PCG 10000 AF™ Pneumatic Gauge Calibrator Fluke PGC AF DAta Sheet Brochure

User Manual: Fluke-PGC-10000-AF-Pneumatic-Gauge-Calibrator-DAta-sheet-Brochure

Open the PDF directly: View PDF PDF.
Page Count: 2

Technical Data
PGC-10000-AF
Pneumatic Gauge
Calibrator
PGC-10000-AF is a special configuration
of DHI’s GPC1 Gas Pressure Controller
and RPM4 A70M/A20M Reference Pres-
sure Monitor (see the GPC1 and RPM4
brochures for details). The PGC-10000-AF
was designed to optimize the testing and
calibration of analog and digital pres-
sure gauges and indicators in ranges from
less than 500 psi (3.5 MPa) to 10 000 psi
(70 MPa). This special configuration was
selected by the United States Air Force
as the next generation high pressure
pneumatic gauge calibration system for
deployment in its Precision Measurement
Equipment Laboratories (PMELs) through-
out the world.
With GPC1 and associated GB-152 Gas
Booster, the operator effortlessly gener-
ates and adjusts pressure to set the device
under test to its cardinal point. The RPM4
then precisely measures the actual pres-
sure applied.
Interactive RPM4 embedded software
auto-matically sets up the calibration
procedure based on entry of the DUT
full scale and tolerance. It then steps the
operator through the test, logging data and
providing real time notification of in or
out-of-tolerance conditions at each point.
A foot pedal “ENTER” switch allows the
operator to trigger readings hands free.
Features
Calibrate any range device in gas from less
than 500 to 10 000 psi (70 MPa) FS with one
compact, integrated system
Effortless operation; no pumping, no weight
lifting
Includes gas booster to boost a bottled air
supply as low as 400 psi (2.75 MPa) to
10 000 psi (70 MPa)
Easily set pressure to DUT cardinal point and
read back actual pressure from the reference
On-board calibration routines with real time
out-of-tolerance notification and data logging
Foot pedal “ENTER” key for hands free sequence
execution
• Completeatypicalhighpressuregaugerunin
about 20 minutes (20 % ascending increments)
Supports 12 units of measure and custom units
Push button switching between gauge and
absolute measurement modes
RS-232 and IEEE-488 interfaces included; com-
patible with COMPASS® for Pressure software
Delivered with connectors and adaptors for
NPT, AN4 and gland and collar DUTs
Includes molded, reusable transit case for
shipping RPM4/HPMS transfer standard for
recalibration
2 Fluke Calibration PCG-10000-AF™ Pneumatic Gauge Calibrator
Specications
General
Power requirements Electrical: 85 V to 264 V ac, 50/60 Hz, 25 VA max consumption and
12 V dc, 1.2A
Pneumatic: 75 psig (500 kPa) shop air at 15 scfm to 75 scfm
(425 slm to 2 125 slm) and 100 psig (700 kPa) clean, dry gas
Test gas supply 400 psi to 3 000 psi (3 MPa to 20 MPa) Maximum high pressure output is
high pressure supply times 25
Operating temperature 18 to 28 °C (64 °F to 82 °F)
Weight RPM4/HPMS: 11.0 kg (22 lb)
GPC1: 24.0 kg (53 lb)
GB-152: 16.4 kg (36 lb)
Total: 51.4 kg (113 lb)
Benchtop system footprint
(GB-152 Gas Booster
mounted under bench) (W x D)
75 cm x 60 cm (30 in x 24 in)
Overall pressure range Gauge mode: 0 psig to 10 000 psig
Absolute mode: atm to 10 000 psia (70 MPa)
Standard test fluid Any non-corrosive gas
Test connection DH500 (gland and collar type for coned and left hand threaded tube
equivalent to AE F250C, HIP HF4, etc.), adaptors to 1/8 in NPT M, 1/4 in
NPT M, 1/4 in NPT F, AN4 M provided
Pressure ranges (all ranges are
gauge and absolute)
Hi Q-RPT: 10 000 psi (70 MPa)
Lo Q-RPT: 3 000 psi (20 MPa)
Pressure measurement
Precision1Hi Q-RPT1: ± 0.0175 % of reading or 0.44 psi (3 kPa), whichever is larger
Lo Q-RPT ± 0.0175 % of reading or 0.13 psi (0.9 kPa), whichever is larger
Predicted stability2± 0.004 % of reading
Measurement uncertainty3Hi Q-RPT: ± 0.02 % of reading or 0.5 psi (3.5 kPa), whichever is larger
Lo Q-RPT: ± 0.02 % of reading or 0.15 psi (1 kPa), whichever is larger
1 Precision: Combined linearity, hysteresis, repeatability.
2 Stability: Change in zero and span, with use of AutoZ feature, over 6 months period for typical RPT used under
typical conditions. As stability can only be predicted and varies from Q-RPT to Q-RPT, stability for a specific Q-RPT
should be established from experience.
3 Maximum deviation of the Q-RPT indication from the true value of applied pressure including precision, predicted
one year stability, temperature effect and calibration uncertainty (assumes calibration reference uncertainty of
± 0.005 % of reading), combined and expanded (k=2) following the ISO “Guide to the Expression of Uncertainty in
Measurement.”
Fluke Calibration
PO Box 9090,
Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (877) 355-3225 or Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or Fax (905) 890-6866
From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116
Web access: http://www.flukecal.com
©2011 Fluke Calibration. Specifications subject to change without notice.
Printed in U.S.A. 11/2011 3798035C D-EN-N
Modification of this document is not permitted without written permission
from Fluke Calibration.
Fluke Calibration. Precision, performance, confidence.
Ordering information
Model
PGC-10000-AF Pneumatic Gauge Calibrator
Includes
GPC1-10000-AF Pneumatic Pressure Controller
RPM4/HPMS A70M/A20M Reference Pressure
Monitor
GB-152-AF Gas Booster
Interconnecting hardware
Test connection adaptor kit
Molded transit case for RPM4/HPMS
Operation and maintenance manuals
A2LA accredited calibration report
System final test report