Biolight Meditech WBP202 Electronic Sphygmomanometer User Manual Electronic Sphygmomanometer

Guangdong Biolight Meditech Co.,Ltd Electronic Sphygmomanometer Electronic Sphygmomanometer

User Manual

       Electronic Sphygmomanometer   User’s Manual         Guangdong Biolight Meditech Co., Ltd.
 Product Information  Product Model:          WBP100/WBP201/WBP202  Product Name:            Electronic Sphygmomanometer  Manufacturer Name:    Guangdong Biolight Meditech Co.,Ltd.  After Service Contact Information: Address: No.2 Innovation First Road, Technical Innovation Coast, Hi-tech Zone, Zhuhai, P.R. China Fax: +86-756-3399919 Post code: 519085 Toll-free consultation hot line: +86-400-8818-233  Service life:5 years  Revision History This  manual  has  a  revision  number.  This  revision  number  changes whenever  the  manual  is  updated  due  to  software  or  technical  specification change. Contents of this manual are subject to change without prior notice.  Document No.: J/WBPCE-A-008  Revision number: V1.0  Release time: Sept. 2015 Copyright  ©  2015  Guangdong  Biolight  Meditech  Co.,  Ltd.  All  rights reserved.  CE mark    EC Representative Name:   Shanghai International Holding Corp. GmbH (Europe) EC Representative Address: Eiffestrasse 80, 20537 Hamburg Germany
 Statement Manufacturer holds the copyright of this manual, and we are also entitled to  deal  with  this  manual  as  confidential  files.  This  manual  is  only  used  for operation, maintenance and service of product, someone else can not publish the manual. This manual contains  exclusive information protected by  copyright laws and  we  reserve  its  copyright.  Without  written  approval  of  manufacturer  no parts  of  this  manual  shall  be  photocopied,  Xeroxed  or  translated  into  other languages. The contents contained in this manual are subject to amendments without notification.  Manufacturer's Responsibility Only under the following circumstances will manufacturer be responsible for the safety, reliability and performance of the instrument:  All  the  installation,  expansion,  readjustment,  renovation  or  repairs  are conducted by the personnel certified by manufacturer.  The  storage  condition,  operation  condition  and  electrical  status  of  the instrument conform to the product specification.  The instrument is used in accordance with the user’s manual.  About this manual This  manual  contains  the  instructions  necessary  to  operate  the  product safely  and  in  accordance  with  its  function  and  intended  use.  Observance  of this  manual  is  a  prerequisite  for  proper  product  performance  and  correct operation.    This manual is based on the maximum configuration and therefore some contents  may  not  apply  to  your  product.  If  you  have  any  question,  please contact us.    This  manual  is  an  integral part  of  the  product.  It  should  always  be  kept close to the equipment so that it can be obtained conveniently when needed.
 Illustrations are for reference only. They may be not in conformity with your products.          Signs in this manual:   Warning: Indicates a potential hazard or unsafe practice that, if   not avoided, will result in death or serious injury. Caution: Indicates a potential hazard or unsafe practice that, if not   avoided, could result in minor personal injury or product/property   damage.       Note: Provides  application  tips  or other useful information to ensure that you get the most from your product.    
Electronic Sphygmomanometer User’s manual I Contents  1、 Intended Use ......................................................................................................................... 1 2、 Application ............................................................................................................................ 1 3、 Safety Information ................................................................................................................ 1 4、 Appearance ........................................................................................................................... 4 5、 Display .................................................................................................................................. 6 6、 Setting ................................................................................................................................... 7 7、 Historical data review function ............................................................................................. 8 8、 Download And Install The Software ..................................................................................... 8 9、 Using Steps ........................................................................................................................... 8 10、 Maintenance and Cleaning .............................................................................................. 11 11、 Accessory and Specifications .............................................................................................. 12 12、 Troubleshooting .............................................................................................................. 14 13、 Equipment Symbols ........................................................................................................ 15 14、 Blood pressure general knowledge ................................................................................. 15 Appendix A Guidance and Manufacturer’s Declaration of EMC ................................................... 17
Electronic Sphygmomanometer User’s manual   1   1、 Intended Use The  device  is  intended  to  measure  blood  pressure  and  pulse  rate  of  adult.  The measurement data is available for clinical reference. It can store, manage the measure data, and transmit the measuring result via wireless signal.   It is intended to be used in clinic, general ward, physical examination department and at home.    Caution:    The  patient  is  the  operator  when  the  device  is  used  at  home  or in  public areas;  The device is intended for use only by clinical professionals and in continual attendance when it’s used in hospital.  2、 Application The  device  is  reusable  and  intended  for  measuring  human  blood  pressure  and pulse rate.    Application population Adult.  Measurement site Place a bare arm through the cuff and position the cuff 1-2cm above the elbow joint.  3、 Safety Information  Warning:  Don’t take measurement immediately after smoking, drinking alcohol, coffee and black tea, or within an hour after a meal;  Magnetic  and  electrical  fields  are  capable  of  interfering  with  the  proper performance  of  the  device.  For  this  reason  make  sure  that  all  external equipments operated in the vicinity of the device comply with the relevant EMC requirements. Mobile phone, X-ray equipment or MRI devices are a possible  source  of  interference  as  they  may  emit  higher  levels  of electromagnetic radiation;  The  device  is  designed  for  adults  and  should  never  be  used  on  infants, 
Electronic Sphygmomanometer User’s manual   2   pregnant, pre-eclampic patients or people with serious arrhythmia. Consult your physician before use on children;  Don’t use the electronic sphygmomanometer in the environment containing inflammable anesthetic gas;  Consult your physician for any of the following situations: a) The application of the cuff over a wound or inflamed area; b)  The  application  of  the  cuff  on  any  limb  with  intravascular  access  or therapy, or an arterio-venous (A-V) shunt; c) The application of the cuff on the side of a mastectomy; d) Simultaneous use with other medical monitoring equipment on the same limb;  The  device  cannot  be used  together  with  high-frequency electrotomes  and defibrillators;  Please do not use the device while it is charging;  Please do not measure immediately after an intense movement or a bath;  This product might not meet its performance specifications if stored or used outside the specified temperature, humidity and altitude ranges;  Please  do  not  share  the  cuff  with  any  infectious  person  to  avoid cross-infection;  Battery replacement should only be performed by personnel who have been confirmed  qualified  by  the  manufacturer.  To  do  otherwise  will  void  your warranty and possible damage your unit;  Any form of modification to this device is forbidden.   Caution:  Make sure that the appropriate cuff size is used;  Cuff  is  replaceable  and  its  replacement  can  only  be  conducted  by manufacturer.  We recommend the users to use the original manufactured cuff and adapter. If not, it might cause measurement error or product failure;  If the adapter is abnormal, please change the adapter;  Too  high  cuff  pressure  or  too  long  charging  time  may  make  users  feel uncomfortable,  and  cause  ischemia  and  neuropathy.  Please  turn  off  the device in time and take off the cuff;  The device and adapter should be considered as a medical system. The power
Electronic Sphygmomanometer User’s manual   3   adapter must be provided or designated by manufacturer (see optional item list). If use another power adapter, it should be verified, otherwise electric shock may occur and cause patient death;  The  effective  wireless  transmission  distance  is 5m  in  a  barrier-free environment. If more than this distance, the product might not work. Please use this product within effective wireless transmission distance;  The charging cable  should  be put beyond children to avoid  the  danger of strangulation;  If children or pets beat,  throw  or tramp on  the device, it  might affect the performance of the device. Please put the device beyond children and pets;  The  main unit,  cable, battery  and  cuff  must  be  disposed  according to  the local regulations at the end of their usage.    Caution:  Changes or modifications not expressly approved by  the party responsible for compliance could void the user’s authority to operate the equipment;  This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency  energy  and,  if  not  installed  and  used  in  accordance  with  the instructions,  may  cause  harmful  interference  to  radio  communications. However,  there  is  no  guarantee  that  interference  will  not  occur  in  a particular installation;  If  this  equipment  does  cause  harmful  interference  to  radio  or  television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: -- Reorient or relocate the receiving antenna. -- Increase the separation between the equipment and receiver. -- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. -- Consult the dealer or an experienced radio/TV technician for help.
Electronic Sphygmomanometer User’s manual   4    Note:  Doctor should make diagnosis on clinical manifestation and symptoms, only with using the device as subsidiary.    Blood pressure can be affected by the position of the cuff, measuring posture and your physiologic condition;  This device complies with Part 15 of the FCC Rules. Operation is subject to the  following  two  conditions:  (1)  this  device  may  not  cause  harmful interference,  and  (2)  this  device  must  accept  any  interference  received, including interference that may cause undesired operation;  The device complies with the standard of IEC 80601-2-30;  If the determined blood pressure (systolic or diastolic) is outside the rated range specified in part specifications, the device will immediately display an error code on screen. In this case, consult a physician or ensure that proper measurement procedures are followed;  Patients  can  use  all  functions  of  the  device  safely  according  to  this  user’s manual;  Any questions about setting, using or maintaining the device, please contact with manufacturer;  If unexpected operation or events appear in use process, please report to the manufacturer;    When measuring, the following conditions may lead to the measured results have differences: a) Talking ;                            b)After defecate or urinate; c) After the bath;                  d) Different measurement site or environment; e) After smoking;                  f) After drinking alcohol, coffee and black tea; g) After the movement;        h) Room temperature changes suddenly i)  Within  an  hour  after  a  meal;      j)  Mood  be  agitated  when  caused  by tension, uneasy.  4、 Appearance    WBP100:     
Electronic Sphygmomanometer User’s manual   5        WBP201/WBP202:    1、 Cuff  2、 Display screen   Display systolic blood pressure, diastolic blood pressure, pulse rate and so on.  3、 Power button;Start/Stop button —  Long press  (press  and  hold  it  more  than  2s)  this  button  to  turn  off  the device; — When the  power  is  off, short  press (press and  hold it  less  than  2s) this button  to  turn  on  the  device.  Otherwise,  short  press  it  to  start/stop  the NIBP measurement.  4、 Charging cable interface Connect to the power supply for charging.  1 3 4 55 6 4 3 2 1
Electronic Sphygmomanometer User’s manual   6   5、     Select users button —  In measurement sate: Short press this button to select users; —  In setting state: Short press this button, the time value would increase one.    6、     Historical data review button —  Not in setting state: Short press this button to review historical data. Long press it to enter setting mode. —  In setting state: Short press this button to switch the setting items. Long press it to save the setting and quit setting mode.  5、 Display  1. Systolic blood pressure; 2. Blood pressure unit; 3. Diastolic blood pressure; 4. Pulse rate; 5. Time; 6. Users name; 7. Pulse rate unit; 8. Signal quality; 9. Bluetooth state; 10. Battery state.  When measuring error, such as transducer failure, loosely wrapped cuff, screen displays the wrong code. The definition of the error code and the possible reason are as follows:  Error code Definition Possible Reason E02 Self-test failed Transducer or other hardware failure. E06 Loose cuff a. Cuff is completely unwrapped; b. No cuff attached; E07 Air leak Air leak in pneumatics, hose, or cuff. E08 Air pressure error Unit cannot maintain stable cuff pressure, e.g. kinked. E09 Weak signal a. Very  weak  patient signal due  to  a loosely wrapped cuff; b. Extremely weak pulse from patient. 10 1 3 4 5 8 7 6 9 2
Electronic Sphygmomanometer User’s manual   7   Error code Definition Possible Reason E10 Range exceeded Measurement range exceeds module specification of 255 mmHg    maximum. E11 Excessive motion a. Too  many  retries  due  to  interference  of  motion artifact; b. Signal  is  too  noisy  during  step-down  to  detect pressure pulse, e.g. subject has severe tremor; c. Irregular pulse rate, e.g. arrhythmia. E12 Overpressure sensed Cuff pressure exceeds software safety limit, 295 mmHg in  Adult  mode.  Could  be  due  to  rapid  squeezing  or bumping of cuff. E13 Signal saturated Large  motion  artifact  that  saturates  the  BP  amplifier’s amplitude handling capability. E14 Air system leak Module  report Air  system  leakage  Failure while  in  the Air System Leakage Test mode. E15 System failure Abnormal events. E19 Time out Measurement took more than 120 seconds in adult.  6、 Setting You can set the time and date of this product. The setting method is as follows: Press the historical data review button “    ”and hold it for 3 seconds to enter the year setting interface. The third number of the year starts to flash. When you press the select users button “    ” once, the third number increase 1 (It means the year increases ten years).   After you finish setting the third number of year, press the historical data review button “    ” to enter the setting interface of the fourth number. The fourth number of the year starts to flash. When you press the select users button “    ” once, the fourth number increase 1 (It means the year increases 1 year). Set month, day, hour, and minute in the same way. After  you finish  setting  time  and  date, press  the  historical  data  review  button “   ” to enter the setting interface of blood pressure units (The default unit is mmHg). When a  flashing  mmHg  appears on  the screen, it  means  the blood pressure unit is mmHg. Press the select users button “    ”, mmHg would disappear and kPa would appear. Press the button “    ” again, kPa would disappear and mmHg would appear. So repeatedly, you can switch between the two units. Press the historical data review button “    ” and hold it for 3 seconds to end the setting after you finish setting unit.
Electronic Sphygmomanometer User’s manual   8   7、 Historical data review function This device can record the blood pressure, pulse rate and time automatically to facilitate the management of blood pressure. Press the historical data review button “    ”, the  latest  measurement result  would  be  displayed on  the screen.  Press  the button “     ” again, the second latest measurement result would be displayed. And so on, you can review all the historical data.  8、 Download And Install The Software   1、Software download methods: Users who use the IOS system device can log in iphone App Store, enter “Biolight” and search for the software of “BP tracker” in the searching results.   Users who use the Android system device can log in the wearcare official website: http://www.wearcare.cn    for download.   2、Download  the  software  of  “BP  tracker”,  click  installation  button  for  free installation.  9、 Using Steps    Preparation before measurement 1、Keep your body  and mood  calm,  take  off the  thicker clothes such as coat, sweater, Etc. When measuring, bare upper arm or wear thinner shirt.  2、Long  press  the  power  button  to  turn  on  the  device.  (Users,  who  use  APP software,  need  to  open  the  Bluetooth  of  the  device  and  terminal,  open  the  APP software, and connect with the device )   Caution:  When  a  call  comes  in  during  the  measurement,  the  measurement  will  be terminated.  It  is  recommended  that  IOS  device  be  set  in  Airplane  mode during measurement to avoid strong magnetism interference.   Using Steps (1)   Apply the cuff
Electronic Sphygmomanometer User’s manual   9    Loosen the calibre of the curly cuff slightly. Place a bare arm through the cuff directly according to the instruction icon “       ”, “      ”. Position the cuff 1-2cm above the elbow joint and ensure the mark “Φ” on the cuff shall lie above artery.           Pull  one  end  of  the  cuff  to  the  lateral  and  tighten.  The  cuff  should  fit comfortably, yet snugly around your arm. You should be able to insert one finger between your arm and the cuff.   Caution:  Make  sure  that  the  appropriate  cuff  size  is  used.  Refer  to  the  cuff circumference range in the Specification section of this manual.   (2) Body posture Stay still and calm for five minutes before taking a blood pressure measurement. Be seated with your feet flat on the floor without crossing your legs. Place your hand palm-side up in front of you on a flat surface such as a desk or table. The cuff should be placed at the same level as your heart.
Electronic Sphygmomanometer User’s manual  10    Note:  Incorrect  measuring  posture  showed  as  follows  might  cause  an  incorrect measurement:   a) Bent over(lean forward); b) Sitting cross-legged; c) Bend the body because of the low table or sitting on sofa.    (3) Measuring  Select users: Press the Select users button to select users, you can select user 1 or user 2;  Start measuring: Press the power button, then the cuff pressure automatically and  the  screen  shows  rising  pressure  values.  Stay  still  and  calm  during measurement. Do not move your arm, body and the device.  (4)   Display the measurement value The device exhausts automatically and displays the measurement results. About the display interface of measurement results, please refer to chapter “5 Display”. Users,  who  use  the  APP  software,  can  read  the  measurement  results  on  the software interface.   (5) Take off the cuff (6) Turn off the device Long press the power button to turn off the device after the measurement.    Note:  While measuring blood pressure, the users must stay still and calm without any talk;  The cuff tied on the limb shall be on the same level as the patient’s heart so as to avoid the reading error resulting from the hydrostatics effect of the blood flow between the heart and cuff. If the cuff position is higher than heart level, the BP  reading will be lower, the measured result shall be added  0.75mmHg (0.1kPa)  for  each  centimeter  higher;  in  case  the  cuff  position  is  lower  than heart level, the BP reading will be higher, the measured result shall be deducted 0.75mmHg (0.1kPa) for each centimeter lower;  If  the  signal  measurement  time  is  more  than  two  minutes  and  the  device doesn’t exhaust automatically, please take off the cuff and turn off the device.   
Electronic Sphygmomanometer User’s manual  11       Caution:  You can stop the measurement process at any time by pressing “Start/Stop button”.    Warning:    Prolonged  and  frequent  NIBP  measurements  are  associated  with  ischemia and neuropathy in the limb wearing the cuff. Please examine the extremities of  the  limb  frequently  for  normal  color,  warmth  and  sensitivity.  If  any abnormality is observed, stop the NIBP measurements.  Too frequent measurements may have influence on the measurement results.   10、 Maintenance and Cleaning   1、 Clean the traces on the host or display with a dry, soft cloth or tissue; 2、 The most commonly used hospital cleaning agent and non-corrosive detergents can be used for cleaning the device, but please be careful that many types of detergents  must  be  diluted  before  use;  Please  use  them  according  to  the directions of the manufacturers of the detergents; 3、 The casing of the device should be kept from the contamination of filth and dirt, and it  can be  wiped  with  non-velvet soft  cloth. When  cleaning, do not spill the liquid onto the instrument. Ensure no liquid is allowed to enter the inside of the device; 4、 It is recommended that if the cuff is used, for example, in hospital or a clinic, it be disinfected twice a week, Wipe the inner side (the side that contacts skin) of the  cuff  with  a  soft  cloth  lightly  moistened  with  Ethyl  alcohol  (75%-90%). Then air dry the cuff. 5、 Please do not clean or wet cuff. Please do not soak the device in liquid; 6、 Please store the host and the cuff in a cool and dry place; 7、 Do not drop the device or subject it to strong impact.     Caution:  Avoid using alcohol-based, amido or acetone-based detergents.  Please  put  the  main  unit  and  its  accessories  beyond  infants  and  young 
Electronic Sphygmomanometer User’s manual  12   children to avoid dangerous accident.  Under normal circumstances, it is unnecessary for the device to have special maintenance, and cautions must be exercised on the following points during the use of the device. ——Please use the device in the environment according to the requirements of the performance criteria; ——Avoid exposure or direct sunlight; ——Avoid excessive radioactive infrared rays or ultraviolet rays; ——Avoid contacts with organic solutions, dusts or corrosive gases.  No  device  component  needs  to  be  maintained  by  the  user.  Electrical schematic diagram and component list are only provided to repair station or personnel which have been confirmed qualified by the manufacturer;  Check the device regularly please.  If you doubt the measurement result is not accurate, please contact the manufacturer.  11、   Accessory and Specifications   11.1   Optional item Items Mode Power adapter (optional) LXCP12-05(Output DC5V,1.0A)   11.2   Technical Specifications   Production model WBP100、WBP201、WBP202 Size 130 mm(L) x 55 mm(W) x 30 mm(H) Weight <350g Measure specifications Measurement way Automatic oscillometry Measurement range          Systolic:          Diastolic:  30 mmHg~255 mmHg(4.0kPa~34.0kPa); 10 mmHg~220 mmHg(1.3kPa~29.3kPa); Pressure accuracy Static: Clinic:    ±3 mmHg   Average error: ±5 mmHg, standard deviation: ≤8 mmHg PR range and accuracy 40 bpm~240 bpm,±5% or ±3bpm, whichever is the greater Blood pressure unit “mmHg” or “kPa”
Electronic Sphygmomanometer User’s manual  13   Software  overpressure protection 295±5 mmHg Measurement time Normally, it is 20s to 45s (depending on HR and moving interference typically) Maximal measurement time: 120s LED indicating lamp   WBP 100 Bluetooth indicating lamp 1 (Blue)   Not connected: The lamp flashes blue; Connected: The lamp lights blue without flashing; (When the device is power off, the lamp is off.) Charging indicating lamp 1 (Orange)   Light up: When the battery is being charged or when the DC power supply is connected; Off: When the battery is fully charged. WBP 201 WBP 202 Bluetooth state indicating Not connected: The indicating flashes; Connected: The indicating lights without flashing; Battery state indicating The electric quantity of battery is divided into four ranks to display the current electricity. Being Charged : The indicating flashes; Not being charged or fully charged: The indicating lights without flashing. Bluetooth   Bluetooth version Bluetooth 4.0 with low power consumption. Transmission frequency   2440 MHz Wireless  transmission distance <5m App software platform IOS system iPhone 4s、iPhone5、iPhone5s、iPhone 6、iPad3+、iPad mini Android (4.3+) system Mobile phone, Palmtop computer Battery Type Rechargeable lithium ion battery Rated voltage 3.7 VDC Capability 600 mAh Operating time ≥24 h   Charge time ≤2 h Cuff
Electronic Sphygmomanometer User’s manual  14   Large cuff 32 cm~42cm(Optional) Medium cuff 22 cm~32cm(With) Small cuff 17 cm~22cm(Optional) Adapter Input 100~240V, 50/60Hz    0.3Amax Output DC 5V, 1A Environment specifications Temperature   Operating: 5℃~+40℃; Transportation and storage: -20℃~+55℃. Humidity   Operating: 25%~95%(non condensing); Transportation and storage: 25%~95% (non condensing). Atmospheric pressure   Operating: 700hPa~1060hPa; Transportation and storage: 500hPa~1060hPa.  12、 Troubleshooting   Symptom   Possible Reason Solution   No  response  or  cannot turn on the device 1. The  battery  power  might  be  at shortage or no electricity; 1、 The  device  itself  might  be damaged. 1、 Please charge the device; 2、 Contact customer service. The  data  cannot  be display  properly  or display reads an abnormal result 1. The  cuff  position  was  not correct  or  it  was  not  properly tightened; 2. Speaking, moving arm or body, being angry, excited or nervous during test. 1. Review  the  cuff application  instructions and retest; 2. Retest  when  calm;  avoid speaking  or  movement during the test. Cannot  make  connection between  device  and receiver 1. Bluetooth  does  not  be  open  on receiver; 2. The  version  of  Android  in  the receiver device isn’t 4.3 or 4.3+; 3. The  version  of  Bluetooth  is  not 4.0; 4. Out of connection range. 1. Open Bluetooth; 2. Upgrade the Android system to version 4.3 or 4.3+; 3. Use the receiver device with Bluetooth 4.0; 4. Close to receiver.
Electronic Sphygmomanometer User’s manual  15   13、 Equipment Symbols   Symbol Symbol Note  Type BF applied part, without defibrillation protected.  Refer to this user’s manual.  Symbol  for  the  marking  of  electrical  and  electronics  devices  according  to Directive 2002/96/EC LOT Batch code SN Serial number IP22 Enclosure degree of ingress protection.    Federal Communications Commission IPX0 Power adapter Degree of protection against ingress of liquid  Fragile.   Show transport package contents fragile, so handling should be handled with care.  Upward. It shows the correct position of the transport package is upright.  Guard against wet. Show packages afraid be wet.  Stacking layer limit. Same packing maximum stacking layers, N represents the number of layers limit. (N is 6). 14、 Blood pressure general knowledge 14.1   Definition of blood pressure Blood  pressure  is  the  pressure  on  the  vessel  wall  side.  When  the  heart  is contracting, the  blood  pressure is  the  highest  at  this time  and is called the  systolic pressure. When the heart is relaxing, the blood pressure is the lowest at this time and is called the diastolic blood pressure.
Electronic Sphygmomanometer User’s manual  16   14.2   Classification of blood pressure   The World Health Organization (WHO) has created the following blood pressure classify table for assessing high blood pressure (without regard age or gender). Please note  that  other  factors  (e.g.  diabetes,  obesity,  smoking,  etc.)  also  need  to  be considered. Consult with your physician for accurate assessment. Scope Systolic  blood pressure (SYS) kPa/mmHg Diastolic  blood pressure (DIA) kPa/mmHg Corresponding measure Hypotension ≤13.3/100 ≤8.0/60 Ask  physician  for measuring. Normal blood pressure 13.3/100~18.7/140 8.0/60~12.0/90 Self measure. Mild Hypertension 18.7/140~21.3/160 12.0/90~13.3/100 Consult physician. Moderate Hypertension 21.3/160~24.0/180 13.3/100~14.7/110 Consult physician. Severe Hypertension ≥24.0/180 ≥14.7/110 Danger!  Please  see  the doctor  as  soon  as possible.     Note:This  table  is  not  intended  to  provide  a  basis  for  any  type  of emergency condition or diagnosis. This table only depicts different classification of  blood  pressure.  Consult  your  physician  for  proper  interpretation  of  blood pressure results.  
 17 Appendix  A  Guidance  and  Manufacturer’s Declaration of EMC    Note:  This product needs special precautions regarding EMC and needs to be installed and put  into  service  according  to  the  EMC  information  provided,  and  this  unit  can  be affected by portable and mobile RF communications equipment.  Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in incorrect operation of the unit.     Caution:    This unit has been thoroughly tested and inspected to assure proper performance and operation.  This machine should not be used adjacent to or stacked with other equipment and that if  adjacent  or  stacked  use  is  necessary,  this  machine  should  be  observed  to  verify normal operation in the configuration in which it will be used.  Guidance and manufacture’s declaration – electromagnetic emission The device is intended for use in the electromagnetic environment specified below. The customer of the user of the device should assure that it is used in such an environment. Emission test Compliance Electromagnetic environment – guidance RF emissions CISPR 11 Group 1 The  device use  RF  energy  only  for  its  internal function.  Therefore,  its  RF  emissions  are  very low and are not  likely  to cause  any interference in nearby electronic equipment. RF emission CISPR 11 Class B The  device  is  suitable  for  use  in  all establishments,  including  domestic establishments  and  those  directly  connected  to the public low-voltage power supply network that supplies buildings used for domestic purposes. Harmonic emissions IEC 61000-3-2 Class A Voltage fluctuations/ flicker emissions IEC 61000-3-3 Complies 
 18  Guidance and manufacture’s declaration – electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The customer or the user of device should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance Electrostatic discharge (ESD) IEC 61000-4-2 ±6 kV contact ±8 kV air ±6 kV contact ±8 kV air Floors should be wood, concrete or  ceramic  tile.  If  floor  are covered with synthetic material, the  relative  humidity  should  be at least 30%. Electrical fast transient/burst IEC 61000-4-4 ±2 kV for power supply lines ±2kV for power supply lines Mains  power  quality  should  be that  of  a  typical  commercial  or hospital environment. Surge IEC 61000-4-5 ± 1 kV line(s) to line(s) ±1 kV differential mode Mains  power  quality  should  be that  of  a  typical  commercial  or hospital environment. Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 <5% UT   (>95% dip in UT) for 0.5 cycle  40% UT (60% dip in UT) for 5 cycles  70% UT (30% dip in UT) for 25 cycles  <5% UT   (>95% dip in UT) for 5 sec <5% UT   (>95% dip in UT) for 0.5 cycle  40% UT (60% dip in UT) for 5 cycles  70% UT (30% dip in UT) for 25 cycles  <5% UT   (>95% dip in UT) for 5 sec Mains power quality should be that of a typical commercial or hospital environment. If the user of the device requires continued operation during power mains interruptions, it is recommended that the device be powered from an uninterruptible power supply or a battery.  Power  frequency (50Hz/60Hz) magnetic  field  IEC 61000-4-8 3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial  or  hospital environment. NOTE    UT is the a.c. mains voltage prior to application of the test level.
 19 Guidance and manufacture’s declaration – electromagnetic immunity The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance        Conducted RF IEC 61000-4-6 3 Vrms 150 kHz to 80 MHz 3 Vrms  Portable  and  mobile  RF  communications  equipment should  be  used  no  closer  to  any  part  of  the  device, including  cables,  than  the  recommended  separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance            Where  P  is  the  maximum  output  power  rating  of  the transmitter  in  watts  (W)  according  to  the  transmitter manufacturer  and  d  is  the  recommended  separation distance in meters (m). Field  strengths  from  fixed  RF  transmitters,  as determined  by  an  electromagnetic  site  survey,a  should be  less  than  the  compliance  level  in  each  frequency range.b Interference  may  occur  in  the  vicinity  of  equipment marked with the following symbol:   Radiated RF IEC 61000-4-3   3 V/m 80 MHz to 2.5 GHz   3 V/m NOTE 1        At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2        These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.  a)    Field  strengths  from  fixed  transmitters,  such  as  base  stations  for  radio  (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted  theoretically  with accuracy. To  assess  the  electromagnetic environment  due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location  in  which  the  device  is  used  exceeds  the  applicable  RF  compliance  level  above,  the  device should  be  observed  to  verify  normal  operation.  If  abnormal  performance  is  observed,  additional measures may be necessary, such as re-orienting or relocating the device. b)  Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
 20  Recommended separation distances between   portable and mobile RF communications equipment and the device . The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining  a  minimum  distance  between  portable  and  mobile  RF  communications  equipment (transmitters) and the device as recommended below, according to the maximum output power of the communications equipment. Rated maximum output power of transmitter (W) Separation distance according to frequency of transmitter (m) 150 KHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz 0.01 0.12 0.12 0.23 0.1 0.38 0.38 0.73 1 1.2 1.2 2.3 10 3.8 3.8 7.3 100 12 12 23 For  transmitters  rated  at  a  maximum  output  power  not  listed  above,  the  recommended  separation distance  d  in  meters  (m)  can  be  estimated  using  the  equation  applicable  to  the  frequency  of  the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.  NOTE 1      At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2      These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.