Boston Scientific CRMN11906 Implantable Defibrillator User Manual Cognis Part 1 Manual

Boston Scientific Corporation Implantable Defibrillator Cognis Part 1 Manual

Cognis Part 1 Manual

Download: Boston Scientific CRMN11906 Implantable Defibrillator User Manual Cognis Part 1 Manual
Mirror Download [FCC.gov]Boston Scientific CRMN11906 Implantable Defibrillator User Manual Cognis Part 1 Manual
Document ID886096
Application IDtrxodP3XZ0FyXXDU8tFrQw==
Document DescriptionCognis Part 1 Manual
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize150.63kB (1882828 bits)
Date Submitted2008-01-04 00:00:00
Date Available2008-05-20 00:00:00
Creation Date2008-01-04 15:49:03
Producing SoftwareAcrobat Distiller 6.0.1 (Windows)
Document Lastmod2008-01-04 15:49:40
Document TitlePrint Preview - C:\WINNT\TEMP\e3temp_5648\.aptcache\aea05648/tfa05648
Document CreatorPScript5.dll Version 5.2.2
Document Author: SYSTEM

SYSTEM GUIDE
COGNIS™ 100-D
CARDIAC RESYNCHRONIZATION THERAPY
HIGH ENERGY DEFIBRILLATOR
REF N118, N119
CAUTION: Federal law
restricts this device to sale by
or on the order of a physician
trained or experienced in
device implant and follow-up
procedures.
- DRAFT -
Part 1 of 2
- DRAFT -
ABOUT THIS MANUAL
Boston Scientific Corporation acquired Guidant Corporation in April 2006.
During our transition period, you may see both the Boston Scientific and
Guidant names on product and patient material. As we work through the
transition, we will continue to offer doctors and their patients technologically
advanced and high quality medical devices and therapies.
The text conventions discussed below are used throughout this manual.
PRM KEYS
The names of Programmer/Recorder/Monitor (PRM)
keys appear in capital letters (e.g., PROGRAM,
INTERROGATE).
1., 2., 3.
Numbered lists are used for instructions that should be
followed in the order given.
•
Bulleted lists are used when the information is not
sequential.
The screen illustrations used in this manual are intended to familiarize you with
the general screen layout. The actual screens you see when interrogating
or programming the pulse generator will vary based on the model and
programmed parameters.
A complete list of programmable options is provided in the appendix
("Programmable Options" on page A-1). The actual values you see when
interrogating or programming the pulse generator will vary based on the model
and programmed parameters.
The following acronyms may be used in this System Guide:
A:
ABM:
AF:
AFib:
AFR:
AGC:
AIVR:
Atrial
Autonomic Balance Monitor
Atrial Fibrillation
Atrial Fibrillation
Atrial Flutter Response
Automatic Gain Control
Accelerated Idioventricular Rhythm
The following are trademarks of Boston Scientific or its affiliates: AUTONOMIC BALANCE
MONITOR, COGNIS, CONTAK, CONTAK CD, CONTAK RENEWAL, EASYTRAK, ENDOTAK,
RENEWAL, RHYTHM ID, SmartDelay, VITALITY, ZIP.
- DRAFT -
AT:
ATP:
ATR:
AV:
BiV:
BCL:
BOL:
CPR:
CRT:
CRT-D:
ECG:
DFT:
EAS:
EF:
EGM:
EMI:
EP:
FCC:
HE:
HRV:
IBP:
IC:
ICD:
LRL:
LV:
LVPP:
LVRP:
MI:
MPR:
MRI:
MSR:
MTR:
NSR:
PAC:
PAT:
PES:
PMT:
PRM:
PSA:
PVARP:
PVC:
RADAR:
RF:
Atrial Tachycardia
Antitachycardia Pacing
Atrial Tachy Response
Atrioventricular
Biventricular
Burst Cycle Length
Beginning of Life
Cardiopulmonary Resuscitation
Cardiac Resynchronization Therapy
Cardiac Resynchronization Therapy Defibrillator
Electrocardiogram
Defibrillation Threshold
Electronic Article Surveillance
Ejection Fraction
Electrogram
Electromagnetic Interference
Electrophysiology; Electrophysiologic
Federal Communications Commission
High Energy
Heart Rate Variability
Indications-Based Programming
Industry Canada
Implantable Cardioverter Defibrillator
Lower Rate Limit
Left Ventricular
Left Ventricular Protection Period
Left Ventricular Refractory Period
Myocardial Infarction
Maximum Pacing Rate
Magnetic Resonance Imaging
Maximum Sensor Rate
Maximum Tracking Rate
Normal Sinus Rhythm
Premature Atrial Contraction
Paroxysmal Atrial Tachycardia
Programmed Electrical Stimulation
Pacemaker-Mediated Tachycardia
Programmer/Recorder/Monitor
Pacing System Analyzer
Post-Ventricular Atrial Refractory Period
Premature Ventricular Contraction
Radio Detection and Ranging
Radio Frequency
- DRAFT -
RV:
RVRP:
SCD:
SRD:
SVT:
TARP:
TENS:
V:
VFib:
VF:
VRP:
VRR:
VT:
VTR:
Right Ventricular
Right Ventricular Refractory Period
Sudden Cardiac Death
Sustained Rate Duration
Supraventricular Tachycardia
Total Atrial Refractory Period
Transcutaneous Electrical Nerve Stimulation
Ventricular
Ventricular Fibrillation
Ventricular Fibrillation
Ventricular Refractory Period
Ventricular Rate Regulation
Ventricular Tachycardia
Ventricular Tachycardia Response
- DRAFT -
- DRAFT -
CONTENTS
INFORMATION FOR USE .................................................................................................
CHAPTER 1
1-1
New or Enhanced Features................................................................................................
1-3
Device Description..............................................................................................................
1-4
Related Information ............................................................................................................
1-5
Indications and Usage ........................................................................................................
1-6
Contraindications................................................................................................................
1-6
Warnings ............................................................................................................................
1-6
Precautions.........................................................................................................................
1-8
Potential Adverse Events .................................................................................................
1-19
Mechanical Specifications ................................................................................................
1-21
Lead Connections.............................................................................................................
1-21
Items Included in Package ...............................................................................................
1-22
Symbols on Packaging .....................................................................................................
1-22
Characteristics as Shipped...............................................................................................
1-23
X-Ray Identifier.................................................................................................................
1-24
Federal Communications Commission (FCC) ..................................................................
1-25
Industry Canada (IC) ........................................................................................................
1-25
Pulse Generator Longevity ...............................................................................................
1-26
Warranty Information ........................................................................................................
1-27
Product Reliability.............................................................................................................
1-27
- DRAFT -
Patient Counseling Information ........................................................................................
Patient Handbook ......................................................................................................
1-28
1-29
USING THE PROGRAMMER/RECORDER/MONITOR.....................................................
CHAPTER 2
2-1
ZOOM LATITUDE Programming System ...........................................................................
2-2
Indications-Based Programming (IBP) ...............................................................................
2-2
Manual Programming .........................................................................................................
2-5
Software Terminology and Navigation ................................................................................ 2-5
Main Screen................................................................................................................. 2-5
PRM Mode Indicator .................................................................................................... 2-6
ECG/EGM Display ....................................................................................................... 2-7
Toolbar ......................................................................................................................... 2-7
Tabs ............................................................................................................................. 2-8
Buttons......................................................................................................................... 2-8
Icons ............................................................................................................................ 2-8
Common Objects ....................................................................................................... 2-10
Use of Color ............................................................................................................... 2-10
Data Management ............................................................................................................
Patient Information.....................................................................................................
Disk Operations .........................................................................................................
Print............................................................................................................................
2-11
2-11
2-11
2-12
Communicating with the Pulse Generator ........................................................................
ZIP Telemetry.............................................................................................................
Starting a Wanded Telemetry Session .......................................................................
Starting a ZIP Telemetry Session...............................................................................
Ending a Telemetry Session ......................................................................................
ZIP Telemetry Security...............................................................................................
2-12
2-12
2-13
2-13
2-14
2-14
DIVERT THERAPY ..........................................................................................................
2-16
STAT SHOCK ...................................................................................................................
2-17
STAT PACE ......................................................................................................................
2-18
Safety Mode .....................................................................................................................
Backup Pacemaker....................................................................................................
2-18
2-19
- DRAFT -
Backup Defibrillator....................................................................................................
Programming the Device Safety Tachy Mode............................................................
2-19
2-20
TACHYARRHYTHMIA DETECTION..................................................................................
CHAPTER 3
3-1
Device Mode.......................................................................................................................
Ventricular Tachy Mode................................................................................................
Electrocautery Protection Mode...................................................................................
3-2
3-2
3-3
Rate Sensing ......................................................................................................................
Calculating Rates and Refractory Periods ...................................................................
Ventricular Rate Thresholds and Zones.......................................................................
CRT Delivery Zone and Tachyarrhythmia Zones ........................................................
Use of Atrial Information ..............................................................................................
3-3
3-4
3-4
3-5
3-5
Ventricular Detection .......................................................................................................... 3-6
Ventricular Detection Enhancement Suites.................................................................. 3-7
Ventricular Redetection.............................................................................................. 3-11
Ventricular Post-shock Detection Enhancements ...................................................... 3-12
Ventricular Detection Details...................................................................................... 3-13
TACHYARRHYTHMIA THERAPY .....................................................................................
CHAPTER 4
4-1
Ventricular Therapy ............................................................................................................
Ventricular Therapy Prescription..................................................................................
Ventricular Therapy Selection ......................................................................................
Ventricular Redetection after Ventricular Therapy Delivery .........................................
Ventricular Redetection after Ventricular ATP Therapy................................................
Ventricular Redetection after Ventricular Shock Therapy ............................................
4-2
4-2
4-3
4-8
4-8
4-9
Antitachycardia Pacing Therapies and Parameters .........................................................
Burst Parameters .......................................................................................................
Coupling Interval and Coupling Interval Decrement ..................................................
Burst Cycle Length (BCL) ..........................................................................................
Minimum Interval........................................................................................................
Burst Scheme ............................................................................................................
Ramp Scheme ...........................................................................................................
Scan Scheme.............................................................................................................
Ramp/Scan Scheme ..................................................................................................
ATP Pulse Width and ATP Amplitude ........................................................................
Ventricular ATP Time-out ...........................................................................................
4-10
4-11
4-12
4-14
4-14
4-15
4-15
4-16
4-17
4-18
4-18
- DRAFT -
QUICK CONVERT ATP .............................................................................................
4-20
Ventricular Shock Therapy and Parameters.....................................................................
Ventricular Shock Vector............................................................................................
Ventricular Shock Energy...........................................................................................
Charge Time ..............................................................................................................
Waveform Polarity......................................................................................................
Committed Shock/Reconfirmation of the Ventricular Arrhythmia ...............................
4-21
4-21
4-21
4-22
4-22
4-23
PACING THERAPIES ........................................................................................................
CHAPTER 5
5-1
Device Programming Recommendations ...........................................................................
5-2
Maintaining CRT.................................................................................................................
5-4
Pacing Therapies................................................................................................................
5-6
Basic Parameters ...............................................................................................................
Brady Mode..................................................................................................................
Lower Rate Limit (LRL) ..............................................................................................
Maximum Tracking Rate (MTR) .................................................................................
Maximum Sensor Rate (MSR) ...................................................................................
Ventricular Pacing Chamber ......................................................................................
Pulse Width................................................................................................................
Amplitude ...................................................................................................................
Sensitivity...................................................................................................................
5-7
5-7
5-10
5-11
5-13
5-14
5-15
5-16
5-16
Post-Therapy Pacing ........................................................................................................
Post-Shock Pacing Delay ..........................................................................................
Post-Therapy Period ..................................................................................................
5-17
5-17
5-18
Temporary Pacing.............................................................................................................
5-18
Sensors and Trending ......................................................................................................
Sensor Trending.........................................................................................................
Adaptive-rate Pacing..................................................................................................
Accelerometer............................................................................................................
5-19
5-19
5-20
5-21
Atrial Tachy Response......................................................................................................
ATR Mode Switch ......................................................................................................
Ventricular Rate Regulation (VRR) ............................................................................
Biventricular Trigger ...................................................................................................
5-27
5-27
5-32
5-33
- DRAFT -
Atrial Flutter Response (AFR)....................................................................................
PMT Termination........................................................................................................
5-34
5-35
Rate Enhancements .........................................................................................................
Tracking Preference...................................................................................................
Rate Hysteresis..........................................................................................................
Rate Smoothing .........................................................................................................
5-35
5-36
5-36
5-38
Lead Configuration ...........................................................................................................
Left Ventricular Electrode Configuration ....................................................................
5-42
5-42
AV Delay...........................................................................................................................
Paced AV Delay .........................................................................................................
Sensed AV Delay .......................................................................................................
SmartDelay Optimization ...........................................................................................
5-46
5-46
5-48
5-50
Refractory .........................................................................................................................
A-Refractory (PVARP) ...............................................................................................
RV-Refractory (RVRP) ...............................................................................................
LV-Refractory (LVRP).................................................................................................
Left Ventricular Protection Period (LVPP) ..................................................................
Blanking and Noise Rejection ....................................................................................
5-52
5-52
5-54
5-55
5-55
5-56
Noise Response ...............................................................................................................
5-60
Ventricular Tachy Sensing Interactions.............................................................................
5-62
SYSTEM DIAGNOSTICS...................................................................................................
CHAPTER 6
6-1
Battery Status .....................................................................................................................
Capacitor Re-formation................................................................................................
Charge Time Measurement .........................................................................................
Last Delivered Ventricular Shock .................................................................................
6-2
6-5
6-5
6-6
Lead Tests ..........................................................................................................................
Intrinsic Amplitude Test ................................................................................................
Lead Impedance Test...................................................................................................
Pace Threshold Test ....................................................................................................
6-6
6-7
6-8
6-8
- DRAFT -
PATIENT DIAGNOSTICS...................................................................................................
CHAPTER 7
7-1
Therapy History ..................................................................................................................
7-2
Trends ................................................................................................................................
7-3
Arrhythmia Logbook ........................................................................................................... 7-5
Histograms................................................................................................................. 7-11
Heart Rate Variability (HRV) ...................................................................................... 7-12
Counters .................................................................................................................... 7-15
Patient Triggered Monitor .................................................................................................
7-17
ELECTROPHYSIOLOGIC TESTING.................................................................................
CHAPTER 8
8-1
EP Test Features ................................................................................................................
Temporary EP Mode ....................................................................................................
Backup Ventricular Pacing During Atrial EP Testing ....................................................
EP Test Screen ............................................................................................................
8-2
8-2
8-2
8-2
Induction Methods ..............................................................................................................
VFib Induction ..............................................................................................................
Shock on T Induction ...................................................................................................
Programmed Electrical Stimulation (PES) ...................................................................
50 Hz/Manual Burst Pacing .........................................................................................
8-4
8-5
8-6
8-7
8-9
Commanded Therapy Methods ........................................................................................
Commanded Shock ...................................................................................................
Commanded ATP.......................................................................................................
8-10
8-10
8-11
IMPLANT INFORMATION .................................................................................................
CHAPTER 9
9-1
Implanting the Pulse Generator..........................................................................................
Step A: Check Equipment .........................................................................................
Step B: Interrogate and Check the Pulse Generator.................................................
Step C: Implant the Lead System..............................................................................
Step D: Take Baseline Measurements ......................................................................
Step E: Form the Implantation Pocket.......................................................................
Step F: Connect the Leads to the Pulse Generator ..................................................
Step G: Evaluate Lead Signals .................................................................................
9-2
9-2
9-3
9-3
9-4
9-5
9-6
9-9
- DRAFT -
Step H: Program the Pulse Generator ....................................................................
Step I: Implant the Pulse Generator ........................................................................
Step J: Complete and Return the Implantation Form ..............................................
9-10
9-11
9-12
POST IMPLANT INFORMATION.....................................................................................
CHAPTER 10
10-1
Follow Up Testing .............................................................................................................
Predischarge Follow Up.............................................................................................
Routine Follow Up......................................................................................................
10-2
10-2
10-2
Post Implant features........................................................................................................
Sensitivity Adjustment................................................................................................
Beeper Feature ..........................................................................................................
Magnet Feature..........................................................................................................
10-3
10-3
10-4
10-5
Explantation......................................................................................................................
10-8
PROGRAMMABLE OPTIONS...........................................................................................
APPENDIX A
A-1
CLINICAL STUDY - COMPANION ....................................................................................
APPENDIX B
B-1
Clinical Study Populations ..................................................................................................
B-1
Summary ............................................................................................................................
B-1
Observed Adverse Events..................................................................................................
Prior History .................................................................................................................
Study Background........................................................................................................
Adverse Event Definitions............................................................................................
Deaths..........................................................................................................................
B-2
B-2
B-2
B-3
B-8
Study Design ......................................................................................................................
B-9
Inclusion/Exclusion Criteria ..............................................................................................
B-10
Endpoints..........................................................................................................................
B-12
Follow-up Schedule ..........................................................................................................
B-15
Demographic Data............................................................................................................
B-15
- DRAFT -
Patient Accountability and Follow-up Duration .................................................................
B-16
Data Analysis and Results for Primary Endpoint and Secondary All-Cause Mortality
Endpoint ...........................................................................................................................
B-20
Additional Study Data .......................................................................................................
B-25
Additional Outcome Measures .........................................................................................
B-25
Data Analysis and Results - CRT-D System Safety .........................................................
B-29
Data Analysis and Results - COMPANION Sub-study .....................................................
B-30
CLINICAL STUDY - DECREASE HF.................................................................................
APPENDIX C
C-1
Summary ............................................................................................................................
C-1
Study Design ......................................................................................................................
C-1
Follow-up Schedule ............................................................................................................
C-2
Inclusion/Exclusion Criteria ................................................................................................
C-2
Demographic Data..............................................................................................................
C-4
Study Results .....................................................................................................................
C-6
CLINICAL STUDY - CONTAK CD .....................................................................................
APPENDIX D
D-1
Clinical Study Populations ..................................................................................................
D-1
Summary ............................................................................................................................
D-1
Observed Adverse Events..................................................................................................
Deaths..........................................................................................................................
D-1
D-5
Study Design ......................................................................................................................
D-5
Inclusion/Exclusion Criteria ................................................................................................
D-6
Follow-up Schedule ............................................................................................................
D-7
- DRAFT -
Demographic Data..............................................................................................................
D-8
Endpoints............................................................................................................................
D-8
Study Results .....................................................................................................................
D-9
Focused Confirmatory Study ............................................................................................
Study Design..............................................................................................................
Demographic Data .....................................................................................................
Inclusion Criteria ........................................................................................................
Major Differences Between CONTAK CD and Focused Confirmatory Study
Patients ......................................................................................................................
Endpoints ...................................................................................................................
Study Results.............................................................................................................
D-30
D-30
D-30
D-31
CLINICAL STUDY - CONTAK RENEWAL ........................................................................
APPENDIX E
E-1
Clinical Study Populations ..................................................................................................
E-1
Summary ............................................................................................................................
E-1
Study Design ......................................................................................................................
E-1
Inclusion/Exclusion Criteria ................................................................................................
E-1
Demographic Data..............................................................................................................
E-2
Ventricular Tachyarrhythmia Detection Time ......................................................................
E-3
Holter Study - CONTAK RENEWAL ...................................................................................
Study Design................................................................................................................
Inclusion/Exclusion Criteria..........................................................................................
Demographic Data .......................................................................................................
Programming Parameters ............................................................................................
Endpoints .....................................................................................................................
Study Results...............................................................................................................
E-3
E-3
E-4
E-4
E-5
E-5
E-6
CLINICAL STUDY - SUMMARY OF CRT OPTIMIZATION ALGORITHM VALIDATION
STUDY ...............................................................................................................................
APPENDIX F
F-1
Clinical Study Design..........................................................................................................
F-1
- DRAFT -
D-32
D-33
D-33
Study Results .....................................................................................................................
F-2
Conclusions ......................................................................................................................
F-10
CLINICAL STUDY - VITALITY..........................................................................................
APPENDIX G
G-1
Clinical Study Populations .................................................................................................
G-1
Chronic Implant Study - VITALITY.....................................................................................
G-1
Acute Study - VITALITY.....................................................................................................
G-3
CLINICAL STUDY - SUMMARY OF GDT1000 SENSING ACUTE STUDY......................
APPENDIX H
H-1
Clinical Study Populations ..................................................................................................
H-1
Study Methods....................................................................................................................
H-1
Study Results .....................................................................................................................
H-3
Conclusions ........................................................................................................................
H-7
- DRAFT -
1-1
INFORMATION FOR USE
CHAPTER 1
This chapter contains the following topics:
•
"New or Enhanced Features" on page 1-3
•
"Device Description" on page 1-4
•
"Related Information" on page 1-5
•
"Indications and Usage" on page 1-6
•
"Contraindications" on page 1-6
•
"Warnings" on page 1-6
•
"Precautions" on page 1-8
•
"Potential Adverse Events" on page 1-19
•
"Mechanical Specifications" on page 1-21
•
"Lead Connections" on page 1-21
•
"Items Included in Package" on page 1-22
•
"Symbols on Packaging" on page 1-22
•
"Characteristics as Shipped" on page 1-23
•
"X-Ray Identifier" on page 1-24
•
"Federal Communications Commission (FCC)" on page 1-25
•
"Industry Canada (IC)" on page 1-25
•
"Pulse Generator Longevity" on page 1-26
•
"Warranty Information" on page 1-27
•
"Product Reliability" on page 1-27
- DRAFT -
1-2
INFORMATION FOR USE
•
"Patient Counseling Information" on page 1-28
- DRAFT -
1-3
NEW OR ENHANCED FEATURES
These pulse generator systems include additional features as compared to
previous products.
Ease of Use
•
•
ZOOMVIEW Programmer Software: the new user interface offers the
following benefits:
–
Clinical focus—features such as patient diagnostic trends and
indications-based programming emphasize the patient’s clinical
condition over device status and parameters.
–
Consistency—ZOOMVIEW software will be available on future pulse
generators, providing the same screens whether you are following a
brady, tachy, or heart failure device.
–
Simplicity—screen complexity is reduced through the use of
progressive disclosure (displaying the information you use frequently
and minimizing the information you only rarely access) and
exception-based reporting.
Indications-Based Programming (IBP): the new ZOOMVIEW feature allows
you to quickly set up programming parameters based on the patient’s
clinical needs and indications.
Tachy Therapy
•
Rhythm ID and Onset/Stability detection: the selection between detection
enhancements provides you the opportunity and flexibility to adjust for
individual patient conditions.
•
QUICK CONVERT ATP: in an attempt to avoid an otherwise scheduled
charge and painful shock for a pace-terminable fast ventricular tachycardia
(VT), the pulse generator delivers one rapid burst of antitachycardia pacing
(ATP) for an episode detected in the ventricular fibrillation (VF) zone.
•
Programmable Shock Vectors: this capability allows you to electronically
change the shocking vectors for added flexibility in treating high
defibrillation thresholds (DFTs).
- DRAFT -
1-4
INFORMATION FOR USE
DEVICE DESCRIPTION
Sensing
•
Sensing is designed to combine the strengths of both implantable
cardioverter defibrillator (ICD) and pacemaker sensing capabilities to
improve detection and therapy by reducing inappropriate mode switching,
pacing inhibition, and shocks.
DEVICE DESCRIPTION
This manual contains information about the COGNIS 100 family of cardiac
resynchronization therapy defibrillators (CRT-Ds) (specific models are listed in
"Mechanical Specifications" on page 1-21).
Therapies
This family of pulse generators has a small, thin, physiologic shape that
minimizes pocket size and may minimize device migration. Pulse generators
within this family provide a variety of therapies, including:
•
Ventricular tachyarrhythmia therapy, which is used to treat rhythms
associated with sudden cardiac death (SCD) such as VT and VF
•
Cardiac Resynchronization Therapy (CRT), which treats heart failure by
resynchronizing ventricular contractions through biventricular electrical
stimulation
•
Bradycardia pacing, including adaptive rate pacing, to detect and treat
bradyarrhythmias and to provide cardiac rate support after defibrillation
therapy
Cardioversion/defibrillation therapies include:
•
A range of low- and high-energy shocks using a biphasic waveform
•
The choice of multiple shock vectors:
–
Distal shock electrode to proximal shock electrode and pulse generator
case (TRIAD electrode system)
–
Distal shock electrode to proximal shock electrode (RV Coil to RA Coil)
–
Distal shock electrode to pulse generator case (RV Coil to Can)
- DRAFT -
INFORMATION FOR USE
RELATED INFORMATION
1-5
Leads
The pulse generator has independently programmable outputs and accepts
the following leads:
•
One IS-11 atrial lead
•
One IS-1 coronary venous pace/sense lead
•
One LV-1 coronary venous pace/sense lead
•
One DF-1/IS-12 cardioversion/defibrillation lead
The pulse generator and the leads constitute the implantable portion of the
pulse generator system.
PRM System
These pulse generators can be used only with the ZOOM LATITUDE
Programming System, which is the external portion of the pulse generator
system and includes:
•
Model 3120 Programmer/Recorder/Monitor (PRM)
•
Model 2868 ZOOMVIEW Software Application
•
Model 6577 Accessory Telemetry Wand
You can use the PRM system to do the following:
•
Interrogate the pulse generator
•
Program the pulse generator to provide a variety of therapy options
•
Access the pulse generator’s diagnostic features
•
Perform noninvasive diagnostic testing
•
Access therapy history data
RELATED INFORMATION
Refer to the lead’s instruction manual for implant information, general warnings
and precautions, indications, contraindications, and technical specifications.
Read this material carefully for implant procedure instructions specific to the
chosen lead configurations.
1.
2.
IS-1 refers to the international standard ISO 5841.3:2000.
DF-1 refers to the international standard ISO 11318:2002.
- DRAFT -
1-6
INFORMATION FOR USE
INDICATIONS AND USAGE
The Physician’s Technical Manual is packaged with the pulse generator. It
provides the technical information needed at implant.
Refer to the PRM system Operator’s Manual for specific information about the
PRM such as setup, maintenance, and handling.
INDICATIONS AND USAGE
Boston Scientific cardiac resynchronization therapy defibrillators (CRT-Ds) are
indicated for patients with moderate to severe heart failure (NYHA III/IV) who
remain symptomatic despite stable, optimal heart failure drug therapy and have
left ventricular (LV) dysfunction (EF ≤ 35%) and QRS duration ≥ 120 ms.
CONTRAINDICATIONS
There are no contraindications for this device.
WARNINGS
General
•
Labeling knowledge. Read this manual thoroughly before implanting the
pulse generator to avoid damage to the system. Such damage can result in
patient injury or death.
•
Avoid shock during handling. Program the pulse generator Tachy
Mode(s) to Off during implant, explant, or postmortem procedures to avoid
inadvertent high voltage shocks.
•
Backup defibrillation protection. Always have sterile external and
internal defibrillation protection available during implant. If not terminated in
a timely fashion, an induced ventricular tachyarrhythmia can result in the
patient’s death.
•
Resuscitation availability. Ensure that an external defibrillator and
medical personnel skilled in CPR are present during post-implant device
testing should the patient require external rescue.
•
Protected environments. Advise patients to seek medical guidance
before entering environments that could adversely affect the operation
of the active implantable medical device, including areas protected by a
warning notice that prevents entry by patients who have a pulse generator.
- DRAFT -
INFORMATION FOR USE
WARNINGS
1-7
•
Magnetic Resonance Imaging (MRI) exposure. Do not expose a patient
to MR device scanning. Strong magnetic fields may damage the device
and cause injury to the patient.
•
Diathermy. Do not subject a patient with an implanted pulse generator
to diathermy since diathermy may cause fibrillation, burning of the
myocardium, and irreversible damage to the pulse generator because of
induced currents.
Programming and Device Operations
•
Atrial tracking modes. Do not use atrial tracking modes in patients with
chronic refractory atrial tachyarrhythmias. Tracking of atrial arrhythmias
could result in VT or VF.
•
Atrial-only modes. Do not use atrial-only modes in patients with heart
failure because such modes do not provide CRT.
•
Ventricular sensing. Left ventricular lead dislodgement to a position near
the atria can result in atrial oversensing and left ventricular pacing inhibition.
•
Slow VT. Physicians should use medical discretion when implanting this
device in patients who present with slow VT. Programming therapy for slow
monomorphic VT may preclude CRT delivery at faster rates if these rates
are in the tachyarrhythmia zones.
Implant Related
•
Do not kink leads. Kinking leads may cause additional stress on the
leads, possibly resulting in lead fracture.
•
Patch leads. Do not use defibrillation patch leads with the pulse generator
system, or injury to the patient may occur.
•
Separate pulse generator. Do not use this pulse generator with another
pulse generator. This combination could cause pulse generator interaction,
resulting in patient injury or a lack of therapy delivery.
- DRAFT -
1-8
INFORMATION FOR USE
PRECAUTIONS
PRECAUTIONS
Clinical Considerations
•
Pacemaker-mediated tachycardia (PMT). Retrograde conduction
combined with a short PVARP might induce PMT.
Sterilization, Storage, and Handling
•
For single use only; do not resterilize devices. Do not resterilize the
device or the accessories packaged with it because the effectiveness of
resterilization cannot be ensured.
•
If package is damaged. The pulse generator blister trays and contents
are sterilized with ethylene oxide gas before final packaging. When the
pulse generator is received, it is sterile provided the container is intact. If
the packaging is wet, punctured, opened, or otherwise damaged, return the
device to Boston Scientific.
•
Storage temperature and equilibration. Recommended storage
temperatures are 0°C–50°C (32°F–122°F). Allow the device to reach a
proper temperature before using telemetry communication capabilities,
programming or implanting the device because temperature extremes may
affect initial device function.
•
Device storage. Store the pulse generator in a clean area away from
magnets, kits containing magnets, and sources of EMI to avoid device
damage.
•
Use by date. Implant the device system before or on the USE BY date on
the package label because this date reflects a validated shelf life. For
example, if the date is January 1, do not implant on or after January 2.
Implantation and Device Programming
•
Lead system. Do not use any lead with this device without first verifying
connector compatibility. Using incompatible leads can damage the
connector and/or result in potential adverse consequences, such as
undersensing of cardiac activity or failure to deliver necessary therapy.
•
Telemetry wand. Make sure the telemetry wand is connected to the
programmer and that it is available throughout the session. Verify that the
wand cord is within reach of the pulse generator.
- DRAFT -
INFORMATION FOR USE
PRECAUTIONS
1-9
•
STAT PACE settings. When a pulse generator is programmed to STAT
PACE settings, it will continue to pace at the high-energy STAT PACE
values if it is not reprogrammed. The use of STAT PACE parameters will
decrease device longevity.
•
Biventricular pacing therapy. This device is intended to provide
biventricular pacing therapy. Programming the device to provide RV-only
pacing, or programming the RV pace amplitude below the pacing threshold
(resulting in LV-only pacing), is not intended for the treatment of heart
failure. The clinical effects of LV-only or RV-only pacing for the treatment of
heart failure have not been established.
•
Pacing and sensing margins. Consider lead maturation in your choice of
pacing amplitude, pacing pulse width, and sensitivity settings.
•
An acute pacing threshold greater than 1.5 V or a chronic pacing
threshold greater than 3 V can result in loss of capture because
thresholds may increase over time.
•
An R-wave amplitude less than 5 mV or a P-wave amplitude less than
2 mV can result in undersensing because the sensed amplitude may
decrease after implantation.
•
Pacing lead impedance should be within the range of 200 Ω and
2000 Ω.
•
Line-powered equipment. Exercise extreme caution if testing leads using
line-powered equipment because leakage current exceeding 10 µA can
induce ventricular fibrillation. Ensure that any line-powered equipment is
within specifications.
•
Proper programming of the lead configuration. If the Lead Configuration
is programmed to Bipolar when a unipolar lead is implanted, pacing will
not occur.
•
Proper programming of the shock vector. If the shock vector is
programmed to RVcoil>>RAcoil and the lead does not have an RA coil,
shocking will not occur.
•
Replacement device. Implanting a replacement device in a subcutaneous
pocket that previously housed a larger device may result in pocket air
entrapment, migration, erosion, or insufficient grounding between the
device and tissue. Irrigating the pocket with sterile saline solution decreases
the possibility of pocket air entrapment and insufficient grounding. Suturing
the device in place reduces the possibility of migration and erosion.
- DRAFT -
1-10
INFORMATION FOR USE
PRECAUTIONS
•
Defibrillation power surge. Defibrillation that causes a power surge
exceeding 360 watt-seconds can damage the pulse generator system.
•
Programming for supraventricular tachyarrhythmias (SVTs).
Determine if the device and programmable options are appropriate for
patients with SVTs because SVTs can initiate unwanted device therapy.
•
AV Delay. To ensure a high percentage of biventricular pacing, the
programmed AV Delay setting must be less than the patient’s intrinsic PR
interval.
•
Adaptive-rate pacing. Adaptive-rate pacing should be used with care in
patients who are unable to tolerate increased pacing rates.
•
Ventricular refractory periods (VRPs) in adaptive-rate pacing.
Adaptive-rate pacing is not limited by refractory periods. A long refractory
period programmed in combination with a high MSR can result in
asynchronous pacing during refractory periods since the combination can
cause a very small sensing window or none at all. Use dynamic AV Delay
or dynamic PVARP to optimize sensing windows. If you are entering a
fixed AV delay, consider the sensing outcomes.
•
Atrial Tachy Response (ATR). ATR should be programmed to On if the
patient has a history of atrial tachyarrhythmias. The delivery of CRT is
compromised because AV synchrony is disrupted if the ATR mode switch
occurs.
•
Threshold test. During the LV threshold test, RV backup pacing is
unavailable.
•
Left ventricular pacing only. The clinical effect of LV pacing alone for
heart failure patients has not been studied.
•
Do not bend the lead near the lead-header interface. Improper insertion
can cause insulation damage near the terminal end that could result in
lead failure.
•
Shock waveform polarity. For IS-1/DF-1 leads, never change the shock
waveform polarity by physically switching the lead anodes and cathodes
in the pulse generator header—use the programmable Polarity feature.
Device damage or nonconversion of the arrhythmia post-operatively may
result if the polarity is switched physically.
- DRAFT -
INFORMATION FOR USE
PRECAUTIONS
1-11
•
Absence of a lead. The absence of a lead or plug in a lead port may
affect device performance. If a lead is not used, be sure to properly insert
a plug in the unused port.
•
Electrode connections. Do not insert a lead into the pulse generator
connector without first visually verifying that the setscrew is sufficiently
retracted to allow insertion. Fully insert each lead into its lead port and then
tighten the setscrew onto the electrodes.
•
Tachy Mode to Off. To prevent inappropriate shocks, ensure that the
pulse generator’s Tachy Mode is programmed to Off when not in use
and before handling the device. For tachyarrhythmia therapy, verify that
the Tachy Mode is activated.
•
Atrial oversensing. Take care to ensure that artifacts from the ventricles
are not present on the atrial channel, or atrial oversensing may result. If
ventricular artifacts are present in the atrial channel, the atrial lead may
need to be repositioned to minimize its interaction.
•
Defibrillation lead impedance. Never implant the device with a lead
system that has less than 15 Ω total shock lead impedance. Device
damage may result. If a shocking lead impedance is less than 20 Ω,
reposition the shocking electrodes to allow a greater distance between the
shocking electrodes.
•
ATR entry count. Exercise care when programming the Entry Count to
low values in conjunction with a short ATR Duration. This combination
allows mode switching with very few fast atrial beats. For example, if the
Entry Count was programmed to 2 and the ATR Duration to 0, ATR mode
switching could occur on 2 fast atrial intervals. In these instances, a short
series of premature atrial events could cause the device to mode switch.
•
ATR exit count. Exercise care when programming the Exit Count to low
values. For example, if the Exit Count was programmed to 2, a few cycles
of atrial undersensing could cause termination of mode switching.
•
Left ventricular lead configuration. Proper programming of the LV
coronary venous lead configuration is essential for proper LV lead
function. Program the lead configuration in accordance with the number of
electrodes on the LV lead; otherwise, erratic LV sensing, loss of LV pacing,
or ineffective LV pacing might occur.
- DRAFT -
1-12
INFORMATION FOR USE
PRECAUTIONS
•
Left Ventricular Protection Period (LVPP). Use of a long LVPP reduces
the maximum LV pacing rate and may inhibit CRT at higher pacing rates.
•
Shunting energy. Do not allow any object that is electrically conductive
to come into contact with the lead or device during induction because it
may shunt energy, resulting in less energy getting to the patient, and may
damage the implanted system.
•
Expected benefits. Determine whether the expected device benefits
outweigh the possibility of early device replacement for patients whose
tachyarrhythmias require frequent shocks.
•
Device communication. Use only the designated PRM and software
application to communicate with this pulse generator.
Environmental and Medical Therapy Hazards
•
•
Avoid electromagnetic interference (EMI). Advise patients to avoid
sources of EMI because EMI may cause the pulse generator to deliver
inappropriate therapy or inhibit appropriate therapy. Examples of EMI
sources are:
•
Electrical power sources, arc welding equipment, and robotic jacks
•
Electrical smelting furnaces
•
Large RF transmitters such as radar
•
Radio transmitters, including those used to control toys
•
Electronic surveillance (antitheft) devices
•
An alternator on a car that is running
Elevated Pressures. Elevated pressures due to hyperbaric chamber
exposure of SCUBA diving may damage the pulse generator. The pulse
generator has been tested to function normally at 1.5 Atmospheres
Absolute (ATA) pressure or 15 ft (4.6 m) depth in sea water. For specific
guidelines prior to hyperbaric chamber exposure, or if the patient is
planning scuba diving activity, contact Technical Services at the number
shown on the back cover of this manual.
- DRAFT -
INFORMATION FOR USE
PRECAUTIONS
1-13
Hospital and Medical Environments
•
Mechanical ventilators. During mechanical ventilation, respiration rate
trending may be misleading; therefore, the Respiratory Sensor should be
programmed to Off.
•
Internal defibrillation. Do not use internal defibrillation paddles or
catheters unless the pulse generator is disconnected from the leads
because the leads may shunt energy. This could result in injury to the
patient and damage to the implanted system.
•
External defibrillation. Use of external defibrillation can damage the
pulse generator.
•
Transcutaneous electrical nerve stimulation (TENS). TENS may
interfere with pulse generator function. If necessary, the following measures
may reduce interference:
1. Place the TENS electrodes as close to each other as possible and as
far from the pulse generator and lead system as possible.
2. Monitor cardiac activity during TENS use.
For additional information, contact Technical Services at the number shown
on the back cover of this manual.
- DRAFT -
1-14
INFORMATION FOR USE
PRECAUTIONS
•
Electrocautery. The use of electrocautery could induce ventricular
arrhythmias and/or fibrillation, cause asynchronous or inhibited pulse
generator operation, or cause the pulse generator to deliver an
inappropriate shock. If electrocautery cannot be avoided, observe the
following precautions to minimize complications:
•
Select Electrocautery Protection Mode. Avoid direct contact with the
pulse generator or leads.
•
Monitor the patient and have temporary pacing equipment, external
defibrillation equipment, and knowledgeable medical personnel
available.
•
Position the ground plate so that the current pathway does not pass
through or near the pulse generator system.
•
Use short, intermittent, and irregular bursts at the lowest feasible
energy levels.
•
Use a bipolar electrocautery system where possible.
Remember to reactivate the Tachy Mode after turning off the electrocautery
equipment.
•
Ionizing radiation therapy. Ionizing radiation therapy may adversely affect
device operation. During ionizing radiation therapy (e.g., radioactive cobalt,
linear accelerators, and betatrons), the pulse generator must be shielded
with a radiation-resistive material, regardless of the distance of the device
to the radiation beam. Do not project the radiation port directly at the
device. After waiting a minimum of one hour following radiation treatment
(to allow for a device memory check to occur), always evaluate device
operation, including interrogation and sensing and pacing threshold testing.
At the completion of the entire course of treatments, perform device
interrogation and follow-up, including sensing and pacing threshold testing
and capacitor re-formation.
- DRAFT -
INFORMATION FOR USE
PRECAUTIONS
•
1-15
Lithotripsy. Lithotripsy may permanently damage the pulse generator if
the device is at the focal point of the lithotripsy beam. If lithotripsy must be
used, avoid focusing near the pulse generator site.
The lithotriptor is designed to trigger off the R-wave on the ECG, resulting
in shock waves being delivered during the VRP.
•
•
If the patient does not require pacing, program the pulse generator
Brady Mode to Off.
•
If the patient requires pacing, program the pulse generator to the VVI
mode because atrial pacing pulses can trigger the lithotriptor.
Ultrasound energy. Therapeutic ultrasound (e.g., lithotripsy) energy may
damage the pulse generator. If therapeutic ultrasound energy must be
used, avoid focusing near the pulse generator site. Diagnostic ultrasound
(e.g., echocardiography) is not known to be harmful to the pulse generator.
- DRAFT -
1-16
INFORMATION FOR USE
PRECAUTIONS
•
Radio frequency ablation. Exercise caution when performing radio
frequency ablation procedures in device patients. If the pulse generator
Tachy Mode is programmed to Monitor + Therapy during the procedure,
the device may inappropriately declare a tachycardia episode and deliver
therapy. Pacing therapy may also be inhibited unless the device is
programmed to Electrocautery mode. RF ablation may cause changes in
pacing thresholds; evaluate the patient’s thresholds appropriately.
Minimize risks by following these steps:
•
Program the Tachy Mode(s) to Electrocautery Protection to avoid
inadvertent tachycardia detection (sensing) or therapy.
•
Monitor the patient and have external defibrillation equipment and
knowledgeable medical personnel available.
•
Avoid direct contact between the ablation catheter and the implanted
lead and pulse generator.
•
Keep the current path (electrode tip to ground) as far away from the
pulse generator and leads as possible.
•
Consider the use of external pacing support for pacemaker-dependent
patients (i.e., using internal or external pacing methods).
•
Monitor pre- and post-measurements for sensing and pacing thresholds
and impedances to determine the integrity of the lead-patient function.
Remember to reactivate the pulse generator after turning off the radio
frequency ablation equipment.
•
Electrical interference. Electrical interference or “noise” from devices
such as electrocautery and monitoring equipment may interfere with
establishing or maintaining telemetry for interrogating or programming the
device. In the presence of such interference, move the programmer away
from electrical devices, and ensure that the wand cord and cables are not
crossing one another. If telemetry is cancelled as a result of interference,
the device should be re-interrogated prior to evaluating information from
pulse generator memory.
- DRAFT -
INFORMATION FOR USE
PRECAUTIONS
•
1-17
Radio frequency (RF) interference. RF signals from devices that
operate at frequencies near that of the pulse generator may interrupt ZIP
telemetry while interrogating or programming the pulse generator. This
RF interference can be reduced by increasing the distance between the
interfering device and the PRM and pulse generator. Examples of devices
that may cause interference include:
•
Cordless phone handsets or base stations
•
Certain patient monitoring systems
•
Remote control toys
Home and Occupational Environments
•
Home appliances. Home appliances that are in good working order and
properly grounded do not usually produce enough EMI to interfere with
pulse generator operation. There have been reports of pulse generator
disturbances caused by electric hand tools or electric razors used directly
over the pulse generator implant site.
•
Magnetic fields. Advise patients that extended exposure to strong (greater
than 10 gauss or 1 mTesla) magnetic fields may trigger the magnet feature.
Examples of magnetic sources include:
•
•
Industrial transformers and motors
•
MRI devices
•
Large stereo speakers
•
Telephone receivers if held within 1.27 cm (0.5 inches) of the pulse
generator
•
Magnetic wands such as those used for airport security and in the
Bingo game
Electronic Article Surveillance (EAS). Advise patients to avoid lingering
near antitheft devices such as those found in the entrances and exits of
department stores and public libraries. Patients should walk through them
at a normal pace because such devices may cause inappropriate pulse
generator operation.
- DRAFT -
1-18
INFORMATION FOR USE
PRECAUTIONS
•
Cellular phones. Advise patients to hold cellular phones to the ear
opposite the side of the implanted device. Patients should not carry a
cellular phone that is turned on in a breast pocket or on a belt within 15 cm
(6 inches) of the implanted device since some cellular phones may cause
the pulse generator to deliver inappropriate therapy or inhibit appropriate
therapy.
Follow-up Testing
•
Conversion testing. Successful VF or VT conversion during
arrhythmia conversion testing is no assurance that conversion will occur
post-operatively. Be aware that changes in the patient’s condition, drug
regimen, and other factors may change the DFT, which may result in
nonconversion of the arrhythmia post-operatively.
•
Pacing threshold testing. If the patient’s condition or drug regimen
has changed or device parameters have been reprogrammed, consider
performing a pacing threshold test to confirm adequate margins for pace
capture.
Explant and Disposal
•
Incineration. Be sure that the pulse generator is removed before
cremation. Cremation and incineration temperatures might cause the pulse
generator to explode.
•
Device handling. Before explanting, cleaning, or shipping the device,
complete the following actions to prevent unwanted shocks, overwriting of
important therapy history data, and audible tones:
•
•
Program the pulse generator Tachy and Brady Modes to Off.
•
Program the Magnet Response feature to Off.
•
Program the Beep When Explant is Indicated feature to Off.
Explanted devices. Return all explanted pulse generators and leads to
Boston Scientific. Examination of explanted pulse generators can provide
information for continued improvement in device reliability and will permit
calculation of any warranty replacement credit due.
Do not implant an explanted pulse generator in another patient as sterility,
functionality, and reliability cannot be ensured.
- DRAFT -
INFORMATION FOR USE
POTENTIAL ADVERSE EVENTS
1-19
POTENTIAL ADVERSE EVENTS
Based on the literature and on pulse generator implant experience, the
following alphabetical list includes the possible adverse events associated with
implantation of a pulse generator system:
•
Air embolism
•
Allergic reaction
•
Bleeding
•
Cardiac tamponade
•
Chronic nerve damage
•
Component failure
•
Conductor coil fracture
•
Death
•
Electrolyte imbalance/dehydration
•
Elevated thresholds
•
Erosion
•
Excessive fibrotic tissue growth
•
Extracardiac stimulation (muscle/nerve stimulation)
•
Failure to convert an induced arrhythmia
•
Foreign body rejection phenomena
•
Formation of hematomas or seromas
•
Inability to defibrillate or pace
•
Inappropriate therapy (e.g., shocks where applicable, ATP, pacing)
•
Incisional pain
•
Incomplete lead connection with pulse generator
•
Infection
•
Insulating myocardium during defibrillation with internal or external paddles
•
Lead dislodgment
•
Lead fracture
•
Lead insulation breakage or abrasion
•
Lead tip deformation and/or breakage
•
Myocardial infarction (MI)
•
Myocardial necrosis
- DRAFT -
1-20
INFORMATION FOR USE
POTENTIAL ADVERSE EVENTS
•
Myocardial trauma (e.g., cardiac perforation, irritability, injury)
•
Myopotential sensing
•
Oversensing/undersensing
•
Pacemaker-mediated tachycardia (PMT)
•
Pericardial rub, effusion
•
Pneumothorax
•
Pulse generator migration
•
Shunting current during defibrillation with internal or external paddles
•
Tachyarrhythmias, which include acceleration of arrhythmias and early,
recurrent atrial fibrillation
•
Thrombosis/thromboemboli
•
Valve damage
•
Venous occlusion
•
Venous trauma (e.g., perforation, dissection, erosion)
•
Worsening heart failure
Patients may develop psychological intolerance to a pulse generator system
and may experience the following:
•
Dependency
•
Depression
•
Fear of premature battery depletion
•
Fear of shocking while conscious
•
Fear that shocking capability may be lost
•
Imagined shocking
In addition to the implantation of a pulse generator system, potential adverse
events associated with the implantation of a coronary venous lead system
include:
•
Allergic reaction to contrast media (i.e., renal failure, etc.)
•
Breakage/failure of implant instruments
•
Prolonged exposure to fluoroscopic radiation
- DRAFT -
INFORMATION FOR USE
MECHANICAL SPECIFICATIONS
1-21
MECHANICAL SPECIFICATIONS
Device mechanical specifications for specific models are listed in the table
below.
Table 1-1.
Mechanical Specifications
Model
Dimensions
W x H x D
Volume
(cm3)
Mass (g)
Connector Type
(RV : LV)
Case
Electrode
Surface Area
(mm²)
N118
6.17 x 7.95 x 0.99
32.5
72.0
IS-1/DF-1 : LV-1
6670
N119
6.17 x 7.95 x 0.99
32.5
72.0
IS-1/DF-1 : IS-1
6670
Models include ZIP telemetry with a nominal RF frequency of 916.5 MHz.
Material specifications are shown below:
•
Case: hermetically sealed titanium
•
Header: implantation-grade polymer
•
Power Supply: lithium-manganese dioxide cell; Boston Scientific; 401988
LEAD CONNECTIONS
Lead connections are illustrated below.
CAUTION: Do not use any lead with this device without first verifying
connector compatibility. Using incompatible leads can damage the connector
and/or result in potential adverse consequences, such as undersensing of
cardiac activity or failure to deliver necessary therapy.
DF-1
IS-1
BI
–
RV
RA
IS-1
BI
–
RV
LV
Figure 1-1.
IS-1
BI
LV-1
LV
DF-1
DF-1
UNI/BI
DF-1
RA
IS-1
BI
IS-1
UNI/BI
Lead connections
- DRAFT -
1-22
INFORMATION FOR USE
ITEMS INCLUDED IN PACKAGE
NOTE: The pulse generator case is used as a defibrillating electrode unless
the pulse generator has been programmed to the Distal Coil to Proximal Coil
(or “Cold Can”) Shock Vector.
ITEMS INCLUDED IN PACKAGE
The following items are included with the pulse generator:
•
One torque wrench
•
Product literature
•
One patient data disk
NOTE: Accessories (e.g., wrenches) are intended for one-time use only.
They should not be resterilized or reused.
SYMBOLS ON PACKAGING
The following symbols may be used on pulse generator packaging and labeling
(Table 1-2 on page 1-22):
Table 1-2.
Symbol
Symbols on packaging
Description
Reference number
Package contents
Pulse generator
Torque wrench
Disk for data storage
Literature enclosed
Serial number
Use by
- DRAFT -
INFORMATION FOR USE
CHARACTERISTICS AS SHIPPED
Table 1-2.
1-23
Symbols on packaging (continued)
Description
Symbol
Lot number
Date of manufacture
Non-ionizing electromagnetic radiation
Sterilized using ethylene oxide
Do not reuse
Dangerous voltage
Consult instructions for use
Temperature limitation
Wand placement indicator for interrogation
Opening instruction
CHARACTERISTICS AS SHIPPED
Refer to the table for pulse generator settings at shipment (Table 1-3 on page
1-23).
Table 1-3.
Characteristics as shipped
Parameter
Setting
Tachy Mode
Storage
Tachy Therapy available
ATP, Shock
Pacing Mode
Storage
- DRAFT -
1-24
INFORMATION FOR USE
X-RAY IDENTIFIER
Table 1-3.
Characteristics as shipped (continued)
Parameter
Setting
Pacing Therapy available
DDDR
Sensor
Accelerometer (MV for respiratory rate trend)
Pace/Sense Configuration
RA: BI/BI
Pace/Sense Configuration
RV: BI/BI
Pace/Sense Configuration
LV: Off
The pulse generator is shipped in a power-saving Storage mode to extend its
shelf life. In Storage mode, all features are inactive except:
•
Telemetry support, which allows interrogation and programming
•
Real-time clock
•
Commanded capacitor re-formation
•
STAT SHOCK and STAT PACE commands
The device leaves Storage mode when one of the following actions occurs;
however, programming other parameters will not affect the Storage mode:
•
STAT SHOCK or STAT PACE is commanded
•
Tachy Mode is programmed to:
–
Off
–
Monitor Only
–
Monitor + Therapy
Once you have programmed the pulse generator out of Storage mode, the
device cannot be reprogrammed to that mode.
X-RAY IDENTIFIER
The pulse generator has an identifier that is visible on x-ray film or under
fluoroscopy. This identifier provides noninvasive confirmation of the
manufacturer and consists of the following:
•
The letters, BOS, to identify Boston Scientific as the manufacturer
•
The number, 112, to identify the Model 2868 PRM software application
needed to communicate with the pulse generator
- DRAFT -
INFORMATION FOR USE
FEDERAL COMMUNICATIONS COMMISSION (FCC)
1-25
The x-ray identifier is embedded in the header of the device at the approximate
location (Figure 1-2 on page 1-25).
Header
X-Ray Identifier
Pulse Generator Case
Figure 1-2.
X-ray identifier
For information on identifying the device via the PRM, refer to the PRM
operator’s manual.
The pulse generator model number is stored in device memory and is shown
on the PRM summary screen once the pulse generator is interrogated.
FEDERAL COMMUNICATIONS COMMISSION (FCC)
This device complies with Title 47, Part 15 of the FCC rules. Operation is
subject to the following two conditions:
•
This device may not cause harmful interference, and
•
This device must accept any interference received, including interference
that may cause undesired operation.
CAUTION: Changes or modifications not expressly approved by Boston
Scientific could void the user’s authority to operate the equipment.
INDUSTRY CANADA (IC)
This device complies with Radio Standards Specification RSS-210. Operation
is subject to the following two conditions:
•
This device may not cause harmful interference, and
•
This device must accept any interference received, including interference
that may cause undesired operation.
CAUTION: Changes or modifications not expressly approved by Boston
Scientific could void the user’s authority to operate the equipment.
- DRAFT -
1-26
INFORMATION FOR USE
PULSE GENERATOR LONGEVITY
PULSE GENERATOR LONGEVITY
Based on simulated studies, it is anticipated that these pulse generators have
average longevity to explant as shown below.
The longevity expectations, which account for the energy used during
manufacture and storage, apply at the conditions shown in the table along
with the following:
•
Assumes 70 ppmLRL; DDDR mode; 100% biventricular pacing; 15%
atrium pacing and 0.4 ms pacing pulse width (RA, RV, LV); RA impedance
500 Ω.
•
Projected longevity is calculated at 6 to 14 maximum energy charging cycles
per year (depending on battery status) with automatic capacitor/battery
management and maximum energy charges, and 3-channel EGM Onset
set to On.
Table 1-4.
Pulse generator life expectancy estimation (implant to explant) for HE models
HE Modelsa
Pacing Amplitude
Longevity (years) at 500 Ω and 700 Ω
Pacing Impedance (RV and LV)
RA/RV
LV
500 Ω
700 Ω
2.5 V
3.0 V
6.6
6.9
2.5 V
3.5 V
6.2
6.6
3.5 V
3.5 V
5.5
5.9
3.5 V
5.0 V
4.6
5.0
a. For RF-enabled models, assumes ZIP wandless telemetry use for 3 hours at implant time and for 20 minutes
during each quarterly follow up.
NOTE: The energy consumption in the longevity table is based upon
theoretical electrical principles and verified via bench testing only.
The pulse generator longevity may increase with a decrease in any of the
following:
•
Pacing rate
•
Pacing pulse amplitude(s)
•
Pacing pulse width(s)
•
Percentage of paced to sensed events
- DRAFT -
INFORMATION FOR USE
WARRANTY INFORMATION
•
1-27
Charging frequency
Longevity is also reduced in the following circumstances:
•
With a decrease in pacing impedance
•
When Patient Triggered Monitor is programmed to On
•
For models with ZIP wandless telemetry, one hour of additional telemetry
reduces longevity by approximately 4 days.
Device longevity may also be affected by:
•
Tolerances of electronic components
•
Variations in programmed parameters
•
Variations in usage as a result of patient condition
An additional maximum-energy shock reduces longevity by approximately
19 days.
Refer to the PRM Summary screen for an estimate of pulse generator longevity
specific to the implanted device.
WARRANTY INFORMATION
A limited warranty certificate for the pulse generator is packaged with the
device. For additional copies, please contact Boston Scientific at the address
and phone number shown on the back cover of this manual.
PRODUCT RELIABILITY
It is Boston Scientific’s intent to provide implantable devices of high quality and
reliability. However, these devices may exhibit malfunctions that may result in
lost or compromised ability to deliver therapy. These malfunctions may include
the following:
•
Premature battery depletion
•
Sensing or pacing issues
•
Inability to shock
•
Error codes
•
Loss of telemetry
- DRAFT -
1-28
INFORMATION FOR USE
PATIENT COUNSELING INFORMATION
Refer to Boston Scientific’s CRM Product Performance Report on
www.bostonscientific.com for more information about device performance,
including the types and rates of malfunctions that these devices have
experienced historically. While historical data may not be predictive of future
device performance, such data can provide important context for understanding
the overall reliability of these types of products.
Sometimes device malfunctions result in the issuance of safety advisories.
Boston Scientific determines the need to issue safety advisories based on
the estimated malfunction rate and the clinical implication of the malfunction.
When Boston Scientific communicates safety advisory information, the
decision whether to replace a device should take into account the risks of the
malfunction, the risks of the replacement procedure, and the performance to
date of the replacement device.
PATIENT COUNSELING INFORMATION
The following topics should be discussed with the patient prior to discharge.
•
•
The patient should:
–
Contact their physician immediately if they hear tones coming from
their pulse generator
–
Contact their physician to have their pulse generator system evaluated
if they receive external defibrillation
–
Understand the signs and symptoms of infection
–
Understand the symptoms that should be reported (e.g., sustained
high-rate pacing requiring reprogramming)
–
Seek medical guidance before entering protected environments such
as areas protected by a warning notice that prevents entry by patients
who have a pulse generator
–
Understand and avoid potential sources of EMI and magnetic fields
in home, work, and medical environments (See Warnings and
Precautions for more detailed information about specific sources)
Persons administering CPR may experience the presence of voltage
(tingling) on the patient’s body surface when the pulse generator delivers
a shock.
- DRAFT -
INFORMATION FOR USE
PATIENT COUNSELING INFORMATION
•
1-29
It is Boston Scientific’s intent to provide implantable devices of high quality
and reliability. However, these devices may exhibit malfunctions that may
result in lost or compromised ability to deliver therapy. When Boston
Scientific communicates safety advisory information, the decision whether
to replace a device should take into account the risks of the malfunction,
the risks of the replacement procedure, and the performance to date of the
replacement device.
Patient Handbook
The Patient Handbook is provided for each device.
It is recommended that you discuss the information in the Patient Handbook
with concerned individuals both before and after implantation so they are fully
familiar with pulse generator operation.
For additional copies, contact your sales representative, or contact Boston
Scientific at the phone number shown on the back cover of this manual.
- DRAFT -
1-30
INFORMATION FOR USE
PATIENT COUNSELING INFORMATION
- DRAFT -
2-1
USING THE PROGRAMMER/RECORDER/MONITOR
CHAPTER 2
This chapter contains the following topics:
•
"ZOOM LATITUDE Programming System" on page 2-2
•
"Indications-Based Programming (IBP)" on page 2-2
•
"Manual Programming" on page 2-5
•
"Software Terminology and Navigation" on page 2-5
•
"Data Management" on page 2-11
•
"Communicating with the Pulse Generator" on page 2-12
•
"DIVERT THERAPY" on page 2-16
•
"STAT SHOCK" on page 2-17
•
"STAT PACE" on page 2-18
•
"Safety Mode" on page 2-18
- DRAFT -
2-2
USING THE PROGRAMMER/RECORDER/MONITOR
ZOOM LATITUDE PROGRAMMING SYSTEM
ZOOM LATITUDE PROGRAMMING SYSTEM
The ZOOM LATITUDE Programming System is the external portion of the
pulse generator system and includes:
•
Model 3120 Programmer/Recorder/Monitor (PRM)
•
Model 2868 ZOOMVIEW Software Application
•
Model 6577 Accessory Telemetry Wand
The ZOOMVIEW software provides advanced device programming and patient
monitoring technology. It was designed with the intent to:
•
Enhance device programming capability
•
Improve patient and device monitoring performance
•
Simplify and expedite programming and monitoring tasks
You can use the PRM system to do the following:
•
Interrogate the pulse generator
•
Program the pulse generator to provide a variety of therapy options
•
Access the pulse generator’s diagnostic features
•
Perform noninvasive diagnostic testing
•
Access therapy history data
You can program the pulse generator using two methods: automatically using
IBP or manually.
INDICATIONS-BASED PROGRAMMING (IBP)
IBP is a tool that provides specific programming recommendations based on
the patient’s clinical needs and primary indications.
IBP is a clinical approach to programming that was developed based on
physician consultation and case studies. The intent of IBP is to enhance patient
outcomes and save time by providing base programming recommendations
that you can customize as needed. IBP systematically presents the specific
features intended for use with the clinical conditions you identify in the IBP user
interface, and allows you to take maximum advantage of the pulse generator’s
capabilities.
IBP can be accessed from the Settings tab on the main application screen
(Figure 2-1 on page 2-3).
- DRAFT -
USING THE PROGRAMMER/RECORDER/MONITOR
INDICATIONS-BASED PROGRAMMING (IBP)
Figure 2-1.
2-3
Indications-based Programming screen
Indications are clustered in general categories as illustrated above. The intent
for each category of indications is described below:
•
•
Sinus Node:
–
If Normal is selected, the intent is to promote intrinsic atrial events
and provide CRT pacing.
–
If Chronotropically Incompetent is selected, the intent is to provide
rate-adaptive CRT pacing.
–
If Sick Sinus Syndrome is selected, the intent is to provide atrial pacing
support and CRT pacing.
AV Node:
–
The intent is for nominal Paced AV delay and Sensed AV delay
settings. The SmartDelay optimization feature may be used to adjust
the AV delay.
NOTE: The selected settings for AF and Sinus Node may affect the suggested
value for the setting of AV Node.
•
Atrial Arrhythmias
–
If Paroxysmal/Persistent is selected, the intent is to avoid tracking
atrial arrhythmias by using dual-chamber pacing mode with ATR Mode
Switch.
–
If Permanent/Chronic AF is selected, the intent is to provide rate
adaptive RV pacing.
- DRAFT -
2-4
USING THE PROGRAMMER/RECORDER/MONITOR
INDICATIONS-BASED PROGRAMMING (IBP)
•
Ventricular Arrhythmias
–
–
–
When History of VF/SCD or Prophylaxis for VT/VF is selected, a
2-zone configuration with the following rate thresholds and therapies
is provided:
–
180 bpm for the VF zone with QUICK CONVERT ATP and
Maximum Energy Shocks enabled
–
160 bpm for the VT zone with therapy disabled (Monitor Only)
When History of VT/VF is selected, a 2-zone configuration with the
following rate thresholds and therapies is provided:
–
200 bpm for the VF zone with QUICK CONVERT ATP and
Maximum Energy Shocks enabled
–
160 bpm for the VT zone with ATP and Maximum Energy Shocks
enabled
–
Onset/Stability enabled
When VF Only is selected, the intent is for a single VF zone of 220 bpm
is provided with only Maximum Energy Shocks enabled.
When you have chosen the patient indications, select the View Recommended
Settings button to view a summary of the programming recommendations
(Figure 2-2 on page 2-5).
NOTE: You must view the settings before you can program them. Selecting
the View Recommended Settings button allows you to view the settings
that are recommended based on the indications that you selected. Viewing
the recommended settings does not overwrite any pending (i.e., not yet
programmed) parameter changes. You must choose to program or reject
the recommended settings after viewing them. If you choose to reject the
recommended settings, all of your pending settings will be restored. If you
choose to program the recommended settings, any pending parameter
changes will be overwritten, with the exception of sensitivity, shock outputs,
and pacing outputs, which are independent of IBP.
- DRAFT -
USING THE PROGRAMMER/RECORDER/MONITOR
MANUAL PROGRAMMING
2-5
Primary
Settings
Summary
Figure 2-2.
Proposed Settings Summary screen
The Proposed Settings Summary screen displays the primary programming
recommendations. Additional details about all changed parameters are
available by selecting the View Changes button from the toolbar. You have the
option to program the proposed settings or reject them, as long as telemetry
is still engaged:
•
Program—select the Program this Profile button to accept the proposed
settings.
•
Reject—select the Reject this Profile button to reject the proposed settings;
this action will return you to the main IBP screen with no changes made.
MANUAL PROGRAMMING
Manual programming controls such as sliders and menus are available to allow
you to individually adjust pulse generator program settings.
Manual programming controls are located on the Settings Summary tab, which
can be accessed from the Settings tab or by selecting the Settings Summary
button on the Summary tab. Refer to other feature descriptions in this manual
for specific manual programming information and instructions. Refer to
"Programmable Options" on page A-1 for detailed listings of available settings.
SOFTWARE TERMINOLOGY AND NAVIGATION
This section provides an overview of the PRM system.
Main Screen
The main PRM screen is shown below, followed by a description of the
components (Figure 2-3 on page 2-6).
- DRAFT -
2-6
USING THE PROGRAMMER/RECORDER/MONITOR
SOFTWARE TERMINOLOGY AND NAVIGATION
Patient name
PRM Mode Indicator
Device name
Details
button
ECG/EGM
display
Tabs
Device
model
Toolbar
Figure 2-3.
Main Screen
PRM Mode Indicator
The PRM Mode Indicator displays at the top of the screen to identify the current
PRM operational mode.
Patient—indicates that the PRM is displaying data obtained by
communicating with a device.
Patient Data Disk—indicates that the PRM is displaying stored
data from a patient data disk.
Demo Mode—indicates that the PRM is displaying sample data
and operating in demonstration mode.
- DRAFT -
USING THE PROGRAMMER/RECORDER/MONITOR
SOFTWARE TERMINOLOGY AND NAVIGATION
2-7
ECG/EGM Display
The ECG area of the screen shows real-time status information about the
patient and the pulse generator. This information is useful in evaluating system
performance:
•
Real-time EGMs can be transmitted from the pace/sense or shocking
electrodes to evaluate lead system integrity and help identify faults such as
lead fractures, insulation breaks, or dislodgments.
•
Annotated event markers identify certain intrinsic cardiac and device-related
events, and provide information such as sensed/paced events, decision of
detection criteria, and therapy delivery.
You can select the Details button to enlarge the ECG/EGM screen. The
following options are available:
•
Show Device Markers—displays annotated markers
•
Enable Surface Filter—minimizes noise on the surface ECG
•
Display Pacing Spikes—shows detected pacing spikes, annotated by a
marker on the surface ECG waveform
You can print a real-time EGM report, which includes annotated event markers,
by performing the following steps:
1. Press one of the print speed keys on the PRM (for example, speed key 25).
2. Press the Paper Form Feed key. The EGM report will begin printing.
You can print a report containing the definitions of all of the annotated markers
by performing the following steps:
1. From the toolbar, click the Reports button. The Reports window displays.
2. Select the Marker Legend checkbox.
3. Click the Print button. The Marker Legend Report is sent to the printer.
Toolbar
The toolbar allows you to perform the following tasks:
•
Select system utilities
•
Generate reports
•
Interrogate and program the pulse generator
•
View pending or programmed changes
- DRAFT -
2-8
USING THE PROGRAMMER/RECORDER/MONITOR
SOFTWARE TERMINOLOGY AND NAVIGATION
•
View attentions and warnings
•
End your PRM session
Tabs
Tabs allow you to select PRM tasks, such as viewing summary data or
programming device settings. Selecting a tab displays the associated screen.
Many screens contain additional tabs, which allow you to access more detailed
settings and information.
Buttons
Buttons are located on screens and dialogs throughout the application. Buttons
allow you to perform various tasks, including:
•
Obtain detailed information
•
View setting details
•
Set programmable values
When a button selection opens a window in front of the Main Screen, a Close
button displays in the upper-right corner of the window to allow you to close the
window and return to the Main Screen.
Icons
Icons are graphic elements that, when selected, may initiate an activity, display
lists or options, or change the information displayed.
Details—opens a window containing detailed information.
Patient—opens a window with patient information details.
Leads—opens a window with details on leads.
- DRAFT -
USING THE PROGRAMMER/RECORDER/MONITOR
SOFTWARE TERMINOLOGY AND NAVIGATION
2-9
Battery—opens a window with details on the pulse generator
battery.
Run—causes the programmer to perform an action.
Check—indicates that an option is selected.
Event—indicates that an event has occurred. When you view the
Trends timeline on the Events tab, event icons display wherever
events have occurred. Selecting an events icon displays details
about the event.
Slider Icons
Horizontal Slider—indicates that a slider object can be clicked
and dragged left or right.
Vertical Slider—indicates that a slider object can be clicked and
dragged up or down.
Sort Icons
Sort Ascending—indicates that Ascending sort is currently
selected on a table column sort button. (e.g., 1, 2, 3, 4, 5)
Sort Descending—indicates that Descending sort is currently
selected on a table column sort button. (e.g., 5, 4, 3, 2, 1)
Increment and Decrement Icons
Increment—indicates that an associated value can be
incremented.
Decrement—indicates that an associated value can be
decremented.
- DRAFT -
2-10
USING THE PROGRAMMER/RECORDER/MONITOR
SOFTWARE TERMINOLOGY AND NAVIGATION
Scroll Icons
Scroll Left—indicates that an associated item can be scrolled left.
Scroll Right—indicates that an associated item can be scrolled
right.
Scroll Up—indicates that an associated item can be scrolled up.
Scroll Down—indicates that an associated item can be scrolled
down.
Common Objects
Common objects such as status bars, scroll bars, menus, and dialogs are used
throughout the application. These operate similarly to the objects found in web
browsers and other computer applications.
Use of Color
Colors are used to highlight buttons, icons, and other objects. The use of
specific color conventions is intended to provide a more consistent user
experience and simplify programming.
Red and yellow are used to provide Warning and Attention indications, as
described below:
•
•
Red—indicates Warning conditions such as the following:
–
The selected device parameter value is not allowed
–
Device and patient diagnostic information that requires serious
consideration
Yellow—indicates Attention conditions such as the following:
–
The selected device parameter value is not recommended, but is
allowed
–
Device and patient diagnostic information that should be addressed
- DRAFT -
USING THE PROGRAMMER/RECORDER/MONITOR
DATA MANAGEMENT
2-11
When a red Warning or yellow Attention button displays in the toolbar, click the
button. The Parameter Interactions screen will display, with information about
corrective action.
DATA MANAGEMENT
The PRM system allows you to manage patient and pulse generator data by
viewing, printing, storing, or retrieving it. This section describes the PRM data
management capabilities.
Patient Information
Information about the patient can be stored in pulse generator memory. The
information is accessible from the Summary screen by selecting the Patient
icon. This information includes, but is not limited to, the following:
•
Patient and physician names
•
Pulse generator serial number
•
Implant date
•
Lead configurations
•
Implant test measurements
The information can be retrieved at any time by interrogating the pulse
generator and viewing it on the PRM screen or printing it as a report.
Disk Operations
The PRM system allows you to store patient data on a removable data disk.
To access the data stored on a disk, first select the Utilities button, then select
the Disk tab. Insert a disk into the disk drive, and select one of the following
disk options:
•
Read Disk—allows you to retrieve saved data from a patient data disk
•
Save All to Disk—allows you to save information to a disk, including the
following:
–
Therapy history
–
Programmed parameter values
–
Trending values
–
HRV
–
Histogram paced/sensed counters
- DRAFT -
2-12
USING THE PROGRAMMER/RECORDER/MONITOR
COMMUNICATING WITH THE PULSE GENERATOR
Print
You can print PRM reports by using the internal printer, or by connecting to an
external printer. To print a report, select the Reports button. Then select the
report you wish to print from the following categories:
•
Follow-up reports
•
Episode reports
•
Other reports (includes device settings, patient data, and other information)
COMMUNICATING WITH THE PULSE GENERATOR
The PRM communicates with the pulse generator using a telemetry wand.
After initiating communication with the wand, some pulse generator models can
use wandless ZIP telemetry (two-way RF communication) to interface with the
PRM. Wanded or ZIP telemetry is required to:
•
Direct commands from the PRM system, such as:
–
INTERROGATE
–
PROGRAM
–
STAT SHOCK
–
STAT PACE
–
DIVERT THERAPY
•
Modify device parameter settings
•
Conduct EP testing
•
Conduct diagnostic tests including the following:
•
–
Pacing impedance tests
–
Pacing threshold tests
–
Intrinsic amplitude tests
Perform manual capacitor re-form
ZIP Telemetry
ZIP telemetry is a wandless, two-way RF communication option that allows the
PRM system to communicate with some pulse generator models. When a
- DRAFT -
USING THE PROGRAMMER/RECORDER/MONITOR
COMMUNICATING WITH THE PULSE GENERATOR
2-13
wanded telemetry session is initiated, the PRM checks the pulse generator’s
telemetry capability. If the PRM detects a pulse generator with ZIP telemetry
capability, a message will display indicating that ZIP telemetry is available and
the wand can be removed. Otherwise, the session will continue with wanded
telemetry.
ZIP telemetry offers the following advantages over traditional wanded telemetry:
•
The faster data transmission speed means less time is required for device
interrogation
•
Data transmission over a longer distance (within 3 meters) minimizes the
need to keep the wand in the sterile field during implant, which may reduce
the risk of infection
•
Continuous telemetry is possible during the entire implant procedure,
allowing monitoring of pulse generator performance and lead integrity
during implant
Regardless of whether ZIP telemetry is being used, wanded communication
is still available.
Starting a Wanded Telemetry Session
Follow this procedure to begin a wanded telemetry communication session:
1. Make sure the telemetry wand is connected to the PRM system and is
available throughout the session.
2. Position the wand over the pulse generator at a distance not greater than 6
cm (2.4 inches).
3. Use the PRM to Interrogate the pulse generator.
4. Retain the wand position whenever communication is required.
Starting a ZIP Telemetry Session
Follow this procedure to begin a ZIP telemetry communication session:
1. Start a wanded telemetry session. Verify that the wand cord is within reach
of the pulse generator to enable the use of wanded telemetry should it
become necessary.
- DRAFT -
2-14
USING THE PROGRAMMER/RECORDER/MONITOR
COMMUNICATING WITH THE PULSE GENERATOR
2. Keep the telemetry wand in position until either a message appears,
indicating that the telemetry wand may be removed from proximity of the
pulse generator, or the ZIP telemetry light illuminates on the PRM system.
Ending a Telemetry Session
Select the End Session button to quit a telemetry session and return to the
startup screen. You can choose to end the session or return to the current
session. Upon ending a session, the PRM system terminates all communication
with the pulse generator.
ZIP Telemetry Security
The pulse generator is a compliant low-power transceiver. The pulse generator
can only be interrogated or programmed by RF signals that employ the
proprietary ZIP telemetry protocol. The pulse generator verifies that it is
communicating with a ZOOMVIEW system before responding to any RF
signals. The pulse generator stores, transmits, and receives individually
identifiable health information in an encrypted format.
ZIP telemetry is possible when all of the following conditions are met:
•
ZIP telemetry setting for the PRM is programmed On
•
The pulse generator has RF communication capabilities
•
The ZIP telemetry channel is available for use
•
The pulse generator is within range of the PRM system
•
The pulse generator has not reached Explant; note that a total of 1.5 hours
of ZIP telemetry will be available after the pulse generator reaches Explant
•
The pulse generator battery capacity is not depleted
In order to meet local communications rules and regulations, ZIP telemetry
should not be used when the pulse generator is outside its normal operating
temperature of 20°C–43°C (68°F–109°F).
The PRM supports communication between two PRMs and two pulse
generators at a time, as two independent sessions. If there are two PRMs
already communicating in the vicinity, a third session will not be allowed to start;
wanded communication will be necessary in this case.
The PRM notifies you if ZIP telemetry is unavailable because of other sessions
already in progress.
- DRAFT -
USING THE PROGRAMMER/RECORDER/MONITOR
COMMUNICATING WITH THE PULSE GENERATOR
2-15
RF signals in the same frequency band used by the system may interfere with
ZIP telemetry communication. These interfering signals include:
•
Signals from other pulse generator/PRM system RF communication
sessions after the maximum number of independent sessions has been
reached. Other nearby pulse generators and PRMs using ZIP telemetry
may prevent ZIP telemetry communication.
•
Interference from other RF sources.
CAUTION: RF signals from devices that operate at frequencies near that
of the pulse generator may interrupt ZIP telemetry while interrogating or
programming the pulse generator. This RF interference can be reduced by
increasing the distance between the interfering device and the PRM and pulse
generator. Examples of devices that may cause interference include:
•
Cordless phone handsets or base stations
•
Certain patient monitoring systems
•
Remote control toys
Radio frequency interference may temporarily disrupt ZIP telemetry
communication. The PRM will normally reestablish ZIP communication when
the RF interference ends or subsides. Because continued RF interference may
prevent ZIP telemetry communication, the system is designed to use wanded
telemetry when ZIP telemetry is not available.
If ZIP telemetry is not available, wanded telemetry communication with the
PRM can be established. The system provides the following feedback to
indicate that ZIP telemetry is not available:
•
The ZIP telemetry indicator light on the PRM turns off
•
If event markers and/or EGMs are activated, transmission of the event
markers and/or EGMs is interrupted
•
If a command or other action has been requested, the PRM displays a
notification indicating the wand should be placed in range of the pulse
generator
ZIP telemetry operates consistently with wanded telemetry—no programming
step can be completed unless the entire programming command has been
received and confirmed by the pulse generator.
- DRAFT -
2-16
USING THE PROGRAMMER/RECORDER/MONITOR
DIVERT THERAPY
The pulse generator cannot be misprogrammed as a result of interrupted ZIP
telemetry. Interruptions of ZIP telemetry may be caused by RF signals that
operate at frequencies near that of the pulse generator and are strong enough
to compete with the ZIP telemetry link between the pulse generator and the
PRM. Significant interference may result in a break or drop-outs of real-time
EGMs. If commands are interrupted, the PRM displays a message to place
the wand on the pulse generator. These situations can be resolved by using
standard wanded telemetry. There will be no interruption of device functionality
or therapy during this period.
NOTE: The PRM operates on a specific frequency range depending on
geography. The PRM determines the ZIP frequency range that the pulse
generator uses based on the specific device model. If the PRM and pulse
generator ZIP frequency ranges do not match, it indicates that the patient has
traveled outside their geography. The PRM will display a message indicating
that ZIP telemetry cannot be used; however, the patient’s pulse generator can
be interrogated by using the wand.
Considerations for Reducing Interference
Increasing the distance from the source of interfering signals may enable the
use of the ZIP telemetry channel.
Repositioning the PRM antenna or repositioning the PRM may improve ZIP
telemetry performance. If ZIP telemetry performance is not satisfactory, the
option of using wanded telemetry is available.
Depending on the environment and PRM orientation relative to the pulse
generator, the system is capable of maintaining ZIP telemetry communication
at distances up to 12 m (40 ft). For optimum ZIP telemetry communication,
position the PRM antenna within 3 m (10 ft) of the pulse generator and remove
any obstruction between the PRM and the pulse generator.
DIVERT THERAPY
When the pulse generator is charging to deliver a shock, the shock delivery
may be diverted from the patient. If diverted, the shock does not count as
one of the total number of shocks that may be delivered during an episode. If
redetection occurs and more shock therapy is required, and if more shocks are
available in the therapy prescription, the pulse generator will charge again to
deliver subsequent shocks.
- DRAFT -
USING THE PROGRAMMER/RECORDER/MONITOR
STAT SHOCK
2-17
Also, the DIVERT THERAPY key can be pressed to divert ATP therapy in
midburst. If redetection occurs, the ATP scheme will not be used again and the
next programmed therapy in the sequence will be initiated.
1. If you are not already in a session, position the telemetry wand within range
of the pulse generator.
2. Press the DIVERT THERAPY key. A message window will appear
indicating that a divert attempt is being made.
3. If using wanded telemetry, maintain the wand position until the message
window disappears, indicating the shock has been diverted. Prematurely
removing the wand (breaking the telemetry link) may allow the pulse
generator to continue charging and to deliver the shock.
NOTE: There is a 500 ms delay between the end of charging and shock
delivery designed to provide a minimum period for the DIVERT THERAPY
command. After this time, pressing DIVERT THERAPY may not divert the
shock.
STAT SHOCK
A nonprogrammable, maximum-output STAT SHOCK can be delivered to the
patient at any time during a communication session. The STAT SHOCK can be
delivered when the pulse generator’s Tachy Mode is programmed to any mode.
This function does not affect the programmed shock sequences (lower-energy
shocks can be delivered following a STAT SHOCK) and does not count as
one of the total number of shocks in a therapy sequence for a given episode.
The output of the STAT SHOCK is at the maximum-output energy and at
the programmed polarity and waveform; STAT SHOCK is always committed
regardless of programmed parameters.
1. If you are not already in a session, position the telemetry wand within range
of the pulse generator.
2. Press the STAT SHOCK key. A message window appears with information
about the shock and instructions to initiate the shock.
3. To initiate the shock, press the STAT SHOCK key again. A different
message window appears indicating that STAT SHOCK is in process.
When the shock has been delivered, the message window disappears.
- DRAFT -
2-18
USING THE PROGRAMMER/RECORDER/MONITOR
STAT PACE
4. Subsequent high-energy STAT SHOCKS may be delivered by repeating
the previous steps.
NOTE:
The STAT SHOCK may be diverted using the DIVERT THERAPY key.
NOTE: Following STAT SHOCK delivery, if the Tachy Mode is programmed
to Monitor Only or Monitor + Therapy, post-shock redetection is initiated
(initial detection criteria and enhancements are not used). If the Tachy Mode
is programmed to Monitor + Therapy and redetection determines that further
therapy is required, the programmed sequence of therapy will be resumed or
initiated, including ATP and/or low-energy shocks.
STAT PACE
Emergency bradycardia pacing using the STAT PACE command sets the
bradycardia operation to parameters intended to ensure capture and keep
the patient stable.
1. If you are not already in a session, position the telemetry wand within range
of the pulse generator.
2. Press the STAT PACE key. A message window displays the STAT PACE
values.
3. Press the STAT PACE key a second time. A message indicates that STAT
PACE is being performed, followed by the STAT PACE values.
4. Select the Close button on the message window.
5. To stop STAT PACE, reprogram the pulse generator.
CAUTION: When a pulse generator is programmed to STAT PACE settings,
it will continue to pace at the high-energy STAT PACE values if it is not
reprogrammed. The use of STAT PACE parameters will decrease device
longevity.
SAFETY MODE
The pulse generator hardware includes a safety core feature. The safety core
is intended to provide life-sustaining therapy in the event of specific failures
within the pulse generator. If the pulse generator is interrogated while safety
core is active, the PRM will indicate that the pulse generator is in Safety Mode.
- DRAFT -
USING THE PROGRAMMER/RECORDER/MONITOR
SAFETY MODE
2-19
Limited inductive telemetry and some device programming are available during
Safety Mode. If you interrogate a pulse generator that is in Safety Mode, the
PRM will display a warning screen directing you to call Technical Services.
Backup Pacemaker
The Safety Mode feature provides a simple VVI pacemaker programmed to
72.5 ppm, with the following BiV output characteristics:
•
0 ms offset
•
5 V amplitude
•
1.0 ms pulse width
Backup Defibrillator
The Safety Mode feature provides a single-zone backup defibrillator, which can
be enabled or disabled using the PRM.
Tachycardia Detection in Safety Mode
In Safety Mode, the tachycardia detection monitors RV senses using a
traditional tachycardia detection window with a rate threshold of 165 bpm.
Tachycardia Therapy in Safety Mode
In Safety Mode, tachycardia therapy consists of maximum energy, committed
shocks, with the following settings:
•
Shock polarity—initial
•
Shock waveform—biphasic
•
Shock vector—V-TRIAD
Within the period of a declared episode, therapy is limited to 5 shocks.
When a magnet is detected, therapy delivery is immediately inhibited although
charging may continue. After the magnet has been applied for 1 second,
the therapy is diverted and detection is inhibited. The magnet must then be
removed for 2 seconds in order to allow detection to continue.
- DRAFT -
2-20
USING THE PROGRAMMER/RECORDER/MONITOR
SAFETY MODE
Programming the Device Safety Tachy Mode
In Safety Mode, you can perform the following steps to program the Safety
Tachy Mode:
1. Select the Tachy Mode button. The Change Device Mode dialog displays.
2. Click to select the required Safety Tachy Mode setting, either Off or Monitor
+ Therapy. Click Apply Changes to apply the new setting, or Cancel
Changes to cancel the new setting.
3. Click Close to dismiss the Change Device Mode dialog.
NOTE: The safety tachy mode automatically will be set to Off if additional
faults are detected while in Safety Mode.
- DRAFT -
3-1
TACHYARRHYTHMIA DETECTION
CHAPTER 3
This chapter contains the following topics:
•
"Device Mode" on page 3-2
•
"Rate Sensing" on page 3-3
•
"Ventricular Detection" on page 3-6
- DRAFT -
3-2
TACHYARRHYTHMIA DETECTION
DEVICE MODE
DEVICE MODE
The Device Mode allows you to program the device to provide the type of
therapy and detection desired.
Ventricular Tachy Mode
The Ventricular Tachy Mode controls the availability of the detection and
therapy functions in the ventricle (Table 3-1 on page 3-2).
You can program the Ventricular Tachy Mode to the following modes:
Table 3-1.
•
Off––disables ventricular tachyarrhythmia detection and automatic
ventricular therapy delivery. This mode is useful during implant or explant,
when connecting the leads to or disconnecting them from the pulse
generator.
•
Monitor Only––enables ventricular tachyarrhythmia detection and episode
storage, but does not automatically deliver therapy to the patient. This
mode is useful in controlled environments, such as during EP testing,
exercise testing, and immediately postoperative, where alternate therapy
(e.g., external defibrillation) is available.
•
Monitor + Therapy––enables the full range of ventricular detection and
ventricular therapy options.
Device feature availability in the Ventricular Tachy Mode settings
Ventricular Tachy Mode
Device features
Rate sensing
Off
Monitor Only
Monitor +
Therapy
Xa
Bradycardia pacing
Ventricular detection/therapy history
Xb
STAT SHOCK
STAT PACE
Real-time annotated EGMs
Ventricular tachyarrhythmia detection
Commanded ventricular ATP
Xc
Commanded ventricular shock
- DRAFT -
TACHYARRHYTHMIA DETECTION
RATE SENSING
Table 3-1.
3-3
Device feature availability in the Ventricular Tachy Mode settings (continued)
Ventricular Tachy Mode
Device features
Off
Monitor Only
Monitor +
Therapy
Xd
Xd
Ventricular EP test
Automatic ventricular tachyarrhythmia therapy
a. In order to enable ventricular sensing when the Ventricular Tachy Mode is programmed to Off, you must program the Brady Mode
to a mode with ventricular sensing.
b. While programmed to Off Mode, the pulse generator will store only STAT SHOCK in history.
c. When the ventricular tachy mode is Monitor + Therapy, the EP Temp V Mode must be programmed to Monitor Only in order to
use the commanded ventricular ATP.
d. Not all forms of EP Tests are available in this mode.
Electrocautery Protection Mode
Electrocautery Protection Mode deactivates the tachyarrhythmia detection and
therapy features of the pulse generator during use of electrocautery equipment.
When Electrocautery Protection is enabled, bradycardia pacing is still
functional; however, the pacing mode switches to an XOO mode (where X
is determined by the programmed pacing mode). Other pacing parameters
remain at the programmed settings.
After cancelling Electrocautery Protection, the following modes will revert to
the previously programmed settings:
•
Ventricular Tachy Mode
•
Brady/CRT Mode
Except for STAT SHOCK and STAT PACE, no commanded therapies,
inductions, or diagnostic tests will be allowed while Electrocautery Protection is
enabled.
Biventricular pacing with LV Offset programmed to zero will be delivered
while Electrocautery Protection Mode is enabled if the programmed mode
is a ventricular pacing mode.
RATE SENSING
Rate sensing is critical to all detection decisions. The pulse generator relies
on the following to determine cardiac cycle length:
•
Bipolar electrodes in the atrium and right ventricle.
- DRAFT -
3-4
TACHYARRHYTHMIA DETECTION
RATE SENSING
•
Automatic gain-controlled sensing circuit for rate sensing. This circuit
ensures proper rate sensing by compensating for changing or diminished
signal amplitudes.
For CRT and bradycardia therapy decisions, rate sensing is based on RV
sensed and ventricular paced events.
Calculating Rates and Refractory Periods
The pulse generator evaluates rate on an interval-by-interval basis. Following
a sensed depolarization, a cycle length is measured and compared to the
programmed detection parameters.
The pulse generator uses refractory periods following paced and sensed
intrinsic events; intrinsic events that fall within these periods are ignored for
detection purposes. The refractory periods, together with noise windows, may
prevent the sensing of nonphysiologic signals and the potential delivery of
unwanted therapy. The nonprogrammable refractory periods are as follows:
•
85 ms atrial refractory following an atrial sensed event
•
150 ms atrial refractory following an atrial pace in DDD(R) and DDI(R)
modes
•
135 ms RV refractory following an RV sensed event or a capacitor charge
•
500 ms refractory following shock delivery (sensing is ignored in all
chambers)
Ventricular Rate Thresholds and Zones
The pulse generator compares each sensed RV cardiac cycle interval against
the programmed Ventricular Tachyarrhythmia Rate Threshold.
A Ventricular Tachyarrhythmia Zone is a range of heart rates defined by at
least one programmed Ventricular Tachyarrhythmia Rate Threshold. You can
program from 1 to 3 Ventricular Tachyarrhythmia Zones, each of which can be
treated by a separate therapy prescription (Table 3-2 on page 3-5, Figure 3-1
on page 3-5).
- DRAFT -
TACHYARRHYTHMIA DETECTION
RATE SENSING
Table 3-2.
3-5
Nominal values for Ventricular Rate Threshold configurations
Ventricular Zone Configuration
VT-1 Zone
VT Zone
VF Zone
1 Zone
– –
– –
200 bpm
2 Zones
– –
160 bpm
200 bpm
3 Zones
140 bpm
160 bpm
200 bpm
Figure 3-1.
Ventricular Tachy Detection settings
•
Rate thresholds in adjacent zones must differ by at least 20 bpm
•
The lowest Ventricular Tachyarrhythmia Rate Threshold must be at least
5 bpm higher than the MTR, MSR, and the MPR
•
The lowest Ventricular Tachyarrhythmia Rate Threshold must be at least
15 bpm higher than the LRL
CRT Delivery Zone and Tachyarrhythmia Zones
The device divides therapy delivery into zones based on heart rate.
•
The programmed LRL and MTR/MSR/MPR define the CRT delivery zone,
or the range over which CRT is delivered.
•
The tachyarrhythmia zones are bounded by the lower rate threshold of
the lowest tachyarrhythmia zone. It is not possible to program the CRT
delivery zone and the tachyarrhythmia zones to overlap. A minimum 5 bpm
difference must exist between the upper limit of the CRT delivery zone and
the lower limit of the tachyarrhythmia zones.
Use of Atrial Information
The atrial rate may be used to:
•
Inhibit ventricular therapy in the presence of atrial fibrillation or atrial flutter
- DRAFT -
3-6
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
•
Bypass ventricular therapy inhibitors if the ventricular rate is faster than
the atrial rate
The pulse generator will respond to atrial sensing regardless of whether an
atrial lead is implanted. If an atrial lead is not implanted or it has a fault,
program the atrial lead from Bipolar to Off (when programmed to Off, atrial
sensing is not performed; atrial pacing will occur). When programming the
atrial lead to Off, change the Brady Mode to prevent atrial sensing or pacing.
Next, on the applicable VT/VT-1 Detection Enhancement screens, program the
following features accordingly to avoid the use of erroneous atrial data:
•
V Rate > A Rate––program to Off (for Onset/Stability detection
enhancement suite)
•
AFib Rate Threshold––program to Off (for Onset/Stability detection
enhancement suite)
•
Atrial Tachyarrhythmia Discrimination––program to Off (for Rhythm ID
detection enhancement suite)
•
Brady/CRT Mode––program to Off, VVI, or VVI(R)
•
Atrial Rate EGM––do not select the atrial trace
NOTE: An atrial EP test should not be performed if the atrial lead is
programmed to Off.
VENTRICULAR DETECTION
Ventricular detection consists of the following components:
•
Initial ventricular detection
•
Reconfirmation/committed shock
•
Redetection and post-shock detection
Initial ventricular detection criteria consist of the programmable parameters
Rate and Duration. The detection criteria may also include one of the following
two detection enhancement suites, which may be used during initial and
post-shock ventricular detection to add specificity beyond Rate and Duration.
•
Onset/Stability
•
Rhythm ID
- DRAFT -
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
3-7
The pulse generator initiates ventricular therapy when it determines that
detection is met. Ventricular detection is met when all of the following occur:
•
A ventricular zone’s detection window becomes and remains satisfied
throughout Duration
•
The ventricular zone’s Duration expires
•
A higher ventricular zone’s detection window is not satisfied
•
Detection enhancements (if programmed to On) indicate therapy
•
The last detected interval is in the ventricular zone
If the above criteria are not met, therapy is not initiated and the pulse generator
continues to evaluate intervals.
Ventricular Detection Enhancement Suites
One of the following ventricular detection enhancement suites may be
programmed to provide specificity beyond Rate and Duration (Table 3-3 on
page 3-7):
•
Rhythm ID
•
Onset/Stability
Detection enhancement suites are not available in the VF zone.
Table 3-3.
Detection enhancement suites available per zone
3-zone configurationa
3-zone configuration (with Monitor Only
zone)b
VT-1 Zone
VT Zone
VF Zone
Rhythm ID
Rhythm ID
None
Onset/Stability
Onset/Stabilityc
None
Rhythm ID
None
Onset/Stability
Rhythm ID
2-zone configuration
None
Onset/Stability
2-zone configuration (with Monitor Only
zone)b
None
None
None
1-zone configuration
a. If the detection enhancement suite is enabled in a 3-zone configuration, it applies to both the VT-1 and VT zones. Detection
enhancement suites cannot be independently enabled for each zone.
b. Detection enhancement suites are not available in the lowest zone of a multi-zone configuration when the zone is used as a
Monitor Only zone (no therapy programmed for that zone).
c. Shock if unstable is the only Onset/Stability detection enhancement available in the VT zone of a 3-zone configuration (applies
only to 3-zone configuration without a Monitor Only zone).
- DRAFT -
3-8
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
NOTE: There is no clinical data to suggest that one detection enhancement
suite is superior to the other for any given patient indication. Therefore,
individual programming and evaluation of detection enhancement specificity
is recommended.
Rhythm ID
Rhythm ID uses Vector Timing and Correlation analysis in addition to atrial and
ventricular interval analysis to determine if a patient’s rhythm should be treated
(VT) or if therapy should be inhibited (SVT).
With Rhythm ID, the pulse generator performs a vector timing and correlation
analysis using the shock EGM and rate EGM. Based on this data, it saves a
reference template of the patient’s normal sinus rhythm.
During Rhythm ID analysis, the pulse generator first determines if the
ventricular rate is greater than the atrial rate. If so, therapy will be initiated. If
the ventricular rate is not greater than the atrial rate, Rhythm ID evaluates the
following criteria to determine if therapy should be inhibited or initiated:
•
Vector Timing and Correlation analysis during initial detection determines
if the rhythm is SVT by comparing it to the previously stored reference
template. If the rhythm is declared SVT, therapy is inhibited.
•
If Vector Timing and Correlation does not declare the rhythm SVT, Stability
and AFib Rate Threshold determine if the ventricular rhythm is unstable
and the atrial rate is fast. If the ventricular rhythm is unstable and the atrial
rate is fast, the rhythm is declared SVT and therapy is inhibited.
Rhythm ID does not consider atrial detection criteria (V Rate > A Rate or A
greater than AFib Rate Threshold) for the following configurations:
•
Dual-chamber devices if Atrial Tachyarrhythmia Discrimination is
programmed to Off
When configured this way, Stability is not evaluated for initial detection. This
may be useful in instances where atrial lead problems have occurred. For
these configurations,therapy is inhibited at initial detection if the rhythm is
declared SVT (correlated based on Vector Timing and Correlation). Otherwise,
therapy is initiated.
Two methods are available for the device to automatically acquire a Rhythm
ID reference template: passive and active. The active method may be useful
for patients who are frequently ventricular paced.
- DRAFT -
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
3-9
If the passive method is enabled, the pulse generator will attempt to collect the
Rhythm ID reference template every two hours using the programmed brady
settings.
If the active method is enabled and seven days have passed since the last
successful collection of a reference template, then every 28 hours the device
automatically analyzes the patient’s intrinsic rhythm by adjusting the brady
parameters. During a Rhythm ID active reference template update, the
following will occur:
1. The device verifies that the patient is at rest (as measured by the
accelerometer input).
2. The device enables a controlled pacing rate decrease to the programmed
Rhythm ID Fallback LRL. During this fallback period, the following occurs:
•
The device temporarily switches the pacing mode to DDI, VDI, VVI,
AAI, or Off (according to the programmed brady mode) and extends
the AV delay up to 400 ms.
•
Biventricular Trigger, Rate Smoothing, ATR, hysteresis, and dynamic
programming (excluding Dynamic VRP) are suspended. The pacing
chamber is set to Biventricular; LV Offset is set to 0.
3. After the Fallback period, pacing parameters are restored to normal
programmed parameters. Fallback periods occur no more than once per
day and will typically last less than one minute.
A method for manually commanding the device to acquire a Rhythm ID
reference template is also available.
NOTE: If Rhythm ID is not enabled, a manual reference template update can
still be performed. However, the acquired template will not be used in analysis
to determine if the patient’s rhythm is VT or SVT.
During a manual Rhythm ID reference template update, the pulse generator will
perform the following tasks:
1. Enable a controlled rate decrease to the programmed Rhythm ID Fallback
LRL. During the fallback period, the following occurs:
•
The device temporarily switches to the programmed Manual Rhythm
ID Brady Mode and extends the AV delay up to 400 ms.
- DRAFT -
3-10
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
•
Biventricular Trigger, Rate Smoothing, ATR, hysteresis, and dynamic
programming (excluding Dynamic VRP) are suspended. The pacing
chamber is set to Biventricular; LV Offset is set to 0.
2. After the Fallback interval, pacing parameters are restored to normal
programmed parameters. This process will typically last one minute.
NOTE: Rhythm ID Fallback LRL settings should be selected such that normal
sinus rhythms are promoted (e.g., normal AV node conduction). Care must be
used when selecting LRL less than 50 ppm (rates that approach the patient’s
ventricular escape rates). Ventricular escape rhythms during Rhythm ID
updates may result in inappropriate therapy decisions.
NOTE: A manual Rhythm ID reference template update should not be
commanded immediately after shock therapy. It may take several minutes for
irregularities in EGM morphology caused by the shock to subside.
Onset/Stability
The Onset/Stability detection enhancement suite analyzes the cardiac cycle
intervals to determine if a patient’s rhythm should be treated (VT) or if therapy
should be inhibited (SVT).
Onset/Stability allows you to program detection enhancements by identifying
the desired type of rhythm discrimination: atrial tachyarrhythmia, sinus
tachycardia, or polymorphic VT (Table 3-4 on page 3-10).
Table 3-4.
Onset/Stability rhythm discrimination available per zone
VT-1 Zone
3-zone Configuration
Atrial Tachyarrhythmia
VT Zone
Polymorphic
VF Zone
VTa
None
Sinus Tachycardia
3-zone Configuration
(with Monitor Only zone)b
None
Atrial Tachyarrhythmia
None
Sinus Tachycardia
Polymorphic VTa
2-zone Configuration
Atrial Tachyarrhythmia
Sinus Tachycardia
Polymorphic VTa
- DRAFT -
None
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
Table 3-4.
3-11
Onset/Stability rhythm discrimination available per zone (continued)
VT-1 Zone
VT Zone
VF Zone
None
None
2-zone Configuration (with Monitor Only zone)b
None
1-zone Configuration
a. Polymorphic VT Discrimination is only available in the VT zone.
b. Rhythm discrimination is not available in the lowest zone of a multi-zone configuration if the zone is used as a Monitor Only
zone (no therapy programmed for that zone).
Reconfirmation/Committed Shock
Reconfirmation refers to the monitoring performed by the device during and
immediately following capacitor charging for a shock. When the Committed
Shock parameter is programmed to Off, the device is allowed to reconfirm that
a shock should be delivered.
Ventricular Redetection
Ventricular Redetection occurs following any:
•
Ventricular therapy delivery
•
Diverted therapy due to reconfirmation analysis (diverted-reconfirm)
•
Manually diverted therapy
•
Therapy not available at Detection Met
Redetection uses the same ventricular detection window process and
programmed tachycardia rate thresholds as initial detection to identify a
tachyarrhythmia.
The primary differences between initial detection and redetection are the
duration parameters used and the detection enhancements that are available:
•
If ventricular shock therapy is delivered, the following will occur:
–
The redetection duration time is determined by the value of the
Post-shock Duration parameter
–
Detection enhancements (except for Vector Timing and Correlation)
are available during redetection
- DRAFT -
3-12
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
•
If ventricular ATP is delivered or if therapy is diverted or unavailable, the
following will occur:
–
The redetection duration time is determined by the Redetection
Duration parameter
–
Detection enhancements (except for Shock if Unstable) are not
available during redetection
Whichever duration is determined to be appropriate, that type of duration
(Redetection or Post-shock) will be in effect in all zones at each zone’s
programmed duration value.
Ventricular Post-shock Detection Enhancements
When programmed to On, the following ventricular post-shock detection
enhancements will be in effect following the Post-shock Duration:
•
Post-shock V Rate > A Rate
•
Post-shock AFib Rate Threshold
•
Post-shock Stability
•
Post-shock SRD
•
Post-shock Rhythm ID (uses AFib Rate Threshold, Stability, V Rate > A
Rate, and SRD)
With the exception of Rhythm ID, all post-shock detection enhancements
perform the same as the corresponding Initial Detection enhancements (with
Rhythm ID, Vector Timing and Correlation is not available post-shock).
Post-shock Stability may be used to prevent shock-induced AF from causing
the pulse generator to deliver undesired additional shocks (Figure 3-2 on
page 3-13.)
The AFib Rate Threshold can be programmed in conjunction with Post-shock
Stability to further discriminate AF and prevent the pulse generator from
delivering undesired ventricular shock therapy.
- DRAFT -
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
Evaluate Stability. If
stable, charge and
shock. If unstable, inhibit
therapy. Post-shock
SRD timer starts.
Shock
Lowest zone
detection window
satisfied.
0s
Post-shock
Duration starts in
lowest zone.
0s
Continue Post-shock Stability
analysis throughout SRD time
as long as the detection
window remains satisfied.
If Stability indicates
therapy, charge and
shock.
5s
Post-shock Duration
expires.
3-13
Post-shock SRD times out.
Charge and deliver shock if
therapy had been inhibited
during SRD.
15 s
20 s
Post-shock Duration = 5 s
Post-shock SRD = 15 s
Figure 3-2.
Post-shock Duration and Post-shock Stability analysis
Ventricular Detection Details
The pulse generator uses the following information to determine appropriate
therapy delivery:
•
Ventricular detection windows
•
Duration parameter
•
Redetection duration and post-shock duration
•
Ventricular episodes
•
Ventricular detection enhancements
Ventricular Detection Windows
Appropriate therapy delivery is dependent upon accurately classifying a
patient’s rhythm. To ensure that appropriate therapy is delivered, the pulse
generator employs detection windows to differentiate tachycardias.
Each ventricular zone has a detection window that consists of the 10 most
recent RV R–R intervals measured by the pulse generator. As each new
interval is measured, it is compared to each zone’s programmed rate threshold
and classified as either fast or slow (i.e., above or below the rate threshold) in
each detection window.
The pulse generator prepares for a potential episode when it counts 3
consecutive fast intervals. The detection window is satisfied and an episode is
declared when 8 out of 10 fast intervals are counted. The detection window
will remain satisfied as long as 6 of 10 intervals remain classified as fast. If the
number of fast intervals falls below 6, the zone’s detection window is no longer
- DRAFT -
3-14
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
satisfied. The zone’s detection window will only become resatisfied when 8 of
10 intervals are again classified as fast (Figure 3-3 on page 3-14).
VF
window
Intervals
in ms
630
600
600
600
380
390
370
380
370
370 350 370
490
500
500
500
500 370
Right
Ventricle
6 of 10 intervals are fast, VF window not satisfied.
7 of 10 intervals are fast, VF window not satisfied.
8 of 10 intervals are fast, VF window is satisfied.
7 of 10 intervals are fast, VF window remains satisfied.
6 of 10 intervals are fast, VF window remains satisfied.
5 of 10 intervals are fast, VF window is no longer satisfied.
6 of 10 intervals are fast, VF
window is not yet resatisfied
One-zone configuration
7 of 10 intervals are fast, VF
window is not yet resatisfied
VF Zone Rate = 150 min-1 (bpm) = 400 ms
S = Slow
Figure 3-3.
8 of 10 intervals are fast, VF
window is resatisfied
F = Fast
Ventricular detection window satisfied
Because Rate Threshold in the higher zones must be programmed at a value
greater than Rate Threshold in lower zones, an interval classified as fast
in a higher window would also be classified as fast in any lower windows
(Figure 3-4 on page 3-15).
- DRAFT -
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
VF Zone Rate = 200 min-1 (bpm) = 300 ms
8 of 10 intervals are fast, VF window is satisfied.
VT Zone Rate = 150 min-1 (bpm) = 400 ms
S = Slow
3-15
F = Fast
7 of 10 intervals are fast, VF window not satisfied.
6 of 10 intervals are fast, VF window not satisfied.
VF window
Intervals in ms
630
600
VT window
600
600
370
380
390
275
275
275 250 250
240
240 240
6 of 10 intervals are fast, VT window not satisfied.
7 of 10 intervals are fast, VT window not satisfied.
8 of 10 intervals are fast, VT window is satisfied.
Figure 3-4.
Interaction of ventricular detection windows, 2-zone configuration
Duration Parameter
The Duration parameter is a timer that measures the length of time in each
zone that a rhythm must be sustained before therapy is delivered.
A Duration timer begins when its respective zone’s detection window is
satisfied. The programmed Duration time is checked following every cardiac
cycle to determine if it has expired.
NOTE: Since the Duration timer is examined synchronously with a cardiac
cycle, the programmed Duration may be exceeded by up to one full cardiac
cycle.
•
As long as the zone’s detection window remains satisfied, the Duration
timer continues to elapse. If the last detected interval is in the zone when
its Duration time expires, detection is considered met and therapy is
initiated (assuming no programmed detection enhancements inhibit therapy
delivery) (Figure 3-5 on page 3-16).
•
If the last detected interval is not in the zone, therapy is not initiated. Each
subsequent interval will be checked until an interval is in the original zone,
or the window is no longer satisfied (Figure 3-6 on page 3-16).
- DRAFT -
3-16
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
•
If at any point during Duration a zone’s detection window detects fewer
than 6 of 10 fast intervals, that zone’s Duration is reset to 0 (Figure 3-7 on
page 3-17). Duration will start again only if the detection window becomes
resatisfied.
VT detection window remains satisfied during
VT Duration.
Right Ventricle
VT detected
VT Duration 1-30 s
VT window satisfied.
VT Duration starts.
VT Duration expires and the
next detected interval is in the
VT zone, detection is met.
ATP therapy is initiated.
Duration starts when a window becomes satisfied and continues to elapse as long as the ventricular detection window
remains satisfied. Detection is met when Duration expires and the next detected interval is in the same ventricular
zone.
Figure 3-5.
Ventricular Duration timer
VT detection window remains satisfied during
VT Duration.
Right ventricle
VT detected
VT Duration 1-30 s
VT window satisfied.
VT Duration starts.
Figure 3-6.
VT Duration expires and the last
detected interval is not in the
same zone, detection in that
zone is not met. Therapy in that
zone is not initiated.
Last detected interval
- DRAFT -
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
3-17
VT detection window is no longer satisfied; fewer
than 6 of 10 intervals are classified as fast.
Right ventricle
VT detected
VT Duration 1-30 seconds
VT window satisfied.
VT Duration starts.
Need 8 of 10 intervals classified as fast to restart VT Duration.
VT Duration resets to zero. VT Duration will start
again when the window becomes resatisfied.
Duration resets when during the Duration period the window is no longer satisfied.
Figure 3-7.
Ventricular Duration reset
A Duration is programmed for each ventricular zone. Different values are
available depending on the configuration programmed (Table 3-5 on page 3-17).
The Duration programmed in lower ventricular rate zones must be greater than
or equal to higher ventricular zones. Longer Durations may be used to prevent
the device from initiating treatment of non-sustained arrhythmias.
Table 3-5.
Duration programmable ranges by ventricular zone and configuration
VT-1 Zonea
Configuration
VT Zonea
VF Zoneb
1 Zone
––
––
1–15 seconds
2 Zones
––
1–30 seconds
1–15 seconds
3 Zones
1–60 seconds
1–30 seconds
1–15 seconds
a. The maximum redetect duration for the VT-1 and VT Zones is 15 seconds.
b. In the VF Zone, the redetect and post-shock duration is fixed at 1 second.
Duration in a Multi-zone Configuration
Duration timers run independently of each other within their respective
ventricular zones.
•
If the arrhythmia is detected in the highest zone, that zone’s Duration timer
takes precedence over the lower zones’ timers; the lower zones’ Duration
timers continue to elapse but are ignored while the higher zone’s Duration
timer runs.
•
If the higher zone’s Duration expires and detection is met, therapy for that
zone will be initiated regardless of whether the lower zones’ Duration
timers have expired.
- DRAFT -
3-18
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
•
If the higher zone’s detection window does not remain satisfied, then the
Duration timers for the lower ventricular zones are no longer ignored.
Programmed therapy for lower ventricular zones will be initiated when a lower
ventricular zone’s duration is met and no higher ventricular zone’s window is
satisfied (Figure 3-8 on page 3-18, Figure 3-9 on page 3-18).
VF window satisfied.
VF Duration starts.
VF window remains satisfied. VF Duration expires
and detection is met. Charging begins. VT Duration
expires during charging, but is ignored.
VF detected
Charging
VF Duration 1 s
VT detected
VT Duration 5 s
VT window satisfied.
VT Duration starts.
Figure 3-8.
Interaction of ventricular Duration, 2-zone configuration, charging
VF window satisfied.
VF Duration starts.
VF window does not remain satisfied.
VF Duration is not satisfied.
VT Detection is met.
VF detected
VF Duration 1 s
VT Duration 5 s
VT detected
Charging
VT window satisfied.
VT Duration starts.
Figure 3-9.
VT Duration expires. Detection is not met
because VF window is satisfied.
Interaction of ventricular Duration, 2-zone configuration, charging delayed
Ventricular Redetection Duration and Post-shock Duration
Duration parameters are used to identify tachyarrhythmias during the ventricular
redetection process.
•
Redetection Duration is applied following delivery of ATP therapy (except
QUICK CONVERT ATP), a diverted-reconfirm, manually diverted therapy,
or if therapy is unavailable at Detection Met (Figure 3-10 on page 3-19).
•
Post-shock Duration is applied following shock therapy delivery (Figure 3-11
on page 3-19).
- DRAFT -
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
3-19
Redetection Duration is programmable in the lower ventricular zones of a
multi-zone configuration. It is nonprogrammable in the VF Zone. Post-shock
Duration can be programmed in the same manner; the values programmed in
the lower ventricular rate zones must be greater than or equal to the values
programmed in the higher zones.
To help minimize time to potential therapy, it is recommended that Redetection
Duration in the VT-1 and VT zones of multi-zone configurations be programmed
at less than or equal to 5 seconds.
It is recommended that Post-shock Duration in the VT-1 and VT zones of
multi-zone configurations also be programmed at less than or equal to 5
seconds. However, you may program for longer durations if shock-induced,
non-sustained, high-rate rhythms, such as accelerated idioventricular rhythm
(AIVR) or AF are evident. The longer durations may allow the rhythm to return
to a lower rate before redetection is met.
Ignored
8 of 10 intervals are
fast, window satisfied
ATP therapy
Initial Detection met
Ignored
Start
Redetection
Redetection
Duration expires.
Detection met.
Start
Redetection
Duration
ATP therapy programmed.
Redetection Duration programmed at 3 s.
Figure 3-10.
ATP therapy
3s
Start Redetection
Redetection following ventricular ATP delivery
Ignored
Ignored
Shock
Charging
Initial Detection met
Shock
8 intervals are fast
Start Redetection
3s
Start Post-shock
Duration
Charging
Post-shock
Duration expires.
Detection met.
Look for 8 of 10 fast intervals
Start Redetection
Shock therapy programmed. Post-shock
Duration programmed at 3 s.
Figure 3-11.
Look for 8 of 10 fast intervals
Redetection following ventricular shock delivery
- DRAFT -
3-20
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
Ventricular Episodes
If three consecutive fast ventricular beats are detected, then the pulse
generator performs the following:
•
Increments the episode number
•
Allocates memory for history data and electrogram storage
•
Starts monitoring for a detection window to be satisfied
When any zone’s detection window becomes satisfied, the start of a ventricular
episode is declared and duration timers begin in those zones where detection
windows are satisfied. The ventricular episode is declared complete when all
detection windows are no longer satisfied and remain unsatisfied for a specified
time.
Each ventricular tachy episodes is classified as treated or non-treated
(Figure 3-12 on page 3-21 through Figure 3-16 on page 3-23).
•
A treated episode is one in which therapy is delivered
•
A nontreated episode is one in which no therapy is delivered
For a treated episode, an End-of-Episode timer starts after therapy is delivered.
For a non-treated episode, an End-of-Episode timer starts at the point that
the pulse generator recognizes that all detection windows are no longer
satisfied. The End-of-Episode time interval is intended to allow the patient
to stabilize before initial detection and initial therapy are used again. The
episode is declared complete if no detection window becomes satisfied for a
specified time following the last therapy attempt (Table 3-6 on page 3-20). If
any window becomes satisfied while the End-of-Episode timer is elapsing, the
End-of-Episode timer is reset to zero. It will start again when either therapy is
attempted or all windows are not satisfied (Figure 3-16 on page 3-23).
Once an episode has been declared complete, the pulse generator will apply
initial detection and therapy to subsequent tachyarrhythmias.
Table 3-6.
End-of-Episode Timer
Episode Classification
Nontreated (no therapy delivered)
Ventricular End-of-Episode Timer
(elapsed time required to declare episode over)
10 seconds
- DRAFT -
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
Table 3-6.
3-21
End-of-Episode Timer (continued)
Episode Classification
Ventricular End-of-Episode Timer
(elapsed time required to declare episode over)
Treated (only ATP therapy delivered)
10 seconds
Treated (any shock therapy delivered)
30 seconds
NOTE: The episode is terminated immediately if the Tachy Mode is
reprogrammed, an induction method or lead test is attempted before the
End-of-Episode time out, or any ventricular detection or ventricular therapy
parameters are reprogrammed.
Coupling
Interval
Detection window does
not become satisfied.
Detection window remains satisfied.
Detection window satisfied.
ATP Burst
Start Duration.
Start episode.
Figure 3-12.
Detection met.
Deliver ATP.
10 s
Treated episode, ventricular mode is Monitor + Therapy and ATP is delivered
Detection window remains satisfied.
Detection window does not become satisfied.
Detection window satisfied.
Shock
30 s
Charging
Start Duration.
Start episode.
Figure 3-13.
End-of-Episode times out.
Episode is over.
Start End-ofEpisode timer.
Detection met.
Start Charging.
Start End-of-Episode timer.
End-of-Episode times out.
Episode is over.
Treated episode, ventricular mode is Monitor + Therapy and shock is delivered
- DRAFT -
3-22
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
Detection window is no longer satisfied; fewer than 6
of 10 intervals are classified as fast.
Detection window does not become satisfied.
Detection window satisfied.
10 s
Start Duration.
Start episode.
Figure 3-14.
expired
Duration did not expire.
Duration resets to zero.
Start End-of-Episode timer.
End-of-Episode times out.
Episode is over.
Nontreated episode, ventricular mode is Monitor + Therapy or Monitor Only, duration is not
F = Fast
S = Slow
Reconfirmation monitors rhythm for five of ten
slow intervals during charging.
Detection window remains satisfied.
Right Ventricle
F S S S S
Detection window
satisfied.
Ta chyarrhythmia sp ontan eou sly conve rts.
10 s
Charging
Start Duration.
Start episode.
Detection met.
Start Charging.
Reconfirmation indicates
tachyarrhythmia not present;
shock not delivered. Start
End-of-Episode timer.
End-of-Episode
times out.
Episode is over.
This example assumes Committed Shock is programmed to Off.
Figure 3-15. Nontreated episode, ventricular mode is Monitor + Therapy and charging is stopped prior
to shock delivery
- DRAFT -
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
3-23
Right Ventricle
Ventricular rate
remains fast.
Shock
Detection window does not become satisfied.
Detection window
satisfied.
Shock
30 s
Charging
Start End-ofEpisode timer.
Start Post-shock
Detection met.
Duration
Start Charging.
End-of-Episode
timer resets to zero.
Restart End-of-Episode timer.
End-of-Episode times
out. Episode is over.
This example illustrates a Treated Episode when Ventricular mode is Monitor + Therapy. The End-of-Episode timer is
reset to 0 when a ventricular detection window becomes satisfied after ventricular therapy delivery, but prior to the
episode time-out being reached. In this example, 2 shocks were delivered in the episode.
Figure 3-16.
Treated episode, ventricular mode is Monitor + Therapy and End-of-Episode timer is reset to 0
Ventricular Detection Enhancements
Ventricular detection enhancements add specificity to the Rate and Duration
detection criteria. You may program ventricular detection enhancements to
perform the following:
•
Delay or inhibit therapy delivery
•
Override therapy inhibition
•
Bypass a sequence of ATP therapy in favor of shock therapy
Ventricular detection enhancements may be programmed to one of the
following:
•
Rhythm ID
•
Onset/Stability
•
Off (i.e., rate only)
If Off is selected, only the ventricular rate and duration are used for therapy
decisions.
- DRAFT -
3-24
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
If either Rhythm ID or Onset/Stability is selected, enhancement parameters are
used in addition to ventricular rate and duration for therapy decisions (Table 3-7
on page 3-24) as follows:
Table 3-7.
•
Vector Timing and Correlation inhibits therapy when the conduction vector
(EGM morphology and timing) during tachyarrhythmia matches a reference
conduction vector of the patient’s normal sinus rhythm.
•
V Rate > A Rate can be used to override the inhibit decision of Onset,
Stability, Vector Timing and Correlation, and/or AFib Rate Threshold. V
Rate > A Rate can be used to deliver ventricular therapy anytime the
ventricular rate is greater than the atrial rate.
•
AFib Rate Threshold can be programmed (together with stability) to inhibit
ventricular therapy if the atrial rhythm is fast.
•
Stability can be programmed to inhibit ventricular therapy delivery if the
ventricular rhythm is unstable.
•
Shock if Unstable can be programmed to bypass the ventricular ATP
therapy and deliver shock therapy if the ventricular rhythm is declared
unstable.
•
Onset can be programmed to inhibit ventricular therapy if the patient’s
heart rate increases gradually.
•
SRD enables the pulse generator to override the Stability, Onset, Vector
Timing and Correlation, and/or AFib Rate Threshold parameters’ decision
to inhibit ventricular therapy if the high rate continues throughout the
programmed time period.
Enhancement parameters available with detection enhancements
Enhancement Parameter
Vector Timing and
Correlationa
Rhythm ID
Onset/Stability
Initial
Post-Shock
Initial
Post-Shock
– –
– –
– –
V Rate > A Rate (dual-chamber devices only)
Xb
Xb
AFib Rate Threshold (dual-chamber devices only)
Xb d
Xb d
Xe
Xe
Stability (to inhibit)
Xf
Xf
Shock if Unstable
– –
– –
– –
- DRAFT -
3-25
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
Table 3-7.
Enhancement parameters available with detection enhancements (continued)
Rhythm ID
Enhancement Parameter
Onset/Stability
Initial
Post-Shock
Initial
Post-Shock
Onset
– –
– –
– –
SRDg
a. This enhancement is not individually programmable.
b. When Rhythm ID is selected, this enhancement is automatically enabled when Atrial Tachyarrhythmia Discrimination is
programmed to On. However, it is not available in single chamber devices or when the Atrial Tachyarrhythmia Discrimination is
programmed to Off in dual chamber devices.
c. This enhancement is not individually programmable when Rhythm ID is enabled.
d. When Rhythm ID is selected, this parameter uses the same value for both initial and post-shock detection. It cannot be
independently enabled or disabled for post-shock detection.
e. When Onset/Stability is selected, this parameter can be enabled and disabled independently for post-shock detection. If enabled,
it uses the same value as the initial detection.
f. When Rhythm ID is enabled and Atrial Tachyarrhythmia Discrimination is programmed to On in dual chamber devices, this
enhancement uses the same value for both initial and post-shock detection. In single chamber devices, or when Atrial
Tachyarrhythmia Discrimination is programmed to Off, this enhancement is automatically disabled for Initial Detection, but is still
enabled for post-shock detection.
g. SRD is available when detection enhancements, which inhibit therapy, are programmed.
Some of these detection enhancement parameters are also independently
programmable as post-shock parameters (Table 3-7 on page 3-24).
The individual detection enhancement parameters that are available depend
on the number of tachy zones that are programmed: 3, 2, or 1 (Table 3-8 on
page 3-25).
Table 3-8.
Individual Ventricular Detection Enhancements available in multizone configurations
VT-1 Zone
3-zone configuration
VT Zone
Vector Timing and
Correlation
Vector Timing and
Correlationa
V Rate > A Rate
V Rate > A Ratea
AFib Rate Threshold
AFib Rate Thresholda
Stability to Inhibit
Stability (to Inhibit)a
Onset
Shock if Unstable
SRD
SRDa
- DRAFT -
VF Zone
––
3-26
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
Table 3-8.
Individual Ventricular Detection Enhancements available in multizone configurations (continued)
VT-1 Zone
3-zone configuration
(with Monitor Only zone)b
––
VT Zone
Vector Timing and Correlation
VF Zone
––
V Rate > A Rate
AFib Rate Threshold
Stability (to Inhibit)
Shock if Unstablec
Onset
SRD
2-zone configuration
Vector Timing and Correlation
––
V Rate > A Rate
AFib Rate Threshold
Stability (to Inhibit)
Shock if Unstablec
Onset
SRD
2-zone configuration (with Monitor Only
zone)b
––
––
––
1-zone configuration
a. Enhancement is available in the middle zone of a 3-zone configuration only when Rhythm ID is enabled.
b. Detection enhancements are not available in the lowest zone of a multi-zone configuration when it is used as a Monitor Only
zone (no therapy programmed for that zone).
c. Shock if Unstable cannot be programmed on in the same zone as other detection enhancements that are programmed to
inhibit therapy (Onset, Stability, and AFib Rate Threshold).
When a specific rhythm discrimination is selected, preselected values are
displayed for the detection enhancements that are suitable for discriminating
that rhythm. However, you can modify those values at your discretion (Table 3-9
on page 3-27).
- DRAFT -
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
Table 3-9.
3-27
Preselected values for initial detection and redetection enhancements
Rhythm ID
Onset/Stability
Parameter
Atrial
Tachyarrhythmia
Discrimination
Sinus
Tachycardia
Discrimination
Polymorphic
VT
Discrimination
Atrial
Tachyarrhythmia
Discrimination
On
Atrial
Tachyarrhythmia
Discrimination
Off
Vector
Timing and
Correlation
– –
– –
– –
Ona
Ona
V Rate >
A Rate
(dual-chamber
models
only)
On
On
– –
Onb
– –
AFib Rate
Threshold
(dual-chamber
models
only)
170 bpm
– –
– –
170 bpm
– –
Stability
(Inhibit)
20 ms
– –
– –
20 ms
30 ms
Onset
(initial
detection
only)
– –
9%
– –
– –
– –
SRD Initial
3:00
minutes:seconds
3:00
minutes:seconds
– –
3:00
minutes:seconds
3:00
minutes:seconds
SRD
Redetection
0:15
minutes:seconds
– –
– –
0:15
minutes:seconds
0:15
minutes:seconds
– –
– –
30 ms
– –
– –
Shock if
Unstable
a. Parameter is not individually programmable.
b. Parameter is not individually programmable when Rhythm ID is enabled.
Vector Timing and Correlation
Vector Timing and Correlation compares EGM signals for an unknown rhythm
with a stored reference template of the EGM signals of a normal sinus rhythm
(NSR). Rhythms that are not similar to the reference template (are not
correlated) are classified as VT. Rhythms that are correlated with the reference
template are classified as SVT.
When an unknown fast rhythm is sensed in the VT or VT-1 zones, each beat
of the fast rhythm is compared to the stored reference template. The pulse
- DRAFT -
3-28
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
generator calculates a Feature Correlation Coefficient and makes the following
therapy decisions based on the calculation:
•
If at least 3 out of 10 beats are correlated, the rhythm is classified as SVT
•
If fewer than 3 out of 10 beats are correlated, the rhythm is classified as VT
Rhythm ID makes its decision to treat or inhibit therapy at the end of Duration. If
the decision is made to inhibit therapy, Rhythm ID (including Vector Timing and
Correlation, V Rate > A Rate, AFib Rate Threshold, and Stability) continues
to be recalculated beat-by-beat throughout SRD.
V Rate > A Rate
The V Rate > A Rate (ventricular rate greater than atrial rate) enhancement
compares the atrial and ventricular rates to classify the type of fast ventricular
rhythm. When the ventricular rate is greater than the atrial rate, therapy will
be initiated regardless of the analysis of the other programmed detection
enhancements.
Analysis is made by comparing the average rate of the last 10 ventricular
intervals prior to the end of Duration to the average rate of the last 10 atrial
intervals prior to the end of Duration (Figure 3-17 on page 3-29). If fewer than
10 atrial intervals are available, those intervals will be used to calculate the
average atrial rate. This analysis is performed using the following criteria:
•
If the average ventricular rate is greater than the average atrial rate by at
least 10 bpm, the ventricular rate is declared to be faster than the atrial
rate (indicated as True on the Episode Detail Report), and therapy will
be initiated.
•
If the average ventricular rate is not greater than the average atrial rate by
at least 10 bpm (indicated as False on the Episode Detail Report), therapy
may continue to be inhibited. The Episode Detail report will indicate the
measured value even though the parameter may be programmed to Off.
If therapy is inhibited, the V Rate > A Rate analysis continues until either the
ventricular rate is greater than the atrial rate or other enhancements indicate
therapy treatment, at which time therapy will be initiated.
NOTE: V Rate > A Rate is not evaluated during redetection following ATP
therapy.
- DRAFT -
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
Number of Intervals
Intervals in
ms
620
600
580
580
620
600
640
560
3-29
600
Atrium
Compute the sum of the atrial intervals.
Average = 600 ms = 100 min-1 (bpm).
Number of
Intervals
Average ventricular rate (150 min-1
[bpm]) is greater than average
atrial rate (100 min-1 [bpm]) by at
least 10 bpm, so delivery therapy.
Compute the sum of 10 most recent ventricular
intervals. Average = 400 ms = 150 min-1 (bpm).
7 8
400 420 380 390 410 400 360 440 410 390
Intervals
in ms
ATP therapy
Ventricle
VT detected
Third fast interval (of the 8 out
of 10 that satisfy the detection
window)
Figure 3-17.
Start Duration.
Evaluate Onset.
Duration expired.
Onset is gradual.
Analyze V Rate > A Rate.
Deliver therapy because V > A.
V Rate > A Rate analysis
V Rate > A Rate can be programmed to bypass inhibitors (Vector Timing and
Correlation, AFib Rate Threshold, Stability, and/or Onset) and initiate therapy in
the event that the ventricular rate is faster than the atrial rate.
NOTE: Refer to "Use of Atrial Information" on page 3-5 for additional
information about device performance when the atrial lead is programmed to
Off.
NOTE: In a Rhythm ID configuration, the evaluation of V Rate > A Rate is
linked to the AFib Rate Threshold. If Atrial Tachyarrhythmia Discrimination is
programmed to Off, the AFib Rate Threshold and V Rate > A Rate detection
enhancements are not evaluated.
AFib Rate Threshold
AFib Rate Threshold analysis identifies AF by comparing the atrial rate to the
programmed AFib Rate Threshold.
AFib Rate Threshold cannot be enabled without also enabling the Stability
detection enhancement. The device analyzes both parameters to determine
whether to withhold or deliver therapy.
If the intrinsic atrial rate is greater than the AFib Rate Threshold and the
ventricular rhythm is classified as unstable, the ventricular rhythm is declared
to be due to AF.
- DRAFT -
3-30
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
The intrinsic atrial rate is declared to be above the AFib Rate Threshold in the
following manner (Figure 3-18 on page 3-30):
•
Atrial analysis begins at initiation of ventricular tachyarrhythmia detection.
Each atrial interval is classified as faster or slower than the AFib Rate
Threshold Interval.
•
When 6 of the last 10 intervals are classified as faster than the AFib Rate
Threshold, the device declares AF to be present.
•
Ventricular stability is then checked. If unstable, therapy is inhibited.
In the event that ventricular therapy is not delivered, the atrial rate continues
to be examined. As long as 4 of 10 intervals remain classified as fast, AF
is considered present. Therapy is inhibited by AFib Rate Threshold/Stability
until any of the following occur:
•
The atrial rate drops below the AFib Rate Threshold
•
The ventricular rhythm becomes stable
•
If programmed to On, V Rate > A Rate is true
•
SRD times out
Only 3 of 10 intervals are
above the AFib Rate Threshold
4 of 10 > AFib Rate Threshold
4 of 10 intervals must remain
> AFib Rate Threshold
6 of 10 atrial intervals are greater
than AFib Rate Threshold
Atrium
ATP therapy
Ventricle
VT detected
Third fast interval (of the 8
out of 10 that satisfy the
detection window)
Figure 3-18.
Detection satisfied.
Atrial rate >
Start duration.
AFib Rate
Start episode.
Threshold.
Start Stability analysis.
Duration expires.
Atrial rate > AFib Rate Threshold.
Rhythm declared unstable.
Therapy is inhibited.
Interaction of AFib Rate Threshold and Stability
- DRAFT -
Rhythm remains unstable.
Atrial rate is less than
AFib Rate Threshold.
Initiate therapy.
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
3-31
When AFib Rate Threshold and Stability are used alone, ventricular therapy
is initiated when a stable rhythm is declared. Ventricular therapy is initiated
for an unstable rhythm when it is determined that the atrial rate is less
than the AFib Rate Threshold (Table 3-10 on page 3-31). When AFib
Rate Threshold and Stability are used with other inhibitor enhancements,
ventricular therapy is not always initiated when no longer inhibited by AFib Rate
Threshold/Stability. Therapy may continue to be inhibited by other programmed
detection enhancements, such as Onset (when the Onset/Stability detection
enhancement suite is enabled) or Vector Timing and Correlation (when the
Rhythm ID detection enhancement suite is enabled).
Consider the following information during these interactions:
•
The AFib Rate Threshold and V Rate > A Rate detection enhancements
are not evaluated if Atrial Tachyarrhythmia Discrimination is programmed to
Off in a Rhythm ID configuration.
•
Because the AFib Rate Threshold is not evaluated during redetection
(following ventricular ATP therapy delivery, any aborted ventricular therapy,
or therapy not available), the Episode Detail report will not display data
for the enhancement during redetection, even though the parameter is
programmed On.
•
The AFib Rate Threshold enhancement is not evaluated for arrhythmia
detection in the following cases; however, the Episode Detail report will still
display the data for the AFib Rate Threshold enhancement based on a
threshold of 170 bpm:
•
Table 3-10.
–
The AFib Rate Threshold is programmed to Off
–
Ventricular Zones is programmed to 1
–
No detection enhancement suite is enabled
An atrial sense event will only be classified as AF while the AFib Rate
Threshold is being evaluated for arrhythmia detection.
AFib Rate Threshold and Stability combinations and resulting therapy
Detected Ventricular Rhythma
Therapy Decisionb
Unstable, A > AFib Rate Threshold
Inhibit
Stable, A > AFib Rate Threshold
Treat
- DRAFT -
3-32
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
Table 3-10.
AFib Rate Threshold and Stability combinations and resulting therapy (continued)
Detected Ventricular Rhythma
Therapy Decisionb
Unstable, A < AFib Rate Threshold
Treat
Stable, A < AFib Rate Threshold
Treat
a. If the detected ventricular rhythm changes, then the appropriate, corresponding row in the table is evaluated.
b. Decisions to inhibit can be overridden by V > A or expiration of SRD.
NOTE: Refer to "Use of Atrial Information" on page 3-5 for additional
information about device performance when the atrial lead is programmed to
Off.
Stability Analysis
Stability analysis distinguishes unstable (irregular) ventricular rhythms from
stable (regular) ventricular rhythms. This is accomplished by measuring the
degree of variability of the tachycardia R–R intervals.
This degree of variability, when used alone, may allow the device to distinguish
conducted AF (which may produce greater R–R variability) from monomorphic
VT (which is typically stable). It also may be used to differentiate MVTs (which
are pace terminable) from polymorphic VTs and VF (which are typically not
pace terminable).
Based on the patient’s needs, you may choose to program Stability as an
inhibitor to prevent therapy for AF, or use stability analysis to direct the type of
therapy to be delivered (Shock if Unstable).
The stability algorithm calculates RV R–R interval differences. These
differences are calculated throughout Duration; an average difference is also
calculated. When Duration expires, rhythm stability is evaluated by comparing
the current average difference to the programmed Stability threshold and/or
Shock If Unstable thresholds. If the average difference is greater than the
programmed thresholds, the rhythm is declared unstable. Independent
thresholds are available for the Stability (to inhibit) or Shock If Unstable
functions; you cannot program both in the same ventricular zone.
The pulse generator performs stability calculations for all episodes (even when
Stability is programmed to Off) and stores the results in therapy history. This
stored data may be used to select an appropriate stability threshold.
- DRAFT -
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
3-33
Stability to Inhibit
The Stability parameter may help you identify rapid rhythms originating in the
atrium, such as AF. These rhythms may result in unstable ventricular rhythms
whose rate exceeds the lowest rate threshold and should not be treated. If a
rhythm is declared stable when Duration expires, programmed therapy will
be delivered. If the rhythm is declared unstable, ventricular therapy will be
inhibited.
At the end of initial Duration, if a tachycardia is declared unstable and
ventricular therapy is inhibited, the pulse generator continues to evaluate for
stability on each new detected interval (Figure 3-19 on page 3-33). Therapy will
not be inhibited by Stability if:
•
V Rate > A Rate declares the ventricular rate greater than the atrial rate
•
The SRD has expired (if programmed to On)
Ventricular therapy is not always initiated when no longer inhibited by
Stability. Therapy may continue to be inhibited by other programmed
detection enhancements, such as Onset (when the Onset/Stability detection
enhancement suite is enabled) or Vector Timing and Correlation (when the
Rhythm ID detection enhancement suite is enabled).
NOTE: Ventricular Therapy can also be inhibited through analysis of the
Stability algorithm as it is used with the AFib Rate Threshold enhancement.
Unstable intervals
Stability analysis continues
Detection window satisfied.
Start Duration.
Stability analysis begins.
Figure 3-19.
Duration expires.
Rhythm declared unstable.
Start SRD.
Therapy inhibited until rhythm stabilizes, V is greater
than A, or SRD times out.
Stability evaluation when Duration expires
- DRAFT -
3-34
TACHYARRHYTHMIA DETECTION
VENTRICULAR DETECTION
Shock if Unstable
When programmed to Shock if Unstable, the stability analysis helps determine
if ventricular ATP therapy should be bypassed in preference for the first
programmed ventricular shock therapy (which may be low- or high-energy) for
the ventricular zone (Figure 3-20 on page 3-34).
Dynamic ventricular arrhythmias such as polymorphic VT or VF may be sensed
at a rate lower than the highest ventricular rate threshold and can be classified
as unstable. Since the sensed rhythm may be detected in a lower ventricular
zone in which ATP may be programmed, the stability analysis may be used
to skip over the programmed ventricular ATP therapies and instead provide
shocks to the patient. Stability is evaluated on each detection/redetection cycle,
including evaluation between bursts of an ATP scheme. Once a ventricular
shock has been delivered in an episode, the Shock If Unstable function no
longer affects therapy selection.
Shock If Unstable may be used only in the VT zone of a 2- or 3-zone
configuration. You cannot program it in a 2-zone configuration if Stability or
Onset is already programmed to On, or if Post V-Shock Stability or AFib Rate
Threshold is programmed to On.
Shock
ATP Burst
Charging
Detection window satisfied.
Start Duration.
Stability analysis begins.
Figure 3-20.
Duration expires.
Rhythm declared stable.
Deliver programmed ATP
therapy.
Start
Redetection.
Detection window
satisfied.
Start Redetection
Duration.
Reanalyze Stability.
Redetection Duration expires.
Rhythm declared unstable.
Skip remaining ATP and
proceed to the first
programmed shock therapy.
Shock if Unstable
Onset
Onset differentiates physiologic sinus tachycardias, which typically begin
slowly, from pathologic tachycardias, which typically begin abruptly. It measures
the rate of transition in the ventricular rhythm from slow rates to tachycardia. If
the rate increase is gradual, it enables the device to inhibit ventricular therapy
in the lowest tachycardia rate zone.
- DRAFT -

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : No
Modify Date                     : 2008:01:04 15:49:40-06:00
Create Date                     : 2008:01:04 15:49:03-06:00
Title                           : Print Preview - C:\WINNT\TEMP\e3temp_5648\.aptcache\aea05648/tfa05648
Author                          : SYSTEM
Creator                         : PScript5.dll Version 5.2.2
Producer                        : Acrobat Distiller 6.0.1 (Windows)
Page Count                      : 100
Mod Date                        : 2008:01:04 15:49:40-06:00
Creation Date                   : 2008:01:04 15:49:03-06:00
Metadata Date                   : 2008:01:04 15:49:40-06:00
Has XFA                         : No
EXIF Metadata provided by EXIF.tools
FCC ID Filing: ESCCRMN11906

Navigation menu