MSOX4104A Mixed Signal Oscilloscope 1 GHz 4 Analog Plus 16 Digital Channels Data Sheet Brochure

DSOX4054A-Oscilloscope-500-MHz-4-Analog-Channels-Data-Sheet-Brochure DSOX4054A-Oscilloscope-500-MHz-4-Analog-Channels-Data-Sheet-Brochure

MSOX4032A-Mixed-Signal-Oscilloscope-350-MHz-2-Analog-Plus-16-Digital-Channels-Data-Sheet-Brochure MSOX4032A-Mixed-Signal-Oscilloscope-350-MHz-2-Analog-Plus-16-Digital-Channels-Data-Sheet-Brochure

MSOX4052A-Mixed-Signal-Oscilloscope-500-MHz-2-Analog-Plus-16-Digital-Channels-Data-Sheet-Brochure MSOX4052A-Mixed-Signal-Oscilloscope-500-MHz-2-Analog-Plus-16-Digital-Channels-Data-Sheet-Brochure

MSOX4054A-Mixed-Signal-Oscilloscope-500-MHz-4-Analog-Plus-16-Digital-Channels-Data-Sheet-Brochure MSOX4054A-Mixed-Signal-Oscilloscope-500-MHz-4-Analog-Plus-16-Digital-Channels-Data-Sheet-Brochure

DSOX4052A-Oscilloscope-500-MHz-2-Analog-Channels-Data-Sheet-Brochure DSOX4052A-Oscilloscope-500-MHz-2-Analog-Channels-Data-Sheet-Brochure

User Manual: MSOX4104A-Mixed-Signal-Oscilloscope-1-GHz-4-Analog-Plus-16-Digital-Channels-Data-Sheet-Brochure

Open the PDF directly: View PDF PDF.
Page Count: 39

Keysight Technologies
InfiniiVision 4000 X-Series Oscilloscopes
Data Sheet
02 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Table of Contents
Oscilloscope Experience Redefined: Experience the Speed,
Usability, and Integration �������������������������������������������������������������������������������������������������������������������3
Oscilloscope Experience Redefined: Experience the Speed ������������������������������������������������������������������������5
Oscilloscope Experience Redefined: Experience the Usability ��������������������������������������������������������������������9
Oscilloscope Experience Redefined: Experience the Integration ��������������������������������������������������������������� 11
Oscilloscope Experience Redefined: Other Key Productivity Tools ������������������������������������������������������������ 14
Oscilloscope Experience Redefined ������������������������������������������������������������������������������������������������������������22
Oscilloscope Experience Redefined: Configuring Your InfiniiVision
4000 X-Series Oscilloscope ������������������������������������������������������������������������������������������������������������24
Oscilloscope Experience Redefined: InfiniiVision 4000 X-Series
Performance Characteristics �����������������������������������������������������������������������������������������������������������26
Oscilloscope Experience Redefined:
License-only Bandwidth Upgrades and Measurement Applications ���������������������������������������������35
Oscilloscope Experience Redefined: Return-to-Keysight Service
Center Bandwidth Upgrades �����������������������������������������������������������������������������������������������������������36
Keysight Oscilloscopes ��������������������������������������������������������������������������������������������������������������������������������37
Evolving Since 1939 �������������������������������������������������������������������������������������������������������������������������������������39
Need more bandwidth, sampling rate, and analysis?
Consider the Infiniium 6000 X-Series
1, 2.5, 4, and 6 GHz
20 GSa/s
Ultra-low noise at 1 mV/div
12.1-inch multi-touch capacitive display with gesture support
Standard color grade, histogram, and enhanced FFT
Optional jitter and real-time eye diagram analysis
See www.keysight.com/find/6000X-Series for more details.
03 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Imagine an oscilloscope that sees everything, triggers on
anything, has the ease-of-use of a tablet device…and grows with
your projects.
The Keysight Technologies, Inc. 4000 X-Series oscilloscopes are
engineered for next-generation performance, delivering waveform
update rates 20 times faster than the competition to display the
most signal detail. An industry-leading 12.1-inch capacitive touch
screen with innovative hardware-based zone touch triggering
provides the most intuitive interface to get you answers faster.
The 4000 X-Series provides maximum investment protection with
fully upgradable 5-instruments-in-1.
Experience the speed
Anomalies and elusive events are the toughest to debug. The
4000 X-Series oscilloscope redefines your debugging experience
with MegaZoom IV smart memory technology. The industry-
leading 1-million-waveforms-per-second update rate, means you
see more of your signal behavior and can feel more confident in
your design.
Experience the usability
You may be surprised just how easy it is to use the InfiniiVision
4000 X-Series. A 12.1-inch capacitive touch screen − the
industry’s largest − works just like your favorite tablet or smart
phone, so debugging your devices is faster than ever before.
Innovative zone touch triggering makes triggering on anything
a snap. Just draw a box around signals of interest and the
oscilloscope triggers on them. So, if you can see it, you can
trigger on it.
Experience the integration
The 4000 X-Series further redefines your oscilloscope experience
by integrating the capabilities of seven instruments in one:
oscilloscope channels, logic channels, digital voltmeter (DVM),
dual-channel WaveGen function/arbitrary waveform generator,
frequency response analyzer, 8-digit hardware counter, and serial
protocol analyzer including USB. All are upgradable, including
bandwidth, for the ultimate investment protection.
Oscilloscope Experience Redefined:
Experience the Speed, Usability, and Integration
4000 X-Series – Oscilloscope experience
redefined
Key features:
Experience the speed:
One million waveforms per second update rate
MegaZoom IV smart memory technology
Standard segmented memory
Experience the usability:
Industry’s first capacitive touch screen
Industry’s largest 12-inch display
Zone touch trigger
Experience the integration:
Industry’s first 5 instruments in 1
Industry’s first fully upgradable including bandwidth
to 1.5 GHz
Industry’s leading application solutions
Figure 1. MegaZoom IV smart memory technology enables the speed,
usability, and integration.
Figure 2. Take advantage of a new oscilloscope application bundle
that will enable ALL software applications (including serial decode and
WaveGen) for one low price (Option DSOX4APPBNDL).
04 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Experience the Speed, Usability, and Integration (Continued)
Overview of Keysight InfiniiVision X-Series oscilloscopes
InfiniiVision 6000 X-Series 4000 X-Series 3000T X-Series 2000 X-Series 1000 X-Series
Analog channels 2 or 4 2 or 4 2 or 4 2 or 4 2
Digital channels 16 (MSO models or
upgrade)
16 (MSO models or
upgrade)
16 (MSO models or
upgrade)
8 (MSO models or
upgrade)
External trigger can be
used as a 3rd digital
channel
Bandwidth 1, 2.5, 4, 6 GHz 200, 350, 500 MHz, 1,
1.5 GHz
100, 200, 350, 500 MHz,
1 GHz
70, 100, 200 MHz 50, 70, 100 MHz
Max sample rate 20 GSa/s 5 GSa/s 5 GSa/s 2 GSa/s 2 GSa/s
Max memory depth 4 Mpts 4 Mpts 4 Mpts 100 kpts (standard)
1 Mpts (option)
1 Mpts
Max waveform update
rate
> 450,000 wfms/sec > 1,000,000 wfms/sec > 1,000,000 wfms/sec > 50,000 wfms/sec > 50,000 wfms/sec
Display 12.1-inch, capacitive
touch
12.1-inch, capacitive
touch
8.5-inch, capacitive
touch
8.5-inch 7 inch
Zone touch trigger Standard Standard Standard No No
Voice control Standard No No No No
WaveGen 20-MHz
function/arbitrary
waveform generator
Dual-channel AWG
(option)
Dual-channel AWG
(option)
Single-channel AWG
(option)
Single-channel function
(option)
Single-channel function
(on G models)
Integrated DVM Standard Standard Standard Standard Free with registration
Integrated hardware
counter (standard)
10-digit frequency,
period, or totalizer
counter
5-digit frequency or
period counter
(8 digits with external
10 MHz clock reference)
8-digit frequency,
period, or totalizer
counter
5-digit frequency counter 5 digits
Search and navigate Standard, lister
supported
Standard, lister
supported
Standard, lister
supported
Serial only No
Segment memory Standard Standard Standard Option Standard on DSO models
Mask/limit test Option Option Option Option Standard on DSO models
Serial protocol
analysis options
I2C/SPI, UART/RS232,
CAN/CAN-dbc/CAN-FD/
LIN/LIN symbolic,
SENT, FlexRay, I2S,
MIL-STD1553, CXPI,
ARINC429, USB 2.0,
Manchester/NRZ, USB
PD
I2C/SPI, UART/RS232,
CAN/CAN-dbc/CAN-FD/
LIN/LIN symbolic,
SENT, FlexRay, I2S,
MIL-STD1553, CXPI,
ARINC429, USB 2.0,
Manchester/NRZ, USB
PD
I2C/SPI, UART/RS232,
CAN/CAN-dbc/CAN-FD/
LIN/LIN symbolic,
SENT, FlexRay, I2S,
MIL-STD1553, CXPI,
ARINC429, Manchester/
NRZ, USB PD
I2C/SPI, UART, CAN/
LIN (will not operate
simultaneously with
digital channels)
I2C/SPI, UART, CAN/LIN
Advanced analysis
options
Power analysis, USB 2.0
signal quality test, HDTV
analysis, FRA
Power analysis, USB 2.0
signal quality test, HDTV
analysis, FRA, NFC
Power analysis, HDTV
analysis, FRA, NFC
No FRA
Color grade Standard No No No No
Histogram Standard No No No No
FFT Standard enhanced FFT Standard enhanced FFT Standard enhanced FFT Standard Standard
Multi-domain analysis Gated FFT Gated FFT Gated FFT No No
Jitter analysis with
clock recovery
Option No No No No
Realtime eye diagram Option No No No No
Advanced math Standard, display four
functions simultaneously
Standard, display one
function
Standard, display one
function
No No
Connectivity Standard USB 2.0, LAN,
video (GPIB option), USB
mouse and keyboard
support
Standard USB 2.0, LAN,
video (GPIB option), USB
mouse and keyboard
support
Standard USB 2.0 (LAN/
video/GPIB option), USB
mouse and keyboard
support
Standard USB 2.0 (LAN/
video/GPIB option), USB
keyboard support
Standard USB 2.0, USB
keyboard support
05 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
What is waveform update rate and why is it important?
As oscilloscopes acquire data, process it, and plot it to the screen, there is inevitable “dead time,” or the time oscilloscopes miss
signals completely. In general, the faster the waveform update rate, the shorter the dead time. The shorter the dead time, the more
likely an oscilloscope is to capture anomalies and infrequent events. This is why it is critical to select an oscilloscope with a fast
waveform update rate.
Figure 3. The 4000 X-Series captures a glitch occurring once in a million
waveform cycles.
One million waveforms per second update rate
If you can’t see the problem, it is hard to troubleshoot it. With an
industry-leading one million waveforms per second update rate,
the InfiniiVision 4000 X-Series gives you the highest probability of
capturing random and infrequent events that you would miss on
an oscilloscope with a lower waveform update rate.
Powered by MegaZoom IV smart memory technology,
the InfiniiVision 4000 X-Series not only lets you see more
waveforms, but it has the uncompromised ability to find the most
difficult problems in your design. Unlike other oscilloscopes,
uncompromised ability means:
Always-fast, responsive operation
No slowdown with logic channels on
No slowdown with protocol decoding on
No slowdown with math functions turned on
No slowdown with measurements turned on
Figure 4. Other vendor’s oscilloscope with 50,000 waveforms/second. A long dead time decreases your chances of
capturing infrequent events.
Figure 5. InfiniiVision 4000 X-Series with 1,000,000 waveforms/second. A short dead time increases your chances of
capturing infrequent events.
Oscilloscope Experience Redefined:
Experience the Speed
06 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
MegaZoom IV ASIC
Analog
signals
Plotter
GUI
Mask
DRAM
WaveGen
Synthesis
Trigger
(InfiniiScan and
traditional)
DRAM
(Acquisition memory)
Acquisition
memory manager
(Segmented memory)
CPU
(Math, I/O, GUI)
Handware
accelerated
(Measurements, Search)
Serial
decoders
16 digital
signals
Keysight achieves this industry-leading waveform update rate with
MegaZoom IV smart memory technology
Traditionally, CPU processing was the major bottleneck for oscilloscope waveform
update rate and responsiveness. Typically, the CPU handles interpolations, logic channel
plotting, serial bus decoding, measurements and more, and the waveform update rate
drops dramatically as these features are turned on.
The InfiniiVision 4000 X-Series requires minimum support from a CPU, as most core
operations are handled by Keysight proprietary technology, the MegaZoom IV smart
memory ASIC. MegaZoom includes hardware serial decoders and hardware mask/
limit testing capability, plots analog and digital data directly to the display, supports
GUI operation, and integrates additional instruments like the dual-channel WaveGen
function/arbitrary waveform generator.
Figure 6. The 4000 X-Series oscilloscopes’ uncompromised responsiveness, speed and waveform update rate is enabled by the MegaZoom IV, smart
memory ASIC. The CPU is not used for core waveform operations.
Oscilloscope Experience Redefined:
Experience the Speed (Continued)
07 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Segmented memory: A smart and efficient way to
capture waveforms
Acquisition memory size is an essential oscilloscope specification
because it determines the amount of data you can capture in a
single acquisition. In general, longer memory is better. However,
no memory is always long enough to capture all the signals you
need, especially when capturing infrequent anomalies, data
bursts, or multiple serial bus packets. Segmented memory
acquisition lets you selectively capture and store important signal
activity without capturing unimportant signal idle time with the
time stamp of each segment relative to the first trigger event.
Segmented memory comes standard in the 4000 X-Series.
Figure 7 shows segmented memory successfully capturing 1,000
events in 3.27274 seconds. Traditional memory architecture
would require 2.7 Gpts of memory to accomplish the same result.
This memory is not available on any scope in the market.
Figure 7. Segmented memory efficiently manages the memory to capture
up to 1,000 segments of interest to you, making it an effective ultra-deep
memory oscilloscope that can easily capture infrequent events and
anomalies.
Segmented memory + serial decode
Segmented memory works in conjunction with serial protocol
decode. For example, by setting the trigger condition to “CAN
serial bus error,” segmented memory captures and stores only
CAN error packets and stitches together each segment for easy
viewing. You can quickly compare time tags in the event lister to
discover time intervals between errors.
Figure 8. Segmented memory being used in conjunction with serial
decode resulting in maximum insight into serial bus.
Oscilloscope Experience Redefined:
Experience the Speed (Continued)
08 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Mask/limit testing (option)
Whether you are performing pass/fail tests to specified
standards in manufacturing or testing for infrequent signal
anomalies, mask/limit testing can be a valuable productivity tool
(DSOX4MASK). The 4000 X-Series features powerful hardware-
based mask testing and can perform up to 270,000 tests per
second. You can select multiple test criteria, including the ability
to run tests for a specific number of acquisitions, a specified time,
or until detection of a failure.
Figure 9. Mask testing evaluated > 22 M waveforms in just 2 minutes.
Figure 10a. The 4000 X-Series was set up to capture data signals with
various rise time edges. Using the search and navigation capability, the
oscilloscope was able to find, mark (white triangles), and quickly navigate
to 16 occurrences of “out of compliance” rise-time edges.
Figure 11. Using the error condition search, the 4000 X-Series quickly
found 5 places with a missing acknowledgment in an I²C serial bus. The
navigation feature moves between the errors and zooms automatically to
show the error packet.
Search and navigation
The parametric and serial bus search and navigation feature
comes standard on the 4000 X-Series oscilloscopes. When you
are capturing long, complex waveforms using an oscilloscope’s
deep acquisition memory, manually scrolling through stored
waveform data to find specific events of interest can be slow and
cumbersome. With automatic search and navigation capability,
you can easily set up specific search criteria and then quickly
navigate to “found and marked” events. Available search criteria
include edges, pulse width (time-qualified), rise/fall times
(time-qualified), runt pulses (time-and level-qualified), frequency
peaks (FFT function, threshold and excursion qualified), and serial
bus frames, packets, and errors.
Oscilloscope Experience Redefined:
Experience the Speed (Continued)
Figure 10b. The 4000 X-Series was set up to capture clock signals for
FFT analysis. Using the search and navigation capability, the scope
found, marked (white triangles) and quickly navigated to the first 11
frequency peaks occurrences. You can sort it in the order of frequency or
amplitude.
9 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Figure 14. Use the Keysight pull-down menu for Windows-like operation.
Figure 13. See 10 measurements, cursor information, and the DVM
simultaneously by dragging the desired docking panel to any open area.
Industry’s largest 12.1-inch display
From the start of product development, we designed every aspect
of this oscilloscope for a touch interface. Large, easily touchable
targets on the industry’s largest 12.1-inch display with capacitive
touch screen technology mean operation is quick and natural, just
like your favorite tablet devices.
Capacitive touch screen technology
Capacitive touch screen technology provides enhanced
productivity. Use the alphanumeric pad for quick annotation,
place waveforms or cursors in exact positions and drag docking
panels across the screen to see more measurement information.
The 4000 X-Series offers three ways to access key menus
and features: touch GUI for those that prefer tablet or smart
phone touch interfaces, front panel keypads for the traditional
oscilloscope users, and Keysight pull down menu for users who
prefer Windows-like operations. The 4000 X-Series also offers a
touch off” button as well as USB mouse and keyboard support.
Redefine your remote Web control oscilloscope experience. The
4000 X-Series not only supports traditional control via a PC Web
browser, but also supports remote control through popular tablet
devices.
Figure 12. The industry’s largest 12.1-inch display and capacitive touch
screen technology with large, touchable targets.
Oscilloscope Experience Redefined:
Experience the Usability
10 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Zone touch trigger
One of the biggest challenges of using an oscilloscope is setting
up an advanced trigger to isolate a signal of interest. While
advanced triggers are powerful features, zone touch trigger
provides a turnkey trigger solution.
You simply observe the signal of interest on the display, and
draw a zone (box) around it. What used to be hours of work can
now take just a few seconds. If you want to move your zones to
another location, just drag them over. The 4000 X-Series can be
set up to easily trigger on one or two zone boxes simultaneously
with either must intersect or must not intersect conditions.
Zone triggering does not compromise the waveform update
rate; the 4000 X-Series will still maintain an ultra-fast 200,000
waveforms per second or more, even with additional features
turned on. In other words, the oscilloscope that sees everything
can easily trigger on anything.
Zone touch trigger + segment memory: A whole
new experience
The combination of the industry’s only hardware-based zone
touch trigger with the 4000 X-Series’ segment memory simplifies
and enhances your debugging experience. In Figure 16, the 4000
X-Series has isolated and captured 1,000 metastable signals,
showing the critical bit errors over a 32-second time span at
5 GS/s sampling rate in the segment memory. The segment
memory also allows you to overlay all segments to identify the
worst-case signal.
Figure 15. When you see anomalies, all you have to do is
draw a zone box to trigger on them.
Figure 16. Combination of the zone touch trigger and
segmented memory.
Oscilloscope Experience Redefined:
Experience the Usability (Continued)
11 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Investment protection through a fully-upgradable 7-in-1 instrument
The InfiniiVision 4000 X-Series redefines the
oscilloscope experience with unprecedented
integration. This 7-in-1 instrument provides:
Oscilloscope
16 digital channels
Serial protocol analyzer
Dual-channel WaveGen 20 MHz function/
arbitrary waveform generator
3-digit voltmeter
Frequency response analysis
8-digit hardware counter with totalizer
Figure 17. The 4000 X-Series provides the capabilities of five instruments
seamlessly integrated into one.
Serial protocol
analysis
Oscilloscope
(analog)
channels
Logic channels
Dual-channel
WaveGen
Digital
voltmeter
Multi-domain analysis: Time-correlate analog,
digital, and frequency domain signals
Viewing the frequency content of waveforms is greatly simplified
by a touch screen operation. Pop up keypads make inputting
start, stop, span and center frequency easy. And the new problem
solving feature called “gated FFT” lets you time correlate the
analog, digital, and frequency domain to aid in analysis and
debug. In addition, there are new capabilities for peak searching,
max and min hold and averaging of FFTs to increase dynamic
range.
When gated FFT is on, the oscilloscope goes into zoom mode.
The FFT analysis shown in the zoomed (bottom) window is taken
from the period of time indicated by the zoom box in the main
(top) window. In the gated FFT mode, touch and flick the zoom
box through the acquisition to investigate how the FFT analysis
changes over time, correlating the RF phenomenon with the
analog and digital phenomenon.
Oscilloscope Experience Redefined:
Experience the Integration
Figure 18. Gated FFT successfully correlated the hopping of the FSK
modulation with the analog and I2C control command.
12 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Figure 21. USB 2.0 trigger, decode and “lister” display.
Figure 20. Dual serial bus CAN and LIN decode and interleaved “lister
display.
Figure 19. Digital channels are captured and displayed time-correlated
with analog channels in MSOs or upgraded DSOs.
Mixed signal oscilloscope (MSO): Integrated
16 digital channels
With an additional 16 integrated digital channels, you now have
up to 20 channels of time-correlated triggering, acquisition and
viewing on the same instrument. This is especially important
in today’s embedded designs with sophisticated digital control
circuitry. Unlike other oscilloscopes in this class, you can buy a
2- or 4-channel DSO and enable the 16 digital channels already in
the instrument at any time to make it an MSO. (DSOXPERFMSO)
Serial protocol analysis: Hardware-based serial
protocol decode and triggering
Keysight InfiniiVision Series, including the new 4000 X-Series,
are the only oscilloscopes to use hardware-based serial
protocol decoding. Other vendors’ oscilloscopes use software
post-processing techniques to decode serial packets ⁄ frames,
and therefore have slow waveform and decode capture rates and
could miss critical events and errors due to a long dead-time.
Faster decoding with hardware-based technology enhances the
probability of capturing infrequent serial communication errors.
After capturing serial bus communication, you can easily
perform a search operation based on specific criteria and then
quickly navigate to bytes/frames of serial data that satisfy that
search criteria. The 4000 X-Series can decode two serial buses
simultaneously using hardware-based decoding, and display the
captured data in a time interleaved “lister”display.
Serial protocol decoding can be used simultaneously with
segmented memory and zone touch triggering.
The 4000 X-Series has: SENT, I2C, SPI, USB 2.0, RS232/UART,
CAN, CAN FD, LIN, FlexRay, CXPI, MIL-STD 1553, ARINC 429, I2S,
user-definable Manchester, user-definable NRZ, and USB PD.
(See page 23)
Oscilloscope Experience Redefined:
Experience the Integration (Continued)
13 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Figure 22. WaveGen sine wave output with and without added AM
modulation.
Figure 24. Dual channel WaveGen output of differential arbitrary signals.
Common mode is shown as a math function.
Dual-channel WaveGen 20-MHz function/arbitrary
waveform generator: Industry-exclusive
The 4000 X-Series offers the industrys only dual-channel,
integrated 20-MHz function/arbitrary waveform generator.
(DSOX4WAVEGEN2) The integrated generator provides stimulus
output of sine, square, ramp, pulse, DC, noise, sine cardinal
(sinc), exponential rise, exponential fall, cardiac, Gaussian pulse
and arbitrary waveforms (AWG) to your device under test. Signal
modulation capability is also available.
With AWG functionality, you can store waveforms from analog
channels or reference memory to the arbitrary memory and
output from WaveGen. Easily create and edit the waveform using
the built-in editor or Keysights free BenchLink Waveform Builder
Basic software: www.keysight.com/find/33503
With dual channels, you can generate differential signals to:
output arbitrary clock and data signals to simulate serial buses,
create complex modulations (more than the standard modulation
feature), output IQ signals and more. The two channels can be
tracked together as well (identical frequency, amplitude, offset
and duty cycle).
3-digit voltmeter
The 4000 X-Series offers a standard integrated 3-digit voltmeter
(DVM) and 5-digit frequency counter inside the oscilloscope. The
voltmeter operates through the same probes as the oscilloscope
channels. However, the DVM measurements are de-coupled from
the oscilloscope triggering system so that both the DVM and
triggered oscilloscope waveform capture can be made with the
same connection. The voltmeter results are always displayed,
keeping these quick characterization measurements at your
fingertips.
Figure 23. WaveGen arbitrary waveform editing screen.
Figure 25. DVM 3-digit
voltage and 5-digit frequency
measurements always at your
fingertips.
Oscilloscope Experience Redefined:
Experience the Integration (Continued)
14 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Figure 26. Power quality measurement, one of many in the power
measurements application.
Figure 27. Triggering on 1080p HDTV signal analysis.
Figure 28. Perform automatic signal quality testing on USB 2.0 low-
speed, full-speed, and hi-speed signals.
USB 2.0 signal quality analysis
With the USB 2.0 signal quality test option (DSOX4USBSQ),
designers of systems with USB interfaces can now perform
automated signal quality testing. This option supports low-speed,
full-speed, and hi-speed applications (hi-speed tests require
1.5 GHz models). The USB 2.0 signal quality test with HTML pass/
fail report generation includes eye-diagram mask testing, jitter
analysis, EOP bit-width, signaling rate, edge monotonicity, and
rise/fall times; all based on official USB-IF algorithms embedded
in the oscilloscope.
See www.keysight.com/find/DSOX4USBSQ for more information.
Oscilloscope Experience Redefined:
Other Key Productivity Tools
Power measurements and analysis
When you are working with switching power supplies and power
devices, the power measurements application (DSOX4PWR)
provides a full suite of power measurements and analysis in the
oscilloscope.
Included with the DSOX4PWR is a license for the U1881A
PC-based power analysis software package, which provides
additional offline measurements and report generation.
See www.keysight.com/find/DSOX4PWR for more information.
HDTV video triggering and analysis
Whether you are debugging consumer electronics with HDTV
or characterizing a design, the HDTV measurement application
(DSOX4VID) provides support for a variety of HDTV standards for
triggering and analysis.
See www.keysight.com/find/DSOX4VID for more information.
15 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Figure 30. The Educators Training Kit helps both students and
experienced users quickly get up to speed on oscilloscope usage.
Educators oscillocope training kit
Teach your students what an oscilloscope is and how to perform
basic measurements with the Educator’s Oscilloscope Training
Kit. This complimentary kit includes training tools created
specifically for electrical engineering and physics undergraduate
students and professors. It contains an array of built-in
training signals, a comprehensive oscilloscope lab guide and
tutorial written specifically for undergraduate students and an
oscilloscope fundamentals PowerPoint® slide set for professors
and lab assistants. Also available is an advanced triggering guide
to help even the most experienced oscilloscope users to get the
most out of their 4000 X-Series oscilloscope.
See www.keysight.com/find/dsoxedk for more information.
Oscilloscope Experience Redefined:
Other Key Productivity Tools (Continued)
Frequency Response Analysis (FRA) option
Frequency Response Analysis (FRA) is an often-critical measurement
used to characterize the frequency response (gain and phase versus
frequency) of a variety of today’s electronic designs, including passive
filters, amplifier circuits, and negative feedback networks of switch mode
power supplies (loop response). InfiniiVision 4000 X-Series oscilloscopes
licensed with the DSOX4FRA option use the oscilloscope’s built-in
waveform generator (WaveGen) to stimulate the circuit under test at
various frequency settings and capture the input and output signals
using two oscilloscope channels. At each test frequency, the oscilloscope
measures, computes, and plots gain (20LogVout/Vin) and phase
logarithmically.
16 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Advanced math analysis provides a variety of additional math
functions and comes standard on the 4000 X-Series. Additionally,
math functions can be nested to provide additional insight into
your designs. You can create up to four math functions, with one
resultant math function displayed at a time.
Operators
Add, subtract, multiply, divide
Transforms
Differentiate, integrate
FFT
Ax + B
Squared, square root
Absolute value
Common logarithm, natural logarithm
Exponential, base 10 exponential
Filters
Low-pass filter, high-pass filter
Averaged value, smoothing, envelope
Visualizations
Magnify
Max hold, min hold
Measurement trend
Chart logic bus timing, chart logic bus state
Figure 32. Four math functions can be created and nested with one
resultant math function.
Figure 31. A variety of advanced math functions are standard in the 4000
X-Series.
36 automatic measurements
Automatic measurements are the essential tool of an
oscilloscope. In order to make quick and efficient measurements,
the 4000 X-Series provides 36 powerful automatic measurements
and can display up to 10 at a time. Measurements can be gated
by auto select, main window, zoom window, or cursors.
Figure 33. Up to 10 automated measurements displayed simultaneously.
Measurements can be gated by cursors.
Oscilloscope Experience Redefined:
Other Key Productivity Tools (Continued)
17 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Figure 34. Store and recall up to four reference waveforms.
Figure 35. N7020A Power Rail Probe is the industry’s only probe
designed and developed to solve your toughest power integrity problems.
Powerful probe solutions and compatibility
Get the most out of your 4000 X-Series scope, by using
Keysight’s complete family of innovative probes and accessories
for your application. The 4000 X-Series supports up to four active
probes simultaneously with its full AutoProbe interface. 1
All 4000 X-Series scopes come standard with a 700 MHz
bandwidth, 10 MΩ input passive probe per each channel and
gives you 700 MHz system bandwidth when used in conjunction
with the 4000 X-Series 1 GHz/1.5 GHz models.Also available
is the N2750A InfiniiMode differential probe and N2795A/96A
single-ended active probe for high signal fidelity measurements
without the high price. For ultra low current measurements,
the N2820A Series high-sensitivity current probes are the
best solution in the industry. For power rail measurement, the
N7020A Power Rail Probe provides the unmatched measurement
accuracy.
For the most up-to-date and complete information about
Keysight’s probes and accessories, visit our Web site at
www.keysight.com/find/scope_probes or refer to the InfiniiVision
Probes and Accessories data sheet with the Keysight literature
number 5968-8153EN.
1� Some restriction may apply. Contact Keysight for more details.
Reference waveforms
Store up to four waveforms in the scope’s non-volatile reference
waveform memory. Compare reference waveforms with live
waveforms, and perform post analysis and measurements on
stored data. You can also store waveforms on a removable
USB memory device in *.h5 format and recall them back into
oscilloscope’s reference waveform memory later. Save and/
or transfer waveforms to a PC as XY data pairs in a comma-
separated values format (*.csv) or store bitmap images and
transfer them to a PC for documentation purposes in a variety of
image formats.
Oscilloscope Experience Redefined:
Other Key Productivity Tools (Continued)
18 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Connectivity and LXI compatibility
Standard USB 2.0 hi-speed host (two on front, one on back) and
device (one on back) ports make PC connectivity easy. Operate
the scope from your PC and save/recall stored waveforms and
setup files via standard LAN (LXI IPv6 Extended Function).
Connect your projector or external monitor through VGA output,
standard with the 4000 X-Series, when sharing and presenting
screen information. An optional external GPIB-to-LAN adapter is
also available (N4865A).
The BV0004B oscilloscope control and automation application
within BenchVue lets you control and visualize the 4000 X-Series
and multiple measurements simultaneously. It lets you build
automated test sequences just as easily as you can with the
front panel. Save time with the ability to export measurement
data to Excel, Word and MATLAB in three clicks. Monitor and
control your 4000 X-Series with a mobile device from anywhere.
Simplify your testing with BenchVue software. Learn more at
www.keysight.com/find/BenchVue
Figure 37. BV0004B BenchVue.
Figure 36. The 4000 X-Series and N2820A Series high-sensitivity current
probe measuring > 500 mA and < 1 mA current simultaneously.
Localized front panel, GUI and help
Operate the oscilloscope in the language most familiar to you.
The graphical user interface, built-in help system, front panel
overlays, and user’s manual are available in 11 languages. During
operation, access the built-in help system just by pressing and
holding any button.
Oscilloscope Experience Redefined:
Other Key Productivity Tools (Continued)
19 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Figure 38. Tablet virtual front panel control.
Virtual front panel
The 4000 X-Series’ innovative capacitive touch screen matches
perfectly with the latest tablet technologies. In addition to the
traditional virtual front panel remote operation through your
favorite PC Web browser, the 4000 X-Series supports remote
oscilloscope control from your tablet devices (and smart phones
with enough resolution). The tablet virtual front panel is identical
to the 4000 X-Series’ touch GUI so you can touch icons, draw
zone touch trigger zones and drag slide panels as if you are
sitting in front of the actual oscilloscope.
Documentation and e-mail
Annotation becomes a simple task. Bring up the annotation menu
and start editing it using the keypad, and then drag it to the
desired location.
Figure 39(a). Annotation and keypad. Figure 39(b). E-mail configuration screen.
Oscilloscope Experience Redefined:
Other Key Productivity Tools (Continued)
Quick e-mail allows you to e-mail the data you want instantly
to your inbox. Send out the screenshot, waveform data, or even
a USB signal quality test report. This removes the hassle of
connecting your PC to your oscilloscope.
20 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Secure erase
The secure erase feature comes standard with all 4000 X-Series
models. At the press of a button, internal non-volatile memory
is clear of all setup, reference waveforms, and user preferences,
ensuring the highest level of security in compliance with National
Industrial Security Program Operation Manual (NISPOM) Chapter
8 requirements.
Figure 40. Infiniium Offline enables a variety of advanced signal analysis
while providing extensive, yet intuitive, waveform documentation.
Infinium Offline oscilloscope analysis software
Keysight’s Infiniium Offline PC-based oscilloscope analysis
software (N8900A) allows you to do additional signal viewing,
analysis and documentation tasks away from your oscilloscope.
Capture waveforms, save to a file, and recall the waveforms into
Infiniium Offline. The application supports a variety of popular
waveform formats from multiple oscilloscope vendors and
includes the following features: navigate, view, measurements,
analyze, view windows, documentation, and optional analysis
upgrades.
For more information, go to: www.keysight.com/find/N8900A
Oscilloscope Experience Redefined:
Other Key Productivity Tools (Continued)
21 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
High-resolution mode for viewing signal details
To build more confidence in your designs, sometimes you need to
look into more signal detail than you can see with the standard
8-bit vertical resolution of the 4000 X-Series.
High-resolution mode offers additional resolution and insight into
the signal, without requiring a repetitive signal. Using real-time
boxcar averaging, high-resolution mode reduces random noise
and effectively increases vertical resolution, up to 12 bits. For
example, it achieves the 113 µVrms noise floor at 1 mV/div,
100 µs/div setting.
Figure 42. Getting 113 uVrms noise floor at 1 mV/div with the high
resolution mode.
Freeze display
Perhaps you need to share with others an infrequent event you
found. With the “freeze display” feature, you can keep intensity
information on the screen while the oscilloscope is stopped or
before saving a screen shot.
Figure 44. The “freeze screen” feature keeps the intensity-grading
information while stopping the waveform acquisition.
Figure 43. Wide array of advanced parametric trigger modes.
Advanced parametric triggering
With today’s more complex signals, you often need to trigger
on complex signal conditions to synchronize the oscilloscope’s
acquisition on specific events. The 4000 X-Series oscilloscope
can trigger on the following conditions: edge, edge then edge,
pulse width (time-qualified), pattern, or, rise/fall time, Nth edge
burst, runt, setup and hold, video, and various serial buses
(optional).
Oscilloscope Experience Redefined:
Other Key Productivity Tools (Continued)
22 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
“Designed for touch.” Industry’s first and
largest 12.1 inch capacitive touch screen
to redefine your oscilloscope experience.
The way an oscilloscope was meant to
be driven, with a designed-for-touch
interface.
The class leading 1.5 GHz upgradeable
bandwidth expands your application
coverage, including USB 2.0 hi-speed
signal integrity testing.
Both USB keyboard and mouse are
supported for additional ease of use.
The new zone touch trigger, if you can see
it, you can trigger on it by just drawing a
box.
5-in-1 instruments redefines the
integration experiences: oscilloscope
channels, digital channels, serial protocol
analysis, dual-channel WaveGen, and
DVM. All features are fully upgradeable,
including bandwidth.
Industry-leading coverage of serial
protocol including USB 2.0 trigger and
decode
Industry’s first dual-channel WaveGen
function/arbitrary generator now allows
you to generate differential, clock and
data, two channel modulation, and IQ
signals. Modulation of any signal is also
included.
Oscilloscope Experience Redefined
23 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Industry-leading 1 million waveform per
second update rate minimizes the dead-time
for maximum probability of capturing infrequent
events and anomalies.
Docking panels with the capacitive touch screen add
a new dimension of usability. See setup summary,
automatic measurements, cursor info, DVM, and
navigation pane in any combination, anywhere on the
screen.
Standard advanced math and four cascade-able
math functions enable even the most sophisticated
signal analysis
Display up to 10 measurements simultaneously,
without compromising other key info. 35 automatic
measurements can be gated by cursors
Not a touch screen fan? Turn off the touch screen
from a front panel button if desired.
Independent knobs per channel for fast operation.
All front panel knobs are push-able for access to
common controls.
Standard segmented memory powered by
MegaZoom IV smart memory technology provides
intelligent capture of just the signal of interest.
Four AutoProbe (active or current probes)
are supported simultaneously for demanding
applications.
Simultaneous 1GHz bandwidth
across all 4 channels.
Industry’s first integrated DVM
Asynchronous from the 4 analog
triggered waveforms.
24 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Oscilloscope Experience Redefined:
Configuring Your InfiniiVision 4000 X-Series Oscilloscope
Step 1. Choose your bandwidth and number of channels
InfiniiVision 4000 X-Series scopes oscilloscopes
4022A 4024A 4032A 4034A 4052A 4054A 4104A 4154A
Bandwidth 1 (–3 dB) 200 MHz 350 MHz 500 MHz 1 GHz 1.5 GHz
Calculated rise time (10 to 90%) ≤ 1.75 ns ≤ 1 ns ≤ 700 ps ≤ 450 ps ≤ 300 ps
Input channels DSOX 2 4 2 4 2 4 4 4
MSOX 2 + 16 4 + 16 2 + 16 4 + 16 2 + 16 4 + 16 4 + 16 4 + 16
1� For example, if you chose 1 GHz, 4+16 channels, the model number will be MSOX4104A.
Step 2. Tailor your oscilloscope with integrated capabilities and measurement applications to save
time and money. After purchase upgrade model numbers are listed below (values in parentheses are
factory-installed option numbers)
Description Model number
Enables all software options listed below DSOX4APPBNDL
Serial protocols
Embedded serial triggering and decode (I²C, SPI) DSOX4EMBD(-EMB)
Computer serial triggering and decode (RS232/UART) DSOX4COMP(-CMP)
USB 2.0 Full/Low Speed serial triggering and decode DSOX4USBFL(-USF)
USB 2.0 Hi-Speed serial triggering and decode DSOX4USBH(-U2H) 1
Automotive serial triggering and decode CAN/CAN-FD/CAN-dbc/LIN/LIN symbolic DSOX4AUTO(-AMS)
FlexRay serial triggering and decode DSOX4FLEX(-FLX)
Single edge nibble transmission triggering and decode (SENT) DSOX4SENSOR(-SEN)
CXPI serial triggering and decode DSOX4CXPI(-CXP)
Audio serial triggering and decode (I²S) DSOX4AUDIO(-SND)
Aerospace and defence serial triggering and decode (MIL-STD 1553, ARINC 429) DSOX4AERO(-AER)
User-definable Manchester/NRZ triggering and decode DSOX4NRZ
USB PD serial triggering and decode DSOX4UPD
Measurement applications
Dual-channel WaveGen 20 MHz arbitrary/function generator DSOX4WAVEGEN2(-WAV)
Power analysis application DSOX4PWR(-PWR)
Frequency Response Analysis (FRA) DSOX4FRA
Mask limit testing DSOX4MASK(-MSK)
Enhanced video/TV application package DSOX4VID(-VID)
USB 2.0 signal quality test option DSOX4USBSQ(-U2Q) 2
NFC trigger and automated test software DSOX4NFC
Productivity tools
Infiniium Offline oscilloscope analysis software N8900A
User-definable Application (UDA) PC-based test automation software N5467B/C
Vector signal analyzer software (version 16 and higher) 89601B
BenchVue Software BV0004B
1� DSOX4USBH is only available for 1 GHz and 1.5 GHz models.
2. USB 2.0 hi-speed tests require 1.5 GHz models.
See page 35 for more detailed upgradability and installation process information.
25 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Step 3. Choose your probes – For a complete list of compatible probes, visit
www.keysight.com/find/scope_probes
Probes 4000 X-Series
N2894A passive probe 700 MHz, 10:1, 10 MΩ Included standard. 1 per channel
N6450-60001 or N2756A 16 digital channel MSO cable Included on MSOX models and DSOXPERFMSO
10076B high-voltage passive probe 250 MHz 4 kV Optional
N2795A active single-ended probe 1-GHz 1-pF 1-MΩ with AutoProbe Optional
N2796A active single-ended probe 2-GHz 1-pF 1-MΩ with AutoProbe Optional
N2750A InfiniiMode differential probe 1.5-GHz 700-fF 200-kΩ with AutoProbe Optional
N2797A extreme temperature active probe 1.5-GHz 1-pF 1-MΩ with AutoProbe Optional
N2790A differential active probe 100 MHz, ± 1.4 kV with AutoProbe Optional
N2791A differential active probe 25 MHz, ± 700 V Optional
N2792A differential active probe 200 MHz, ± 20 V Optional
N2793A differential active probe 800 MHz, ± 15 V Optional
1147B AC/DC current probe 50 MHz 15 A with AutoProbe Optional
N2893A AC/DC current probe 100 MHz 15 A with AutoProbe Optional
N2820A 2-channel high-sensitivity current probe 50 uA to 5 A Optional
N7020A power rail probe 2-GHz, 1:1, 50 k, ± 24 V offset range Optional
N2805A high voltage differential probe, 200 MHz, ± 100 V (DC + peak AC), 50:1, 4-MΩ, 4 pF Optional
N2804A high voltage differential probe, 300 MHz, ± 300 V (DC + peak AC), 100:1, 4-MΩ, 4 pF Optional
N7040A 23 MHz, 3 kA, AC current probe Optional
N7041A 30 MHz, 600 A, AC current probe Optional
N7042A 30 MHz, 300 A, AC current probe Optional
N7026A 150 MHz, 40 Apk, AC/DC high-sensitivity current probe with AutoProbe Optional
Step 4. Choose your accessories
Recommended accessories 4000 X-Series
GPIB-to-LAN external adapter N4865A
Rack mount kit N2763A
Soft carrying case N2733B
Hard copy manual N6455A
Oscilloscope Experience Redefined:
Configuring Your InfiniiVision 4000 X-Series Oscilloscope (Continued)
Step 5. Calibration plans
Calibration
D/MSOX4000-A6J ANSI Z540-1-1994 calibration Optional
D/MSOX4000-AMG Calibration + Uncertainties + Guardbanding (Accredited) Optional
26 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
DSO and MSO 4000 X-Series oscilloscopes
4000 X-Series specification overview
4022A 4024A 4032A 4034A 4052A 4054A 4104A 4154A
Bandwidth 1 (–3 dB) 200 MHz 350 MHz 500 MHz 1 GHz 1.5 GHz 3
All-channel real-time bandwidth 200 MHz 350 MHz 500 MHz 1 GHz 1 GHz
Calculated rise time (10 to 90%) ≤ 1.75 ns ≤ 1 ns ≤ 700 ps ≤ 450 ps ≤ 300 ps
Input channels DSOX 2 4 2 4 2 4 4 4
MSOX 2 + 16 4 + 16 2 + 16 4 + 16 2 + 16 4 + 16 4 + 16 4 + 16
Maximum sample rate 5 GSa/s half channel, 2.5 GSa/s all channel
Maximum memory depth Standard 4 Mpts, standard segmented memory
Display size and type 12.1-inch high-definition capacitive touch display
Waveform update rate > 1 million waveforms per second
System bandwidth with N2894A standard
passive probe
200 MHz 200 MHz 350 MHz 350 MHz 500 MHz 500 MHz 700 MHz 700 MHz
System analog channels
Hardware bandwidth limits Approximately 20 MHz (selectable)
Input coupling AC, DC
Input impedance Selectable: 1 MΩ ± 1% (16 pF), 50 Ω ± 1.5%
Input sensitivity range 200 MHz ~ 500 MHz models: 1 mV/div to 5 V/div 2 (1 MΩ and 50 Ω)
1 and 1.5 GHz models: 1 mV/div to 5 V/div 2 (1 MΩ), 1 mV/div to 1 V/div (50 Ω)
Vertical resolution 8 bits (measurement resolution is 12 bits with averaging)
Maximum input voltage 1 MΩ 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk. Frequency derating (assumes sine
wave input): 400 Vpk until 40 kHz. Then derates at 20 db/dec until 6 Vpk
50 Ω 50 Ω: ≤ 5 Vrms max
DC vertical gain accuracy 1± 2.0% full scale 2
DC vertical offset accuracy ± 0.1 div ± 2 mV ± 1% of offset setting
Channel-to-channel isolation 200 MHz~1 GHz ≥ 40 dB from DC to maximum specified bandwidth of each model
1.5 GHz ≥ 40 dB from DC to 1 GHz, ≥ 35 dB from 1 to 1.5 GHz
Offset range ± 5 V (< 10 mV/div), ± 20 V (10 to 200 mV/div), ± 75 V (> 200 mV/div)
Vertical system digital channels
Digital input channels 16 digital (D0 to D15. Pod 1: D7 ~ D0, Pod 2: D15 ~ D8)
Thresholds Threshold per pod
Threshold selections TTL (+1.4 V), 5 V CMOS (+2.5 V), ECL (–1.3 V), user-defined (selectable by pod)
User-defined threshold range ± 8.0 V in 10 mV steps
Maximum input voltage ± 40 V peak CAT I
Threshold accuracy 1± (100 mV + 3% of threshold setting)
Maximum input dynamic range ± 10 V about threshold
Minimum voltage swing 500 mVpp
Input impedance 100 kΩ ± 2% at probe tip
Input capacitance ~8 pF
Vertical resolution 1 bit
1� Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration
temperature.
2. 1 mV/div and 2 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 1 mV/div and 2 mV/div sensitivity
setting�
3� 1.5 GHz real time bandwidth in half-channel mode or full channel equivalent time mode.
Oscilloscope Experience Redefined:
InfiniiVision 4000 X-Series Performance Characteristics
27 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Horizontal system analog channels
4022A 4024A 4032A 4034A 4052A 4054A 4104A 4154A
Time base range 2 ns/div to 50 s/div 1 ns/div to 50 s/div 500 ps/div to 50 s/div
Time base accuracy 1± 10 ppm
Time base delay time range Pre-trigger Greater of 1 screen width or 200 μs (400 μs in interleaving mode)
Post-trigger 1 to 500 s
Channel-to-channel deskew range ± 100 ns
Δ Time accuracy (using cursors) ± 0.001% of reading ± 0.16% screen width ± 30pS
Modes Main, zoom, roll, XY
XY On channels 1 and 2 only. Z Blanking on Ext Trigger Input, 1.4 V threshold
Bandwidth: Maximum bandwidth. Phase error at 1 MHz: < 0.5 degree
Time base: 200 ns/div to 50 ms/div
Horizontal system digital channels
Minimum detectable pulse width 2 ns
Channel-to-channel skew 2 ns (typical); 3 ns (maximum)
Acquisition system
4022A 4024A 4032A 4034A 4052A 4054A 4104A 4154A
Maximum analog channels sample rate 5 GSa/s half channel interleaved, 2.5 GSa/s all channels
Analog channels equivalent sample rate N/A 128 Gsa/s
Maximum analog channels record length 4 Mpts half channel interleaved, 2 Mpts all channel
Maximum digital channels sample rate 1.25 GSa/s
Maximum digital channels record length 2 Mpts (with digital channels only)
Modes Normal Default mode
Peak detect Capture glitches as narrow as 200 ps at all time base settings
Averaging Selectable from 2, 4, 8, 16, 64, ... to 65,536
High
resolution
Real-time boxcar averaging reduces random noise and effectively increases vertical resolution
12 bits: ≥ 50 μs/div
11 bits: ≥ 20 μs/div
10 bits: ≥ 10 us/div
9 bits: ≥ 5 us/div
Segmented Segmented memory optimizes available memory for data streams that have long dead times
between activity. Maximum segments = 1000. Re-arm time = 1 μs (minimum time between trigger
events). Re-arm time when used with the zone touch trigger = 65 μs or faster (typical)
Roll Displays the waveform moving across the screen from right to left. Available at the time base
50 ms/div or slower
Equivalent
time
1 GHz and 1.5 GHz models only. 7.8 ps fine interpolator resolution yields a maximum effective
sample rate of 128 GSa/s
1� Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration
temperature.
Oscilloscope Experience Redefined:
InfiniiVision 4000 X-Series Performance Characteristics (Continued)
28 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Trigger system
Trigger sources Analog channel (1 ~ 4), digital channel (D0 ~ D15), line, external, WaveGen (1, 2, or Mod)
(FM/FSK)
Trigger modes Normal Requires trigger event for oscilloscope to trigger
Auto Triggers automatically in absence of trigger event
Single Front panel button that triggers only once on a trigger event. Press [Single] button again
for oscilloscope to find another trigger event, or press [Run] front-panel button to trigger
continuously in either auto or normal mode
Force Front panel button that forces a trigger
Trigger coupling DC DC coupled trigger
AC AC coupled trigger, cutoff frequency: < 10 Hz (internal); < 50 Hz (external)
HF reject High-frequency reject, cutoff frequency ~ 50 kHz
LF reject Low-frequency reject, cutoff frequency ~ 50 kHz
Noise reject Adds hysteresis to the trigger circuitry. Selectable OFF or ON, decreases sensitivity 2x
Trigger holdoff range 40 ns to 10.00 s
Trigger sensitivity (internal) 1200 MHz ~ 1 GHz < 10 mV/div: greater of 1 div or 5 mV; ≥ 10 mV/div: 0.6 div
1.5 GHz DC to 1 GHz: < 10 mV/div: Greater of 1 div or 5 mV; ≥ 10 mV/div: 0.6 div
1 to 1.5 GHz: < 10 mV/div: Greater of 1.5 div or 5 mV; ≥ 10 mV/div: 1.0 div
Trigger sensitivity (external) 1± 1.6 V 40 mVpp DC to 100 MHz, 70 mVpp 100 to 200 MHz
± 8 V 200 mVpp DC to 100 MHz, 350 mVpp 100 to 200 MHz
Trigger level range Any channel ± 6 div from center screen
External 8 V range = ± 8 V, 1.6 V range = ± 1.6 V
1� Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration
temperature.
Oscilloscope Experience Redefined:
InfiniiVision 4000 X-Series Performance Characteristics (Continued)
29 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Trigger type selections
InfiniiScan Zone (HW zone qualifier) Trigger on user-defined zones drawn on the display. Applies to one analog channel at a time. Specify zones as
either “must intersect” or “must not intersect.” Up to two zones. > 200,000 wfm/sec update rate.
Supported modes: normal, peak detect, high resolution.
Also works simultaneously with the serial decodes and mask/limit test.
Edge Trigger on a rising, falling, alternating, or either edge of analog channels, digital channels, or an external signal.
Trigger on a rising or falling of a line signal.
Edge then edge (B trigger) Arm on a selected edge, wait a specified time, then trigger on a specified count of another selected edge.
Minimum 4 ns
Pulse width Trigger on a pulse on a selected channel, whose time duration is less than a value, greater than a value, or inside a
time range
Minimum duration setting: 2 ns (500 MHz, 1 GHz, 1.5 GHz), 4 ns (350 MHz), 6 ns (200 MHz)
Maximum duration setting: 10 s
Range minimum: 10 ns
Pattern Trigger when a specified pattern of high, low, and don’t-care levels on any combination of analog, digital, or
trigger channels is [entered | exited]. Pattern must have stabilized for a minimum of 2 ns to qualify as a valid
trigger condition.
Minimum duration setting: 2 ns (500 MHz, 1 GHz, 1.5 GHz), 4 ns (350 MHz), 6 ns (200 MHz)
Maximum duration setting: 10 s
Or Trigger on any selected edges from available sources (analog and digital channels only up to 500 MHz)
Rise/fall time Trigger on rise-time or fall-time edge speed violations (< or >) based on user-selectable threshold. Select from
(< or >) and time settings range between
Minimum: 1 ns (500 MHz, 1 GHz, 1.5 GHz model), 2 ns (350 MHz model), 3 ns (200 MHz model)
Maximum: 10 s
Nth edge burst Trigger on the Nth (1 to 65535) edge of a pulse burst. Specify idle time (10 ns to 10 s) for framing
Runt Trigger on a positive runt pulse that fails to exceed a high-level threshold. Trigger on a negative runt pulse that
fails to exceed a low-level threshold. Trigger on either polarity runt pulse based on two threshold settings. Runt
triggering can also be time-qualified (< or >) with a minimum time setting of 2 ~ 6 ns and maximum time setting of
10 s
Minimum time setting: 2 ns (500 MHz, 1 GHz, 1.5 GHz), 4 ns (350 MHz), 6 ns (200 MHz)
Setup and hold Trigger on setup/hold violations. Setup time can be set from 7 to 10 s. Hold time can be set from 0 s to 10 ns.
Minimum window (setup time + hold time) must be 3 ns or greater
Video Trigger on all lines or individual lines, odd/even or all fields from composite video, or broadcast standards (NTSC,
PAL, SECAM, PAM-M)
Enhanced video (HDTV) (Option) Trigger on lines and fields of enhanced and HDTV standards (480p/60, 567p/50, 720p/50, 720p/60, 1080p/24,
1080p/25, 1080p/30, 1080p/50, 1080p/60, 1080i/50, 1080i/60)
ARINC429 (Option) Trigger and decode on ARINC429 data. Trigger on word start/stop, label, label + bits, label range, error conditions
(parity, word, gap, word or gap, all), all bits (eye), all 0 bits, all 1 bits
CAN (Option) Trigger on CAN (controller area network) version 2.0A,2.0B, and CAN-FD (Flexible Data-rate) signals. Trigger on
the start of frame (SOF), the end of frame (EOF), data frame ID, data frame ID and data (non-FD), data frame ID
and data (FD), remote frame ID, remote or data frame ID, error frame, acknowledge error, from error, stuff error,
CRC error, spec error (ack or form or stuff or CRC), all errors, BRS Bit (FD), CRC delimiter bit (FD), ESI bit active
(FD), ESI bit passive (FD), overload frame., message, message and signal (non-FD), message and signal (FD, first
8 bytes only)
FlexRay (Option) Trigger on frame ID or specific error condition, along with cycle-base and repetition-cycle filtering. Can also
trigger on specific events such as BSS, TSS, FES, and wake up
C (Option) Trigger at a start/stop condition or user defined frame with address and/or data values. Also trigger on missing
acknowledge, address with no acq, restart, EEPROM read, and 10-bit write
S (Option) Trigger on 2’s complement data of audio left channel or right channel (=, ≠, <, >, > <, < >, increasing value, or
decreasing value)
LIN (Option) Trigger on LIN (local interconnect network) sync break, sync frame ID, frame ID and data, parity error, or
checksum error
Oscilloscope Experience Redefined:
InfiniiVision 4000 X-Series Performance Characteristics (Continued)
30 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Waveform measurements and cursors
DC vertical accuracy/cursors 2Single cursor accuracy: ± [DC vertical gain accuracy + DC vertical offset accuracy + 0.21% full
scale]
Dual cursor accuracy: + [DC vertical gain accuracy + 0.42% full scale] 1
Cursors 2 pairs of XY cursors
Automatic measurement of positions, ∆X, 1/X, ∆Y, and ∆Y/∆X
Automatic measurements Measurements continuously updated with statistics. Cursors track last selected measurement.
Select up to four measurements from the list below:
Voltage Peak-to-peak, maximum, minimum, amplitude, top, base, overshoot, pre-shoot, average- N
cycles, average- full screen, DC RMS- N cycles, DC RMS- full screen, AC RMS- N cycles, AC
RMS- full screen (standard deviation), ratio (RMS1/RMS2)
Time Period, frequency, counter, + width, - width, burst width, duty cycle, bit rate, rise time, fall time,
delay, phase, X at min Y, X at max Y
Count Positive pulse count, negative pulse count, rising edge count, falling edge count
Mixed Area- N cycles, area- full screen
Automatic measurement logging Available via BenchVue
Counter Built-in frequency counter
Source Any analog or digital channel
Resolution 5 digits. Up to 8 digits with an external reference 10 MHz input
Max frequency Bandwidth of oscilloscope
Mask limit test option Standard mask/limit test capability that provides easy, fast pass/fail comparison of a signal
under test to a predefined mask template or auto-mask template. Predefined mask templates or
edits to an auto-mask template can be made via a simple text editor. > 270,000 mask tests per
second (waveform update rate)
Waveform math
Number of math functions Four (display one at a time)
Arithmetic Add, subtract, multiply, divide, differentiate, integrate, FFT, Ax + B, squared, square root, absolute, common log,
natural log, exponential, base 10 exponential, LP filter, HP filter, averaged value, smoothing, envelope, magnify,
max hold, min hold, measurement trend, chart logic bus (timing or state)
FFT Up to 64 kpts resolution. FFT window types: Hanning, flat top, rectangular, Blackman-Harris, Bartlett
1� Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ± 10 °C from firmware calibration
temperature.
2. 1 mV/div and 2 mV/div is a magnification of 4 mV/div setting. For vertical accuracy calculations, use full scale of 32 mV for 1 mV/div and 2 mV/div sensitivity
setting�
3� Suspend, resume, reset, bit stuff error, and SE1 error are USB 2.0 low and full speed only.
Trigger type selections
CXPI (Option) Trigger on the start of frame (SOF), the end of frame (EOF), PTYPE, frame ID, data and info frame ID, data and info
frame ID (long frame), CRC field error, parity error, inter-byte space error, inter-frame space error, framing error,
data length error, sample error, all errors, sleep frame, wakeup pulse
MIL-S TD155 3 (Opt ion) Trigger on MIL-STD 1553 signals on data word start/stop, command/status start/stop, RTA, RTA + 11 bits, and
error conditions (parity, sync, Manchester)
SPI (Option) Trigger on SPI (serial peripheral interface) data pattern during a specific framing period. Supports positive and
negative chip select framing as well as clock Idle framing and user-specified number of bits per frame. Supports
MOSI and MISO data
UART/RS232/422/485 (Option) Trigger on Rx or Tx start bit, stop bit, data content, or parity error
USB (Option) Trigger on start of packet (SOP), end of packet (EOP), suspend 3, resume 3, reset 3, packets (token, data,
handshake, or special), and errors (PID, CRC5, CRC16, glitch, bit stuff 3, SE1 3). Supports USB 2.0 low speed, full
speed, and hi-speed (hi-speed is supported on 1 GHz and 1.5 GHz models only)
SENT (Option) Trigger on SENT bus. start of fast channel message, start of slow channel message, fast channel SC and data,
slow channel message ID, slow channel message ID and data, tolerance violation, fast channel CRC error, slow
channel CRC error, all CRC errors, pulse period error, successive sync pulses error (1/64)
User-definable Manchester/NRZ
(Option)
Trigger on start-of-frame (SOF), bus value, and Manchester errors
USB PD (Option) Trigger on preamble, EDP, ordered sets, preamble errors, CRC errors, header content (control messages, data
messages, extended messages and value in HEX)
Oscilloscope Experience Redefined:
InfiniiVision 4000 X-Series Performance Characteristics (Continued)
31 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Display characteristics
Display 12.1-inch capacitive touch/gesture enabled color TFT LCD
Resolution 800 (H) x 600 (V) pixel format (screen area)
Graticules 8 vertical divisions by 10 horizontal divisions with intensity controls
Format YT, XY and Roll
Maximum waveform update rate > 1,000,000 wfm/s
Persistence Off, infinite, variable persistence (100 ms to 60 s)
Intensity gradation 64 intensity levels
Connectivity
USB 2.0 hi-speed host port USB 2.0 hi-speed host ports x3, two front and one real panel. Supports memory devices, printers, keyboards and mice
USB 2.0 hi-speed device port
One USB 2.0 hi-speed device port on rear panel. USB Test and Measurement Class (USBTMC) compatible
LAN port 10/100Base-T port on rear panel. LXI IPv6 extended function
Web remote control VNC Web interface (via major Web browsers)
Video out port SVGA out on rear panel. Connect oscilloscope display to an external monitor or projector
GPIB port N4865A GPIB-to-LAN adapter (optional)
10 MHz out/in BNC connector on the rear panel. Supported modes: Off, 10 MHz out, or reference signal mode (10 MHz in)
Trigger out BNC connector on the rear panel. Supported modes: triggers, mask, waveform generator 1 sync pulse, and waveform
generator 2 sync pulse
Dual-channel WaveGen built-in function/arbitary waveform generator (specifications are typical)
WaveGen outputs Two (front-panel BNC connectors)
Both waveform generator outputs can be frequency tracked, amplitude tracked, or completely tracked.* A generator’s
output can be inverted to create a differential signal
Waveforms Sine, square, ramp, pulse, DC, noise, sine cardinal (sinc), exponential rise, exponential fall, cardiac, Gaussian pulse, and
arbitrary
Modulation Modulation is available on channel 1 only. Modulation is not available when tracking mode is enabled.
Modulation types: AM, FM, FSK
Carrier waveforms: Sine, Ramp, Sine Cardinal, Exponential Rise, Exponential Fall, and Cardiac.Modulation Source:
Internal (no external modulation capability)
AM:
Modulation: Sine, Square, Ramp
Modulation frequency: 1 Hz to 20 kHz
Depth: 0% to 100%
FM:
Modulation: Sine, Square, Ramp
Modulation frequency: 1 Hz to 20 kHz
Minimum carrier frequency: 10 Hz
Deviation: 1 Hz to carrier frequency or (2e12/carrier frequency), whichever is smaller
FSK:
Modulation: 50% duty cycle square wave
FSK rate: 1 Hz to 20 kHz
Hop frequency: 2 x FSK rate to 10 MHz
1� Only the following combination of wave shapes can be frequency tracked or completely tracked:
1) Sine, ramp, sine cardinal, cardiac, and Gaussian pulse.
2) Square wave and pulse.
3) Exponential rise and exponetial fall.
4) Arbitrary.
Oscilloscope Experience Redefined:
InfiniiVision 4000 X-Series Performance Characteristics (Continued)
32 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Dual-channel WaveGen built-in function/arbitary waveform generator (specifications are typical)
Sine Frequency range 0.1 Hz to 20 MHz
Amplitude flatness ± 0.5 dB (relative to 1 kHz)
Harmonic distortion 40 dBc
Spurious (non harmonics) 40 dBc
Total harmonic distortion 1%
SNR (50-Ω load, 500-MHz BW) 40 dB (Vpp ≥ 0.1 V); 30 dB (Vpp < 0.1 V)
Square wave/pulse Frequency range 0.1 Hz to 20 MHz
Duty cycle 20 to 80%
Duty cycle resolution Larger of 1% or 10 ns
Rise/fall time 19 ns (10 to 90%)
Overshoot < 2%
Asymmetry (at 50% DC) ± 1% ± 5ns
Jitter (TIE RMS) 500 ps
Ramp/triangle wave Frequency range 0.1 Hz to 200 kHz
Linearity 1%
Variable symmetry 0 to 100%
Symmetry resolution 1%
Pulse Frequency range 0.1 Hz to 10 MHz
Pulse width 20 ns minimum
Pulse with resolution 10 ns
Edge time Fixed at 19 ns (not variable)
Overshoot < 2%
Noise Bandwidth 20 MHz typical
Sine cardinal (Sinc) Frequency range 0.1 Hz to 1.0 MHz
Exponential rise/fall Frequency range 0.1 Hz to 5.0 MHz
Cardiac Frequency range 0.1 Hz to 200.0 kHz
Gaussian pulse Frequency range 0.1 Hz to 5.0 MHz
Arbitrary Waveform length 1 to 8,192 points
Amplitude resolution 10 bits (including sign bit) 2
Repetition rate 0.1 Hz to 12 MHz
Sample rate 100 MSa/s
Filter bandwidth 20 MHz
Frequency Sine wave and ramp accuracy 130 ppm (frequency < 10 kHz)
50 ppm (frequency > 10 kHz)
Square wave and pulse accuracy [50 + frequency/200] ppm (frequency < 25 kHz)
50 ppm (frequency ≥ 25 kHz)
Resolution 0.1 Hz or 4 digits, whichever is larger
Amplitude Range: Minimum 20 mVpp if |offset| ≤ 0.5 Vpp into Hi-Z 1
10 mVpp if |offset| ≤ 0.5 Vpp into 50 Ω 1
Range: Maximum 10 Vpp except, 9 Vpp if Sinc or Cardiac, 7.5 Vpp if Gaussian
pulse into Hi-Z; 5 Vpp/4.5 Vpp into 50 Ω
Resolution 100 μV or 3 digits, whichever is higher
Accuracy 1.5% (frequency = 1 kHz)
1� Sinc, cardiac and Gaussian pulse: ± 1.25 V into Hi-Z; ± 625 mV into 50 Ω.
2. Full resolution is not available at output due to internal attenuator stepping.
Oscilloscope Experience Redefined:
InfiniiVision 4000 X-Series Performance Characteristics (Continued)
33 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Dual channel WaveGen - built-in function/arbitary waveform generator
DC offset Range ± 5 V into Hi-Z, except ± 4 V if sine wave, ± 2.5 V if sinc, cardiac, or
Gaussian pulse into Hi-Z
± 2.5 V into Hi-Z, except ± 2 V if sine wave, ± 1.25 V if sinc, cardiac, or
Gaussian pulse into 50 Ω
Resolution Larger of 250 uV or 3 digits
Accuracy (waveform modes) ± 1.5% of offset setting ± 1% of amplitude ± 1 mV
Accuracy (DC mode) ± 1.5% of offset setting ± 3 mV
Main output Impedance 50 Ω typical
Isolation Not available, main output BNC is grounded
Protection Overload automatically disables output
Trigger output Trigger output available on trig-out BNC
Digital voltmeter (specifications are typical)
Source Analog channels only (1-4)
Functions ACrms, DC, DCrms, frequency
Resolution ACV/DCV: 3 digits
Counter frequency: 5.5 digits
Measuring rate 100 times/second
Autoranging Automatic adjustment of vertical amplification to maximize the dynamic range of measurements
Range meter Graphical display of most recent measurement, plus extrema over the previous 3 seconds
General and environmental characteristics
Power line consumption Maximum 120 W
Power voltage range 100 to 120 V, 50/60/400 Hz; 100 to 240 V, 50/60 Hz ± 10% auto ranging
Temperature Operating 0 to +50 °C
Non-operating –30 to +70 °C
Humidity Operating 50% to 95% RH at 40 °C for 5 days
Non-operating 90% RH at 65 °C for 24 hr
Altitude Operating Up to 3,000 m
Non-operating Up to 15,300 m
Electromagnetic compatibility Meets EMC Directive (2004/108/EC), meets or exceeds IEC 61326-1:2005/EN
61326-1:2006 Group 1 Class A requirement
CISPR 11/EN 55011
IEC 61000-4-2/EN 61000-4-2
IEC 61000-4-3/EN 61000-4-3
IEC 61000-4-4/EN 61000-4-4
IEC 61000-4-5/EN 61000-4-5
IEC 61000-4-6/EN 61000-4-6
IEC 61000-4-11/EN 61000-4-11
Canada: ICES-001:2004
Australia/New Zealand: AS/NZS
Safety UL61010-1 2nd edition, CAN/CSA22.2 No. 61010-1-04
Vibration Meets IEC60068-2-6 and MIL-PRF-28800; class 3 random
Oscilloscope Experience Redefined:
InfiniiVision 4000 X-Series Performance Characteristics (Continued)
34 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
General and environmental characteristics
Shock Meets IEC 60068-2-27 and MIL-PRF-28800; class 3 random; (operating 30 g, ½ sine. 11 ms duration, 3 shocks/
axis along major axis, total of 18 shocks)
Dimensions (W x H x D) 454 mm x 275 mm x 156 mm
Weight Net: 6.3 kg (13.9 lbs), shipping: 11.4 kg (25 lbs)
Kensington style lock Rear-panel security slot connects to standard Kensington-style lock
Nonvolatile storage
Reference waveform display 4 internal waveforms or USB thumb drive
Save formats Setup (*.scp), 8- or 24-bit bitmap image (*.bmp), PNG 24-bit image (*.png), CSV data (*.csv), ASCII XY data
(*.csv), binary data (*.bin), lister data (*.csv), reference waveform data (*.h5), multi-channel waveform data (*.h5),
mask (*.mask), arbitrary waveform data (*.csv), power harmonics data (*.csv), USB signal quality (*.html & *.bmp)
Max USB flash drive size Supports industry-standard flash drives
Set ups without USB flash drive 10 internal setups
Set ups with USB flash drive Limited by size of USB drive
Included standard with oscilloscope
Calibration Certificate of Calibration, 2-year calibration interval
Probes One per channel N2894A 700 MHz passive probe (10:1 attenuation)
N6450-60001 16-digital-channel MSO cable (1 per oscilloscope included on all MSO models and
DSOXPERFMSO)
Localized interface English, Chinese (simplfied and traditional), French, German, Italian, Japanese, Korean, Portuguese, Russian, and
Spanish localized front panel overlays, interface, and built-in help system
Power cord Localized power cord
Front panel protection Front panel cover
Documentation CD containing localized user’s guide, service guide, and programmer’s manual
For MET/CAL procedures, visit the Cal Lab Solutions website: http://www.callabsolutions.com/products/Keysight/
Related literature
Publication title Publication number
Oscilloscope Memory Architectures – Why All Acquisition Memory is Not Created Equal - Application Note 5991-1024EN
Triggering on Infrequent Anomalies and Complex Signals using Zone Trigger - Application Note 5991-1107EN
Switch Mode Power Supply Measurements - Application Note 5991-1117EN
Characterizing Hi-Speed USB 2.0 Serial Buses In Embedded Designs - Data Sheet 5991-1148EN
Oscilloscope Experience Redefined:
InfiniiVision 4000 X-Series Performance Characteristics (Continued)
35 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
4000 X-Series bandwidth upgrade models
Model number Description Type
DSOX4B3T52U 350 MHz ~ 500 MHz, 2 ch License only
DSOX4B3T54U 350 MHz ~ 500 MHz, 4 ch License only
Process description
1 Place order for a license-only upgrade or measurement application product with a Keysight sales partner.
2For measurement applications, you will receive a paper or electronic .pdf entitlement certificate. For bandwidth upgrades only, you will receive a
stick-on label document indicating upgraded bandwidth specification in addition to a paper entitlement certificate.
3Use entitlement certificate containing instructions and certificate number needed to generate a license file for a particular 4000 X-Series
oscilloscope model number and serial number unit.
4Receive the licensed file and installation instructions via email.
5Copy license file (.lic extension) from email to a USB drive and follow instructions in email to install the purchased bandwidth upgrade or
measurement application on the oscilloscope.
6 For bandwidth upgrades only, attach bandwidth upgraded stick-on labels to front and rear panels of the oscilloscope. Model number and serial
number of the oscilloscope do not change.
Oscilloscope Experience Redefined:
License-only Bandwidth Upgrades and Measurement Applications
Model number Description
DSOX4EMBD Embedded serial triggering and analysis (I²C, SPI)
DSOX4COMP Computer serial triggering and analysis
(RS232/UART)
DSOX4USBFL USB 2.0 Full/Low Speed serial decode and
triggering
DSOX4USBH 1USB 2.0 Hi-Speed serial decode and triggering
DSOX4USBSQ USB 2.0 signal quality test
DSOX4UPD USB PD serial triggering and decode
DSOX4AUTO Automotive serial triggering and analysis
(CAN/CAN-dbc/LIN)
DSOX4FLEX FlexRay serial triggering and analysis
DSOX4CXPI CXPI serial triggering and decode
DSOX4AUDIO Audio serial triggering and analysis (I²S)
DSOX4AERO Aerospace and defence serial triggering and decode
(MIL-STD 1553 and ARINC 429)
DSOX4NRZ User-definable Manchester and NRZ triggering and
decode
DSOX4WAVEGEN2 Dual-channel WaveGen 20 MHz arbitrary/function
generator
DSOX4PWR Power analysis application
DSOX4FRA Frequency Response Analysis (FRA)
DSOX4MASK Mask Limit Testing
DSOX4VID Enhanced video/TV application package
DSOXPERFMSO 4000 X-Series oscilloscope MSO upgrade
(N6450-60001 MSO cable is shipped separately)
DSOX4NFC NFC triggering and automated test software
1� DSOX4USBH is available only for 1 GHz and 1.5 GHz models.
36 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
4000 X-Series bandwidth upgrade models
Model number Description Type
DSOX4B2T32U 200 MHz ~ 350 MHz , 2 ch Service center
DSOX4B2T34U 200 MHz ~ 350 MHz , 4 ch Service center
DSOX4B5T104U 500 MHz ~ 1 GHz , 4 ch Service center
DSOX4B1T154U 1 GHz ~ 1.5 GHz , 4 ch Service center
DSOX4B2T52U 200 MHz ~ 500 MHz , 2 ch Service center
DSOX4B2T54U 200 MHz ~ 500 MHz , 4 ch Service center
DSOX4B2T104U 200 MHz ~ 1 GHz , 4 ch Service center
DSOX4B2T154U 200 MHz ~ 1.5 GHz , 4 ch Service center
DSOX4B3T104U 350 MHz ~ 1 GHz , 4 ch Service center
DSOX4B3T154U 350 MHz ~ 1.5 GHz , 4 ch Service center
DSOX4B5T154U 500 MHz ~ 1.5 GHz , 4 ch Service center
Process description
1 Place order for a return-to-Keysight service center bandwidth upgrade product to a Keysight sales partner. Serivce Center installation,
calibration, shipment costs are in addition to bandwidth upgrade product price.
2 Keysight Business Center will contact you regarding process and timing of the service center installation. Continue to use oscilloscope until
contacted again later when parts are available at service center.
3 Ship the oscilloscope per provided instructions to service center.
4 Service center ships back upgraded oscilloscope with stick-on labels applied to front and rear panels indicating upgraded bandwidth
specification. Model number and serial number of the oscilloscope do not change.
Oscilloscope Experience Redefined:
Return-to-Keysight Service Center Bandwidth Upgrades
37 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
Keysight Oscilloscopes
Multiple form factors from 20 MHz to > 90 GHz | Industry-leading specs | Powerful applications
38 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
www.axiestandard.org
AdvancedTCA® Extensions for Instrumentation and Test (AXIe) is an
open standard that extends the AdvancedTCA for general purpose and
semiconductor test. The business that became Keysight was a founding
member of the AXIe consortium. ATCA®, AdvancedTCA®, and the ATCA logo
are registered US trademarks of the PCI Industrial Computer Manufacturers
Group.
www.lxistandard.org
LAN eXtensions for Instruments puts the power of Ethernet and the Web
inside your test systems. The business that became Keysight was a founding
member of the LXI consortium.
www.pxisa.org
PCI eXtensions for Instrumentation (PXI) modular instrumentation delivers a
rugged, PC-based high-performance measurement and automation system.
Learn more at
www.keysight.com/find/software
Start with a 30-day free trial.
www.keysight.com/find/free_trials
Download your next insight
Keysight software is downloadable
expertise. From first simulation through
first customer shipment, we deliver the
tools your team needs to accelerate from
data to information to actionable insight.
Electronic design automation (EDA)
software
Application software
Programming environments
Productivity software
39 | Keysight | InfiniiVision 4000 X-Series Oscilloscopes - Data Sheet
This information is subject to change without notice.
© Keysight Technologies, 2013 - 2017
Published in USA, November 7, 2017
5991-1103EN
www.keysight.com
Evolving Since 1939
Our unique combination of hardware, software, services, and people can help you
reach your next breakthrough. We are unlocking the future of technology.
From Hewlett-Packard to Agilent to Keysight.
myKeysight
www.keysight.com/find/mykeysight
A personalized view into the information most relevant to you.
http://www.keysight.com/find/emt_product_registration
Register your products to get up-to-date product information and
find warranty information.
Keysight Services
www.keysight.com/find/service
Keysight Services can help from acquisition to renewal across your
instrument’s lifecycle. Our comprehensive service offerings—one-
stop calibration, repair, asset management, technology refresh,
consulting, training and more—helps you improve product quality
and lower costs.
Keysight Assurance Plans
www.keysight.com/find/AssurancePlans
Up to ten years of protection and no budgetary surprises to ensure
your instruments are operating to specification, so you can rely on
accurate measurements.
Keysight Channel Partners
www.keysight.com/find/channelpartners
Get the best of both worlds: Keysight’s measurement expertise and
product breadth, combined with channel partner convenience.
www.keysight.com/find/4000X-Series
For more information on Keysight
Technologies’ products, applications or
services, please contact your local Keysight
office. The complete list is available at:
www.keysight.com/find/contactus
Americas
Canada (877) 894 4414
Brazil 55 11 3351 7010
Mexico 001 800 254 2440
United States (800) 829 4444
Asia Pacific
Australia 1 800 629 485
China 800 810 0189
Hong Kong 800 938 693
India 1 800 11 2626
Japan 0120 (421) 345
Korea 080 769 0800
Malaysia 1 800 888 848
Singapore 1 800 375 8100
Taiwan 0800 047 866
Other AP Countries (65) 6375 8100
Europe & Middle East
Austria 0800 001122
Belgium 0800 58580
Finland 0800 523252
France 0805 980333
Germany 0800 6270999
Ireland 1800 832700
Israel 1 809 343051
Italy 800 599100
Luxembourg +32 800 58580
Netherlands 0800 0233200
Russia 8800 5009286
Spain 800 000154
Sweden 0200 882255
Switzerland 0800 805353
Opt. 1 (DE)
Opt. 2 (FR)
Opt. 3 (IT)
United Kingdom 0800 0260637
For other unlisted countries:
www.keysight.com/find/contactus
(BP-9-7-17)
DEKRA Certified
ISO9001 Quality Management System
www.keysight.com/go/quality
Keysight Technologies, Inc.
DEKRA Certified ISO 9001:2015
Quality Management System

Navigation menu