Creative Industry IPC240 Video Baby Monitor User Manual IPC240 Baby Monitor

Shenzhen Creative Industry Co., Ltd. Video Baby Monitor IPC240 Baby Monitor

User manual

OT240M Baby Monitor   User Manual  Components  Built-in MP3+4G TF card with songs and stories; User-friendly menu; Two-way talkback to comfort baby. One User Manual One parent unit  One rechargeable Ni-MH battery for the parent unit  One baby unit  One power adapter  One USB connected adapter  One Pulse Oximeter (including one SpO2 probe) Two batteries for Pulse Oximeter  Settings  Step  1:  Connect  the  baby  unit  to  the  power  and insert  battery  pack  into  parent  unit  with  correct polarity (+/-) as indicated.
 Step 2: Toggle the key on the baby unit, press and hold  the  parent key until  the  green LED  is  on.  The parent  unit  is  now  powered  on  and  the  parent  unit will search for and link itself to the baby monitor.     Baby unit settings       1. Play the next lullaby        2. Pause/play 3. Play previous lullaby        4. Nightlight 5. Volume up                        6. Volume down 7. Power switch                    8. MP3 TF slot 9. USB power jack              10. Power indicator
11. Press to access menu settings, press and hold to power on/off. 12. Remote  control  the  camera  of  the  baby  unit (turns left/right or up/down). 13. Turn on/off LED Press and hold to talk with your baby. Tips:   To turn off  LCD  screen by pressing button 13 for  power  saving  and  more  restful  sleep  at night.   Talk  with  baby  by  pressing  and  holding  the button 14.   Press button 11 to open menu when parent unit and  baby  unit  turned  on.  Constant  LCD light mode  and  power  saving  mode  are  available  in second item.   Menu structure of parent unit   Voice & volume Screen always on or power saving   Brightness   Zoom  MP3   MP3    Temperature
  Power  Signal SpO2 Value     Pulse Rate Value 1. Volume settings Press menu button to access main menu when parent unit is turned on, choose this icon and then increase or  decrease  the  volume  of  parent  unit  pressing  the left or right direction keys. (Default is Mute) Constant LCD light/Power saving Press menu button to access main menu, choose this icon to set the display mode by pressing left or right direction keys. (The light under Power saving mode will  self-adjust  according  to  the  environmental sounds). 2. Brightness setting Press menu button to access main menu, choose this icon and  select  the  desired  brightness  level  (from 1~6) using the left or right direction keys. 3. Zoom Press menu button to access main menu, choose this icon to change the zoom x1 or x2 by pressing right direction key. 4. MP3 kid's learning Method  1:  Press  menu  button  of  parent  unit  to access  main  menu,  choose this icon  and  turn on/off MP3  by  pressing  right  direction  key  and  select desired lullaby left key. Method 2: To start MP3 press the bottom key on the baby unit.
Technical parameters  Model OT240M Modulation system FSK Frequency 2.4G Channel 19 Screen material 2.4” LCD SpO2 Range 35% - 100% SpO2 Alarm Low limit 90% Pulse Rate Range 30bpm – 240bpm  Pulse Rate Alarm High limit 160bpm Low limit 70bpm Baby unit battery Dc 6v/2a Parent unit sensitivity Aa1800mA 3.6V NI-MH Parent unit adapter ≤85dBm Parent unit battery DCV / 500mA Battery cycle life Aa1800mA 3.6V NI-MH Battery charging time 14 Hours Operating 12 Hours Temperature range 32-122F (0-50C) Open area range 820ft(250m) Lullaby Mp3 +4g TF
Technical features    Low power indication: Should battery voltage fall below 3.5V, parent unit  will alarm and  will  shut off if below 3.0V.  Squelch:      automatic      filtering      environment background noise.   Automatic  sound  trigger:  If  the  sound  exceeds 50dB, the baby unit will transmit sound so as to prevent parent from noise interference.   Alarming:  A  10  second “beep-beep”  alarm  will sound  to  alert  parent  when  the  distance  between the parent unit and baby unit has been exceeded; When  the  measured  SpO2  or  Pulse Rate  value exceeds  the  preset  alarm  limit,  then  the  parent unit  will  activate  the  audible  and  visual  alarm indications.   Automatic  channel  search:  the  parent  unit  will search the  proper channel to  connect to the baby monitor.   Sweet   nightlight:  A   nightlight   switch   at   the bottom  of  baby  unit  provides  convenience  for baby care at night.   Infrared night vision: Detect images in the dark. (changes to black/white).   Charging indicator: power indicator flashing.   MP3  player:   built-in  MP3  with   lullabies   and children  stories  in  baby  unit,  which can  be operated by parent unit. Additional lullabies may be uploaded. Cautions:   This  product  is  intended  as  an  aid.  It  is  not  a
substitute  for  responsible  and  proper  audit supervision and not intended for use as a medical monitor.   Please do not use power adapters that may not fit this product configuration.   Please do not expose the baby monitor to extreme cold, heat, wet or direct sunlight.   Make sure the baby unit and the adapter cord are always out of reach of the baby, at least 3 feet/1 meter  away  from  the  baby  or  crib,  to  avoid  a possible strangulation hazard.   Never cover  the  baby  unit  or  parent  unit  with anything such as a towel or a blanket. Clean units only when when powered off.  Troubleshooting  Problem: Power indicator light not working. Possible cause: Does  not  switch  on;  Power  is  not getting to the unit; Low batteries. Solution: Turn on the units; Make sure the units are connected   to   the    power   supply;   Replace   the batteries. Problem: Power indicator flashing. Possible  cause:  Not  connected  well  with  adapter; Low batteries. Solution:  Make  sure  well  connected;  Replace  the batteries. Problem: Parent unit makes loud noises. Possible cause: Units are too close together; Volume dial on parent unit may be too high. Solution: Move the parent unit at least 3 feet/1meter
away from the baby unit; Turn down volume on the parent unit. Problem: Weak sound or mute coming from parent unit. Possible cause: Volume on parent unit or set too low or muted. Solution: Turn up the units closer to each other. Problem: Units may be too far from each other. Possible cause: No signals between units. Solution: Move the units closer to each other.  Instructions for Pulse Oximeter  Connection    Insert  the  sensor  tips  into  the  slots  on  the  probe holders. Place the sensor tip labeled “LED” into the stationary probe holder on the silicone wrap belt, and place  the  sensor  tip  labeled  “PD”  into  the  sliding probe holder, as illustrated in Figure 1 and Figure 2.         Figure 1 Probe tip and holder
  Figure 2 Y-type SpO2 Sensor with Rubber Wrap  NOTE: Make sure that the silicone button seat fully into the hole in the back of the probe holder. Wrap  the  oximeter  on  the  leg  near  ankle  with  its belts and position the front panel  to a proper place for convenient operation, as shown in Figure 4. Wrap up  the  Y-type  sensor  onto  the  baby’s  foot  by the rubber wrapper,  as illustrated  in  Figure  3. Make sure  the  alignment  of  the  light-emitting  tip  (LED) and the photo- detecting tip (PD) for aiming at each other,  then  thread  the  end  of  wrap  belt  through  the outside  slots  on  the  probe  holders  and  tighten  the rubber belt with proper force. The final placement is shown in Figure 4.         Figure 3 Sensor wrapping up
  Y-type  SpO2 Sensor  with rubber wrapper   Pulse Oximeter Figure 4 Placement of the Oximeter and the Sensor  It's  recommended  that  wrapping  up  the  sensor  onto the  baby's  first  toe  if  the  foot  is  too  thick  to  be tighten, as shown in figure 5.            Figure 5  IMPORTANT:  The  sensor  connection  is  critical  for the signal strength and quality. Try to make LED and PD  facing  each  other  in  opposite  side,  so  that  the light  beam  is  as  vertically  transmitted as  possible and the light path is as short as possible.
Instructions for Operation  After all connection  are  completed,  press the Power Button ( ) for 3 seconds to power on the oximeter, referring  to  below  figure.  When  the  oximeter  is powered  up,  the  Power  Indicator  (yellow  lamp    ) will light constantly. If the heart beat is detected, the Heartbeat  Indicator  (red  lamp )  flashes  according with  the  heart  rate.  At  the  same  time  the  oximeter sends  the  measured  data  (SpO2  and  PR)  to  Baby Monitor  in  a  given  time  interval  via  wireless communication, and the SpO2  and PR value will be displayed on the terminal side of Baby Monitor.   NOTES: Appearance of the Pulse Oximeter 1. The oximeter has no automatic power-off function. Press the Power Button for 3 seconds to turn it off after the end of each use. 2. If the Power Indicator flashes it means the battery voltage is low for the oximeter, please replace the batteries.
          Distributed by:  eH4H LLC  1200  Abernathy  Road,  Suite  1700,  Atlanta, GA 30328 Website: www.Dormebaby.com  Tel: 888-741-2224 ext. 2  Fax: 832-442-4000  A division of Tele-H-for H Holdings, LLC  FCC Caution.   §15.19 Labelling requirements. This device complies with part 15 of the FCC Rules.  Operation  is  subject  to  the  following
two conditions: (1) This device may not cause harmful interference, and (2) this device must accept  any  interference  received,  including interference  that  may  cause  undesired operation.   §15.21 Information to user. Any  Changes  or  modifications  not  expressly approved  by  the  party  responsible  for compliance  could  void  the  user's  authority to operate the equipment.   §15.105 Information to the user. Note:  This  equipment  has  been  tested  and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules.  These  limits  are  designed  to  provide reasonable  protection  against  harmful interference  in  a  residential  installation.  This equipment   generates   uses   and   can   radiate radio  frequency  energy  and,  if  not  installed and  used  in  accordance  with  the  instructions, may  cause  harmful  interference  to  radio communications.  However,  there is  no guarantee that interference will not occur in a
particular  installation.  If  this  equipment  does cause  harmful  interference  to  radio  or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures: -Reorient or relocate the receiving antenna. -Increase    the    separation    between    the equipment and receiver. -Connect  the  equipment  into  an  outlet  on  a circuit   different   from   that   to   which   the receiver is connected. -Consult    the    dealer    or    an    experienced radio/TV technician for help. *RF warning for Mobile device: This equipment complies with FCC radiation exposure limits set forth for an uncontrolled environment. This equipment sh ould be installed and operated with minimum distance 20cm between the radiator & your body.