Q-TECH™ PROGRAMMERA COMPONENT OF THE S-ICD®SYSTEMUSER’S MANUALModel 2020102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 1
Copyright©2008 Cameron Health, Inc., San Clemente, CA USAAll rights reserved.Limited Software License and Equipment Use.S-ICD®is a registered trademark of Cameron Health, Inc. SQ-RX™, Q-TRAK™, Q-GUIDE™ and Q-TECH™ are all trademarks ofCameron Health, Inc. Manuals or other written documentation may not be copied or distributed without Cameron Health, Inc.authorization. Cameron Health, Inc.905 Calle AmanecerSuite 300San Clemente, CA  92673USATel:  1 949 498 5630Free:  1 877 SICD 4111 877 742 3411Fax:  1 949 498 5932URL:  www.cameronhealth.comCameron Health BVWorld  Trade CenterNieuwe Stationsstraat 106811 KS ArnhemThe NetherlandsTel:  31 26 3550260Free:  800 SICD 4 YOU800 7423 4 968Fax:  31 26 3550269URL:  www.cameronhealth.com102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 2
Table of ContentsGeneral Description 6Description 6Indications for Use 6Contraindications 6Programmer Warnings and Cautions 6General 6Batteries  7Electrostatic Discharge (ESD) 7S-ICD System Warnings and Cautions 7General 8Use of Other Medical Therapies/Diagnostic Procedures 8Electromagnetic Interference (EMI) Outside the Hospital Environment 8Potential Adverse Events 9Operation 10Setup of the Q-TECH Programmer 10Packaging 10Inserting the Main Battery Pack 10Inserting the Backup Batteries 11Charging the Programmer 12Using the Q-TECH Programmer 12Turning the Programmer On 12Placing the Programmer in Suspend Mode 13Using the Programmer Touch Screen 13Stylus  13Keyboard 14Using the Wand 14Navigation 14Screen Header 15Navigation Bar 15Reset the Programmer 17Configuring the Q-TECH Programmer 17Configuring Programmer Settings 17Date and Time Format 18Time Zone  18Language Preference 19Printer Selection 19Programmer Software Version 213102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 3
Modes of Operation for the Q-Tech Programmer 22Online Behavior 22Offline Behavior 22Stored Patient Sessions  22Modes of Operation for the SQ-RX Device 22Shelf Mode 23Therapy On Mode 23Therapy Off Mode 23Connecting and Disconnecting from the SQ-RX Device 23Scanning For Devices 23Connecting to the SQ-RX Device  24Ending a Patient Session 25Programming the SQ-RX Device at Implant 26Entering Electrode Information 26Create the Patient Chart 28Automatic Setup  29Programming Therapy Parameters 32Defibrillation Testing  34Performing an SQ-RX Follow-up 36Viewing Device Status 36Viewing Stored Episodes 37Printing Reports from the Programmer 38Printing Reports 38Summary Report 38Captured S-ECG Report 39Episodes Report 39S-ECG Features 40S-ECG Rhythm Strip Markers 40S-ECG Scale Settings 41Capture and View S-ECG Strips 41Utilities Menu 42Acquire Reference S-ECG 43Beeper Control  44Manual Setup 44Smart Charge 46Additional Programmer Functions 47Rescue Shock 47Magnet Use 484Table of Contents (Continued)102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 4
Programmer Maintenance 49Cleaning and Maintenance 49Programmer End of Life 49Possible Problems and Solutions 50Inability to Print 50No Printer Available 50Loss of Communication with Printer 50Inability to Communicate with the SQ-RX Device 50Compliance Statements 51FCC 51EMI/RFI 51Declaration Tables 52Declaration Electromagnetic Emissions 52Declaration Electromagnetic Immunity Part 1 52Declaration Electromagnetic Immunity Part 2 53Recommended Separation Distances 54Specifications 55Product Guidelines 55Specifications 55Definitions of Package Label Symbols 56Warranty 57Cameron Health Q-TECH Programmer Warranty 57Service 57Disclaimer 575Table of Contents (Continued)102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 5
DescriptionThe Q-TECH Programmer (the ”programmer”), Model 2020, is a component of the Cameron Health S-ICD System, which is prescribed for patients when cardiac arrhythmia management is warranted. The other components of the S-ICD System include theSQ-RX Pulse Generator (the “SQ-RX device”), Q-TRAK Subcutaneous Electrode (the “electrode”) and the Q-GUIDE SubcutaneousElectrode Insertion Tool (the “EIT”). The programmer communicates wirelessly with the SQ-RX device to enable adjustment of programmable settings and data collection. The S-ICD System is designed for ease of use and simplicity of patient management. TheS-ICD System has a number of automatic functions designed to reduce the amount of time required for implantation, initial programming and patient follow-up.   Indications for UseThe S-ICD System is intended to provide defibrillation therapy for the treatment of life-threatening ventricular tachyarrhythmias.ContraindicationsThe S-ICD System is contraindicated for patients with symptomatic bradycardia, incessant ventricular tachycardia and patients withdocumented spontaneous, frequently recurring ventricular tachycardia that is reliably terminated with anti-tachycardia pacing. Unipolar pacemakers are contraindicated for use with the S-ICD System.Programmer Warnings and CautionsGeneral• External defibrillation equipment should be available for immediate use during the implantation procedure and follow-up. • To avoid risk of electric shock, the programmer must only be connected to a supply mains with protective earth.• Mains isolation is achieved by disconnecting the AC or DC connector of the power supply. Do not position the programmerso that it is difficult to disconnect the power connections.• The US power cord is for connection to 110 VAC supply mains. All other power cords are for connection to 230 VAC supplymains.• Confirm that the programmer is in communication with the intended implanted SQ-RX Pulse Generator.• The programmer is intended for use by healthcare professionals only.  The programmer may cause radio interference ormay disrupt the operation of nearby equipment.  It may be necessary to take mitigation measures., such as reorienting orrelocating the programmer or shielding the location.• By design, the programmer emits radio frequencies (403.5 MHz and 2.5 GHz). This may interfere with nearby equipment.Closely monitor medical electrical equipment in the vicinity when using the programmer.• The presence of other devices in the same frequency band (403.5 MHz for the SQ-RX Pulse Generator and 2.5 GHz for theprinter) may interfere with communication. If communication problems occur, refer to the Possible Problems andSolutions section of this manual. • If the programmer is used near other medical or office equipment, all equipment should be observed to verify normaloperation. Other equipment may interfere with the programmer, even if that other equipment complies with theInternational Special Committee on Radio Interference (CISPR) emission requirements.• Use of accessories other than those specified by Cameron Health may result in increased emissions or decreased functionality of the programmer and may cause unintended operational behavior of the programmer.• Do not attempt to remove the wireless communication card or wireless printer card. Use of non-approved Cameron Health components will void the warranty and may cause unpredictable SQ-RX Pulse Generator operation. • No modification of the programmer is allowed.• Use only the Cameron Health programmer and appropriate software for communicating with, and programming theSQ-RX Pulse Generator.• Contact your local Cameron Health representative or Customer Service department for assistance or a replacement part.6Q-TECH PROGRAMMER GENERAL DESCRIPTION102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 6
BatteriesThe following warnings pertain to safe handling of the batteries and power supply used with the programmer. To prevent the batteryfrom leaking, generating heat or exploding, observe the following precautions:• Use the battery only with an approved Cameron Health power adapter.• Do not discard the battery in a fire, incinerate, or subject battery cells to temperatures that exceed 100° C/212° F. Thismay result in harmful emissions or explosion.• The battery incorporates built-in safety devices. To ensure their proper function, do not disassemble or alter any parts ofthe battery.• As a fire precaution, do not use rechargeable AAA batteries. Insert only alkaline AAA batteries into the programmer.• Do not short-circuit the battery by directly connecting any of the exposed terminals with metal objects. Do not transportor store the battery together with metal objects, such as necklaces, hairpins, etc.• Do not use or leave the battery near a heat source, such as a fire or heater.• Do not immerse the battery in water.• Do not crush, puncture, open, dismantle, or otherwise mechanically interfere with the batteries.• Do not directly solder the battery.• Do not put the battery into a microwave or pressurized container.• Do not use the battery in combination with primary batteries, such as dry-cell batteries, or batteries of different capacities or brands.• Immediately remove the battery from the programmer and stop use if the battery gives off an odor, generates heat,becomes discolored or deformed, or in any way appears abnormal during use.• Do not continue charging the battery if it does not recharge within the specified charge time.• If the battery should leak, do not allow the contents to come into contact with the skin or eyes. If they do, wash immediately with cold water and seek medical advice.• Do not store the battery in any location having temperatures outside the storage range (-10° C to 55° C/14° F to 131° F).Exposure to high temperatures may cause the battery to overheat or ignite, and may possibly reduce the performanceand service life of the battery.• Batteries may be considered hazardous waste and are to be disposed of in accordance with the manufacturer’s instructions, deposited in the appropriate local battery recycling facility, or returned to Cameron Health for proper disposal.• Keep batteries out of the reach of children.Note: If the programmer is not likely to be used for an extended period, remove the main battery. A reset of the Time Zone is required when the programmer returns to active communicationElectrostatic Discharge (ESD)The programmer may be affected by ESD. If ESD occurs and the programmer’s functionality is affected, attempt to reset the programmer or contact your local Cameron Health representative or Customer Service Department for instructions.• Do not touch or connect the Wand to the Programmer unless ESD precautionary procedures are used.S-ICD System Warnings and Cautions Before using the S-ICD System, read and follow all warnings and cautions provided in this manual. Refer to the SQ-RX PulseGenerator, Q-TRAK Electrode or Q-GUIDE Electrode Insertion Tool User’s Manuals for information on implanting and explanting theS-ICD System.The S-ICD System contains sterile products for single use only. Do not resterilize. Handle the S-ICD System with care at all timesand maintain proper sterile technique.7Q-TECH PROGRAMMER GENERAL DESCRIPTION102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 7
General• Placing a magnet over the SQ-RX Pulse Generator suspends arrhythmia detection and therapy response. Removing themagnet resumes arrhythmia detection and therapy response. • Battery depletion will eventually cause the SQ-RX Pulse Generator to stop functioning. Defibrillation and excessive numbers of charging cycles shorten the SQ-RX Pulse Generator’s battery longevity. Use of Other Medical Therapies/Diagnostic Procedures• External defibrillation or cardioversion may damage the implanted device. Current flow through the SQ-RX PulseGenerator may be minimized by avoiding the placement of defibrillation paddles directly over the device.• Do not expose a patient with an implanted S-ICD System to diathermy. The interaction of diathermy therapy with animplanted SQ-RX Pulse Generator can damage the SQ-RX Pulse Generator and cause patient injury.• Do not expose the patient to MRI scanning. MRI scanning can damage the SQ-RX Pulse Generator and cause patientinjury.• Electrical interference or “noise” from sources such as electrosurgical and monitoring equipment can interfere with thecommunication between the programmer and SQ-RX Pulse Generator or cause inappropriate therapy. If interferenceoccurs, move and reposition the programmer away from the source of the interference.• Ionizing radiation therapy, such as radioactive cobalt, linear accelerators, and betatrons may adversely affect the S-ICDSystem operation. Therapeutic ionizing radiation cannot be immediately detected; however, it can damage the electroniccomponents of the SQ-RX Pulse Generator. Follow these conditions to minimize the risks of ionizing radiation:§Shield the SQ-RX Pulse Generator with a radiation-resistant material, regardless of the distance between the SQ-RXPulse Generator and the radiation beam.§Do not project the radiation port directly at the SQ-RX Pulse Generator.§Always evaluate the S-ICD System operation following each radiation treatment.• Lithotripsy and other therapeutic forms of ultrasound can damage the SQ-RX Pulse Generator. If required, avoid directflow of the pulse waves near the site of the implanted SQ-RX Pulse Generator.• Use caution during ablation procedures. Program the S-ICD System to Therapy Off. Keep the current path (electrode tipto ground) as far away as possible from the implanted SQ-RX Pulse Generator and electrode.• The programmer is not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen ornitrous oxide.Electromagnetic Interference (EMI) Outside the Hospital EnvironmentExposure to EMI or static magnetic fields may suspend tachycardia detection and possibly cause temporary inhibition of therapydelivery. EMI may also trigger delivery of a shock in the absence of a tachycardia. Automatic sensing and detection of tachycardiaswill resume when the patient moves away from the EMI or static magnetic field source.• To minimize the risk, advise patients to avoid sources of EMI or static magnetic fields having strengths of >10 gauss or1 mTesla. Sources of EMI include, but are not limited to:§High-voltage power lines§Arc welding equipment§Electrical smelting furnaces§Large radio-frequency transmitters (such as radar)§Alternators on running engines in automobiles§Communications equipment (such as high-power radio transmitters)8Q-TECH PROGRAMMER GENERAL DESCRIPTION102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 8
Sources of strong static magnetic fields may include the following:§Industrial transformers and motors§Large stereo speakers§Magnetic wands, such as those used for airport securityPatients should seek medical guidance from their physician before entering an area where a posted sign prohibits patients with animplantable cardioverter defibrillator or pacemaker.Potential Adverse EventsPotential adverse events related to implantation of the S-ICD System may include, but are not limited to, the following:• Acceleration of arrhythmia• Allergic reaction• Bleeding• Conductor fracture• Cyst formation• Death • Electrode dislodgement• Electrode insulation failure• Electrode deformation and/or breakage• Erosion/extrusion• Improper electrode connection to the device • Inappropriate shock delivery• Infection• Hematoma• Hemothorax• Keloid formation • Migration or dislodgement• Potential morbidity due to inability to defibrillate and pace• Muscle stimulation• Nerve damage• Pneumothorax• Postoperative discomfort• Premature battery depletion• Random component failures• Tissue necrosis• Ventricular arrhythmiaIf any adverse events occur, invasive corrective action and/or S-ICD System modification or removal may be required.Patients who receive an S-ICD System may also develop psychological disorders that include, but are not limited to, the following:• Depression• Fear of shocks• Phantom shocks9Q-TECH PROGRAMMER GENERAL DESCRIPTION102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 9
Setup of the Q-TECH ProgrammerPackagingProgrammer components include:• Model 2020 Programmer with pre-loaded software• Touch screen stylus• Lithium ion rechargeable battery• Non-rechargeable alkaline AAA backup batteries• Wireless printer card• Cameron Health Programmer wireless communication card with model 4510 telemetry wand• AC adapter power cord and power supply for battery chargingVisually inspect the packaging to ensure the contents are complete. Do not use if there is:• Evidence of damage or cracking• Evidence of battery leakage or corrosionReturn the product to Cameron Health if any of these conditions exist. Contact your local Cameron Health representative orCustomer Service Department for instructions and return packaging.Inserting the Main Battery PackThe programmer is powered by a rechargeable lithium ion battery pack, or through AC power with the AC adapter power cord.Note: If connected to AC power, make sure the programmer is in Suspend Mode before insertion of the battery pack. SeePlacing the Programmer in Suspend Mode section.The main battery is located on the back of the programmer base. To insert the main battery pack:1. Slide the battery so the outside latch is toward the rear of the programmer (Figure 1).10Q-TECH PROGRAMMER OPERATIONMain BatteryFigure 1: Main battery cover102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 10
2.  Insert the battery pack into place (Figure 2).3.  Secure the battery pack with the slide cover by latching it in place.Inserting the Backup BatteriesTwo internal AAA batteries serve as a backup power source for situations when the programmer is not connected to an externalpower source, the main battery pack is removed, or the main battery pack charge supply is low. The user can remove and replacethe two AAA alkaline batteries as needed.To install or replace the two internal batteries:1.  Place the programmer in Suspend Mode by holding down the FN and ESC On/Off keys simultaneously.2.  Open the battery compartment (Figure 3). The backup battery cover is located on the top left of the keyboard. Angle the programmer screen forward to provide access to the cover. Apply downward pressure to the center of the battery cover and slide it toward the front of the programmer.11Q-TECH PROGRAMMER OPERATIONFigure 2: Inserting battery packFigure 3: Inserting backup batteries102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 11
3. Remove the depleted AAA alkaline batteries.4. Insert two AAA batteries, observing the polarity indication.5. Close the battery compartment by sliding the cover over the top of the batteries, while applying downward pressure to the centerof the cover.Charging the ProgrammerThe programmer is primarily intended to be AC-powered, but can also be operated on battery power provided that the internal batteries are adequately charged.The main battery pack is recharged automatically whenever the programmer is powered using the AC adapter. A battery pack isfully charged within 4 hours. The Battery Status indicator located on the upper right corner of the screen displays the status of themain battery power when the unit is in use:• All four bars are illuminated (green) - The battery is 100% charged.• Three bars are illuminated (green) - The battery is 75% charged.• Two bars are illuminated (yellow) - The battery is 50% charged.• One bar is illuminated (flashing red) - The battery is 25% charged.The programmer displays one of the following alert screens as battery power gets progressively lower.   • Programmer Battery Low• Programmer Battery Critical• Out Of PowerTo charge the programmer:1. Ensure that the main battery pack is inserted into the programmer.2. Connect the AC adapter power cord to the programmer.3. Insert the power cord into an AC power outlet.    Using the Q-TECH ProgrammerTurning the Programmer OnThe programmer can usually be started by simply opening the programmer lid. If the programmer does not start, press the ESCOn/Off key in the upper left corner of the keyboard (Figure 4).12ESC On/OffFigure 4: Turning the programmer On/Off102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 12
Placing the Programmer in Suspend ModeThe programmer has a Suspend Mode, which is activated automatically to conserve power. When this mode is in effect, the displaywill be blank.The programmer enters Suspend Mode whenever:• The programmer is inactive for >15 minutes • The Power Off keys (FN and ESC On/Off) are pressed• The touch screen display is closed • The Power Off button is selected from the Main Menu If the programmer enters Suspend Mode during active communication with the SQ-RX device, the telemetry link will be disconnected. However, the programmer will maintain the current display screen and writes the ongoing session to memory. Pressing the ESC On/Off button or opening the touch screen from the folded down position ends Suspend Mode operation.Using the Programmer Touch ScreenThe programmer is equipped with an LCD touch screen. By using the stylus and keyboard, the user can select onscreen options and enter patient information. Stylus The stylus is used to enter commands, change parameters, and navigate through the programmer screens.  Note: Do not substitute any type of ink pen for the stylus; this may cause a permanent stain and damage to the displayscreen.To use the stylus:1.  Push the square stylus eject button on the right side of the programmer base. The stylus is ejected from the holder (Figure 5).2.  Use the stylus to lightly tap the appropriate functions on the touch screen, or press and hold the stylus to move a scroll bar.13Figure 5: Removing the stylus102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 13
3.  Push the stylus back into its holder when finished.KeyboardThe compact computer keyboard, with 58 full-size keys, is used to enter text and numeric data. Using The WandUse only the Cameron Health Model 4510 wand (the “wand”) to communicate with the SQ-RX device.   Note: The wand is a non-sterile device. DO NOT STERILIZE. The wand must be contained in a sterile barrier to be usedin the sterile field.To connect the wand to the programmer, insert the connector cable in the communication card located on the side of the programmer as shown in Figure 6.To obtain optimal telemetry, the wand should be placed directly over the implanted SQ-RX device. Although the programmer mayreceive S-ECG telemetry at greater distances, programming should be performed with the wand placed over the SQ-RX device.When telemetry loss occurs, the display screen will turn yellow and a message will appear with the text “Communication Loss” toalert the user. Reposition the wand to establish communication. The programmer will display a screen informing the user if thedevice was found and programming can continue.Note: If communication cannot be reestablished, the session should be ended and restarted by scanning for the SQ-RXdevice.NavigationThe programmer’s graphic user interface (GUI) facilitates management and control of the S-ICD System. The Navigation Bar andoval on-screen buttons allow the user to navigate programming software screens. In addition, a continuous subcutaneous electrocardiogram (S-ECG) is displayed along the bottom of the screen during Online (active) communication with the SQ-RXdevice. 14Figure 6: Connecting the wand102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 14
Screen HeaderWhen the programmer is Offline (inactive communication), the screen header displays the Battery Status Indicator.When viewing Offline Stored Sessions, the screen header displays:Patient nameTherapy On/OffBattery Status IndicatorWhen the programmer is online (active communication), the screen header displays:Therapy On/OffPatient namePatient heart rateProgrammer Battery and Telemetry status indicatorScreen titleRescue shock iconNavigation BarThe Navigation Bar is the primary method for navigating the Online programmer screens. The bar is located along the left side ofthe programmer screen and the corresponding screen appears with the selected icon highlighted. Table 1 displays a list of the programmer icons and their corresponding descriptions.15102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 15
16Q-TECH PROGRAMMER OPERATIONTable 1: Icon DescriptionsIcon Description User ApplicationMain Menu Button Allows user to return to the main menuAuto Setup Button Allows user to access the auto setup menuDevice Settings  ButtonAllows user to access the SQ-RX device settings screenDevice Status Button(open folder andclosed folder)Allows user to access the SQ-RX device status screen. User can view number    of shocks delivered since the last update as well as the SQ-RX device batterylife.Patient Chart Button Allows user to access the patient chart screenCaptured EpisodesS-ECG ButtonAllows user to access captured S-ECG and stored episode screensInduction Test Button Allows user to access induction screenManual Shock Button(open hand and closed hand)Allows user to access the manual shock screenBattery & Telemetry  MeterLeft side of the meter allows user to view the programmer’s battery status.  The right side of the meter allows viewing of telemetry signal.Capture S-ECG  Allows user to capture a live S-ECGS-ECG Display SettingsAllows user to modify the zoom and sweep speed on the live S-ECGHeart Rate Icon Allows user to view current heart rateRescue Button Allows user to administer a shock from any programmer screenToggle Switch Allows user to turn functions on or off102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 16
17Q-TECH PROGRAMMER OPERATIONReset the ProgrammerPeriodically the programmer may need to be reset if:• You cannot exit a screen• The operating system stops responding• You are unable to turn the programmer On using the ESC On/Off keys• You see a red alert screen instructing you to press CTL-ALT-DELThe programmer can be reset by pressing and holding CTL-ALT-DEL for 5 seconds. If the programmer does not respond to the resetprocess, contact your local Cameron Health representative or Customer Service Department for assistance.Configuring the Q-TECH ProgrammerConfiguring Programmer SettingsThe programmer should be configured before communication with an SQ-RX device is attempted. This includes setting the date andtime format, time zone, language and printer. Once these settings are configured during the initial setup process, they will becomethe default parameters and will not normally need to be changed with each session.  Note: If power loss occurs and the battery is depleted, these settings will need to be reconfigured.To configure the programmer settings:1.   Select the Programmer Settings button on the Startup screen (Figure 7), to display the Programmer Settings screen (Figure 8).2.  Select the corresponding line to access each setting. The settings that can be configured include:Date and time formatTime zoneLanguagePrinterFigure 7: Startup screen Figure 8: . Programmer Settings screen102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 17
Date and Time FormatTo set the date and time format:1.  Select Set Date and Time Format on the Programmer Settings screen (Figure 9). The Date and Time Settings screen appears(Figure 10).   2.  Select the desired date format.  3.  Select the Automatically Adjust for Daylight Savings button to initiate the daylight savings option. A green checkmark willappear when this feature is active.4.  Select the Save button to save the changes and return to the Programmer Settings screen, or select Cancel to return to theProgrammer Settings screen without saving the changes.Time Zone To set the time zone in which the programmer will be used:1.  Select Set Time Zone on the Programmer Settings screen. The Set Time Zone screen appears.2.  Select the country/time zone button for the zone in which the programmer will be used. A green checkmark will appear in theselected button (Figure 11). The SQ-RX device line frequency of 50 Hz or 60 Hz is automatically programmed based on the country/time zone selected.18Q-TECH PROGRAMMER OPERATIONFigure 9: Programmer Settings screen Figure 10: Date and Time Settings screen102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 18
3.  Select the Save button to save the changes and return to the Programmer Settings screen, or select Cancel to return to theProgrammer Settings screen without saving the changes.Language PreferenceTo set the language preference:1.  Select Set Language on the Programmer Settings screen. The Language Settings screen appears.2.  Select the Save button to save the changes and return to the Programmer Settings screen, or select Cancel to return to theProgrammer Settings screen without saving the changes.Printer SelectionTo select the printer to be used with the programmer:  1.  Insert the wireless communication card into the printer’s USB port.2.  Select Printer Setup on the Programmer Settings screen. A previously configured printer becomes the default printer and will bedisplayed at this time. If a default printer has not already been selected and configured, the programmer will scan the area tolocate wireless printers. A window will appear informing the user that the programmer is currently scanning for printers. 3.  Select the printer of choice (Figure 12). If no printers are found, a window will appear stating that there are no printers. Selectthe Scan Again button or the Cancel button to stop the scanning process. If a printer was selected during another session, thePrinter Setup screen is displayed.19Q-TECH PROGRAMMER OPERATIONFigure 11: Time Zone selection screen102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 19
4.  Once the scan is complete, the Printer Setup screen appears, which displays a list of Cameron Health-approved printers. Up to6 printers can be configured. Select the button next to the desired printer. The Printer Setup screen appears (Figure 13). Note: Refer to Possible Problems and Solutions for printer problems.20Q-TECH PROGRAMMER OPERATIONFigure 12: Printer Selection screenFigure 13: Printer Setup screen102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 20
215.  Select the desired printer name and enter the name using the keyboard (up to 15 characters) (Figure 14). A unique printer identi-fier should appear with the printer selection.6.  Select the Save button to save the changes and return to the Programmer Settings screen, or select Cancel to return to theProgrammer Settings screen without saving the changes.Programmer Software VersionTo view the programmer’s software version:1.  Select About Programmer on the Programmer Settings screen. The Programmer Software Version Info screen appears (Figure 15).  Q-TECH PROGRAMMER OPERATIONFigure 14: Use the keyboard to enter a name for the selected printer.Figure 15: Programmer Software Version Info screen102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 21
2.  This screen displays the current version of the programmer software. Select the Back button to return to the ProgrammerSettings screen.Note: The patient printed report contains the programmer software version.Modes of Operation for the Q-TECH Programmer Online BehaviorThe programmer’s interface varies according to whether the programmer is Online (actively communicating) or Offline (not communicating) with a selected SQ-RX device.An Online session occurs when the programmer establishes a telemetry link with a specific SQ-RX device. Online and Offline display screens have a grey background appearance. During active communication, if the telemetry signal is lost between the programmer and the SQ-RX device for more than five seconds, a yellow alert screen is displayed notifying the user. This may occur if the wand is moved out of the telemetry communications range, or if noise or interfering objects inhibit communication.Programming commands, including Rescue Shocks, will not be available until telemetry is reestablished.Telemetry reconnection may occur automatically within one minute when the SQ-RX device and wand are within telemetry range.Upon disconnection caused by the loss of telemetry, the last 60 seconds of S-ECG can be viewed through the Captured S-ECG programmer display screen. Offline BehaviorWhen the programmer is not actively communicating with an SQ-RX device, it is Offline. Stored patient sessions can be viewed andprinted and programmer settings can be accessed during Offline sessions. Offline screens have a grey background.Stored Patient Sessions During a patient follow-up visit, the programmer will retrieve data from the SQ-RX device’s memory. The programmer can store upto ten sessions per patient. When the eleventh session occurs, the programmer will automatically replace the oldest stored sessionwith the new data. A stored session includes the following information:•  Captured ECG Reports•  Episode History (including any downloaded episodes)•  Patient Data•  Programmed Device SettingsTo view stored patient sessions:1.  From the Startup screen, select Stored Patient Sessions.2.  Select the desired patient session.Modes of Operation for the SQ-RX DeviceThe SQ-RX device has three modes of operation:•  Shelf •  Therapy On•  Therapy Off22Q-TECH PROGRAMMER OPERATION102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 22
Shelf ModeThe Shelf mode is a low power consumption state intended for storage only. When communication is initiated between the SQ-RXdevice and the programmer, a full-energy capacitor reformation is automatically performed and the SQ-RX device is prepared forset-up. Once the SQ-RX device is taken out of Shelf mode, it cannot be reprogrammed back into Shelf mode.Therapy On ModeThe Therapy On mode is the primary operating mode of the SQ-RX device, allowing automatic detection of, and response to, ventricular tachyarrhythmias.  Note: The SQ-RX device must be programmed out of Shelf mode before being programmed to Therapy On.Therapy Off ModeThe Therapy Off mode disables automatic therapy delivery and enables manual control of shock delivery. Programmable parametersmay be viewed and adjusted via the programmer. Also, the subcutaneous electrogram (S-ECG) may be displayed or printed.  The SQ-RX device automatically defaults to Therapy Off mode when the SQ-RX device is taken out of Shelf mode.Note: Manual and rescue shock therapy are available once the initial Setup process is complete. Connecting and Disconnecting from the SQ-RX DeviceThis section provides the information necessary for selecting, connecting to, and disconnecting from the SQ-RX device.Scanning For Devices1.  Select the Scan For Devices button on the Startup Screen (Figure 16). The Device List screen appears with a Scan Progress Bardisplayed during the scanning process. Select the Cancel button at any time to end the scanning process.23Q-TECH PROGRAMMER OPERATIONFigure 16: The Scan for Devices button is on the Startup screen.102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 23
2.  When the scanning process is complete, a list of all Cameron Health SQ-RX devices detected (up to 15) will be displayed on theDevice List (Figure 17). The SQ-RX devices that are in shelf mode will be displayed as “Not Implanted.” Any SQ-RX devices thatwere previously taken out of shelf mode are displayed either as “Implanted” or with the stored patient name. Note: The Device List screen can also contain the text “Boot Loader Device” indicating that an alert has occurred. Do notproceed if this occurs. Contact your local Cameron Health representative or Customer Service Department for instruc-tions.   3. If the desired SQ-RX device is not listed, select the Scan Again button to re-initiate the scanning process. Select the Cancel but-ton to return to the Main Menu screen.Note: Refer to the Inability to Communicate with the SQ-RX Device section for further assistance.Connecting to the SQ-RX Device 1.  From the Device List screen, select the desired SQ-RX device to initiate the communication session.Note: Regardless of how many devices are located by a scan, the user must select an SQ-RX device from the list to beginactive communication.Connecting to an SQ-RX Device in Shelf Mode:1. When the SQ-RX device selection is made, the programmer connects to the selected SQ-RX device. A window will appear indicating connection is in process. The Cancel button can be selected at any time to end the connection and return to theDevice List screen.2. Once communication is established with the SQ-RX device, the Device Identification screen appears (Figure 18). 24Q-TECH PROGRAMMER OPERATIONFigure 17: Device List screen102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 24
25Q-TECH PROGRAMMER OPERATIONNote: The Device Identification screen is visible only while connecting to an SQ-RX device in Shelf Mode.3.  The SQ-RX device model and serial numbers are automatically acquired and displayed during the initial scanning process.Select Continue to remove the SQ-RX device from Shelf Mode and prepare for implantation, or select Cancel to return to theDevice List screen.  Ending a Patient SessionTo end an online patient session and return the programmer to its offline operation mode:1. Select the Main Menu button on the Navigation Bar. The Main Menu screen appears.2. Select the End Session button (Figure 19).Figure 18: Device Identification screenFigure 19: Select the End Session button to end a patient session.102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 25
26Q-TECH PROGRAMMER OPERATION3.  A warning message will be displayed to the user if any of the following conditions exists:•  Automatic Therapy is programmed to Off•  Reference S-ECG has not be acquired•  Automatic Setup Optimization has not been completed. This message is typically displayed following the implant procedure as Setup Optimization was not performed (Figure 20).4.  Select the Continue button to end the patient session and return to the Startup Menu screen, or select Cancel to remain onlineand return to the Main Menu screen Note: Once the Continue button is selected, the session is stored in the SQ-RX device and communication is terminated.  Note: In order to confirm that Automatic Therapy is programmed On upon disconnection, always use the End Sessionprocess and review all displayed warning messages.Programming the SQ-RX Device at Implant This section provides the information necessary for programming the SQ-RX device during an implant setting.Entering Electrode InformationThe programmer maintains information on the implanted electrode. To record this information for a patient’s new or replacementelectrode: 1.  Select the Main Menu button.2.  Select the Implant button.3.  Select the Automatic Setup button in the Navigation Bar.4.  Select Set Electrode ID button (Figure 21).Figure 20: Session incomplete message102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 26
27Q-TECH PROGRAMMER OPERATION5.  Enter the electrode model and serial number (Figure 22).6.  Select the Program button to save the information. A confirmation screen will appear during communication with the SQ-RXdevice (Figure 23). Select Cancel to cancel information storage and return to the Automatic Setup screen.Figure 21: Select the Set Electrode ID button to enter electrode information.Figure 22: Entering electrode information102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 27
28Q-TECH PROGRAMMER OPERATIONCreate the Patient ChartThis chart contains reference information for the patient. To set up the patient chart:1. Select the Main Menu button on the Navigation Bar.  2. Select the Implant button (Figure 24).Figure 23: Confirmation screen showing communication in progressFigure 24: Select the Implant button to create a patient chart.102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 28
3. Select the Patient View button to access the Patient Menu screen.  4. The SQ-RX device model and serial numbers appear on the first line of the chart. The electrode model and serial numbersappear on the second line of the chart. The implant date appears on the third line of the chart (Figure 25). Using the keyboard,enter the following patient information:Patient Name - up to 25 charactersPhysician Name - up to 25 charactersPhysician Information - up to 25 charactersNotes - up to 50 characters Note: The Notes field will automatically wrap the text with the presence of a space between any characters within thefirst line.    5. Select the Save button to update the SQ-RX device with the patient information. Note: Failure to save the new patient information will result in loss of the entered data.Automatic SetupBefore the SQ-RX device can be activated, it must go through an initial Automatic Setup Process at the time of the implant. The Automatic Setup Process is initiated as follows:1. Select the Main Menu button.2. Select the Implant button.3.  Select the Auto Setup button on the Navigation Bar. The Automatic Setup screen appears.  4. Select the Patient Exhibits Long QT button if the patient’s  QT interval is greater than 250 ms (Figure 26).  Select Continue if thepatient’s heart rate is less than 130 bpm.  For rates greater than 130 bpm, select the Cancel button and refer to the instructionsfor Manual Setup.29Q-TECH PROGRAMMER OPERATIONFigure 25: Entering information into the patient chart102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 29
5.  Once started the Automatic Setup will:•  Perform the shock electrode integrity check (Figure 27) to measure electrode impedance. Normal sub-threshold impedance range is < 400 Ohms.•  Select the best sensing configuration. The sense electrode configuration appears on the printed report and can beviewed via the Manual Setup process.• Select the appropriate gain selection. The selected sense gain appears on the Printed Report and can be viewed via theManual Setup process.30Q-TECH PROGRAMMER OPERATIONFigure 26: Automatic Setup screenFigure 27: Measuring electrode impedance102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 30
The progress of the Automatic Setup process is shown in the status bar. When each function is complete, the arrow next to thefunction moves to a down position.6. The Automatic Setup Sensing Optimization screen will be displayed. The programmer will display a message requesting that thepatient sit up; however, this process can be omitted during the implantation procedure by selecting the Skip button (Figure 28).7. Following the optional optimization process, the Acquire Reference S-ECG screen is displayed (Figure 29). Select the Continuebutton to acquire a reference S-ECG.31Q-TECH PROGRAMMER OPERATIONFigure 28: Automatic Setup screenFigure 29: Acquire Reference S-ECG screen102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 31
8. Once the Reference S-ECG acquisition process begins, a status screen appears. The process may take up to one minute, duringwhich the patient should remain still. During this process, a template of the patient’s baseline QRS complex is stored in theSQ-RX device. Select Cancel at any time to end Reference S-ECG acquisition.9. When acquisition is complete, select the Continue button to finish the Automatic Setup process. A confirmation screen willappear when Automatic Setup is complete (Figure 30).Programming Therapy ParametersOnce Automatic Setup has been completed, the SQ-RX device therapy parameters may be selected.To set the therapy parameters:1. Select the Main Menu button on the Navigation Bar.  2. Select the Implant button.3. Select the Device Settings button on the Navigation Bar to display the Device Settings screen (Figure 31). 32Q-TECH PROGRAMMER OPERATIONFigure 30: Confirmation of successful Automatic SetupFigure 31: Device Settings screen102099 003_rev a sept 2 2008.qxp  9/3/2008  11:25 AM  Page 32
33Q-TECH PROGRAMMER OPERATION4. Select the On/Off Therapy switch to set the desired therapy mode (Figure 32).5. Select and drag the Conditional Shock Zone (yellow) and Shock Zone (red) slider bars to set the desired zone configuration.•  The Shock Zone is programmable between 170 and 250 bpm in steps of 10 bpm.•  The Conditional Shock Zone is programmable between 170 and 240 bpm in steps of 10 bpm. Enhanced Detection Criteriaare automatically enabled when the Conditional Shock Zone is programmed.•  When programming both the Shock Zone and Conditional Shock Zone, maintain at least a 10 bpm difference between thetwo zones. If the Conditional Shock Zone slider (yellow) is dragged over the Shock Zone slider (red), the two sliders willmerge to create a single Shock Zone.6. If post-shock pacing is desired, toggle the Post Shock Pacing button to On. (Post-shock bradycardia pacing occurs at a non-programmable rate of 50 ppm for up to 30 seconds. Pacing is inhibited if the intrinsic rate is greater than 50 bpm.)7. Select the Program button to permanently program the SQ-RX device. A screen will appear to confirm that the SQ-RX device settings were programmed (Figure 33). If the SQ-RX device does not accept the SQ-RX device programming, a Program Devicealert screen will appear. Select the Try Again button to return to the Device Settings screen.  Figure 32: Setting the On/Off Therapy switch102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 33
8. Once programming is confirmed, select the Continue button to proceed to the next operation.Note: If any of the SQ-RX device settings are changed from the Device Settings screen and not subsequently programmed,the Pending Program Changes screen will appear. You may select Cancel to return to the Device Settings screen or selectContinue to cancel all SQ-RX device setting changes.    Defibrillation TestingOnce the SQ-RX device is implanted and Automatic Therapy is programmed On, defibrillation testing may be conducted.Note: Defibrillation testing at the time of implantation is at the discretion of the physician.To induce VF and test the S-ICD System:1. Select the Programmer Main Menu button on the Navigation Bar to access the Main Menu.2. Select the Patient Test button (Figure 34) to display the Manual Shock Test screen.34Q-TECH PROGRAMMER OPERATIONFigure 33: Confirmation of programmingFigure 34: Select the Patient Test Button to access the Manual Shock Test screen.102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 34
3. Select the Induction Test button on the Navigation Bar.4. Select either standard (STD) or reverse (REV) polarity.5. Select and drag the red marker to set the desired shock energy for the first delivered shock (Figure 35). The shock energy maybe programmed from 10 to 80 J. 6. Select the Continue button to display the induction test screen or select Cancel to return to the initial Induction Test screen. Note: Episode data associated with rescue shocks, manual shocks and induction testing is not stored in the SQ-RX device.7. Select and hold the Hold To Induce button for the desired duration (Figure 36). 35Q-TECH PROGRAMMER OPERATIONFigure 36: Induction test screenFigure 35: Setting the desired first shock energy for defibrillation testing102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 35
36Q-TECH PROGRAMMER OPERATIONThe following functions occur during the test:•  The S-ICD System induces ventricular fibrillation using 200 mA alternating current (AC) at 50 Hz. Induction continuesuntil the Hold To Induce button is released (up to a maximum of 10 seconds per attempt).•  Arrhythmia detection and the Live S-ECG are suspended during AC induction. Once the Induction button is released, theprogrammer displays the patient’s rhythm.•  Upon detection and confirmation of an induced arrhythmia, the S-ICD System automatically delivers a shock at the programmed energy output and polarity.•  If the shock fails to convert the arrhythmia, re-detection occurs and subsequent shocks are delivered at the SQ-RXdevice’s maximum energy output (80 J).Note: The SQ-RX device can deliver a maximum of five shocks per episode. At any time, an 80 J rescue shock can bedelivered by pressing the Rescue Shock button.8. At any time prior to therapy delivery, the programmed energy may be aborted by selecting the red Abort button.9. Select the Exit button to exit the induction process and return to the initial Induction Test screen.Performing an SQ-RX Follow-up Viewing Device StatusOnce communication is established, the programmer displays the Device Status screen with information regarding the currentepisodes and battery status of the SQ-RX device.To view this information:1. Select the Main Menu button.2. Select the Follow Up button.3. Select the Device Status button on the Navigation Bar. 4. The Device Status screen will appear showing an overview of all SQ-RX device activity since the last communication session(Figure 37). The Device Status overview reports:•  Date of the last follow-up session•  Total number of shocks delivered since the last follow-up session•  Total number of treated episodes since the last follow-up session•  Total number of untreated episodes since the last follow-up session•  Remaining SQ-RX device battery lifeFigure 37: Device status screen102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 36
37Q-TECH PROGRAMMER OPERATIONViewing Stored EpisodesThe SQ-RX device stores up to 24 treated and 20 untreated tachycardia episodes, which can be viewed during a patient’s follow-upsession. When the maximum number of episodes is exceeded, the most recent episode replaces the oldest stored episode; however,the first treated episode will not be overwritten.Note: If a spontaneous episode occurs during a patient’s follow-up session while the SQ-RX device is wirelessly communicating with the programmer, the episode will not be stored. To view stored episodes:1. Select the Main Menu button.2. Select the Follow Up button.3. Select the Captured & Stored Episodes S-ECG icon from the Navigation bar.4. Select the Episodes toggle switch to access the episodes pick list.5. Select an episode (treated or untreated) from the list (Figure 38). The selected Episodes screen appears.6. Each selected Episode screen also displays the programmed parameters and the stored S-ECG data at the time of episodedeclaration. 7. Select the Continue button to return to the Episodes List screen.The following details are available for each episode:Treated EpisodesUp to 96 seconds of S-ECG data is stored for each Treated Episode:•  Pre-episode S-ECG: Up to 12 seconds•  First shock: Up to 24 seconds of pre-shock S-ECG and up to 12 seconds of post-shock S-ECG•  Subsequent shocks: A minimum of 6 seconds of pre-shock and a minimum of 6 seconds post-shock S-ECGFigure 38: Stored Episodes screen102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 37
Untreated EpisodesAn Untreated Episode is defined as any high-rate episode that spontaneously terminates during the charging process, before ashock is delivered.Up to 36 seconds of S-ECG data is stored for each Untreated Episode:• 12 seconds of pre-episode S-ECG• Up to 24 seconds of tachycardia S-ECG dataPrinting Reports from the ProgrammerPrinting ReportsPatient reports can be printed before or after a patient session is ended. It is recommended that a final report be printed immediately following the implant procedure. There are three patient reports:• Summary Report• Captured S-ECG Report• Episode ReportsTo print patient reports from either an Online or Offline session:1. Select the Main Menu button to display the Main Menu screen.2. Select the Print Reports button.  3. Select the desired report type. A green checkmark will appear next to the selected report. Report types are described below.4. Select the Print button to print the selected reports. 5. Select the Cancel button to return to the previously accessed screen.Summary ReportThe Summary Report includes the following information:• Patient Name• Date of Current Follow-Up• Date of Last Follow-Up• Printed Report Date• Device Model/Serial Number• Subcutaneous Electrode Model/Serial Number• Implant Date• Therapy Parameters• Programmed Gain Settings and Sensing Configuration• Initial Shock Polarity Configuration• Episode Summary: Since Last Follow-Up and Since Initial Implant• Battery Status • Subcutaneous Electrode Impedance Measurement• Device Integrity Check, if applicable• Programmer Software VersionTo view and print a summary report, select the Summary Report button on the Select Episode S-ECG Report screen.38Q-TECH PROGRAMMER OPERATION102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 38
Captured S-ECG ReportTo print a Captured S-ECG report:1. Select the Captured S-ECG Report button from the Printed Report Type list screen. A list of the captured S-ECG strips is displayed. Select the button next to the Captured S-ECG(s) to be printed. A green checkmark appears next to the selectedCaptured S-ECG (Figure 39).2. Select Continue to return to the Printed Report Type pick list screen. 3. Select the Print button to print the selected report.4. Select the Cancel button to return to the previously accessed screen.Episodes ReportTo print an Episode Report:1. Select the Episodes Report button on the Printed Report Type list screen. The Episode List screen appears showing a list of thestored episodes (Figure 40). Select the button next to the episode(s) to be printed. A green checkmark appears next to theselected episode(s).39Q-TECH PROGRAMMER OPERATION Figure 39: Selecting Captured S-ECG Reports for printing102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 39
2. Select Continue to return to the Printed Report Type pick list screen. 3. Select the Print button to print the selected report.4. Select the Cancel button to return to the previously accessed screen.S-ECG FeaturesThe programmer provides the capability to view, adjust, and capture the streaming S-ECG from the SQ-RX device.S-ECG Rhythm Strip MarkersThe system provides annotations to identify specific events on the S-ECG. These markers are shown in Table 2 and an example isshown in Figure 41.40Q-TECH PROGRAMMER OPERATIONFigure 40: Stored Episode List screenFigure 41: Sample S-ECG markersTable 2: S-ECG Rhythm Strip MarkersDescription MarkerDisplayScreenPrinted ReportsCharging CX33Sensed Beat S3333Noisy Beat N3333Paced Beat P3333Tachy Detection T3333Discard Beat •3333Shock 3333Return to NSR X33102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 40
S-ECG Scale SettingsTo adjust the real-time S-ECG amplitude and display speed scale settings:1. Select the S-ECG Display Settings icon located to the right of the Live S-ECG window. The S-ECG Settings screen is displayed.2. Select and drag the Gain or Sweep Speed Scale bars as desired (Figure 42). The S-ECG scale will change according to the selected setting. The gain setting controls the visual gain. The programmer defaults to Full range for devices with an SQ-RXdevice gain setting of x1 and to Half range for devices with an SQ-RX device gain setting of x2. The Sweep Speed slider controlsthe display speed of the scrolling Live S-ECG. The nominal sweep speed setting is 25 mm/sec.Note: Amplitude settings and display speed adjustments on scrolling real-time S-ECG and Captured S-ECGs affect thedisplay screen settings only and have no impact on the SQ-RX device settings for sensing.Capture and View S-ECG StripsThe programmer can display, capture, and store real-time S-ECG rhythm strips. The programmer saves a maximum of five,12-second, Captured S-ECG recordings comprised of: • 8.5 seconds before activation of the Capture S-ECG button• 3.5 seconds after activation of the Capture S-ECG buttonIf an additional recording is required, then the oldest previous recording is replaced with the new recording.Note: When the programmer goes Offline during a follow-up session the previous 60 seconds of S-ECG data are availablefor review in the Offline Stored Patient Sessions display.41Q-TECH PROGRAMMER OPERATIONFigure 42: Adjusting Gain and Sweep Speed102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 41
42Q-TECH PROGRAMMER OPERATIONTo capture a new S-ECG rhythm strip:1. Select the Capture S-ECG button located to the left of the Live S-ECG window (Figure 43). The S-ECG will scroll across thedisplay screen. Calipers appear below the Captured S-ECG rhythm strip (Figure 44). Each 12-second recording is date and timestamped according to the programmer’s date and time setting. 2. Select and move the calipers across the S-ECG strip to measure intervals as desired.3. Select the Continue button to return to the previously accessed screen.To view a previously Captured S-ECG:1. Select the Main Menu button.2. Select the Follow Up button.3. Select the Captured & Stored Episode S-ECG button.4. Select Captured on the toggle button. The Captured S-ECG pick list screen appears.5. Select one Captured S-ECG from the list. The Captured S-ECG Details screen appears.6. Select and drag the calipers to view details.7. Select the Continue button to return to the Captured S-ECG list screen.Utilities MenuThe programmer Utilities menu provides access to additional SQ-RX device features. These include Acquire Reference S-ECG,Beeper Control, Manual Setup and Smart Charge.To access the Utilities menu during an Online session:1. Select the Main Menu button to display the Main Menu screen.2. Select the Utilities button. The Utilities screen appears (Figure 45).Figure 43: Selecting the Capture S-ECG button Figure 44: Captured S-ECG rhythm strip102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 42
43Q-TECH PROGRAMMER OPERATIONAcquire Reference S-ECGTo acquire a manual Reference S-ECG:1. Select the Acquire Reference S-ECG button on the Utilities menu to access the Acquire Reference S-ECG screen.2. Select Continue to acquire a Reference S-ECG. The programmer will begin acquiring the Reference S-ECG. A message willappear requesting that the patient remain still (Figure 46). The reference S-ECG QRS template is recorded and stored in theSQ-RX device. Select the Continue button to complete the process and return to the Utilities screen. The Cancel button can beused at any time to end S-ECG acquisition and return to the Utilities screen.Figure 45: Utilities screenFigure 46: The patient should remain still while the Reference S-ECG is acquired.102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 43
Beeper Control The SQ-RX device has an internal alert system that emits an audible beeping tone when an alert condition occurs. Alert conditionsmay include Elective Replacement Indicator (ERI), End of Life (EOL), subcutaneous electrode impedance out of range, prolongedcharge times, and failed Device Integrity Check. Beeper Control allows activation or suppression of the beeper.Once an alert condition is triggered, audible tones beep for 16 seconds every nine hours until the alert condition is resolved.Beeper Control is automatically enabled when the SQ-RX device is taken out of Shelf mode.To reset the Beeper Control:Note: The Beeper Control display screen is enabled only when an alert condition occurs. If an alert condition is activated a notification screen will appear upon connection.1. Select the Main Menu button to display the Main Menu.2. Select the Utilities button to display the Utilities Menu.3. Select the Beeper Control button. 4. Select the Reset Beeper button to suspend audible beeping tones triggered by an alert condition (Figure 47). If the alert condition is not corrected, the audible beeping tones will be reactivated during the next automatic S-ICD System self-check.5. The disable Beeper feature is enabled when ERI is reached or EOL occurs. If desired, select disable Beeper when ERI or EOLoccurs. Beeper tones are permanently disabled.Note: If beeper is disabled alert conditions will not trigger audible tones.  Manual SetupManual Setup enables the user to perform the electrode integrity test and select the electrode sensing configuration and gain setting in the SQ-RX device.1. From the Utilities Menu, select the Manual Setup button. The Impedance Test screen appears.2. Select the Test button to perform the electrode integrity test (Figure 48). 44Q-TECH PROGRAMMER OPERATIONFigure 47: Resetting the Beeper102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 44
3. Select Continue to proceed or select Test to retry. The sensing configuration can be evaluated by temporarily activating any ofthe vectors or gain settings. The Live S-ECG and markers will reflect the temporary selections for evaluation purposes.4. There are three available sensing vectors that can be manually selected:• Primary: Sensing from the proximal electrode ring on the subcutaneous electrode to the surface of the active SQ-RXdevice• Secondary: Sensing from the distal sensing electrode ring on the subcutaneous electrode to the surface of the activeSQ-RX device• Alternate: Sensing from the distal sensing electrode ring to the proximal sensing electrode ring on the subcutaneouselectrodeThe gain setting adjusts the sensed S-ECG signal sensitivity. It may be manually selected via the Select Gain toggle button(Figure 49).45Q-TECH PROGRAMMER OPERATIONFigure 48: Impedance test screenFigure 49: Gain adjustment102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 45
• 1x Gain (± 4 mV) should be selected when the signal amplitude is large enough to cause clipping when the 2x gain isselected.• 2x Gain (± 2 mV) should be selected when the signal amplitude is small enough to allow use of a more sensitive settingwithout causing clipping of the captured signal. The 2x gain selection amplifies the signal twice as much as the 1x gainselection.To program the manually selected sense configuration:1. Select the Program button to save the sense vector and gain settings.2. The Acquire Reference S-ECG process is automatically enabled during the Manual Setup Process. Select the Continue button toacquire a reference S-ECG. A confirmation screen will appear when the captured reference S-ECG is acquired (Figure 50).3. Select the Continue buttonSmart ChargeThrough the Smart Charge feature, the SQ-RX device charge initiation sequence adapts to the occurrence of non-sustained ventricular arrhythmia episodes by delaying capacitor charging. This conserves battery longevity and may prevent unnecessaryshocks for non-sustained arrhythmias.Smart Charge is enabled automatically when an untreated ventricular arrhythmia episode is recorded. To reset the Smart Chargefeature:1. Select the Main Menu button.2. Select the Utilities button on the Main Menu to display the Utilities Menu.3. Select the Smart Charge button. The Reset Smart Charge screen appears (Figure 51).46Q-TECH PROGRAMMER OPERATIONFigure 50: Captured Reference S-ECG confirmation102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 46
4. Select the Reset button to reset the Smart Charge to zero or press Cancel to return to the Utilities Menu without resetting theSmart Charge. 5. A confirmation window will appear with the message “Smart Charge successfully reset.” Press the Continue button to return tothe Utilities screen. Additional Programmer Functions Rescue ShockThe Rescue Shock button is available on the programmer display when the SQ-RX device Setup Process is complete and an SQ-RXdevice is actively communicating with the programmer. During active communication, a maximum (80 J) rescue shock can be delivered upon programmer command. To deliver a rescue shock:1. Select the red Rescue Button at the top of the programmer screen. The Rescue Shock screen appears (Figure 52).47Q-TECH PROGRAMMER OPERATIONFigure 51: Reset Smart Charge screenFigure 52: Rescue Shock screen102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 47
2. Select the Shock button to begin charging the SQ-RX device for a Rescue shock. A red background screen with the word“Charging” will appear. Alternatively, selecting the Cancel button will prevent delivery of a rescue shock and will return to theprevious menu screen. The shock can be aborted by selecting the Abort button.3. A confirmation screen will appear with notification that the shock was delivered successfully along with the corresponding shockimpedance (Figure 53).If for any reason the shock could not be delivered, a red background screen will appear with a message stating “The Shock couldnot be delivered.”Note: In the event telemetry is lost, SQ-RX device commands, including Rescue Shocks, will not be available untiltelemetry is reestablished.Magnet UseThe Cameron Health Model 4520 magnet is a non-sterile accessory used to inhibit the delivery of therapy from the SQ-RX device.Placing the magnet against the skin directly over the implanted SQ-RX device will temporarily inhibit rhythm detection, abort high- voltage capacitor charging and delivery of post-shock pacing. Removal of the magnet will return the device to normal operation. If the magnet is applied during an episode, the episode will not be stored in the SQ-RX device memory.Note: A manual Rescue Shock can be delivered during magnet application.48Q-TECH PROGRAMMER OPERATIONFigure 53: Confirmation of Rescue Shock delivery102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 48
Cleaning and MaintenanceThere are no user-serviceable parts in the programmer system with the exception of the wand and batteries.  To clean the programmer and wand as needed:1. Turn the Programmer Off by pressing the FN key and the ESC On/Off key simultaneously.2. Gently wipe the screen and keyboard with a soft, clean, dry cloth.3. Wipe the plastic portions of the programmer and the wand with an isopropyl alcohol-moistened cloth. 4. Dry the programmer immediately to remove residue.Programmer End of LifeThe programmer and accessories are designed to provide years of service under typical use. To dispose of, return or exchange a programmer, contact your local Cameron Health representative or Customer Service Department for instructions and return packaging.  49Q-TECH PROGRAMMER MAINTENANCE102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 49
This section presents potential programmer issues and the possible solutions to resolve them. Contact your Cameron HealthRepresentative or Customer Service for additional assistance.Inability to PrintIf unable to print, follow the steps below:1. Ensure that the printer is turned On and that it contains paper and a sufficient ink supply.2. Check printer feed for paper jam.3. Ensure the wireless printer card is inserted into the programmer.4. Ensure the wireless printer card is inserted into the printer.No Printer AvailableThe No Printer Available screen will appear if a printer was not set up. Refer to the Printer Selection section for instructions.Loss of Communication with PrinterWhen communication between the programmer and the printer fail, a Printing Error screen will appear with a message stating“Error communicating with the printer; would you like to try again?”. If this occurs:1. Select the Try Again button to reconnect to the printer. Verify that the selected printer has a Cameron Health-approved wirelessadapter inserted in the programmer.2. Move the programmer closer to the printer.3. Move any devices, and the associated cables, that may be interfering with the RF communication.Inability to Communicate With the SQ-RX DeviceIf the programmer is unable to communicate with the SQ-RX device, follow the steps below:1. Attempt to reposition the wand.2. Select Scan For Devices from the Main Menu or select Scan Again from the Device List screen to locate the desired device.3. Move any equipment and associated cables that may be interfering with RF communication.50Q-TECH PROGRAMMER POSSIBLE PROBLEMS AND SOLUTIONS102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 50
Federal Communications Commission (FCC) ComplianceThis transmitter is authorized by rule under the Medical Implant Communications Service (part 95 of the FCC Rules) and must notcause harmful interference to stations operating in the 400.150 - 406.00 MHz band in the Meteorological Aids (i.e., transmitters andreceivers used to communicate weather data), the Meteorological Satellite, or the Earth Exploration Satellite Services and mustaccept interference that may be caused by such aids, including interference that may cause undesired operation. This transmittershall be used only in accordance with the FCC Rules governing the Medical Implant Communications Service. Analog and digitalvoice communications are prohibited. Although this transmitter has been approved by the Federal Communications Commission,there is no guarantee that it will not receive interference or that any particular transmission from this transmitter will be free frominterference.1999/5/EC Compliance (R&TTE Directive)The S-ICD System contains radio equipment in the frequency range 402 MHz to 405 MHz for ultra low power active medicalimplants.  The radio equipment in the S-ICD System complies with the applicable harmonized standards and essentialrequirements of the R&TTE Directive.  For additional information contact Cameron Health, Inc. direct by the telephone numberslisted on the inside front cover or back page of this manual.  EMI/RFI This equipment has been tested and found to comply with the applicable limits for medical devices, IEC 60601-1-2:2001 or ActiveImplantable Medical Device Directive 90/385/33c.  Although this testing shows the device to provide reasonable protection against harmful interference in a typical medical installation, there is no guarantee that interference will not occur in a particular installation.  If the device does cause harmfulinterference the user is encouraged to try and correct the interference by the following measures:• Reorient or relocate the device• Increase the separation between the devices• Connect the equipment to an outlet on a different circuit• Contact your Cameron Health Representative or Cameron Health Customer Service51Q-TECH PROGRAMMER COMPLIANCE STATEMENTSWireless ConnectionFrom ProgrammerFrequency Bandwidth FCC Designation R&TTE ClassProgrammer WirelessAdapter and PulseGenerator403.5 MHz < 300 KHz CFR 47, Part 951,MICSClass 1Subclass 47Printer Adapter 2.400 - 2.4835 GHz < 1000 KHz Maximum CFR 47, Part 15,Class BClass 2Subclass 21Table 3: EMI/RFI Information102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 51
52Table 4: Declaration Electromagnetic EmissionsThe Q-TECH Model 2020 programmer is intended for use in the electromagnetic environment specified below.  The customer or the user of the Q-TECH Model 2020 should assure that it is used in such an environment.Emissions Test Compliance Electromagnetic Environment GuidanceRF Emissions CISPR 11 Group 1 The Q-TECH Model 2020 programmer uses RF energy only for its internal func-tion.  Therefore, its RF emissions are very low and are not likely to cause anyinterference in nearby electronic equipment.RF Emissions CISPR 11 Class A The Q-TECH Model 2020 programmer is suitable for use in all establishmentsother than domestic and those directly connected to the public low-voltagepower supply network that supplies buildings used for domestic purposes.Harmonic Emissions IEC 61000-3-2 Class AVoltage Fluctuations/Flicker Emissions IEC 6100-3-3CompliesThe Q-TECH Model 2020 programmer is intended for use in the electromagnetic environment specified below.  The customer or the user of the Q-TECH Model 2020 should assure that it is used in such an environment.Immunity Test IEC 60601 Test LevelCompliance Level Electromagnetic Environment GuidanceElectrostatic Discharge(ESD)IEC 61000-4-2± 6 kV contact± 8 kV air± 6 kV contact± 8 kV airFloors should be wood, concrete, or ceramic tile.  If floors are covered with synthetic material, then the relative humidity should be at least 30%.Electrical Fast Transient/BurstIEC 61000-4-4± 2 kV for powersupply lines ± 1 kV for input/output lines± 2 kV for powersupply lines ± 1 kV for input/output linesMains power quality should be that of a typical commercial or hospital environment.SurgeIEC 61000-4-5± 1 kV line(s) to line(s)± 2 kV line(s) toearth± 1 kV line(s) to line(s)± 2 kV line(s) toearthMains power quality should be that of a typical commercial or hospital environment.Voltage Dips, Short interruptions, and voltage variations onpower supply input linesIEC 61000-4-11<5 % UT(>95% dip in UT)for 0.5 cycles40% UT(60% dip in UT)for 5 cycles70% UT(30% dip in UT)for 25 cycles<5% UT(>95% dip in UT)for 5 sec<5 % UT(>95% dip in UT)for 0.5 cycles40% UT(60% dip in UT)for 5 cycles70% UT(30% dip in UT)for 25 cycles<5% UT(>95% dip in UT)for 5 secMains power quality should be that of a typical commercial or hospital environment.If the user of the Q-TECH Model 2020 programmer requires continued operationduring power mains interruptions, it is recommended that the Q-TECH Model 2020programmer be powered from an uninterruptible power supply or a battery.Power Frequency(50/60 Hz) Magnetic FieldsIEC 61000-4-83 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of  a typical location in a typical commercial or hospital environment.NOTE:  UTis the a.c. mains voltage prior to application of the test level.Table 5: Declaration Electromagnetic Immunity Part 1Q-TECH PROGRAMMER DECLARATIONS102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 52
53Table 6: Declaration Electromagnetic Immunity Part 2The Q-TECH Model 2020 programmer is intended for use in the electromagnetic environment specified below.  The customer or  the user of the Q-TECH Model 2020 should assure that it is used in such an environment.Immunity Test IEC 60601 Test LevelCompliance Level Electromagnetic Environment GuidanceConducted RFIEC 61000-4-63 Vrms150 kHz to 80 MHz3 V Portable and mobile RF communications equipment should be used no closer to anypart of the Q-TECH Model 2020 programmer, including cables, than the recommended separation distance calculated from the equation applicable to thefrequency of the transmitter.Recommended Separation Distancewhere P is the maximum output power rating of the transmitter in watts (W)according to the transmitter manufacturer and d is the recommended separationdistance in meters (m).Field strengths from fixed RF transmitters, as determined by an electromagneticsite survey,a  should be less than the compliance level in each frequency range.bInterference may occur in the vicinity of equipment marked with the following symbol:Radiated RFIEC 61000-4-33 V/m80 Mhz to2.5 GHz3 V/maField strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.  To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.  If the measured field strength in the location in which the Q-TECH Model 2020 is used exceeds the applicable RF compliance level above, the Q-TECH Model 2020 should be observed to verify normal operation.  If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Q-TECH Model 2020.bOver the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.Note 1  At 80 MHz and 800 MHz, the higher frequency range applies.Note 2  These guidelines may not apply in all situations.  Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.Q-TECH PROGRAMMER DECLARATIONS102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 53
54Table 7: Recommended Separation Distances Recommended separation distances between portable and mobile RF communications equipment and the Q-TECH Model 2020 programmerThe Q-TECH Model 2020 programmer is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.  The customeror the user of the Q-TECH Model 2020 programmer can help prevent electromagnetic interference by maintaining a minimum distance between portable andmobile RF communications equipment (transmitters) and the Q-TECH Model 2020 programmer as recommended below, according to the maximum outputpower of the communications equipment.0.01 0.117 0.117 0.2330.1 0.369 0.369 0.36911.17 1.17 1.1710 3.69 3.69 7.38100 11.7 11.7 23.3For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equa-tion applicable to the frequency of the transmitter, where p is the maximum output power rating of the transmitter in watts (W) according to the transmittermanufacturer.  Note 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.Note 2 These guidelines may not apply in all situations.  Electromagnetic propagation is affected by absorption and reflection from structures, objects andpeople.Q-TECH PROGRAMMER DECLARATIONS102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 54
55Q-TECH PROGRAMMER SPECIFICATIONSComponent RequirementDC PowerBattery Pack Type 2200 mAh 12.6 volt lithium-ion battery packCharge Time Approximately 4 hoursBattery Capacity Gradually declines approximately 15 - 20% of initial capacity over 400 charge/dischargecyclesStorage Battery capacity declines with age (storage)End Of Life Battery capacity has declined 50% of battery pack’s initial capacity, approximately 1000 charge/discharge cyclesTypical use is 4 yearsBack-Up Batteries Two AAA alkaline batteriesPower SupplyInput 100 to 240 VAC, 47 - 63 Hz, 0.7AOutput 15 VDC, 1.6 APower:  24 WManufacturer/Model Elpac Power Systems MW2415Environment  Operating StorageTemperature 15° C to +38° C+59° F to +100° F-10° C to + 55° C+14° F to + 131° FRelative Humidity 5% to 93% maximum at 40° C, non-condensing5% to 93% maximum at 40° C, non-condensingTable 8: Product GuidelinesComponent SpecificationDimensionsWidth x Depth x Height24 cm x 18 cm x 4.0 cm9.3 in x 7.2 in x 1.4 inWeight 1.1 kg, 3.0 lbsIncludes 2200 mAh battery packStandard Screen Display High Quality SVGA600 x 800 pixels, 16-bit color display with backlight, stylus and touch screen sensitiveTable 9: Specifications102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 55
56Q-TECH PROGRAMMER DEFINITIONS OF PACKAGE LABEL SYMBOLSTable 10: Packaging SymbolsSymbol Definition Symbol DefinitionType BF Applied Part Symbol - Indicates    connection of  Type BF applied part.Federal Caution - Federal Law (USA) restrictsthis SQ-RX device to sale by or on the order of a physician.PLimitation European Community Represented -   Authorized Representative in the EU community. MManufacturer - Identification of manufacturer Part Number - Component number. Storage Temperature - Product is stored with temperature limitations.Do Not Dispose - Do not dispose of this productin the unsorted municipal waste stream.  Disposeof this product  according to local regulations.Hazardous Voltage - Caution - dangerousvoltage.hReference Number - Finished goods stock  number.Radio - Radio frequency.fSerial Number - Serial number of the SQ-RX device. iInstructions - Consult instructions before use.Intended Use Symbol - In the USA Federal Law restricts the SQ-RX device to sale by or on the order of a physician.Risk Of Shock - Indicates the risk of electricshock.Reserved - Symbols. Fragile: Handle with Care -  Transport and store with care.Reserved - Symbols. Keep Dry - Ship and store in a dry place.Flap, Open - Identifies control to open lid orflap. NDate Of Manufacture - Date on which the SQ-RX device was  manufactured. Power Plug - Identifies the storage place. Electrostatic Discharge - Identifies the potential of a sudden flow of electric current through a material that is normally an insulator. 102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 56
Cameron Health Q-TECH Programmer WarrantyCameron Health, Inc. offers a full lifetime warranty on the Q-TECH Programmer to the user or using facility. Please contact yourlocal Cameron Health office or local representative for details. The warranty on the Q-TECH Programmer does not extend to anyproduct that has been tampered with, altered, or repaired by any person other than an employee of Cameron Health organization.  ServiceWhen requesting service, please provide information concerning the nature of the failure and the manner in which the equipmentwas used when the failure occurred. The model number and serial number should also be provided. Before returning any productsto the factory, call Cameron Health Customer Service or your local Cameron Health Representative for a Return Authorization num-ber.DisclaimerEvery effort has been made to make the material in the Q-TECH Programmer User’s Manual complete, accurate, and up-to-date.Cameron Health reserves the right to make changes without notice and shall not be responsible for any damages, including but notlimited to, consequential damages caused by reliance on the material presented, including but not limited to typographical errors.57Q-TECH PROGRAMMER WARRANTY102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 57
102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 58
59102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 59
60PN 102099-003 Rev A 2008/08 Cameron Health, Inc.905 Calle AmanecerSuite 300San Clemente, CA  92673USATel:  1 949 498 5630Free:  1 877 SICD 4111 877 742 3411Fax:  1 949 498 5932URL:  www.cameronhealth.comCameron Health BVWorld Trade CenterNieuwe Stationsstraat 106811 KS ArnhemThe NetherlandsTel:  31 26 3550260Free:  800 SICD 4 YOU800 7423 4 968Fax:  31 26 3550269URL:  www.cameronhealth.com102099 003_rev a sept 2 2008.qxp  9/3/2008  11:26 AM  Page 60