DP Flow Configuration Data Sheet Emerson Process Management Electric Co. Carbon Monoxide Alarm 2051CF 00806 0100 4010

User Manual: Emerson Process Management - Emerson Electric Co. Carbon Monoxide Alarm 2051CF

Open the PDF directly: View PDF PDF.
Page Count: 4

DownloadDP Flow Configuration Data Sheet Emerson Process Management - Electric Co. Carbon Monoxide Alarm 2051CF 00806-0100-4010
Open PDF In BrowserView PDF
Configuration Data Sheet
00806-0100-4010, Rev FA
August 2012

DP Flow CDS

DP Flow Configuration Data Sheet
All sections are required on this form.
★
= Default Value

 Select only one of the items provided
 One or more of the listed items can be selected

Customer Information
Customer: ____________________________________________________ Contact Name: ___________________________________________________
P.O./Reference No: _____________________________________________ Fax No./Email: __________________________________________________
Phone No.: ____________________________________________________ P.O. Line Item: ___________________________________________________
Model No.______________________________________________________________________________________________________________________
Customer Signoff: ______________________________________________________________________________________________________________

Tagging
Hardware Tag: _________________________________________________________________________________________________
Service:_______________________________________________________________________________________________________

Instrument Selection
Select DP Instrument and complete appropriate additional information.
Annubar®:
_ 3051SFA
___ Fully Compensated
___ Pressure Compensated
___ Temperature Compensated
 DP Compensated(1)
___ DP Only
 3095MFA
 3051CFA
 2051CFA
 485
 585

Compact:
_ 3051SFC
___ Fully Compensated
___ Pressure Compensated
___ Temperature Compensated
___ DP Compensated(1)
___ DP Only
 3095MFC
 3051CFC
 2051CFC
 405

Integral Orifice:
_ 3051SFP
___ Fully Compensated
___ Pressure Compensated
___ Temperature Compensated
___ DP Compensated(1)
___ DP Only
 3095MFP
 3051CFP
 2051CFP
 1195

Orifice Plates:
 _1595 Conditioning Plate
_ 1495 Standard Plate
_____ Concentric Square Edged
________ ISO 5167-2 2003
________ AGA Report #3 2003
________ ASME MFC 3M 2004
________ Drain/Vent (ISO TR 15377)
____ Restriction Orifice
____ Alt. Bore
Type:______________

Additional Information

Additional Information

Additional Information

Additional Information

Sensor Size (if known): __________

Primary Element Technology:
 Conditioning (405C)
 Standard (405P)
Beta Ratio (if known): ____________
 Compact Annubar (405A)

Bore Size (if known): ____________

Bore Size (if known): _______________
or
Beta Ratio (if known): ______________

Connection Type:
For 3051SFA, 3095MFA, 3051CFA,
2051CFA, 485
 Pak-Lok
 Flange w/ Opposite Side Support
__ Flange Rating:_______________
 Flange-Lok
__ Flange Rating:_______________
Gear Drive Flo-Tap:
___ Threaded
___ Flanged
__ Flange Rating:_______________
Manual Drive Flo-Tap:
___ Threaded
___ Flanged
__ Flange Rating:_______________

Connection Type:
Pipe Ends
 Flanged
__ Flange Rating:_______________
 Beveled
 Threaded
Body Only
 Threaded
 Socket-Weld

Plate Type:
 Paddle★
 Paddle – Spiral Wound (1495 only)
 Universal for RTJ Type Flange
___ With Plate Holder
Tap Type:
 Flange
 Corner
 Pipe – D & D/2
Flange Rating:_____________________

For 585
 Flange w/ Opposite Side Support
__ Flange Rating:_______________
 Gear Drive Flo-Tap
__ Flange Rating:_______________
 Main Steam Annubar w/ Opposite
__ Side Support
 Non-Rosemount Primary Element: ________________________________________________________________
Note: Please submit Primary Element Manufacturer’s Calculation Data Sheet.
(1) Compensates for varying discharge coefficient and gas expansion factor based on Reynold’s Number, assuming a fixed pressure and temperature.

www.rosemount.com

Configuration Data Sheet
00806-0100-4010, Rev FA
August 2012

DP Flow CDS
Fluid Selection
Steam:

 Superheated

 Saturated - Pressure Based

 Saturated - Temperature Based

Liquid:

 Water

 Methanol

 Ethanol

 Ammonia

 Other Database Liquid ________________________ (see page 4)

 Air

 Nitrogen

 Oxygen

 Other Database Gas ________________________ (see page 4)

Gas:

 Hydrogen

Natural Gas:

 Please complete Natural Gas Data Sheet (document number 00806-0300-4803) or submit gas analysis report.

Custom(1):

Name_________________________________________________

 Gas

Specific Gravity / Molecular Weight __________________________

Viscosity ______________________

Density / Compressibility @ Flowing Conditions _________________________

@ Base Conditions ___________________

Isentropic exponent ______________________________________
 Liquid

Density @ Flowing Conditions _______________________________________

@ Base Conditions ___________________

Vapor Pressure _________________________________________
(1) Please provide data at normal flowing conditions. For MultiVariable flowmeter configuration, a Custom Gas Data Sheet (document number 00806-0200-4716)
or a Custom Liquid Data Sheet (document number 00806-0300-4716) is required.

Application Data

Sch: __________________

For Duct Mount Only
Recommended

Circular

Square or Rectangular

Duct ID_____________ in.★  mm

Duct Span(1)_____________  in.★  mm

Wall Thickness_______ in.★  mm

OR
Pipe I.D. _____________ in.★  mm

Duct
Span

Line Size: ____________ in.★  mm

Duct Width_______________ in.★  mm
Wall Thickness___________  in.★  mm

Width
Width

Duct
Span

Wall Thickness ________ in.★  mm

316 SST
Primary Element Material:_________________
(316 SST★)
Pipe Orientation:

CS
Pipe/duct Material:________________________________
(CS★)

Horizontal★


 Vertical - Flow Up
 Vertical - Flow Down

(1) Duct Span denotes the dimension that the Annubar primary element will span.

Process Information (Gray boxes are required values)
Units

Minimum

Normal

Maximum

Full Scale / Design

Flow Rate:
Pressure:
Process Temp:
Atmospheric Pressure: ______________________
(14.696 psia★) (used to convert gage pressure to absolute pressure or absolute pressure to gage pressure)
14.696 psia

Base Conditions (Required only if base volumetric flow rate units are used)
 Standard★

 Normal (ISO Standard)

 Standard - Natural Gas (AGA)

 User Defined

P = 14.696 psia/ 101.325 kPaa

P = 14.696 psia/ 101.325 kPaa

P = 14.73 psia/ 101.53 kPaa

P = _____________________________

T = 60 °F/ 15.56 °C

T = 32 °F/ 0 °C

T = 60 °F/ 15.56 °C

T = _____________________ °F  °C

MultiVariable Flowmeter Calibration
For MultiVariable flowmeters, please specify each sensor’s Lower Trim Value (LTV) and Upper Trim Value (UTV).(1)
Differential Pressure:

LTV=_______________________

UTV=_____________________

Units=____________________

Static Pressure:

LTV=_______________________

UTV=_____________________

Units=____________________

Process Temperature:

LTV=_______________________

UTV=_____________________

Units=____________________

(1) If left blank, trim values will be determined from process conditions entered on page 2.

2

Configuration Data Sheet
00806-0100-4010, Rev FA
August 2012

DP Flow CDS

Flowmeter Configuration Defaults(1)
Flowmeter Models: 3051SFA / 3051SFC / 3051SFP
3051S MultiVariable

3051S Single Variable

Compensated Mass and Energy Flow
(Measurement Type: 1-4)

Direct Process Variable
(Measurement Type: 5-7)

Differential Pressure
(Measurement Type: D)

4 mA = 0, 20 mA = Full Scale Flow

4 mA = 0, 20 mA = URL inH2O

4 mA = 0, 20 mA = Full Scale Flow

Process Variable Assignment
PV: Mass Flow
2V: Differential Pressure
3V: Static Pressure (if available)
4V: Process Temp. (if available)

Process Variable Assignment
PV: Square Root of DP
2V: Static Pressure (if available)
3V: Process Temp. (if available)
4V: Module Temp.

Process Variable Assignment
PV: Scaled Variable Representing Flow
2V: Differential Pressure
3V: Module Temp.

Protocol: HART
Burst Mode: Off
Write Protect: Off
Alarm: High
LCD: Flow, DP, P, T

Protocol: HART
Burst Mode: Off
Write Protect: Off
Alarm: High
LCD: DP, P, T

Protocol: HART
Burst Mode: Off
Write Protect: Off
Alarm: High
LCD: Flow

Flowmeter Models: 3051CFA / 3051CFC / 3051CFP

Flowmeter Models: 2051CFA / 2051CFC / 2051CFP

Flowmeter Models: 3095MFA / 3095MFC / 3095MFP

Differential Pressure

Differential Pressure

Compensated Mass Flow

4 mA = 0, 20 mA = URL inH2O

4 mA = 0, 20 mA = URL inH2O

4 mA = 0, 20 mA = Full Scale Flow

Process Variable Assignment
PV: Square Root of DP

Process Variable Assignment
PV: Square Root of DP

Protocol: HART
Burst Mode: Off
Write Protect: Off
Alarm: High
LCD: Flow

Protocol: HART
Burst Mode: Off
Write Protect: Off
Alarm: High
LCD: Flow

Process Variable Assignment
PV: Mass Flow
2V: Differential Pressure
3V: Static Pressure
4V: Process Temp.
Protocol: HART
Burst Mode: Off
Write Protect: Off
Alarm: High
LCD: Flow

(1) If device settings other than default are required, please complete the appropriate Configuration Data Sheet: 00806-0100-4801 for 3051S,
00806-0100-4803 for 3051SMV, 00806-0100-4001 for 3051C, 00806-0100-4101 for 2051C, and 00806-0100-4716 for 3095.

3

Configuration Data Sheet
00806-0100-4010, Rev FA
August 2012

DP Flow CDS
Fluid Database List
 1~1~2~2TETRAFLUORO-ETHANE

 2-METHYL-1-PENTENE

 ETHANE

 METHYL ETHYL KETONE

 PROPANE

 1~1~2TRICHLOROETHANE

 ACETIC ACID

 ETHANOL

 N-BUTYRALDEHYDE

 PROPYLENE

 1~2~4TRICHLOROBENZENE

 ACETONE

 ETHYLAMINE

 METHYL VINYL ETHER

 PYRENE

 1~2-BUTADIENE

 ACETONITRILE

 ETHYLBENZENE

 N-BUTANE

 STYRENE

 1~3~5TRICHLOROBENZENE

 ACETYLENE

 ETHYLENE

 N-BUTANOL

 SULFUR DIOXIDE

 1~3-BUTADIENE

 ACRYLONITRILE

 ETHYLENE GLYCOL

 N-BUTYRONITRILE

 TOLUENE

 1~4-DIOXANE

 AIR

 ETHYLENE OXIDE

 N-DECANE

 TRICHLOROETHYLENE

 1~4-HEXADIENE

 ALLYL ALCOHOL

 FLUORENE

 N-DODECANE

 VINYL ACETATE

 1-BUTENE

 AMMONIA

 FURAN

 NEON

 VINYL CHLORIDE

 1-DECANAL

 ARGON

 HELIUM-4

 NEOPENTANE

 VINYLCYCLOHEXENE

 1-DECANOL

 BENZALDEHYDE

 HYDRAZINE

 N-HEPTADECANE

 WATER

 1-DECENE

 BENZENE

 HYDROGEN

 N-HEPTANE

 1-DODECANOL

 BENZYL ALCOHOL

 HYDROGEN CHLORIDE

 N-HEXANE

 1-DODECENE

 BIPHENYL

 HYDROGEN CYANIDE

 NITRIC ACID

 1-HEPTANOL

 CARBON DIOXIDE

 HYDROGEN PEROXIDE

 NITRIC OXIDE

 1-HEPTENE

 CARBON MONOXIDE

 HYDROGEN SULFIDE

 NITROBENZENE

 1-HEXADECANOL

 CARBON TETRACHLORIDE

 ISOBUTANE

 NITROETHANE

 1-HEXENE

 CHLORINE

 ISOBUTENE

 NITROGEN

 1-NONANAL

 CHLOROPRENE

 ISOBUTYLBENZENE

 NITROMETHANE

 1-NONANOL

 CHLOROTRIFLUOROETHYLENE

 ISOPENTANE

 NITROUS OXIDE

 1-OCTANOL

 CARBON DIOXIDE

 ISOPRENE

 N-NONANE

 1-OCTENE

 CYCLOHEPTANE

 ISOPROPANOL

 N-OCTANE

 1-PENTADECANOL

 CYCLOHEXANE

 M-CHLORONITROBENZENE

 N-PENTANE

 1-PENTANOL

 CYCLOPENTANE

 M-DICHLOROBENZENE

 OXYGEN

 1-PENTENE

 CYCLOPENTENE

 METHANE

 PENTAFLUOROETHANE

 1-UNDECANOL

 CYCLOPROPANE

 METHANOL

 PHENOL

 2~2-DIMETHYLBUTANE

 DIVINYL ETHER

 METHYL ACRYLATE

 PROPADIENE

Standard Terms and Conditions of Sale can be found at www.rosemount.com/terms_of_sale
The Emerson logo is a trade mark and service mark of Emerson Electric Co.
Rosemount, Annubar, ProPlate, and the Rosemount logotype are registered trademarks of Rosemount Inc.
Engineering Assistant is a registered trademark of AMS Suite.

Emerson Process Management
Rosemount Measurement
8200 Market Boulevard
Chanhassen MN 55317 USA
Tel (USA) 1 800 999 9307
Tel (International) +1 952 906 8888
Fax +1 952 906 8889

00806-0100-4010 Rev FA, 8/12

Emerson Process Management
Blegistrasse 23
P.O. Box 1046
CH 6341 Baar
Switzerland
Tel +41 (0) 41 768 6111
Fax +41 (0) 41 768 6300

Emerson FZE
P.O. Box 17033
Jebel Ali Free Zone
Dubai UAE
Tel +971 4 811 8100
Fax +971 4 886 5465

Emerson Process Management Asia Pacific
Pte Ltd
1 Pandan Crescent
Singapore 128461
Tel +65 6777 8211
Fax +65 6777 0947
Service Support Hotline : +65 6770 8711
Email : Enquiries@AP.EmersonProcess.com



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : No
Create Date                     : 1999:02:04 16:49:40Z
Modify Date                     : 2013:04:01 14:23:11-05:00
Has XFA                         : No
XMP Toolkit                     : Adobe XMP Core 5.2-c001 63.139439, 2010/09/27-13:37:26
Format                          : application/pdf
Creator                         : 
Title                           : DP Flow Configuration Data Sheet
Creator Tool                    : FrameMaker 10.0.1
Metadata Date                   : 2013:04:01 14:23:11-05:00
Producer                        : Acrobat Distiller 10.1.4 (Windows)
Document ID                     : uuid:1890de36-2609-4aa0-8136-a695fe84d83c
Instance ID                     : uuid:ca924075-4587-42a8-b73b-4706fcedc0e9
State                           : 1
Version                         : 1.1
Page Mode                       : UseOutlines
Page Count                      : 4
EXIF Metadata provided by EXIF.tools

Navigation menu