Sample Setup Guide
Sampling with ULTRA Passive (Diffusive) Samplers
Passive sampling is the collection of airborne gases and vapors at a rate controlled by a physical
process such as diffusion, without the use of an air sampling pump. Passive (diffusive) samplers
rely on the movement of contaminant molecules across a concentration gradient (i.e., molecules
diffuse from an area of high concentration in air to an area of low concentration onto the sampler).
This rate of diffusion can be calculated mathematically and determined experimentally for individual
chemicals. SKC provides diffusion or sampling rates for many organic vapors that can be sampled
with SKC ULTRA Passive Samplers. This Sample Setup Guide demonstrates the basic operation
of SKC ULTRA Passive Samplers and defines the critical sampling information that should be sent
to a laboratory for analysis.

Introduction

SKC offers ULTRA Passive Samplers with a choice
of sorbents. Choose the sampler with the sorbent
best suited for the chemical(s) of interest. Refer to
the SKC Passive Sampling Guide on the SKC website,
www.skcinc.com, for more information. The sorbents and
analysis for ULTRA Passive Samplers are the same as those
specified in the active EPA Method TO-17. ULTRA samplers
provide for low parts per billion (ppb) to parts per trillion (ppt)
detection of organic vapors in ambient and indoor air.1 Consult
the SKC Passive Sampling Guide for sampling rates and
minimum and maximum sampling times.
ULTRA Passive Samplers are offered in two configurations.
Select the configuration that best fits your application.
Prefilled ULTRA Passive Samplers
The ULTRA sampler housing is prefilled with purged sorbent. The
compartment access port on the back of the sampler provides for
easy and direct transfer of sorbent to a thermal desorption tube in a
laboratory.

1. Sample Collection
Remove the sampler from the
resealable pouch.

Record sample ID number,
sampling location, and date on
the pouch.
Sampler ID label

Prepare the passive sampler for sampling.
For Prefilled ULTRA: Slide sampler cover to open position
(diffusion holes visible). Write the sample start time on the
pouch label.

User-filled ULTRA Passive Samplers
Sorbent for the ULTRA Sampler is provided separately in glass vials
with PTFE lids to maintain low sorbent background. The sampler's
compartment access port and transfer funnel accessory allow the
user to transfer purged sorbent from the vial directly into the housing
before sampling. After sampling, the user either ships the sample
sorbent in the sampler or transfers the sorbent back into the vial for
shipment. Samplers or vials are shipped to a laboratory for thermal
desorption and analysis.

Diffusion holes (sample
sorbent compartment)
Clip
Blank/
correction
sorbent
compartment

Sliding cover in
open position

Publication 1744 Rev 1503

SKC Inc. 724-941-9701

SKC South 434-352-7149

SKC Gulf Coast 281-859-8050
www.skcinc.com

SKC-West 714-992-2780

Sampling with ULTRA Passive (Diffusive) Samplers
For User-filled ULTRA:
Using the Transfer Funnel
accessory, transfer sorbent
from vial to sampler. Slide
sampler cover to open
position (diffusion holes
visible). Write the sample
start time on pouch label.

2. Sampler Storage and Shipment to Analytical Laboratory
Before Sampling:
Prefilled:
30 days (except charcoal - 6 months) in a
clean, organic solvent-free environment
User-filled: 60 days (except charcoal - 1 year) in a
clean, organic solvent-free environment
After Sampling: 21 days at ≤ 39.2 F (4 C) in a clean, organic
solvent-free environment

When sampling is completed, slide sampler cover to closed
position.
Sample compartment

3. Critical Information to Include in Sample Shipment
The laboratory will need to know the air volume for the
chemical of interest. Calculate as follows:

Clip

Blank/
correction
compartment

Carefully package sealed pouch(es) containing sampler(s) or
sorbent vial(s) and blank(s). Use expedited shipping to ship
sample(s) and Chain of Custody Form(s) to a laboratory for
analysis.

Chemical sampling rate
supplied by SKC (ml/min) X Sampling time (min) = Air volume (ml or L)

Sliding cover in
closed position

For Prefilled ULTRA: Write the sample stop time on the
pouch label. Ensure ID number on sampler matches ID
number on pouch and place sealed sampler in resealable
pouch.

The laboratory will determine the total micrograms (μg) found
on the sampler for the chemical of interest. The air volume
will be used to report the chemical concentration in air as
ppb or ppt.

For User-filled ULTRA: Write the sample stop time on the
pouch label.
a. Ensure ID number on sampler matches ID number on
pouch and place sealed sampler in resealable pouch.
Or
b. Transfer sorbent from the sampler body into a vial
and place vial in envelope.

4. Analysis
Thermal desorption (TD) followed by analysis using gas
chromatography (GC) with a detector specified in the method
for the compound of interest. See Analysis Instructions at
www.skcinc.com/instructions/40125.pdf or www.skcinc.com/
instructions/40127.pdf (charcoal).
References
1

SKC Update to EPA Method TO-17 available at www.skcinc.com

Copyright 2007 - 2015

Notice: This publication is intended for general information only and should not be used as a substitute for reviewing applicable government regulations, equipment operating instructions, or legal standards. The
information contained in this document should not be construed as legal advice or opinion nor as a final authority on legal or regulatory procedures.

Publication 1744 Rev 1503

SKC Inc. 724-941-9701

SKC South 434-352-7149

SKC Gulf Coast 281-859-8050
www.skcinc.com

SKC-West 714-992-2780



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : Yes
Encryption                      : Standard V4.4 (128-bit)
User Access                     : Print, Copy, Extract, Print high-res
Author                          : SKC Inc.
Create Date                     : 2015:03:30 14:29:28-04:00
Keywords                        : Sampling, with, ULTRA, Passive, (Diffusive), Samplers, Sample, Setup, Guide, Publication, 1744, PDF, Document
Modify Date                     : 2015:03:30 14:36:01-04:00
XMP Toolkit                     : Adobe XMP Core 5.2-c001 63.139439, 2010/09/27-13:37:26
Format                          : application/pdf
Title                           : Sampling with ULTRA Passive (Diffusive) Samplers Sample Setup Guide Publication 1744 PDF Document
Creator                         : SKC Inc.
Description                     : Sampling with ULTRA Passive (Diffusive) Samplers Sample Setup Guide Publication 1744 PDF Document
Subject                         : Sampling with ULTRA Passive (Diffusive) Samplers Sample Setup Guide Publication 1744 PDF Document
Creator Tool                    : PScript5.dll Version 5.2.2
Metadata Date                   : 2015:03:30 14:36:01-04:00
Producer                        : Acrobat Distiller 10.1.13 (Windows)
Document ID                     : uuid:50f23f3f-bf50-438b-b526-a084b6dc96f8
Instance ID                     : uuid:d15a4e58-5e72-4b26-a6fb-e213708c4dcf
Page Count                      : 2
EXIF Metadata provided by EXIF.tools

Navigation menu