Beijing M and B Electronic Instruments MB-MSA100BT Digital Transmission System User Manual

Beijing M&B; Electronic Instruments Co., LTD Digital Transmission System

User Manual

BeijingM&BElectronicInstrumentsCo.,Ltd.
April2016
MSA100BTPeakFlowMeter
InstructionManual
Version:V1.2
PLEASEREADALLOFTHEINFORMATIONINTHEPACKAGEINSERTBEFOREUSINGTHEPEAKFLOWMETER.
IFYOUDONOTUNDERSTANDTHEINSTRUCTIONS,CONSULTM&B,CALL(+86)1061253803(WorkingTime:
fromMondaytoFriday,8:30~17:30),ORWRITETOmarket@mbelec.com.
MSA100BTPeakFlowMeterInstructionManualV1.2

I
Date MeasurementsRecommendationSignatureofthe
physician
PEFFEV1
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
MSA100BTPeakFlowMeterInstructionManualV1.2

II
Aboutthismanual
Firstly,thankyouverymuchforusingourM&Bproduct.
ToensurecorrectuseofMSA100BTPeakFlowMeter(hereinbelowmentionedasMSA100BT
orPeakFlowMeterortheMeter),readthefollowingpointscarefullyandadheretothem.After
readingthismanual,keepitinasafeplacewhereitcanbereferredtoanytimeaquestionarises.
ModelNumber:MSA100BT
ProductName:PeakFlowMeter
Manufacturer:BeijingM&BElectronicInstrumentsCo.,Ltd.
Address:Room6319,Building1,No.27,YongwangRoad,DaxingBioengineeringandMedicine
IndustryBase,ZhongguancunSciencePark,DaxingDistrict,Beijing
Issueddate:April,2016,Version:V1.2
Intellectualproperty
TheintellectualpropertyofMSA100BTanditsmanualbelongtoBeijingM&BElectronic
InstrumentsCo.,Ltd.(hereinbelowmentionedasM&B).Copying,reproductionortranslationof
alloranypartofthecontentsofthismanualwithoutM&B’spermissionisstrictlyprohibited.All
rightsreservedbyM&B.
Statement
M&Bhasthefinalrightstointerpretationforthismanual.
M&Bisresponsiblefortheproductssafety;reliabilityandperformanceifbelowconditions
aresatisfied:
Assembly,extensions,modificationsorrepairarecarriedoutbypersonsauthorizedby
M&B.
TheotherrelatedelectricalequipmentusedtogetherwiththeMeterconformstoCE
standards.
MSA100BTisusedinaccordancewiththeinstructionsmanual.
MaintenanceService
Freeservicerange:
FreeservicesareofferedinaccordancewithM&Bwarrantyregulations.
Feebasedservicerange:
MSA100BTPeakFlowMeterInstructionManualV1.2

III
BeyondM&Bwarrantyregulations.
Duringthewarrantyperiod,warrantyshallnotextendtothefollowingconditions:
1) Itisdamagedfromimproperuse.
2) Batteryvoltageisbeyondthescopeofproductspecification.
3) Naturaldisasters.
4) Replacedaccessories,consumableswithoutM&Bauthorization.
5) ModificationorrepairbyanyoneexceptforBeijingM&Bauthorizedpersonorcompany.
Contactus
Servicetelephoneandproductsupport:
PleasecontactM&Bifyouhaveanyquestionsregardingproductsupport.
+861061253803(WorkingTime:Monday~Friday8:30~17:30);
Orcontacttheregionalservicepartnerordistributor.
Orderconsumablesandrepairparts
Pleasepurchaserepairparts(mouthpieceetc.)fromtheM&Bservicepartnerordistributorin
yourcountry.
Otherquestions
Pleasecontactusasbeloworvisitourwebsitewww.mbelec.com.
Headquarters
Company:BeijingM&BElectronicInstrumentsCo.,Ltd.
Address:Room6319,Building1,No.27,YongwangRoad,DaxingBioengineeringandMedicine
IndustryBase,ZhongguancunSciencePark,DaxingDistrict,Beijing
InformationforTheEuropeanUnionrepresentative:
EuropeanUnionRepresentativeName:ShanghaiInternationalHoldingCorp.GmbH(Europe).
EuropeanUnionRepresentativeAddress:Eiffestrasse80,20537Hamburg,Germany.
MSA100BTPeakFlowMeterInstructionManualV1.2

IV
CAUTION:Whenthepeakflowmeterisusedtowatchlungconditionssuchasasthma,theuser
shouldbeunderthecareofaphysicianorotherlicensedhealthcareprofessional.Alicensed
healthcareprofessional’sadviceisrequiredtounderstandthemeaningandimportanceofthe
measuresyougetwithyourpeakflowmeterandtodecideonanappropriatetreatmentplan.
Thetreatmentplangiventoyoubyyourphysicianorotherlicensedhealthcareprofessionalwill
tellyouwhatactiontotakewhentherearechangesinyourpeakflownumber.
Nomatterwhatyourpeakflowmeasuresare,ifyouhavesignsandsymptomssuchaschest
tightness,shortnessofbreath,coughingorwheezingyoushouldfollowyourlicensedhealthcare
professional’sadviceforcontactinghimorher.
Instructionsforusingthepeakflowmetermustbefollowedcarefullyinordertogetcorrect
measureofairflow.
Ifyouareunabletoobtainareadingyoushouldcontactyourphysicianimmediately.
Thepeakflowmetershouldnotbeusedbymorethanoneperson.
Ifyouhaveanyquestionsaboutyourpeakflowmeteranditsuse,talktoyourlicensedhealthcare
professional.
MSA100BTPeakFlowMeterInstructionManualV1.2

I
Content
1BRIEFINSTRUCTION.....................................................................................................................................‐1‐
1.1FunctionDescription........................................................................................................................‐1‐
1.2IntendedUse....................................................................................................................................‐1‐
1.3IndicationsforUse...........................................................................................................................‐2‐
1.4MeasuringPrinciple.........................................................................................................................‐2‐
1.5Step‐by‐stepInstructions...............................................................................................................‐2‐
1.6TechnicalSpecification.....................................................................................................................‐3‐
1.7OtherSpecification..........................................................................................................................‐3‐
1.8CompositionofPackage..................................................................................................................‐3‐
1.9TransportationandStorageConditions..........................................................................................‐4‐
1.10OperatingCondition........................................................................................................................‐4‐
1.11DeviceComposition.........................................................................................................................‐4‐
2SAFETYPRECAUTIONS.................................................................................................................................‐5‐
2.1SafetyInformation...........................................................................................................................‐5‐
2.1.1TermsandSymbols..............................................................................................................‐5‐
2.1.2Environment.........................................................................................................................‐5‐
2.1.3Preparation..........................................................................................................................‐5‐
2.1.4OtherNotices.......................................................................................................................‐5‐
2.2TransportationandStorage.............................................................................................................‐6‐
2.3InterferenceInstruction...................................................................................................................‐6‐
2.4Classification....................................................................................................................................‐6‐
2.5SafeOperationandProcessingConditions.....................................................................................‐6‐
2.6StandardsandRegulationCompliance...........................................................................................‐7‐
3MAINTENANCE............................................................................................................................................‐8‐
3.1BatteryAssemblage.........................................................................................................................‐8‐
3.2AssembleMouthpieceandTurbineintoMainUnit.......................................................................‐8‐
3.3Cleaning............................................................................................................................................‐8‐
4OPERATION................................................................................................................................................‐10‐
4.1LCDScreenDisplayIntroduction...................................................................................................‐10‐
4.2FirstTimeUse................................................................................................................................‐11‐
4.3TimeSetting...................................................................................................................................‐11‐
4.4Poweron/Poweroff......................................................................................................................‐12‐
4.5Measuring......................................................................................................................................‐12‐
4.6ImportantInformationforMeasuringandMonitoringofPeakFlowValues..............................‐14‐
4.7SimpleAssessment(TrafficLightScheme)....................................................................................‐16‐
4.8RecordsPlayback...........................................................................................................................‐16‐
4.9Deleterecords................................................................................................................................‐17‐
4.10DataTransmission..........................................................................................................................‐17‐
5TROUBLESHOOTING..................................................................................................................................‐19‐
6GUIDANCEANDMANUFACTURE’SDECLARATIONS.................................................................................‐20‐
7TECHNICALSPECIFICATION........................................................................................................................‐23‐
MSA100BTPeakFlowMeterInstructionManualV1.2
‐1‐
1 BRIEFINSTRUCTION
1.1 FunctionDescription
MSA100BTPeakFlowMeterisahandheldpulmonaryfunctionmeasuringmedicaldevice
thatmeasuresyourmaximumpossibleexhalationwhichiscalledpeakexpiratoryflow(PEF)and
forcedexpiratoryvolumein1second(FEV1).TheaccuracymeetsAmericanThoracicSociety(ATS)
Standard2005Revision.
Apeakflowmeterisusedtomeasureaperson’s“peakexpiratoryflow”,whichisasimple
measureofairflowthatcantellyouhowwellyouarebreathing.Ittellsyouhowwellairismoving
throughtheairwaysinyourlungs.ForcedExpiratoryVolume(FEV1)isameasureofthevolumeof
airexpelledin1second.Ifyouhaveabreathingconditionsuchasasthma,C.O.P.D.,orchronic
bronchitis,yourphysician(orotherlicensedhealthcareprofessional)mayrecommendthatyou
useapeakflowmetertowatchforchangesinyourairflow.Yourphysician(orlicensedhealthcare
professional)willgiveyouatreatmentplanthatwilltellyouwhatactionstotakewhenyouhavea
changeinairflow.Inaddition,youshouldrecordyourpeakflowmeasuresasrecommendedby
yourphysician(orotherlicensedhealthcareprofessional).Reviewingpeakflowmeasurescan
helpyouandyourphysician(orlicensedhealthcareprofessional)checkscloselyonyourasthma
orC.O.P.D.toprovidethebesttreatmentforyou.
MSA100BTissuitableforchildrenthroughadultswhoarecapableoffollowingthe
instructionsforuse.BecausetheMeterhasanautomaticmemorythreehundredsreadings,you
cantaketheMeterwithyouthenexttimeyouvisityourdoctorforareviewofmanyreadings.The
doctormakesanappropriatetreatmentplanbasedonthem,andhelpspatientstorecuperate
health.
TheMSA100BTcanconnecttocellphonebyBluetooth.Youcantransfertherecordsto
phoneviaBLE4.0,thenusingapptohelpsaveandtracklongterm.
Thepeakflowmetershouldnotbeusedbymorethanoneperson.
1.2 IntendedUse
MSA100BTisintendedformonitoringPEF(PeakExpiredFlowRate)andFEV1(Forced
ExpiratoryVolume)forpatientuseathome.Thedeviceisdesignedforpediatrictoadultpatients.
WhentheMeterisusedtomonitorlungconditionssuchasasthmaorC.O.P.D.,theusershouldbe
MSA100BTPeakFlowMeterInstructionManualV1.2

‐2‐
underthecareofalicensedhealthcareprofessional.Alicensedhealthcareprofessional’sadviceis
requiredtounderstandthemeaningandimportanceofmeasurementsreportedbytheMeterand
howtodecideonanappropriatetreatmentplan.Thistreatmentplanwilltellthepatientwhat
actiontotakewhentherearechangesintheirpeakflowmeasurements.
1.3 IndicationsforUse
ThisdeviceisintendedformonitoringPEF(PeakExpiredFlowRate)andFEV1(Forced
ExpiratoryVolumeinonesecond)forpatienthomeuse.Thedeviceisdesignedforpediatricto
adultpatients.Thedeviceisintendedformonitoringrespiratoryconditionssuchasasthma.
1.4 MeasuringPrinciple
Whenairflowpassesthroughturbinesensor,itmakestheleafofturbinerotate.Rotary
speedoftheleafislinearrelationshipwithairflowrate.Bythetechnologyofphotoelectric
conversion,turningoftheleafisconvertedtobepulsesignal,whichisprocessedbythemain
controlleraftersampling.ThePEFandFEV1valuearecalculatedbyanalgorithmformulaand
displayedontheLCD.
1.5 Step‐by‐stepInstructions
PLEASEREADALLOFTHEINFORMATIONINTHEPACKAGEINSERTBEFOREUSINGTHEPEAKFLOWMETER.
IFYOUDONOTUNDERSTANDTHEINSTRUCTIONS,CONSULTM&B,CALL(+86)1061253803(WorkingTime:
fromMondaytoFriday,8:30~17:30),ORWRITETOmarket@mbelec.com.
ASKYOURPHXSICIANOROTHERLICENSEDHEALTHCAREPROFESSIONALTOWATCHYOUUSETHEPEAKFlOW
METER.THISWILLHELPASSURETHATYOUAREUSINGITCORRECTLY.
NOMATTERWHATYOURPEAKFLOWMEASURESARE,IFYOUHAVESIGNSANDSYMPTOMSSUCHASCHEST
TIGHTNESS,SHORTNESSOFBREATH,COUGHINGORWHEEZINGYOUSHOULDFOLLOWYOURLICENSED
HEALTHCAREPROFESSIONAL'SADVICEFORCONTACTINGHIMORHER.
IFYOUAREUNABLETOOBTAINAREADINGYOUSHOULDCONTACTYOURPHYSICIANIMMEDIATELY.
a. Getthepeakflowmeterready(takeoutofcase,insertmouthpiece,positionindicator;
directionswilldifferaccordingtodevice)
b. Standup
c. Takeadeepbreathin
MSA100BTPeakFlowMeterInstructionManualV1.2

‐3‐
d. Placethemeterinthemouthandcloselipsaroundthemouthpiece.Makesureyourlips
formatightsealaroundthemouthpiece.Donotputyourtongueintothemouthpiece.Do
notbendyourneck.
e. Blowoutashardandfastasyoucan.
f. Readthenumbernexttotheindicator.Thisisyourpeakflowmeasure.Writeitdown.
g. Afteryoumovetheindicatorbacktozero,repeatthestepstwomoretimestogettwo
morepeakflowmeasures.
h. Choosethehighestnumberofyourthreepeakflowmeasures.Writethenumberinyour
“PeakPlowDiary."Directionsforthediaryarewrittenonthediary.
i. Howtostoreyourpeakflowmeter(directionswilldifferaccordingtodevice).
1.6 TechnicalSpecification
1. Measuringmethod
FlowTurbinesensor
VolumeFlowIntegration
2. Measuringrang
Volume0.01L~9.99L
Flow50L/min~900L/min
3. Accuracy
Volume±3%or±0.05L,whicheverisgreater
Flow±10%or±20L/min,whicheverisgreater
4. DisplaySegmentliquidcrystaldisplay
5. Memory300measurementswithdataandtime
1.7 OtherSpecification
TypeBFappliedpart
Caution
Refertoinstructionmanual/booklet
ModelMSA100BT
PowersourceAAA1.5V×2alkalinebatteries
1.8 CompositionofPackage
No.NameUnitQuantityNote
MSA100BTPeakFlowMeterInstructionManualV1.2

‐4‐
1Mainequipmentpiece1MSA100BT
2Turbinesensorpiece1
3Mouthpiecepiece1
4Instructionmanualpiece1
1.9 TransportationandStorageConditions
1. Environmenttemperatureof‐10degC~+55degC;
2. Relativehumidityof≤93%;
3. Atmosphericpressureof500hPa~1060hPa.
1.10 OperatingCondition
1. Environmenttemperatureof+10DegC~+40DegC;
2. Relativehumidityof25%~85%;
3. Atmosphericpressureof860hPa~1060hPa.
1.11 DeviceComposition
MSA100BTPeakFlowMeterisdividedintopartsbelowlikeFigure11.
1. Mainequipment.
2. AAA1.5×2alkalinebatteries.
3. Mouthpieceandturbinesensor.
Figure11MakeupofMSA100BTPeakFlowMeter
MSA100BTPeakFlowMeterInstructionManualV1.2
‐5‐
2 SAFETYPRECAUTIONS
2.1 SafetyInformation
2.1.1 TermsandSymbols
Asthemanufacturerofmedicaldevices,thecompanystrivestoensurethesecurity,reliability
andperformanceoftheinstrument.Forcorrectuseofthisinstrument,readthefollowingpoints
carefullyandadheretothem.Afterreadingthismanual,keepitinasafeplacewhereitcanbe
referredtoanytimeaquestionarises.
2.1.1.1 TermsandDefinition
Thismanualisusingthetermof"ATTENTION",etc.frombeginningtoend,whichaimsto
indicatethedangerandspecifiedseveritydegreeorlevel.Pleasebefamiliarwithitsdefinitionand
importance.
ATTENTION:Denotespotentialdangerorunsafetreatmentwhichmayresultinslightinjury
ordamagetothedeviceorotherproperty.Thenoteprovidesapplicationhintsor
otherusefulinformation.Readthenotecarefullytoensuresafeandcorrectuse.
2.1.1.2 ExplanationofSafetySymbols
SymbolsExplanation
Denotesanoteorwarningcontent;
Denotesaprohibitedoperation.Theoperationmustneverbeperformed;
Performtheoperationaccurately.Theinstructionmustbestrictlyadheredto.
2.1.2 Environment
ATTENTION:Ambienttemperatureexceedingtherangeoftechnicalspecificationwill
affecttheMeter’saccuracy,whichwillcausethemachinedamagedormachine
lifereduced.
2.1.3 Preparation
ATTENTION:Beforeusingthedevice,operatormustverifywhetherworkproceduresand
conditionsareappropriatefortheuseofthisinstrument.
2.1.4 OtherNotices
MSA100BTPeakFlowMeterInstructionManualV1.2

‐6‐
ATTENTION:Inordertoavoidelectricshockordetectorfailure,donotallowanyliquidto
permeateintothedevice.Ifliquidhaspermeatedintoit,pleasecontactyour
localservicepartnerordistributorofM&Btoinspectthemeter.
ATTENTION:Donotusethedetectornearflammableanestheticgases,steamorliquid.
ATTENTION:Forsafeoperationofthedevice,theusermustfollowthisinstruction
manual.However,theseinstructionsdonotsupersedeestablishedmedical
proceduresconcerningrespiratorysystemexamination.
ATTENTION:Donotreplaceaccessories,consumableswithoutM&Bauthorization.
2.2 TransportationandStorage
ATTENTION:PuttheMeterintothespecifiedstoragecaseduringtransportation.
2.3 InterferenceInstruction
ATTENTION:TheMSA100BThasfunctionsthatitdoesnotaffectnormaloperationof
otherdevices,orisnotaffectedbyotherdevices.
2.4 Classification
TypeofprotectionagainstelectricshockInternalpowerdevice
DegreeofprotectionagainstelectricshockTypeBFappliedpart
Degreeofprotectionagainstliquidpenetration
IP22,Protectedagainstsolidforeign
objectsof12,5mmandgreaterand
Protectionagainstverticallyfalling
waterdropswhenENCLOSUREtiltedup
to15°
ATTENTION:Theclassificationoftheelectricalshock,firedisaster,machineryandother
specialrisksisdoneaccordingtoEN606011.
2.5 SafeOperationandProcessingConditions
Thesterilizationordisinfectionmethod
recommendbymanufacturer
Notapplicable
Electromagneticinterference
Avoidtheusingofwirelesstelephonesorotherstrong
interferenceequipmentneartheMeter
MSA100BTPeakFlowMeterInstructionManualV1.2

‐7‐
WorkModeInterval
ElectricsurgicalinterferencedamageNodamage
M&BElectronicInstrumentsCo.,Ltddoesnotacceptliabilityorresponsibilityforincorrect
useoroperationoftheMSA100BT.
ATTENTION:Ifindoubt,consultyourPhysicianorHealthcareexpert
ATTENTION:Nomodificationofthisequipmentisallowed.
ATTENTION:Donotmodifythisequipmentwithoutauthorizationofthemanufacturer.
ATTENTION:Ifthisequipmentismodified,appropriateinspectionandtestingmustbe
conductedtoensurecontinuedsafeuseoftheequipment.
2.6 StandardsandRegulationCompliance
EN6060112006:MedicalElectricalEquipmentGeneralRequirementsforSafety;
IEC60601122004:MedicalElectricalEquipmentSafetyRequirementsEMC;
ATSStandardizationsofspirometry2005Revision.
MSA100BTPeakFlowMeterInstructionManualV1.2
‐8‐
3 MAINTENANCE
3.1 BatteryAssemblage
Showasfiguire31,slidethecoverdown,inserttwoAAAbatteriesintheshellfollowing
symbolinside,andclosethecover.
Figure31MSA100BE0batteryassemble
ATTENTION:IfdonotuseMeterforlongtime,pleasetakeoutbatteriesfrombatterycase
toavoidbatteryleakage.
ATTENTION:Pleaseselectbatterieswithregularfactoryinsteadoffakeandshoddy
batteries.
3.2 AssembleMouthpieceandTurbineintoMainUnit.
Insertmouthpiecetoturbinesensor,thentheassemblageofmouthpieceandturbinescrew
clockwiseintomainunit.Seetheprocedureatfigure32.
Figure32Assemblemouthpieceandturbine
3.3 Cleaning
MSA100BTPeakFlowMeterInstructionManualV1.2

‐9‐
ATTENTION:Pleasecleanmouthpiecewhenfirsttimeuse,preventfromdustbecauseof
transportorstorage.
TheMetershouldbecleanedwithin30minutesaftereachuse.Separatemouthpieceand
turbinefrommainbody.Disconnectmouthpieceandturbine.Immersethemouthpieceand
turbineinapreparedsoapsolution.Cleanthoroughlybyswirling.Thoroughlyrinsethe
mouthpiecewithwater(distilledwaterisrecommended).Allowthedevicetoairdry.
Figure33Disassembleturbineandmouthpiece
ATTENTION:Don’tusehighpressurewaterorrunningwatertorinseturbine,which
maybeleadtodamageturbine,andmakeaccuracydeteriorative.
Don’tuseboilingwatertorinseturbine.
Whendried,assemblemouthpieceandturbine,andinsertitintomainunit.
UsecleananddampclothcleanmainbodyofMSA100BT,butneverputmainbodyintowater.
Showasfigure34.
Figure34cleaningtomainunit
MSA100BTPeakFlowMeterInstructionManualV1.2
‐10‐
4 OPERATION
4.1 LCDScreenDisplayIntroduction
Figure41LCDdemonstration
Figure41showsallsegmentofLCDdisplaying,andthecharactersandsymbolsmeaningis
describedbelow.
:Fourrowsdigitdisplay,indicatingindividuallydate,time,measurementvalue,andrecord
number.
Day:Indicatenoworhistoricalmeasurementrecordday.
Month:Indicatenoworhistoricalmeasurementrecordmonth.
T:Indicatenoworhistoricalmeasurementrecordtime.
Y:Indicatenoworhistoricalmeasurementrecordyear.
AM:Indicatenoworhistoricalmeasurementrecordtimeisbefore12:00.
PM:Indicatenoworhistoricalmeasurementrecordtimeisafter12:00.
PEF:IndicatenoworhistoricalmeasurementPEFvalue.
FEV1:IndicatenoworhistoricalmeasurementFEV1value.
:Symbolofinsufficientbatteryvolume
L/min:UnitofPEFvalue.
L:UnitofFEV1value.
MR:RecordsnumberorID.
:Indicatortotrafficcolouroutsidesurfaceofmainunit,thepositionofarrowmeansthe
degreeofPEFnormalordeclined.
MSA100BTPeakFlowMeterInstructionManualV1.2

‐11‐
:IndicateBluetoothisopeningandenterthedatatransferinterface.
4.2 FirstTimeUse
Firstlyassemblemouthpieceandturbineaccordingsection3.2.
Secondlyinsertbatteriesaccordingtosection3.1.PoweronMSA100BT,themainunitenters
timesettingstate.
4.3 TimeSetting
Afternewbatterieshavebeeninserted,theLCDdisplayisshownasfigure42.Whiletheyear
“2012”isblinking,itindicatestheyearwaitingforsetting.Systemdefaultsettingisyear:2012,
month:1,day:1,time12:00amo’clock.Assumetargetyearis2013,settingmethodsarebelow.
Figure42Initialdisplayafterpoweron
Pressleftkeyorrightkeytochangeyear.Ifpressleftkeyonce,theyeardisplayingdecrease1.
Ifpressrightkeyonce,theyeardisplayingincrease1.Set2013,youshouldpressrightkeyonce.
Thenpresspower/confirmkeyonceturntonextsetting,shownasfigure43.Pleasenote:ifyou
holddownthekeymorethan2seconds,itspeedsup.
ATTENTION:Ifdateandtimeisnotset,youcan’tusetomeasure.Ifdateandtimeisnot
setcorrectly,yourrecordsdatemaybewronglater,whichwillmakehistorical
recordtimeincorrect.
MSA100BTPeakFlowMeterInstructionManualV1.2

‐12‐
Figure43yearsetting
Nowtheyearcharactersarenotblinking,meaningtheyearhassetover.Followingthemonth
“1”beginsblinking,youshouldsetmonth.Thesettingwaylikemonth,day,hourandminute,is
thesameasthatofyear.Whenalldateandtimeissetover,thedeviceentersintomeasurestatus.
4.4 Poweron/Poweroff
Withthemeteroff,presspower/confirmkeyonce;themeterwillpoweron.
Withthemeteron,presspower/confirmkeyuntilthebuzzerbeginbeep,thenreleasethe
key,themeterwillpoweroff.Beforethemeterswitchesoff,thehighestvalesofthecurrent
measurementsessionanditsrelatedmemoryposition“MRXXXisdisplayed.
Afterpoweron,thesystemcankeepactivestatuswithinoneminute,thenitwillpoweroff
automatically.Ifpressanykeyormeasureoncemore,thesystemwillprolong1minutetopower
off.
4.5 Measuring
Whenthesystempoweron,thelastresultisshown,andthenthedeviceindicates“READY”
forameasurementbytwoshortbeeps.Inhalecompletelyandholdyourbreathformoment.
Coverthemouthpiecetightlywithyourlips.Blowintoturbineashardandasfastasyoucan.
Showasfigure44.Alongbeepconfirmsthattheresultisrecorded.PEFisdisplayedforabout3
secondsfollowedbyFEV1andthenthemonitorisreadyforanewmeasurementwhichis
indicatedbytwoshortbeeps.Showasfigure45.
Itisrecommendedtoperformthreeormoremeasurementssequentially.Thedeviceshall
memorythebestvalueasthetestresult.
Aftereachmeasurement,atfirsttheactualreadingisshowninthedisplayandthenit
switchesautomaticallytothehighestreadingofyourcurrentmeasurementsession.
MSA100BTPeakFlowMeterInstructionManualV1.2

‐13‐
StoretheMeterinacleananddustfreeplace.
ATTENTION:Askyourphysician(orotherlicensedhealthcareprofessional’s)towatchyou
usethispeakflowmeterbeforeyourelyonanymeasurements.Readall
informationinthisinstructionbookletbeforeusing.
ContactM&Bat8008100520,+8601061253747+86010612537168020+86
010612537168001/80028AM5PMMondayFridaywithquestions;orcontact
withM&Bservicepartnerordistributor.
ATTENTION:Ifyouhavesignsandsymptomssuchaschesttightness,shortnessofbreath,
coughingorwheezingfollowyourphysician’s(orotherlicensedhealthcare
professional’s)adviceforcontactingthemnomatterwhatreadingsyouobtain.
Ifyoudonotgetareadingcontactsyourphysicianimmediately.
ATTENTION:DonotmeasurePEFandFEV1understrongsunlight.
Figure44Peakexpiremeasuringmethod
Figure45PEFandFEV1measurementvaluedisplay
Pleasenote:MSA100BTsavesthehighestPEFreadingfromswitchingthedeviceonuntil
MSA100BTPeakFlowMeterInstructionManualV1.2

‐14‐
switchingthedeviceoff.
4.6 ImportantInformationforMeasuringandMonitoringofPeakFlow
Values
Itisimportanttoregularlymonitoryourpeakflowvalues.Doctorsgenerallyrecommend
takingonemeasurementinthemorningandoneintheevening,everydaynormallybefore
inhalingyourprescribeddrugs.Carryoutadditionalmeasurementwheneveryoufeelunwellor
perceiveshortnessofbreath.
Pleasenotethatasthmaattacksareindicatedinadvancebylowpeakflowvaluesbeforeyou
feelit!Pleasecontactyourdoctorincasessuchasthese.Alsodosoifyouhavesignsand
symptomssuchaschesttightness,shortnessofbreathcoughingorwheezing.
Selfmeasurementmeanscontrol,notdiagnosisortreatment.Inanyevent,pleasebesureto
discussyourmeasuredvalueswithyourdoctor.Yourdoctorwillalsoexplainwhichvaluesare
normalforyou.
Youshouldneveralterthedosagesofanymedicationwithouttalkingtoyourdoctor.
Pleasenotethatthedevicestoresupto300measuredvalueswithdateandtime.Whenthe
memoryisfull,shownasfigure45,eraseallrecordsyourself,orvisityourdoctorintime,let
he/shetoanalyzeyourdata.
Pleaseconfirmthatdateandtimesettingsarealwaysactual.
DonotmeasurePEFandFEV1underoutdoorstrongsunlight.
TheperformanceoftheMetercanbeaffectedbyextremesoftemperatureandhumidity.
Pleasereferto“technicalspecifications”fordetails.
*Table41NormalPEFvaluesforMales(CaucasianAmerican)
AgeHeight(cm)
(Years) 120130 140 150 160 170 180 190200
8171208249292339388441496554
10 185222262306352402454510568
15 246283324367414463516571629
20 347384425468515564616672730
25 354392432475522571624679738
30 357395435479525575627682741
35 357394435478525574626682740
40 352390430474520570622677736
45 344382422465512561614669727
50 332369410453500549601657715
MSA100BTPeakFlowMeterInstructionManualV1.2

‐15‐
55 316353394437483533585641699
60 296333373417463513565621679
65 272309349393439489541597655
70 244281322365411461513569627
75 212249290333380429482537595
80 176214254298344393 446501560
85 137174215258304354406462520
90 93131171215261310363418477
*Table42NormalPEFvalueforFemales(CaucasianAmerican)
AgeHeight(cm)
(Years) 120130 140 150 160 170 180 190200
8170198228261295332371413456
10 207235265297332369408449493
15 262290320353387424463505548
20 275303333365400437476517561
25 282310340372407444483524568
30 286313344376411448487528572
35 286314344377411448487529572
40 284312342374409446485526570
45 278306336369403440479521564
50 270298328360395432471512556
55 258286316349383420459501544
60 243271301334368405444486529
65 225253283316351387427468511
70 204232263295330366406447490
75 180208238271305342381423466
80 153181211244278315354396439
85 123151181213248285324365409
90 90118148180215252291332376
*SpirometricReferenceValuesfromaSampleoftheGeneralU.S.Population,JohnL.Hankinson,JohnR.Odencrantz,andKathleen
B.Fedan,et.al.,DivisionofRespiratoryDiseaseStudies,NationalInstituteforOccupationalSafetyandHealth,CentersforDisease
ControlandPrevention,Morgantown,WestVirginia,1999.TheThirdNationalHealthAndNutritionExaminationSurvey(NHANES
III).
Figure45Memoryisfull
MSA100BTPeakFlowMeterInstructionManualV1.2

‐16‐
4.7 SimpleAssessment(TrafficLightScheme)
Usinggreen,yellowandredtrafficlighttoexpressthedegreeofyourillnessisastraightway.
Greenindicatesillnessisundercontrol,don’trequirehighermedicationdosage.Yellowindicates
youmustcaution,ifyourmeasuredvaluesfrequentlybelocatedinthisarea,increaseyour
medicationdosageasdiscussedwithyourphysician.Redindicatesitisdangerous.Actasdiscussed
withyourphysicianorseekemergencymedicaltreatment.
Simpleassessmentsettingsneedtoconsultwithyourdoctors(orotherlicensedhealthcare
professional’s).Thegeneralsettingmethodisbelow.
(1)MeasureyourpersonalbestvalueusingMSA100BT,whichisthehighestvalueyouachieve
underoptimumconditions(freeofsymptoms,feelingwell).OrchooseNHANESIIIorother
effectivepredictiontocalculateanormalvalue,itistobeasabestvalue.
(2)PressandholdPowerKeyforalongtime,whileinstallthebattery.DonotloosenPower
Key,untilyoucanseepredictedvalueofPEFblinked.Itistrafficlightschemebasicvalueor
predictionvalue.
(3)ThenuseRightKeyorLeftKeytoincreaseordecreasevalue,enteryourbestvalueor
“normalPEFvalueshownasTable4,
(4)PressPowerKeyonceagain,thendisplay”,settingsuccesses.
(5)Thegreenarealowerlimitusuallyreachesupto80%ofthepredictedvalue,theyellow
arealowerlimitusuallyreachesupto50%ofthepredictedvalue,thepercentisnotallowedto
edit,anditisdefaultintheMeter.
4.8 RecordsPlayback
Whenthesystementersintomeasuringstatus,presskeyleftorright,youcanwatch
historicalrecords.Ifyoukeeppressingthebuttonyoucanmovequicklythroughtherecords.At
thesametimerecordnumberwillrefreshwithincreasingordecreasingnumber.Therecords
containmeasureddate,time,PEFvalue,FEV1valueandpositioninmemory.Presspower/confirm
keyonceorwaitthecurrentrecorddisplayover,thesystemwillentersintomeasuringstate.
MSA100BTPeakFlowMeterInstructionManualV1.2

‐17‐
Figure46Recordsplayback
4.9 Deleterecords
Inmeasuringstatuspressleftandrightkeyforfivesecondslong,andallrecordswillbe
deleted.Thehintafterdeletedisshownasfigure47.
Figure47Allrecordshasdeleted
4.10 DataTransmission
1. OpeningtheAPPinthephone,theAPPwillconnecttoMSA100BTautomatically
2. IfMSA100BTnotconnectedtotheAPP,the"bLE"symbolatthebottomflashes;ifMSA100BTconnectedto
theAPP,the"bLE"symbolatthebottomalwaysshow.
3. IfMSA100BTconnecttoAPPsuccessfully,MSA100BTwilluploadthehistoryandcurrentmeasurementsto
APPautomaticallywhenyoufinishedameasurement.
4. Afterthedatatransmissioncompleted,MSA100BTwilldeletethehistoryandcurrentmeasurements,and
returntothemeasurementinterfaceautomatically.
MSA100BTPeakFlowMeterInstructionManualV1.2

‐18‐
Figure4—8MSA100BTenterstheBluetoothdatatransferinterface
ATTENTION: MSA100BTonlysupportsBLEversionofBluetooth
MSA100BTPeakFlowMeterInstructionManualV1.2

‐19‐
5 TROUBLESHOOTING
MessageCauseRemedy
ThePEFresultishigherthan900L/min,or
FEV1resultishigherthan9.99L
thisisaverygoodresult
Meansrecordsaddressexceedslimit,or
recorddateandtimeiserror,orsoftware
isfailure
Restartsystem
Recordsdonotexistnorecords
MemoryisfullItisbettertodeleterecords,
ornewrecordisnotstored
batteryinsufficientItisbettertochangenew
batteries
NOdisplayon
screen
Batteryislow,orbatterypoleisreversed. Changenewbatteries,and
notepoledirection
Measuringvalue
error
Inspectwhethertheleafof
turbineisrunfreely.Andthe
leafisdirty.Orconsult
whetherthevalueisnormal
withyourdoctor
OtherquestionpleasecontactwithM&Bservicepartnerordistributor.
MSA100BTPeakFlowMeterInstructionManualV1.2
‐20‐
6 GUIDANCEANDMANUFACTURE’SDECLARATIONS
ElectromagneticEmissions
TheMSA100BTpeakflowmeterisintendedforuseintheelectromagneticenvironmentspecified
below.Thecustomerortheuserofthedeviceshouldensurethatitisusedinsuchanenvironment.
EmissionstestComplianceElectromagneticenvironmentguidance
RFemissions
CISPR11
Group1
TheMSA100BTpeakflowmeterdetectorusesRFenergyonly
foritsinternalfunction.Therefore,itsRFemissionsarevery
lowandarenotlikelytocauseanyinterferenceinnearby
electronicequipment.
RFemissions
CISPR11 ClassB
TheMSA100BTpeakflowmeterdetectorissuitableforusein
allestablishments,otherthandomesticestablishmentsand
powersupplyisinternal.
ElectromagneticImmunity
TheMSA100BTpeakflowmeterdetectorisintendedforuseintheelectromagneticenvironment
specifiedbelow.Thecustomerortheuserofthedeviceshouldensurethatitisusedinsuchan
environment.
ImmunitytestIEC60601testlevelCompliancelevel
Electromagnetic
environment–guidance
Electrostatic
discharge(ESD)
IEC6100042
±2kVcontact
±4kVcontact
±6kVcontact
±8kVcontact
±2kVair
±4kVair
±8kVair
±15kVair
±2kVcontact
±4kVcontact
±6kVcontact
±8kVcontact
±2kVair
±4kVair
±8kVair
±15kVair
Floorsshouldbewood,
concreteorceramictile.If
floorsarecoveredwith
syntheticmaterial,the
relativehumidityshouldbe
atleast30%.
Powerfrequency
(50/60Hz)
magneticfield
IEC6100048
10A/m 10A/mPowerfrequencymagnetic
fieldsshouldbeatlevels
characteristicofatypical
locationinatypical
commercialorhospital
environment.
RadiatedRFIEC
6100043
3V/m
80MHzto2.5GHz
3V/m P2.3d 800MHz
to2,5GHz
wherePisthemaximum
outputpowerratingofthe
transmitterinwatts(W)and
distherecommended
separationdistancein
MSA100BTPeakFlowMeterInstructionManualV1.2

‐21‐
meters(m)b.
FieldstrengthsfromfixedRF
transmitters,asdetermined
byanelectromagneticsite
survey,ashouldbelessthan
thecompliancelevelineach
frequencyrange.b
Interferencemayoccurin
thevicinityofequipment
markedwiththefollowing
symbol:
aFieldstrengthsfromfixedtransmitters,suchasbasestationsforradio(cellular/cordless)telephones
andlandmobileradios,amateurradio,AMandFMradiobroadcastandTVbroadcastcannotbe
predictedtheoreticallywithaccuracy.ToassesstheelectromagneticenvironmentduetofixedRF
transmitters,anelectromagneticsitesurveyshouldbeconsidered.Ifthemeasuredfieldstrengthin
thelocationinwhichMSA100BTpeakflowmeterdetectorisusedexceedstheapplicableRF
compliancelevelabove,MSA100BTpeakflowmeterdetectorshouldbeobservedtoverifynormal
operation.Ifabnormalperformanceisobserved,additionalmeasuresmaybenecessary,suchas
reorientingorrelocatingMSA100BTpeakflowmeterdetector.
bOverthefrequencyrange150kHzto80MHz,fieldstrengthsshouldbelessthan3V/m.
EMCDeclaration
Withmanyelectronicdevicessuchasmicrowave,TV,wirelessrouter,PCs,andmobiles,medical
devicesinusemaybesusceptibletoelectromagneticinterferencefromthesedevices.Thismay
resultinincorrectoperationofthemedicaldeviceandcreateapotentiallyunsafesituation.
Medicaldevicesshouldbekeptdistance1mfromthoseelectronicdevices.
InordertoregulatetherequirementsforEMCwiththeaimtopreventunsafeproductsituations,
theEN6060112standardhasbeenimplemented.
Thisstandarddefinesthelevelsofimmunitytoelectromagneticinterferencesaswellasmaximum
levelsofelectromagneticemissionsformedicaldevices.
MedicaldevicesmanufacturedbyM&BconformtothisEN6060112standardforbothimmunity
andemissions.
Themedicaldevicesshouldnotbeusedadjacenttoorstackedwithotherequipment.Ifadjacent
orstackeduseisunavoidable,themedicaldeviceshouldbeseentooperatenormallyasitshould
inthissituation.
FCCDeclaration
Note:ThisequipmenthasbeentestedandfoundtocomplywiththelimitsforaClassBdigital
MSA100BTPeakFlowMeterInstructionManualV1.2

‐22‐
device,pursuanttoPart15oftheFCCRules.Theselimitsaredesignedtoprovidereasonable
protectionagainstharmfulinterferenceinaresidentialinstallation.Thisequipmentgenerates,
uses,andcanradiateradiofrequencyenergy,andifnotinstalledandusedinaccordancewiththe
instructions,maycauseharmfulinterferencetoradiocommunications.However,thereisno
guaranteethatinterferencewillnotoccurinaparticularinstallation.Ifthisequipmentdoescause
harmfulinterferencetoradioortelevisionreception,whichcanbedeterminedbyturningthe
equipmentoffandon,theuserisencouragedtotrytocorrecttheinterferencebyoneormoreof
thefollowingmeasures:
Reorientorrelocatethereceivingantenna.
Increasetheseparationbetweentheequipmentandreceiver.
Connecttheequipmentintoanoutletonacircuitdifferentfromthattowhichthe
receiverisconnected.
Consultthedealeroranexperiencedradio/TVtechnicianforhelp.
Changesormodificationsnotexpresslyapprovedbythepartyresponsibleforcompliancecould
voidtheusersauthoritytooperatetheequipment.
ThisdevicecomplieswithPart15oftheFCCRules.Operationissubjecttothefollowingtwo
conditions:
(1)Thisdevicemaynotcauseharmfulinterference,and
(2)Thisdevicemustacceptanyinterferencereceived,includinginterferencethatmaycause
undesiredoperation.
MSA100BTPeakFlowMeterInstructionManualV1.2
‐23‐
7 TECHNICALSPECIFICATION
MSA100BTPeakFlowMeterisahandholdpulmonaryfunctionmeasuringmedicaldevice
thatmeasuresyourmaximumpossibleexhalationwhichiscalledpeakexpiratoryflow(PEF)and
forcedexpiratoryvolumein1second(FEV1).TheparametersaredisplayedontheLCDscreen.
MSA100BTalsocanrecordmeasurementsupto300pieces,andsupportsBluetoothconnection.
Whenmeasurementrecordsadduptomaximumorotherpatientwoulduse,youcantransferthe
recordstophoneviaBluetooth,thenasoftwarehelptosaveandtrackforlongtime.
General
SensortypeTurbine
MeasurementparametersFEV1/PEF
Memorysize300
BTPSPEFandFEV1resultsdisplayedatBTPSconditions
Accuracy
MeetATS2005Revisionaccuracyrequirement
Volumerange0.01L~9.99LAirflowrange50L/min~900L/min
Volumeaccuracy±0.050Lor±3%Airflowaccuracy±10%or±20L/min
Resolution
Volumeresolution0.01LAirflowresolution 1L/min
Resistancetoflow
Backpressure@660L/min<0.11KPa/sec
Backpressure@900L/min<0.15KPa/sec
Repeatability
PEF±5%or±10L/min
FEV1±3%or±0.05L
Displayscreen
Size1.8inches
TypeLCD
Control
Standardpower/confirmkey,left/rightkey
Environmentspecification
Powersource AAA1.5×2batteries
DesignInterval,ordinarydevice
BatteryworktimeMaximumoneyear
ConnectionWherefitted,Bluetoothforphoneconnection.
Operation
Temperatu re
Humidity
10oC40oC
25%~85%
MSA100BTPeakFlowMeterInstructionManualV1.2

‐24‐
Atmosphere860hPa1060hPa
Storage
Temperatu re
Humidity
Atmosphere
10oC55oC
15%~93%
500hPa1060hPa
Zoning
Green/Yellowthreshold80%ofreference
Green/Yellowthreshold50%ofreference
Physicalspecification
Height12.6centimeter
Wide5.4centimeter
Thickness2.8centimeter
Weight120gcontainbatteries
andmouthpiece
Maintenancewarranty
StandardOneyear
OptionProvideotheroptions.
ThedetailpleasecontactwithM&Banddistributors.

Navigation menu