A Emerson 1066 Liquid Analytical Transmitter Liq Manual 51
User Manual: Emerson 1066 Liquid Analytical Transmitter 1066
Open the PDF directly: View PDF .
Page Count: 134
Download | |
Open PDF In Browser | View PDF |
Instruction Manual Rosemount™ 1066 Smart-Enabled, 2-Wire Transmitter LIQ-MAN-1066 Rev. J April 2017 Essential Instructions Read this page before proceeding Emerson designs, manufactures, and tests its Rosemount products to meet many national and international standards. Because these instruments are sophisticated technical products, you must properly install, use, and maintain them to ensure they continue to operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program when installing, using, and maintaining Rosemount products. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation. • Read all instructions prior to installing, operating, and servicing the product. If this Instruction Manual is not the correct manual, telephone 1-800-854-8257 and the requested manual will be provided. Save this Instruction Manual for future reference. • If you do not understand any of the instructions, contact your Emerson representative for clarification. • Follow all warnings, cautions, and instructions marked on and supplied with the product. • Inform and educate your personnel in the proper installation, operation, and maintenance of the product. • Install your equipment as specified in the Installation Instructions of the appropriate Instruction Manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources. • To ensure proper performance, use qualified personnel to install, operate, update, program, and maintain the product. • When replacement parts are required, ensure that qualified people use replacement parts specified by Rosemount. Unauthorized parts and procedures can affect the product’s performance and place the safe operation of your process at risk. Look alike substitutions may result in fire, electrical hazards, or improper operation. • Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury. WaRNING: ExplOSION hazaRD DO NOT OpEN WhIlE CIRCuIT IS lIvE. ONly ClEaN WITh DaMp ClOTh. NOTICE If a 475 universal haRT® Communicator is used with these transmitters, the software within the 475 may require modification. If a software modification is required, please contact your local Emerson Service Group or National Response Center at 1-800-654-7768. Electrostatic ignition hazard. Special condition for safe use (when installed in hazardous area) 1. The plastic enclosure, excepting the front panel, must only be cleaned with a damp cloth. The surface resistivity of the non-metallic enclosure materials is greater than one gigaohm. Care must be taken to avoid electrostatic charge build-up. The 1066 Transmitter must not be rubbed or cleaned with solvents or a dry cloth. 2. The panel mount gasket has not been tested for type of protection IP66 or Class II and III. Type of protection IP66 and Class II, III refer the enclosure only. Essential Instructions I 3. The surface resistivity of the non-metallic enclosure materials is greater than one gigaohm. Care must be taken to avoid electrostatic charge build-up. The Model 1066 Transmitter must not be rubbed or cleaned with solvents or a dry cloth. 4. Special Condition of Use of 1066 C FF/FII5 and 1066T FF/FII5. For use with simple apparatus model series 140, 141, 142, 150, 400, 401, 402, 402VP, 403, 403VP, 404, and 410VP contacting conductivity sensors and model series 222, 225, 226, 228 toroidal sensors. About this document This manual contains instructions for installation and operation of the 1066 Smart Transmitter. The following list provides notes concerning all revisions of this document. Rev. level A Date 1/2012 B 3/2012 C 9/2012 This product manual version adds specifications and instrument instructions for Contacting Conductivity, Toroidal Conductivity, Chlorine, Oxygen, and Ozone measurements. This product manual version adds FM agency approval. D 3/2013 Updated CSA Intrinsically Safe Installation drawings. E 7/2013 F 9/2013 Updated CSA test Standards and Intrinsically Safe installation drawings and update CE certificates. Added FM temperature specifications to Non-Incendive Hazardous Location Approval. Added Section 10: HART® Communications G 11/2014 Changed agency water exposure testing description to “Type”. J 04/2017 Updated the Address and Emerson logo. Also, updated the FM, H II Notes This is the initial release of the product manual. The manual has been reformatted to reflect the Emerson documentation style and updated to reflect any changes in the product offering. 05/2015 FM approvals updated. CSA installation drawings and CE Declarations. Instruction Manual Table of Contents LIQ-MAN-1066 April 2017 Contents Section 1: Quick Start Guide 1.1 Quick start guide..........................................................................................................1 2.1 Features and Applications...........................................................................................3 2.3 pH/ORP ........................................................................................................................4 Section 2: Description and Specifications 2.2 2.4 2.5 2.6 Specifications - General ................................................................................................4 2.3.1 2.2.2 2.4.1 2.4.2 2.5.1 2.5.2 Recommended Sensors for Conductivity .......................................................8 Performance Specifications.............................................................................8 Recommended Sensors for Conductivity........................................................9 Chlorine (Codes - L)......................................................................................................9 2.6.1 Free and Total Chlorine....................................................................................9 2.6.3 Recommended Sensors ..................................................................................9 2.6.5 2.9 Performance Specifications.............................................................................7 Toroidal Conductivity (Codes - T) .................................................................................8 2.6.4 2.8 Performance Specifications - Transmitter (ORP input) ....................................6 Contacting Conductivity (Codes - C) ............................................................................7 2.6.2 2.7 Performance Specifications - Transmitter (pH input) ......................................6 2.6.6 Performance Specifications.............................................................................9 Monochloromine ............................................................................................9 Performance Specifications...........................................................................10 Recommended Sensors ................................................................................10 Dissolved Oxygen (Codes - DO) .................................................................................10 2.7.1 2.7.2 Performance Specification ............................................................................10 Recommended Sensors ................................................................................10 Dissolved Oxygen (Codes - DO)..................................................................................10 2.8.1 2.8.2 Performance Specification ............................................................................10 Recommended Sensors ................................................................................10 Ordering Information .................................................................................................11 Section 3: Installation 3.1 Unpacking and Inspection..........................................................................................13 3.3 Preparing Conduit Openings ......................................................................................13 3.2 Installation – General Information..............................................................................13 Section 4: Wiring 4.1 4.2 Table of Contents General ...................................................................................................................... 17 4.1.1 4.1.2 General Information ......................................................................................17 Digital Communication.................................................................................17 Power Supply/Current Loop – 1066 HT ......................................................................17 III Table of Contents Instruction Manual April 2017 LIQ-MAN-1066 4.2.1 Power Supply and Load Requirements ..........................................................17 4.2.3 Current Output Wiring ..................................................................................19 4.2.2 4.3 4.4 Power Supply-Current Loop Wiring...............................................................18 Power Supply Wiring For 1066 FF ...............................................................................20 4.3.1 Power Supply Wiring .....................................................................................20 Sensor Wiring to Main Board ......................................................................................21 Section 5: Intrinsically Safe Installation 5.1 All Intrinsically Safe Installations ................................................................................27 6.1 User Interface.............................................................................................................33 6.3 Main Display ...............................................................................................................34 Section 6: Display and operation 6.2 6.4 Instrument Keypad.....................................................................................................33 Menu System..............................................................................................................35 Section 7: programming – Basics 7.1 7.2 7.3 7.4 General.......................................................................................................................37 Changing the Startup Settings ...................................................................................37 7.2.1 7.2.2 7.6 7.7 7.8 Procedure......................................................................................................38 Choosing Temperature Units and Automatic/Manual Temperature Compensation ..38 7.3.1 Purpose .........................................................................................................38 7.4.1 Purpose .........................................................................................................38 7.4.3 Procedure: Configure Outputs ......................................................................38 Configuring and Ranging Current Outputs ................................................................38 7.4.2 7.5 Purpose .........................................................................................................37 7.4.4 Definitions.....................................................................................................38 Procedure: Ranging the Current Outputs......................................................38 Setting a Security Code ..............................................................................................38 7.5.1 7.5.2 Purpose .........................................................................................................39 Procedure......................................................................................................39 Security Access...........................................................................................................40 7.6.1 7.6.2 How the Security Code Works.......................................................................40 Procedure......................................................................................................40 Using Hold..................................................................................................................40 7.7.1 7.7.2 Purpose .........................................................................................................40 Using the Hold Function................................................................................40 Resetting Factory Default Settings .............................................................................41 7.8.1 7.8.2 Purpose .........................................................................................................41 Procedure......................................................................................................41 Section 8: programming – Measurements 8.1 IV Introduction ..............................................................................................................44 Table of Contents Instruction Manual Table of Contents LIQ-MAN-1066 April 2017 8.2 pH Measurement Programming ................................................................................44 8.2.1 Description....................................................................................................44 8.2.3 Preamp..........................................................................................................44 8.2.2 8.2.4 8.2.5 8.2.6 8.2.7 8.3 8.2.8 Solution Temperature Correction .................................................................45 Temperature Coefficient ...............................................................................45 Resolution .....................................................................................................45 Filter ..............................................................................................................45 Reference Impedance....................................................................................45 ORP Measurement Programming ..............................................................................45 8.3.1 Measurement................................................................................................46 8.3.3 Filter ..............................................................................................................46 8.3.2 8.4 Measurement................................................................................................44 8.3.4 Preamp..........................................................................................................46 Reference Impedance....................................................................................46 Contacting Conductivity ............................................................................................47 8.4.1 Description....................................................................................................47 8.4.3 Measure ........................................................................................................48 8.4.2 8.4.4 8.4.5 8.4.6 8.4.7 8.4.8 8.4.9 Sensor Type ...................................................................................................47 Range ............................................................................................................48 Cell Constant.................................................................................................48 RTD Offset.....................................................................................................48 RTD Slope......................................................................................................48 Temp Comp ..................................................................................................48 Slope .............................................................................................................49 8.4.10 Reference Temp ............................................................................................49 8.4.11 Filter ..............................................................................................................49 8.4.12 Custom Setup ...............................................................................................49 8.5 8.4.13 Cal Factor ......................................................................................................49 Toroidal Conductivity Measurement Programming...................................................50 8.5.1 Description....................................................................................................50 8.5.3 Measure ........................................................................................................51 8.5.2 8.5.4 8.5.5 8.5.6 8.5.7 8.5.8 8.5.9 8.6 Table of Contents Sensor Type ...................................................................................................50 Range ............................................................................................................51 Cell Constant.................................................................................................51 Temp Comp ..................................................................................................51 Slope .............................................................................................................52 Reference Temp ............................................................................................52 Filter ..............................................................................................................52 8.5.10 Custom Setup ...............................................................................................52 Chlorine Measurement Programming........................................................................53 8.6.1 Free Chlorine Measurement Programming...................................................53 8.6.1.1 Measure..........................................................................................54 8.6.1.2 Units ...............................................................................................54 V Table of Contents Instruction Manual April 2017 LIQ-MAN-1066 8.6.1.3 Filter................................................................................................54 8.6.1.4 Free Chlorine pH Correction ...........................................................54 8.6.1.5 Manual pH Correction ....................................................................54 8.6.2 8.6.1.6 Resolution.......................................................................................54 Total Chlorine Measurement Programming..................................................55 8.6.2.1 Description .....................................................................................55 8.6.2.2 Measure..........................................................................................55 8.6.2.3 Units ...............................................................................................55 8.6.2.4 Filter................................................................................................55 8.6.3 8.6.2.5 Resolution.......................................................................................55 Monochloramine Measurement Programming ............................................56 8.6.3.1 Measure: Monochloramine.............................................................56 8.6.3.2 Units ...............................................................................................56 8.6.3.3 Filter................................................................................................57 8.7 8.6.3.4 Resolution.......................................................................................57 Oxygen Measurement Programming.........................................................................57 8.7.1 Oxygen Measurement Application .................................................58 8.7.3 Partial Press.....................................................................................58 8.7.2 8.7.4 8.7.5 8.8 8.7.6 Salinity............................................................................................58 Filter................................................................................................58 Pressure Units .................................................................................58 Ozone Measurement Programming...........................................................................59 8.8.1 Units ...............................................................................................59 8.8.3 Resolution.......................................................................................59 8.8.2 Section 9: Calibration 9.1 Units ...............................................................................................58 Filter................................................................................................59 Introduction ..............................................................................................................67 9.2 Calibration..................................................................................................................67 9.2.1 Auto Calibration .........................................................................................................68 9.2.2 Manual Calibration – pH................................................................................68 9.2.4 Standardization – pH.....................................................................................69 9.2.3 9.3 9.4 9.2.5 SMART sensor auto calibration upload – pH..................................................69 ORP and Redox Calibration.........................................................................................70 Contacting Conductivity Calibration..........................................................................71 9.4.1 Entering the Cell Constant.............................................................................72 9.4.3 Calibrating the Sensor in a Conductivity Standard (in process cal) ................72 9.4.2 9.4.4 9.4.5 VI Entering a Known Slope Value – pH ..............................................................68 Zeroing the Instrument .................................................................................72 Calibrating the Sensor To A Laboratory Instrument (meter cal) ....................73 Cal Factor ......................................................................................................73 Table of Contents Instruction Manual Table of Contents LIQ-MAN-1066 April 2017 9.5 Toroidal Conductivity Calibration...............................................................................74 9.5.1 Entering the Cell Constant.............................................................................74 9.5.3 Calibrating the Sensor in a Conductivity Standard (in process cal) ................75 9.5.2 9.6 Zeroing the Instrument .................................................................................75 Calibration – Chlorine.................................................................................................76 9.6.1 9.6.2 Calibration – Free Chlorine ............................................................................76 9.6.1.1 Zeroing the Sensor..........................................................................77 9.6.1.2 In Process Calibration......................................................................77 Calibration – Total Chlorine...........................................................................77 9.6.2.1 Zeroing the Sensor..........................................................................78 9.6.2.2 In Process Calibration......................................................................78 9.6.3 Calibration – Monochloromine ..................................................................................79 9.6.4 9.7 9.6.5 9.9 In Process Calibration ....................................................................................80 Calibration – Oxygen..................................................................................................80 9.7.1 Zeroing the Sensor ........................................................................................82 9.7.3 Calibrating the Sensor Against A Standard Instrument (in process cal) .........83 9.7.2 9.8 Zeroing the Sensor ........................................................................................80 Calibrating the Sensor in Air ..........................................................................82 Calibration – Ozone....................................................................................................83 9.8.1 9.8.2 Zeroing the Sensor ........................................................................................84 In Process Calibration ....................................................................................84 Calibrating Temperature ............................................................................................85 9.9.1 Calibration.....................................................................................................85 Section 10: haRT® Communications 10.1 Introduction ...............................................................................................................93 10.2 Physical Installation and Configuration ......................................................................94 10.3 Measurements Available via HART .............................................................................96 10.4 Diagnostics Available via HART...................................................................................96 10.5 HART Hosts ................................................................................................................97 10.6 Wireless Communication using the 1066 ................................................................100 10.7 Field Device Specification (FDS) ...............................................................................100 APPENDIX 10.1 Device Variables .......................................................................................101 APPENDIX 10.2 Additional Transmitter Status – Command 48 Status Bits........................103 APPENDIX 10.3 1066 HART Configuration Parameters......................................................108 APPENDIX 10.4 475 Menu Tree for 1066 HART 7...............................................................115 Section 11: Return of Material 11.1 General.....................................................................................................................121 11.2 Warranty Repair .......................................................................................................121 11.3 Non-Warranty Repair ...............................................................................................121 EC Declarations of Conformity..........................................................................................................123 Table of Contents VII Table of Contents April 2017 VIII Instruction Manual LIQ-MAN-1066 Instruction Manual LIQ-MAN-1066 Section 1: Quick Start Guide April 2017 Section 1: Quick Start Guide 1.1 Quick Start Guide 1. For mechanical installation instructions, see page 14 for panel mounting and page 15 for pipe or wall mounting. 2. Wire the sensor to the main circuit board. See pages 21-23 for wiring instructions. Refer to the sensor instruction sheet for additional details. Make loop power connections. 3. Once connections are secured and verified, apply DC power to the transmitter. 4. When the transmitter is powered up for the first time, Quick Start screens appear. Quick Start operating tips are as follows: a. A highlighted field shows the position of the cursor. b. To move the cursor left or right, use the keys to the left or right of the ENTER key. To scroll up or down or to increase or decrease the value of a digit use the keys above and below the ENTER key. Use the left or right keys to move the decimal point. c. Press ENTER to store a setting. Press EXIT to leave without storing changes. Pressing EXIT during Quick Start returns the display to the initial start-up screen (select language). 5. Choose the desired language and press ENTER. 6. Choose measurement and press ENTER. a. For pH, choose preamplifier location. Select Analyzer to use the integral preamplifier in the transmitter; select Sensor/J-Box if your sensor is SMART or has an integral preamplifier or if you are using a remote preamplifier located in a junction box. 5. If applicable, choose units of measurement. 6. For contacting and toroidal conductivity, choose the sensors type and enter the numeric cell constant using the keys. 7. Choose temperature units: °C or °F. 8. After the last step, the main display appears. The outputs are assigned to default values. 9. To change output settings, to scale the 4-20 mA current outputs, to change measurementrelated settings from the default values, and to enable pH diagnostics, press MENU. Select Program and follow the prompts. Refer to the appropriate menu. 10. To return the transmitter to the factory default settings, choose Program under the main menu, and then scroll to Reset. 11. Please call the Rosemount Customer Support Center at 1-800-854-8257 if you need further support. 1 Section 2: Description and Specifications April 2017 2 Instruction Manual LIQ-MAN-1066 Description and Specifications Instruction Manual LIQ-MAN-1066 Section 2: Description and Specifications April 2017 Section 2: Description and Specifications 2.1 Features and applications This loop-powered multi-parameter unit serves industrial, commercial and municipal applications with the widest range of liquid measurement inputs available for a two-wire liquid transmitter. The 1066 Smart transmitter supports continuous measurement of one liquid analytical input. The design supports easy internal access and wiring connections. analytical Inputs: Ordering options for pH/ORP, Resistivity/Conductivity, % Concentration, Total Chlorine, Free Chlorine, Monochloramine, Dissolved Oxygen, and Ozone. large Display: The high-contrast LCD provides live measurement readouts in large digits and shows up to four additional variables or diagnostic parameters. Digital Communications: HART 7 and FOUNDATION Fieldbus options. Menus: Menu screens for calibrating and programming are simple and intuitive. Plain language prompts and help screens guide the user through the procedures. All menu screens are available in eight languages. Live process values are displayed during programming and calibration. Quick Start programming: Popular Quick Start screens appear the first time the unit is powered. The instrument prompts the user to configure the sensor loop in a few quick steps for immediate commissioning. user help Screens: Fault and warning messages include help screens similar to PlantWeb™ alerts that provide useful troubleshooting tips to the user. These on-screen instructions are intuitive and easy to use. Diagnostics: The transmitter continuously monitors itself and the sensor for problems. A display banner on the screen alerts Technicians to Fault and/or Warning conditions. languages: Emerson extends its worldwide reach by offering eight languages – English, French, German, Italian, Spanish, Portuguese, Chinese and Russian. Current Outputs: HART units include two 4-20 mA electrically isolated current outputs giving the ability to transmit the live measurement value and the process temperature reported from the sensor. Input Dampening: is automatically enabled to suppress noisy process readings. Smart-Enabled ph: Rosemount SMART pH capability eliminates field calibration of pH probes through automatic upload of calibration data and history. automatic Temperature Compensation: Most measurements require temperature compensation. The 1066 will automatically recognize Pt100, Pt1000 or 22k NTC RTDs built into the sensor. Smart Wireless Thum adaptor Compatible: Enable wireless transmissions of process variables and diagnostics from hard-to-reach locations. Specifications 3 Section 2: Description and Specifications Instruction Manual April 2017 LIQ-MAN-1066 Specifications - General 2.2 Case: Polycarbonate. IP66 (CSA, FM), Type 4X (CSA) Dimensions: Overall 155 x 155 x 131mm (6.10 x 6.10 x 5.15 in.). Cutout: 1/2 DIN 139mm x 139mm (5.45 x 5.45 in.) Conduit openings: Six. Accepts PG13.5 or 1/2 in. conduit fittings Display: Monochromatic graphic liquid crystal display. No backlight. 128 x 96 pixel display resolution. Active display area: 58 x 78mm (2.3 x 3.0 in.). All fields of the main instrument display can be customized to meet user requirements. ambient temperature and humidity: -20 to 65 °C (-4 to 149°F), RH 5 to 95% (non-condensing). Storage Temperature: -20 to 70 °F (-4 to 158 °F) haRT® Communications: PV, SV, TV, and 4V assignable to measurement, temperature and all live HART diagnostics. RFI/EMI: EN-61326-1 Complies with the following Standards: CSA: C22.2 No 0 – 10; C22.2 No 0.4 – 04; C22.2 No. 25-M1966: , C22.2 No. 94-M91: , C22.2 No.142-M1987: , C22.2 No. 157-M1992: , C22.2 No. 213-M1987: , C22.2 No. 60529:05. UL: 50:11th Ed.; 508:17th Ed.; 913:7th Ed.; 1203:4th Ed.. ANSI/ISA: 12.12.10-2013. ATEX: EN 60079-0:2012+A11:2013, 60079-11:2012 IECEx: IEC 60079-0: 2011 Edition: 6.0, I EC 60079-11 : 2011-06 Edition: 6.0 FM: 3600: 2011, 3610: 2010, 3611: 2004, 3810: 2005, IEC 60529:2004, ANSI/ISA 60079-0: 2009, ANSI/ISA 60079-11: 2009 hazardous location approvals Intrinsic Safety (with appropriate safety barrier): Class I, II, III, Div. 1* Groups A-G T4 Tamb = -20 °C to 65 °C Enclosure 4X, IP66 For Intrincically Safe Installation, see drawing 1400669 aTEx 1180 II 1 G Baseefa11ATEX0195X Ex ia IIC T4 Ga T4 Tamb = -20 °C to 65 °C Non-Incendive: Class I, Div. 2, Groups A-D* Dust Ignition Proof Class II & III, Div 1, Groups EFG Class II & III, Div. 1, Groups E-G Type 4/4X Enclosure T4 Tamb = -20 °C to 65 °C IECEx BAS 11.0098X Ex ia IIC T4 Ga T4 Tamb = -20 °C to 65 °C Class I, II & III, Division 1, Groups A-G T4 Tamb = -20 °C to 65 °C IP66 enclosure Class I, Zone 0, AEx ia IIC T4 Tamb = -20°C to 65°C For Intrinsically Safe Installation, see drawing 1400670 Class I, Division 2 Groups A-D Dust Ignition proof Class II & III, Div 1, Groups EFG Class II & III, Division 1, Groups E-G Tamb = -20°C to 65°C, IP66 enclosure For Non-Incendive Field Wiring Installation, see drawing 1400670 For Non-Incendive Field Wiring Installation, see drawing 1400669 *Additionally approved as a system with models 140,141,142, 150, 400, 400VP, 401, 402, 402VP, 403,403VP, 404 & 410VP contacting conductivity sensors and models 222, 225, 226 & 228 inductive conductivity sensors. 4 Specifications Instruction Manual Section 2: Description and Specifications LIQ-MAN-1066 April 2017 Input: One isolated sensor input. Measurement choices of pH/ORP, resistivity/conductivity/TDS, % concentration, total and free chlorine, monochloramine, dissolved oxygen, dissolved ozone, and temperature. For contacting conductivity measurements, temperature element can be a PT1000 RTD or a PT100 RTD. Other measurements (except ORP) and use PT100 or PT1000 RTDs or a 22k NTC (D.O. only). power and load Requirements: Supply voltage at the transmitter terminals should be at least 12.7 Vdc. Power supply voltage should cover the voltage drop on the cable plus the external load resistor required for HART communications (250 Ω minimum). Minimum power supply voltage is 12.7 Vdc. Maximum power supply voltage is 42.4 Vdc (30 Vdc for intrinsically safe operation). The graph shows the supply voltage required to maintain 12 Vdc (upper line) and 30 Vdc (lower line) at the transmitter terminals when the current is 22 mA. FIGuRE 2-1. load/power Supply Requirements 1500 1364 ohms 1250 Load, ohms 1000 750 545 ohms with HART communication 500 250 without HART communication 0 12 18 24 30 36 42 Power supply voltage, Vdc HART option analog Outputs: Two-wire loop powered (Output 1 only). Two 4-20 mA electrically isolated current outputs (Output 2 must be externally powered). Superimposed HART digital signal on Output 1. Fully scalable over the operating range of the sensor. Weight/Shipping Weight: 2 lbs/3 lbs (1 kg/1.5 kg) Specifications 5 Section 2: Description and Specifications April 2017 Instruction Manual 2.3 ph/ORp (Codes – p) 2.3.1 Performance Specifications - Transmitter (pH input) 2.3.2 Performance Specifications - Transmitter (ORP input) LIQ-MAN-1066 For use with any standard pH or ORP sensor. SMART pH sensor with SMART pre-amplifiers from Rosemount. Measurement choices are pH, ORP, or Redox. The automatic buffer recognition feature uses stored buffer values and their temperature curves for the most common buffer standards available worldwide. The transmitter will recognize the value of the buffer being measured and perform a self stabilization check on the sensor before completing the calibration. Manual or automatic temperature compensation is menu selectable. Change in pH due to process temperature can be compensated using a programmable temperature coefficient. Measurement Range [ph]: 0 to 14 pH accuracy: ±0.01 pH Buffer recognition: NIST, DIN 19266, JIS 8802, and BSI. Input filter: Time constant 1 - 999 sec, default 4 sec. Response time: 5 seconds to 95% of final reading Recommended Sensors for ph: All standard pH sensors. Supports SMART pH sensors from Rosemount. Measurement Range [ORp]: -1400 to +1400 mV accuracy: ± 1 mV Input filter: Time constant 1 - 999 sec, default 4 sec. Response time: 5 seconds to 95% of final reading Recommended Sensors for ORp: All standard ORP sensors FIGuRE 2-2. General purpose and high performance ph sensors 3900, 396pvp and 3300hT 6 Specifications Instruction Manual Section 2: Description and Specifications LIQ-MAN-1066 April 2017 Contacting Conductivity (Codes – C) 2.4 Measures conductivity in the range 0 to 600,000 µS/cm (600mS/cm). Measurement choices are conductivity, resistivity, total dissolved solids, salinity, and % concentration. In addition, the “Custom Curve” feature allows users to define a three to five point curve to measure ppm, %, or a no unit variable. The % concentration selection includes the choice of five common solutions (012% NaOH, 0-15% HCl, 0-20% NaCl, and 0-25% or 96-99.7% H2SO4). The conductivity concentration algorithms for these solutions are fully temperature compensated. Three temperature compensation options are available: manual slope (X% / °C), high purity water (dilute sodium chloride), and cation conductivity (dilute hydrochloric acid). Temperature compensation can be disabled, allowing the transmitter to display raw conductivity. For more information concerning the use of the contacting conductivity sensors, refer to the product data sheets. 2.4.1 Note: The 410VP 4-electrode high-range conductivity sensor is compatible with the 1066. Performance Specifications ENDURANCE series of conductivity sensors TM Temperature specifications: Temperature range Temperature Accuracy, Pt-1000, 0-50 °C Temperature Accuracy, Pt-1000, Temp. > 50 °C 0-200 °C ± 0.1 °C ± 0.5 °C Input filter: Time constant 1 - 999 sec, default 2 sec. Response time: 3 seconds to 95% of final reading using the default input filter Salinity: Uses Practical Salinity Scale Total Dissolved Solids: Calculated by multiplying conductivity at 25 °C by 0.65 Table 2-1. performance Specifications: Recommended Range – Contacting Conductivity Cell 0.01S/cm Constant 0.01 0.1µS/cm 1.0µS/cm 10µS/cm 100µS/cm 0.1µS/cm to 2000µS/cm 1.0 10mS/cm 20mS/cm to 200mS/cm 4-electrode Specifications 1000mS/cm 2000µS/cm to 20mS/cm 1 µS/cm to 20mS/cm linearity for Standard Cable ≤ 50 ft (15 m) 100mS/cm 200µS/cm to 2000µS/cm 0.01µS/cm to 200µS/cm 0.1 1000µS/cm 2µS/cm to 1400mS/cm ±0.6% of reading in recommended range ±2% of reading outside high recommended range ±5% of reading outside low recommended range ±4% of reading in recommended range 7 Section 2: Description and Specifications Instruction Manual April 2017 2.4.2 2.5 2.5.1 LIQ-MAN-1066 Recommended Sensors for Conductivity All Rosemount 400 series conductivity sensors (Pt 1000 RTD) and 410VP 4-electrode sensor. Toroidal Conductivity (Codes – T) Measures conductivity in the range of 1 µS/cm to 2,000,000 µS/cm (2 S/cm). Measurement choices are conductivity, resistivity, total dissolved solids, salinity, and % concentration. The % concentration selection includes the choice of five common solutions (0-12% NaOH, 0-15% HCl, 0-20% NaCl, and 025% or 96-99.7% H2SO4). The conductivity concentration algorithms for these solutions are fully temperature compensated. For other solutions, a simple-to-use menu allows the customer to enter his own data. The transmitter accepts as many as five data points and fits either a linear (two points) or a quadratic function (three to five points) to the data. Reference temperature and linear temperature slope may also be adjusted for optimum results. Two temperature compensation options are available: manual slope (X% / °C) and neutral salt (dilute sodium chloride). Temperature compensation can be disabled, allowing the transmitter to display raw conductivity. For more information concerning use of the toroidal conductivity sensors, refer to the product data sheets. Performance Specifications High performance 225 Toroidal & 226 Conductivity sensors Temperature specifications: Temperature range -25 to 210 °C (-13 to 410 °F) Temperature Accuracy, Pt-100, -25 to 50 °C ± 0.5 °C Temperature Accuracy, Pt-100, 50 to 210 °C ± 1 °C Repeatability: ±0.25% ±5 µS/cm after TaBlE 2-2. performance Specifications: Recommended Range – Toroidal Conductivity Model 1µS/cm 10µS/cm 226 225 & 228 242 222 (1in & 2in) loop performance (Following Calibration) 100µS/cm 1000µS/cm 10mS/cm 100mS/cm 1000mS/cm 2000mS/cm 50µS/cm to 500mS/cm 500mS/cm to 2000mS/cm 50µS/cm to 1500mS/cm 1500mS/cm to 2000mS/cm 100µS/cm to 2000mS/cm 500µS/cm to 2000mS/cm 226: ±1% of reading ±5µS/cm in recommended range 225 & 228: ±1% of reading ±15µS/cm in recommended range 222, 242: ±4% of reading ±5mS/cm in recommended range 225, 226 & 228: ±5% of reading outside high recommended range 8 Specifications Instruction Manual LIQ-MAN-1066 2.5.2 2.6 2.6.1 2.6.2 2.6.3 2.6.4 Section 2: Description and Specifications April 2017 zero cal Input filter: time constant 1 - 999 sec, default 2 sec. Response time: 3 seconds to 95% of final reading Salinity: Uses Practical Salinity Scale Total Dissolved Solids: Calculated by multiplying conductivity at 25 °C by 0.65 Recommended Sensors for Conductivity All Rosemount submersion/immersion and flow-through toroidal sensors. Chlorine (Codes – Cl) Free and Total Chlorine The 1066 is compatible with the 499ACL-01 free chlorine sensor and the 499ACL-02 total chlorine sensor. The 499ACL-02 sensor must be used with the TCL total chlorine sample conditioning system. The 1066 fully compensates free and total chlorine readings for changes in membrane permeability caused by temperature changes. For free chlorine measurements, both automatic and manual pH correction are available. For automatic pH correction select an appropriate pH sensor. For more information concerning the use and operation of the amperometric chlorine sensors and the TCL measurement system, refer to the product data sheets. Performance Specifications Resolution: 0.001 ppm or 0.01 ppm – selectable Input Range: 0nA – 100 µA automatic ph correction for Free Chlorine: (user selectable for code -CL): 6.0 to 10.0 pH Temperature compensation: Automatic (via RTD) or manual (050 °C). Input filter: Time constant 1 - 999 sec, default 5 sec. Response time: 6 seconds to 95% of final reading 499ACL-01 Chlorine sensor Recommended Sensors Chlorine: 499ACL-01 Free Chlorine or 499ACL-02 Total Residual Chlorine ph: These pH sensors are recommended for automatic pH correction of free chlorine readings: 3900-02-10, 3900-01-10, and 3900VP-02-10. Monochloramine The 1066 is compatible with the 499A CL-03 Monochloramine sensor. The 1066 fully compensates readings for changes in membrane permeability caused by temperature changes. Because monochloramine measurement is not affected by pH of the process, no pH sensor or correction is required. For more information concerning the use and operation of the amperometric chlorine sensors, refer to the product data sheets. Specifications 9 Section 2: Description and Specifications April 2017 2.6.5 Performance Specifications 2.6.6 Recommended Sensors 2.7 2.7.1 2.7.2 2.8 2.8.1 2.8.2 10 Instruction Manual LIQ-MAN-1066 Resolution: 0.001 ppm or 0.01 ppm – selectable Input Range: 0nA – 100µA Temperature compensation: Automatic (via RTD) or manual (0-50 °C). Input filter: Time constant 1 - 999 sec, default 5 sec. Response time: 6 seconds to 95% of final reading Rosemount 499ACL-03 Monochloramine sensor Dissolved Oxygen (Codes –DO) The 1066 is compatible with the 499ADO, 499ATrDO, Hx438, Gx438 and Bx438 dissolved oxygen sensors and the 4000 percent oxygen gas sensor. The 1066 displays dissolved oxygen in ppm, mg/L, ppb, µg/L, % saturation, % O2 in gas, ppm O2 in gas. The transmitter fully compensates oxygen readings for changes in membrane permeability caused by temperature changes. Automatic air calibration, including salinity correction, is standard. The only required user entry is barometric pressure. For more information on the use of amperometric oxygen sensors, refer to the product data sheets. Performance Specifications Resolution: 0.01 ppm; 0.1 ppb for 499A TrDO sensor (when O2 <1.00 ppm); 0.1% Input Range: 0nA – 100μA Temperature Compensation: Automatic (via RTD) or manual (0-50 °C). Input filter: Time constant 1 - 999 sec, default 5 sec. Response time: 6 seconds to 95% of final reading Recommended Sensors Dissolved Oxygen 499ADO sensor with Variopol connection Rosemount amperometric membrane and steam-sterilizable sensors listed above Ozone (Codes –Oz) The 1066 is compatible with the 499AOZ sensor. The 1066 fully compensates ozone readings for changes in membrane permeability caused by temperature changes. For more information concerning the use and operation of the amperometric ozone sensors, refer to the product data sheets. Performance Specifications Resolution: 0.001 ppm or 0.01 ppm – selectable Input Range: 0nA – 100μA Temperature Compensation: Automatic (via RTD) or manual (0-35 °C) Input filter: Time constant 1 - 999 sec, default 5 sec. Response time: 6 seconds to 95% of final reading Recommended Sensors Rosemount 499A OZ ozone sensor Dissolved Ozone 499AOZ sensors with Variopol connection Specifications Instruction Manual Section 2: Description and Specifications LIQ-MAN-1066 2.9 April 2017 Ordering Information The 1066 2-Wire Transmitter is intended for the continuous determination of pH, ORP (Redox), conductivity, (both contacting and toroidal), and for measurements using membrane-covered amperometric sensors (oxygen, ozone, free and total chlorine, and monochloramine). For free chlorine measurements, which often require continuous pH correction a second input for a pH sensor is available. Two 4-20mA analog outputs are standard on HART units. The 1066 is compatible with SMART pH sensors from Rosemount. HART digital communications is standard and FOUNDATION® fieldbus digital communication is offered as an option. Communication with the 1066 is through: Local keypad interface 475 HART® and FOUNDATION fieldbus Communicator HART protocol version 7 FOUNDATION fieldbus AMS (Asset Management Solutions) Aware SMART Wireless THUM™ Adapter TaBlE 2-3. Ordering Information Description 1066 ph/ORp, Conductivity, Chlorine, Oxygen, and Ozone 2-Wire Transmitter Measurement P pH/ORP T Toroidal Conductivity C CL DO OZ Communication Contacting Conductivity Chlorine Dissolved Oxygen Ozone HT HART® Digital Communication Superimposed on 4-20mA Output FI FOUNDATION™ fieldbus Digital Output with FISCO FF FOUNDATION™ fieldbus Digital Output agency approval 60 None Required 67 69 73 Specifications FM Approved, Intrinsically Safe (appropriate sensor & safety barrier required), and Non-Incendive CSA Approved , Intrinsically Safe (appropriate sensor & safety barrier required), and Non-Incendive ATEX/IECEx Approved, Intrinsically Safe (safety barrier required) 11 Section 2: Description and Specifications April 2017 12 Instruction Manual LIQ-MAN-1066 Specifications Instruction Manual LIQ-MAN-1066 Section 3: Installation April 2017 Section 3: Installation 3.1 unpacking and inspection 3.2 Installation – General Information Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions. Save the box. If there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present. If items are missing, notify Rosemount immediately. 1. Although the transmitter is suitable for outdoor use, installation is direct sunlight or in areas of extreme temperatures is not recommended unless a sunshield is used. 2. Install the transmitter in an area where vibration and electromagnetic and radio frequency interference are minimized or absent. 3. Keep the transmitter and sensor wiring at least one foot from high voltage conductors. Be sure there is easy access to the transmitter. 4. The transmitter is suitable for panel, pipe, or surface mounting. 5. The transmitter case has six 1/2-inch (PG13.5) conduit openings. Use separate conduit openings for the power/output cable, the sensor cable, and the other the sensor cable as needed (pH input for free chlorine with continuous pH correction). 3.3 6. Use weathertight cable glands to keep moisture out to the transmitter. If conduit is used, plug and seal the connections at the transmitter housing to prevent moisture from getting inside the instrument. preparing Conduit Openings There are six conduit openings in all configurations of 1066. Conduit openings accept 1/2-inch conduit fittings or PG13.5 cable glands. To keep the case watertight, block unused openings with Type 4X or IP66 conduit plugs. To maintain ingress protection for outdoor use, seal unused conduit holes with suitable conduit plugs. NOTE: Use watertight fittings and hubs that comply with your requirements. Connect the conduit hub to the conduit before attaching the fitting to the transmitter. Electrical installation must be in accordance with the National Electrical Code (aNSI/NFpa-70) and/or any other applicable national or local codes. Installation 13 Section 3: Installation April 2017 Instruction Manual LIQ-MAN-1066 FIGuRE 3-1. panel Mounting Dimensions 14 Installation Instruction Manual LIQ-MAN-1066 Section 3: Installation April 2017 FIGuRE 3-2. pipe and wall mounting dimensions (Mounting bracket pN: 23820-00) Installation 15 Section 3: Installation April 2017 16 Instruction Manual LIQ-MAN-1066 Installation Instruction Manual Section 4: Wiring LIQ-MAN-1066 April 2017 Section 4: Wiring 4.1 4.1.1 General General Information The 1066 is easy to wire. All wiring connections are located on the main circuit board. The front panel is hinged at the bottom. The panel swings down for easy access to the wiring locations. 4.1.2 Digital Communication 4.2 power Supply/Current loop – 1066 - hT 4.2.1 HART and FOUNDATION fieldbus communications are available as ordering options for 1066. HART units support Bell 202 digital communications over analog 4-20mA current output 1. Power Supply and Load Requirements Refer to Figure 4-1. The supply voltage must be at least 12.7 Vdc at the transmitter terminals.. The power supply must be able to cover the voltage drop on the cable as well as the load resistor (250 Ω minimum) required for HART communications. The maximum power supply voltage is 42.0 Vdc. For intrinsically safe installations, the maximum power supply voltage is 30.0 Vdc. The graph shows load and power supply requirements. The upper line is the power supply voltage needed to provide 12.7 Vdc at the transmitter terminals for a 22 mA current. The lower line is the power supply voltage needed to provide 30 Vdc for a 22 mA current. The power supply must provide a surge current during the first 80 milliseconds of startup. The maximum current is about 24 mA. For digital communications, the load must be at least 250 ohms. To supply the 12.7 Vdc lift off voltage at the transmitter, the power supply voltage must be at least 17.5 Vdc. FIGuRE 4-1. load/power Supply Requirements 1500 1364 ohms 1250 Load, ohms 1000 750 545 ohms with HART communication 500 250 without HART communication 0 12 18 24 30 36 42 Power supply voltage, Vdc HART option Wiring 17 Section 4: Wiring April 2017 4.2.2 Instruction Manual LIQ-MAN-1066 Power Supply-Current Loop Wiring Refer to Figure 4-2. Run the power/signal wiring through the opening nearest TB-2. For optimum EMI/RFI protection: 1. Use shielded power/signal cable and ground the shield at the power supply. 2. Use a metal cable gland and be sure the shield makes good electrical contact with the gland. 3. Use the metal backing plate when attaching the gland to transmitter enclosure. The power/signal cable can also be enclosed in an earth-grounded metal conduit. Do not run power supply/signal wiring in the same conduit or cable tray with loop power lines. Keep power supply/signal wiring at least 6 ft (2 m) away from heavy electrical equipment. FIGuRE 4-2. haRT Communications 18 Wiring Instruction Manual Section 4: Wiring LIQ-MAN-1066 4.2.3 April 2017 Current Output wiring The 1066 HART units are shipped with two 4-20mA current outputs. Current Output 1 is loop power; it is the HART communications channel. Current output 2 is available to report process temperature measured by the temperature sensing element or RTD within the sensor. Wiring locations for the outputs are on the main board which is mounted on the hinged door of the instrument. Wire the output leads to the correct position on the main board using the lead markings (+/positive, -/negative) on the board. FIGuRE 4-3. 1066 haRT loop power Wiring INSTALL PLUGS IN ALL OTHER OPENINGS AS NEEDED _ + OUTPUT 2 +24V GND GND THUM 1 TB6 2 HINGE SIDE OF FRONT PANEL SENSOR WIRING (OUTPUT1) LOOP PWR +24V TB7 HINGED PANEL 4-20mA / -24VDC RETURN 4-20mA / +24VDC INNER ENCLOSURE TB5 TB3 TB2 ANODE RTN CATHODE SNS SHLD RTD IN GND SOL REF +V SHLD pH -V TB4 TB1 1066 HART CIRCUIT BOARD (pH/CL/DO/OZ) ASSY 24406-xx 1 TB7/OUTPUT 2 REQUIRES EXTERNAL DC POWER. 2 TB6/ THUM TERMINAL IS USED ONLY FOR DWG NO. WIRELESS THUM ADAPTOR INSTALLATIONS. Wiring 40106613 19 Section 4: Wiring April 2017 4.3 4.3.1 Instruction Manual LIQ-MAN-1066 power Supply Wiring For 1066-FF Power Supply Wiring Run the power/signal wiring through the opening nearest TB2. Use shielded cable and ground the shield at the power supply. To ground the transmitter, attach the shield to TB2-3. Note: For optimum EMI/RFI immunity, the power supply/output cable should be shielded and enclosed in an earth-grounded metal conduit. Do not run power supply/signal wiring in the same conduit or cable tray with loop power lines. Keep power supply/signal wiring at least 6 ft (2 m) away from heavy electrical equipment. FOuNDaTION Fieldbus Figure 4-4 shows a 1066PFF being used to measure and control pH and chlorine levels in drinking water. The figure also shows three ways in which Fieldbus communication can be used to read process variables and configure the transmitter. FIGuRE 4-4. Configuring 1066p Transmitter with FOuNDaTION fieldbus FIGuRE 4-5. Typical Fieldbus Network Electrical Wiring Configuration 20 Wiring Instruction Manual LIQ-MAN-1066 4.4 Section 4: Wiring April 2017 Sensor Wiring to Main Board Wire the correct sensor leads to the main board using the lead locations marked directly on the board. Rosemount SMART pH sensors can be wired to the 1066 using integral cable SMART sensors or compatible VP8 pH cables. After wiring the sensor leads, carefully take up the excess sensor cable through the cable gland. Keep sensor and output signal wiring separate from loop power wiring. Do not run sensor and power wiring in the same conduit or close together in a cable tray. FIGuRE 4-6. ph/ORp sensor wiring to the 1066 printed circuit board DWG NO. 40106612 Wiring 21 Section 4: Wiring April 2017 Instruction Manual LIQ-MAN-1066 FIGuRE 4-7. Contacting and Toroidal Conductivity sensor wiring to the 1066 circuit board DWG NO. 40106615 22 Wiring Instruction Manual Section 4: Wiring LIQ-MAN-1066 April 2017 FIGuRE 4-8. Chlorine, oxygen, ozone sensor wiring to 1066 printed circuit board (1066-Cl, 1066-DO, 1066-Oz) HINGE SIDE OF FRONT PANEL OUTPUT 2 (OUTPUT1) LOOP PWR TB5 +24V +24V ANODE GND GND CATHODE TB6 THUM TB7 SENSOR WIRING TB3 TB2 RTN REF SNS SHLD RTD IN GND SOL +V SHLD pH -V TB4 TB1 1066 CIRCUIT BOARD ASSY 24406-xx CHLORINE, OXYGEN, OZONE SENSOR WIRING (FOLLOW RECOMMENDED ORDER) 1) TB5/ANODE & CATHODE ANODE CATHODE 2) TB3/RTD RETURN SENSE RTD IN 3) TB2/ SOLUTION GROUND NO CONNECTION NO CONNECTION SOLUTION GROUND DWG NO. 40106611 NOTE: A) TB1, TB4, TB6 AND TB7 NOT USED FOR OXYGEN AND OZONE SENSOR WIRING B) TB1, TB2 AND TB4 MAY BE USED FOR pH SENSOR WIRING IF FREE CHLORINE MEASUREMENT REQUIRES LIVE pH INPUT. Wiring 23 Section 4: Wiring Instruction Manual April 2017 LIQ-MAN-1066 FIGuRE 4-9. power/Current loop wiring with wireless ThuM adaptor INSTALL PLUGS IN ALL OTHER OPENINGS AS NEEDED LOOP POWER/THUM + 4-20mA / -24VDC RETURN _ 4-20mA / +24VDC INNER ENCLOSURE YELLOW / OUTPUT 1 +24V BLACK / OUTPUT 1 GND SENSOR WIRING (OUTPUT1) OUTPUT 2 LOOP PWR +24V +24V GND GND TB7 TB6 1 TB5 TB3 TB2 ANODE RTN REF CATHODE SNS SHLD RTD IN GND SOL THUM RED / +VDC 3 GREEN / GROUND SCREW HINGE SIDE OF FRONT PANEL 2 +V SHLD pH -V WIRE NUT WHITE TB4 TB1 1066 HART CIRCUIT BOARD (pH/CL/DO/OZ) ASSY 24406-xx 1 TB7/OUTPUT 2 REQUIRES EXTERNAL DC POWER. 2 TB6/ THUM TERMINAL IS USED ONLY FOR WIRELESS THUM ADAPTOR INSTALLATIONS. 250 OHM RESISTOR IS PRE-INSTALLED IN-CIRCUIT. 3 SPLICE CONNECTOR - PROVIDED BY END USER. WIRELESS THUM ADAPTOR DWG NO. 40106614 24 Wiring Instruction Manual Section 4: Wiring LIQ-MAN-1066 April 2017 FIGuRE 4-10. haRT loop power Wiring INSTALL PLUGS IN ALL OTHER OPENINGS AS NEEDED _ + OUTPUT 2 +24V +24V GND GND THUM 1 TB6 2 HINGE SIDE OF FRONT PANEL SENSOR WIRING (OUTPUT1) LOOP PWR TB7 HINGED PANEL 4-20mA / -24VDC RETURN 4-20mA / +24VDC INNER ENCLOSURE TB5 TB3 TB2 ANODE RTN CATHODE SNS SHLD RTD IN GND SOL REF +V SHLD pH -V TB4 TB1 1066 HART CIRCUIT BOARD (pH/CL/DO/OZ) ASSY 24406-xx 1 TB7/OUTPUT 2 REQUIRES EXTERNAL DC POWER. 2 TB6/ THUM TERMINAL IS USED ONLY FOR DWG NO. WIRELESS THUM ADAPTOR INSTALLATIONS. Wiring 40106613 25 Section 4: Wiring April 2017 26 Instruction Manual LIQ-MAN-1066 Wiring Intrinsically Safe Installation 15 16 17 18 FIELD DEVICE INPUT ASSOCIATED APPARATUS OUTPUT SO LONG AS THE CAPACITANCE AND INDUCTANCE OF THE LOAD CONNECTED TO THE SENSOR TERMINALS DO NOT EXCEED THE VALUES SPECIFIED IN TABLE I TABLE III 30 30 17.5 1066...HT... ANALOG OUTPUT 2 SIGNAL TERMINALS TB7 -1 & -2 1066...FF/FI... ENTITY LOOP POWER SIGNAL TERMINALS TB6 -1 & -2 1066...FI... FISCO LOOP POWER SIGNAL TERMINALS TB6 -1 & -2 Vmax IN: Vdc 30 MODEL NO. Pmax IN: W 1.0 Imax IN:mA 200 (475 INSTALLATION DRAWING IS 00475-1130) 375 OR 475 ENTITY PARAMETERS: REMOTE TRANSMITTER INTERFACE 30 1066...HT...LOOP POWER SIGNAL TERMINALS TB6 -1, -2 & -3 MODEL NO. Vmax (VDC) 6.04 0.0 Ci (µF) 0.0 DWG NO 1.9 Voc max OUT: Vdc 5.32 1.3 0.9 0 0 0 0 Li (mH) 32 Isc max OUT:µA 0 0 0 0 Ci (nF) Lo 0.9 3762 pF 151.95 nH Co Pmax (W) 231 mW 11.88 V 153.4 mA Io Po Uo MODEL 1066 TB1-1 THRU 12 TABLE II (FOR 1066-P/CL/DO/OZ...) OUTPUT PARAMETERS Li (mH) 380 300 200 200 Imax (mA) 12.08 9.39 38.5 C D La (mH) 1.51 Ca (µF) OUTPUT PARAMETERS 1.44 A, B GAS GROUPS 1066 SUPPLY ENTITY PARAMETERS TABLES I AND II ARE FOR pH, CHLORINE, DISSOLVED OXYGEN AND OZONE OPTIONS TABLE I (FOR 1066-P/CL/DO/OZ...) APPROVED MODELS 1066-AA-BB-CC XMTR E REV 5.1 CERTIFICATION AGENCY SUBMITTAL / APPROVAL AGENCY CONTROLLED DOCUMENT Instruction Manual LIQ-MAN-1066 Section 4: Wiring April 2017 Section 5: Intrinsically Safe Installation FIGuRE 5-1. CSa Installation all Intrinsically Safe Installations 27 EITHER OR BOTH MAY BE INSTALLED 1066-CL/DO/OZ...ONLY (SENSOR AND SECOND ANALOG OUTPUT CONNECTION UNCHANGED FROM ABOVE) ALTERNATE POWER CONNECTION IF SMART THUM WIRELESS ADAPTER IS USED AMPEROMETRIC SENSOR CSA APPROVED DEVICE OR SIMPLE APPARATUS SMART THUM WIRELESS ADAPTER WHITE YELLOW RED GREEN BLACK ROSEMOUNT MODEL 375 OR 475 FIELD COMMUNICATOR REMOTE TRANSMITTER INTERFACE FOR USE IN CLASS I AREA (SEE NOTE 3 AND TABLE III) N/C N/C SHLD SHLD GND SOL REF SENSOR WIRING TB3 TB2 RTN TB1 OPTIONAL CSA APPROVED PREAMP THAT MEETS REQUIREMENTS OF NOTE 4 TB5 TB4 SNS RTD IN PH SENSOR CSA APPROVED DEVICE OR SIMPLE APPARATUS OR ANODE CATHODE RECOMMENDED CABLE PN 9200273 (UNPREPED) PN 23646-01 (PREPPED) 10 COND, 2 SHIELDS, 24 AWG. SEE NOTE 2 SHLD GND SOL SHLD SENSOR WIRING TB3 TB2 REF TB1 1066-CL/DO/OZ...ONLY TB4 SNS RTN RTD IN ANODE AMPEROMETRIC SENSOR CSA APPROVED DEVICE OR SIMPLE APPARATUS GND (OUTPUT1) LOOP PWR TB6 SHLD SHLD GND SOL REF SENSOR WIRING TB3 TB2 RTN TB1 CATHODE TB5 TB5 TB4 SNS RTD IN EITHER OR BOTH MAY BE INSTALLED THUM GND (OUTPUT1) LOOP PWR TB6 ANODE CATHODE OR THUM GND (OUTPUT1) LOOP PWR TB6 THUM SPLICE CONNECTOR ROSEMOUNT MODEL 375 OR 475 FIELD COMMUNICATOR REMOTE TRANSMITTER INTERFACE FOR USE IN CLASS I AREA (SEE NOTE 3 AND TABLE III) ANALOG OUTPUT 2 ONLY AVAILABLE ON 1066...HT... ROSEMOUNT MODEL 375 OR 475 FIELD COMMUNICATOR REMOTE TRANSMITTER INTERFACE FOR USE IN CLASS I AREA (SEE NOTE 3 AND TABLE III) ANALOG OUTPUT 2 ONLY AVAILABLE ON 1066...HT... NI CLASS I, DIV 2 GRPS A-D CLASS II, DIV 2 GRPS E-G IS CLASS I, GRPS A-D CLASS II, GRPS E-G CLASS III + _ SAFETY BARRIER (SEE NOTES 10 & 11) SAFETY BARRIER (SEE NOTES 2 & 8 FOR FISCO, SEE NOTES 2, 8, 9 & 10 FOR ALL OTHER OPTIONS.) SAFETY BARRIER (SEE NOTES 2 & 8 FOR FISCO, SEE NOTES 2, 8, 9 & 10 FOR ALL OTHER OPTIONS.) SAFETY BARRIER (SEE NOTES 2 & 8 FOR FISCO, SEE NOTES 2, 8, 9 & 10 FOR ALL OTHER OPTIONS.) SAFETY BARRIER (SEE NOTES 2 & 8 FOR FISCO, SEE NOTES 2, 8, 9 & 10 FOR ALL OTHER OPTIONS.) LOAD LOAD LOAD LOAD LOAD UNSPECIFIED POWER SUPPLY 30 VDC MAX FOR IS 24V TYPICAL D DWG NO UNSPECIFIED POWER SUPPLY 30 VDC MAX FOR INTRINSIC SAFETY (24 VDC TYPICAL) 17.5 VDC MAX. FOR FISCO OPTION UNSPECIFIED POWER SUPPLY 30 VDC MAX FOR INTRINSIC SAFETY (24 VDC TYPICAL) 17.5 VDC MAX. FOR FISCO OPTION UNSPECIFIED POWER SUPPLY 30 VDC MAX FOR INTRINSIC SAFETY (24 VDC TYPICAL) 17.5 VDC MAX. FOR FISCO OPTION UNSPECIFIED POWER SUPPLY 30 VDC MAX FOR INTRINSIC SAFETY (24 VDC TYPICAL) 17.5 VDC MAX. FOR FISCO OPTION E REV February 2017 OUTPUT 2 GND OUTPUT 2 GND OUTPUT 2 GND PH SENSOR CSA APPROVED DEVICE OR SIMPLE APPARATUS TB7 TB7 TB7 28 WARNING- TO PREVENT IGNITION OF FLAMMABLE OR COMBUSTIBLE ATMOSPHERES, DISCONNECT POWER BEFORE SERVICING. WARNING- SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY OR SUITABILITY FOR DIVISION 2. Section 4: Wiring Instruction Manual LIQ-MAN-1066 FIGuRE 5-2. CSa Installation Intrinsically Safe Installation DRV_B RCV_B RCV_A RSHLD APPROVED MODEL 222, 225, 226 OR 228 TOROIDAL CONDUCTIVITY SENSOR OR MODELS 140, 141, 142, 150, 400, 400VP, 401, 402, 402VP, 403, 403VP, 404 & 410VP CONTACTING CONDUCTIVITY SENSOR OR ANY SIMPLE APPARATUS WITH LESS THAN 200 FEET OF ATTACHED CABLE 17 17 18 18 RCV_B RCV_A RSHLD DRV_B DRV_A 1066-CL/DO/OZ...ONLY TB1 DSHLD AMPEROMETRIC SENSOR CSA APPROVED DEVICE OR SIMPLE APPARATUS SHLD SHLD GND SOL REF SENSOR WIRING TB3 TB2 RTN TB1 EITHER OR BOTH MAY BE INSTALLED TB4 SNS RTD IN 18 OR TB2 TB5 TB1 17 THUM +24V GND LOOP PWR RTN SENSE RTDIN SHLD ANODE CATHODE TB2 RTN PH SENSOR CSA APPROVED DEVICE OR SIMPLE APPARATUS TB6 GND (OUTPUT1) LOOP PWR TB6 THUM LOOP PWR 1066C... ENTITY OUTPUT PARAMETERS: Voc = 4.75V, Isc = 676.93mA, Pmax = 258mW THE 1066C IS CERTIFIED AS A SYSTEM FOR USE WITH MODELS 140, 141, 142, 150, 400, 400VP, 401, 402, 402VP, 403, 403VP, 404 & 410VP CONTACTING CONDUCTIVITY SENSORS, OR ANY SIMPLE APPARATUS WITH < 200 FEET OF CABLE. THE 1066T... IS CERTIFIED AS A SYSTEM FOR USE WITH THE MODELS 222, 225, 226 & 228 TOROIDAL CONDUCTIVITY SENSORS, OR ANY SIMPLE APPARATUS WITH < 200 FEET OF CABLE. OUTPUT2 +24V GND OUTPUT 2 GND TB6 DSHLD DRV_A APPROVED MODEL 222, 225, 226 OR 228 TOROIDAL CONDUCTIVITY SENSOR OR MODELS 140, 141, 142, 150, 400, 400VP, 401, 402, 402VP, 403, 403VP, 404 & 410VP CONTACTING CONDUCTIVITY SENSOR OR ANY SIMPLE APPARATUS WITH LESS THAN 200 FEET OF ATTACHED CABLE TB7 +24V OUTPUT2 SHLD RTDIN SENSE Intrinsically Safe Installation TB7 TB7 GND WARNING- SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY OR SUITABILITY FOR DIVISION 2. WARNING- TO PREVENT IGNITION OF FLAMMABLE OR COMBUSTIBLE ATMOSPHERES, DISCONNECT POWER BEFORE SERVICING. 16 16 15 16 15 15 16 15 17 17 17 17 ROSEMOUNT MODEL 375 OR 475 FIELD COMMUNICATOR REMOTE TRANSMITTER INTERFACE FOR USE IN CLASS I AREA (SEE NOTE 3 AND TABLE III) ANALOG OUTPUT 2 ONLY AVAILABLE ON 1066...HT... NI CLASS I, DIV 2 GRPS A-D CLASS II, DIV 2 GRPS E-G IS CLASS I, GRPS A-D CLASS II, GRPS E-G CLASS III LOAD LOAD ANALOG OUTPUT 2 ONLY AVAILABLE ON 1066....HT... D DWG NO UNSPECIFIED POWER SUPPLY 30 VDC MAX FOR INTRINSIC SAFETY (24 VDC TYPICAL) 17.5 VDC MAX. FOR FISCO OPTION UNSPECIFIED POWER SUPPLY 30 VDC MAX FOR INTRINSIC SAFETY (24 VDC TYPICAL) 17.5 VDC MAX. FOR FISCO OPTION ANALOG OUTPUT 2 ONLY AVAILABLE ON 1066...HT... SAFETY BARRIER (SEE NOTES 2 & 8 FOR FISCO, SEE NOTES 2, 8, 9 & 10 FOR ALL OTHER OPTIONS.) SAFETY BARRIER (SEE NOTES 2 & 8 FOR FISCO, SEE NOTES 2, 8, 9 & 10 FOR ALL OTHER OPTIONS.) E REV LIQ-MAN-1066 THUM Instruction Manual Section 5: Intrinsically Safe Installation April 2017 FIGuRE 5-3. CSa Installation 29 30 NON-INCENDIVE FIELD WIRING METHODS MAY BE USED FOR CONNECTING SENSORS TO THE INSTRUMENT. ATTACHED SENSORS MUST BE FM APPROVED DIVISION 2 WIRING METHOD PER THE NEC (EXCLUDING NONINCENDIVE FIELD WIRING). METAL CONDUIT IS NOT REQUIRED FOR INTRINSICALLY SAFE INSTALLATIONS. HOWEVER, IF CONDUIT IS USED, BONDING BETWEEN CONDUIT IS NOT AUTOMATIC AND MUST BE PROVIDED AS PART OF THE INSTALLATION. NO REVISION TO DRAWING WITHOUT PRIOR FM APPROVAL. AGENCY CONTROLLED DOCUMENT. THE ASSOCIATED APPARATUS MUST BE FM APPROVED AND MUST BE RESISTIVELY LIMITED HAVING LINEAR OUTPUTS. CONTROL EQUIPMENT CONNECTED TO ASSOCIATED APPARATUS MUST NOT USE OR GENERATE MORE THAN 250 Vrms OR Vdc. THE INTRINSICALLY SAFE ENTITY CONCEPT ALLOWS INTERCONNECTION OF INTRINSICALLY SAFE DEVICES WITH ASSOCIATED APPARATUS WHEN THE FOLLOWING IS TRUE: FIELD DEVICE INPUT ASSOCIATED APPARATUS OUTPUT RESISTANCE BETWEEN INTRINSICALLY SAFE GROUND AND EARTH GROUND MUST BE LESS THAN 1.0 Ohm. DUST-TIGHT CONDUIT SEAL MUST BE USED WHEN INSTALLED IN CLASS II AND CLASS III ENVIRONMENTS. CONTACTING CONDUCTIVITY SENSORS, AMPEROMETRIC AND pH SENSORS WITHOUT PREAMPS SHALL MEET THE REQUIREMENTS OF SIMPLE APPARATUS AS DEFINED IN ANSI/ISA RP12.6 AND THE NEC, ANSI/NFPA 70. THEY CAN NOT GENERATE NOR STORE MORE THAN 1.5 V, 100 mA, 25 mW OR A PASSIVE COMPONENT THAT DOES NOT DISSIPATE MORE THAN 1.3W. WHEN CONNECTIONS ARE MADE TO 1066...HT... OPTION ANALOG OUTPUT 2 (TB7-1 & -2), SEPARATE WIRING AND A SECOND BARRIER ARE REQUIRED. THE WIRING FROM EACH BARRIER MUST BE INSTALLED AS SEPARATE INTRINSICALLY SAFE CIRCUITS IN ACCORDANCE WITH THE NATIONAL ELECTRICAL CODE. INTRINSICALLY SAFE APPARATUS (MODEL 1066, SMART THUM WIRELESS ADAPTER, MODEL 375, 475) AND ASSOCIATED APPARATUS (SAFETY BARRIER) 16 15 14. 13. 12. 11. 9. 8. 7. 6. 4. 3. NOTES: UNLESS OTHERWISE SPECIFIED ANY SINGLE SHUNT ZENER DIODE SAFETY BARRIER APPROVED BY FM HAVING THE FOLLOWING OUTPUT PARAMETERS: SUPPLY/SIGNAL TERMINALS TB6-1 AND 2 FOR FIELDBUS OPTION OR TB6-1, 2 AND 3 FOR HART OPTION. ALSO TB7-1 AND 2 IF ANALOG OUTPUT 2 IS USED. TO THE SENSOR TERMINALS DO NOT EXCEED THE VALUES SPECIFIED IN TABLE I WHERE: 1066 MODELS WITH P/L/D/Z OPTIONS HAVE OUTPUT ENTITY PARAMETERS WHICH ALLOW THE USE OF VARIOUS SENSORS WHICH MAY THE MODEL 1066-C/T HAS SYSTEM APPROVAL FOR USE WITH MODELS 222, 225, 226 & 228 TOROIDAL SENSORS OR 140, 141, 142, 150, 400, 401, 402, 20. FF/FI MODEL NO. 2.22 15.5 68.9 HT 5.86 23.5 47 FF SIGNAL SIGNAL SIGNAL 0.795 3.17 6.36 HT OUTPUT PARAMETERS FI TABLE III Uo (Vt) Volts OUTPUT PARAMETERS HT FF 10.69 11.33 211.2 77.77 250 110.12 FI SCALE: 1:1 D SIZE DWG NO WEIGHT: SHEET 1 OF 3 1400670 ONE OF THE ALLOWED TERMINATIONS MIGHT ALREADY BE INTEGRATED IN THE ASSOCIATED APPARATUS. THE NUMBER OF PASSIVE APPARATUS CONNECTED TO THE BUS SEGMENT IS NOT LIMITED DUE TO I.S. REASONS. IF THE ABOVE RULES ARE RESPECTED, UP TO A TOTAL LENGTH OF 1000 m (SUM OF TRUNK AND ALL SPUR CABLES) OF CABLE IS PERMITTED. THE INDUCTANCE AND THE CAPACITANCE OF THE CABLE WILL NOT IMPAIR THE INTRINSIC SAFETY OF THE INSTALLATION. AT EACH END OF THE TRUNK CABLE AN APPROVED INFALLIBLE LINE TERMINATION WITH THE FOLLOWING PARAMETERS IS SUITABLE: F THE CABLE USED TO INTERCONNECT DEVICES NEEDS TO HAVE THE PARAMETERS IN THE FOLLOWING RANGE: IN EACH SEGMENT ONLY ONE ACTIVE DEVICE, NORMALLY THE ASSOCIATED APPARATUS, IS ALLOWED TO PROVIDE THE NECESSARY ENERGY FOR THE FIELDBUS SYSTEM. THE VOLTAGE Uo (OR Voc OR Vt) OF THE ASSOCIATED APPARATUS IS LIMITED TO A RANGE OF 14 V TO 24 Vdc. ALL OTHER EQUIPMENT CONNECTED TO THE BUS CABLE HAS TO BE PASSIVE, MEANING THAT THEY ARE NOT ALLOWED TO PROVIDE ENERGY TO THE SYSTEM, EXCEPT A LEAKAGE CURRENT OF 50 FOR EACH CONNECTED DEVICE. SEPARATELY POWERED EQUIPMENT NEEDS GALVANIC ISOLATION TO ASSURE THAT THE INTRINSICALLY SAFE FIELDBUS CIRCUIT REMAINS PASSIVE. DELIVERED BY THE ASSOCIATED APPARATUS, CONSIDERING FAULTS AND APPLICABLE FACTORS. IN ADDITION, THE MAXIMUM UNPROTECTED CAPACITANCE (Ci) AND THE INDUCTANCE (Li) OF EACH APPARATUS (OTHER THAN THE TERMINATION) CONNECTED TO THE FIELDBUS MUST BE LESS THAN OR EQUAL TO 5nF and 10 H RESPECTIVELY. THE FISCO CONCEPT ALLOWS INTERCONNECTION OF INTRINSICALLY SAFE APPARATUS TO ASSOCIATED APPARATUS NOT SPECIALLY EXAMINED MODEL NO. II C II B II A GAS GROUPS E REV April 2017 1. 2. INSTALLATION TO BE IN ACCORDANCE WITH THE NATIONAL ELECTRICAL CODE. 17 OR GREATER THAN THE VOLTAGE (Voc OR Vt) AND CURRENT (Isc OR It) WHICH CAN BE DEVELOPED BY THE ASSOCIATED APPARATUS (SAFETY BARRIER). IN ADDITION, THE MAXIMUM UNPROTECTED CAPACITANCE (Ci) AND INDUCTANCE (Li) OF THE INTRINSICALLY SAFE APPARATUS, INCLUDING INTERCONNECTING WIRING, MUST BE EQUAL OR LESS THAN THE CAPACITANCE (Ca) AND INDUCTANCE (La) WHICH CAN BE SAFELY CONNECTED TO THE APPARATUS. (REF. TABLES I, II AND III). A 1066 WITH THE -FI OPTION MAY BE INSTALLED PER THE FISCO INSTRUCTIONS OR PER THE INTRINSICALLY SAFE INSTRUCTIONS ON THIS DRAWING. 19. AGENCY CONTROLLED DOCUMENT ANY CHANGE WILL REQUIRE CERTIFICATION AGENCY SUBMITTAL/APPROVAL APPROVED MODELS 1066-AA-BB-CC XMTR Section 5: Intrinsically Safe Installation Instruction Manual LIQ-MAN-1066 FIGuRE 5-6. FM installation Intrinsically Safe Installation NOTES: UNLESS OTHERWISE SPECIFIED MODEL 775 SMART THUM WIRELESS ADAPTER WHITE YELLOW RED GREEN ROSEMOUNT MODEL 375 OR 475 FIELD COMMUNICATOR REMOTE TRANSMITTER INTERFACE FOR USE IN CLASS I AREA (SEE NOTE 3 AND TABLE III) REF SHLD GND SOL SHLD pH (SENSOR AND SECOND ANALOG OUTPUT CONNECTION UNCHANGED FROM ABOVE) TB2 ALTERNATE POWER CONNECTION IF SMART THUM WIRELESS ADAPTER IS USED SENSOR WIRING TB3 RTN TB1 1066-CL/DO/OZ... ONLY TB5 TB4 -V +V SNS RTD IN AMPEROMETRIC SENSOR FM APPROVED DEVICE OR SIMPLE APPARATUS OR ANODE CATHODE EITHER OR BOTH MAY BE INSTALLED OPTIONAL FOR FIELDBUS /FISCO INSTALLATIONS +24V GND (OUTPUT1) LOOP PWR TB6 SHLD REF SHLD GND SOL OPTIONAL FM APPROVED PREAMP THAT MEETS REQUIREMENTS OF NOTE 9 SPLICE CONNECTOR OUTPUT 2 +24V GND TB2 pH PH SENSOR FM APPROVED DEVICE OR SIMPLE APPARATUS TB7 TB1 SENSOR WIRING TB3 SNS RTN TB5 TB4 -V +V ANODE RTD IN CATHODE RECOMMENDED CABLE PN 9200273 (UNPREPED) PN 23646-01 (PREPPED) 10 COND, 2 SHIELDS, 24 AWG. SEE NOTE 2 GND +24V (OUTPUT1) LOOP PWR TB6 OPTIONAL FOR FIELDBUS /FISCO INSTALLATIONS OUTPUT 2 +24V GND AMPEROMETRIC SENSOR FM APPROVED DEVICE OR SIMPLE APPARATUS TB7 SHLD REF SHLD GND SOL 1066-CL/DO/OZ... ONLY TB2 pH EITHER OR BOTH MAY BE INSTALLED TB1 SENSOR WIRING TB3 SNS RTN TB5 TB4 -V +V ANODE RTD IN CATHODE OR THUM (OUTPUT1) LOOP PWR +24V GND PH SENSOR FM APPROVED DEVICE OR SIMPLE APPARATUS OUTPUT 2 TB6 THUM GND +24V Intrinsically Safe Installation TB7 ANALOG OUTPUT 2 ONLY AVAILABLE ON 1066...HT... MULTIPLE DEVICES (NUMBER IS LIMITED BY THE REQUIREMENT TO MEET ALL OTHER FIELDBUS I.S. & FISCO REQUIREMENTS FOR THE NETWORK.) ANY FM APPROVED TERMINATOR ANY FM APPROVED FF/FISCO DEVICE ANY FM APPROVED FF/FISCO DEVICE MULTIPLE DEVICES (NUMBER IS LIMITED BY THE REQUIREMENT TO MEET ALL OTHER FIELDBUS I.S. & FISCO REQUIREMENTS FOR THE NETWORK.) ROSEMOUNT MODEL 375 OR 475 FIELD COMMUNICATOR REMOTE TRANSMITTER INTERFACE FOR USE IN CLASS I AREA (SEE NOTE 3 AND TABLE III) ANY FM APPROVED TERMINATOR ANY FM APPROVED FF/FISCO DEVICE ANY FM APPROVED FF/FISCO DEVICE ROSEMOUNT MODEL 375 OR 475 FIELD COMMUNICATOR REMOTE TRANSMITTER INTERFACE FOR USE IN CLASS I AREA (SEE NOTE 3 AND TABLE III) ANALOG OUTPUT 2 ONLY AVAILABLE ON 1066...HT... NI CLASS I, DIV 2 GRPS A-D CLASS II, DIV 2 GRPS E-G IS CLASS I, GRPS A-D CLASS II, GRPS E-G CLASS III HAZARDOUS AREA WARNING- SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY OR SUITABILITY FOR DIVISION 2. WARNING- TO PREVENT IGNITION OF FLAMMABLE OR COMBUSTIBLE ATMOSPHERES, DISCONNECT POWER BEFORE SERVICING. SAFETY BARRIER (SEE NOTES 10 & 11) SAFETY BARRIER (SEE NOTES 2,8,20 FOR FISCO INSTALLATIONS, SEE NOTES 2, 8, 9 & 10 FOR ANY OTHER) SAFETY BARRIER ( SEE NOTES 2, 8, 9 & 10 FOR ALL OTHER OPTIONS.) SAFETY BARRIER (SEE NOTES 2,8,20 FOR FISCO INSTALLATIONS, SEE NOTES 2, 8, 9 & 10 FOR ANY OTHER) SAFETY BARRIER ( SEE NOTES 2, 8, 9 & 10 FOR ALL OTHER OPTIONS.) LOAD D SCALE: 1:1 SIZE DWG NO WEIGHT: SHEET 2 OF 3 1400670 E REV UNSPECIFIED POWER SUPPLY 30 VDC MAX FOR INTRINSIC SAFETY (24 VDC TYPICAL) 17.5 VDC MAX. FOR FISCO OPTION UNSPECIFIED POWER SUPPLY 30 VDC MAX FOR INTRINSIC SAFETY (24 VDC TYPICAL) UNSPECIFIED POWER SUPPLY 30 VDC MAX FOR INTRINSIC SAFETY (24 VDC TYPICAL) 17.5 VDC MAX. FOR FISCO OPTION UNSPECIFIED POWER SUPPLY 30 VDC MAX FOR IS 24V TYPICAL LOAD LOAD LOAD LOAD UNSPECIFIED POWER SUPPLY 30 VDC MAX FOR INTRINSIC SAFETY (24 VDC TYPICAL) NON-HAZARDOUS AREA LIQ-MAN-1066 THUM Instruction Manual Section 5: Intrinsically Safe Installation April 2017 FIGuRE 5-7. FM installation 31 NOTES: UNLESS OTHERWISE SPECIFIED 1066-CL/DO/OZ... ONLY APPROVED FOR USE WITH ROSEMOUNT MODELS: 222, 225, 226 & 228 TOROIDAL CONDUCTIVITY SENSORS OR ROSEMOUNT MODELS: 140,141, 142, 150, 400, 401, 402, 402VP, 403, 403VP, 404 & 410VP CONTACTING CONDUCTIVITY SENSORS. NON-INCENDIVE FIELD WIRING CONNECTIONS FOR CLASS1, DIVISION 2, GROUPS ABCD 16 16 17 17 RSHLD AMPEROMETRIC SENSOR FM APPROVED DEVICE OR SIMPLE APPARATUS OR TB2 DSHLD REF SHLD GND SOL SHLD pH EITHER OR BOTH MAY BE INSTALLED RTN TB2 17 SHLD RTDIN SENSE TB1 SENSOR WIRING TB3 SNS RTN TB5 TB4 -V +V ANODE RTD IN CATHODE 16 RSHLD PH SENSOR FM APPROVED DEVICE OR SIMPLE APPARATUS THUM GND THUM LOOP PWR TB2 OPTIONAL FOR FIELDBUS /FISCO INSTALLATIONS THUM +24V GND (OUTPUT1) LOOP PWR TB6 GND OUTPUT2 GND OUTPUT 2 +24V GND RTN DSHLD APPROVED FOR USE WITH ROSEMOUNT MODELS: 222, 225, 226 & 228 TOROIDAL CONDUCTIVITY SENSORS OR ROSEMOUNT MODELS: 140,141, 142, 150, 400, 401, 402, 402VP, 403, 403VP, 404 & 410VP CONTACTING CONDUCTIVITY SENSORS. NI CLASS I, DIV 2 GRPS A-D CLASS II, DIV 2 GRPS E-G IS CLASS I, GRPS A-D CLASS II, GRPS E-G CLASS III 17 17 15 17 15 15 17 15 ANY FM APPROVED TERMINATOR ANY FM APPROVED FF/FISCO DEVICE MULTIPLE DEVICES (NUMBER IS LIMITED BY THE REQUIREMENT TO MEET ALL OTHER FIELDBUS I.S. & FISCO REQUIREMENTS FOR THE NETWORK.) ROSEMOUNT MODEL 375 OR 475 FIELD COMMUNICATOR REMOTE TRANSMITTER INTERFACE FOR USE IN CLASS I AREA (SEE NOTE 3 AND TABLE III) ANALOG OUTPUT 2 ONLY AVAILABLE ON 1066...HT... ANY FM APPROVED FF/FISCO DEVICE TB7 LOOP PWR SHLD RTDIN SENSE 32 OUTPUT2 GND WARNING- SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY OR SUITABILITY FOR DIVISION 2. WARNING- TO PREVENT IGNITION OF FLAMMABLE OR COMBUSTIBLE ATMOSPHERES, DISCONNECT POWER BEFORE SERVICING. LOAD LOAD D WEIGHT: SHEET 3 OF 3 E REV April 2017 SCALE: 1:1 SIZE DWG NO ANALOG OUTPUT 2 ONLY AVAILABLE ON 1066...HT... 1400670 UNSPECIFIED POWER SUPPLY 30 VDC MAX FOR INTRINSIC SAFETY (24 VDC TYPICAL) 17.5 VDC MAX. FOR FISCO OPTION UNSPECIFIED POWER SUPPLY 30 VDC MAX FOR INTRINSIC SAFETY (24 VDC TYPICAL) ANALOG OUTPUT 2 ONLY AVAILABLE ON 1066...HT... SAFETY BARRIER (SEE NOTES 2,8,20 FOR FISCO INSTALLATIONS, SEE NOTES 2, 8, 9 & 10 FOR ANY OTHER) SAFETY BARRIER ( SEE NOTES 2, 8, 9 & 10 FOR ALL OTHER OPTIONS.) Section 5: Intrinsically Safe Installation Instruction Manual LIQ-MAN-1066 FIGuRE 5-8. FM installation Intrinsically Safe Installation Instruction Manual LIQ-MAN-1066 Section 6: Display and Operation April 2017 Section 6: Display and Operation 6.1 6.2 user Interface The 1066 has a large display which shows the measurement readout and temperature in large digits and up to four additional process variables or diagnostic parameters concurrently. The displayed variables can be customized to meet user requirements. This is called display Format. The intuitive menu system allows access to Calibration, Hold (of current outputs), Programming, and Display functions. In addition, a dedicated DIAG button is available to provide access to useful operational information on installed sensor(s) and any problematic conditions that might occur. The display flashes Fault and/or Warning when these conditions occur. Help screens are displayed for most fault and warning conditions to guide the user in troubleshooting. During calibration and programming, key presses cause different displays to appear. The displays are selfexplanatory and guide the user step-by-step through the procedure. Instrument Keyboard There are four Function keys and four Selection keys on the instrument keypad. Function Keys: The MENu key is used to access menus for programming and calibrating the instrument. Four toplevel menu items appear when pressing the MENU key: • Calibrate: calibrate the attached sensor and analog output(s). • Hold: Suspend current output(s). • Program: Program outputs, measurement, temperature, security and reset. • Display: Program display format, language, warnings, and contrast Pressing MENu from the main (live values) screen always causes the main menu screen to appear. Pressing MENU followed by EXIT causes the main screen to appear. Pressing the DIaG key displays active Faults and Warnings, and provides detailed instrument information and sensor diagnostics including: Faults, Warnings, Sensor information, Out 1 and Out 2 live current values, model configuration string e.g. 1066-P-HT-60 and Instrument Software version. Pressing DIAG provides useful diagnostics and information (as applicable): Measurement, Sensor Type, Raw signal value, Cell constant, Zero Offset, Temperature, Temperature Offset, selected measurement range, Cable Resistance, Temperature Sensor Resistance, software version. The ENTER key. Pressing ENTER stores numbers and settings and moves the display to the next screen. The ExIT key. Pressing EXIT returns to the previous screen without storing changes. Display and Operation 33 Section 6: Display and Operation April 2017 Instruction Manual LIQ-MAN-1066 Selection Keys: Surrounding the ENTER key, four Selection keys – up, down, right and left, move the cursor to all areas of the screen while using the menus. Selection keys are used to: 6.3 1. 2. 3. 4. 5. Select items on the menu screens Scroll up and down the menu lists Enter or edit numeric values Move the cursor to the right or left Select measurement units during operations Main Display The 1066 displays the primary measurement value and temperature, and up to four secondary measurement values, a fault and warning banner, and a digital communications icon. process Measurements: One process variable and process temperature is displayed by default. For all configurations, the Upper display area shows the live process variable and the Center display area shows the Temperature (default screen settings). Secondary values: Up to four secondary values are shown in display quadrants at the bottom half of the screen. All four secondary value positions can be programmed by the user to any displayable parameter available. Fault and Warning Banner: If the transmitter detects a problem with itself or the sensor the word Fault or Warning will appear at the bottom of the display. A fault requires immediate attention. A warning indicates a problematic condition or an impending failure. For troubleshooting assistance, press Diag. Formatting the Main Display The main display screen can be programmed to show primary process variables, secondary process variables and diagnostics. 1. Press MENU 2. Scroll down to Display. Press ENTER. 3. Main Format will be highlighted. Press ENTER. 4. The sensor 1 process value will be highlighted in reverse video. Press the selection keys to navigate down to the screen sections that you wish to program. Press ENTER. 5. Choose the desired display parameter or diagnostic for each of the four display sections in the lower screen. 6. Continue to navigate and program all desired screen sections. Press MENu and ExIT. The screen will return to the main display. 34 Display and Operation Instruction Manual Section 6: Display and Operation LIQ-MAN-1066 6.4 April 2017 The default display shows the live process measurement in the upper display area and temperature in the center display area. The user can elect to disable the display of temperature in the center display area using the Main Format function. See Figure 4-1 to guide you through programming the main display to select process parameters and diagnostics of your choice. Menu System The 1066 uses a scroll and select menu system. Pressing the MENU key at any time opens the top-level menu including Calibrate, Hold, Program and Display functions. To find a menu item, scroll with the up and down keys until the item is highlighted. Continue to scroll and select menu items until the desired function is chosen. To select the item, press ENTER. To return to a previous menu level or to enable the main live display, press the EXIT key repeatedly. To return immediately to the main display from any menu level, simply press MENU then EXIT. The selection keys have the following functions: • • • • The Up key (above ENTER) increments numerical values, moves the decimal place one place to the right, or selects units of measurement. The Down key (below ENTER) decrements numerical values, moves the decimal place one place to the left, or selects units of measurement The Left key (left of ENTER) moves the cursor to the left. The Right key (right of ENTER) moves the cursor to the right. To access desired menu functions, use the Quick Reference. During all menu displays (except main display format and Quick Start), the live process measurement and temperature value are displayed in the top two lines of the Upper display area. This conveniently allows display of the live values during important calibration and programming operations. Menu screens will time out after two minutes and return to the main live display. Display and Operation 35 Section 6: Display and Operation April 2017 36 Instruction Manual LIQ-MAN-1066 Display and Operation Instruction Manual Section 7: Transmitter programming LIQ-MAN-1066 April 2017 Section 7: Programming the Transmitter – Basics 7.1 General 7.2 Changing Startup Settings 7.2.1 Typical programming steps include the following listed procedures. Each of these programming functions are easily and quickly accomplished using the intuitive menu system. • Changing the measurement type, measurement units and temperature units. • Choose temperature units and manual or automatic temperature compensation mode • Configure and assign values to the current outputs • Set a security code for two levels of security access • Accessing menu functions using a security code • Enabling and disabling Hold mode for current outputs • Resetting all factory defaults, calibration data only, or current output settings only Purpose To change the measurement type, measurement units, or temperature units that were initially entered in Quick Start, choose the Reset analyzer function or access the Program menus for the sensor. The following choices for specific measurement type, measurement units are available for each sensor measurement board. TaBlE 7-1. Measurements and Measurement units Signal board pH/ORP – P Contacting conductivity - C Toroidal conductivity - T Chlorine - CL Oxygen - DO Ozone - OZ Transmitter Programming available measurements pH, ORP, Redox Conductivity, Resistivity, TDS, Salinity, NaOH (0-12%), HCl (0-15%), Low H2SO4, High H2SO4, NaCl (0-20%), Custom Curve Conductivity, Resistivity, TDS, Salinity, NaOH (0-12%), HCl (0-15%), Low H2SO4, High H2SO4, NaCl (0-20%), Custom Curve Free Chlorine, Total Chlorine, Monochloramine Oxygen (ppm), Trace Oxygen (ppb), Percent Oxygen in gas Ozone Measurements units: pH, mV (ORP, Redox) μS/cm, mS/cm, S/cm % (concentration) μS/cm, mS/cm, S/cm % (concentration) ppm, mg/L ppm, mg/L, ppb, µg/L % Sat, Partial Pressure, % Oxygen In Gas, ppm Oxygen In Gas ppm, mg/L, ppb, μg/L 37 Section 7: Transmitter programming April 2017 Instruction Manual LIQ-MAN-1066 7.2.2 Procedure 7.3 Choosing Temperature units and automatic/Manual Temperature Compensation 7.3.1 7.4 7.4.1 7.4.2 7.4.3 38 Follow the Reset Analyzer procedure (Section 7.8) to reconfigure the transmitter to display new measurements or measurement units. To change the specific measurement or measurement units for each measurement type, refer to the Program menu for the appropriate measurement (Section 6.0). Purpose Most liquid analytical measurements (except ORP and Redox) require temperature compensation. The 1066 performs temperature compensation automatically by applying internal temperature correction algorithms. Temperature correction can also be turned off. If temperature correction is off, the 1066 uses the temperature entered by the user in all temperature correction calculations. 1.234 µS/cm 25.0 ºC Temperature Units: °C Temp Comp: Auto Manual: +25.0 °C Configuring and Ranging Current Outputs Purpose The 1066 has two analog current outputs. Ranging the outputs means assigning values to the low (4 mA) and high (20 mA) outputs. This section provides a guide for configuring and ranging the outputs. ALWAYS CONFIGURE THE OUTPUTS FIRST. Definitions 1. CURRENT OUTPUTS. The transmitter provides a continuous output current (4-20 mA) directly proportional to the process variable or temperature. The low and high current outputs can be set to any value. 2. ASSIGNING OUTPUTS. Assign a measurement or temperature to Output 1 or Output 2. 3. DAMPEN. Output dampening smooths out noisy readings. It also increases the response time of the output. Output dampening does not affect the response time of the display. 4. MODE. The current output can be made directly proportional to the displayed value (linear mode) or directly proportional to the common logarithm of the displayed value (log mode). Procedure: Configure Outputs Under the Program/Outputs menu, the adjacent screen will appear to allow configuration of the outputs. Follow the menu screens in Figure 7-1 to configure the outputs. 1.234 µS/cm 25.0 ºC OutputM Configure Assign: Meas Scale: Linear Dampening: 0sec Fault Mode: Fixed Fault Value: 21.00 mA Transmitter Programming Instruction Manual Section 7: Transmitter programming LIQ-MAN-1066 April 2017 FIGuRE 7-1. Configuring and Ranging the Current Outputs 1.234µS/cm Program MAIN MENU Range 1.234µS/cm 25.0ºC 1.234µS/cm Output Range O1 SN 4mA: O1 SN 20mA: O2 SN 4mA: S2 20mA: 0.000 20.00 0ºC 100ºC 25.0ºC Output Range OM 4mA: 1.000 OM 20mA: 1.000% OM 4mA: 1.000% OM 20mA: 7.0 pH 25.0ºC 1.234µS/cm Program Outputs Outputs Measurement Temperature 25.0ºC 1.234µS/cm Outputs Range Configure Simulate Security Diagnostic Setup Reset Analyzer 25.0ºC Output Configure Configure Output1 Output2 1.234µS/cm 1.234µS/cm 25.0ºC OutputM Configure Assign: Range: Scale: Dampening: Simulate 25.0ºC OutputM Assign S1 Measurement S1 Temperature S1 Meas 4-20mA Linear 0sec 1.234µS/cm 25.0ºC OutputM Range 1.234µS/cm 25.0ºC 1.234µS/cm Simulate Output 1 Output 2 4-20mA 0-20mA 25.0ºC OutputN Hold At 12.00mA 1.234µS/cm 25.0ºC Scale Linear Log 7.4.4 Procedure: Ranging the Current Outputs 7.5 Setting a Security Code 7.5.1 7.5.2 The adjacent screen will appear under Program/Output/Range. Enter a value for 4 mA and 20 mA for each output. Follow the menu screens in Figure 7-1 to assign values to the outputs. 1.234µS/cm O1 O1 O2 O2 SN SN SN SN 25.0 ºC Output 4mA: 20mA: 4mA: 20mA: Range 0.000 µS/cm 20.00 µS/cm 0 °C 100 °C Purpose The security codes prevent accidental or unwanted changes to program settings, displays, and calibration. The 1066 has two levels of security code to control access and use of the instrument to different types of users. The two levels of security are: • all: This is the Supervisory security level. It allows access to all menu functions, including Programming, Calibration, Hold and Display. • Calibration/hold: This is the operator or technician level menu. It allows access to only calibration and Hold of the current outputs. Procedure 1. Press MENU. The main menu screen appears. Choose program. 2. Scroll down to Security. Select Security. 3. The security entry screen appears. Enter a three digit security code for each of the desired security levels. The security code takes effect two minutes after the last key stroke. Record the security code(s) for future access and communication to operators or technicians as needed. 4. The display returns to the security menu screen. Press EXIT to return to the previous screen. To return to the main display, press MENU followed by EXIT. Figure 7-2 displays the security code screens. Transmitter Programming 39 Section 7: Transmitter programming Instruction Manual April 2017 LIQ-MAN-1066 Program MAIN MENU FIGuRE 7-2. Setting a Security Code 1.234 µS/cm 25.0 ºC Program Outputs Measurement Temperature Security Diagnostic Setup Reset Analyzer 7.6 7.6.1 How the Security Code Works When entering the correct access code for the Calibration/Hold security level, the Calibration and Hold menus are accessible. This allows operators or technicians to perform routine maintenance. This security level does not allow access to the Program or Display menus. When entering the correct access code for All security level, the user has access to all menu functions, including Programming, Calibration, Hold and Display. Procedure 7.7 using hold 40 25.0 ºC Security Calibration/Hold: 000 All: 000 Security access 7.6.2 7.7.1 1.234 µS/cm 1. If a security code has been programmed, selecting the Calibrate, Hold, Program or Display top menu items causes the security access screen to appear 2. Enter the three-digit security code for the appropriate security level. 1.234 µS/cm 25.0 ºC Security Code 3. If the entry is correct, the appropriate menu screen appears. If the 000 entry is incorrect, the Invalid Code screen appears. The Enter Security Code screen reappears after 2 seconds. Purpose The transmitter output is always proportional to measured value. To prevent improper operation of systems or pumps that are controlled directly by the current output, place the analyzer in hold before removing the sensor for calibration and maintenance. Be sure to remove the transmitter from hold once calibration is complete. During hold, both outputs remain at the last value. Once in hold, all current outputs remain on Hold indefinitely. Transmitter Programming Instruction Manual Section 7: Transmitter programming LIQ-MAN-1066 April 2017 7.7.2 7.8 7.8.1 7.8.2 Hold MAIN MENU FIGuRE 7-3. using hold 1.234 µS/cm Hold: Hold 25.0 ºC No 1.234 µS/cm 25.0 ºC Hold outputs? No Yes Using the Hold Function To hold the outputs, 1. Press MENU. The main menu screen appears. Choose hold. 2. The hold Outputs? screen appears. Choose Yes to place the analyzer in hold. Choose No to take the analyzer out of hold. 3. The Hold screen will then appear and hold will remain on indefinitely until hold is disabled. See Figure 7-3. Resetting Factory Default Settings Purpose This section describes how to restore factory calibration and default values. The process also clears all fault messages and returns the display to the first Quick Start screen. The 1066 offers three options for resetting factory defaults. 1. Reset all settings to factory defaults 2. Reset sensor calibration data only 3. Reset analog output settings only Procedure To reset to factory defaults, reset calibration data only or reset analog outputs only, follow the Reset Analyzer flow diagram. Transmitter Programming 41 Section 7: Transmitter programming April 2017 Instruction Manual LIQ-MAN-1066 FIGuRE 7-4. Resetting Factory Default Settings 42 Transmitter Programming Instruction Manual LIQ-MAN-1066 Section 8: programming Measurements April 2017 Section 8: Programming Measurements 8.1 Introduction The 1066 automatically recognizes the measurement input upon first power-up and each time the transmitter is powered. Completion of Quick Start screens upon first power up enable measurements, but additional steps may be required to program the transmitter for the desired measurement application. This section covers the following programming and configuration functions: 1. Selecting measurement type or sensor type (all sections) 2. Identifying the preamp location (pH-see Section 8.2) 3. Enabling manual temperature correction and entering a reference temperature (all sections) 4. Enabling sample temperature correction and entering temperature correction slope (selected sections) 5. Defining measurement display resolution (pH and amperometric) 6. Defining measurement display units (all sections) 7. Adjusting the input filter to control display and output reading variability or noise (all sections) 8. Selecting a measurement range (conductivity – see Section 8.4, 8.5) 9. Entering a cell constant for a contacting or toroidal sensor (see Section 8.4, 8.5) 10. Creating an application-specific concentration curve (conductivity-see Section 8.4, 8.5) 11. Enabling automatic pH correction for free chlorine measurement (Section 8.6.1) To fully configure the transmitter, you may use the following: 1. Reset Transmitter function to reset factory defaults and configure to the desired measurement. Follow the Reset Transmitter menu to reconfigure the transmitter to display new measurements or measurement units. 2. Program menus to adjust any of the programmable configuration items. Use the following configuration and programming guidelines for the applicable measurement. Programming Measurements 43 Section 8: programming Measurements Instruction Manual April 2017 8.2 8.2.1 LIQ-MAN-1066 ph Measurement programming Description The section describes how to configure the 1066 transmitter for pH measurements. The following programming and configuration functions are covered. 1. Measurement type: pH Select pH, ORP, Redox. 2. Preamp location: Transmitter Identify preamp location (automatic detection for SMART pH sensors) 3. Filter: 4 sec Override the default input filter, enter 0-999 seconds 4. Reference Z: Low Select low or high reference impedance 5. Resolution: 0.01pH Select 0.01pH or 0.1pH for pH display resolution To configure pH: 1. Press MENu 2. Scroll down to program. Press ENTER. 3. Scroll down to Measurement. Press ENTER. The adjacent screen format will appear (factory defaults are shown). To program any function, scroll to the desired item and press ENTER. 1.234 µS/cm 25.0 ºC SN Configure Measure: pH Preamp: Analyzer Sol’n Temp Corr: Off Resolution: 0.01pH Filter: 4 sec Reference Z: Low The following sub-sections provide you with the initial display screen that appears for each configuration function. Use the flow diagram for pH programming and the 1066 live screen prompts for each function to complete configuration and programming. 8.2.2 8.2.3 Measurement The display screen for selecting the measurement is shown. The default value is displayed in bold t ype. Refer to the pH/ORP Programming flow diagram to complete this function. 25.0 ºC SN Measurement pH ORP Redox Preamp The display screen for identifying the Preamp location is shown. The default value is displayed in bold type. Refer to the pH/ORP Programming flow diagram to complete this function. Note: Sensor/JBox must be selected to support SMART pH sensors from Rosemount. 44 1.234 µS/cm 1.234 µS/cm 25.0 ºC SN Preamp Analyzer Sensor/JBox Programming Measurements Instruction Manual LIQ-MAN-1066 8.2.4 Solution Temperature Correction 8.2.5 Temperature Coefficient The display screen for selecting the Solution temperature correction algorithm is shown. The default value is displayed in bold type. Refer to the pH/ORP Programming flow diagram to complete this function. The display screen for entering the custom solution temperature coefficient is shown. The default value is displayed in bold type. Refer to the pH/ORP Programming flow diagram to complete this function. 8.2.6 Resolution 8.2.7 Filter 8.2.8 8.3 Section 8: programming Measurements The display screen for selecting 0.01pH or 0.1pH for pH display resolution is shown. The default value is displayed in bold type. Refer to the pH/ORP Programming flow diagram to complete this function. The display screen for entering the input filter value in seconds is shown. The default value is displayed in bold type. Refer to the pH/ORP Programming flow diagram to complete this function. Reference Impedance The display screen for selecting Low or High Reference impedance is shown. The default value is displayed in bold type. Refer to the pH/ORP Programming flow diagram to complete this function. April 2017 1.234 µS/cm 25.0 ºC SN Sol’n Temp Corr. Off Ultra Pure Water High pH Custom 1.234µS/cm 25.0ºC SN Sol’n Temp Coeff. - 0.032pH/ºC 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN Resolution 0.01pH 0.1pH SN Input filter 04 sec SN Reference Z Low High ORp Measurement programming The section describes how to configure the 1066 transmitter for ORP measurements. The following programming and configuration functions are covered: 1. Measurement type: pH Select pH, ORP, Redox. 2. Preamp location: Transmitter Identify preamp location 3. Filter: 4 sec Override the default input filter, enter 0-999 seconds 4. Reference Z: Low Select low or high reference impedance 5. Sensor wiring scheme: Normal or Reference to Ground Programming Measurements 45 Section 8: programming Measurements April 2017 To configure ORP: 1. Press MENu 2. Scroll down to program. Press ENTER. 3. Scroll down to Measurement. Press ENTER. Instruction Manual LIQ-MAN-1066 1.234 µS/cm 25.0 ºC SN Configure Measure: pH Preamp: Analyzer Flter: 4 sec Reference Z: Low The adjacent screen format will appear (factory defaults are shown). To program any displayed function, scroll to the desired item and press ENTER. The following sub-sections provide you with the initial display screen that appears for each configuration function. Use the flow diagram for ORp programming at the end of Sec. 6 and the 1066 live screen prompts for each function to complete configuration and programming. 8.3.1 Measurement 8.3.2 Preamp 8.3.3 8.3.4 46 The display screen for selecting the measurement is shown. The default value is displayed in bold type. Refer to the pH/ORP Programming flow diagram to complete this function. The display screen for identifying the Preamp location is shown. The default value is displayed in bold type. Refer to the pH/ORP Programming flow diagram to complete this function. Filter The display screen for entering the input filter value in seconds is shown. The default value is displayed in bold type. Refer to the pH/ORP Programming flow diagram to complete this function. Reference Impedance The display screen for Selecting Low or high Reference impedance is shown. The default value is displayed in bold type. Refer to the pH/ORP Programming flow diagram to complete this function. 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN Measurement pH ORP Redox SN Preamp Analyzer Sensor/JBox SN Input filter 04 sec SN Reference Z Low High Programming Measurements Instruction Manual LIQ-MAN-1066 8.4 8.4.1 Section 8: programming Measurements April 2017 Contacting Conductivity Measurement programming Description The section describes how to configure the 1066 transmitter for conductivity measurements using contacting conductivity sensors. The following programming and configuration functions are covered. 1. Measure: Conductivity Select Conductivity, Resistivity, TDS. Salinity or % conc 2. Type: 2-Electrode Select 2-Electrode or 4-Electrode type sensors 3. Cell K: 1.00000/cm Enter the cell Constant for the sensor 4. Measurement units 5. Filter: 2 sec Override the default input filter, enter 0-999 seconds 6. Range: Auto Select measurement Auto-range or specific range 7. Temp Comp: Slope Select Temp Comp: Slope, Neutral Salt, Cation or Raw 8. Slope: 2.00% / °C Enter the linear temperature coefficient 9. Ref Temp: 25.0 °C Enter the Reference temp 10. Cal Factor: default=0.95000/cm Enter the Cal Factor for 4-Electrode sensors from the sensor tag 1.234 µS/cm To configure the contacting conductivity: 1. Press MENU 2. Scroll down to Program. Press ENTER. 3. Scroll down to Measurement. Press ENTER. The adjacent screen format will appear (factory defaults are shown). To program any displayed function, scroll to the desired item and press ENTER. 25.0 ºC SN Configure Type: 2-Electrode Measure: Cond Range: Auto Cell K: 1.00000/cm RTD Offset: 0.00 ºC RTD Slope: 0 Temp Comp: Slope Slope: 2.00% / °C Ref Temp: 25.0 °C Filter: 2 sec Custom Setup The following sub-sections provide you with the initial display screen that appears for each configuration function. Use the flow diagram for contacting conductivity programming at the end of Section 8 and the 1066 live screen prompts for each function to complete configuration and programming. 8.4.2 Sensor Type The display screen for selecting 2-Electrode or 4-Electrode type sensors is shown. The default value is displayed in bold type. Refer to the contacting conductivity Programming flow diagram to complete this function. Programming Measurements 1.234 µS/cm 25.0 ºC SN Type 2-Electrode 4-Electrode 47 Section 8: programming Measurements April 2017 8.4.3 Measure The display screen for selecting the measurement is shown. The default value is displayed in bold type. Refer to the contacting conductivity Programming flow diagram to complete this function. Instruction Manual LIQ-MAN-1066 1.234 µS/cm 25.0 ºC SN Measurement Conductivity Resistivity TDS Salinity NaOH (0-12%) HCl (0-15%) Low H2SO4 High H2SO4 NaCl (0-20%) Custom Curve 8.4.4 Range 8.4.5 Cell Constant 8.4.6 RTD Offset 8.4.7 RTD Slope 8.4.8 48 The display screen for Selecting Auto-ranging or a specific range is shown. The default value is displayed in bold type. Note: Ranges are shown as conductance, not conductivity. Refer to the contacting conductivity Programming flow diagram to complete this function. The display screen for entering a cell Constant for the sensor is shown. The default value is displayed in bold type. Refer to the contacting conductivity Programming flow diagram to complete this function. The display screen for Entering the RTD Offset for the sensor is shown. The default value is displayed in bold type. Refer to the contacting conductivity Programming flow diagram to complete this function. The display screen for entering the RTD slope for the sensor is shown. The default value is displayed in bold type. Refer to the contacting conductivity Programming flow diagram to complete this function. Temp Comp The display screen for Selecting Temperature Compensation as Slope, Neutral Salt, Cation or Raw is shown. The default value is displayed in bold type. Refer to the contacting conductivity Programming flow diagram to complete this function. 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN Range Auto 50 µS 500 µS 2000 µS 20 mS 200 mS 600 mS SN Cell Constant 1.00000 /cm SN RTD Offset 0.00 °C 1.234µS/cm 25.0ºC SN RTD Slope 2.00%/ºC 1.234 µS/cm 25.0 ºC SN Temp Comp Slope Neutral Salt Cation Raw Programming Measurements Instruction Manual LIQ-MAN-1066 8.4.9 Slope 8.4.10 Reference Temp 8.4.11 8.4.12 8.4.13 Section 8: programming Measurements The display screen for Entering the conductivity/temp Slope is shown. The default value is displayed in bold type. Refer to the contacting conductivity Programming flow diagram to complete this function. The display screen for manually entering the Reference temperature is shown. The default value is displayed in bold type. Refer to the contacting conductivity Programming flow diagram to complete this function. Filter The display screen for entering the input filter value in seconds is shown. The default value is displayed in bold type. Refer to the contacting conductivity Programming flow diagram to complete this function. Custom Setup April 2017 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN Slope 2.00 %/ºC SN Ref Temp (25.0ºC normal) +25.0ºC 1.234 µS/cm 25.0 ºC SN Input filter 02 sec 1.234µS/cm 25.0ºC The display screens for creating a custom curve for converting conductivity to concentration is shown. Refer to the contacting conductivity Programming flow diagram to complete this function. SN Custom Curve Configure Enter Data Points Calculate Curve When the custom curve data entry is complete, press ENTER. The display will confirm the determination of a custom curve fit to the entered data by displaying this screen: SN Calculate Curve Custom curve fit completed. In Process Cal recommended. If the custom curve fit is not completed or is unsuccessful, the display will read as follows and the screen will return to the beginning custom curve screen. SN Calculate Curve Failure Cal Factor Upon initial installation and power up, if 4-electrode was selected for the sensor type in the Quick Start menus, the user enters a Cell Constant and a “Cal Factor” using the instrument keypad. The cell constant is needed to convert measured conductance to conductivity as displayed on the transmitter screen. The “Cal Factor” entry is needed increase the accuracy of the live conductivity readings, especially at low conductivity readings below 20 uS/cm. Both the Cell Constant and the “Cal Factor” are printed on the tag attached to the 4-electrode sensor/cable. Programming Measurements 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC 1.234µS/cm 25.0 ºC SN Cal Factor 0.95000/cm 49 Section 8: programming Measurements April 2017 8.5 8.5.1 Instruction Manual LIQ-MAN-1066 Toroidal Conductivity Measurement programming Description The section describes how to configure the 1066 transmitter for conductivity measurements using inductive/toroidal sensors. The following programming and configuration functions are covered: 1. Measure: Conductivity Select Conductivity, Resistivity, TDS. Salinity or % conc 2. Sensor Model: 228 Select sensor type 3. Measurement units 4. Range selection 5. Cell K: 3.00000/cm Enter the cell Constant for the sensors 6. Temp Comp: Slope Select Temp Comp: Slope, Neutral Salt, Cation or Raw 7. Slope: 2.00% / °C Enter the linear temperature coefficient 8. Ref Temp: 25.0 °C Enter the Reference temp 9. Filter: 2 sec Override the default input filter, enter 0-999 seconds To configure toroidal conductivity: 1. Press MENU 2. Scroll down to Program. Press ENTER. 3. Scroll down to Measurement. Press ENTER. The adjacent screen format will appear (factory defaults are shown). To program any displayed function, scroll to the desired item and press ENTER. 1.234 µS/cm 25.0 ºC SN Configure Model: 228 Measure: Cond Range: Auto Cell K: 3.00000/cm RTD Offset: 0.00 ºC RTD Slope: 0 Temp Comp: Slope Slope: 2.00% / °C Ref Temp: 25.0 °C Filter: 2 sec Custom Setup The following sub-sections provide you with the initial display screen that appears for each configuration function. Use the flow diagram for toroidal conductivity programming at the end of Section 8 and the 1066 live screen prompts for each function to complete configuration and programming. 8.5.2 50 Sensor Type The display screen for selecting the sensor model is shown. The default value is displayed in bold type. Refer to the toroidal conductivity Programming flow diagram to complete this function. 1.234 µS/cm 228 225 226 247 25.0 ºC SN Model Other Programming Measurements Instruction Manual LIQ-MAN-1066 8.5.3 8.5.4 Section 8: programming Measurements Measure The display screen for selecting the measurement is shown. The default value is displayed in bold type. Refer to the toroidal conductivity Programming flow diagram to complete this function. Range The display screen for Selecting Auto-ranging or a specific range is shown. The default value is displayed in bold type. Note: Ranges are shown as conductance, not conductivity. Refer to the toroidal conductivity Programming flow diagram to complete this function. NOTE: when manually changing ranges, a zero calibration and recalibration in a solution of known conductivity must be performed with the toroidal sensor wired to the instrument. Refer to Section 9.5.2 zeroing the Instrument and Section 9.5.3 Calibrating the Sensor in a Conductivity Standard. 8.5.5 1.234 µS/cm 25.0 ºC Measurement Conductivity Resistivity TDS Salinity NaOH (0-12%) HCl (0-15%) Low H2SO4 High H2SO4 NaCl (0-20%) Custom Curve 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN Range Auto 50-600 µS 0.5-100 mS 90-1500 mS Cell Constant The display screen for entering a cell Constant for the sensor is shown. The default value is displayed in bold type. Refer to the toroidal conductivity Programming flow diagram to complete this function. NOTE: When manually changing ranges, the Cell Constant may change through the calibration process. 8.5.6 April 2017 Temp Comp The display screen for Selecting Temperature Compensation as Slope, Neutral Salt, or Raw is shown. The default value is displayed in bold type. Refer to the toroidal conductivity Programming flow diagram to complete this function. Programming Measurements Cell Constant 2.7000 /cm Temp Comp Slope Raw 51 Section 8: programming Measurements April 2017 8.5.7 Slope 8.5.8 Ref Temp 8.5.9 Filter The display screen for Entering the conductivity/temp Slope is shown. The default value is displayed in bold type. Refer to the toroidal conductivity Programming flow diagram to complete this function. The display screen for manually Entering the Reference temperature is shown. The default value is displayed in bold type. Refer to the toroidal conductivity Programming flow diagram to complete this function. The display screen for entering the input filter value in seconds is shown. The default value is displayed in bold type. Refer to the toroidal conductivity Programming flow diagram to complete this function. using the highest range (90 mS to 1500 mS) in very low conductivity processes below 100 µS (conductance) might generate a high noise value relative to the actual process value. In these cases, it is recommended to increase the input filter setting above the default value of 2 sec. to suppress the effect of noise. 8.5.10 Custom Setup The display screens for creating custom curves for converting conductivity to concentration is shown. Refer to the toroidal conductivity Programming flow diagram to complete this function. When the custom curve data entry is complete, press ENTER. The display will confirm the determination of a custom curve fit to the entered data by displaying this screen: If the custom curve fit is not completed or is unsuccessful, the display will read as follows and the screen will return to the beginning custom curve screen. 52 Instruction Manual LIQ-MAN-1066 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN Slope 2.00%/ºC SN Ref Temp (25.0ºC normal) +25.0ºC SN Input filter 02 sec 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN Custom Curve Configure Enter Data Points Calculate Curve SN Calculate Curve Custom curve fit completed. In Process Cal recommended. SN Calculate Curve Failure Programming Measurements Instruction Manual Section 8: programming Measurements LIQ-MAN-1066 8.6 April 2017 Chlorine Measurement programming The 1066 can measure any of three variants of Chlorine: • Free Chlorine • Total Chlorine • Monochloramine The section describes how to configure the 1066 transmitter for Chlorine measurements. 8.6.1 Free Chlorine Measurement Programming This Chlorine sub-section describes how to configure the 1066 transmitter for Free Chlorine measurement using amperometric chlorine sensors. Automatic temperature compensation is available using Auto or Manual pH correction. For maximum accuracy, use automatic temperature compensation. The following programming and configuration functions are covered: 1. Measure: Free Chlorine Select Free Chlorine, Total Cl, Monochloramine 3 Resolution: 0.001 Select display resolution 0.01 or 0.001 2. 4. 5. 6. Units: ppm Select units ppm or mg/L Free Cl Correct: Live Select Live/Continuous pH correction or Manual Manual pH: 7.00 pH For Manual pH correction, enter the pH value Filter: 5sec Override the default input filter, enter 0-999 seconds To configure chlorine for free chlorine: 1. Press MENU 3. Scroll down to Measurement. Press ENTER. 2. Scroll down to Program. Press ENTER. The adjacent screen format will appear (factory defaults are shown). To program any displayed function, scroll to the desired item and press ENTER. 1.234 µS/cm 25.0 ºC SN Configure Measure: Free Chlorine Units: ppm Filter: 5sec Free Cl Correct: Live Manual pH: 7.00 pH Resolution: 0.001 The following sub-sections provide you with the initial display screen that appears for each configuration function. Use the flow diagram for chlorine programming at the end of Sec. 8 and the 1066 live screen prompts for each function to complete configuration and programming. Programming Measurements 53 Section 8: programming Measurements Instruction Manual 8.6.1.1 Measure 1.234 µS/cm 8.6.1.2 Units April 2017 8.6.1.3 The display screen for selecting the measurement is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. The display screen for selecting units as ppm or mg/L is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. Filter The display screen for entering the input filter value in seconds is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. 8.6.1.4 Free Chlorine pH Correction 8.6.1.5 Manual pH Correction 8.6.1.6 Resolution 54 The display screen for Selecting Live/Continuous pH correction or Manual pH correction is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. The display screen for manually entering the pH value of the measured process liquid is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. The display screen for selecting display resolution as 0.001 or 0.01 is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. LIQ-MAN-1066 25.0 ºC SN Measurement Free Chlorine Total Chlorine Monochloramine 1.234µS/cm ppm mg/L 25.0ºC SN Units 1.234 µS/cm 25.0 ºC SN Input filter 05 sec 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN Free Cl pH Correction Live/Continuous Manual SN Manual pH 07.00 pH SN Resolution 0.001 0.01 Programming Measurements Instruction Manual Section 8: programming Measurements LIQ-MAN-1066 8.6.2 8.6.2.1 April 2017 Total Chlorine Measurement Programming Description This Chlorine sub-section describes how to configure the 1066 transmitter for Total Chlorine measurement using amperometric chlorine sensors. The following programming and configuration functions are covered: 1. Measure: Free Chlorine Select Free Chlorine, pH Ind. Free Cl. Total Cl, Monochloramine 3. Resolution: 0.001 Select display resolution 0.01 or 0.001 2. 4. Units: ppm Select units ppm or mg/L Filter: 5sec Override the default input filter, enter 0-999 seconds To configure chlorine measurement for total chlorine: 1. Press MENU 3. Scroll down to Measurement. Press ENTER. 2. Scroll down to Program. Press ENTER. 1.234 µS/cm 25.0 ºC SN Configure Measure: Free Chlorine Units: ppm Filter: 5sec Resolution: 0.001 The adjacent screen format will appear (factory defaults are shown). To program any displayed function, scroll to the desired item and press ENTER. The following sub-sections provide you with the initial display screen that appears for each configuration function. Use the flow diagram for chlorine programming at the end of Sec. 6 and the 1066 live screen prompts for each function to complete configuration and programming. 8.6.2.2 Measure 8.6.2.3 Units 8.6.2.4 The display screen for selecting the measurement is shown. The default value is displayed in bold type. Refer to the chlorine Programming flow diagram to complete this function. The display screen for selecting units as ppm or mg/L is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function Filter The display screen for entering the input filter value in seconds is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. Programming Measurements 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN Measurement Free Chlorine Total Chlorine Monochloramine SN Units ppm mg/L 1.234 µS/cm 25.0 ºC SN Input filter 05 sec 55 Section 8: programming Measurements Instruction Manual April 2017 LIQ-MAN-1066 8.6.2.5 Resolution 8.6.3 Monochloramine Measurement Programming 1.234 µS/cm 25.0 ºC SN Resolution 0.001 0.01 The display screen for selecting display resolution as 0.001 or 0.01 is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. This Chlorine sub-section describes how to configure the 1066 transmitter for Monochloramine measurement using amperometric chlorine sensors. The following programming and configuration functions are covered: 1. Measure: Free Chlorine Select Free Chlorine, pH Ind. Free Cl. Total Cl, Monochloramine 3. Resolution: 0.001 Select display resolution 0.01 or 0.001 2. 4. Units: ppm Select units ppm or mg/L Filter: 5 sec Override the default input filter, enter 0-999 seconds To configure chlorine for monochloramine: 1. Press MENU 3. Scroll down to Measurement. Press ENTER. 2. Scroll down to Program. Press ENTER. 1.234 µS/cm 25.0 ºC SN Configure Measure: Free Chlorine Units: ppm 5 sec Filter: Resolution: 0.001 The following screen format will appear (factory defaults are shown). To program any displayed function, scroll to the desired item and press ENTER. The following sub-sections provide you with the initial display screen that appears for each configuration function. Use the flow diagram for chlorine programming at the end of Sec. 8 and the 1066 live screen prompts for each function to complete configuration and programming. 8.6.3.1 Measure: Monochloramine 8.6.3.2 Units 56 The display screen for selecting the measurement is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. The display screen for selecting units as ppm or mg/L is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN Measurement Free Chlorine Total Chlorine Monochloramine ppm mg/L SN Units Programming Measurements Instruction Manual LIQ-MAN-1066 8.6.3.3 Section 8: programming Measurements Filter The display screen for entering the input filter value in seconds is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. 8.6.3.4 Resolution 8.7 Oxygen Measurement programming The display screen for selecting display resolution as 0.001 or 0.01 is shown. The default value is displayed in bold type. Refer to the Chlorine Programming flow diagram to complete this function. April 2017 1.234 µS/cm 25.0 ºC SN Input filter 05 sec 1.234 µS/cm 25.0 ºC SN Resolution 0.001 0.01 This section describes how to configure the 1066 transmitter for dissolved and gaseous oxygen measurement using amperometric oxygen sensors. The following programming and configuration functions are covered: 1. Sensor type: Select Water/Waste, Trace. BioRx, BioRx-Other, Brew, %O2 In Gas 2. Measure type: Select Concentration, % Saturation, Partial Pressure, Oxygen in Gas 3. Units: ppm Select ppm, mg/L, ppb, g/L, % Sat, %O2-Gas, ppm Oxygen-Gas 4. Pressure Units: bar Select pressure units: mm Hg, in Hg, Atm, kPa, mbar, bar 5. Salinity: 00.0 %. Enter Salinity as %. 6. Filter: 5 sec Override the default input filter, enter 0-999 seconds 7. Partial Press: mmHg Select mm Hg, in Hg. atm, kPa, mbar or bar for Partial pressure To configure Oxygen: 1. Press MENU 2. Scroll down to Program. Press ENTER. 3. Scroll down to Measurement. Press ENTER. The adjacent screen format will appear (factory defaults are shown). To program any displayed function, scroll to the desired item and press ENTER. 1.234µS/cm 25.0ºC SN Configure Type: Water/Waste Units: ppm Partial Press: mmHg Salinity: 00.0‰ Filter: 5sec Pressure Units: bar Use Press: At Air Cal Custom Setup The following sub-sections show the initial display screen that appears for each configuration function. Use the flow diagram for oxygen programming at the end of Sec. 6 and the 1066 live screen prompts for each function to complete configuration and programming. Programming Measurements 57 Section 8: programming Measurements April 2017 8.7.1 8.7.2 Oxygen Measurement application The display screen for programming the measurement is shown. The default value is displayed in bold type. Refer to the Oxygen Programming flow diagram to complete this function. Units The display screen for selecting units as ppm , mg/L, ppb, µg/L, % Saturation, %Oxygen in Gas, or ppm Oxygen in Gas is shown. The default value is displayed in bold type. Refer to the Oxygen Programming flow diagram to complete this function. Instruction Manual LIQ-MAN-1066 1.234 µS/cm 25.0ºC SN Type Water/Waste Trace Oxygen BioRx-Rosemount BioRx-Other Brewing Oxygen In Gas 1.234 µS/cm ppm mg/L ppb µg/L 25.0 ºC SN Units % Saturation Partial Pressure % Oxygen In Gas ppm Oxygen In Gas 8.7.3 Partial Press 8.7.4 Salinity The display screen for selecting pressure units for Partial pressure is shown. This selection is needed if the specified measurement is Partial pressure. The default value is displayed in bold type. Refer to the Oxygen Programming flow diagram to complete this function. The display screen for Entering the Salinity (as parts per thousand) of the process liquid to be measured is shown. The default value is displayed in bold type. Refer to the Oxygen Programming flow diagram to complete this function. 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0ºC 1.234 µS/cm 25.0 ºC SN Partial Press mm Hg in Hg atm kPa mbar bar SN Salinity 00.0 ‰ Enter Salinity as ‰ 8.7.5 8.7.6 58 Filter The display screen for entering the input filter value in seconds is shown. The default value is displayed in bold type. Refer to the Oxygen Programming flow diagram to complete this function. Pressure Units The display screen for selecting pressure units for atmospheric pressure is shown. This selection is needed for the display of atmospheric pressure. The default value is displayed in bold type. The user must enter a known value for local atmospheric pressure. Refer to the Oxygen Programming flow diagram to complete this function. SN Input filter 05 sec 1.234 µS/cm 25.0 ºC Pressure Units mm Hg in Hg atm kPa mbar bar Programming Measurements Instruction Manual LIQ-MAN-1066 8.8 Section 8: programming Measurements April 2017 Ozone Measurement programming This section describes how to configure the 1066 transmitter for ozone measurement using amperometric ozone sensors. The following programming and configuration functions are covered: 1. Units: ppm Select ppm, mg/L, ppb, µg/L 2. Resolution: 0.001 Select display resolution 0.01 or 0.001 3. Filter: 5sec Override the default input filter, enter 0-999 seconds To configure Ozone: 1. Press MENU 2. Scroll down to Program. Press ENTER. 3. Scroll down to Measurement. Press ENTER. 1.234 µS/cm 25.0ºC SN Configure Units: ppm Filter: 5 sec Resolution: 0.001 The adjacent screen format will appear (factory defaults are shown). To program any displayed function, scroll to the desired item and press ENTER. The following sub-sections show the initial display screen that appears for each configuration function. Use the flow diagram for ozone programming at the end of Sec. 6 and the 1066 live screen prompts for each function to complete configuration and programming. Note: Ozone measurement boards are detected automatically by the analyzer. No measurement selection is necessary. 8.8.1 Units 8.8.2 Filter 8.8.3 The display screen for selecting measurement units is shown. The default value is displayed in bold type. Refer to the Ozone Programming flow diagram to complete this function. The display screen for entering the input filter value in seconds is shown. The default value is displayed in bold type. Refer to the Ozone Programming flow diagram to complete this function. Resolution The display screen for selecting display resolution as 0.001 or 0.01 is shown. The default value is displayed in bold type. Refer to the Ozone Programming flow diagram to complete this function. Programming Measurements 1.234 µS/cm ppm mg/L ppb µg/L 25.0 ºC SN Units 1.234 µS/cm 25.0 ºC SN Input filter 05 sec 1.234 µS/cm 25.0 ºC SN Resolution 0.001 0.01 59 Section 8: programming Measurements April 2017 Instruction Manual LIQ-MAN-1066 FIGuRE 8-1. Configuring ph/ORp Measurements 60 Programming Measurements Programming Measurements M E N U M A I N P r o g r a m 25.0ºC Reset Analyzer Frequency Security Outputs Measurement Temperature Program 1.234µS/cm 25.0ºC Configure? 1.234µS/cm Measurement contacting 25.0ºC 25.0ºC 25.0ºC 25.0ºC 25.0ºC SN Calculate Curve Custom curve fit completed. In Process Cal recommended. 1.234µS/cm progress … Curve fit in SN Calculate Curve 1.234µS/cm Configure Enter Data Points Calculate Curve Pt1: Pt1: Pt2: Pt2: Pt3: Pt3: Pt4: Pt4: Pt5: Pt5: 25.0ºC 1.000 ppm 1.000 µS/cm 1.000 ppm 1.000 µS/cm 1.000 ppm 1.000 µS/cm 1.000 ppm 1.000 µS/cm 1.000 ppm 1.000 µS/cm SN Data Points 1.234µS/cm SN Custom Config Units: ppm # of Points: 2 Ref Temp: 25.0 ºC Slope: 2.00 %/°C 25.0ºC 2.00 %/ºC 1.234µS/cm 25.0ºC SN Slope 1.234µS/cm Auto 0-50 µS 50-500 µS 500-2000 µS 2000 µS - 20 mS 20 mS - 200 mS 200 mS - 600 mS SN Range 25.0ºC 25.0ºC 1.234µS/cm SN Type 2-Electrode 4-Electrode 1.234µS/cm 1.234µS/cm 25.0ºC 25.0ºC 25.0ºC 25.0ºC 1.000 ppm SN PointM 1.234µS/cm % ppm mg/L g/L 25.0ºC SN Units 1.234µS/cm (25.0ºC normal) +25.0ºC SN Ref Temp 1.234µS/cm Slope Neutral Salt Cation Raw SN Temp Comp 1.234µS/cm 25.0ºC SN RTD Slope 1.234µS/cm 0.00°C 2.00%/ºC 25.0ºC SN RTD Offset 1.234µS/cm 1.00000 /cm SN Cell Constant LIQ-MAN-1066 SN Custom Curve 1.234µS/cm Cell K: 1.00000/cm RTD Offset: 0.00ºC RTD Slope: 0 Temp Comp: Slope Slope: 2.00%/°C Ref Temp: 25.0°C Filter: 2 sec Custom Setup Measure: Cond Type: 2-Electrode Range: Auto SN Configure 1.234µS/cm SN Measurement Conductivity Resistivity TDS Salinity 1.234µS/cm Instruction Manual Section 8: programming Measurements April 2017 Figure 8-2. Configure Contacting Measurements 61 62 M E N U M A I N P r o g r a m 25.0ºC Reset Analyzer Frequency Security Outputs Measurement Measurement Temperature Program 1.234µS/cm 25.0ºC Configure? 1.234µS/cm Measurement Toroidal 25.0ºC 228 Cond Auto 25.0ºC 25.0ºC 25.0ºC Custom curve fit completed. In Process Cal recommended. SN Calculate Curve 25.0ºC Pt1: Pt1: Pt2: Pt2: Pt3: Pt3: Pt4: Pt4: Pt5: Pt5: 1.000 ppm 1.000 µS/cm 1.000 ppm 1.000 µS/cm 1.000 ppm 1.000 µS/cm 1.000 ppm 1.000 µS/cm 1.000 ppm 1.000 µS/cm SN Data Points 25.0ºC Units: ppm # of Points: 2 Ref Temp: 25.0 ºC slope: 2.00 %/°C SN Custom Config 25.0ºC 2.000 %/ºC 1.234µS/cm 25.0ºC SN Slope 1.234µS/cm 200µS Auto 50-600 µS 0.5-100 mS 90-1500 mS 1.234µS/cm 25.0ºC SN Range 1.234µS/cm Conductivity Resistivity TDS Salinity NaOH (0-12%) HCl (0-15%) Low H2SO4 High H2SO4 NaCl (0-20%) 25.0ºC 25.0ºC 25.0ºC 1.000 ppm SN PointM 1.234µS/cm SN Units 1.234µS/cm 25.0ºC (25.0ºC normal) SN Ref Temp 25.0ºC Slope Neutral Salt Raw SN Temp Comp 1.234µS/cm 2.00%/ºC 1.234µS/cm % ppm mg/L g/L 25.0ºC SN RTD Slope 1.234µS/cm 0.00°C +25.0ºC 25.0ºC SN RTD Offset 1.234µS/cm 3.00000 /cm SN Cell Constant 1.234µS/cm April 2017 1.234µS/cm SN Calculate Curve Curve fit in progress … 1.234µS/cm Configure Enter Data Points Calculate Curve SN Custom Curve 1.234µS/cm Cell K: 3.00000/cm RTD Offset: 0.00ºC RTD Slope: 0 Temp Comp: Slope Slope: 2.00%/°C Ref Temp: 25.0°C Filter: 2 sec Custom Setup Model: Measure: Range: SN Configure 1.234µS/cm 247 228 225 226 25.0ºC SN Model 1.234µS/cm 1.234µS/cm SN Measurement Section 8: programming Measurements Instruction Manual LIQ-MAN-1066 Figure 8-3. Configure Toroidal Measurements Programming Measurements Programming Measurements M E N U M A I N P r o g r a m 25.0ºC Reset Analyzer Frequency Security Outputs Alarms Measurement Measurement Temperature Program 1.234µS/cm 25.0ºC Configure? 1.234µS/cm Measurement Oxygen 25.0ºC 25.0ºC 25.0ºC At Air Cal mA Input 25.0ºC SN Resolution 1.234µS/cm mm Hg in Hg atm kPa mbar bar 0.001 0.01 SN Use Pressure? 25.0ºC SN Partial Press 1.234µS/cm ppm mg/L ppb µg/L % Saturation Partial Pressure % Oxygen In Gas ppm Oxygen In Gas 25.0ºC 25.0ºC 25.0ºC Presssure Units 1.234µS/cm 05 sec SN Input filter 1.234µS/cm mm Hg in Hg atm kPa mbar bar 00.0 ‰ SN Salinity 1.234µS/cm LIQ-MAN-1066 1.234µS/cm Salinity: 00.0‰ Filter: 5sec Pressure Units: bar Resolution: 0.001 Use Press: At Air Cal Type: Water/Waste Units: ppm Partial Press: mmHg SN Configure 1.234µS/cm Water/Waste Trace Oxygen BioRx-Rosemount BioRx-Other Brewing Oxygen In Gas SN Type 1.234µS/cm 25.0ºC SN Units 1.234µS/cm Instruction Manual Section 8: programming Measurements April 2017 Figure 8-4. Configure Oxygen Measurements 63 64 25.0ºC Security Diagnostic Setup Reset Analyzer Frequency Outputs Measurement Measurement Temperature Program 1.234µS/cm 25.0ºC Configure? 1.234µS/cm Measurement Chlorine 25.0ºC 25.0ºC Filter: 5sec Dual Cal: Disable Free Cl Correct: Live Manual pH: 7.00 pH Resolution: 0.001 Measure: Free Chlorine Units: ppm SN Configure 1.234µS/cm Free Chlorine Total Chlorine Monochloramine SN Measurement 1.234µS/cm 25.0ºC 25.0ºC SN Resolution 1.234µS/cm 05 sec SN Input filter 1.234µS/cm 0.001 0.01 ppm mg/L 25.0ºC SN Units 1.234µS/cm 25.0ºC 25.0ºC 07.00 pH SN Manual pH 1.234µS/cm Live/Continuous Manual SN Free Cl pH Correction 1.234µS/cm Section 8: programming Measurements April 2017 Instruction Manual LIQ-MAN-1066 FIGuRE 8-5. Configuring Chlorine Measurements Programming Measurements Program MAIN MENU Programming Measurements 25.0ºC Reset Analyzer Frequency Security Outputs Measurement Measurement Temperature Program 1.234µS/cm 25.0ºC Configure? 1.234µS/cm Measurement Ozone Filter: Resolution: Units: 25.0ºC 5sec 0.001 ppm SN Configure 1.234µS/cm 0.001 0.01 25.0ºC 25.0ºC 05 sec SN Input filter 1.234µS/cm LIQ-MAN-1066 SN Resolution 1.234µS/cm ppm mg/L ppb µg/L 25.0ºC SN Units 1.234µS/cm Instruction Manual Section 8: programming Measurements April 2017 FIGuRE 8-5. Configuring Ozone Measurements 65 Program MAIN MENU Section 8: programming Measurements April 2017 66 Instruction Manual LIQ-MAN-1066 Programming Measurements Instruction Manual LIQ-MAN-1066 Section 9: Calibration April 2017 Section 9: Calibration 9.1 Introduction Calibration is the process of adjusting or standardizing the transmitter to a lab test or a calibrated laboratory instrument, or standardizing to some known reference (such as a commercial buffer). The auto-recognition feature of the transmitter will enable the appropriate calibration screens to allow calibration for any configuration of the transmitter. Completion of Quick Start upon first power up enables live measurements but does not ensure accurate readings in the lab or in process. Calibration should be performed with each attached sensor to ensure accurate, repeatable readings. This section covers the following calibration functions: 1. Auto buffer cal for pH (pH Cal - Section 9.2) 2. Manual buffer cal for pH (pH Cal - Section 9.2) 3. Set calibration stabilization criteria for pH (pH Cal - Section 9.2) 4. Standardization calibration (1-point) for pH, ORP and Redox (pH Cal - Section 9.2 and 9.3) 5. SMART sensor auto calibration upload 9.2 Calibration New sensors must be calibrated before use. Regular recalibration is also necessary. Use auto calibration instead of manual calibration. Auto calibration avoids common pitfalls and reduces errors. The transmitter recognizes the buffers and uses temperature-corrected pH values in the calibration. Once the 1066 successfully completes the calibration, it calculates and displays the calibration slope and offset. The slope is reported as the slope at 25 °C. To calibrate the pH loop with a connected pH sensor, access the Calibration screen by pressing ENTER from the main screen and press ENTER. The following calibration routines are covered: 1. Auto Calibration - pH 2 point buffer calibration with auto buffer recognition 2. Manual Calibration - pH 2 point buffer calibration with manual buffer value entry 3. Standardization - pH 1 point buffer calibration with manual buffer value entry 4. Entering A Known Slope Value - pH Slope calibration with manual entry of known slope value 5. SMART sensor auto calibration – auto detection and upload of cal data To calibrate pH: 1. Press the MENu button 2. Select Calibrate. Press ENTER. 3. Select ph. Press ENTER. Calibration The following sub-sections show the initial display screen that appears for each calibration routine. Use the flow diagram for ph calibration at the end of Section 7 and the live screen prompts to complete calibration. 1.234 µS/cm 25.0 ºC SN Calibrate? pH Temperature 67 Section 9: Calibration April 2017 9.2.1 Auto Calibration This screen appears after selecting ph calibration. Instruction Manual LIQ-MAN-1066 1.234 µS/cm 25.0 ºC SN pH Cal Buffer Cal Standardize Slope: 59.16mV/pH Offset: 600 mV Note that pH auto calibration criteria can be changed. The following criteria can be adjusted: • • • Stabilization time (default 10 sec.) Stabilization pH value (default 0.02 pH) Type of Buffer used for AUTO CALIBRATION (default is Standard, non-commercial buffers). The following commercial buffer tables are recognized by the analyzer: • • • • Standard (NIST plus pH7) DIN 19267 Ingold Merck This screen will appear if the auto cal is successful. The screen will return to the pH Buffer Cal Menu. The following screens may appear if the auto cal is unsuccessful. 1. A High Slope Error will generate this screen display: 2. A Low Slope Error will generate this screen display: 3. An Offset Error will generate this screen display: 68 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0ºC 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN Setup Stable Time: 10 sec Stable Delta: 0.02 pH Buffer: Standard SN pH Auto Cal Slope: 59.16 mV/pH Offset: 60 mV SN pH Auto Cal High Slope Error Calculated: 62.11 mV/pH Max: 62.00 mV/pH Press EXIT SN pH Auto Cal Low Slope Error Calculated: 39.11mV/pH Min: 40.00 mV/pH Press EXIT SN pH Auto Cal Offset Error Calculated: 61.22mV Max: 60.00mV Press EXIT Calibration Instruction Manual LIQ-MAN-1066 9.2.2 April 2017 Manual Calibration – pH New sensors must be calibrated before use. Regular recalibration is also necessary. Use manual calibration if non-standard buffers are being used; otherwise, use auto calibration. Auto calibration avoids common pitfalls and reduces errors. The adjacent appears after selecting Manual pH calibration. 9.2.3 Entering a Known Slope Value – pH 9.2.4 Standardization – pH If the electrode slope is known from other measurements, it can be entered directly in the 1066 transmitter. The slope must be entered as the slope at 25 °C. 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN pH Manual Cal Buffer 1 Buffer 2 SN pH Slope@25ºC 59.16 mV/pH The pH measured by the 1066 transmitter can be hanged to match the 1.234 µS/cm 25.0 ºC reading from a second or referee instrument. The process of making SN Enter Value 07.00pH the two readings agree is called standardization. During standardization, the difference between the two pH values is converted to the equivalent voltage. The voltage, called the reference offset, is added to all subsequent measured cell voltages before they are converted to pH. If a standardized sensor is placed in a buffer solution, the measured pH will differ from the buffer pH by an amount equivalent to the standardization offset. This screen may appear if pH Cal is unsuccessful. An Offset Error will generate this screen display: If the pH Cal is successful, the screen will return to the Cal sub-menu. 9.2.5 Section 9: Calibration SMART sensor auto calibration upload – pH 1.234 µS/cm 25.0 ºC SN Standardize Offset Error Calculated: 96mV Max: 60mV Press EXIT All calibration data including slope (mV/pH unit), offset (mV), glass impedance (MegOhms), and reference impedance (kOhms) is automatically downloaded to the SMART sensor upon successful calibration. This data transfer to the sensor is transparent and does not require any user action. Calibrated SMART sensors will be loop-calibrated when wired or attached (via VP8 cable connection) to any SMART-enabled Rosemount instrument. To calibrate any SMART sensor, choose any available calibration method. Note that new SMART sensors upon first shipment from Emerson are pre-calibrated and do not require buffer calibration or standardization to be used in process immediately. Calibration 69 Section 9: Calibration April 2017 9.3 Instruction Manual LIQ-MAN-1066 ORp and Redox Calibration For process control, it is often important to make the measured ORP or Redox agree with the ORP or Redox of a standard solution. During calibration, the measured ORP or Redox is made equal to the ORP or Redox of a standard solution at a single point. To calibrate the ORP loop with a connected ORP sensor, access the Calibration screen by pressing ENTER from the main screen and press ENTER. The following calibration routine is covered: 1. Standardization ORP 1 point buffer calibration with manual buffer value entry. To calibrate ORP: 1. Press the MENU button. 2. Select Calibrate. Press ENTER. 3. Select ORP and Redox. Press ENTER. 1.234 µS/cm 25.0 ºC SN Calibrate? ORP Temperature The following sub-sections show the initial display screen that appears for each calibration routine. Use the flow diagram for ORp calibration at the end of Section 8 and the live screen prompts to complete calibration. 9.3.1 Standardization – ORP and Redox For process control, it is often important to make the measured ORP and Redox agree with the ORP and Redox of a standard solution. During calibration, the measured ORP and Redox is made equal to the ORP and Redox of a standard solution at a single point. This screen appears after selecting ORP and Redox calibration: 1.234 µS/cm 25.0 ºC SN Enter Value +0600 mV If the ORP and Redox Cal is successful, the screen will return to the Cal sub-menu. The following screen may appear if ORP and Redox Cal is unsuccessful. 70 1.234 µS/cm 25.0 ºC SN Standardize Offset Error Calculated: 61.22mV Max: 60.00mV Press EXIT Calibration Instruction Manual LIQ-MAN-1066 9.4 Section 9: Calibration April 2017 Contacting Conductivity Calibration New conductivity sensors rarely need calibration. The cell constant printed on the label is sufficiently accurate for most applications. CalIBRaTING aN IN-SERvICE CONDuCTIvITy SENSOR After a conductivity sensor has been in service for a period of time, recalibration may be necessary. There are three ways to calibrate a sensor. a. Use a standard instrument and sensor to measure the conductivity of the process stream. It is not necessary to remove the sensor from the process piping. The temperature correction used by the standard instrument may not exactly match the temperature correction used by the 1066. To avoid errors, turn off temperature correction in both the transmitter and the standard instrument. b. Place the sensor in a solution of known conductivity and make the transmitter reading match the conductivity of the standard solution. Use this method if the sensor can be easily removed from the process piping and a standard is available. Be careful using standard solutions having conductivity less than 100 µS/cm. Low conductivity standards are highly susceptible to atmospheric contamination. Avoid calibrating sensors with 0.01/cm cell constants against conductivity standards having conductivity greater than 100 µS/cm. The resistance of these solutions may be too low for an accurate measurement. Calibrate sensors with 0.01/cm cell constant using method c. c. To calibrate a 0.01/cm sensor, check it against a standard instrument and 0.01/cm sensor while both sensors are measuring water having a conductivity between 5 and 10 µS/cm. To avoid drift caused by absorption of atmospheric carbon dioxide, saturate the sample with air before making the measurements. To ensure adequate flow past the sensor during calibration, take the sample downstream from the sensor. For best results, use a flow-through standard cell. If the process temperature is much different from ambient, keep connecting lines short and insulate the flow cell. To calibrate the conductivity loop with a connected contacting conductivity sensor, access the Calibration screen by pressing ENTER from the main screen and press ENTER. The following calibration routines are covered: 1. Zero Cal Zero the transmitter with the sensor attached 2. In Process Cal Standardize the sensor to a known conductivity 3. Cell K: 1.00000/cm Enter the cell Constant for the sensor 4. Meter Cal Calibrate the transmitter to a lab conductivity instrument 5. Cal Factor: 0.95000/cm Enter the Cal Factor for 4-Electrode sensors from the sensor tag To calibrate contacting conductivity: 1. Press the MENU button 2. Select Calibrate. Press ENTER. 3. Select Conductivity. Press ENTER. 1.234 µS/cm 25.0 ºC SN Calibrate? Conductivity Temperature The adjacent screen will appear. To calibrate Conductivity or Temperature, scroll to the desired item and press ENTER. Calibration 71 Section 9: Calibration April 2017 The following sub-sections show the initial display screen that appears for each calibration routine. Use the flow diagram for Conductivity calibration at the end of Section 7 and the live screen prompts for each routine to complete calibration. 9.4.1 The adjacent screen appears after selecting Conductivity calibration: Entering the Cell Constant New conductivity sensors rarely need calibration. The cell constant printed on the label is sufficiently accurate for most applications. The cell constant should be entered: • When the unit is installed for the first time Instruction Manual LIQ-MAN-1066 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN Calibration Zero Cal In Process Cal Meter Cal Cell K: 1.00000/cm SN Cell Constant 1.00000 /cm • When the probe is replaced 9.4.2 The display screen for entering a cell Constant for the sensor is shown. The default value is displayed in bold type. Zeroing the Instrument This procedure is used to compensate for small offsets to the conductivity signal that are present even when there is no conductivity to be measured. This procedure is affected by the length of extension cable and should always be repeated if any changes in extension cable or sensor have been made. Electrically connect the conductivity probe as it will actually be used and place the measuring portion of the probe in air. Be sure the probe is dry. The adjacent screen will appear after selecting Zero Cal from the Conductivity Calibration screen: The adjacent screen will appear if zero Cal is successful. The screen will return to the conductivity Cal Menu. The adjacent screen may appear if zero Cal is unsuccessful. 9.4.3 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN Zero Cal In Air In Water SN Zero Cal Sensor Zero Done SN Zero Cal Sensor Zero Fail Offset too high Press EXIT Calibrating the Sensor in a Conductivity Standard (in process cal) This procedure is used to calibrate the sensor and transmitter against a solution of known conductivity. This is done by submerging the probe in the sample of known conductivity, then adjusting the displayed value, if necessary, to correspond to the conductivity value of the sample. Turn temperature correction off and use the conductivity of the standard. Use a calibrated thermometer to measure temperature. The probe must be cleaned before performing this procedure. The adjacent screen will appear after selecting In Process Cal from the Conductivity Calibration screen: The adjacent screen will appear if In Process Cal is successful. The screen will return to the conductivity Cal Menu. The adjacent screen may appear if In Process Cal is unsuccessful. The screen will return to the conductivity Cal Menu. 72 1.234 µS/cm 25.0 ºC SN InProcess Cal Wait for stable reading. Calibration Instruction Manual LIQ-MAN-1066 The adjacent screen will appear if In Process Cal is successful. The screen will return to the conductivity Cal Menu. The adjacent screen may appear if In Process Cal is unsuccessful. The screen will return to the conductivity Cal Menu. Section 9: Calibration April 2017 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN InProcess Cal Updated cell constant: 1.00135/cm SN InProcess Cal Calibration Error Press EXIT 9.4.4 Calibrating the Sensor To A Laboratory Instrument (meter cal) This procedure is used to check and correct the conductivity reading of the 1066 using a laboratory conductivity instrument. This is done by submerging the conductivity probe in a bath and measuring the conductivity of a grab sample of the same bath water with a separate laboratory instrument. The 1066 reading is then adjusted to match the conductivity reading of the lab instrument. The adjacent screen will appear after selecting Meter Cal from the Conductivity Calibration screen: After pressing ENTER, the display shows the live value measured by the sensor If the meter cal is successful the screen will return to the conductivity Cal Menu. The adjacent screen will appear if Meter Cal is unsuccessful. The screen will return to the conductivity Cal Menu. 9.4.5 Calibration 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN Meter Cal Use precision resistors only SN Enter Value xx.xx kΩ SN Meter Cal Calibration Error Press EXIT Cal Factor 1.234 µS/cm 25.0 ºC Upon initial installation and power up, if 4-electrode was selected for the SN Cal Factor sensor type in the Quick Start menus, the user enters a Cell Constant and 0.95000 /cm a “Cal Factor” using the instrument keypad. The cell constant is needed to convert measured conductance to conductivity as displayed on the transmitter screen. The “Cal Factor” entry is needed increase the accuracy of the live conductivity readings, especially at low conductivity readings below 20uS/cm. Both the Cell Constant and the “Cal Factor” are printed on the tag attached to the 4-electrode sensor/cable. 73 Section 9: Calibration April 2017 9.5 Instruction Manual LIQ-MAN-1066 Toroidal Conductivity Calibration Calibration is the process of adjusting or standardizing the transmitter to a lab test or a calibrated laboratory instrument, or standardizing to some known reference (such as a conductivity standard). This section contains procedures for the first time use and for routine calibration of the 1066 transmitter. To calibrate the conductivity loop with a connected contacting conductivity sensor, access the Calibration screen by pressing ENTER from the main screen and press ENTER. The following calibration routines are covered: 1. Zero Cal Zero the transmitter with the sensor attached 2. In Process Cal Standardize the sensor to a known conductivity 3. Cell K: 1.00000/cm Enter the cell Constant for the sensor 4. Meter Cal Calibrate the transmitter to a lab conductivity instrument To calibrate toroidal conductivity: 1. Press the MENU button 2. Select Calibrate. Press ENTER. 3. Select Conductivity. Press ENTER. The adjacent screen will appear. To calibrate Toroidal Conductivity or Temperature, scroll to the desired item and press ENTER The following sub-sections show the initial display screen that appears for each calibration routine. Use the flow diagram for Conductivity calibration at the end of Section 7 and the live screen prompts to complete calibration. The adjacent screen appears after selecting Conductivity calibration: 9.5.1 Entering the Cell Constant 1.234 µS/cm 25.0 ºC Calibration Zero Cal In Process Cal Cell K: 2.7000/cm New toroidal sensors always need to be calibrated. The cell constant provided on the sensor label is a nominal value and does not need to be entered. This procedure sets up the transmitter for the probe type connected to the transmitter. Each type of probe has a specific cell constant The display screen for entering a cell constant for the sensor is shown. The default value is displayed in bold type. 74 1.234 µS/cm 25.0 ºC Cell Constant 2.7000/cm Calibration Instruction Manual LIQ-MAN-1066 9.5.2 April 2017 Zeroing the Instrument This procedure is used to compensate for small offsets to the conductivity signal that are present even when there is no conductivity to be measured. This procedure is affected by the length of extension cable and should always be repeated if any changes in extension cable or sensor have been made. Electrically connect the conductivity probe as it will actually be used and place the measuring portion of the probe in air. The adjacent screen will appear after selecting Zero Cal from the Conductivity Calibration screen: The adjacent screen will appear if zero Cal is successful. The screen will return to the conductivity Cal Menu. The adjacent screen may appear if zero Cal is unsuccessful. 9.5.3 Section 9: Calibration 1.234 µS/cm 25.0 ºC SN Zero Cal In Air In Water 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN Zero Cal Sensor Zero Done SN Zero Cal Sensor Zero Fail Offset too high Press EXIT Calibrating the Sensor in a Conductivity Standard (in process cal) This procedure is used to check and correct the conductivity reading of the 1066 to ensure that the reading is accurate. This is done by submerging the probe in the sample of known conductivity, then adjusting the displayed value, if necessary, to correspond to the conductivity value of the sample. The probe must be cleaned before performing this procedure. The temperature reading must also be checked and standardized if necessary, prior to performing this procedure. The adjacent screen will appear after selecting In Process Cal from the Conductivity Calibration screen: The following screen will appear if In Process Cal is successful. The screen will return to the conductivity Cal Menu. This screen may appear if In Process Cal is unsuccessful. The screen will return to the conductivity Cal Menu. 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN InProcess Cal Wait for stable reading. SN InProcess Cal Updated cell constant: 3.01350/cm SN InProcess Cal Calibration Error Press EXIT Calibration 75 Section 9: Calibration April 2017 9.6 Instruction Manual LIQ-MAN-1066 Calibration – Chlorine The 1066 can measure three variants of Chlorine: • Free Chlorine • Total Chlorine • Monochloramine The section describes how to calibrate any compatible amperometric chlorine sensor. The following calibration routines are covered in the family of supported Chlorine sensors: • Air Cal • Zero Cal • In Process Cal 9.6.1 Calibration – Free Chlorine A free chlorine sensor generates a current directly proportional to the concentration of free chlorine in the sample. Calibrating the sensor requires exposing it to a solution containing no chlorine (zero standard) and to a solution containing a known amount of chlorine (full-scale standard). The zero calibration is necessary because chlorine sensors, even when no chlorine is in the sample, generate a small current called the residual current. The transmitter compensates for the residual current by subtracting it from the measured current before converting the result to a chlorine value. New sensors require zeroing before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. To calibrate the chlorine sensor, access the Calibration screen by pressing ENTER from the main screen and press ENTER. The following calibration routines are covered: 1. Zero Cal Zeroing the sensor in solution with zero free chlorine 2. Grab Cal Standardizing to a sample of known free chlorine concentration To calibrate free chlorine: 1. Press the MENU button 2. Select Calibrate. Press ENTER. 3. Select Free Chlorine. Press ENTER. The adjacent screen will appear. To calibrate Free Chlorine or Temperature, scroll to the desired item and press ENTER. 1.234 µS/cm 25.0 ºC SN Calibrate? Free Chlorine Temperature The following sub-sections show the initial display screen that appears for each calibration routine. Use the flow diagram for Chlorine calibration at the end of Section 7 and the live screen prompts to complete calibration. The adjacent screen appears after selecting Free Chlorine calibration: 76 1.234 µS/cm 25.0 ºC SN Calibration Zero Cal In Process Cal Calibration Instruction Manual LIQ-MAN-1066 9.6.1.1 Section 9: Calibration April 2017 Zeroing the Sensor The adjacent screen will appear if In Zero Cal is successful. The screen will return to the Amperometric Cal Menu. In Process Calibration The adjacent screen will appear prior to In Process Cal 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN Zero Cal Sensor zero failed Press EXIT 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN InProcess Cal Wait for stable reading. If the In Process Cal is successful, the screen will return to the Cal sub-menu. The adjacent screen may appear if In Zero Cal is unsuccessful. The screen will return to the Amperometric Cal Menu. 25.0 ºC SN Zero Cal Sensor zero done The adjacent screen may appear if In Zero Cal is unsuccessful. The screen will return to the Amperometric Cal Menu. 9.6.1.2 1.234 µS/cm SN Zero Cal Zeroing Wait The adjacent screen will appear during Zero Cal. Be sure sensor has been running in zero solution for at least two hours before starting zero step. SN InProcess Cal Calibration Error Press EXIT 9.6.2 Calibration Calibration – Total Chlorine Total chlorine is the sum of free and combined chlorine. The continuous determination of total chlorine requires two steps. First, the sample flows into a conditioning system (TCL) where a pump continuously adds acetic acid and potassium iodide to the sample. The acid lowers the pH, which allows total chlorine in the sample to quantitatively oxidize the iodide in the reagent to iodine. In the second step, the treated sample flows to the sensor. The sensor is a membrane-covered amperometric sensor, whose output is proportional to the concentration of iodine. Because the concentration of iodine is proportional to the concentration of total chlorine, the transmitter can be calibrated to read total chlorine. Because the sensor really measures iodine, calibrating the sensor requires exposing it to a solution containing no iodine (zero standard) and to a solution containing a known amount of iodine (full-scale standard). The Zero calibration is necessary because the sensor, even when no iodine is present, generates a small current called the residual current. The transmitter compensates for the residual current by subtracting it from the measured current before converting the result to a total chlorine value. New sensors require zeroing before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. The best zero standard is deionized water. The purpose of the In Process Calibration is to 77 Section 9: Calibration April 2017 Instruction Manual LIQ-MAN-1066 establish the slope of the calibration curve. Because stable total chlorine standards do not exist, the sensor must be calibrated against a test run on a grab sample of the process liquid. Several manufacturers offer portable test kits for this purpose. To calibrate the chlorine sensor, access the Calibration screen by pressing ENTER from the main screen, and press ENTER. The following calibration routines are covered: 1. Zero Cal Zeroing the sensor in solution with zero total chlorine 2. Grab Cal Standardizing to a sample of known total chlorine concentration To calibrate total chlorine: 1. Press the MENU button 2. Select Calibrate. Press ENTER. 3. Select Total Chlorine. Press ENTER. The adjacent screen will appear. To calibrate Total Chlorine or Temperature, scroll to the desired item and press ENTER 1.234 µS/cm 25.0 ºC SN Calibrate? Total Chlorine Temperature The following sub-sections provide you with the initial display screen that appears for each calibration routine. Use the flow diagram for Chlorine calibration at the end of Sec. 9 and the live screen prompts to complete calibration. 1.234 µS/cm 25.0 ºC This adjacent screen appears after selecting Total Chlorine calibration: 9.6.2.1 Zeroing the Sensor The adjacent screen will appear during Zero Cal. Be sure sensor has been running in zero solution for at least two hours before starting zero step. The adjacent screen will appear if In Zero Cal is successful. The screen will return to the Amperometric Cal Menu. The adjacent screen may appear if In Zero Cal is unsuccessful. The screen will return to the Amperometric Cal Menu. SN Calibration Zero Cal In Process Cal 1.234 µS/cm 25.0 ºC SN Zero Cal Zeroing Wait 1.234µS/cm 25.0ºC SN Zero Cal Sensor zero done 1.234 µS/cm 25.0 ºC SN Zero Cal Sensor zero failed Press EXIT 9.6.2.2 In Process Calibration The adjacent screen will appear prior to In Process Cal If the In Process Cal is successful, the screen will return to the Cal submenu. The adjacent screen may appear if In Process Cal is unsuccessful. The screen will return to the Amperometric Cal Menu. 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN InProcess Cal Wait for stable reading. SN InProcess Cal Calibration error Press EXIT 78 Calibration Instruction Manual LIQ-MAN-1066 9.6.3 Section 9: Calibration April 2017 Calibration – Monochloromine A monochloramine sensor generates a current directly proportional to the concentration of monochloramine in the sample. Calibrating the sensor requires exposing it to a solution containing no monochloramine (zero standard) and to a solution containing a known amount of monochloramine (full-scale standard). The zero calibration is necessary because monochloramine sensors, even when no monochloramine is in the sample, generate a small current called the residual or zero current. The transmitter compensates for the residual current by subtracting it from the measured current before converting the result to a monochloramine value. New sensors require zeroing before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. The best zero standard is deionized water. The purpose of the In Process calibration is to establish the slope of the calibration curve. Because stable monochloramine standards do not exist, the sensor must be calibrated against a test run on a grab sample of the process liquid. Several manufacturers offer portable test kits for this purpose. To calibrate the chlorine sensor, access the Calibration screen by pressing ENTER from the main screen, and press ENTER. The following calibration routines are covered: 1. Zero Cal Zeroing the sensor in solution with zero total chlorine 2. Grab Cal Standardizing to a sample of known monochloramine concentration To calibrate monochloramine: 1. Press the MENU button 2. Select Calibrate. Press ENTER. 3. Select Monochloramine. Press ENTER. The adjacent screen will appear. To calibrate Monochloramine or Temperature, scroll to the desired item and press ENTER. 1.234 µS/cm 25.0 ºC SN Calibrate? Monochloramine Temperature The following sub-sections provide you with the initial display screen that appears for each calibration routine. Use the flow diagram for Chlorine calibration at the end of Sec. 9 and the live screen prompts to complete calibration. The adjacent screen appears after selecting Monochloramine calibration: Calibration 1.234 µS/cm 25.0 ºC SN Calibration Zero Cal In Process Cal 79 Section 9: Calibration April 2017 9.6.4 Zeroing the Sensor The adjacent screen will appear during Zero Cal. Be sure sensor has been running in zero solution for at least two hours before starting zero step. The adjacent screen will appear if In Zero Cal is successful. The screen will return to the Amperometric Cal Menu. The adjacent screen may appear if In Zero Cal is unsuccessful. The screen will return to the Amperometric Cal Menu. Instruction Manual LIQ-MAN-1066 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN Zero Cal Zeroing Wait SN Zero Cal Sensor zero done SN Zero Cal Sensor zero failed Press EXIT 9.6.5 In Process Calibration The adjacent screen will appear prior to In Process Cal If the In Process Cal is successful, the screen will return to the Cal submenu. The adjacent screen may appear if In Process Cal is unsuccessful. The screen will return to the Amperometric Cal Menu. 9.7 80 Calibration – Oxygen 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN InProcess Cal Wait for stable reading. SN InProcess Cal Calibration Error Press EXIT Oxygen sensors generate a current directly proportional to the concentration of dissolved oxygen in the sample. Calibrating the sensor requires exposing it to a solution containing no oxygen (zero standard) and to a solution containing a known amount of oxygen (full-scale standard). The Zero Calibration is necessary because oxygen sensors, even when no oxygen is present in the sample, generate a small current called the residual current. The transmitter compensates for the residual current by subtracting it from the measured current before converting the result to a dissolved oxygen value. New sensors require zeroing before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. The recommended zero standard is 5% sodium sulfite in water, although oxygen-free nitrogen can also be used. The 499A TrDO sensor, used for the determination of trace (ppb) oxygen levels, has very low residual current and does not normally require zeroing. The residual current in the 499A TrDO sensor is equivalent to less than 0.5 ppb oxygen. The purpose of the In Process Calibration is to establish the slope of the calibration curve. Because the solubility of atmospheric oxygen in water as a function of temperature and barometric pressure is well known, the natural choice for a full-scale standard is air-saturated water. However, air-saturated water is difficult to prepare and use, so the universal practice is to use air for calibration. From the point of view of the oxygen sensor, air and air- Calibration Instruction Manual LIQ-MAN-1066 Section 9: Calibration April 2017 saturated water are identical. The equivalence comes about because the sensor really measures the chemical potential of oxygen. Chemical potential is the force that causes oxygen molecules to diffuse from the sample into the sensor where they can be measured. It is also the force that causes oxygen molecules in air to dissolve in water and to continue to dissolve until the water is saturated with oxygen. Once the water is saturated, the chemical potential of oxygen in the two phases (air and water) is the same. Oxygen sensors generate a current directly proportional to the rate at which oxygen molecules diffuse through a membrane stretched over the end of the sensor. The diffusion rate depends on the difference in chemical potential between oxygen in the sensor and oxygen in the sample. An electrochemical reaction, which destroys any oxygen molecules entering the sensor, keeps the concentration (and the chemical potential) of oxygen inside the sensor equal to zero. Therefore, the chemical potential of oxygen in the sample alone determines the diffusion rate and the sensor current. When the sensor is calibrated, the chemical potential of oxygen in the standard determines the sensor current. Whether the sensor is calibrated in air or air-saturated water is immaterial. The chemical potential of oxygen is the same in either phase. Normally, to make the calculation of solubility in common units (like ppm DO) simpler, it is convenient to use water-saturated air for calibration. Automatic air calibration is standard. The user simply exposes the sensor to water-saturated air. The transmitter monitors the sensor current. When the current is stable, the transmitter stores the current and measures the temperature using a temperature element inside the oxygen sensor. The user must enter the barometric pressure. From the temperature the transmitter calculates the saturation vapor pressure of water. Next, it calculates the pressure of dry air by subtracting the vapor pressure from the barometric pressure. Using the fact that dry air always contains 20.95% oxygen, the transmitter calculates the partial pressure of oxygen. Once the transmitter knows the partial pressure of oxygen, it uses the Bunsen coefficient to calculate the equilibrium solubility of atmospheric oxygen in water at the prevailing temperature. At 25 °C and 760 mm Hg, the equilibrium solubility is 8.24 ppm. Often it is too difficult or messy to remove the sensor from the process liquid for calibration. In this case, the sensor can be calibrated against a measurement made with a portable laboratory instrument. The laboratory instrument typically uses a membrane-covered amperometric sensor that has been calibrated against water-saturated air. To calibrate the oxygen sensor, access the Calibration screen by pressing ENTER from the main screen, select and press ENTER. The following calibration routines are covered: 1. Zero Cal Zeroing the sensor in a medium with zero oxygen 2. Air Cal Calibrating the sensor in a water-saturated air sample 3. In Process Cal Standardizing to a sample of known oxygen concentration 4. Sen@ 25 °C:2500 nA/ppm Entering a known slope value for sensor response To calibrate oxygen: 1. Press the MENU button 2. Select Calibrate. Press ENTER. 3. Select Oxygen. Press ENTER. The adjacent screen will appear. To calibrate Oxygen or Temperature, scroll to the desired item and press ENTER Calibration 1.234 µS/cm 25.0 ºC SN Calibrate? Oxygen Temperature 81 Section 9: Calibration Instruction Manual April 2017 LIQ-MAN-1066 The following sub-sections provide you with the initial display screen that appears for each calibration routine. Use the flow diagram for Oxygen calibration at the end of Section 9 and the live screen prompts for each routine to complete calibration. The adjacent screen appears after selecting Oxygen calibration: Air calibration criteria can be changed. The following criteria can be adjusted: • Stabilization time (default 10 sec.) 1.234 µS/cm 25.0 ºC SN Calibration Air Cal Zero Cal In Process Cal Sen@ 25 °C:2500nA/ppm Zero Current: 1234 nA • Stabilization pH value (default 0.05 ppm) • Salinity of the solution to be measured (default 00.0 parts per thousand) The adjacent screen will appear to allow adjustment of these criteria: 9.7.1 Zeroing the Sensor The adjacent screen will appear during Zero Cal. The adjacent screen will appear if In Zero Cal is successful. The screen will return to the Amperometric Cal Menu. The adjacent screen may appear if In Zero Cal is unsuccessful. The screen will return to the Amperometric Cal Menu. 9.7.2 Calibrating the Sensor in Air The adjacent screen will appear prior to Air Cal The adjacent screen will appear if In Air Cal is successful. The screen will return to the Amperometric Cal Menu. The adjacent screen may appear if In Air Cal is unsuccessful. The screen will return to the Amperometric Cal Menu. 1.234 µS/cm 25.0 ºC SN Setup Stable Time: 10 sec Stable Delta: 0.05 ppm Salinity: 00.0 ‰ 1.234 nA SN Zero Cal Zeroing Wait 1.234 nA SN Zero Cal Sensor zero done 1.234 nA SN Zero Cal Sensor zero failed Press EXIT 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC 1.234 µS/cm 25.0 ºC SN Air Cal Start Calibration Setup Done SN Air Cal SN Air Cal Failure Check Sensor Press EXIT 82 Calibration Instruction Manual LIQ-MAN-1066 9.7.3 Section 9: Calibration April 2017 Calibrating the Sensor Against A Standard Instrument (in process cal) The adjacent screen will appear prior to In Process Cal If the In Process Cal is successful, the screen will return to the Cal sub-menu. The adjacent screen may appear if In Zero Cal is unsuccessful. The screen will return to the Amperometric Cal Menu. 9.8 Calibration – Ozone An ozone sensor generates a current directly proportional to the concentration of ozone in the sample. Calibrating the sensor requires exposing it to a solution containing no ozone (zero standard) and to a solution containing a known amount of ozone (full-scale standard). The Zero Calibration is necessary because ozone sensors, even when no ozone is in the sample, generate a small current called the residual or zero current. The transmitter compensates for the residual current by subtracting it from the measured current before converting the result to an ozone value. New sensors require zeroing before being placed in service, and sensors should be zeroed whenever the electrolyte solution is replaced. The best zero standard is deionized water. The purpose of the In Process Calibration is to establish the slope of the calibration curve. Because stable ozone standards do not exist, the sensor must be calibrated against a test run on a grab sample of the process liquid. Several manufacturers offer portable test kits for this purpose. To calibrate the ozone sensor, access the Calibration screen by pressing ENTER from the main screen, select and press ENTER. The following calibration routines are covered: 1. Zero Cal Zeroing the sensor in solution with zero total chlorine 2. Grab Cal Standardizing to a sample of known ozone concentration To calibrate ozone: 1. Press the MENU button 2. Select Calibrate. Press ENTER. 3. Select Ozone. Press ENTER. The adjacent screen will appear. To calibrate Ozone or Temperature, scroll to the desired item and press ENTER. Calibration 1.234 µS/cm 25.0 ºC SN Calibrate? Ozone Temperature 83 Section 9: Calibration April 2017 Instruction Manual LIQ-MAN-1066 The following sub-sections provide you with the initial display screen that appears for each calibration routine. Use the flow diagram for Ozone calibration at the end of Section 9 and the live screen prompts to complete calibration. The adjacent screen appears after selecting Ozone calibration: 9.8.1 Zeroing the Sensor The following screen will appear during Zero Cal The following screen will appear if In Zero Cal is successful. The screen will return to the Amperometric Cal Menu. The following screen may appear if In Zero Cal is unsuccessful. The screen will return to the Amperometric Cal Menu. 1.234 µS/cm 25.0 ºC SN Calibration Zero Cal In Process Cal 1.234 nA SN Zero Cal Zeroing Wait 1.234 nA SN Zero Cal Sensor zero done 1.234 nA SN Zero Cal Sensor zero failed Press EXIT 9.8.2 In Process Calibration The following screen will appear after selecting In Process Cal If the In Process Cal is successful, the screen will return to the Cal submenu. The following screen may appear if In Zero Cal is unsuccessful. The screen will return to the Amperometric Cal Menu. 84 1.234 µS/cm 25.0 ºC SN InProcess Cal Wait for stable reading. 1.234 µS/cm 25.0 ºC SN InProcess Cal Calibration Error Press EXIT Calibration Instruction Manual LIQ-MAN-1066 9.9 Section 9: Calibration April 2017 Calibrating Temperature Most liquid analytical measurements require temperature compensation (except ORP and Redox). The 1066 performs temperature compensation automatically by applying internal temperature correction algorithms. Temperature correction can also be turned off. If temperature correction is off, the 1066 uses the manual temperature entered by the user in all temperature correction calculations. To calibrate temperature, access the Calibration screen by pressing ENTER from the main screen, select Temperature and press ENTER. The following calibration routine is covered: 1. Temperature with manual temperature entry To calibrate temperature: 1. Press the MENU button 2. Select Calibrate. Press ENTER. 3. Select Temperature. Press ENTER. The adjacent screen will appear. 9.9.1 1.234 µS/cm 25.0 ºC SN Calibrate +0 25.0 °C The following sub-section provides you with the initial display screen that appears for temperature calibration. Use the flow diagram for Temp calibration at the end of Sec. 7 to complete calibration. Calibration The adjacent screen will appear during Temperature Cal. If the sensor Temperature offset is greater than 5 °C from the default value, the following screen will appear: 1.234 µS/cm 25. 0 ºC 1.234 µS/cm 25.0 ºC SN Calibrate Cal in progress. Please wait. You may continue by selecting Yes or suspend this operation by selecting No. SN Temp Offset > 5 °C If the Temp Cal is successful, the screen will return to the Cal Menu. No Yes Continue? Note: To select automatic or manual temp compensation or to program temperature units as °C or °F, refer to Section 7.3 – Programming Temperature in this manual. Calibration 85 86 1.234µS/cm 25.0ºC 25.0ºC Changing slope overrides buffer Cal 25.0ºC Changing offset overrides buffer Cal 1.234µS/cm 07.00pH 25.0ºC 25.0ºC 25.0ºC 25.0ºC 060 mV SN pH Offset 25.0ºC 59.16 mV/pH SN pH Slope@25ºC 1.234µS/cm Cal in progress. Please wait. SN Standardize 1.234µS/cm 07.00pH SN Setup 25.0ºC Standard DIN 19267 Ingold Merck 1.234µS/cm 25.0ºC 25.0ºC Offset Error Calculated: Max: Press EXIT 96mV 60mV SN Standardize 1.234µS/cm Cal in progress. Please wait. SN Manual Cal 1.234µS/cm 1.234µS/cm 1.234µS/cm 25.0ºC 25.0ºC Place Sensor in Buffer 1 Press ENTER SN pH Auto Cal Slope: 59.16mV/pH Offset: 60 mV SN pH Manual Cal 25.0ºC SN Buffer 1.234µS/cm Stable Time: 10 sec Stable Delta: 0.02 pH Buffer: Standard 25.0ºC Manual BufferM 1.234µS/cm Start AutoCal Setup 1.234µS/cm Buffer 1 Buffer 2 1.234µS/cm TDS Salinity NaOH HCl Low H2SO4 High H2SO4 NaCl Resistivity Custom Conc’n Temperature SN pH Auto Cal SN pH Manual Cal SN Enter Value 1.234µS/cm Auto Manual 25.0ºC SN Calibrate? 1.234µS/cm 25.0ºC 25.0ºC 25.0ºC 25.0ºC 25.0ºC 25.0ºC 25.0ºC Slope: 59.16 mV/pH Offset: 60 mV SN pH Auto Cal 1.234µS/cm Cal in progress. Please wait. SN pH Auto Cal 1.234µS/cm 10.01pH SN pH Auto Buffer 2 1.234µS/cm Wait SN pH Auto Buffer 2 1.234µS/cm Place Sensor in Buffer 1 Press ENTER SN pH Auto Cal 1.234µS/cm 07.01pH SN pH Auto Buffer 1 1.234µS/cm Wait SN pH Auto Buffer 1 April 2017 1.234µS/cm 25.0ºC Free Chlorine Total Chlorine Chloramine Ozone Oxygen pH ORP Redox Conductivity 1.234µS/cm SN pH Buffer Cal 1.234µS/cm 25.0ºC Calibrate? 1.234µS/cm 25.0ºC SN pH Cal 1.234µS/cm C a l i b r a t e Buffer Cal Standardize Slope: 59.16mV/pH Offset: 600 mV M E N U M A I N Section 9: Calibration Instruction Manual LIQ-MAN-1066 FIGuRE 9-1. Calibrate ph Calibration Calibration M E N U M A I N C a l i b r a t e 25.0ºC Calibrate? 1.234µS/cm 25.0ºC Use precision resistors only SN Meter Cal 25.0ºC 25.0ºC 25.0ºC xx.xx kΩ SN Enter Value 1.234µS/cm 10.00 uS/cm SN Enter Value 1.234µS/cm 25.0ºC 25.0ºC 25.0ºC 25.0ºC Cal in progress. Please wait. SN Meter Cal 1.234µS/cm Cal in progress. Please wait. SN InProcess Cal 1.234µS/cm Press ENTER if reading is stable. SN InProcess Cal 1.234µS/cm 25.0ºC Zeroing Wait SN Zero Cal 25.0ºC Updated cell constant: 1.00135/cm SN InProcess Cal 25.0ºC SN Zero Cal 1.234µS/cm Sensor must be dry and in air 1.234µS/cm 25.0ºC SN Zero Cal 1.234µS/cm Sensor Zero Done 1.234µS/cm 25.0ºC SN Zero Cal Zeroing Wait 1.234µS/cm Sensor must be In 0% solution SN Zero Cal 1.234µS/cm In Air In Water 25.0ºC 25.0ºC SN Zero Cal 1.234µS/cm SN InProcess Cal 1.234µS/cm Wait for stable reading. Zero Cal In Process Cal Meter Cal Cell K: 1.00000/cm SN InProcess Cal Take sample; Press ENTER. 25.0ºC SN Calibration 1.234µS/cm LIQ-MAN-1066 1.234µS/cm Temperature Conductivity 1.234µS/cm TDS Salinity NaOH HCl Low H2SO4 High H2SO4 NaCl Resistivity’ Resistivity’ CustomConc’n Conc’n Custom SN Calibrate? 25.0ºC Free Chlorine pH Independ. Free Cl Total Chlorine Chloramine Ozone Oxygen pH ORP Redox 1.234µS/cm Instruction Manual Section 9: Calibration April 2017 Figure 9-2. Calibrate Contacting and Toroidal Conductivity 87 88 M E N U M A I N C a l i b r a t e 25.0ºC Calibrate? 1.234µS/cm TDS Salinity NaOH HCl Low H2SO4 High H2SO4 NaCl Resistivity’ Custom Conc’n Temperature Conductivity SN Calibrate? 25.0ºC Free Free Chlorine Chlorine pH Independ. Free Cl Total Chlorine Total Chlorine Chloramine Chloramine Ozone Oxygen pH ORP Redox 1.234µS/cm 25.0ºC 25.0ºC Wait for stable reading SN InProcess Cal 1.234µS/cm Zero Cal In Process Cal SN Calibration 1.234µS/cm 25.0ºC 25.0ºC Press ENTER if reading is stable. SN InProcess Cal 1.234µS/cm Zeroing Wait SN Zero Cal 1.234µS/cm 25.0ºC 25.0ºC Cal in progress. Please wait. SN InProcess Cal 1.234µS/cm 10.00 ppm SN Enter Value 1.234µS/cm Take sample; Press ENTER. SN InProcess Cal 1.234µS/cm Sensor zero done 25.0ºC 25.0ºC SN Zero Cal 1.234µS/cm Section 9: Calibration April 2017 Instruction Manual LIQ-MAN-1066 Figure 9-3. Calibrate Free Chlorine, Total Chlorine, and Monochloramine Calibration Calibration M E N U M A I N C a l i b r a t e 25.0ºC Calibrate? 1.234µS/cm 25.0ºC Wait for stable reading SN InProcess Cal 25.0ºC 25.0ºC SN InProcess Cal 1.234µS/cm Press ENTER if reading is stable. Zeroing Wait 25.0ºC Take sample; Press ENTER. 25.0ºC SN Air Cal 1.234µS/cm 25.0ºC 25.0ºC 25.0ºC 10.00 ppm SN Enter Value 1.234µS/cm Stable Time: 10 sec Stable Delta: 0.05 ppm Salinity: 00.0 ‰ SN Setup 1.234µS/cm Start Calibration Setup SN InProcess Cal 1.234µS/cm Sensor zero done SN Zero Cal 1.234µS/cm Zero Current: 1234nA Air Cal Zero Cal In Process Cal Sen@ 25°C:2500nA/ppm 25.0ºC SN Zero Cal 25.0ºC SN Calibration 1.234µS/cm 1.234µS/cm 25.0ºC 25.0ºC 25.0ºC 25.0ºC 25.0ºC Cal in progress. Please wait. SN InProcess Cal 1.234µS/cm 00.0 ‰ SN Salinity 1.234µS/cm 0.05 ppm SN Stable Delta 1.234µS/cm 10 sec SN Stable Time 1.234µS/cm Done 25.0ºC 25.0ºC SN Air Cal 1.234µS/cm Wait Air Cal 1.234µS/cm 760 mm Hg SN Air Pressure LIQ-MAN-1066 1.234µS/cm Conductivity 1.234µS/cm Temperature Free Chlorine Total Chlorine Chloramine Ozone Oxygen pH ORP Redox SN Calibrate? TDS Salinity NaOH HCl Low H2SO4 High H2SO4 NaCl Resistivity CustomConc’n 1.234µS/cm Instruction Manual Section 9: Calibration April 2017 Figure 9-4. Calibrate Oxygen 89 90 M E N U M A I N C a l i b r a t e 25.0ºC Calibrate? 1.234µS/cm 25.0ºC TDS Salinity NaOH HCl Low H2SO4 High H2SO4 NaCl Resistivity Custom Conc’n Temperature Conductivity SN Calibrate? Free Chlorine Total Chlorine Chloramine Ozone Oxygen pH ORP Redox 1.234µS/cm 25.0ºC 25.0ºC Wait for stable reading SN InProcess Cal 1.234µS/cm Zero Cal In Process Cal SN Calibration 1.234µS/cm 1.234µS/cm SN InProcess Cal Press ENTER if reading is stable. Zeroing Wait 25.0ºC 25.0ºC SN Zero Cal 1.234µS/cm 25.0ºC 25.0ºC Cal in progress. Please wait. SN InProcess Cal 1.234µS/cm 10.00 ppm SN Enter Value 1.234µS/cm Take sample; Press ENTER. SN InProcess Cal 1.234µS/cm Sensor zero done 25.0ºC 25.0ºC SN Zero Cal 1.234µS/cm Section 9: Calibration April 2017 Instruction Manual LIQ-MAN-1066 Figure 9-5. Calibrate Ozone Calibration Calibration M E N U M A I N C a l i b r a t e 25.0ºC Calibrate? 1.234µS/cm Temperature Total Chlorine Chloramine Ozone Oxygen pH ORP Redox Conductivity SN Calibrate? 25.0ºC TDS Salinity NaOH HCl Low H2SO4 High H2SO4 NaCl Resistivity CustomConc’n Free Chlorine 1.234µS/cm 25.0ºC 25.0ºC Stable Time: 10sec Stable Delta: 1 mV SN Setup 1.234µS/cm 25.0ºC 1.234µS/cm 25.0ºC 25.0ºC 25.0ºC 0001 mV 25.0ºC SN Standard M 1.234µS/cm 0000ppm SN Stable Delta 1.234µS/cm SN Standard M 1.234µS/cm Stabilizing Wait SN Stable Time 010 sec 25.0ºC SN Standardize Cal in progress. Please wait. 1.234µS/cm SN 2-Point Cal 1.234µS/cm Standard 1 Standard 2 +0600 mV Standardize 2-Pt Calibration Slope: 0mV/decade Offset: 0 mV Setup SN ISE Cal 1.234µS/cm 25.0ºC SN Enter Value 1.234µS/cm Instruction Manual LIQ-MAN-1066 Section 9: Calibration April 2017 FIGuRE 9-6. Calibrate ORp 91 92 M E N U M A I N C a l i b r a t e 25.0ºC Calibrate? 1.234µS/cm Free Chlorine Total Chlorine Chloramine Ozone Oxygen pH ORP Redox Conductivity 1.234µS/cm 25.0ºC Temperature Temperature TDS Salinity NaOH HCl Low H2SO4 High H2SO4 NaCl Resistivity Custom Conc’n SN Calibrate? 25.0ºC +025.0°C SN Calibrate 1.234µS/cm 25.0ºC Cal in progress. Please wait. SN Calibrate 1.234µS/cm 25.0ºC SN Temp Offset > 5°C Continue? No Yes 1.234µS/cm Section 9: Calibration April 2017 Instruction Manual LIQ-MAN-1066 FIGuRE 9-7. Calibrate Temperature Calibration Instruction Manual Section 10: haRT Communication LIQ-MAN-1066 April 2017 Section 10: HART Communications ® 10.1 Introduction The 1066 transmitter can communicate with a HART host using HART Revision 5 or HART Revision 7. The revision of HART used by the 1066 can be selected using the keypad/display or a HART master such as AMS or the 475 Handheld Communicator. The default version of HART is Revision 5. Since some HART hosts cannot accommodate HART 7, the choice of HART Revision should be based on the capabilities of the host, and should be chosen as a first step in configuration. If HART Revision 7 can be used with the host, it does offer a number of advantages over Revision 5, including long tag name, time stamped data, and measurement status, and enhanced burst mode. When haRT 5 is chosen, the Device Revision of the 1066 is Device Revision 1; when haRT 7 is chosen the Device Revision is Revision 2 (or greater). The Device Revision of the DD (Device Description) and install files for AMS and DeltaV used should be the same as the Device Revision of the 1066. A single HART 5 (Device Revision 1) or HART 7 (Device Revision 2 or greater) DD (Device Description) is used for all model codes of the 1066, which include the pH/ORP, conductivity, oxygen, chlorine, and ozone transmitters. All 1066 transmitters have the same HART device identification, as outlined below: haRT 5 Device Identification (1066 Revision 1): Manufacturer Name: Rosemount Model Name: 1066 Manufacturer ID: 46 (0x2E) Device Type Code: 33 (0x0021) haRT protocol Revision: 5.1 Device Revision: 1 haRT 7 Device Identification (1066 Revision 2): Manufacturer Name: Rosemount Model Name: 1066 Manufacturer ID: 46 (0x2E) Device Type Code: 11809 (0x2E21) haRT protocol Revision: 7.3 Device Revision: 2 HART Communications 93 Section 10: haRT Communication April 2017 10.2 physical Installation and Configuration 10.2.1 HART Wiring and Output Configuration 10.2.2 HART Multidrop (Bus) Configuration 10.2.3 Instruction Manual LIQ-MAN-1066 HART communications is superimposed on Analog Output 1 for all of the measurements and parameters of the 1066. The HART Polling Address should be left at its default value of “0”, unless the 1066 is used in a Multidrop configuration with up to 14 other transmitters. When the Polling Address is greater than “0”, the 4-20 mA output is held at 4 mA or below, and does not change in response to changes in the measurement in HART 5. In HART 7, Loop Current Mode should be set to “Off” to hold the current output to a minimum value. HART Configuration To access the HART Configuration screens, select the “HART” menu item in the Main Menu. If HART 7 is chosen (Univ Cmd Rev = 7), the following controls are available: FIGuRE 10-1. haRT 7 Configuration Screen: Basic Definitions • univ Cmd Rev – toggles between HART version 5 and HART version 7. If the HART host being used can accommodate HART 7, HART 7 should be chosen due to its larger feature set. If the host can only use HART 5, then HART 5 must be chosen. Note: If the 1066 is connected to a HART host and the HART version is changed, the host will likely detect the transmitter as a new transmitter with a different device revision number. 94 • polling address – Choose “0” unless Multidrop is being used. If Multidrop is being used, each transmitter should have its own polling address of from 1 to 15. • loop current mode – Set Output 1 current to a minimum value for multidrop applications (HART 7 only). • Find Device Cmd – Setting Find Device to “On”, enables the 1066 to be identified by the host. The transmitter returns identity information including device type, revision level, and device ID (HART 7 only). HART Communications Instruction Manual Section 10: haRT Communication LIQ-MAN-1066 • April 2017 Burst Message 0, 1, 2 – Toggles burst messages 0, 1, and/or 2 on or off (HART 7 only). See the end of section 10.2 for the HART burst commands available. If HART 5 is chosen (Univ Cmd Rev = 5), the following controls are available: FIGuRE 10-2. haRT 5 Configuration Screen: Basic Definitions • • • univ Cmd Rev – toggles between HART version 5 and HART version 7. If the HART host being used can accommodate HART 7, HART 7 should be chosen due to its larger feature set. If the host can only use HART 5, then HART 5 must be chosen. Note: If the 1066 is connected to a HART host and the HART version is changed, the host will likely detect the transmitter as a new transmitter with a different device revision number. polling address – Choose “0” unless Multidrop is being used. If Multidrop is being used, each transmitter should have its own polling address of from 1 to 15. Burst Mode – Toggles the single HART 5 burst message on or off. See below for the HART burst commands available. Burst Commands available in haRT 5 and haRT 7 If burst messages are enabled by setting the burst messages to on, the information in the burst message can be selected using a HART host from the following commands: • Burst command: Off – Turns burst mode off Cmd 1 – Bursts the Primary Value Cmd 2 – Bursts Loop Current + % of range of the Primary Value Cmd 3 – Bursts Dynamic Variables (PV, SV, TV, & QV) + Loop Current Cmd 9 – Bursts up to 8 Device Variables with time stamp and status and Cmd 48 Additional Transmitter Status (HART 7 only) Cmd 33 – Bursts 4 Device Variables Cmd 48 – Bursts Additional Transmitter Status Bits (HART 7 only) Cmd 93 – Bursts Trend Data (HART 7 only) HART Communications 95 Section 10: haRT Communication April 2017 10.3 Instruction Manual LIQ-MAN-1066 Measurements available via haRT A number of live measurements are made available by HART in addition to the main measurements such as pH or Conductivity. All of these measurements are called Device Variables, which can be mapped to the Dynamic Variables PV, SV, TV, and QV for regular reading by the typical HART host. Each 1066 transmitter type, 1066P, 1066C, etc. will have its own set of Device Variables, based on the secondary measurements used in making the main measurement. appendix 10.1 shows the Device Variable for the each transmitter type, and the Dynamic Variables, which they can be mapped to. 10.4 Diagnostics available via haRT 10.4.1 Status Information – Device Status Bits Bit 0 primary variable out of limits: Bit 1 Non-primary variable out of limits: Bit 2 loop Current Saturated: Bit 3 loop Current Fixed: Bit 4 More Status available: This bit is set when PV is out of its limits. This bit is set when any active device variable other than the Primary variable is out of its limits. This bit is set when Analog Output 1 is not fixed, and it is less than 3.8 mA or greater than 22.0 mA. This bit is set when Analog Output 1 is being simulated, calibrated, or when a device failure is detected and the Analog Output 1 is configured to output a fixed value. The “more status available” bit will be set when the device status condition occurs (i.e. bit goes from 0 to 1) on at least one of the Additional Transmitter Status bits are set. Bit 5 Cold Start: Bit 6 Bit 7 96 This bit is set when a Master Reset is performed either by Command 42, or a power cycle. 2 bits are maintained internally, for primary and secondary masters. Configuration Changed: This bit is set when a configuration or calibration parameter is changed either through a write command or a local interface command. 2 bits are maintained internally, for primary and secondary masters. Field Device Malfunction: This bit is set when a fault condition is detected in the device electronics or sensor. HART Communications Instruction Manual Section 10: haRT Communication LIQ-MAN-1066 10.4.2 10.4.2 10.5 10.5.1 April 2017 Status Information – Extended Device Status Bits (HART 7 only) Bit 0 Maintenance Required: Bit 1 Device variable alert: Bit 2 Critical power Failures: This bit is set when a device fault is detected. This bit is set when any enabled device variable status is not good. This bit is not supported and will always be cleared on 1066. Additional Transmitter Status (Command 48) Additional Transmitter Status provides diagnostic status bits specific to the condition of sensors, electronics, and the memory of the 1066. Calibration errors and notification of events, such as calibration in progress and relay activation are also indicated by status bits. appendix 10.2 shows these bits organized according to the 1066 transmitter measurement type. haRT hosts A HART host can access live measurements, diagnostic messages, and provide a tool for configuring the measurement and calibrating the 1066. The configuration parameters for the 1066 transmitter are listed in appendix 10.3. Two examples of HART hosts are shown below. AMS Intelligent Device Manager The AMS Device Intelligent Device Manager is member of the AMS Suite of asset management applications, which provides tools for configuration, calibration, diagnosing, and documenting transmitters and valves. The following AMS windows are examples of these functions: FIGuRE 10-3. Main Measurement and Overall Status HART Communications 97 Section 10: haRT Communication April 2017 Instruction Manual LIQ-MAN-1066 FIGuRE 10-4. Device variables and Dynamic variables FIGuRE 10-5. Diagnostic Messages (additional Transmitter Status) 98 HART Communications Instruction Manual LIQ-MAN-1066 Section 10: haRT Communication April 2017 FIGuRE 10-6. Configuration FIGuRE 10-7. Calibration HART Communications 99 Section 10: haRT Communication Instruction Manual April 2017 LIQ-MAN-1066 10.5.2 475 Field Communicator 10.6 Wireless Communication using the 1066 HART (and Fieldbus) devices can be accessed in the field using the 475, which provides the same basic functionality as the AMS Intelligent Device Manager. Asset management information can be uploaded into the AMS database from the 475 for a common database for asset management data. The 475 uses a color menu driven display. The 475 Menu for the 1066 appears in Appendix 10.4. The 1066 can communicate by Wireless HART using the Smart Wireless THUM Adaptor and the 1420 Smart Wireless Gateway. All the information available with the wired device can be accessed wirelessly, making it possible to have the measurements and benefits of HART communication in locations where running cable would be difficult or prohibitively expensive. FIGuRE 10-8. 475 Field Communicator FIGuRE 10-9. Wireless Communication using the 1066 Although HART 5 or HART 7 can burst the Dynamic Variables (PV, SV, TV, & QV), HART 7 should be used with the THUM because up to 8 Device Variables can be continually burst using Command 9. 10.7 100 Field Device Specification (FDS) For more details on the implementation of HART in the 1066 and its command structure, the Field Device Specification for the relevant Device Revision should be consulted. They can be downloaded from our website. HART Communications Instruction Manual Section 10: haRT Communication LIQ-MAN-1066 April 2017 appENDIx 10.1 – Device variables 1066 ph Device variables Device variable Name primary value Type: pH (1) ORP (2) Redox (3) Other Device variables: Temperature Sensor mV input Sensor Glass impedance Sensor Reference impedance assignable to Dynamic variables variable Range PV, SV, TV or QV PV, SV, TV or QV PV, SV, TV or QV 0 to 14 pH -1500 to 1500 mV -1500 to 1500 mV SV, TV or QV -15 to 200 °C 5 to 360 °F -750 to 750 mV 0 to 2000 Mµ 0 to 10000 kµ TV or QV TV or QV TV or QV 1066C and 1066T Device variables Device variable Name primary value Type: Conductivity (7) Resistivity (8) % Concentration: NaOH (9) HCl (10) Low H2SO4 (11) High H2SO4 (12) NaCl (13) Custom Concentration (14) assignable to Dynamic variables variable Range PV, SV, TV or QV PV, SV, TV or QV 0 to 2000000 µS/cm 0 to 50000000 µ -cm PV, SV, TV or QV PV, SV, TV or QV PV, SV, TV or QV PV, SV, TV or QV PV, SV, TV or QV PV, SV, TV or QV 0 to 12 % 0 to 15 % 0 to 25 % 96 to 99.7 % 0 to 25 % 0 to 1000 ppm 0 to 1000 mg/L 0 to 100 mg/L 0 to 100 % 0 to 1000 None 0 to 10000 ppm 0 to 36 ppth TDS (15) Salinity (16) other Device variables: Temperature PV, SV, TV or QV PV, SV, TV or QV Temperature resistance Conductance Input resistance Raw Conductivity Raw Resistivity TV or QV TV or QV TV or QV TV or QV TV or QV HART Communications SV, TV or QV -25 to 200 °C -13 to 360 °F 0 to 100000 µ 0 to 2000000 µS 0 to 500 kµ 0 to 2000000 µS/cm 0 to 50000000 ohm-cm 101 Section 10: haRT Communication Instruction Manual April 2017 1066DO Device variables Device variable Name primary value Type: LIQ-MAN-1066 assignable to Dynamic variables variable Range PV, SV, TV or QV 0 to 100 ppm 0 to 1000 ppb Oxygen (17) 0 to 100 mg/L 0 to 1000 µg/L 0 to 300 % Saturation 0 to 760 mmHg 0 to 30 inHg 0 to 1 bar 0 to 1000 mbar 0 to 100 kPa 0 to 1 atm Other Device variables: Temperature PV, SV, TV or QV Temperature resistance Sensor input current TV or QV PV, SV, TV or QV -15 to 200 °C 5 to 360 °F 0 to 100000 Ω 0 to 100000 nA assignable to Dynamic variables variable Range 1066Oz Device variables Device variable Name primary value Type: Ozone (18) PV, SV, TV or QV 0 to 20 ppm 0 to 1000 ppb 0 to 20 mg/L 0 to 1000 μg/L Other Device variables: Temperature PV, SV, TV or QV -15 to 200 °C Temperature resistance Sensor input current TV or QV PV, SV, TV or QV 0 to 100000 μ Device variable Name assignable to Dynamic variables variable Range Chlorine (19) PV, SV, TV or QV 0 to 20 ppm 1066Cl Device variables primary value Type: 5 to 360 °F 0 to 100000 nA 0 to 1000 ppb 0 to 20 mg/L Other Device variables: 0 to 1000 µg/L Temperature PV, SV, TV or QV -15 to 200 °C Temperature resistance TV or QV 0 to 100000 Ω Sensor input current PV, SV, TV or QV pH (free chlorine pH compensation) SV, TV or QV 102 5 to 360 °F 0 to 100000 nA 0 to 14 pH HART Communications Instruction Manual Section 10: haRT Communication LIQ-MAN-1066 April 2017 appENDIx 10.2 – additional Transmitter Status – Command 48 Status Bits 1066 ph Device variables Byte / Bit 1/0 1/1 1/2 1/3 1/4 3/0 3/1 Meaning / Class Device Status Bits Set Cpu Error / Error The software checksum is not as expected. The CPU memory has been corrupted. 4 – More Status available 7 – Field Device Malfunction Factory Data Error / Error An error was detected in the factory segment of the non-volatile memory. At least one factory configuration parameter has been corrupted. 4 – More Status available 7 – Field Device Malfunction Self-Test Fail / Error An electronic component is out of specification. 4 – More Status available 7 – Field Device Malfunction hardware/Software Mismatch / Error The software is not compatible with the hardware. 4 – More Status available 7 – Field Device Malfunction Internal Communications Error / Error The analog input electronics is non-responsive. Keypad Error / Warning At least one key in the device keypad is stuck. This condition makes the local operator interface unuseable. If no other alerts are present, the device can still perform its other functions normally. 4 – More Status available 7 – Field Device Malfunction 4 – More Status available user Data Error / Warning An error was detected in the non-volatile memory. One or more user configuration parameter may be corrupted. Reset analyzer to factory defaults and re-configure the device. If the problem persists, replace device. 4 – More Status available Software Mismatch / Warning The input CPU software is not fully compatible with the main CPU software. 4 – More Status available Maintenance Required / Other 4 – More Status available 4 – More Status available 8/1 Simulation active / Mode A device variable is being simulated. Non-volatile Memory Defect / Warning 8/3 Watchdog Reset Executed / Other 3/2 3/3 3/7 6/6 6/1 8/0 8/2 8/4 8/5 8/6 Need Factory Calibration / Warning The device's non-volatile memory has been corrupted. The device measurements may be out of specification. 4 – More Status available Reset In progress / Other The transmitter's configuration is being reset to factory defaults. 4 – More Status available Device variable alert / Other 4 – More Status available volatile Memory Defect / Error power Supply Condition Out Of Range / Warning Environmental Condition Out Of Range / Warning Electronic Defect / Error Reset device or turn power off then on. If problem persists, replace device. HART Communications 4 – More Status available 4 – More Status available 4 – More Status available 4 – More Status available 4 – More Status available 4 – More Status available 103 Section 10: haRT Communication Instruction Manual April 2017 LIQ-MAN-1066 1066 ph Device variables continued Byte / Bit 8/7 10 / 0 10 / 1 13 / 0 13 / 1 Meaning / Class Device locked / Mode Locked device prevents all host modifications. Unlock device to make changes to the device. 4 – More Status available analog Channel 2 Saturated / Warning The secondary variable is outside the analog output range. 1. Check the secondary value. 2. Check the analog output scaling. 4 – More Status available analog Channel 2 Fixed / Mode Output 2 is either being tested, calibrated, or accidentally left on hold. A fault condition could also set the analog output to a fixed value. If there is no active fault, wait for test or calibration to end or take Output 2 out of hold mode. 4 – More Status available analog Channel 1 Saturated / Warning The primary variable is outside the analog output range. 1. Check the primary value. 2. Check the analog output scaling. 4 – More Status available analog Channel 1 Fixed / Mode Output 1 is either being tested, calibrated, or accidentally left on hold. A fault condition could also set the analog output to a fixed value. If there is no active fault, wait for test or calibration to end or take Output 1 out of hold mode. 4 – More Status available Temperature Status Bits Byte / Bit 0/0 2/0 2/1 2/2 5/1 104 Device Status Bits Set Meaning / Class Temperature Error / Error The temperature measuring circuit is open or shorted. Check the wiring and the temperature element in the sensor. If the temperature element in the sensor is open or shorted, replace the sensor. Temperature high / Warning The measured temperature is above the temperature range of the transmitter and can damage the sensor. The temperature limits are: 1066P / 1066DO, CL, and OZ: Temperature > 150 °C 1066C and 1066T: Temperature > 300 °C 1. Check process temperature. 2. Check sensor and its wiring. Temperature low / Warning The measured temperature is below the temperature range of the transmitter and can damage the sensor. The temperature limits are: 1066P / 1066DO, CL, and OZ: Temperature < -15 °C 1066C and 1066T: Temperature -25 °C 1. Check process temperature. 2. Check sensor and its wiring. RTD Sense line Open / Warning The sense line of the temperature sensor is not connected. 1. Check sensor wiring. 2. If a 2-wire RTD is being used for temperature compensation, use wire jumper to connect sense and return terminals. Temperature Calibration In progress / Other A temperature calibration is or has been performed. Device Status Bits Set 4 – More Status available 7 – Field Device Malfunction 1 – Non-primary variable out of limits 4 – More Status available 1 – Non-primary variable out of limits 4 – More Status available 4 – More Status available 4 – More Status available HART Communications Instruction Manual Section 10: haRT Communication LIQ-MAN-1066 1066ph/ORp Status Bits Byte / Bit 0/1 0/2 0/3 2/3 4/0 4/1 4/2 4/3 5/4 5/5 5/6 April 2017 Meaning/Class Reference Impedance Too high / Error The reference impedance is above the high fault setpoint. The reference electrode may be coated or plugged. 1. Clean or replace the sensor. 2. Check sensor wiring. 3. Increase the setpoint value. 4. Set the reference impedance level to high. Glass Impedance Too high / Error The glass impedance is above the high fault setpoint. The glass electrode may be severely coated. 1. Clean or replace sensor. 2. Check sensor wiring. 3. Increase the glass impedance high fault setpoint. Broken ph Glass / Error The glass impedance is too low. The glass electrode of the pH sensor may be cracked. 1. Check sensor. Replace sensor if cracks are present. 2. Check sensor wiring. 3. Check preamplifier location configuration. Device Status Bits Set 4 – More Status available 7 – Field Device Malfunction 4 – More Status available 7 – Field Device Malfunction 4 – More Status available 7 – Field Device Malfunction ph voltage Too high / Warning The sensor voltage is outside the expected range for a pH measurement. 1. Check sensor wiring. 2. Replace sensor. 4 – More Status available ph Slope Too low / Warning The pH slope calculated during buffer calibration was below the minimum slope limit. The pH electrode may be worn out, damaged, or coated. 1. Check and clean sensor, then retry buffer calibration. 2. Decrease minimum slope limit (default is 40 mV/pH). Note that a pH sensor with a slope less than 50 mV/pH is usually near the end of its useful life. 3. Replace sensor. 4 – More Status available Calibration Error / Warning 4 – More Status available Buffer Calibration In progress / Other A pH buffer calibration is being or has been performed. 4 – More Status available ph Slope Too high / Warning The pH slope calculated during buffer calibration exceeded the maximum slope limit. 1. Check the buffers used and retry buffer calibration. 2. Increase the maximum slope limit (default is 62 mV/pH). Note: A slope of 62 mV/pH or greater indicates that there has been an error made during calibration or a faulty pH sensor. 3. Replace sensor. 4 – More Status available zero Offset Error / Warning The zero offset from a buffer calibration or single point standardization has exceeded the limit. The reference electrode may be poisoned or plugged. 1. Check and clean sensor, then retry buffer calibration. 2. Increase maximum offset limit (default is 60 mV). Note that a pH sensor with an offset of 60 mV or greater is likely poisoned and has to be replaced. 3. Replace sensor. 4 – More Status available ph Standardization In progress / Other A pH standardization is being or has been performed. 4 – More Status available Stabilization In progress / Other A pH sensor is stabilizing or has been stabilizing. HART Communications 4 – More Status available 105 Section 10: haRT Communication Instruction Manual April 2017 LIQ-MAN-1066 1066C and 1066T Status Bits continued Byte / Bit 2/5 2/6 2/7 4/3 4/4 5/2 5/3 5/4 5/5 106 Meaning / Class Need zero Calibration / Warning The sensor offset is too high resulting in a negative reading. This trigger point for this alert is dependent upon the conductivity measurement technology: 1066C: (conductance - zero offset) < -2 µS 1066T: (conductance - zero offset) < -50 µS A sensor zero calibration should be performed. Concentration Out Of Range / Warning The measured concentration is outside the conductivity range where a valid concentration can be derived for the following 5 concentrations: 0 to 20% NaCl 0 to 12% NaOH 0 to 15% HCl 0 to 25% H2SO4 96 to 99.7% H2SO4. If there are no other fault conditions check process temperature and check that the actual concentration is outside the range for which curve is defined. Input Out Of Range / Warning The input is outside the device measurement range. the range limits are: 1066C (2- electrode conductivity): Conductance > 500 mS 1066CT (4- electrode conductivity): (Conductance > 3000 mS) or (Vcond < 0mV) 1066T: Conductance > 1500 mS 1. Check sensor wiring. 2. Replace sensor. Calibration Error / Warning An error occurred in the last calibration procedure. Check sensor and repeat the calibration procedure. Sensor zero Error / Warning An error occurred in the last sensor zero procedure. Check sensor and repeat sensor zero procedure. Sensor zero In progress / Other A sensor zero is being or has been performed. zero Cal In Water In progress / Other A zero calibration in water is being or has been performed. Meter Calibration In progress / Other A meter calibration is being or has been performed. Curve Fit In progress / Other A curve fit is being or has been performed. Device Status Bits Set 4 – More Status available 4 – More Status available 4 – More Status available 4 – More Status available 4 – More Status available 4 – More Status available 4 – More Status available 4 – More Status available 4 – More Status available HART Communications Instruction Manual Section 10: haRT Communication LIQ-MAN-1066 April 2017 1066DO, 1066Cl, 1066Oz Status Bits Byte / Bit 2/4 2/5 4/3 4/4 5/2 5/3 5/6 Meaning / Class SENSOR_CuRRENT > 300 na for OXYGEN BioRx - Other SENSOR_CURRENT > 800 nA for OXYGEN Brewing SENSOR_CURRENT > 106 uA for OXYGEN Oxygen in Gas SENSOR_CURRENT > 350 nA for Ozone SENSOR_CURRENT > 5 uA for Free Chlorine SENSOR_CURRENT > 40 uA for Total Chlorine SENSOR_CURRENT > 20 uA for Monochloramine 1. Make sure the device configuration matches the sensor being used. 2. Check sensor wiring. 3. Replace sensor. Device Status Bits Set 4 – More Status available Need zero Calibration / Warning The sensor offset is too high resulting in a negative reading. The sensor zero limits are: PV < -0.5 if unit is ppm or mg/L PV < -50 if unit is ppb or ug/L PV < -2.0 % for % Saturation PV < -2.0 % for Concentration in Gas PV < -20 ppm for Concentration in Gas PV < -30 mmHg for Partial Pressure Perform sensor a zero calibration. Calibration Error / Warning There has been a calibration error. Check the sensor. 4 – More Status available Sensor zero In progress / Other A sensor zero is being or has been performed. 4 – More Status available Stabilization In progress / Other The sensor is stabilizing or has stabilized. 4 – More Status available Sensor zero Error / Warning An error occurred in the last sensor zero procedure. Check sensor and repeat sensor zero procedure. air Cal In progress / Other An air calibration is being or has been performed. HART Communications 4 – More Status available 4 – More Status available 107 Section 10: haRT Communication Instruction Manual April 2017 LIQ-MAN-1066 appENDIx 10.3 – 1066 haRT Configuration parameters parameters Common to all 1066 Transmitters input_filter_type Selects between continuous and adaptive filtering of the measurement. pv_damping_value Provides the filter time constant. instrument_software_version The software version of the main processor. input_software_version The software version of the input processor. Display parameters: loi_language Selects the language to be displayed by the 1066. Type Enumeration / Range Read/Write Default FLOAT 0.0 -- 99.0 seconds R/W 0 ENUM FLOAT FLOAT ENUM loi_warnings Enables or disables the display of transmitter warnings. loi_configuration_code Locks and unlocks access to configuration from the keypad display of the 1066. "000" disables. loi_calibration_code Locks and unlocks access to calibration from the keypad display of the 1066. "000" disables. loi_main_display_upper Selects the measurement displayed on the upper portion of the main display. loi_main_display_center Selects the measurement displayed on the center portion of the main display. ENUM loi_main_display_left Selects the measurement or output displayed on the left portion of the display. loi_main_display_lower_left Selects the measurement or output displayed on the lower left portion of the display. loi_main_display_right Selects the measurement or output displayed on the right portion of the display. loi_main_display_lower_right Selects the measurement or output displayed on the lower right portion of the display. UINT (2) R/W R Adaptive R English Italiano Français Português Español Chinese Deutsch Russian Enable Disable UINT (2) R/W English R/W Enable R/W 0 R/W Assign PV R/W ENUM Assign PV Blank ENUM ENUM Compensating pH (free Cl) R / W Temperature PV Blank Valid List per 1066 Model R / W ENUM Valid List per 1066 Model R/W ENUM Valid List per 1066 Model R/W ENUM Valid List per 1066 Model R/W 1066ph/ORp valid list Input mv Glass Impedance Ref Impedance Slope zero Offset 1066C/T 108 Adaptive Continuous valid list Raw Conductivity Raw Resistivity 1066DO/Cl/Oz valid list Input Current ph (if free chlorine) all 1066 valid list aO 1 Current aO 1 % Output aO 2 Current aO 2 % Output 0 Temperature HART Communications Instruction Manual Section 10: haRT Communication LIQ-MAN-1066 April 2017 1066 Temperature parameters rtd_offset The temperature offset resulting from a temperature calibration. Type FLOAT rtd_slope FLOAT The slope (dimensionless) resulting from a two point temperature calibration. (Valid for cell constants < 0.02 1/cm) temp_comp_mode Selects automatic or manual temperature compensation. manual_temperature If manual temperature compensation is chosen, provides a constant temperature value used by the transmitter. ENUM FLOAT 1066 ph/ORp Enumeration / Range Manual Automatic Type Enumeration / Range Read/Write Default R/W 1 R/W Automatic R/W 25.0 °C R/W 0 Read/Write Default R/W 0.0 mV zero_offset The zero offset of a pH sensor resulting from a calibration. ph_slope The slope of a pH sensor resulting from a buffer calibration. FLOAT min_ph_slope The minimum value of the slope allowed for a successful calibration. max_ph_slope The maximum value of the slope allowed for a successful calibration. ph_stabilization_time The period of time used to determine stability of a pH sensor during automatic calibration. ph_stabilization_value The pH change used to determine stability of a pH sensor during automatic calibration. ph_buffer_standard The types of pH buffers available for automatic calibration. FLOAT 40 -- 62 mV/pH R / W 40 mV/pH FLOAT 40 -- 62 mV/pH R / W 62 mV/pH FLOAT 0 -- 99 seconds R/W 10 seconds FLOAT 0.01 -- 1.0 pH R/W 0.02 pH ENUM Standard / Nist DIN 19267 Ingold Merck Fisher ENUM analyzer sensor / J-box R/W Standard / Nist zero_offset_limit The maximum value of the zero offset allowed for a successful calibration. preamp_location The location of the preamplifier in a pH measurement. FLOAT 40 -- 62 mV/pH R / W 59.16 mV/pH FLOAT 60 R/W R/W reference_impedance_level ENUM Low R/W Configures the transmitter to use reference electrodes with High impedances less than 500 kohm (low) or greater than 500 kohm (high). ph_solution_temperature_correction ENUM Off Configures the transmitter to correct the pH measurement for changes ultra pure water in the pH of the solution with temperature. high pH ammonia custom ph_solution_temperature_coefficient The coefficient used to correct solution pH when using custom solution pH temperature correction. HART Communications FLOAT -9.999 -- 9.999 pH / °C R/W Low Off 0.000 pH / °C 109 Section 10: haRT Communication Instruction Manual April 2017 1066 ph/ORp continued ph_display_resolution Changes the resolution of the displayed pH value. ph_sensor_isopotential The pH at which the millivolt output of the pH sensor remains constant with temperature changes; virtually always 7.00 pH. refz_high_fault_setpoint The high setpoint that triggers a reference electrode mpedance alert. LIQ-MAN-1066 Type ENUM Read/Write Default R/W 7.00 pH FLOAT 0.01 pH 0.1 pH 0.00 -- 14.00 pH FLOAT 0.0 -- 9,999.0 kohm R/W 500.0 kohm 0.0 -- 9,999.0 Mohm R/W 1,500 Mohm Off On R/W Off Off On R/W Off Advanced Basic R/W Advanced glassz_high_fault_setpoint FLOAT The high setpoint that triggers a glass pH electrode impedance alert. diagnostics_switch ENUM Turns the glass and reference electrode impedance measurements on and off. glassz_temperature_correction ENUM Turns temperature correction of the glass impedance measurement on and off. glassz_measurement_type Toggles between a basic glass impedance measurement, and an advanced impedance measurement that is more accurate. Enumeration / Range ENUM R/W FLOAT R/W calculated_slope FLOAT The preliminary slope calculated by a pH calibration before it is accepted by the transmitter. R/W calculated_offset The preliminary zero offset calculated by a pH calibration before it is accepted by the transmitter. 110 0.01 pH HART Communications Instruction Manual Section 10: haRT Communication LIQ-MAN-1066 1066C and 1066T April 2017 Type Enumeration / Range Read/Write cell_constant The cell constant of the conductivity sensor being used, typically determined by calibration. FLOAT cable_correction Selects automatic or manual cable resistance correction. µS/cm mS/cm R/W 0.0001-- 100.0 /cm R/W 1.00 /cm ENUM Manual Automatic R/W Automatic manual_cable_resistance The known cable resistance used in manual cable correction. FLOAT 0.0 -- 99.99 ohm R/W 0.0 ohm FLOAT R only 0.0 µS zero_offset_in_soln The zero offset determined by a zero calibration in a solution. FLOAT R only 0.00% R/W 3 conductivity_unit The conductivity unit used by the measurement. zero_offset_in_air The zero offset determined by a zero calibration in air. ENUM Default custom_curve_num_data_points ENUM The number of conductivity and concentration points to be used to calculate a custom concentration curve. Custom Curve Concentration Data points: 2 3 4 5 custom_curve_concentration_1 FLOAT R/W custom_curve_concentration_3 FLOAT R/W custom_curve_concentration_2 custom_curve_concentration_4 custom_curve_concentration_5 Custom Curve Conductivity Data points: FLOAT FLOAT FLOAT R/W R/W R/W custom_curve_conductivity_1 FLOAT R/W custom_curve_conductivity_3 FLOAT R/W custom_curve_conductivity_2 custom_curve_conductivity_4 custom_curve_conductivity_5 HART Communications FLOAT FLOAT FLOAT µS/cm R/W R/W R/W 111 Section 10: haRT Communication Instruction Manual April 2017 1066C LIQ-MAN-1066 Type electrode_type ENUM Type of contacting conductivity sensor being used by the transmitter. cond_range Selects a particular range for the conductivity measurement or automatic ranging for 2 and 4 electrode sensors. series_cap_correction Turns capacitance correction on and off. ENUM 4-Electrode: Automatic 0-42μS 36-200μS 180-1000μS 0.9-10mS/cm 9-600mS/cm ENUM FLOAT temp_comp_slope Provides a slope value used in linear temperature compensation. FLOAT 1066T toroidal_sensor_model Selects the model of toroidal sensor being used for enhanced accuracy. cond_range Selects a particular range for the conductivity measurement or automatic ranging for toroidal sensors. temp_comp_type The type of temperature compensation used by the transmitter. temp_comp_slope Provides a slope value used in linear temperature compensation. reference_temperature The temperature that temperature compensation corrects to. It is usually 25 °C. 112 2-Electrode 4-Electrode Read/ Write Default R / W 2-Electrode R / W Automatic 2-Electrode: Automatic 0-50 µS 40-500 µS 400-2000 µS 1.8-20 mS 18-200 mS cell_factor The second calibration constant used by a 4 electrode sensor in addition to cell constant. temp_comp_type The type of temperature compensation used by the transmitter. reference_temperature The temperature that temperature compensation corrects to. It is usually 25 °C. Enumeration / Range ENUM Off On R / W Off Linear slope Neutral salt Cation Raw / None R / W Linear slope R / W 0.95 /cm 0.2 -- 9.99 % / °C FLOAT Type R / W 2.0% / °C R / W 25 °C Enumeration / Range Read/Write Default ENUM 228 225 226 247 Other R/W 228 ENUM Automatic 50-600 µS 0.5-100 mS 90-1500 mS R/W Automatic ENUM Linear slope Raw / None R/W Linear slope FLOAT R/W 2.0% / °C FLOAT R/W 25 °C HART Communications Instruction Manual Section 10: haRT Communication LIQ-MAN-1066 1066Oz/Cl/DO April 2017 Type polar_voltage FLOAT The polarization voltage used by the transmitter; automatically set by sensor selection. temp_coeff FLOAT The temperature coefficient used to compensate temperature; automatically set by sensor selection. amp_sensor_sensitivity FLOAT The response of the sensor to changes in concentration, determined by calibration. amp_sensor_zero_current FLOAT The current output of the sensor when the concentration is 0. It is determined by a zero calibration. 1066Oz chlorine_type The resolution of the displayed ozone measurement. 1066Cl chlorine_type The type of chlorine being measured. chlorine_resolution The resolution of the displayed chlorine measurement. ph_correction_mode Turns automatic pH compensation of free chlorine measurements on and off. Read/Write Default R/W 0 0.1 -- 1,000,000.0 nA/ppm R/W 2,500 nA/ppm -999.9 -- 999.9 nA R/W 0.0 nA R/W 0 Type Enumeration / Range Read/Write Default Type Enumeration / Range Read/Write Default ENUM 0.001 ppm 0.01 ppm R/W 0.001 ppm On Off R/W Off 0.00 -- 14.00 pH R/W 7.00 pH ENUM ENUM ENUM manual_ph FLOAT Provides a constant pH value, if manual pH compensation of a free chlorine measurement is used. HART Communications Enumeration / Range 0.001 0.01 Free Chlorine Total Chlorine Chloramine R/W R/W 0.001 Free Chlorine 113 Section 10: haRT Communication Instruction Manual April 2017 1066DO LIQ-MAN-1066 Type oxygen_sensor_type The type of oxygen sensor used. ENUM oxygen_measurement_type The type of oxygen measurement being made. ENUM oxygen_salinity The salinity of the process solution, which is used by the transmitter to correct for the effect of salinity on the oxygen measurement. FLOAT pressure_units The units of pressure used by the oxygen measurement. ENUM oxygen_process_pressure The process pressure used by the transmitter to calculate percent oxygen in gas or % saturation. FLOAT Enumeration / Range Read/Write Default Water/Waste Trace Oxygen BioRx-Rosemount BioRx-Other Brewing Oxygen In Gas R/W Water/Waste Concentration in R/W Liquid Percent Saturation Partial Pressure Concentration in Gas Concentration in Liquid 0.0 - 99.9 ppth R/W 0.0 ppth mmHg inHg psi atm kPa mbar bar R/W mmHg R/W 760.0 mmHg R/W 760.0 mmHg oxygen_air_pressure FLOAT The barometric pressure used by the transmitter during an air calibration. oxygen_units ENUM The available units used for the various oxygen measurement types. Concentration units: in liquid: ppm ppb mg/l �g/l percent Saturation partial pressure Oxygen In Gas air_cal_stabilization_time FLOAT The period of time used to determine stability of an oxygen sensor during an air calibration. air_cal_stabilization_value FLOAT The change in the oxygen measurement used to determine stability of an oxygen sensor during an air calibration. air_cal_stabilization_value_unit FLOAT The unit used during an air calibration. 114 units: % units: mmhg units: ppm % 0 -- 99 seconds 0.001 -- 9,999.0 R/W ppm % kpa inhg mbar psi % R/W 10 seconds R/W 0.05 R/W ppm HART Communications Instruction Manual LIQ-MAN-1066 Section 10: haRT Communication April 2017 appENDIx 10.4 – 475 Menu Tree for 1066 haRT 7 HART Communications 1. Overview 1.1. Device Status 1.2. Comm Status 1.3. PV 1.4. SV 1.5. Loop Current 1.6. % of Range 1.7. Find Device 1.8. Device Information 1.8.1. Identification 1.8.1.1. Tag 1.8.1.2. Long Tag 1.8.1.3. Model 1.8.1.4. Serial Number 1.8.1.5. Date 1.8.1.6. Descriptor 1.8.1.7. Message 1.8.2. Revision Numbers 1.8.2.1. Universal 1.8.2.2. Field Device 1.8.2.3. DD Revision 1.8.2.4. Hardware 1.8.2.5. Instrument Software 1.8.2.6. Input Software 2. Configure 2.1. Guided Setup 2.1.1. Basic Setup 2.2. Manual Setup 2.2.1. Device Information 2.2.1.1. Tag 2.2.1.2. Long Tag 2.2.1.3. Descriptor 2.2.1.4. Message 2.2.1.5. Date 2.2.1.6. Current Date 2.2.1.7. Current Time 2.2.1.8. Set Date and Time 2.2.2. Measurement (1066 ph/ORp) 2.2.2.1. PV is 2.2.2.2. Preamp Location 2.2.2.3. Soln Temp Correct’n (pH only) 2.2.2.4. Temp Coefficient (PV is pH and Soln Temp Correct’n is Custom) 2.2.2.5. Resolution (pH only) 2.2.2.6. Sensor Isopotential (pH only) 2.2.2.7. Input Filter 2.2.2.8. Filter Type 2.2.2.9. Reference Imp 2.2.3. Measurement (1066 DO) 2.2.3.1. Sensor Type 2.2.3.2. Measurement Type 2.2.3.3. PV Unit 2.2.3.4. Salinity 2.2.3.5. Pressure Unit 2.2.3.6. Input Filter 2.2.3.7. Filter Type 2.2.3.8. Polar Voltage 2.2.3.9. Temp Coefficient 115 Section 10: haRT Communication April 2017 Instruction Manual LIQ-MAN-1066 2.2.4. Measurement (1066 Ozone) 2.2.4.1. PV Unit 2.2.4.2. Resolution 2.2.4.3. Input Filter 2.2.4.4. Filter Type 2.2.4.5. Polar Voltage 2.2.4.6. Temp Coefficient 2.2.5. Measurement (1066 Chlorine) 2.2.5.1. Measurement Type 2.2.5.2. PV Unit 2.2.5.3. Resolution 2.2.5.4. pH Correction (Free Chlorine Measurement Type only) 2.2.5.4.1. pH Correction If pH Correction is manual 2.2.5.4.2. Manual pH If pH Correction is Live/Continuous 2.2.5.4.3. Preamp Location 2.2.5.4.4. Resolution 2.2.5.4.5. Sensor Isopotential 2.2.5.4.6. Soln Temp Correct’n 2.2.5.5. Input Filter 2.2.5.6. Filter Type 2.2.5.7. Polar Voltage 2.2.5.8. Temp Coefficient 2.2.6. Measurement (1066 Contacting & Torridal Conductivity) 2.2.6.1. Measurement Type 2.2.6.2. PV Unit 2.2.6.3. Sensor Config 2.2.6.3.1. Sensor Type (Contacting Model) 2.2.6.3.2. Sensor Model (Toroidal Model) 2.2.6.3.3. Range 2.2.6.3.4. Cell Constant 2.2.6.3.5. Cell Factor (4-Electrode) 2.2.6.3.6. RTD Offset (Cell Constant < 0.02) 2.2.6.3.7. RTD Slope (Cell Constant < 0.02) 2.2.6.4. Setup Custom Curve (Custom Curve Measurement) 2.2.6.5. Temp Comp 2.2.6.5.1. Temp Comp Type 2.2.6.5.2. Temp Slope 2.2.6.5.3. Reference Temp 2.2.6.6. Cable Correction (2-Electrode or Toroidal) 2.2.6.6.1. Series Cap Corr 2.2.6.6.2. Cable R. Correction 2.2.6.6.3. Cable Resistance 2.2.6.7. Input Filter 2.2.6.8. Filter Type 2.2.7. Temperature 2.2.7.1. Temperature Unit 2.2.7.2. Temp Comp 2.2.7.3. Manual Value (If Temp Comp is Manual) 2.2.8. analog Outputs 2.2.8.1. Output 1 2.2.8.1.1. Primary Variable 2.2.8.1.2. PV URV 2.2.8.1.3. PV LRV 2.2.8.1.4. Scale 2.2.8.1.5. Dampening 2.2.8.1.6. Fault Mode 2.2.8.1.7. Fault Value 116 HART Communications Instruction Manual LIQ-MAN-1066 Section 10: haRT Communication April 2017 2.2.8.2. Output 2 2.2.8.2.1. Secondary Variable 2.2.8.2.2. SV URV 2.2.8.2.3. SV LRV 2.2.8.2.4. Scale 2.2.8.2.5. Dampening 2.2.8.2.6. Fault Mode 2.2.8.2.7. Fault Value 2.2.9. Communications 2.2.9.1. Burst Message 1 2.2.9.1.1. Burst Message 2.2.9.1.2. Message Content 2.2.9.1.3. Update Rate 2.2.9.1.3.1. Trigger Mode 2.2.9.1.3.2. Trigger Level 2.2.9.1.3.3. Trigger Level Unit 2.2.9.1.3.4. Classification 2.2.9.1.3.5. Update Rate 2.2.9.1.3.6. Default Update Rate 2.2.9.2. Burst Message 2 2.2.9.2.1. Burst Message 2.2.9.2.2. Message Content 2.2.9.2.3. Update Rate 2.2.9.2.3.1. Trigger Mode 2.2.9.2.3.2. Trigger Level 2.2.9.2.3.3. Trigger Level Unit 2.2.9.2.3.4. Classification 2.2.9.2.3.5. Update Rate 2.2.9.2.3.6. Default Update Rate 2.2.9.3. Burst Message 3 2.2.9.3.1. Burst Message 2.2.9.3.2. Message Content 2.2.9.3.3. Update Rate 2.2.9.3.3.1. Trigger Mode 2.2.9.3.3.2. Trigger Level 2.2.9.3.3.3. Trigger Level Unit 2.2.9.3.3.4. Classification 2.2.9.3.3.5. Update Rate 2.2.9.3.3.6. Default Update Rate 2.2.9.4. Event Notification 2.2.9.4.1. Event Message 2.2.9.4.2. Unack Update Rate 2.2.9.4.3. Default Update Rate 2.2.9.4.4. Debounce Interval 2.2.9.4.5. Pending Events 2.2.9.4.6. Acknowledge Event (If there are pending events) 2.2.9.5. 12-Point Sample 2.2.9.5.1. Data Sampling 2.2.9.5.2. Device Variable 2.2.9.5.3. Sample Interval 2.2.9.5.4. 12-Point Sample (Data Sampling is not Off) 2.2.9.6. Variable Mapping 2.2.9.6.1. PV is 2.2.9.6.2. SV is 2.2.9.6.3. TV is 2.2.9.6.4. QV is 2.2.9.7. Multidrop 2.2.9.7.1. Polling Address 2.2.9.7.2. Loop Current Mode HART Communications 117 Section 10: haRT Communication April 2017 Instruction Manual LIQ-MAN-1066 2.2.10. Device Display 2.2.10.1. Main Display Format 2.2.10.1.1. Upper 2.2.10.1.2. Center 2.2.10.1.3. Left 2.2.10.1.4. Lower Left 2.2.10.1.5. Right 2.2.10.1.6. Lower Right 2.2.10.2. Display Language 2.2.10.3. Warnings 2.2.10.4. Display Contrast 2.2.11.Security 2.2.11.1. HART Lock 2.2.11.1.1. Lock State 2.2.11.1.2. Lock/Unlock 2.2.11.2. Local Operator Interface 2.2.11.2.1. Configuration Code 2.2.11.2.2. Calibration Code 2.3. alert Setup 2.3.1. Diagnostics (1066 ph/ORp/Redox only) 2.3.1.1. Reference Imp 2.3.1.1.1. High Fault Setpoint 2.3.1.2. Glass Impedance (pH only) 2.3.1.2.1. High Fault Setpoint 2.3.1.2.2. Temp Correction 2.3.1.2.3. Measurement Type 3. Service Tools 3.1. Alerts 3.2. Variables 3.3. Trends 3.4. Maintenance 3.4.1. 1066 DO (1066 Dissolved Oxygen Only) 3.4.1.1. Oxygen 3.4.1.2. Temperature 3.4.1.3. Input Current 3.4.1.4. Air Calibration 3.4.1.5. Air Calibration Setup 3.4.1.6. Zero Calibration 3.4.1.7. In-Process Cal 3.4.1.8. Sensitivity@25 °C 3.4.1.9. Zero Current 3.4.2. 1066 Cl /1066 Oz (1066 Chlorine /1066 Ozone Only) 3.4.2.1. Chlorine/Ozone 3.4.2.2. Temperature 3.4.2.3. Input Current 3.4.2.4. Zero Calibration 3.4.2.5. In-Process Cal 3.4.2.6. Sensitivity@25 °C 3.4.2.7. Zero Current 118 HART Communications Instruction Manual LIQ-MAN-1066 Section 10: haRT Communication April 2017 3.4.3. 1066 ph/ORp (ph Only) 3.4.3.1. pH 3.4.3.2. pH Buffer Calibration 3.4.3.3. Standardize pH 3.4.3.4. pH Slope 3.4.3.5. Zero Offset 3.4.3.6. Calibration Setup 3.4.3.6.1. Limits 3.4.3.6.1.1. Minimum Slope 3.4.3.6.1.2. Maximum Slope 3.4.3.6.1.3. Maximum Offset 3.4.3.6.2. Automatic Buffer Recognition 3.4.3.6.2.1. Buffer Standard 3.4.3.6.2.2. Stable Time 3.4.3.6.2.3. Stable Delta 3.4.4. 1066 ph/ORp (ORp/Redox Only) 3.4.4.1. ORP/Redox 3.4.4.2. Calibrate ORP/Redox 3.4.4.3. Zero Offset 3.4.4.4. Maximum Offset 3.4.5. 1066 C/1066 T (1066 Contacting/Torridal Only) 3.4.5.1. Conductivity 3.4.5.2. Raw Conductivity 3.4.5.3. Temperature 3.4.5.4. Zero Calibration 3.4.5.5. In-Process Calibration 3.4.5.6. Cell Constant 3.4.5.7. Cell Factor (4 electrode) 3.4.5.8. Zero Offset 3.4.5.9. Soln Offset (%Concentration) 3.4.6. Temperature 3.4.6.1. Temperature 3.4.6.2. Calibrate Temp 3.4.6.3. RTD Offset 3.4.7. analog Outputs 3.4.7.1. Output 1 3.4.7.1.1. Loop Current 3.4.7.1.2. Calibrate Output 1 3.4.7.2. Output 2 3.4.7.2.1. SV Current 3.4.7.2.2. Calibrate Output 2 3.4.8. Meter 3.4.8.1. Input Resistance 3.4.8.2. Meter Calibration 3.4.9. Reset/Restore 3.4.9.1. Reset Device 3.4.9.2. Load Default Configuration 3.4.9.3. Reset Configuration Changed 3.5. Simulate 3.5.1. PV (show current PV type label) 3.5.2. Temperature 3.5.3. End Variable Simulation (If any variable is being simulated) 3.5.4. Output 1 3.5.5. Output 2 HART Communications 119 Section 10: haRT Communication April 2017 120 Instruction Manual LIQ-MAN-1066 HART Communications Instruction Manual Section 11: Return of Material LIQ-MAN-1066 April 2017 Section 11: Return of Material 11.1 11.2 General To expedite the repair and return of instruments, proper communication between the customer and the factory is important. Before returning a product for repair, call 1-949-757-8500 for a Return Materials Authorization (RMA) number. Warranty Repair The following is the procedure for returning instruments still under warranty: 1. Call Emerson for authorization. 2. To verify warranty, supply the factory sales order number or the original purchase order number. In the case of individual parts or sub-assemblies, the serial number on the unit must be supplied. 3. Carefully package the materials and enclose your “Letter of Transmittal”. If possible, pack the materials in the same manner as they were received. 4. Send the package prepaid to: Emerson 8200 Market Blvd, Chanhassen, MN 55317 Attn: Factory Repair RMA No. ____________ Mark the package: Returned for Repair Model No. ___________ IMpORTaNT 11.1 please see second section of “Return of Materials Request” form. Compliance with the OSha requirements is mandatory for the safety of all personnel. MSDS forms and a certification that the instruments have been disinfected or detoxified are required. Non-Warranty Repair The following is the procedure for returning for repair instruments that are no longer under warranty: 1. Call Emerson for authorization. 2. Supply the purchase order number, and make sure to provide the name and telephone number of the individual to be contacted should additional information be needed. 3. Do Steps 3 and 4 of Section 10.2. NOTICE Consult the factory for additional information regarding service or repair. Return of Materials 121 EC Declaration of Conformity April 2017 122 Instruction Manual LIQ-MAN-1066 Return of Material Instruction Manual LIQ-MAN-1066 EC Declaration of Conformity April 2017 Note: Please see website for most recent Declaration. EC Declaration of Conformity 123 LIQ-MAN-1066 Rev. J April 2017 www.Emerson.com/Rosemountliquidanalysis Youtube.com/user/Rosemount Analyticexpert.com Twitter.com/Rosemount_News facebook.com/Rosemount Emerson 8200 Market Blvd. Chanhassen, MN 55317, USA Tel +1 800 999 9307 Fax +1 952 949 7001 liquid.CSC@Emerson.com ©2017 Emerson. All rights reserved. The Emerson logo is a trademark and service mark of Emerson Electric Co. Rosemount is a mark of one of the Emerson family of companies. All other marks are the property of their respective owners. The contents of this publication are presented for information purposes only, and while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice.
Source Exif Data:
File Type : PDF File Type Extension : pdf MIME Type : application/pdf PDF Version : 1.4 Linearized : Yes Has XFA : No XMP Toolkit : Adobe XMP Core 5.2-c001 63.139439, 2010/09/27-13:37:26 Create Date : 2017:04:21 14:33:02+05:30 Creator Tool : QuarkXPress(R) 11.1 Modify Date : 2017:07:13 00:21:06+08:00 Metadata Date : 2017:07:13 00:21:06+08:00 Producer : QuarkXPress(R) 11.1 Format : application/pdf Title : a X Press Private : %%DocumentCustomColors: (PANTONE 288 C).%%+ (RTF R(0)G(0)B(0)).%%+ (PANTONE 144 4c).%%DocumentProcessColors: Cyan Magenta Yellow Black.%%CMYKCustomColor: 0 .52 1 0 (PANTONE 144 4c).%%QRKLabCustomColor: .18 .53 .303 (PANTONE 288 C).%%RGBCustomColor: 0 .114 .446 (PANTONE 288 C).%%+ 0 0 0 (RTF R(0)G(0)B(0)).%%EndComments Document ID : uuid:466c53e6-18fb-4416-8326-508078f8e5e1 Instance ID : uuid:c4f2ad36-ca2e-4c67-9437-82cb82daaf42 Page Count : 134 Creator : QuarkXPress(R) 11.1EXIF Metadata provided by EXIF.tools