Sample Setup Guide Sampling Train — Multiple Sorbent Tubes Publication 1451 PDF

User Manual: 1451

Open the PDF directly: View PDF PDF.
Page Count: 2

Sampling Train — Multiple Sorbent Tubes
Sample Setup Guide
SKC Inc. 724-941-9701 SKC South 434-352-7149 SKC Gulf Coast 281-859-8050 SKC-West 714-992-2780
www.skcinc.com
Publication 1451 Rev 1707
Multiple-tube sampling provides many bene ts when collecting hazardous gases and vapors
from air. This method reduces time in the eld and uses fewer sample pumps because multiple
contaminants requiring different sample tubes can be sampled with one pump. Multiple-tube
sampling also allows collection of Time-Weighted Averages (TWAs) and Short-Term Exposure
Limits (STELs) and side-by-side simultaneous sampling at different ow rates. This Sample Setup
Guide demonstrates how to set up a Sampling Train Using Multiple Sorbent Tubes.
Required Equipment
1. An air sampling pump capable of sampling at the
recommended ow rate with the sampling medium in line,
such as:
• SKC Pocket Pump® TOUCH with Constant Pressure
Controller Cat. No. 224-26CPC-10 (for use in constant
ow mode only)
• SKC AirChek® Series with Constant Pressure Controller
Cat. No. 224-26-CPC and Adjustable Multiple Tube
Holder Cat. No. 224-26 Series
2. An air ow calibrator such as:
SKC chek-mate Calibrator Cat. No. 375-07550N
TSI 4146 Calibrator Cat. No. 740-4146
3. An Adjustable Multiple-tube Low Flow Holder:
Dual (for 2 tubes) Cat. No. 224-26-02
Tri (for 3 tubes) Cat. No. 224-26-03
Quad (for 4 tubes) Cat. No. 224-26-04
4. The sorbent sample tubes speci ed in the method
5. The appropriate size protective covers for the
Adjustable Low Flow Holder:
for 6-mm OD x 70-mm L tubes Cat. No. 224-29A
for 8-mm OD x 110-mm L tubes Cat. No. 224-29B
for 10-mm OD x 150-mm L tubes Cat. No. 224-29C
for 10-mm OD x 220-mm L tubes Cat. No. 224-29D
Optional Equipment
1. SKC Tube Breaker:
Cat. No. 226-03-055
Figure 1. Multiple-tube sampling train with AirChek TOUCH
Introduction
To determine the correct ow rate for the chemical of interest,
refer to the appropriate analytical method. See the operating
instructions for the pump to ensure that it can sample at the
correct ow rate.
1. Preparing the Sorbent Tubes
Using a tube breaker, break off both ends of each sorbent
tube to provide an opening of at least one-half the internal
diameter. These tubes will be used for calibrating the ow
and not for collecting the sample.
2. Setting Up the Calibration Train with Adjustable
Multiple-tube Low Flow Holder — Figure 1
If using a Pocket Pump TOUCH in constant ow mode or any
of the AirChek Series pumps, use exible tubing to connect
the CPC to the pump inlet. Using exible tubing, connect
the adjustable low ow holder to the CPC (or to the pump
inlet if using a Pocket Pump TOUCH in constant pressure
mode). Place the sorbent tubes into the black rubber sleeves
of the adjustable low ow holder. The printed arrow on the
sorbent tube shows the direction of the air ow and should
point toward the holder. If there is no arrow printed on the
tube, insert the end of the tube with the smallest sorbent
section (backup section) into the holder.
Pump
Inlet
Tygon
Tubing
Quad Adjustable
Low Flow Holder
Sample
Tubes
with
Covers
CPC
Notice: This publication is intended for general information only and should not be used as a substitute for reviewing applicable government regulations, equipment operating instructions, or legal standards. The
information contained in this document should not be construed as legal advice or opinion nor as a nal authority on legal or regulatory procedures.
SKC Inc. 724-941-9701 SKC South 434-352-7149 SKC Gulf Coast 281-859-8050 SKC-West 714-992-2780
www.skcinc.com
Publication 1451 Rev 1707
Sampling Train — Multiple Sorbent Tubes
Figure 2. Multiple-tube calibration train with
Pocket Pump TOUCH (constant ow mode)
Figure 3. Top view of a quad adjustable low ow holder
Figure 4. Dual-tube sampling train on worker
3. Calibrating the Flow Rate — Figure 2
Allow the pump to equilibrate from one temperature
extreme to another and to run for 5 minutes before
calibrating. Note the ow rates speci ed by each sampling
method and add them together. Set the pump to a ow
rate at greater than or equal to 15% higher than that
sum. Calibrate the ow rate through each tube. The ow
rate through any one tube cannot exceed 500 ml/min.
To calibrate the ow rate, connect the open end of one
sorbent tube to an external calibrator.
Adjust the ow adjust screw on the corresponding port
on the tube holder until the desired ow rate is achieved.
Turn the screw clockwise to decrease the ow. For tri
and quad models, rst rotate each anti-tamper cover
to expose the ow adjust screw (Figure 3). Disconnect
the tube from the calibrator and calibrate the next tube.
Changing the ow on one tube will not affect the ow
through tubes already set. When using a multiple tube
holder, ensure that all ports contain sorbent tubes. If
sampling with fewer tubes than the number of ports,
insert unopened sorbent tubes in the empty ports to seal
them. Calibrate the ow rate for each tube as speci ed in
the analytical method for the chemical of interest. When
the ow rate has been calibrated, remove the calibrator
and set aside the sorbent tubes used to calibrate the
ow. These tubes will be used after sampling to verify
the ow rate. Record the pre-sample ow rate.
4. Sampling — Figure 4
When ready to start sampling, break off both ends of new
sorbent tubes in the same manner used for calibrating
the ow. Insert the sorbent tubes into the rubber sleeves
of the adjustable low ow holder. Ensure unused ports
contain unopened tubes to "seal" each port. Place a
protective cover over each sorbent tube. Attach the clips
to a worker’s collar and the pump to the worker’s belt.
The sorbent tubes should be placed in a vertical position
during sampling. Turn on the pump and note the start
time and any other sampling information. If one tube is
removed from a multiple tube holder during sampling,
immediately replace it with an unopened tube for the
remainder of the sampling period.
5. After Sampling
At the end of the sampling period, turn off the pump and
note the ending time. Remove the sorbent tubes, seal
the ends of each tube with the caps provided, and record
pertinent sample information.
Using a calibrator, calibrate the ow rate with
representative sorbent tubes in line (the sorbent tubes
originally used to calibrate the ow) to verify that the ow
has not changed by more than 5%.
Submit eld blanks from the same lot number as the
sample tubes. Field blanks should be subjected to
exactly the same handling as the samples (open, seal,
and transport) except that no air is drawn through them.
Pack the sample sorbent tubes, eld blanks, and all
pertinent information securely for shipment to a laboratory
for analysis.
Copyright 1998 - 2017
CPC
Pump
Inlet
Dual Adjustable
Low Flow Holder
Flow Adjust
Screws
Sample Tubes
Tygon
Tubing
chek-mate
Tygon Tubing
Sample Pump
Anti-tamper
Cover
Flow Adjust
Screws
Dual Adjustable
Low Flow Tube
Holder
Tygon Tubing
CPC
Inlet
Sample Pump
Protective Covers
with Clips

Navigation menu