Hantek 100MHz 2 Channel Digital
Storage Oscilloscope - DSO5102P
Sold and supported in the United States, buy the Hantek DSO5102P at CircuitSpecialists.com

DSO5000P Series
Digital Storage Oscilloscope
User Manual
(Version 1.1)

Contents

Contents
Contents .................................................................................................................................................. i
Chapter 1

Safety Tips ....................................................................................................................... 1

1.1

General Safety Summary ......................................................................................................... 1

1.2

Safety Terms and Symbols....................................................................................................... 2

1.3

Terms on Product ..................................................................................................................... 2

1.4

Symbols on Product ................................................................................................................. 2

1.5

Product Scrapping .................................................................................................................... 2

Chapter 2

Overview .......................................................................................................................... 3

2.1

Brief Introduction to DSO 5000P Series ................................................................................... 3

2.2

Help System ............................................................................................................................. 3

Chapter 3
3.1

Getting Started Guide ..................................................................................................... 5

Installation ................................................................................................................................ 5

3.1.1

Power Supply .................................................................................................................... 5

3.1.2

Power Cord ....................................................................................................................... 5

3.2

Functional Check...................................................................................................................... 5

3.2.1

Power on the oscilloscope ................................................................................................ 5

3.2.2

Connect the oscilloscope .................................................................................................. 5

3.2.3

Observe the waveform...................................................................................................... 6

3.3

Probe Examination ................................................................................................................... 6

3.3.1

Safety ................................................................................................................................ 6

3.3.2

Use of Probe Check Wizard ............................................................................................. 7

3.4

Manual Probe Compensation ................................................................................................... 7

3.5

Probe Attenuation Setting ........................................................................................................ 8

3.6

Self Calibration ......................................................................................................................... 8

Chapter 4

Main Feature Description ............................................................................................... 9

4.1

Oscilloscope Setup................................................................................................................... 9

4.2

Trigger ...................................................................................................................................... 9

4.3

Data Acquisition .......................................................................................................................11

4.4

Waveform Scaling and Positioning..........................................................................................11

4.5

Waveform Measurement ........................................................................................................ 12

Chapter 5
5.1

Display Area ........................................................................................................................... 14

5.1.1
5.2

XY Format ....................................................................................................................... 16

Horizontal Controls ................................................................................................................. 16

5.2.1
5.3

Basic Operation ............................................................................................................ 14

Scan Mode Display (Roll Mode) ..................................................................................... 19

Vertical Controls ..................................................................................................................... 19

DSO5000P Series Digital Storage Oscilloscope User Manual

i

Contents

5.3.1

Math FFT ........................................................................................................................ 21

5.4

Trigger Controls ...................................................................................................................... 26

5.5

Menu and Option Buttons....................................................................................................... 32

5.5.1

Save/Recall ..................................................................................................................... 32

5.5.2

Measure .......................................................................................................................... 33

5.5.3

Acquire ............................................................................................................................ 35

5.5.4

Utility ............................................................................................................................... 37

5.5.5

Cursor ............................................................................................................................. 37

5.5.6

Display ............................................................................................................................ 38

5.6

Fast Action Buttons ................................................................................................................ 39

5.6.1

Autoset ............................................................................................................................ 39

5.6.2

Help................................................................................................................................. 40

5.6.3

Default Setup .................................................................................................................. 40

5.7

Multi-functional Knobs and Buttons ........................................................................................ 42

5.8

Signal Connectors .................................................................................................................. 43

Chapter 6

Application Examples .................................................................................................. 44

6.1

Example 1: Taking Simple Measurements ............................................................................. 44

6.2

Example 2: Taking Cursor Measurements ............................................................................. 46

6.3

Example 3: Analyzing Input Signals to Eliminate Random Noise .......................................... 49

6.4

Example 4: Capturing Single-shot Signal............................................................................... 50

6.5

Example 5: Using X-Y Mode .................................................................................................. 51

6.6

Example 6: Triggering on Pulse Width ................................................................................... 52

6.7

Example 7: Triggering on Video Signal .................................................................................. 53

6.8

Example 8: Using Slope Trigger to Capture Particular Slope Signal ..................................... 55

6.9

Example 9: Using Overtime Trigger to Measure Long Pulse Signal ...................................... 55

6.10

Example 10: Using Math Functions to Analyze Waveforms................................................... 56

6.11

Example 11: Measuring Data Propagation Delay .................................................................. 57

Chapter 7
7.1

Troubleshooting............................................................................................................ 59

Problem Settlement ................................................................................................................ 59

Chapter 8

Specifications ............................................................................................................... 60

8.1

Technical Specifications ......................................................................................................... 60

8.2

Accessories ............................................................................................................................ 66

Chapter 9

General Care and Cleaning .......................................................................................... 67

9.1

General Care .......................................................................................................................... 67

9.2

Cleaning ................................................................................................................................. 67

Appendix A Harmful and Poisonous Substances or Elements ...................................................... 68

DSO5000P Series Digital Storage Oscilloscope User Manual

ii

Safety Tips

Chapter 1 Safety Tips
1.1 General Safety Summary
Read the following safety precautions to avoid injury and prevent damage to this product or any
products connected to it. To evade potential hazards, use this product only as specified.
Only qualified personnel should perform maintenance.
Avoid fire or personal injury.
Use suitable power cord. Use only the power cord specified for this product and certified for the
country of use.
Connect and disconnect properly. Connect a probe with the oscilloscope before it is connected
to measured circuits; disconnect the probe from the oscilloscope after it is disconnected from
measured circuits.
Ground the product. This product is grounded through the grounding conductor of the power
cord. To avoid electric shock, the grounding conductor must be connected to earth ground. Before
making connections to the input or output terminals of the product, ensure that the product is
properly grounded.
Connect the probe in a right way. The probe ground lead is at ground potential. Do not connect
the ground lead to an elevated voltage.
Check all terminal ratings. To avoid fire or shock hazard, check all ratings and markings on the
product. Refer to the product manual for detailed information about ratings before making
connections to the product.
Do not operate without covers. Do not operate this product with covers or panels removed.
Avoid exposed circuitry. Do not touch exposed connections and components when power is
present.
Do not operate with suspected failures. If you suspect there is damage to this product, have it
inspected by qualified service personnel.
Assure good ventilation.
Do not operate in wet/damp environments.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.

DSO5000P Series Digital Storage Oscilloscope User Manual

1

Safety Tips

1.2
.2 Safety Terms and Symbols
The following terms may appear in this manual:
WARNING. Warning statements point out conditions or practices that could result in injury
or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage
to this product or other property.

1.3 Terms on Product
The following terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you read the marking.
CAUTION indicates a possible hazard to this product or other property.

1.4 Symbols on Product
The following symbols may appear on the product:

Protective
Ground
(Earth)
Terminal

Mains
Disconnected
OFF (Power)

Measurement
Ground
Terminal

Mains
Connected
ON (Power)

CAUTION
Refer to Manual

Measurement
Input Terminal

High Voltage

1.5 Product Scrapping
Device Recycling
We need extract and utilize natural resources to produce this device. If you do not reclaim the
device in a proper way, some substances it contains may become harmful or poisonous to
environments or human bodies. To avoid them being released outside and to minimize the waste
of natural resources, we suggest you reasonably call back this device to ensure proper recovery
and recycling of most materials within it.

DSO5000P Series Digital Storage Oscilloscope User Manual

2

Overview

Chapter 2 Overview
2.1 Brief Introduction to DSO 5000P Series
Model
DSO5072P
DSO5102P
DSO5202P

Channels
2
2
2

Bandwidth

Sample Rate

70MHz
1GS/s
100MHz
1GS/s
200MHz
1GS/s
Table 2-1 Model List of DSO5000P Series

LCD
7 inch color
7 inch color
7 inch color

DSO5000P Series oscilloscopes cover the bandwidths from 70MHz to 200MHz, and provide
the real-time and equivalent sample rates respectively up to 1GSa/s and 25GSa/s. In addition,
they have 7 inch color TFT LCD as well as WINDOWS-style interfaces and menus for easy
operation.
What’s more, the plenty menu information and the easy-to-operate buttons allow you to gain
information as much as possible in measurement; the multifunctional knobs and the powerful
shortcut keys help you save a lot of time in operation; the Autoset function lets you detect sine and
square waves automatically; the Probe Check Wizard guides you to adjust the probe
compensation and set the Probe option attenuation factor. By using the three methods the
oscilloscope provides (context-sensitive, hyperlinks, and an index), you may master all operations
on the device in quite a short time so as to greatly improve your efficiency in production and
development.

2.2 Help System
This oscilloscope has a Help system with topics covering all of its features. You can use the Help
system to display several kinds of information:
General information about understanding and using the oscilloscope, such as Using the Menu
System.
Information about specific menus and controls, such as the Vertical Position Control.
Advice to problems you may come across while using an oscilloscope, such as Reducing
Noise.
The Help system provides three methods for you to find the information you want:
context-sensitive, hyperlinks, and an index.
Context-Sensitive
Push the HELP front-panel button and the oscilloscope displays information about the last menu
displayed on the screen. The HELP SCROLL LED lights beside the HORIZONTAL POSITION
DSO5000P Series Digital Storage Oscilloscope User Manual

3

Overview

knob indicate the alternative function of the knob. If a topic uses more than one page, turn the
HELP SCROLL knob to move from page to page within the topic.
Hyperlinks
Most of the help topics contain phrases marked with angle brackets, such as . They are
links to other topics. Turn the HELP SCROLL knob to move the highlight from one link to another.
Push the Show Topic option button to display the topic corresponding to the highlighted link. Press
the Back option button to return to the previous topic.
Index
Push the front-panel HELP button, and then press the Index option button. Push the Page Up or
Page Down option button until you find the index page which contains the topic you want to view.
Turn the HELP SCROLL knob to highlight a help topic. Press the Show Topic option button to
display the topic.
NOTE: Press the Exit option button or any menu button to remove the Help text from the
screen and return to displaying waveforms.

DSO5000P Series Digital Storage Oscilloscope User Manual

4

Getting Started Guide

Chapter 3 Getting Started Guide
3.1 Installation
To keep proper ventilation of the oscilloscope in operation, leave a space of more than 5cm away
from the top and the two sides of the product.

3.1.1Power Supply
Use a power supply that delivers 90 to 240 VRMS, 45 to 440 Hz.

3.1.2Power Cord
Use only power cords designed for this product. Refer to 8.2 Accessories for specific standards.

3.2 Functional Check
Follow the steps below to perform a quick functional check to your oscilloscope.

3.2.1Power on the oscilloscope
Plug in the oscilloscope and press the ON/OFF button. Then push the DEFAULT SETUP button.
The default Probe option attenuation setting is 10X.

The Default Setup button

3.2.2Connect the oscilloscope
Set the switch on the probe to 10X and connect the probe to Channel 1 on the oscilloscope. First,
align the slot in the probe connector with the protuberance on the CH1 BNC and push to connect;
then, turn to right to lock the probe in place; after that, connect the probe tip and reference lead to
the PROBE COMP connectors. There is a mark on the panel: Probe COMP ~5V@1KHz.

DSO5000P Series Digital Storage Oscilloscope User Manual

5

Getting Started Guide

CH1: to connect with the probe

PROBE COMP

3.2.3Observe the waveform
Press the AUTOSET button and you should see within a few seconds a square wave of about 5V
peak-to-peak at 1kHz in the display. Press the CH1 MENU button twice to remove Channel 1.
Push the CH2 MENU button and repeat Step 2 and Step 3 to observe Channel 2.

3.3 Probe Examination
3.3.1Safety
When using the probe, keep your fingers behind the guard on the probe body to avoid electric
shock. Do not touch metallic portions of the probe head while it is connected to a voltage source.
Connect the probe to the oscilloscope and connect the ground terminal to ground before you start
any measurements.

DSO5000P Series Digital Storage Oscilloscope User Manual

6

Getting Started Guide

3.3.2Use of Probe Check Wizard
Every time you connect a probe to an input channel, you should use the probe check wizard to
verify that this probe is operating correctly. There are two ways to do this:
1) Use the vertical menu (for example, push the CH1 MENU button) to set the Probe option
attenuation factor.
2) Press the PROBE CHECK button to use the Probe Check Wizard and configure the probe
option attenuation factor properly following menu prompts.

3.4 Manual Probe Compensation
Upon the first connection of a probe and an input channel, you should manually perform this
adjustment to match the probe to the input channel. Uncompensated or miscompensated probes
may lead to errors or faults in measurement. To adjust the probe compensation, follow the steps
below.
1. Set the Probe option attenuation in the channel menu to 10X. Set the switch on the probe to
10X and connect the probe to Channel 1 on the oscilloscope. If you use the probe hook-tip,
ensure it is firmly inserted onto the probe. Attach the probe tip to the PROBE COMP
~5V@1KHz connector and the reference lead to the PROBE COMP Ground connector.
Display the channel and then press the AUTOSET button.
2. Check the shape of the displayed waveform.

Compensated correctly

Overcompensated

Undercompensated

3. If necessary, use a nonmetallic screwdriver to adjust the variable capacity of your probe until
the shape of the waveform turns to be the same as the above figure. Repeat this step as
necessary. See the figure below for the way of adjustment.

DSO5000P Series Digital Storage Oscilloscope User Manual

7

Getting Started Guide

3.5 Probe Attenuation Setting
Probes are of various attenuation factors which affect the vertical scale of the signal. The Probe
Check function is used to verify if the Probe attenuation option matches the attenuation of the
probe.
As an alternative method to Probe Check, you can push a vertical menu button (such as the CH 1
MENU button) and select the Probe option that matches the attenuation factor of your probe.
Make sure that the Attenuation switch on the probe matches the Probe option in the oscilloscope.
Switch settings are 1X and 10X.
When the Attenuation switch is set to 1X, the probe limits the bandwidth of the oscilloscope to
6MHz. To use the full bandwidth of the oscilloscope, be sure to set the switch to 10X.

3.6 Self Calibration
The self calibration routine helps optimize the oscilloscope signal path for maximum measurement
accuracy. You can run the routine at any time but should always run it if the ambient temperature
changes by 5℃ or more. For a more accurate calibration, please power on the oscilloscope and
wait for 20 minutes until it has adequately warmed up.
To compensate the signal path, disconnect any probes or cables from the front-panel input
connectors. Then, push the UTILITY button, select the Do Self Cal option and follow the directions
on the screen.
.

DSO5000P Series Digital Storage Oscilloscope User Manual

8

Main Feature Description

Chapter 4 Main Feature Description
This chapter provides some general information that you need to learn before using an
oscilloscope. It contains:
1. Oscilloscope setup
2. Trigger
3. Data acquisition
4. Waveform scaling and positioning
5. Waveform measurement

4.1 Oscilloscope Setup
While operating the oscilloscope, you may often use three features: Autoset, saving a setup and
recalling a setup. Hereinafter they are introduced one by one.
Autoset: This function can be used to adjust the horizontal and vertical scales of the oscilloscope
automatically and set the trigger coupling, type, position, slope, level and mode, etc., to acquire a
stable waveform display.
Saving a Setup: By default, the oscilloscope will save the setup each time before being closed,
and automatically recall the setup once being turned on. (Note: If you modify the setup, please
wait for more than 5 seconds before turning off the oscilloscope to ensure the proper
storage of new settings.) You can save 10 settings permanently in the oscilloscope and reset
them as necessary.
Recalling a Setup: The oscilloscope can recall any of your saved setups or the default factory
setup.
Default Setup: The oscilloscope is preset for normal operations when it is shipped from the
factory. This is the default setup. You may recall this setup at any time for your requirements. To
view the default settings, refer to Section 5.6.3.

4.2 Trigger
The trigger determines when the oscilloscope begins to acquire data and display a waveform.
Once a trigger is properly set up, the oscilloscope can convert unstable displays or blank screens
to meaningful waveforms. Here introduce some basic concepts about trigger.
Trigger Source: The trigger can be generated with multiple sources. The most common one is the
input channel (alternative between CH1 and CH2). Whether the input signal is displayed or not, it
can trigger normal operations. Also the trigger source can be any signal connected to an external
DSO5000P Series Digital Storage Oscilloscope User Manual

9

Main Feature Description

trigger channel or the AC power line (only for Edge triggers). The source with the AC power line
shows the frequency relationship between the signal and the AC commercial power.
Trigger Type: The oscilloscope has six types of triggers: Edge, Video, Pulse Width, Slope,
Overtime and Swap.
Edge Trigger uses the analog or digital test circuits for triggering. It happens when the
input trigger source crosses a specified level in a specified direction.
Video Trigger performs a field or line trigger through standard video signals.
Pulse Width Trigger can trigger normal or abnormal pulses that meet trigger conditions.
Slope Trigger uses the rise and fall times on the edge of signal for triggering.
Overtime Trigger happens after the edge of signal reaches the set time.
Swap Trigger, as a feature of analog oscilloscopes, gives stable displays of signals at
two different frequencies. Mainly it uses a specific frequency to switch between two
analog channels CH1 and CH2 so that the channels will generate swap trigger signals
through the trigger circuitry.
Trigger Mode: You can select the Auto or Normal mode to define how the oscilloscope acquires
data when it does not detect a trigger condition. Auto Mode performs the acquisition freely in
absence of valid trigger. It allows the generation of untriggered waveforms with the time base set
to 80ms/div or slower. Normal Mode updates the displayed waveforms only when the
oscilloscope detects a valid trigger condition. Before this update, the oscilloscope still displays the
old waveforms. This mode shall be used when you want to only view the effectively triggered
waveforms. In this mode, the oscilloscope displays waveforms only after the first trigger. To
perform a single sequence acquisition, push the SINGLE SEQ button.
Trigger Coupling: Trigger Coupling determines which part of the signal will be delivered to the
trigger circuit. This can help to obtain a stable display of the waveform. To use trigger coupling,
push the TRIG MENU button, select an Edge or Pulse trigger, and then select a Coupling option.
Trigger Position: The horizontal position control establishes the time between the trigger position
and the screen center.
Slope and Level: The Slope and Level controls help to define the trigger. The Slope option
determines whether the trigger point is on the rising or falling edge of a signal. To perform the
trigger slope control, press the TRIG MENU button, select an Edge trigger, and use the Slope
button to select rising or falling. The TRIGGER LEVEL knob controls the trigger point is on which
position of the edge.

Trigger level can be
adjusted vertically
Rising Edge Falling Edge
Trigger slope can be rising or falling

DSO5000P Series Digital Storage Oscilloscope User Manual

10

Main Feature Description

4.3 Data Acquisition
When you acquire an analog signal, the oscilloscope will convert it into a digital one. There are
two kinds of acquisition: Real-time acquisition and Equivalent acquisition. The real-time acquisition
has three modes: Normal, Peak Detect, and Average. The acquisition rate is affected by the
setting of time base.
Normal: In this acquisition mode, the oscilloscope samples the signal in evenly spaced intervals to
establish the waveform. This mode accurately represents signals in most time. However, it does
not acquire rapid variations in the analog signal that may occur between two samples, which can
result in aliasing and may cause narrow pulses to be missed. In such cases, you should use the
Peak Detect mode to acquire data.
Peak Detect: In this acquisition mode, the oscilloscope gets the maximum and minimum values of
the input signal over each sample interval and uses these values to display the waveform. In this
way, the oscilloscope can acquire and display those narrow pulses that may have otherwise been
missed in Normal mode. However, noise will appear to be higher in this mode.
Average: In this acquisition mode, the oscilloscope acquires several waveforms, averages them,
and displays the resulting waveform. You can use this mode to reduce random noise.
Equivalent Acquisition: This kind of acquisition can be utilized for periodic signals. In case the
acquisition rate is too low when using the real-time acquisition, the oscilloscope will use a fixed
rate to acquire data with a stationary tiny delay after each acquisition of a frame of data. After
repeating this acquisition for N times, the oscilloscope will arrange the acquired N frames of data
by time to make up a new frame of data. Then the waveform can be recovered. The number of
times N is related to the equivalent acquisition rate.
Time Base: The oscilloscope digitizes waveforms by acquiring the value of an input signal at
discrete points. The time base helps to control how often the values are digitized. Use the
SEC/DIV knob to adjust the time base to a horizontal scale that suits your purpose.

4.4 Waveform Scaling and Positioning
The display of waveforms on the screen can be changed by adjusting their scale and position.
Once the scale changes, the waveform display will increase or decrease in size. Once the position
changes, the waveform will move up, down, right, or left.
The channel reference indicator (located on the left of the graticule) identifies each waveform on
the screen. It points to the ground level of the waveform record.
Vertical Scale and Position: The vertical position of a waveform can be changed by moving it up
and down on the screen. To compare data, you may align a waveform over another. When you
push the VOLTS/DIV button to change the vertical scale of a waveform, the waveform display will
contract or expand vertically to the ground level.

DSO5000P Series Digital Storage Oscilloscope User Manual

11

Main Feature Description

Horizontal Scale and Position: Pretrigger Information
You can adjust the HORIZONTAL POSITION control to view waveform data before the trigger,
after the trigger, or some of each. When you change the horizontal position of a waveform, you are
actually changing the time between the trigger position and the screen center.
For example, if you want to find out the cause of a glitch in your test circuit, you should trigger on
the glitch and make the pretrigger period long enough to capture data before the glitch. Then you
can analyze the pretrigger data and perhaps find the cause. You are allowed to change the
horizontal scale of all the waveforms by turning the SEC/DIV knob. For example, you may want to
see just one cycle of a waveform to measure the overshoot on its rising edge. The oscilloscope
shows the horizontal scale as time per division in the scale readout. Since all active waveforms
use the same time base, the oscilloscope only displays one value for all the active channels.

4.5 Waveform Measurement
The oscilloscope displays graphs of voltage versus time and can help to measure the displayed
waveform. There are several ways to take measurements, using the graticule, the cursors or
performing an automatic measurement.
Graticule: This method allows you to make a quick, visual estimate and take a simple
measurement through the graticule divisions and the scale factor.
For example, you can take simple measurements by counting the major and minor graticule
divisions involved and multiplying by the scale factor. If you counted 6 major vertical graticule
divisions between the minimum and maximum values of a waveform and knew you had a scale
factor of 50mV/division, you could easily calculate your peak-to-peak voltage as follows:
6 divisions x 50mV/division = 300mV.
Cursor: This method allows you to take measurements by moving the cursors. Cursors always
appear in pairs and the displayed readouts are just their measured values. There are two kinds of
cursors: Amplitude Cursor and Time Cursor. The amplitude cursor appear as a horizontal broken
line, measuring the vertical parameters. The time cursor appear as a vertical broken line,
measuring the horizontal parameters.
When using cursors, please make sure to set the Source to the waveform that you want to
measure on the screen. To use cursors, push the CURSOR button.

DSO5000P Series Digital Storage Oscilloscope User Manual

12

Main Feature Description

Cursor
Cursor

Automatic Measurement: The oscilloscope performs all the calculations automatically in this
mode. As this measurement uses the waveform record points, it is more precise than the graticule
and cursor measurements. Automatic measurements show the measurement results by readouts
which are periodically updated with the new data acquired by the oscilloscope

DSO5000P Series Digital Storage Oscilloscope User Manual

13

Basic Operation

Chapter 5 Basic Operation
The front panel of the oscilloscope is divided into several functional areas. In this chapter we will
give a quick overview of all control buttons and knobs on the front panel as well as the displayed
information on the screen and relative testing operations. The figure below illustrates the front
panel of the DSO5000P series digital oscilloscope.

Front Panel of DSO5000P Series

5.1 Display Area
1

23

4

5

6

7

8

19
9
18

17 16 15 14

13

12 11

DSO5000P Series Digital Storage Oscilloscope User Manual

10
14

Basic Operation

1. Display Format:
: YT

: XY

: Vectors

: Dots

: Gray indicates auto persistence; Green means persistence display is enabled. When
the icon is set to green, the time for persistence display will be shown behind it.
2. Acquisition Mode: Normal, Peak Detect or Average
3. Trigger Status:
The oscilloscope is acquiring pretriggered data.
All pretriggered data have been acquired and the oscilloscope is ready to accept a trigger.
T

The oscilloscope has detected a trigger and is acquiring the posttrigger information.
The oscilloscope works in auto mode and is acquiring waveforms in the absence of triggers.
The oscilloscope is acquiring and displaying waveform data continuously in scan mode.
The oscilloscope has stopped acquiring waveform data.

S

The oscilloscope has finished a single sequence acquisition.

4. Tool Icon:

: If this icon lights up, it means the keyboard of the oscilloscope is locked by the host computer
via USB control.
: If this icon lights up, it means the USB disk has been connected.
: This icon lights up only when the USB slave interface is connected with the computer.
5.
6.
7.
8.
9.
10.
11.
12.

Readout shows main time base setting.
Main Time Base Window
Display of window’s position in data memory and data length.
Window Time Base
Operating Menu shows different information for different function keys.
Readout shows frequency count.
Readout points out horizontal waveform position
Trigger Type:
: Edge trigger on the rising edge.
: Edge trigger on the falling edge.
: Video trigger with line synchronization.
: Video trigger with field synchronization.

DSO5000P Series Digital Storage Oscilloscope User Manual

15

Basic Operation

: Pulse Width trigger, positive polarity.
: Pulse Width trigger, negative polarity.
13.
14.
15.
16.

Pop-up Prompt
Readout tells trigger level.
Icon indicates whether the waveform is inverted or not.
20M Bandwidth Limit. If this icon lights up, it means the bandwidth limit is enabled, otherwise
disabled.
17. Icon indicates channel coupling.
18. Channel Marker
19. Window displays waveform.

5.1.1XY Format
The XY format is used to analyze phase differences, such as those represented by Lissajous
patterns. The format plots the voltage on CH1 against the voltage on CH2, where CH1 is the
horizontal axis and CH2 is the vertical axis. The oscilloscope uses the untriggered Normal
acquisition mode and displays data as dots. The sampling rate is fixed at 1 MS/s.
The oscilloscope can acquire waveforms in YT format at any sampling rate. You may view the
same waveform in XY format. To perform this operation, stop the acquisition and change the
display format to XY.
The table below shows how to operate some controls in XY format.
Controls

Usable or not in XY format

CH1 VOLTS/DIV and VERTICAL POSITION controls

Set the horizontal scale and position

CH2 VOLTS/DIV and VERTICAL POSITION controls

Continuously set the vertical scale and
position

Reference or Math

Unusable

Cursors

Unusable

Autoset (display format reset to YT)

Unusable

Time base controls

Unusable

Trigger controls

Unusable

5.2 Horizontal Controls
Use the horizontal controls to change the horizontal scale and position of waveforms. The
horizontal position readout shows the time represented by the center of the screen, using the
trigger time as zero. When you change the horizontal scale, the waveform will expand or contract
to the screen center. The readout near the upper right of the screen shows the current horizontal
position in second. M represents ‘Main Time Base’, and W indicates ‘Window Time Base’. The
DSO5000P Series Digital Storage Oscilloscope User Manual

16

Basic Operation

oscilloscope also has an arrow icon at the top of the graticule to indicate the horizontal position.

1. HORIZONTAL POSITION Knob: Used to control the trigger position against the screen center.
Push this button to reset the trigger point back to the screen center.
AN: Used to set the horizontal position as zero.
2. Each option in HORIZ MENU is described as follows.
Options

Settings

Comments

Window Control

Major Window
Minor Window

Selects the major or minor window in dual-window
mode. The window is highlighted once selected.
Press this option button in single-window mode to
enter the daul-window mode.

Mark

Right arrow
Left arrow
Set/Clear
Clear All

This function is usable only in dual-window mode. It
sets marks at some waveform record locations that
users are interested in, and searches for these marks
by right and left arrows. Then it positions the window
to this mark for further observation.

Holdoff

None

Select this menu and turn the multi-functional knob to
adjust the trigger holdoff time within the range of
100ns-10s. Select this menu and push the
multi-functional knob to reset the holdoff time with the
starting value 100ns.

Autoplay

None

This function is usable in dual-window mode. Push
this menu button and auto move it from left to right at
a specified speed. In the expanded window will
display corresponding waveforms until it stops once
reaching the rightmost side of the major scan window

DSO5000P Series Digital Storage Oscilloscope User Manual

17

Basic Operation

Single-window Mode

Dual-window Mode (Full Screen)
Location of expanded window data in memory

Major Window

Minor Window
(Expanded Window)

Multi-functional
Knob

3.SEC/DIV Knob: Used to change the horizontal time scale so as to magnify or compress the
waveform horizontally. If the waveform acquisition is stopped (by using the RUN/STOP or SINGLE
SEQ button), the SEC/DIV control will expand or compress the waveform. In dual-window mode,
push this knob to select major or minor window. When the major window is selected, this knob
provides the same functions as it provides in single-mode window. When the minor window is
selected, turn this knob to scale the waveform whose magnification is up to 1000.

DSO5000P Series Digital Storage Oscilloscope User Manual

18

Basic Operation

Notes:
1. For more information of the trigger holdoff, see Section Trigger Controls.
2. In single-window mode, press F0 to hide or show the menus on the right side. The
dual-window mode does not support the menu hiding function.

5.2.1Scan Mode Display (Roll Mode)
With the SEC/DIV control set to 80ms/div or slower and the trigger mode set to Auto, the
oscilloscope works in the scan acquisition mode. In this mode, the waveform display is updated
from left to right without any trigger or horizontal position control.

5.3 Vertical Controls
Vertical controls can be used to display and remove waveforms, adjust vertical scale and position,
set input parameters and perform math calculations. Each channel has a separate vertical menu
to set. See below for menu description.
1. VERTICAL POSITION Knob: Move the channel waveform up and down on the screen. In
dual-window mode, move the waveforms in both windows at the same time in a same direction.
Push this knob to return waveforms to the vertical center position on the screen. Two channels
correspond to two knobs.

2. Menu (CH1, CH2): Display vertical menu options; turn on or off the display of channel
waveforms.

DSO5000P Series Digital Storage Oscilloscope User Manual

19

Basic Operation

Options

Settings

Coupling

DC
AC
Ground

20MHz Bandwidth
Limit

Unlimited
Limited

VOLTS/DIV

Coarse
Fine

Probe Attenuation
Invert

1X
10X
100X
1000X
Off
On

Comments
DC passes both DC and AC components of the input signal.
AC blocks the DC component of the input signal and
attenuates signals below 10Hz.
Ground disconnects the input signal.
Limits the bandwidth to reduce display noise; filters the
signal to eliminate noise and other unnecessary HF
components.
Selects the resolution of the VOLTS/DIV knob.
Coarse defines a 1-2-5 sequence. Fine changes the
resolution to small steps between the Coarse settings.
Selects a value according to the probe attenuation factor so
as to ensure correct vertical readouts. Reduce bandwidth to
6MHz when using a 1X probe.
Inverts the waveform relative to the reference level.

Ground Coupling
Used to display a zero-volt waveform. Internally, the channel input is connected with a zero-volt
reference level.
Fine Resolution
In the fine resolution setting, the vertical scale readout displays the actual VOLTS/DIV setting. The
vertical scale changes only after you adjust the VOLTS/DIV control and set to coarse.
Remove Waveform Display
To remove a waveform from the screen, first push the menu button to display the vertical menu,
then push again to remove the waveform. A channel waveform which is unnecessary to be
displayed can be used as a trigger source or for math operations.
3. VOLTS/DIV Knob
Control the oscilloscope to magnify or attenuate the source signal of the channel waveform. The
vertical size of the display on the screen will change (increase or decrease) to the ground level.
Also you may use this knob to switch between coarse and fine.
4. MATH MENU: Display the waveform math operations. See the table below for details.
The MATH menu contains source options for all math operations.
Operations
+

Source Options
CH1+CH2
CH1-CH2

-
CH2-CH1
X
/
FFT

CH1xCH2
CH1/CH2
CH2/CH1
CH1 or CH2

Comments
Add Channel 1 to Channel 2.
Subtract the Channel 2 waveform from the Channel 1
waveform.
Subtract the Channel 1 waveform from the Channel 2
waveform.
Channel 1 multiply channel 2
Channel 1 divide channel 2
Channel 2 divide channel 1
Three types of window available for selection: Hanning,
Flattop, Rectangular.
Zoom: Use the FFT Zoom button to adjust the window size.
Scale: x1, x2, x5, x10.

Note: All selected menus are highlighted in orange.
DSO5000P Series Digital Storage Oscilloscope User Manual

20

Basic Operation

5.3.1Math FFT
This chapter elaborates how to use the Math FFT (Fast Fourier Transform). You can use the Math
FFT mode to convert a time-domain (YT) signal into its frequency components (spectrum), and to
observe the following types of signals:
Analyze harmonics in power cords;
Measure harmonic content and distortion in systems;
Characterize noise in DC power supplies;
Test impulse response of filters and systems;
Analyze vibration.
To use the Math FFT mode, perform the following tasks:
Set the source (time-domain) waveform;
Display the FFT spectrum;
Choose a type of FFT window;
Adjust the sample rate to display the fundamental frequency and harmonics without aliasing;
Use zoom controls to magnify the spectrum;
Use cursors to measure the spectrum.

5.3.1.1

Setting Time-domain Waveform

It is necessary to set the time-domain (YT) waveform before using the FFT mode. Follow the steps
below.
1. Push the AUTOSET button to display a YT waveform.
2. Turn the VERTICAL POSITION knob to vertically move the YT waveform to the center (zero
division) so as to ensure the FFT will display a true DC value.
3. Turn the HORIZONTAL POSITION knob to position the part of the YT waveform to be
analyzed in the center eight divisions of the screen. The oscilloscope uses the 2048 center
points of the time-domain waveform to calculate the FFT spectrum.
4. Turn the VOLTS/DIV knob to ensure the entire waveform remains on the screen. If the entire
waveform is invisible, the oscilloscope may display wrong FFT results by adding
high-frequency components.
5. Turn the SEC/DIV knob to provide the resolution you need in the FFT spectrum.
6. If possible, set the oscilloscope to display multiple signal cycles.
If you turn the SEC/DIV knob to select a faster setting (fewer cycles), the FFT spectrum will display
a larger frequency range and reduce the possibility of FFT aliasing.
To set the FFT display, follow the steps below.
1. Push the MATH MENU button;
DSO5000P Series Digital Storage Oscilloscope User Manual

21

Basic Operation

2. Set the Operation option to FFT;
3. Select the Math FFT Source channel.
In many situations, the oscilloscope can also generate a useful FFT spectrum despite the YT
waveform not being triggered. This is especially true if the signal is periodic or random (such as
noise).
Note: You should trigger and position transient or burst waveforms as close as possible to
the screen center.
Nyquist Frequency
The highest frequency that any real-time digital oscilloscope can measure without errors is half of
the sample rate, which is called the Nyquist frequency. Frequency information beyond the Nyquist
frequency is undersampled which brings about the FFT aliasing. The math function can convert
the center 2048 points of the time-domain waveform to an FFT spectrum. The resulting FFT
spectrum contains 1024 points from DC (0Hz) to the Nyquist frequency. Usually, the screen
compresses the FFT spectrum horizontally to 250 points, but you can use the FFT Zoom function
to expand the FFT spectrum so that you can clearly view the frequency components at each of the
1024 data points in the FFT spectrum.
Note: The oscilloscope’s vertical response is a little bit larger than its bandwidth (70MHz,
100MHz or 200MHz, depending on the model; or 20MHz when the Bandwidth Limit option is
set to Limited). Therefore, the FFT spectrum can display valid frequency information above
the oscilloscope bandwidth. However, the amplitude information near or above the
bandwidth will not be accurate.

5.3.1.2

Displaying FFT Spectrum

Push the MATH MENU button to display the Math menu. Use the options to select the Source
channel, the Window algorithm and the FFT Zoom factor. Only one FFT spectrum can be
displayed at a time.
Math
Options
Source
Window
FFT Zoom

FFT

Settings

Comments

CH1, CH2
Hanning,
Flattop,
Rectangular

Choose a channel to be the FFT source.
Select a type of the FFT window. For more
information, refer to Section 5.3.1.3.
Change the horizontal magnification of the FFT
display. For detailed information, refer to Section
5.3.1.6.

X1, X2, X5, X10

DSO5000P Series Digital Storage Oscilloscope User Manual

22

Basic Operation

1

Fundamental frequency component
Frequency component

5

2

3

4

1. Frequency at the center graticule line
2. Vertical scale in dB per division (0dB=1VRMS)
3. Horizontal scale in frequency per division
4. Sample rate in number of samples per second
5. FFT window type

5.3.1.3

Selecting FFT Window

Using windows can eliminate the spectral leakage in the FFT spectrum. The FFT algorithm
assumes that the YT waveform repeats all the time. When the number of cycles is integral (1, 2,
3 ...), the YT waveform starts and ends at the same amplitude and there are no discontinuities in
the signal shape.
If the number of cycles is nonintegral, the YT waveform starts and ends at different amplitudes and
transitions between the start and end points will cause discontinuities in the signal that introduces
high-frequency transients.

DSO5000P Series Digital Storage Oscilloscope User Manual

23

Basic Operation

Applying a window to the YT waveform changes the waveform so that the start and stop values
are close to each other, which reduces the discontinuities.

The Math FFT function has three FFT Window options. There is a trade-off between frequency
resolution and amplitude accuracy for each type of window. You shall determine which one to
choose according to the object you want to measure and the source signal characteristics.
Window
Hanning
Flattop
Rectangular

Measurement
Periodic
Waveform
Periodic
Waveform
Pulse or Transient
Waveform

Characteristics
Better frequency, poorer amplitude accuracy than Flattop
Better amplitude, poorer frequency accuracy than Hanning
Special-purpose window applicable to discontinuous
waveforms. This is actually the same as no windows.

DSO5000P Series Digital Storage Oscilloscope User Manual

24

Basic Operation

5.3.1.4

FFT Aliasing

Problems occur when the time-domain waveform acquired by the oscilloscope contains frequency
components higher than the Nyquist frequency. The frequency components above the Nyquist
frequency will be undersampled and displayed as lower frequency components that ‘fold back’
from the Nyquist frequency. These erroneous components are called aliases.

5.3.1.5

Eliminating Aliases

To eliminate aliases, use the following methods.
Turn the SEC/DIV knob to set a faster sample rate. Because the Nyquist frequency increases
as you increase the sample rate, the aliased frequency components will be displayed correct.
If too many frequency components appear on the screen, you may use the FFT Zoom option
to magnify the FFT spectrum.
If there is no need to observe the frequency components above 20MHz, set the Bandwidth
Limit option to Limited.
Filter the signal input from outside and limit the bandwidth of the source waveform to lower
than the Nyquist frequency.
Identify and ignore the aliased frequencies.
Use zoom controls and cursors to magnify and measure the FFT spectrum.

5.3.1.6

Magnifying and Positioning FFT Spectrum

You may scale the FFT spectrum and use cursors to measure it through the FFT Zoom option
which enables the horizontal magnification. To vertically magnify the spectrum, use the vertical
controls.
Horizontal Zoom and Position
You can use the FFT Zoom option to magnify the FFT spectrum horizontally without changing the
sample rate. The available zoom factors are X1(default), X2, X5 and X10. When the zoom factor is
set to X1 and the waveform is located at the center graticule, the left graticule line is at 0Hz and
the right is at the Nyquist frequency.
You magnifies the FFT spectrum to the center graticule line when you change the zoom factor.
That is, the axis for horizontal magnification is the center graticule line. Turn the Horizontal
Position knob clockwise to move the FFT spectrum to the right. Push the SET TO ZERO button to
position the center spectrum at the center of the graticule.
Vertical Zoom and Position
When the FFT spectrum is being displayed, the channel vertical knobs become the zoom and
position controls corresponding to their respective channels. The VOLTS/DIV knob provides the
following zoom factors: X1(default), X2, X5 and X10. The FFT spectrum is magnified vertically to
the marker M (math waveform reference point on the left edge of the screen). Turn the VERTICAL
POSITION knob clockwise to move up the spectrum.
DSO5000P Series Digital Storage Oscilloscope User Manual

25

Basic Operation

5.3.1.7

Using Cursors to Measure FFT Spectrum

You may use cursors to take two measurements on the FFT spectrum: amplitude (in dB) and
frequency (in Hz). Amplitude is referenced to 0db that equals 1VRMS here. You may use cursors
to measure at any zoom factor.
Push the CURSOR button, choose the Source option and then select Math. Press the Type option
button to select between Amplitude and Frequency. Click the SELECT CURSOR option to choose
a cursor. Then use the V0 knobs to move Cursor S and Cursor E. Use the horizontal cursor to
measure the amplitude and the vertical cursor to measure the frequency. Now the display at the
DELTA menu is just the measured value, and the values at Cursor S and Cursor E.
Delta is the absolute value of Cursor S minus Cursor E.

Frequency Cursors

Amplitude Cursors

5.4 Trigger Controls
The trigger can be defined through the Trigger Menu and front-panel controls. There are six types
of trigger: Edge, Video, Pulse Width, Swap, Slope and Overtime. Refer to the following tables to
find a different set of options for each type of trigger.

DSO5000P Series Digital Storage Oscilloscope User Manual

26

Basic Operation

1. Level
It sets the amplitude level the signal must cross to cause an acquisition when using the Edge or
Pulse Width trigger.
2. Set to 50%
The trigger level is set to the vertical midpoint between the peaks of the trigger signal.
3. Force Trigger
Used to complete an acquisition regardless of an adequate trigger signal. This button becomes
useless if the acquisition is already stopped.
4. TRIG MENU
Push this button to display trigger menus. The edge trigger is in common use. See the table below
for details.
Options
Trigger Type
Edge Video
Pulse Slope
Swap Overtime

Settings

Source

CH1
CH2
EXT
EXT/5
AC Line

Mode

Auto
Normal

Coupling

AC
DC
HF Reject
LF Reject

Comments
By default the oscilloscope uses the edge trigger which triggers
the oscilloscope on the rising or falling edge of the
input signal when it crosses the trigger level (threshold).
Select the input source as the trigger signal.
CH1, CH2: No matter the waveform is displayed or not, a
certain channel will be triggered.
EXT: Does not display the trigger signal and allows a trigger
level range of +1.6V to -1.6V.
EXT/5: Same as EXT option, but attenuates the signal by a
factor of 5 and allows a trigger level range of +8V to -8V.
AC Line: Uses a signal derived from the power cord as the
trigger source.
Select a trigger mode.
By default, the oscilloscope uses the Auto mode. In this mode,
the oscilloscope is forced to trigger when it does not detect a
trigger within a certain amount of time based on the SEC/DIV
setting. The oscilloscope goes into the scan mode at 80ms/div
or slower time base settings.
In the Normal mode, the oscilloscope updates the display only
when it detects a valid trigger condition. New waveforms are not
displayed until they replace old ones. Use this mode to just view
valid triggered waveforms. Only after the first trigger does the
display appear.
Select the components of the trigger signal applied to the trigger
circuitry.
AC: Blocks DC components and attenuates signals below
10Hz.
DC: Passes all components of the signal.
HF Reject: Attenuates the high-frequency components above
80kHz.
LF Reject: Blocks DC components and attenuates the
low-frequency components below 8kHz.

NOTE: Trigger coupling only affects the signal passed through the trigger system. It does
not affect the bandwidth or coupling of the signal displayed on the screen.
DSO5000P Series Digital Storage Oscilloscope User Manual

27

Basic Operation

Video Trigger
Options

Settings

Comments
With Video highlighted, an NTSC, PAL or SECAM standard
video signal will be triggered. The trigger coupling is preset to
AC.

Video

Source
Polarity

Sync

Standard

CH1
CH2
EXT
EXT/5
Normal
Inverted
All Lines
Line Number
Odd Field
Even Field
All Fields
NTSC
PAL/SECAM

Select the input source as the trigger signal. Ext and Ext/5 use
the signal applied to the EXT TRIG connector as the source.
Normal: Triggers on the negative edge of the sync pulse.
Inverted: Triggers on the positive edge of the sync pulse.
Choose a proper video sync. When selecting Line Number for
the Sync option, you may use the User Select knob to specify a
line number.
Choose a video standard for sync and line number count.

Note: When you choose Normal Polarity, the trigger always occurs on negative-going sync
pulses. If the video signal contains positive-going sync pulses, use the Inverted Polarity
option.
Pulse Width Trigger
You can use it to trigger on aberrant pulses.
Options

Settings

Comments
With Pulse highlighted, the trigger occurs on pulses that meet
the trigger condition (defined by the Source, When and Set
Pulse Width options).

Pulse

Source

When
Set
Pulse
Width
Polarity
Mode
Coupling
More

CH1
CH2
EXT
EXT5
=
<
>
20ns
10.0sec
Positive
Negative
Auto
Normal
AC
DC
HF Reject
LF Reject

Select the input source as the trigger signal.

Select the trigger condition.
to

With Set Pulse Width highlighted by pressing F4, turn the
multi-functional knob to set the pulse width.
Select to trigger on positive or negative pulses.
Select the type of trigger. The Normal mode is best for most
pulse width trigger applications.
Select the components of the trigger signal applied to the
trigger circuit.
Switch between submenu pages.

Trigger When: The pulse width of the source must be 5ns so that the oscilloscope can detect the
pulse.

DSO5000P Series Digital Storage Oscilloscope User Manual

28

Basic Operation

Triggers when pulse is
less than width setting

Triggers when pulse is
greater than width setting

Threshold level

Triggers when pulse is not
equal to width setting ±5%

Triggers when pulse is
equal to width setting ±5%

Threshold level

Tolerance
= Trigger Point

Tolerance

=, : Within a ±5% tolerance, triggers the oscilloscope when the signal pulse width is equal to
or not equal to the specified pulse width.
<, >: Triggers the oscilloscope when the source signal pulse width is less than or greater
than the specified pulse width.
Slope Trigger: Judges trigger according to the time for rising or falling, more flexible and accurate
than the Edge trigger.
Options

Settings

Comments

Slope

Source

CH1
CH2
EXT
EXT5

Select the input source as the trigger signal.

Slope

Rising
Falling

Select the slope type of signal.

Mode

Auto
Normal

Select the type of trigger. The Normal mode is best for
most pulse width trigger applications.

Coupling

AC
DC
Noise Reject
HF Reject
LF Reject

Selects the components of the trigger signal applied to
the trigger circuitry.

Next Page
Vertical

When

Time

V1
V2
=
<
>
20ns to 10.0sec

Adjust the vertical window by setting two trigger levels.
Select this option and press F3 to choose V1 or V2.
Select the trigger condition.
With this option highlighted by pressing F4, turn the
multi-functional knob to set the time span.

DSO5000P Series Digital Storage Oscilloscope User Manual

29

Basic Operation

Swap Trigger: As a feature of analog oscilloscopes, it gives stable displays of signals at two
different frequencies. Mainly it uses a specific frequency to switch between two analog channels
CH1 and CH2 so that the channels will generate swap trigger signals through the trigger circuitry.
Options

Settings

Comments

Swap
Trigger
Mode

Auto
Normal

Channel

CH1
CH2

Select the type of trigger.
Push an option such as CH1, select the channel trigger
type and set the menu interface.

Below list options in submenus. Swap Trigger allows CH1 and CH2 to select different trigger
modes and to display waveforms on a same screen. That is, both channels can choose the
following four trigger modes.
Type

Edge

Slope

Rising
Falling

Coupling

AC
DC
HF Reject
LF Reject

Type
Polarity
Standard

Sync

Type
Polarity

When

Push F3 or F4 to select the components of the trigger signal
applied to the trigger circuitry.

Video
Normal
Inverted
NTSC
PAL/SECAM
All Lines
Line Number
Odd Field
Even Field
All Fields

Select by F4, F5.

Pulse
Positive
Negative
=
<
>

Select by F3.

Set Pulse
Width

Pulse Width

Push F4 to select. Adjust the multi-functional knob V0 to set
the pulse width.

Coupling

AC
DC
Noise Reject
HF Reject
LF Reject

Select by F5.

DSO5000P Series Digital Storage Oscilloscope User Manual

30

Basic Operation

Type

Slope

Slope

Rising
Falling

Select the slope type of signal.

Mode

Auto
Normal

Select the type of trigger. The Normal mode is best for most
pulse width trigger applications.

Coupling

AC
DC
Noise Reject
HF Reject
LF Reject

Selects the components of the trigger signal applied to the
trigger circuitry.

Vertical

V1
V2

Adjust the vertical window by setting two trigger levels.
Select this option and press F3 to choose V1 or V2.

=
When

Time

<
>
20ns to 10.0sec

Select the trigger condition.

Press F4 to select this option. Turn the multi-functional
knob to set the time span.

Overtime Trigger: In Pulse Width trigger, you may sometimes be puzzled with the long time for
trigger, as you do not need a complete pulse width to trigger the oscilloscope, but want the trigger
occurs just upon the overtime point. This is called Overtime Trigger.
Options

Settings

Comments

Type

OT

Source

CH1
CH2

Polarity

Positive
Negative

Mode

Auto
Normal

Overtime

t

Press F5 to select Overtime option and adjust V0 to set the
time.

Coupling

AC
DC
HF Reject
LF Reject

Selects the components of the trigger signal applied to the
trigger circuitry.

Select the trigger source.
Select to trigger on positive or negative pulses.

Holdoff: To use Trigger Holdoff, push the HORIZONTAL Menu button and set the Holdoff Time
option by pressing F4. The Trigger Holdoff function can be used to generate a stable display of
complex waveforms (such as pulse trains). Holdoff is the time between when the oscilloscope
detects one trigger and when it is ready to detect another. During the holdoff time, the oscilloscope
will not trigger. For a pulse train, the holdoff time can be adjusted to let the oscilloscope trigger only
on the first pulse in the train.

DSO5000P Series Digital Storage Oscilloscope User Manual

31

Basic Operation

Acquisition Interval

Acquisition Interval

Trigger Level
Indicates
Trigger Points
Holdoff

Holdoff

5.5 Menu and Option Buttons
As shown below, these six buttons at the top of the front panel are used mainly to recall relative
setup menus.

SAVE/RECALL: Displays the Save/Recall menu for setups and waveforms.
MEASURE: Displays the Measure menu.
ACQUIRE: Displays the Acquire menu.
UTIILITY: Displays the Utility menu.
CURSOR: Displays the Cursor menu.
DISPLAY: Displays the Display menu.

5.5.1Save/Recall
Press the SAVE/RECALL button to save or recall oscilloscope setups or waveforms.
The first page shows the following menu.
Options

Settings

Comments

Waveforms
Source

CH1
CH2 off
MATH off

REF

RefA
RefB

Select the reference location to store or recall a waveform.

Save

Save the source waveform to the selected reference location.

Operation

Ref on
Ref off

Select a waveform display to store.

Display or remove the reference waveform on the screen.

DSO5000P Series Digital Storage Oscilloscope User Manual

32

Basic Operation

Press ‘Next Page’ to recall the following menu.
Options
Setups

Settings

Comments

Operation
Source

Flash memory
USB disk

Store the current setups to the USB disk or the memory of the
oscilloscope.

Memory

0 to 9
Save

Operation

Recall

Specify the memory location in which to store the current
waveform settings or from which to recall the waveform settings.
Use the V0 knob to select.
Complete the saving operation.
Recall the oscilloscope settings stored in the location selected in
the Setup field. Push the Default Setup button to initialize the
oscilloscope to a known setup.

See below for waveform menus.

The white waveforms on the menu is
the recalled RefA waveform

At most 9 groups of
setups can be stored

Note: The oscilloscope will save the current settings if you wait 5 seconds after the last
modification, and it will recall these settings the next time you power on the oscilloscope.

5.5.2Measure
Push the MEASURE button to perform auto measurements. There are 32 types of measurements
and up to 8 can be displayed at a time.
Turn the V0 knob to select an unspecified option. Press V0 or F6 when the red arrow icon stops on
it. Then the following menu appears.
Options
Source

Settings
CH1
CH2

Comments
Select the measure source.

Measurement Type
1
2
3

Frequency
Period
Mean

4

Pk-Pk

5

CRMS

6

PRMS

Calculate the waveform frequency by measuring the first cycle.
Calculate the time of the first cycle.
Calculate the arithmetic mean voltage over the entire waveform.
Calculate the absolute difference between the greatest and the
smallest peaks of the entire waveform.
Calculate the Root Mean Square voltage over the entire waveform.
Calculate the actual RMS measurement of the first complete cycle in
the waveform.

DSO5000P Series Digital Storage Oscilloscope User Manual

33

Basic Operation

7

Min

8

Max

9

Rising

10

Falling

11

+ Width

12

- Width

13

+ Duty

14

- Duty

15
16
17
18

Base
Top
Middle
Amplitude

19

Overshoot

20

Preshoot

21

PMean

22
23
24

FOVShoot
RPRESoot
BWidth

25

Delay 1-2 ↑

26

Delay 1-2 ↓

27

LFF

28

LFR

29

LRF

30

LRR

31

FFR

32

FRF
Off

The most negative peak voltage measured over the entire
waveform.
The most positive peak voltage measured over the entire waveform.
Measure the time between 10% and 90% of the first rising edge of
the waveform.
Measure the time between 90% and 10% of the first falling edge of
the waveform.
Measure the time between the first rising edge and the next falling
edge at the waveform 50% level.
Measure the time between the first falling edge and the next rising
edge at the waveform 50% level.
Measure the first cycle waveform. Positive Duty Cycle is the ratio
between positive pulse width and period.
Measure the first cycle waveform. Negative Duty Cycle is the ratio
between positive pulse width and period.
Measure the highest voltage over the entire waveform.
Measure the lowest voltage over the entire waveform.
Measure the voltage of the 50% level from base to top.
Voltage between Vtop and Vbase of a waveform.
Defined as (Base - Min)/Amp x 100 %, Measured over the entire
waveform.
Defined as (Max - Top)/Amp x 100 %, Measured over the entire
waveform.
Calculate the arithmetic mean voltage over the first cycle in the
waveform.
Defined as (Vmin-Vlow)/Vamp after the waveform falling.
Defined as (Vmin-Vlow)/Vamp before the waveform falling.
The duration of a burst measured over the entire waveform.
The time between the first rising edge of source 1 and the first rising
edge of source 2.
The time between the first falling edge of source 1 and the first
falling edge of source 2.
The time between the first falling edge of source 1 and the last
falling edge of source 2.
The time between the first falling edge of source 1 and the last rising
edge of source 2.
The time between the first rising edge of source 1 and the last falling
edge of source 2.
The time between the first rising edge of source 1 and the last
rising edge of source 2.
The time between the first falling edge of source 1 and the first
rising edge of source 2.
The time between the first rising edge of source 1 and the first
falling edge of source 2.
Do not take any measurement.

Use the knob V0 or the functional keys F3, F4 to select the type of measurement.

DSO5000P Series Digital Storage Oscilloscope User Manual

34

Basic Operation

The readouts in big font size on
the menu are just results of the
corresponding measurements.
Taking Measurements: For a single waveform (or a waveform divided among multiple
waveforms), up to 8 automatic measurements can be displayed at a time. The waveform channel
must stay in an ‘ON’ (displayed) state to facilitate the measurement. The automatic measurement
can not be performed on reference or math waveforms, or in XY or Scan mode.

5.5.3Acquire
Push the ACQUIRE button to set the acquisition parameter.
Options
Category

Settings
Real Time
Equ-Time

Comments
Acquire waveforms by real-time digital technique.
Rebuild waveforms by equivalent sample technique.
Acquire and accurately display most waveforms.
Detect glitches and eliminate the possibility of aliasing.
Reduce random or uncorrelated noise in signal display.
The number of averages is selectable.

Mode
(Real Time)

Normal
Peak Detect
Average

Averages
(Real Time)

4
16
64
128

Select the number of averages by pressing F3 or F4.

Memory Depth
(Real Time)

4K, 20K, 40K

Select the memory depth for different board models.

Normal: For the oscilloscope model with the bandwidth of 100MHz, the maximum sample rate is
1GS/s. For time base with insufficient sample rate, you may use the Sine Interpolation Algorithm to
interpolate points between sampled points to produce a complete waveform record (4K by
default).
DSO5000P Series Digital Storage Oscilloscope User Manual

35

Basic Operation

Normal Acquisition Intervals

1

2

3

4

5

6

7

8

9

10

Sample Points
Normal Mode Acquires a Single Sample Point in Each Interval

Peak Detect: Use this mode to detect glitches within 10ns and to limit the possibility of aliasing.
This mode is valid at the SEC/DIV setting of 4µs/div or slower. Once you set the SEC/DIV setting
to 4µs/div or faster, the acquisition mode will change to Normal because the sample rate is fast
enough that Peak Detect is unnecessary. The oscilloscope does not display a message to tell you
that the mode has been changed to Normal.
Average: Use this mode to reduce random or uncorrelated noise in the signal to be displayed.
Acquire data in Normal mode and then average a great number of waveforms. Choose the
number of acquisitions (4, 16, 64 or 128) to average for the waveform.
Stopping the Acquisition: When you are running the acquisition, the waveform display is live.
Stop the acquisition (press the RUN/STOP button) to freeze the display. In either mode, the
waveform display can be scaled or positioned by vertical and horizontal controls.
Equivalent Acquisition: Just repeat the Normal acquisition. Use this mode to take a specific
observation on repeatedly displayed periodic signals. You can get a resolution of 40ps, i.e.
25GSa/s sample rate, which is much higher than that obtained in real-time acquisition.
The acquisition principle is as follows.

Input repeated signals
First Acquisition
Second Acquisition
Third Acquisition
Fourth Acquisition

As shown above, acquire input signals (cycle repeatable) for more than once at a slow sample rate,
arrange the sample points by the time they appear, then recover waveforms.

DSO5000P Series Digital Storage Oscilloscope User Manual

36

Basic Operation

5.5.4Utility
Push the UTILITY button to display the Utility Menu as follows.
Options

Comments

System Info

Display the software and hardware versions, serial number and some other
information about the oscilloscope.

Update
Program

Insert a USB disk with upgrade program and the disk icon at the top left corner is
highlighted. Press the Update Program button and the Software Upgrade dialog
pops up. Push F6 to upgrade or push F2 to cancel.

Save
Waveform

Insert a USB disk and the disk icon at the top left corner is highlighted. Click on
this icon and you can see the waveform pause for a while, being saved. You can
find the saved waveform data under the pic_x folder in the USB disk. Here X
represents how many times you press the button. Each press generates a
corresponding folder. For example, press once and a folder pic_1 is generated;
press twice and two folders pic_1, pic_2 are generated.

Self
Calibration

Press this option and the Self Calibration dialog pops up. Press F6 to perform
the self calibration. Press F4 to cancel.

Advance

Buzzer and time setups
Press this button to see the dialog for buzzer and time setups. Turn the knob V0
to select buzzer or time (with a black frame). Press V0 and the black frame turn
to be red. Turn V0 again to set ON/OFF or time. Press V0 again to exit or save
the setups.

Self Calibration: The self calibration routine can optimize the precision of the oscilloscope to fit
the ambient temperature. To maximize the precision, you should perform the self calibration once
the ambient temperature changes by 5℃ or more. Follow the instructions on the screen.
Tip: Press any menu button on the front panel to remove the status display and enter a
corresponding menu.

5.5.5Cursor
Push the CURSOR button to display the Cursor Menu.
Options

Settings

Comments
Select a measurement cursor and display it.
Voltage measures amplitude while Time measures time and
frequency.

Type

Voltage
Time

Source

CH1, CH2
MATH
REFA, REFB

Select a waveform to take the cursor measurement.
Use the readouts to show the measurement.

Select
Cursor

S
E

S indicates Cursor 1. E indicates Cursor 2.
A selected cursor is highlighted, which can be moved freely.
Both cursors can be selected and moved at the same time. The
box behind the cursor displays the location of the cursor.

Delta

Display
the
difference
(delta) between
the cursors.

Display the measurement in the box under this option.

DSO5000P Series Digital Storage Oscilloscope User Manual

37

Basic Operation

Moving Cursors: Press the key near Select Cursor to select a cursor and turn V0 to move it.
Cursors can be moved only when the Cursor Menu is displayed.

Voltage Cursor

Time Cursor

5.5.6Display
The waveform display is affected by settings of the oscilloscope. A waveform can be measured
once it is captured. The different styles to display a waveform on the screen give significant
information about it.
There are two modes to display waveforms Single-window and Dual-window. Refer to Horizontal
Controls for more information.
Push the DISPLAY button and the following menu appears.
Options

Settings

Type

Vectors
Dots

Persist

OFF
0.2S-8S selectable
Infinite

Format

YT
XY

Contrast

Grid
Grid Intensity

Dotted line
Real line
OFF

Comments
Vectors fills up the space between adjacent sample
points in the display; Dots only displays the sample
points.
Sets the time length to display each displayed sample
point.
YT format shows the vertical voltage in relation to time
(horizontal scale); XY format displays a dot between
CH1 and CH2 each time when a sample is acquired,
where the voltage or current of CH1 determines the X
coordinate of the dot (horizontal) and the voltage or
current of CH2 determines the Y coordinate (vertical).
For detailed information, refer to descriptions on XY
format in the following text.
0-15 16 ranks adjustable, with a progress bar to
display; Press F5 to select this option. Turn the
multi-functional knob to adjust.
Next Page
Off only displays the horizontal and vertical
coordinates at the center graticule on the screen.
0-15 16 ranks adjustable, with a progress bar to
display.

DSO5000P Series Digital Storage Oscilloscope User Manual

38

Basic Operation

5.6 Fast Action Buttons

AUTOSET: Automatically set the oscilloscope controls to generate a usable display of the input
signals. Refer to the following table for relative content.
SINGLE SEQ: Acquire a single waveform and then stop the acquisition.
RUN/STOP: Continuously acquire waveforms or stop the acquisition.
HELP: Display the Help menu.
DEFAULT SETUP: Recall the factory setup. Refer to the following table for relative content.
Save to USB Disk: Save all displays on the screen to a USB device, just like the screen capturing
function of a computer.

5.6.1Autoset
Autoset is one of the advantages digital oscilloscopes have. When you push the AUTOSET button,
the oscilloscope will identify the type of waveform (sine or square wave) and adjust controls
according to input signals so that it can accurately display the waveform of the input signal.
Functions
Acquire Mode
Cursor

Settings
Adjusted to Normal or Peak Detect
Off

Display Format

Set to YT

Display Type
Horizontal Position
SEC/DIV
Trigger Coupling
Trigger Holdoff
Trigger Level
Trigger Mode
Trigger Source
Trigger Slope
Trigger Type
Trigger Video Sync
Trigger Video Standard
Vertical Bandwidth

Set to Vectors for an FFT spectrum; otherwise, unchanged
Adjusted
Adjusted
Adjusted to DC, Noise Reject or HF Reject
Minimum
Set to 50%
Auto
Adjusted; Autoset can not be used for the EXT TRIG signal
Adjusted
Edge
Adjusted
Adjusted
Full
DC (if GND was chosen before); AC for the video signal; otherwise,
Vertical Coupling
unchanged
VOLTS/DIV
Adjusted
The Autoset function examines all channels for signals and displays corresponding waveforms.
Autoset determines the trigger source according to the following conditions.
DSO5000P Series Digital Storage Oscilloscope User Manual

39

Basic Operation

If multiply channels get signals, the oscilloscope will use the channel with the lowest
frequency signal as the trigger source.
If no signals are found, the oscilloscope will use the lowest-numbered channel displayed
in Autoset as the trigger source.
If no signals are found and no channels are displayed, the oscilloscope will display and
use Channel 1 as the trigger source.
Sine Wave:
When you use the Autoset function and the oscilloscope determines that the signal is similar to a
sine wave, the oscilloscope displays the following options.
Sine Wave Options
Multi-cycle Sine
Single-cycle Sine
FFT
Undo Setup

Details
Display multiple cycles that have appropriate vertical and horizontal
scales.
Set the horizontal scale to display about one cycle of the waveform.
Convert the input time-domain signal to its frequency components
and display the result as a graph of frequency versus amplitude
(spectrum). Since it is a math calculation, see Section 5.3.1 Math FFT
for more information.
Let the oscilloscope recall the previous setup.

Square Wave or Pulse:
When you use the Autoset function and the oscilloscope determines that the signal is similar to a
square wave or pulse, the oscilloscope displays the following options.
Square Wave Options
Multi-cycle Square
Single-cycle Square
Rising Edge
Falling Edge
Undo Setup

Details
Display multiple cycles that have appropriate vertical and
horizontal scales.
Set the horizontal scale to display about one cycle of the
waveform. The oscilloscope displays Min, Mean and Positive
Width automatic measurements.
Display the rising edge.
Display the falling edge.
Let the oscilloscope recall the previous setup.

5.6.2Help
Push the HELP button to display the Help menu which provides topics covering all the menu
options and controls of the oscilloscope. For detailed information about the Help system, refer
to Section 2.2.

5.6.3Default Setup
When you push the DEFAULT SETUP button, the oscilloscope will display the CH1 waveform and
remove all the others. When you are at the default setup, press F1 to Undo Preset. Then the
oscilloscope returns to the status before default setup. The table below gives the options, buttons
and controls that change settings at default setup.
DSO5000P Series Digital Storage Oscilloscope User Manual

40

Basic Operation

Menu or System
Acquire

Cursor

Display

Horizontal

Math

Option, Button or Knob
(Three mode options)

Default Setting
Normal

Averages

16

Run/Stop

Run

Type

Off

Source

CH1

Horizontal (amplitude)

±3.2div

Vertical (time)

±4div

Type

Vectors

Persist

Off

Format

YT

Window Mode

Single-window

Trigger Knob

Level

Position

0.00s

SEC/DIV

200 s

Operation

—

Source

CH1-CH2

Position

0div

Vertical Scale

20dB

FFT Operation
FFT

Measure

Trigger (Edge)

Trigger (Video)

Trigger (Pulse)

Trigger (Slope)

Source

CH1

Window

Hanning

FFT Zoom

X1

Source

CH1

Type

None

Type

Edge

Source

CH1

Slope

Rising

Mode

Auto

Coupling

DC

Level

0.00v

Polarity

Normal

Sync

All lines

Standard

NTSC

When

=

Set Pulse Width

1.00ms

Polarity

Positive

Mode

Auto

Coupling

DC

Slope

Rising

Mode

Auto

Coupling

DC

When

=

DSO5000P Series Digital Storage Oscilloscope User Manual

41

Basic Operation

CH1

Trigger (Swap)

Trigger (OT)

Vertical System,
All Channels

Type

Edge

Slope

Rising

Mode

Auto

Coupling

DC

Level

0.00v

CH2
Type

Edge

Slope

Rising

Mode

Auto

Coupling

DC

Level

0.00v

Source

CH1

Polarity

Positive

Mode

Auto

Time

20ns

Coupling

DC

Bandwidth Limit

Unlimited

VOLTS/DIV

Coarse

Probe

Voltage

Voltage Probe Attenuation

10X

Invert

Off

Position

0.00div (0.00V)

VOLTS/DIV

1.00V

The following settings do not change when you push the DEFAULT SETUP button.
Language Option
Saved Settings
Saved Reference Waveforms
Display Contrast
Calibration Data

5.7 Multi-functional Knobs and Buttons
V0: Multi-functional knob. Under different menu options, it supports selecting menu
options (MEASURE), moving cursors and levels (Slope Trigger).
Press this knob to reset data (trigger holdoff, overtime of the overtime trigger and
slope trigger), select menu
options and so on. Easy to operate.
F7: Push this button in single-window mode to switch between dotted line display
and cross display. Push it in dual-window mode to perform autocruise.

DSO5000P Series Digital Storage Oscilloscope User Manual

42

Basic Operation

F0: Hide/Show button. Push it to hide the menu options on the right side of the
screen and give a full screen display of waveforms. Push it again to show the menu
options.
F1-F5: These five buttons are all multi-functional. They are in charge of selecting
corresponding menu options on the screen in different menu modes. For example, in
the UTILITY menu, F1-F5 respectively correspond to ‘System Info’ – ‘Advance’.

F6: This functional button is mainly used to turn pages and confirm a selection, such
as ‘next page’, ‘previous page’, and ‘press F6 to confirm’ appearing when you push
Self Calibration option.

5.8 Signal Connectors
See the figure below to find the three signals connectors and a pair of metal electrodes at the
bottom of the oscilloscope panel.

1. CH1, CH2: Input connectors for waveform display, through which to connect and input the
signal to be measured.
2. EXT TRIG: Input connector for an external trigger source, though with to connect and input the
external trigger signal.
3. Probe Compensation: Voltage probe compensation output and ground, used to electrically
match the probe to the oscilloscope input circuit. The probe compensation ground and BNC
shields connect to earth ground and are considered to be ground terminals. To avoid damages,
do not connect a voltage source to any of these ground terminals.

DSO5000P Series Digital Storage Oscilloscope User Manual

43

Application Examples

Chapter 6 Application Examples
This chapter unfolds a further description on main features of the oscilloscope by giving eleven
simplified application examples for reference to help solve your own test problems.
1.

Taking simple measurements
Using AUTOSET
Using the Measure menu to take auto measurements

2.

Taking cursor measurements
Measuring ring frequency and ring amplitude
Measuring pulse width
Measuring rise time

3.

Analyzing input signals to eliminate random noise
Observing a noisy signal
Eliminating random noise

4.

Capturing a single-shot signal

5.

Using X-Y mode

6.

Triggering on a pulse width

7.

Triggering on a video signal
Observing triggers on video fields and video lines

8.

Using Slope Trigger to capture a particular slope signal

9.

Using Overtime Trigger to measure a long pulse signal

10. Using math functions to analyze waveforms
11. Measuring data propagation delay

6.1 Example 1: Taking Simple Measurements
When you want to observe an unknown signal in a certain circuit without having its amplitude and
frequency parameters, you may use this function to take a fast measurement on the frequency,
period and peak-to-peak amplitude of the signal.
Follow the steps below.
1. Set the switch on the oscilloscope probe to 10X;
2. Push the CH1 MENU button and set the Probe option attenuation to 10X;
3. Connect the CH1 probe to the test point of the circuit;
4. Press the AUTOSET button.
The oscilloscope will automatically set the waveform to a best display. If you want to further
optimize the waveform display, you may manually adjust the vertical and horizontal controls until
the waveform meets your particular requirement.

DSO5000P Series Digital Storage Oscilloscope User Manual

44

Application Examples

Taking Auto Measurements
The oscilloscope can display most signals by automatic measurements. To measure such
parameters as the signal frequency, period, peak-to-peak amplitude, rise time and positive width,
follow the steps below.
1. Push the MEASURE button to see the Measure menu.
2. Turn the knob V0 to select the first ‘unspecified’ option (marked by red arrow), press V0 or F6
to enter the submenu.
3. Select CH1 for the Source option. Then repeatedly push F3 or F4 to select measure items in
the Type menu. Push the back menu to go back to the measure interface. Or turn and push
V0 to select a measure item and go back to the measure interface. The corresponding box
under the measure item shows the measurements.
4. Repeat Step 2 and Step 3. Then select other measure items. Totally 8 measure items can be
displayed.
Note: All readouts change with the measured signals.
The figure below shows three measure items as an example. The boxes under them display the
measurements in large fonts.

DSO5000P Series Digital Storage Oscilloscope User Manual

45

Application Examples

6.2 Example 2: Taking Cursor Measurements
You can use the cursor to quickly measure the time and amplitude of a waveform.
Measuring Ring Time (convertible to Frequency) and Amplitude on Rising Edge of Pulse
To measure the ring time on the rising edge of the pulse, follow the steps below.
1.

Push the CURSOR button to view the Cursor menu.

2.

Push F1 the Type option button and select Time.

3.

Push F2 or F3 the Source option button and select CH1.

4.

Push F4 to select a cursor. If S is selected, turn V0 to move Cursor S on the screen; if E is
selected, turn V0 to move Cursor E; if both are selected, turn V0 to move them at the same
time.

5.

Put Cursor S on the first peak of the ring.

6.

Put Cursor E on the second peak of the ring.

7.

At Delta displays the measured time and at Cursor S an Cursor E display the positions of
these two cursors.

8.

Push the Type option button and select Voltage.

9.

Put Cursor S on the highest peak of the ring.

10. Put Cursor E on the lowest point of the ring. The amplitude of the ring will be displayed at
Delta.
See figures below for better understanding.

DSO5000P Series Digital Storage Oscilloscope User Manual

46

Application Examples

Measuring Pulse Width
To analyze a pulse signal and to know its width, follow the steps below.
1.

Push the CURSOR button to view the Cursor menu.

2.

Push F1 the Type option button and select Time.

3.

Push F2 or F3 the Source option button and select CH1.

4.

Push F4 to select a cursor. If S is selected, turn V0 to move Cursor S on the screen; if E is
selected, turn V0 to move Cursor E; if both are selected, turn V0 to move them at the same
time.

5.

Place Cursor S on the rising edge of the pulse and Cursor E on the falling edge.

6.

Thus at Delta displays the measured time and at Cursor S and Cursor E displays the time
relative to the trigger.

See the figure below for better understanding.

DSO5000P Series Digital Storage Oscilloscope User Manual

47

Application Examples

Measuring Rise Time of Pulse
You may need to measure the rise time of the pulse in many application environments, usually, to
measure the rise time between the 10% and 90% levels of the pulse waveform. To do so, follow
the steps below.
1.

Turn the SEC/DIV knob to display the rising edge of the waveform.

2.

Turn the VOLTS/DIV and VERTICAL POSITION knobs to adjust the waveform amplitude to
about 5 divisions.

3.

Push the CH1 MENU button.

4.

Push the VOLTS/DIV option button and select Fine. Turn the VERTICAL POSITION knob to
accurately separate the waveform into 5 divisions.

5.

Turn the VERTICAL POSITION knob to center the waveform. Position the waveform baseline
to 2.5 divisions below the center graticule.

6.

Press the CURSOR button.

7.

Push the Type option button and select Time. Push the Source option button to select CH1.

8.

Select Cursor S and turn V0 to place it at the 10% level of the waveform.

9.

Select Cursor E and turn V0 to place it at the 90% level of the waveform.

10. The Delta readout in the Cursor Menu is the rise time of the pulse.
See the figure below for better understanding.

5 divisions

DSO5000P Series Digital Storage Oscilloscope User Manual

48

Application Examples

6.3 Example 3: Analyzing Input Signals to Eliminate
Random Noise
In certain circumstances, to display a noisy signal on the oscilloscope and to get its details, you
may follow the steps below to analyze this signal.

Observing Noisy Signal
1.

Press the ACQUIRE button to see the Acquire menu.

2.

Push the Type option button and select Real Time.

3.

Push the Peak Detect option button.

4.

If necessary, push the DISPLAY button and set the Contrast option to view the noise more
clearly.

See the figure below for better understanding.

DSO5000P Series Digital Storage Oscilloscope User Manual

49

Application Examples

Eliminating Random Noise
1.

Press the ACQUIRE button to see the Acquire menu.

2.

Push the Type option button and select Real Time.

3.

Push the Average option button.

4.

Push the Averages option button and adjust the number of running averages to watch the
change in the waveform display.

Note: Averaging reduces random noise and let you view the signal details more easily.
See the figure below for better understanding

6.4 Example 4: Capturing Single-shot Signal
You may refer to the following example to easily capture some aperiodic signals like pulses and
glitches.
To set for a single-shot acquisition, follow the steps below.
1.

First, set up the oscilloscope probe and the attenuation factor of CH1.

2.

Turn the vertical VOLTS/DIV and horizontal SEC/DIV knobs to a proper position for a better
examination of the signal.

3.

Press the ACQUIRE button to see the Acquire menu.

4.

Push the Peak Detect option button.

5.

Push the TRIG MENU button and select Rising for the Slope option. Then adjust the trigger
level properly.

6.

Push the SINGLE SEQ button to start the acquisition.

DSO5000P Series Digital Storage Oscilloscope User Manual

50

Application Examples

Using this feature can help you to capture occasional events more easily. This is an advantage of
the digital storage oscilloscope.

6.5 Example 5: Using X-Y Mode
Viewing Phase Differences between Two Channel Signals
For example, you need to measure the change in a phase across a circuit network.
Connect the oscilloscope with circuitry and view the input and output of the circuit in XY mode.
Follow the steps below.
1.

First, prepare two oscilloscope probes and set the switches to 10X on both probes.

2.

Push the CH1 MENU button and set the Probe option attenuation to 10X; push the CH2
MENU button and set the Probe option attenuation to 10X.

3.

Connect the CH1 probe to the input of the network, and connect the CH2 probe to the output.

4.

Push the AUTOSET button.

5.

Turn the VOLTS/DIV knobs to display approximately the same amplitude signals on each
channel.

6.

Push the DISPLAY button to see the Display menu.

7.

Push the Format option button and select XY.

8.

Now the oscilloscope displays a Lissajous pattern to characterize the input and output of the
circuit.

9.

Turn the VOLTS/DIV and VERTICAL POSITION knobs to properly scale the waveform
display.

10. Use the Lissajous's oscillographic method to observe and calculate the phase differences
following the formula below.
DSO5000P Series Digital Storage Oscilloscope User Manual

51

Application Examples

As sin =A/B or C/D, in which is the phase difference angle between channels and A, B, C, D
represent what shown in the figure below, you can get the value of the phase difference angle by
the formula: =±arcsin(A/B) or ±arcsin(C/D).
If the principal axes of the ellipse are in the first and third quadrants, the phase difference angle
should be in the first and fourth quadrants, i.e. within (0~ /2) or (3 /2~2 ). If the principal axes of
the ellipse are in the second and fourth quadrants, the phase difference angle should be in the
second and third quadrants, i.e. within ( /2~ ) or ( -3 /2). See the figure below for better
understanding.
Signal Horizontal
Centering

A

D

B

C

6.6 Example 6: Triggering on Pulse Width
Triggering on a Specific Pulse Width
While testing the pulse width of a signal in a circuit, you may need to verify the pulse width is
consistent with the theoretic value. Or even if the edge triggering shows that your signal has the
same pulse width with the specific signal, you still doubt about the result. Then you can follow the
steps below.
1.

Set the Probe option attenuation to 10X.

2.

Push the AUTOSET button to trigger a stable waveform display.

3.

Push the Single Cycle option button in the Autoset menu and read out the signal pulse width.

4.

Push the TRIG MENU button.

5.

Push F1 to select Pulse for the Type option; push F2 to select CH1 for the Source option; turn
the TRIGGER LEVEL knob to set the trigger level at the bottom of the signal.

6.

Push F6 to enter the next page. Select the When option button and Push F4 to select ‘=’.

7.

Push the Set Pulse Width option button. Turn V0 to set the pulse width to the value read out in
Step 3.

DSO5000P Series Digital Storage Oscilloscope User Manual

52

Application Examples

8.

Turn the TRIGGER LEVEL knob to set the pulse width to the value read out in Step 3.

9.

Push the More option button and select Normal for the Mode option. Once triggering on
normal pulses, the oscilloscope can give a stable waveform display.

10. If the When option is set to >, < or and there appear any aberrant pulses that meet the
specified condition, the oscilloscope will trigger. For example, the signal contains such
aberrant pulses as shown below, you may select ‘ ’ or ‘<’ to trigger on the pulse.

As shown in the above figure, you can get a stable waveform display if inputting a square wave at
the frequency of 1KHz, with pulse width set to 500 s.

6.7 Example 7: Triggering on Video Signal
Assume that you are monitoring the video signals of a television to see if they are input normally,
and the video signal is of an NTSC system. You can get a stable display by using the video trigger.
Triggering on Video Fields
To trigger on the video fields, follow the steps below.
1.

Push the TRIG MENU button to see the Trigger menu.

2.

Push F1 to select Video for the Type option.

3.

Push the Source option button to select CH1; push the Polarity option button to select Normal;
push the Standard option button to select NTSC.

4.

Push the Sync option button to select Odd Field, Even Field or All Fields.

5.

Turn the Trigger Level knob to adjust the trigger level and stabilize video signals.

6.

Turn the horizontal SEC/DIV and the Vertical Position knobs to display on the screen a
complete video signal triggering on a video field.

The figure below shows a stable signal triggering on a video field.
DSO5000P Series Digital Storage Oscilloscope User Manual

53

Application Examples

Triggering on Video Lines
To trigger on the video lines, follow the steps below.
1.

Push the TRIG MENU button to see the Trigger menu.

2.

Push F1 to select Video for the Type option.

3.

Push the Source option button to select CH1; push the Polarity option button to select Normal;
push the Standard option button to select NTSC; push the Sync option button to select Line
Number.

4.

Turn the Trigger Level knob to adjust the trigger level and stabilize video signals.

5.

Turn V0 to adjust the line number (NTSC: 0-525 lines).

6.

Turn the horizontal SEC/DIV and the vertical VOLTS/DIV knobs to display on the screen a
complete video signal triggering on a video line. See the figure below.

DSO5000P Series Digital Storage Oscilloscope User Manual

54

Application Examples

6.8 Example 8: Using Slope Trigger to Capture Particular
Slope Signal
In many occasions, we are not only concerned about the edge of the signal, but also want to know
the rise and fall times of the signal. To better observe this kind of signals, we bring in the slope
trigger. Follow the steps below.
1.

Push the TRIG MENU button to see the Trigger menu.

2.

Push F1 to select Slope for the Type option.

3.

Push the Source option button to select CH1; push the Slope option button to select Rising;
push the Mode option button to select Auto; push the Coupling option button to select DC.

4.

Click the ‘Next Page’ button and select Vertical. Turn the V0 knob to adjust V1 and V2 to
proper locations. Select the When option button and set it to ‘=’.

5.

Select ‘Time’ and turn V0 to adjust the time until you get a stable display of waveforms. See
the figure below.

6.9 Example 9: Using Overtime Trigger to Measure Long
Pulse Signal
It is not easy to observe some part of a long pulse signal by using the edge or pulse width trigger.
In such case, you can use the overtime trigger by following steps.
1.

Push the TRIG MENU button to see the Trigger menu.

2.

Push F1 to select OT for the Type option; push the Polarity option button to select Normal;
push the Mode option button to select Auto; push the Coupling option button to select DC.

DSO5000P Series Digital Storage Oscilloscope User Manual

55

Application Examples

3.

Turn the Trigger Level knob to adjust the trigger level and stabilize video signals.

4.

Turn V0 to adjust the line number (NTSC: 0-525 lines).

5.

Turn the horizontal SEC/DIV and the vertical VOLTS/DIV knobs to display on the screen a
complete video signal triggering on a video line. See the figure below.

Note: The difference between the overtime and the delay triggers is that the overtime
trigger can identify the pulse you need according to your set time and trigger on any point
of the pulse. In the other word, the overtime trigger occurs based on pulse identification. It
is similar to the > mode of the pulse width trigger, but not the same.

6.10 Example 10: Using Math Functions to Analyze
Waveforms
Using math functions to analyze input waveforms is another advantage of the digital oscilloscope.
For example, you want to get the instantaneous difference between two channel waveforms. By
using the math function of the oscilloscope, you can get a better representation of the waveform
on the screen. To observe this signal, follow the steps below.
1.

Set the Probe option attenuation to 10X.

2.

Open CH1 and CH2 at the same time, both with the attenuation of 10X.

3.

Push the AUTOSET button to trigger a stable waveform.

4.

Push the MATH MENU button to see the Math menu.

5.

Push the Operation option button and select ‘CH1+CH2’.

6.

Turn the horizontal SEC/DIV and the vertical VOLTS/DIV knobs to properly scale the
waveform for easy check.

In addition, the oscilloscope also supports the - and FFT functions. For a detailed analysis on FFT,
DSO5000P Series Digital Storage Oscilloscope User Manual

56

Application Examples

refer to Chapter 5.3.1 Math FFT.
Note: You should compensate both probes before performing the math operation;
otherwise, differences in probe compensation will appear as errors in the differential
signal.

As illustrated in the above figure, input a 1KHz sine wave from CH1 and a 1KHz square wave from
CH2.
Follow the above steps to set up the Math menu, and observe the subtracted waveform as shown
in the figure below.

Those in pink are added waveforms.

6.11 Example 11: Measuring Data Propagation Delay
When you doubt that there appear instabilities in a serial data propagation control circuit, you can
set the oscilloscope to measure the propagation delay between the enable signal and the transfer
data.
DSO5000P Series Digital Storage Oscilloscope User Manual

57

Application Examples

To set the propagation delay measurement, follow the steps below.
1.

Connect two oscilloscope probes respectively to the CS (chip-select) pin and the DATA pin on
the chip.

2.

Set the Probe option attenuation to 10X for both probes.

3.

Open CH1 and CH2 at the same time, both with the attenuation of 10X.

4.

Push the AUTOSET button to trigger a stable waveform display.

5.

Adjust the horizontal and vertical controls to optimize the waveform display.

6.

Push the CURSOR button to view the Cursor menu.

7.

Push the Type option button and select Time.

8.

Select Cursor S and turn V0 to place it on the active edge of the enable signal.

9.

Select Cursor E and turn V0 to place it on the data output transition (See the figure below).

10. Read the data propagation delay in the Delta readout.

CH2

CH1

DATA

CS

CS

DATA

DSO5000P Series Digital Storage Oscilloscope User Manual

58

Troubleshooting

Chapter 7 Troubleshooting
7.1 Problem Settlement
1. If the oscilloscope does not start up at power on, follow these steps:
1) Check the power cord to verify it has been connected properly;
2) Check the power on/off button to ensure it has been pushed;
3) Then restart the oscilloscope.
2. If there is no display of waveforms on the screen when the oscilloscope is turned on,
follow these steps:
1) Check the probe to assure its proper connection to the input BNC;
2) Check the channel switch (such as CH1, CH2 menu buttons) to make sure it has been turned
on;
3) Check the input signal to verify it has been connected to the probe correctly;
4) Affirm that all measured circuits have signals to output;
5) Turn up the magnitude for DC signals with large magnitude;
6) In addition, you may press the Auto Measure button to perform an automatic detection of
signals at first.
3. If the waveform of the input signal is distorted seriously, follow these steps:
1) Check the probe to assure its proper connection to the channel BNC;
2) Check the probe to assure its good connection to the measured object;
3) Check the probe to verify it has been well calibrated. Otherwise, refer to the content about
calibration described in this manual.
4. If the waveform is rolling continuously on the screen but can not be triggered, follow
these steps:
1) Check the trigger source to make sure it consistent with the input channel;
2) Check the trigger level to assure its correct adjustment. You may push the TRIGGER LEVEL
knob or press the SET TO 50% button to reset the trigger level back to the center of the signal;
3) Check the trigger mode to confirm it is a right choice for the input signal. The default trigger
mode is edge trigger. However, it is not suitable for all kinds of input signals.

DSO5000P Series Digital Storage Oscilloscope User Manual

59

Specifications

Chapter 8 Specifications
8.1 Technical Specifications
All specifications herein mentioned apply to the DSO5000P series oscilloscopes. Before checking
an oscilloscope from your seller to see if it complies with these specifications, make sure it meets
the following conditions:
The oscilloscope must have been operating continuously for twenty minutes under the
specified operating temperature.
The Do Self Cal operation must be performed through the Utility menu if the operating
temperature changes by more than 5℃.
The oscilloscope must be within the factory calibration interval.
All specifications are guaranteed unless noted ‘typical’.

Oscilloscope Specifications
Horizontal
Sample Rate Range

1GS/s

Waveform Interpolation

(sin x)/x

Record Length

Maximum 40K samples per single-channel; maximum 20K
samples per dual-channel (4K, 20K optional)
DSO5072P DSO5102P

SEC/DIV Range
Sample Rate and
Delay Time Accuracy
Delta Time Measurement
Accuracy
(Full Bandwidth)

4ns/div to 40s/div, in a
sequence

DSO5202P
2, 4, 8

2ns/div to 40s/div, in a 2, 4,
8 sequence

±50ppm over any 1ms time interval
Single-shot, Normal mode
± (1 sample interval +100ppm × reading + 0.6ns)
>16 averages
± (1 sample interval + 100ppm × reading + 0.4ns)
Sample interval = s/div ÷ 200
DSO5072P DSO5102P

Position Range

20ns/div to 80 s/div

(-8div × s/div) to 40ms

200 s/div to 40s/div

(-8div × s/div) to 400s

DSO5202P
2ns/div to 10ns/div

DSO5000P Series Digital Storage Oscilloscope User Manual

(-4div × s/div) to 20ms

60

Specifications

Vertical
A/D Converter

8-bit resolution,
each channel sampled simultaneously

VOLTS/DIV Range

2mV/div to 5V/div at input BNC

Position Range

2mV/div to 200mV/div, ±2V
>200mV/div to 5V/div, ±50V

Analog Bandwidth in
Normal and Average
modes at BNC or with probe, DC
Coupled

2mV/div to 20mV/div, ±400mV
50mV/div to 200mV/div, ±2V
500mV/div to 2V/div, ±40V
5V/div, ±50V

Selectable Analog Bandwidth Limit,
typical

20MHz

Low Frequency Response (-3db)
Rise Time at BNC, typical

10Hz at BNC
DSO5072P

DSO5102P

DSO5202P

≤5.0ns

<3.5ns

<1.8ns

±3% for Normal or Average acquisition mode, 5V/div to
10mV/div
±4% for Normal or Average acquisition mode, 5mV/div
to 2mV/div

DC Gain Accuracy

Measurement Type: Average of 16 waveforms with
vertical position at zero
Accuracy: ± (3% × reading + 0.1div + 1mV) when
10mV/div or greater is selected
DC Measurement Accuracy,
Average Acquisition Mode

Volts Measurement Repeatability,
Average Acquisition Mode

Measurement Type: Average of 16 waveforms with
vertical position not at zero
Accuracy: ± [3% × (reading + vertical position) + 1% of
vertical position + 0.2div]
Add 2mV for settings from 2mV/div to 200mV/div; add
50mV for settings from 200mV/div to 5V/div
Delta volts between any two averages of 16
waveforms acquired under same setup and ambient
conditions

Note: Bandwidth reduced to 6MHz when using a 1X probe.
Trigger
Coupling

Sensitivity
Source

Trigger Sensitivity
(Edge Trigger Type)

CH1
CH2
DC
EXT

DSO5072P, DSO5102P

DSO5202P

1div from DC to
10MHz;
1.5div from 10MHz to
Full

1.5div from 10MHz
to 100MHz;
2div from 100MHz
to Full

200mV from DC to
100MHz

200mV from DC to
100MHz; 350mV
from 100MHz to
200MHz

DSO5000P Series Digital Storage Oscilloscope User Manual

61

Specifications

EXT/5

Trigger Level Range

Trigger Level
Accuracy, typical
(Accuracy is for
signals having rise
and fall times 20ns)
Set Level to 50%,
typical

1V from DC to 100MHz

1V from DC to
100MHz;
1.75V from
100MHz to
200MHz

AC

Attenuates signals below 10Hz

HF Reject

Attenuates signals above 80kHz

LF Reject

Same as the DC-coupled limits for frequencies above
150kHz; attenuates signals below 150kHz

Source

Range

CH1, CH2

±8 divisions from center of screen

EXT

±1.2V

EXT/5

±6V

Source

Accuracy

CH1、CH2

0.2div × volts/div within ±4 divisions from center of
screen

EXT

± (6% of setting + 40mV)

EXT/5

± (6% of setting + 200mV)

Operates with input signals 50Hz

Note: Bandwidth reduced to 6MHz when using a 1X probe.
Video Trigger Type

Source

Range

CH1, CH2

Peak-to-peak amplitude of 2 divisions

EXT

400mV

EXT/5

2V

Signal Formats and
Field Rates, Video
Trigger Type

Supports NTSC, PAL and
SECAM broadcast systems for
any field or any line

Holdoff Range

100ns to 10s

Pulse Width Trigger
Pulse Width Trigger
Mode

Trigger when < (Less than), > (Greater than), = (Equal), or
Positive pulse or Negative pulse

Pulse Width Trigger
Point

Equal: The oscilloscope triggers when the trailing edge of the pulse
crosses the trigger level.
Not Equal: If the pulse is narrower than the specified width, the trigger
point is the trailing edge. Otherwise, the oscilloscope triggers when a
pulse continues longer than the time specified as the Pulse Width.
Less than: The trigger point is the trailing edge.
Greater than (also called overtime trigger): The oscilloscope triggers
when a pulse continues longer than the time specified as the Pulse
Width.

Pulse Width Range

Selectable from 20ns to 10s

DSO5000P Series Digital Storage Oscilloscope User Manual

(Not Equal);

62

Specifications

Slope Trigger
Slope Trigger Mode

Trigger when < (Less than), > (Greater than), = (Equal), or
Positive slope or Negative slope

(Not Equal);

Slope Trigger Point

Equal: The oscilloscope triggers when the waveform slope is equal to
the set slope.
Not Equal: The oscilloscope triggers when the waveform slope is not
equal to the set slope.
Less than: The oscilloscope triggers when the waveform slope is less
than the set slope.
Greater than: The oscilloscope triggers when the waveform slope is
greater than the set slope.

Time Range

Selectable from 20ns to 10s

Overtime Trigger

The leading edge: Rising edge or Falling edge; Time Setting: 20-10s

Swap Trigger
CH1

Internal Trigger: Edge, Pulse Width, Video, Slope

CH2

Internal Trigger: Edge, Pulse Width, Video, Slope

Trigger Frequency
Counter
Readout Resolution

6 digits

Accuracy (typical)

±30ppm (including all frequency reference errors and ±1 count errors)

Frequency Range

AC coupled, from 4Hz minimum to rated bandwidth

Signal Source

Pulse Width or Edge Trigger modes: all available trigger sources
The Frequency Counter measures trigger source at all times, including
when the oscilloscope acquisition pauses due to changes in the run
status, or acquisition of a single shot event has completed.
Pulse Width Trigger mode: The oscilloscope counts pulses of significant
magnitude inside the 1s measurement window that qualify as triggerable
events, such as narrow pulses in a PWM pulse train if set to < mode and
the width is set to a relatively small time.
Edge Trigger mode: The oscilloscope counts all edges of sufficient
magnitude and correct polarity.
Video Trigger mode: The Frequency Counter does not work.

Acquisition
Acquisition Modes

Normal, Peak Detect, and Average

Acquisition Rate,
typical

Up to 2000 waveforms per second per channel (Normal acquisition
mode, no measurement)

Single Sequence

Acquisition Mode

Acquisition Stop Time

Normal,
Peak Detect

Upon single acquisition on all channels
simultaneously

Average

After N acquisitions on all channels simultaneously,
N can be set to 4, 8, 16, 32, 64 or 128

DSO5000P Series Digital Storage Oscilloscope User Manual

63

Specifications

Inputs
Inputs
Input Coupling

DC, AC or GND

Input Impedance,
DC coupled

1M ±2% in parallel with 20pF±3pF

Probe Attenuation

1X, 10X

Supported Probe
Attenuation Factors

1X, 10X, 100X, 1000X

Maximum Input
Voltage

Overvoltage Category

Maximum Voltage

CAT I and CAT II

300VRMS
Category

CAT III

150VRMS (1×)

(10×),

Installation

Installation Category II: derate at 20dB/decade above 100kHz to 13V
peak AC at 3MHz* and above. For non-sinusoidal waveforms, peak
value must be less than 450V. Excursion above 300V should be of less
than 100ms duration. RMS signal level including all DC components
removed through AC coupling must be limited to 300V. If these values
are exceeded, damage to the oscilloscope may occur.

Measurements

Cursors

Voltage difference between cursors: △V
Time difference between cursors: △T
Reciprocal of △T in Hertz (1/ß T)

Automatic
Measurements

Frequency, Period, Mean, Peak-to-peak, Cycle RMS, Minimum,
Maximum, Rise Time, Fall Time, Positive Width, Negative Width

Linux Feature
Kernel Version

Linux2.6.13

Supported File system

Yaffs, Fat32

Drivers

Buzzer Driver, DMA Driver, FPGA Driver, I2C Driver,SPI Driver,
IO-bank Driver, USB Host Driver, LCD Driver, USB massstorage、
gadget Driver

Linux Applications

busybox1.18.4, gnupg1.4.11

U_boot Version

Vivi_2.6.13

DSO5000P Series Digital Storage Oscilloscope User Manual

64

Specifications

General Specifications
Display
Display Type

7 inch 64K color TFT (diagonal liquid crystal)

Display Resolution

800 horizontal by 480 vertical pixels

Display Contrast

Adjustable (16 gears) with the progress bar

Probe Compensator Output
Output Voltage,
typical

About 5Vpp into 1M

Frequency, typical

1kHz

load

Power Supply
Supply Voltage

100-120VACRMS(±10%), 45Hz to 440Hz, CATⅡ
120-240VACRMS(±10%), 45Hz to 66Hz, CATⅡ

Power Consumption

<30W

Fuse

2A, T rating, 250V

Environmental
Temperature
Cooling Method
Humidity
Altitude

Mechanical Shock

Operating: 32℉ to 122℉ (0℃ to 50℃)
Nonoperating: -40℉ to 159.8℉ (-40℃ to +71℃)
Convection
+104℉ or below (+40℃ or below): 90% relative humidity
106℉ to 122℉ (+41℃ to 50℃): 60% relative humidity
Operating and Nonoperating

3,000m (10,000 feet)

Random Vibration

0.31gRMS from 50Hz to 500Hz, 10
minutes on each axis

Nonoperating

2.46gRMS from 5Hz to 500Hz, 10
minutes on each axis

Operating

50g, 11ms, half sine

Length

313mm

Height

142mm

Depth

108mm

exclusive of packing and
accessories

2.08Kg

Length

385mm

Width

200mm

Height

245mm

inclusive of all accessories

约 2.5Kg

Mechanical
Size

Weight

Packing
Gross Weight

DSO5000P Series Digital Storage Oscilloscope User Manual

65

Specifications

8.2 Accessories
All the following accessories are available.
Standard Accessories
Sketch

Description
X1, X10 two passive probes. The passive probes have a 6MHz
bandwidth (rated 100Vrms CAT III) when the switch is in the X1
position, and a maximum bandwidth (rated 300Vrms CAT II) when
the switch is in the X10 position. Each probe consists of all
necessary fittings.
A USB A-B line, used to connect external devices with USB-B
interface like a printer or to establish communications between PC
and the oscilloscope.
A power cord special for this product. In addition to the power cord
shipped with your instrument, you may purchase another one
certified for the country of use.
A software installation CD. It contains the user manual of
DSO5000P, giving particular descriptions on the DSO5000P series
oscilloscopes.

DSO5000P Series Digital Storage Oscilloscope User Manual

66

General Care and Cleaning

Chapter 9 General Care and Cleaning
9.1 General Care
Do not put or leave the device in a place where the LCD display will be exposed to direct sunlight
for long periods of time.
Note: To avoid damage to the oscilloscope or probes, do not expose them to sprays, liquids, or
solvents.

9.2 Cleaning
Examine the oscilloscope and probes as often as operating conditions require. To clean the
exterior surface, perform the following steps:
1) Use a lint-free cloth to remove floating dust on the outside of the oscilloscope and probes.
Take care to avoid scratching the glabrous display filter.
2) Use a soft cloth dampened with water to clean the oscilloscope. For more efficient cleaning,
you may use an aqueous solution of 75% isopropyl alcohol.
Note: To avoid damage to the surface of the oscilloscope or probes, do not use any
corrosive or chemical cleaning agents.

DSO5000P Series Digital Storage Oscilloscope User Manual

67

General Care and Cleaning

Appendix A Harmful and Poisonous
Substances or Elements
Harmful and poisonous substances or elements 1
Component2

Pb

Hg

Cd

Cr(Vi)

PBB

PBDE

Shell and Chassis

X

0

0

X

0

0

Display Module

X

X

0

0

0

0

Circuit Board

X

0

0

X

0

0

Power Supply

X

0

0

X

0

0

Electric Wire and Cable Assembly

X

0

0

0

0

0

Connector

X

0

0

X

0

0

Fastener and Installed Hardware

X

0

X

X

0

0

Other Accessories (including
probes)

X

0

0

X

0

0

Others

0

0

0

0

0

0

‘X’ means that at least the content of this poisonous and harmful substance in a homogeneous
material of this component exceeds the limit specified in the SJ/T 11363-2006 standard.
‘0’ indicates that the content of this poisonous and harmful substance in all homogeneous
materials of this component is refrained under the limit stated in the SJ/T 11363-2006 standard.
This component list contains components approved in the file ‘Management Measures’.

DSO5000P Series Digital Storage Oscilloscope User Manual

68

