     Control unit 300 Series  mounting + operating
   October 31, 2012    Page 2 of 10  1 Overview This control unit is designated to operate an electrically driven bed with optional comfort functions.  •  Power supply is provided via SMPS so the control box has protection class III.  •  The system is operated with an RF remote. The RF receiver is integrated into the control unit.  Bed view from beneath  Here for example you can see a control unit with two actuators and two massage motors connected  Actuator 1 Actuator 2 Control unitMassage motor 1 2Massage motor 2
   October 31, 2012    Page 3 of 10 Control Unit 258-2 CU258-2 2  Mounting the control unit  This mounting bracket is fixed on the flipside of the bed   These two bolts fix the position of the control unit between the mounting bracket.  The clip snaps in the holding fixtures of the control unit
   October 31, 2012    Page 4 of 10
   October 31, 2012    Page 5 of 10   Performance data control unit Duty cycle  2/18 min. or 10 %, max, 5 switching cycles per minute Output voltage  29 V / 6A Further features   Housing colour  black RAL 9005 Housing material  ABS + PC Ports  drives: 2 actuators / 2 vibration motors power input 1 / multifunction output 1 Indicators + buttons  Reset + pairing button / Pairing LED for RF Power LED Protection type  IP20 Protection class  III Relative humidity  30% - 75% Ambient temperature  +50°F - +104°F Power supply  external    Integrated RF receiver                                                      1   Key Function 1 Pairing enable  Performance data RF receiver modulation GFSK Frequency Range  2.403~2.480GHz Transmitted Power  0 dBm Battery  30V supply voltage  Working Current  <60mA Standby Current  <20mA Ambient Temperature -40—85Ԩ PCB Material & Size  FR4  1.6mm  120*120mm
   October 31, 2012    Page 6 of 10   2.1  Power supply PD12 65444   Connecting cable73178 to control box  Power supply cable 3.00.401.151.30 to mains power  Performance data Output voltage 29 V DC constant voltage output Components  Indicators LED operating status indication (green) Colour black RAL 9005 Further data  Protection type IP20 Protection class II Relative humidity 30% - 75% Ambient temperature +50°F - +104°F Power supply wide input voltage range 100-240 V AC Safety NTC temperature sensor, primary fuse, overload protection, overcurrent switch-off, short-circuit-proof output
   October 31, 2012    Page 7 of 10  Remote Control 258B    2.2  RF258B Keys for fixed positions        Choose between comfortable positions Manual adjustment keys                    Move head & foot rest up and down manually  Massage keys    X  With the timer key you program duration and stop the massage function. X  The LEDs signal the programmed duration, the key below starts the massage. With the Head & Foot keys you can select the massage zones. Key Function 1  Massage Head ON 2  Adjust Massage timer 3  Massage Foot ON 4  Move bed in TV position 5  Start both Massage (Head and Foot) 6  Move bed in Lounge position 7  Relay 1 on 8  Move bed in Flat position 9  Relay 3 on 10  Relay 2 on 11  Move bed in Zero G position 12  Relay 4 on Frontside                    1          2          3                         4          5         6                 7          8          9                 10        11        12          Pairing button  & battery box  (flipside)  Insert 3x AAA                                 13           13 Pairing
   October 31, 2012    Page 8 of 10 Performance data remote (RF transmitter) Modulation GFSK Frequency Range  2.403~2.480GHz Battery AAA Working Current  <40mA Standby Current  <8uA Ambient Temperature -40—85Ԩ RC Dimension  145*50*14mm PCB Material & Size  FR4  1.6mm  79*42mm Housing Material  ABS+PC
   October 31, 2012    Page 9 of 10  3  Board Schematics (remote)  3.1 Remote   3.2 RX module Pls insert diagram       _
   October 31, 2012    Page 10 of 10 4 Functional Description  Take cable 3.00.401.151.30 and connect it to mains supply.  The system is delivered paired! You can test the functions of the system with the RF remote. Pairing instruction (in case of trouble) Pairing the remote with the control box.  The distance between remote and control box for pairing should be less than 2 meters! 1.  Remove the battery cover on the flipside of the remote.  2.  Insert/check existing 3x AAA batteries. 3.  Push the reset button on the control box twice with something spiky.  The green pairing LED adjoining is lit. Now you have 10 s to pair remote and receiver. 4.  On the flipside of the remote is a silicone push button positioned above the batteries (13).  A blue LED starts blinking as the button is pressed.  Keep it pressed until the blue LED is lit continuously, then pairing is completed, the blue LED light fades away.                       This device complies with Industry Canada licence- exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device. 2. Changes or modifications not expressly approved by the party responsible for  compliance could void the user's authority to operate the equipment.  The users manual or instruction manual for an intentional or unintentional radiator shall caution the user that changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. In cases where the manual is provided only in a form other than paper, such as on a computer disk or over the Internet, the information required by this section may be included in the manual in that alternative form, provided the user can reasonably be expected to have the capability to access information in that form.   Le présent appareil est conforme aux CNR d'Industrie Canada applicables aux appareils radio exempts de licence. L'exploitation est autorisée aux deux conditions suivantes : (1) l'appareil ne doit pas produire de brouillage, et (2) l'utilisateur de l'appareil doit accepter tout brouillage radioélectrique subi, même si le brouillage est susceptible d'en compromettre le fonctionnement."  RSS-102 Statement: This equipment complies with Industry Canada radiation exposure limits set forth for an uncontrolled environment. Déclaration CNR-102: Cet équipement est conforme à l'exposition aux rayonnements Industry Canada limites établies pour un environnement non contrôlé.