Preloaded Coated Filters
Cat. No. 225-9027
SKC coated lters are shipped preloaded in the cassettes with end plugs and
shrink bands in place. No assembly is required; they are ready to use.
Method & Chemical:
OSHA PV2055 for 1,3,5-triglycidyl isocyanurate (TGI)
Filter & Coating: Glass ber lter coated with hydrobromic acid and cellulose
pad. Two-piece 37-mm cassette. Closed-face sampling con guration
Prior to Sampling: Store at 39.2 F (4 C) for up to 6 months.
Sample Stability: Store samples at 39.2 F (4 C) for up to 7 days.
For information on other available
coated lters, request SKC Publication 1219.
Operating Instructions
863 Valley View Road, Eighty Four, PA 15330
Tel: 724-941-9701 • www.skcinc.com
How to Use the SKC Preloaded Coated Filters
Ensure that lter cassettes are at room temperature before sampling.
1. Select one coated lter cassette for calibrating the ow rate. A red plug
secures the inlet, which is clearly marked “inlet,” and a blue plug secures the
outlet. Set up the sampling train for calibration with the representative lter
cassette in line. For “closed-face” sampling, remove the plugs and connect the
cassette to the sampling train. For “open-face” sampling, remove the outlet
plug and the cassette inlet section and then connect the cassette to the sampling
train. For details on setting up a sampling train, refer to the SKC Sample Setup
Guide “Sampling Train - Filters” (SKC Publication 1922).
2. Calibrate the pump to the desired ow rate using a calibrator.
3. Replace the cassette used for calibration with a fresh coated lter cassette
for sample collection. Attach the cassette to a worker’s collar, as close to the
breathing zone as possible, by using a lter cassette holder (SKC Cat. No.
225-1). Sample for the speci ed time interval and record the time.
4. Remove the lter cassette at the end of the speci ed sampling period and
replace both end plugs (and the inlet if necessary). Recheck the ow rate using
the same cassette and calibrator used for calibration (in Step 2) to ensure that
the ow rate has not changed by more than 5%.
5. Store samples at 39.2 F (4 C) for up to 7 days. Appropriately package
samples, blanks, and all pertinent data to send to a lab for analysis. Expedite
shipping of samples to laboratory. Friday shipments are not recommended.