Agilent 4155C Semiconductor
Parameter Analyzer
Agilent 4156C Precision Semiconductor
Parameter Analyzer
Data Sheet

Introduction
Agilent 4155C and 4156C
Basic Functions
• Set measurement and/or stress
conditions
• Control measurement and/or stress
execution
• Perform arithmetic calculations
• Display measured and calculated
results on the LCD display
• Perform graphical analysis
• Store and recall measurement
setups, and measurement and
graphical display data
• Dump to printers or plotters for
hardcopy output
• Perform measurement and analysis
with built-in instrument BASIC
• Self test, Auto calibration

Configuration
The 4155C and 4156C both come
standard with I/CV 2.1 Lite
automation software. A PC-based
instrument controller with I/CV Lite
preinstalled and an Agilent 82357A
USB/GPIB interface are also included
with the standard configuration. You
have the option of deleting the
controller and cable from your order,
but I/CV Lite is always included with

both instruments. If you want the full
version of I/CV 2.1, you can request
the E5240BU upgrade kit when you
order a 4155C or 4156C. For more
information about the differences
between I/CV 2.1 Lite and I/CV 2.1,
please refer to the Agilent I/CV 2.1
Technical Overview, publication
number 5988-8474EN.
4155C

4156C

4xMPSMU

4xHRSMU

2xVMU
2xVSU
I/CV 2.1 Lite

2xVMU
2xVSU
I/CV 2.1 Lite

Standard PC-based controller
and USB/GPIB interface

SMU: Source Monitor Unit
Display resolution: 6 digits at each
current range (0.01fA display resolution at 10pA range) 2
HRSMU: High Resolution SMU
(1fA/2µV to 100mA/100V)
MPSMU: Medium Power SMU
(10fA/2µV to 100mA/100V)
HPSMU: High Power SMU
(10fA/2µV to 1A/200V)
VMU: Voltage Monitor Unit
(0.2µV resolution in differential
mode)
VSU: Voltage Source Unit
PGU: Pulse Generator Unit (1 channel)
GNDU: Ground Unit
1

Minimum number of installable MPSMU or
PGU is two.

2

Accuracy not guaranteed. Minimum
guaranteedresolution is 1fA at 10pA range.

41501B (Optional)

2xPGU (Option)1
GNDU

HPSMU (Option) or
2xMPSMU (Option)1

S1

Hardware
Specification Condition
The “supplemental” information and
“typical” entries in the following specifications are not warranted, but provide
useful information about the functions
and performance of the instruments.

The measurement and output accuracy
are specified at the rear panel connector terminals when referenced to the
Zero Check terminal under the following con-ditions:
1. 23° C ±5° C (double between 5° C to
18° C, and 28° C to 40° C if not noted
otherwise)
2. After 40 minutes warm-up

3. Ambient temperature change less
than ±1° C after auto calibration
execution.
4. Integration time: medium or long
5. Filter: ON (for SMUs)
6. Kelvin connection (for HRSMU,
HPSMU, and GNDU)
7. Calibration period: 1 year

Agilent 4156C Precision Semiconductor Parameter Analyzer
HRSMU (High Resolution SMU) Specifications

Output terminal/connection:

Voltage Range, Resolution, and Accuracy (HRSMU)

Dual triaxial connectors, Kelvin
(remote sensing)

Voltage
Range
±2V
±20V
±40V
±100V
1
2

Set.
Reso.
100µV
1mV
2mV
5mV

Set.
Accuracy
±(0.02%+400µV)
±(0.02%+3mV)
±(0.025%+6mV)
±(0.03%+15mV)

Meas.
Reso.
2µV
20µV
40µV
100µV

Meas.
Accuracy
±(0.01%+200µV)
±(0.01%+1mV)
±(0.015%+2mV)
±(0.02%+5mV)

Max.
Current
100mA
100mA
1
2

100mA (Vout £20V), 50mA (20V
Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.3
Linearized                      : No
Create Date                     : 2003:04:17 09:59:41
Producer                        : Acrobat Distiller 4.0 for Windows
Creator                         : Windows NT 4.0
Title                           : 4156C_TO_Final421.pdf
Modify Date                     : 2003:04:17 09:59:43-07:00
Page Count                      : 12
EXIF Metadata provided by EXIF.tools

Navigation menu