Hitachi V1050F (V 1050F) Oscilloscope Operating Guide Operation Manual Schematic Diagram H_V1050 H V1050
User Manual: H_V1050
Open the PDF directly: View PDF .
Page Count: 52
Download | |
Open PDF In Browser | View PDF |
Model V-1050 OSCI.... LLOSCOPE , . OPERATION rYlANUAL Contents 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . composition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . How to produce the bright line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Method for Connecting Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ( 1) When us ing a probe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (2) When a di rect connect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (3) Caution for using the measured signal as an external trigger signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Measuring Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (1) DC voltage measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (2) AC voltage measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (3) Period measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (4) Measurement of time difference . . . . . . . . . . . . . . . . . . . . . , . . . . . . . . . . . . . (5) Measurement of rise (fall) time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (6) Measurement of single-shot signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (7) Trigger of complexed waveform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (8) Phase shift between two signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (9) Measurement by X-Y operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (10) How to use TV exclusive synchronization . . . . . . . . . . . . . . . . . . . . . . . . . . (11) Operating procedure of delayed sweep ., . . . . . . . . . . . . . . . . . . . . . . . . . . . (12) How to use TRIG VIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Panel Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ( 1) Power Supply and CRT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (2) Vertical deflection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (3) Horizontal defleqtion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (4) A triggering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (5) B triggering . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (6) Miscellaneous . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Users Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Schematic Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Panel Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1 2 5 8 8 9 10 11 11 12 12 13 14 14 15 15 16 18 20 24 25 25 26 28 30 33 33 35 39 40 49 I. Features (7) Trigger view: Hitachi Model V-1050 oscilloscope is a Displays the selected internal or external portable-type, advanced-class oscilloscope with trigger signal. a bandwidth of DC to 100 MHz Designed by putting signal is selected, you can correlate the If an external trigger special emphasis on operability and ruggedness, time between the trigger signal and input this oscilloscope has the following features: signals. (1) Wide bandwidth: The instrument has a bandwidth from DC to 100 MHz 2. Composition (2) High sensitivity: Sensitivity is 500 (8)Auto focusing Focusing shift is automatically corrected. Composition of Model V-1050 oscilloscope is ~V/div. (3) Large 6" screen: Employment of a large square CRT makes waveforms easier to observe. (4) Internal graticule: Employment of internal graticule CRT permits waveforms observation to be made without parallax error. (5) TV synchronization: as follows. (1) Model V-1050 oscilloscope unit ............ 1 (2) Probe (AT-10AE 1.5) . . . . . . . . . . . . . . . . . . . . . . . 2 (3) Fuse (2A for 100V and l20V set or lA for 220 and 240V set) ......... l (4) Dust proof cover . . . . . . . . . . . . . . . . . . . . . . . . . . 1 (5) Sc.rew driver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 (6) Power supply cord . . . . . . . . . . . . . . . . . . . . . . . . . l (7) Operation ManuaL . . . . . . . . . . . . . . . . . . . . . . . . . 1 Employment of a new TV sync/separator circuit allows instrument to observe TV signals stably. (6) Delayed sweep. A portion of the signal can be magnified before observation. 1 3. Precautions such as a vase on the oscilloscope. Precautions to be observed to lengthen the * service life of this instrument. Do not place the instrument in a place where vibration is strong. Avoid using the instrument at a place vibrating violently. Since the oscillo- C o. * scope is a precision instrument, excessively Avoide installing instrument in an extremely hot or cold place. ..... strong vibrations may cause damage. * o Avoid placing this instrument in a place "-'" Do not place the instrument near a magnet or magnetic body. An oscilloscope is an equipment exposed to sunlight for a long period of time, using electron beam. in a closed car in midsummer, or near a room Therefore, do not bring a magnet close to the instrument or do not use the heating device such as a stove. instrument near an equipment generating strong o The operating maximum ambient temperature is magnetic force. 50°C. * Do not use instrument that has been left outdoors on a cold winter day. The operating ambient tem- * Do not put a heavy objects on the oscilloscope. Do not block the ventillation holes. * * Do not apply a heavy shock to the oscilloscope. Do not insert a wire, pin, etc. through the * Do not drag the set, leaving the probe attached * Do not leave a hot soldering iron on the cabinet perature is O°C or more. * Avoid moving the instrument rapidly from a hot place to a cold place of vice versa, or condensation may form on inside of the instrument. * ventillation hole. Keep the instrument away from damp air, water, and dust. Unexpected trouble may be cau~ed when the instrument is placed in a damp or dusty place. The operating ambient humidity is 35-85%. Since an accidental instrusion of water may also cause troubles, .do not place a water-filled containers 2 to it. or the screen. * Do not try to turn the instrument upside down. Otherwise, knobs may be broken. * Do not use the instrument upright, leaving BNC C:,' cable connected to EXT BLANKING, GATE OUT PUT terminals on the rear panel. NOTE: When opening the case, disconnect the power Otherwise, supply plug beforehand without fail. the cable may be damaged. When cleaning the inside, insure beforehand that no electricity remains in the condensers of the power supply circuit. Recheck the operating procedure and if problem persists, contact a nearly service station or * Cleaning of CRT Dirty surface of CRT screen tends to cause measur- agent. ing errors. Remove the stains on CRT and filter by using a * Removal of stain from the case. clean and soft cloth, paying attention not to o When the outside of the case is stained, impair them. When the stain is extremely heavy, wash them with remove the stain by first wiping it lightly with a cloth moistened with neutral washing neutral washing agent and then leave them stand agent and then wipe the surface with a dry until the moisture is remov-ed naturally. cloth. o If the screen is installed while it is moistened, water rings may be formed and the waveform may o When the panel surface is stained, remove the stain in similar way with a clean, soft cloth. be blurred to become hard to observe. Pay When heavy stains are present, first remove attention not to leave finger prints on it. the stains by wiping the surface lightly with a cloth moistened with diluted neutral washing agent and then wipe thoroughly with a dry cloth. o When dust has accumulated on the in side, --,:-) remove it by using a dry brush, or by using the * Check the line voltage. The operating voltage range of this oscilloscope is as shown below. Check the line voltage without fail before turning on the power switch. exhaust of a compressor or a vacuum cleaner. 3 Rating length.) Line Voltage (50/60Hz) AC100V AC 90V - llOV AC120V AC108V - l32V AC220V AC198V - 242V Shape (Diameter x length)mm AC2l6V - 264V AC240V Nominal volts ~5% JIS type name 2A 6.35~ x 31. 8 MF61NM250V 2A AC lA 6.35~ x 31.8 MF61NM250V lA AC Do not increase the brightness too much. In the case of normal shipment, the voltage Do not increase the brightness of the spot and selector will be set convenient for user up. trace too much. When it is intended to use the oscilloscope on volt ages other rating, voltage selector can be turned. (Rated are indicated on the volt~ges (' J at 400Hz. * * The standards if the fuses are as follows. Your eyes may be strained and the fluorescent surface of CRT may be burnt. * Do no apply an excessive voltage. The input withstand voltage of each in put con- rear panel of the oscilloscope.) nector and probe input is as follows. Use only specified fuses. a voltav,e higher than specified. Never apply In order to protect the circuit against overcurrent, a 2A (make use of ACIOOV or AC120V) or lA (make use of AC220V or AC240V) is used on the primary side of the power supply. When this fuse is below out, check thoroughly the cause, repair any faulty point present, and then replace with a. specified fuse. Do not try to use the fuse other than the specified ones. Otherwise, fault may be caused or danger may be invited. (Particularly, do not use a fuse different from the specified one in current capacity and in 4 INPUT direct When probe is used EXT TRIG INPUT EXT BLANKING 250V(DC + AC peak at 1 kHz) 500V(DC + AC peak at 1 kHz) 250V(DC + AC peak) 20V(DC + AC peak) I Calibration Interval 1 To maintain jnstrument accuracy, perform the calibration of the V-I050 at least every 1000 hours of operation, or every six months if used infrequently. (.•...::) ". 4. How to produce the bright line Unless handled erroneously, this instrument VAR HOLD OFF MIN (full c.c.w.) B TRIG MODE AUTO will never become faulty by ordinary operation. Set all the levers of the switches either to Before turning ON the POWER switch, insure the power supply voltage is within the range of l08V ~ the left side or to the upper side. After ending all the settings mentioned above, 132V for AC 120V set, 198V - 242V for AC 220V set, and 216V - 264V for AC 240V set. turn ON the POWER. and, 15 seconds later, rotate Refer to the indication on the rear panel of the the INTENSITY knob clockwise. instrument for other voltages. bright line will appear. If observation is to be started immediately, Connect the plug of the power cord on the rear panel into the power supply wall socket and set the FOC(TS knob a.t a point where the bright set the knobs as follows. line is sharpest. POWER/INTENSITY If the instrument is not used with the power IOFFI th~ Midrange supply turned on rotate AC-GND-DC GND clockwise to reduce the brightness and also blur HPOSITION Midrange FOC(TS V.MODE CHI BW LIMIT 20MHz Full bandwidth (button out) D1SPLAY A INTENSITY counter- the FOCUS. INOTEI For usual observation, leave the following non-calibrating function section set to "CAL" Midrange position. A TRIG MODE AUTO VARIABLE A TRIG COUPING AC In this case the VOLTS/DIY is A TRIG SOURCE CHI calibrated to its indicating value. A TIME/DIV O.5ms/DIV B TIME/DIV Arbitrary ~ -+- <') Then the sweep POSITION PULL SWP VAR Rotate in the direction of arrow. Push in the knob or rotate in the direction of arrow. In this case the TIME/DIV is calibrated to its indicating value. 5 Align the bright line with the horizontal line appears even when no signal is present or scale line at the center of the screen by operat- when AC-GND-DC switch js at GNP position. ing CHI POSITION. means that the measurement of DC voltage can be In some cases the bright line This may be oblique to the scale slightly by the measured. effect of earth magnetism. when observing low frequency signals of about In this case, bring the bright line untjl it lies on the horizontal scale line at the center of the screen by properly adjusting the semi-fixed variable resistor TRACE ROTATION on the front panel. The following switching is needed 30 Hz or less. NORM A TRIG MODE DC A TRIG COUPLING - General measurement - Triggering can be effected by operating LEVEL (1) In the case of observing a single waveform knob under this setting. Use CHI or CH2 when not observing the phase difference between two waveforms or when engag- settings. ing in a operation other than X-Y operation. MODE Switch of Vertical Axis CH2 Make the following settings when using CHI. A TRIG SOURCE CH2 MODE Switch of Vertical Axis CHI MODE Switch of TRIG AUTO TRIG SOURCE CHI AC-GND-DC AC or DC Under these settings almost all the repetitive signals of about 30 Hz or more applied to CHI can be triggered and observed by adjusting A TRIG LEVEL. (When, the range of TIME/DIV is between 2mS/DIV and 20ns/DIV) Sjnce the MODE of horizontal axis is at AUTO position, the bright 6 When using only CH2, use the instrument after making the following (~. (2) When observing two waveforms .. ) (3) When observing three or four waveforms. Observation of two waveforms can be made easily This instrument can be observed the triggering by setting the MODE switch of vertical axis to signals (CH3 as A TRIG, CH4 as B TRIG) at ALT or CHOP. When observing two waveforms of depressed position of TRIG VIEW . high repetition frequencies set the MODE switch The table shows the relation between setting to ALT and, in the case of low frequencies, set of vertical mode and traces on CRT. it to CHOP. Normally, ALT and CHOP are recom- mended by selection of TIME/DIV. V. MODE CHI ALT; 0.2ms/DIV 20ns/DIV CHOP; 0.5ms/DIV 0.5s/DIV No. of display CHl, CH3 CH2 CH2, CH3 ALT CHI, CH2, CH3, CH4 When measuring the phase difference, measure CHOP CHI, CH2, CH3, CH4 after effecting triggering with leading phase ADD CHI + CH2, CH3, CH4 signal. The displaying position of CH3 (A TRIG) is located near the 100%'s sub-graduation, and possible to adjust by A TRIG LEVEL knob. CH4 is fixed at O%'s sub-graduation. 7 5. Method for Connecting Signals (4) When observing waveform with X-Y Set the MODE switch of vertical axis to CH2 (X-Y) and DISPLAY switch to X-Y. Then the instrument works as an X-Y oscilloscope. Each input is applied to the instrument as The first step of measurement is introduce the signal desired to measure to the oscilloscope properly. Do it with utmost care. (1) When using a probe follows. Use the attached probe, AT-10AE 1.5, when X-axis signal measuring a high frequency wave with high (horizontal axis signal) CHl INPUT accuracy. CH2 INPUT input signal is attenuated by this probe to Y-axis signal It should be noted, however, that since the (vertical axis signal) In this case leave the horizontal axis magnification switch (PULL x 10MAG inner shaft knob) at depressed position, and the AC-GND-DC of X-axis (CH1) to AC, also TRIG VIEW and BW LIMIT 20MHz are undepressed position (button out). 1/10 before it is input to the oscilloscope the use of the probe is disadvantageous for low signals, and that at the same time the measuring range is extended by that amount for high signals.U -"'Lt . - - -- - - - 10%~~-+-rY-+-~hcl-+-L '0 --\ - - - -- - - - --- 1 O%~~-+-r~+-r-~~ Zero level (reference line) (3) Period measurement This will be explained taking the drawing at follows as an example (2) AC voltage measurement The same as paragraph 6 (1), "DC voltage mea- 100% 90% -- -- surement", but here those is no need of matching the zero level with the scale line. Move the zero level at will to a position easy to - -- - - -- - - - - - 111 ill /\ 11\ /\ J ilL i I \ I \ I \ IJ I \ \1 10% 0% -- - - - - - - \1 j - - - - - observe. In the drawing as follows, VOLTSjDIV is IVjDIV, IVjDIV x 5 = 5 Vp-p [50 Vp-p at using the probe AT-IO'AE 1. 5( 10: 1) 1. When magnifying and observ- ing a small-amplitude signal superinposing as increases, set AC-GND-DC switch to AC. voltage is cut off and AC voltage can be observed by increasing sensitivity. 12 The DC Time A Time B One period covers the time A and time B, which are separated from each other by 2.0 DIV on the CRT. When the sweep time is ImsjDIV, the period is given by 1 msjDIV x 2.0 2.0 ms 2.0 x 10-3 s ( .) This means that CHI is used as the triggering Accordingly, the frequency is 1/(2.0 x 10- 3 ) = 500 Hz signal when investigating the length of time (However, when the knob MAG x 10 is at pulled by which the signal of CH2 is delayed from the out position, TIME/DIV must be converted to signal of CHI. 1/10 since the sweep is magnified.) case. (4) Measurement of time difference CH2 is used in the reversed In other words, the signal leading in phase is selected as the triggering signal Triggering signal source "SOURCE" is selected source. as offering reference signal when measuring the If this process is reversed, the portion to be time difference between two signals. measured may sometimes not appear on the screen. that pulse trains as shown in (a). Assume Then (b) shows the case when CHI is taken as the trig- Thereafter, equalize the amplitudes of the two signals appearing on the screen or superimpose gering signal source and (c) the case where one another. CH2 is taken. Read the time difference by the interval CHI between 50% amplitude points of the two signals. ': ~ I Sometimes the superimposing method is more con- I 1Pu-u--u-L venient from the point of view of procedure. I ,I , caz, I ) ,, (a) l CHI CH I ", ,'-+------,, I I I - '" Time difference " CHZ '\ - -' rn.eft difference Equalize ~ ~ amplitudes o CHZ _____ J.::::-V by VAR ! Equal amplitude measuring method (b) Superposition measuring method (c) 13 The true rise time is given by {Cautions ') Since the pulsed wave contains many high- Trx = ~ Tro2 - Trs2 frequency wave components (higher harmonics) depending on its width or period, pay the same attention as given to high frequency signals when handling it. Accordingly, use a probe or coaxial cable and shorten the earth lead wire as much as Moreover, in general, in a circuit free from waveform distortion such the following relationship is established between frequency band and rise time. possible. fc (5) Measurement of rise (fall) time where, fc To measure the rise time pay attention not only to the above mentioned items but also to measurement error. / a~ overshoot and sag, (' tr x tr = 0.35 Frequency band (Hz) Rise time (s) The rise time and fall time are determined by the time elapsed between the 10% to 90% values The following relationship exists between the rise time Trx of the waveform to be measured, the rise time Trs of oscilloscope, and the rise time Tro displayed on the screen. of pulse width. (6) Measurement of single-shot signal Single sweep is conveniently used in the measurement and photography of single-shot signal, waveform of remarkably non-uniform repetition (such When the rise time of the pulse going to be as impulse waves, sound waves, switch noise measured is sufficiently longer than the rise waves). time of the oscilloscope (3.5ns in our case), the effect of the rise time of the oscilloscope on the measurement can be neglected. However, if both are close to each other, measurement error may be caused. 14 First set A TRIG MODE on the front panel to NORM and the effect synchronization by using a signal or repetitive waveform of about the same level and by rotating LEVEL. ('" Next, set A TRIG MODE to SINGLE and depress RESET button to insure that sweep is made on and only one time. Then remove the vertical axis input signal (by, for example, setting AC-GNDDC switch to GND) , depress SINGLE RESET and , "') ~,;-, ;-. insure that READY LAMP goes on. B/\ H " Nf'\ A7Ci0c;d}i~ Y' Trigger level Y setting line Ca) Signal waveform When apply the input signal, sweep is made for one time and READY LAMP goes out. Since sweep is also made at no signal time depending on the level, do not rotate LEVEL once SINGLE RESET is completed. (c) When the trigger setting level is y' Triggering of complexed waveform (7) Triggering of complexed waveform In the case shown in the Fig. and advances to C, E, F,····· and which allows (a) below where waveforms greatly different in amplitude alternate, the waveform is doubled if the trigger level is not set properly. In the case where the trigger level is selected as Y line two waveforms, one starting with A and advancing to B, C, D, E, F,····· and the other starting with E and advancing to F, G, H, I ..... , will appear alternately on the screen. They will be doubled as shown in Fig. (b) When the trigger setting level is Y (b), for which no triggering can be taken. (8) Phase shift between two signals To measure a phase shift between two signals of the same frequency, the dual trace display system can be utilized up to the upper limit frequency of the amplifier. First, position both signals on the cent er line of the scale, e.g. just 4 div, as shown in the following figure by means of the VARIABLE and horizontal positioning knobs. Next, set the distance where the center of the In such a case, rotate LEVEL clockwise until the trigger level comes to Y' line. synchronization. Then the waveform of the base channel intersects with that of the scale to 8 div horizontally. waveform on the screen becomes the one is shown in Fig. (c) follow which start with B 15 (9) Measurement by X-Y operation The phase shift between two signals of the same CHI (Bas e) ~ CH2 J frequency can also be measured using a V. I/V /~/ "~ ~ -I 'l \ 1,1. \. \1\ ~\. '/ '/ Lissajou's figure by X-Y operation. A sine wave input is applied to the audio circuit being tested. The same sine wave input is applied to the vertical input of the oscilloscope, and the output of the tested circuit is applied to the horizontal input of the oscil- -----'> __ Phase di ffe ~nce 0.7 div 8dlv (360") ------') loscope. The amount of phase difference between the two signals can be calculated from the resulting wave form. As shown in the above figure, set 1 cycle, 360° to 8 div. 360° Then, 8div = 45°/div. Accordingly, the phase difference in the above example can be calculated as follows: 1. Using an audio signal generator with a pure sinusoidal signal, apply a sine wave test signal at the desired test frequency to the audio network being tested. 2. Set the signal generator output for the normal Horizontal distance on the screen:0.7 div Phase difference = 45°/DIV x 0.7 div = 31.5° If desired, the circuit's output may be observed If the portion of the phase difference is much on the oscilloscope. smaller, use the MAGNIFIER at the xlO position over-driven, the sine wave display on the oscil- in the above setting. At this time, 360° is displayed in 8 div xlO. 360° Then, 8 divxlO = 4.5°/DIV (0.2 div=0.9°) 16 operating level of the circuit being tested. If the test circuit is loscope is clipped and the signal level must be reduced. 3. Connect the Channel 2 probe to the output of the test circuit. ~:'.' 4. Set the DISPLAY to X-Y position. 5. Connect the Channel 1 INPUT probe to the input SINE of the test circuit. (The input and output test B ~ A connections to the vertical and horizontal WHERE oscilloscope inputs may be reversed.) PHASE 6. Adjust the Channel 1 and 2 gain controls for a ANGLE suitable viewing size. 7. Some typical results are shown in Fig.(b). If the two signals are in phase, the oscilloscope trace is a straight diagonal line. If the ver- tical and horizontal gains are properly adjusted, this line is at a 45° angle. A 90° phase shift produces a circular oscilloscope pattern. Phase shift of less (or more) than 90° produces an elliptical oscilloscope pattern. ~ The amount of phase shift can be calculated from the oscilloscope trace as shown Fig. (a) Phase shift calculation / ~ NO AMPLITUDE DISTORTION NO PHASE SHIFT AMPLITUDE DISTORTION NO PHASE SHIFT ~ 0 180 0 OUT OF PHASE in Fig.(a). 0 NO AMPLITUDE DISTORTION PHASE SHIFT 0 AMPLITUDE DISTORTION 90° OUT OF PHASE PHASE SHIFT Flg. (b) Typical phase measurement oscilloscope displays. 17 (10) How to use TV exclusive synchronization Cl) Difference in the circuits On the image waveform of TV Exclusive circuit for conventional oscillograph l SiV~ EXCIU.. (~rinciple f - - - - - . - - - - - - , - - - - - - - - - - . __ General circlllt i-,iffiPl€ sync~r0111z1ng I ;t t.": . circuit for this. lDstrument drawing) • TV exclusive synchronizing Clrcult separator circuit -------4------------------Video slgnal si Synchronizing signal pulse (SYNC pulse) To trigger circuit I To trigger circuit I ~ ~----------------lV(vertical) ." ~I " ~I~ In the work concerned with TV, complexed signals containing video signal, blanking pedestal I " u I I I SYNC Peak Detector signal, and synchronizing signal are often However, since the waveform is commeasured. plexed, a special circuit is needed to effect a stable triggering with vertical waveform. I SLOPE ! II Hard to synchronize I Synchronization is more I because video signal I easily effected than in I is applied directly ,the circuit shown at i left, because the signal I as trigger signal. I i~ integrated to remove I L-...L_ _ _ _ _ _ _ _ hlgh frequency com- I pOTIents. I triggering~ de;:~ted I Stable obtained, after the peak of SYNC pulse I then the trigger level in feed back roap of operational amplifier should be fixed its SYNC tip. ( 18 I .. , ® Operation To observe vertical signal I1 I I I1 I1 To observe horizontal signal ---, ijm millIII III"I IU A TRIG MODE: TV COUPLING: TV-V LEVEL: PULL TV(-) TIME: O.lms/DIV O.2S/DIV Will IIlllll I I A TRIG MODE: TV COUPLING: TV-H LEVEL: PULL TV(-) TIME: 50~s/DIV 20ns/DIV Generaly, video signal of TV has negative synchronizing signal. Use at pull out position which is A TRIG LEVEL knob. If the video signal is inverted then A TRIG LEVEL is depressed. signal ',) 4-synchronizing signal Example of Example of (-) synchroniging signal (+) synchronizing signal If the sync and blanking pulses are positive, set the switch to TV(+). 19 (11 ) Operating procedure of delayed sweep INTEN A TIME/DIV Choose A TIME/DIV properly There are two kinds of time delay sweep; one is B TIME/DIV Set B TIME/DIV at a more rapid AUTO time delay sweep (continuous time delay sweep time than the one set by sweep) and the other TRIG time delay sweep (trig- A TIME/DIV gering time delay sweep). These are selected by two, the INT (CHI, CH2 internal triggering time delay sweep) and EXT (external triggering time delay sweep). mode. Usually, the instrument is used in AUTO Although the. AUTO time delay sweep is easy to operate the maximum magnification factor is limited by delayed jitter (rolling) (to a few hundred times). On the other hand, since no jitter is generated in TRIG time delay sweep, this sweep has the feature of being enabled to increase the magnification factor. However, the magnification factor is limited by the brightness of CRT (to a few thousand times). At time of AUTO (continuous time delay sweep) Effect triggering by A sweep and set the knobs as follows. Then the high brightness portion of A sweep will appear without fail (if not, adjust INTENSITY). TRIG time delay sweep is further classified into 20 AUTO DISPLAY MODE switch of B TRIG. (i) B TRIG MODE (Used to magnify and observe any portion of a complexed waveform in horizontal direction. ) Rotate DLY TIME POSITION knob (the cent er FINE is for fine adjustment). The high brightness portion will move continuously. Bring this high brightness portion to the position desired to be magnified, switch DISPLAY to B. Then the high brightness portion is magnified to occupy the full area of the screen. The sweep time is the indicated value of B TIME/DIV. i-----:::==========t-Bright portion A INTEN ALT B~ ) (~ fail. state.) (This state is called the B-triggered Rotate DLY TIME POSITION knob. The high brightness portion will move continuously. Further rotation will bring this portion to the next "peak". Therefore, bring this high bright portion to a position desired to be magnified and then set DISPLAY at B. The high brightness portion is magnified to occupy the full area of the screen. A INTEN ALT Blx-vl ~) ( (D The sweep time for this case is the value indicated by B TIME/DIV. The gen- locking is fixed at AC. A INTEN B X-Y At time of TRIG (TRIGGERING time delay sweep) Effect TRIGGERING with A sweep and set the knobs c_----""'~ __~) as follows. B TRIG MODE TRIG (INT or EXT) DISPLAY INTEN A TIME/DIV Chose A TIME/DIV properly. B TIME/DIV Set B TIME/DIV at a more rapid sweep time than the one set by A , ,: ) TIME/DIV. / SLOPE Set SLOPE either to + or to - Thereafter, rotate LEVEL. The high bright- ness portion of A sweep will appear without A INTEN B X - Y (-----=~=----~) * The case in which SLOPE is + is shown 21 ® ALT sweep This ALT sweep is displayed A sweep and B (delayed) sweep alternatively. (d) Only displayed at intensive part Belowing figures are showing when CHI to be put triangle waveform f. A IETEN ALT B X-Y and CH2 to be put rectangular waveform. ( (a) Normal observation lI#J ) A INTEN ALT B X-Y (1)lI ) (b) Prepared to magnify When DISPLAY switch is set the ALT, displaying signal of B sweep can be control the position CHI A sweep A INTEN (lI#J ALT to easily observe by TRACE SEP. The range of position be shift to upper and lower three B X-Y ) CH2 A sweep divisions compere the A sweep. Then depress the TRIG VIEW switch, 4 traces come to screen in totally 8 traces to observe. (c) (b) and intensive display in alternatively sweep These new 4 traces are A and B sweep of CH3 and CH4. I7"I-AAA.q,....,.~ -- '- f- ~CH3 Input Signal ~CHl Input Signal ~CH2 Input Signal ~CH4 Input Signal f '. '~'..'. ,) ;1 7. Panel Descriptions (1) Power Supply and CRT Operating voltage and fuse This model can be operated from either a 100 :) volt, a 120-volt, a 220-volt or a 240-volt nominal line voltage source. The Line Voltage Selector assembly on the rear panel converts the instrument from one operating range to the other. In addition, this assembly changes the primary connections of the power transformer to allow selection of one of four regulating ranges. also includes the line fuse. The assembly Use the following procedure to convert this instrument between C!) nominal line voltages or regulating ranges. brightness of the CRT display. 1. Disconnect the instrument from the power source. 2. To convert from lOO-volts nominal to 220-volts nominal line voltage or vice versa, pull out the Voltage Selector switch turn it, and plug it back into the remaining holes. Change the ,,") 3. Before apply power to the instrument, check that the indicating tabs on the rear panel. Clockwise adjust- ment increases brightness. @) ® line-cord power plug to match the power-source receptacle. POWER/INTENSITY Turns the power on or off and controls the Pilot lamp Lights when the instrument power is on. FOCUS Provides adjustment for optimum display definition. @ TRACE ROTATION Corrects slight tilting of trace caused by external magnetic fields. 25 ® @ SCALE ILLUM Controls graticule illumination. area, photographing. the input signal fed to CHI. CHI OR [K] On X-Y operation, the kaob functions to change the.sensitivity of the X-axis. (CH2 or [YJ sensitivity switch) This is a knob for switching the sensitivity of INPUT This is an input plug for use with the CHI the input signal fed to CH2. vertical amplifier and X-axis (horizontal axis) is accomplished in ten steps from 5mV/DIV to 5V/ amplifier during X-Y operation. DIV. CH2 OR [YJ INPUT This is an input plug for use with the CH2 change the sensitivity of the Y-axis. Switching action On X-Y operation, the knob functions to To mea- sure by the use of the indicated voltage sensi- vertical amplifier and Y-axis (vertical axis) tivity, be sure to set each of the VARIABLE to amplifier during X-Y operation. CAL position by turning fully clockwise. AC-GND-DC (Alternating Current-Ground Switch-Direct signal is applied to the input terminal by the are ten times the indicated voltage. Switches the coupling of the signal fed to the vertical axis input. the DC position. If the use of a 1/10 low capacitance probe, the values Current) 26 Switching action is accomplished in ten steps from 5mV/DIV to 5V/ DIV. dently of display portion or applied signals. (2) Vertical deflection ® sensitivity switch) This is a knob for switching the sensitivity of Compress display within graticule area, indepen- ® 0 illuminate the graticule when viewing in a dark ~ TRACE FINDER (j) VOLTS/DIY (CHI or Useful to DC coupling is obtained on On AC position, the direct @ VARIABLE This is a vertical axis sensitivity fine adjustment which is capable of attenuating to less current component is blocked by a capacitor. than 1/2.5 by indication of each range of VOLTS/ The GND position grounds the input of the ampli- DIV. fiers and opens the input terminal sensitivity indicated by VOLTS/DIY, turn the To measure a voltage by the use of voltage c VARIABLE clockwise fully to CAL. @ ) a .. Put a high level of signal sinchronized with POSITION, PULL xlO MAG the observing signal into the A EXT TRIG CHI # (Vertical position adjustment) With the knob turned clockwise, the waveforms trigger signal source. of CH-l move upward. terminal or CH2 terminal (when using CHI) as When the knob is turned counterclockwise, the waveforms move downward. b. CH2 it (Vertical position adjustment) Clockwise rotation will move pattern up, and c. In case of observing signals over 4kHz, select "ALT" of vertical MODE. counterclockwise rotation will move pattern down When the knob is pulled, the vertical axis sensitivity at each range of VOLTS/DIY is In case of observing low frequency signals under 4kHz, select "HF REJ" of A TRIG COUPLING. ~ MODE o CHI Only the input signal applied to CHI is increased by 10 times. displayed. INote I 1. When measuring at 5 mY/DIY o CH2, I X-YI ~ 5 V/DIY range, use the xl (PULL xlO MAG switch is depressed posi t ion). 2. Otherwise the S /N and frequency- Only the input signal applied to CH2 is displayed. o ALT bandwidth of instrument will be decreased. CHI and CH2 signals are displayed alternately In case of the observation of signals in the on consecutive sweeps. high-sensitivity range of 500~V to 2mV/DIV and o CHOP in the CHOP MODE, it is sometimes difficult to CHI and CH2 signals are displayed simultaneously get stable synchronization due to noises caused by switching between channels at about 250 kHz rate. by the high amplification of vertical amplifier. In these cases: 27 o ADD (3) Horizontal deflection Displays the algebraic sum of the channel I and channel 2 input signals. If the channel 2 dis- play is inverted (press CH2 INV), an CHI minus CH2 display results. @ BW @ DISPLAY This switch is used to select the operation mode of the horizontal axis. o A The BW LIMIT 20MHz switch provides a method of o INTEN Although the sweep on the screen is A sweep it indicates B sweep (delay time quency signals when viewing low-frequency signals. sweep) by intensity modulation. o ALT Signals applied respectively to A sweep deflection system is limited to about 20MHz. and B sweep appear on the screen alter- Reduces the bandwidth of channel I and channel natively each sweep. 2 to approximately 20MHz. o B INV Inverts the polarity of the channel 2. occupy the full area of the screen. The sweep time at this time is B. Useful to comparison of two signals of opposite polarities, and observation of a differential signal of CHI and CH2 along with ADD mode. TRIG VIEW This switch is used when desiring to observe the waveform triggering signal on the screen. The intensity modulated portion in INTEN mentioned above is magnified to @ CH2 o X-Y This position is used when using the instrument as an X-Y oscilloscope. X direction signal is input to CHI and Y direction signal to CH2. The vertical deflection sensitivity at this time is It is also used when measuring the phase dif- read on CH2 VOLTS/DIV and horizontal ference between the synchronizing signal and axis sensitivity on CHI VOLTS/DIV. other input signal. Vertical position is set by CH2 POSITION t ion. ) 28 This reducing interference from unwanted high-freWhen set it -3dB bandwidth point of the vertical @ A sweep appears on the screen. setting is used in normal cases. LIMIT 20MHz (quad phenomena observa- and horizontal position by : POSITION. @ words. TRACE SEP B sweep of the position adjustment knob when Normally, leave this knob rotated to MIN. ALT sweep mode. L';) INot;~e instrument has a special control system to easy observe at high speed sweep range when DISPLAY is set to ALT or B. Therefore, the traces is still remained when decrease the INTENSITY to minimum. ~ A TIME/DIV (Sweep speed selection) The outer knob controls the A (Main) sweep rate, which has from 20 ns/DIV to 0.5 s/DIV selects 23 fixed sweep speeds. @lB TIME/DIV The inner knob control the B (Delayed)-Sweep rate. @ PULL SWP VAR Provides continuous adjustment of A sweep TIME/ DIV between calibrated positions, when the inner shaft is at pull. Counterclockwise rota- tion to the full delays the sweep by 2.5 times or more. Normally, the inner shaft is left depressed. VAR HOLD OFF Increases the time between sweeps and aids triggering on complex displays such as high-frequency signal, Rotate the VAR HOLD OFF slightly to obtain a stabilized triggering. irre~ular signal and digital @ FINE PULL xlO MAG POSITION This knob is used to adjust position of the display horizontally. It is indespensable in the measurement of the time of waveform. Display is moved toward right when the knob is rotated clockwise and toward left with counterclockwise rotation. The outer shaft is for course adjustment and inner shaft for fine adjustment. A and B sweep are magnified 10 times by pulling out FINE knob (inner shaft) of POSITION. In this case the sweep time is 1/10 of the value indicated by TIME/DIV. Bring the portion of the waveform desired to be magnified observed to the outer of the scale by operating POSITION of the horizontal axis. Next switch xlO MAG switch to PULL (pulled out state). Then the waveform placed at the cent er is magnified in right and left directions. The sweep time in this case is 10 times the sweep speed obtained by TIME/DIV, in other words, the reading is 1/10 of the sweep time indicated. 29 (4) A Triggering @ SOURCE This switch is used to select the triggering 100% 90% signal source A Sweep. - c·· Used when triggering is made by observing ( .. signal for CH2. \ / I1 o CH2 1\ I 100% 90% Used when triggering is made by observing signal for CHI \ 1/ o CHI o LINE Used when observing a signal triggered with the frequency of mains (AC POWER). o EXT Used when triggering is made by a signal applied to EXT input connector, indepen- / dently from observation signal. o EXTdO .. ...... Attenuates external TRIG INPUT signal by a I factor of 10. @TRIG INPUT (CH3) Input terminal for use for external triggering @DLY TIME MULT signal of A sweep. This control is used to set the delay time of @COUPLING B sweep starting point with respect to A sweep This switch is used to select the coupling mode starting point. of A sweep triggering signal. When the above mentioned DISPLAY is set in INTEN or B. o AC At this setting both the DC component and C') the very low frequency of triggering, signal are cut off. 30 @ oHF REJ A TRIG LEVEL Among the AC components of triggering signal, This knob is used to decide at which portion the high frequency components of about 4 kHz of the waveform should the sweep the started This knob is also by setting trigger level. or more are attenuated. A stabilized trigger- ing unaffected by noises of about 4 kHz or enabled to switch SLOPE. (normal state) is for + more can be obtained. oLF REJ Depressed position SLOPE and PULL position (state in which the knob is protrud- Among the AC components of the triggering ing) is for SLOPE. signal the low frequency components of less than about 4 kHz are attenuated. A stabilized *Explanation of trigger polarity SLOPE. triggering free from noises of less than about 4kHz can be obtained. oDC Push Triggering signal is amplified unchanged. This setting is used when triggering with a very low frequency signal or when effecting DC like triggering. at time of 8 SLOPE oTV-H This setting is used when observing the entire horizontal picture of video signal. oTV-V This setting is used when observing the entire Pull at time of 8 SLOPE vertical picture of video signal. INote I Both setting will be combinated with TV position on A TRIG MODE switch. 31 '*' Explanation @ of trigger level LEVEL. A TRIG MODE o AUTO The instrument is brought into automatically triggering sweep in which sweep is always conducted. In the presence of trig(. , > IIF.NABLE ~[NGLE I+5V) SCHEMATIC DIAGRAM ~ [TRIGGER (2/2)) 43 n. IOK ,lln6 lS~015 B~WP('Xlr. ~12\ R5JO I.J2~lF I 2\\ f ." '" '" ,fg fiG "' 11.SKD RI 1.-lBW 2 I R SCHEMATIC DIAGRAM ~ 44 [SWEEP GENERATOR] _m' O POSITION I Pl'LL X10 PEF-501 (RV-2) SCHEMATIC DIAGRAM ~ [CALIBRATOR & FRONT PANEL CONTROLS) 45 I~'''~I~---~------- tJ\'l':lm :~~;: SCHEMATIC DIAGRAM~ 46 [HORIZONTAL AMPLIFIER, Z-AXIS AMPLIFIER & HIGH VOLTAGE CIRCUITS] JPI107 +lOOV'------I<< , I TI401 ) 240V 2<' 220V 2~ '" 120 120V ~i lOOV (RV 1301.il1J1 '00 +""----_« 2 «3 J +l2V' _ _ _ _ _ ,oP1OOl «. -12"'----~« ; ? ~ ~ ~ ~ "' <3 HZV ~ .; GND RIl2S IOKF <; -12V ') LJNETRIG +Cii32 41/16V SCHEMATIC OIAGRA M~ (POWER SUPPL YJ ( I PEF-4!l9!- 47 PEF -495 (\,ERTIC~L PRE~Mf'LlFIER &. \ ERTICAL CH~~~EL SWITCH: Pll06 6 'f'O\\'ER>,I'PPIY& HORIZO.\TAi l:kAO I Fl~LlER ~357 PEF-5m :RV-2J Ll4Ql ~ 1.1402 ~ DIAGRAM~ 1.1403 ~ 48 PEF531 (CONNECTORS] (2) 50 Rear panel 11. (1) Panel controls Front panel 49
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.6 Linearized : No Create Date : 2004:04:22 11:53:48+03:00 Modify Date : 2012:03:24 17:39:40+02:00 Has XFA : No XMP Toolkit : Adobe XMP Core 4.2.1-c041 52.342996, 2008/05/07-20:48:00 Producer : Adobe Acrobat 9.0 Paper Capture Plug-in Metadata Date : 2012:03:24 17:39:40+02:00 Creator Tool : Acrobat 5.0 Scan Plug-in for Windows Format : application/pdf Title : Hitachi V1050F (V-1050F) Oscilloscope Operating Guide - Operation Manual - schematic diagram Creator : Document ID : uuid:2a61a7aa-a98f-413d-aa9a-db137c27b2af Instance ID : uuid:d031821d-66ba-40e9-8d7d-cf42e0075aa0 Web Statement : http://vikiwat.com Page Count : 52 Warning : [Minor] Ignored duplicate Info dictionaryEXIF Metadata provided by EXIF.tools