THq Manual

User Manual:

Open the PDF directly: View PDF PDF.
Page Count: 8

DownloadTHq Manual
Open PDF In BrowserView PDF
THq
Digital Tissue Homogenizer

User Manual
OMNI THq

This manual is a guide for the use of the Omni International THq, Digital Tissue Homogenizer and accessories.
Data herein has been verified and validated. It is believed adequate for the intended use of the instrument. If the instrument or
procedures are used for purposes over and above the capabilities specified herein, confirmation of the validity and suitability
should be obtained, otherwise Omni International does not guarantee results and assumes no obligation or liability. This
publication is not a license to operate under, or a recommendation to infringe upon, any process patents. Notes, cautions, and
warnings within the text of this manual are used to emphasize important and critical instructions.
This Omni International product is warranted to be free from defects in material and workmanship for a period of THREE
YEARS from the date of delivery. Omni International will repair or replace and return free of charge any part which is returned
to its factory within said period, transportation prepaid by user, and which is found upon inspection to have been defective in
materials or workmanship. For the first 90 days, both parts and service are without charge. For the balance of the period, parts
will be provided but service will be charged at established labor rates. This warranty does not include normal wear from use; it
does not apply to any instrument or parts which have been altered by anyone other than an employee of Omni International
nor to any instrument which has been damaged through accident, negligence, failure to follow operating instructions, the
use of electric currents or circuits other than those specified on the plate affixed to the instrument, misuse, or abuse. Omni
International reserves the right to change, alter, modify, or improve any of its instruments without any obligation whatsoever
to make corresponding changes to any instrument previously sold or shipped.
THE FORGOING OBLIGATION IS IN LIEU OF ALL OBLIGATIONS AND LIABILITIES INCLUDING NEGLIGENCE AND ALL WARRANTIES
OF MERCHANTABILITY OR OTHERWISE, EXPRESSED OR IMPLIED IN FACT OR BY LAW, AND STATE OUR ENTIRE AND EXCLUSIVE
LIABILITY AND BUYERS EXCLUSIVE REMEDY FOR ANY CLAIM OF DAMAGES IN CONNECTION WITH THE SALE OR FURNISHING
OF GOODS OR PARTS, THEIR DESIGN, SUITABILITY FOR USE, INSTALLATION, OR OPERATION. OMNI INTERNATIONAL WILL
IN NO EVENT BE LIABLE FOR ANY SPECIAL OR CONSEQUENTIAL DAMAGES WHATSOEVER, AND THEIR LIABILITY UNDER NO
CIRCUMSTANCES WILL EXCEED THE CONTRACT PRICE FOR THE GOODS FOR WHICH LIABILITY IS CLAIMED.
- READ ALL INSTRUCTIONS BEFORE USING.

The Omni THq has been engineered for economical functionality as well as safety; however, basic safety precautions
and common sense must always be demonstrated when using any electrical product. DO NOT attempt to modify
any part of the Omni THq. If you experience problems with or have questions about your Omni THq, contact your
authorized dealer or call Omni International at 1-800-776-4431 or 1-770-421-0058.
- DO NOT allow the machine to be submerged in any liquid.
- DO NOT use in any setting other than an indoor laboratory.
- DO NOT plug power cord into an incorrect outlet.
To reduce the risk of burns, electrocution, fire, or injury:
- Use this product only for its intended purpose as described in this booklet. DO NOT use attachments not
recommended by the manufacturer.
- DO NOT operate the product if it is damaged in any way.
- Keep this product away from heated surfaces.
WARNING: The tip of the saw tooth generator probe is sharp.
WARNING: Reduce the risk of unintentional starting; make sure that the power switch is in the OFF position prior to
plugging in the unit.
WARNING: DO NOT process pathogenic material in an open container, since aerosols created during normal processing
could be inhaled by the operator. Please call for assistance in processing pathogens or other material which require
sealed enclosures.
WARNING: Keep all housings in place and in working order.
WARNING: Remove all tools from the generator probe before turning the motor on.
WARNING: DO NOT use the motor in a dangerous environment.
WARNING: Disconnect the motor before servicing, and when changing the generator probe.
WARNING: DO NOT modify the plug or cord that is provided. If the plug will not fit the outlet, have the proper outlet
installed by a qualified electrician.
WARNING: Reduce the risk of unintentional starting; make sure the power switch is in the OFF position before
plugging in the motor.
WARNING: Damaged or worn power cords should be repaired or replaced immediately by a qualified electrician.
WARNING: Improper connection of the equipment can result in a risk of electric shock.
DECONTAMINATION
Should an instrument or component that has been used with radioactive or pathogenic material require factory or
field service, comply with the following procedure to ensure the safety of service personnel:
Clean the parts to be serviced of all encrusted material and decontaminate them. There must be no radioactivity
detectable by survey equipment. Obtain a Decontamination Certificate from Omni International. Complete the
certificate and attach to the instrument or parts being returned. If no Decontamination Certificate is attached, and a
potential radioactive or biological hazard is detected or suspected by Omni International, the equipment will not be
serviced until proper decontamination and certification is complete. The sender will be contacted for instructions as to
the disposition of the equipment. Disposition costs will be borne by the sender.
WARNING: It is a violation of federal law to transport biologically hazardous or radioactive materials without proper
packaging, labeling, and appropriate warnings.
Should this product ever require service, please contact
Omni International at 1-800-776-4431

Overview
OMNI THq

1

5

2

6

3

7

4

8

1. Power cord port
2. Run/Stop LED indicators
3. Power
4. Generator Probe coupling

Clamp

5. Digital Readout Screen
6. Run button
7. Stop button
8. Increase/Decrease RPM

The Omni THq consists of the following:

12-118

Description

Part Number

Quantity

Motor Drive Unit

12-200-1

1

Clamp

12-118

1

Tool Kit

T1001

1

Soft Tissue Omni Tip

30007L

6

Hard Tissue Omni Tip

30007HT

6

Instruction Manual

03-225

1

Optional Stand Available (sold separately) - Order No. S1000

Stand Mounting

Clamp
Release Button

Push to lock
into Clamp

1. Slide the THq motor into the opening
on the end of the post clamp
assembly and push it down firmly.
The unit will lock in place.

Clamp
12-118

2. To release the THq, press the black
button on the back of the opening
of the post clamp assembly and lift
the motor from the cradle.
Stand

3. Slide the post clamp assembly
down over the end of the post until
the clamp is at the desired height
and lock in place.

S1000

(sold separately)

Generator Probes and Omni TipsTM
1. Insert Generator
Probe into coupling

2. Quarter turn
clockwise to lock

3. Ready to use

*always submerge probe
in liquid before starting

1/4 Turn

*Never run dry

IMPORTANT INFORMATION
The blue cap must be removed from the end of the
generator probe before operating.
1.

2.

3.
Ready to homogenize.

Stainless Steel Generator Probes

Omni Tips®
Plastic
Disposable
Generator
Probes

Size

Processing
Range

Order No.

5 mm x 75 mm

.2 mL - 5 mL

B5-075

5 mm x 75 mm (6 pack)

.2 mL - 5 mL

B5-075-6

7 mm x 110 mm

.25 mL - 30 mL

B7-110ST

7 mm x 110 mm (6 pack)

.25 mL - 30 mL

B7-110ST-6

10 mm x 110 mm

1.5 mL - 100 mL

B10-110ST

10 mm x 110 mm (6 pack)

1.5 mL - 100 mL

B10-110ST-6

Type
Hard Tissue

Soft Tissue

Size

Quantity

Order No.

7 mm x 110 mm

25 Pack

30750H

7 mm x 110 mm

50 Pack

32750H

7 mm x 110 mm

25 Pack

30750

7 mm x 110 mm

50 Pack

32750

Operation
1. Run button

1
2
3
4
5

2. Stop button
3. Increase RPM
4. Decrease RPM
5. Power button

General Operation
Power
- Plug in the power cord and press the power button
		
Set Speed
- Adjust speed using the Increase/Decrease speed
buttons
Start
- Press the Run button
Stop
- Press the Stop button

Troubleshooting
PROBLEM

CORRECTIVE ACTION

The THq is plugged in and turned on but
is not functioning.

- Check power cord connectors
- Check wall socket for power to the
outlet

Teflon bearing wears quickly

- Fluid level in the sample may be too low
- Immerse the probe further into the fluid

Execessive splashing in the sample

- Fluid level is too low for the tube size

Motor does not spin.
Green light flashes.

- Rotor is locked. Remove the obstruction

935C Cobb Place Blvd.
Kennesaw, GA 30144
800.776.4431 • 770.421.0058
www.omni-inc.com

PN: 03-225 Rev. C



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : Yes
Language                        : en-US
Tagged PDF                      : Yes
XMP Toolkit                     : Adobe XMP Core 5.6-c137 79.159768, 2016/08/11-13:24:42
Create Date                     : 2017:10:05 15:56:14-04:00
Metadata Date                   : 2017:10:05 15:56:16-04:00
Modify Date                     : 2017:10:05 15:56:16-04:00
Creator Tool                    : Adobe InDesign CC 2017 (Windows)
Instance ID                     : uuid:194b7a8b-1e92-48ca-9854-7e3eac484373
Original Document ID            : xmp.did:6b2849db-2743-4f91-aca1-1ce5db3fbee8
Document ID                     : xmp.id:6eaf857b-e8e4-1443-9f4f-08b4f55fc9b3
Rendition Class                 : proof:pdf
Derived From Instance ID        : xmp.iid:a2b1bf71-5bdd-d149-910a-b0fca0ff75dc
Derived From Document ID        : xmp.did:a2b1bf71-5bdd-d149-910a-b0fca0ff75dc
Derived From Original Document ID: xmp.did:6b2849db-2743-4f91-aca1-1ce5db3fbee8
Derived From Rendition Class    : default
History Action                  : converted
History Parameters              : from application/x-indesign to application/pdf
History Software Agent          : Adobe InDesign CC 2017 (Windows)
History Changed                 : /
History When                    : 2017:10:05 15:56:14-04:00
Format                          : application/pdf
Producer                        : Adobe PDF Library 15.0
Trapped                         : False
Page Count                      : 8
Creator                         : Adobe InDesign CC 2017 (Windows)
EXIF Metadata provided by EXIF.tools

Navigation menu