PRODUCT INFORMATION

RDU 40 & 2210 DU

Remote Display Unit RDU 40 & 2210 DU
BENEFITS

• Remote read-out
• 100 m (328 ft) distance from gauge
• Easy to install
• Easy to use
• Robust, weather / dirt protection
• Excellent readability
• Provides temperature inputs for TankRadar Pro
The display unit is available in two versions; RDU 40 for use with TankRadar
Rex and 2210 DU for TankRadar Pro.
Its robust design makes it withstand many years of outdoor use under harsh
environmental conditions. The display functions are software controlled and
interacts with the connected TankRadar gauge.
It is easy to work with via the 4-key display keyboard.
The functionality can be customized to meet the specific requirements from
the users of TankRadar Pro and TankRadar Rex.

Level

Level

6.767

m

6.767

m

RDU 40 for TankRadar Rex
The RDU 40 is remotely mounted, connected by a 3-wire cable. Up to two
units can be connected to one TankRadar Rex gauge.
It can display calculated data, such as level, average temperature, volume,
signal strength, etc.
The operator can set up a user-defined window where the most useful information is presented. This window will be shown as the default view.
The RDU 40 can view up to six spot temp elements connected to TankRadar
Rex.

2210 DU for TankRadar Pro
The 2210 DU can be remotely mounted, connected by a 4-wire cable, or factory mounted at the TankRadar Pro’s enclosure. If remotely mounted, up to six
PT100 elements can be connected to the optional temperature measurement
card in the 2210 DU.
The TankRadar Pro is fully configurable from the 2210 DU. User-friendly
menues and an installation wizard will guide through the complete start-up
procedure.
All measured and calculated values such as level, spot temperature, volume,
signal strength etc are displayed as numeric values or shown as bargraphs. A
user-defined window can be set up by the operator.

Level

6.767

m

RDU 40 & 2210 DU
Specification Remote Display Unit
View/Software:

RDU 40 (for TankRadar Rex)

2210 DU (for TankRadar Pro)

Available data

Level, ullage, volume, level rate,
signal strength, average temperature

Level, ullage, volume, level rate,
signal strength, spot temperature

Configuration of RTG from display

No

Yes

Display: single view

Easy readable with character size 25 mm (1 inch)

Easy readable with character size 25 mm (1 inch)

Display: standard view

Maximum 7 text lines, 16 characters/line
can be displayed in each screen view

Maximum 7 text lines, 16 characters/line
can be displayed in each screen view

Display: user-defined view

The customer can set up a view where
the most useful information is scrolled.

The customer can set up a view where
the most useful information is scrolled.

Electrical:

RDU 40 (for TankRadar Rex)

2210 DU (for TankRadar Pro)

Display type

Graphic LCD display 128 x 64 pixels

Graphic LCD display 128 x 64 pixels

Ambient temperature

-20 to 55° C (-4 to 130° F)

-20 to 55° C (-4 to 130° F)

Ex class

ATEX: II 2 (I) G, EEx ib IIC T4
IECEx: Ex ib IIC T4
FM: Intrinsically safe, Class 1, Div 1,
Group A, B, C, D

ATEX: II 2 (I) G, EEx ib [ia] IIC T4
IECEx: Ex ib IIC T4
FM: Intrinsically safe, Class 1, Div 1,
Group A, B, C, D

Max cable length

100 m (328 ft) (total length if two units
are connected to the same gauge

100 m (328 ft)

Cable requirement

3-wire, shielded instrument cable, min 0.5 mm²
(AWG 20), max 2.5 mm² (AWG 14)
Max Ø 14 mm (0.55 inch)
Min insulation thickness for each wire
0.25 mm (0.00985 inch)

4-wire, shielded instrument cable, Min 0.5 mm²
(AWG 20), Max 2.5 mm² (AWG 14)
Max Ø 14 mm (0.55 inch)
Min insulation thickness for each wire
0.25 mm (0.00985 inch)

Temperature option:

RDU 40 (for TankRadar Rex)

2210 DU (for TankRadar Pro)

Measuring range

Not applicable! Temperature elements are connected
directly to TankRadar Rex. See separate specification.

-50 to 250° C (-58 to 480° F)

Number of elements

3 x 3 wire PT100 spot or
6 x PT100 spot with commun return

Ex class

Ex ia IIC T4

Ambient temperature

-40 to 70° C (-40 to 158° F

Accuracy

± 0.5° C

Resolution

0.1° C

Cable requirement

Shielded instrument cable, Min 0.5 mm² (AWG 20),
Max 2.5 mm² (AWG 14)
Max Ø 18 mm (0.7 inch)
Min insulation thickness for each wire
0.25 mm (0.00985 inch)

Mechanical:

RDU 40 (for TankRadar Rex)

2210 DU (for TankRadar Pro)

Material housing

Die-cast aluminium

Die-cast aluminium

Dimensions

150 x 120 x 70 mm (6 x 4 x 3 inch)

150 x 120 x 70 mm (6 x 4 x 3 inch)

Weight

1.2 kg (2.6 lb)

1.2 kg (2.6 lb)

Cable entry

2 x M20, 1 x M25
Optional: ½ inch NPT and ¾ inch NPT
by external adapters

2 x M20, 1 x M25
Optional: ½ inch NPT and ¾ inch NPT
by external adapters

Protection class

IP 67, NEMA 4

IP 67, NEMA 4

Weather/dirt protection lid

Included

Included

Emerson Process Management
Rosemount Tank Gauging
Box 130 45, SE-402 51 Göteborg, SWEDEN
T: +46 31 3370 000 F: +46 31 25 30 22
E-mail: sales.srt@emersonprocess.com
www.rosemount-tg.com

© Rosemount Tank Radar AB. Fifth Edition. Septebmer 2006. Ref.no: 106029En



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : Yes
XMP Toolkit                     : 3.1.1-111
Instance ID                     : uuid:aec4d25e-3055-4787-b5d7-9ab94909154e
Document ID                     : adobe:docid:indd:2cfb135f-4d13-11da-b5ab-a79d86000865
Rendition Class                 : proof:pdf
Derived From Instance ID        : 61e26f5c-4d11-11da-b5ab-a79d86000865
Derived From Document ID        : adobe:docid:indd:3ecd0d30-3632-11da-b040-befc7385f3f0
Manifest Link Form              : ReferenceStream, ReferenceStream, ReferenceStream, ReferenceStream
Manifest Placed X Resolution    : 72.00, 300.00, 72.00, 72.00
Manifest Placed Y Resolution    : 72.00, 300.00, 72.00, 72.00
Manifest Placed Resolution Unit : Inches, Inches, Inches, Inches
Manifest Reference Instance ID  : uuid:97EA0D71B5ABDB119D848D1CC14494C4, uuid:42B6FD5262B0DB11BF3CD3CF295413EE, uuid:97EA0D71B5ABDB119D848D1CC14494C4
Manifest Reference Document ID  : uuid:96EA0D71B5ABDB119D848D1CC14494C4, adobe:docid:photoshop:cfb638f4-70e0-11d8-957a-eef08d2e3ec7, uuid:41B6FD5262B0DB11BF3CD3CF295413EE, uuid:96EA0D71B5ABDB119D848D1CC14494C4
Create Date                     : 2007:01:30 14:16:08+01:00
Modify Date                     : 2007:01:30 14:16:16+01:00
Metadata Date                   : 2007:01:30 14:16:16+01:00
Creator Tool                    : Adobe InDesign CS2 (4.0.2)
Thumbnail Format                : JPEG
Thumbnail Width                 : 256
Thumbnail Height                : 256
Thumbnail Image                 : (Binary data 10120 bytes, use -b option to extract)
Format                          : application/pdf
Producer                        : Adobe PDF Library 7.0
Trapped                         : False
Page Count                      : 2
Creator                         : Adobe InDesign CS2 (4.0.2)
EXIF Metadata provided by EXIF.tools

Navigation menu