Operation Manual
BM1000 Pulse Oximeter
BERRYMED LIMITED
Version: 1.0
Product Description
BM1000 Fingertip Pulse Oximeter is an important and common device to check oxygen saturation
(SpO2) and pulse rate. It’s a small, compact, simple, reliable and durable physiological monitoring
device. Include the mainboard, OLED display and dry batteries. It is suitable for monitoring adults
and children. It can be used in the hospital’s operation room, ICU, clinic section office, out-patient
department, sickroom and emergency treatment. It can also be used in the recovery and health care
organizations, th e community m edical treatments, th e o xygen bars, t he fam ily n ursing, the
physical care in sports (you can use it before or after the sport, but it is not recommended to use it
during the sport).
Measurement Principle
An experience formula of data process is established taking use of Lambert Beer Law according to
Spectrum A bsorption Characteristics of hemoglobin ( Hb) and O xyhemoglobin ( HbO2) in glow
and near-infrared zo nes. O peration principle of th e i nstrument is Ph otoelectric O xyhemoglobin
Inspection T echnology is ad opted in a ccordance wi th Capacity Pu lse Sc anning an d R ecording
Technology, s o t hat tw o beams of dif ferent w avelength of lights (66 0nm glow a nd 9 40nm near
infrared light) can be focused onto human nail tip through perspective clamp finger-type sensor.
Then measured signal can be obt ained by a photosensitive element, information acquired through
which w ill be show n o n tw o gro ups of LED s th rough pr ocess in e lectronic c ircuits a nd
microprocessor.
Arterial oxygen s aturation is m easured by a m ethod called p ulse oximetry. It i s a continuous,
non-invasive method based on the different spectra absorption of hemoglobin and oxyhemoglobin
(called spectrophotometer principle). It measures how much light, sent from light sources on the
other side.
Diagram of Operation Principle
Safety Information
The person who uses the pulse oximeter must receive adequate training before use.
The pulse oximeter is intended only as adjunct in patient assessment. It must be used in
z
conjunction with clinical signs and symptoms. It is not intended as a device used for
treatment purposes.
When using the pulse oximeter together with the electrical surgery equipment, the user
should pay attention to and guarantee safety of the patient being measured.
EXPLOSION HAZARD: Do not use the pulse oximeter in the presence of flammable
anesthetics, explosive substances, vapors or liquids.
Make sure not to use the pulse oximeter during MRI (magnetic resonance imaging)
scanning or CT (Computed Tomography) environment because induced current could
potentially cause burns.
The pulse oximeter is without alarm function. Continuous monitoring for a long time is
not suitable.
No modification of this product is allowed. Maintenance should be operated by
professional maintenance personnel who are approved by manufacturers.
Please shut off the power before clean the pulse oximeter. Never permit high-pressure
and high-temperature disinfection of the device. Never use cleaning agents/disinfectants
other than the recommended.
The product is commonly seal product. Keep its surface dry and clean, and prevent any
liquid from infiltrating it.
The pulse oximeter is precision and fragile. Avoid pressure, knock, strong vibration or
other mechanical damage. Hold it carefully and lightly. If it is not in use, it should be
appropriately placed.
For disposal of pulse oximeter and accessories, follow local regulations or your
hospital’s policy regarding disposal of such pulse oximeter and accessories. Do not
dispose randomly.
Use AAA alkaline batteries. Do not use carbon or poor quality batteries. Remove the
batteries if the product is not to be used for a long time.
If patient is an intended operator, you must read the operation manual carefully and
understand deeply or consult with the doctor and manufacturer before using. If you
have any discomfort in use, please stop using immediately and go to the hospital.
A functional tester can’t be used to assess the accuracy.
The patient is an intended operator and can perform the maintenance the equipment.
Product Feature
1. Simple and convenient usage of product, simple one-touch operation.
2. Small volume, light weight, convenient to carry.
3. Lower consumption, original two AAA batteries can continuously work for 24 hours.
4. Low voltage rem inder show s i n scre en w hen there’s l ow bat tery, m ay i nfluence th e no rmal
working.
5. The machine will automatically power off when there’s no signal generated.
6. Daily maintenance and calibration is unnecessary.
7. Communication c an be real ized between the pr oduct and m obile phone w ith i ts w ireless
Bluetooth.
Display Introduction
Figure 1
Battery Installation
1. Hold the product in one hand with the front panel facing the palm. Put the other hand’s big
finger on the battery cabinet lid’s press sign, press downwards and push the lid open at the
same time. The battery cabinet is opened as shown in Figure 2.
2. Install batteries into the slots per the “+” and “-” symbols as shown in Figure 3.
Cover the lid onto the cabinet and push it upwards to make it close well.
z The positive and negative electrodes of batteries should be installed correctly. Otherwise
the device will be damaged.
z When install or remove batteries, please follow the correct operation sequence to operate.
Otherwise the battery compartment will be damaged.
Figure 2
Figure 3
Hang lace Installation
1. Thread thinner end of the hang lace through the hanging hole.
2. Thread thicker end of the lace through the threaded end before pulling it tightly.
Directions for use
1. Press Clip’s press sign in the Figure 1 and open the clip. Let the testee’s finger put into the
rubber cushions of the clip, make sure the finger is in the right position as shown in Figure 4,
and then clip the finger.
2. Press the “POWER/FUNCTION" switch but ton on the front pa nel to turn on the pro duct.
Using first fin ger, m iddle f inger or ri ng f inger w hen doing test. D on’t sh ank t he finger and
keep the testee at case during the process. The readings will be displayed on the OLED screen
a moment later as shown in Figure 5.
Display screen data refresh time for one second.
Figure 4
Figure 5
NOTE:
z Don’t put the product on extremities with arterial catheter or venous syringe.
z Cover the sensor with opaque material under the condition of strong light. Failure to do
so will result in inaccurate measurement.
z Try to keep the patient still (specially the arm) and avoid the measured site suffering
excessive motion.
z Don’t use the product to measure patients whose pulse rate is lower than 30bpm, which
may cause incorrect results.
z Make sure no contamination or scar exists in the size where the product is placed.
Otherwise, the measured result may be incorrect because the signal received by the
sensor is affected.
z When used on different patients, the product is prone to crossed contamination, which
should be prevented and controlled by the user. Disinfection is recommended before
using the product on other patients.
z Prolonged use or the patient’s condition may require changing the sensor site
periodically. Change sensor site and check skin integrity, circulatory status, and correct
alignment at least even 2 hours.
z Don’t perform SpO2 monitoring and NIBP measurements on the same arm
simultaneously. Obstruction of blood flow during NIBP measurements may adversely
affect the reading of the SpO2 value.
z Testee’s fingernail can’t be too long. Otherwise the finger can’t be inserted into the
sensor to a suitable depth and the SpO2 measurements may be inaccurate.
z Make sure to place the product on the finger in a correct direction. The LED part of the
sensor should be at the backside of the patient hand and photodetector part at the inside.
Make sure to insert the finger to suitable depth into the sensor so that the fingernail is
z
just opposite to the light emitted from the sensor.
The highest temperature of sensor contacts with patient’s skin don’t be allowed more
than 41℃.
Shock, anemia, hypothermia and the application of vasoconstriction drug may decrease
arteria blood flow to an unmeasurable level.
Pigment, or deep color (for example: nail polish, artificial nails, dye or pigmented cream)
may cause inaccurate measurements.
Function Description
a. When the data has been displayed on the screen, change the display direction by pressing the
“POWER/FUNCTION” button again. (as shown in Figure 6)
b. When th e product is powered o n, long press the “ POWER/FUNCTION” but ton, Bluetooth
function will be started. The Bluetooth indicator light on the top of display will flicker. (as shown
in Figure 7)
c. The product will a utomatically be powered off when no fi nger is in the device for longer than
10 seconds. And switch to another display mode. (as shown in Figure 8)
Figure 6
Figure 7
Figure 8
Bluetooth Communication Function
a. Open the mobile phone, and double-click SpO2 software icon “
following interface. (as shown in Figure 9)
”. And enter the
Figure 9
b. Click the “Add new userss” and enter thhe following interface. (ass shown in Fiigure 10)
Figure 10
c.
hown
Input the useer name, clickk the “Details” and jump the followingg informationn table. (as sh
in Figure 11))
Figure 11
d. Input personnal informatioon, then clickk “Save” and enter the folllowing interfface. (as show
wn in
Figure 12)
Figure 12
e.
(as shown in Figgure 13)
Click “berry”” and enter thhe following interface.
Figure 13
f.
Click“
”, search equuipment and jump
the folllowing interfa
face. (as show
wn in Figure 14)
Figure 14
c lick“BerryMe
ed” and enteer the follo w
wing in terfacee. (as
g. Wait to searrch out the equipment,
shown in Figgure 15)
Figure 15
h. Click“Pair”, to p air wi thh“BerryMed”” a nd e nter the
t follow ingg test interf aace. (a s sho wn
w in
Figure 16)
Figure 16
i.
l escape key
k of your mobile pho ne a nd jum p the
If you wan t to e xit the software, c lick
following wiindow. And thhen click“OK
K”, exit the so
oftware. (as shhown in Figu
ure 17)
Figure 17
Cleaning
and
Disinfe
ection
Cleaaning
1. Clean the prooduct with cootton or soft cloth
moisteneed with waterr.
2. After cleaninng, wipe off thhe water withh a soft cloth..
3. Allow the product to air dry.
Disinfection
The recommendeed disinfectannts include: etthanol70%,
isopropanol70%, glutaraldehyyde( 2%) soluution disinfectants.
1. Clean the prooduct as instrructed above.
w on e of tthe recomm ended
with
2. Disinfect th e prod uct wiith cot ton orr soft cloth moistened
disinfectants.
f th e disinfeectant lef t onn t he p roduct with a soft cloth
3. After dis infeection, be s urre to wi pe o ff
moistened with
w water.
4. Allow the sennsor to air drry.
Packing
List
The standard config
guration
Finggertip pulse oximeter
1pc
Hang lace
1pc
The operation maanual
1pc
Technical
Specificat
tions
de: OLED
1. Display mod
2. SpO2:
3.
4.
5.
6.
Measurement range: 35~100%
Accuracy: ±2%(80%~100%); ±3%(70%~79%)
Pulse Rate:
Measurement range: 25~250bpm
Accuracy: ±2bpm
Electrical specifications:
Working voltage: D.C.2.2 V~D.C.3.4V
Battery Type: Two 1.5V AAA alkaline batteries
Power consumption: smaller than 50mA
Product specifications:
Size: 58 (H) × 34 (W) × 30(D) mm
Weight: 50g (include two AAA batteries)
Environment requirements:
Temperature:
Operation: +5~+40℃
Transportation and storage: -10~+50℃
Humidity:
Operation: 15%~80%(noncondensing)
Transportation and storage: 10%~90%(noncondensing)
Atmospheric pressure:
Operation: 860hPa~1060hPa
Transportation and storage: 700hPa~1060hPa
Troubleshooting
Trouble
Possible reason
solution
The SpO2 and P R
can’t be displayed
normally an
the v
alue
disappeared.
1. The fi nger is n ot properly
positioned.
2. The p atient’s SpO2 is to o l ow
to be detected.
3. Bluetooth signal is interrupted.
1. Please the finger pr operly and try
again.
2. Try ag ain; Go to a ho spital fo r a
diagnosis if you are sure the device
works all right.
3. Check t he Bluetooth co nnection
and reconnect.
The SpO2 and P R
display instable.
1. The f inger is not plac ed i nside
enough.
2. The fi nger is sh aking o r the
testee is moving.
1. Place t he f inger properly and try
again.
2. Let the testee keep calm.
The device can’t
be powered on.
1. The batt eries are drai ned or
almost drained.
2. The ba tteries are not i nserted
properly.
3. The device’s malfunction.
1. Change batteries.
2. Reinstall batteries.
3. Please contact the supplier.
The scre en is
suddenly off.
1. The product is a utomatically
powered of f when n o si gnal is
1. Normal.
2. Replace the batteries.
detected longer than
10
seconds.
2. Power qua ntity of the ba tteries
is exhausted.
Symbol Meaning
Symbol
Meaning
“CAUTIOUS”! Please refer to the operation manual.
Type BF Equipment.
The product does not contain alarm function.
When the end-user wishes to discard this product, it must be sent to separate
collection facilities for recovery and recycling.
Information of manufacture, including name and address.
Date of manufacture.
European Union for approval.
Serial Number.
Batch Code.
Type Number.
The European Union authorized.
IP21
The product i s prote cted a gainst harmful effects of dripping w ater pe r IEC
60529.
Shanghai Berry Electronic Tech Co., Ltd
Suit C, 1st Floor, 7th Building, N0.1188 Lianhang Road, Minhang District,
Shanghai, China 201112
TEL: 0086-21-5853 1958 FAX: 0086-21-5853 0468
www.berry-med.com
Kingsmead Service limited (UK)
145-157 St John Street, London, EC1V 4PY (UK)
TEL :044-20-7193 9159 F AX: 044-20-7193 9159
www.kingsmead-service.com
If you need additional information, please contact with the company.
FCC warning statements:
This equipment has been tested and found to comply with the limits for a Class B digital device,
pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection
against harmful interference in a residential installation. This equipment generates, uses and can
radiate radio frequency energy and, if not installed and used in accordance with the instructions, may
cause harmful interference to radio communications. However, there is no guarantee that interference
will not occur in a particular installation. If this equipment does cause harmful interference to radio or
television reception, which can be determined by turning the equipment off and on, the user is
encouraged to try to correct the interference by one or more of the following measures:
‧ Reorient or relocate the receiving antenna.
‧ Increase the separation between the equipment and receiver.
‧ Connect the equipment into an outlet on a circuit different from that to which the receiver is
connected.
‧ Consult the dealer or an experienced radio/TV technician for help.
Caution: Any changes or modif cations to this device not explicitly approved by manufacturer could
void your authority to operate this equipment.
This device complies with part 15 of the FCC Rules. Operation is subject to the following two
conditions: (1) This device may not cause harmful interference, and (2) this device must accept any
interference received, including interference that may cause undesired operation.
IC warning statements:
-English Warning Statement:
“This device complies with Industry Canada licence-exempt RSS standard(s).
Operation is subject to the following two conditions: (1) this device may not
cause interference, and (2) this device must accept any interference, including
interference that may cause undesired operation of the device."
The digital apparatus complies with Canadian CAN ICES-3 (B)/NMB-3(B).
-French Warning Statement:
“Le présent appareil est conforme aux CNR d'Industrie Canada applicables
aux appareils radio exempts de licence. L'exploitation est autorisée aux deux
conditions suivantes : (1) l'appareil ne doit pas produire de brouillage, et (2)
l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique subi, même
si le brouillage est susceptible d'en compromettre le fonctionnement."
Cet appareil numérique de la classe B est conforme à la norme NMB-003 du
Canada.
This equipment complies with IC radiation exposure limits set forth for an
uncontrolled environment and meets RSS-102 of the IC radio frequency (RF)
Exposure rules. This equipment has very low levels of RF energy that are deemed to
comply without testing of specifc absorption ratio (SAR).Cet équipement est
conforme aux limites d’exposition aux rayonnements énoncées pour un
environnement non contrôlé et respecte les règles d’exposition aux fréquences
radioélectriques (RF) CNR-102 de l’IC. Cet équipement émet une énergie RF très
faible qui est considérée conforme sans évaluation du débit d’absorption spécifque
(DAS).

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : No
XMP Toolkit                     : Adobe XMP Core 4.2.1-c041 52.342996, 2008/05/07-20:48:00
Producer                        : Acrobat Distiller 7.0 (Windows)
Creator Tool                    : PScript5.dll Version 5.2
Modify Date                     : 2016:01:07 14:17:30+08:00
Create Date                     : 2015:12:10 11:31:55+08:00
Metadata Date                   : 2016:01:07 14:17:30+08:00
Format                          : application/pdf
Title                           : Microsoft Word - User Manual
Creator                         : Administrator
Document ID                     : uuid:ff91d7a1-5524-4368-91d1-bbaa14faf1e3
Instance ID                     : uuid:f507fd9a-32d4-4f61-b2d8-155143b1c319
Has XFA                         : No
Page Count                      : 16
Author                          : Administrator
EXIF Metadata provided by EXIF.tools
FCC ID Filing: 2AG2CBM1000B

Navigation menu