchscreen controls for an extended period of time but, unless the treadle is
depressed, the current elapsed time will be discarded and will not be added to the
step’s cumulative time. If the step’s cumulative time is zero, the step will not be
shown in the step summary list.
Metrics Tabs
The End Case Metrics screens are accessed through the labeled tabs next to the
Summary panel. Tabs are provided for Anterior, Posterior, Laser, and Consumables.
When selected, each tabs displays the associated metrics for that tab as shown by the
Anterior tab in Figure 2-71. The Consumables tab displays all consumable items that
have been detected by the system.
Setup Form
The End Case Form Setup screen shown in Figure 2-72 is displayed by pressing the
Setup Form button on the End Case Screen. The screen shows a picture of the End
Case Form print-out and provides controls for navigating to other pages in the print-
out and for customizing the layout/content of the print-out.
The buttons at the bottom of the Form Setup screen perform the following actions:
Page arrow buttons - Pressing these buttons allows the user to move through each
page of the printout.
Close button - Pressing the Close button causes the system to save the various
layout and content customization settings and return to the End Case screen.
Cancel button - Returns the system back to the End Case screen without saving
changes made to content and layout.
Pressing the Edit Form button allows
the user to customize the layout and
content of the printout.
Figure 2-72 End Case: Setup Form Screen
Edit
Button
2.96 8065751025
Constellation® Vision System
Setup Form: Editing the Form
To modify the form, press the Edit Form button. When pressed, the “smoked glass”
panel is retracted allowing access to the various customization controls. The editing
controls on the dialog provide the following functionality:
Cell navigation buttons including up, down, left, right.
Cell text edit button.
Drop list to permit the specification of a cell data fields.
Cell appearance controls including font, point size, bold/italics, text alignment,
and cell border lines (on/off).
Table insertion/deletion controls
Table row height, insertion, and deletion controls.
Table column width, insertion, and deletion controls.
Setup Form: Editing the Form - Customizing Table Cells
The form is composed of a number of tables which are in turn composed of cells
arranged in rows and columns. Each table cell provides various attributes that can be
customized with respect to their appearance and content displayed. A cell can contain
static text or it can be defined to include a data field whose text is determined at
runtime. An example of a data field would be Surgeon Name.
To customize a cell follow these steps:
Select a table on the form by pressing on it. The table is then shown with
selection handles. The active cell is shown in red.
Select the cell to be customized by using the Up, Down, Left, Right cell
navigation buttons located in the Cell Selection & Edit panel.
To specify static text for a cell, press the Edit button (see Figure 2-72). When
pressed, the keyboard popup is displayed allowing text to be entered by the user.
To specify a cell data field, press the Data Field drop list to display a list of
available data field types including, but not limited to, Surgeon Name, Date,
Procedure Name, etc. After selecting the desired data field, the associated data is
automatically entered into the cell.
Specify the font and point size for a cell’s text using the Font, Font Style, and
Font Size drop list controls. The Font Style drop list control is used to apply bold
and italics formatting to the cell’s text.
Select the desired text alignment by pressing the Left Alignment, Center
Alignment, or Right Alignment toggle buttons. The selected alignment will be
highlighted.
Select the desired cell borders to show or hide by pressing the Bottom, Top, Left,
or Right Border toggle buttons. If a border is on, it will be highlighted.
1.
2.
3.
4.
5.
6.
7.
8065751025 2.97
Constellation® Vision System
Setup Form: Editing the Form - Creating New Tables
To begin the customization process, the user must first select a table on the form by
pressing on it. The table is then shown with selection handles and the active cell is
shown in red.
To insert a new table after the selected table, follow these steps:
In the Table section of the Edit Form panel, press the Insert button (Ins). A table
with two rows and three columns appears below the selected table and the new
table is now the selected table. To delete the table, press the Delete button (Del).
The Table Type is selected from the Type dropdown list in the Table section of the
Edit Form panel. The default table type for new tables is Custom which contains
no entries or formatting (see Customizing Table Cells for information on editing
individual cells).
Select the type of table desired from the Type dropdown list. There are five
preformatted tables available from the list which automatically pull data from the
procedures performed. The Custom selection allows the user to create a custom
table to meet their particular requirements. Preformatted tables may also be
customized according to the users requirements.
To increase the height of the selected row, press the Height% dropdown button
and select an entry between 1 and 20%. The default row height is 2, so selecting 1
will result in a smaller row height and all selections above 2 will result in a larger
row height. The height of the selected row is displayed on the Height% button.
To insert or delete a row, press the Ins (insert) or Del (delete) button in the Row
section of the Edit Form panel.
To increase the width of the selected column, press the Width% dropdown button
and select an entry between 5 and 90%. The default column width is 30, so
selecting 5 through 25 will result in a smaller column width and all selections
above 30 will result in a larger column width. The width of the selected column is
displayed on the Width% button.
To insert or delete a column, press the Ins (insert) or Del (delete) button in the
Column section of the Edit Form panel.
1.
2.
3.
4.
5.
6.
2.98 8065751025
Constellation® Vision System
PROBES AND HANDPIECES
Different probes and handpieces are required for each operating mode of the
Constellation® Vision System. Following is a representative selection of probes and
handpieces with a general description of each. See the Accessories and Parts section
of this manual, or consult your Alcon representative, for a complete selection of all
probes, handpieces, and handpiece tips available.
Vitrectomy Probes
20, 23, and 25 Ga UltraVit Probe
20 and 23 Ga AVIT UltraVit
Pneumatic Scissors
The Pneumatic Hanpiece is a multi-purpose tool that is designed to interface with
a broad range of disposable DSP and reusable Quick Lock connector scissors and
forceps tips to provide footpedal-driven membrane dissection and membrane/tissue
manipulation. The console can drive these in multi-cut and proportional modes.
Bimanual mode enables forceps activation during the first portion of footpedal travel
and scissors activation during the second portion of footpedal travel
Figure 2-73 UltraVit Probes
Figure 2-74 Pneumatic Scissors
8065751025 2.99
Constellation® Vision System
Fragmentation Handpiece
The Fragmentation handpiece is configured to provide simultaneous vacuum and
fragmentation, or vacuum only, depending upon the console setup. The handpiece has
a stainless steel shell for improved reliability and durability. Other than attaching and
removing the needle and aspiration line, no assembly or disassembly is required.
Figure 2-75 Fragmentation Handpiece
Phaco Ultrasound Handpieces
Alcon's phaco handpieces integrate irrigation, aspiration and emulsification. The three
functions of the lens extraction step enable the surgeon to simultaneously maintain or
inflate the anterior chamber, emulsify the lens, and aspirate the lens material from the
eye.
These handpieces require no disassembly other than removal of the disposable tubing,
the ultrasonic tip, and the infusion sleeve with bubble suppression insert.
Figure 2-76 Infiniti™* Ultrasonic (U/S) Handpiece
Infiniti™* Ultrasonic (U/S) Handpiece - This handpiece is used for ultrasonic
applications on the Constellation® Vision System with 1.1 mm TurboSonics® tips
or 0.9 mm TurboSonics® tips, including flared and/or ABS® tips.
CAUTIONS
Do not test or operate U/S handpieces unless the tip is immersed in BSS® sterile
irrigating solution or distilled water or is in surgical use. Irreparable damage to
the handpiece and tip can result if run dry.
Ensure that test chamber is filled with BSS® sterile irrigating solution before tuning
U/S handpieces. Tuning a handpiece dry may result in premature tip failure and
breakage.
2.100 8065751025
Constellation® Vision System
WARNINGS!
Use of an ultrasonic handpiece other than the U/S, or use of a handpiece repaired
without Alcon authorization, is not permitted, and may result in patient injury,
including potential shock hazard to patient and/or operator.
Use of the U/S handpiece in the absence of irrigation flow and/or in the presence of
reduced or lost aspiration flow can cause excessive heating and potential damage
to the cornea and other tissues.
8065751025 2.101
Constellation® Vision System
TurboSonics® Family of Tips
U/S tips are made of medical grade titanium alloy, and are attached to an U/S
handpiece to deliver mechanical energy to the lens, assisting in its removal by
aspiration. Depending on the needs and technique preferred by the surgeon, various
styles of tips and tip bevels are available (see Figure 2-77). Various U/S tip styles are
color coded.
1.1 mm U/S Tips - The standard ultrasonic tips are the original 1.1 mm
TurboSonics® tips. They are designed for use only with 1.1 mm infusion sleeves.
0.9 mm U/S Tips - The 0.9 mm ultrasonic tips are designed to allow entry through
a smaller incision. They are designed for use only with 0.9 mm infusion sleeves.
Aspiration Bypass System - Tips with the ABS® feature contain a small hole in
the distal portion of the tip’s wall. This helps to maintain flow through the system
even during occlusion of the tip’s main port.
WARNINGS!
Use 0.9 mm tips with 0.9 mm infusion sleeves. Use 1.1 mm tips with 1.1 mm
infusion sleeves. Mismatch of consumable components and use of settings not
specially adjusted for a particular combination of consumable components may
create a patient hazard.
Read all package label material printed on the consumable paks prior to their use.
Standard U/S Tips - The 1.1 mm TurboSonics®tip
with the round shaft is the original, classical U/S
tip shape. The 0.9 mm has a smaller diameter shaft.
The Aspiration Bypass System - Tips with
the ABS®feature contain a small hole in the distal
portion of the tip's wall.
Kelman®Tips - The Kelman®tips have a bent shaft
which generates transverse ultrasound motion, in addition
to the conventional longitudinal motion, to enhance cutting
efficiency. In addition, the bend allows better visibility
during the surgical procedure.
Tapered Tip - The tapered ABS®tip is a combination
of the 0.9 mm tip and the flared ABS®tip. The shaft
inner and outer diameters is equivalent to straight tips,
while the distal end is comparable to flared tips. The
tapered ABS®tip has the improved holding force of a
flared tip, and the same aspiration flow characteristics
as a straight tip.
ABS®hole
ABS®hole
ABS®hole
Flared ABS®Tips - The flared tips have a larger proximal
port, providing increased holding force. They narrow in the
middle of the shaft, thus allowing smaller incisions and
improving occlusion breaks by reducing outflow from the
anterior chamber, following occlusion breaks. Flared tips
also have the Aspiration Bypass System feature, to further
enhance performance.
Figure 2-77 TurboSonics® Tips - Shown here are samples of U/S tips used with the
U/S handpieces.
2.102 8065751025
Constellation® Vision System
Figure 2-78 Infiniti™* U/S Handpiece shown with Infusion Sleeve and Bubble
BUBBLE SUPPRESSION INSERT (BSI)INFUSION SLEEVE
MicroSmooth™* Infusion Sleeves
Infusion sleeves cover the tip of the handpiece to provide irrigation to the anterior
chamber of the eye during surgery (see Figure 2-78). Infusion sleeves are used with
the Infiniti™* U/S handpiece, and with some Ultraflow™* I/A handpieces. Infusion
sleeves used with Infiniti™* U/S handpieces require a BSI (bubble suppression
insert). Infusion sleeves must be correctly matched to the specific tip type (see the
following descriptions).
Depending on the needs and technique preferred by the surgeon, various styles of
infusion sleeves are available as listed in Table 2-2.
Table 2-2. MicroSmooth™* Infusion Sleeves
Infusion Sleeves Sleeve Color Recommended
Incision Size
0.9 mm 1.1 mm
High Infusion Semi-transparent Semi-transparent
blue 3.2 mm
Standard -- Blue 3.0 mm
Micro Purple Blue/green 2.75 mm
Ultra Red Green 2.2 mm
8065751025 2.103
Constellation® Vision System
Ultraflow™* Handpieces and Tips
The Ultraflow™* handpiece is used in I/A mode to maintain chamber pressure with
irrigation while removing cortical material via aspiration. (See Figure 2-79 and note
the band markings on the tips that identify size of tip aperture.) Some configurations
of the Ultraflow™* IT and SP handpieces also use infusion sleeves. The following
Ultraflow™* I/A handpieces and tips are available:
Ultraflow™* IT Handpiece and Interchangeable Tips - The Ultraflow™* IT
consists of a handpiece body that accepts interchangeable tips. These tips do not
require an adapter or infusion sleeve as they contain a built-in metal infusion
sleeve.
Ultraflow™* IT Handpiece and Threaded Tip Adapter - Reusable I/A tips with
TurboSonics® silicone infusion sleeves can be used with the Ultraflow™* IT
handpiece with threaded tip adapter.
Ultraflow™* SP Handpiece (Single-Piece with fixed tips) - The Ultraflow™* SP
consists of a single-piece handpiece with irrigation tip, threaded tip adapter, or I/A
tip with a built-in metal infusion sleeve. Various tip configurations are available.
WARNINGS!
Use of non-Alcon surgical reusable or disposable I/A handpieces that do not meet
Alcon surgical specifications, or use of an Alcon handpiece not specified for use
with the Constellation® Vision System, may result in a fluidic imbalance. This, in
turn, may cause a shallowing or collapsing of the anterior chamber.
Exceeding the recommended level of 100 mmHg with a 0.5 mm or larger I/A tip
may cause anterior chamber shallowing and/or incarceration or tearing of the
posterior capsule.
I/A tips are not to be used with U/S handpieces.
2.104 8065751025
Constellation® Vision System
Figure 2-79 Ultraflow™* IT handpiece and tips
Figure 2-80 Ultraflow™* IT handpiece with infusion sleeve, reusable I/A tip, and
threaded tip adapter
Figure 2-81 Ultraflow™* O-ring tool with large and small O-rings
Figure 2-82 Ultraflow™* SP handpiece (handpiece shown with .3 mm 45° tip)
O-Rings
Straight Tip
120°Tip 90°Tip 45°Tip Curved Tip
O-Rings
0.5 mm I/A Tip
Threaded Tip Adapter
0.2 mm I/A Tip
0.3 mm I/A Tip
O-Ring Extractor
Small O-Rings
Large O-Rings
8065751025 2.105
Constellation® Vision System
Diathermy/Coagulation Handpieces
Single use Bipolar Coagulation Brushes are available in a wide variety of
configurations: straight, curved, 20-gauge, 23-gauge, tapered, and widestroke. All
single use bipolar accessories are available with and without cables. Also available
are reusable and single-use bipolar cables.
Bipolar Coagulation Forceps are lightweight and ergonomically designed to reduce
hand fatigue, as well as to provide precise control and safety. They are available in
high-conductive non-stick alloy, titanium, or single-use configurations. They are also
available with a wide variety of tip styles.
Figure 2-83 Single use bipolar brush