User Manual
Preface
32
Preface
Thank you for selecting the GE006 Blood Glucose Monitoring System. This manual provides all the
information you need to operate this product for accurate test results. Please read this entire manual
before you start any testing.
For people living with diabetes, it is important to regularly monitor blood glucose levels to effectively
reduce complications from the disease. The easy-to-use GE006 Blood Glucose Monitoring System
provides accurate, reliable test results to help you better manage your diabetes.
The GE006 Blood Glucose Monitoring System is designed for in vitro diagnostic use only (for self-testing
by a single user outside the body). Testing requires only a small amount of fresh capillary whole blood
from either the fingertip, palm or forearm.
The GE006 Blood Glucose Monitoring System is manufactured and supported by Bionime Corporation. If
you have any questions or concerns, please contact your local Bionime Customer Service or call us at
886 0800 241 688 GMT+08:00, 08:30~17:30.
The GE006 Blood Glucose Monitoring System is a personal blood glucose monitoring system to be used
for self-testing only. It is not recommended for multiple users.
A healthcare professional should be contacted when Customer Service is not available.
Caution
54
Precaution
- Before using the GE006 Blood Glucose Monitoring System to test your blood glucose, please read all of
the instructions and conduct all of the tests including the quality control test (Refer to page 37).
- Please perform the quality control test regularly to make sure the test results are accurate.
- The GE006 Blood Glucose Meter can only be used with GE006 Blood Glucose Test Strips. Other brands’
test strips should not be used under any circumstances. The use of other brands’ strips may give
inaccurate results.
- The GE006 Blood Glucose Monitoring System is intended for in vitro diagnostic use only. The GE006
Blood Glucose test results using fresh capillary whole blood samples from the fingertip, palm and
forearm are calibrated to be the equivalent to that of plasma testing.
- The GE006 Blood Glucose Monitoring System is intended for self-testing. It should not be used to
screen for or diagnose diabetes mellitus.
- If the GE006 Blood Glucose Meter and Test Strips are exposed to a substantial change in temperature,
please wait 30 minutes before measurement.
- Follow all environmental protection regulations when disposing of batteries.
- Avoid contact with dripping or splashing liquids.
- The minimum blood sample size to test using the GE006 Blood Glucose Monitoring System is 0.75µL:
( )
Blood sample size above 3.0µL might contaminate the test strip port and the meter.
Blood sample size below 0.75µL may cause an inaccurate result or may prevent a meter reading. An Er4
reading will be displayed if the sample size is too small. In this case, repeat the test with a new test strip.
Sample Size Example 0.75µL 1.0 1.5µL 2.0µLµL 3.0µL
Important Safety Notes:
- The GE006 Blood Glucose Meter and Lancing Device are for single patient use. Do not
share them with anyone including other family members.
- All parts of the kit are considered biohazardous and can potentially transmit infectious
diseases, even after following the cleaning and disinfecting procedures.
- Users should wash their hands thoroughly with soap and water before and after handling
the meter, lancing device, or test strips.
Please refer to the
section "Cleaning and Disinfecting Procedures" on page 47.
Table of Contents
76
Table of Contents
About the Blood Glucose Monitoring System
The Blood Glucose Monitoring System ........ ................................... ................
The GE006 Blood Glucose Meter ................................................. ..................................
The GE006 Blood Glucose Test Strip ......................................................... .................
Before Testing
Meter Activation and Battery Change ............................................................................... .......
Setting Up Your Meter - Setting the Date, Time and Volume ............................................... ................
Turning On/Off the Meter ........................................................................................... ....................
Turning On/Off the Bluetooth .............................................................................. .......................
Handling the GE006 Blood Glucose Test Strip ................................................ ..................................
Testing Procedure
Getting Ready for Testing ......................................................................................... .....................
Performing a Blood Glucose Test .............................................. .............................................
Setting Markers .......
Alternative Site Testing ......................................................... ..........................................................
View Window Appearance .................................................... ......................................................
Understanding Test Results and Messages ......................................................... ....................................................
GE006
GE006 ................................................. .........
.............................................................
...............................................................
................................................
.........................
................................................
................................................
...................................
................................................
...................................................
...........................................................................................................................................................................
................................................
.................................................
.............
Quality Control
About Quality Control Testing ......................................................................................................................................................
Performing a Quality Control Test .............................................................................................................................................
Understanding Control Test Results ..........................................................................................................................................
.....................................................
.....................................................
.....................................................
......................................................
.....................................................
..................................................
....................................................................................................................................................................
.........................................................................................
NCC 警語 ...............................................................................................................................................................................................
..................................................
..................................................
.................................................
.................................................
.................................................
.................................................
.................................................
..................................................
Test Memory
Recalling Test Results ......................................................... .........................................................
Recalling Average Test Results ....................................................... ........................................
Caring for Your Meter .............................................................................. ....................................
Cleaning and Disinfecting Procedures .............................................. .................................
Error Messages and Troubleshooting ................................................. ................................
Limitations ................................................................................................. ..........................................
Real Time Synchronize
Specifications ................................................................................................ ....................................
Warranty ......................................................................................................... .....................................
Customer Service .................................................................................................................. ............
Expected Blood Glucose Values Without Diabetes ........................................... ................
Component Manufacturer Information ........................................................... .......................
Log Book .......................................................................................................................... .....................
Warranty Card ...................................................................................................................... .............
Emergency Card ..................................................................................................... ..........................
FEDERAL COMMUNICATIONS COMMISION (FCC) STATEMENT
40
42
43
45
46
47
51
55
56
58
60
61
63
64
65
66
67
82
85
8
10
14
16
18
21
22
23
25
26
32
34
35
31
37
9
10 *
53 *2
6 *
1
8
4 *
7 *
8 *
9 *
The GE006 Blood Glucose Monitoring System
Contents
Your GE006 Blood Glucose Monitoring System consists of several items. Please identify each item, learn
its name and how it is used.
Below are the items included in your GE006 Blood Glucose Monitoring System:
1. Getting Started Guide
2. User's Manual (includes Log Book, Warranty Card, Emergency Card )
3. GE006 Blood Glucose Test Strip Package Insert *
4. Rightest® Control Solution GC550 Package Insert *
5. GE006 Blood Glucose Meter ( with 2 AAA batteries installed )
6. GE006 Blood Glucose Test strips ( 10/25 pcs )*
7. *
8. Lancing Device *
9. Clear Cap *
10. Disposable Sterile Lancets ( 10 pcs ) *
11. Carrying Case ( not shown )*
12. Instruction for the lancing device ( not shown )*
(* Different packages have different bundled items. Some of packages might not include * items.)
*
*
Rightest® Control Solution GC550
Rightest® GD500
The GE006 Blood Glucose Monitoring System
If you want to purchase optional items, please contact the Customer Support Center at
866-0800-241-688.
1110
The Blood Glucose MeterGE006 The Blood Glucose MeterGE006
Left Button ( )
Main Button ( )
Display Window
Right Button ( )
Test Strip Port
Meter Battery
Product name
Battery Cover
Meter Serial Number
1312
Indicates a test result stored in memory
Indicates a test result not included into the Average
function of this meter
Indicates the average result
Indicates a control solution test result
Unit of test result
Test result
Warns when the battery is low or must be replaced
Indicates when to apply the blood sample
Indicates the time in 12H format
The GE006 Blood Glucose Meter The GE006 Blood Glucose Meter
Appears when an error occurs
Appears after you insert test strip into meter
Current date under time mode or testing date under
memory mode
Warns when the operational temperature limit is
exceeded during testing
Displays current time under time mode or testing time
under memory mode
Manufacturing use only
Indicates a meal marker with test result
Indicates that the Bluetooth is turned on
Indicates that the Bluetooth is turned off
Indicates the time period of test results
1514
The GE006 Blood Glucose Test Strip The GE006 Blood Glucose Test Strip
The GE006 Blood Glucose Meter can only be used with GE006 Blood Glucose Test Strips and Rightest™
Control Solution GC550. The use of other test strips or control solutions can lead to incorrect results.
Sample Port
Apply blood/control solution here.
Noble Metal Electrodes
Electrochemical sensor
View WIndow
PRECAUTION
- Close the GE006 Blood Glucose Test Strip vial immediately after removing a test strip.
- Do not reuse GE006 Blood Glucose Test Strips.
- Do not use expired GE006 Blood Glucose Test Strips.
- When you open a new vial of GE006 Blood Glucose Test Strips, record the date on the vial.
Discard the vial of test strips after 3 months from opening.
- Store the GE006 Blood Glucose Test Strip between 4-30°C (or 39-86°F) and in a location (<
90% relative humidity). Do not expose to direct sunlight or heat.
- If the GE006 Blood Glucose Meter and Test Strips are exposed to a substantial change in
temperature, please wait 45 minutes before measurement.
- For detailed information, please refer to the GE006 Blood Glucose Test Strip Package
Insert.
Electrode Contacts
Hold here to insert
test strip into
meter.
17
16
Meter Activation and Battery Change
Your Blood Glucose Meter comes with two AAA, batteries installed. Two new batteries will
provide power to perform approximately 800 tests under normal use with one data transfer. Press the
main button or insert a strip to activate your meter.
GE006
4. The GE006 Blood Glucose Meter
5. Press any button to exit the self-test and enter Setting Mode.
will enter Self-Testing Mode automatically when replacing the battery
(all symbols will appear on the screen).
6. You must set the time and date when replacing the battery. See Chapter “Setting Up Your Meter -
Setting the Date, Time, and Volume" on page 18. Test results are still stored in the memory.
1. Turn the
over. Press and slide the
battery cover to open.
GE006 Blood Glucose
Meter
2. Install the batteries. Be sure to
put the batteries in the correct
direction.
3. Slide the battery cover
back until it snaps into
place.
Meter Activation and Battery Change
CAUTION
- Please follow the local regulation to properly discard a used battery.
- Risk of explosion if battery is replaced by an incorrect type.
You can enter the Setting Mode two ways.
1. Replace the Battery
2. With Battery Inserted
After removing the battery, press the main button several times until there is no signal on screen,
then follow the battery installation steps to replace battery. The GE006 Blood Glucose Meter will
perform a self-test. Press the main button to exit the self-test and enter the Setting Mode.
Press the main button to turn on the GE006 Blood Glucose Meter. Hold down the main button for 7
seconds. During this time the screen will go blank until you hear a beep. After the beep, the meter will
turn on into the Setting Mode. The display screen will show setting data.
NOTE
- Use the Left and Right buttons to select your setting data and press the main button to
confirm each selection. After confirming all of the settings, you will return to the Time Mode.
1918
1)
2)
3)
4a) 4b)
5) 6)
1. Year setting
, press the Left or Right button until you see the
current year. Press the Main button to confirm.
2. Month setting
With the month blinking, press the Left or Right button until you see the
current month. Press the Main button to confirm.
With the year format blinking
Once the year is confirmed, the
Month Setting will appear.
Once the month is
confirmed, the Day Setting will appear.
Once the day is confirmed, the Time
Setting will appear.
3. Day setting
With the day blinking, press the Left or Right button until you see the current
day. Press the Main button to confirm.
Setting Up Your Meter - Setting the Date, Time and Volume Setting Up Your Meter - Setting the Date, Time and Volume
4. Time format 12/24H selection
With the time format blinking, press the Left or Right
button to switch between 12H and 24H. Press the Main
button to confirm.
5. Hour setting
With the hour blinking, press the Left or Right button until
you see the current hour. Press the Main button to
confirm.
6. Minute setting
With the minutes blinking, press the Left or Right button
until you see the current minute. Press the Main button to
confirm.
Once confirmed, the Hour Setting will
appear.
Once the hour is confirmed, the Minute Setting
will appear.
Once the minute is confirmed, the Volume
Setting will appear.
2120
7a) 7b)
NOTE
- If you do not change any settings during Meter Set-up for over 2 minutes, the GE006 Blood
Glucose Meter will leave setting mode and power off automatically.
7. Volume Setting
With the volume blinking, press the Left or Right button to
turn the volume on or off. Press the Main button to confirm
and finish the settings.
8. Setting the Unit of Measurement
With milligrams per deciliter (mg/dL) or millimoles per liter
(mmol/L) blinking, press the Left or Right button to switch.
Then press the Main button to confirm it and finish the
settings.
9. Ending setting
After measurement of unit setting, for confirmation you will
hear a beep (if volume is turned on). All the settings are saved
and completed and will return to time screen.
Setting Up Your Meter - Setting the Date, Time and Volume
1. How to turn on the GE006 Blood Glucose Meter
Press the Main button or Insert a test strip.
2. Manual Power off
To power off the GE006 Blood Glucose Meter, press and hold the Main button for 4 seconds.
3. Auto Power off
The GE006 Blood Glucose Meter will power off automatically after 2 minutes if no buttons are pressed
or no strip is inserted.
Turning On / Off the Meter
8a) 8b)
9)
23
22
Turning On/ Off the Bluetooth
1. Auto turn on the Bluetooth
After leaving the Setting mode, the Bluetooth will turn on automatically and the
" " symbol blinks on the screen.
2. Turn off the Bluetooth
In Time mode, press and hold the Right or Left button for two seconds. The
blinking " " symbol will switch to the “ “ symbol with a beep (If Volume is
turned on).
3. Manual turn on the Bluetooth
In Time mode, press and hold the Right or Left button for two seconds. The
“ “ symbol will switch to the blinking “ “ symbol with a beep (If Volume is
turned on).
1)
2)
NOTE
1.The blinking symbol should turn to steady immediately when
the Bluetooth connection is established.
2. Blinking back and forth “ “ symbol on the screen means the
meter is transferring data.
“ “ 12
Handling the GE006 Blood Glucose Test Strip
How to handle the GE006 Blood Glucose Test Strip.
Inserting the GE006 Blood Glucose Test Strip:
1. Hold the GE006 Blood Glucose Test Strip between your thumb and middle
finger with the view window " " facing up.
2. Put your forefinger on the side of the strip as shown.
3. Insert the GE006 Blood Glucose Test Strip into test strip port until it clicks
and firmly stops.
1
2
3
25
24
Handling the GE006 Blood Glucose Test Strip
Removing the GE006 Blood Glucose Test Strip:
1. Hold the GE006 Blood Glucose Test Strip as shown.
2. Rotate the GE006 Blood Glucose Test Strip counterclockwise and pull up
simultaneously.
3. Take the GE006 Blood Glucose Test Strip out of the test strip port.
Please follow your health care professional's direction and discard used
strips properly.
2
3
Getting Ready for Testing
Before performing a blood glucose test, prepare the
items below :
- GE006 Blood Glucose Meter
- GE006 Blood Glucose Test Strips
(Please check the expiration date on the test strip vial. Do
not use expired test strips)
- Rightest™ GD500 Lancing device
- Sterile lancet
- Alcohol swab (optional)
PlungerBase
Lancet carrier
Release button
Depth adjustable cap
1
10
9
7
27
26
6
Performing a Blood Glucose Test
1. Hold the depth adjustable cap in one hand and hold the base
with the other hand. Bend the cap down. When a gap
appears between the cap and base, pull the cap and base in
opposite directions.
2. Pull off the depth adjustable cap.
3. Insert a new disposable lancet firmly into lancet carrier.
4. Twist off and set aside the protective cover of the disposable
lancet.
5. Replace the depth adjustable cap.
6. Choose a depth of penetration by rotating the top portion of
the depth adjustable cap until your desired setting is visible
in the window. Settings are based on skin type " " for
soft or thin skin; " " for average skin; " " for thick or
calloused skin.
12
34
5
Performing a Blood Glucose Test
7. Hold the base in one hand and pull on the plunger with the
other hand. The device will be cocked. Release the plunger
and it will automatically move back to its original position
near the base.
8. Wash your hands with warm soapy water and dry
thoroughly.
9. Take one GE006 Blood Glucose Test Strip from the vial.
Close the vial cap immediately.
10. Insert the strip into the test strip port of the GE006 Blood
Glucose Meter with the view window facing up.
11. Once the strip is inserted, all symbols will appear on your
meter display and will be accompanied by a beep (if
volume is turned on).
8
NOTE
- The will automatically detect the Code
number on the strip. You do not have to check the Code number on
the meter display and strip vial.
GE006 Blood Glucose Meter
11
29
28
Performing a Blood Glucose Test
12. A blood drop icon will appear on the display window and
will be accompanied by a beep (if volume is turned on).
Apply the blood sample within 2 minutes.
13. Place the lancing device against your fingertip and press
the release button.
14. Touch and hold the blood drop to the edge of sample
port until the view window is filled with blood. If the view
window is not completely filled with blood the test will
not start. If the blood sample was insufficient, discard
the test strip and repeat with a new GE006 Blood
Glucose Test Strip.
15. Countdown Mode will begin on the display window. After
5 seconds, your test result will appear.
15
12
14c
14a 14b
13
Performing a Blood Glucose Test
16. Pull off the depth adjustable cap. Without touching the used
disposable lancet, insert the lancet tip into the protective
lancet cover.
17. Hold the release button in one hand (see picture 17, step 1)
and pull on the plunger with the other hand (see picture 17,
step 2) to safely eject the used disposable lancet.
18. Discard the used disposable lancet into an appropriate
puncture-proof or biohazard container.
19. Replace the depth adjustable cap after finishing the test.
16
1
2
17
18 19
31
30
CAUTION
- Do not apply your blood drop to the sample port on the strip until you see the " " appear.
The GE006 Blood Glucose Meter is performing an internal test and will display " " and
“ " if you apply blood too soon. If this occurs, please repeat the test with a new test
strip.
- Record the opening date of a new test strip vial. Discard the vial of test strips
3 months after opening.
- Always keep the metal contacts of the test strip port clean. If any dust or
impurities are present, please clean with a small, soft brush.
- The GE006 Blood Glucose Monitoring System and Rightest™ GD500 Lancing
Device are intended for a single user only. Do not share these devices with anyone,
including other family members.
- All parts of this kit are considered biohazards and can potentially transmit infectious
diseases, even after you have performed the cleaning and disinfecting procedure.
- Users should wash hands thoroughly with soap and water after handling the meter,
lancing device, and test strips.
- Please refer to the section "Cleaning and Disinfecting Procedures "on page 47 for cleaning
and disinfecting instructions.
Performing a Blood Glucose Test Setting Markers
According to actual life status, you can set meal markers with test result.
You have to perform meal marker setting immediately after the meter displays the reading of your test
result.
1. After test, the test result is displayed on the screen with time, date and flashing “ “.
2. Press the Right or Left button to choose among “ “ (before meal), “ “ (after meal), ” ” (no
marker) and “ “. Then press the Main button to confirm and save this record.
Before meal: You execute this blood glucose test before a meal.
After meal: You execute this blood glucose test after a meal.
(Blank) no marker: regular test or definition by user.
No average: You will exclude this test result from average calculation.
Please refer to "Excluding test results from the average calculation "on page 45 for
more information.
NOTE
If no button pressed, the marker will be set blank automatically after 15 seconds.
33
32
Pressing for
a few seconds
Releasing
button
Alternative site testing: palm or forearm blood sampling
- To perform alternative site testing, install the clear cap for your
.
( For detailed information, refer to the instruction manual for the
. )
- To increase the blood flow, massage the intended puncture area of your palm or
forearm for a few seconds.
- Immediately after massaging the intended puncture area, press and hold the
lancing device with the clear cap against palm or forearm.
- Press the release button.
- Continue holding the lancing device against your palm or forearm and gradually
increase pressure for a few seconds until the blood sample size is sufficient. DO
NOT test on the palm or forearm if you are testing for hypoglycemia (Low blood
glucose).
Rightest™ GD500
Lancing Device
Rightest™ GD500
Lancing Device
(
Palm
Forearm
Alternative Site Testing Alternative Site Testing
CAUTION
- Test results may vary if blood samples are taken from different sites or under certain
conditions where glucose levels are changing rapidly such as: following a drink, a meal, an
insulin dose or exercise. In these cases, only the fingertip should be used.
- DO NOT test on the palm or forearm if you are testing for hypoglycemia (Low blood glucose).
- Fingertip samples can show the rapid change in glucose faster than palm or forearm
samples.
- Consult your healthcare professional before sampling from your palm or forearm.
- Since the blood flow in the forearm and palm is slower than in the fingertip,
sites other than
fingertip.
- If you use the normal lancing device for your palm or forearm, the blood sample may not be
sufficient for meter operation.
- Periodically compare the test system to a laboratory test system that is known to be well
maintained and monitored by a healthcare provider.
use the clear
cap provided with the GD500 Lancing Device when testing Rightest™
34 35
View Window Appearance
Make sure your blood sample covers the
whole area of the view window to get an
accurate test result. An insufficient blood
sample will result in an error message ("Er4").
If this occurs, repeat the test with a new test
strip.
Insufficient blood sample Sufficient blood sample
CAUTION
- Check the expiration date printed on the strip vial every time you use a test strip. Do not use
expired GE006 Blood Glucose Test Strips.
- Use each GE006 Blood Glucose Test Strip immediately after removing it from the vial.
- Do not reuse GE006 Blood Glucose Test Strips.
-
- Apply the blood sample only on the sample port of the test strip.
Do not perform testing immediately after moving from one ambient temperature to another
(e.g. after coming indoors from the outside). Allow 45 minutes for the temperature of the
meter and the test strips to equilibrate.
Understanding Test Results and Messages
If your blood glucose result is unusually high or low, or if you question your test results, repeat the test
with a new GE006 Blood Glucose Test Strip.
You can also run a Quality Control Test with the Control Solution GC550 to check your GE006
Blood Glucose Meter and GE006 Test Strip (Refer to “Performing a Quality Control Test” on page 40) .
If the test result still remains unusually high or low, contact your healthcare professional immediately.
If you are experiencing symptoms that are not consistent with your blood glucose test results and you
have made sure to follow all instructions in this manual, contact your healthcare professional
immediately.
Consult your health care provider for appropriate disposal of used test strips and lancets.
Rightest™
37
36
Understanding Test Results and Messages
The Meter displays results between 10 and 600 mg/dL.
If the test result is below 10 mg/dL , “ ” will appear on the screen. Please
repeat your test with by a new test strip.
If you still get a “ ” result, contact your healthcare professional.
If the test result is above 600 mg/dL, “ ” will appear on the screen. Please
repeat your test again with a new test strip.
If you still get a “ ” result, contact your healthcare professional.
GE006 Blood Glucose
About Quality Control Testing
What is a Quality Control Test ?
To ensure proper monitoring function, it is necessary to regularly perform a quality control test.
Use one of the Control Solution GC550 when testing your GE006 Blood Glucose Monitoring
System in the Control Solution Mode. If the test result is within the Control Solution Range printed on the
strip vial label, the GE006 Blood Glucose Monitoring System passes the Quality Control Test and your
GE006 Blood Glucose Monitoring System is working properly.
Rightest™
Control Solution Range:
Example of Control Solution Range printed on
your test strip vial label.
Control Solution Range mg/dL
Level 2: 107-145
Level 4: 315-426
NOTE
- Level 2 of is provided in your starter kit. If you want to
purchase or another control solutions,
Rightest™ Control Solution GC550
Level 4 Level 2 please contact Bionime Customer
Service at 886 0800 241 688.
3938
About Quality Control Testing
When should a Quality Control Test Be Performed ?
- To ensure that your GE006 Blood Glucose Meter and GE006 Blood Glucose Test Strip are
working properly.
- To confirm that you are following the correct testing procedures.
- To prepare for your initial blood glucose test.
- To check the GE006 Blood Glucose Test Strip when you open a new vial of strips.
-
- To practice testing.
Required Items for Quality Control Tests
To perform a quality control test, prepare the items below :
- GE006 Blood Glucose Meter
- GE006 Blood Glucose Test Strips
-
To check your GE006Blood Glucose Meter after it has been dropped, damaged or exposed to liquids.
- If you suspect that your test results are inaccurate, or if your test results are not consistent with the
way you feel.
Control Solution GC550Rightest™
CAUTION
- Each time you open a new bottle of control solution, write the expiration date on the label.
The is good for 3 months after opening the bottle, or
until the expiration date printed on the label, whichever comes first.
Rightest™ Control Solution GC550
- Wipe the bottle cap with a clean tissue before tightly closing the bottle of control solution.
- Close the bottle of control solution tightly immediately after using.
- Check the expiration date before use. (Refer to the
Package Insert).
- Keep control solution bottles out of reach of children.
Rightest™ Control Solution GC550
Example
Exp. Date 2014-11
About Quality Control Testing
41
40
Performing a Quality Control Test
1. Take one GE006 Blood Glucose Test Strip from the vial and close
the vial cap immediately.
2. Insert the GE006 Blood Glucose Test Strip horizontally (not
vertically) with the colored view window facing up into the test
strip port.
3. While the blood drop icon is flashing on the display window, press
and hold the Main button for at least 3 seconds until the " "
symbol appears.
4. You will see the blinking " " icon and " " icon on the
screen, prompting you to apply
.
5. Shake the bottle of well before
opening the cap. Place the cap on a flat surface.
6. Place a drop of control solution onto the top of the cap.
7. Gently touch the sample port of the strip with the control solution
from the top of the cap.
8. The screen will display the count time starting from 5 (you will
hear a beep if the volume is turned on).
Rightest™
Rightest™
Control Solution
GC550
Control Solution GC550
8
OPEN
5
6
12
3
Performing a Quality Control Test
9. Tightly replace the cap on the bottle.
10. The control solution result will appear. Compare your Quality Control Test result
to the Control Solution Range printed on the GE006 Blood Glucose Test Strip vial
label.
Rightest™ Control Solution GC550 CLOSE
9
CAUTION
- Your Control Solution Test results will not be included in the average calculations, however,
they can still can be recalled and viewed. The Control Solution Test result will be shown with
the " " icon on the screen.
- The Control Solution Test should be conducted between 15-40°C (59-104°F).
- Before " " and " " appear, do not touch the control solution to the sample port on the
strip. The GE006 Blood Glucose Meter is performing an internal check. Touching the control
solution to the sample port before prompted will result in an error message: " " and
“ " .
- Do not drip the control solution on to the sample port of the test strip directly.
The reagent on the strip might leak into the bottle of control solution and may
cause the degeneration of the control solution. This could contaminate the meter
via the test strip port.
- Keep the test strip port clean and dry. Clean immediately if the test strip port is
stained or is overly exposed to moisture.
- Do not touch the tip of the control solution bottle. If the tip is touched, clean with water.
and be accompanied by beeps (if volume is turned on)
7
43
42
Understanding Control Test Results
Your control solution test results should fall within the control solution range. If the results are within the
range, the GE006 Blood Glucose Monitoring System is working correctly.
Possible reasons your Control Solution results are out of the range :
- Your Rightest™ Control Solution GC550 is expired or was first opened more than 3 months ago.
- Your GE006 Blood Glucose Test Strip has expired.
- You left the cap of the Blood Glucose Test Strip vial or the control solution off for a period of time.
- You did not perform the test procedure correctly.
- The GE006 Blood Glucose Meter or GE006 Blood Glucose Test Strip have malfunctioned.
If Rightest™ Control Solution GC550 results are out of range, your GE006 Blood Glucose Monitoring System
may not be working properly. Repeat the Quality Control Test. If your control solution results are still out of
range, do not use GE006 Blood Glucose Meter to test your blood glucose. please contact Bionime
Customer Service at 886 0800 241 688.
Control Solution Range
88~120 mg/dL 250~340 mg/dL
Level 2 Level 4
Recalling Test Results
The Blood Glucose Meter is able to automatically store a maximum of 500 test results with time
and date. If your meter has stored 500 results, the newest test result will replace the oldest one.
To recall your test memory, start the meter without a test strip inserted.
GE006
1. Press the Main button to switch from the Time Mode to Memory Mode. You will
see the “mem” icon in the upper left corner of the display. The latest test result will
appear with the number “1”.
2. Use either the Left or Right button on the side to review all previous results with
date and time. Results display from the most recent (Sequence no. “1”) to the
oldest (Sequence no. “500”) in the lower right hand corner of the screen.
3. The Quality Control Test result can be recalled from the saved data. When " "
appears under the data, the test was performed using the
. The control solution test result will not be used when recalling
test average results.
Rightest™ Control
Solution GC550
45
44
Recalling Test Results
1. Press the Main button to switch the screen to Average Mode.
2. On the average screen, use either the Left or Right button to view 1-day, 7-days, 14-days, 30-days,
60-days or 90-days averages.
3. The number shown in the lower right hand corner indicates how many test results have been
calculated.
4. Excluding test results from the average calculation:
You may exclude the test result from average calculation right after each test.
You have to perform this exclusion immediately after the meter displays the reading of your test result.
Press the Right or Left button to choose " " then press the main button to confirm. The test
result is now excluded from averaging.
The GE006 Blood Glucose meter provides several average test results. View the 1-day, 7-days, 14-days,
30-days, 60-days and 90-days average test results for better blood glucose monitoring.
Recalling Average Test Results
5. Quick Searching:
To view test results in sequential order automatically, enter the Memory Mode. Press the Right or Left
button for 2 seconds. (The Right button is for viewing from the most current to the oldest test results.
The Left button is for viewing from the oldest to the most current test results). Release the button to
stop at any particular test result.
NOTE
- Test results must exist
during the desired time interval. For example: To get a 14-day average on 1/30/14, you
must have test results dated between 1/17 and 1/30. If no test results exist during that time
frame you will not receive an average.
- To exclude test results from average calculation, you need to remain in Test Mode and
perform the exclusion right after you get the test result. Once you move to MEM Mode,
Average Mode or Time Mode, you will not be able to perform the exclusion.
- The "Lo", "Hi" results, control solution test results and test results under abnormal
temperature < 50°F (10°C), > 104°F (40°C) will be excluded from average calculations.
The Average function requires that the correct time and date is set.
47
46
Caring for Your Meter
Maintenance
Keep your meter and test strip free of dust, water or any other liquid. Store the meter in the carrying
case when not in use. If your meter is dropped or damaged, perform a quality control test with the
control solution before performing a blood glucose test.
Cleaning Meter
Clean the outside of the meter with a damp cloth and mild soap/detergent. Do not get the test strip port
wet.
Cleaning Test Strip Port
If your test strip port is stained with blood, control solution or any liquid, please use a dry tissue or
alcohol swab to clean it immediately. Do not use anything wet to clean it. Perform a quality control test
to ensure the Blood Glucose Meter is working properly. GE006
Cleaning and Disinfecting Procedures
Indirect transmission of Human Immunodeficiency Virus (HIV), Hepatitis B Virus (HBV) and Hepatitis C
Virus (HCV) during the delivery of healthcare services has been increasingly reported. Persons using
blood glucose monitoring systems have been identified as one risk group due to the shared use of
fingerstick (lancing) devices and point of care blood testing devices.
The cleaning procedure is to remove dust, blood and body fluid from the surface and should be
performed whenever the meter or lancing device is visibly dirty. Performing the cleaning procedure
once per week is recommended. The disinfecting procedure is necessary to kill pathogens such as HIV,
HBV and HCV on the device. This procedure should be performed periodically; once per week is
recommended.
If the meter is being operated by a second person who is providing testing assistance to the user, the
meter and lancing device should be decontaminated prior to use by the second person. NOTE: the
cleaning procedure can only remove visible contaminates from surfaces. Only the disinfecting
procedure can eliminate non-visible pathogens.
The meter and lancing device should be cleaned before being disinfected.
DisCide® ULTRA DISINFECTING TOWELETTES may be used to clean and disinfect the meter and lancing
device. Users may purchase the disinfecting wipes from the manufacturer (Palmero Health Care),
distributors or major on line retail sites such as amazon.com. You can contact Palmero Health Care
customer service at 1-800-344-6424 and customerservice@Palmerohealth.com.
49
48
Cleaning and Disinfecting Procedures
To clean the meter:
1. Throughly wipe the entire surface of the meter with disinfecting wipes
listed above to clean any possible dirt, dust, blood and other body
fluids.
To disinfect the meter:
2. Take another disinfecting wipe and wipe the meter thoroughly. (Note: All
blood and body fluids should be cleaned from surface before
performing the disinfecting procedure)
3. Allow the surface to remain wet for 1 minute.
4. Allow to air dry.
NOTE
- Clean and disinfect the outside of the device only. Do not remove battery cover when
cleaning and disinfecting.
NOTE
- Your GE006 Blood Glucose Meter has been tested to ensure that there is no change in the
performance or external materials of the device after 550 cleaning cycles and 550
disinfecting cycles. The testing simulates 2 cleaning and disinfecting cycles per week over
the typical life of the meter (5 years).
1
Allow to remain
wet for 1 minute
Cleaning and Disinfecting Procedures
To clean the lancing device:
1. Throughly wipe the entire surface of the lancing device with
disinfecting wipes listed above to clean any possible dirt, dust, blood
and other body fluids.
To disinfect the lancing device:
2. Take another disinfecting wipe and wipe the Lancing device
thoroughly. (Note: All blood and body fluids should be cleaned from
surface before performing disinfecting procedure)
3. Allow the surface to remain wet for 1 minute.
NOTE
- Your GD500 Lancing Device has been tested to ensure that there is no
change in the performance of the device after 550 cleaning cycles and 550 disinfecting
cycles. The testing simulates 2 cleaning and disinfecting cycles per week over the typical
life of the meter (5 years).
Rightest™
2
3
Allow to remain
wet for 1 minute
1
Cleaning and Disinfecting Procedures
CAUTION
- Users should wash their hands thoroughly with soap and water after handling the meter,
lancing device or test strips.
- Please examine your LCD screen, test strip port, buttons and surface of your meter and
lancing device after cleaning and disinfecting cycles. Stop using the meter and/or lancing
device if any of the following occur:
- Thin, sliver streaks appear on the screen,
- The screen becomes cracked, soft, dissolved, brittle or swollen.
- You are unable to turn on/off your meter, operate the left/ right button, the lancing device
release button or depth adjustable cap.
- You are unable to enter meter settings, function modes or recall your testing results.
If you have any questions or concerns, please contact your GE006 Blood Glucose
Monitoring System authorized representative or call the Customer Service at
886 0800 241 688.
51
50
Error Messages and Troubleshooting
Temperature Error
1. In order to get accurate testing results, please test between 50-104 °F ( 10-40 °C).
2. When the temperature is below 43 °F (6 °C) or over 111 °F (44 °C), the meter will
not function and the " " symbol will blink.
3. If the GE006 Blood Glucose Meter and Test Strips are exposed to a substantial
change in temperature, please wait 45 minutes before measurement.
Error Messages and Troubleshooting
Battery Error
1. The " " symbol blinks when the battery power is low. Please change the
battery as soon as possible. You can still test.
and " “ symbols blink when the battery is too low. Meter
cannot test. Please change the battery immediately.
2. The " "
Strip Error
1. When the test strip is inserted incorrectly, the meter will beep 4 times while
the “ ” symbol flashes. Do not apply a sample on the test strip.
Please re-insert the unused test strip correctly (refer to “Handling the
GE006 Blood Glucose Test Strip“ on Page 24 for more information).
2. If, after performing the required steps, the “ ” and flashing “ ”
symbols appear on the screen, it means the GE006 Test strip has been
inserted incorrectly more than twice. Please, re-insert the unused test strip
again after reviewing the steps on how to insert a test strip correctly on p.
24 or call Customer Service for support. This error message can also
indicate that you may have used the wrong test strips. Please check the
test strip vial to ensure you are using the GE006 Test strips.
Sampling Error
Please do not apply the blood to the sample port of the test strip before the
meter displays ” ”. Discard the test strip If the meter shows “ “.
1a 1b
2
12
53
52
Error Messages and Troubleshooting
Er1 -
strip.
Er2 - The meter has malfunctioned. Do a Quality Control Test or reinstall the
batteries to check if the meter works properly.
Er3 -The signal transmission is disrupted. Repeat the test using a new test strip.
Er4 - The blood sample volume is insufficient. Repeat the test using a new test
strip.
Er5 - An issue calibrating the Meter has occurred. Please follow the steps
below:
(1) Remove the test strip from the meter.
(2) Turn off the meter (press and hold the Main button for 3 seconds).
(3) Press the main button to turn ON the meter.
If after turning on the Meter, you do not see Er5, your meter is functioning
properly and able to perform a test.
If one of the above error messages still appears, please contact Customer
Service at 886-0800-241-688.
The inserted test strip has been used or damaged. Please use a new test
Meter Malfunction
If the GE006 Blood Glucose Meter will not turn on, please follow the steps below:
1. Open the battery cover and remove the batteries.
2. Wait for 5 minutes and reinsert the batteries as described as instructed on page 16, (Meter Activation and
Changing the Battery).
The meter should work normally after finishing the above steps. If the meter still does not work, please
contact Customer Service at 886-0800-241-688.
Error Messages and Troubleshooting
55
54
CAUTION
A blood sample should ONLY be applied to the test strip after the test strip has been inserted
correctly and the Meter's screen is showing the image of the test strip and a blood drop
flashing symbol. If the blood drop flashing symbol is NOT showing on the screen, do not
apply a sample to the test strip. Please re-insert the unused test strip correctly. The screen
MUST show the image of the test strip accompanied by the blood drop symbol flashing (this
symbol take 3 seconds to show after correctly inserting the test strip) in order for a sample to
be applied. Please consult your User Manual and/or contact Customer Service for support on
how to correctly insert a GE006 Test strip.
Limitations
- GE006 Blood Glucose Monitoring System is not intended for serum or plasma testing.
- Inaccurate test results may be obtained at altitudes greater than 10,000 feet (3048 meters) above sea
level.
- Severe dehydration and excessive water loss may cause inaccurately low results.
- High concentrations of Uric acid 10.0 mg/dL, L-Dopa 3 mg/dL, Dopamine 2 mg/dL, Ascorbic
acid (Vitamin C) 5 mg/dL may result in inaccurate test results.
- Not for neonatal use.
- For over-the-counter use. For single-patient use only. For in vitro diagnostic use only.
- Not for screening or diagnosis of diabetes mellitus.
- Not for use on critically ill patients in shock, dehydrated patients or hypo-osmolar patients.
- Alternative site testing (AST) should only be performed during steady-state times (when glucose is not
changing rapidly).
- Alternative Site Testing should not be used to calibrate continuous glucose monitoring systems
(CGMS).
- Alternative site testing should not be performed if testing for hypoglycemia (Low blood glucose).
- Results from alternative site testing should not be used in insulin dose calculation.
≧ ≧ ≧
≧
Real Time Synchronize
57
56
Real Time Synchronize
The GE006 Blood Glucose Meter can be used with the HeartSquare App, a mobile application. After
pairing the GE006 Blood Glucose Meter with the mobile phone by Bluetooth, the HeartSquare App will
automatically detect any activities on the GE006 Blood Glucose Meter. Upon detection or activation of
the GE006 Blood Glucose Meter, the HeartSquare App will display the current status of the GE006 Blood
Glucose Meter, as well as a step by step walk through of performing a new blood glucose test.
A blood drop icon will appear on the display window of Meter. And the indication of applying blood will
also appear on the screen of your mobile phone.
Countdown Mode will begin both on the display window of the meter and the screen of your mobile
phone.
NOTE
- For more information of HeartSquare App, please visit www.heartsquare.com.
- The screenshots in this User's Manual were taken from a mobile phone using Android 4.3+.
Appearance may vary slightly depending on the operating system and manufacturer of
your mobile phone.
FEDERAL COMMUNICATIONS COMMISION (FCC) STATEMENT
59
58
FEDERAL COMMUNICATIONS COMMISION (FCC) STATEMENT
15.21
You are cautioned that changes or modifications not expressly approved by the part responsible for
compliance could void the user's authority to operate the equipment.
15.105(b)
This equipment has been tested and found to comply with the limits for a Class B digital device,
pursuant to part 15 of the FCC rules. These limits are designed to provide reasonable protection against
harmful interference in a residential installation. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the instructions, may cause harmful
interference to radio communications. However, there is no guarantee that interference will not occur in
a particular installation. If this equipment does cause harmful interference to radio or television
reception, which can be determined by turning the equipment off and on, the user is encouraged to try
to correct the interference by one or more of the following measures:
„Reorient or relocate the receiving antenna.
„Increase the separation between the equipment and receiver.
„Connect the equipment into an outlet on a circuit different from that to which the receiver is
connected.
„Consult the dealer or an experienced radio/ TV technician for help.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
1) this device may not cause harmful interference and
2) this device must accept any interference received, including interference that may cause undesired
operation of the device.
61
60
SpecificationNCC警語
50 ~ 104 10 ~ 40°F °C ( )
10 - 90%
Measurement Technology
Sample
Minimum Sample Volume
Measuring Range
Test Time
Memory Capacity
10-600 mg/dL
5 seconds
500 blood glucose test results with date and time
Oxidase Electrochemical Sensor
Capillary whole blood
0.75 microliters
Operating Temperature
Operating Relative Humidity
Power Saving Turns off automatically 2 minutes after last user
action. To turn off manually, press themain button
“ " for 3 seconds.
A) 根據NCC低功 電波 射性電機管 辦法第十條規定:率 輻 理
第十二條
第十四條
經型式認證合格之低功率射頻電機,非經許可,公司、商號或使用者均不得擅
自變更頻率、加大功率或變更原設計之特性及功能。
低功率射頻電機之使用不得影響飛航安全及干擾合法通信;經發現有干擾現象
時,應立即停用,並改善至無干擾時方得繼續使用。 前項合法通信,指依電信
法規定作業之無線電通信。低功率射頻電機須忍受合法通信或工業、科學及醫
療用電波輻射性電機設備之干擾。
63
62
Warranty
Specification
Bionime Corporation warrants that this product will be free from defects in materials and workmanship
for five years from the date of purchase.
GE006
Different models have different specifications. This warranty applies only to the GE006 Blood Glucose
Monitoring System; other models are not covered with this warranty card.
GE is a trademark of General Electric Company and is used under license by Bionime Corporation, No.
100, Sec. 2, Daqing St., South Dist., Taichung City 40242, Taiwan
This warranty does not apply to the performance of a GE006 Blood Glucose Meter that has been
altered, misused, tampered with or abused in any way.
This warranty applies only to the original purchaser of the Blood Glucose Monitoring System.
Please complete and return the enclosed warranty card.
NOTE
- During blood glucose measurement, the GE006 Blood Glucose Meter itself may come into
contact with blood. All parts of the GE006 Blood Glucose Monitoring System are considered
biohazardous and can potentially transmit infectious diseases. Please follow your health
care professional's direction to properly dispose of the used GE006 Blood Glucose
Monitoring System after removing the batteries.
Hematocrit
Power Supply
Battery Life
Meter Dimension
Meter Weight
30 - 60%
Two AAA Batteries
Approximately 800 tests, each test with one data
transfer
85mm x 57mm x 22.5 mm
81.0 ± 5g with batteries
Monitor
Display Area
Meter Storage/ Transportation Conditions
Test Strip Storage/ Transportation Conditions
LCD
( )
( ), < 90% relative humidity
39 mm x 39.5 mm
14 ~ 140°F -10 ~ 60°C
39 ~ 86°F 4 ~ 30°C
64 65
Expected glucose values without diabetes ⁽¹⁾
Customer Service
References
1) American Diabetes Association: Standard of Medical Care in Diabetes 2014, Diabetes Care, Vol.37,
supplement 1, S14-S80, January 2014.
Status
Fasting
Two Hours after meals
Range (mg/dL)
<100
<140
Expected Glucose Values Without Diabetes
For in vitro diagnostic use
Temperature limitation
Manufacturer
LO T
IV D
Lot number Expiry date
For single use only
Method of sterilization using irradiation
Description of used symbols
EC Representive
CE-mark (with No. of notified body)
Consult the instruction for use
Please review all of the instructions to make sure you are performing the steps correctly. If you have
any questions or concerns, please contact your Bionime authorized representative. We will make every
effort to provide assistance and solutions for you.
67
66
Component Manufacturer Information Log Book
Name:
Address:
Home Phone:
Doctor:
Pharmacy:
Insulin/Pills:
In case of emergency contact:
Work Phone
Doctor's Phone
Pharmacy Phone
Log book date From: To:
Blood Glucose Meter, Test Strip, Control Solution
Manufacturer: Bionime Corporation
Product complies with In Vitro Diagnostic Medical Device Directive 98/79/EC. ( CE0197 )
Lancing Device
Manufacturer: Bionime Corporation
Product complies with Medical Device Directive 93/42/EEC
Disposable Sterile Lancets
Manufacturer: SteriLance Medical (SuZhou) Inc.
No.68 LiTangHe RD, XiangCheng, SuZhou, JiangSu 215133, P.R. China
Product complies with Medical Device Directive 93/42/EEC. ( CE0197 )
EC- Rep: EMERGO EUROPE, Molenstraat 15, 2513 BH The Hague, The Netherlands
No. 100, Sec. 2, Daqing St., South Dist., Taichung City 40242, Taiwan
EC-Rep: Bionime GmbH, Tramstrasse 16, 9442 Berneck / Switzerland
E-mail: info@bionime.ch
No. 100, Sec. 2, Daqing St., South Dist., Taichung City 40242, Taiwan
EC-Rep: Bionime GmbH, Tramstrasse 16, 9442 Berneck / Switzerland
E-mail: info@bionime.ch
69
68
Log Book
Breakfast
Lunch
Dinner
Bedtime
Other
Comments
DATE M T W T F S S
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Breakfast
Lunch
Dinner
Bedtime
Other
Comments
DATE M T W T F S S
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Log Book
71
70
Log Book
Breakfast
Lunch
Dinner
Bedtime
Other
Comments
DATE M T W T F S S
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Breakfast
Lunch
Dinner
Bedtime
Other
Comments
DATE M T W T F S S
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Log Book
73
72
Log Book
Breakfast
Lunch
Dinner
Bedtime
Other
Comments
DATE M T W T F S S
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Breakfast
Lunch
Dinner
Bedtime
Other
Comments
DATE M T W T F S S
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Log Book
74 75
Log Book
Breakfast
Lunch
Dinner
Bedtime
Other
Comments
DATE M T W T F S S
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Breakfast
Lunch
Dinner
Bedtime
Other
Comments
DATE M T W T F S S
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Log Book
76 77
Log Book
Breakfast
Lunch
Dinner
Bedtime
Other
Comments
DATE M T W T F S S
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Breakfast
Lunch
Dinner
Bedtime
Other
Comments
DATE M T W T F S S
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Log Book
78 79
Log Book
Breakfast
Lunch
Dinner
Bedtime
Other
Comments
DATE M T W T F S S
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Breakfast
Lunch
Dinner
Bedtime
Other
Comments
DATE M T W T F S S
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Log Book
80
Log Book
Breakfast
Lunch
Dinner
Bedtime
Other
Comments
DATE M T W T F S S
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Breakfast
Lunch
Dinner
Bedtime
Other
Comments
DATE M T W T F S S
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Blood
Glucose
Insulin/
Medication
Log Book
81
83
82
Distributor name and address
PLACE
STAMP
HERE
Warranty CardWarranty Card
Name Tel
Address
Serial No Model
Date of purchase
(Please present this card for replacement purpose)
Distributor name and address
Limited warranty:
1. Bionime warrants to the original purchaser only, that the GE006 Blood Glucose Meter shall be free of any defects in
materials or workmanship and, provided it is not modified, altered or misused, will perform in accordance with
specifications for a period of five (5) years from the original date of purchase.
2. Bionime guarantees the performance of the GE006 Blood Glucose Meter only if used as directed and provided that the
failure to perform has not been caused in whole or in part by the use of test strips that are not GE006 Blood Glucose
Test Strips manufactured by Bionime. Use only GE006 Blood Glucose Test Strips in your GE006 Blood Glucose Meter.
3. The sole obligation of Bionime under this warranty shall be to replace any defective GE006 Blood Glucose Meter. No
other warranties, express or implied, are made. Bionime shall not be responsible for any incidental or consequential
damages.
4. You must contact your distributor for assistance and/or instructions for obtaining a replacement GE006 Blood Glucose
Meter.
85
84
Emergency Card
Warranty Card
Do you have □ Type I □ Type II □ Gestational Diabetes ?
Have you owned a blood glucose monitoring system before ? □ Yes □ No
Which brand/s were you most recently using ?
Will the GE006 Blood Glucose Monitoring System be your primary system ? □ Yes □ No
How often do you test your blood glucose ? Times per day per week
Do you use insulin ? □ Yes □ No Oral medication ? □ Yes □ No
How did you hear about the GE006 Blood Glucose Monitoring System ?
Thank you for answering these questions and for your purchase of the GE006 Blood Glucose Monitoring System.
Thank you for purchasing our product. Please complete and mail this warranty card within 30 days of purchase of your GE006
Blood Glucose Monitoring System.
Name Male/Female Date of Birth
Address
City Country Postal Code
Phone Number
Healthcare Professional Who Recommended City Country
Store/Pharmacy Name Where Purchased City Country
Date of Purchase Model No: Serial/Lot No.
*Please fill this card and carry with you at anytime.
EMERGENCY CARD
GE006 Blood Glucose Monitoring System
• User Name:
• User Phone No.:
• Blood Type:
• Doctor/Hospital:
I am a diabetes patient. If you
find me in a coma or stupor,
please take me to the hospital
on left side. Or call :