Bird 4421 Thruline RF Power Meter BIRD_4421

User Manual: BIRD_4421

Open the PDF directly: View PDF PDF.
Page Count: 59

DownloadBird 4421 Thruline RF Power Meter BIRD_4421
Open PDF In BrowserView PDF
INSTRUCTION BOOK

THRULINE® RF POWER METER MODEL 4421
AND
THRULINE® DIRECTIONAL
RF POWER SENSORS
4020 SERIES, 4027A SERIES,
4027F SERIES, AND 4028 SERIES

Bird® Electronic Corporation
30303 Aurora Road
Cleveland (Solon), Ohio 44139
Sales & Technical Support:

440-248-1200
866-695-4569 toll free

Sales email:

sales@bird-technologies.com

Technical Support email:

techsupport@bird-technologies.com

©Copyright 2011 by Bird Electronic Corporation
Instruction Book Part Number 920-4421 Rev U
Thruline® is a Registered Trademark
of Bird Electronic Corporation

Safety Precautions

The following are general safety precautions that are not necessarily related to
any specific part or procedure, and do not necessarily appear elsewhere in this
publication. These precautions must be thoroughly understood and apply to all
phases of operation and maintenance.
WARNING
Keep Away From Live Circuits
Operating Personnel must at all times observe general safety precautions.
Do not replace components or make adjustments to the inside of the test
equipment with the high voltage supply turned on. To avoid casualties,
always remove power.
WARNING
Shock Hazard
Do not attempt to remove the RF transmission line while RF power is present.
WARNING
Do Not Service Or Adjust Alone
Under no circumstances should any person reach into an enclosure for the
purpose of service or adjustment of equipment except in the presence of
someone who is capable of rendering aid.
WARNING
Safety Earth Ground
An uniterruptible earth safety ground must be supplied from the main
power source to test instruments. Grounding one conductor of a two
conductor power cable is not sufficient protection. Serious injury or death
can occur if this grounding is not properly supplied.
WARNING
Chemical Hazard
Dry cleaning solvents for cleaning parts may be potentially dangerous. Avoid
inhalation of fumes or prolonged contact with skin.
WARNING
Resuscitation
Personnel working with or near high voltages should be familiar with
modern methods of resuscitation.
WARNING
Remove Power
Observe general safety precautions. Do not open the instrument with the power on.

Safety Symbols
WARNING
Warning notes call attention to a procedure which, if not correctly
performed, could result in personal injury.
CAUTION
Caution notes call attention to a procedure which, if not correctly performed,
could result in damage to the instrument.
The caution symbol appears on the equipment indicating there is
important information in the instruction manual regarding that
particular area.

 Note: Calls attention to supplemental information.
iii

Warning Statements
The following safety warnings appear in the text where there is danger to
operating and maintenance personnel and are repeated here for emphasis.
WARNING
Never attempt to connect or disconnect RF equipment from the transmission line
while RF power is being applied. Leaking RF energy is a potential health hazard.
Refer to page 5, 35, and 41.
WARNING
To avoid personal injury, disconnect the power cord from the AC line before
performing any maintenance, including fuse replacement or changing the
line voltage setting.
Refer to page 35, 38, 39, and 41.
WARNING
Use only Nickel Metal Hydride (NiMH) batteries that have a minimum
capacity of 4500 milliamper hours (mAh). Do not install batteries that are
not Nickel Metal Hydride. Do not install NiMH batteries that have less than
4500 mAh capacity. Failure to comply may result in damage to the batteries,
damage to the instrument, and injury to personnel from battery chemicals.
Refer to page 40.
WARNING
Heavy load. Do not attempt to lift unaided.
Refer to page 41.

Caution Statements
The following equipment cautions appear in the text whenever the equipment
is in danger of damage and are repeated here for emphasis.
CAUTION
Changing the sensor’s connectors will invalidate calibration data, and may
reduce the maximum power rating of the unit.
Refer to page 2, 41, 43, 45, and 49.
CAUTION
The Bird 4421 must be powered off when connecting or disconnecting the
power sensor from the power meter.
Refer to page 5 and 41.
CAUTION
Do not use the power sensor with a load VSWR greater then 2:1. Damage to
the power meter, power sensor, or the RF power source could occur.
Refer to page 5.

iv

CAUTION
Long-term storage of this instrument can affect battery performance and
reduce battery life. Do not store the instrument for long periods of time
without recharging the batteries (refer to page 40). Failure to comply may
result in reduced battery charge and shortened battery life.
Refer to page 7
CAUTION
During remote operation, periodically monitor the bus service request line.
Failure to detect a service request could result in equipment damage.
Refer to page 11 and 16.
CAUTION
Due to the complexity of the Bird Power Sensor, field repairs beyond general
maintenance should not be attempted. Removal or disturbance of the power
sensor cover can result in cancellation of lifetime warranty.
Refer to page 35 and 41.
CAUTION
Failure to install the properly rated fuse may result in equipment damage or
nuisance failures.
Refer to page 39.

v

Safety Statements
USAGE

ANY USE OF THIS INSTRUMENT IN A MANNER NOT SPECIFIED
BY THE MANUFACTURER MAY IMPAIR THE INSTRUMENT’S
SAFETY PROTECTION.
USO

EL USO DE ESTE INSTRUMENTO DE MANERA NO
ESPECIFICADA POR EL FABRICANTE, PUEDE ANULAR LA
PROTECCIÓN DE SEGURIDAD DEL INSTRUMENTO.
BENUTZUNG

WIRD DAS GERÄT AUF ANDERE WEISE VERWENDET ALS VOM
HERSTELLER BESCHRIEBEN, KANN DIE GERÄTESICHERHEIT
BEEINTRÄCHTIGT WERDEN.
UTILISATION

TOUTE UTILISATION DE CET INSTRUMENT QUI N’EST PAS
EXPLICITEMENT PRÉVUE PAR LE FABRICANT PEUT
ENDOMMAGER LE DISPOSITIF DE PROTECTION DE
L’INSTRUMENT.
IMPRIEGO

QUALORA QUESTO STRUMENTO VENISSE UTILIZZATO IN
MODO DIVERSO DA COME SPECIFICATO DAL PRODUTTORE LA
PROZIONE DI SICUREZZA POTREBBE VENIRNE
COMPROMESSA.

vi

SERVICE

SERVICING INSTRUCTIONS ARE FOR USE BY SERVICETRAINED PERSONNEL ONLY. TO AVOID DANGEROUS
ELECTRIC SHOCK, DO NOT PERFORM ANY SERVICING UNLESS
QUALIFIED TO DO SO.
SERVICIO

LAS INSTRUCCIONES DE SERVICIO SON PARA USO
EXCLUSIVO DEL PERSONAL DE SERVICIO CAPACITADO. PARA
EVITAR EL PELIGRO DE DESCARGAS ELÉCTRICAS, NO
REALICE NINGÚN SERVICIO A MENOS QUE ESTÉ CAPACITADO
PARA HACERIO.
WARTUNG

ANWEISUNGEN FÜR DIE WARTUNG DES GERÄTES GELTEN
NUR FÜR GESCHULTES FACHPERSONAL.
ZUR VERMEIDUNG GEFÄHRLICHE, ELEKTRISCHE SCHOCKS,
SID WARTUNGSARBEITEN AUSSCHLIEßLICH VON
QUALIFIZIERTEM SERVICEPERSONAL DURCHZUFÜHREN.
ENTRENTIEN

L’EMPLOI DES INSTRUCTIONS D’ENTRETIEN DOIT ÊTRE
RÉSERVÉ AU PERSONNEL FORMÉ AUX OPÉRATIONS
D’ENTRETIEN. POUR PRÉVENIR UN CHOC ÉLECTRIQUE
DANGEREUX, NE PAS EFFECTUER D’ENTRETIEN SI L’ON N’A
PAS ÉTÉ QUALIFIÉ POUR CE FAIRE.
ASSISTENZA TECNICA

LE ISTRUZIONI RELATIVE ALL’ASSISTENZA SONO PREVISTE
ESCLUSIVAMENTE PER IL PERSONALE OPPORTUNAMENTE
ADDESTRATO. PER EVITARE PERICOLOSE SCOSSE
ELETTRICHE NON EFFETTUARRE ALCUNA RIPARAZIONE A
MENO CHE QUALIFICATI A FARLA.

vii

UNITS ARE EQUIPPED WITH RECHAREABLE BATTERIES. THESE ARE TO
BE REPLACED BY AUTHORIZED SERVICE PERSONNEL ONLY!!!
LAS UNIDADES VIENEN EQUIPADAS CON BATERIAS RECARGABLES. ¡¡¡Y
SOLAMENTE EL PERSONAL DE SERVICIO AUTORIZADO PUEDE
REEMPLAZARLAS!!!
GERÄTE SIND MIT WIEDER AUFLADBAREN BATTERIEN BESTÜCKT.
BATTERIEN SIND NUR VON QUALIFIZIERTEM SERICE PERSONAL
AUSZUWECHSELN!!!
CES DISPOSITIFS SONT ÉQUIPÉS DE BATTERIES RECHARGEABLES.
SEUL LE PERSONNEL D’ENTRETIEN AUTORISÉ EST HABILITÉ À LES
REMPLACER !
LE UNITÀ SONO DOTATE DI BATTERIE RICARICABILI, CHE DEVONO DA
COME SPECIFICATO DAL PRODUTTORE LA PROTEZIONE DI SICUREZZA
POTREBBE VENIRNE COMPROMESSA.

viii

BE SURE THE 115/230V AC VOLTAGE SELECTOR IS SET TO THE PROPER
LINE VOLTAGE, AND THE CORRECT AC LINE FUSE IS INSTALLED BEFORE
AC POWER IS APPLIED.
S’ASSURER QUE LE SÉLECTEUR DE TENSION 115/230V C.A. EST BIEN
RÉGLÉ POUR LA TENSION DU RÉSEAU ET QUE LE FUSIBLE DE LIGNE
C.A. CORRECT EST EN PLACE AVANT DE METTRE SOUS TENSION C.A.
CERCIORESE QUE EL SELECTOR DE VOLTAJE DE 115/230V CA ESTE
COLOCADO A LA LINEA DE VOLTAJE APROPIADA Y QUE EL FUSIBLE
ESTE INSTALADO A LA LINEA CA ANTES DE APLICAR LA CORRIENTE
ALTERNA.
VOR EINSCHALTEN DER WECHSELSTROMZUFUHR SICHERSTELLEN,
DASS DER 115/230V WECHSELSPANNUNGS-SELEKTOR AUF DIE
VORSCHRIFTSMÄSSIGE LEITUNGSSPANNUNG EINGESTELLT UND DIE
RICHTIGE WECHSELSTROM-HAUPTSICHERUNG EINGESETZT IST.
PRIMA DI EROGARE CORRENTE, ASSICURARSI CHE IL SELETTORE DI
VOLTAGGIO 115/230 V.C.A. SIA REGOLATO CORRETTAMENTE E CHE IL
FUSIBLE ADATTO ALLA LINEA DI ALIMENTAZIONE C.A. SIA INSTALLATO.

ix

About This Manual

This manual covers the Bird 4421 RF Power Meter and the its sensors. This manual covers the operating and maintenance instructions for the following models:
Power Meter

4421
4021

4022

4024

4025

4027A250K
4027A4M
4027A35M
4027F2M

4027A400K
4027A10M
4027A60M
4027F10M

4027A800K
4027A12M
4027A100M
4027F60M

4027A2M
4027A25M
4027A150M

4028A Series
Sensors

4028A250K
4028A4M

4028A400K
4028A10M

4028A2M
4028A25M

4028A3M

4028B Series
Sensors

4028B10M

4028B3M

4028C Series
Sensors

4028C10M

4020 Series
Sensors
4027A Series
Sensors
4027F Series
Sensors

Changes to this Manual
We have made every effort to ensure this manual is accurate. If you should
discover any errors, or if you have suggestions for improving this manual,
please send your comments to our factory. This manual may be periodically
updated. When inquiring about updates to this manual refer to the part number and revision level on the title page.

Contents
Chapter Layout
Introduction - Describes the purpose and function of the Power Meter as well
as a general overview of the product.
Installation - Describes the installation instructions for the power meter models
covered in this manual.
Operating Instructions - Describes the features of the power meters and provides power-up instructions.
IEEE-488 GPIB Interface - Describes the features of the IEEE-488 GPIB Interface.
RS-232 Interface - Describes the features of the RS-232 Interface.
Maintenance - Contains preventative maintenance information, troubleshooting section, and a list of replacement parts with part numbers.

x

Table of Contents
Safety Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .iii
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Warning Statements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Caution Statements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Safety Statements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

iii
iv
iv
vi

About This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .x
Changes to this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . x
Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . x
Chapter Layout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . x
Chapter 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Items Supplied . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Optional Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Power Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4020 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4027A Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4027F Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4028 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Frequency and Power Ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4020 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4027A Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4027F Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4028 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1
1
1
2
2
2
2
2
4
4
4
4
4

Chapter 2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Sensor Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
RF Line Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Panel Mounting the 4421 Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Handle Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
115/230V AC Input Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
AC Line Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5
5
5
6
6
6
7

Chapter 3 Operating Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Push Button Functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Audible Warning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Chapter 4 IEEE-488 GPIB Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cable Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Interface Capabilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Dip Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Talker-Only Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Command Syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
General Bus Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Device Dependent Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

11
11
12
12
12
12
13
13
13
17
xi

Chapter 5 RS-232 Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cable Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
DIP Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Auto Baud . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Talker-Only Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Command Syntax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
General Bus Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Device Dependent Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

25
25
25
26
26
27
27
27
27
29

Chapter 6 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Functional Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Push Button Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Front Panel (Cal Cart only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cord Reel (Cal Cart only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Replacing Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Replacing Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Long Term Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Power Sensor (Cal Cart only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Casters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Handle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Storage and Shipment (Cal Cart only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Specifications, Cal Cart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Max. Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Specifications, 4421 Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Specifications, Power Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Specifications, RS-232 Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Specifications, IEEE-488 GPIB Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Customer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

35
35
37
37
38
38
39
39
40
40
41
41
41
41
42
42
42
43
43
47
47
50

Limited Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51

xii

Chapter 1

Introduction

Power Meter
The Bird 4421 RF Power Meter is one component of a complete RF power measurement system. An RF power sensor such as a Bird 4021 is also required.
The system can be controlled with the front panel buttons, or remotely
through an RS-232 connection or a GPIB-488 connection.

Items Supplied
z
z
z
z

Bird 4421 RF Power Meter
AC Power Cord
Sensor Cable
Instruction Manual

Optional Accessories
Panel Mount Kit (P/N 4421-250) - Allows the Bird 4421 to be installed in a
standard 19" panel for rack mount applications.
Null Modem Kit (P/N 4380-250) - Contains the hardware necessary to allow
the 4421 to be remotely controlled by controllers with different wiring
arrangements. Requires an RS-232 interface module.
Figure 1

Bird 4421 Meter Outline Drawing
10-1/8” (257 mm)
MODEL 4421 RF POWER METER

1. Operating Push
Buttons

RANGE
AUTO

4-1/4”
(108 mm)

UP

2. LCD

DOWN

FWD

RFL

SWR

MIN

MAX

dBm

LIGHT

3. ON/OFF Push Button

ON / OFF

4. Handle
1

3

2

FRONT

5. Central Button
6. Fuse Access Door
7. AC Line Module

12-1/4” (312 mm)

8. Master ON/OFF
Switch
9. GPIB connector
10. GPIB DIP switches
11. RS-232 connector

5

4

SIDE
8

7

12. RS-232 DIP switches
9

10

12

11

13. Power Sensor Socket

6
I ON
O OFF

AC SUPPLY
REMOVE AC POWER BEFORE
ATTEMPTING TO SERVICE THIS
INSTRUMENT

POWER
SENSOR

VOLTAGE
115/230 V
FUSE RATING T630mA/T315mA
FREQUENCY 50/60 Hz
MAXIMUM POWER 60W

13

REAR

1

Bird 4421 RF Power Meter

Power Sensors
Power sensors are available with a variety of connectors; see "Available Connectors" on page 49 for a complete list. Since the accuracy is critically dependent on the connectors used at calibration, do not remove or change the
connectors.
CAUTION
Changing the sensor’s connectors will invalidate calibration data, and may
reduce the maximum power rating of the unit.
4020 Series
Bird 4020 Series Power Sensors are designed for lab or field use and are accurate to within ±3%(1σ) of reading.
4027A Series
Bird 4027A Series Power Sensors are designed for use in semiconductor processing and calibration applications. Stringent calibration provides long-term
unit-to-unit repeatability, allowing consistent amounts of RF energy to be
applied to the etch process over many etch cycles. 4027A Sensors are accurate
to ±1%(1σ) at specified calibration frequencies and power levels.
4027F Series
Bird 4027F Series Power Sensors are similar to the 4027A series. However,
additional filtering allows the 4027F to ignore harmonics of the signal being
measured. The 4027F is also less sensitive to AM components of the signal.
4027F Sensors are accurate to ±1%(2σ) at specified calibration frequencies
and power levels.
Figure 2 Power Sensor Outline Drawing, 4020, 4027A, 4027F, & 4028A Series
3-3/4”
(95 mm)

3/4”
(19 mm)

SENSOR PLUG
SOCKET

2-1/2”
(64 mm)

3-1/4”
(83 mm)

DIRECTIONAL POWER SENSOR
MODEL 4022
0.3 - 1000 WATTS
25 - 1000 MHz

1-3/8”
(35 mm)

-CAUTION-

SOURCE

LOAD

4028 Series
Bird 4028 Series Power Sensors are high power sensors otherwise similar to
the 4027A series. 4028A sensors are based on a 7/8” line, 4028B sensors use a
1-5/8” line, and 4028C sensors use a 3-1/8” line. 4028 Sensors are accurate to
±2% at specified calibration frequencies and power levels.

2

Introduction

Figure 3

Power Sensor Outline Drawing 4028B Series only
3.75"
(96 mm)

SOURCE

SENSOR PLUG
SOCKET

LOAD

4.75"
(121 mm)
3.5"
(89 mm)

6.75"
(172 mm)

Figure 4

Power Sensor Outline Drawing 4028C Series only

3

Bird 4421 RF Power Meter

Frequency and Power Ranges
4020 Series
Model

Frequency Range

RF Power Range

4021
4022
4024
4025

1.8 – 32 MHz
25 – 1000 MHz
1.5 – 32 MHz
100 kHz – 2.5 MHz

300 mW – 1 kW
300 mW – 1 kW
3 W – 10 kW
3 W – 10 kW

Model

Frequency Range

RF Power Range

4027A250K
4027A400K
4027A800K
4027A2M
4027A4M
4027A10M
4027A12M
4027A25M
4027A35M
4027A60M
4027A100M
4027A150M

250 – 400 kHz
400 – 550 kHz
800 – 950 kHz
1.5 – 2.5 MHz
3 – 5 MHz
10 – 15 MHz
10 – 15 MHz
25 – 30 MHz
35 – 45 MHz
45 – 65 MHz
95 – 105 MHz
150 – 170 MHz

3 W – 10 kW
3 W – 10 kW
3 W – 10 kW
3 W – 10 kW
3 W – 10 kW
3 W – 10 kW
300 mW – 1 kW
3 W – 9 kW
3 W – 7.5 kW
3 W – 6 kW
3 W – 4 kW
3.75 W – 3.75kW

Model

Frequency Range

RF Power Range

4027F2M
4027F10M
4027F60M

1.8 – 2.2 MHz
12 – 15 MHz
57 – 63 MHz

100 W – 10 kW
3 W – 4 kW
3.75 W – 3.75 kW

Model

Frequency Range

RF Power Range

4028A250K
4028A400K
4028A2M
4028A3M
4028A4M
4028A10M
4028A25M
4028B3M
4028B10M
4028C10M

250 – 400 kHz
400 – 550 kHz
1.5 – 2.5 MHz
2.5 – 3.5 MHz
3.5 – 4.5 MHz
10 – 15 MHz
25 – 30 MHz
2.5 – 4 MHz
10 – 15 MHz
10 – 15 MHz

1 kW – 20 kW
1 kW – 20 kW
1 kW – 25 kW
1 kW – 25 kW
1 kW – 25 kW
1 kW – 25 kW
1 kW – 25 kW
1 kW – 25 kW
1 kW – 25 kW
500W – 50 kW

4027A Series

4027F Series

4028 Series

4

Chapter 2

Installation

This chapter provides information about preparing the Bird 4421 for use.

Sensor Connection
CAUTION
Changing the sensor’s connectors will invalidate calibration data, and may
reduce the maximum power rating of the unit.
CAUTION
The Bird 4421 must be powered off when connecting or disconnecting the
power sensor from the power meter.
1. Turn OFF the ON/OFF rocker switch on the meter’s rear panel.
2. Align the latch on the cable with the notch of the “Power Sensor” socket on
the power meters rear panel.
3. Insert the cable until it clicks into place.
4. Connect the other end of the sensor cable to the sensor.

RF Line Connection
WARNING
Never attempt to connect or disconnect RF equipment from the transmission
line while RF power is being applied.
Leaking RF energy is a potential health hazard.
CAUTION
Do not use the power sensor with a load VSWR greater then 2:1. Damage to
the power meter, power sensor, or the RF power source could occur.
Connect the end of the power sensor labeled “SOURCE” to the RF source. Connect the end labeled “LOAD” to the load or antenna. Reversing these connections will cause measurement errors.

Panel Mounting the 4421 Power Meter
You can install the 4421 Power Meter in an equipment rack if you have the
optional panel mount kit (refer to Optional Accessories, page 1). The panel
mount kit includes complete installation instructions. Figure shows the overall dimensions and mounting points for a 4421 Power Meter installed in a
panel mount kit.

 Note: The power supply interrupt switch for the 4421 Power Meter
is located on the rear panel. When you install a unit in a panel mount
kit, you need to provide a means to interrupt the power supply that is
easily accessible to the user (such as a switch mounted in the panel).

5

Bird 4421 RF Power Meter

Figure 5

Panel Mounting Dimensions

4421

RF POWER METER

POWER

SENSOR

FWDRFL SWRMIN MAXdBmLIGHT

FUSE

I ON
O OFF

POWER
SENSOR

WARN I NG

REMOVE AC POWER BEFORE ATTEMPTING
TO SERVICE THIS INSTRUMENT.

AC SUPPLY

FUSE
VOLTAGE
100 - 132 VAC
T .250A, 250V
178 - 264 VAC
T .125A, 250V
AT 47.5 - 66 Hz
MAXIMUM POWER 15W

Handle Operation
The handle on the Bird 4421 can be set to four different positions (see
Figure 6). To adjust the handle, press the center buttons on both sides.
Releasing the buttons will lock the handle into position.
Figure 6

Handle Positions

115/230V AC Input Power
The internal power supply can operate at 115 VAC or 230 VAC. There is no
user setting to select the input voltage because the power supply automatically senses the input voltage. However, you must make sure that you have
the correct power cord plug and fuses installed for the line voltage you are
using (refer to Replacing Fuses, page 39).
AC Line Connectors
To make the AC line cord compatible with non-U.S. voltages, users must
install the appropriate connector on the power cord.
6

Installation

Batteries
The Bird 4421 RF Power Meter is completely portable and is powered from
internal rechargable nickel metal hydride batteries.
The batteries are shipped in a low charge state. It is, therefore, recommended
that you chrage the instrument for approximately 28 hours before using it for
continous operation.

 Note: Battery charging at temperatures greater than 45° C (113°
F) can result in reduced operational time.

CAUTION
Long-term storage of this instrument can affect battery performance and
reduce battery life. Do not store the instrument for long periods of time
without recharging the batteries (refer to page 40). Failure to comply may
result in reduced battery charge and shortened battery life.

7

Chapter 3

Operating Instructions
This chapter describes operator controls and indicators on the Bird 4421 RF
Power Meter. For remote operation using a GPIB or RS-232 controller, refer to
the instructions in Chapter 4 or Chapter 5 respectively.

Push Button Functions
Figure 7

Push Buttons
MODEL 4421 RF POWER METER
AUTO

FWD
RFL
SWR

MW nW
KW µW
W mW
dBm

LO BAT
REMOTE

FWD

LISTEN

RFL

SWR

TALK

MIN

LOCAL

MAX

AUTO

dBm

LIGHT

UP
DOWN

LOCKOUT

Push Button
FWD, RFL

RANGE

ON / OFF

Description
Press to measure forward or reflected RF power. FWD or RFL
indicator and current unit of measure turn on.

SWR

Press to measure standing wave ratio. SWR indicator turns on. Value
displayed will be between 1.0 and 199.9

MIN, MAX

Used after pressing FWD, RFL, SWR, or dBm. Displays the minimum
(maximum) measured value of the previous function as long as MIN
(MAX) is held down.

dBm

Used after pressing FWD or RFL. dBm indicator turns on. Power is
displayed in dBm units.

 Note: Used after pressing SWR. Return loss is displayed.
LIGHT

Press to turn on or turn off the display’s backlight. If left on, the light
automatically shuts off after 30 minutes.

AUTO

Press to automatically set the scale. AUTO turns on.

UP, DOWN

Press to select the next higher (lower) scale. If the scale is too high for
the power sensor, an error will be displayed.

 Note: Used while AUTO indicator is on. Stops automatic
scaling. AUTO indicator turns off.

ON/OFF

Press to turn the power meter on or off. This switch will not work if the
master power switch on the rear panel is OFF.

 Note: The display flashes for about 30 seconds when the
unit is first turned on.

9

Bird 4421 RF Power Meter

Error Codes
The Bird 4421 displays error codes when the RF power is either below the
selected range (underrange) or above the selected range (overrange). Table 4-1
displays the error codes and Table 4-2 lists the function limits.

Table 4-1 Error Codes
Symbol

Explanation
Value greater than overrange limit of function
Value less than underrange limit of function

Table 4-2 Function Limits
Function

Limit

Error

FWD, RFL

Power > 199.9% of full scale or 120% of
top range

Overrange

FWD dBm, RFL dBm

Power > 120% of full scale

Overrange

Power < 3% of low range

Underrange

FWD < 20% of low range

Underrange

FWD – RFL = 0

Overrange

FWD < 20% of low range

Underrange

RFL < 20% of low range

Underrange

Return Loss > 40 dB

Underrange

SWR

Return Loss

Audible Warning
If the RF power level exceeds 120% of the power sensor’s maximum power
capability, the power meter will sound a warning buzzer.

10

Chapter 4

IEEE-488 GPIB Interface
This chapter discusses setup of the IEEE-488 interface feature and describes
the IEEE commands that apply to the Bird 4421. Operators should understand IEEE standard 488-1978 and have basic computer programming skills
before attempting to write any programs.
CAUTION
During remote operation, periodically monitor the bus
service request line. Failure to detect a service request
could result in equipment damage.

Description
The Bird 4421 IEEE-488 (GPIB) interface has an eight-position DIP switch
that sets operational conditions and interface addresses. The bottom line of
the display indicates the current bus status.
Cable Connector
The interface uses a standard IEEE-488 cable connector. Pin assignments are
listed in Table 5-3.

Table 5-3 IEEE-488 Pin Assignments
Pin

Designation

Type

1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24

D101
D102
D103
D104
EOI
DAV
NRFD
NDAC
IFC
SRQ
ATN
SHIELD
D105
D106
D107
D108
REN
GROUND
GROUND
GROUND
GROUND
GROUND
GROUND
GROUND, LOGIC

Data
Data
Data
Data
Management
Handshake
Handshake
Handshake
Management
Management
Management
Ground
Data
Data
Data
Data
Management
Ground
Ground
Ground
Ground
Ground
Ground
Ground

11

Bird 4421 RF Power Meter

Interface Capabilities

Table 5-4 IEEE-488 Interface Module Capabilities
Code Name
SH1
AH1
T5

L4
SR1
RL1
PP0
DC1
DT1
C0
E1
TE0
LE0

Description

Source Handshake Can handshake data or command bytes when the unit is
acting as a source.
Acceptor
Can handshake the bus when it is acting as the acceptor of
data or commands.
Talker
Can send data over the bus to other devices. This capability
exists only after the instrument has been addressed to talk, or
after a reading in talk-only mode.
Listener
Can receive device-dependent data over the bus. This capability
exists only after the unit has been addressed to listen.
Service Request
Can request service from the controller.
Remote-Local
Can be placed in remote or local mode.
Parallel Poll
Does not have parallel polling capability.
Device Clear
Can be reset to factory settings.
Device Trigger
Can have its readings triggered.
Controller
Does not have controller capability.
Bus Driver Type
Has open-collector bus drivers.
Extended Talker
Does not have extended talker capability.
Extended Listener Does not have extended listener capability.

Indicators
The bottom line of the power meter’s display shows indicators describing the
status of the Bird 4421 when used with the IEEE interface. These are:
REMOTE - When REMOTE is displayed, the power meter is being controlled
through the interface. Measurements, units of measure, and certain other
parameters may be changed from a remote location.
LISTEN - When LISTEN is displayed, the power meter is receiving data.
TALK - When TALK is displayed, the power meter is transmitting data. This
is always shown when the unit is in “talker-only” mode.
LOCAL LOCKOUT - When LOCAL LOCKOUT is displayed, the push buttons
are disabled and the power meter’s functions are being remotely controlled.

Setup
Dip Switch
Set the interface to ADDR (Addressable) by positioning DIP switch 1 to ON.
This makes the 4421 respond to controller commands.

 Note: There are several button styles on DIP switches (slide, rocker,
lever). Examine the DIP switch to determine the ON and OFF positions.

Set the primary address using the DIP switches. The primary address is factory set to 6, but can be set to any value between 1 and 31 (0 is reserved for
the controller) To set the primary address, turn OFF switches 4 – 8 so that the
sum of the bits turned off equals the desired primary address. In Figure 8
below, the address is set to 6 (the off switches have values of 4 and 2).
Figure 8

IEEE Interface Default DIP Switch Settings

 Note: The primary address is used by the

controller to refer to specific devices on the bus.
When programming the controller, the address in
the program must be the same as the address set
on the interface module. Each device on the bus
must have a different primary address.

12

IEEE-488 GPIB Interface

Talker-Only Mode
The Bird 4421 can be set up for manual operation while automatically sending
data to an output device (Talker-Only Mode). To do so, turn DIP switch 1 OFF
and cycle the power. TALK will be displayed.
In Talker-Only mode, pressing a button on the meter triggers a measurement.
When the measurement is complete, the information is sent to the bus and
LISTEN turns on momentarily. A listen-only device on the bus, such as a
printer, can read the value. The power meter is then ready to accept another
button press.

Command Syntax
The Bird 4421 accepts two types of commands. General bus commands are
commands, such as Device Clear (DCL), that apply to any IEEE interfaced
device. Device-dependent commands are specific to the 4421.
If an invalid command is sent to the unit, an error condition is placed in the
serial poll byte and the offending command is not executed.
A group of device-dependent commands can be sent as a single string as long
as like command categories are not repeated, for example: “PNFCFDT3TRG”.
This string sets up the 4421 to send no prefixes, read forward dBm, make one
reading on “TRG”, and triggers a measurement.

 Note: Commands can be entered in either upper or lower case.
 Note: Only the last command entered of each category will be executed. As a command string is processed by the 4421, each category of
command is stored in a separate location. Two commands of the same
category will be stored in the same location, so that the second will
overwrite and erase the first one.

General Bus Commands
The general bus commands supported by the IEEE-488 interface feature are
listed in Figure 5-5. The syntax for executing general commands varies among
controllers; check the documentation supplied with your controller for the
proper command structure.

Table 5-5 IEEE-488 General Bus Commands
Command

Effect on Bird 4421

IDN?

Gives product identification

REN

Goes into remote mode when next addressed

GTL

Cancels remote mode, restores local operation

LLO

Locks out local operation

IFC

Goes into talker and listener-idle status

DCL

Returns to default conditions

SDC

Returns to default conditions

GET

Triggers reading in T2 and T3 modes

SPE, SPD

Puts the status byte on the bus

13

Bird 4421 RF Power Meter

IDentificatioN

(IDN?)

Function

Product identifies itself

Remarks

Replies command same as U2

Remote ENable

(REN)

Function

Enables remote operation.

Remarks

The unit must be addressed to listen after setting REN true.

The REMOTE indicator turns on when this command is received.

Go To Local

(GTL)

Function

Returns device to local operation.

Remarks

Issuing a GTL command while the device is in Local Lockout mode does not
clear the lockout condition.
The REMOTE indicator turns off.
The LISTEN indicator remains on.

Local LOckout

(LLO)

Function

Disables local operation of all devices on the bus.

Remarks

REN must be true to use LLO.
LLO is cleared by setting REN false.

InterFace Clear

(IFC)

Function

Terminates all bus activity and passes control to the system controller.

Remarks

All devices are set to talker and listener idle states.
LISTEN mode is cancelled, and its indicator is turned off.

14

IEEE-488 GPIB Interface

Device CLear

(DCL)

Function

Resets the status of all devices to an initialized state.

Remarks

Does not change the current interface mode.
The 4421 returns to the factory default condition listed in Figure 5-6.

Table 5-6 IEEE-488 Default Conditions
Default Condition

Related Command

Forward Carrier Wave

FC

Auto Range ON

RYY

Two Terminators (CR LF)

YT

Prefixes YES

PY

Trigger One Shot on Talk Address

T1

All SRQ’s OFF

M00

Send EOI with last byte of message

KY

Selective Device Clear

(SDC)

Function

Resets the status of a selected device to an initialized state.

Remarks

Only the device addressed will be cleared.
The 4421 returns to the factory default condition.

Group Execute Trigger

(GET)

Function

Initiates a measurement for all devices set to trigger on GET.

Remarks

The 4421 must be already set to trigger on GET.
Used to synchronize measurements of multiple instruments.

15

Bird 4421 RF Power Meter

Serial Polling Enable/Disable
Function

(SPE/SPD)

Enables or disables the serial polling sequence.
CAUTION
During remote operation, periodically monitor the bus service request line.
Failure to detect a service request could result in equipment damage.

Remarks

The SPE command puts all devices in serial poll mode waiting to be addressed.
The SPD command clears the SRQ bit (bit 6) and ends the polling sequence.
When addressed a device sends its status byte to the controller. A value of 1
for a bit means that the device condition that bit refers to is true. A value of 0
means that the condition is false.
The 4421 does not use all bits of the status byte. Table 5-7 lists the bits used,
along with a description and how to reset them.

Table 5-7 IEEE Status Byte Description
Bit

16

Name

6

SRQ

3

Measurement
Complete

2

Reading
Underflow

1

Reading Overflow

0

Error

Condition
Set if a service request is generated by the 4421. If an SRQ
has been received by the controller and this bit is cleared,
other instruments on the bus should be checked to
determine where the SRQ occurred.
Cleared by a serial poll of the 4421.
Set when the power meter has completed a reading.
Cleared by requesting a reading over the bus.
Set when the RF power is underrange and a reading has
been completed.
Cleared by requesting a reading over the bus.
Set when the RF power is overrange and a reading has
been completed.
Cleared by requesting a reading over the bus.
Set if an illegal device-dependent command (IDDC) or
illegal device-dependent command option (IDDCO) was
received, or if the power meter fails the self test.
Cleared by reading the U1 status word. The U1 word
contains details on the error, see “Status” on page 21.

IEEE-488 GPIB Interface

Device Dependent Commands
The device-dependent commands used by the 4421 Power Meter are listed in
Table 5-8, organized by category.

 Note: The programming card also has a complete command list.
Table 5-8 IEEE-488 Device Dependent Command Summary
Category
Measurement

Range

Terminators

Prefixes
Triggers

Serial Polling

Status

Command
FC

Description
Forward carrier wave

FD

Forward dBm

RC

Reflected carrier wave

RD

Reflected dBm

SW

Standing wave ratio

RL

Return loss

MN

Minimum value

MX

Maximum value

RYY

Auto range on

R00 to R17

Manual ranges

RNN

Auto range off, stay at present range

YT

Two terminators: CR, LF

YO

One terminator: CR

YN

No terminator

PY

Prefix YES

PN

Prefix NO

T0

Continuous on TALK

T1

One shot on TALK

T2

Continuous on GET

T3

One shot on GET

T4

Continuous on measurement command

T5

One shot on measurement command

M00

Do not generate SRQ

M01

Generate SRQ on error

M02

Generate SRQ on measurement overrange

M04

Generate SRQ on measurement underrange

M08

Generate SRQ on operation complete

U0

Send back current machine state

U1

Send back error conditions

U2

Send back revision levels

Self-Test

J0

Run self-test

EOI Response

KY

Send EOI on last byte

KN

Do not send EOI on last byte

WXXXXXX

Place XXXXXX in RAM

Writable Store

17

Bird 4421 RF Power Meter

Forward Carrier Wave
Forward dBm
(FD)
Reflected Carrier Wave
Reflected dBm
(RD)

(FC)
(RC)

Function

Selects forward or reflected RF power measurement mode.

Remarks

Measurement results are returned in Watts or dBm.

Standing Wave Ratio
Return Loss
(RL)

(SW)

Function

Selects SWR or return loss match measurement mode.

Remarks

Measurement results are returned in VSWR or dB.

MiNimum Value
MaXimum Value

(MN)
(MX)

Function

Selects minimum or maximum measurement mode.

Remarks

Another measurement must be selected before selecting min or max.
Returns the minimum (or maximum) value of the previous measurement type.

18

IEEE-488 GPIB Interface

Range

(Rxx)
Function

Selects a measurement range listed in Table 5-9.

Remarks

If the selected range is outside the range of the connected power sensor, the
command is ignored.

Table 5-9 Measurement Ranges
Command
RYY
RNN
R17
R16
R15
R14
R13
R12
R11
R10
R09
R08
R07
R06
R05
R04
R03
R02
R01
R00

Terminators

Power Range
Turn auto range on
Turn auto range off. Keep present range
18.0 – 199.9 MW
1.80 – 19.99 MW
0.180 – 1.999 MW
18.0 – 199.9 kW
1.80 – 19.99 kW
0.180 – 1.999 kW
18.0 – 199.9 W
1.80 – 19.99 W
0.180 – 1.999 W
18.0 – 199.9 mW
1.80 – 19.99 mW
0.180 – 1.999 mW
18.0 – 199.9 µW
1.80 – 19.99 µW
0.180 – 1.999 µW
18.0 – 199.9 nW
1.80 – 19.99 nW
0.180 – 1.999 nW

(Yx)

Function

Selects the characters that follow the end of a data string. Set x to:
z
z
z

Remarks

“T” for two terminators; a carriage return(CR) and a line feed(LF).
“O” for one terminator; a carriage return(CR).
“N” for no terminator; message can be terminated by EOI.

Many controllers use the terminator sequence to recognize the end of an input
string. Using incorrect terminators can lock the bus.

19

Bird 4421 RF Power Meter

Prefixes

(Px)

Function

Turns the prefix mode on or off. Set x to:
z
z

Remarks

“Y” to enable prefixes.
“N” to disable prefixes.

Prefixes are sent over the bus with the measurement, and indicate the status
of the current measurement (see Table 5-10 for examples):
z
z
z
z
z

“FC”,”FD”,”RC”,”RD”,”SW”,”RL”,”MN”,”MX” indicates the measurement type.
“U” indicates underflow; the value sent is “.000”.
“O” indicates overflow; the value sent is “199.9”.
“N” indicates normal; the value sent is a normal on-scale reading.
“4421” indicates the Bird model number.

Table 5-10

Prefix Examples

Data String
NFC.0.123W(CR)(LF)

Normal (N) forward carrier wave (FC), prefixes on

OFC 199.9W(CR)(LF)

Overflowed (O) forward carrier wave (FC), prefixes on

199.9W(CR)(LF)
URD

Triggers

Description

.000W(CR)(LF)

Overflowed forward carrier wave, prefixes off
Underflowed (U) reflected dBm (RD), prefixes on

(Tx)

Function

Selects the condition which will trigger a reading (see Table 5-11).

Remarks

Failure to trigger device before requesting a reading will lock the bus.
T1 halts the bus until a reading is available.
T0 and T1 do not set the measurement complete SRQ.
T2, T3, T4, and T5 set a SRQ when the measurement is complete.
Fastest reading rate is 2.4 readings/second; slowest is 1 reading/sec.

Table 5-11

Trigger Conditions

Command

20

Trigger Condition

T0

Continuous on talk

T1

One shot on talk

T2

Continuous on GET

T3

One shot on GET

T4

Continuous on measurement command (FC,FD,
RC, RD, SW, RL, MN, MX)

T5

One shot on measurement command (FC, FD,
RC, RD, SW, RL, MN, MX)

IEEE-488 GPIB Interface

SRQ Mask

(Mxx)

Function

Selectively masks status bits to prevent unwanted service requests.
Set xx to the sum of the binary values of the desired SRQ trigger bits. For
example, M12 would set the SRQ for both operation complete and underrange
(values 8 and 4). Set xx to “00” to never generate an SRQ.

Remarks

If a status bit is masked (bit set to 0), SRQs won’t be generated for that condition.

Table 5-12

SRQ Mask Bits

Binary Value

Bit Number

Message

1

0 (LSB)

2

1

Measurement overrange

4

2

Measurement underrange

8

3

Operation complete

4

Not Used

5

Not Used

6

Can’t mask

Error (IDDC, IDDCO, self-test fail)

7 (MSB)

Status

Not Used

(Ux)
Function

Reads a status word and returns the information as a string. Set x to:
“0” for machine status.
“1” for error status.
“2” for revision history.

z
z
z

Remarks

After sending the status command, a status word is sent the next time the
unit is addressed to talk. To ensure the correct status is transmitted, the status word should be requested as soon as possible after the command is sent.
Machine Status Word (U0) - The format of the machine status word is shown
in Figure 9. The default values are also shown.
Figure 9
–

4

Machine Status Word Format
4

2

0

–

F

C

R

Y

Y

P

Y

Y

T

T

1

M

0

0

K

Y TERM

Header
Measurement Function
Range
Prefix
Terminator
Trigger
SRQ Mask
EOI Response

21

Bird 4421 RF Power Meter

Error Status Word (U1) - The format of the error status word and the possible
error messages are shown in Figure 10. When an error occurs, an error is also
flagged in the status (serial poll) byte, and a SRQ may be generated (See “SRQ
Mask” on page 21). All flags will revert to their non-error states after the U1
command is sent.
Figure 10

Error Status Word Format
–

4

4

0

2

–

V

C

M

C

V

0

F

L

TERM

Header
Command Status (shown valid)
Command Option Status (shown valid)
Self-test results (shown fail)

Status

Meaning

Description

ICM

Invalid Command

Set when an illegal device-dependent command
(IDDC) such as V2 is received. (V is illegal)

VCM

Valid Command

Set when no IDDC is received.

ICO

Invalid Command Option

Set when an illegal device-dependent command
option (IDDCO) such as T6 is received. (6 is illegal)

VCO

Valid Command Option

Set when no IDDCO is received.

PS

Self-Test Pass

Set when a self-test has been initiated by the
J0 command and the test result is acceptable.

FL

Self-Test Fail

Set when the self-test has failed. (This is the
default condition.)

Revision History Word (U2) - The format of the revision history word is
shown in Figure 11.
Figure 11
–

4

Revision History Word Format
4

2

0

–

X

X

X

X

X

X

Y

*

Y

Z

**

Z

7

8

0

6

TERM

Header
Writable stores
Software Rev level
Hardware Rev level
IEEE-488 1978 compliance
IEEE-488 Bus address
* YY is replaced with current software revision level.
** ZZ is replaced with current hardware revision level.

 Note: If Writable stores parameter has not been set, this command
responds only with -4420-.

22

IEEE-488 GPIB Interface

Self Test

(J0)

Function

Initiates a hardware and software test.

Remarks

Results are stored in the U1 status word (see “Status” on page 21).

“J0” must be sent each time before reading the result.

End Or Identify (Kx)
Function

Enables or disables the End or Identify (EOI) signal. Set x to:
z
z

Remarks

“Y” to enable.
“N” to disable.

Disabling EOI can cause some controllers to lock unless another terminator is used.
When enabled, EOI is only asserted at the end of a multiple byte string.

Writable Store

(Wxxxxxx)

Function

Storage for six bytes of ASCII data.

Remarks

Data stored is lost when the 4421 is turned off.
Data is sent back as part of the U2 status word.

23

Chapter 5

RS-232 Interface
This chapter discusses setup of the RS-232 interface feature and describes the
RS-232 commands that apply to the Bird 4421. Operators should understand
EIA Standard RS-232-C and have basic computer programming skills before
writing any programs.

Description
The Bird 4421 RS-232 interface feature is an integral part of the I/O hub circuit board inside of the mm21. An eight-position DIP switch is used to set
operational conditions such as baud rate, parity, and stop bits. The bottom
line of the display indicates the current bus status.
Cable Connector
The interface uses a standard 25-pin RS-232 connector. Pin assignments are
listed in Table 6-13. If the controller uses a different wiring arrangement, do
not attempt to rewire the interface module’s connector. A null modem kit
should be used for rewiring instead.

Table 6-13
Pin

RS-232 Pin Assignments

Designation

Notes

1

Protection Ground

Chassis Ground

2

Transmit Data

3

Receive Data

4

Request to Send

(Output) Set true after module power up

5

Clear to Send

(Input) Set by input device. When true, it enables the
module to transmit. When false, it disables transmission.

6

Data Set Ready

(Input) Set internally true by module

7

Signal Ground

Return path for data and control signals

8

Receive Signal DET

(Input) Set true by module

20

Data Terminal Ready

(Output) Set true after module power up

Indicators
The bottom line of the power meter’s display shows indicators describing the
status of the Bird 4421 when used with the RS-232 interface. These are:
TALK - When TALK is displayed, the power meter is transmitting data. This
is always shown when the unit is in “talker-only” mode.
LISTEN - When LISTEN is displayed, the power meter is receiving data.

25

Bird 4421 RF Power Meter

Setup
DIP Switch
Set the DIP switches according to application needs and the requirements of
the controller. Available settings and factory defaults are listed in Figure 12.

 Note: There are several button styles on DIP switches (slide, rocker,

lever). Examine the DIP switch to determine the ON and OFF positions.

Figure 12

RS-232 Interface Default DIP Switch Settings
1

2

3

4

5

6

7

8

ON
 Note: If you change switch settings when the unit is ON, the new
settings will not become effective unitl you turn power OFF and then
back ON.

Switch

Function

Description

1

Stop Bit

ON 1 Stop Bit
OFF2 Stop Bits*

2

Command
Mode

ON ENT command needed before sending reading to
controller*
OFFTrigger automatically sends reading
Reading also sent when front panel push button pressed

3

Word Length

ON 8 Data Bits*
OFF7 Data Bits

Parity

(4) (5)
ON ON
ON OFF
OFFON
OFFOFF

No Parity*
Odd Parity
Even Parity
Mark Parity

(6) (7) (8)
ON ON ON
ON OFFON
ON OFFOFF
OFFON ON
OFFON OFF
OFFOFFON
OFFOFFOFF

Auto Baud†
300
600
1200
2400*
4800
9600

4,5

6,7,8

Baud Rate

* Default factory setting
† For more information see “Auto Baud”

26

RS-232 Interface

Auto Baud
Auto Baud is used to automatically determine the correct transmission rate.
After setting DIP switches 1 through 5, follow the steps below to use auto baud:
1.
2.
3.
4.

Set DIP switches 6, 7, and 8 to ON.
Connect a controller to the power meter.
Turn the power meter on.
Send the character U (hexadecimal 55) from the controller to the power
meter.
5. Wait 1 second.
6. If the power meter’s front panel displays LISTEN, the baud rate has been
determined. If not, repeat steps 4 and 5.
If, after 30 seconds, the module cannot absolutely determine a baud rate, it
will choose a rate based on data acquired during the test.

 Note: Auto Baud is the only automatically chosen setting. Other
items such as parity and stop bits must be manually selected.

Talker-Only Mode
The Bird 4421 can be set up for manual operation while automatically sending
data to an output device (Talker-Only Mode). To do so, turn DIP switch 2 OFF
and cycle the power. TALK will be displayed.
In Talker-Only mode, pressing a button on the meter triggers a measurement.
When the measurement is complete, the information is sent to the bus.

Command Syntax
The Bird 4421 accepts two types of commands. General bus commands are
commands, such as Initialize (INT), that apply to any RS-232 interfaced
device. Device-dependent commands are specific to the 4421.
If an invalid command is sent to the unit, an error condition is placed in the
serial poll byte and the offending command is not executed.
A group of device-dependent commands can be sent as a single string as long
as like command categories are not repeated. For example: “PNFCFDT3TRG”.
This string sets up the 4421 to send no prefixes, read forward dBm, make one
reading on “TRG”, and triggers a measurement.

 Note: Commands can be entered in either upper or lower case.
 Note: Only the last command entered of each category will be executed. As a command string is processed by the 4421, each category of
command is stored in a separate location. Two commands of the same
category will be stored in the same location, so that the second will
overwrite and erase the first one.

General Bus Commands
The general bus commands supported by the RS-232 interface module are
listed in Table 6-14.

Table 6-14

RS-232 General Bus Commands

Command
INT
ENT
TRG
B1 to B7
XO/XF

Effect on Bird 4421
Returns to default conditions
Sends a reading to the controller
Triggers reading in T3 mode
Selects a baud rate
Enables/disables software handshake

27

Bird 4421 RF Power Meter

INiTialize

(INT)

Function

Resets the Bird 4421 and returns it to the factory defaults.

Remarks

If INT is linked with any other command within a string, it must be separated
from that command by a space.

ENTer

(ENT)
Function

Makes the power meter transmit a reading to the controller.

Remarks

A measurement must have already been triggered, placing a reading in the
output buffer.
To send a reading whenever a measurement is triggered, set DIP switch 2 to
OFF. The ENT command will not need to be sent.

TRiGger

(TRG)

Function

Baud Select

Initiates a measurement if the power meter is in trigger mode (T3).

(Bx)

Function

Selects a baud rate listed in Table 6-15.

Remarks

When the meter recognizes a valid Bx command, its baud rate is immediately
changed. (The controller is assumed to be transmitting at the new rate; otherwise sending commands would not be possible.)
This command overrides the DIP switch setting.

Table 6-15

Xmission Flow Control
Function

28

Command

Baud Rate

B2
B3
B4
B5
B6
B7

300
600
1200
2400
4800
9600

(XO/XF)

Enables or disables the XON/XOFF flow control.
z
z

Remarks

Baud Rates

XO enables flow control.
XF disables flow control.

When data is being sent from the power meter to the computer and flow control is enabled, data transmission will be suspended when the XOFF character
(hexadecimal 13) is sent by the computer. Transmission will resume when
XON (hex 11) is sent by the computer.
When data is being sent from the computer to the power meter, XOFF will be
sent to the computer when the input buffer fills up. XON will be sent to the
computer when the buffer has emptied.

RS-232 Interface

Device Dependent Commands
The device-dependent commands used by the 4421 Power Meter are listed in
Table 6-16, organized by category.

Table 6-16 RS-232 Device Dependent Command Summary
Category
Measurement

Range

Terminators

Prefixes
Triggers

Status

Self-Test
Writable Store

Forward Carrier Wave
Forward dBm
(FD)
Reflected Carrier Wave
Reflected dBm
(RD)

Command
FC
FD
RC
RD
SW
RL
MN
MX
RYY
R00 to R17
RNN
YT
YO
YN
PY
PN
T0
T1
T3
T5
U0
U1
U2
J0
WXXXXXX

Description
Forward carrier wave
Forward dBm
Reflected carrier wave
Reflected dBm
Standing wave ratio
Return loss
Minimum value
Maximum value
Auto range on
Manual ranges
Auto range off, stay at present range
Two terminators: CR, LF
One terminator: CR
No terminator
Prefix YES
Prefix NO
Continuous on ENT
One shot on ENT
One shot on TRG
One shot on measurement command
Send back current machine state
Send back error conditions
Send back revision levels
Run self-test
Place XXXXXX in RAM

(FC)
(RC)

Function

Selects forward or reflected RF power measurement mode.

Remarks

Measurement results are returned in Watts or dBm.

Standing Wave Ratio
Return Loss
(RL)

(SW)

Function

Selects SWR or return loss match measurement mode.

Remarks

Measurement results are returned in VSWR or dB.

29

Bird 4421 RF Power Meter

MiNimum Value
MaXimum Value

(MN)
(MX)

Function

Selects minimum or maximum measurement mode.

Remarks

Another measurement must be selected before selecting min or max.
Returns the minimum (or maximum) value of the previous measurement type.

Range

(Rxx)
Function
Remarks

Selects a measurement range listed in Figure 6-17.
If the selected range is outside the range of the connected power sensor, the
command is ignored.

Table 6-17

Measurement Ranges

Command
RYY
RNN
R17
R16
R15
R14
R13
R12
R11
R10
R09
R08
R07
R06
R05
R04
R03
R02
R01
R00

Terminators

Selects the characters that follow the end of a data string. Set x to:
z
z
z

Remarks

“T” for two terminators; a carriage return(CR) and a line feed(LF).
“O” for one terminator; a carriage return(CR).
“N” for no terminator.

Many controllers use the terminator sequence to recognize the end of an input
string. Using incorrect terminators can lock the bus.

(Px)

Function

Turns the prefix mode on or off. Set x to:
z
z

30

Turn auto range on
Turn auto range off. Keep present range
18.0 – 199.9 MW
1.80 – 19.99 MW
0.180 – 1.999 MW
18.0 – 199.9 kW
1.80 – 19.99 kW
0.180 – 1.999 kW
18.0 – 199.9 W
1.80 – 19.99 W
0.180 – 1.999 W
18.0 – 199.9 mW
1.80 – 19.99 mW
0.180 – 1.999 mW
18.0 – 199.9 µW
1.80 – 19.99 µW
0.180 – 1.999 µW
18.0 – 199.9 nW
1.80 – 19.99 nW
0.180 – 1.999 nW

(Yx)

Function

Prefixes

Power Range

“Y” to enable prefixes.
“N” to disable prefixes.

RS-232 Interface

Remarks

Prefixes are sent over the bus with the measurement, and indicate the status
of the current measurement (see Table 6-18 for examples):
z
z
z
z
z

“FC”,”FD”,”RC”,”RD”,”SW”,”RL”,”MN”,”MX” indicates the measurement type.
“U” indicates underflow; the value sent is “.000”.
“O” indicates overflow; the value sent is “199.9”.
“N” indicates normal; the value sent is a normal on-scale reading.
“4421” indicates the Bird model number.

Table 6-18

Prefix Examples

Data String
NFC.0.123W(CR)(LF)

Normal (N) forward carrier wave (FC), prefixes on

OFC 199.9W(CR)(LF)

Overflowed (O) forward carrier wave (FC), prefixes on

199.9W(CR)(LF)
URD

Triggers

Description

.000W(CR)(LF)

Overflowed forward carrier wave, prefixes off
Underflowed (U) reflected dBm (RD), prefixes on

(Tx)

Function

Selects the condition which will trigger a reading (see Table 6-19).

Remarks

Failure to trigger device before requesting a reading will lock the bus.
Fastest reading rate is 2.4 readings/second; slowest is 1 reading/sec.

Table 6-19

Trigger Conditions

Command

Status

Trigger Condition

T0

Continuous on ENT

T1

One shot on ENT

T3

One shot on TRG

T5

One shot on measurement command (FC, FD,
RC, RD, SW, RL, MN, MX)

(Ux)
Function

Reads a status word and returns the information as a string. Set x to:
z
z
z

Remarks

“0” for machine status.
“1” for error status.
“2” for revision history.

After sending the status command, a status word is sent the next time the
unit is addressed to talk. To ensure the correct status is transmitted, the status word should be requested as soon as possible after the command is sent.

31

Bird 4421 RF Power Meter

Machine Status Word (U0) - The format of the machine status word is shown
in Figure 13. The default values are also shown.
Figure 13
–

Machine Status Word Format
4

4

2

0

–

C

F

R

Y

Y

P

Y

Y

T

T

1

X

0

TERM

Header
Measurement Function
Range
Prefix
Terminator
Trigger
XON / XOFF

 Note: This command sets the unit to the settings that were last

issued remotely (through RS-232 or GPIB). Therefore, it will over ride
any manual changes selected at the front panel of the unit.

Error Status Word (U1) - The format of the error status word and the possible
error messages are shown in Figure 14. All flags will revert to their non-error
states after the U1 command is sent.
Figure 14

Error Status Word Format
–

4

4

0

2

–

C

V

M

V

C

0

F

TERM

L

Header
Command Status (shown valid)
Command Option Status (shown valid)
Self-test results (shown fail)

Status

Meaning

Description

ICM

Invalid Command

VCM
ICO

Valid Command
Invalid Command
Option

VCO
PS

Valid Command Option
Self-test Pass

FL

Self-test Fail

Set when an illegal device-dependent command
(IDDC) such as V2 is received. (V is illegal)
Set when no IDDC is received.
Set when an illegal device-dependent command
option (IDDCO) such as T6 is received. (6 is
illegal)
Set when no IDDCO is received.
Set when a self-test has been initiated by the J0
command and the test result is acceptable.
Set when the self-test has failed. (This is the
default condition.)

Revision History Word (U2)The format of the revision history word is shown
in Figure 15.
Figure 15
–

Revision History Word Format
4

4

2

0

–

X

X

X

X

X

X

Header
Writable stores
Software Rev level
Hardware Rev level
Interface standard
* YY is replaced with current software revision level.
** ZZ is replaced with current hardware revision level.

32

Y

*

Y

Z

**

Z

RS-232 TERM

RS-232 Interface

Self Test

(J0)

Function

Initiates a hardware and software test.

Remarks

Results are stored in the U1 status word (see “Status” on page 31).
“J0” must be sent each time before reading the result.

Writable Store

(Wxxxxxx)

Function

Storage for six bytes of ASCII data.

Remarks

Data stored is lost when the 4421 is turned off.
Data is sent back as part of the U2 status word.

33

Chapter 6

Maintenance
This chapter describes routine maintenance, along with troubleshooting
instructions for the power meter and power sensor. Disassembly instructions
for the Bird Cal Cart are also provided. For service beyond this level, return
the unit to a qualified service center.
WARNING
To avoid personal injury, disconnect the power cord from the AC line before
performing any maintenance, including fuse replacement or changing the
line voltage setting.
WARNING
Never attempt to connect or disconnect RF equipment from the transmission
line while RF power is being applied.
Leaking RF energy is a potential health hazard.
WARNING
The Bird 4421 contains no user-serviceable parts. Do not remove its cover.
The Bird 4421 Power Meter requires only simple, routine maintenance.
z
z
z

Wipe off dust and dirt regularly.
Check the connectors and cables for damage.
Clean the connector contacts with alcohol or dry cleaning solvent.

Cleaning
CAUTION
Do not use harsh or abrasive detergents for cleaning.
Clean the Bird 4421 Power Meter and its display with a soft cloth dampened
with mild detergent and water only. Clean sensors with a dry cleaning solvent
that leaves no residue.

Troubleshooting
Since the power meter and power sensor can only work together, the first step
is to determine which is malfunctioning. Connect the power sensor to the
meter and perform the functional test on page 37. If the power meter is malfunctioning, refer to the troubleshooting table below. If the power sensor is
malfunctioning, return it for service.
CAUTION
Due to the complexity of the Bird Power Sensor, field repairs beyond general
maintenance should not be attempted.
Removal or disturbance of the power sensor cover can result in cancellation
of lifetime warranty.
This manual cannot list all malfunctions that may occur, or corrective actions.
If a malfunction is not listed or is not corrected by the listed corrective actions,
contact a qualified service center.

 Note: For power sensors in Cal Carts, following the test procedure
will check the sensor not covered by the safety cover. To test the other
sensor slide the safety cover over, then proceed as normal.

35

Bird 4421 RF Power Meter

PROBLEM
Power meter has
no power

POSSIBLE CAUSE
Have the batteries been
charged?

Recharge the batteries.

Is the power meter’s ac
power cord connected?

Connect the power cord.

Is the ON/OFF rocker switch
on the rear panel set to
OFF?

Set the switch to ON.

(Cal Cart only)
Is the Cal Cart’s AC power
cord connected to the AC
line?

Connect AC power.

Blown fuse?

Check fuse rating and replace fuse
(See “Replacing Fuses” on
page 39).

Is the sensor cable
connected to both the power
meter and power sensor?

Connect sensor cable.

Is the sensor cable
defective?

Replace sensor cable.

Display blank or
not updating

Have the batteries been
charged?

NO: Recharge battery
YES: Return meter for service.

Power meter turns
off while on battery
power

Is “LO BAT” displayed?

YES: Recharge battery
NO: Return meter for service.

Push buttons do
not respond

Test the push buttons (See
“Push Button Test” on
page 37). Are they
defective?

Return meter for service.

Dash moves
across the display

Every segment on
the display is lit
IEEE-488
Interface does not
respond to the
interface link

36

CORRECTION

Return meter for service.
Are the address in the
interface program and the
address setting on the DIP
switches on the interface
module the same?

Change the program or DIP switch
setting so that the addresses are
the same (See “Dip Switch” on
page 12).

Send the “J0” self test
command, then check the
status. Does the power
meter fail the self test?

Replace I/O hub PCB.

Is the IEEE cable defective?

Replace IEEE interface cable.

RS-232 Interface
does not respond
to the interface
link. Fails J0 selftest command.

Are the DIP switches set
correctly?

Set DIP switches.

Is the RS-232 cable
defective?

Replace RS-232 interface cable.

With the RS-232,
the power meter
displays “TALK”
but does not
operate as
expected.

Is DIP switch 2 set correctly?

For “talk/listen” operation, set this
switch to ON. For “talk always” set
this switch to OFF.

Is DIP switch 1 set correctly?

Set DIP switch 1 as indicated in
Figure 8.

Maintenance

Functional Test
1. Turn the power meter off. The ON/OFF switch on the rear panel should be
OFF and the ac power cable should be connected.
2. Turn ON the ON/OFF switch on the rear panel of the meter.
3. While holding down the FWD and SWR push buttons, press the ON/OFF
button on the front panel of the power meter. Immediately release all three.

 Note: The unit tests the display on power up cycling through and

activating each segment two times and then activating all segments at
once two times.

4. The power meter’s model number and revision date should scroll across the
display. If a dash “–” is displayed instead, then the meter is malfunctioning.
5. The power sensor’s model number and revision date should scroll across
the display. If a dash is displayed after the power meter data, then the
power sensor is malfunctioning.
Figure 16

Test Display, No Malfunction
– . . 4 4 2 1 . . – 0 7 . 0 1 . 0 2 – . . 4 0 2 1 . . – 0 7 . 0 1 . 0 2 –
Meter’s
Model
Number

Meter’s
Software
Rev. Date

Sensor’s
Model
Number

Sensor’s
Software
Rev. Date

Push Button Test

 Note: If a push button is malfunctioning, return the power meter.
1. Disconnect the power sensor.
2. Turn the power meter ON.

 Note: After the power up display disappears, a dash “–” should
scroll across the display.

3. Turn the power meter OFF.
4. Connect the power senson.
5. Turn the power meter ON.

 Note: “AUTO” and “FWD” should be displayed, and a reading of
“0.00 W” (or a very low number).

6. Press RFL.

 Note: “FWD” should change to “RFL” on the display. The reading
should remain the same.

7. Press SWR.

 Note: “RFL” should change to “SWR”. “0.00 W” should change to
“

” (underrange error).

8. Hold down MIN.

 Note: “

” should change to “

” (overrange error).

” should change to “

”.

9. Release MIN.

 Note: “
10. Press dBm.

 Note: “SWR” should turn OFF and “dBm” should turn ON..
37

Bird 4421 RF Power Meter

11. Press dBm.

 Note: “dBm” should change to “SWR”.
12. Press FWD.

 Note: “SWR” should change to “FWD” and “ ” to “.0000 W” (or a
very low number).

13. Hold down MAX.

 Note: “.0000” should change to “1.0000”.
14. Release MAX.

 Note: “1.0000” should change to “.0000” (or a very low number).
15. Press LIGHT.

 Note: The back-light should turn on.
16. Press LIGHT.

 Note: The back-light should turn off.
17. Press s (up).

 Note: The power meter should change ranges each time it is
pressed until it reaches “0.000 KW”.

18. Press t (down).

 Note: The power meter should change ranges each time it is
pressed until it reaches “0.00 W”.

19. Turn the power meter OFF.

Repair
WARNING
To avoid personal injury, disconnect the power cord from the AC line before
performing any maintenance, including fuse replacement or changing the
line voltage setting.
Front Panel (Cal Cart only)

 Note: If the power meter is mounted in a Cal Cart, to access the

power sensor or the ac connections, it will be necessary to remove the
front panel of the Cal Cart.

1. Remove both screws on the front panel, between the RF connector and the
handle.
2. Pull on the handle set into the base of the front panel to remove it.

 Note: When making AC connections, refer to the following wiring
diagram.

Figure 17

38

Cal Cart Wiring Schematic

Maintenance

Cord Reel (Cal Cart only)
1. Disconnect the cord reel from the terminal strip.
2. Unscrew the AC connector.
3. Unscrew and remove the cord reel.
4. Screw the replacement cord reel into place.
5. Thread the AC connector through the grommet, and connect the wires on
the other end to the terminal strip (See Figure 18).
Replacing Fuses
WARNING
To avoid personal injury, disconnect the power cord from the ac line before
performing any maintenance, including fuse replacement.
CAUTION
Failure to install the properly rated fuse may result in equipment damage or
nuisance failures.
1. Gently pry the fuse drawer out of the ac module. The fuse holder does not
detach from the AC module.
2. Install the replacement fuses then close and secure the fuse holder. Fuse 1
is a spare fuse and fuse 2 is the active fuse.
AC Line Voltage

Fuse Rating

115 Vac

T630 mA, 5x20 mm Time Lag Fuse

230 Vac

T315 mA, 5x20 mm Time Lag Fuse

Figure 18

AC Line Fuse

AC Module
Fuse
Drawer
Fuse 1

Fuse 2

39

Bird 4421 RF Power Meter

Replacing Batteries
WARNING
Exposed AC line voltage (115 VAC or 230 VAC). Disconnect the power cord
from the AC line before replacing the batteries. Failure to comply may
result in severe electrical shock or death.
WARNING
Use only Nickel Metal Hydride (NiMH) batteries that have a minimum
capacity of 4500 milliamper hours (mAh). Do not install batteries that are
not Nickel Metal Hydride. Do not install NiMH batteries that have less than
4500 mAh capacity. Failure to comply may result in damage to the batteries,
damage to the instrument, and injury to personnel from battery chemicals.
1. Disconnect the AC power cord from the ac mains supply.
2. Remove the four screws that secure the top cover.
3. Lift the top cover to access the attached ground wire then disconnect the
ground wire and remove the cover.
4. Unfasten the battery retaining belts then remove the battery tubes
(Figure 19). Be sure to note the polarity and orientation of the battery
tubes before removing them.
5. Remove the batteries from each tube and insert replacement batteries. Be
sure to note the polarity (positive and negative) arrangement of the batteries.
6. Install the battery tubes into the unit and secure with the retaining belts.
Be sure to position the retaining belts as they were before removal and
tighten them securely.
7. Connect the ground wire to the top cover.
8. Install the top cover and secure it with the four screws removed earlier.
Figure 19

Batteries

X4

X4

Item
1
2

Description
Battery retaining belt
Battery tube (batteries inside)

Long Term Storage
Do not store the instrument for long periods of time without recharging the batteries. When the instrument is stored for long periods of time without use, the
batteries will loose their charge and also loose the ability to reach full charge
when put into service. To restore the battery charge, perform a full charge for 28
hours. If the batteries are not fully charged after 28 hours, completely discharge
the batteries then perform a full charge again. If necessary, repeat this discharge and charge cycle up to three times. If the batteries do not remain
charged after three discharge-charge cycles, replace the batteries.
40

Maintenance

Power Sensor (Cal Cart only)
WARNING
To avoid personal injury, disconnect the power cord from the AC line before
performing any maintenance, including fuse replacement or changing the
line voltage setting.
WARNING
Never attempt to connect or disconnect RF equipment from the transmission line
while RF power is being applied. Leaking RF energy is a potential health hazard.
CAUTION
The Bird 4421 must be powered off when connecting or disconnecting the
power sensor from the power meter.
CAUTION
Changing the sensor’s connectors will invalidate calibration data, and may
reduce the maximum power rating of the unit.
CAUTION
Due to the complexity of the Bird Power Sensor, field repairs beyond general
maintenance should not be attempted. Removal or disturbance of the power
sensor cover can result in cancellation of lifetime warranty.
1.
2.
3.
4.
5.
6.
7.

Disconnect the RF line from the Cal Cart.
Remove the Cal Cart front panel (see "Front Panel (Cal Cart only)" on page 38).
Disconnect the sensor cable from the power sensor.
Disconnect the RF cable from the sensor output port.
Remove the screws on the sensor mounting bracket.
Remove the sensor from the mounting bracket.
Put the new sensor in the bracket and screw it into place.

 Note: Make sure the arrow on the side of the sensor points towards the load,
and that the end labeled “SOURCE” points towards the front of the Cal Cart.

8. Connect the RF cable to the sensor end labeled “LOAD”. Connect the sensor cable.
9. Replace the Cal Cart front panel.
Load
WARNING
Heavy load. Do not attempt to lift unaided.
1.
2.
3.
4.
5.

Disconnect all RF cables from the RF connectors on the load.
Unplug the blower assembly on blower-equipped loads.
Replace the vent plugs with the shipping plugs.
Remove the mounting bolts connecting the load to the Cal Cart.
Remove the load.

1.
2.
3.
4.

Remove the load (See "Load" on page 41).
Flip the cart over.
Unscrew the defective caster.
Screw the new caster into place and replace the load.

Casters

Handle
1. Unscrew the hadle
2. Remove the old handle.
3. Screw the new handle into place.
41

Bird 4421 RF Power Meter

Storage and Shipment (Cal Cart only)
Cover the unit before storing to keep out dust and dirt. It is not necessary to
install the shipping plug. Store in a dry, dust-free environment where the
ambient temperature will remain between –20 and +50 °C (–4 to +122 °F).
Before shipping the Bird Cal Cart, take the following precautions:
z

Remove the vent plugs and replace them with the shipping plugs.
Wrap the vent plugs with padding and tape them to the side of the
load for protection.

 Note: With the shipping plugs installed, it is not necessary to
empty out the coolant.

z

Repack in the original carton or contact Bird for a transit case.

Specifications, Cal Cart
To determine the model numbers of the Cal Cart components, refer to "About This
Manual" on page x. For load specifications, refer to the load manual included.

 Note: Certain sensors or connectors may be incompatible with

some models. Refer to "About This Manual" on page x for a list of available components, or contact Customer Service.

Max. Power
Under normal operating conditions, the Cal Cart’s maximum power is the
MINIMUM of the max. power of the sensor and the max. power of the load.
The following modifiers apply:
z Above 65 MHz, derate the load’s max. power. See "High Frequency
Derating 1" on page 47.
z Due to the wide variety of available connectors, frequency range and
maximum power may be reduced. Insertion loss is specified with female
N connectors. Choose connectors appropriate for the frequency and power
of operation.
z Derate RF power by 2.5% for every 305m (1,000 ft.) above 1,520m (5,000 ft.).

Table 7-20

Frequency Range

Sensor dependent, 30 MHz max.

Power Range

Dependent on sensor, sensor connectors, and load

Connectors

Customer specified (See "Available Connectors" on page 49)

Dimensions
Short Cart
Long Cart

42”L x 20”W x 42”H (1067 x 508 x 1067 mm)
52”L x 20”W x 42”H (1321 x 508 x 1067 mm)

Weight, Nominal
Short Cart
Long Cart

175 lbs. (80 kg)
290 lbs. (114 kg)

Temperature Range
Operating
Storage

0 to +35 °C (32 to +95 °F)
–20 to +70 °C (–4 °F to +158 °F)

Altitude

1520 m (5000 ft.)

Humidity

85% noncondensing max

Calibration Cycle,
Nominal

1 year

CE

42

Bird Cal Cart

CE Compliant. Refer to Declaration of Conformity for specific
standards.

Maintenance

Specifications, 4421 Power Meter
Specifications, Power Sensors

Table 7-21

Bird 4421 RF Power Meter

Frequency Range

Sensor dependent

Power Range

Sensor dependent

VSWR Display

1.0 – 199.9 max

Return Loss Display

0 to 40 dB max

Display Accuracy

± 1 on least significant digit

AC Power

115/230 Vac @ 50/60 Hz; 60 W

Batteries

8 Nickel Metal Hydride 1.2 v cells, rechargeable

Battery Life

Approximately 8 hours continuous usage
(reduced after high temperature charging)

Battery Charger

Built-in battery charger. Drained batteries require
approximately 28 hours to recharge.

Display

LCD, 4½ digit display. Indicates mode, measurement units,
battery condition, remote status, and signal increase/
decrease. Self contained backlight.

Optional Interfaces

IEEE-488 GPIB
RS-232

Fuse Rating

IEC (5 x 20 mm) Time Lag Type T

115 Vac
230 Vac

T630 mA
T315 mA

CE

CE Compliant. Refer to Declaration of Conformity for
specific standards.

Humidity

95% max. (noncondensing)

Altitude

Up to 10,000 feet (3,000 m)

Temperature Range
Operating
Storage

0 to 50 °C (32 to 122 °F)
–20 to +50 °C (–4 to +122 °F)

Dimensions

15.5”L x 12.25”W x 4.25”H (393 x 311 x 108 mm)

Weight, Nominal

9.5 lbs. (4.3 kg)

CAUTION
Changing the sensor’s connectors will invalidate calibration data, and may
reduce the maximum power rating of the unit.

43

Bird 4421 RF Power Meter

Table 7-22

Specifications Common to all Sensors

Impedance, Nominal

50 ohms

Max. Allowable Terminating
VSWR

2.00:1

Calibration Technique

Frequency-specific calibration factors stored in
nonvolatile memory in each sensor.
Sensor output corrected for frequency and
temperature within specified ranges.

Calibration Cycle, Nominal

4027A and 4027F, 6 months
All other models, 1 year

Accuracy, Reflected

Calculated from FWD accuracy and FWD power
FWD Power
RFL Accuracy = FWD Accuracy + -------------------------------10 Directivity / 10

Accuracy, VSWR

Calculated from FWD and RFL power
P R⎞ ⎛
P R⎞
⎛
VSWR = ⎜ 1 + ------⎟ ⁄ ⎜ 1 – ------⎟
P F⎠ ⎝
P F⎠
⎝

Sampling Rate, Nominal

2 readings/second

Operating Power

Supplied by power meter via sensor cable

Connectors
4028B10M
4028C10M
4028A Series
All other models
CE

CE Compliant. Refer to Declaration of Conformity
for specific standards.

Humidity, Max.

95% (noncondensing)

Altitude, Max.

10,000 feet (3,000 m)

Temperature Range
Operating
Storage

0 to 50 °C (32 to 122 °F)
–20 to +70 °C (–4 to +158 °F)

Dimensions
4028B3M, 4028B10M
4028A Series
All other models

6.75”L x 3.5”W x 4.75”H (175 x 89 x 121 mm)
4.7”L x 3.2”W x 3.8”H (120 x 82 x 97mm)
5.2”L x 2.5”W x 3.25”H (137 x 64 x 83 mm)

Weight, Nominal
4028B3M, 4028B10M,
4028C10M

5 lb. 2 oz. (2.33 kg)

4028A2M, 3M, 4M,
and 25M

3 lb. 5 oz. (1.5 kg)

All other models

44

1-5/8” EIA Flanged
3-1/8” EIA Flanged
7-16 DIN, LC, HN, or 7/8”
Customer specified from QC list, appropriate for
frequency and power.

10M,
1 lb. 13 oz. (0.8 kg)

Maintenance

Table 7-23

Bird 4020 Series RF Power Sensors

RF Power Range
4021, 4022
4024, 4025
Frequency Range
4021
4022
4024
4025
Accuracy, Fwd, Best Case*
VSWR, Max.
4021, 4024, 4025
4022
Insertion Loss, Max.
4021, 4024, 4025
4022
Directivity, Min.
4021, 4022
4024
4025
*

300 mW – 1 kW
3 W – 10 kW
1.8 – 32 MHz
25 MHz – 1 GHz
1.5 – 32 MHz
100 kHz – 2.5 MHz
± 3% (1σ)
1.05:1
1.05:1, 25 – 512 MHz and 1.10:1, 512 MHz – 1 GHz
0.05 dB
0.05 dB, 25 – 512 MHz and 0.13 dB, 512 MHz – 1 GHz
30 dB
28 dB, 1.5 – 2.5 and 25 – 32 MHz and 30 dB, 2.5 – 25 MHz
28 dB, 100 – 125 kHz and 30 dB, 125 – 2500 kHz

For rated accuracy, no more than 1% AM; Harmonics –50 dBc or less
Derate accuracy by 3.0% (1σ) below 15 °C and above 35 °C

CAUTION
Changing the sensor’s connectors will invalidate calibration data, and may
reduce the maximum power rating of the unit.

Table 7-24

Bird 4027A Series RF Power Sensors

Frequency Range
4027A250K
250 – 400 kHz
4027A400K
400 – 550 kHz
4027A800K
800 – 950 kHz
4027A2M
1.5 – 2.5 MHz
4027A4M
3 – 5 MHz
4027A10M
10 – 15 MHz
RF Power Range
4027A12M
300 mW – 1 kW
4027A25M
3 W – 9 kW
4027A35M
3 W – 7.5 kW
4027A60M
3 W – 6 kW
Accuracy, Fwd, Best Case*
Calibration Frequencies, Typical (MHz)†
4027A250K
0.25, 0.40
4027A400K
0.40
4027A800K
0.90
4027A2M
1.8, 2.0, 2.17
4027A4M
4.0, 5.0
4027A10M
10.0, 13.56, 15.0
Calibration Power, Typical
4027A12M
All other models
VSWR, Max.
Insertion Loss, Max.
Directivity, Min.
4027A12M
All other models
*
†

4027A12M
4027A25M
4027A35M
4027A60M
4027A100M
4027A150M

10 – 15 MHz
25 – 30 MHz
35 – 45 MHz
45 – 65 MHz
95 – 105 MHz
150 – 170 MHz

4027A100M
4027A150M
All other models

3 W – 4 kW
3.75 W – 3.75 kW
3 W – 10 kW

± 1.0% (1σ)
4027A12M
4027A25M
4027A35M
4027A60M
4027A100M
4027A150M

10.0, 13.56, 15.0
25.76, 27.12, 28.48
40.68
55.0, 60.0
95.0, 100.0
162.0

700 W
1.7 kW
1.05:1
0.05 dB (with female “N” connectors)
30 dB
28 dB

For rated accuracy, no more than 1% AM; Harmonics –50 dBc or less
Derate accuracy by 1% (1σ) outside cal. power or cal. frequency
Derate accuracy by 1% (1σ) below 15 °C and above 35 °C
Other calibration frequencies available upon request

45

Bird 4421 RF Power Meter

Table 7-25

Bird 4027F Series RF Power Sensors

Frequency Range
4027F2M
4027F10M
4027F60M
RF Power Range
4027F2M, 4027F10M
4027F60M
Accuracy, Fwd, Best Case
Calibration Frequencies, Typical*
4027F2M
4027F10M
4027F60M
Calibration Power, Typical
Harmonic Rejection, Min.
4027F2M
4027F10M
4027F60M
Low Frequency Rejection, Min.
4027F10M
4027F60M
Max Error Induced by 10% AM
4027F2M, 4027F10M
4027F60M
VSWR, Max.
Insertion Loss, Max.
Directivity, Min.
*

1.8 – 2.2 MHz
12 – 15 MHz
57 – 63 MHz
0.1 – 10 kW
0.1 – 3 kW
± 1.0% (2σ)
1.8, 2.0, 2.17 MHz
12.0, 12.5, 13.56, 14.0, 15.0 MHz
57.0, 58.5, 60.0, 61.5, 63.0 MHz
1.7 kW
26 dB @ 3.6 – 3.8 MHz, 30 dB @ > 3.8 MHz
30 dB @ > 25 MHz
30 dB @ > 114 MHz
30 dB @ < 1 MHz
30 dB @ < 15 MHz
0.2% @ < 5 kW, 1.0% @ 5 – 10 kW
0.2% @ < 1.5 kW, 1.0% @ 1.5 – 3 kW
1.05:1
0.05 dB (with female “N” connectors)
28 dB

Other calibration frequencies available upon request

Table 7-26 Bird 4028A and 4028B Series RF Power Sensors
Frequency Range
4028A250K
4028A400K
4028A2M
4028A3M
4028A4M
4028A10M, 4028B10M,
4028C10M
4028A25M
4028B3M
RF Power Range
4028A250K, 4028A400K
4028C10M
All other models
Accuracy, Fwd, Best Case*
Calibration Frequencies, Typical
(MHz)†
4028A250K
4028A400K
4028A2M
4028A3M
4028A4M
4028A10M, 4028B10M,
4028C10M
4028A25M
4028B3M
Calibration Power, Typical
VSWR, Max.
Insertion Loss, Max.
Directivity, Min.
*
†

46

250 – 400 kHz
400 – 550 kHz
1.5 – 2.5 MHz
2.5 – 3.5 MHz
3.5 – 4.5 MHz
10 – 15 MHz
25 – 30 MHz
2.5 – 4 MHz
1 kW – 20 kW
500W – 50 kW
1 kW – 25 kW
± 2.0% (2σ)
0.25, 0.40
0.40
1.8, 2.0, 2.17
2.5, 3.2, 3.5
3.5, 4.0
10.0, 13.56, 15.0
25.76, 27.12, 28.48
2.5, 3.2, 3.5, 4.0
3.5 kW
1.05:1
0.05 dB (with female 7-16 DIN connectors)
28 dB

For rated accuracy, no more than 1% AM; Harmonics –50 dBc or less
Derate accuracy by 2% (2σ) outside cal. power or cal. frequency
Derate accuracy by 2% (2σ) below 15 °C and above 35 °C
Other calibration frequencies available upon request

Maintenance

Specifications, RS-232 Interface
Table 7-27

RS-232 Interface Module

Logic Levels
Modes of Operation
Talk Always
Addressable
Connector
Humidity, Max
Altitude, Max
Temperature Range
Operating
Storage
Dimensions
Weight

Meets all EIA Standard RS-232-C specifications
Switch and bus selectable
Allows the 4421 to send to the bus keyboard-initiated
measurements only
Allows the 4421 to be commanded by an RS-232 interface controller
RS-232 Interface Type
95% (non-condensing)
10,000 feet (3,000 m)
0 to 50º C (32 to 122º F)
–20 to +50º C (–4 to +122º F)
6.5”L x 4.5” W (165 x 115 mm)
0.5 lbs (0.23 kg) nominal

Specifications, IEEE-488 GPIB Interface
Table 7-28 IEEE-488 Interface
Logic Levels
Modes of Operation
Talk Only
Addressable
Connector
Humidity, Max
Altitude, Max
Temperature Range
Operating
Storage
Dimensions
Weight, Nominal

Figure 20

Meets all IEEE Standard 488-1978 specifications
Switch and bus selectable
Allows the 4421 to send to the bus keyboard-initiated measurements
only
Allows the 4422 to be addressed as talker or listener under the
command of an IEEE-488 bus controller
Standard IEEE-488 bus type
95% (non-condensing)
10,000 feet (3,000 m)
0 to 50º C (32 to 122º F)
–20 to +50º C (–4 to +122º F)
6.5”L x 4.5” W (165 x 115 mm)
0.5 lbs (0.23 kg)

High Frequency Derating 1
Load Derating Above 65 MHz
5 & 10 kW Loads

10
9
8

Max. Power (kW)

7
6
5
4
3
2
1
0
10

100

1000

Frequency (MHz)

47

Bird 4421 RF Power Meter

Figure 21

High Frequency Derating 2
Load Derating Above 65 MHz
1 & 2.5 kW Loads

2.5

Max. Power (kW)

2.0

1.5

1.0

0.5

0
10

100

1000

Frequency (MHz)

Table 7-29

Replacement Parts
Description

48

Qty

Part Number

Fuse, IEC (5 x 20 mm) Time Lag Type T
115 Vac, T630 mA
230 Vac, T315 mA

1

Cord, AC Power
115 Vac
230 Vac Harmonized

1

Plug, 115 Vac

1

5A2626

Cable, Sensor

1

4421-038

Grommet

1

4421A372

Cordreel

1

4421A383

Casters

4

4421A384

Handle

1

4421A385

IEEE-488 Bus Interface Module

1

4421-489-2

Cable, IEEE-488
2m
1m

1

RS-232 Bus Interface Module

1

Cable, RS-232
10 ft.
5 ft.

1

Null Modem Kit (RS-232 only)

1

4380-250

Panel Mount Kit

1

4421-250

Battery, C size, NiMH

8

5A1230

5A2257-14
5A2257-11
5-1286
5A2416

5-1317-2
5-1317-1
4421-233-1
5-1662-2
5-1662-1

Maintenance

Available Connectors
CAUTION
Changing the sensor’s connectors will invalidate calibration data, and may
reduce the maximum power rating of the unit.

Connector

P/N

HN (F)

4240-268

LC (F)

4240-031

BNC-Female

4240-125

BNC-Male

4240-132

C-Female

4240-100

C-Male

4240-110

HN-Female

4240-268

HN-Male

4240-278

LC-Female

4240-031

LC-Male

4240-025

Open Term. # 10-32 Nut

4240-080

N (F)

4240-062

7/16 Jack, IEC Type 169-4

4240-344

LT-Female

4240-018

LT-Male

4240-012

N-Female

4240-062

N-Male

4240-063

SC-Female

4240-090

SMA-Female

4240-336

SMA-Male

4240-334

7/16 Jack, IEC Type 169-4

4240-344

7/16 Plug, IEC Type 169-4

4240-363

TRU 6934 (F)

4240-371

TRU 7958 (F)

4240-372

Mini UHF-Female

4240-346

UHF-Female

4240-050

UHF-Male

4240-179

1-5/8” EIA Fixed

4240-096

1-5/8” EIA Swivel

4240-208

7/8” EIA

4240-002

TNC-Female

4240-156

TNC-Male

4240-160

49

Bird 4421 RF Power Meter

Customer Service
If you need to return the unit for any reason, contact the Bird Service Center
for a return authorization. All instruments returned must be shipped prepaid
and to the attention of Bird Service Center.
Bird Service Center
30303 Aurora Road
Cleveland (Solon), Ohio 44139-2794
Phone: (440) 519-2298
Fax: (440) 519-2326
E-mail: bsc@bird-technologies.com
For the location of the sales office nearest you, give us a call or visit our Web
site at:
http://www.bird-electronic.com

50

Limited Warranty
All products manufactured by Seller are warranted to be free from defects in material and workmanship for a period of one (1) year, unless otherwise specified, from date of shipment and to conform to
applicable specifications, drawings, blueprints and/or samples. Seller’s sole obligation under these
warranties shall be to issue credit, repair or replace any item or part thereof which is proved to be
other than as warranted; no allowance shall be made for any labor charges of Buyer for replacement
of parts, adjustment or repairs, or any other work, unless such charges are authorized in advance by
Seller.
If Seller’s products are claimed to be defective in material or workmanship or not to conform to specifications, drawings, blueprints and/or samples, Seller shall, upon prompt notice thereof, either examine the products where they are located or issue shipping instructions for return to Seller
(transportation-charges prepaid by Buyer). In the event any of our products are proved to be other
than as warranted, transportation costs (cheapest way) to and from Seller’s plant, will be borne by
Seller and reimbursement or credit will be made for amounts so expended by Buyer. Every such claim
for breach of these warranties shall be deemed to be waived by Buyer unless made in writing within
ten (10) days from the date of discovery of the defect.
The above warranties shall not extend to any products or parts thereof which have been subjected to
any misuse or neglect, damaged by accident, rendered defective by reason of improper installation or
by the performance of repairs or alterations outside of our plant, and shall not apply to any goods or
parts thereof furnished by Buyer or acquired from others at Buyer’s request and/or to Buyer’s specifications. Routine (regularly required) calibration is not covered under this limited warranty. In addition, Seller’s warranties do not extend to the failure of tubes, transistors, fuses and batteries, or to
other equipment and parts manufactured by others except to the extent of the original manufacturer’s
warranty to Seller.
The obligations under the foregoing warranties are limited to the precise terms thereof. These warranties provide exclusive remedies, expressly in lieu of all other remedies including claims for special
or consequential damages. SELLER NEITHER MAKES NOR ASSUMES ANY OTHER WARRANTY
WHATSOEVER, WHETHER EXPRESS, STATUTORY, OR IMPLIED, INCLUDING WARRANTIES
OF MERCHANTABILITY AND FITNESS, AND NO PERSON IS AUTHORIZED TO ASSUME FOR
SELLER ANY OBLIGATION OR LIABILITY NOT STRICTLY IN ACCORDANCE WITH THE
FOREGOING.

Special Lifetime Warranty - Series 4020, Series 4027A, Series 4027F, and Series
4028 Power Sensor Head
In addition to its standard warranty, the Bird Electronic Corporation warrants its Series 4020, Series
4027A, Series 4027F, and Series 4028 Thruline® Power Sensor Heads for lifetime to original purchaser. This extended warranty is against burnout. For the warranty to apply, the Sensor Head must
be used with the correct Bird Electronic Corporation Display Unit, the maximum power rating of the
Sensor must not be exceeded, the Sensor RF circuit must be properly terminated and the Sensor not
subjected to physical abuse.
Bird Electronic Corporation, at its option, will repair or replace the defective Sensor at its world
Headquarters at 30303 Aurora Road, Solon, Ohio 44139.
The customer is responsible to pay transportation charges to return the defective sensor to Bird.

51



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : No
Page Count                      : 59
Page Mode                       : UseOutlines
Has XFA                         : No
XMP Toolkit                     : XMP toolkit 2.9.1-13, framework 1.6
About                           : uuid:4a2eaa52-f7e1-4bed-bbc2-8080c638bf58
Producer                        : Acrobat Distiller 8.1.0 (Windows)
Keywords                        : 4421 4020 4027 thruline sensor meter precision 4021 4022 4024 4025 filter 4028 high power
Create Date                     : 2011:11:03 13:40:26Z
Creator Tool                    : FrameMaker 8.0
Modify Date                     : 2014:08:27 16:15:19+05:30
Metadata Date                   : 2014:08:27 16:15:19+05:30
Document ID                     : uuid:ac9e9a79-b178-438f-b927-58d64a174edb
Instance ID                     : uuid:e74bbbf0-9cf5-47b8-a33b-1bf30ea5f810
Format                          : application/pdf
Title                           : Bird 4421 Thruline RF Power Meter
Creator                         : Bird Electronic Corporation
Description                     : 8/4/04
Author                          : Bird Electronic Corporation
Subject                         : 8/4/04
EXIF Metadata provided by EXIF.tools

Navigation menu