KAB45 1 M1 1373 Microclave Vs Invision Plus Comparative Matrix Rev01 Web

User Manual: KAB45-1

Open the PDF directly: View PDF PDF.
Page Count: 1

Straight fluid path
Flat swabbing
surface
Minimal residual
volume
Split-septum with
internal blunt cannula
MicroClave by ICU Medical Inc. InVision-Plus by RyMed Technologies, Inc.
MicroClave® and InVision-Plus® Comparative Matrix
Performance data on le at ICU Medical Inc. San Clemente, CA 92673. Reference ENG-433 InVision-Plus and RyMed are trademarks of RyMed Technologies, Inc.
1. Evaluation: Needleless Connectors. ECRI Institute Health Devices, Sept. 2008, Volume 37, Number 9
2. Ryder M, James G, Pulchini E, Bickle L, Parker A. Presented at the Infusion Nursing Society Meeting, May 2011. Dierences in bacterial
transfer and fluid path colonization through needlefree connector-catheter systems in vitro.
©2013 ICU Medical Inc. | M1-1373 Rev. 01 800.824.7890 | www.icumed.com
1.26”
0.45”
1.807 g (MicroClave Clear)
1.874 g (MicroClave)
1.745 g
1.27”
0.43”
PRODUCT
PERFORMANCE MICROCLAVE TECHNOLOGY INVISIONPLUS TECHNOLOGY
Base Technology
Internal cannula and silicone compression seal split-septum. Internal
cannula windows are exposed by the insertion of a male luer, and
cannula enters the male luers internal space to achieve flow.
Internal cannula and silicone compression seal split-septum. Internal cannula
windows are exposed by the insertion of a male luer, and cannula enters the male
luers internal space to achieve flow.
Displacement Neutral: 0 to -0.01 mL
-0.0049 mL and -2.2 cm Published1
Neutral: 0 to -0.01 mL
-0.0043 mL and -1.9 cm Published1
Residual Volume 0.04 mL 0.03 mL
Fluid Path Straight through polycarbonate cannula. Straight through polycarbonate cannula.
Moving Parts in
Fluid Path No No
Number of
Assembly Parts 3, of which 1 moves on luer access. 5, of which 3 move on luer access.
Fluid Residual
External on Disconnect Minimal Minimal
Clamping Sequence None required None required
Flow Rate 165 mL/min 131 mL/min
Clear Available Yes Yes
Antimicrobial
Available Yes Yes
Bacterial Transfer
Performance The least amount of bacterial transfer of any connector tested.2Exhibits a higher bacterial transfer rate than MicroClave.2
Clear housing
Split-septum with
internal blunt cannula
Minimal residual
volume
Straight fluid path
Flat swabbing
surface

Navigation menu