1MA212 MA 212 0e Conducted EMS And EMI Meas

User Manual: MA 212

Open the PDF directly: View PDF PDF.
Page Count: 35

Download1MA212 MA 212 0e Conducted EMS And EMI Meas
Open PDF In BrowserView PDF
Conducted EMS and EMI
Measurements with R&S®EMC32
Application Note
Products:
ı

R&S SMB100A

®

ı

R&S ESR

ı

R&S BBA100

®

ı

R&S ESRP

ı

R&S NRP-Z91

®

ı

R&S ESU

ı

R&S OSP120

®

ı

R&S ESCI

ı

R&S ENV216

®

ı

R&S ESPI

ı

R&S ESL

®
®
®
®
®
®

Ottmar Gerlach
07.2013-1MA212_0e

Application Note

This application note shows how to configure,
calibrate and perform conducted EMS (Electro
Magnetic Susceptibility) measurements according
to IEC / EN 61000-4-6 and EMI (Electro Magnetic
Interference) measurements according to CISPR
®
16-2-1 with the R&S EMC32 software tool.

Table of Contents
Table of Contents................................................................................ 2
1 Introduction ......................................................................................... 3
2 EMC32 Software Configuration ......................................................... 4
2.1

Overview .......................................................................................................................4

2.2

Installation ....................................................................................................................4

3 Conducted EMS .................................................................................. 6
3.1

IEC / EN61000-4-6.........................................................................................................6

3.2

Hardware Configuration ..............................................................................................7

3.2.1

CDN (Coupling Decoupling Network) ............................................................................7

3.2.2

EM-Clamp ......................................................................................................................7

3.2.3

BCI (Bulk Current Injection) ...........................................................................................8

3.2.4

Interlock Connection ......................................................................................................8

3.3

EMS Configuration in EMC32 .....................................................................................9

3.4

EUT Monitoring ............................................................................................................9

3.5

Example ......................................................................................................................11

3.5.1

Calibrating Signal Paths for CDN with External Amplifier............................................11

3.5.2

EMS Scan Template ....................................................................................................13

3.5.3

EUT Monitoring ............................................................................................................15

4 Conducted EMI .................................................................................. 19
4.1

CISPR 16-2-1...............................................................................................................19

4.2

EMI Auto Test Template ............................................................................................19

4.2.1

LISN Settings for Preview Measurement .....................................................................21

4.2.2

Preview Measurements ...............................................................................................21

4.2.3

Data Reduction ............................................................................................................22

4.2.4

Frequency Zoom (I) .....................................................................................................26

4.2.5

Maximization Measurements .......................................................................................27

4.2.6

Frequency Zoom (II) ....................................................................................................27

4.2.7

Final Measurements ....................................................................................................28

4.2.8

Report ..........................................................................................................................28

4.3

Example ......................................................................................................................29

4.3.1

Hardware Configuration ...............................................................................................29

4.3.2

Select Devices .............................................................................................................29

4.3.3

Start Measurement ......................................................................................................30

5 Ordering information ........................................................................ 33

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

2

1 Introduction
The goal of this application note is to describe the hardware and software configuration
for a typical conducted disturbance measurement setup, its calibration and an example
for an EMS and EMI measurement.
The application note consists of two main parts:
EMS – Electromagnetic Susceptibility test standards for commercial equipment
(IEC/EN 61000-4-6), hardware configuration for CDN (Coupling / Decoupling Network),
EM-clamp and BCI (Bulk Current Injection), EMC32 Software installation and
configuration for CDN calibration and example measurement. EMS measurements
have the goal of determining whether the EUT (Equipment Under Test) is immune to
electromagnetic interference signals or fields. This means that the functionality of the
EUT is not disturbed or only disturbed to a permissible limit. The certification of the
susceptibility is a precondition in the EC to get the CE label for an electronic device.
EMI – Electromagnetic Interference test standards for commercial equipment (CISPR
16-2-1), hardware configuration with e.g. LISN (Line Impedance Stabilization Network)
or current probe, EMC32 configuration and example measurement. EMI
measurements have the goal of determining whether the electromagnetic interference
produced by the EUT does not exceed a defined limit value. This ensures that the
functionality of other electrical devices is not adversely affected. The certification that
the interference radiation is below a certain limit is for example, a precondition in the
European Community (EC) to obtain the CE label for an electronic device.
Without a CE label it is not allowed to place any electrical or electronic devices on the
European Market.
The setups, especially the equipment are recommendations based on long term
experience in design and implementation of EMC systems and have been optimized
for best performance. The use of other components than the recommended ones may
result in significant performance variations.
In addition the achievable system performance depends not only on the test
equipment, but also on the environment e.g. the anechoic chamber performance. The
configurations described in this application note can therefore not guarantee the
fulfillment of the respective standards in any case.
®

The following abbreviations are used in the following text for R&S test equipment:
ı
ı
ı
ı
ı
ı
ı
ı
ı
ı
ı
ı
ı
ı

®

The R&S SMB100A Signal Generator is referred to as SMB100A.
®
The R&S BBA100 Broadband Amplifier is referred to as BBA100.
®
The R&S NRP2 Power Meter is referred to as NRP2.
®
The R&S NRP-Z91 Power Sensor is referred to as NRP-Z91.
®
The R&S ENV216 Two-Line V-Network is referred to as ENV216.
®
The R&S OSP120 Switch Matrix is referred to as OSP120.
®
The R&S ESU EMI Test Receiver is referred to as ESU.
®
The R&S ESL EMI Test Receiver is referred to as ESL.
®
The R&S ESCI EMI Test Receiver is referred to as ESCI.
®
The R&S ESPI EMI Test Receiver is referred to as ESPI.
®
The R&S ESR EMI Test Receiver is referred to as ESR.
®
The R&S ESRP EMI Test Receiver is referred to as ESRP.
®
R&S EMC32 EMC Software is referred to as EMC32.
®
R&S refers to Rohde & Schwarz GmbH und Co KG

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

3

2 EMC32 Software Configuration
2.1 Overview
EMC32 offers various features allowing the experienced user to configure and perform
a wide range of EMC tests. Many of the configurations are pre-configured and do not
need to be changed for standard test described in this application note.
The following chapters guide through the configuration step by step. Please refer to the
EMC32 Getting Started Tutorial or use the online help by pressing F1 for further
information.
The EMC32 software must be installed and configured before first use. The main steps
are:
ı

Software and driver installation. The drivers of the standard R&S devices in the
selection lists are contained in EMC32.

ı

Configuration of hardware

ı

Calibration of signal paths

ı

Calibration of transducers

ı

Carrying out the first EMS or EMI test

2.2 Installation
After inserting the EMC32 CD-ROM the installer is either executed via auto-start, or
needs to be started manually by double-clicking on SHOW AUTOSTART.BAT. The
following screen appears.

Fig. 2-1: EMC32 Start Screen
Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

4

Note:
If you have a Windows 7 64 bit OS you must execute
Setup_EMC32_AMS32.exe as administrator. This is done by right
clicking on the file name and selecting “Execute as Administrator” in
the Windows-Explorer.
Installing/Updating Drivers

Fig. 2-2: Install Drivers

EMC32 has a driver installer tool that allows you to install / update the required drivers.
Configuring Hardware
ı

Start EMC32

ı

Choose Configuration Wizard in the start window.

Fig. 2-3: Configuration Wizard

The Configuration Wizard can also be started in the menu EXTRAS  WIZARDS 
CONFIGURATION WIZARD.
Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

5

3 Conducted EMS
3.1 IEC / EN61000-4-6
IEC/EN 61000-4-6 describes conducted EMS-tests for commercial equipment. To carry
out these tests detailed knowledge of the standard is necessary. The short overview
shows the main parameters influencing a test system.
Frequency
Range

150 kHz to 80 MHz

Modulation

80% Modulation with 1 kHz

Note: 1.8 times higher interference voltage (80% modulation) is necessary to
achieve 3.24 times more amplifier power.
Harmonics

< 15 dBc

Coupling
Methods

CDN (Coupling Decoupling Network, preferred method)
EM-Clamp
BCI (Bulk Current Injection, current clamp)

Impedance

150  (at output of coupling network)

Test
Environment

Shielded room

Test Setup

Test table with conducting surface and grounding connection
Cable length EUT – coupling device: 0,1 to 0,3 m
Cabling height: 0,1 m over conducting surface (with non-conducting supports)
All coupling devices have to be grounded properly.

 1 %, i.e. at least 100 steps. The measurement time per step depends on the
EUT, but should not take longer as 66.7 seconds per decade.
Table 3-1: Influencing Parameters
Frequency
step

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

6

3.2 Hardware Configuration
For conducted EMS measurements according to IEC/EN 61000-4-6 there are three
possible methods of coupling into the EUT.

3.2.1 CDN (Coupling Decoupling Network)
The most common test method is coupling via CDN. It requires the least power and is
always used when appropriate CDNs for the type of signal are available (AC power
line, analog or digital input signal). The block diagram shows the basic configuration for
measurements according IEC / EN 61000-4-6 with a CDN.

BBA100 Power Amplifier
SMB100A Signal Generator
Shielded Room

LAN,
GPIB
or
USB

CDN

RF

RF

Attenuator
6dB

NRP-Z91 Power Sensor

USB

AC mains,
analog or digital
input signal

EUT
PC with R&S®EMC32
software
Fig. 3-1: Configuration for CDN

3.2.2 EM-Clamp
The EM-clamp is used when a CDN is not possible, e.g. for shielded or complex
cables. The clamp and the separate decoupling clamp are placed around the cable.
BBA100 Power Amplifier
SMB100A Signal
Generator

Shielded Room
RF Injection Clamp

Decoupling Clamp
RF in
F1

LAN, GPIB or USB

OSP120 Open
Switch and Control
Unit

RF Out

EUT

®

PC with R&S EMC32
Software

NRP-Z91
Power
Sensor
USB

Fig. 3-2: Configuration for EM-Clamp

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

7

3.2.3 BCI (Bulk Current Injection)
Used similarly as EM-clamp. A monitoring of the injected current is strongly
recommended. Alternatively, two NRP-Z91 power sensors can be used instead of the
OSP120 + NRP-Z91 combination. In case the measurement equipment is located in
remote location, it is convenient to use an NRP2 Power Meter with a LAN connector
plus 2 NRP-Z91 power sensors. There are inexpensive solutions for converting
electrical LAN cables to optical fibres which can be fed more easily into a break box of
a shielded room.
BBA100 Power Amplifier
SMB100A Signal
Generator

Shielded Room
BCI Clamp Monitoring
Clamp

EUT
RF in

OSP120 Open
Switch and Control
Unit

RF Out

F1

NRP-Z91
Power
Sensor

®

PC with R&S EMC32
Software

USB
Fig. 3-3: BCI

3.2.4 Interlock Connection
The interlock is a safety feature, which ensures, that the RF power is only switched on,
when the doors to the measurement site are closed. Contacts on the entrance doors of
the anechoic room are required to implement the interlock.
Note:
The Interlock is a safety feature. It makes sure, that no person is
exposed to hazardous fields or voltages. Therefore a proper
installation of an interlock loop is strongly recommended.

Four interlock switches can be connected to the BBA100 Broadband Amplifier. The
Device Interlock reduces the RF output to secure the amplifier or other connected
devices. The three GROUP INTERLOCKS are used for the operators’ safety, e.g. by
turning the RF power off in case a door of the shielded room is open. Tests can only be
carried out as long as this loop is closed.

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

8

3.3 EMS Configuration in EMC32

Fig. 3-4: EMS configuration window

Select the EMS conducted checkbox and choose the devices. You may either control
the power sensors via USB like in this example or together with an NRP2 power meter.

3.4 EUT Monitoring
In order to classify the susceptibility of a DUT (Device Under Test) or EUT it is
necessary to detect the effects that the disturbing frequencies have on the physical
output signals (U, I, f, etc.), TTL I/O or signal monitored by a video camera. Usually the
output signals are disturbed by certain frequencies so it is necessary to record the
output signal synchronized with the current disturbing frequency. The following picture
shows the possible monitoring routes that can be handled by EMC32.

EMC32-S

LAN (TCP/IP)

CANoe™
Application

LAN (TCP/IP)
RS232

Customer‘s
Monitoring PC

LAN
RS232
GPIB, LAN
RS232, VISA

LAN (TCP/IP)

Signal Processing

CAN, MOST,
LIN, FlexRay

Customer Specific

EUT

Video Monitoring
System

Camera

Measurement or
Stimulus Device

Physical Parameters
(U,I,f,T,...)

TTL I/O
Converter
Signal Conversion

TTL

Signal Source

Fig. 3-5: EUT Monitoring Routes
Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

9

The following figure shows the data flow of the monitoring process.

Monitoring
Process

LAN (TCP/IP)
or
RS232
GPIB, LAN, RS232, VISA

DMM

Oscilloscope

EUT
Monitoring PC

Fig. 3-6: EUT Monitoring data flow

EMC32 offers a generic monitoring device for adapting custom EUT monitoring
devices such as oscilloscopes, digital voltmeters, video monitors etc.

Fig. 3-7: Generic Monitoring

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

10

3.5 Example
This example shows how to configure, calibrate and measure EMS of a EUT with the
CDN method shown in 3.2.1. The NRP-Z91 Power Sensors in this configuration are
connected to the NRP2 Power Meter which is connected to the PC via GPIB or LAN.
They could also be connected directly to the PC via USB depending on the location of
your controller PC.

3.5.1 Calibrating Signal Paths for CDN with External Amplifier
One NRP-Z91 Power Sensor measures the forward power of the BBA100 Broadband
Amplifier, while the second NRP-Z91 measures the voltage injected on the AC mains
via Lüthi CDN L-801 M2/M3 with a Lüthi CR-100A Calibration Adapter.
BBA100 Power Amplifier
SMB100A Signal Generator
Shielded Room
Attenuator 6dB

RF
CDN L-801

LAN

Mains

GPIB
NRP2
Power
Meter

NRP-Z91 Power Sensor

PC with
R&S®EMC32
software

CR-100A
Cal
Adapter

NRP-Z91 Power Sensor

Fig. 3-8: Hardware configuration for CDN calibration

In the following example, the NRP-Z91 Power Sensors are connected to a NRP2
Power Meter which is remotely controlled via GPIB.

Fig. 3-9: Software configuration for CDN calibration

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

11

CDN impedance mismatch adjustment:
The BNC input of the CDN has 50 ohms while the CAL adapter output has 150 ohms.
This results in 15.6 dB (U(CAL Adapter) = U(0) / 6) attenuation must be compensated
with an offset in the CAL Adapter-PM Sensor control.

Fig. 3-10: 50 / 150 Ohm attenuation adjustment

EMC32 shows the measurement results in the Test Components window following
window after the test has been performed.

Fig. 3-11: Test Components

Following System Parameters are monitored

Fig. 3-12: System Parameters

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

12

ı

Generator Output in dBm – Power fed from the generator to the amplifier.

ı

Transducer Forward Power in W – Power fed to the CDN.

ı

Immunity Level in V – Interfering voltage inducted by the CDN.

3.5.2 EMS Scan Template
The EMS Scan Template gives an overview on the hardware configuration and the
used standards.

Fig. 3-13: EMS Scan Template

In the LEVELING MODE menu select the leveling method, the calibration table and the
location of the power sensor.

Fig. 3-14: Leveling Mode

In the Level menu select either a CONSTANT IMMUNITY LEVEL, e.g. 10.0 Volts or an
IMMUNITY SHAPE TABLE and the UPPER and LOWER LEVELING TOLERANCE, e.g. 0.2 dB.

Fig. 3-15: Immunity Level and Tolerance

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

13

The DEVICE SETUPS menu shows the compressed view of the hardware setup in Fig. 36.

Fig. 3-16: Device Setup

The ACTIONS sub-menu allows you to add sub-ranges, custom steps etc.

Fig. 3-17: Actions

The MONITOR SYSTEM PARAMETERS table gives a compressed overview of the location
and function of the used power sensors.

Fig. 3-18: System Monitoring

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

14

3.5.3 EUT Monitoring
EUT Monitoring allows monitoring the reaction of the EUT synchronized with the
inducted EMS frequency scan in form of a voltage, current or digital result. The
application PCMON Simulator included with this application note simulates a EUT
monitoring program running a TCP server which communicates with the EMC32
Generic Monitoring driver. In the following example a simple TCP client console is
used to demonstrate how the commands are sent to the PCMON Simulator and
measurement values read back.

Fig. 3-19: TCP Communication with EUT Monitor

In EMC32 the commands for communication with an EUT monitor are defined in the
Device List  Generic Monitoring device.

Fig. 3-20: Generic Monitoring device

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

15

1.

In the General menu define INTERFACE TYPE = LAN and the IP address and Port of
the TCP Server (PCMON Simulator) = e.g. 172.22.1.1:7777. Make sure the
PCMON Simulator is in listening mode (LED turns green). Then set the Generic
Monitoring STATE = PHYSICAL.

Fig. 3-21: General Settings

2.

Select the GENERAL COMMANDS tab and set the commands as follows.

Fig. 3-22: General Commands

3.

Select the MEASUREMENT QUERIES tab and set the command to read the
measurement result (from PCMON Simulator) as shown below.

Fig. 3-23: Measurement Queries

4.

Select the EMS INFORMATION tab and define the commands as follows. The
ANTENNA and TURNTABLE POSITION parameters are only needed for radiated
EMS.

Fig. 3-24: EMS Information

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

16

The DEVICE PROGRAMMING tab allows custom commands for the EUT Monitoring
device to be defined but is not necessary for the PCMON Simulator example. Press
OK to close the GENERIC MONITORING window and OK again to leave the DEVICE LIST
menu.
In the CHANNEL tab the detecting sensor, e.g. VOLTAGE is selected.

Fig. 3-25: EUT Monitor – Channel

In the HARDWARE tab select the PCMON device.

Fig. 3-26: EUT Monitor – Hardware

Click on the device icon to open the PCMON menu and select the SETTINGS tab. It
displays all available commands which can be deactivated if not needed, for instance
ANT. POSITION and TT POSITION (turntable) which are not necessary for conducted
EMS.

Fig. 3-27: PCMON - Generic Monitoring

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

17

Select the NOGO tab to define NOGO TYPE, LIMIT VALUE and VALUE RANGE. This
means, that if the measured voltage returned by the EUT monitoring system (PCMON
Simulator) invoked by the READ? command is above 8,000000 V or out of the
6,000000 V to 10,000000 V range will set a NoGo flag. The ACTIONS tab (Figure 3-36)
allows to define which actions will be taken in case of a NoGo flag.

Fig. 3-28: EUT-Monitor – NoGo

The ACTIONS tab defines the further actions to be taken in case of a Trigger, NoGo
limits have been exceeded or the Go conditions have been met.

Fig. 3-29: EUT-Monitor Actions

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

18

4 Conducted EMI
4.1 CISPR 16-2-1
The CISPR 16-2-1 specification deals with conducted disturbance measurements
concerning effects of cable bundling and specifies the methods of measurement of
conducted disturbance phenomena in the frequency range 9 kHz to 30 MHz. Annex B
of CISPR 16-2-1 to 16-2-3 contains a table of the minimum sweep times (fastest scan
rates), from which the minimum sweep times for the CISPR band for conducted EMI
measurement (Band B = 0.15 MHz to 30 MHz) for different detector types can be
calculated. A peak detector scans this range in 2.985s (100 ms/MHz) while the quasipeak detector would need 1:39 h (200 s/MHz). In order to reduce the measurement to
a reasonable duration, a fast preview measurement with a peak detector is performed
first, the critical frequencies, either exceeding an absolute limit, or a relative limit above
the noise floor are stored for final, precise (re)measurement with a QP detector. This
method is called data reduction (see 4.2.3 for further details).

4.2 EMI Auto Test Template
Open the SYSTEM  TEST TEMPLATES  EMI AUTO TEST  VOLTAGE WITH 2-LINELISN. The EMI Auto Test editor requires the EMC32-K10 option to be installed. The
automated sequence always consists of preview measurements, data reduction,
maximization measurements, final measurements and reporting.

Fig. 4-1: Emi Auto Test Template

The EMI Auto Test of Voltage with 2-Line-LISN has following submenus:
ı

GENERAL SETTINGS – Defines according start and stop frequencies and the limit
lines of a certain Hardware Setup (e.g. Voltage with 2-Line LISN).

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

19

Fig. 4-2: General Settings

ı

TEST SPECIFIC PARAMETERS – Enables / disables the HP 150 kHz filter which not
only suppresses frequencies under 150 kHz but more important, their associated
harmonics that falsify the scan result.

Fig. 4-3: HP Filter

ı

ACTIONS – Allows to add user defined actions (“Program a Device”, “Remote
Action”, etc.) before and after each measurement step.

Fig. 4-4: Actions

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

20

ı

LIMIT LINES FOR EVALUATION – Defines which limit lines are used to determine the
frequencies for the final measurement.

Fig. 4-5: Limit Lines

4.2.1 LISN Settings for Preview Measurement
Select which line should be measured, L1, N or both.

Fig. 4-6: LISN Settings for Preview Measurement

4.2.2 Preview Measurements
Quick, complete test over the whole frequency range repeated for all necessary
accessory combinations (EMI conducted: 150 kHz to 30 MHz either for one or for both
LISN lines). At each combination an EMI scan or sweep according to the test template
is performed and the output of all scans/sweeps is merged into result tables with the
information on relevant accessory settings for the larger of the measured values at all
frequencies. The number of temporary generated tables for internal use and the
resulting output tables is depending on the settings in the hardware setup (number of
sub-ranges) and in the scan/sweep test template (number of detectors).

Fig. 4-7: Preview Scan / Sweep Template

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

21

4.2.3 Data Reduction
The CISPR 16-2-1 specification stipulates 9 kHz resolution bandwidth (RBW). The step
size for the time-domain scan is ¼ RBW = 9 kHz / 4 = 2.25 kHz. The resulting number
of frequency points is > 10000 and too large for final measurements with the quasipeak detector. The data reduction sequence step avoids spending unnecessary
measurement time for uninteresting frequencies during ZOOM, M AXIMIZATION,
ADJUSTMENT and FINAL MEASUREMENT. The number of measured frequencies is
reduced to a list of critical frequencies used as input at least for the Final
Measurement.

CISPR

Fig. 4-8: Data Reduction Algorithm

The reduction of frequency points from the preview for final measurement is achieved
in two paths (optionally 3 paths).

Fig. 4-9: Data Reduction in EMC32
Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

22

Evaluation Path1:

Fig. 4-10: Evaluation Path1

st

The 1 evaluation path contains a combination of PEAK SEARCH with relative threshold
level and M AXIMA LIMITATION. This path helps to find the highest narrowband signals in
the measured frequency range and reduce them to a reasonable number for further
evaluation.
PEAK SEARCH is used to find all (preferably) narrowband interference peaks depending
on a custom decision level. The decision level determines how far a local maximum
must exceed the adjacent lower test points to be considered as a peak and to appear
in the output file.
In case a limit line is additionally specified, not the absolute level characteristic of the
trace is examined, but the distance between the test point and the respective limit
value.

Fig. 4-11: Limit Lines for Evaluation

With 0 dB decision level, all local maxima are detected as peaks, with 20 dB for
instance, only distinctive narrowband signals are detected. This function should only be
used on non-intermittent curves, i.e. it should be the first one when combined with
other data reduction functions.
M AXIMA LIMITATION is used to transfer a defined number (max. 1000) of maximum
levels from the input to the output file. It is possible to focus e.g. on the 10 most critical
points of a measurement result.

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

23

Evaluation Path 2:

Fig. 4-12: Evaluation Path 2

nd

In the 2 evaluation path the SUBRANGE M AXIMA function is applied. Optionally
determine n points per subrange (normally n=1). The highest local maxima will be
collected from each subrange.
The SUBRANGE M AXIMA function is used to split the input file trace into a defined
number of subranges (max. 1000). The maximum for each subrange is determined and
transferred to the output file. If a limit line is specified, the distance between the test
points and the according limit value is examined instead of the absolute level
characteristic of the trace. The subrange limits are equidistant for linear or logarithmic
scales. This function is suitable for detecting broadband and narrowband signals over
the complete measured spectrum.

REDUCING THE NUMBER OF M AXIMA FROM PATH 1 AND PATH 2 EVALUATION:

Fig. 4-13: Reduce Maxima

Path 1 and path 2 peak search results will be combined and the number of maxima
reduced by following functions:
For the ACCEPTANCE ANALYSIS a limit line for each detector can optionally be selected
which will also be used for the level evaluation in the final measurements. Each limit
line may be optionally shifted by the defined offset value (thus EMC32 will add the
offset value to the copy of the original limit line in the current test.
The ADDITIONAL LIMIT LINES FOR GRAPHIC are only displayed in the result graphic but
are not used for data reduction or evaluation purpose.
Besides, the acceptance offset has to be defined (an offset of e.g. - 10 dB means that
all level points which are higher than 10 dB below the limit will be accepted).

Fig. 4-14: Acceptance Analysis with Fixed Acceptance Offset
Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

24

Alternatively to a fixed acceptance offset, you can directly let filter the data with an
arbitrary acceptance line for each detector (the acceptance offset will then be ignored).
This acceptance line is a user defined limit line (frequency range and level unit have to
match the actual limit line) and is normally defined such that it runs closely above the
noise floor. This is helpful in situations where the noise floor comes relatively close to
the limit.

Fig. 4-15: Acceptance Analysis with Arbitrary Acceptance Line

The parameter 'Display titles of limit lines in graphic' allows enabling or disabling the
display of the limit line name label. Optionally two 'additional limit lines' can be defined
which will be shown in the measurement graphics.

Fig. 4-16: Additional Limit Lines for Graphic

The number of measurement points that will remain after the data reduction can be
limited through a further maxima setting.

Evaluation Path 3:

Fig. 4-17: Evaluation Path 3

rd

Optionally a 3 evaluation path (ACCEPTANCE ANALYSIS) can be used to ensure that
any limit exceeding point will be kept. In rare cases a spike may dominate over a true
Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

25

interferer which therefore may not be kept (although exceeding the limit), depending on
the parameters described above. An offset can be applied (e.g. - 3 dB means that all
points which are higher than 3 dB below the limit will be added).

EVALUATION TRACEABILITY AFTER TEST
At the end of a test you will get the data reduction results, combined from all frequency
ranges. This helps for the traceability of the results from the data reduction.

ADDITIONAL OPTIONS FOR THE TEST FLOW CONTROL:
The following buttons on the right side of this data reduction dialog provide additional
user interaction and evaluation during the test flow:
ı

AFTER DATA REDUCTION:
INTERACTIVE DATA REDUCTION – When enabled the results from the data reduction
can optionally be revised (interactive data reduction). For this purpose the test
will be interrupted after the data reduction has been completed so that the list of
critical frequencies can be edited or expanded (this will be indicated by the
PAUSE sign in the test control toolbar).
MERGE RESULTS – When enabled the following additional evaluation after the data
reduction is done:
▪
If the preview results for only the 1st detector are generated, but the final
measurements are defined with two detectors, then this preview result list will
nd
be copied to the 2 detector and also be used for the final measurement with
nd
the 2 detector.
▪
If results for two detectors are generated, then both result lists can
optionally be merged into a single list which will be used with the 1st detector
only in the consecutive test phases. Especially with DC motors it may be
interesting to extract maxima from a preview MaxPeak curve (broadband
interferers) as well as maxima from a preview AV curve (NB interferers) and
re-measure all these points with the same (Quasipeak) detector.

ı

BEFORE FINAL MEAS:
INTERACTIVE DATA REDUCTION – When enabled the results from the latest test
phase (typically from the maximization) can optionally be revised in another
interactive data reduction before the final measurements. For this purpose the
test will be interrupted so that the list of critical frequencies can be edited
(normally removal of irrelevant points).

4.2.4 Frequency Zoom (I)
The accuracy of the remaining frequencies after data reduction can be increased by
partial scans or sweeps. If the preview measurements are sampled / collected by
means of a sweep template, then a template must be selected, because the frequency
resolution might not be sufficient. When doing preview measurements with scans, then
this feature is optional but may be necessary if the disturbance signals are drifting.

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

26

Fig. 4-18: Frequency Zoom (I)

4.2.5 Maximization Measurements
The (optional) maximization is performed with single measurements based on the
selected scan template.

Fig. 4-19: Maximization Measurements

The maximization measurements will be performed on all LISN setups activated in this
sub-dialog, and maximized in a way so that in the end for each frequency the
maximum level and the corresponding LISN setup will be stored.
Note:
After the maximization phase, the test phases FREQUENCY ZOOM (II), ADJUSTMENT and
FINAL MEASUREMENTS will be performed sequentially (i.e. in one block) before tuning to
the next critical frequency (see next chapter).

4.2.6 Frequency Zoom (II)
This additional test phase for increasing the frequency accuracy is optional (not
available with all automatic test flows !), but may be helpful with interferers of which the
frequency is drifting over time.
Note:
nd

st

The "2 zoom" measurements will not be performed in one block (like "1 zoom"
measurements). Those measurements will be performed just before the corresponding
adjustment / final measurement. As a consequence there is a short delay between the
zoom and subsequent single measurements (all referring to the same interferer)
helping with drifting interferers.

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

27

4.2.7 Final Measurements
The FINAL MEASUREMENTS are performed as single measurements based on the
selected scan template. The scan template should fulfill all requirements of the EMC
standard (detectors, IF bandwidth, measurement time). The results will be evaluated
against the corresponding limit line (selected in the Data Reduction sub-dialog).

Fig. 4-20: Final Measurements

With conducted disturbance measurement, the LISN setup will be used and stored
along with the result before each final measurement.

4.2.8 Report
The automatic report generation at the end of the test can be activated either as a
printer hardcopy or a file (HTML: an HTML file plus one WMF file for each graphics
diagram; RTF: a generic MS Word format; PDF: ADOBE Acrobat Reader). The file
report will be stored in the sub-directory
\Report\
with the name
Report1
(ascending numbers).
Note that a report setup can be defined with place holders so that certain results of an
automatic test flow will be automatically included (typically the final results). The
purpose of the Report Settings dialog is to reference a standard report template which
is copied into a newly created test. This template can be modified in the TEST NEW
dialog.

Fig. 4-21: Report Settings

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

28

4.3 Example
4.3.1 Hardware Configuration
Shielded Room
ENV216 Two-Line
V-Network (LISN)

R&S® ESRP EMI Test Receiver

EUT
e.g. DIY
Computer

PE

®

PC with R&S EMC32
Software

LAN
or
GPIB

Fig. 4-22: Conducted EMI Hardware Configuration

Note:
Do not connect the power mains of the ENV216 Two-Line-V-Network
to a plug secured by a leakage current circuit breaker. Due to the
high error current, the circuit breaker will immediately trigger. Also
be sure to connect a cable from the rear PE (Protective Earth)
connecter to electrical earth before plugging in the mains.

4.3.2 Select Devices
ı

Open the EXTRAS  DEVICE LIST or press F9 to define desired EMI Test Receiver,
e.g. ESRP7.

Fig. 4-23: ESRP7 Configuration

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

29

ı

and LISN.

Fig. 4-24: LISN Parameters

Choose the appropriate CORRECTION TABLE for the N and L1 test. In case the test
receiver has an according 25-pin Sub-D remote connected, select LISN CONTROL
 REMOTE CONTROL WITH THE RECEIVER for an automated scan.

4.3.3 Start Measurement
Switch to Measurement Mode by clicking on the SWITCH TO MEASUREMENT MODE (F4)
icon,

Fig. 4-25: Switch to Measurement Mode icon

selecting the MEASUREMENT MODE (F4) menu item

Fig. 4-26: Measurement Mode menu item

or pressing the F4 button.
In order to start the measurement, press the Start button.

Fig. 4-27: Start button

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

30

A PREVIEW-SCAN of the complete frequency range (150 kHz to 30 MHz) will be
performed first. Depending on the receiver type, this part can preferably be performed
with the FFT-based time domain scan (ScanFast mode in EMC32), available on the
ESRP, ESR and ESU, which only takes only a few seconds and is by magnitudes
faster than the conventional stepped frequency scan (ScanLin mode) which is
supported by default by all receiver types.

Fig. 4-28: Final Measurement Results

The FINAL TEST result reveals that 29 test points were critical and needed to be remeasured according to the procedure described above. The green values passed after
the final measurement, while the red ones (Margin < 0.0 dB) exceed the limit.

Fig. 4-29: Critical Frequencies
Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

31

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

32

5 Ordering information
EMC32 Software
EMC32
EMC32-K10

EMI Auto Test for EMC32-EB

1117.6840.02

EMC32-K2

Meas. Functions “audio break through”, “Spurious Emissions”

1147.5506.02

EMS Measurements
SMB100A

Signal Generator

1406.6000.02

SMB-B101

Frequency Range 9kHz to 1.1GHz

1407.2509.02

BBA100A

Broadband Amplifier

5354.9000.50

BBA9K250M

Single Band 9 kHz – 250 MHz

5354.2506.02

NRP2

Power meter, control unit with color display, GPIB, USB, LAN

1144.1374.02

NRP-Z91

Power sensor 9 kHz – 6 GHz

1168.8004.02

NRP-Z4

USB adapter for NRP-Z sensors

1146.8001.02

EMI Measurements
ESR3

EMI Test Receiver 9 kHz – 3.6 GHz

1316.3003.03

ESR7

EMI Test Receiver 9 kHz – 7 GHz

1316.3003.07

ESRP3

EMI Test Receiver 9 kHz – 3.6 GHz

1316.4500.03

ESRP7

EMI Test Receiver 9 kHz – 7 GHz

1316.4500.07

ESU8

EMI Test Receiver 20 Hz – 8 GHz

1302.6005.08

ESU26

EMI Test Receiver 20 Hz – 26 GHz

1302.6005.26

ESU40

EMI Test Receiver 20 Hz – 40 GHz

1302.6005.40

ESCI

EMI Test Receiver 9 kHz – 3 GHz

1166.5950.03

ESCI7

EMI Test Receiver 9 kHz – 7 GHz

1166.5950.07

ESL3

EMI Test Receiver 9 kHz – 3 GHz

1300.5001.03

ESL6

EMI Test Receiver 9 kHz – 6 GHz

1300.5001.06

ESPI3

EMI Test Receiver 9 kHz – 3 GHz

1164.6407.03

ESPI7

EMI Test Receiver 9 kHz – 7 GHz

1164.6407.07

Z11MA212_0e

Rohde & Schwarz Conducted EMS and EMI Measurements with R&S®EMC32

33

About Rohde & Schwarz

Regional contact

Rohde & Schwarz is an independent group of
companies specializing in electronics. It is a leading
supplier of solutions in the fields of test and
measurement, broadcasting, radiomonitoring and
radiolocation, as well as secure communications.
Established more than 75 years ago, Rohde &
Schwarz has a global presence and a dedicated
service network in over 70 countries. Company
headquarters are in Munich, Germany.

Europe, Africa, Middle East
+49 89 4129 12345
customersupport@rohde-schwarz.com

Environmental commitment
ı

Energy-efficient products

ı

Continuous improvement in environmental
sustainability

ı

ISO 14001-certified environmental management
system

North America
1-888-TEST-RSA (1-888-837-8772)
customer.support@rsa.rohde-schwarz.com
Latin America
+1-410-910-7988
customersupport.la@rohde-schwarz.com
Asia/Pacific
+65 65 13 04 88
customersupport.asia@rohde-schwarz.com
China
+86-800-810-8228 /+86-400-650-5896
customersupport.china@rohde-schwarz.com
This application note and the supplied programs
may only be used subject to the conditions of use
set forth in the download area of the Rohde &
Schwarz website.

R&S® is a registered trademark of Rohde & Schwarz GmbH &
Co. KG; Trade names are trademarks of the owners.



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.5
Linearized                      : No
Page Count                      : 35
Language                        : en-US
Tagged PDF                      : Yes
Title                           : 1MA212
Author                          : Ottmar Gerlach
Subject                         : Conducted EMS and EMI Measurements with R&S®EMC32
Keywords                        : EMC32, EMI, EMS, EMV, ESR, ESRP, ESCI, ESL, BBA100, ENV216
Creator                         : Microsoft® Word 2010
Create Date                     : 2013:07:19 12:08:16+02:00
Modify Date                     : 2013:07:19 12:08:16+02:00
Producer                        : Microsoft® Word 2010
EXIF Metadata provided by EXIF.tools

Navigation menu