Data Sciences XC1000 External Programmer for Implant User Manual

Data Sciences International Inc External Programmer for Implant

users manual

Download: Data Sciences XC1000 External Programmer for Implant User Manual
Mirror Download [FCC.gov]Data Sciences XC1000 External Programmer for Implant User Manual
Document ID521138
Application IDadqDT10ieAWZomtn/JgFFw==
Document Descriptionusers manual
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize49.12kB (614045 bits)
Date Submitted2005-03-07 00:00:00
Date Available2005-03-07 00:00:00
Creation Date2005-03-04 16:06:20
Producing SoftwareAcrobat Distiller 6.0.1 (Windows)
Document Lastmod2005-03-04 16:10:01
Document Titleuntitled
Document CreatorAcrobat PDFMaker 6.0 for Word

LVP Programming System
User Guide
Acknowledgements
Dataquest®, the DSI Logo, and PhysioTel® are registered trademarks of Transoma
Medical in the United States and certain other foreign countries.
A.R.T™ and Multiplus™ are trademarks of Transoma Medical.
LVP Program Utility Version 1.0
Copyright© Transoma Medical 2005
All Rights Reserved
Printed in the U.S.A.
Reorder Number
001710-001
DSI, A Division of Transoma Medical
4211 Lexington Ave North • St. Paul, MN 55126
Telephone: 1(651) 481-7400 • 1-800-262-9682
Fax: 1(651) 481-7404
Web Site: www.datasci.com
Table Of Contents
The LVP Programming System......................................................................................... 1
About the LVP Programming System............................................................................ 1
About the LVP-1000 ...................................................................................................... 2
Features..................................................................................................................... 3
Specifications............................................................................................................. 3
About the Data Monitor.................................................................................................. 4
About the Remote Transceiver...................................................................................... 5
Upgrading Firmware ...................................................................................................... 6
Getting Started .................................................................................................................. 8
Computer Hardware and Software Requirements......................................................... 8
Installing the LVP Program Utility .................................................................................. 8
Programming the LVP-1000............................................................................................ 14
Using the Program Wizard .......................................................................................... 14
Choosing the Transmission Schedule ......................................................................... 14
Programming the LVP-1000 ........................................................................................ 15
Using the Log Viewer ...................................................................................................... 25
Dataquest A.R.T. and the LVP-1000............................................................................... 27
Configuration ............................................................................................................... 27
Configuring a Receiver ............................................................................................ 27
Configuring an LVP-1000......................................................................................... 28
Completing the Configuration .................................................................................. 32
Data Collection ............................................................................................................ 33
Selecting a Sampling Protocol in Dataquest A.R.T.................................................. 33
Defining the Size of Data Files................................................................................. 34
Selecting Data Types to Collect............................................................................... 35
Starting Duty-Cycled Sampling ................................................................................ 37
Stopping Duty-Cycled Sampling .............................................................................. 39
Starting Continuous Sampling ................................................................................. 39
Stopping Continuous Sampling................................................................................ 40
Starting Scheduled Sampling................................................................................... 40
Stopping Scheduled Sampling................................................................................. 41
Receivers..................................................................................................................... 41
RMC-433 Specifications .......................................................................................... 41
RMC-916 Specifications .......................................................................................... 42
Left Ventricular Pressure ............................................................................................. 43
Obtaining DSI Technical Services................................................................................... 45
Error Messages............................................................................................................... 47
Glossary .......................................................................................................................... 51
Index ............................................................................................................................... 53
This guide provides information on the intended use of the DSI LVP-1000 Telemetry
Transmitter and the LVP Programming System. It also contains information relating to
the requirements of 1999/5/EC (R&TTE Directive), FCC compliance and transmitter
disposal. Please read this guide carefully and retain for future reference.
Intended Use:
The DSI LVP-1000 Telemetry Transmitter is intended for measuring physiologic
parameters in laboratory animals. This device measures left ventricular pressure and
temperature and should be placed subcutaneously, no more than 1.5 inches from the
surface of the skin. The LVP Programming System is used to set the transmission
parameters of LVP-1000 devices.
Declaration of Conformity:
LVP-1000-433 Transmitters and the XC-1000 (Remote Transceiver) conform to the
requirements of 1999/5/EC (R&TTE Directive). The declaration of conformity may be
viewed on the Internet at www.datasci.com.
Use of these devices may be subject to licensing in the following countries. For licensing
options please contact the national spectrum management authorities
Austria
Belgium
Denmark
France
Greece
Iceland
Ireland
Italy
Liechtenstein
Luxembourg
The Netherlands
Norway
Portugal
Spain
Switzerland
FCC Statement:
LVP-1000-916 TX FCC ID MHALVP916
XC-1000 TX FCC ID MHAXC1000
IC: IC:5681A-XC1000
These devices comply with part 15 of the FCC rules and RSS210 of the Canadian rules..
Operation is subject to the following two conditions: (1) This device may not cause
harmful interference and (2) this device must accept any interference that may cause
undesired operation.
Caution:
Changes or modifications not expressly approved by the party responsible for compliance could
void the users authority to operate the equipment.
Transmitter Disposal:
Transmitters contain a primary battery cell. The battery cell may be removed by
destructively dismantling the transmitter. Recycle the battery cell and remaining
electronics in accordance with applicable federal, state, and local regulations.
The LVP Programming System
The LVP Programming System
About the LVP Programming System
The LVP Programming System encompasses the software and hardware components
needed to program the LVP-1000 transmitter. The hardware components are connected
to a computer that has the LVP Program Utility installed.
Remote
Transceiver
Data Monitor
Computer
with LVP
Program
Utility
The LVP Programming System consists of the following components:
LVP Program Utility
The LVP Program Utility is a simple software program installed on a computer. The LVP
Program Utility allows you to read and change the data transmission schedule for your
LVP-1000 transmitters. It also keeps a log of all of the LVP-1000s and the configurations
you have programmed.
Remote Transceiver - Model XC-1000
The Remote Transceiver sends and receives programming signals to and from the
implanted transmitter.
Data Monitor - Model DM-1000
The Data Monitor converts the programming signals received from the Remote
Transceiver into a digital format and provides feedback about system operation via the
front panel display. The Data Monitor is connected to a serial port on the computer.
The LVP Programming System
Data Monitor Power Supply
The Data Monitor power supply provides power to the Data Monitor and Remote
Transceiver.
Magnet
The magnet manually activates the LVP-1000 transmitter during programming. It
also can be used to manually initiate a data transmission to the Dataquest A.R.T.
System. Keep the magnet in a convenient storage location so it is readily
accessible and to prevent inadvertent activation of the LVP-1000.
About the LVP-1000
The LVP-1000 measures left ventricular (LV) pressure and temperature from laboratory
animals. The LVP-1000 was originally developed for human clinical use, and provides
high fidelity left ventricular pressure measurements. It is ideal for cardiac-specific
applications such as congestive heart failure models where highly accurate parameters
of heart function are required.
Remote Sensor
Assembly (RSA)
Transmitter
Multi-Conductor Leads
There are 2 models of this transmitter, the LVP-1000-433, which is sold for use in
Europe; and the LVP-1000-916, which is sold for use in the United States and Canada.
Both models are used and function in the same manner, except for the radio
transmission frequency. In DSI manuals and product literature, the shortened model
name, LVP-1000, refers to both models.
The LVP-1000 is the first in a series of future generation transmitters providing
automated configuration, confirmation of transmitter ID, and programmed Duty-Cycling
The LVP Programming System
to extend battery life. The LVP-1000 uses an extremely stable pressure sensor located
in the Remote Sensor Assembly (RSA) to minimize pressure drift over extended implant
durations. The RSA is located separate from the Transmitter to minimize the effects of
fluid head pressure. The RSA and Transmitter are permanently connected via two MultiConductor Leads.
Features
•
•
•
•
•
•
•
Hermetic packaging for extended implant life
Very low pressure drift
"Smart" transmitter identification
• Automated configuration
• Positive unique identification during sampling
High fidelity pressure waveforms
Separate transmission frequency allows simultaneous use with a PhysioTel
transmitter model in a single animal
Compatible with Dataquest A.R.T.™ 3.0 and later versions
Measures LV pressure and body temperature
Specifications
Frequency Response
Initial Accuracy
Maximum Pressure Drift
Temperature Accuracy
Temperature Drift
Duty-cycle battery life
Continuous Battery Life
100Hz
+/- 3 mmHg
1 mmHg/month
+/- 0.5 °C
0.1 °C/month
> 2 years - depends on usage
6 weeks
The LVP Programming System
About the Data Monitor
The Data Monitor, DM-1000, is a component of the LVP Programming System's
hardware. The Data Monitor is connected to a serial port on the computer. It must also
be connected to the power supply provided, which must be plugged into an electrical
outlet. The Remote Transceiver, XC-1000, is connected to the Data Monitor.
Front of the Data Monitor
Back of the Data Monitor
Status Lights
The front of the Data Monitor has several status lights and a Transfer button. In general,
these lights are not used with the LVP Programming system. Here is a brief explanation
of these status lights.
• The 'Ready' light will be solidly illuminated when the LVP Program Utility is in
use.
• At all other times, the 'Ready' light will flash when the Data Monitor is receiving
power.
• All other lights may be illuminated or flash temporarily when the Data Monitor is
first powered up and when the LVP Program Utility is opened. These flashes do
not indicate any problem or condition.
• The 'Transfer' button does not perform any function in the LVP Programming
System.
Note
•
You will not be able to use the Program Wizard of the LVP Program Utility
if all of the LVP Programming System's hardware is not connected to the
computer.
•
Always disconnect the Data Monitor from your computer or disconnect
the Data Monitor power supply before starting your computer. Reconnect
the hardware after the computer fully boots. If it is connected when your
computer boots, the Windows operating system may mistakenly
recognize it as mouse. You will notice your mouse pointer moving
randomly around the screen, not under the control of your mouse. If this
occurs, simply disconnect the Data Monitor.
The LVP Programming System
About the Remote Transceiver
The Remote Transceiver, XC-1000, is a component of the LVP Programming System's
hardware.
The Remote Transceiver is connected to the back of the Data Monitor. It is used to send
and receive programming signals from the LVP-1000 during programming.
The Remote Transceiver has a red-colored circle on the front label. When programming,
hold the circle directly over the LVP-1000 transmitter. (see figure below).
The LVP Programming System
The Remote Transceiver should be held within 2 inches of the LVP-1000 during
programming. To program, you will first activate the LVP-1000 transmitter by passing the
magnet over it. Pass the magnet and then move the Remote Transceiver into place
within 2 seconds to start receiving the signals.
Upgrading Firmware
Occasionally, DSI may recommend that you upgrade the firmware of your Data
Monitor to improve its function. If this occurs, DSI will provide you with a
computer file and instructions for installing the firmware. The instructions
provided below are a general outline of the procedure.
There are 2 sets of instructions here, depending on how the firmware file is
provided, either on CD or electronically via email or the DSI website.
If the Firmware file is provided on a CD.
1. Insert the CD containing the firmware file into the computer's CD-ROM drive.
2. Start the LVP Program Utility.
3. From the toolbar, select Actions --> Upgrade Firmware.... The Upgrade Data
Monitor Firmware dialog appears.
4. Click the Browse... button to locate the CD-ROM drive and the upgrade file.
The upgrade file type will be *.s19.
5. Click Open. The file location and extension will appear in the File field.
6. Click OK. The Status Area of the Main Screen will show a message stating
"Applying Upgrade... Please wait..." This process may take a few minutes.
7. The Status Area will display the message "Upgrade completed successfully."
when finished.
8. You may then use the LVP Program Utility as normal.
The LVP Programming System
If the Firmware file is provided via email or downloaded from the DSI website.
1. Save the Firmware file on the computer.
2. Start the LVP Program Utility.
3. From the toolbar, select Actions --> Upgrade Firmware.... The Upgrade Data
Monitor Firmware dialog appears.
4. Click the Browse... button to navigate to where the firmware file was saved.
The firmware file type will be *.s19.
5. Click Open. The file location and extension will appear in the File field.
6. Click OK. The Status Area of the Main Screen will show a messages stating
"Applying Upgrade... Please wait..." This process may take a few minutes.
7. The Status Area will display the message "Upgrade completed successfully."
when finished.
8. You may then use the LVP Program Utility as normal.
Getting Started
Getting Started
Computer Hardware and Software Requirements
The LVP Programming System is designed for use with a personal computer.
The LVP Program Utility will run on any computer system that meets the
minimum requirements for the Windows 2000 or Windows XP operating systems
and has a serial port. Installing the LVP Program Utility on a computer that has
Dataquest A.R.T. or OpenART installed will ensure that these requirements are
met.
At a minimum, a computer must have:
•
•
•
•
•
Windows 2000 (Service Pack 2) or Windows XP Professional
One serial port available for the Data Monitor connection
A hard drive with 7 megabytes of hard drive storage to install the LVP Program
Utility
A CD-ROM drive
128 MB RAM
Installing the LVP Program Utility
Follow this procedure to install the LVP Program Utility software.
Assumptions
The computer meets the minimum requirements as specified in Computer Hardware and
Software Requirements.
Before Installation
1. Log in to the computer as a system administrator.
2. Ensure that the time zone is properly set and that Windows is set to
automatically adjust clock for daylight saving changes.
To install the software
1. Insert the LVP Program Utility CD into the CD-ROM drive of your computer.
If your CD-ROM drive is configured to run automatically, the InstallShield Preparing Setup window appears. Otherwise, double click on the My
Computer icon on the desktop, double click on the CD-ROM drive, and double
click on Setup.exe to start the InstallShield Wizard.
Getting Started
The Welcome window appears.
Getting Started
2. Click Next > to display the Software License Agreement window.
3. Read the License Agreement and then select “I accept the terms of the license
agreement" to accept. Click Next > to continue.
10
Getting Started
4. The Choose Destination Location window appears. To change the installation
location of the LVP Program Utility files, click Change... and use the browser to
change the folder. (Use the default setting if possible.) Click Next > to continue
to the next window.
5. The Ready to Install the Program window appears. Click Install to begin the
installation.
11
Getting Started
6. The Setup Status window appears and shows the file copy status.
7. The InstallShield Wizard Complete window appears when the software is
installed. Click Finish to complete the setup.
12
Getting Started
Note
•
•
•
Consult your IT staff or call DSI Technical Services for assistance with the
installation procedure.
You may install the LVP Program Utility on a computer that already has
Dataquest A.R.T. installed.
The computer’s Windows operating system must be set to automatically adjust
the clock for daylight saving changes to install and operate the LVP
Programming Utility. If you are using a computer with Dataquest A.R.T. 3.0 also
installed, you will need to turn this setting off to operate the Dataquest A.R.T.
software.
13
Programming the LVP-1000
Programming the LVP-1000
Using the Program Wizard
The Program Wizard allows you to define the LVP-1000 transmission schedule and
settings.
Consider the following when using the Program Wizard:
• You must have the LVP Programming System hardware connected to your
computer to start the Program Wizard. The Data Monitor, Remote Transceiver,
and power supply must all be correctly connected.
• Animals may be implanted with both an LVP-1000 and another DSI transmitter
model, such as a D70-PCT. In this case, turn off the second transmitter before
attempting to program the LVP-1000.
• Do not try to collect data from the LVP-1000 using Dataquest A.R.T.™ during
programming. Before starting the LVP Program Utility, stop sampling from all
LVP-1000 devices in Dataquest A.R.T.™
• Determine how you wish to program the LVP-1000 for your studies, depending
on how you wish to collect data. See Choosing the Transmission Schedule for
more information on the differences between the transmission schedules.
• The LVP-1000 will retain this transmission schedule until it is reprogrammed with
different settings.
Choosing the Transmission Schedule
The LVP-1000 can be programmed for one of three modes of transmission: Manual
Transmission, Continuous Transmission, or Automatic (duty-cycle) Transmission. These
each have different benefits and uses.
Manual Transmission
•
•
•
•
•
•
The LVP-1000 must be activated with a magnet to collect a data transmission.
The LVP-1000 is OFF unless activated.
The transmission length is programmed to a discreet length.
The LVP-1000 will turn OFF when the transmission is complete.
Use this mode when the animal is not actively being studied.
Use the Duty-cycle sampling mode in Dataquest A.R.T.
Continuous Transmission
•
•
•
•
•
•
•
14
The LVP-1000 can continuously transmit data in this mode.
After programming, the LVP-1000 is turned OFF. Turn it ON by activating with
the magnet.
When ON, the LVP-1000 will transmit until you use a magnet to turn it OFF. (This
is similar to other DSI transmitters.) When OFF the LVP-1000 will not transmit
until you use a magnet to turn it ON.
A transmission length is not selected.
The LVP-1000 must be reprogrammed to stop data transmissions.
Use either Continuous or Scheduled sampling modes in Dataquest A.R.T.
This mode uses the most battery power: the battery will last 6 weeks in this
mode.
Programming the LVP-1000
Automatic (Duty-Cycle) Transmission
•
•
•
•
•
•
•
The LVP-1000 transmits data according to the programmed schedule.
The LVP-1000 automatically turns ON and transmits for the programmed
transmission length. It is OFF at all other times.
A magnet can be used to activate the LVP-1000 to collect a transmission when it
is OFF.
The LVP-1000 can be programmed to transmit only during certain parts of the
day, using a Start and End time; or to transmit periodically 24 hours a day.
The transmission length is programmed to a discreet length. All transmissions
will be this length.
Use the Duty-cycle sampling mode in Dataquest A.R.T.
The battery life will depend on the transmission schedule. The battery can last for
over 2 years in this mode.
Programming the LVP-1000
The Program Wizard allows you to set the LVP-1000 transmission schedule and
configuration settings. The Program Wizard guides you through the following
programming procedures.
1. Starting the Wizard and reading the current configuration
2. Enabling automatic LVP-1000 transmissions
3. Reviewing the updated configuration settings and programming the LVP-1000
To start the Program Wizard and read the configuration
1. From the Main Window, click on the Program icon or select Actions ->
Program .... The LVP Program Utility will verify that the Data Monitor is
connected and functioning properly. The Program Wizard window appears.
15
Programming the LVP-1000
Note: The Program Wizard Window contains the following warning:
WARNING: An LVP-1000 programmed using this LVP Program Utility software
must not be used with the LVP-1000 Pressure Monitoring System.
Explanation: The LVP-1000 Pressure Monitoring System is a product of
Transoma Medical and is designed for human clinical use. Do not use the LVP
Program Utility to program LVP-1000 devices if you are collecting data using
the clinical LVP-1000 Pressure Monitoring System because data may not be
collected properly.
The LVP Program Utility should only be used to program LVP-1000 devices that
are implanted in research animals and used in conjunction with Dataquest
A.R.T. or Dataquest OpenART for data collection.
16
Programming the LVP-1000
2. Click Next >. The Read the LVP-1000 Configuration window appears
3. Hold the Remote Transceiver within 2 inches of the LVP-1000 transmitter.
4. Pass the magnet over the transmitter to start reading the configuration. The
Program Wizard reads the current LVP-1000 configuration and displays the
message, LVP-1000 configuration reading complete.
5. Remove the Remote Transceiver from the transmitter.
6. Click Next > to view any status or warning messages.
17
Programming the LVP-1000
7. On the View Configuration Status window, note all status messages. Normally,
the message "No special conditions were detected." will appear.
•
•
18
If the configuration status messages are acceptable, click Next > to
continue programming the LVP-1000.
If any of the messages are unacceptable, click Cancel to leave the
Program Wizard and locate another LVP-1000, if appropriate. Refer to
Error Messages and/or contact Technical Services if you need
assistance.
Programming the LVP-1000
8. The View Initial Configuration window shows the current program configuration
of the LVP-1000. Click Next > to begin changing these settings.
19
Programming the LVP-1000
To enable LVP-1000 transmissions
1. From the Set LVP-1000 Transmission window, choose one of the following
actions:
•
•
20
Select a transmission length from the Transmission Length field.
Proceed to Step 2.
Place a check mark next to Continuous transmission. This will disable
the Transmission Length field. Click Next >. The Program Wizard
displays the View Updated Configuration window. Continue with Step 1 of
the To view the updated configuration and program the LVP-1000
procedure.
Programming the LVP-1000
2. Click Next >. The Enable Automatic Transmission window appears.
3. Take one of the following actions:
•
Place a check mark next to Enable automatic transmissions to
schedule transmissions at a particular time of the day. Proceed to Step 6.
• If you do not want to enable automatic transmissions, clear the Enable
automatic transmissions checkbox. The rest of the fields in this window
become disabled.
Click Next >. The Program Wizard displays the View Updated
Configuration window. Go to Step 1 of the To view the updated
configuration and program the LVP-1000 procedure to continue
programming the LVP-1000.
Note: If automatic transmissions are not enabled, you must pass a magnet
over the LVP-1000 each time you need to begin a transmission.
6. Select a transmission start and end time from the Start and End transmission
time of day fields on the Enable Automatic Transmission window. During this
time period, the LVP-1000 will transmit data to the system at defined intervals. If
24-hour monitoring is desired, set these two fields to the same time.
7. Select an interval between LVP-1000 transmissions from the Interval between
transmissions box. The interval cannot be larger than the time span between
the Start and End transmission times selected in the previous step.
21
Programming the LVP-1000
8. Click Next > to accept all automatic transmission settings and view the updated
configuration settings.
To view the updated configuration and program the LVP-1000
1. From the View Updated Configuration window, examine the updated
configuration. If any of the listed settings are unacceptable, use the Back button
to return to the appropriate window to make changes.
22
Programming the LVP-1000
2. Click Next > to begin the programming process. The Start LVP-1000
Programming window appears.
3. Hold the Remote Transceiver within 2 inches of the LVP-1000 transmitter.
4. Pass the magnet over the transmitter so the LVP Program Utility can begin
programming the new configuration. The LVP Program Utility programs the
LVP-1000 with the settings defined in the updated configuration and displays
the message, LVP-1000 programming complete when complete.
5. Remove the Remote Transceiver from the transmitter.
6. Click Finish to complete the programming process and to return to the LVP
Program Utility Main window.
23
Programming the LVP-1000
Note
•
•
•
24
The time zone an LVP-1000 Transmitter will use for automatic transmissions
(duty-cycle transmissions) is the time zone setting of the computer used for
programming.
The LVP-1000 Transmitter does not automatically adjust for Daylight Savings
Time (DST, also known as Summer Time). The automated transmission
schedule is based on the Daylight Savings status of the computer when
programmed. Transmissions will be offset by one hour when the computer clock
is adjusted for DST. For example, if the LVP-1000 is programmed to transmit at
3:00 AM during Standard Time (Winter Time), the transmission will occur at 4:00
AM during DST. Adjust the LVP-1000 transmission schedule by reprogramming.
The LVP-1000 cannot transmit data until it has been programmed.
Using the Log Viewer
Using the Log Viewer
Use the Log Viewer to review the LVP-1000s you have programmed, as well as the
configurations programmed.
To locate this dialog select Actions --> View the Log ... or click the Log Viewer
Icon.
The Log Viewer opens in an Internet Explorer Window. You must select the LVP1000 Serial Number of the LVP-1000 log you wish to view, or you can view all
Serial Numbers (All).
25
Using the Log Viewer
After making a Serial Number selection, the log will be displayed. The log entries
are listed in chronological order with the most recent entry at the top of the list
and the earliest entry at the bottom of the list.
26
Dataquest A.R.T. and the LVP-1000
Dataquest A.R.T. and the LVP-1000
The LVP-1000 is designed to be used with Dataquest A.R.T. and/or Dataquest
OpenART. The following sections of the manual show how to configure an RMC-433 or
RMC-916 receiver and an LVP-1000, as well as how to acquire data using Duty-cycle
sampling in Dataquest A.R.T.
This information is also available in the Dataquest A.R.T. and OpenART manuals. This
manual assumes that the user is generally familiar with Dataquest A.R.T. or Dataquest
OpenART. If you are a new user of one of these software packages, it is recommended
that you read the manuals included with them for information that is not included here.
LVP-1000 Features in Dataquest A.R.T.
Compared to PhysioTel transmitters, the LVP-1000 has some new and exciting features:
• Automated Configuration - The serial number and calibration values are stored in
the LVP-1000. During configuration, you simply place a receiver near the LVP1000 to collect this data. See Using the New LVP-1000 Wizard for details.
• Duty-cycle Sampling - The LVP-1000 can be programmed to transmit on a
regular schedule to prolong battery life.
• Different Transmission Frequency - The LVP-1000 can be implanted along with a
PhysioTel transmitter in a single animal to maximize the amount of data
collected.
Configuration
Configuring a Receiver
To configure an RMC-433 or RMC-916
1. Connect the RMC-433, or an RMC-916 to a Data Exchange Matrix.
2. Select Hardware - Verify from the Configuration window. The Hardware
Verification Errors window appears stating "Differences that Configuration can
resolve automatically."
27
Dataquest A.R.T. and the LVP-1000
3. Click Accept to have the program add the receiver model and its serial number
to the configuration tree. The New Transmitter window appears.
4. Click Yes to add a new transmitter and continue with Configuring an LVP-1000.
Configuring an LVP-1000
Assumption
You have configured your Data Exchange Matrix and an RMC-433 or RMC-916.
1. Highlight the receiver you would like to assign the transmitter to, right click and
choose New Transmitter. The New LVP-1000 Wizard dialog will appear.
28
Dataquest A.R.T. and the LVP-1000
2. Use a magnet to activate the LVP-1000 transmitter.
3. Physically place the LVP-1000 transmitter on or near an RMC-433 or RMC-916
receiver.
4. Once the transmitter signal is detected, the New LVP-1000 Wizard - Step 1
window will be replaced with the New LVP-1000 Wizard - Step 2.
5. Confirm the transmitter model and serial number. The serial number must be
unique.
29
Dataquest A.R.T. and the LVP-1000
6. Click Next > to continue. The New LVP-1000 Wizard - Step 3 window appears.
7. Select the species of your animal to ensure proper algorithm selection. A default
species automatically appears based on the transmitter that you selected.
If your species is not available, choose a species from the drop down list whose
heart rate is similar to the expected heart rate of your animal. When debating
between two species with different heart rates, DSI recommends selecting the
species that has a higher heart rate than the expected heart rate for your
animal. This ensures that the processing algorithm will not 'miss' any trends.
8. Enter a unique animal ID of your choice.
9. Click Finish to complete the transmitter configuration process and return to the
Configuration window.
30
Dataquest A.R.T. and the LVP-1000
10. Configure additional transmitters by repeating steps 1-9. After configuring all
transmitters, continue with Completing the Configuration.
Note
•
•
•
•
Factory calibration values are automatically detected by the software.
After configuring an LVP-1000 transmitter, you can move it to the Inventory and
use the same receiver to configure another LVP-1000 transmitter.
To move a configured transmitter to Inventory, highlight it, right click, and choose
Move to Inventory.
Transmitters in Inventory retain their calibration values. Click and drag a
transmitter out of Inventory to a receiver when you wish to collect data from the
transmitter.
31
Dataquest A.R.T. and the LVP-1000
Completing the Configuration
Your final configuration step is to verify the hardware configuration.
To verify your hardware configuration
1. From the Acquisition program, select Hardware - Verify to make sure the
hardware and configuration match. You should receive a message stating
"Hardware and settings verified successfully."
2. Click OK.
3. If any changes were made or errors are detected, the Hardware Verification
Errors window appears. The section entitled "Differences that Configuration can
resolve automatically" contains connecting or disconnecting messages about a
receiver, Data Exchange Matrix, or converter.
4. Click Accept to have the Configuration module resolve these differences.
5. If there are errors listed in the "Differences requiring user action" section,
choose Edit to return to the Configuration window and correct the configuration.
The messages listed in this section require user interaction. For example if
calibrations were not entered for a transmitter, a message about adjusting
calibrations would appear.
32
Dataquest A.R.T. and the LVP-1000
6. Before running Hardware - Verify again, be sure to resolve any of the user
action differences. Once all of the differences have been resolved, a message
window appears confirming hardware and settings verification. The software will
add any changes to the configuration tree.
7. Exit Configuration.
Note
•
•
•
•
•
This action automatically identifies new generation hardware (Data Exchange
Matrices, APR-1, RPC-1, RMC-1, RMC-1HF, RMC-433, and RMC-916). It will
also check for signal calibrations.
The Hardware - Verify action runs automatically the very first time you open a
configuration and can perform much of the maintenance by recognizing the
hardware connection locations.
If a Data Exchange Matrix or new generation receiver is moved or disconnected,
all of the associated configuration items are moved or disconnected with it.
When new generation hardware is disconnected, all devices and associated
configuration information in the database will be moved into the Recycle Bin
located on the Configuration window for retrieval.
Old generation hardware is not automatically moved.
Data Collection*
Selecting a Sampling Protocol in Dataquest A.R.T.
There are two ways to sample data: continuously or on a schedule. Both types of
sampling require an understanding of sampling slots - see your Dataquest A.R.T.
manual for more information on sampling slots.
Each type of sampling method is described here. For more information on a specific step
see the following headings.
Duty-Cycle Sampling
Duty-cycled sampling is for use with LVP-1000 transmitters. However, additional
PhysioTel transmitters can also be sampled using duty-cycle sampling if they are
grouped with an LVP-1000 transmitter. Each duty-cycled animal fills a sampling slot
until its sampling is turned off by the user. An LVP-1000 should be programmed to
enable automatic transmissions.
Perform the following steps for duty-cycle sampling
1.
2.
3.
4.
5.
Program the LVP-1000 transmitter - see Programming the LVP-1000
Define Source Settings
Define the Size of Data Files
Start Duty-Cycled Sampling
Stop Duty-Cycled Sampling
If you are using Dataquest OpenART, refer to your OpenART 3rd party software user guide for
further instructions.
33
Dataquest A.R.T. and the LVP-1000
Continuous Sampling
Continuous sampling is just what the name implies - sampling one or more transmitters
for an extended period without breaks in the data. A user manually starts continuous
sampling, and it continues uninterrupted for minutes, hours, or days until a user stops
the process. Each animal that is sampled continuously will fill a sampling slot until its
sampling is terminated by the user. An LVP-1000 should be programmed to transmit
continuously for continuous sampling.
Perform the following steps to sample continuously
1.
2.
3.
4.
5.
Program the LVP-1000 transmitter - see Programming the LVP-1000
Define Source Settings
Define the Size of Data Files
Start Continuous Sampling
Stop Continuous Sampling
Scheduled Sampling
Scheduled sampling allows the sampling of many transmitters sequentially and at
intervals defined by the user in the Acquisition software. Each animal that is sampled on
a schedule will fill a sampling slot only during its scheduled time of sampling. An LVP1000 should be programmed to transmit continuously for scheduled sampling.
Perform the following steps for a scheduled sampling protocol
1.
2.
3.
4.
5.
6.
Program the LVP-1000 transmitter - see Programming the LVP-1000
Define the Size of Data Files
Select Data Types to Collect
Set Up Scheduled Sampling
Start Scheduled Sampling
Stop Scheduled Sampling
Defining the Size of Data Files
After selecting which parameters and waveforms you wish to collect from each animal,
define the maximum size of your data files.
To define the size of the data files
1. From the Acquisition window, select Control - Preferences. The Acquisition
Preferences Window appears.
34
Dataquest A.R.T. and the LVP-1000
2. Click the Storage tab.
3. Enter a value of 1 to 32 megabytes as the maximum size of the collected data
files. The default value is 1 MB.
4. Click OK to accept the settings and return to the Acquisition window.
Notes
•
•
DSI recommends increasing the maximum file size if collecting waveform data
continuously over a long period of time.
Make sure that your back up method (floppy disk, Zip disk, CD, DVD, network,
etc.) can accommodate the size of your individual data files.
Selecting Data Types to Collect
After creating a configuration and study folders, select the data types to collect from
each animal. Designate which parameters and waveforms you wish to collect from each
animal, by marking them online. Parameters and waveforms for each animal need to be
marked online in order to be collected.
To select data types to collect
1. From the Acquisition window, select an animal. Double click on the animal icon
to display the Source Settings window.
35
Dataquest A.R.T. and the LVP-1000
2. Enter the following information for each animal.
•
•
•
Segment Duration - Enter a value of 1 to 300 seconds to specify the length of
the scheduled sample. This value is used to break continuous sampling into
contiguous segments for parameter extracting, browsing, and analysis. If using
an LVP-1000 transmitter, the segment duration must be less than or equal to the
length.
Waveforms - Click each of the waveform check boxes to select them as online
for sampling. A check mark indicates waveforms for this data signal will be
sampled and saved to the designated study folder.
Parameters - Click on each of the parameter check boxes to select them as
online for sampling. A check mark indicates parameters for this data signal will
be sampled and saved to the designated study folder.
Schedule
• Enabled for scheduled sampling - Place a check mark in the box to enable the
animal ID for scheduled sampling according to the sampling schedule you
defined in the previous section.
• Save Waveforms Every - Enter a numeric number using the spin dial to select
how often you would like to save a waveform. The computer will automatically
calculate the sampling interval between waveform samples based on the entered
number. This option is only available when the animal ID is set for scheduled
sampling.
36
Dataquest A.R.T. and the LVP-1000
3. Click OK to accept the settings and return to the Acquisition window.
Note
If you have many animals using the same transmitter model, you can copy and paste the
Source Settings from one animal to another. See "Copy/Paste Source Settings" in the
Dataquest A.R.T. Manual for more information.
Starting Duty-Cycled Sampling
Assumptions
LVP-1000 transmitters are turned on and programmed for duty-cycled sampling.
To start duty-cycled sampling
1. From the Acquisition window, select the animal icon.
2. Right-click and choose Duty-Cycled Sampling... to display the Duty Cycled
Sampling window.*
3. Select one of the three duty-cycled sampling options:
•
Save and Trace - Begins duty-cycled sampling and saving of data to the
hard drive while displaying waveform traces in the graph window when a signal
is received from the LVP-1000 transmitter. The selected waveforms and
parameters defined in the Source Settings window will be saved.
•
Trace - Begins duty-cycled sampling and displays the waveform traces in the
graph window when a signal is received from the LVP-1000 transmitter. It will
not save the displayed graphs.
•
Save - Begins duty-cycled sampling and saving of data to the hard drive
when a signal is received from the LVP-1000 transmitter. The selected
Only appears in Dataquest A.R.T. 3.1 or later versions. If Dataquest A.R.T. 3.0 or 3.01 is used,
duty-cycled sampling only saves the data and begins as soon as Duty-Cycled Sampling… is
selected; no trace options are available.
37
Dataquest A.R.T. and the LVP-1000
waveforms and parameters defined in the Source Settings window will be saved.
A graph window will not be shown.
4. Click OK. The animal icon will turn yellow until it receives a signal from the LVP1000 transmitter and green when a signal is received.
To start duty-cycled sampling of an LVP-1000 and another transmitter
A PhysioTel transmitter (non-LVP-1000) can be selected to sample at the same time as
an LVP-1000 that is transmitting in Duty-cycle mode. For example, this is useful if an
animal is implanted with both an LVP-1000 and a D70-PCTP transmitter.
1. Set both animal IDs to the same group and priority number.
To set priority and group
From the Acquisition window, select the animal icon, right click, and
choose Set Priority. The Set Priority window appears.
Assign priority and group to the selected animal icon.
• Level – The priority level, with 99 being the highest priority.
The system samples scheduled sources in priority order from
highest to lowest. Therefore, the system samples an animal with
a priority of 99 before an animal with a priority of 98 or 97.
• Group – Forces all sources with the same priority to be
sampled together. For example, all sources with 99 are sampled
together, if possible. Click this box to place a checkmark in it and
identify this animal as a member of a group. This is typically
used to ensure simultaneous scheduled sampling of related
sources such as a TA11PA-C40 transmitter and a flow probe
converter.
Repeat steps to assign each animal ID a priority and group.
2. From the Acquisition window, select both animal IDs by holding down the Ctrl
key and selecting the animal icons with the mouse.
3. Right-click and choose Duty-Cycled Sampling... to display the Duty Cycled
Sampling window*.
4. Select one of the three duty-cycled sampling options as described in the
previous section.
5. The animal icons will turn yellow until Acquisition receives a signal from the
LVP-1000 transmitter. Both icons will then turn green and begin sampling when
a signal is received.
Only appears in Dataquest A.R.T. 3.1 or later versions. If Dataquest A.R.T. 3.0 or 3.01 is used,
duty-cycled sampling only saves the data and begins as soon as Duty-Cycled Sampling… is
selected; no trace options are available.
38
Dataquest A.R.T. and the LVP-1000
Note
•
•
•
Duty-cycled sampling uses the sampling slots the same way as continuous
sampling.
To start sampling multiple animals with LVP-1000 transmitters at the same time,
hold the Ctrl key and select the animal icons with the mouse so they are
highlighted. Then follow steps 2-3.
The Duty Cycled Sampling Window will only appear in Dataquest A.R.T. 3.1 or
later. If Dataquest A.R.T. 3.0 or 3.01 is used, duty-cycled sampling only saves
the data and begins as soon as Duty-Cycled Sampling… is selected; no trace
options are available.
Stopping Duty-Cycled Sampling
To stop duty-cycled sampling
1. From the Acquisition window, select the appropriate animal icon or multiple
animal icons by holding the Ctrl button.
2. Right click and choose Sampling Off.
3. Alternatively, you can choose from the menu bar, Control - Stop All Sampling
to stop continuous, duty-cycled, and scheduled sampling. Choosing this option
results in the following message: "Are you sure you want to stop ALL sampling
now?"
Choosing Yes will stop all sampling; choosing No will not affect data sampling.
Starting Continuous Sampling
Assumptions
Transmitters are turned on.
To start continuous sampling
1. From the Acquisition window, select the animal icon.
2. Right-click and choose Continuous Sampling… to display the Continuous
Sampling window.
3. Select one of the four continuous sampling options:
•
Save and Trace - Begins continuous sampling and saving of data to the hard
drive while displaying waveform traces in the graph window. The selected
waveforms and parameters defined in the Source Settings window will be saved.
•
Trace - Begins continuous sampling and displays the waveform traces in the
graph window. It will not save the displayed graphs.
39
Dataquest A.R.T. and the LVP-1000
•
Save - Begins continuous sampling and saving of data to the hard drive. The
selected waveforms and parameters defined in the Source Settings window will
be saved. A graph window will not be shown.
•
Diagnostic - Begins a continuous diagnostic trace. Use Diagnostic tracing to
view unconverted data, or to individually set axis limits, sampling rates, and filter
cutoffs. Use the Settings button located on the Continuous Sampling window to
change Diagnostic settings.
Notes
•
•
•
To start sampling multiple animals at the same time, hold the Ctrl button and
select the animal icons with the mouse so they are highlighted. Follow steps 2-3.
To enable or disable data types for tracing, or to set Y-axis limits, select the
Trace tab in the Acquisition Preferences window.
For more information about the Diagnostics option, please refer to Diagnostic
Trace in the Acquisition section of the Dataquest A.R.T. manual.
Stopping Continuous Sampling
To stop continuous sampling
1. From the Acquisition window, select the appropriate animal icon or multiple
animal icons by holding the Ctrl button.
2. Right click and choose Sampling Off.
3. Alternatively, you can choose from the menu bar, Control – Stop All Sampling
to stop both continuous and scheduled sampling. Choosing this option results in
the following message: "Are you sure you want to stop ALL sampling now?"
Choosing Yes will stop all sampling; choosing No will not affect data sampling.
Starting Scheduled Sampling
Assumptions
Transmitters are turned on.
To start scheduled sampling
1. From the Acquisition window, select Control – Scheduled Sampling ON. Data
sampling begins for each animal defined for scheduled sampling.
2. (Steps 2 - 5 are Optional) From the Acquisition window, select the animal icon.
3. Right-click and choose Continuous Sampling… to display the Continuous
Sampling window.
4. Select one of the two available continuous sampling options to view a waveform
trace:
40
Dataquest A.R.T. and the LVP-1000
5. Trace - Begins continuous sampling and displays the waveform traces in the
graph window. It will not save the displayed graphs.
6. Diagnostic - Begins a continuous diagnostic trace. Use Diagnostic tracing to
view unconverted data, or to individually set axis limits, sampling rates and filter
cutoffs. Use the Settings button located on the Continuous Sampling window to
change diagnostic settings.
5. Click OK to view the Real-Time Graphs window containing the waveform traces
for the selected animal.
Note
•
•
Each Trace window in the Real-Time Graphs window uses an additional
sampling slot.
The Trace windows only display waveforms checked Online in the Source
Settings window.
Stopping Scheduled Sampling
To stop scheduled sampling
1. From the Acquisition window, choose Control – Scheduled Sampling OFF.
2. A warning message appears: "Are you sure you want to stop scheduled
sampling?"
3. Click on the Yes button to end data sampling or click on No to continue
sampling data.
4. Alternatively, choosing Control – Stop All Sampling will stop both continuous
and scheduled sampling. Choosing this option results in the following message:
"Are you sure you want to stop ALL sampling now?"
Choosing Yes will stop all sampling; choosing No will not affect data sampling.
Receivers
There are two models of receivers, the RMC-433 and RMC-916. The receiver model
number should match the model number of LVP-1000, either LVP-1000-433 or LVP1000-916.
RMC-433 Specifications
This receiver meets the following mechanical specifications:
• External chassis dimensions - The Height x Width x Length dimensions are
3.15"(8.0cm) x 2.5"(6.4cm) x 7.68"(19.5 cm).
• Weight - 1.41 lbs. (640 g)
• Connector - One 8 pin Fischer watertight connector.
41
Dataquest A.R.T. and the LVP-1000
•
I/O cable interface - Use a Category 5 twisted pair LAN cable or better that is
less than 100 meters in length to connect an RMC-433 to a Data Exchange
Matrix.
Notes
•
•
•
•
Only the LVP-1000-433 transmitter model can be used with this receiver model.
The LVP-1000-433 transmitter model is available in Europe (EU and EFTA).
The Distributed Receiver Array function is available for this receiver model.
Activity parameter collection is not available with this receiver model.
RMC-916 Specifications
This receiver meets the following mechanical specifications:
• External chassis dimensions - The Height x Width x Length dimensions are
3.15"(8.0cm) x 2.5"(6.4cm) x 7.68"(19.5 cm).
• Weight - 1.41 lbs. (640 g)
• Connector - One 8 pin Fisher watertight connector.
• I/O cable interface - Use a Category 5 twisted pair LAN cable or better that is
less than 100 meters in length to connect an RMC-916 to a Data Exchange
Matrix.
Notes
•
•
•
•
42
Only the LVP-1000-916 transmitter model can be used with this receiver model.
The LVP-1000-916 transmitter model is available in the U.S.A. and Canada.
The Distributed Receiver Array function is available for this receiver model.
Activity parameter collection is not available with this receiver model.
Dataquest A.R.T. and the LVP-1000
Left Ventricular Pressure
The above graph displays a marked left ventricular blood pressure waveform and the
table on the following page describes the feature marks. The graph below shows the
same waveform segment, but is scaled to show the waveform clearly.
43
Dataquest A.R.T. and the LVP-1000
Data Series
Abbreviation
IBI
Diastole
AdPdt
Inter-beat
interval
End Diastolic
pressure
dP/dt at
pressure A
BdPdt
dP/dt at
pressure B
CdPdt
dP/dt at
pressure C
Contractility
Index
Contractility
Index
DSP (est)
Estimated
systemic
diastolic
pressure
Mean
Mean
pressure
Systole
Systolic
pressure
Max dP/dt
dPdt
Min dP/dt
Tau
Vmax
44
Parameter
Minimum
dP/dt (max
negative
dP/dt)
Relaxation
index
Contractility
index
Description
Interval between beats registered at
position of max dP/dt.
Pressure immediately before ventricular
contraction.
Pressure over time at fixed pressure
relative to diastolic pressure value.
Default pressure value = 40mmHg.
Pressure over time at fixed pressure
relative to diastolic pressure value.
Default pressure value = 60mmHg.
Pressure over time at fixed pressure
relative to diastolic pressure value.
Default pressure value = 80mmHg.
The maximum change in pressure over
time divided by the pressure at that
point (max dP/dt)/P. P is defined as the
absolute pressure minus end diastolic
pressure.
Estimated systemic diastolic pressure.
This is the pressure at the point of max
dP/dt, sometimes referred to as
P@maxdP/dt. This should reflect the
point where the valve opens in the left
ventricle of the heart.
Mean pressure is calculated based on
averaging the pressure over 20 subsegments of the segment length. The
default reduced mean calculation
discards outliers in the sub-segments in
the calculation.
Peak pressure after detected max
dP/dt.
Maximum change in pressure over time
between any two sample points on the
rising edge of a pressure cycle.
Maximum negative change in pressure
over time between any two sample
points on the falling edge of a pressure
cycle.
The relaxation time constant from min
dP/dt to the point where the pressure
has dropped 66% of the distance from
systolic to diastolic pressure.
This is the extrapolation of dP/dt vs
pressure extrapolated back to P=0. This
is the contractility of the heart
unaffected by loading.
Troubleshooting
Obtaining DSI Technical Services
Contact DSI Technical Services for help if you experience a problem with any part of the
LVP Program Utility, the Dataquest A.R.T. system, or with the LVP-1000 transmitters.
Before contacting DSI Technical Services
•
•
Check that you correctly performed any procedures that correspond to the
problem.
Locate your assigned customer number. Whenever you contact DSI Technical
Services, you need to provide your:
• Name
• Company name
• Phone number
• Fax number
• DSI customer number
To obtain system installation or software-related assistance
Provide the following information:
• A printed copy or the exact error message that was received including the error
code number.
• The name and version of the software you are using. You also need to indicate
whether your Dataquest A.R.T. or OpenART software is the Platinum, Gold,
Silver, or Analysis Only version.
• The computer's operating system.
• The computer brand.
• Your name, DSI customer number, company name, phone number, email
address, and fax number.
To receive answers to surgical procedure questions
Provide the following information:
• Model name and serial number of the transmitter.
• Placement of the transmitter (subcutaneous or intraperitoneal) and catheter.
• Your name, DSI customer number, company name, phone number, email
address, and fax number.
To obtain information about transmitters or hardware equipment
Provide the following information:
• The hardware or transmitter model name and serial number.
• The name and version of the software you are using. You also need to indicate
whether your Dataquest A.R.T. or OpenART software is the Platinum, Gold,
Silver, or Analysis Only version.
• Your computer's operating system.
• Your name, DSI customer number, company name, phone number, email
address, and fax number.
45
Troubleshooting
DSI Technical Services Contact Information
Web Site:
Phone:
Phone:
Fax:
46
www.datasci.com
1-651-481-7400
1-800-262-9687 (U.S. and Canada)
1-651-481-7404
Troubleshooting
Error Messages
These error messages are for the LVP Programming System only. If the error you
experienced is not listed here, contact DSI Technical Services. Consult the Dataquest
A.R.T. Manual for error messages experiences while using that software.
Errors associated with programming
Unable to start reading LVP-1000 settings. Please retry.
Pop-up dialog during the Program Wizard.
Cause: The LVP-1000 was out of range of the Remote Transceiver or there is electrical
noise interference.
Solution: Make sure the Remote Transceiver is properly positioned over the transmitter
of the LVP-1000. The red area of the Remote Transceiver must be positioned facing the
transmitter. Pass the magnet over the transmitter again to retry programming.
Unable to start programming LVP-1000 settings. Please retry.
Pop-up dialog during the Program Wizard.
Cause: The LVP-1000 was out of range of the Remote Transceiver or there is electrical
noise interference.
Solution: Make sure the Remote Transceiver is properly positioned over the transmitter
of the LVP-1000. The red area of the Remote Transceiver must be positioned facing the
transmitter. Pass the magnet over the transmitter again to retry programming.
CRC data error. The LVP-1000 may be out of range. Please try again.
Pop-up error dialog
Cause: The LVP-1000 is too far away from the Remote Transceiver.
Solution: Retry. Pass the magnet over the telemetry unit and reposition the red area of
the Remote Transceiver within 2 inches of the LVP-1000.
Programming has timed out. Please cancel and retry.
Appears in the Program Wizard dialogs.
Cause: No signal was received from the LVP-1000 when the Program Wizard was
waiting for a signal to either read the configuration or start programming a new
configuration.
Solution: Cancel and restart the Program Wizard.
Unable to communicate with the LVP-1000. Please try again.
Appears as a pop-up dialog.
Cause: An unspecified error occurred during programming of the LVP-1000.
Solution: Close the Program Wizard and retry.
Errors associated with hardware
Failure during hardware detection. Please check the hardware connection and try again.
Pop-up dialog appears when trying to open the Program Wizard.
Cause: The LVP Program Utility encountered problems during the hardware check prior
to open the Program Wizard.
47
Troubleshooting
Solution: Ensure that the hardware, Data Monitor, Remote Transceiver, and power
supply are all correctly connected. Next, check the Settings dialog to make sure that the
Serial Port selected is indeed where the Data Monitor is connected.
Please check the hardware connection and try again.
Appears as pop-up dialog.
Cause: The Data Monitor or Remote Transceiver became disconnected or failed in the
middle of programming.
Solution: Check the connections and retry.
Hardware failure. Please check the hardware connection and restart the software.
A pop up dialog appears when trying to open the Program Wizard.
Cause: There was an unexpected error during the hardware check.
Solution: Close and restart the LVP Program Utility. Also ensure that all the hardware is
properly connected.
Could not get hardware info.
Appears when trying to view the Hardware Info dialog.
Cause: The software experienced a problem communicating with the Data Monitor or
Remote Transceiver.
Solution: Check all connections on the Data Monitor and the connection to the
computer. Retry.
Errors associated with the Serial Port
Could not connect to the Serial Port.
Appears when trying to start the Program Wizard.
Cause: The software cannot communicate with the Serial Port selected.
Solution: Use the Settings dialog to select a different Serial Port. Check the connection
to the Data Monitor.
Invalid Serial Port
Appears as a pop-up dialog from the Settings Dialog.
Cause: An invalid Serial Port number was entered.
Solution: Select or enter a valid Serial Port number. Most computers will use COM1.
Errors Associated with the Log File
LVP Program Utility was unable to write a log entry because the log file was corrupted.
Appears in the Main Window status area.
Cause: This message appears if information cannot be written to the log file due to a
corrupted log file. This would occur when the Program Wizard is finished or is canceled
by the user.
Solution: Using Windows Explorer, browse to the LVP Program Utility folder. (The
default location is C:\Program Files\LVP Program Utility 1.0.) Locate the file 'log.xml'.
Move this file to another folder, and make sure all copies are deleted from the LVP
Program Utility folder. The LVP Program Utility will create a new log file the next time the
Program Wizard is used.
48
Troubleshooting
Contact DSI Technical Services if you do not feel comfortable performing this solution by
yourself. Technical Services may also be able to determine the cause of the corruption
in the log file and may be able to repair it for you.
LVP Program Utility was unable to read the log file because the log file was corrupted.
Appears when the Log Viewer is opened.
Cause: The log file has been corrupted.
Solution: Using Windows Explorer, browse to the LVP Program Utility folder. (The
default location is C:\Program Files\LVP Program Utility 1.0.) Locate the file 'log.xml'.
Move this file to another folder, and make sure all copies are deleted from the LVP
Program Utility folder. The LVP Program Utility will create a new log file the next time the
Program Wizard is used.
Contact DSI Technical Services if you do not feel comfortable performing this solution by
yourself. Technical Services may also be able to determine the cause of the corruption
in the log file and may be able to repair it for you.
The log file is corrupt. No log record was created.
Cause: The log file is corrupt and the new record could not be added.
Solution: Contact DSI Technical Services for information on how to proceed.
The log file is corrupt. The log viewer will not be opened.
Cause: The log file is corrupt and cannot be viewed.
Solution: Contact DSI Technical Services for information on how to proceed.
Logging error. No log record was created.
Cause: The log file is corrupt and the new record could not be added.
Solution: Contact DSI Technical Services for information on how to proceed.
Unable to save the log file.
Cause: The log file is corrupt and the new record could not be added.
Solution: Contact DSI Technical Services for information on how to proceed.
Miscellaneous Error Messages
Error displaying dialog
Appears when trying to open a dialog.
Cause: An internal error of the LVP Program Utility.
Solution: Close the LVP Program Utility, disconnect the Data Monitor power supply, and
reboot the computer. Reconnect the Data Monitor power supply and restart the LVP
Program Utility. If the problem recurs, contact DSI Technical services.
Automatic adjustment for daylight savings time is disabled. Please change this in Date
and Time Properties.
Cause: The Program Wizard will not open if the Windows operating system function to
automatically adjust the clock for daylight savings changes has been disabled. This may
occur if the computer also has Dataquest A.R.T. 3.0 installed.
Solution: Enable the daylight savings adjustment in the Windows Date and Time
Properties. Retry the Program Wizard.
49
Troubleshooting
Unable to open Internet Explorer.
Appears when trying to open the Log Viewer.
Cause: The software was unable to open the Log Viewer in Internet Explorer.
Solution: Disconnect the Data Monitor power supply and reboot the computer.
Reconnect the power supply and retry.
If the problem persists, the software should be re-installed. You may save the log.xml file
to a new location prior to the re-installation, and copy it into the LVP Program Utility
folder after the installation. This will preserve the log file. Contact DSI Technical Services
if you are uncomfortable performing this procedure without assistance.
Unexpected exception. Please restart the software.
Cause: An internal error has occurred in the software.
Solution: Close the software and restart. If this does not solve the problem, disconnect
the Data Monitor power supply and reboot the computer. Reconnect the power supply
and retry. If the problem persists, contact DSI Technical Services.
Errors associated with firmware files
Unknown firmware file.
Cause: The LVP Program Utility did not recognize the file specified when the Upgrade
Firmware function was used.
Solution: Firmware Upgrades should only be performed with a file and instructions from
DSI. Verify the location of the firmware upgrade file provided. Retry the action with this
file. If problems persist, contact DSI Technical Services.
Error in the firmware file.
Cause: An error was encountered when the software tried to upgrade the firmware. The
firmware file may have been corrupted.
Solution: Firmware Upgrades should only be performed with a file and instructions from
DSI. Contact DSI Technical Services for a new copy of the firmware file.
Unable to upgrade the firmware.
Cause: An error occurred during the upgrade process and the upgrade was not
completed.
Solution: Firmware Upgrades should only be performed with a file and instructions from
DSI. Retry the procedure. If the problem persists, contact DSI Technical Services.
Could not copy the firmware file.
Appears during a firmware upgrade.
Cause: The *.s19 file was not copied properly during the upgrade process.
Solution: Retry the upgrade procedure. If this fails, contact DSI Technical Services for
further assistance.
50
Glossary
Glossary
continuous: A transmission setting of the LVP-1000. The LVP-1000 is always sending
telemetry data in this setting. Using this setting uses the most battery power.
Data Monitor: Hardware that converts the data received from the remote transceiver
into a digital format and transfers the data to the computer. The Data Monitor is
connected to the computer via a serial port.
duty-cycle: A transmission setting for the LVP-1000. The LVP-1000 follows a schedule
to automatically turn on, measure the LV Pressure and transmit the data to the
receiver. Can occur once or multiple times per day. This mode consumes the
least battery power, but consumption will depend on the schedule selected.
end diastolic pressure (LVEDP): End of the diastolic period and start of the systolic
period, typically characterized by a sharp upturn in pressure, before the climb to
LVSP through max LV dP/dt.
left ventricle (LV): The chamber on the left side of the heart that receives arterial blood
from the left atrium and pumps it into the aorta.
Left Ventricular Pressure (LVP): Blood pressure measured in the heart's left ventricle.
The LVP-1000 catheter is inserted through the ventricle wall into the LV.
log: See system log.
LV: Left ventricle
LVP: see Left Ventricular Pressure
LVP-1000: Implanted telemetry transmitter that measures the LV pressure and transmits
the pressure data to the Dataquest system using a radio telemetry signal. It has
two main parts: the Remote Sensor Assembly (RSA) and the Telemetry Unit
(TU), connected by Multi-conductor leads.
max LV -dP/dt: The maximum negative derivative (dP/dt) value between LVSP and
LVBDP of the following cycle that is calculated using 2 or more data points.
max LV +dP/dt: The maximum positive derivative (dP/dt) value between LVEDP and
LVSP that is calculated using 2 or more adjacent data points.
mean diastolic pressure (LVMDP): The average of the LV pressure waveform between
and including LVBDP and LVEDP.
Multi-Conductor Lead: Connects the Remote Sensor Assembly to the transmitter of the
LVP-1000
51
Glossary
Program wizard: The Program wizard allows you to set the LVP-1000 transmission
schedule and configuration settings.
Remote Sensor Assembly: Part of the LVP-1000. It is comprised of the Pressure
Transmitting Catheter and pressure sensor. It is surgically attached to the
epicardium by the suture skirt.
Remote Transceiver: A device that sends and receives programming signals to and
from the implanted LVP-1000 and is connected to the Data Monitor.
RSA: see Remote Sensor Assembly
serial port: A connection interface on the computer. It has 9 pins organized in 2 rows
and usually located on the back of the computer. Some computers may have
more than one serial port.
system log: The record of all LVP Program Utility attempts to program LVP-1000
transmitters. It also records the configurations programmed.
systolic pressure (LVSP): End of the systolic period, determined by a maximum
pressure value.
telemetry data: Data that is collected by the LVP-1000 and transmitted to the Dataquest
System receiver via radio signals.
Telemetry Unit: Refers to the transmitter portion of the LVP-1000. See transmitter for
more details.
transmission: A telemetry data communication from the LVP-1000.
transmission length: The period of time that a transmission will last. All transmissions,
both scheduled and manual, will last this period of time. Use the Program Wizard
to set the transmission length.
transmitter: Part of the LVP-1000. It transmits the telemetry data to the receiver. It also
sends and receives programming information from the Remote Transceiver. The
transmitter is implanted in a subcutaneous pocket and is connected to the
Remote Sensor Assembly via two Multi-Conductor leads.
wizard: Refers to an interactive utility that guides you through a particular task . The
Program Wizard is the only wizard in the LVP Program Utility.
52
Index
Acquisition program ..................... 33, 34
sampling
continuous................................... 39
duty-cycled.................................. 39
scheduled.................................... 40
selecting a protocol ..................... 33
Actions ......................................... 15, 25
activate......................................... 14, 15
Automatic ..................................... 14, 15
transmission ................................... 21
CD ........................................................ 6
Configuration
creating new configurations............ 27
new LVP-1000 ................................ 28
Configuration module ......................... 32
Continuous ................................... 14, 15
Sampling............................. 34, 39, 40
transmission ................................... 20
Data Exchange Matrix........................ 27
data files
size ................................................. 34
Data Monitor....................... 1, 4, 5, 6, 47
power supply ................................ 1, 4
Data Types......................................... 35
Dataquest A.R.T......... 1, 2, 8, 14, 27, 45
Daylight Saving Time ................... 15, 24
DM-1000 .............................................. 4
Duty-Cycle.......................................... 14
Sampling............................. 33, 37, 39
transmission ................................... 21
Enable Automatic Transmission... 15, 21
Error Messages............................ 45, 47
Firmware Upgrades........................ 6, 47
Help.................................................... 45
installation ............................................ 8
Internet Explorer Window................... 25
Interval ............................................... 15
Left Ventricular Pressure.................... 43
Log ..................................................... 25
View................................................ 25
Log Viewer.......................................... 25
LV Pressure detection ........................ 43
LVP Program Utility ................ 1, 4, 8, 45
installation ......................................... 8
LVP Programming System ........... 1, 4, 5
LVP-1000.............. 1, 2, 5, 14, 15, 37, 39
LVP-1000-433 ................................... 2
LVP-1000-916 ................................... 2
Magnet.................................................. 2
Manual Transmission ......................... 14
Multi-Conductor Leads ......................... 2
OFF .................................................... 14
ON ...................................................... 14
parameters ......................................... 35
pressure .......................................... 43
PhysioTel transmitter.......................... 38
Program.............................................. 15
Program Wizard.................................. 14
Programming ................................ 14, 15
receiver................................... 27, 41, 42
configuring ...................................... 27
features ..................................... 41, 42
RMC-433......................................... 27
RMC-916......................................... 27
Remote Sensor Assembly .................... 2
Remote Transceiver ................. 1, 2, 4, 5
s19 file ............................................ 6, 47
sampling
Continuous...................................... 35
Duty-Cycle ...................................... 35
Scheduled ....................................... 35
Scheduled........................................... 14
Sampling ............................. 34, 40, 41
Serial Number..................................... 25
Serial Port......................................... 1, 4
software installation .............................. 8
Source Settings window ..................... 35
Status Area........................................... 6
Status Lights......................................... 4
Technical Services ....................... 45, 47
time zone ............................................ 24
transmission
Index
automatic ........................................ 14
continuous ...................................... 14
duty-cycle ....................................... 14
manual............................................ 14
Transmission Length.............. 14, 15, 20
Transmission Schedule...................... 14
Transmission Time
End ................................................. 15
Start ................................................ 15
54
Transmitter ..................................... 2, 15
Upgrade Firmware .......................... 6, 47
waveforms .......................................... 35
www.datasci.com................................ 45
XC-1000 ........................................... 4, 5

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : No
Page Count                      : 58
XMP Toolkit                     : XMP toolkit 2.9.1-14, framework 1.6
About                           : uuid:2be6e2f5-f163-4c97-8f51-6dab0a4c27c7
Producer                        : Acrobat Distiller 6.0.1 (Windows)
Creator Tool                    : Acrobat PDFMaker 6.0 for Word
Modify Date                     : 2005:03:04 16:10:01-05:00
Create Date                     : 2005:03:04 16:06:20-05:00
Metadata Date                   : 2005:03:04 16:10:01-05:00
Document ID                     : uuid:56b9fd72-5c30-44c7-b375-2bcb79eab29b
Format                          : application/pdf
Title                           : untitled
Creator                         : 
Author                          : 
EXIF Metadata provided by EXIF.tools
FCC ID Filing: MHAXC1000

Navigation menu