Hitachi V1050F (V 1050F) Oscilloscope Operating Guide Operation Manual Schematic Diagram H_V1050 H V1050

User Manual: H_V1050

Open the PDF directly: View PDF PDF.
Page Count: 52

DownloadHitachi V1050F (V-1050F) Oscilloscope Operating Guide - Operation Manual Schematic Diagram H_V1050 H V1050
Open PDF In BrowserView PDF
Model

V-1050

OSCI.... LLOSCOPE
,

.

OPERATION rYlANUAL

Contents
1.
2.
3.
4.
5.

6.

7.

8.
9.
10.
11.

Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
composition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
How to produce the bright line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Method for Connecting Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
( 1) When us ing a probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(2) When a di rect connect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(3) Caution for using the measured signal as an external
trigger signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measuring Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(1) DC voltage measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(2) AC voltage measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(3) Period measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(4) Measurement of time difference . . . . . . . . . . . . . . . . . . . . . , . . . . . . . . . . . . .
(5) Measurement of rise (fall) time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(6) Measurement of single-shot signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(7) Trigger of complexed waveform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(8) Phase shift between two signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(9) Measurement by X-Y operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(10) How to use TV exclusive synchronization . . . . . . . . . . . . . . . . . . . . . . . . . .
(11) Operating procedure of delayed sweep ., . . . . . . . . . . . . . . . . . . . . . . . . . . .
(12) How to use TRIG VIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Panel Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
( 1) Power Supply and CRT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(2) Vertical deflection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(3) Horizontal defleqtion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(4) A triggering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(5) B triggering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
(6) Miscellaneous . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Users Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Schematic Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Panel Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.
.

1
1

2
5
8
8
9

10
11
11

12
12
13
14
14
15
15
16
18
20
24
25
25
26
28
30
33
33
35
39

40
49

I. Features

(7) Trigger view:

Hitachi Model V-1050 oscilloscope is a

Displays the selected internal or external

portable-type, advanced-class oscilloscope with

trigger signal.

a bandwidth of DC to 100 MHz Designed by putting

signal is selected, you can correlate the

If an external trigger

special emphasis on operability and ruggedness,

time between the trigger signal and input

this oscilloscope has the following features:

signals.

(1) Wide bandwidth:
The instrument has a bandwidth from DC to
100 MHz

2. Composition

(2) High sensitivity:
Sensitivity is 500

(8)Auto focusing
Focusing shift is automatically corrected.

Composition of Model V-1050 oscilloscope is
~V/div.

(3) Large 6" screen:
Employment of a large square CRT makes waveforms easier to observe.
(4) Internal graticule:
Employment of internal graticule CRT permits
waveforms observation to be made without
parallax error.
(5) TV synchronization:

as follows.
(1) Model V-1050 oscilloscope unit ............ 1
(2) Probe (AT-10AE 1.5) . . . . . . . . . . . . . . . . . . . . . . . 2
(3) Fuse (2A for 100V and l20V set
or lA for 220 and 240V set) ......... l
(4) Dust proof cover . . . . . . . . . . . . . . . . . . . . . . . . . . 1
(5) Sc.rew driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
(6) Power supply cord . . . . . . . . . . . . . . . . . . . . . . . . . l
(7) Operation ManuaL . . . . . . . . . . . . . . . . . . . . . . . . . 1

Employment of a new TV sync/separator
circuit allows instrument to observe TV
signals stably.
(6) Delayed sweep.
A portion of the signal can be magnified
before observation.

1

3. Precautions

such as a vase on the oscilloscope.

Precautions to be observed to lengthen the

*

service life of this instrument.

Do not place the instrument in a place where
vibration is strong. Avoid using the instrument at
a place vibrating violently. Since the oscillo-

C

o.

*

scope is a precision instrument, excessively
Avoide installing instrument in an extremely hot
or cold place.

.....

strong vibrations may cause damage.

*

o Avoid placing this instrument in a place

"-'"

Do not place the instrument near a magnet or
magnetic body. An oscilloscope is an equipment

exposed to sunlight for a long period of time,

using electron beam.

in a closed car in midsummer, or near a room

Therefore, do not bring a

magnet close to the instrument or do not use the

heating device such as a stove.

instrument near an equipment generating strong

o The operating maximum ambient temperature is

magnetic force.

50°C.

*

Do not use instrument that has been left outdoors
on a cold winter day. The operating ambient tem-

*

Do not put a heavy objects on the oscilloscope.
Do not block the ventillation holes.

*
*

Do not apply a heavy shock to the oscilloscope.
Do not insert a wire, pin, etc. through the

*

Do not drag the set, leaving the probe attached

*

Do not leave a hot soldering iron on the cabinet

perature is O°C or more.

*

Avoid moving the instrument rapidly from a hot
place to a cold place of vice versa, or condensation may form on inside of the instrument.

*

ventillation hole.

Keep the instrument away from damp air, water,
and dust. Unexpected trouble may be

cau~ed

when

the instrument is placed in a damp or dusty
place.
The operating ambient humidity is 35-85%. Since
an accidental instrusion of water may also cause
troubles, .do not place a water-filled containers
2

to it.
or the screen.

*

Do not try to turn the instrument upside down.
Otherwise, knobs may be broken.

*

Do not use the instrument upright, leaving BNC

C:,'

cable connected to EXT BLANKING, GATE OUT
PUT terminals on the rear panel.

NOTE: When opening the case, disconnect the power

Otherwise,

supply plug beforehand without fail.

the cable may be damaged.

When cleaning the inside, insure beforehand
that no electricity remains in the condensers
of the power supply circuit.

Recheck the operating procedure and if problem
persists, contact a nearly service station or

*

Cleaning of CRT
Dirty surface of CRT screen tends to cause measur-

agent.

ing errors.
Remove the stains on CRT and filter by using a

*

Removal of stain from the case.

clean and soft cloth, paying attention not to

o When the outside of the case is stained,

impair them.
When the stain is extremely heavy, wash them with

remove the stain by first wiping it lightly
with a cloth moistened with neutral washing

neutral washing agent and then leave them stand

agent and then wipe the surface with a dry

until the moisture is remov-ed naturally.

cloth.

o If the screen is installed while it is moistened,
water rings may be formed and the waveform may

o When the panel surface is stained, remove the
stain in similar way with a clean, soft cloth.

be blurred to become hard to observe.

Pay

When heavy stains are present, first remove

attention not to leave finger prints on it.

the stains by wiping the surface lightly with
a cloth moistened with diluted neutral washing
agent and then wipe thoroughly with a dry cloth.
o When dust has accumulated on the in side,
--,:-)

remove it by using a dry brush, or by using the

*

Check the line voltage.
The operating voltage range of this oscilloscope
is as shown below.

Check the line voltage without

fail before turning on the power switch.

exhaust of a compressor or a vacuum cleaner.

3

Rating

length.)

Line Voltage (50/60Hz)

AC100V

AC 90V - llOV

AC120V

AC108V - l32V

AC220V

AC198V - 242V

Shape
(Diameter x length)mm

AC2l6V - 264V

AC240V

Nominal volts

~5%

JIS type name

2A

6.35~

x 31. 8

MF61NM250V 2A AC

lA

6.35~

x

31.8

MF61NM250V lA AC

Do not increase the brightness too much.

In the case of normal shipment, the voltage

Do not increase the brightness of the spot and

selector will be set convenient for user up.

trace too much.

When it is intended to use the oscilloscope on
volt ages other rating, voltage selector can be
turned.

(Rated

are indicated on the

volt~ges

(' J

at 400Hz.

*

*

The standards if the fuses are as

follows.

Your eyes may be strained and the

fluorescent surface of CRT may be burnt.

*

Do no apply an excessive voltage.
The input withstand voltage of each in put con-

rear panel of the oscilloscope.)

nector and probe input is as follows.

Use only specified fuses.

a voltav,e higher than specified.

Never apply

In order to protect the circuit against overcurrent, a 2A (make use of ACIOOV or AC120V) or
lA (make use of AC220V or AC240V) is used on the
primary side of the power supply.

When this

fuse is below out, check thoroughly the cause,
repair any faulty point present, and then replace
with a. specified fuse.

Do not try to use the

fuse other than the specified ones.

Otherwise,

fault may be caused or danger may be invited.
(Particularly, do not use a fuse different from
the specified one in current capacity and in
4

INPUT

direct

When probe is used
EXT TRIG INPUT
EXT BLANKING

250V(DC + AC peak at 1 kHz)
500V(DC + AC peak at 1 kHz)
250V(DC + AC peak)
20V(DC + AC peak)

I Calibration Interval 1
To maintain jnstrument accuracy, perform the
calibration of the V-I050 at least every 1000
hours of operation, or every six months if
used infrequently.

(.•...::)
".

4. How to produce the bright line
Unless handled erroneously, this instrument

VAR HOLD OFF

MIN (full c.c.w.)

B TRIG MODE

AUTO

will never become faulty by ordinary operation.
Set all the levers of the switches either to

Before turning ON the POWER switch, insure
the power supply voltage is within the range of
l08V

~

the left side or to the upper side.
After ending all the settings mentioned above,

132V for AC 120V set, 198V - 242V for

AC 220V set, and 216V - 264V for AC 240V set.

turn ON the POWER. and, 15 seconds later, rotate

Refer to the indication on the rear panel of the

the INTENSITY knob clockwise.

instrument for other voltages.

bright line will appear.
If observation is to be started immediately,

Connect the plug of the power cord on the
rear panel into the power supply wall socket and

set the FOC(TS knob a.t a point where the bright

set the knobs as follows.

line is sharpest.

POWER/INTENSITY

If the instrument is not used with the power

IOFFI

th~

Midrange

supply turned on rotate

AC-GND-DC

GND

clockwise to reduce the brightness and also blur

HPOSITION

Midrange

FOC(TS

V.MODE

CHI

BW LIMIT 20MHz

Full bandwidth (button out)

D1SPLAY

A

INTENSITY counter-

the FOCUS.
INOTEI
For usual observation, leave the following
non-calibrating function section set to "CAL"

Midrange

position.

A TRIG MODE

AUTO

VARIABLE

A TRIG COUPING

AC

In this case the VOLTS/DIY is

A TRIG SOURCE

CHI

calibrated to its indicating value.

A TIME/DIV

O.5ms/DIV

B TIME/DIV

Arbitrary

~

-+-

<')

Then the sweep

POSITION

PULL SWP VAR

Rotate in the direction of arrow.

Push in the knob or rotate in the
direction of arrow. In this case
the TIME/DIV is calibrated to its
indicating value.
5

Align the bright line with the horizontal

line appears even when no signal is present or

scale line at the center of the screen by operat-

when AC-GND-DC switch js at GNP position.

ing CHI POSITION.

means that the measurement of DC voltage can be

In some cases the bright line

This

may be oblique to the scale slightly by the

measured.

effect of earth magnetism.

when observing low frequency signals of about

In this case, bring

the bright line untjl it lies on the horizontal
scale line at the center of the screen by
properly adjusting the semi-fixed variable
resistor TRACE ROTATION on the front panel.

The following switching is needed

30 Hz or less.
NORM

A TRIG MODE

DC

A TRIG COUPLING

- General measurement -

Triggering can be effected by operating LEVEL

(1) In the case of observing a single waveform

knob under this setting.

Use CHI or CH2 when not observing the phase
difference between two waveforms or when engag-

settings.

ing in a operation other than X-Y operation.

MODE Switch of Vertical Axis

CH2

Make the following settings when using CHI.

A TRIG SOURCE

CH2

MODE Switch of Vertical Axis

CHI

MODE Switch of TRIG

AUTO

TRIG SOURCE

CHI

AC-GND-DC

AC or DC

Under these settings almost all the repetitive
signals of about 30 Hz or more applied to CHI
can be triggered and observed by adjusting
A TRIG LEVEL.

(When, the range of TIME/DIV is

between 2mS/DIV and 20ns/DIV)

Sjnce the MODE of

horizontal axis is at AUTO position, the bright

6

When using only CH2,

use the instrument after making the following

(~.

(2) When observing two waveforms

..

)

(3) When observing three or four waveforms.

Observation of two waveforms can be made easily

This instrument can be observed the triggering

by setting the MODE switch of vertical axis to

signals (CH3 as A TRIG, CH4 as B TRIG) at

ALT or CHOP.

When observing two waveforms of

depressed position of TRIG VIEW .

high repetition frequencies set the MODE switch

The table shows the relation between setting

to ALT and, in the case of low frequencies, set

of vertical mode and traces on CRT.

it to CHOP.

Normally, ALT and CHOP are recom-

mended by selection of TIME/DIV.

V. MODE
CHI

ALT;

0.2ms/DIV

20ns/DIV

CHOP;

0.5ms/DIV

0.5s/DIV

No. of display
CHl, CH3

CH2

CH2, CH3

ALT

CHI, CH2, CH3, CH4

When measuring the phase difference, measure

CHOP

CHI, CH2, CH3, CH4

after effecting triggering with leading phase

ADD

CHI + CH2, CH3, CH4

signal.
The displaying position of CH3 (A TRIG) is
located near the 100%'s sub-graduation, and
possible to adjust by A TRIG LEVEL knob.
CH4 is fixed at O%'s sub-graduation.

7

5. Method for Connecting Signals

(4) When observing waveform with X-Y
Set the MODE switch of vertical axis to CH2
(X-Y) and DISPLAY switch to X-Y.

Then the

instrument works as an X-Y oscilloscope.
Each input is applied to the instrument as

The first step of measurement is introduce the
signal desired to measure to the oscilloscope
properly.

Do it with utmost care.

(1) When using a probe

follows.

Use the attached probe, AT-10AE 1.5, when

X-axis signal

measuring a high frequency wave with high

(horizontal axis signal)

CHl INPUT

accuracy.

CH2 INPUT

input signal is attenuated by this probe to

Y-axis signal

It should be noted, however, that since the

(vertical axis signal)

In this case leave the horizontal axis magnification switch (PULL

x

10MAG inner shaft knob)

at depressed position, and the AC-GND-DC of
X-axis (CH1) to AC, also TRIG VIEW and BW LIMIT
20MHz are undepressed position (button out).

1/10 before it is input to the oscilloscope the
use of the probe is disadvantageous for low
signals, and that at the same time the measuring range is extended by that amount for high
signals.


U

-"'Lt . - -

-- - - -

10%~~-+-rY-+-~hcl-+-L

'0

--\

- - - -- - - -

---

1

O%~~-+-r~+-r-~~

Zero level (reference line)

(3) Period measurement
This will be explained taking the drawing at
follows as an example

(2) AC voltage measurement

The same as paragraph 6 (1), "DC voltage mea-

100%
90%

-- --

surement", but here those is no need of matching the zero level with the scale line.

Move

the zero level at will to a position easy to

- -- - - -- - -

- - -

111 ill /\ 11\ /\
J ilL i I \ I \ I \

IJ

I

\

\1

10%
0% -- - - - - - -

\1 j

- - -

-

-

observe.
In the drawing as follows, VOLTSjDIV is IVjDIV,
IVjDIV

x

5

=

5 Vp-p [50 Vp-p at using the probe

AT-IO'AE 1. 5( 10: 1) 1.

When magnifying and observ-

ing a small-amplitude signal superinposing as
increases, set AC-GND-DC switch to AC.
voltage is cut off and AC voltage can be
observed by increasing sensitivity.

12

The DC

Time A

Time B

One period covers the time A and time B, which
are separated from each other by 2.0 DIV on the
CRT.
When the sweep time is ImsjDIV, the period is
given by
1 msjDIV

x

2.0

2.0 ms
2.0 x 10-3 s

(

.)

This means that CHI is used as the triggering

Accordingly, the frequency is
1/(2.0 x 10- 3 ) = 500 Hz

signal when investigating the length of time

(However, when the knob MAG x 10 is at pulled

by which the signal of CH2 is delayed from the

out position, TIME/DIV must be converted to

signal of CHI.

1/10 since the sweep is magnified.)

case.

(4) Measurement of time difference

CH2 is used in the reversed

In other words, the signal leading in

phase is selected as the triggering signal

Triggering signal source "SOURCE" is selected

source.

as offering reference signal when measuring the

If this process is reversed, the portion to be

time difference between two signals.

measured may sometimes not appear on the screen.

that pulse trains as shown in (a).

Assume
Then (b)

shows the case when CHI is taken as the trig-

Thereafter, equalize the amplitudes of the two
signals appearing on the screen or superimpose

gering signal source and (c) the case where

one another.

CH2 is taken.

Read the time difference by the interval

CHI

between 50% amplitude points of the two signals.

':

~
I

Sometimes the superimposing method is more con-

I

1Pu-u--u-L

venient from the point of view of procedure.

I

,I
,
caz,
I

)

,,

(a)

l

CHI

CH I ",

,'-+------,,

I
I

I

-

'"

Time
difference

"

CHZ

'\

- -'

rn.eft

difference Equalize
~ ~
amplitudes

o

CHZ

_____ J.::::-V by

VAR

!
Equal amplitude measuring method
(b)

Superposition measuring method

(c)

13

The true rise time is given by

{Cautions ')
Since the pulsed wave contains many high-

Trx

= ~ Tro2 - Trs2

frequency wave components (higher harmonics)
depending on its width or period, pay the same
attention as given to high frequency signals when
handling it.

Accordingly, use a probe or coaxial

cable and shorten the earth lead wire as much as

Moreover, in general, in a circuit free from
waveform distortion such

the following relationship is established
between frequency band and rise time.

possible.

fc

(5) Measurement of rise (fall) time

where, fc

To measure the rise time pay attention not only
to the above mentioned items but also to measurement error.

/

a~ overshoot and sag, ('

tr

x

tr = 0.35
Frequency band (Hz)
Rise time (s)

The rise time and fall time are determined by
the time elapsed between the 10% to 90% values

The following relationship exists between the
rise time Trx of the waveform to be measured,
the rise time Trs of oscilloscope, and the
rise time Tro displayed on the screen.

of pulse width.
(6) Measurement of single-shot signal
Single sweep is conveniently used in the measurement and photography of single-shot signal, waveform of remarkably non-uniform repetition (such

When the rise time of the pulse going to be

as impulse waves, sound waves, switch noise

measured is sufficiently longer than the rise

waves).

time of the oscilloscope (3.5ns in our case),
the effect of the rise time of the oscilloscope
on the measurement can be neglected.

However,

if both are close to each other, measurement
error may be caused.

14

First set A TRIG MODE on the front

panel to NORM and the effect synchronization by
using a signal or repetitive waveform of about
the same level and by rotating LEVEL.

('"

Next, set A TRIG MODE to SINGLE and depress
RESET button to insure that sweep is made on and

only one time.
Then remove the vertical axis
input signal (by, for example, setting AC-GNDDC switch to GND) , depress SINGLE RESET and
,

"')

~,;-, ;-.

insure that READY LAMP goes on.

B/\

H "

Nf'\

A7Ci0c;d}i~

Y' Trigger level
Y

setting line

Ca) Signal waveform

When apply the input signal, sweep is made for
one time and READY LAMP goes out.

Since sweep

is also made at no signal time depending on
the level, do not rotate LEVEL once SINGLE
RESET is completed.

(c) When the trigger
setting level is y'

Triggering of complexed waveform

(7) Triggering of complexed waveform
In the case shown in the Fig.

and advances to C, E, F,····· and which allows

(a) below where

waveforms greatly different in amplitude alternate, the waveform is doubled if the trigger
level is not set properly.

In the case where

the trigger level is selected as Y line two
waveforms, one starting with A and advancing to
B, C, D, E, F,····· and the other starting with
E and advancing to F, G, H,

I ..... , will appear

alternately on the screen.

They will be

doubled as shown in Fig.

(b) When the trigger
setting level is Y

(b), for which no

triggering can be taken.

(8) Phase shift between two signals
To measure a phase shift between two signals of
the same frequency, the dual trace display
system can be utilized up to the upper limit
frequency of the amplifier.
First, position both signals on the cent er line
of the scale, e.g. just 4 div, as shown in the
following figure by means of the VARIABLE and
horizontal positioning knobs.
Next, set the distance where the center of the

In such a case, rotate LEVEL clockwise until
the trigger level comes to Y' line.

synchronization.

Then the

waveform of the base channel intersects with
that of the scale to 8 div horizontally.

waveform on the screen becomes the one is
shown in Fig.

(c) follow which start with B

15

(9) Measurement by X-Y operation
The phase shift between two signals of the same
CHI (Bas e)

~

CH2

J

frequency can also be measured using a

V. I/V

/~/ "~

~

-I
'l

\

1,1.

\.
\1\
~\.

'/

'/

Lissajou's figure by X-Y operation.
A sine wave input is applied to the audio
circuit being tested.

The same sine wave input

is applied to the vertical input of the oscilloscope, and the output of the tested circuit
is applied to the horizontal input of the oscil-

-----'>

__ Phase di ffe ~nce

0.7 div
8dlv (360")

------')

loscope.

The amount of phase difference between

the two signals can be calculated from the
resulting wave form.

As shown in the above figure, set 1 cycle, 360°
to 8 div.

360°
Then, 8div = 45°/div.

Accordingly, the phase difference in the above
example can be calculated as follows:

1. Using an audio signal generator with a pure
sinusoidal signal, apply a sine wave test
signal at the desired test frequency to the
audio network being tested.
2. Set the signal generator output for the normal

Horizontal distance on the screen:0.7 div
Phase difference = 45°/DIV x 0.7 div = 31.5°

If desired, the circuit's output may be observed

If the portion of the phase difference is much

on the oscilloscope.

smaller, use the MAGNIFIER at the xlO position

over-driven, the sine wave display on the oscil-

in the above setting.

At this time, 360° is

displayed in 8 div xlO.
360°
Then, 8 divxlO = 4.5°/DIV (0.2 div=0.9°)

16

operating level of the circuit being tested.
If the test circuit is

loscope is clipped and the signal level must be
reduced.
3. Connect the Channel 2 probe to the output of
the test circuit.

~:'.'

4. Set the DISPLAY to X-Y position.
5. Connect the Channel 1 INPUT probe to the input
SINE

of the test circuit. (The input and output test

B

~

A

connections to the vertical and horizontal
WHERE

oscilloscope inputs may be reversed.)

PHASE

6. Adjust the Channel 1 and 2 gain controls for a

ANGLE

suitable viewing size.
7. Some typical results are shown in Fig.(b).

If

the two signals are in phase, the oscilloscope
trace is a straight diagonal line.

If the ver-

tical and horizontal gains are properly
adjusted, this line is at a 45° angle.
A 90° phase shift produces a circular oscilloscope pattern.

Phase shift of less (or more)

than 90° produces an elliptical oscilloscope
pattern.

~

The amount of phase shift can be

calculated from the oscilloscope trace as shown

Fig. (a)

Phase shift calculation

/

~

NO AMPLITUDE DISTORTION
NO PHASE SHIFT

AMPLITUDE DISTORTION
NO PHASE SHIFT

~

0

180 0 OUT OF PHASE

in Fig.(a).

0

NO AMPLITUDE DISTORTION
PHASE SHIFT

0

AMPLITUDE DISTORTION
90° OUT OF PHASE
PHASE SHIFT
Flg. (b)
Typical phase measurement
oscilloscope displays.

17

(10) How to use TV exclusive synchronization

Cl)

Difference in the circuits

On the image waveform of TV
Exclusive circuit for conventional
oscillograph

l

SiV~

EXCIU..

(~rinciple

f - - - - - . - - - - - - , - - - - - - - - - - . __

General circlllt

i-,iffiPl€ sync~r0111z1ng

I ;t
t.": .

circuit for this.
lDstrument

drawing)

• TV exclusive synchronizing

Clrcult

separator circuit

-------4------------------Video slgnal

si

Synchronizing
signal pulse
(SYNC pulse)

To trigger circuit I

To trigger circuit

I

~

~----------------lV(vertical)

."

~I

"

~I~

In the work concerned with TV, complexed signals
containing video signal, blanking pedestal

I

"
u

I

I

I

SYNC Peak
Detector

signal, and synchronizing signal are often
However, since the waveform is commeasured.
plexed, a special circuit is needed to effect a
stable triggering with vertical waveform.

I

SLOPE

!

II

Hard to synchronize I Synchronization is more I
because video signal I easily effected than in I
is applied directly ,the circuit shown at
i left, because the signal I
as trigger signal.
I i~ integrated to remove
I

L-...L_ _ _ _ _ _ _ _

hlgh frequency com-

I pOTIents.

I

triggering~
de;:~ted I

Stable
obtained, after
the peak of SYNC pulse
I
then the trigger level in
feed back roap of operational amplifier should
be fixed its SYNC tip.

(

18

I

..

,

®

Operation
To observe vertical
signal

I1

I
I

I1
I1

To observe horizontal
signal
---,

ijm millIII III"I

IU
A TRIG MODE: TV
COUPLING: TV-V
LEVEL: PULL TV(-)
TIME: O.lms/DIV O.2S/DIV

Will

IIlllll

I
I

A TRIG MODE: TV
COUPLING: TV-H
LEVEL: PULL TV(-)
TIME: 50~s/DIV 20ns/DIV

Generaly, video signal of TV has negative
synchronizing signal.

Use at pull out

position which is A TRIG LEVEL knob.

If

the video signal is inverted then A TRIG
LEVEL is depressed.
signal

',)

4-synchronizing
signal

Example of

Example of

(-) synchroniging signal (+) synchronizing signal

If the sync and blanking pulses are
positive, set the switch to TV(+).
19

(11 ) Operating procedure of delayed sweep

INTEN

A TIME/DIV

Choose A TIME/DIV properly

There are two kinds of time delay sweep; one is

B TIME/DIV

Set B TIME/DIV at a more rapid

AUTO time delay sweep (continuous time delay

sweep time than the one set by

sweep) and the other TRIG time delay sweep (trig-

A TIME/DIV

gering time delay sweep).

These are selected by

two, the INT (CHI, CH2 internal triggering time
delay sweep) and EXT (external triggering time delay
sweep).
mode.

Usually, the instrument is used in AUTO
Although the. AUTO time delay sweep is

easy to operate the maximum magnification factor
is limited by delayed jitter (rolling) (to a few
hundred times).

On the other hand, since no

jitter is generated in TRIG time delay sweep,
this sweep has the feature of being enabled to
increase the magnification factor.

However, the

magnification factor is limited by the brightness
of CRT (to a few thousand times).
At time of AUTO (continuous time delay sweep)
Effect triggering by A sweep and set the knobs
as follows.

Then the high brightness portion of A sweep will
appear without fail (if not, adjust INTENSITY).

TRIG time delay sweep is further classified into

20

AUTO

DISPLAY

MODE switch of B TRIG.

(i)

B TRIG MODE

(Used to magnify and observe any portion of a
complexed waveform in horizontal direction. )

Rotate DLY TIME POSITION knob (the cent er FINE
is for fine adjustment).

The high brightness

portion will move continuously.

Bring this high

brightness portion to the position desired to
be magnified, switch DISPLAY to B.

Then the

high brightness portion is magnified to occupy
the full area of the screen.

The sweep time is

the indicated value of B TIME/DIV.

i-----:::==========t-Bright

portion

A INTEN ALT B~

)

(~

fail.
state.)

(This state is called the B-triggered
Rotate DLY TIME POSITION knob.

The high brightness portion will move continuously.

Further rotation will bring this

portion to the next "peak".

Therefore, bring

this high bright portion to a position
desired to be magnified and then set DISPLAY
at B.

The high brightness portion is

magnified to occupy the full area of the
screen.

A INTEN ALT Blx-vl

~)

(

(D

The sweep time for this case is the

value indicated by B TIME/DIV.

The gen-

locking is fixed at AC.
A INTEN B X-Y

At time of TRIG (TRIGGERING time delay sweep)
Effect TRIGGERING with A sweep and set the knobs

c_----""'~

__~)

as follows.
B TRIG MODE

TRIG (INT or EXT)

DISPLAY

INTEN

A TIME/DIV

Chose A TIME/DIV properly.

B TIME/DIV

Set B TIME/DIV at a more rapid
sweep time than the one set by A

, ,:

)

TIME/DIV.

/

SLOPE

Set SLOPE either to + or to -

Thereafter, rotate LEVEL.

The high bright-

ness portion of A sweep will appear without

A INTEN B X - Y

(-----=~=----~)
* The case in which SLOPE
is + is shown

21

®

ALT sweep
This ALT sweep is displayed A sweep and B
(delayed) sweep alternatively.

(d) Only displayed at
intensive part

Belowing figures

are showing when CHI to be put triangle waveform

f.

A IETEN ALT B X-Y

and CH2 to be put rectangular waveform.

(

(a) Normal observation

lI#J )

A INTEN ALT B X-Y

(1)lI

)

(b) Prepared to magnify

When DISPLAY switch is set the ALT, displaying
signal of B sweep can be control the position
CHI A sweep

A INTEN

(lI#J

ALT

to easily observe by TRACE SEP.

The range of

position be shift to upper and lower three

B X-Y

)

CH2 A sweep

divisions compere the A sweep.

Then depress

the TRIG VIEW switch, 4 traces come to screen
in totally 8 traces to observe.
(c) (b) and intensive
display in alternatively sweep

These new 4

traces are A and B sweep of CH3 and CH4.
I7"I-AAA.q,....,.~--

'-

f-

~CH3

Input Signal

~CHl

Input Signal

~CH2

Input Signal

~CH4

Input Signal

f

'.

'~'..'.

,)

;1

7. Panel Descriptions
(1) Power Supply and CRT
Operating voltage and fuse
This model can be operated from either a 100
:)

volt, a 120-volt, a 220-volt or a 240-volt
nominal line voltage source.
The Line Voltage Selector assembly on the
rear panel converts the instrument from one
operating range to the other.

In addition,

this assembly changes the primary connections
of the power transformer to allow selection of
one of four regulating ranges.
also includes the line fuse.

The assembly
Use the following

procedure to convert this instrument between

C!)

nominal line voltages or regulating ranges.

brightness of the CRT display.

1. Disconnect the instrument from the power source.
2. To convert from lOO-volts nominal to 220-volts
nominal line voltage or vice versa, pull out
the Voltage Selector switch turn it, and plug
it back into the remaining holes.

Change the

,,")

3. Before apply power to the instrument, check
that the indicating tabs on the rear panel.

Clockwise adjust-

ment increases brightness.

@)

®

line-cord power plug to match the power-source
receptacle.

POWER/INTENSITY
Turns the power on or off and controls the

Pilot lamp
Lights when the instrument power is on.
FOCUS
Provides adjustment for optimum display definition.

@

TRACE ROTATION
Corrects slight tilting of trace caused by
external magnetic fields.

25

®

@

SCALE ILLUM
Controls graticule illumination.
area, photographing.

the input signal fed to CHI.

CHI OR

[K]

On X-Y operation, the kaob functions to

change the.sensitivity of the X-axis.
(CH2 or

[YJ

sensitivity switch)

This is a knob for switching the sensitivity of

INPUT

This is an input plug for use with the CHI

the input signal fed to CH2.

vertical amplifier and X-axis (horizontal axis)

is accomplished in ten steps from 5mV/DIV to 5V/

amplifier during X-Y operation.

DIV.

CH2 OR [YJ INPUT
This is an input plug for use with the CH2

change the sensitivity of the Y-axis.

Switching action

On X-Y operation, the knob functions to
To mea-

sure by the use of the indicated voltage sensi-

vertical amplifier and Y-axis (vertical axis)

tivity, be sure to set each of the VARIABLE to

amplifier during X-Y operation.

CAL position by turning fully clockwise.

AC-GND-DC
(Alternating Current-Ground Switch-Direct

signal is applied to the input terminal by the
are ten times the indicated voltage.

Switches the coupling of the signal fed to the
vertical axis input.
the DC position.

If the

use of a 1/10 low capacitance probe, the values

Current)

26

Switching action

is accomplished in ten steps from 5mV/DIV to 5V/
DIV.

dently of display portion or applied signals.
(2) Vertical deflection

®

sensitivity switch)

This is a knob for switching the sensitivity of

Compress display within graticule area, indepen-

®

0

illuminate the graticule when viewing in a dark

~ TRACE FINDER

(j)

VOLTS/DIY
(CHI or

Useful to

DC coupling is obtained on

On AC position, the direct

@

VARIABLE
This is a vertical axis sensitivity fine adjustment which is capable of attenuating to less

current component is blocked by a capacitor.

than 1/2.5 by indication of each range of VOLTS/

The GND position grounds the input of the ampli-

DIV.

fiers and opens the input terminal

sensitivity indicated by VOLTS/DIY, turn the

To measure a voltage by the use of voltage

c

VARIABLE clockwise fully to CAL.

@

)

a .. Put a high level of signal sinchronized with

POSITION, PULL xlO MAG

the observing signal into the A EXT TRIG

CHI # (Vertical position adjustment)
With the knob turned clockwise, the waveforms

trigger signal source.

of CH-l move upward.

terminal or CH2 terminal (when using CHI) as

When the knob is turned

counterclockwise, the waveforms move downward.

b.

CH2 it (Vertical position adjustment)
Clockwise rotation will move pattern up, and

c. In case of observing signals over 4kHz,
select "ALT" of vertical MODE.

counterclockwise rotation will move pattern
down
When the knob is pulled, the vertical axis
sensitivity at each range of VOLTS/DIY is

In case of observing low frequency signals
under 4kHz, select "HF REJ" of A TRIG COUPLING.

~

MODE
o CHI
Only the input signal applied to CHI is

increased by 10 times.

displayed.
INote I
1. When measuring at 5 mY/DIY

o CH2, I X-YI
~

5 V/DIY range,

use the xl (PULL xlO MAG switch is depressed
posi t ion).

2.

Otherwise the S /N and frequency-

Only the input signal applied to CH2 is
displayed.
o ALT

bandwidth of instrument will be decreased.

CHI and CH2 signals are displayed alternately

In case of the observation of signals in the

on consecutive sweeps.

high-sensitivity range of

500~V

to 2mV/DIV and

o CHOP

in the CHOP MODE, it is sometimes difficult to

CHI and CH2 signals are displayed simultaneously

get stable synchronization due to noises caused

by switching between channels at about 250 kHz
rate.

by the high amplification of vertical amplifier.
In these cases:

27

o ADD

(3) Horizontal deflection

Displays the algebraic sum of the channel I and
channel 2 input signals.

If the channel 2 dis-

play is inverted (press CH2 INV), an CHI minus
CH2 display results.

@ BW

@

DISPLAY
This switch is used to select the operation
mode of the horizontal axis.
o A

The BW LIMIT 20MHz switch provides a method of

o INTEN Although the sweep on the screen is A
sweep it indicates B sweep (delay time

quency signals when viewing low-frequency signals.

sweep) by intensity modulation.
o ALT

Signals applied respectively to A sweep

deflection system is limited to about 20MHz.

and B sweep appear on the screen alter-

Reduces the bandwidth of channel I and channel

natively each sweep.

2 to approximately 20MHz.

o B

INV
Inverts the polarity of the channel 2.

occupy the full area of the screen. The
sweep time at this time is B.

Useful to comparison of two signals of opposite
polarities, and observation of a differential
signal of CHI and CH2 along with ADD mode.
TRIG VIEW
This switch is used when desiring to observe
the waveform triggering signal on the screen.

The intensity modulated portion in
INTEN mentioned above is magnified to

@ CH2

o X-Y

This position is used when using the
instrument as an X-Y oscilloscope.

X

direction signal is input to CHI and Y
direction signal to CH2.

The vertical

deflection sensitivity at this time is

It is also used when measuring the phase dif-

read on CH2 VOLTS/DIV and horizontal

ference between the synchronizing signal and

axis sensitivity on CHI VOLTS/DIV.

other input signal.

Vertical position is set by CH2 POSITION

t ion. )

28

This

reducing interference from unwanted high-freWhen set it -3dB bandwidth point of the vertical

@

A sweep appears on the screen.
setting is used in normal cases.

LIMIT 20MHz

(quad phenomena observa-

and horizontal position by : POSITION.

@

words.

TRACE SEP
B sweep of the position adjustment knob when

Normally, leave this knob rotated to MIN.

ALT sweep mode.

L';)

INot;~e

instrument has a special control system to

easy observe at high speed sweep range when
DISPLAY is set to ALT or B.

Therefore, the

traces is still remained when decrease the
INTENSITY to minimum.
~ A TIME/DIV (Sweep speed selection)

The outer knob controls the A (Main) sweep rate,
which has from 20 ns/DIV to 0.5 s/DIV selects
23 fixed sweep speeds.
@lB TIME/DIV
The inner knob control the B (Delayed)-Sweep rate.

@

PULL SWP VAR
Provides continuous adjustment of A sweep TIME/
DIV between calibrated positions, when the
inner shaft is at pull.

Counterclockwise rota-

tion to the full delays the sweep by 2.5 times
or more.
Normally, the inner shaft is left depressed.
VAR HOLD OFF
Increases the time between sweeps and aids triggering on complex displays such as high-frequency signal,

Rotate the VAR HOLD OFF slightly to

obtain a stabilized triggering.

irre~ular

signal and digital

@

FINE
PULL xlO MAG
POSITION
This knob is used to adjust position of the
display horizontally.

It is indespensable in

the measurement of the time of waveform.
Display is moved toward right when the knob is
rotated clockwise and toward left with counterclockwise rotation.

The outer shaft is for

course adjustment and inner shaft for fine
adjustment.

A and B sweep are magnified 10

times by pulling out FINE knob (inner shaft) of
POSITION.

In this case the sweep time is 1/10

of the value indicated by TIME/DIV.

Bring the

portion of the waveform desired to be magnified
observed to the outer of the scale by operating
POSITION of the horizontal axis.
Next switch xlO MAG switch to PULL (pulled out
state).

Then the waveform placed at the cent er

is magnified in right and left directions.
The sweep time in this case is 10 times the
sweep speed obtained by TIME/DIV, in other words,
the reading is 1/10 of the sweep time indicated.
29

(4) A Triggering

@ SOURCE
This switch is used to select the triggering

100%
90%

signal source A Sweep.

-

c··

Used when triggering is made by observing

(

..

signal for CH2.

\

/
I1

o CH2

1\

I
100%
90%

Used when triggering is made by observing
signal for CHI

\

1/

o CHI

o LINE Used when observing a signal triggered with
the frequency of mains (AC POWER).
o EXT

Used when triggering is made by a signal
applied to EXT input connector, indepen-

/

dently from observation signal.
o EXTdO
..

......

Attenuates external TRIG INPUT signal by a

I

factor of 10.
@TRIG INPUT (CH3)
Input terminal for use for external triggering
@DLY TIME MULT

signal of A sweep.

This control is used to set the delay time of

@COUPLING

B sweep starting point with respect to A sweep

This switch is used to select the coupling mode

starting point.

of A sweep triggering signal.

When the above mentioned

DISPLAY is set in INTEN or B.

o AC

At this setting both the DC component and

C')

the very low frequency of triggering, signal
are cut off.
30

@

oHF REJ

A TRIG LEVEL

Among the AC components of triggering signal,

This knob is used to decide at which portion

the high frequency components of about 4 kHz

of the waveform should the sweep the started
This knob is also
by setting trigger level.

or more are attenuated.

A stabilized trigger-

ing unaffected by noises of about 4 kHz or

enabled to switch SLOPE.
(normal state) is for +

more can be obtained.
oLF REJ

Depressed position
SLOPE and PULL

position (state in which the knob is protrud-

Among the AC components of the triggering

ing) is for

SLOPE.

signal the low frequency components of less
than about 4 kHz are attenuated.

A stabilized

*Explanation of trigger polarity SLOPE.

triggering free from noises of less than about
4kHz can be obtained.
oDC

Push

Triggering signal is amplified unchanged.
This setting is used when triggering with a
very low frequency signal or when effecting DC
like triggering.

at time of

8

SLOPE

oTV-H
This setting is used when observing the entire
horizontal picture of video signal.
oTV-V
This setting is used when observing the entire

Pull
at time of

8

SLOPE

vertical picture of video signal.
INote I Both setting will be combinated with TV
position on A TRIG MODE switch.

31

'*' Explanation

@
of trigger level LEVEL.

A TRIG MODE
o AUTO

The instrument is brought into automatically triggering sweep in which sweep is
always conducted.
In the presence of trig(. 
,
>

IIF.NABLE

~[NGLE

I+5V)

SCHEMATIC DIAGRAM ~
[TRIGGER (2/2))

43

n.

IOK
,lln6
lS~015

B~WP('Xlr.

~12\

R5JO
I.J2~lF

I 2\\

f

."
'"
'"
,fg

fiG

"'

11.SKD

RI

1.-lBW

2

I

R

SCHEMATIC DIAGRAM ~

44

[SWEEP GENERATOR]
_m'

O

POSITION

I

Pl'LL X10

PEF-501
(RV-2)

SCHEMATIC DIAGRAM ~
[CALIBRATOR & FRONT PANEL CONTROLS)

45

I~'''~I~---~-------

tJ\'l':lm

:~~;:

SCHEMATIC DIAGRAM~

46

[HORIZONTAL AMPLIFIER,
Z-AXIS AMPLIFIER
& HIGH VOLTAGE CIRCUITS]

JPI107

+lOOV'------I<< ,

I
TI401

)

240V

2<'

220V

2~ '"
120

120V

~i

lOOV

(RV 1301.il1J1

'00

+""----_« 2
«3

J

+l2V' _ _ _ _ _

,oP1OOl

«.
-12"'----~«

;

?
~

~
~
~
"'

<3

HZV

~
.;

GND

RIl2S
IOKF

<;

-12V

')

LJNETRIG

+Cii32

41/16V

SCHEMATIC OIAGRA M~
(POWER SUPPL YJ

(

I

PEF-4!l9!-

47

PEF -495

(\,ERTIC~L PRE~Mf'LlFIER

&. \ ERTICAL

CH~~~EL

SWITCH:

Pll06

6

'f'O\\'ER>,I'PPIY& HORIZO.\TAi

l:kAO

I

Fl~LlER
~357

PEF-5m
:RV-2J

Ll4Ql

~
1.1402

~

DIAGRAM~

1.1403

~

48

PEF531

(CONNECTORS]

(2)

50

Rear panel

11.
(1)

Panel controls
Front panel

49



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : No
Create Date                     : 2004:04:22 11:53:48+03:00
Modify Date                     : 2012:03:24 17:39:40+02:00
Has XFA                         : No
XMP Toolkit                     : Adobe XMP Core 4.2.1-c041 52.342996, 2008/05/07-20:48:00
Producer                        : Adobe Acrobat 9.0 Paper Capture Plug-in
Metadata Date                   : 2012:03:24 17:39:40+02:00
Creator Tool                    : Acrobat 5.0 Scan Plug-in for Windows
Format                          : application/pdf
Title                           : Hitachi V1050F (V-1050F) Oscilloscope Operating Guide - Operation Manual - schematic diagram
Creator                         : 
Document ID                     : uuid:2a61a7aa-a98f-413d-aa9a-db137c27b2af
Instance ID                     : uuid:d031821d-66ba-40e9-8d7d-cf42e0075aa0
Web Statement                   : http://vikiwat.com
Page Count                      : 52
Warning                         : [Minor] Ignored duplicate Info dictionary
EXIF Metadata provided by EXIF.tools

Navigation menu