Programmable DC Power Supply
KD3000-6000 Series User Manual

SHENZHEN KORAD TECHNOLOGY CO. LTD.
Address:F3, Building A4, District B, Songbai Industrial Park,
Guangming New District, Shenzhen, 518106, P.R.China.
Tel: 86-755-61535026
Fax: 86-755-61535025
E-mail: adam@koradtechnology.com
www.koradtechnology.com
www.korad.com.cn

User Manual

KD3000/6000- Series User Manual

Main Features
Low noise
Cooling fan controlled by heatsink temperature
Constant voltage / constant
Digital panel control
4 digits display
Software calibration
Over Current Protection
Button lock function
USB/RS232 for remote control

1

KD3000/6000- Series User Manual

7. OUT
Description:Turns on or off the output.
Boolean:0 OFF,1 ON
Example: OUT1 Turns on the output
8. STATUS?
Description:Returns the POWER SUPPLY status.
Contents 8 bits in the following format
Bit
Item
Description
0
CH1 0=CC mode, 1=CV mode
1,2,3,4,5
N/A
6
Output 0=Off, 1=On
7
N/AN/A
9. *IDN?
Description:Returns the KA3005P identification.
Example *IDN?
Contents KORAD KD3005P V2.0 (Manufacturer, model
name,).
10. RCL
Description:Recalls a panel setting.
NR1 1 5: Memory number 1 to 5
Example RCL1 Recalls the panel setting stored in
memory number 1
11. SAV
Description:Stores the panel setting.
NR1 1 5: Memory number 1 to 5
Example : SAV1 Stores the panel setting in memory
number 1
12. OCP
Description:Over current
Example :OCP1 OCP ON

KD3000/6000- Series User Manual

SAFETY INSTRUCTION
Safety Guidelines
•Do not block or obstruct the cooling fan vent opening.
•Avoid severe impacts or rough handling that leads to
damage.
•Do not discharge static electricity .
•Do not disassemble unless you are qualified as
servi ce perso nnel.

AC INPUT

!

•AC Inut Voltage: 110V / 120V / 220V / 230V , 50 / 60 Hz
•Connect the protective grounding conductor of the AC
power cord to an earth ground, to avoid electrical
shock.

•Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (note below)
Operation
•
Environment Relative Humidity: < 80%
•Altitude: < 2000m
•
•Temperature: 32–104°F
Storage
environment

FUSE
!

•Location: Indoor
•Relative Humidity: < 70%
•Temperature:-10-70℃

Mode l
KD 300 3

110/ 120 VV
T4A/ 250 V

220 /230 V
T2A/ 250 V

KD 300 5
KD 600 3

T5A/ 25 0V
T5A/ 25 0V

T3A/ 250 V
T3A/ 250 V

•To ensure fire protection, replace the fuse only with the specified type
and rating.
•Disconnect the power cord before fuse replacement.
•Make sure the cause of fuse blowout is fixed before fuse replacement.

9

2

KD3000/6000- Series User Manual

KA Series Remote Control Syntax V2.0

KD3005D DIGITAL_CONTROL DC POWER SUPPLY

OCP

Over Current Protection Active

C .C

Constant Current Mode

C .V

Constant Voltage Mode

LO CK

Panel LOCK Indicator

V

Display of voltage
& current value
A

VOLTAGE

Adjustment Knob
for Voltage

CURRENT
Push/Hold

Push/Hold
LOCK / UNLOCK

OCP ON/OFF
PUSH

PUSH

FI NE /CO AR SE

FI NE /CO AR SE

POWER

-

+

Power
ON / OFF Button

Adjustment Knob
for Current

GND

Output TerminalS

FNSZ1057-00-00
S/N NO:011062300000000-0000

Fan

WARNI NG

VOLTAGE RATING: AC 220V
FUSE RATING: 2A / 250V

TO AVOID ELECTRIC SHOCK, THE POWER CORD PROTECTIVE
GROUNDING CONDUCTOR MUST BE CONNECTED TO GROUND.
FOR CONTINUED FIRE PROTECTION, REPLACE FUSE WITH 250V FUSE OF THE
SPECIFIED TYPE AND RATING.NO OPERATION SERVICEABLE COMPONENTS INSIDE

Label on Vol tage
& Fuse

DO NOT REMOVE COVERS; REFER
SERVICING TO QUALIFIED PERSONNEL.

Power Socket

USB Interface
Rs232 Interface

DISPLAY
Voltage level

Current level

V

A

Voltmeter displays the setup value of output
voltage .

Displays the setup value of output current .

3

KD3000/6000- Series User Manual

Command format :VSET:
1. VSET: command header
2. X: output channel
3. : separator
4. NR2: parameter
Command Details:
1. ISET:
Description: Sets the output current.
Example:ISET1:2.225
Response time 50ms
Sets the CH1 output current to 2.225A
2. ISET?
Description: Returns the output current setting.
Example: ISET1?
Returns the CH1 output current setting.
3. VSET:
Description:Sets the output voltage.
Example VSET1:20.50
Sets the CH1 voltage to 20.50V
4. VSET?
Description:Returns the output voltage setting.
Example VSET1?
Returns the CH1 voltage setting
5. IOUT?
Description:Returns the actual output current.
Example IOUT1?
Returns the CH1 output current
6. VOUT?
Description:Returns the actual output voltage.
Example VOUT1?
Returns the CH1 output voltage

8

KD3000/6000- Series User Manual

KD3000/6000- Series User Manual

REMOTE CONTROL
COM setting

Condition Indicators - LED Panel Lights

Set up the COM port inside the PC according to the
following list.

OCP

Over Current Protection indicator. When the power supply
is in OCP mode this light is on.

C .C

C.C indicates constant current. When the power supply is
in constant current mode, this light is on.

C .V

C.V indicates constant voltage. When the power supply is
in constant voltage mode, this light is on.

LOC K

Panel LOCK Indicator

• Baud rate: 9600
• Parity bit: None
• Data bit: 8
• Stop bit: 1
• Data flow control: None

Voltage and Current Adjustment Knob Operation
There are 3 adjustment modes for the voltage and current levels, that is, Mode 1
and Mode 2. Mode 1: Before setting, push the knobs to adjust the voltage and
current levels. Mode 2: adjust directly, no need to push the knobs. And these 2
modes can be shifted by pushing the voltage adjustment knob and the current
adjustment knob at the same time and holding for 2 seconds.
Mode 3: remote control mode (programmable control mode).

RS232 Interface Definition

2
3

2
3

5

5

Mode1 LOCK Adjustment Mode
VOLTAGE
Push/Hold
LOCK / UNLOCK

PC

KD3005P DC POWER SUPPLY

PUSH
FINE/COARSE

CURRENT

Functionality Run this query command via the terminal
check
application such as MTTTY (Multi-threaded TTY).

*DIN?
This should return the identification information:
Manufacturer, model name, software version.
KORAD KD3003P Vx.xx

7

Push/Hold
OCP ON/OFF
PUSH
FINE/COARSE

Voltage Adjustment Knob: Push the voltage
adjustment knob and then the voltage meter will
flicker, when voltage output can be changed by
adjusting the knob. T hen the resolution of the knob
rotation can be changed. Push it to change the
resolution of voltage adjustment;

Current Adjustment Knob: Push the voltage
adjustment knob and then the voltage meter will
flicker, when voltage output can be changed by
adjusting the knob. And push the knob again when
the meter flikers, then the resolution of the knob
rotation can be changed.
Will be closed.

4

KD3000/6000- Series User Manual

KD3000/6000- Series User Manual

Specifications
Mode 2 Continuous Adjustment Mode
In mode 2,rotate the adjustment knobs to adjust the
voltage and current values. The default of the
voltage initial settings is 1v while that of the current
is 100mA. The voltage and current levels can be
changed by pushing the knobs.

Note: The specifications below are tested under the conditions of
temperation 25℃+-5℃ and the warm-up for 20 minutes.
Mode ls
Voltage
Current

KD300 3P

KD300 5P

KD600 3P

0-30V
0-5A

0-60V
0-3A

Voltage
≤0.01 %+2m v
Current
≤0.1%+5m A
Line Regulation

≤0.01 %+2m v
≤0.1%+10m A

≤0.01 %+2m v
≤0.1%+5m A

Voltage
≤0.01 %+3m v
Current
≤0.1%+3m A
Setup Resolution

≤0.01 %+3m v
≤0.1%+3m A

≤0.01 %+3m v
≤0.1%+3m A

0-30V
0-3A

Load Regulation

Operation of Over Current Protection

CURRENT

press and hold for 3 seconds to start OCP mode, where
the output will be cut off when the output current reaches
the set value. In the OCP mode, rotate this knob to recover
the output. Press and hold for 3 seconds again and then the
OCP functi on

Push/Hold
OCP ON/OFF
PUSH
FINE/COARSE

Voltage

10m V
1 mA
Setup Accuracy(25℃+-5℃)

10m V
1 mA

Current

Mode 3

Remote Control Mode
Push and hold the VOLTAGE knob for 3 seconds to lock
the VOLTAGE and CURRENT adjustment knobs. Then the
output of the power supply will be off. At this time, the
CURRENT adjustment knob becomes the output knob;
push the CURRENT knob and then the output of the power
supply will be ON and OFF accordingly. Push and hold the
VOLTAGE knob again for 3 seconds and the VOLTAGE
and CURRENT adjustment knobs will be unlocked .

VOLTAGE
Push/Hold
LOCK / UNLOCK
PUSH
FINE/COARSE

Power Switch and Output Terminals
POWER

On / Off main power.

≤0. 5%+20m V

≤0. 5%+30m V

≤0. 5%+10m A

≤0. 5%+5m A

≤1 m V rms
Current
≤3 mArms
Temp. Coefficient

≤2 m V rms
≤3 mArms

≤1 m V rms
≤3 mArms

≤150 ppm
≤150 ppm
Read Back Resolution

≤15 0ppm
≤150 ppm

≤150 ppm
≤150 ppm

10 m V
1 mA

10 m V
1 mA

Voltage

≤0. 5%+20m V
Current
≤0. 5%+5m A
Ripple(20-20M)
Voltage

Voltage

Current

10 m V
1 mA
Read Back Temp. Coefficient

Voltage
Current

Voltage
Current

+

outputs voltage and current.

Connects the ground (earth ) terminal.

5

10m V
1 mA

≤150 ppm
≤150 ppm

≤150 ppm
≤150 ppm

Accessories
User manual *1, Power cord*1
Weight and Dimension
KD3003,KD3005,KD6003:110mm(W)*156mm(H)*260(D)
KD3003x4Kg,KD3005x4.8Kg

6

≤150 ppm
≤15 0 ppm



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.3
Linearized                      : No
Page Count                      : 6
Create Date                     : D:20141007092027
Modify Date                     : D:20141007092027
Producer                        : Corel PDF Engine Version 1.0.0.458
Creator                         : CorelDRAW 版本 12.0
Title                           : KD3000PV3.0
EXIF Metadata provided by EXIF.tools

Navigation menu