User manual

Facial rejuvenation and body
contouring in a single device

INTRODUCTION

The purpose of this manual is to provide the safyre user with the information on the instructions for use,
treatments, controls, safety precautions, installation and maintenance of the equipment, aiding use and
care of the system.

Adequate training by qualified staff is indispensable to apply treatments using this equipment.
The manufacturer provides training courses that are essential in order to be able to use the safyre
device.

This user manual includes the following information:
 Safety and regulatory: Contains explanations and indications about the necessary security
measures in order to operate the safyre device.
 Storage, handling and transportation: Recommendations for the storage and handling of the
product, in order to avoid damage.
 Initial verification: Considerations to receive the device and list of components and accessories
safyre.
 General description: Overview of the various safyre subsystems.
 Installation: Explains the electrical requirements, space and environmental required for
installation of safyre device.
 Operation: Explains how to use safyre.
 Alerts: Description of the different alarms with counting system.
 Description of procedure: safyre procedure and treatments that can be also made protocol
and contraindications.
 Maintenance: It provides a detailed explanation of the procedures for cleaning and maintenance
of device and accessories.
 Specifications: Contains a summary sheet with all device specifications.

Use of the safyre equipment is only authorised for people with sufficient training and
knowledge to guarantee its correct use and handling within the safety parameters
established.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 43 de 119
MKT-07-M1/20160623/MAN00159

SAFETY AND REGULATORY

Regulation

The CE mark accredits compliance with the European Directive applicable to the device.

Electrical safety requisites
The safyre equipment is designed and intended for continuous operation, allowing it to operate on
universal mains voltage of 100 – 240 V, 50 – 60 Hz.

safyre is protected from electric shock by earthling protection. The unit is earthed through the
conductor on the power cable.

WARNING: In order to avoid the risk of electric shock, this equipment must only be
connected to a power grid with earthling protection. The power lines must be free of
transitory, voltage or current peaks, voltage drops or surges.

WARNING: Avoid connecting the safyre to the same electric line to which a sensitive
apparatus is connected.

There are high currents inside the equipment that may be dangerous, so no housing panel must be
removed other than by personnel authorized by the manufacturer. The safyre must never be left
unattended when switched on.

The dispositive protection device is a timed fuse with ceramic encapsulation of 5A (T5AH250V).

WARNING: The equipment has 1 fuse installed. If it blows, it is mandatory to replace it
with one that has the same rating to guarantee safety and avoid fire risk.

Electromagnetic safety requisites

safyre has been designed so it does not emit electromagnetic disturbances that may affect radio
services or the essential operation of other devices. Likewise, safyre is designed to provide adequate
immunity to offer basic safety and normal functioning in the presence of electromagnetic disturbances.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 44 de 119
MKT-07-M1/20160623/MAN00159

safyre is an equipment that intentionally applies RF electromagnetic energy to perform its intended
function, marked with the symbol IEC 60417-5140 (See regulatory marking and labelling).

safyre requires special precautions in relation to EMC and needs to be installed and
commissioned according to the information on EMC provided. Portable and mobile RF
communications equipment may affect the safyre during its normal use.

Only the cables and accessories supplied by the manufacturer must be used. Use of
accessories, transducers and cables other than those specified may cause an increase in
emissions or a decrease in the immunity of safyre.

safyre must not be used next to or stacked up with other equipment. Should that
happen, the safyre must be observed to check its correct operation.

General warnings and precautionary techniques

1. In case of dropping one of the handpieces, under no circumstance should you reuse it, since
it might be damaged. Contact the Technical Support.
2. If an accessory shows any kind of defect, you should stop using it immediately and contact
the Technical Support.
3. For security reasons, before cleaning the device, disconnect the power cable from the outlet.
4. Access to the internal parts of the device should be performed by authorized service personnel.
For reparations or other information you should contact High Technology Products, S.L.
5. For a correct usage of the handpiece, make sure to handle it carefully, avoiding accidental
impacts that could damage it or reduce its effectiveness.
6. In case of accident (e.g. accidental fall of the applicator in a container of water) unplug the
device and contact our Technical Support.
7. Do not use accessories which are not original: they could damage the device and void the
warranty. High Technology Products S.L, declines all responsibility for damage caused by nonoriginal spare parts.
8. The scope of the device is professional aesthetic medicine; treatment should be performed
always under medical supervision, therefore it should not be used for other applications or by
unqualified personnel. Warranty is void in case of breaching this observation.
9. It is essential to take the device-training course.
10. The operating instructions must be kept near the device.
11. Check periodically that the accessories wires are in perfect condition.
12. Do not use electrical extenders for the power supply connection wire.

13. Any tampering by unauthorized personnel voids the warranty.
14. The device shall not be exposed to the elements (heat, cold, rain...).

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 45 de 119
MKT-07-M1/20160623/MAN00159

15. No responsibility is assumed for the misuse of the device.
16. The manufacturer and distributors will only be responsible for the safety of the product, if the
repairs, modifications and adjustments are carried out by expressly authorized personnel, and if
the device is used by qualified personnel and in accordance with the instructions.
17. The device should never be used in the presence of flammable gases or liquids.
18. Only personnel specially authorized by the manufacturer may provide technical service,
especially inside the protective covers. This includes any adjustment in the power supplies,
power amplifiers and control, since within the device dangerous electrical voltages are present.
19. For security reasons, the network cable is fitted with a plug with protective ground connection.
Use only a proper grounded AC outlet. Otherwise, this could cause damage to the appliance or
to people. If there are any specific questions, please contact our Technical Support at High
Technology Products, S.L.
20. You must only use glycerin or products and fluids specified by the manufacturer;
otherwise the warranty will be void.

Other safety requirements

IMPORTANT: For any questions on health status, pregnancy, lactation, etc., both the
person who should receive treatment and you should do it, check before starting
treatment to a specialist.

Regulation for labeling

Symbol S1 - Information
Information based on Art.13of Legislative Decree. 151/05 on 25/07/2005.
“The Implementation of Directive 2002/96/EC on the reduction of hazardous substances in electrical and
electronic device and waste wear”.
Present on the label of the technical features indicating that the product should not be disposed of as
urban waste but must be the subject of a separate collection.
If the waste is to be removed in an unsuitable way, it is possible that some parts of the product may
have potentially negative effects on the environment and human health.
In case of faulty product abuse, sanctions are foreseen.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 46 de 119
MKT-07-M1/20160623/MAN00159

Symbol IEC 60417- 5333 - Information
Type BF Applicable Part.
Floating applicable parts. Not allowed to use it in combination with defibrillators.
Present on the main label the device at the bottom rear part.

Symbol IEC 60417- 5007 - Information
“ON” (power).
Present at the unit switch. It indicates that the device is on and consuming energy.

Symbol IEC 60417- 5008 - Information
“OFF” (Energy).
Present at the unit switch. Indicates that the device is off and no power is being consumed.

Symbol IEC 60417- 5134 - Information
Electromagnetic Interference.
Present in the supporting documents according to the requirements of the harmonized standard
IEC60601-1-2:2007. Electromagnetic interference may occur if the device is placed nearby some device
marked with this symbol.

Symbol ISO 7010- W001 - Information
General warning.
Present in the user interface, the appearance of this symbol during a normal usage of the device means
that you should consult the user manual before any further action.

Symbol 5.12 EN280 - Information
Symbol manufacturer.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 47 de 119
MKT-07-M1/20160623/MAN00159

It indicates the manufacturer's data. Present in the main tag device at the back.

Symbol S2 - Information
Fuse.
Present on the main label of the device at the bottom rear, indicating its features.

Brand 1 – Information
Brand attesting the conformity of device to the legal obligations of the European Community directives
are applicable.
Present on the main label of the device at the bottom rear.

Technical general warnings

SAFYRE
100 – 240 V 50 - 60 Hz
350 W
High Technology Products
S.L.
Passatge Masoliver, 24-28
Barcelona (Spain)
MADE IN SPAIN

External Labeling of the Device - Information
Label present in the rear of the unit next to the power connector in accordance with the requirements of
IEC 60601-1 harmonized standard for the basic safety of medical electrical device. Symbols for sanitary
device according IEC980:2009 harmonized standard.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 48 de 119
MKT-07-M1/20160623/MAN00159

Almacenamiento
Storage

Temperatura

2ºC – 50ºC

Temperature

Humedad
Humidity

(Sin condensación)

< 90%

(Non‐condensing)

Outer packaging labeling of the Device - Information
Label present in the outer packaging of the device in compliance with the applicable requirements of the
harmonized standard IEC60601-1 for operating conditions.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 49 de 119
MKT-07-M1/20160623/MAN00159

STORAGE, HANDLING AND TRANSPORTATION

Location and transport of the device
Due to the weight of the device, safyre may cause injury if not moved with the necessary care. safyre is
designed

to work with a mobile stand, although it may be used without that stand as table top

equipment.

If used as table top equipment, it must always be transported between two persons and placed on the
work table at an adequate height (approximately 1 metre from the floor) to guarantee its correct handling
and visibility.

If you attempt to transport the equipment without adequate care, the personnel may
suffer harm.

Place the equipment away from sources of heat. Keeps enough space around the device to guarantee
its correct ventilation, above all making sure never to obstruct the ventilation grilles of any of the
panels of the device.

It is recommended that a minimum of 40cm. separation between the back of the unit and
20cm. on the sides. The device must rely on their feet (if used in desktop mode).

Work and storage conditions

In order to ensure that the safyre equipment works in an optimum way, it is recommendable to maintain
the workplace at between 18ºC and 30ºC, with a relative humidity under 80%.

For correct transport and storage of the equipment, it is recommended to keep the ambient temperature
between 2 ºC and 50 ºC.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 50 de 119
MKT-07-M1/20160623/MAN00159

INITIAL VERFICATION DEVICE AND ACCESSORIES

It is very important to check your device when you receive it, in the presence of the carrier and with the
highest degree of attention, because the damage due to dropping or shocks may not be noticed at first.

Although the manufacturer has taken the utmost care in packaging, this might not be sufficient to
guarantee that the merchandise has not suffered damage.
You should check:
 The material received matches the list of components and accessories herein.
 The outer packaging conditions of humidity, external damage, or any abnormalities.
 That each of the components and accessories are in good condition.

Keep the original packaging of the device during the first 30 days of use as changes shall
not be accepted without delivery of the equipment in the original packaging during that
period.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 51 de 119
MKT-07-M1/20160623/MAN00159

List of components and accessories

Description

Nº

Description

Nº

Description

Nº

Description

Nº

safyre device

1

SD electrode holder
body

1

Multipolar applicator
body 2x2

1

Multipolar applicator
body 3x3

1

Monopole
applicator body 70

1

Monopole applicator
facial 35

1

Massager

1

SD electrode holder
facial

Massager accessory

Glycerin (dermic gel
RF)

1

Pedal

1

User and packaging
manuals

English and Castellan
protocols

2

Display safyre

1

Passive plate

1

Guarantee
certificate

Conformity
declaration

1

1

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 52 de 119
MKT-07-M1/20160623/MAN00159

GENERAL DESCRIPTION OF THE DEVICE AND HANDPIECES

External description of the device
The safyre device is comprised of a central unit and a series of radio frequency heads, with extractable
accessories and mechanical massage. safyre also includes accessories for patient safety during the
treatment. The following is a description of the main components of the device.

Accessory
cradle
TFT – Touch screen

RF accessory
connector

Mechanical massage
accessory connector

Passive plate
connector

Pedal

Main view - safyre

Air ventilation
Power
switch – I/O
Mains cord
socket

Rear view - safyre

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 53 de 119
MKT-07-M1/20160623/MAN00159

Description of the applicators and accessories
safyre includes a series of radio frequency heads, both for bodily as well as facial use, with multiple
extractable accessories, and an head to perform the mechanical massage function just as described
below.

It is essential to clean the extractable radio frequency accessories and the mechanical
massage heads using a damp cloth to avoid gel accumulation at the end of each
treatment. Also perform cleaning during the treatment if necessary.

To pay special attention to the internal electrical connection, both the accessories as
well as the base, in case remains of gel or liquids are observed during handling.

Radio frequency body head

Connector head

Internal connector
Applicator (electric)
Internal connector
Applicator (mechanical)

RF body electrode holder

 RF body electrode holder: Base of the body radio frequency head that allows manual
connection / disconnection of the different body accessories easily and comfortably.
 Connector head: Type of connection that allows connection / disconnection of the RF head to
the central unit.
 Internal head connector (electrical): Type of connection (located inside the base of the head)
that connects electronically with the different extractable body accessories, guaranteeing
detection of the head and active RF emission when the connection is complete.
 Internal connector head (mechanical): Centring piece (x2) that mechanically connects the
base of the head to the different extractable body accessories.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 54 de 119
MKT-07-M1/20160623/MAN00159

Extractable body accessory – multipolar 6 pins (3+3)
The 6 pin multipolar accessory for body treatments uses multipolar radio frequency technology, so the
radio frequency current circulates between the pins without the need for a passive plate.
Accessory connector
(Mechanical)
RF emitters

Joint rings

Multipolar 6 pin accessory

Accessory connector
(Electrical)

Extractable body accessory – multipolar 4 pins (2+2)
The 4 pin multipolar accessory for body treatments uses multipolar radio frequency technology the
same way as the 6 pin multipolar accessory, so the radio frequency current circulates among the pins
without the need for the passive plate.

Accessory connector
(Mechanical)

RF emitters

Joint rings
Accessory connector
(Electrical)

Accesorio multipolar 4 pin

Extractable body accessory – monopolar 70
The monopolar 70 accessory for body treatments uses monopolar radio frequency technology, so the
radio frequency current requires use of the passive plate in contact with the body to be able to perform
the treatment.
Accessory connector
(Mechanical)
RF emitter

Joint rings
Accessory connector
(Electrical)

Monopole Accessory

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 55 de 119
MKT-07-M1/20160623/MAN00159

The following details the parts that comprise the extractable accessories for body radio frequency:
 Accessory connector (electric): Type of connection of the radio frequency accessories that are
electronically connected with the internal connector head (electric) located in the electrode
holder that guarantees detection of the head and the active RF emission when the connection is
correct.
 Accessory connector (mechanical): Female centering piece (x2) that mechanically connects
the different extractable body accessories with the electrode holder for body radio frequency.
 Joint rings: Two fixing elements that guarantee the seal tightness of the connection of
extractable accessories, mainly avoiding liquids or remains of gel entering during its use.
 RF emitter/s: Applicable part in contact with the patient through which the radio frequency
energy is emitted once the electric connection of the accessories is correct (as long as the
radio frequency emission is activated).

It is very important to always guarantee a complete contact of the RF emitters with
the patient’s skin, along with a sufficient amount of gel, to avoid false contacts while the
radio frequency is activated. Guaranteeing good contact of those emitters avoids any
kind of adverse effect on the patient during application of the treatment.

Radio frequency facial head

Internal connector
head (electric)

Internal connector head
(mechanical)

Joint rings

RF facial electrode holder
Parts that comprise the facial electrode holder:

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 56 de 119
MKT-07-M1/20160623/MAN00159

 Facial RF head: Base of the facial radio frequency head that allows manual connection /
disconnection of the different facial accessories easily and comfortably.
 Connector head: Type of connection that allows connection / disconnection of the RF head to
the central unit.
 Internal connector head (electric): Type of connection (located in the interior base of the
electrode holder) that is electronically connected with the different extractable facial
accessories, guaranteeing detection of the head and the active RF emission when the
connection is complete.
 Internal connector head (mechanical): Centering piece that mechanically connects the base of
the head with the different extractable facial accessories.
 Joint rings: Two fixing elements that guarantee the seal tightness of the connection of
extractable accessories, mainly avoiding liquids or remains of gel entering during its use.

Monopolar facial extractable accessory
The monopolar accessory for facial treatments uses monopolar radio frequency technology, so the
radio frequency current requires the use of a passive plate in contact with the body to close the circuit.

Accessory connector
(mechanical)
RF emitter
Accessory connector
(electric)

Facial accessory monopole

Parts that comprise the facial electrode holder:
 Accessory connector (electric): Type of connection for the radio frequency accessories
connected electronically to the internal head connector (electric) located on the base of the
head, guaranteeing the direction of the head and of the active RF emission when the
connection is complete.
 Accessory connector (mechanical): Female centring piece that mechanically connects the
different extractable facial accessories to the body radio frequency head base.
 RF emitter/s: Applicable part in contact with the patient through which radio frequency energy
is emitted once the electric connection to the accessories is correct (as long as the radio
frequency emission is activated).

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 57 de 119
MKT-07-M1/20160623/MAN00159

It is highly important to as well as guarantee a complete contact between the RF
emitters and the patient’s skin, along with a sufficient amount of gel, to avoid false
contact while the radio frequency is activated. Guarantee good contact of those emitters
to avoid any kind of adverse effect to the patient during application of the treatment.

Mechanical massage head

The following details the parts comprising the mechanical massage head:

Mechanical massage
base

Applicator connector
Push button

Massage
accessory

Mechanical massage head

 Base of the mechanical massage head: Base of the mechanical massage head that allows
manual connection / disconnection of the massage roller, easily and comfortably.
 Massage accessory: Rotational applicable (extractable) part that connects to the base of the
head and performs the body massage when that function is activated, by user interface or the
push button on the head.
 Head connector: Type of connection that allows connection / disconnection of mechanical
massage head base from the central unit.
 Push button: Head push button that allows activation /deactivation of the massage and
modification of the turning direction of the roller.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 58 de 119
MKT-07-M1/20160623/MAN00159

Operation of the mechanical massage head push button
Long press

With massage inactive, keeping it pressed activates the rotation of the roller.
(Once it is active, work without continuing to press).

Long press

With the massage active, keeping it pressed stops the rotation of the roller
(pause the treatment).

Short press

With the active massage, a short press changes the roller turning direction.

The push button must not be kept pressed during use of the mechanical massage
head. The aim of the push button is to start or stop the treatment and to change the
turning direction (if considered necessary), without the need to use the user interface,
easily and comfortably.

The mechanical massage head is not used with the pedal.

Accessories
Pasive plate: The passive plate is the neutral element that is used in contact with the body in treatments
using monopole radio frequency, to facilitate passage of the current provided by the heads.

The facial and body monopole accessories require a passive plate for normal operation.
It is important to make sure this accessory is in perfect contact with the patient’s body
during the whole treatment to avoid defective current contact. Guaranteeing good contact
of those emitters avoids any kind of adverse effect on the patient during application of the
treatment.

Surface in contact
with the patient

Passive plate

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 59 de 119
MKT-07-M1/20160623/MAN00159

Pedal: Keeping the pedal continuously pressed enables radio frequency emission by the different
heads to perform the treatment. It is recommendable to perform the RF treatments with the pedal.

Pedal activation

Pedal

With the equipment in play mode, the radio frequency emission is activated by keeping
one’s foot pressed on the pedal.

Both when placing as well as on lifting the heads from the treatment area on the patient, it
is very important that the radio frequency emission is disabled (pedal not pressed). This
avoids current false contacts, guaranteeing a safe treatment.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 60 de 119
MKT-07-M1/20160623/MAN00159

INSTALLATION AND START UP

Placement of the device in the support

The equipment has a cradle for accessories located on the rear (see Exterior description of the
equipment). The accessories must always be placed in their cradle, when starting up the equipment and
when its use has finished.

The remaining accessories must be kept duly in their respective boxes to avoid blows and
falls that may give rise to malfunctioning of the equipment. Never use a defective head. In
such a case, contact with the technical service.

Connecting accessories and applicators to the equipment

The following scheme details connection of the different accessories and heads to the SAFYRE
equipment, before beginning the treatment:

RF head
connector

Passive plate
connection

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Mechanical massage
head connector

Pedal connector

Página 61 de 119
MKT-07-M1/20160623/MAN00159

 Use the connector for Mechanical massage head connection to connect the mechanical
massage accessory.
 Use the Radio frequency head connection connectors to connect any radio frequency head
(monopole or multipolar).

The passive plate must be connected to work with the monopole heads (corporal y facial).
Use the passive plate connector for that purpose.

Use the pedal connection to connect the pedal once the heads to be used are connected, in order to
work in that mode.

To connect /disconnect the heads and accessories correctly:

Connect the head:

Place the head connector in the port on the equipment, pressing
lightly for it to make contact. Line up the connector slot with the
slot on the equipment port to join them. Turn the thread shown in
the image clockwise until it is perfectly locked.

Disconnect the head:

Turn the thread in the image anti-clockwise (the opposite
direction to the image) until it unlocks from the machine port.
Never stretch the connector without having turned the thread to
release it. After it is unlocked, remove the connector.

Head recognition

The equipment includes an extractable head recognition system that recognises correct connection of
the radio frequency heads, as well as the mechanical massage head.
 In the event of any radio frequency or mechanical massage head not being duly connected,
the equipment shall not recognise it and the recognition button will not appear on the
interface. In the case of disconnection of the accessory or the head during treatment, the
recognition button will disappear.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 62 de 119
MKT-07-M1/20160623/MAN00159

The passive plate may remain connected to the equipment although monopole treatment
is not being performed. It recognizes the pedal connection, so it may work indistinctly
with the pedal or from the user interface. It is highly recommendable to use the pedal
during radio frequency treatments.

Connecting extractable accessories
For correct connection of extractable radio frequency accessories on the relevant facial or body
electrode holder, one must proceed as follows:
Mechanical connector (x2)

Joint rings

Electric connector

Align the mechanical connectors of the electrode holder and the accessory, male and female,
respectively, and fit the accessory into the head until the connectors lock together. At that point, the
electric connector will automatically be connected by slight pressure until it is fixed in place.
Mechanical connecto
Joint rings

Electric connector

The function of the joint rings is to guarantee the seal tightness of the head, to avoid
liquid or gel getting inside the head cavity and being able to cause malfunctioning of the
head electric connections.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 63 de 119
MKT-07-M1/20160623/MAN00159

Clean the extractable accessories and electrode holder, especially the connection points,
every time a treatment is completed. Clean the head during the treatment is also
advisable.

Proceed in a similar way to that described above to correctly connect the extractable massage
accessory to the base. Thus:

Mechanical connector (x4)

Align the mechanical connectors of the electrode holder and the accessory, male and female,
respectively, and fit the accessory into the head until the connectors lock together, pressing lightly until
fixing them.

Never press the mechanical massage head with remains of gel on your hands, to avoid it
malfunctioning.

Switching on the equipment

Before switching on the equipment for the first time, carefully read section safety
requisites that detail the electric safety requisites, fundamental electromagnetic safety
requisites and requisites of location, storage and transport that must be maintained to
avoid malfunctioning of the equipment causing damage to the actual equipment, operator
or patient.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 64 de 119
MKT-07-M1/20160623/MAN00159

In order to start up the equipment:

Power switch
I/O

Ensure that the power switch is in position Off.

Position I – On

Take the end of the power cable installed in the
Position O – Off

holder * and plug it into the power connector on
the bottom rear of the unit. Connect the other end
of the cord into the electrical outlet in the room.

Power Jack

Connect the power switch I / O, passing the On

Internal safety fuse

position. At this point, the power switch will go
into a luminous state indicating that it is properly
fed. Wait a few seconds for the TFT touch screen
and the screen lights up computer startup appear.
* If the electrical outlet is different from the cable
supplied with the computer, you must first
connect the cable will be delivered as an
accessory.

TFT – Touchscreen
The safyre interacts with the operator through a touchscreen. Softly press the interactive buttons that
appear on the screen to browse the different menus and submenus of the system.

The touchscreen resistance system is touch sensitive in the different areas, so contact with treatment
gels and other products must be avoided to minimize damage and wear of the screen.

See maintenance to ascertain the conditions of cleanliness to avoid operating errors of
the touchscreen.

Switching off the equipment

The equipment is designed for continuous use (see safety requisites) so it may remain on between the
different treatments.

To disconnect the equipment, place the power switch I/O in position O – Off and disconnect the power
cable both from the equipment as well as the power socket.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 65 de 119
MKT-07-M1/20160623/MAN00159

Inspect and clean both the equipment casing as well as the accessories, after switching off the
equipment. To do so, see maintenance that explains how to clean and sterilise the equipment and
accessories.

Place the accessories in their holder, as described in section Placing the accessories in the holder.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 66 de 119
MKT-07-M1/20160623/MAN00159

OPERATION OF THE DEVICE
Treatment selection screen
Once the equipment is switched on, the treatment screen appears, just as shown in the bottom image.
The treatment time is chosen on the treatment screen, selecting the work mode (see Radio frequency
mode and Mechanical massage mode). Once the work mode is selected, the work parameters are
selected on the same screen and the treatment commences (see Radio frequency mode and
Mechanical massage mode).

Restore

Increase (+) / decrease (-)
treatment time

Codes and
licenses

To select the common treatment time for both work modes (radio-frequency mode and mechanical
massage mode):
 Use the keys increase time (+) and decrease time (-) of the user interface, to program the desired
treatment time.
 Use the restore button to restart the treatment time. The equipment will return to the last time
selected.

If both a radio frequency head and the mechanical massage head have been connected,
the treatment time is common to both modes.

Use the code and license button to access the code screen. It is only for technical service.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 67 de 119
MKT-07-M1/20160623/MAN00159

Radio frequency mode
Jack 1 RF Jack 2 RF

RF recognition
buttons
Start /Pause RF
treatment
Increase/ Decrease RF
power

Output 1 RF
Output 2 RF

Before using the radio frequency mode:
 Connect the Radio frequency head to be used, correctly as indicated in Connection of
accessories and heads to the equipment.
 Connect the extractable radio frequency accessory to be used, correctly as indicated in
Connection of extractable accessories.
 At this point the equipment shall recognise the head and accessories selected so the screen
shall display the recognition between for that selection.

Recognition button

The facial recognition button is displayed on screen after a facial head is connected.

The body recognition button is displayed on screen after a body head is connected.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 68 de 119
MKT-07-M1/20160623/MAN00159

In the case of using a head with a monopole accessory, both the head as well as the
passive plate must be connected.

In the case of connecting heads to the radio frequency jacks on the equipment (jack 1
and jack 2), the equipment shall display the recognition button for both jacks. Otherwise,
it shall only show the recognition button of the jack the head has been connected to.

To use the equipment with the pedal, connect the Pedal correctly as indicated in connection of
accessories and heads to the equipment.
To select the work parameters of the radio frequency mode:
 To work in radio frequency mode, it is only necessary to program the power value (expressed as
a %) to work with. To do so, use the keys increase power (+) and decrease power (-) of the radio
frequency mode.
 The radio frequency power values may be adjusted from 0% to 100%.
To start / stop the treatment:
 Press the required RF recognition button (jack 1 or jack 2) and the required radio frequency jack
will be enabled.
RF recognition
button

 To start the treatment using the Start/pause RF treatment button on the user interface, and for
it to be marked. Press the Start/pause RF treatment button again on the user interface and it
will be marked. Press the Start/pause RF treatment button again to pause the treatment.

RF treatment starts/ pause
button

If the pedal is used, activate/deactivate the pedal to start/pause the radio frequency
treatment.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 69 de 119
MKT-07-M1/20160623/MAN00159

Active treatment screen:

RF Jack enabled
Treatment ACTIVE

Active power
emission on the
patient

 During active treatment, the treatment button remains enabled. Likewise, the recognition
button is enabled with the jack established.
 During active treatment, the power emission button displayed is enabled and the active power
of RF delivered is shown on screen expressed in W, in real time.

Mechanical massage mode

Massage recognition
button
Start / Pause
massage

Speed selector

Rotation selector

Before using the mechanical massage mode:
 Connect the Mechanical massage head along with its rotating accessory, correctly as
indicated in Connection of accessories and heads to the equipment.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 70 de 119
MKT-07-M1/20160623/MAN00159

 At this point, the equipment will recognise the head and accessories selected de so the
recognition button of that selection will appear on screen.

Recognition button

The mechanical massage recognition button is displayed on screen when the head is
correctly connected.
To select the work parameters of the mechanical massage mode:
 Use the speed selector to select the head rotation speed according to 3 possible speeds: low,
medium and high.

Low speed

High speed

Average speed

 Use the rotation selector to select the rotation direction of the mechanical massage accessory.

Rotation left

Rotation right

To start/stop the treatment:
 Press the mechanical massage recognition button to select work in that mode. This button
shall be enabled on the user interface.
Massage recognition
button

 Begin the treatment from the user interface using the Start/pause mechanical massage
treatment button and it will be marked. Press the Start/pause mechanical massage
treatment button again to stop the treatment.
Mechanical massage treatment
Start / pause button

 Start the treatment using the head trigger by a long press to activate the mechanical
massage (the Start/pause mechanical massage treatment button is activated no the user
interface). A long press on the trigger is again is to stop the mechanical massage (disabling
the Start/pause mechanical massage treatment button on the user interface).

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 71 de 119
MKT-07-M1/20160623/MAN00159

Mechanical massage gel
applicator trigger

The push button must not be held down when using the mechanical massage head. The
object of the push button is to start or pause the treatment and to change the rotating
direction (if considered necessary), without the need to use the user interface, in an easy,
comfortable way.

When the massage is activated, giving a short press changes the rotation direction of the
massage head.
The mechanical massage head is not used with the pedal.

Active treatment screen:

Mechanical massage
Jack enabled

ACTIVE
treatment

Active rotation
indicator

 During active treatment, the treatment button remains enabled. Likewise, the recognition
button is enabled.
 During the active treatment, the rotation indicator is enabled.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 72 de 119
MKT-07-M1/20160623/MAN00159

DESCRIPTION OF THE safyre PROCEDURE
Intended use
safyre combines radio frequency and mechanical massage technologies for body contouring and facial
rejuvenation treatments, obtaining visible rejuvenation results on the skin, cellulite elimination,
strengthening the cells and greater firmness of the tissue.

Operation description

safyre uses radio frequency technology to generate an electromagnetic radio frequency current at 1
MHz that acts on the organism, causing a localised hyperaemia. That increased temperature inside the
tissues activates

liberation of thermal shock proteins, that protect the collagen, avoiding its

denaturalisation and recovering that affected. In addition to that process there is an increase in the
circulatory response in the zone treated that facilitates oxygen and nutrition contribution to the tissues,
as well as greater elimination of toxins. The main result is rejuvenation of the skin, cell strengthening and
a greater firmness of the tissue.

safyre also includes a mechanical massage so, using a speed and direction adjustable rotational head,
it achieves a massage-tensing effect on the skin that helps with the tissue reduction and skin restoration
by the radio frequency technology.

Radio frequency treatments

Radio frequency generates an electromagnetic current that penetrates through the tissues in contact
with the heads, working from the liquid surface layers of tissue to the deeper layer of the epidermis and
dermis. When it runs through the tissues, the current generates a resistance that is transformed into
movement and that movement generates an increase in temperature in the tissue treated.

At the moment when the organism detects a higher temperature than normal, it intensifies the blood flow
to the area to cool it. The abundance of arterial blood is positive for the tissues, increasing the additional
contribution of oxygen, nutrients and other oligo-elements. It is due to this effect that there is also an
elimination of intercellular toxins. Finally, the blood circulation transports the waste generated to the
evacuation points from the organism.

As a result, the cell is strengthened and prepared to perform its function correctly and at the maximum
of its capacity. Due to this, a cell of muscle tissue shall be more intensely firmed up after taking
advantage of the effect of the current, while a surface cutaneous cell will accelerate its mitosis or cellular
division, to provide increased rejuvenation and greater protection of the skin against the external
elements.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 73 de 119
MKT-07-M1/20160623/MAN00159

The Monopole and Multipolar accessories allow work at different depths, in order to treat different
tissues and types of cellulitis, facial laxity and other areas.

Radio frequency causes a deep heating that affects the skin and subcutaneous fatty tissue. That heating
will favour:
 Lymphatic drainage that will allow a decrease in the liquids and toxins that are embedded in the
adipose of the tissue affected by cellulite.
 An increased circulation in the area will allow the metabolism to be improved, both of the
subcutaneous tissue as well as improved appearance of the accompanying skin.
 Forming a new collagen, both of the skin as well as the subcutaneous tissue, allowing all the
tissue to acquire firmness thanks to reorganisation of the fibrous spaces and thickening of the
overlying dermal layer.
 Lastly, after the controlled thermal lesion with retraction of the tissue, there is an inflammatory
response that is accompanied by migration of fibroblasts that shall reinforce the collagen
structure even more, resulting in rejuvenation of the area treated.

Effects of radio frequency

The procedure by which the current penetrates the tissues and generates the biological effect consists
of 4 phases:

Initial status.

Radio frequency current
penetration.

Increased internal
temperature.

Skin reddening after the
treatment.

General warnings
 The product must always be applied on the skin. Never apply the product on the accessories.
 One must always use the recommended product supplied by the manufacturer. Never use
any other kind of product.
 When performing treatments, one must always begin at a low power to heat the treatment zone.
Once this first heating is performed, the power may be raised progressively by intensifying the
treatment. The final power must always be adapted to the client’s response to the treatment.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 74 de 119
MKT-07-M1/20160623/MAN00159

WARNING: In the case of the electrode suffering any damage or having any anomaly, immediately stop
working with it. Application of a damaged electrode may harm the client.

Facial treatments

The advantages of the treatment are:
 Facial firming up

 Wrinkle prevention

 Collagen regeneration

 Scar attenuation

 Reduction of expression lines

 Reduction of acne marks

 Facial lifting

 Improved skin luminosity

 Facial toning and firmness

 Oxygenation and toxin elimination

The recommended time for such treatments is provided by the equipment treatment protocol. The
treatment is performed quite simply as follows:
 First heat the treatment area for a brief space of time massaging it at low power (depending on
the client’s tolerance of the initial heat).
 After heating up, the power is progressively raised (until completing the treatment) reaching a
temperature on the skin that allows the desired effect for the treatment to be obtained. Always
work at a power according to the client response to the heat generated.
 The movements during the treatment must be upward with pressure in the area and downward
without pressure.

Body treatments

The accessories are specially designed for firming up and fat reduction treatments:


Flaccidness reduction



Firming up



Stretch mark attenuation



Scar attenuation



Body contouring



Reduction of the adipose tissue



Improvement of cellulite

The time for body treatments is set by the equipment treatment protocol. To perform the treatment
simply, proceed as follows:
 First heat the treatment area for a brief space of time, massaging it at a low power (depending
on the client’s tolerance of the initial heat).

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 75 de 119
MKT-07-M1/20160623/MAN00159

 After heating up, the power must be progressively raised (until completing the treatment)
reaching a temperature on the skin that allows the desired effect for treatment to be obtained.
Always work at a power according to the client response to the heat generated.

To work with the multipolar 6 pin or multipolar 4 pin body accessory:
 The movements may be using the body multipolar accessory for reduction movements and
according to the area treated, circular with pressure or up and down, always with pressure, in
order to obtain a beneficial reduction of cellulite and contouring.

To work with the monopole body accessory:
 To work with the monopole accessory (always with the passive plate placed perfectly in contact
with the patient’s body) it is recommended to work, depending on the treatment zone, or circular
with pressure, or work alternating upward movements with pressure and downward without
pressure, in order to achieve contouring and firming up of the beneficial tissue. Each one of the
treatment images details the recommended massage mode.

Contraindications

Precaution
 Women who use IUD
 Metal implants
 Decreased thermal sensitivity due to lack of cutaneous sensitivity
 Do not use creams with cold or heat effect

Contraindications
 Pregnant women
 Lactating periods
 Risk of thrombosis
 Neoformations (cancer)
 Kidney disease
 Do not apply mentholated products, alcohol or products that accelerate cellular regeneration
(glycolic or fruit acid).
 Electric and electronic implants, either internal or connected to the exterior, with batteries or
controlled by radio, such as: pacemaker, neurostimulators, drug dosage devices, cochlear
hearing implants or external monitoring devices
 Internal haemorrhage processes in acute phase

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 76 de 119
MKT-07-M1/20160623/MAN00159

 Patients undergoing decogulation treatment
 In the first 48 h. of post-surgery for certain pathologies
 Bearers of metallic prostheses
 Patients with malign neoplasy
 Persons who are not temperature sensitive
 Active infectious processes
 Persons who suffer from uncompensated arthropathies

Secondary effects

 Redness on the area treated
 Feeling of high heat in the area treated for 30 minutes after the treatment

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 77 de 119
MKT-07-M1/20160623/MAN00159

DEVICE MAINTENANCE
Considerations for cleaning and maintenance



For safety reasons, before performing cleaning device, disconnect the power cord from the
wall outlet.



For parts of methacrylate, it should be up care in the use of cleaning products, if the material
comes into contact with substances such as alcohol, solvents, ammonia or other aggressive
products (thinners, acetone... ) reacts to form a " cracks " inside that they do not occur after
cleaning the device, but in the following days.



To clean this material is advisable to use a product -specific methacrylate as the "Vetril” or
use a cloth wrung after being submerged in water with diluted neutral soap.



To clean the device always uses cloth or "soft" roles especially for acrylic surfaces.



Cleaning and sanitizing applicators after each treatment device is encouraged. To do this,
remove remaining product (gel) deposited on the applicator after each treatment and clean and
disinfect the applicator before using again. Do not use products containing alcohol for

cleaning applicators.
 After cleaning the outside of the housing, dry all parts thoroughly, using synthetic fabrics before
starting the device or applicators.


For cleaning TFT screen must use a cotton cloth dampened with water previously drained.

 Completely avoid the use of sprays since they penetrate inside the device.
 Do not enter the ventilation slots on the device any cleaning product, or attempt to clean the
internal parts that look through them. For no reason the device should be disassembled for
cleaning: no need to clean the inside, and in the event that it was necessary cleaning should
only be performed by authorized for High Technology Products, SL specialized technical staff.
 Do not immerse the handpiece in liquid.

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 78 de 119
MKT-07-M1/20160623/MAN00159

TECHNICAL SPECIFICATIONS safyre

BASIC OPERATION
TYPE OF DEVICE

Radio frequency equipment and mechanical massage

RF EMISION

Monopolar / Multipolar Capacitive

TECHNICAL CHARACTERISTICS
MAXIMUM POWER

300 W

RADIO FREQUENCY POWER

150 W

RADIO FREQUENCY POWER

1 MHz

POWER
POWER

Single phase, 100-240 V

MAINS FREQUENCY

50 – 60 Hz.

MAXIMUM POWER

300W

REGULATION
Class to IEC 60601-1

Class I, Type BF

DIRECTIVE 2006/95/EEC (LVD) on low voltage
DIRECTIVE 2994/108/EEC (EMC) on electromagnetic compatibility
DIRECTIVE 2002/95/EEC (RoHS)
DIRECTIVE 2002/96/EEC (RAEE)
PHYSICAL CHARACTERISTICS
DIMENSIONS

430 x 300 x 400 mm

DIMENSIONS WITH STAND (optional)

403 x 1015 x 500 mm

WEIGHT

20 Kg

SCREEN

Touchscreen 10, 4 “

WORK RANGE

18-30 °C
2-50 °C
<90% Humidity (without condensation)

safyre - Manual de usuario/User manual/Manuel de l’utilisateur

Página 79 de 119
MKT-07-M1/20160623/MAN00159



Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.6
Linearized                      : No
Author                          : marketing4
Create Date                     : 2016:07:05 16:13:10+02:00
Modify Date                     : 2016:07:05 16:34:19+02:00
XMP Toolkit                     : Adobe XMP Core 5.2-c001 63.139439, 2010/09/27-13:37:26
Format                          : application/pdf
Creator                         : marketing4
Title                           : Microsoft Word - 1.doc
Creator Tool                    : PScript5.dll Version 5.2.2
Metadata Date                   : 2016:07:05 16:34:19+02:00
Producer                        : Acrobat Distiller 10.1.16 (Windows)
Document ID                     : uuid:8f1a64a8-6bd1-4d8f-86eb-8ed224ff4120
Instance ID                     : uuid:7e8eedd1-6660-4ca5-a533-9fc958a5456b
Page Count                      : 38
EXIF Metadata provided by EXIF.tools

Navigation menu