5790B AC Measurement Standard Extended Specifications Data Sheet Brochure
User Manual: 5790B-AC-Measurement-Standard-data-Sheet-Brochure
Open the PDF directly: View PDF .
Page Count: 15
5790B
AC Measurement
Standard
Extended Specifications
2 Fluke Calibration Fluke 5790B Extended Specifications
General Specifications
Warm-up Time ............................................ 30 minutes
Relative Humidity
Operating ................................................ 45 % to 50 °C 75 % to 45 °C 95 % to 30 °C
Storage .................................................... <95 % non-condensing
Altitude
Operating ................................................ 3,050 meters (10,000 feet)
Non-Operating ........................................ 12,200 meters (40,000 feet)
Temperature
Operating ................................................ 0 °C to 50 °C
Calibration .............................................. 15 °C to 35 °C
Storage .................................................... -40 °C to 70 °C
Electromagnetic Compatibility (EMC)
International ........................................... IEC 61326-1: Controlled Electromagnetic Environment
CISPR 11: Group 1, Class A
Group 1: Equipment has intentionally generated and/or uses
conductively-coupled radio frequency energy that is
necessary for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments
other than domestic and those directly connected to a low-
voltage power supply network that supplies buildings used
for domestic purposes. There may be potential difficulties in
ensuring electromagnetic compatibility in other environments
due to conducted and radiated disturbances.
Emissions that exceed the levels required by CISPR 11 can
occur when the equipment is connected to a test object.
Korea (KCC) ............................................. Class A Equipment (Industrial Broadcasting & Communication
Equipment)
Class A: Equipment meets requirements for industrial
electromagnetic wave equipment and the seller or user
should take notice of it. This equipment is intended for use in
business environments and not to be used in homes.
USA (FCC) ................................................ 47 CFR 15 subpart B. This product is considered an exempt device
per clause 15.103.
Surge ........................................................... ANSI C62.41-1980, Category A
Reliability .................................................... MIL-T-2880D, paragraph 3.13.3
Size
Height ..................................................... 17.8 cm (7 in) standard rackmount + 1.5 cm (0.6 in)
Width ...................................................... 43.2 cm (17 in)
Depth ....................................................... 63 cm (24.8 in)
Maximum Power Requirements
5790B ...................................................... 95 VA
With Wideband Option ............................ 120 VA
Weight
5790B ...................................................... 24 kg (53 lb)
With Wideband ....................................... 24.5 kg (54 lb)
Line Power .................................................. 50 Hz/60 Hz; 100 V - 120 V, 220 V - 240 V
Safety .......................................................... IEC 61010-1: Overvoltage Category II, Pollution Degree 2
IEC 61010-2-030: Measurement 1000 V CAT O
Remote Interfaces ....................................... RS-232, IEEE-488
Confidence Level ......................................... 99 % unless otherwise specified.
DC Zero Cal .................................................. Perform the dc zero calibration every 30 days. In addition, perform
the dc zero calibration after powering up the unit the first time
after unpacking following a shipment or if exposed to an
environmental change of greater than 5 °C.
3 Fluke Calibration Fluke 5790B Extended Specifications
Resolution and Range Limits
Voltage Range Autoran
g
e Limits [1] Resolution
U
pp
er Lower Filter Fast Filter Med/Slow
2.2 mV 2.2 mV 600
μ
V0.1
μ
V0
.1
μ
V
7 mV 7 mV 1.9 mV 0.1
μ
V0
.1
μ
V
22 mV 22 mV 6 mV 0.1
μ
V0
.1
μ
V
70 mV 70 mV 19 mV 0.1
μ
V0
.1
μ
V
220 mV 220 mV 60 mV 0.1
μ
V0
.1
μ
V
700 mV 700 mV 190 mV 1.0
μ
V0
.1
μ
V
2.2 V 2.2 V 600 mV 1.0
μ
V0
.1
μ
V
7 V 7 V 1.9 V 10
μ
V1
.0
μ
V
22 V 22 V 6 V 10
μ
V1
.0
μ
V
70 V 70 V 19 V 100
μ
V10
μ
V
220 V 220 V 60 V 100
μ
V10
μ
V
700 V 700 V 190 V 1.0 mV 100
μ
V
1000 V 1050 V 600 V 1.0 mV 100
μ
V
[1] In locked ranges, readings may be made approximately 1 % beyond the autorange limits.
Electrical Specifications
The Product specifications describe the Absolute Instrumental Uncertainty of the Product. The Product
specifications include stability, temperature, and humidity; within specified limits, linearity, line and load
regulation, and the reference standard measurement uncertainty. The Product specifications are provided at a 99
% confidence level, k=2.58, normally distributed, unless otherwise stated.
The relative specifications are provided for enhanced accuracy applications. To calculate an enhanced absolute
specification from the relative specification, it is necessary to combine the uncertainty of your external standards
with the pertinent relative specifications. Specifications are valid after allowing a warm-up period of 30 minutes,
or twice the time the Product has been turned off.
AC Absolute Voltage
Voltage
Range Frequency Range
Absolute ±5 °C of Calibration Temperature
AC/DC Transfer
Mode ± ppm
2 Years
Measurement Mode ± (ppm of Reading + μV)
90 Days 1 Year 2 Years
2.2 mV
10 Hz - 20 Hz
20 Hz- 40 Hz
40 Hz - 20 kHz
20 kHz - 50 kHz
50 kHz - 100 kHz
100 kHz - 300 kHz
300 kHz - 500 kHz
500 kHz - 1 MHz
1700 + 1.3
740 + 1.3
420 + 1.3
810 + 2.0
1200 + 2.5
2300 + 4.0
2400 + 6.0
3200 + 6.0
1700 + 1.3
740 + 1.3
420 + 1.3
810 + 2.0
1200 + 2.5
2300 + 4.0
2400 + 8.0
3500 + 8.0
1700 + 1.3
740 + 1.3
420 + 1.3
820 + 2.0
1200 + 2.5
2300 + 4.0
2600 + 8.0
5000 + 8.0
7 mV
10 Hz - 20 Hz
20 Hz - 40 Hz
40 Hz - 20 kHz
20 kHz - 50 kHz
50 kHz - 100 kHz
100 kHz - 300 kHz
300 kHz - 500 kHz
500 kHz - 1 MHz
850 + 1.3
370 + 1.3
210 + 1.3
400 + 2.0
600 + 2.5
1200 + 4.0
1300 + 6.0
2000 + 6.0
850 + 1.3
370 + 1.3
210 + 1.3
400 + 2.0
600 + 2.5
1200 + 4.0
1300 + 8.0
2300 + 8.0
850 + 1.3
370 + 1.3
210 + 1.3
410 + 2.0
610 + 2.5
1200 + 4.0
1400 + 8.0
3600 + 8.0
22 mV
10 Hz - 20 Hz
20 Hz - 40 Hz
40 Hz - 20 kHz
20 kHz - 50 kHz
50 kHz - 100 kHz
100 kHz - 300 kHz
300 kHz - 500 kHz
500 kHz - 1 MHz
290 + 1.3
180 + 1.3
110 + 1.3
210 + 2.0
310 + 2.5
810 + 4.0
860 + 6.0
1400 + 6.0
290 + 1.3
190 + 1.3
110 + 1.3
210 + 2.0
310 + 2.5
810 + 4.0
890 + 8.0
1700 + 8.0
290 + 1.3
190 + 1.3
110 + 1.3
210 + 2.0
310 + 2.5
820 + 4.0
1000 + 8.0
2600 + 8.0
70 mV
10 Hz - 20 Hz [1]
20 Hz - 40 Hz
40 Hz - 20 kHz
20 kHz - 50 kHz
50 kHz - 100 kHz
100 kHz - 300 kHz
300 kHz - 500 kHz
500 kHz - 1 MHz
240 + 1.5
120 + 1.5
64 + 1.5
120 + 2.0
260 + 2.5
510 + 4.0
660 + 6.0
1100 + 6.0
240 + 1.5
120 + 1.5
65 + 1.5
130 + 2.0
260 + 2.5
510 + 4.0
670 + 8.0
1100 + 8.0
240 + 1.5
130 + 1.5
69 + 1.5
130 + 2.0
260 + 2.5
530 + 4.0
680 + 8.0
1300 + 8.0
4 Fluke Calibration Fluke 5790B Extended Specifications
220 mV
10 Hz - 20 Hz
[1]
210 210 + 1.5 210 + 1.5 210 + 1.5
20 Hz - 40 Hz 82 84 + 1.5 85 + 1.5 87 + 1.5
40 Hz - 20 kHz 34 37 + 1.5 38 + 1.5 43 + 1.5
20 kHz - 50 kHz 67 69 + 2.0 69 + 2.0 73 + 2.0
50 kHz - 100 kHz 160 + 2.5 160 + 2.5 160 + 2.5
100 kHz - 300 kHz 240 + 4.0 250 + 4.0 280 + 4.0
300 kHz - 500 kHz 360 + 6.0 380 + 8.0 400 + 8.0
500 kHz - 1 MHz 940 + 6.0 1000 + 8.0 1200 + 8.0
700 mV
10 Hz - 20 Hz [1] 210 210 + 1.5 210 + 1.5 210 + 1.5
20 Hz - 40 Hz 73 75 + 1.5 76 + 1.5 78 + 1.5
40 Hz - 20 kHz 27 31 + 1.5 33 + 1.5 38 + 1.5
20 kHz - 50 kHz 47 50 + 2.0 51 + 2.0 56 + 2.0
50 kHz - 100 kHz 79 + 2.5 79 + 2.5 84 + 2.5
100 kHz - 300 kHz 160 + 4.0 180 + 4.0 210 + 4.0
300 kHz - 500 kHz 300 + 6.0 300 + 8.0 340 + 8.0
500 kHz - 1 MHz 900 + 6.0 960 + 8.0 1200 + 8.0
2.2 V
10 Hz - 20 Hz [2] 200 200 200 200
20 Hz - 40 Hz 63 65 66 69
40 Hz - 20 kHz 18 22 24 29
20 kHz - 50 kHz 43 45 46 52
50 kHz - 100 kHz 70 71 76
100 kHz - 300 kHz 150 160 200
300 kHz - 500 kHz 250 260 310
500 kHz - 1 MHz 840 900 1200
7 V
10 Hz - 20 Hz [2] 200 200 200 200
20 Hz - 40 Hz 63 66 67 70
40 Hz - 20 kHz 18 22 24 29
20 kHz - 50 kHz 44 46 48 53
50 kHz - 100 kHz 80 81 88
100 kHz - 300 kHz 180 190 220
300 kHz - 500 kHz 380 400 470
500 kHz - 1 MHz 1100 1200 1500
22 V
10 Hz - 20 Hz [2] 200 200 200 200
20 Hz - 40 Hz 63 66 67 70
40 Hz - 20 kHz 21 25 27 31
20 kHz - 50 kHz 44 46 48 53
50 kHz - 100 kHz 80 81 85
100 kHz - 300 kHz 180 190 220
300 kHz - 500 kHz 380 400 470
500 kHz - 1 MHz 1100 1200 1500
70 V [3]
10 Hz - 20 Hz [2] 200 200 200 200
20 Hz - 40 Hz 63 67 68 72
40 Hz - 20 kHz 25 30 32 39
20 kHz - 50 kHz 55 56 57 63
50 kHz - 100 kHz 91 94 110
100 kHz - 300 kHz 190 200 220
300 kHz - 500 kHz 400 410 510
500 kHz - 1 MHz 1100 1200 1500
220 V [3]
10 Hz - 20 Hz 200 200 200 200
20 Hz - 40 Hz 63 67 68 72
40 Hz - 20 kHz 23 29 31 38
20 kHz - 50 kHz 63 67 69 77
50 kHz - 100 kHz 96 98 110
100 kHz - 300 kHz 210 210 260
300 kHz - 500 kHz 440 500 700
5 Fluke Calibration Fluke 5790B Extended Specifications
700 V
10 Hz - 20 Hz [4] 200 200 200 200
20 Hz - 40 Hz 92 96 99 110
40 Hz - 20 kHz 36 39 41 47
20 kHz - 50 kHz 120 130 150
50 kHz - 100 kHz 400 500 850
1000 V
10 Hz - 20 Hz 200 200 200 200
20 Hz - 40 Hz 92 96 99 110
40 Hz - 20 kHz 33 37 38 44
20 kHz - 50 kHz
[5
]
120 130 150
50 kHz - 100 kHz
[
5
]
400 500 850
[1] For 9.5 to 10 Hz, the specifications is ±(1000 ppm of reading + 1.5 μV)
[2] For 9.5 to 10 Hz, the specifications is ±(1000 ppm of reading)
[3] Inputs >100 kHz and with a V*Hz product >2.2E7 are typical.
[4] Typical specification, as determined by sourcing with the Fluke 5205A Precision Power Amplifier.
[5] Inputs >30 kHz and >750 V are typical, as determined by sourcing with the Fluke 5205A Precision Power Amplifier.
Note: The Product is to be used in controlled environments. For disturbances on the mains power supply of >0.5 Vrms from
10 MHz to 40 MHz, add 5 ppm to the 2.2 V range.
AC Relative Voltage
Voltage
Range Frequency Range
Relative Uncertaint
y
±5 °C of Calibration Tem
p
erature
AC/DC Transfer
Mode ± ppm 2
Years
Measurement Mode ±(
pp
m of Readin
g
+
μ
V)
90 Days 1 Year 2 Years
2.2 mV
10 Hz - 20 Hz 100 + 1.3 110 + 1.3 110 + 1.3
20 Hz - 40 Hz 54 + 1.3 64 + 1.3 68 + 1.3
40 Hz - 20 kHz 44 + 1.3 57 + 1.3 61 + 1.3
20 kHz - 50 kHz 57 + 2.0 67 + 2.0 110 + 2.0
50 kHz - 100 kHz 79 + 2.5 86 + 2.5 120 + 2.5
100 kHz - 300 kHz 190 + 4.0 230 + 4.0 390 + 4.0
300 kHz - 500 kHz 590 + 6.0 720 + 8.0 1200 + 8.0
500 kHz - 1 MHz 2200 + 6.0 2600 + 8.0 4400 + 8.0
7 mV
10 Hz - 20 Hz 80 + 1.3 83 + 1.3 86 + 1.3
20 Hz - 40 Hz 33 + 1.3 39 + 1.3 45 + 1.3
40 Hz - 20 kHz 29 + 1.3 36 + 1.3 42 + 1.3
20 kHz - 50 kHz 40 + 2.0 44 + 2.0 63 + 2.0
50 kHz - 100 kHz 53 + 2.5 57 + 2.5 72 + 2.5
100 kHz - 300 kHz 110 + 4.0 130 + 4.0 210 + 4.0
300 kHz - 500 kHz 370 + 6.0 450 + 8.0 740 + 8.0
500 kHz - 1 MHz 1600 + 6.0 2000 + 8.0 3400 + 8.0
22 mV
10 Hz - 20 Hz 69 + 1.3 72 + 1.3 75 + 1.3
20 Hz - 40 Hz 34 + 1.3 40 + 1.3 46 + 1.3
40 Hz - 20 kHz 30 + 1.3 36 + 1.3 43 + 1.3
20 kHz - 50 kHz 40 + 2.0 45 + 2.0 64 + 2.0
50 kHz - 100 kHz 53 + 2.5 57 + 2.5 73 + 2.5
100 kHz - 300 kHz 97 + 4.0 110 + 4.0 160 + 4.0
300 kHz - 500 kHz 310 + 6.0 380 + 8.0 610 + 8.0
500 kHz - 1 MHz 1200 + 6.0 1500 + 8.0 2500 + 8.0
70 mV
10 Hz - 20 Hz 60 + 1.5 61 + 1.5 62 + 1.5
20 Hz - 40 Hz 27 + 1.5 30 + 1.5 37 + 1.5
40 Hz - 20 kHz 22 + 1.5 25 + 1.5 34 + 1.5
20 kHz - 50 kHz 34 + 2.0 36 + 2.0 44 + 2.0
50 kHz - 100 kHz 53 + 2.5 54 + 2.5 62 + 2.5
100 kHz - 300 kHz 110 + 4.0 120 + 4.0 170 + 4.0
300 kHz - 500 kHz 270 + 6.0 290 + 8.0 320 + 8.0
500 kHz - 1 MHz 910 + 6.0 970 + 8.0 1200 + 8.0
220 mV
10 Hz - 20 Hz 55 60 + 1.5 61 + 1.5 62 + 1.5
20 Hz - 40 Hz 20 27 + 1.5 29 + 1.5 35 + 1.5
40 Hz - 20 kHz 17 22 + 1.5 24 + 1.5 31 + 1.5
20 kHz - 50 kHz 17 22 + 2.0 24 + 2.0 33 + 2.0
50 kHz - 100 kHz 51 + 2.5 52 + 2.5 59 + 2.5
100 kHz - 300 kHz 100 + 4.0 120 + 4.0 170 + 4.0
300 kHz - 500 kHz 260 + 6.0 290 + 8.0 310 + 8.0
500 kHz - 1 MHz 890 + 6.0 950 + 8.0 1200 + 8.0
6 Fluke Calibration Fluke 5790B Extended Specifications
700 mV
10 Hz - 20 Hz 55 60 + 1.5 61 + 1.5 62 + 1.5
20 Hz - 40 Hz 20 27 + 1.5 29 + 1.5 34 + 1.5
40 Hz - 20 kHz 15 22 + 1.5 24 + 1.5 31 + 1.5
20 kHz - 50 kHz 15 22 + 2.0 24 + 2.0 33 + 2.0
50 kHz - 100 kHz 51 + 2.5 52 + 2.5 59 + 2.5
100 kHz - 300 kHz 100 + 4.0 120 + 4.0 170 + 4.0
300 kHz - 500 kHz 260 + 6.0 270 + 8.0 310 + 8.0
500 kHz - 1 MHz 890 + 6.0 950 + 8.0 1200 + 8.0
2.2 V
10 Hz - 20 Hz 55 60 61 62
20 Hz - 40 Hz 19 26 28 34
40 Hz - 20 kHz 15 20 22 27
20 kHz - 50 kHz 15 21 23 33
50 kHz - 100 kHz 49 50 57
100 kHz - 300 kHz 92 110 160
300 kHz - 500 kHz 220 230 280
500 kHz - 1 MHz 830 890 1200
7 V
10 Hz - 20 Hz 55 60 61 62
20 Hz - 40 Hz 19 27 29 36
40 Hz - 20 kHz 15 20 22 27
20 kHz - 50 kHz 18 23 26 35
50 kHz - 100 kHz 62 64 73
100 kHz - 300 kHz 140 150 180
300 kHz - 500 kHz 360 380 450
500 kHz - 1 MHz 1100 1200 1500
22 V
10 Hz - 20 Hz 55 60 61 62
20 Hz - 40 Hz 19 28 30 37
40 Hz - 20 kHz 15 20 22 27
20 kHz - 50 kHz 18 23 26 35
50 kHz - 100 kHz 62 64 69
100 kHz - 300 kHz 140 150 180
300 kHz - 500 kHz 360 380 450
500 kHz - 1 MHz 1100 1200 1500
70 V [1]
10 Hz - 20 Hz 55 60 62 63
20 Hz - 40 Hz 19 29 31 39
40 Hz - 20 kHz 15 23 25 34
20 kHz - 50 kHz 22 25 27 39
50 kHz - 100 kHz 64 68 85
100 kHz - 300 kHz 140 150 180
300 kHz - 500 kHz 370 390 490
500 kHz - 1 MHz 1100 1200 1500
220 V [1]
10 Hz - 20 Hz 55 61 62 64
20 Hz - 40 Hz 19 30 32 40
40 Hz - 20 kHz 15 23 25 34
20 kHz - 50 kHz 24 30 34 49
50 kHz - 100 kHz 66 69 83
100 kHz - 300 kHz 160 170 220
300 kHz - 500 kHz 410 480 680
700 V
10 Hz - 20 Hz [2] 55 62 63 65
20 Hz - 40 Hz 19 31 33 41
40 Hz - 20 kHz 19 24 25 31
20 kHz - 50 kHz 100 110 140
50 kHz - 100 kHz 390 500 850
1000 V
10 Hz - 20 Hz [2] 55 62 63 65
20 Hz - 40 Hz 19 31 33 41
40 Hz - 20 kHz 19 24 25 31
20 kHz - 50 kHz
[
3]
100 110 140
50 kHz - 100 kHz
[
3]
390 500 850
[1] Inputs >100 kHz and with a V*Hz product >2.2E7 are typical.
[2] Typical specification, as determined by sourcing with the Fluke 5205A Precision Power Amplifier.
[3] Inputs >30 kHz and >750 V are typical, as determined by sourcing with the Fluke 5205A Precision Power Amplifier.
7 Fluke Calibration Fluke 5790B Extended Specifications
AC Non-sine Absolute Voltage
Range [1] Frequency Range 1-Year Absolute, tcal ±5 °C,
±(% of output)
2.2 mV
10 Hz - 45 Hz 0.1
45 Hz - 1 kHz 0.1
1kHz -20kHz 0.17
20 kHz - 100 kHz 0.5
7 mV
10 kHz - 45 Hz 0.1
45 Hz - 1 kHz 0.1
1kHz -20kHz 0.17
20 kHz - 100 kHz 0.5
22 mV
10 Hz - 45 Hz 0.1
45 Hz - 1 kHz 0.1
1kHz -20kHz 0.17
20 kHz - 100 kHz 0.5
70 mV
10 Hz - 45 Hz 0.1
45 kHz - 1 kHz 0.1
1 kHz - 20 kHz 0.17
20 kHz - 100 kHz 0.5
220 mV
10 Hz - 45 Hz 0.1
45 Hz - 1 kHz 0.1
1 kHz - 20 kHz 0.17
20 kHz - 100 kHz 0.5
700 mV
10 Hz - 45 Hz 0.1
45 Hz - 1 kHz 0.1
1kHz -20kHz 0.17
20 kHz - 100 kHz 0.5
2.2 V [2]
10 Hz - 45 Hz 0.1
45 Hz - 1 kHz 0.1
1kHz -20kHz 0.17
20 kHz - 100 kHz 0.5
7 V
10 Hz - 45 Hz 0.1
45 Hz - 1 kHz 0.1
1kHz -20kHz 0.17
20 kHz - 100 kHz 0.5
22 V [2]
10 Hz - 45 Hz 0.1
45 Hz - 1 kHz 0.1
1kHz -20kHz 0.17
20 kHz - 100 kHz 0.5
70 V
10 Hz - 45 Hz 0.1
45 Hz - 1 kHz 0.1
1kHz -20kHz 0.17
20 kHz - 100 kHz 0.5
[1] Specifications apply for non-sinusoidal inputs with crest factor <3.0 and with harmonic content band-limited to <1 MHz.
[2] Crest factor limited to <2.3 for signals greater than 75 % of full scale RMS.
8 Fluke Calibration Fluke 5790B Extended Specifications
DC Absolute Voltage
Voltage Range
Absolute ±5 °C of Calibration Tem
p
erature
Measurement Mode ±(
pp
m of Readin
g
+
μ
V)
90 Da
y
s 1 Year 2 Years
220 mV 37 + 1.5 38 + 1.5 43 + 1.5
700 mV 31 + 1.5 33 + 1.5 38 + 1.5
2.2 V 22 24 29
7 V 22 24 29
22 V 25 27 31
70 V 30 32 39
220 V 29 31 38
700 V 39 41 47
1000 V 37 38 44
Note: DC specification valid only when dc input signal is averaged with an equal and opposite dc input signal to eliminate dc
offset errors. The use of Input 1 for dc inputs is not recommended due to the inherent thermal EMFs in a “N”
connector. See Operators Manual for details.
DC Relative Voltage
Voltage Range
Relative ±5 °C of Calibration Tem
p
erature
Measurement Mode ±(
pp
m of Readin
g
+
μ
V)
90 Da
y
s 1 Year 2 Years
220 mV 22 + 1.5 24 + 1.5 31 + 1.5
700 mV 22 + 1.5 24 + 1.5 31 + 1.5
2.2 V 20 22 27
7 V 20 22 27
22 V 20 22 27
70 V 23 25 34
220 V 23 25 34
700 V 24 25 31
1000 V 24 25 31
Note: DC specification valid only when dc input signal is averaged with an equal and opposite dc input signal to eliminate dc
offset errors. The use of Input 1 for dc inputs is not recommended due to the inherent thermal EMFs in a “N”
connector. See Operators Manual for details.
9 Fluke Calibration Fluke 5790B Extended Specifications
Secondary Electrical Specifications
Secondary performance specifications and operating characteristics are included in uncertainty specifications. They are provided for
special calibration requirements such as stability or linearity tests.
AC Secondary Performance
Voltage
Range Frequency Range
24 Hour AC
Stability ± 1 °C
Slow Filter Peak-
Peak ± μV
Tem
p
erature Coefficient [1]
Input
Resistance [2]
10 °C to 40 °C 0 °C to 10 °C
40 °C to 50 °C
pp
m / °C
2.2 mV
10 Hz - 20 Hz 0.4 50 50
20 Hz - 40 Hz 0.4 50 50
40 Hz - 20 kHz 0.4 50 50
20 kHz - 50 kHz 0.4 50 50 10 MΩ
50 kHz - 100 kHz 0.8 75 75
100 kHz - 300 kHz 1.5 100 100
300 kHz - 500 kHz 3.0 150 150
500 kHz - 1 MHz 4.5 200 200
7 mV
10 Hz - 20 Hz 0.4 15 15
20 Hz - 40 Hz 0.4 15 15
40 Hz - 20 kHz 0.4 15 15
20 kHz - 50 kHz 0.4 15 15 10 MΩ
50 kHz - 100 kHz 0.8 25 25
100 kHz - 300 kHz 1.5 60 60
300 kHz - 500 kHz 3.0 80 80
500 kHz - 1 MHz 4.5 125 125
22 mV
10 Hz - 20 Hz 0.4 5 5
20 Hz - 40 Hz 0.4 5 5
40 Hz - 20 kHz 0.4 5 5
20 kHz - 50 kHz 0.4 5 5 10 MΩ
50 kHz - 100 kHz 0.8 8 8
100 kHz - 300 kHz 1.5 10 10
300 kHz - 500 kHz 3.0 40 40
500 kHz - 1 MHz 4.5 100 100
± (ppm of
Readin
g
)
70 mV
10 Hz - 20 Hz 18 5 5
20 Hz - 40 Hz 18 5 5
40 Hz - 20 kHz 18 5 5
20 kHz - 50 kHz 18 5 5 10 MΩ
50 kHz - 100 kHz 24 8 8
100 kHz - 300 kHz 24 10 10
300 kHz - 500 kHz 48 30 30
500 kHz - 1 MHz 150 75 75
220 mV
10 Hz - 20 Hz 12 1.5 3.0
20 Hz - 40 Hz 81.5 3.0
40 Hz - 20 kHz 81.5 3.0 10 MΩ
20 kHz - 50 kHz 82.0 3.0
50 kHz - 100 kHz 18 5.0 8.0
100 kHz - 300 kHz 24 10.0 10.0
300 kHz - 500 kHz 36 20.0 20.0
500 kHz - 1 MHz 120 50.0 50.0
700 mV
10 Hz - 20 Hz 8 1.5 3.0
20 Hz - 40 Hz 61.5 3.0
40 Hz - 20 kHz 61.5 3.0 10 MΩ
20 kHz - 50 kHz 62.0 3.0
50 kHz - 100 kHz 12 5.0 8.0
100 kHz - 300 kHz 18 10.0 10.0
300 kHz - 500 kHz 36 20.0 20.0
500 kHz - 1 MHz 96 50.0 50.0
2.2 V
10 Hz - 20 Hz 8 1.5 3.0
20 Hz - 40 Hz 51.5 3.0
40 Hz - 20 kHz 51.5 3.0
20 kHz - 50 kHz 52.0 3.0 10 MΩ
50 kHz - 100 kHz 10 5.0 8.0
100 kHz - 300 kHz 18 10.0 10.0
300 kHz - 500 kHz 30 20.0 20.0
500 kHz - 1 MHz 90 50.0 50.0
10 Fluke Calibration Fluke 5790B Extended Specifications
Voltage
Range Frequency Range
24 Hour AC
Stability ± 1 °C
Slow Filter Peak-
Peak ± μV
Tem
p
erature Coefficient [1]
Input
Resistance [2]
10 °C to 40 °C 0 °C to 10 °C
40 °C to 50 °C
pp
m / °C
7 V
10 Hz - 20 Hz 8 1.5 3.0
20 Hz - 40 Hz 51.5 3.0
40 Hz - 20 kHz 51.5 3.0
20 kHz - 50 kHz 52.0 3.0 50 kΩ
50 kHz - 100 kHz 10 5.0 8.0
100 kHz - 300 kHz 18 15.0 15.0
300 kHz - 500 kHz 30 30.0 30.0
500 kHz - 1 MHz 90 65.0 65.0
22 V
10 Hz - 20 Hz 8 1.5 3.0
20 Hz - 40 Hz 51.5 3.0
40 Hz - 20 kHz 51.5 3.0
20 kHz - 50 kHz 52.0 3.0 50 kΩ
50 kHz - 100 kHz 10 5.0 8.0
100 kHz - 300 kHz 18 15.0 15.0
300 kHz - 500 kHz 30 30.0 30.0
500 kHz - 1 MHz 90 65.0 65.0
70 V [3]
10 Hz - 20 Hz 8 1.5 3.0
20 Hz - 40 Hz 51.5 3.0
40 Hz - 20 kHz 51.5 3.0
20 kHz - 50 kHz 52.0 3.0 50 kΩ
50 kHz - 100 kHz 18 5.0 8.0
100 kHz - 300 kHz 36 15.0 15.0
300 kHz - 500 kHz 48 40.0 40.0
500 kHz - 1 MHz 120 75.0 75.0
220 V [3]
10 Hz - 20 Hz 8 1.5 3.0
20 Hz - 40 Hz 51.5 3.0
40 Hz - 20 kHz 51.5 3.0
20 kHz - 50 kHz 52.0 3.0 50 kΩ
50 kHz - 100 kHz 18 5.0 8.0
100 kHz - 300 kHz 36 15.0 15.0
300 kHz - 500 kHz 48 40.0 40.0
700 V
10 Hz - 20 Hz [4] 81.54.0
20 Hz - 40 Hz 51.5 4.0
40 Hz – 20 kHz 51.5 4.0 500 kΩ
20 kHz – 50 kHz 18 5.0 7.0
50 kHz - 100 kHz 36 15.0 15.0
1000 V
10 Hz - 20 Hz [4] 81.54.0
20 Hz - 40 Hz 51.5 4.0
40 Hz - 20 kHz 51.5 4.0 500 kΩ
20 kHz - 50 kHz
[
5]
18 5.0 7.0
50 kHz - 100 kHz
[
5]
36 15.0 15.0
[1] Add to uncertainty when more than 5 °C from calibration temperature.
[2] Input capacitance approximately 100 pF.
[3] Inputs with a V*Hz product >2.2 E7 are unspecified.
[4] Typical specification, as determined by sourcing with the Fluke 5205A Precision Power Amplifier.
[5] Inputs that are >30 kHz and >750 V are typical, as determined by sourcing with the Fluke 5205A Precision Power
Amplifier.
11 Fluke Calibration Fluke 5790B Extended Specifications
DC Secondary Performance
Voltage Range
Tem
p
erature Coefficient [1]
Input
Resistance [2]
10 °C to 40 °C 0 °C to 10 °C
40 °C to 50 °C
ppm / °C
220 mV 1.5 3.0 10 MΩ
700 mV 1.5 3.0 10 MΩ
2.2 V 1.5 3.0 10 MΩ
7 V 1.5 3.0 50 kΩ
22 V 1.5 3.0 50 kΩ
70 V 1.5 3.0 50 kΩ
220 V 1.5 3.0 50 kΩ
700 V 1.5 4.0 500 kΩ
1000 V 1.5 4.0 500 kΩ
[1] Add to uncertainty when more than 5 °C from calibration temperature.
[2] Input capacitance approximately 100 pF.
Note: DC specification valid only when dc input signal is averaged with an equal and opposite dc input signal to eliminate dc
offset errors. The use of Input 1 for dc inputs is not recommended due to the inherent thermal EMFs in a “N” connector.
See Operators Manual for details.
12 Fluke Calibration Fluke 5790B Extended Specifications
O
perating Characteristics
Maximum Non-destructive Input ..................... 1200 V rms
Guard Isolation ................................................ 10 V peak
Volt-Hertz Product .......................................... 1 x 108
Frequency Accuracy (from 0 °C to 50 °C)
10 Hz - 120 Hz ............................................ 100 ppm + 10 digits
Above 120 Hz .............................................. 100 ppm + 2 digits
Frequency Resolution ..................................... 1.00 Hz to 119.99 Hz
0.1200 kHz to 1.1999 kHz
1.200 kHz to 11.999 kHz
12.00 kHz to 119.99 kHz
0.1200 MHz to 1.0000 MHz
1.0000 MHz to 1.1999 MHz (Wideband only)
1.200 MHz to 11.999 MHz (Wideband only)
12.00 MHz to 30.00 MHz (Wideband only)
30.00 MHz to 50.00 MHz (Wideband 5790B/5 & 5790B/AF only)
Reading Rate
<40 Hz ........................................................ 2 seconds per reading
40 Hz ........................................................... 2 seconds decreasing linearly to 1 second at 200 Hz
>200 Hz ...................................................... 1 second per reading
Maximum Settling Time to Full Specifications (in range lock)
Filter Off ...................................................... 1 sample
dc ............................................................ 6 seconds
<200 Hz .................................................. 8 seconds
>200 Hz .................................................. 4 seconds
Filter Fast .................................................... 4 averaged samples
dc ............................................................ 10 seconds
<200 Hz .................................................. 16 seconds
>200 Hz .................................................. 8 seconds
Filter Medium .............................................. 16 averaged samples
dc ............................................................ 22 seconds
<200 Hz .................................................. 32 seconds
>200 Hz .................................................. 16 seconds
Filter Slow ................................................... 32 averaged samples
dc ............................................................ 40 seconds
<200 Hz .................................................. 64 seconds
>200 Hz .................................................. 32 seconds
Filter Buffer Restart Limits:
Fine: Fast: 10 counts
Medium/Slow
<220 mV ............................................. 10 counts
>220 mV ............................................. 100 counts
Medium: Fast: 100 counts
Medium/Slow
<220 mV ............................................. 100 counts
>220 mV ............................................. 1000 counts
Course: Fast: 1000 counts
Medium/Slow
<220 mV ............................................. 1000 counts
>220 mV ............................................. 10000 counts
Specified for sinewave with THD less than 1 %
A
UX Input Characteristics
The AUX input can be used with the Fluke A40/A40A Series Current Shunts to make relative ac current
measurements. The 5790A-7001 A40/A40A Current Shunt Adapter and Cable are required. For optimal current
measurements using shunts,
see the Operators Manual.
Input Resistance .......................................... 91 Ω ± 1 %
Operating Input Voltage .............................. 250 mV to 500 mV
Maximum Non-Destructive Input ................. 50 V rms
13 Fluke Calibration Fluke 5790B Extended Specifications
Wideband Specifications (5790B/3,
5790B/5, and 5790B/AF)
Voltage [1]
Range Frequency Range
Flatness [2]
1 year ±
3 °C ± (% of
Reading
+ μV)
Flatness [3]
Temperature
Coefficient
ppm / °C
Absolute Uncertainty 0 °C to 50 °C [4]
± (% of Reading + μV)
Resolution
90 Days 1 Year 2 Years
2.2 mV
10 Hz - 30 Hz 0.10 + 0 75 0.5 + 1.2 0.6 + 1.5 0.8 + 2
30 Hz - 120 Hz 0.05 + 0 75 0.5 + 1.2 0.6 + 1.5 0.8 + 2
120 Hz - 1.2 kHz 0.05 + 0 75 0.5 + 1.2 0.6 + 1.5 0.8 + 2
1.2 kHz - 120 kHz 0.05 + 0 75 0.5 + 1.2 0.6 + 1.5 0.8 + 2
120 kHz - 500 kHz 0.07 + 1 75 0.5 + 1.2 0.6 + 1.5 0.8 + 2 0.1
μ
V
500 kHz - 1.2 MHz 0.07 + 1 75
1.2 MHz - 2 MHz 0.07 + 1 100
2 MHz - 10 MHz 0.17 + 1 200
10 MHz - 20 MHz 0.30 + 1 200
20 MHz - 30 MHz 0.70 + 2 400
30 MHz - 50 MHz
[
5]
1.00 + 2 400
7 mV
10 Hz - 30 Hz 0.10 + 0 75 0.4 + 5 0.5 + 7 0.7 + 8
30 Hz - 120 Hz 0.05 + 0 75 0.4 + 5 0.5 + 7 0.7 + 8
120 Hz - 1.2 kHz 0.05 + 0 75 0.4 + 5 0.5 + 7 0.7 + 8
1.2 kHz - 120 kHz 0.05 + 0 75 0.4 + 5 0.5 + 7 0.7 + 8
120 kHz - 500 kHz 0.07 + 1 75 0.4 + 5 0.5 + 7 0.7 + 8 0.1
μ
V
500 kHz - 1.2 MHz 0.07 + 1 75
1.2 MHz - 2 MHz 0.07 + 1 100
2 MHz - 10 MHz 0.1 + 1 200
10 MHz - 20 MHz 0.17 + 1 200
20 MHz - 30 MHz 0.37 + 1 300
30 MHz - 50 MHz
[
5]
0.5 + 1 300
22 mV
10 Hz - 30 Hz 0.10 75 0.4 + 10 0.5 + 13 0.7 + 16
30 Hz - 120 Hz 0.05 75 0.4 + 10 0.5 + 13 0.7 + 16
120 Hz - 1.2 kHz 0.05 75 0.4 + 10 0.5 + 13 0.7 + 16
1.2 kHz - 120 kHz 0.05 75 0.4 + 10 0.5 + 13 0.7 + 16
120 kHz - 500 kHz 0.07 75 0.4 + 10 0.5 + 13 0.7 + 16 0.1
μ
V
500 kHz - 1.2 MHz 0.07 75
1.2 MHz - 2 MHz 0.07 75
2 MHz - 10 MHz 0.1 100
10 MHz - 20 MHz 0.17 100
20 MHz - 30 MHz 0.37 200
30 MHz - 50 MHz
[
5]
0.6 200
70 mV
10 Hz - 30 Hz 0.10 40 0.4 + 20 0.5 + 30 0.6 + 40
30 Hz – 120 Hz 0.05 40 0.4 + 20 0.5 + 30 0.6 + 40
120 Hz - 1.2 kHz 0.05 40 0.4 + 20 0.5 + 30 0.6 + 40
1.2 kHz - 120 kHz 0.05 40 0.4 + 20 0.5 + 30 0.6 + 40
120 kHz - 500 kHz 0.05 40 0.4 + 20 0.5 + 30 0.6 + 40 1.0
μ
V
500 kHz - 1.2 MHz 0.05 40
1.2 MHz - 2 MHz 0.05 75
2 MHz - 10 MHz 0.1 100
10 MHz - 20 MHz 0.15 100
20 MHz - 30 MHz 0.35 200
30 MHz - 50 MHz
[
5]
0.6 200
220 mV
10 Hz - 30 Hz 0.10 40 0.3 + 60 0.4 + 80 0.5 + 100
30 Hz - 120 Hz 0.04 40 0.3 + 60 0.4 + 80 0.5 + 100
120 Hz - 1.2 kHz 0.04 40 0.3 + 60 0.4 + 80 0.5 + 100
1.2 kHz - 120 kHz 0.04 40 0.3 + 60 0.4 + 80 0.5 + 100
120 kHz - 500 kHz 0.04 40 0.3 + 60 0.4 + 80 0.5 + 100 1.0
μ
V
500 kHz - 1.2 MHz 0.05 40
1.2 MHz - 2 MHz 0.05 75
2 MHz - 10 MHz 0.1 100
10 MHz - 20 MHz 0.15 100
20 MHz - 30 MHz 0.35 200
30 MHz - 50 MHz [5] 0.6 200
14 Fluke Calibration Fluke 5790B Extended Specifications
700 mV
10 Hz - 30 Hz 0.10 40 0.3 + 200 0.4 + 300 0.5 + 400
30 Hz - 120 Hz 0.03 40 0.3 + 200 0.4 + 300 0.5 + 400
120 Hz - 1.2 kHz 0.03 40 0.3 + 200 0.4 + 300 0.5 + 400
1.2 kHz - 120 kHz 0.03 40 0.3 + 200 0.4 + 300 0.5 + 400
120 kHz - 500 kHz 0.03 40 0.3 + 200 0.4 + 300 0.5 + 400 10.0
μ
V
500 kHz - 1.2 MHz 0.05 40
1.2 MHz - 2 MHz 0.05 75
2 MHz - 10 MHz 0.1 100
10 MHz - 20 MHz 0.15 100
20 MHz - 30 MHz 0.35 200
30 MHz - 50 MHz
[
5]
0.6 200
2.2 V
10 Hz - 30 Hz 0.10 40 0.3 + 300 0.35 + 400 0.4 + 500
30 Hz - 120 Hz 0.03 40 0.3 + 300 0.35 + 400 0.4 + 500
120 Hz - 1.2 kHz 0.03 40 0.3 + 300 0.35 + 400 0.4 + 500
1.2 kHz - 120 kHz 0.03 40 0.3 + 300 0.35 + 400 0.4 + 500
120 kHz - 500 kHz 0.03 40 0.3 + 300 0.35 + 400 0.4 + 500 10.0
μ
V
500 kHz - 1.2 MHz 0.05 40
1.2 MHz - 2 MHz 0.05 75
2 MHz - 10 MHz 0.1 100
10 MHz - 20 MHz 0.15 100
20 MHz - 30 MHz 0.35 200
30 MHz - 50 MHz
[
5]
0.6 200
7 V
10 Hz - 30 Hz 0.10 40 0.3 + 500 0.35 + 800 0.4 + 1000
30 Hz - 120 Hz 0.03 40 0.3 + 500 0.35 + 800 0.4 + 1000
120 Hz - 1.2 kHz 0.03 40 0.3 + 500 0.35 + 800 0.4 + 1000
1.2 kHz - 120 kHz 0.03 40 0.3 + 500 0.35 + 800 0.4 + 1000
120 kHz - 500 kHz 0.03 40 0.3 + 500 0.35 + 800 0.4 + 1000 100.0
μ
V
500 kHz - 1.2 MHz 0.05 40
1.2 MHz - 2 MHz 0.05 75
2 MHz - 10 MHz 0.1 100
10 MHz - 20 MHz 0.15 100
20 MHz - 30 MHz 0.35 200
30 MHz - 50 MHz
[5],[6] 0.6 200
[1] Range limits same as INPUT 1 or INPUT 2.
[2] Relative to 1 kHz, for 2-year specification multiply by 1.5.
[3] Add to flatness specifications when more than 3 °C from calibration temperature.
[4] At input connector.
[5] Applies to 5790B/5 & 5790B/AF only.
[6] Maximum amplitude is limited to 3.5 V.
Wideband Characteristics
Maximum Non-Destructive Input .................... 200 V rms
Guard Isolation ................................................ 0.5 V peak
Input Impedance
1 kHz ........................................................... 50 Ω (± 0.5 %)
VSWR .......................................................... <1.05 typical
5790B/AF
The 5790B/AF absolute specification is ±0.23 % of voltage reading (1 year, 23 °C ±3 °C, 95 % confidence level
(k=2), normally distributed). Specification applies to 50 MHz, 223.61 mV, referenced to the end of the provided
serialized 0.91 meter (3 ft) cable. When using the cable and 50 MHz Cable Correction, other ranges and
frequencies can be measured but the Product is only specified within ±4 % of frequency and amplitude stated.
15 Fluke Calibration Fluke 5790B Extended Specifications
Ordering information
Fluke Calibration. Precision, performance, confidence.™
Fluke Calibration
PO Box 9090,
Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (877) 355-3225 or Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or Fax (905) 890-6866
From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116
Web access: http://www.flukecal.com
©2015 Fluke Calibration. Specifications subject to change without notice.
Printed in U.S.A. 5/2015 6005014a-en
Pub-ID 13404-eng
Modification of this document is not permitted without written permission
from Fluke Calibration.
Models
5790B AC Measurement Standard
5790B/03 AC Measurement Standard
+30 MHZ Wideband Option
5790B/05 AC Measurement Standard
+ 50 MHz Wideband Option
5790B/AF AC Measurement Standard
+ 50 MHz Wideband Option Cali-
brated with WB cable
Accessories
5790B/5 WB Cable Cable, 50 Ohm Wideband Cable
for 50 MHz Measurements
A40B/Set Complete set of A40B shunts,
transit case and accessories
A40B-001MA 1 milliamp Current Shunt
A40B-010MA 10 milliamp Current Shunt
A40B-020MA 20 milliamp Current Shunt
A40B-050MA 50 milliamp Current Shunt
A40B-100mA 100 milliamp Current Shunt
A40B-200mA 200 milliamp Current Shunt
A40B-500mA 500 milliamp Current Shunt
A40B-1A 1 amp Current Shunt
A40B-2A 2 amp Current Shunt
A40B-5A 5 amp Current Shunt
A40B-10A 10 amp Current Shunt
A40B-20A 20 amp Current Shunt
A40B-50A 50 amp Current Shunt
A40B-100A 100 amp Current Shunt
A40B-LEAD/4MM Cable N-Type connector to
Dual Terminal Plugs
A40B-LEAD/N Cable N-Type connector to
N-Type Connector
The broadest range
of calibration solutions
Fluke Calibration provides the broadest range of cali-
brators and standards, software, service, support and
training in electrical, temperature, pressure, RF and
flow calibration.
Visit www.flukecal.com for more information about
Fluke Calibration products and services.