Cleveland Medical Devices 0103 QUANTATIVE MOTOR ASSESSMENT SYSTEM User Manual USERS MANUAL 2

Cleveland Medical Devices Inc. QUANTATIVE MOTOR ASSESSMENT SYSTEM USERS MANUAL 2

Contents

USERS MANUAL 2

 Telephone: (216) 791-6720 or Toll-free 1-877-CleveMed (1-877-253-8363) Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  1    9:00 a.m. - 5:00 p.m. EST     Monday - Friday Fax:    (216) 791-6739  E-Mail:  Customer Support:      Sales:  Web:       Mailing Address: Cleveland Medical Devices Inc.       4415 Euclid Avenue, Fourth Floor       Cleveland, Ohio 44103  © Cleveland Medical Devices Inc. 2006                 Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  2TABLE OF CONTENTS   Introduction....................................................................................................................... 4 Intended Use ................................................................................................................... 4 Contraindications ........................................................................................................... 4 Warnings ......................................................................................................................... 4 Package Contents and Warranty Information ................................................................ 6 Chapter 1: About ParkinSense........................................................................................ 7 General Device Description ........................................................................................... 7 User Unit......................................................................................................................... 8 Sensors ............................................................................................................................ 9 User Interface Software ................................................................................................ 10 Chapter 2: Setting Up ParkinSense............................................................................... 11 Installing the ParkinSense Program............................................................................. 11 Charging ParkinSense User Unit ................................................................................. 11 Installing the ParkinSense Hardware........................................................................... 11 Starting the ParkinSense Program ............................................................................... 12 Obtaining USB data communication port: ..................... Error! Bookmark not defined. Chapter 3: About ParkinSense Software...................................................................... 13 Running the Application ............................................................................................... 13 Using the Application ................................................................................................... 13 Accessing the Patient Database.................................................................................... 14 Adding a Patient ........................................................................................................... 14 Updating a Patient........................................................................................................ 15 System Setup.................................................................................................................. 15 Collecting Data Wirelessly Using Bluetooth ................................................................ 18 Starting Data Acquisition ............................................................................................. 19 Acquired Data Display ................................................................................................. 19 Saving Acquired Data ................................................................................................... 19 Stopping Data Acquisition ............................................................................................ 19 Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  3Collecting Data From Memory..................................................................................... 20 Switching Data Modes .................................................................................................. 21 Erasing Memory............................................................................................................ 21 Reviewing Saved Data .................................................................................................. 22 Printing Current Graph ................................................................................................ 24 Graph Color and Scale Setup ....................................................................................... 25 Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  4Introduction Intended Use  ParkinSense is not FDA cleared to market.  It is not intended for use in experimentation that involves human testing without specific IRB approval.  The  ParkinSense  movement  disorders  quantification  system  is  intended  for  use  by researchers  analyzing  quantitative  features  of  movement  disorders  which  require recording motion and/or electrical activity of muscle.     THIS DEVICE COMPLIES WITH PART 15 OF THE FCC RULES.  OPERATION IS SUBJECT  TO  THE  FOLLOWING  TWO  CONDITIONS:  (1)  THIS  DEVICE  MAY NOT CAUSE HARMFUL INTERFERENCE, AND (2) THIS DEVICE MUST ACCEPT ANY  INTERFERENCE  RECEIVED,  INCLUDING  INTERFERENCE  THAT  MAY CAUSE UNDESIRED OPERATION.  NOTE: THE MANUFACTURERE IS NOT RESPONSIBLE FOR ANY RADIO OR TV INTERVERENCE  CAUSED  BY  UNAUTHORIZED  MODIFICATIONS  TO  THIS EQUIPMENT.    SUCH  MODIFICATIONS  COULD  VOID  THE  PATIENT’S AUTHORITY TO OPERATE THE EQUIPMENT.   Contraindications  Interference may occur in the vicinity of equipment marked with the following symbol:     This device complies with CFR 47 – Part 15, 15.109(b) and 15.247.   Warnings  !!!! Do NOT  connect the  system to  a  computer  via the mini-USB  port at the  same time a subject is connected to the electromyography electrode leads.  !!!!  Do NOT expose the system to water.  Water exposure may permanently damage the unit.     Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  5  Minimum System Requirements  Personal computer with Pentium III™ 500 MHz or higher processor (or equivalent); Microsoft® Windows 98/NT 4/2000/XP; 256 MB of installed memory; Minimum 500 MB free hard disk space (1 GB recommended); CD-ROM Drive; 1024x768 or greater display resolution; Mouse or other pointing device; One USB port Bluetooth™ Radio (Included with ParkinSense kit, however other Bluetooth adapters may be used)   The ParkinSense system communicates with your computer using a Bluetooth Radio link. There  is  a  Bluetooth  radio  embedded  inside  the  ParkinSense  unit  which  will communicate  with  a  Bluetooth  radio  connected  to  your  computer.    A  USB  Bluetooth adapter is provided with the ParkinSense system, however other Bluetooth adapters may be used. Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  6Package Contents and Warranty Information  Cleveland Medical Devices Inc. thanks you for your recent product purchase.  CleveMed offers phone technical support (9 AM-5 PM EST) and warrants the ParkinSense (parts and labor) for  90  days from the  date  of  purchase.   Technical support  after the 90  day period  will  be  charged  hourly.  Call  CleveMed  for  hourly  rates.    For  your  benefit,  we recommend that  you record the  pertinent details  below.    If  necessary, this  information will allow us to better serve your needs.  We highly recommend that you staple a copy of the sales receipt to the blank pages in the back of this manual.  Please  check  to  make  sure  your  kit  has  the  required  components  and  records  the requested data.   Data of Purchase:___________________________   S/N:_______________________________   (1) ParkinSense Unit (1) Data Cable, USB (1) USB Bluetooth Adapter (1) EMG snap lead cable (1) Flexible wrist band (1) ParkinSense software (1) User’s Guide; this document     Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  7Chapter 1:  About ParkinSense  General Device Description  The  ParkinSense  movement  disorder  assessment  system  can  be  utilized  for  a  wide  variety of applications to quantify motion and electrical muscle activity.  The system includes a user worn unit, a radio for wireless data transmission, and software to collect, manage, and analyze data.  The ParkinSense user unit is typically worn on the wrist and finger of a subject (Figure 1).  The wrist module slips over a comfortable band worn around the wrist or forearm.  A sensor module then slips over the middle finger of a subject. ParkinSense provides a total of eight channels of data including three channels of accelerometry, three channels of gyroscopes, and two channels of  electromyography  (EMG).    The  system  measures  three  dimensional  motion  using  three orthogonal accelerometers and three orthogonal gyroscopes located in the sensor module.  The system also  measures  two  channels of EMG.  The lemo connector input  on  the wrist module provides connection for the two channel EMG snap leads which come with the system.    Figure 1. The user unit is worn by the subject and consists of finger and wrist worn components.   Table 1.  ParkinSense system specifications. Sensor Module Size  0.57 x 0.80 x 1.03 in. Sensor Module Weight  0.7 oz. / 20 g. Command Module Size  2.88 x 1.56 x 0.66 in. Command Module Weight  3.25 oz. / 92 g. Electromyography  Lemo Connector with 5 leads (2 differential EMG leads and one ground) Wireless Link  Bluetooth™ Radio Data Transmission Range  ~ 20 feet (line of sight) Data Rate  57.6 kbps, 4800 bytes/s throughput Transmission Frequency  2.400-2.483 GHz Transmission Bandwidth  1 MHz Batteries  Rechargeable Lithium Polymer 850 mA-Hr Battery Life  12 Hrs  Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  8User Unit  The user unit (Figure 1) is worn by the subject and consists of two parts connected by a thin, flexible grey cable.  To put the user unit on, the subject should first slip the thin, flexible wrist band over their forearm just below their wrist.  Next, the sensor module of the ParkinSense User Unit should be place over their middle finger with the grey cable facing toward the wrist.  The clip of the wrist module should then be slid over the wrist band.   Figures 2 – 5.  From left to right.   2Add pictures here to show setup. Fig 2. Wrist band alone, Fig 3. Sliding over the finger, Fig 4. Sliding over wrist band.  Fig 5. Complete System.   The ParkinSense user unit integrates the following features:   Bluetooth  Radio – A Bluetooth Radio is integrated in the ParkinSense system.  This provides real-time data transmission  up  to  a computer with another Bluetooth  radio.  The transmission range is 30 feet line of sight.  On Board Memory – In addition to data transmission over the wireless link, data is continually stored to on board flash memory whenever the system is turned on.  The system can store up to 12 hours of data.  Mini  USB  Connector  –  The  mini  USB  connector  provides  a  method  to  directly  connect  to  a computer with the ParkinSense software installed.  This provides three important functions: -  Data download from the user unit memory -  Erasing the user unit memory -  Recharging the user unit battery  Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  9 Figure 6. The ParkinSense wrist module includes both a mini USB and lemo connector.  Electromyography  –  The  user  unit  provides  a  lemo  connector  input  for  two  channels  of electromyography  (EMG).  The  electromyography  data  transmitted  over  the  wireless  link  and saved in memory is processed data.  The user unit samples EMG at 2 KHz.  However, the root mean square (RMS) value of the EMG signal is processed every 7.8 milliseconds and stored to memory or sent over the wireless link. This provides data synchronization between the EMG and the mechanical sensors which are sampled at 128Hz (or every 7.8 milliseconds). (Table 2)  Gyroscopes – The finger worn sensor module includes three orthogonally positioned gyroscopes.  This provides three axes of angular rotation data. (Table 2)  Accelerometers  -  The  finger  worn  sensor  module  includes  three  orthogonally  positioned accelerometers.  This provides three axes of linear acceleration data. (Table 2)  Sensors  ParkinSense integrates motion sensing and EMG in a single device.  Eight channels of data are providing  including  three  orthogonal  accelerometers,  three  orthogonal  gyroscopes,  and  two channels of differential EMG.  The exact sensors specifications are shown below (Table 3).   Table 3.  ParkinSense system sensor specifications.   Angular Rate  Acceleration  Electromyography Sensor Type  MEMS Gyroscopes  MEMS Accelerometers  Snap electrodes Number of Channels  3 Orthogonal  Channels  3 Orthogonal  Channels  2 differential channels Input Range  ±1200 deg/sec  ±5 g  ±5 mV Min. Freq (3dB)  0Hz  0 Hz  30 Hz Max. Freq (3dB)  15 Hz  15 Hz  1024 Hz Input Noise  <1.6 deg/sec RMS  <2 mg RMS  <15 µV p-p CMRR  N/A  N/A  >60 dB A/D resolution  12-bit  12-bit  12-bit Sampling Rates  128 sps  128 sps  2048 sps per channel DC offset rejection  N/A  N/A  300 mV Input Impedance  N/A  N/A  >20 Mohm  Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  10 Accelerometer - axis  Add a picture here to show the orientation of the sensors. (Fig 7)    User Interface Software  ParkinSense user interface software provides several functions including:  -  Real-time data display -  Saving real-time data to file -  Patient database and file management -  Saved data review -  Downloading data from the user unit -  Erasing the memory of the user unit -  Frequency analysis of saved data files  The software is explained in greater detail in Chapter 3.   Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  11Chapter 2: Setting Up ParkinSense Installing the ParkinSense Program  1.  Insert the ParkinSense installation CD-ROM into your computer’s CD-ROM drive. 2.  Double-click on the ParkinSense Installer program. 3.  Follow the on-screen prompts to install the software. 4.  Once the software has been installed, start the ParkinSense software. Charging ParkinSense User Unit  1.  Locate the ParkinSense User Unit. 2.  Plug the large end of the USB data cable into your computer’s USB port. Plug the small end of the USB cable into the ParkinSense User Unit. 3.  The green LED should come on, indicating the device is charging.  Installing the ParkinSense Hardware 1.  Make sure the device is fully charged before first use. 2.  Turn on the ParkinSense User Unit. A series of beeps indicates that the device is on and ready for wireless data transmission. 3.  Install the Bluetooth driver from the disc provided with the Bluetooth unit. Follow the directions in the package. 4.  Plug the Bluetooth Device into an open USB port. 5.  Go to your Bluetooth Devices Manager by going to  <Start Menu > < Settings> <Control Panel> < Bluetooth Configuration>   6.  Select Bluetooth Wizard, and complete the Wizard process. Select the following services: File Transfer, Network Access, Bluetooth Serial Port. 7.  When the Wizard has completed, double click on My Bluetooth Places. Select the Bluetooth Setup Wizard, and check the box for “I want to find a specific Bluetooth device…”  8.  Windows will now find all available Bluetooth devices. Double click on Parkinsense.    Figure 3. Add Bluetooth Device Wizard  9.  The wizard will now prompt you for a PIN Code for the device. Enter four zero (0000 and click on “Initiate Pairing.” Check “A7 Serial Port.” Click on “Configure.” The port number will be displayed. Record this number for later use.  `    Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  1210. After the device is installed, the wizard will display the new port assigned to your              device. The new port for your device is label as “A7 Serial Port on Parkinsense:”          11. Go to “My Bluetooth Places.” Double click on “A7 Serial Port on Parkinsense:”The icon will light up and the port will be displayed.   Note  :  When using an external Bluetooth device, attaching the USB Bluetooth device to a different USB port will change the radio port of your ParkinSense User Unit.   Starting the ParkinSense Program (if it isn’t already active)  1.  Click the Start Button 2.  Point to Programs 3.  Click on ParkinSense                              Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  13Chapter 3: About ParkinSense Software Running the Application  When the ParkinSense Data Collection Package has been installed, it can be run from the Windows Start menu. Click Start  Programs  ParkinSense  Using the Application  Once the program is running, a main menu will appear.     Figure 7.  The ParkinSense Main Menu  From this menu, the patient database can be accessed, new data can be collected, and old data can be reviewed.    To exit the program, click the Exit button.   Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  14  Accessing the Patient Database  To access the patient database, click on the Patient Database button.  The following menu is displayed.  The Patient Database stores all patients’ personal information, as well as their data results.   Figure 8.  Patient Database  In the patient database, the user can: Add a patient Update a patient’s information Remove a patient from the database   Adding a Patient  Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  15To add a patient, fill in the necessary information into the text boxes.  Once complete, click on the Add Patient button to submit the information.   Updating a Patient    To change a patient’s information in the database, click on the patient’s name on the database list.  Change the necessary data, and then click Update Patient.     Removing a Patient  Highlight a patient’s name on the Patient List.  Then click Remove Patient.  This will remove all the patient’s information in the database, as well as the patient’s recorded data.  To exit the Patient Database, click Exit.  System Setup    From the main screen click on the System Setup button to access the System Setup menu. Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  16The System Setup menu allow you to set up the proper ports for wireless and USB data acquisition between the ParkinSense user unit and the software package.  COM Port (Radio): This is the port required for wireless data acquisition via Bluetooth. Enter the port obtained during the ParkinSense User Unit setup here. It can also be detected automatically by doing a Scan.  Scan For Radio Ports  This function allows you to scan for COM Port (Radio) of a ParkinSense User Unit. Follow the following step to obtain a Radio port by scanning: 1.  Make sure ParkinSense user unit is on and unplugged from the USB data           cable. 2.  Click on the “Scan For Ports” button and the status indicator should change from blank to Scanning… 3.  When the status indicator displays “Done!,” use the pull down menu of  the COM Port (Radio) to select the Radio port. When one has more than one ParkinSense User Unit on, the scan will return the ports of all the devices turned on during the scan.   To exit System Setup and save the changes that you have made, click on OK.   Patient Test Mode System Setup allowed you to change the mode of operation of the software, from regular data acquisition to Patient Test operation. In Patient Test mode, the software will guide the patient through a series of activities and record those data to the patient’s profiles.  To switch to Patient Test mode, follow these steps:  1.  Make sure the proper COM Port (Radio) is setup for wireless data acquisition.   2.  Under Patient Test, check the box right next to “Test Mode Enabled”. 3.   Click on “Select Test Patient” to select a patient for which the test is perform on from the drop down menu. After you have selected a patient, click “Load Patient” button to continue. Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  17 4.  At the main screen of System Setup, click OK to save the changes you have made. 5.  You will now return to the main screen of the ParkinSense Software, hit the space bar to begin the Patient Test. On screen direction will guide you through a set of activities, while recording the result acquired from these activities in the patient ID folder.   6. To exit out of Patient Test operation mode, simply go back into System Setup and uncheck the checkbox  label as “Test Mode Enabled”.                             Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  18Collecting Data Collecting Data Wirelessly Using Bluetooth  Make sure that you have set up the proper port setting in System Setup before you start collecting data.   Click on the “Collect Data” button to open the data collection mode.    A window will pop up to Load a Patient.  Use the pull down menu to choose the correct patient, then click Load Patient.  Once the patient is loaded, the data screen appears.    Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  19Starting Data Acquisition  To begin data acquisition, ensure the ParkinSense arm device is switched to ON, and click Start Data on the tool bar.  When communication with the ParkinSense arm device is established, data will begin scrolling across the graph display.  Acquired Data Display  The graph area of the “Collect Data” window in ParkinSense Research ToolBox displays real-time data as it is acquired by the User Unit and transmitted to the PC.  By default, there are eight channels visible scaled at 10 seconds.   To change the axis scales and number of visible channels, click Channels/Time on the tool bar. Check the channels you would like to be visible, and uncheck those channels that you would like to hide. Chose the time scale most suitable for your application and click OK to save.    Saving Acquired Data      To begin saving the received data, click the Start Recording button at any time.  This will bring up a window called Enter File Name.  Enter a pertinent file name to save the patient’s data for this trial.  After entering a file name, click Start Saving.  This will begin the recording process.  Stopping Data Acquisition  Click the Stop Data button on the tool bar to end data acquisition. If data is recording, click Stop Recording to end recording.    Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  20Collecting Data From Memory  When the ParkinSense User Unit is on, the device immediately starts recording data to its internal memory. To download data from this memory at anytime, go through the following steps.   1. From the main screen, click on the Collect Data button to open the data collection mode.                      2. A window will pop up to Load a Patient.  Use the pull down menu to choose the correct patient and click Load Patient.  3. Once the patient is loaded, a data screen appears.     Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  21Make sure that you have setup the proper port setting in System Setup before you start downloading data from memory. Begin by turning on your ParkinSense User Unit. While the ParkinSense User Unit is on, plug the large end of the USB data cable into your computer’s USB port. Plug the small end of the USB cable into the User Unit.     Click the Download Memory button to start downloading the data from memory into your desktop. This will bring up a window called Enter File Name.  Enter in a pertinent file name to save the patient’s data for this trial.  After entering a file name, click Start Saving.  This will begin the transferring process.   While downloading from memory, you can observe the data being transferred in the graph area. When the computer has downloaded all the data from memory, a long beep will indicate the download process has finished.  An “Erase Data” window will appear asking if you want to erase the memory from the User Unit. To erase the memory, click “Yes,” otherwise click “No”. When you click “Yes” to erase the memory on the device, a series of beeps will indicate that the memory is being erased.  The deletion is complete when the beeps stop. Switching Data Modes  To switch back to wireless data acquisition from USB, you must turn off and remove the USB cable from the device. When you turn on the device again, you should hear a series of beep indicating the startup of the device and transition into wireless mode.   To switch to USB data aquisiton from wireless mode, you must first turn on the device. There will be a series of beeps during the start-up routine.  Wait for these to stop, then plug in the USB data cable into your device. You should hear two beeps after you plug in the cable. The beeps indicate that the device is switching from wireless mode to USB data mode. Erasing Memory  While you are in USB data mode, you can erase the memory at anytime by clicking on “Erase Memory” button on the tool bar menu. A window dialog appears to ensure you want to erase the memory. Clicking “Yes” will continue the process of erasing the memory. A series of beeps will follow, indicating the memory is erasing. The process is finished when the beep pattern stops. Clicking “No” will cancel the process. Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  22Reviewing Saved Data  Saved data can be reviewed at a later time by clicking the “Review Saved Data” button.  Clicking this will bring up a Load Patient window.  Use the pull down menu to find the patient’s name and then click Load Patient.  After clicking Load Patient, a second window appears prompting you to select a file.  Choose the file to review and click “Analyze.”    After  indicating  the  file  to  be  reviewed,  the  recorded data  will  appear for  analysis.  The  data display format that  you  are  observing  is  the  “Time  Graph”.  Use  the scroll bar to examine the data. The scroll bar will scroll across hours/minutes/seconds of the data you collected.     The data you collected can be display in three different formats by using the toolbar menu of Review Saved Data. The data can be display as time graph, frequency graph, and as power graph.  Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  23     To view the data you collected as frequency graph, simply click on “Frequency Graph” of the toolbar menu. There’s a brief wait period before the data can be display as frequency graph. The wait period depend on how much data is to be processed. For 12 hours of data, the expected wait time would be around 1.5 minutes to 3 minutes.  In “Frequency Graph” display format, you can scroll across hours/minutes/seconds to view the peak frequency of the subject’s movement over time.  The peak frequency is calculated over 30 second windows and plotted in the graph. The image display below is an example of what you would see when you are in “Frequency Graph” display option.    The data you collected can also be displayed as a Power Graph. To display the data collected as Power Graph, simply click on “Power Graph” on the toolbar menu. . There’s a brief wait period before the data can be display as Power Graph. The wait period depend on how much data is to be processed. For 12 hours of data, the expected wait time would be around 1.5 minutes to 3 minutes. An example of what you would see is shown below.    Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  24Once your data is processed for “Frequency Graph” or “Power Graph”, you can switch between the three display formats without processing the collected data again. You can use the toolbar menu to switch between Time Graph, Frequency Graph, and Power Graph. If you exit “Review Saved Window”, your data will have to be processed again for “Frequency Graph” and “Power Graph”, the next time you use these functions.  To exit Review Saved Data window, click on Exit on the toolbar.    Printing Current Graph You can print the current view of the graph display in both “Review Saved Data” and “Collect Data” window.  To print the current graph in “Collect Data” or “Review Saved Data”, simply click on “Print” on the toolbar menu.       A new window will appear upon clicking on “Print” called Page Setup. From Page Setup, you can select the paper size, margins, paper source, and orientation of the print. After you have selected the pint option, hit OK to continue.  After you have clicked OK from the previous window, a new window will appear for you to select a printer of your choice. The image below is an example of what you would see. After you have selected the desired printer to print the graph, click OK to proceed with the printing. To cancel the print process, click on Cancel of this window.             Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  25Graph Color and Scale Setup  The graph color and scale can be changed in both the Collect Data window and the Review Saved Data window.  To access this menu, click on the graph title button to the right of the graph.  This will bring up the Color and Scales window.    Different color and scale options are available.  Click on the desired color or scale number, and then click “OK” to return to the data window.        Channels / Time  The number of visible graphs and their scales can be changed in both the Collect Data window and the Review Saved Data window. To access this menu, click on the “Channels/Time” button on the toolbar. This will pull up the Channels/Time Scales window. Different visible channels and time scale options are available. The checked the channels will be visible.  Choose your desired visible channels and time scale. Click “OK” to save your changes and return to the graph window.              Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  26Miscellaneous Information:  Obtaining a Bluetooth port manually: 1.  Turn on the ParkinSense User Unit. A series of beeps indicates that the device is on and ready for wireless data transmission. 2.  Plug the Bluetooth Device into the USB port 3.  Go to your blue tooth Devices Manager by going to         Start  Control Panel  Bluetooth Devices                            4.  Double click on Bluetooth Devices to enter the Bluetooth Devices Manager. 5.  From here, click on COM Ports tab at the top.      6.  The system will do a scan and return a list of devices found, as well as the associated ports.  Your ParkinSense radio is found in any ports labeled “A7 Serial Port.”   Clevemed Proprietary Information
ParkinSense User’s Guide   v. 1.0  27          Clevemed Proprietary Information