BEWELL CONNECT BW-X07HD HANDHELD VITALSIGNS MONITORING SYSTEM User Manual

BEWELL CONNECT CORP HANDHELD VITALSIGNS MONITORING SYSTEM

User Manual

1 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017  BewellConnect Logo Picture cover REF PRODUCT+CE+pictos         VisioCheck  By Visiomed     DIAGRAMS Diagram A                 1) LED (Flash)  2) HD Camera 3) Indicating light 4) Display screen 5) Home button 6) Infrared probe 7) Earphone port 8) SIM card tray 9) SpO2 port 10) BPM port
2 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 11) Speaker 12) On/Off button 13) Up/down wheel (OK Button) 14) ECG port 15) USB port      Diagram B       Diagram C      Diagram D    Diagram E     Diagram F   Accessories picture ECG Connexion  Diagram  Thermometer Diagram
3 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017   + rating label      SpO2 Connexion  Diagram   BPM Connexion  Diagram
4 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017
5 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017     TECHNICAL DEFINITIONS ECG or EKG : Electrocardiogram SpO2: Blood oxygen saturation MRI: Magnetic resonance imaging RF: Radio Frequency EMC: Electromagnetic Compatibility PVC: Polyvinyl chloride ABS: Acrylonitrile Butadiene Styrene TPU: Thermoplastic Polyurethane  ISM device: Industrial, scientific and medical device CISPR  11:  International  Standard  for  electromagnetic  emissions  (disturbances)  from  industrial, scientific and medical (ISM) equipment MAM: Mean Arterial Measurement MAP: Mean arterial pressure SBP: Systolic Blood Pressure  DBP: Diastolic Blood Pressure  PI : Perfusion index
6 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017  Language Logo BewellConnect COVER VisioCheck Picture of product CE + Pictos  + Ref product By Visiomed   This product is a medical device, class IIa, compliant with European directive and part 15 of the FCC rules.  TABLE OF CONTENTS  1. Notes on safety (general + each device) 1.1 ECG cable/Lead wires and SpO2 adapter cable 1.2 Reusable blood pressure cuff 1.3 Finger clip SpO2 sensor for adult 1.4 FCC  statement 2. Information 3. Features  4. Description 5. Precautions before use 5.1 ECG 5.2 Oximeter 5.3 Blood pressure monitor 5.4 Thermometer 6. User guide  6.1 General functions 6.2 Taking a measurement 6.2.1 ECG 6.2.2 Oximeter 6.2.3 Blood pressure monitor 6.2.4 Thermometer
7 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 7. HD Camera 8. Storage and maintenance 9. Specifications 9.1 Technical specifications 9.2 Blood pressure 9.3 SpO2 9.4 ECG 9.5 Temperature 9.6 Wireless communication 9.7 Key and interface 9.8 Power parameter 9.9 Appearance 9.10 Accessories included 9.11 Other Accessories  10. Troubleshooting  Log onto our website to find out how to use VisioCheck  www.bewell-connect.com/install The  manufacturer  reserves  the  right  to  change  the  product’s  technical specifications  without  prior notice.  Dear customer, Thank you for purchasing  the HandHeld VitalSigns Monitoring System VisioCheck. Please read this guide prior to device use.  This device is not intended for use as a conventional diagnostic tool, but it is a healthcare tool which can provide a doctor the recorded data as a reference to help evaluate the patient’s health. It must be used in conjunction with other devices to evaluate symptoms and clinical signs.  For further information about the health or specific measurements, PLEASE CONTACT A DOCTOR.
8 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 1. Notes on safety  - The equipment must be used per the requirements of this manual. Failure to properly use the  equipment  may  result  in  degraded  performance,  loss  of  functionality,  measurement abnormalities  and  equipment  damage.  The  manufacturer  assumes  no  responsibility  for personal injury or device damage sustained by or through the use of this product.  - This device is intended to monitor patient’s physical index at home, hospital, hospital facility or other outdoor situation. Do not use for any other purpose. - This product should not encourage self-medication or adaptation of the treatment. - Always consult the doctor if the patient has any questions or he believes he has abnormal measurements. The measurement results are given for information - Do not take measurements in a moving vehicle. - Stop use of the device in case of anomalies or malfunction.  Visiomed and Bewell Connect Corp are not held liable for any problems resulting from failure to observe the installation and operating rules stated in these directions for use. Visiomed and Bewell Connect Corp are not held liable of the use with incompatible components which can result in degraded performance.  Warning  - This device must always be placed in a clean and dry place.  - This  device  is  not  designed  for  use  by  people  (including  children)  with  reduced  physical, emotional or mental abilities or  those lacking experience or  knowledge unless used in  the presence  of  someone  responsible  for  their  security,  under  supervision  or  following instructions about how to use the device. - Do not use the ECG function while charging the device because of the risk of electric shock.  - Do not use the ECG with a cardiac pacemaker or defibrillator. - The protective glass over the lens is the most fragile part of the thermometer. Do not touch the glass of the infrared lens with your fingers. - Do not insert other objects into any holes. - Do not use accessories, detachable parts and materials not described in the instructions for use - Do not use sharp tools to operate any of the buttons. - Do not interconnect this device with other equipment not described in the instructions for use - Do  not  open,  disassemble,  repair  or  modify  the  device.  In  case  of  problems,  contact customer service. - Do  not  expose  the  device  to  static  electricity.  Disperse  static  electricity  from  your  body before handling the device. - Do not use near flammable anesthetics, or near pressurized oxygen such as in a hyperbaric chamber, ultraviolet sterilizer or oxygen tent. - Do  not  expose  the  device  to  strong  shocks,  vibrations,  electric  shocks,  sunlight  or  water. Never drop the device. - Do not operate the device if it has been immersed in liquid - Follow the maintenance instructions specified in this manual.
9 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 - This  electrical  medical  equipment  requires  specific  precautions  regarding  electromagnetic compatibility.  It  must  be  installed  and  used  in  accordance  with  EMC  (Electromagnetic Compatibility) Information. - Mobile  and  portable  RF  communications  equipment  may  affect  the  equipment.  The equipment  is  intended  for  use  in  an  electromagnetic  environment  in  which  radiated  RF disturbances are  controlled.  The  customer or  the  user of the equipment can help prevent electromagnetic  interference  by  maintaining  a  minimum  distance  between  portable  and mobile RF  communications equipment (transmitters) and the equipment as  recommended below, according to the maximum output power of the communications equipment.  - The  received  signal  may  be  interfered/disrupted  by  other  equipment  operating  in  this frequency band (i.e. Bluetooth, other non-associated access points, microwave ovens, ISM devices, etc)  even  if  the  other  equipment complies  with  its  applicable  regulations  such as CISPR 11.  - Use this device in an ambient temperature range of 5 to 40°C / 41 to 104°F.  - Avoid using the device immediately after a significant change in temperature or humidity.  - Do not expose this device to extreme temperature conditions > 50°C (122°F) or < -10°C (14°F).  - Do not use this device at a relative humidity of more than 95%. - Do not use the device out of the required condition to avoid inaccurate measurement.  - The  values  displayed  by  the  device  are  the  ones  obtained  at  the  time  of  measurement. Medical conditions can change suddenly. If you notice any change in your condition, consult a doctor, regardless of the measured results.  - Do not use the device for measurements when the battery power is lower than 15% or 3.5V. - Be careful about the small parts broken off the device, as children may swallow and it could lead to suffocation.  - When the screen displays the battery symbol  , this indicates that the battery is low and you should charge the product through our adapter as soon as possible.  1.1. ECG cable/lead-wires and SpO2 adapter cable   - As  with  all  medical  equipment,  carefully  place  cables  to  reduce  the  possibility  of  patient entanglement or strangulation. - Do not use the cables during MRI scanning. Conducted current could cause burns. - Do not use the ECG function during charge of the device. Risk of electric shock   - Do not use the ECG with a cardiac pacemaker or defibrillator. - Do not use lotion or oil prior to the test. Skin must be clean and dry. - This product is not suitable for PVC, TPU and ABS plastic allergy patients and users. - The ECG cable/lead-wires and SpO2 adapter cable are designed for use with VisioCheck - Do  not  use  the  ECG  cable/lead-wires  and  SpO2  adapter  cable  if  the  cable,  wires  and connector are broken or damaged.  1.2. Reusable Blood Pressure Cuff for adults   - When the cuff is not connected to  monitor, always fit cap to  tubing to prevent accidental entry of liquid. - Periodically inspect and replace damaged or deteriorated cuffs.
10 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 - Only  use  the  appropriate  size  cuff  when  the  artery  index  marker  falls  within  the  printed range, Otherwise erroneous readings may result.  Principle of operation This device adopts the oscillometric technology with Fuzzy Algorithm to measure the arterial blood pressure and pulse rate. The cuff is wrapped around the arm and automatically inflated by the air pump. The sensor of the device catches weak fluctuation of the arm in response to each heartbeat. The amplitude of the pressure waves is measured, converted in millimeters of the mercury column, and is displayed by digital value.  1.3. Finger Clip SpO2 Sensor for Adult  - Probes are designed for use with VisioCheck. Do not use them with other device.  - Using the sensor in the presence of direct sunlight may result in inaccurate measurements. In such cases, put  the  receiving surface of  the sensor  back  to  ambient light sources (such as fluorescent lamps) or cover the sensor with a black cloth after removing the sensor from the patient's  body  in  order  for  the  monitor  to  displaySpO2  value,  pulse  rate  value  and  their waveforms. - Reusable sensors must be moved to a new site on the patient at least every 4 hours. Because individual skin condition affects the ability of the skin to tolerate sensor placement, it may be necessary  to  change  the  sensor  site  more  frequently  with  some  patients.  If  skin  integrity changes, move the sensor to another site. - Intravascular dyes may lead to inaccurate measurements. - The performance of the sensor is compromised by motion.  - Do not apply tape to secure the sensor in place or to tape it shut; venous pulsations may lead to inaccurate saturation measurements. - As  with  all  medical  equipment,  carefully  place  cables  to  reduce  the  possibility  of  patient entanglement or strangulation. - Do not use the  sensor  or  other oximeter sensors  during  MRI  scanning. Conducted  current may cause burns. Also, the sensor may affect the MRI image, and the MRI unit may affect the accuracy of oximeter measurements. - Do not alter or modify the sensor. Alterations or modifications may affect the performance or accuracy. - Do not use if the wire or the shell of connector is broken or damaged. - Sterilize using ethylene oxide only. - The MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided at the end of this user manual. - Portable  and  mobile  RF  communications  equipment  can  affect  MEDICAL  ELECTRICAL EQUIPMENT. - The equipment is without a manual sensitivity adjustment. The minimum amplitude or value of patient physiological signal is 70% SpO2. Operation of the EQUIPMENT or SYSTEM below this amplitude or value may cause inaccurate results. - The use of accessories, transducer and cables other than those specified, with the exception of  those  sold  by  the  manufacturer  of  the  equipment  or  system  as  replacement  parts  for internal  components,  may  result  in  increased  EMISSION  or  decreased  IMMUNITY  of  the EQUIPMENT or SYSTEM.
11 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 - Be  aware  that  following  removal  of  the  sensor  from  the  patient,  it  is  possible  that environmental light may cause the monitor to continue to display a waveform or data values but these are not a basis for a clinical diagnosis. - Make sure the application site for the sensor is not deeply pigmented, or deeply colored. For example,  nail  polish,  artificial  nails,  dye  or  pigmented  cream  may  cause  inaccurate measurements. In any of these cases reposition the sensor or choose an alternative sensor for use at a different site. - Protect the sensor and connector from becoming soaked with any liquid. Do not immerse the sensor or  connector  in  any  cleaning  solution,  disinfectant,  or  other liquid  (the  sensor  and connector may be wiped with cleaner and disinfectant but never immersed). - Where  possible,  the  application  site  for  the sensor should be an  extremity free  of  arterial catheters, blood pressure cuffs, or intra-vascular infusion lines. - At  elevated  ambient  temperatures,  patient  skin  could  become  severely  burned  after prolonged sensor application on sites with poor blood circulation. To prevent this condition, be sure to check patient application sites frequently. All listed sensors operate without risk of exceeding  41℃  /105.8°F  on  the  skin  if  the  initial  skin  temperature  does  not  exceed 35℃/95°F. - The waste of the pulse oximeter probe must not be disposed of as unsorted municipal waste and  must  be  collected  separately.  Please  contact  an  authorized  representative  for information concerning the disposal of your equipment - A warning to the effect that misapplication of a pulse oximeter probe with excessive pressure for prolonged periods can induce pressure injury.   1.4. FCC statement changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.  This  device  complies  with  Part  15  of  the  FCC  Rules.  Operation  is  subject  to  the  following  two conditions:  (1) this device may not cause harmful interference, and (2)  this  device  must  accept  any  interference  received,  including  interference  that  may  cause undesired operation.  This  equipment  has  been  tested  and  found  to  comply  with  the  limits  for  a  Class  B  digital  device, pursuant to  Part  15  of  the  FCC  Rules. These  limits  are  designed to  provide reasonable  protection against harmful  interference in  a  residential  installation. This  equipment generates,  uses,  and  can radiate radio frequency energy, and if not installed and used in accordance with the instructions, may cause  harmful  interference  to  radio  communications.  However,  there  is  no  guarantee  that interference  will  not  occur  in  a  particular  installation.  If  this  equipment  does  cause  harmful interference to radio or television reception, which can be determined by turning the equipment off and on,  the user is encouraged to  try to correct the interference by one or  more of the  following measures: – Reorient or relocate the receiving antenna. – Increase the separation between the equipment and receiver. –Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
12 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 – Consult the dealer or an experienced radio/TV technician for help. This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. End user must follow the specific operating instructions for satisfying RF exposure compliance. This transmitter must not be co-located or operating in conjunction with any other antenna or transmitter. The device is designed to meet the requirements for exposure to radio waves established by the Federal Communications Commission (USA). These requirements set a SAR limit of 1.6 W/kg averaged over one gram of tissue. The highest SAR value reported under this standard during product certification for use when properly worn on the body is 1.250W/kg.  For body operation, this device has been tested and meets FCC RF exposure guidelines when used with any accessory that contains no metal and that positions a minimum of 5mm from the body. Use of other accessories may not ensure compliance with FCC RF exposure guidelines.  2. Information VisioCheck® is a four in one handheld, smart and connected medical device, designed for the remote monitoring  of  vitalsigns  and  telemedicine,  by  health  professionals  and  patients.  The  product combines a blood pressure monitor, a pulse oximeter, an electrocardiograph (5 electrodes, 7 leads), and a non-contact infrared thermometer.   VisioCheck software is an embedded software which provides multiple services. It is applicable for home care, hospital and other remote and mobile situations. The software can be used by healthcare professional providers or patients, to monitor vitals. Medical data can be stored locally and a reports can be sent through WiFi, 3G or 4G. It is designed to be easy to use to take measurements and share results accross long distances.  VisioCheck  allows  the  user  to  communicate  with  their  health  professional  through  an  HD  video conferencing  system  but  cannot  be  used  as  a  phone  (no  voice-only  IP  connection,  only  M2M  sim cards)   3. Features  The devices used with the application include: ->  Non-Contact  Thermometer  based  on  a  miniature  version  of  Visiomed’s  ThermoFlash® technology -> Blood Pressure Monitor (and pulse rate) -> 5 electrodes and 7 leads ECG (and pulse rate) -> Pulse Oximeter (and pulse rate)   4. Description View diagram A page 2  Accessories View diagram B page 2 1) BPM Cuff *1 2) SpO2 Adapter Cable *1
13 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 3) SpO2 universal sensor connector *1 4) ECG 5-Electrode Cable*1 5) Infrared Probe Cover*1 6) USB Cable*1 7) Adapter*1  5. Precautions before use  Before using the VisioCheck, - Do not confuse self-monitoring with self-diagnosis. Measurement should only be interpreted by a health professional who is familiar with your medical history. - If you are taking medication, consult with your physician to determine the most appropriate time to measure your vital signs.  - This device should not  encourage self-medication or  adaptation of your treatment. NEVER change a prescribed medication without first consulting with your physician - This device is designed for use by adults only. If using the device on a child, toddler or elderly patient, please consult with your physician first.  - Check that the accessories are intact and clean.  - Always consult your doctor with any medical questions.  Adhere to the temperature ranges and humidity levels shown in the specifications.  5.1. Precaution before use ECG  - Low battery may affect the accuracy of heart rate measurements.  - Do not use this device near an Electrosurgical Unit (ESU), since it may function incorrectly as a result. - Do not use the ECG while charging VisioCheck because of the risk of electric shock. - The electrodes pads are disposable and for single use only  Duration of use: it is recommended to replace the ECG cable every 6 months. If it is necessary to use, please evaluate the period of use, it should be regularly inspected.  5.2. Precaution before use Oximeter  - Do not use this device near an Electrosurgical Unit (ESU), since it may function incorrectly as a result. -  The  pulse  oximeter  must  be  placed  so  that  the  pulse  can  be  recorded  correctly  and  the  main objective is to evaluate the oxygen saturation (SpO2) level. Check that nothing is interfering with the readings. - Do not use the device in an MRI (Magnetic Resonance Imaging).  - Do not use the device in an explosive atmosphere. - Frequently check (every 30 minutes) on the application zone of the device sensor (checking for signs of skin sensitivity, change in position, etc.). - The device must not be used near or on top of another device. - Inaccurate readings may result from:
14 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 -> Nail-polish (colorimetric interference) or artificial nails -> High levels of dysfunctional hemoglobin (carboxyhemoglobin or methemoglobin) -> Intravascular dyes (indocyanine green or methylene blue) -> Bright light. If necessary, protect the sensor positioning area -> Excessive movement of the patient -> Defibrillators and interference from high-frequency electrosurgical devices -> Venous pulse ->  Placing  the  sensor  on  a  patient  with  a  blood-pressure  cuff,  an  arterial  or  intravascular catheter -> High blood pressure, severe vasoconstriction, acute anemia or hypothermia -> Heart failure or stress -> Poor pulse quality (low blood flow) -> Low levels of hemoglobin.  Duration  of  use:  It  is  recommended  to  replace  the  SpO2  adapter  cable  every  6  months.  If  it  is necessary to use, please evaluate, and the period of use should be thoroughly inspected.  5.3. Precaution before use Blood pressure monitor  Blood  pressure  measurements  determined  by  this  monitor  are  equivalent  to  those  obtained  by  a trained observer using the cuff/stethoscope auscultation method, within the limits prescribed by the American National Standard Institute, electronic or automated sphygmomanometers. Reference: Association for the Advancement of Medical Instrumentation. American National Standard. Manual, electronic or automated sphygmomanometers ANSI/AAMI SP10-2002.  The followings texts are adapted for a personal use of the VisioCheck  Choice of the arm for the blood pressure measurement:   For the first use, it’s recommended to take the measurement in both arms. If one arm consistently has  higher  blood  pressure  than  the  other  (>  10  mmHg  in  US  or  >  20mmHg  in  Europe),  that  arm should be used to measure your blood pressure except in the event of contraindications (e.g. fistula, ganglion) in this arm. If there is no difference in readings, the readings are often more easily taken in the non-dominant arm (left for right-handed, right for left-handed). Reference: American Heart Association. How to Monitor and Record Your Blood Pressure.  Haute  Autorité  de  Santé.  «  Prise  en  charge  des  patients  adultes  atteints  d’hypertension  artérielle essentielle » - Recommandations pour la pratique clinique - juillet 2005    Self-measurement of blood pressure    -Repetition  is  essential  in  self-monitoring  blood  pressure  as  single  measurements  are  often unreliable for a number of reasons due to huge variations in blood pressure and its susceptibility to factors such as: recent activity, stress and even breathing. Taking 3 consecutive measurements in a row and averaging them provides a more reliable home measurement by a non-professional.  This method comes recommended by  the  American Heart  Association  and the European  Society  of Hypertension.
15 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017  - The rule of three: 3 measurements recommended while seated, at rest, 3 days in a row: In  the  morning  before  taking  medications,  3  consecutives  measurements  in  a  row  (MAM);  In  the evening after taking medications, 3 measurements in a row (MAM).  A single measurement may not be significant especially in terms of blood pressure self-monitoring.  - Blood pressure self-measurement: the ideal targets  for a  person with high blood  pressure  when self-measuring at home are different and lower than those of the American Heart Association (AHA) or European Society of Hypertension (ESH) ). These targets depend on age and can vary depending on your health.  General case: SBP* < 135 mmHg and DBP* < 85 mmHg  * SBP: Systolic Blood Pressure - DBP: Diastolic Blood Pressure  The aforementioned targets are provided for reference only and need to be rechecked by a doctor.  It is important to check your targets in the event of serious disease or pregnancy.   Before taking measurements, - Avoid eating, smoking or exercising for 30 minutes prior. Relax for 15 minutes before taking blood pressure.  - Stress increases blood pressure. Avoid taking measurements during a stressful period - You must take a blood pressure measurement in a calm environment and you must be relaxed and seated with your arm on a table.  - Remain stationary and do not speak while taking your measurement. Note: For more reliable results, take measurements on the same arm. Speak to your doctor if you have any questions about using the blood pressure monitor  Duration of use: The Blood pressure cuff is reusable for 1 year.  5.4. Precaution before using of Thermometer  This device is pre-set at the factory. It is not necessary to calibrate it when starting it up. - In order to obtain reliable and stable results: Each time there is a significant change in the ambient temperature  (due  to  a  change  in  environment),  you  are  advised  to  allow  the  VisioCheck thermometer  to  acclimate  to this  ambient temperature  for  15  to  20  minutes before  using  it.  It  is important to allow a one-minute interval between two measurements.   Practical considerations when taking a temperature: -  In  order  to  ensure  that  precise  and  accurate  temperature  measurements  are  obtained,  it  is essential  that  each  user  has  received  adequate  information  and  training  in  the  temperature measurement technique when using such a device. - It is essential to remember that although procedures such as taking a temperature may be simple, they must be followd correctly. - The patient must not have undertaken vigorous physical activity prior to taking his/her temperature and the room temperature must be moderate.
16 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 - Be aware of physiological variations in temperature which must be taken into consideration when evaluating the results: temperature increases by 0.5C° /1.6°F between 6am and 3pm. Women have a temperature  that  is  higher,  on  average,  by  around  0.2C°/  0.4°F.  Their  temperature  also  varies  in accordance with their ovarian cycle. It rises by 0.5C°/ 1.6°F in the second half of the cycle and at the early stages of pregnancy. - When sitting, temperature is lower by about 0.3° to 0.4°C/ 0.5 to 0.7 °F than when standing. Reference: Haut  Conseil de Santé Publique : Recommandations sanitaires du Plan  national canicule 2014. P123 (French Public Health Council: Sanitary recommendations of the National Heatwave Plan 2014)   6. User guide  6.1. General functions   6.1.1. Start/Shutdown the Device:   Start: Long press the On/Off button until the display screen is on. Shutdown: Long press the On/Off button until the display screen is off   6.1.2. Modules inspection   After  turning  on  the  device,  it  performs  a  module  inspection  which  allow  to  verify  each  module  functioning before launching the application.                                                                                          The status Failed means that the module doesn’t functions properly, you should not use it.    Please contact the customer service.   The status Succeeded means that the module functions properly, it can be used.      1.1.1. Connect the device to internet   You can connect the device to the internet using a M2M sim card or a WiFi connection.
17 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017  Note: Connect the device to internet is required for data sharing (see page 20)  1.1.1.1. Sim card Installation   Insert a small  paper  clip  or the  sim  card eject tool  provided into  the  opening  of  the  SIM  tray,  located  on  the  side  of VisioCheck®. See Diagram A, page 2. Press lightly to bring out the SIM tray and place the sim card as indicated. Close carefully.  Note: M2M sim card provided by your retailer.      6.1.2.2  Wifi Connection     Go to the settings parameters of the system. In wireless and networks > WLAN, turn the button ON. The list of available networks will be displayed. Select your secured network and enter the password.  Go back to the Application, the device is now connected.                                                6.1.3. VisioCheck Settings  The VisioCheck settings are defined for all accounts (admin, user)  6.1.3.1. General In the general settings, you can choose the units in which you want to display your measurements. The Admin can also erase all the previous data saved in the application by clicking on “Erase app data button”.
18 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017     6.1.3.2. Measures In  the  Measures  setting,  you  can  select  the  mode  you  want  to  use  for  the  measurements.  For personal use, we advise you to consult your physician prior choosing the measurements modes.                                                                         For the Blood Pressure Monitor, you can choose between a simple standard measurement or a triple measurement (Mean Arterial Measurement). You can also set a static mode to simulate a range of normal, hypertensive and hypotensive dynamic  noninvasive  BP  in  order  to  verify  if  the  VisioCheck’s  blood  pressure  monitor functions correctly and that the blood pressures values measured are reliable.  You’ll need a blood pressure simulator and tester for that (not provided).
19 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017                                                                 For the ECG, you can select the duration of the measurement by choosing manual (control the recording time is controlled during the measurement) or timer mode (the recording time is preset for 30 seconds).  You can also set the filters in three different positions to control the interference of electric current during the tracing ECG: Off, 50Hz (mostly for France and Europe) or 60Hz (mostly for US)                    For the Oximeter, you can choose between three different modes: Single (the measurement is instantaneous), Manual (you can control the recording time during the measurement) or Timer (the recording time is preset for 30 seconds).
20 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017                                                                  6.1.4. Accounts  The  professional  use  allows  you to  create  and  manage several accounts  (organization,  admin and users accounts)   6.1.4.1. Organization At the first use you need to provide your organization’s informations.                                                        6.1.4.2. Admin account
21 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 You  need  to  create  an  admin  account  which  allow  you  to  create,  modify  or  delete  other  user accounts. To be connected to your admin account, you will need to provide your email and password.                                      6.1.4.3. User accounts As Admin, you can add new user accounts                               To connect to his account, the user will need his login and password.  As admin, you can also modify or delete patient’s files related to other user accounts.  6.1.5. Patient’s file and patient’s profile  The Admin or any user account can create a patient’s file.
22 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 The patient’s file include profile, dashboard, history and share options.                                        You  can  also  have  an  unidentified  patient  file,  allowing  the  user  to  make  measurement  without assigning the result to a specified profile. It allows to use the medical device directly and simply.   To switch to unidentified patient or another patient’s file you can click on the top bar.   6.1.5.1. Patient’s profile  Before  starting  to  monitor  a  patient,  you  can  create  his  patient  file,  and complete his profile with his personal data (medical history, allergy, contact) and choose the visibility of the file. Don’t forget to save all the information to have a complete file.   Thereby all the measurements can be saved in the patient’s file.
23 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 6.1.5.2. Dashboard The dashboard is the last measurements, done for the patient. It also allows to view oximeter and ECG recordings simultaneously                                                                   6.1.5.3. View history  The data history can be viewed by day, week, month or year, by selecting the specific date and the specific device (ECG, BPM, OXI or TEMP)                                           6.1.5.4. Data sharing
24 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 You can share the data of a patient with another health professional, a family member, caregiver. For that, the contact information should be pre-saved in the patient profile.                                                           6.2. Taking a measurement   To take a measurement, go to the main screen of your account and click on the stethoscope icon. Then select the icon of the device you want to use.                                                                           6.2.1. ECG See diagram C page 2  Before taking a measurement
25 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017  It is recommended to be in a comfortable position (lying flat or sitting), your back straight and in a relaxed  position  for  the  full  duration  of  the  measurement.  Make  sure  that  your  chest  is  free  of tension.  Prepare the skin by cleansing and drying the areas where the electrodes will be placed. Typically, 70% alcohol is used for this purpose. Be sure to dry the skin.  If necessary, shave the skin before placing the electrodes. Avoid skin with scars, wounds, lesions, folds or swelling.  It is recommended placing the electrodes facing a visible bone structure. The electrodes pads are disposable and for single use only  WARNING: It is strictly prohibited to attach the lead cable electrodes on the patient’s body if the lead cable is not connected with the device.  Before you begin a measurement, first connect the lead cable with the device, next attach the lead cable  electrodes  to  your body. After taking a  measurement,  first  remove the lead  cable  electrode from the body, second disconnect the lead cable from the device.  1) Connect the ECG cable to VisioCheck  2) Stick the electrodes pads to the correct positions (see picture below).  3) Connect the lead-wires set to the electrodes pads.     -> Red electrode: Right shoulder or right wrist  -> Yellow electrode: Left shoulder or left wrist  -> Green electrode: Left ankle or left spine iliac  -> White electrode: Two fingers under the nipple (for the woman, if the breast is voluminous, under the fold of the breast opposite the left nipple) -> Black electrode: right ankle or right spine iliac
26 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017   Taking a Measurement  Follow the steps to properly measure and record your data 1) Press “start measure” on the screen of VisioCheck to launch the measurement  2)  Click  on  “stop  measure”  to  stop  the  measurement  and  then  “save  measure”,  to  save  the measurement. Remain  still,  and  do  not  flex  or  move  until  the  recording  is  complete.  If  the  device  senses  poor contact  or  the  measurement  is  interrupted  by  a  hand  adjustment,  you  will  need  to  repeat  the recording. After a measurement, you should first remove the lead-wires from the body, second the cable should be disconnected to the device.  You can take ECG and SpO2 measurements simultaneously, and view them in real-time from the Dashboard.     WARNING: When disconnecting the cable, pull on the plug itself not on the cable. Scale:  Calibration of ECG Signal: 1 mV=10 mm ± 1.0 (in ordinates) Graph paper unwinding speed: 1s = 25 mm (in abscissas)
27 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017  Notes about the results  The values for Pulse given by VisioCheck are between 40-180mmHg. Outside this interval, the device display OUT OF RANGE and the measurement should be retake.   6.2.2. SpO2 See diagram D page 2  Before taking a measurement Avoid eating, smoking or exercising for 30 minutes before measuring your oxygen saturation. Remain calm for 15 minutes before the measurement  1) Plug the adapter cable into VisioCheck SpO2 port 2) Connect the SpO2 adapter cable to the SpO2 Sensor connector. 3) Insert an index finger over the sensor window with the fingernail facing upwards and the fingertip against the end. 4) Spread open the rear tabs of the sensor to provide even force over the length of the pads.  5)  Check  the  position  of  the  sensor.  If  an  index  finger  cannot  be  positioned  correctly  or  is  not available, a smaller finger can be used. Do not use on the thumb, toe, or across a child’s hand or foot. Note:  When  selecting  a  sensor  site,  priority  should  be  given  to  an  extremity  free  of  an  arterial catheter, blood pressure cuff, or intravascular infusion line. The sensor should be oriented in such a way that the cable is positioned along the top of the hand. See picture below.
28 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017                                         Taking a Measurement  Follow the steps to properly measure and record your data 1) Press “start measure” on the screen of VisioCheck to launch the measurement. 2)  Click  on  “stop  measure”  to  stop  the  measurement  and  then  on  “save  measure”,  to  save  the measurement. Remain still, and do not speak, flex or move until the recording is complete. If the device senses poor contact  or  the  measurement  is  interrupted  by  a  hand  adjustment,  you  will  need  to  repeat  the recording. For the value to be reliable,  the  blood  supply index (signal strength of the arterial pulse) must  be superior to 0.3%.  You can take ECG and SpO2 measurements simultaneously, and view them in real-time from the Dashboard.    WARNING: When disconnecting the cable, pull on the plug itself not on the cable.   Note: If the sensor does not track the pulse reliably, the blood flow may be too thick, thin, or deeply pigmented to permit appropriate light transmission. If any of these situations occur, reposition the sensor or choose an alternate sensor.
29 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017  Notes about results: Many factors can modify hemoglobin’s affinity for oxygen and reduce oxygen saturation in the blood. Measuring oxygen saturation is vital for individuals who may be at risk of desaturation, a condition that could result in the need for medical treatment to address the deficiency.  Warning:  This  table  DOES  NOT  apply  to  individuals  already  suffering  from  certain  illnesses  (e.g. asthma, heart disease, respiratory diseases) or who are at altitudes above 1500 meters. If you are already suffering from a disease, always consult your doctor to analyze your results.
30 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 The values for oxygen saturation given by VisioCheck are between 70-99 mmHg. Outside this interval, the device display OUT OF RANGE and the measurement should be retake. The values for Pulse given by VisioCheck are between 40-180mmHg. Outside this interval, the device display OUT OF RANGE and the measurement should be retake.   6.2.3. Blood pressure See diagram E page 2  Before taking a measurement  - Sit down on a chair with your feet flat on the floor (do not cross your legs). - Apply cuff to arm so that the orange mark on the cuff is over branchial artery. The cuff must be 0.4 – 0.8 inches / 1-2 cm from elbow crease. - Wrap cuff around arm. Make sure the white side line folds within the arrow box, if it does not, change to a bigger or smaller cuff that fits better. - Close the cuff with the Velcro (not too tight). - Put your arm on the table, palm up so that the blood pressure cuff is level with the heart.  - Remain seated and calm for 5 to 15 minutes before taking measurement - Do not take measurements over clothing. Remove jewelry from the arm you are measuring before taking measurement (watch, ring, bracelet etc.).  - Avoid eating, smoking or  exercising 30 minutes  before taking  measurements.  Relax for  15 minutes before taking blood pressure
31 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017                                       Taking a Measurement in real time  1) Press “start measure” on the screen of VisioCheck to launch the measurement.  The  armband  inflates  automatically.  Do  not  remove  the  armband  before  the  end  of  the measurement.  2) At the end of the measurement, click on “save measure”, to save the measurement.   Notes:  - For more reliable results, always take measurements on the same arm for the same patient - Only use the appropriate size cuff when the artery  index marker falls within the printed range. Otherwise erroneous readings may result -The measurement will be too low if the armband is higher than the heart or too high if it is lower than the heart.    Notes about results:
32 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 In 2013, the American Heart Association (AHA) defined a blood pressure classification shown in the table below. Nevertheless, this classification is just a general guide, as blood pressure varies from one person to another, based on age, weight, and health. This blood pressure chart reflects categories defined by the American Heart Association     The values for Blood Pressure given by VisioCheck are between 0-297 mmHg. Outside this interval, the device display OUT OF RANGE and the measurement should be retake. The values for Pulse given by VisioCheck are between 40-180mmHg. Outside this interval, the device display OUT OF RANGE and the measurement should be retake.   6.2.4. Body temperature See diagram F page 2  Before taking a measurement
33 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 Please take the following precautions before any measurement to ensure a stable and reliable result: - Push back hair from the forehead. - Wipe away any perspiration from the forehead. - Avoid any drafts (e.g. from nasal specs, air conditioning, etc.). - Each  time  there  is  a  significant  change  in  the  ambient  temperature  due  to  a  change  in environment, allow the thermometer to acclimate to this ambient temperature for at least 15 minutes before using it. - Do not put the thermometer in contact with an open wound - Do not drink hot or cold drinks, or exercise while taking a temperature measurement  Taking a Measurement   1) Aim at  the forehead, over the  region center, from a  distance  of  about 2-5cm  / 1-2 inches,  see picture below 2) Press and release the “OK” button on the right side of VisioCheck.  3) Wait for the BEEP sonore without moving. The temperature is instantly displayed on the screen. 4) Click on “save measure” to save the measurement.                                                 The reliability of the measurement cannot be guaranteed if the temperature is measured over another part of the body (e.g. arm, torso…).  Note about results  Diagnostic of temperature readings
34 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 The values for temperature given by VisioCheck are between 32°-43°C /89.6 -109.4 °F. Outside this interval, the device display OUT OF RANGE and the measurement should be retake.   7. HD Camera  To be defined   8. Storage and maintenance  Do not pour or spray any liquid on the surface of the device or its accessories; Make sure that liquid does not come into contact with any of the device openings. Do not  sterilize the device or its accessories in an  autoclave, with ethylene oxide, by radiation, by steam,  by  high pressure  or  by  immersing  it  in  a  liquid.  The  device  or  its  accessories  must  not  be sterilized.  Protect the device or its accessories from UV radiation WARNING: Never immerse the device or its accessories in any liquid.   VisioCheck - Avoid  extreme  changes  in  temperature  and  humidity.  Do  not  use  this  device  in  locations subject to high or low temperatures or humidity.  - Do not store the device in ambient conditions such as exposed to direct sunlight, with high temperatures and/or high humidity, that are wet or damp or where water may get on the device, that  are  dusty  near fires  or  open  flames  or  exposed  to  strong  vibration,  or  strong electromagnetic fields.
35 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 - Dispose of the device in accordance with applicable local regulations. - This device is designed to be compliant with the rules and regulations in locations which it is sold and will be labeled as required. - Any changes or  modifications to this  device, not  expressly approved by the manufacturer, will void the user’s authority to operate the equipment   Storage Conditions -10°C~+50°C (14°F ~122°F), 10~95%RH, 50kPa~106kPa(Non condensing)  Working conditions 5°C~+40°C (41°F~104°F), 15%~95%, 70kPa~106kPa(Non condensing)   -  Clean the device using a medical disinfectant wipe only.  - Do  not  use  water,  or  substances  such  as  benzene,  gasoline,  paint  thinner,  concentrated alcohol, or detergents. Do not clean other parts of the device with any liquid. - Do not drop or hit this device. -  Do not disassemble the device.  ECG cable/lead-wires, SpO2 adapter cable  Cleaning 1) Disconnect the cables from the monitoring device. WARNING: When disconnecting pull on the plug itself not on the cable. 2) Dampen  a  clean  cloth  or  gauze  pad  with  an  appropriate  cleaning  solution  and  wipe  all exposed surfaces. 3) Dampen a clean cloth or gauze pad with sterile or distilled water and again wipe all exposed surfaces. 4)  Dry all exposed surfaces with a clean, dry cloth or gauze pad.  NOTES:  Do  not  treat  the  cables  with  oils  (e.g.  household  or  lubricating)  or  aggressive  fluids  or solvents (e.g. acetone).  The cable is not made with natural rubber latex.  Disinfection Disinfect with a chemical disinfectant, such as ethanol, propanol, phenolic disinfectant.   Pulse oximeter probe  Cleaning The sensor may be surface cleaned by wiping it with a solution such as 70% isopropyl alcohol.   To clean or care for the sensor: 1) Disconnect the pulse oximeter probe from the SpO2 adapter cable 2) Soak a clean, dry gauze pad with the cleaning solution. Wipe all surfaces of the sensor and cable with this gauze pad. 3) Soak another clean, dry gauze pad with the sterile or distilled water. Wipe all surfaces of the sensor and cable with this gauze pad. 4) Dry the sensor and cable by wiping all surfaces with a clean, dry gauze pad
36 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 Disinfecting If low level disinfection is required, use a 1:10 bleach solution. Do not use undiluted bleach (5%-5.25% sodium  hypochlorite)  or  any  cleaning  solution  other  than  those  recommended  here  because permanent damage to the sensor could occur.   Blood Pressure cuff  Cleaning  1) Disconnect the BP cuff from the monitoring device 2) Clean the cuff by wiping a damp cloth with mild soapy and water  Warnings: Do not sterilize cuffs. Do not use bleach. Wash and rinse; temperatures should not exceed 104 F/ 40°C  Disinfecting The cuff can be disinfected by 70% isopropyl alcohol. After disinfection, the cuff cover should be left out to air dry. Warnings: 1)  When  cuff  is  not  connected  to  monitor,  always  fit  the  cap  to  the  tubing  to  prevent accidental entry of liquid. 2) Periodically inspect and replace damaged or deteriorated cuffs.    Thermometer  Cleaning - Clean the surface of the glass over the lens with a cotton ball lightly moistened with 70% isopropyl alcohol.  9. Specifications   1.1. Technical specifications Vital Signs Temperature (Non-contact), SpO2 (Probe type-internal), BP (Inflation), ECG (5 leads-internal) Model BW-XO7HD  Brand Bewell Connect  Classification  Class IIa (for Europe) / Class II (for US) CPU Chipset  Qualcomm MSM8909  Type  Cortex A7  Number of cores  Quad Core  Frequency  1.1GHz  GPU Chipset  Adreno 304  Number of cores  Quad Core   Frequency  400MHz
37 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 RAM Capacity  1GB  ROM Capacity  8GB  Operating system OS Version  Android 5.1  Language  Multi-language  LCD Type  HFFS  Screen size  10“/ 3.97inch  Color  16M  Brightness  400cd/m²  Screen resolution  480*800  Touch panel Type  5 Points capacitive touch screen  Material  G+F+F  Camera Front Camera  5.0MP AF  Rear Camera  No  Flash light  Yes  Speaker Built-in  8Ω/0.5W×1  MIC Built-in  Yes  Receiver Built-in  No  G-sensor Built-in  Yes Gyro-Sensor Built-in  No  Light-Sensor Built-in  No  Proximity Sensor Built-in  No  Grade of Waterproofing IP22 1.2. Blood Pressure Measuring Mode Upper arm type  Measuring Method Inflation+MAP (Mean arterial pressure)  Measurement Range Pressure: 0-295mmHg (0-39.3kPa)  Pulse Rate: 40-180 times/min Measurement Accuracy Pressure: ±3mmHg (±0.4kPa)  Pulse Rate: ±5%  1.3. SpO2 Measuring Mode Finger type (Internal)  Measuring Method Infrared & Absorption Spectrometry  Measurement Range Oxygen Saturation: 0%~100%  Pulse Rate:25~250 times/min PI: 0.05%~20%  Measurement Accuracy Oxygen Saturation: 70%-100%: ±2% 0%-69%: Accuracy is not defined  Pulse Rate: ±3 times/min PI: Accuracy is not defined  1.4. ECG Measuring Mode 5 Electrodes,7-Lead ECG  ECG Lead Ⅰ、Ⅱ、Ⅲ、 aVR、aVL、aVF、V  DC offset Voltage Range ±300mV  Differential  Voltage Measurement Range ±5mV  Voltage  Measurement Accuracy ±5%
38 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 Anti-electrosurgical Capability No  Defibrillation Protection No  PACE No  1.5. Temperature Measurement range Body temperature:32.0°C~43.0°C / 89°F ~109.4°F Room temperature: 0°C~50°C / 32°F~122°F Surface temperature: 0°C~90°C / 32°F~194°F Measurement speed 1s-2s   Measurement site Forehead Measurement distance 2-5 cm / 1-2 inches Display resolution  0.1°C  / 0.1°F Measurement Accuracy +/- 0.2°C (35°C-42°C) / +/- 0.2°F (95°F-107.6°F) in body temperature +/-0.3°C (32°C -34.9°C, 42.1°C-43°C) / +/- 0.3°F (89.6°F-93.2°F, 107.6°F-109.4°F) in body temperature. +/- 0.2°C (22°C -40°C ) / +/- 0.2°F (71,6°F-104°F); others +/-2°C  / +/-2°F in surface temperature. Unit °C/°F 1.6. Wireless Communication  Wi-Fi  Built-in  Yes  Cryptographic algorithms  WFA, WPA/WPA2  Version  802.11b/g/n  Bluetooth  Built-in  Yes  Version  4.0  GPS  Built-in  Yes  A-GPS  Yes  3G(WCDMA)  Frequency band  B1, B2  LTE-FDD  Frequency band  B1, B2, B3, B4, B7, B12, B20 Data transmission speed  Upload  50Mbps (4G mode)  Download  150Mbps (4G mode)  Phone feature  Phone calls & message  No  FM Radio  No  NFC  No  1.7. Key and Interface Earphone  3.5mm CTIA standard headphone jack  Buttons  Home Key; Power key; Up/Down Wheel (Up/Down, one click confirm)  Micro USB  Charge  BP port  Yes
39 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 SpO2 probe port  Yes  ECG cable port  Yes  Non-contact  temperature sensor  Yes  Micro Sim Card  Yes  Micro SD  No  1.8. Power Parameter  Battery Type  Li-ion battery  Capacity  3.7V/2500mAh  Time on stand by  3 days  Performance capacity  4 hours  Charge lamp Blue/Green Adapter Information Temperature Operating: The adapter is capable to operate from 32° to 104° Fahrenheit / 0°C to 40°C.  Non-  Operating:  The  adapter  is  capable  to  store from -4° F - 140°F / -20°C to 60°C Humidity Operating: The adapter is capable to operate from 10% to 90% RH.  Non-  Operating:  The  adapter  is  capable  to  store from 5% to 90% RH.  INPUT: 100-240V~50/60Hz,500mA OUTPUT: 5.0V  2.0A 1.9. Appearance Body Material PC+ABS  Color White  Size 172*87*30 mm (length*width*height) About 6.77in*3.43in*1.16 inches Weight 298g / 10.5oz (without accessories) 1.10. Life time Life time of button 50 000 times pressed Life time of BPM Bump and Valve 30 000 times Battery life 500 times of full Charging (The capacity will decrease 10% each year) Continuous usage With Battery 100% Charged, BPM: 50 times; SPO2: 5 hours; ECG: 5.5 housrs. Period of  5 years 1.11. Accessories included  - Carrying pouch, including:  o VisioCheck BW-XO7HD o Cable and cuff for BPM (22~42 cm / 8.5~16.5 in) o Pulse oximeter SpO2 probe  o SpO2 adapter cable (50cm / 20 inches) o ECG cable 5 electrodes / 7 leads
40 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 o AC Adaptor 5V 2A with EU/US/UK plug, USB cable 1 m / 3.3 ft - Pouch of 10 disposable adhesive electrodes for ECG - PP bag including: o Quickstep o User manual o Warranty card with serial number o Sim card removal tool o Earphone with microphone 1.12. Other Accessories (not included) Item NO. DESIGNATION BW-XA8-13 cuff for BPM-infant (8cm-13cm) BW-XA12-19 cuff for BPM- Children (12cm-19cm) BW-XA17-25 cuff for BPM-Medium (17cm-25cm) BW-XA22-36 cuff for BPM-standard(22cm-36cm) BW-XA22-42 cuff for BPM-large(22-42cm) BW-XA32-52 cuff for BPM-X-large(32cm-52cm) BW-XSPOP Pediatric Pulse oxymeter SpO2 probe with extension cable (50cm / 20 inches) BW-XSPOS Pulse oxymeter SpO2 probe  BW-XSPOC SpO2 adapter cable (50cm / 20 inches) BW-XDC2A AC Adaptor 5V 2A with EU/US/UK plug, USB cable 1 m BW-ECG5/7 ECG cable 5 electrodes / 7 leads BW-XTM1 Professionnal table stand for VisioCheck BW-XCASE  Silicone Protection holster with lanyard and cart stand BW-XTP1 Table stand for VisioCheck   BW-XTCART VisioCheck professional cart with wheel BW-XT001 Telescopic  mobil  stand  with  2  International  AC  adaptor,  1  USB  - power cord 2 m, wheel with security lock BW-XT002 Basket for cart BW-XT003 Handle for cart BW-XT004 Tray for cart BW-XT005 Cart stand for VisioCheck   No part of this manual may be copied, reproduced or translated into another language without prior written consent  of Visiomed.  Visiomed reserves the  right to revise this  guide at  any time without prior notice.
41 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017    2.  Troubleshooting   Main device 1.  Device can’t be turned on.  System is abnormal.  Battery has low power. Device is damaged. Restart the device. Charge the device. Contact customer service. 2.  Module Inspection Failed The module concerned has a problem Device malfunction Restart the device  If restarting can’t solve the problem, contact customer service.  3.  After turning on the device, 4G/Wi-Fi signal is abnormal. Network signal is weak. Wi-Fi hotspot is configured incorrectly.  Device malfunction. Be patient and wait. If there is still no signal, try to restart the device. If restarting can’t solve the problem, contact customer service. If the communication mode is Wi-Fi, check whether the Wi-Fi hotspot (router) is configured correctly. 4.  Device crash Device malfunction Stop measuring and unplug the BPM cuff, SpO2 cable and ECG cable. Long press the on/off button to shut down. If the device can’t be shutdown normally, try to press the on/off button for 15s to force restart. Contact customer service. 5.  Device doesn’t recognize SIM card.    SIM card is connected poorly.    Device malfunction.   Check if SIM card is plugged in correctly   Take out the SIM card and plug in again. Contact customer service. 6.  Speaker has no sound.    The setting of volume is too low.    Device malfunction.   Reset the volume in system setting.   Contact customer service. 7.  Earphone is abnormal (no   Earphone has a poor connection     Plug the earphone correctly.
42 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 sound/ with noise/ abnormal recording)     Using earphones that are not in match with the device.    Use the earphone matching to the device. Contact customer service. Oximeter module 8.  SpO2 probe malfunction  Probe is not entirely plug in the device. Probe has a problem. Device malfunction Take out the probe and plug in again, then re-measure. Change with a new probe and re-measure If the problem persists, restart the device. If problem can’t be solved, contact customer service. 9.  No pulse rate found  The measured site has novital signs. Measuring in wrong operation. Device malfunction Re-measure according to the user manual If still no vital signs after several tries, contact customer service. 10.  Weak perfusion index PI of the measured site is low, lower than 0.3% may cause inaccuracy. Change measuring site, and re-measure according to the user manual. 11.  Light interference  The test environment has strong light  Re-measuring in weak light environment. 12.  Oximeter module fails to start. Firmware malfunction. Restart the device. If the problem persists, contact customer service BPM module 13.  Measure overtime  cuff being worn the wrong way. Device malfunction. Remeasure according to the user manual. If the problem persists, contact customer service 14.  Measured result is low Device malfunction Remeasuring according to the user manual. If the problem persists, contact customer service 15.  Pressurize overtime Cuff dropping, loose or deflating Restart the device If the problem persists, contact customer service 16.  Deflate overtime Valve damage Restart the device If the problem persists, contact   customer service 17.  Pulse detecting overtime  Wearing the cuff the wrong way. The measured person is wearing thick clothing. Device malfunction. Wear the cuff according to the user manual. Undress the thick clothes and do the measure a second time.  If the problem persists, contact
43 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 customer service 18.  Interference in measuring process The measured person moves or talks.  Do again the measuring according to the user manual. 19.  BPM module fails to start Firmware malfunction Restart the device If the problem persists, contact customer service ECG monitor 20.  ECG malfunction Hardware malfunction (chip does not respond for 5s) Chip error of ECG  Data frames lost Restart the device If the problem persists, contact customer service  21.  Weak ECG signal  Electrodes stuck in the wrong position. Re-connect the cable with device according to the user manual, and restart the device for measurement. 22.  Measuring result of heart over the limit Measuring result of heart over the limit To measure again according to user manual If still not normal, please contact your doctor. 23.  ECG interference Interference signal exist. Be quite and breath smoothly during measurement. Check and remove the interference signal around the cable and electrodes. Re-connect the cable with device and measure again. 24.  ECG fails to start Hardware malfunction Restart the device If the problem persists, contact customer service Body temperature module 25.  Temperature module malfunction  Open circuit or bad connect on the cable of infrared sensor. Infrared sensor is abnormal. Battery voltage lower than 3.2V. Communication protocol is wrong. Reading data overtime Restart the device Please charge the device and     restart. If the problem persists, contact customer service 26.  Abnormal body temperature measurement. Big temperature difference between room temperature and device temperature.  Room temperature beyond the range of 41 – 104 Fahrenheit / 5℃-40℃ Please allow the device to acclimatize the room temperature for at least 30min before using it. Please measure the body temperature within 41-104
44 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017 during body temperature measuring, Fahrenheit / 5℃- 40℃ of room temperature. 27.  Temperature module fails to start Hardware malfunction Restart the device If the problem persists, contact customer service
45 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017
46 BewellConnect - VisioCheck - BW-X07HD – User manual – 022017