Anritsu MS4630B User Manual
User Manual: Anritsu MS4630B User Manual
Open the PDF directly: View PDF
.
Page Count: 222
MS4630B
Network Analyzer
Operation Manual
Vol.1
Panel Operation
10th Edition
For safety and warning information, please read this
manual before attempting to use the equipment.
Keep this manual with the equipment.
ANRITSU CORPORATION
Document No.: M-W1534AE-10.0
Safety Symbols
To prevent the risk of personal injury or loss related to equipment malfunction, Anritsu Corporation uses the following
safety symbols to indicate safety-related information. Ensure that you clearly understand the meanings of the
symbols BEFORE using the equipment. Some or all of the following symbols may be used on all Anritsu
equipment. In addition, there may be other labels attached to products that are not shown in the diagrams in this
manual.
Symbols used in manual
DANGER
This indicates a very dangerous procedure that could result in serious injury or
death if not performed properly.
WARNING
This indicates a hazardous procedure that could result in serious injury or death if
not performed properly.
CAUTION
This indicates a hazardous procedure or danger that could result in light-to-severe
injury, or loss related to equipment malfunction, if proper precautions are not taken.
Safety Symbols Used on Equipment and in Manual
The following safety symbols are used inside or on the equipment near operation locations to provide information
about safety items and operation precautions. Ensure that you clearly understand the meanings of the symbols
and take the necessary precautions BEFORE using the equipment.
This indicates a prohibited operation.
symbolically in or near the barred circle.
The prohibited operation is indicated
This indicates an obligatory safety precaution.
indicated symbolically in or near the circle.
The obligatory operation is
This indicates a warning or caution. The contents are indicated symbolically in or
near the triangle.
This indicates a note. The contents are described in the box.
These indicate that the marked part should be recycled.
MS4630B
Network Analyzer
Operation Manual Vol.1 Panel Operation
10
17
November 1998 (First Edition)
December 2007 (10th Edition)
Copyright © 1998-2007, ANRITSU CORPORATION.
All rights reserved. No part of this manual may be reproduced without the prior written permission of the
publisher.
The contents of this manual may be changed without prior notice.
Printed in Japan
ii
For Safety
WARNING
1. ALWAYS refer to the operation manual when working near locations
at which the alert mark shown on the left is attached. If the advice in
the operation manual is not followed there is a risk of personal injury
or reduced equipment performance. The alert mark shown on the
left may also be used with other marks and descriptions to indicate
other dangers.
2. IEC 61010 Standard
The IEC 61010 standard specifies four categories to ensure that an
instrument is used only at locations where it is safe to make
measurements. This instrument is designed for measurement
category I (CAT I). DO NOT use this instrument at locations
specified as category II, III, or IV as defined below.
Measurement category I (CAT I):
Secondary circuits of a device that is not directly connected to a
power outlet.
Measurement category II (CAT II):
Primary circuits of a device that is directly connected to a power outlet,
e.g., portable tools or home appliance.
Measurement category III (CAT III):
Primary circuits of a device (fixed equipment) to which power is
supplied directly from the distribution panel, and circuits running from
the distribution panel to power outlet.
Measurement category IV (CAT IV):
Building service-line entrance circuits, and circuits running from the
service-line entrance to the meter or primary circuit breaker
(distribution panel).
Electric Shock
3. To ensure that the instrument is earthed, always use the supplied 3pin power cord, and insert the plug into an outlet with an earth
terminal. If power is supplied without earthing the equipment, there
is a risk of receiving a severe or fatal electric shock or causing
damage to the internal components.
iii
For Safety
WARNING
iv
Repair
4. This equipment cannot be repaired by the operator. DO NOT attempt
to remove the equipment covers or unit covers or to disassemble
internal components.
Only qualified service personnel with a
knowledge of electrical fire and shock hazards should service this
equipment. There are high-voltage parts in this equipment presenting
a risk of severe injury or fatal electric shock to untrained personnel. In
addition, there is a risk of damage to precision components.
Calibration
5. The performance-guarantee seal verifies the integrity of the equipment.
To ensure the continued integrity of the equipment, only Anritsu service
personnel, or service personnel of an Anritsu sales representative,
should break this seal to repair or calibrate the equipment. If the
performance-guarantee seal is broken by you or a third party, the
performance of the equipment cannot be guaranteed. Be careful not
to break the seal by opening the equipment or unit covers.
Falling Over
6. This equipment should always be positioned in the correct manner.
If the cabinet is turned on its side, etc., it will be unstable and may be
damaged if it falls over as a result of receiving a slight mechanical
shock.
Always set up the equipment in a position where the power switch
can be reached without difficulty.
Battery Fluid
7. DO NOT short the battery terminals and never attempt to disassemble
the battery or dispose of it in a fire. If the battery is damaged by any of
these actions, the battery fluid may leak. This fluid is poisonous.
DO NOT touch the battery fluid, ingest it, or get in your eyes. If it is
accidentally ingested, spit it out immediately, rinse your mouth with
water and seek medical help. If it enters your eyes accidentally, do
not rub your eyes, rinse them with clean running water and seek
medical help. If the liquid gets on your skin or clothes, wash it off
carefully and thoroughly.
LCD
8. This instrument uses a Liquid Crystal Display (LCD). DO NOT subject
the instrument to excessive force or drop it. If the LCD is subjected to
strong mechanical shock, it may break and liquid may leak.
This liquid is very caustic and poisonous.
DO NOT touch it, ingest it, or get in your eyes. If it is ingested
accidentally, spit it out immediately, rinse your mouth with water and
seek medical help. If it enters your eyes accidentally, do not rub
your eyes, rinse them with clean running water and seek medical help.
If the liquid gets on your skin or clothes, wash it off carefully and
thoroughly.
For Safety
CAUTION
Fuse Replacement
1. Always remove the mains power cable from the power outlet before
replacing blown fuses. There is a risk of electric shock if fuses are
replaced with the power cable connected. Always use new fuses of
the type and rating specified on the rear panel of the instrument.
There is a risk of fire if a fuse of a different rating is used.
T5.0A indicates a time-lag fuse.
Cleaning
2. Keep the power supply and cooling fan free of dust.
• Clean the power inlet regularly. If dust accumulates around the
power pins, there is a risk of fire.
Keep
the cooling fan clean so that the ventilation holes are not
•
obstructed. If the ventilation is obstructed, the cabinet may
overheat and catch fire.
3. NEVER touch parts to which the label shown on the left is attached.
These parts are hotter than 55°C and there is a risk of receiving a
burn.
Check Terminal
4. Never input a signal of more than the indicated value between the
measured terminal and ground. Input of an excessive signal may
damage the equipment.
v
For Safety
CAUTION
Replacing Memory
Back-up Battery
This equipment uses a Poly-carbomonofluoride lithium battery to backup
the memory. This battery must be replaced by service personnel when
it has reached the end of its useful life; contact the Anritsu sales section
or your nearest representative.
Note: The battery used in this equipment has a maximum useful life of
7 years. It should be replaced before this period has elapsed.
External
Storage Media
This equipment uses memory cards as external storage media for
storing data and programs.
If this media is mishandled or becomes faulty, important data may be lost.
To prevent this chance occurrence, all important data and programs
should be backed-up.
Anritsu will not be held responsible for lost data.
Pay careful attention to the following points.
• Never remove the memory card from the instrument while it is being
accessed.
• The memory card may be damaged by static electric charges.
• The back-up battery in SRAM memory cards has a finite life.
Replace the battery periodically. For details, refer to the explanation
on the memory card later in this manual.
• Anritsu has thoroughly tested all external storage media shipped with
this instrument. Users should note that external storage media not
shipped with this instrument may not have been tested by Anritsu, thus
Anritsu cannot guarantee the performance or suitability of such media.
Use in a residential
environment
This instrument is designed for an industrial environment.
In a residential environment this instrument may cause radio interference
in which case the user may be required to take adequate measures.
vi
Equipment Certificate
Anritsu Corporation certifies that this equipment was tested before
shipment using calibrated measuring instruments with direct traceability
to public testing organizations recognized by national research
laboratories, including the National Institute of Advanced Industrial
Science and Technology, and the National Institute of Information and
Communications Technology, and was found to meet the published
specifications.
Anritsu Warranty
Anritsu Corporation will repair this equipment free-of-charge if a
malfunction occurs within one year after shipment due to a manufacturing
fault, under the condition that this warranty is void when:
• The fault is outside the scope of the warranty conditions described in
the operation manual.
• The fault is due to mishandling, misuse, or unauthorized modification
or repair of the equipment by the customer.
• The fault is due to severe usage clearly exceeding normal usage.
• The fault is due to improper or insufficient maintenance by the
customer.
• The fault is due to natural disaster including fire, flooding, earthquake,
etc.
• The fault is due to use of non-specified peripheral equipment,
peripheral parts, consumables, etc.
• The fault is due to use of a non-specified power supply or in a nonspecified installation location.
In addition, this warranty is valid only for the original equipment
purchaser. It is not transferable if the equipment is resold.
Anritsu Corporation shall assume no liability for injury or financial loss of
the customer due to the use of or a failure to be able to use this equipment.
Anritsu Corporation Contact
In the event that this equipment malfunctions, contact an Anritsu Service
and Sales office. Contact information can be found on the last page of
the printed version of this manual, and is available in a separate file on
the CD version.
vii
Notes On Export Management
This product and its manuals may require an Export License/Approval by
the Government of the product's country of origin for re-export from your
country.
Before re-exporting the product or manuals, please contact us to confirm
whether they are export-controlled items or not.
When you dispose of export-controlled items, the products/manuals need
to be broken/shredded so as not to be unlawfully used for military purpose.
viii
Crossed-out Wheeled Bin Symbol
Equipment marked with the Crossed-out Wheeled Bin Symbol complies
with council directive 2002/96/EC (the “WEEE Directive”) in European
Union.
For Products placed on the EU market after August 13, 2005, please
contact your local Anritsu representative at the end of the product's
useful life to arrange disposal in accordance with your initial contract and
the local law.
ix
CE Conformity Marking
Anritsu affixes the CE conformity marking on the following product(s) in
accordance with the Council Directive 93/68/EEC to indicate that they
conform to the EMC and LVD directive of the European Union (EU).
CE marking
1. Product Model
Model:
MS4630B Network Analyzer
2. Applied Directive
EMC:
LVD:
Directive 2004/108/EC
Directive 2006/95/EC
3. Applied Standards
• EMC:Emission: EN 61326-1: 2006 (Class A)
Immunity:EN 61326-1: 2006 (Table 2)
IEC 61000-4-2 (ESD)
IEC 61000-4-3 (EMF)
IEC 61000-4-4 (Burst)
IEC 61000-4-5 (Surge)
IEC 61000-4-6 (CRF)
IEC 61000-4-8 (RPFMF)
IEC 61000-4-11 (V dip/short)
Performance Criteria*
B
A
B
B
A
A
B, C
*: Performance Criteria
A: During testing, normal performance within the
specification limits.
B: During testing, temporary degradation, or loss of
function or performance which is self-recovering.
C: During testing, temporary degradation, or loss of
function or performance which requires operator
intervention or system reset occurs.
x
Harmonic current emissions:
EN 61000-3-2: 2006 (Class A equipment)
• LVD: EN 61010-1: 2001 (Pollution Degree 2)
4. Authorized representative
Name:
Address, city:
Country:
Loic Metais
European Quality Manager
ANRITSU S.A. France
16/18 Avenue du Québec SILIC 720 Zone de
Courtaboeuf
91951 Les Ulis Cedex
France
xi
C-tick Conformity Marking
Anritsu affixes the C-tick mark on the following product(s) in accordance
with the regulation to indicate that they conform to the EMC framework
of Australia/New Zealand.
C-tick marking
1. Product Model
Model:
MS4630B Network Analyzer
2. Applied Standards
EMC:Emission: EN 61326-1: 2006 (Class A equipment)
xii
Power Line Fuse Protection
For safety, Anritsu products have either one or two fuses in the AC power
lines as requested by the customer when ordering.
Single fuse:
A fuse is inserted in one of the AC power lines.
Double fuse:
A fuse is inserted in each of the AC power lines.
Example 1: An example of the single fuse is shown below:
Fuse Holder
Example 2: An example of the double fuse is shown below:
Fuse Holders
xiii
xiv
About This Manual
Lineup of Operation Manuals
Two different Operation Manuals (Vol. 1 and Vol. 2) come standard with the MS4630B. Refer to these manuals as
required.
Vol.1
MS4630B
Lineup of Operation
Manuals
Panel Operation
Remote Operation
Vol.2
(GPIB and PTA Remote Control)
Panel Operation:
This manual provides general information about the MS4630B
and detailed information about preparations before use, and
panel operations excluding automatic measurement such as performance tests.
Remote Operation (GPIB Control and PTA Control):
This manual explains IEEE488.2-based GPIB remote control as
the MS4630B conforms to IEEE488.2.
This manual also explains PTA, a function for programming
high-speed control and processing using a high-level language
PTL. It is performed by the personal computer incorporated in
the MS4630B, promoting automatic measurement along with
the GPIB.
I
Table of Contents
For Safety ...........................................................
iii
About This Manual ................................................
I
Section 1
1.1
1.2
1.3
1.4
1.5
Product Overview ..................................................................
Organization of This Manual .................................................
System Configuration ............................................................
Application Parts and Peripheral Devices .............................
Specifications ........................................................................
Section 2
2.1
2.2
2.3
2.4
2.5
4.1
4.2
4.3
4.4
II
2-2
2-3
2-6
2-7
2-10
Basic Operations .............................. 3-1
Turning ON the Power ...........................................................
Displaying the Self-Test Result .............................................
Data Displayed on the Screen ...............................................
Presetting ..............................................................................
Symbols .................................................................................
List of Soft Key Menu Functions ...........................................
The Basics of Key Operations ...............................................
Section 4
1-2
1-3
1-4
1-5
1-6
Preparations before Use .................. 2-1
Environmental Conditions of Installation Site ........................
Safety Measures ....................................................................
Mounting on a Rack and Stacking with Other Units ..............
Preparations to Make before Turning ON the Power ............
Notes on Handling Storage Medium .....................................
Section 3
3.1
3.2
3.3
3.4
3.5
3.6
3.7
Introduction ...................................... 1-1
3-2
3-4
3-6
3-9
3-10
3-12
3-14
Selecting Measurement Items
(Channels Group) ............................. 4-1
Selecting a Measurement Channel (Menu) ...........................
Active Channel (Active) .........................................................
Selecting an Analysis Port (Meas) ........................................
Selecting Measurement Items (Format) ................................
4-2
4-4
4-4
4-6
Section 5
5.1
5.2
5.3
5.4
Setting a Frequency (Frequency, Span) ...............................
Settings Related to Input/Output (Out/Input) .........................
Improving the S/N Ratio and Setting an Delay .....................
Aperture (Group Delay Measurement) (Avg) ........................
Settings Related to Sweep ....................................................
Section 6
6.1
6.2
6.3
6.4
6.5
8.1
8.2
8.3
8.4
9.1
9.2
9.3
6-2
6-3
6-5
6-8
6-9
7-2
7-5
Calibration (Calibration Group) ....... 8-1
Explanation ............................................................................
Normalization (X-S) ...............................................................
Vector Errors Calibration (Cal) ..............................................
Interpolating Calibration Data ................................................
Section 9
5-7
5-10
Marker (Marker Group)..................... 7-1
Setting, Moving, and Displaying Markers (Marker) ...............
Useful Marker Functions (Fctn) .............................................
Section 8
5-2
5-3
Display (Display Group)................... 6-1
Active Trace (Active) .............................................................
Setting a Scale (Scale) ..........................................................
Setting the Display Screen (Trace) .......................................
Displaying All Items at the Touch of a Key (Display All) .......
Turning the Backlight ON/OFF (Backlight ON/OFF) .............
Section 7
7.1
7.2
Selecting Measurement Parameters
(Measurement Group) ...................... 5-1
8-2
8-3
8-4
8-9
Analysis of Measurement Data
(Utility Group) ................................... 9-1
Filter Analysis Function (Filter) ..............................................
Resonator Analysis Function (Resonator) ............................
Limit Test (Limit) ....................................................................
9-2
9-5
9-8
III
Section 10 Hard Copy and Remote Control
(Copy/Memory Group) ..................... 10-1
10.1 Outputting a Hard Copy of Screen Data (Copy) ...................
10.2 Making Settings Related to the Hard Copy and ....................
Remote Control (Copy Control, Local) ..................................
10.3 Save/Recall ...........................................................................
10-2
10-2
10-6
Section 11 System .............................................. 11-1
11.1 Switching between Marker Setting Modes
(Frequency/Point) ..................................................................
11.2 Selecting an Impedance Measurement Method
(Bridge/Transfer Method) ......................................................
11.3 Setting the Date and Time .....................................................
11.4 Checking Options ..................................................................
11.5 Setting a Display Color ..........................................................
11.6 Displaying a Self-test Result .................................................
11-2
11-2
11-3
11-3
11-4
11-4
Section 12 Measurements .................................. 12-1
12.1
12.2
12.3
12.4
12.5
Analyzing a Network ..............................................................
Basics of Measurement .........................................................
Measuring Transfer Characteristics ......................................
Measuring Reflection Characteristics ....................................
Measuring Transfer and Reflection Characteristics
Simultaneously (Option 12) ...................................................
12.6 Measuring a Filter ..................................................................
12.7 Measuring a Resonator .........................................................
12-2
12-6
12-7
12-16
12-24
12-27
12-29
Section 13 Performance Tests ........................... 13-1
13.1 Cases where Performance Tests are Required ....................
13.2 List of Measuring Devices Used for Performance Tests .......
13.3 Performance Test ..................................................................
13-2
13-3
13-4
Section 14 Maintenance ..................................... 14-1
14.1 Cleaning the Cabinet .............................................................
14.2 Notes on Storage ...................................................................
14.3 Repackaging and transporting the System to Return ...........
IV
14-2
14-3
14-4
Appendixes....................................................... App-1
Appendix A Performance Test Result Form ...................................
Appendix B List of Defaults .............................................................
Appendix C List of Controls and Connectors on Front and
Rear Panels .................................................................
Appendix D List of Error Message ..................................................
Appendix E Example of Settings for Video Plotter VP1500 II ........
A-1
B-1
C-1
D-1
E-1
V
VI .
Section 1 Introduction
This section provides a brief overview of the product and explains the manual structure, system configuration, application
parts, peripheral devices, and specifications of this system.
1.1
1.2
1.3
1.4
1.5
Product Overview ......................................................................
Organization of This Manual .....................................................
System Configuration ................................................................
1.3.1 Standard configuration...................................................
1.3.2 Options ..........................................................................
Application Parts and Peripheral Devices .................................
Specifications ............................................................................
1-2
1-3
1-4
1-4
1-4
1-5
1-6
1-1
Section 1 Introduction
1.1
Product Overview
This system is a network analyzer used to measure transmission and reflection (external reflection bridges are also used)
characteristics of electronic parts and circuits over a wide frequency range of 10 Hz to 300 MHz.
This system has a 6.5-inch flat display (640 × 480 dots). This display is a color LCD based on an active matrix drive
system.
The display shows a soft key menu, various measurement graphs, and measurement parameters to allow you to specify
parameters, make measurements, and output measurement results to the printer and plotter.
This system incorporates measurement ports R, TA, and TB (option), allowing you to measure port-to-port ratios and
absolute levels.
This system has a GPIB interface as standard, so you can configure an automatic measurement system easily by connecting a personal computer and other measuring devices. This system also has a PTA (Personal Test Automation) function
as standard, so you can use this system as a controller to construct an automatic measurement system without using an
external personal computer.
■ Applications
This system can be widely used when developing, adjusting, and inspecting electronic parts and devices in the communications market (for mobile and optical communications, etc.) and the AV market (for high-quality TVs and satellite
broadcasting). Especially, the macro analysis function for filters and piezoelectric resonators facilitates and accelerates
measurements.
1-2
1.2
1.2
Organization of This Manual
Organization of This Manual
This manual consists of 14 sections and five appendixes. These sections and appendixes are briefly described below.
Section
Section 1
Introduction
Section 2
Preparations before Use
Section 3
Basic Operations
Section 4
Section 5
Selecting Measurement
Items
Selecting Measurement
Parameters
Section 6
Displays
Section 7
Markers
Section 8
Calibrations
Analysis of measurement
data
Hard copy and Remote
Section 10
control
Section 9
Section 11 System
Section 12 Measurements
Section 13 Performance tests
Section 14 Maintenance
Appendix A
Appendix B
Appendix C
Appendix D
Appendix E
Description
This section provides a brief overview of the product. It also explains the
manual structure, system configuration, application parts, and specifications of this system.
This section explains the safety measures and preparations that must be
made before using (turning on) this system.
This section explains basic operations to the operators who use this system
for the first time.
This section explains to select measurement items.
This section explains to set measurement conditions.
This section explains to set the measurement waveforms for easy observation.
This section explains to read the measurement data by the marker and
explains other marker functions.
This section explains to calibrate the measurement system for eliminating
errors.
This section explains the analysis of filter and resonator and the limit test.
This section explains to output a hard copy of screen data to the printer and
save/recall measurement conditions and to set the GPIB remote control etc.
This section explains the important setting which is not frequently changed
of this equipment.
This section explains the typical measurement.
This section explains the measuring devices, setup, and procedures for
performance tests.
This section explains daily maintenance, long-period storage, repackaging,
and transportation.
Performance Test Result Form
List of Defaults
List of Control and Connectors on Front and Rear Panels
Type and Contents of the Error Message
Example of Settings for Video Plotter VP1500 II
1-3
Section 1 Introduction
1.3
System Configuration
1.3.1
Standard configuration
The table below shows the components included in the standard configuration.
Item
Main unit
Accessories
Type name*1 or
symbol*1
MS4630B
Product name*1
Quantity*1
Network analyzer
Power cord
1
1
F0013
Fuse
2
W1534AE
W1535AE
Operation
Manuals
1 set
Remarks
Two 5 A fuses, for AC line
T5 A250V
Panel Operation
Remote Operation
NOTE:
*1
When you order a system component, let us know its type name (or symbol), product name, and quantity.
1.3.2
Options
Options for this system are as follows:
Option No.
01
02
10
12
13
14
1-4
Product name
PMC interface
RS-232C, Centronics interface
Output attenuator
3-ch receiver
High stability reference oscillator
3-branch output
Quantity
1
1
1
1
1
1
Remarks
FUJISOKU memory card interface
0 to 70 dB, Variable in steps of 10 dB
1.4
1.4
Application Parts and Peripheral Devices
Application Parts and Peripheral Devices
The table below lists the application parts and peripheral devices for the MS4630B. All are optionally available.
Application Parts and Peripheral Devices
Type name*1
or symbol*1
J0007
J0008
P0005
P0006
P0007
P0008
P0009
J0079
J0395
B0334C
B0329C
B0331C
B0333C
MC3305A
MC3306A
VP-1500II
Z0047
62BF50
62B50
62BF75
62B75
MA2201A
MA2202A
MA2203A
MA2301A
MA2302A
MA2303A
MA2204A
MA2403A
MA414A
MA1506A
MA4605A
ME010 Seriese
NOTE:
*1
Product name*1
GPIB connection cable, 1 m
GPIB connection cable, 2 m
Memory card (32K bytes)
Memory card (64K bytes)
Memory card (128K bytes)
Memory card (256K bytes)
Memory card (512K bytes)
Fixed attenuator for high power
Fixed attenuator for high power
Carrying case
Protection cover
Front handles
Rack mount kit
PTA keyboard
PTA keyboard
Video plotter
Paper for UA-455A
Reflection bridge
Reflection bridge
Reflection bridge
Reflection bridge
Reflection bridge
Reflection bridge
Reflection bridge
Reflection bridge
Reflection bridge
Reflection bridge
Impedance probe
Impedance probe
Impedance measuring kit
π-circuit fixture
Impedance convesion adapter
test fixture
Remarks
408JE-101
408JE-102
BS32F1-C-172,
Battery life: About 5 years
BS64F1-C-173,
Battery life: About 5 years
BS128F1-C-174, Battery life: About 4.3 years
BS256F1-C-1175, Battery life: About 2.2 years
BS512F1-C-1176, Battery life: About 1.1 years
DC to 8 GHz, 30 dB, 25 W
DC to 9 GHz, 30 dB, 30 W
With protection cover and casters
A set of two handles
JIS type
ASCII type
A set of 5 rolls
10 to 1000 MHz, BNC-P, 50 Ω, unbalance
10 to 1000 MHz, BNC-J, 50 Ω, unbalance
10 to 1000 MHz, BNC-J, 75 Ω, unbalance
10 to 1000 MHz, BNC-P, 75 Ω, unbalance
10 Hz to 250 kHz, 600 Ω, balance, MA214 terminal
10 Hz to 250 kHz, 600 Ω, balance, MA214 terminal
10 Hz to 250 kHz, 900 Ω, balance, MA214 terminal
2 kHz to 2 MHz, 75 Ω, balance, MA214 terminal
2 kHz to 2 MHz, 135 Ω, balance, MA214 terminal
2 kHz to 2 MHz, 150 Ω, balance, MA214 terminal
30 Hz to 300 kHz, 2 Ω to 1 MΩ
30 kHz to 30 MHz, 2 Ω to 1 MΩ
for MA2403A
DC to 125 MHz, for resonator mesurement
DC to 300 MHz, 50 Ω : 75 Ω unbalance
PIN, AMD, chip inductor, etc.
When you order a product, lest us know its type name (or symbol), product name, and quantity.
1-5
Section 1 Introduction
1.5
Specifications
Measurement
item
Frequency
Input
Transmission characteristic (measurement of ratios):
Magnitude, phase, and group delay
Reflection/impedance characteristics: Magnitude and phase (An external transducer is also used.)
Level characteristic:
Absolute magnitude
Range:
10 Hz to 300 MHz
Resolution:
0.01 Hz
Accuracy (standard)
Aging rate:
≤±10–6/day (15 min. after power-on)
Temperature characteristic: ≤±5 × 10–6 (0 to 50˚C)
Accuracy (Option 13: High-stability reference resonator)
Aging rate:
≤±2 × 10–8/day (24 hours after power-on)
Temperature characteristic: ≤±5 × 10–8/day (0 to 50˚C)
Number of channels
Standard:
2 (R, TA), Option 12: 3 (R, TA, TB)
Impedance:
Switchable between 50 Ω and 1MΩ (between 75 Ω and 1 MΩ
when the MA4605A is used)
Input range (IRG):
0/+20 dBm
Maximum input power
AC: +20 dBm
DC: ±2.2 V (50 Ω)
AC: 0 dBm
DC: ±20 V (1 M Ω)
Connector:
BNC-J
Probe source:
+12 ±1 V, 100 mA (with a circuit for protection against shorts)
Average noise
level
≤–120 dBm (RBW : 1 kHz, 1 to 300 MHz), ≤–110 dBm (RBW : 1 kHz, 80 kHz to 1 MHz)
Cross talk
Between channels:
≥120 dB (80 kHz to 300 MHz), ≥110 dB (to 80 kHz)
Between transmitter and receiver circuits:
≥125 dB
Resolution
bandwidths
3, 10, 30, 100, 300, and 500 Hz, 1, 2, 3, 4, 5, 10, and 20 kHz, and automatically set bandwidth
Output level range
Output A:
Output B (standard):
Output B (Option 14):
Output
Output resolution:
Output level accuracy:
Output level linearity:
Output level deviation:
Step error:
Output impedance:
Connector:
1-6
0 to +21 dBm, Option 10: –70 to +21 dBm
–6 to +15 dBm , Option 10: –76 to +15 dBm
–9.5 to +11.5 dBm, Option 10: –79.5 to +11.5 dBm
(take 6 dB from above value, when uses with MA4605A
75 ohm adapter)
0.01 dB
≤±1.0 dB (Frequency: 100 MHz, A output: +10 dBm)
≤0.5 dB
(at 0 dBm, Frequency: 100 MHz, A output: 0 to +21 dBm)
≤1.5 dB (A output: +10 dBm at 100 MHz)
±0.5 dB (option 10)
50 Ω (75 Ω when the MA4605A is used)
BNC-J
1.5
Magnitude
measurement
Phase
measurement
Group delay
measurement
Calibration
Specifications
Measurement range:
≥120 dB
Measurement resolution:
0.001 dB
Display scale:
0.01 dB/div to 50 dB/div (1-2-5 sequence)
Dynamic accuracy
10 kHz to 300 MHz
80 kHz to 100 MHz
Level relative to IRG
±0.2 dB
±0.2 dB
0 to –10 dB
±0.05 dB
±0.05 dB
–10 to –60 dB
±0.30 dB
±0.10 dB
–60 to –70 dB
±1.00 dB
±0.30 dB
–70 to –80 dB
±4.00 dB
±1.20 dB
–80 to –90 dB
–
±4.00 dB
–90 to –100 dB
Measurement range:
Measurement resolution:
Display scale:
Dynamic accuracy
Level relative to IRG
0 to –10 dB
–10 to –60 dB
–60 to –70 dB
–70 to –80 dB
–80 to –90 dB
–90 to –100 dB
±180˚
0.001˚
0.01 to 50˚/div (1-2-5 sequence)
Measurement range:
DRG:
∆θ:
∆F:
1 ps to 400 ms
∆θ/(360 × ∆F)
Phase measurement range
Frequency span × Smoothing aperture (%)
80 kHz to 100 MHz
±1.5˚
±0.3˚
±0.8˚
±2.0˚
±6.0˚
±20.0˚
10 kHz to 300 MHz
±1.5˚
±0.3˚
±2.0˚
±6.0˚
±20.0˚
–
2
× 100 (%)
number of measurement points
Measurement resolution:
2.78 × 10–5/∆F
Display scale:
1 ps/div to 50 ms/div
Dynamic accuracy:
Phase measurement accuracy/(360 × Aperture frequency)
Calibration type: RESPONSE, 1PORT, 1PATH 2PORT, RESPONSE/ISOLATION, π-NET
Calibration data interpolation: When the measurement frequency or number of measuremenpoints is changed, new calibration data can be obtained by
interpolation according to the old data (except when a log frequency is set or the number of measurement points is 1001).
Normalization:
X-S
Electric length correction range:
0 to ±999999.9999999 m, Resolution: 100 nm
Phase offset range:
±180˚
Smoothing aperture:
20 to
1-7
Section 1 Introduction
Frequency sweep:
Sweep
Display
Marker
Trace data
computation
1-8
Level sweep:
Number of measured points:
Number of breakpoints:
Sweep time:
LIN (CENTER/SPAN, START/STOP),
LOG (START/STOP)
LIN (START/STOP/STEP)
11, 21, 51, 101, 251, 501, 1001
Any number between 1 and 1001
150 us/point, 38 ms/250 points, all sweep (RBW: 20 kHz,
normalization, 1 trace)
1 ms to 27.5 h
Setting range:
Sweep function
Sweep range:
All sweep or partial sweep (between markers), List sweep
Sweep control:
REPEAT/SINGLE, STOP/CONT
Sweep trigger:
INT/EXT (RISE, FALL, LEVEL)
Maximum number of screens displayed:
2 channels, 4 traces
Analysis formats:
LOG MAG (M), PHASE (P), DELAY (D), M/P, M/D, LIN
MAG (LIN), LIN/P, LIN/D, REAL(R), IMAG (I), R/I, Z, Z/θ,
Q, Z/Q, POLAR, VSWR, IMPD (Z∠θ, Rs/Ls, Q/D, R+jx),
ADMT (Y∠θ, Rp/Lp, Cp, Q/D, G+jB)
Display unit:
640 × 480 dots, 6.5-inch, color LCD
Marker function:
NORMAL MKR, ∆MKR, 0MKR, MKR→MAX, MKR→
MIN, MKR→CF, ∆→SPAN, MKR→+PEAK, MKR→–
PEAK, MKR TRACK+PEAK, MKR TRACK–PEAK, MKR
CHANGE, MKR→OFFSET
Setting:
Set each marker position with a frequency or point.
Multi-marker:
A maximum of ten markers can be set for each trace.
Filter functions:
F0, IL, pass band (L, R), attenuation band (L, R), Ripple, Q,
SF
Resonator functions
RESON1:
Fr, Fa, Zr ,Za (0 PHASE) and Fm, Fn, Fm, Fn (MAX/MIN)
RESON2:
Fs, Fr, Fa, Zr, Za, Q, equivalent constants (R1, L1, C1, CO)
Averaging function
System:
SUM, MAX, MIN
Number of times:
1 to 1000
Measurement data memory (Each memory has up to 1001 points and has the same data format
as the anarysis format.)
Main trace (MT) memory: Two for channel 1 and two for channel 2 (XMEM)
Calibration S memory:
Two for channel 1 and two for channel 2 (SMEM)
Image memory:
Two for channel 1 and two for channel 2 (IMEM)
Subtrace (ST):
The following computations can be performed between MT
and ST. (Computations are performed on the data which is the
same as the display data.)
MT→ST, MT = MT–ST, MT = ST
Limit line:
A single limit line or a segment (10) can be set. Judgement
can be passed according to the limit line.
1.5
Measurement
parameter automatic
setting
Resolution bandwidth and sweep time:
Data storage and fetching:
Function memory
Auxiliary
storage device
Hard copy
Input/output
connectors on
rear panel
Specifications
The resolution bandwidth is automatically set for the set sweep
time.
The minimum sweep time is automatically set for the set receive bandwidth.
Able to store the measuring conditions, measurement data, calibration data and PTA application program in the internal
memory, FD and PMC and fetch them.
Up to 100 functions (however, the maximum number of functions depends on the storage capacity)
Storage device and capacity
Internal memory:
1
Capacity:
512 KB (Non-volatile)
3.5-inch FDD:
1 unit
Capacity:
720 KB (2 DD), 1.44 MB (2 HD), MS-DOS format
PMC (Option 01):
1 slot
Capacity:
32 to 512 KB
Hard copy output to video plotte, printer and FD
Reference resonator input
Frequency:
5/10 MHz –10 ppm
Level:
≥0.7 Vp-p (AC connection)
Input impedance:
50 Ω (BNC-J connector)
Reference resonator output
Frequency:
10 MHz
Level:
TTL (DC connection, BNC-J connector)
External trigger input:
TTL level (BNC-J connector)
GPIB:
Conforms to IEEE488.2 (24-pin Amphenol connector)
I/O port:
Parallel interface for PTA (36-pin Amphenol connector)
RGB output:
For external monitor (15-pin D-sub connector)
Video output:
Separate (round DIN 8-pin)
Centronics (option 02):
Parallel interface for printer (25-pin D-sub connector)
RS-232C (option 02):
Serial interface (9-pin D-sub connector)
External
control
Standard:
Power supply
100 to 110/200 to 240 VAC (—15/10%, max. 250 V, auto switching between 100 V and 200V),
47.5 to 63 Hz, ≤180 VA (max.)
Dimensions
and weight
Operating
temperature
EMC
GPIB and PTA, Option 02: RS-232C
426 (W) × 177 (H) × 451 (D) mm, ≤15 kg
0 to +50°C (+4 to 50°C during FDD operation)
Conducted disturbance:
Radiation disturbanc:
Harmonic Current Emission:
Electrostatic Discharge:
Electromagnetic Field Immunity:
Fast Transient / Burst:
Surge:
Conducted RF:
Voltage Dips / Short Interruptions:
EN 61326-1: 2006 (Class A)
EN 61326-1: 2006 (Class A)
EN 61000-3-2: 2006 (Class A)
EN 61326-1: 2006 (Table 2)
EN 61326-1: 2006 (Table 2)
EN 61326-1: 2006 (Table 2)
EN 61326-1: 2006 (Table 2)
EN 61326-1: 2006 (Table 2)
EN 61326-1: 2006 (Table 2)
1-9
Section 1 Introduction
1-10 .
Section 2 Preparations before Use
This section explains the safety measures that must be taken before using the system and the preparations that must also
be made before using the system. These safety measures must be taken to protect the human body and equipment. The
preparations that must be made before use include installation of this system in a rack, piling up this system along with
other units, protective grounding, fuse replacement, and storage medium handling. For how to connect the GPIB cable
and set addresses, see the Operation Manual (GPIB Remote Control).
2.1
2.2
2.3
2.4
2.5
Environmental Conditions of Installation Site ............................
2-2
2.1.1 Places to avoid ..............................................................
2-2
2.1.2 Distance from fan...........................................................
2-2
Safety Measures .......................................................................
2-3
2.2.1 General safety measures related to power supply ........
2-3
2.2.2 Maximum operation level of input terminals ..................
2-4
2.2.3 When the fun stops ........................................................
2-5
Mounting on a Rack and Stacking with Other Units ..................
2-6
2.3.1 Mounting on a rack ........................................................
2-6
2.3.2 Stacking with other units ................................................
2-6
Preparations to Make before Turning ON the Power ................
2-7
2.4.1 Connecting the Power Cord ..........................................
2-8
2.4.2 Replacing fuses .............................................................
2-9
Notes on Handling Storage Medium ......................................... 2-10
2.5.1 Floppy disk (FD) ............................................................ 2-11
2.5.2 Plug-in memory card (PMC) .......................................... 2-13
2-1
Section 2 Preparations before Use
2.1
Environmental Conditions of Installation Site
2.1.1
Places to avoid
This system operates normally within the ambient temperature range of 0 to 50 ˚C(4 to 50 ˚C ,when Floppy disc drive is
operating). To get the most from this system, avoid using it in the following places:
•
Place exposed to strong vibrations
•
Humid or dusty place
•
Place exposed to direct sunlight
•
Place exposed to active gases
To assure stable operation for an extended period, it is recommended that the system be used in a place where the supply
voltage fluctuates little.
CAUTION
Using this unit at room temperatures after using or leaving it at low temperatures for a long period could cause the internal circuits to short because of
condensation.
To prevent this, be sure to turn ON the POWER switch after the system has
been allowed to dry fully.
2.1.2
;
;;;
Distance from fan
This system has a fan on the back panel to prevent the internal temperature from rising excessively. Ensure that the
system is installed with a minimum clearance of 10 cm from the rear vent to the wall, another unit, and so forth.
10 cm or more
Built-in fan
Do not use in the vertical position
2-2
2.2
2.2
Safety Measures
Safety Measures
This section explains the safety measures that must be taken to prevent personal and system damage and interruption of
system operation.
2.2.1
General safety measures related to power supply
WARNING
• Before turning ON the power:
Always establish a protective ground. Turning ON the
power without establishing a protective ground could cause
electric shock which might result in severe injury or loss of
life. Also check the supply voltage. If a high voltage exceeding the specified value is applied, system damage or a fire
could result.
• While turning ON the power:
During maintenance, it is sometimes necessary to check or
adjust the inside of the system with the upper, lower, or side
cover open. Touching a high voltage circuit inside the system carelessly could cause electric shock which might result in severe injury or loss of life. Ask qualified service personnel to perform maintenance.
2-3
Section 2 Preparations before Use
2.2.2
Maximum operation level of input terminals
The maximum operation level of the input terminals of this system is +20 dBm. Do not exceed this level. The maximum
output level of the output terminals is 21 dBm (A output).
CAUTION
The maximum operation level of the input terminals of this system is +20 dBm.
Supplying a signal exceeding this level could burn the input ATT and input
mixer.
is a symbol used to alert operators to a specific operating procedure that, if
not followed, may result in severe damage to the operators and system.
If supplying a signal including direct current component could burn the internal circuit.
These signal could not be measured. Take out direct current component of the
signal before measure it.
MS4630B
Network Analyzer
10Hz–300MHz
Channels
1
Menu
F1
Meas
Shift
Marker
2
Active
Marker
Format
Fctn
7
8
9
BS
4
5
6
MHz
µsec
1
2
3
kHz
msec
mm
•
+/–
Preset
F2
Measurement
Frequency
Span
F3
System
Avg
Remote
Sweep
Out/Input
0
sec
Enter m
F4
Calibration
Utility
Copy/Memory
Control
Limit
Local
F5
Cal
PTA
A
F6
Scale
Backlight Off
Battery
Busy
Display All
X→S
X–S
Filter
Resonator
Active
Trace
Probe
Source
Window Close
Outputs
PMC
Save/
Recall
Copy
Display
B
Keyboard
A (Opt B)
B
R
+12V
100mA
Inputs
TA
TB
Max +20dBm
R
Inputs
TA
Max +20dBm
2-4
TB
2.2
2.2.3
Safety Measures
When the fun stops
WARNING
The device has a fan installed at the rear panel to prevent the interior temperature from rising excessively. When the fan stops for some reasons, the temperature within the unit frame rise to such an extent that it may cause a fire.
For this reason, the device displays a message on the screen warning against
the stoppage of the fan at the rear panel (See (3) message concerning the device abnormality of Appendix D Error Message). When this message is displayed, immediately turn off the power of the device and take out the power
cord from the power outlet. As the unit frame is considered to have reached a
high temperature in this instance, take due precautions when handling it.
2-5
Section 2 Preparations before Use
2.3
Mounting on a Rack and Stacking with Other Units
2.3.1
Mounting on a rack
To mount this system in a rack, a rack mount kit B0333C (option) is required. Mount the system according to the
illustration included in the rack mount kit.
2.3.2
Stacking with other units
To pile up this system and other units with the same width and depth, use linkage plates B0332 (option) for secure
linkage.
As these linkage plates are provided with an auto lock mechanism, the upper and lower units are locked automatically.
2-6
2.4
2.4
Preparations to Make before Turning ON the Power
Preparations to Make before Turning ON the Power
This system operates normally from a 100 Vac +10/–15 % power source. AC power must be supplied after taking
measures to prevent the following:
•
•
•
Injury and loss of life resulting from electric shock
Damage to internal circuits which results from abnormal voltage
Troubles resulting from ground current
The WARNING and CAUTION on the rear panel are used for safety related information to prevent the risk of the
operator's injury.
CAUTION
NO OPERATOR SERVICEABLE PARTS INSIDE.
REFER SERVICING TO
QUALIFIED PERSONNEL.
FOR CONTINUED FIRE
PROTECTION REPLACE
ONLY WITH SPECIFIED
TYPE AND RATED FUSE.
WARNING
CAUTION
Users should not repair the system.
Never open covers and disassemble internal parts. To perform
maintenance, call Anritsu service
personnel, who have undergone
specific training and are thoroughly
acquainted with fire and electric
shocks. Touching a high voltage
circuit inside the system carelessly
could cause electric shock which
might result in severe injury or loss
of life, or it could damage precision
parts.
Be sure to replace fuses
with ones of the specified
type and rating. Using
fuses of a different type or
rating could cause a fire.
Observe the precautions given on the following pages.
2-7
Section 2 Preparations before Use
2.4.1
Connecting the Power Cord
Check that the POWER switch on the front panel is turned off (switched to the (O) side).
Insert the power plug into an outlet, and connect the other end to the power inlet on the rear panel. To ensure that the
instrument is grounded, always use the supplied 3-pin power cord, and insert the plug into an outlet with a ground
terminal. (2) Grounding using a conversion adapter
WARNING
If the power cord is connected without the instrument grounded, there is a risk
of receiving a fatal electric shock. In addition, the peripheral devices connected to the instrument may be damaged.
When connecting to the power supply, DO NOT connect to an outlet without a
ground terminal. Also, avoid using electrical equipment such as an extension
cord or a transformer.
CAUTION
If an emergency arises causing the instrument to fail or malfunction, disconnect the instrument from the power supply by either turning off the POWER
switch on the front panel (switch to the (O) side), or by pulling out the power
cord or the power inlet.
When installing the instrument, place the instrument so that an operator may
easily operate the POWER switch.
If the instrument is mounted in a rack, a power switch for the rack or a circuit
breaker may be used for power disconnection.
2-8
2.4
2.4.2
Preparations to Make before Turning ON the Power
Replacing fuses
WARNING
•
Replacing a fuse with the power on could cause electric shock. Before
replacing a fuse, turn OFF the POWER switch and unplug the power cord
from the wall outlet.
•
Turning ON the power without establishing a protective ground could
cause electric shock.
An abnormal AC supply voltage could damage the internal circuits. After
replacing a fuse, establish the protective ground by one of the methods
explained before, make sure the AC supply voltage is normal, then turn ON
the POWER switch.
CAUTION
If you do not have a spare fuse, use a fuse of the same type and voltage and
current ratings as those of the blown fuse.
•
A fuse of a different type or rating could make it difficult to remove or attach, cause poor contact or delayed blowing.
•
A fuse with too high a voltage or current rating may not blow and thus
cause a fire upon reoccurrence of a fault.
Two 5 A fuses come standard with the system as shown in Section 1.3.1.
Set these fuses in the fuse holders.
If a fuse is blown due to any problem, determine and remedy the cause before replacing the fuse.
Then, replace the fuse following the procedure below.
Step
Operation
1
Set the POWER switch on the front panel to the
wall outlet.
(OFF) position, then unplug the power cord from the
2
Turning the fuse holder cap counterclockwise with a standard screwdriver allows you to remove it along
with the fuse.
3
Remove the fuse from the fuse cap and insert a spare fuse into it (any end first).
4
Replace the fuse cap in the fuse holder and tighten it by turning it clockwise with a standard screwdriver.
2-9
Section 2 Preparations before Use
2.5
Notes on Handling Storage Medium
CAUTION
Do not remove a storage medium, such as a plug-in memory card or floppy
disk, while it is being accessed. While a storage medium is being accessed,
the BUSY lamp is lit. If a storage medium is removed while it is being accessed, the data recorded on it could be destroyed.
This system uses a floppy disk (or an optional plug-in memory card) to store setting data and programs.
Abusing a storage medium could erase the data on it. It is recommended that backup copies of important data be created.
We assume no responsibility for loss of the data stored on storage media.
Notes on handling floppy disks and the plug-in memory card are provided in the following section.
2-10
2.5
2.5.1
Notes on Handling Storage Medium
Floppy disk (FD)
This section explains how to handle 3.5-inch floppy disks, a storage medium used in this system.
Front
Rear
Fig. 2-1
(1)
3.5-inch floppy disk
Handling notes
The plastic case of a 3.5-inch floppy disk has a shutter, which protects the surface of the internal disk. When the
floppy disk is set in the floppy disk drive, the shutter opens automatically, exposing part of the internal disk. Do
not touch (slide) the shutter.
Observe the following precautions:
(a) When the LED on the floppy disk drive is lit, never remove the floppy disk. If removed, the data on it could
be destroyed or the floppy disk drive could break.
(b) Do not touch the magnetic disk surface with a finger or any object.
(c) Do not leave floppy disks in a dusty or humid place.
(d) Do not place floppy disks near a magnetic object.
(e) Do not leave floppy disks in a place exposed to direct sunlight or near a heater.
(f) Store floppy disks in a place where the temperature is 4 to 53 °C and the humidity is 8 to 90 % (take measures
against condensation, as necessary).
2-11
Section 2 Preparations before Use
(2)
Write protect switch
A 3.5-inch floppy disk has a write protect tab so that its contents are not modified or deleted by mistake. To turn
ON write-protection, slide the write protect tab in the direction of the arrow as shown below. (If an attempt is
made to write data with write-protection on, an error will result.)
Fig. 2-2
(3)
Write protect switch of 3.5-inch floppy disk
Setting and removing a 3.5-inch floppy disk
To set a floppy disk in the floppy disk drive, insert it in the direction of the arrow with the top surface left until it
clicks.
To remove a floppy disk from the floppy disk drive, press the EJECT button. Make sure the LED is OFF before
removing the floppy disk.
EJECT button
Shift
Shift
7
8
9
BS
4
5
6
MHz
µsec
1
2
3
kHz
msec
mm
System
0
8
9
BS
4
5
6
MHz
µsec
1
2
3
kHz
msec
mm
•
+/–
System
•
+/–
sec
Enter m
0
Copy/Memory
Control
Copy
Fig. 2-3
sec
Enter m
Copy/Memory
Control
Save/
Recall
Front
2-12
7
Copy
Save/
Recall
Rear
Setting and removing a 3.5-inch floppy disk
LED lamp
2.5
2.5.2
Notes on Handling Storage Medium
Plug-in memory card (PMC)
This section explains the following points about the plug-in memory card. (hereafter called the PMC)
• Inserting in the slot in the main unit
• Handling notes
• Handling the cap
• Loading and replacing a battery
• Using the write protect switch
(1)
Inserting the PCM into the slot in the main unit
CAUTION
Inserting the PCM by force could damage the electrodes. Insert it properly as
shown below.
MS4630B
Network Analyzer
10Hz–300MHz
Channels
1
Menu
F1
Meas
Shift
Marker
2
Active
Marker
Format
Fctn
7
8
9
BS
4
5
6
MHz
µsec
1
2
3
kHz
msec
mm
•
+/–
Preset
F2
Measurement
Frequency
Span
F3
System
Avg
Remote
Sweep
Out/Input
0
sec
Enter m
F4
Calibration
Utility
Copy/Memory
Control
Limit
Local
F5
Cal
PTA
F6
Scale
Backlight Off
Battery
Busy
Display All
X→S
X–S
Filter
Resonator
Active
Trace
Probe
Source
Window Close
Outputs
Keyboard
PMC
Save/
Recall
Copy
Display
A
B
A (Opt B)
B
R
+12V
100mA
Inputs
TA
TB
Max +20dBm
Battery
Busy
PMC
PLUG-IN MEMORY CARD
1)
Hold the PMC with the surface marked
with a green triangle up.
2)
Holding the PMC horizontally, align the
triangle printed on the PMC with the
triangle printed on the front panel of
the main unit.
3)
Slowly insert the PMC straight into the
slot until it stops.
(PMC)
SRAM
Note:
Inserting the PCM by force could damage the connector.
Insert the PMC into the slot in the main unit properly as mentioned above.
2-13
Section 2 Preparations before Use
(2)
Notes on handling the PMC
(a) Do not subject the PMC to strong shock by dropping or bending it.
(b) Do not expose it to water.
(c) Do not expose it to high temperature, high humidity, or direct sunlight.
(d) Do not insert tweezers or the like into the PMC connector.
(e) Protect the PMC connector from dust.
(f) Do not insert any unspecified PMC into the PMC slot.
(g) 128-KB, 256-KB, and 512-KB plug-in memory cards are shipped with a battery that is not loaded. Load it
before use.
(h) Lives of batteries at room temperatures are listed below. When the battery has been used up, the data stored in
the PMC disappears. Replace the battery before it is used up. On the back of the PMC a “Battery replacement
schedule” section is provided. Immediately after loading a new battery, enter the scheduled battery replacement date in this section according to the table below.
PMC形名
BS32F1-C-172
BS64F1-C-173
BS128F1-C-174
BS256F1-C-1175
BS512F1-C-1176
Memory
capacity
32 KB
64 KB
128 KB
256 KB
512 KB
Battery life
Battery
CAUTION
●電池寿命
(32Kバイト)
:約5年(常温)
About 5 years
About 5 years
About 4.3 years
About 2.2 years
About 1.1 years
Battery life:About 5 years
(at room temperature)
BR2325
●機器電源をONにして,
プラグイン状態で電池を交換して
ください。
Battery replacement must be done by inserting
the card into the instrument while the power is on.
●電池はBR2325を使用してください。
Use only BR2325 battery.
●強いショックを与えたり,
折り曲げないこと。
Do not drop or bend.
●高温高湿・直射日光にさらさないこと。
Do not expose to extreme temperature
or wetness.
Immediately after loading a new
battery, enter the scheduled battery replacement date in this section.
次回電池交換予定日
Battery replacement Schedure
Date :
ANRITSU CORP.
MADE IN JAPAN
When the red “Battery” lamp lights up, this indicates that the battery voltage is too low. Replace the battery as soon as
possible.
2-14
2.5
(3)
Notes on Handling Storage Medium
Handling the cap
The cap prevents the PMC from being inserted upside down or right side left. Usually, use the PMC with the cap attached.
However, remove the cap in the following cases:
•
When the shape of the PMC insertion slot in the main unit is as shown below:
Reference: The PMC with a cap can
be inserted into these slots.
PMC
PMC
PMC
PMC
•
The cap must be removed when replacing the battery.
C
A
U
TI
O
N
[Removing the cap]
The cap can be removed easily with its back (“CAUTION” is printed) up.
Cap
While raising the Cap slightly (up to the
position indicated by the dotted line),
slide it in the direction of the arrow.
Tab
2-15
Section 2 Preparations before Use
(4)
Loading and replacing a battery
■ Loading a battery initially (SRAM only)
Do not forget to load the enclosed lithium battery before using the PMC. When loading the battery, place the
PMC with its back (“CAUTION” is printed) up.
Remove the cap from the PMC.
2)
Unlock the battery holder and
draw out the battery holder.
3)
Put the battery in the battery
holder with the + mark up.
4)
Insert the battery holder into the
slot in the PMC, lock it, then
attach the cap.
Battery
C
A
U
TI
O
N
1)
Battery holder
LO
CK
Battery holder lock
LOCK
LOCK
UNLOCK
LOCK
Cap
■ Replacing the battery (SRAM only)
Replace the battery while the system is powered on and with the PMC set in the system.
If a battery is replaced while the system power is OFF or the PMC is removed from the system, the data stored
in the PMC could disappear.
Step
1
2
3
4
5
6
7
2-16
Operation
Have a lithium battery ready.
Turn ON the system.
Remove the cap from the PMC and insert the PMC into the PMC slot (align the green triangle with the
triangle on the front panel of the main unit).
Unlock the battery holder.
Draw out the battery holder and replace the battery with a new one.
Insert the battery holder into the PMC and lock the battery holder.
Remove the PMC from the slot and attach the cap.
2.5
Notes on Handling Storage Medium
1)
Turn ON the POWER switch.
2)
Insert the PMC as mentioned on page 2-13.
3)
Place your finger nail in the groove in the battery
holder and draw out the battery holder.
4)
Put a new battery (BR2325) in the battery holder
and insert the battery holder into the PMC.
PMC slot in main unit
LO
CK
PMC (Plug-in Memory Card)
Battery holder
Battery (BR2325)
(+ mark down)
■ Locking the battery holder
To lock the battery holder, slide the LOCK switch to the left using a ballpoint pen or the like.
LOCK Switch
LOCK
Battery holder
(5)
Using the write protect switch (SRAM only)
The write protect switch is factory-set at the OFF position. To turn ON write-protection, slide the write protect
switch to the ON position using a ballpoint pen or the like.
If write-protection can be turned ON by software, keep the write protect switch set at the OFF position.
Protedted
ON
PROTECT
Not Protected
ON
PROTECT
2-17
Section 2 Preparations before Use
2-18.
Section 3 Basic Operations
This section explains the basic operations of this network analyzer for operators who use it for the first time.
Note that this section covers only the basic operations which are necessary to quickly and easily check the status of the
basic operations and performance of this system.
For advanced operations, see Sections 4 and later.
3.1
3.2
3.3
3.4
3.5
3.6
3.7
Turning ON the Power ...............................................................
Displaying the Self-Test Result .................................................
Data Displayed on the Screen ...................................................
Presetting ..................................................................................
Symbols .....................................................................................
List of Soft Key Menu Functions ................................................
The Basics of Key Operations ...................................................
3.7.1 Hard keys (master keys) ................................................
3.7.2 Soft keys ........................................................................
3.7.3 Operation windows ........................................................
3.7.4 Operations to Perform in Windows ................................
3.7.5 Channels and Traces ....................................................
3-2
3-4
3-6
3-9
3-10
3-12
3-14
3-14
3-14
3-14
3-15
3-16
3-1
Section 3 Basic Operations
3.1
Turning ON the Power
Before turning ON the system, establish a protective ground according to Section 2.2 and then plug the enclosed power
cord into the wall outlet.
WARNING
Turning ON the power without establishing a protective ground could cause
electric shock which might result in severe injury or loss of life. If you do not
have a 3-pole (ground-type, 2-pole) wall outlet, be sure to connect the FG terminal on the rear panel or the ground terminal of the enclosed power cord to
the ground level before turning ON the system.
Std Osc
Ext
Input(5/10MHz)
(50Ω,≥7Vp-p)
Buffer
Output(10MHz)
(TTL Level)
Ext Trigger
(TTL Level)
Centronics
∼LINE Input 47.5-63HZ,150VA Max
100-240V
T 5.0A
RS232C
WARNING
CAUTION
∼LINE Input 47.5-63HZ,150VA Max
100-240V
T 6.3A
Module Bus
PTA I/O Port
RGB
Separate
Video
GPIB
SH1 AH1 T6 L4 SR1 RL1
PPO DC1 DT1 C1-4 C24 E2
Rear panel
Frame ground (FG) terminal: To prevent electric shock,
connect this terminal to the
ground level.
CAUTION
An abnormal AC supply voltage could damage the internal circuits.
Before turning ON the system, make sure that the AC supply voltage as specified (nominal value +10/–15 %, 250 V or lower).
3-2
3.1
Turning ON the Power
The POWER switch is located at the position shown below. The l and o marks above the POWER switch indicate the
power ON and OFF states, respectively.
MS4630B
Network Analyzer
10Hz–300MHz
Channels
1
Menu
F1
Meas
Shift
Marker
2
Active
Marker
Format
Fctn
7
8
9
BS
4
5
6
MHz
µsec
1
2
3
kHz
msec
mm
•
+/–
Preset
F2
Measurement
Frequency
Span
F3
System
Avg
Remote
Sweep
Out/Input
0
sec
Enter m
F4
Calibration
Utility
Copy/Memory
Control
Limit
Local
F5
Cal
PTA
Scale
Backlight Off
Battery
Busy
Display All
X→S
A
F6
X–S
Filter
Resonator
Active
Trace
Probe
Source
Window Close
Outputs
PMC
Save/
Recall
Copy
Display
B
Keyboard
A (Opt B)
B
R
+12V
100mA
Inputs
TA
TB
Max +20dBm
Front panel
POWER switch
Power ON
When the pushbutton is pressed so that it is in the lowered position, the system is powered. Power is supplied to all circuits in
the system. The system is then ready for use.
Power OFF
When the pushbutton is pressed a second time so that it is in
the raised position, the system is turned off.
Step
1
2
3
4
5
Operation
Connect the FG terminal on the rear panel
to the ground level.
Measure the AC supply voltage at the wall
outlet using an AC voltmeter.
Set the POWER switch on
the front panel to the OFF
position.
Insert the jack at one end of the power
cord into the AC inlet on the rear panel.
Insert the plug at the other end of the power cord into the wall outlet.
6
Set the POWER switch on the front panel
to the l position.
Remarks
When you plug a 3-pole power cord with a ground terminal,
the FG terminal need not be connected to the ground level.
The measured value must be with +10/–15 % of the rated
voltage. In Japan, the rated voltage is 100 V.
Press the POWER switch so that it is in the raised position.
Insert the power cord jack as far as it will go.
Press the POWER switch so that it is in the lowered position. Power is supplied to all circuits in the system. The
system is ready for use.
The display goes on, showing the self-test result. (See the
next page.)
The fan on the rear panel starts rotating to expel hot air.
3-3
Section 3 Basic Operations
3.2
Displaying the Self-Test Result
This system performs a self-test on the internal hardware at power-on and displays the result on the display screen along
with the software version information.
The following modules are tested and the result is displayed:
• MAIN CPU
• DISP CPU
• DSP
• LOCAL
• OUTPUT
• RECEIVER (R)
• RECEIVER (TA)
• RECEIVER (TB/opt)
Information about the versions of the following software is displayed:
• OVERALL
• SYSTEM
• MAIN
• DISP
• MEAS
• DSP
• PTA
• PTA USER
Example: The output section is faulty and other sections are normal.
SELF TEST INFORMATION screen
3-4
3.2
NOTE:
1.
Displaying the Self-Test Result
If the self-test result shows any problem or nothing is displayed on the screen, contact Anritsu for repair.
The contact addresses and telephone numbers are given at the end of this manual.
When you contact Anritsu for repair, please provide the following information:
•
•
•
2.
Model and serial number indicated on the rear panel
Symptom
Name of the person in charge and telephone number
Turning on the power with the Preset key pressed down allows you to reset the backup data stored in the
internal memory to the settings on shipping.
If backup data has been destroyed due to an incidental problem and therefore the machine does not start
operating after power-on, following the above procedure may start the machine.
3-5
Section 3 Basic Operations
3.3
Data Displayed on the Screen
If the self-test result is OK, press the WINDOW CLOSE key on the front panel to close the SELF TEST INFORMATION
window and display an ordinary measurement screen.
Let’s take a look at the data displayed on the screen.
Example 1: Single channel measurement screen
1
Indicates that trace A of channel 1 is active and the frequency at the corresponding active marker point is 150.005
MHz.
2
Indicates that the active marker of trace A of channel 1 is active marker 0, that it is at the 250th point, and that the
value measured at this point (frequency) is –13.5915 dB.
3
Indicates that the active marker of trace B of channel 1 is active marker 0, that it is at the 250th point, and that the
value measured at this point (frequency) is –17.0609 deg.
4
The waveform display scale of trace A of channel 1 is calibrated in 10 dB and the line marked with ▲ indicates
–50.000 dB.
5
The waveform display scale of trace B of channel 1 is calibrated in 50 dB and the line marked with ▲ indicates
0.000 deg.
1
2
3
5
4
7
8
9
10
6
Measurement screen 1
6
Indicates that the start frequency is 10 kHz and the stop frequency is 300 MHz.
7
Indicates that a ratio computation (TA/R) is performed.
8
Indicates that the waveform of trace A of channel 1 indicates a logarithmic magnitude (MAG).
9
Indicates that the waveform of trace B of channel 1 indicates a phase (PHA).
10
Indicates that the resolution bandwidth is set automatically and the current setting is 20 kHz.
3-6
3.3
Data Displayed on the Screen
Example 2: Dual channel measurement screen
1
Indicates that trace A of channel 1 is active and the frequency at the corresponding active marker point is 150.005
MHz.
2
Indicates that the active marker of trace A of channel 1 is active marker 0, that it is at the 250th point, and that the
value measured at this point (frequency) is –0.0020 dB.
3
Indicates that the active marker of trace B of channel 1 is active marker 0, that it is at the 250th point, and that the
value measured at this point (frequency) is –72 ps.
4
The waveform display scale of trace A of channel 1 is calibrated in 10 dB and the line marked with ▲ indicates
–50.000 dB.
5
The waveform display scale of trace B of channel 1 is calibrated in 100 ns and the line marked with ▲ indicates
0.000 us.
6
Indicates that the start frequency of channel 1 is 10 kHz and the stop frequency is 300 MHz.
7
Indicates that a ratio computation (TA/R) is performed in channel 1.
8
Indicates that the waveform of trace A of channel 1 indicates a logarithmic magnitude (MAG).
9
Indicates that the waveform of trace B of channel 1 indicates a group delay (DLY).
10
Indicates that the resolution bandwidth of channel 1 is set automatically and the current setting is 20 kHz.
11
Indicates that calibration is performed in channel 1.
1
2 3
5
4
7
8
9
10
11
6
12
17
14
13
15
18
19
20
16
Measurement screen 2
3-7
Section 3 Basic Operations
12
Indicates that the active marker of trace A of channel 2 is active marker 0, that it is at the 250th point, and that the
value measured at this point (frequency) is –13.5819 dB.
13
Indicates that the active marker of trace B of channel 2 is active marker 0, that it is at the 250th point, and that the
value measured at this point (frequency) is –16.9570 ps.
14
The waveform display scale of trace A of channel 2 is calibrated in 10 dB and the line marked with ▲ indicates
–50.000 dB.
15
The waveform display scale of trace B of channel 2 is calibrated in 50 deg. and the line marked with ▲ indicates
0.000 us.
16
Indicates that the start frequency of channel 2 is 10 kHz and the stop frequency is 300 MHz.
17
Indicates that a ratio computation (TA/R) is performed in channel 2.
18
Indicates that the waveform of trace A of channel 2 indicates a logarithmic magnitude (MAG).
19
Indicates that the waveform of trace B of channel 2 indicates a group delay (DLY).
20
Indicates that the resolution bandwidth of channel 2 is set automatically and the current setting is 20 kHz.
3-8
3.4
3.4
Presetting
Presetting
Pressing the ( Preset ) key and selecting “Yes” from the soft key menu resets the measurement parameters of this system to
default values (see Appendix B), excluding the following parameters:
(1)
GPIB interface conditions
•
•
•
•
•
(2)
Printer/plotter parameters
•
•
•
•
(3)
Active Port
Copy Device
GPIB Address
Form Feed
Save/recall parameter
•
(4)
GPIB My Address
Control Function
Enable Register All
Terminator
Time Out
Drive
Other system parameters
•
•
•
Marker setting mode
Impedance measurement method
Screen color
Table 3-4
Preset
Yes
------
Function
Performs initialization.
NOTE:
Turning on the power with the Preset key pressed down allows you to reset the backup data stored in the internal
memory to the settings on shipping.
If backup data has been destroyed due to an incidental problem and therefore the machine does not start operating
after power-on, following the above procedure may start the machine.
3-9
Section 3 Basic Operations
3.5
Symbols
Symbol
0MKR
A
ADM
ADMT
AUTO
AVG
B
BS
CAL
CALC
CF
CH1
CH2
CNT
CONT
deg
DIR
DLY
EL
EXT
Fctn
FREQ
FREQ−TB
F1 to F6
GND
IL
IMAG
IMP
IMPD
ISLN
I/F
LEVEL−TB
LIN
LIN
LOG
LOGMAG
MAG
MAX
MEAS
MIN
MK
3-10
FULL
Zero Marker
Output A
Admitance
Admitance
Automatic Setting
Averaging for S/N Improvement
Output B
Back Space
Calibration
Calculate
Center Frequency
Channel 1
Channel 2
Center Frequency
Sweep Continue
Degree
Directry
Delay
Electric length
External
Function
Frequency
Frequency Table
Function Keys No.1 to 6
Ground
Insertion Loss
Imaginary part
Impedance
Impedance
Isolation
Interface
Level Table
Linear
Linear Magnitude
Logarithmic
Logarithmic Magnitude
Logarithmic Magnitude
Maximum
Measure
Minimum
Marker
3.5
Symbol
MKR
MP
MSG
MT
NWA
OFS
OSC
OSL
PHA
PMC
PARAM
PRMS
PTA
R
RB
RBW
RCL
REAL
REF
RESON
RESP
RPL
RSV
RTL
SEPA
SPAN
ST
STA
St d
STO
SWT
TA
TB
TRK
UNL
UNT
VSWR
ΔMKR
Symbols
FULL
Marker
Measurement Point
Message
Main Trace
Network Analyzer
Offset
Oscillator
Open-Short-Load
Phase
Plug-in Memory Card
Paramater
Paramaters
Personal Test Automation
Reference port
Resolution Bandwidth
Resolution Bandwidth
Recall
Real Part
Reference
Resonator
Response
Ripple
Request Service
Return to Local
Separated type Video Signal
Frequency Span
Sub Trace
Start Frequency
Standard
Stop Frequency
Sweep Time
Test Port A
Test Port B
Tracking
Unlisten
Untalk
Voltage Standing Wave Ratio
∆Marker
3-11
Section 3 Basic Operations
3.6
List of Soft Key Menu Functions
The following table lists the functions that can be performed by pressing soft keys (F1 to F6):
Panel key
Preset
Local
PTA
Menu
Meas
Format
Marker
Fctn
Frequency
Span
Avg
Major function
Initialization
Setting related
to GPIB
PTA function
Measurement
channel selection
Analysis port
selection
Analysis format
selection
Marker setting
Marker function
Center/start frequency setting
Span/stop frequency setting
Setting related
to S/N improvement
Sweep
Setting related
to sweep
Out/input
Setting related
to input/output
Cal
Scale
Trace
3-12
Soft key menu
YES
GPIB
←
RS232C
←
CH1
CH2
CH1&CH2
TA/R
TA
LOGMAG
POLAR
ON
TB/R
TB
PHASE
IMPD
CHART
SCROLL
TB/TA
R
DELAY
ADMT
CHART
OFF
NORMAL
ΔMKR
0MKR
MK→
MAX
MKR→
+PEAK
CENTER
MK→
MIN
MKR→
-PEAK
START
HARD
←COPY
No
MKR→
CF
TRK
+PEAK
LOG
START
STOP
LOG
STOP
SMOOTH- DELAY
ING
APERTURE
REPEAT
SINGLE
LIST
SWEEP
LIST
POWER
LIST
SWEEP
EDIT
setup
RANGETA
*****
RANGETB
*****
SPAN
AVERAGE
Setting related
to calibration
Display scale set- AUTO
ting
SCALE
SPLIT
Setting related
DISP
to display
SUB
TRACE
STOP
/CONT
LIST
SWEEP
GUARD
POWER
←SWEEP
RANGE-R
MAG&PHA MAG&DLY
VSWR
more
←
REF MKR COUPLED
No
MAKER
MKR
MKR
CHANGE
LIST
∆→SPAN
MK→
OFS
TRK
TRK OFF
-PEAK
RBW
SWEEP
TIME
*****
OUTPUT
A/B
INPUT
←IMPD
*****
SCALE
OFFSET
OFS LINE
STORAGE
OVER
WRITE
MT = ST
GRID
←
MT =
MT-ST
AVERAGE
FORMAT
←
setup
←
FULL/MKR
COUPLED
CHANNEL
etc
etc
etc
etc
etc
etc
etc
etc
etc
etc
etc
REF
IMPD
setup
←
EL
etc
CAL ON
DISPLAY
MT→ST
←ITEM
etc
etc
3.6
Panel key
Filter
Limit
Resonator
Major function
Filter analysis
function
Limit test setting
List of Soft Key Menu Functions
Soft key menu
ANALYSIS
LINE
ENTRY
←
ANALYSIS
Resonator analysis function
Hard copy setting GPIB
SETUP
←
LIMIT
TEST
SETUP
←
Control
RS232C
(Opt)
←
Save/Recall Saving/recalling INDEX RCL RECALL
setting
←
TEXT SAVE TITLE
←
←
System parameter USER
System
CLOCK
setting
←
PRESET
←
BEEP
RESON1
RESON2
SAVE
←
HARD
COPY
←
MANAGE
←
BIT MAP
←
DRIVE
←
etc
etc
OPTION
←
COLOR
←
SELF
TEST
←
3-13
Section 3 Basic Operations
3.7
The Basics of Key Operations
Three types of keys are used to operate the MS4630B Network Analyzer.
3.7.1
Hard keys (master keys)
Pressing hard keys (master keys) on the front panel performs the associated functions and displays a soft key menu.
Pressing marker keys and hard keys in the Measurement section allows you to enter numeric values such as measurement
conditions.
NOTE:
Usually, use numeric keys or the rotary knob to enter numeric values.
3.7.2
Soft keys
A soft key menu appears at the right of the LCD.
Pressing the soft key (F1 to F6) corresponding to a desired menu option performs the associated function.
NOTE:
When there are two or more menu pages, the top label of each menu page indicates a menu page number (e.g., 1/2).
Since the system has a menu learning function, pressing a hard key allows you to access the menu page that you
accessed most recently.
3.7.3
Operation windows
Keys for displaying windows.
Selecting a menu label containing “←” displays an operation window (hereafter simply called a window).
Setting items that are rarely changed and items that require you to select many parameters are displayed in these windows.
GPIB
Indicates that selecting this key displays a window.
3-14
3.7
3.7.4
(1)
The Basics of Key Operations
Operations to Perform in Windows
Selecting setting items
A reverse cursor appears in the window.
Move the reverse cursor to the setting item you want to change.
Reverse cursor
A character string or value displayed in reverse video can be changed now.
Move the reverse cursor to the setting item you want to change.
To move the reverse cursor, use the arrow keys below the rotary knob.
UP and DOWN arrow keys:
Move the reverse cursor to the preceding or next setting item.
LEFT and RIGHT arrow keys:
Moves the reverse cursor to the preceding or next option in a
setting item.
Indicates that this item is a setting item. To move the reverse cursor to another setting item use the UP and DOWN arrow
keys.
GPIB MY ADDRESS
:
[
1]
CONTROL FUNCTION:
1. DEVICE
2. CONTROLLER
ENABLE REGISTER ALL
1. OFF
2. ON
TERMINATER (for TALKER)
1. LF&EOI
2. CR/LF&EOI
TIME OUT
[
20 sec]
ACTIVE PORT for HARD COPY
1. GPIB
2. RS232C
3. CENTRONICS
To move the reverse cursor to another option in a setting item, use the LEFT and RIGHT arrow keys.
The underlined options are current choices.
(2)
Making a new setting take effect
1) Numeric value
Move the reverse cursor to a desired setting item, then enter a numeric value with numeric keys or rotary knob.
2) Option
Move the reverse cursor to a desired option, then press the Enter key or move the reverse cursor to another
option.
Only when an item is enclosed in < > can you move the cursor and make the choice take effect at the same time
by entering the item number using numeric keys.
When an item takes effect, it is underlined.
3-15
Section 3 Basic Operations
(3)
Closing a window
You can close a window by one of the following methods:
•
Press another hard key.
•
Press the
•
Move the reverse cursor to the upper right corner of the window and press the Enter key.
(If the window is in a deep layer, the window in a layer which is one layer shallower than the current will
appear.)
3.7.5
Window Close
under the soft key.
Channels and Traces
This system has two measurement channels: channel 1 (CH1) and channel 2 (CH2).
“Channel” refers to a set of internal hardware resources necessary for performing measurements. For example, settings of
a measurement frequency and an analysis port are included in the settings of a channel. Since there are two channels, two
measurements can be performed using one measuring device. *1
One channel has two traces: trace A (TR_A) and trace B (TR_B).
“Trace” refers to the displayed result of measurement performed in a channel. Two traces appear only when a measurement result is displayed with respect to two parameters such as “magnitude” and “phase,” allowing you to make various
settings for each trace.
For example, a marker setting is included in a trace setting.
As stated above, this system allows you to obtain a maximum of four traces for two channels. *2
NOTE:
*1
*2
3-16 .
When a channel contains only one hardware resource, two measurements cannot be performed at the same
time. They must be performed one at a time (e.g., frequency, then output level).
However, the limit test function may come under certain restrictions (See 6.3.1).
Section 4
Selecting Measurement Items (Channels Group)
This section explains how to operate keys for selecting measurement items. These keys are in the Channels section on the
front panel.
MS4630B
Channels
Network Analyzer
10Hz–300MHz
Channels
1
F1
Meas
Shift
Marker
2
Menu
Active
Marker
Format
Fctn
1
7
8
9
BS
4
5
6
MHz
µsec
1
2
kHz
msec
mm
2
Preset
F2
Measurement
Frequency
Span
F3
3
System
Avg
Remote
Sweep
Out/Input
0
•
+/–
sec
Enter m
Menu
Active
Meas
Format
F4
Calibration
Utility
Copy/Memory
Control
Limit
Local
F5
Cal
PTA
X→S
A
F6
Backlight Off
Display All
Filter
Resonator
Copy
Display
B
Scale
Battery
Busy
X–S
Active
Trace
Probe
Source
Window Close
Outputs
Keyboard
PMC
Save/
Recall
A (Opt B)
B
R
+12V
100mA
Inputs
TA
TB
Max +20dBm
Front panel
Major functions
• Measurement channels (CH1, CH2)
•
Selection of active channel
•
Setting of analysis port
•
Selection of analysis (display) format
4.1
4.2
4.3
4.4
Selecting a Measurement Channel (Menu) ...............................
4.1.1 Selecting a measurement channel ................................
4.1.2 Coupled Channel ...........................................................
Active Channel (Active) .............................................................
Selecting an Analysis Port (Meas) ............................................
4.3.1 Measuring a ratio (TA/R, TB/R, TB/TA) .........................
4.3.2 Measuring a level (TA, TB, R) .......................................
4.3.3 Notes on level measurement .........................................
Selecting Measurement Items (Format) ....................................
4.4.1 Selecting an analysis format ..........................................
4.4.2 Selecting an impedance marker display mode ..............
4.4.3 Selecting an admittance marker display mode ..............
4.4.4 Phase offset ...................................................................
4-2
4-2
4-3
4-4
4-4
4-4
4-5
4-5
4-6
4-6
4-7
4-8
4-8
4-1
Section 4 Selecting Measurement Items (Channels Group)
4.1
Selecting a Measurement Channel (Menu)
Menu
Set a measurement channel.
MENU
CH1
CH2
CH1&CH2
--COUPLED
CHANNEL
4.1.1
Press the
4-2
Description of function
Selects CH1 as a measurement channel.
Selects CH2 as a measurement channel.
Selects CH1 and CH2 as measurement channels (Dual channel measurement.)
Couples measurement conditions of CH1 and CH2 (when set ON.)
Selecting a measurement channel
Menu
key and select a measurement channel from the soft key menu.
4.1
4.1.2
Selecting a Measurement Channel (Menu)
Coupled Channel
Press the Menu key and select a channel coupling mode from the soft key menu “COUPLED CHANNEL.” When the
channels are coupled, setting an interlinked item for the channel in the active status will also result in setting of the same
item for the channel in the non-active status. The table below shows the setup items which can be interlinked by this
function by key group.
Channels group
Measurement group
Display group
Calibration group
Utility group
Analysis port
Analysis format
Phase offset
Frequency
Output level
Power sweep
Input range
Input impedance
Averaging
RBW
Smoothing
Delay aperture
Sweep time
Sweep control
Marker sweep
Number of measurement points
Break point
Active trace
Electrical length
Sub-trace
CAL ON/OFF
CAL element data fetching start
Normalize ON/OFF
Start of fetching of the reference data for
normalization
Filter analysis function
Resonator analysis function
The Marker, Copy & Memory and System groups have no target functions.
4-3
Section 4 Selecting Measurement Items (Channels Group)
4.2
Active Channel (Active)
Select an active channel using the “Active” key
1
2
Select an active channel whose settings (measurement conditions, etc.) are effective.
Active
Single channel (CH1 or CH2)
B
A
Pressing the
Active
key toggles between CH1 and CH2.
Dual channel (CH1 and CH2)
A
Pressing the
B
Active
key selects a channel (active channel) whose settings (measurement conditions, etc.) are effec-
tive.
NOTE:
A channel whose settings (measurement conditions, etc.) are effective is called an active channel. The lamp lit
A
above the
4.3
Press the
4.3.1
Meas
B
Active
key indicates the active channel.
Selecting an Analysis Port (Meas)
Meas
key and select an analysis port from the soft key menu.
Measuring a ratio (TA/R, TB/R, TB/TA)
Select an analysis port.
Meas
Select an analysis port.
MEAS (1/2)
TA/R
TB/R
TB/TA
--etc.
4-4
Description of function
Measures the ratio (TA/R) of the signal input to input port R to the signal input to input port TA.
Measures the ratio (TB/R) of the signal input to input port R to the signal input to input port TB.
Measures the ratio (TB/TA) of the signal input to input port TA to the signal input to input port TB.
Switches to the next menu.
4.3
4.3.2
Selecting an Analysis Port (Meas)
Measuring a level (TA, TB, R)
MEAS (2/2)
TA
TB
R
--etc.
Description of function
Measures the level of the signal input to input port TA.
Measures the level of the signal input to input port TB.
Measures the level of the signal input to input port R.
Returns to the previous menu.
NOTE:
The input port TB is an optional unit.
When this unit is not installed, setting related to input port TB cannot be made.
The setting of analysis format and display scale can be made independently for each analysis port selected.
4.3.3
Notes on level measurement
When TA, TB, or R is selected, meaningful waveform data is not displayed if an analysis format other than LOGMAG
and LINMAG is selected.
The phase of signal is not measured on the level measurement, then the function which needs the phase information
(Averaging and CAL except Normalize) is not operated.
4-5
Section 4 Selecting Measurement Items (Channels Group)
4.4
Selecting Measurement Items (Format)
Press the
4.4.1
Press the
(1)
Format
key to select measurement items from the soft key menu.
Selecting an analysis format
Format
key to select measurement items from the soft key menu and MORE window.
Selection using the soft key menu.
Format
Select an analysis (display) format.
Format (1/2)
LOG MAG
PHASE
DELAY
MAG & PHA
MAG & DLY
etc.
Format (2/2)
POLAR
IMPD CHART
ADMT CHART
VSWR
more
etc.
Description of function (unit)
Displays the analysis result with a logarithmic magnitude. (dB)
Displays the analysis result with a phase. (deg)
Displays the analysis result with a group delay. (sec)
Displays the analysis result with a logarithmic magnitude and phase. (dB), (deg)
Displays the analysis result with a logarithmic magnitude and group delay. (dB), (sec)
Switches to the next menu.
Description of function (unit)
Displays the analysis result with a polar chart (polar coordinates). (unit less number)∠(deg)
Displays the analysis result with an impedance chart.
Displays the analysis result with an admittance chart.
Displays the analysis result with VSWR. (unit less number)
Displays the window for selecting another format.
Returns to the previous menu.
NOTE :
• For the unit of measure of marker display values, see IMPD MKR FORMAT and ADMT MKR FORMAT.
The same waveform as that of the POLAR format is displayed.
• No correct measurement result can be obtained for the analysis format of group delay with the horizontal axis
being the logarithm frequency (LOG).
4-6
4.4
(2)
Selecting Measurement Items (Format)
Selection using the window.
Select other analysis format.
key
Effective key: Ten-key, rotary knob
-MORE- WINDOW
FORMAT:
01.
Displays the linear magnitude (unit less number).
02.
Displays both the linear magnitude and phase (unit less number), (deg).
03.
Displays both the linear magnitude and group delay (unit less number), (sec).
04.
Displays the real number component (unit less number).
05.
Displays the imaginary number component (unit less number).
06.
Displays the real and imaginary number components (unit less number), (unit
less number).
07.
Displays the impedance in logarithm (Ω).
08.
Displays the impedance and impedance phase (Ω), (deg).
09.
Displays Q (unit less number).
10.
Displays the impedance in logarithm and Q (Ω), (unit less number).
IMPD MKR FORMAT for IMPD CHART:
ADMT MKR FORMAT for ADMT HART:
PHASE OFFSET:
Note:
Enter two-digit numeric to make a selection using the ten-key.
Example) Enter “0” and “7” to select LOG Z.
Pressing the Enter key is not required when using the ten-key for data entry.
4.4.2
Selecting an impedance marker display mode
When the analysis format is “IMPD CHART”, set the display mode of the marker value.
Effective key: Rotary knob
key
MORE- WINDOW
FORMAT~
IMPD MKR FORMAT for IMPD CHART:
Z∠θ:
The absolute value and phase angle of impedance
Rs/Ls, Cs:
Equivalent series resistance and equivalent series inductance, or equivalent
series capacity
Q/D:
Q and tan δ of the resonance circuit
ADMT MKR FORMAT for ADMT CHART:~
PHASE OFFSET:~
4-7
Section 4 Selecting Measurement Items (Channels Group)
4.4.3
Selecting an admittance marker display mode
When the analysis format is “ADMT CHART”, set the display mode of the marker value.
Effective key: Rotary knob
key
-MORE- WINDOW
FORMAT~
IMPD MKR FORMAT for IMPD CHART:~
ADMT MKR FORMAT for ADMT CHART:
Y∠θ:
The absolute value and phase angle of admittance
Rp/Lp, Cp:
Equivalent parallel resistance and equivalent parallel inductance, or
equivalent parallel capacity
Q/D:
Q and tan δ of the resonance circuit
PHASE OFFSET:~
4.4.4
Press the
Phase offset
Format
key and set an offset value in the MORE window.
Effective key: Ten-key, rotary knob
-MORE- WINDOW
FORMAT~
IMPD MKR FORMAT for IMPD CHART:~
ADMT MKR FORMAT for ADMT CHART:~
PHASE OFFSET:
When the format is “PHASE” or the impedance phase is “θ,” set a phase
offset.
Setting range:
±180 deg
4-8 .
Section 5 Selecting Measurement Parameters
(Measurement Group)
This section explains how to select measurement parameters.
MS4630B
Network Analyzer
10Hz–300MHz
Channels
1
Menu
F1
Meas
Measurement
Shift
Marker
2
Active
Format
Marker
Fctn
7
8
9
BS
4
5
6
MHz
µsec
1
2
3
kHz
msec
mm
Preset
F2
Measurement
Frequency
Span
F3
Frequency
Span
System
Avg
Remote
Sweep
Out/Input
0
•
+/–
sec
Enter m
F4
Calibration
Utility
Copy/Memory
Control
Limit
Local
F5
Cal
PTA
Scale
Backlight Off
Battery
Busy
Display All
X→S
A
F6
X–S
Filter
Resonator
Active
Trace
Probe
Source
Window Close
Outputs
PMC
Save/
Recall
Copy
Display
B
Keyboard
A (Opt B)
B
R
+12V
100mA
Inputs
TA
TB
Avg
Sweep
Out/Input
Max +20dBm
Front panel
Major functions
• Frequency setting
• Averaging, smoothing, resolution bandwidth, and delay aperture setting
• Setting of sweep control, sweep time, number of sweep points, list sweep
• Setting of input range and impedance
• Setting of output power and power sweep
5.1
5.2
5.3
5.4
Setting a Frequency (Frequency, Span) ...................................
Settings Related to Input/Output (Out/Input) .............................
5.2.1 Output level....................................................................
5.2.2 Switching between output ports A and B .......................
5.2.3 Power sweep .................................................................
5.2.4 Input range ....................................................................
5.2.5 Input impedance ............................................................
5.2.6 Reference impedance ...................................................
Improving the S/N Ratio and Setting an Delay Aperture
(Group Delay Measurement) (Avg) ...........................................
5.3.1 Averaging ......................................................................
5.3.2 RBW ..............................................................................
5.3.3 Smoothing......................................................................
5.3.4 Delay aperture ...............................................................
Settings Related to Sweep ........................................................
5.4.1 Sweep time ....................................................................
5.4.2 Sweep control ................................................................
5.4.3 Number of measurement points ....................................
5.4.4 Breakpoints ....................................................................
5.4.5 Sweep by external trigger ..............................................
5.4.6 List sweep ......................................................................
5.4.7 Marker sweep ................................................................
5-2
5-3
5-3
5-4
5-4
5-5
5-5
5-6
5-7
5-7
5-8
5-8
5-9
5-10
5-10
5-10
5-11
5-11
5-12
5-13
5-19
5-1
Section 5 Selecting Measurement Parameters (Measurement Group)
5.1
Setting a Frequency (Frequency, Span)
Set a measurement frequency using the
Effective key: Ten-key, rotary knob
Frequency
Frequency
and
Span
keys.
Set a center or start frequency.
FREQ
CENTER
START
LOG START
Description of function
Selects the center span setting mode, allowing you to set a center frequency.
(0 to 300 MHz)
Selects the start-stop setting mode, allowing you to set a start frequency.
(0 to 300 MHz)
Selects the log frequency setting mode, allowing you to set a log start frequency. *1
----
Span
Set a frequency span or stop frequency.
SPAN
SPAN
STOP
LOG STOP
Description of function
Selects the center span setting mode, allowing you to set a frequency span.
(0 to 300 MHz)
Selects the start-stop setting mode, allowing you to set a stop frequency.
(0 to 300 MHz)
Selects the log frequency setting mode, allowing you to set a log stop frequency. *1
----
NOTE:
*1
5-2
In the log frequency setting mode, frequencies that can be set are 10 Hz, 100 Hz, 1 kHz, 10 kHz, 100 kHz,
1 MHz, 10 MHz, 100 MHz, and 300 MHz.
A frequency range can be set independently for each setting mode.
5.2
5.2
Settings Related to Input/Output (Out/Input)
Press the
Out/Input
key and make input-/output-related settings using the soft key menu and window.
Out/Input
Set an output power level and input range.
5.2.1
(1)
Settings Related to Input/Output (Out/Input)
Output level
Setting of the output level
Use the soft key menu shown below to set the output level.
Effective key: Ten-key, rotary knob
OUTPUT (1/2)
POWER
setup
POWER
SWEEP
OUTPUT A/B
-etc.
Function Detail
Enables setting of the output level. *1
Switches the display to the next menu.
NOTE:
*1 The output level depends on the output level related setting shown below. See NOTE 1 of 5.2.1 (2).
(2)
Output level related setting
Use the soft key menu or the window shown below to execute the output level related setting.
Effective key: Ten-key, rotary knob
OUTPUT (1/2)
POWER
setup
POWER
SWEEP
OUTPUT A/B
-etc.
Function Detail
Displays the window to exevute other output level related setting. *1
Switches the display to the next menu.
-OUTPUT PWR SETUP- WINDOW
SOURCE POWER:
Allows you to set the internal signal source level. (0 to +21 dBm)
OUTPUT ATT:
Allows you to set the attenuation of the output attenuator (optional unit).
This is not displayed when no optional unit is installed.
OFFSET:
Allows you to set the apparent offset of the output power.
The actual output power does not change.
5-3
Section 5 Selecting Measurement Parameters (Measurement Group)
NOTE:
*1 •
•
5.2.2
Press the
The relationship among the above values can be expressed as follows (when output port A is selected):
Output power = (Source power) – (Output attenuation) + (Offset)
When output port B (2-branch output) is selected, 6 dB is subtracted from the output power. When output
port B (3-branch output : option 14) is selected, 9.5 dB is subtracted from the output power.
When the output attenuator (optional unit) is not installed, the output attenuation is assumed to be 0 dB.
Switching between output ports A and B
Out/Input
key and select OUTPUT A/B from the following soft key menu to switch between output ports A and B:
OUTPUT (1/2)
POWER
setup
POWER
SWEEP
OUTPUT A/B
-etc.
5.2.3
Description of function
Allows you to switch between output ports A and B (branch outputs) on the front panel.
Switches to the next menu.
Power sweep
Press the Out/Input key and select POWER SWEEP from the following window:
Effective key: Ten-key, rotary knob
OUTPUT (1/2)
POWER
setup
POWER
SWEEP
OUTPUT A/B
-etc.
Description of function
Displays the window for setting power sweep.
Switches to the next menu.
-POWER SWEEP- WINDOW
POWER SWEEP:
Turns ON/OFF power sweep.
START POWER:
Allows you to set the output power at the sweep start frequency.
END POWER:
Allows you to set the output power at the sweep stop frequency.
STEP:
Allows you to set an output power change step.
The start power changes to the stop power in specified steps.
5-4
5.2
Settings Related to Input/Output (Out/Input)
STOP POWER
STEP
START POWER
5.2.4
Input range
Press the Out/Input key and make the settings related to the input range using the following soft key menu and window.
Effective key: Ten-key, rotary knob
INPUT (2/2)
RANGE-TA
RANGE-TB
RANGE-R
INPUT IMPD
REF IMPD
etc.
Description of function
Allows you to set the input range of input port TA. (0/+20 dBm) *1
Allows you to set the input range of input port TB. (0/+20 dBm) *1
Allows you to set the input range of input port R. (0/+20 dBm) *1
Returns to the previous menu.
NOTE:
Input port TB is an optional unit. When it is not installed, settings related to input port TB cannot be made.
*1 Input impedance is 1 MΩ, use 0 dBm range.
5.2.5
Input impedance
Press the Out/Input key and set an input impedance using the following soft key menu and window:
Effective key: Ten-key, rotary knob
key
INPUT (2/2)
RANGE-TA
RANGE-TB
RANGE-R
INPUT IMPD
REF IMPD
etc.
Description of function
Displays the window for setting an input impedance.
Returns to the previous menu.
-INPUT IMPEDANCE- WINDOW
IMPEDANCE(TA)
IMPEDANCE(TB)
IMPEDANCE(R)
Set the input impedance of each input port to 50/75 Ω or 1 MΩ.
The 75 Ω input requires a conversion adapter (MA4605A).
5-5
Section 5 Selecting Measurement Parameters (Measurement Group)
5.2.6
Reference impedance
Press the Out/Input key and select REF IMPD from the following soft key menu to set the reference impedance:
Effective key: Ten-key, rotary knob
INPUT (2/2)
RANGE-TA
RANGE-TB
RANGE-R
INPUT IMPD
REF IMPD
etc.
5-6
Description of function
Allows you to set a reference impedance. (0.1 to 10000.0 Ω)
Returns to the previous menu.
5.3 Improving the S/N Ratio and Setting an Delay Aperture (Group Delay Measurement)(Avg)
5.3
Press
Improving the S/N Ratio and Setting an Delay
Aperture (Group Delay Measurement) (Avg)
Avg
key and make settings using the following soft key menu and window:
Avg
Set averaging, smoothing, delay, aperture, and resolution bandwidth.
5.3.1
Press the
Averaging
Avg
key and make averaging-related settings using the following soft key menu and window:
(1)
Set the frequency of sweep required to exevute the averaging process.
Effective key: Ten-key, rotary knob
AVG
AVERAGE No
SMOOTHING
DELAY
APERTURE
RBW
AVERAGE
FORMAT
--
(2)
Description of function
Allows you to set a number of sweeps to be performed for averagingn (1 to 1000.)
Displays the window for setting an averaging type.
Set the type of the averaging process.
Effective key: Ten-key, rotary knob
key
-AVG FORMAT- WINDOW
AVERAGE TYPE
SUM:
Averages measured points for each sweep. *1
MAX:
Holds the maximum measured point. (MAX HOLD) *2
MIN:
Holds the minimum measured point. (MIN HOLD) *2
NOTE:
*1
*2
Real parts and imaginary parts of measured values as vector quantity are averaged as many times as the
number of measurement to get real part data and imaginary part data.
When the analysis port is TA, TB, or R (level measurement), the signal phase cannot be measured properly,
resulting in an indefinite average and abnormal operation.
The measured value whose absolute value is maximum (minimum) is used as a vector quantity.
Normal operation is assured if the analysis port is TA, TB, or R (level measurement).
5-7
Section 5 Selecting Measurement Parameters (Measurement Group)
5.3.2
RBW
Set the resolution bandwidth.
Effective key: Ten-key, rotary knob
AVG
AVERAGE No
SMOOTHING
DELAY
APERTURE
RBW
AVERAGE FORMAT
--
Description of function
Allows you to set a resolution bandwidth. *1
NOTE:
*1 One of the following values can be set for “RBW”:
3 Hz, 10 Hz, 30 Hz, 100 Hz, 300Hz, 500 Hz, 1 kHz, 2 kHz, 3 kHz, 4 kHz, 5 kHz, 10 kHz, 20 kHz, and AUTO
In the RBW auto-setting mode, the measurable minimum resolution bandwidth is automatically selected according to the current sweep time.
To automatically set the RBW, type [0] and [Enter].
5.3.3
Smoothing
Execute the moving average process required to smooth the waveform.
Effective key: Ten-key, rotary knob
AVG
AVERAGE No
SMOOTHING
DELAY
APERTURE
RBW
AVERAGE FORMAT
--
Description of function
Performs moving averaging for smoothing waveforms (0 to 50%.)
NOTE:
The measurement result around the start and stop frequencies, for which the moving averaging process cannot be
performed, is displayed as invalid data.
The invalid data is usually displayed at the bottom of the screen.
However, if the analysis format is POLAR, IMPD CHART, or ADMT CHART, the invalid data is displayed at the
center of a circle.
5-8
5.3 Improving the S/N Ratio and Setting an Delay Aperture (Group Delay Measurement)(Avg)
5.3.4
Delay aperture
Set the aperture for the group delay measurement.
Effective key: Ten-key, rotary knob
AVG
AVERAGE No
SMOOTHING
DELAY
APERTURE
RBW
AVERAGE
FORMAT
--
Description of function
Allows you to set an aperture for delay (group) delay measurement. *1
NOTE:
*1 Specify an aperture with a ratio (%) to the frequency span. The setting range is from 0.2 % to 20 %. The
minimum value and resolution depend on the following expression:
Minimum value and resolution = 2/(Number of measure points) *100 (%)
NOTE:
The group delay measurement is calculated using the measurement data obtained at the two measurement points
specified in the delay aperture parameter.
Therefore, the measurement result is displayed only for frequencies at which these two measurement points can
co-exist. Since this condition is not met around the start and stop frequencies, the result is displayed as invalid
data.
This invalid data becomes the minimum value of the measurement data.
5-9
Section 5 Selecting Measurement Parameters (Measurement Group)
5.4
Press the
Settings Related to Sweep
Sweep
key and make settings related to sweep using the displayed soft key menus and windows.
Sweep
5.4.1
Set sweep control, sweep time, and the number of sweep points.
Sweep time
Press the Sweep key and select SWEEP TIME from the soft key menu to set a sweep time.
Effective key: Ten-key, rotary knob
SWEEP (1/2)
REPEAT
SINGLE
STOP/CONT
SWEEP TIME
setup
etc
•
Description of function
Allows you to set a sweep time (0 to 27.5 hours.)
To specify “AUTO” for “SWEEP TIME,” enter 0 (sec).
When AUTO is specified for “SWEEP TIME,” the fastest sweep time is automatically set according to the current
RBW, number of sweep points, and number of traces.
5.4.2
Press the
Sweep control
Sweep
key and make settings related to sweep control using the displayed soft key menu and window.
SWEEP (1/2)
REPEAT
SINGLE
STOP/CONT
SWEEP TIME
setup
etc
5-10
Description of function
Sweeps repetitively.
Sweeps only once.
Stops (STOP) or restarts (CONT) sweeping.
5.4
5.4.3
Settings Related to Sweep
Number of measurement points
Press the Sweep key and specify a number of measurement points using the displayed soft key menu and window.
Effective key: Ten-key, rotary knob
SWEEP (1/2)
REPEAT
SINGLE
STOP/CONT
SWEEP TIME
setup
etc
Description of function
Displays the window for setting sweep control parameters.
-SETUP- WINDOW
MEASURE POINTS:
Set the number of measure points. Specify one of 11, 21, 51, 101, 251, 501,
and 1001.
BREAK POINTS:
~
TRIGGER SOURCE:
~
EXT TRIGGER:
~
EXT TRIGGER SLOPE:
~
5.4.4
Breakpoints
Press the Sweep key and make sweep-related settings using the displayed soft key menu and window.
Effective key: Ten-key, rotary knob
SWEEP (1/2)
REPEAT
SINGLE
STOP/CONT
SWEEP TIME
setup
etc
Description of function
Displays the window for setting sweep control parameters.
-SETUP- WINDOW
MEASURE POINTS:
~
BREAK POINTS:
Set the number of points at which sweep is interrupted. (1 to 1001)
TRIGGER SOURCE:
~
EXT TRIGGER:
~
EXT TRIGGER SLOPE:
~
5-11
Section 5 Selecting Measurement Parameters (Measurement Group)
5.4.5
Press the
Sweep by external trigger
Sweep
key and make sweep-related settings using the displayed soft key menu and window.
Effective key: Ten-key, rotary knob
SWEEP (1/2)
REPEAT
SINGLE
STOP/CONT
SWEEP TIME
setup
etc
key
Description of function
Displays the window for setting sweep control parameters.
-SETUP- WINDOW
MEASURE POINTS:
~
BREAK POINTS:
~
TRIGGER SOURCE:
Specify an event (trigger) that causes sweep to start.
When an internal trigger is specified, the next sweep is started by an event
that occurs in the system. When an external trigger is specified, the next
sweep is started by the control signal input to the EXT TRIG IN terminal.
EXT TRIGGER:
Specify a type of sweep started by an external trigger.
NORMAL:
Sweeps only once when an event is caused by the control signal.
STEP:
Measures one point when an event is caused by the control signal.
STATE:
Stops or restarts sweep according to the state of the control signal.
Low level input: Stop High level input: Restart
EXT TRIGGER SLOPE:
Specify an edge (leading or trailing) of the control signal that causes an
event.
5-12
5.4
5.4.6
Settings Related to Sweep
List sweep
What is list sweep?
The device stores in itself the measurement conditions corresponding to the measurement points on the frequency axis
(horizontal axis) in the measurement screen, as data tables. In the actual sweep measurement, the device refers to the data
in these tables to set the conditions.
There are the following four types of data tables that store such measurement conditions:
•
•
•
•
Frequency data table
Level data table (used at the time of power sweep)
RBW data table
User wait data table
NOTE:
These data tables exist independently for each measurement channel.
If both measurement channels are to be used in the same measurement conditions, i.e., if the coupled channel
function is turned ON, only the data tables for the active channel are used for the measurement.
If a measurement parameter is changed, a value calculated inside the device according to the value of the said measurement parameter is reflected in these data tables.
If, for example, a frequency parameter such as the center frequency or the frequency span is set, a calculation is carried
out based on these frequency parameters to equalize the frequency intervals between the measurement points in the case
of linear sweep, the result of which is reflected at each point in the frequency data table.
Thus, changing the measurement parameters normally creates these data tables. However, regardless of the values of
these measurement parameters, you can change the measurement condition at any given point in a data table to carry out
the measurement.
List sweep refers to sweeping according to a condition that a user has arbitrarily set for each measurement point by
changing the said condition data in a data table.
The user does not need to set whether list sweep is done or not.
Modifying, changing, and registering data in the data table edit screen (LIST SWEEP EDIT menu) will cause the sweep
to be performed using the new measurement conditions in the next measurement.
These data tables can be saved to or recalled as files from auxiliary storage media such as floppy disks, internal memory,
and memory cards.
5-13
Section 5 Selecting Measurement Parameters (Measurement Group)
The following explains the details of measurement condition data tables.
(1)
Frequency data table
A frequency data table stores the data of measurement frequencies corresponding to each measurement point (Xaxis).
Normally, when a measurement parameter related to frequencies such as the center frequency or the frequency
span is set, the data for all the measurement points at that time is automatically created inside the device.
(2)
Level data table
A level data table stores the data of output power corresponding to each measurement point (Y-axis).
The data in this table is used to carry out power sweep.
Normally, when a measurement parameter related to the power sweep function is set, the data for all the measurement points at that time is automatically created inside the device.
(3)
RBW data table
A RBW data table stores the RBW values corresponding to each measurement point.
Normally, when a measurement parameter such as RBW and the sweep time is set, the data for all the measurement points at that time is automatically created based on the same RBW value.
(4)
User wait table
A user wait table stores the user wait time corresponding to each measurement point.
Normally, when a measurement parameter such as the sweep time and RBW is set, the wait time elapsing between
switching the measurement frequency at each measurement point and starting the measurement is set as the same
value for between all the measurement points.
The user wait function allows you to set any additional wait time for any given point in addition to the preset wait
time.
All the initial values of the user wait time are set to 0 seconds.
NOTE:
Changing a measurement parameter automatically updates the contents of these data tables that generally store the
measurement conditions. Therefore, changing a data table for the purpose of list sweep and then setting a related
measurement parameter again will update the data tables.
To prevent these tables from being updated when a measurement parameter is set, use the F3: LIST SWEEP
GUARD function in the SWEEP (2/2) menu.
5-14
5.4
SWEEP (2/2)
LIST SWEEP
LIST
LIST SWEEP
EDIT
LIST SWEEP
GUARD
-FULL/MKR
etc.
Settings Related to Sweep
Description of function
Displays the contents of tables of measurement conditions for list sweep in a list.
Returns to the previous menu.
-LIST SWEEP LIST- WINDOW
<< LIST SWEEP TABLE (CH*) >>
Point
Frequency
RBW
User Wait
POWER
0
10.000 000kHz
20kHz
0us
**********
1
20.000 000kHz
20kHz
0us
**********
2
30.000 000kHz
20kHz
0us
**********
•
•
•
•
•
•
•
•
•
•
•
•
•
•
•
Description of display contents
The measurement condition tables for list sweep are displayed in a list for each measurement point number.
Point:
Indicates the measurement point number on the frequency axis in the measurement screen. The leftmost point of the
screen represents Point 0. The rightmost point of the screen represents (the number of measurement points - 1).
To check the number of measurement points, press the F5: Setup key in the SWEEP (1/2) menu to open a window.
The initial value of the number of measurement points is 501.
Frequency:
Indicates the measurement frequency corresponding to each measurement point.
RBW:
Indicates the RBW (resolution bandwidth) corresponding to each measurement point.
User Wait:
Indicates the user wait time corresponding to each measurement point.
5-15
Section 5 Selecting Measurement Parameters (Measurement Group)
Power:
Indicates the output power corresponding to each measurement point.
This item is used if the power sweep function is turned ON. Otherwise, this item indicates ******** and cannot be
changed.
Description of operation
The list screen, when brought up, shows the data entry area, “LIST CURRENT POINT:” (in which a numeric value can
be entered) at the bottom.
This data entry area shows the current point (the measurement point number at which the measurement conditions can be
changed).
The current point also matches the measurement point number enclosed inside a square
in the list screen.
Use the data knob, Up/Down keys, and ten-key pad to change the current point. Then, the indication in the list screen
changes according to this change.
The current point may be set in the range from 0 through 1000.
5-16
5.4
SWEEP (2/2)
LIST SWEEP
LIST
LIST SWEEP
EDIT
LIST SWEEP
GUARD
-FULL/MKR
etc.
Settings Related to Sweep
Description of function
Brings up a window to change the contents of tables of measurement conditions
for list sweep.
Returns to the previous menu.
-LIST SWEEP EDIT- WINDOW
POINT:
Sets the current point.
CURRENT FREQUENCY:
Registers the frequency at the current point.
STEP:
Adds the frequency set in this item (the increment) to the frequency at the
current point, and registers it as the frequency at the next point.
The frequencies up to the one shown in the “END” item below can be repetitiously registered in the same way. (NOTE 1)
END:
Specifies the last frequency in the repetitious registration described above.
RBW:
Registers the RBW at the current point.
The RBW is not automatically specified.
USER WAIT:
Registers the user wait time at the current point.
0.01msec to 7200sec
POWER:
Registers the output power at the current point.
This registration is enabled only when the power sweep is turned ON.
ENTRY:
Updates the data tables according to the registered conditions.
NOTE1:
If the “STEP” frequency is set to 0 Hz, only the frequency at the current point is updated.
NOTE:
Unless the update is performed after the conditions are registered, the registered contents are not reflected in the
data tables.
5-17
Section 5 Selecting Measurement Parameters (Measurement Group)
The following describes the various ways of performing the update.
1. REPLACE
Replaces the condition at the current point with a new ones.
2. INSERT
Inserts a new condition at the current point. The old data at the current point as well as the data at any following points
is accordingly moved down.
The data at the final point will be discarded.
3. DELETE
Deletes the data at the current point. The data at any following points is accordingly moved up.
The data at the new final point is set to the same value as the data at the old final point.
4. READY
Indicates that the device is ready to perform the edit.
SWEEP (2/2)
LIST SWEEP
LIST
LIST SWEEP
EDIT
LIST SWEEP
GUARD
-FULL/MKR
etc.
Description of function
Prevents the measurement condition tables for list sweep from being updated
even if the measurement parameters are changed.
Returns to the previous menu.
LIST SWEEP GUARD function
ON: Does not update the measurement condition tables even if the measurement parameters are changed.
The screen shows “GUARD ON”.
OFF: Updates the measurement condition tables accordingly if the measurement parameters are changed (Initial state).
5-18
5.4
5.4.7
Settings Related to Sweep
Marker sweep
Press the SWEEP switch to display the software key menu to switch between marker sweep and full sweep.
SWEEP (2/2)
LIST SWEEP
LIST
LIST SWEEP
EDIT
LIST SWEEP
GUARD
-FULL/MKR
etc.
Description of function
Switches between full-band sweep (FULL) and marker sweep (MKR).
Returns to the previous menu.
The marker sweep sweeps the frequency band between the active and reference markers.
5-19
Section 5 Selecting Measurement Parameters (Measurement Group)
5-20 .
Section 6 Display (Display Group)
This section explains various functions for displaying easy-to-read measurement waveforms.
MS4630B
Network Analyzer
10Hz–300MHz
Channels
1
Menu
F1
Meas
Shift
Marker
2
Active
Format
Marker
Fctn
7
8
9
BS
4
5
6
MHz
µsec
1
2
3
kHz
msec
mm
•
+/–
Preset
F2
Measurement
Frequency
Span
F3
A
System
Avg
Remote
Sweep
Out/Input
0
sec
Enter m
F4
Calibration
Utility
Copy/Memory
Control
Limit
Local
F5
Cal
F6
Scale
Backlight Off
Battery
Busy
Display All
X→S
X–S
Filter
Resonator
Active
Scale
Trace
Probe
Source
Window Close
Outputs
Keyboard
PMC
Save/
Recall
Copy
Display
B
A
PTA
A (Opt B)
B
Display
B
R
Active
Trace
+12V
100mA
Inputs
TA
TB
Max +20dBm
Front panel
Major functions
• Setting a trace scale
•
Setting trace and scale grid display types
•
Selecting display items
•
Subtrace function
6.1
6.2
6.3
6.4
6.5
Active Trace (Active) .................................................................
Setting a Scale (Scale) ..............................................................
6.2.1 Scale ..............................................................................
6.2.2 Offset .............................................................................
6.2.3 Offset line.......................................................................
6.2.4 Auto scale ......................................................................
6.2.5 Setting an electric length ...............................................
Setting the Display Screen (Trace) ...........................................
6.3.1 Split display....................................................................
6.3.2 Waveform storage .........................................................
6.3.3 Waveform overwrite .......................................................
6.3.4 Selecting a display grid ..................................................
6.3.5 Selecting the items to display or clear ...........................
6.3.6 Subtrace ........................................................................
Displaying All Items at the Touch of a Key (Display All) ............
Turning the Backlight ON/OFF (Backlight ON/OFF) .................
6-2
6-3
6-3
6-3
6-4
6-4
6-4
6-5
6-5
6-5
6-6
6-6
6-7
6-8
6-8
6-9
6-1
Section 6 Display (Display Group)
6.1
Active Trace (Active)
A
Press the
A
B
key to select an active trace.
Active
B
Select a trace whose settings are effective (active trace).
Active
You can use this key only when two traces are displayed for one measurement channel.
A
Pressing the
6-2
B
Active
key toggles between A (trace A) and B (trace B).
6.2 Setting a Scale (Scale)
6.2
Setting a Scale (Scale)
Press the
Scale
key to make scale-related settings.
Scale
6.2.1
Set the scale of the displayed trace.
Scale
Press the Scale key and select AUTO SCALE or SCALE from the soft key menu to set the screen scale.
Effective key: Ten-key, rotary knob
SCALE
AUTO SCALE
SCALE
OFFSET
OFS LINE
EL
--
NOTE:
•
•
6.2.2
Description of function
Automatically sets a Y-axis scale best suited to the measurement data.
Allows you to set a Y-axis scale (display magnification) manually.
When two traces are displayed for one measurement channel, a scale and an offset can be set for each trace.
The settings made for the active trace are made effective.
If the analysis format is LOGZ, the width of Y-axis is set in decade units (1, 2, 4, 5, 8, or 10 decades).
Offset
Press the Scale key and select OFFSET from the soft key menu to set a screen display offset.
Effective key: Ten-key, rotary knob
key
SCALE
AUTO SCALE
SCALE
OFFSET
OFS LINE
EL
--
NOTE:
•
•
•
•
Description of function
Allows you to set a Y-axis offset (display bias) manually.
When two traces are displayed for one measurement channel, a scale and an offset can be set for each trace.
The settings made for the active trace are made effective.
If the analysis format is LOGZ, a scale can be set from 1 mΩ to 100 M Ω in one to ten steps.
An offset can be set by
or
key (except LOGZ). It can be changed one tenth step of Y-axis width by each
press of the key.
If the analysis format is POLAR, IMPD CHART, or ADMT CHART, an offset cannot be set.
6-3
Section 6 Display (Display Group)
6.2.3
Offset line
Press the Scale key and select OFS LINE from the soft key menu to set a reference line for screen display offset.
Effective key: Ten-key, rotary knob
SCALE
AUTO SCALE
SCALE
OFFSET
OFS LINE
EL
--
Description of function
Allows you to set a reference line for offset.
NOTE:
A reference line is fixed at the bottom line if analysis format is LOGZ.
6.2.4
Press the
Auto scale
Scale
key and select AUTO SCALE from the soft key menu to set a scale and offset automatically.
SCALE
AUTO SCALE
SCALE
OFFSET
OFS LINE
EL
--
Description of function
Allows you to set a Y-axis scale best suited to the measurement data.
NOTE:
The auto scale is disabled if the analysis format is LOGZ.
6.2.5
Setting an electric length
Press the Scale key and select EL from the soft key menu to set an electric length.
Effective key: Ten-key, rotary knob
SCALE
AUTO SCALE
SCALE
OFFSET
OFS LINE
EL
--
6-4
Description of function
Allows you to set a calibration value used to calibrate electric lengths.
6.3
6.3
Press the
Setting the Display Screen (Trace)
Trace
Trace
key to make the setting related to the display screen.
Set a trace display type and scale grid display type, select display items, and make subtrace-related settings.
6.3.1
Press the
Split display
Trace
key and select SPLIT DISP from the soft key menu to display split views.
TRACE (1/2)
SPLIT DISP
STORAGE
OVER WRITE
GRID
DISPLAY
ITEM
etc.
NOTE:
*1 •
•
•
•
6.3.2
Press the
Setting the Display Screen (Trace)
Description of function
Displays a screen split into upper and lower views. *1
Switches to the next menu.
When one trace is displayed for one channel, the display screen cannot be split.
When the screen display is split is the one channel, two traces mode, trace A is displayed in the upper half
and trace B is displayed in the lower half.
When traces are displayed for two channels, traces of channel 1 are shown in the upper view and traces of
channel 2 are shown in the lower view.
If the screen display is not split when at least three waveform are displayed, this will disable the use of the
limit test function.
Waveform storage
Trace
key and select STORAGE from the soft key menu to turn ON/OFF the waveform storage function.
TRACE (1/2)
SPLIT DISP
STORAGE
OVER WRITE
GRID
DISPLAY
ITEM
etc.
Description of function
Allows you to turn ON/OFF the waveform storage function.
Switches to the next menu.
6-5
Section 6 Display (Display Group)
6.3.3
Press the
Waveform overwrite
Trace
key and select OVER WRITE from the soft key menu to turn ON/OFF the waveform overwrite function.
TRACE(1/2)
SPLIT DISP
STORAGE
OVER WRITE
GRID
DISPLAY
ITEM
etc.
6.3.4
Description of function
Allows you to turn ON/OFF the waveform overwrite function.
Switches to the next menu.
Selecting a display grid
Press the Trace key and select GRID from the soft key menu to select a grid type.
Effective key: Ten-key, rotary knob
key
TRACE (1/2)
SPLIT DISP
STORAGE
OVER WRITE
GRID
DISPLAY
ITEM
etc.
Description of function
Displays the window for selecting a grid type.
Switches to the next menu.
-GRID- WINDOW
GRID TYPE:
Select a scale grid display type.
ALL:
Displays the frame and all vertical/horizontal lines.
CENTER:
Displays the frame and only vertical/horizontal center lines.
FRAME:
Displays only the frame.
6-6
6.3
6.3.5
Setting the Display Screen (Trace)
Selecting the items to display or clear
Press the Trace key and select the items to display or clear using the following soft key menu and window.
Select an item to be deleted on the window. The selected item constitutes the deletion target.
TRACE (1/2)
SPLIT DISP
STORAGE
OVER WRITE
GRID
DISPLAY
ITEM
etc.
Description of function
Displays the window to select the displayed item.
Switches to the next menu.
Effective key: Ten-key, rotary knob
key
-DISPLAY ITEM- WINDOW
SELECT ITEM:
01.
A line to display the marker value, scale value and other data of Trace A.
02.
A line to display the marker value, scale value and other data of Trace B.
03.
Selection to display the measuring conditions such as the analysis for mat and
RBW (common to Trace A and B).
04.
A line (lines) to display the range of the frequencies measured.
05.
Source Exif Data:
File Type : PDF
File Type Extension : pdf
MIME Type : application/pdf
PDF Version : 1.4
Linearized : Yes
Encryption : Standard V1.2 (40-bit)
User Access : Print, Copy, Fill forms, Extract, Assemble, Print high-res
Create Date : 2003:02:10 10:27:13Z
Modify Date : 2014:09:05 09:13:58+08:00
Has XFA : No
XMP Toolkit : Adobe XMP Core 5.2-c001 63.139439, 2010/09/27-13:37:26
Producer : Acrobat Distiller 4.0 for Macintosh
Metadata Date : 2014:09:05 09:13:58+08:00
Document ID : uuid:468469a4-e06d-48f8-b658-2bbf280cce0a
Instance ID : uuid:b5255d50-e70d-44aa-a84f-30f5d0c3a0f8
Format : application/pdf
Page Mode : UseOutlines
Page Count : 222
EXIF Metadata provided by EXIF.tools