Technical Data
8508A Reference Multimeter
accessory guide
Introduction
The 8508A Reference Multimeter is
complemented by a comprehensive
range of accessories, all of which are
designed to help you get the most
from your reference multimeter.
This datasheet explains in more
detail the make-up of each accessory
and offers technical specifications
where required.
Accessories currently available for
the 8508A Reference Multimeter
include:
•8508A-LEAD
Comprehensive Measurement
Lead Kit
•8508-7000K
Calibration Kit
•8508A-SPRT
Standard Platinum Resistance
Thermometer, 25.5 ohm
•8508A-PRT
Platinum Resistance Thermometer,
100 W
•Y8508
Rack Mount Kit
•Y8508S
Rack Mount Kit Slides
•NVLAP
NVLAP Accredited Calibration
Ordering Information Item Number
4022363
•UKAS
UKAS Accedited Calibration
Ordering Information Item Number
1883673
8508A-LEAD
Qty. Description
41.5 m two core screened low thermal cable with gold plated
6 mm copper spade terminals
11 m screened 322/0.1 copper core (30 Amp rating) with 6
mm gold plated coper spade terminals
1General purpose probe kit and wallet includes:
2x test leads with 4 mm safety shroud plugs (red and black)
2x test probes (red and black)
2x crocodile clips (red and black)
2x 4-wire zero PCB
The 8508A Reference Multime-
ter has excellent performance
specifications. But stability and
accuracy can often be compro-
mised by the use of inappropriate
connecting leads and associ-
ated processes when performing
measurements.
The 8508A-LEAD set includes
high quality, low leakage,
low dielectric absorption PTFE
insulated cables, terminated
with 6 mm gold plated copper
spade terminals. Their insula-
tion properties make them
ideal for resistance and current
measurements minimizing leak-
age current and charge storage
problems, their low capacitance
makes them ideal for ac applica-
tions, and the 6 mm gold plated
copper spade terminals ensure
low thermal emfs for dc voltage
and resistance work.
A general purpose lead kit
which includes test probes
and alligator clips provides the
convenience of making rapid
connections for less demanding
applications and troubleshooting.
It also includes two four-wire
shorting PCBs (used to null front
and rear inputs).
8508A-7000K
The Fluke 8508A-7000K Cali-
bration Kit contains a specially
designed lead set, 1 GW stan-
dard resistor and two four-wire
shorting P.C.B.s for performance
verification and calibration
adjustment of the 8508A. The
lead set allows the 8508A to be
connected to a 5720A calibra-
tor and a 5725A amplifier in
the configuration described in
the 8508A User Manual Calibra-
tion and Verification chapter,
enabling users to exactly dupli-
cate the setup used for factory
calibration.
The leads are constructed from
low capacitance, low leakage
dielectric material (PTFE), with
low thermal emf 6 mm gold
plated copper spade termina-
tions. The 1 GW standard resistor
plugs directly into the 8508A
front input terminals without any
inter-connecting leads, minimis-
ing errors from lead effects and
noise pickup. It is specified for
stability with time (100 ppm/yr)
and temperature (25 ppm/ °C),
and is supplied with a calibration
certificate stating its measured
value to an uncertainty of
110 ppm.
8508A-LEAD kit contents
2 Fluke Calibration 8508A Reference Multimeter accessory guide
8508A-SPRT
Measure temperatures as high
as 661 °C
Inconel™ and platinum sheaths
guard against contamination
Less than 10 mK/year drift
Fifth wire provides shield
ground
8508A-PRT
Range to 660 °C
Meets all ITS-90 requirements
for resistance ratios
Rtp drift < 30 mK after 500
hours at 660 °C
Strain-free sensor design
The Fluke 8508A-SPRT Extended
Range Metal-Sheath SPRT and
8508A-PRT Secondary Standard
PRT Temperature Sensors have
8508A-7000K (cont.)
Quantity Description
11 GΩ resistor – 100 ppm/yr and 25 ppm/°C stability,
factory calibration uncertainty 110 ppm (95 %CL)
1Screened lead set that connects 8508A Reference
Multimeter to 5720A and 5725A Precision Calibrator
and Amplifier
2Four-wire shorting P.C.B. (used to null front and rear
input)
1Carry case
been designed and manufactured
to complement the Fluke 8508A
Reference Multimeter’s precision
PRT temperature function.
Fluke Calibration
PO Box 9090,
Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (877) 355-3225 or Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or Fax (905) 890-6866
From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116
Web access: http://www.flukecal.com
© 2003, 2010, 2012 Fluke Corporation. Specifications subject to change without notice.
Printed in U.S.A. 1/2012 2091295C A-EN-N Pub-ID 10754-eng
Modification of this document is not permitted without written permission from Fluke Corporation.
Fluke Calibration. Precision, performance, confidence.
8508A-SPRT 8508A-PRT
Temperature range -200 °C to 661 °C -200 °C to 660 °C
Nominal RTFW 25.5 Ω (± 0.5 Ω) 100 Ω (± 1 Ω)
Calibration
uncertainty ± 2.1 mK @ 660 °C ± 15 mK @ 661 °C
Sensitivity 0.1 Ω/°C 0.4 Ω/ °C
Drift rate <10 mK per year
max. 2-3 mK per year
typical
< 30 mK per
500 hours at 660°C
Stability ± 3 mK ± 3 mK
Protective sheath Iconel™ 600
5.56 mm dia. x
482 mm L (0.219 in
dia. x 19 in L)
Iconel™ 600
6.35 mm dia. x
381 mm L (0.25 in
dia. x 15 in L)