Dr ger Safety and KGaA AAC00XX area monitor User Manual 9033112 02 2009 de

Dräger Safety AG & Co. KGaA area monitor 9033112 02 2009 de

UserMan

Download: Dr ger Safety and KGaA AAC00XX area monitor User Manual 9033112 02 2009 de
Mirror Download [FCC.gov]Dr ger Safety and KGaA AAC00XX area monitor User Manual 9033112 02 2009 de
Document ID1258668
Application IDgm1GJTMpAHcPQQQUubCF+g==
Document DescriptionUserMan
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize240.78kB (3009714 bits)
Date Submitted2010-03-30 00:00:00
Date Available2010-03-30 00:00:00
Creation Date2010-02-08 09:02:55
Producing SoftwareAcrobat Distiller 8.1.0 (Windows)
Document Lastmod2010-02-08 15:17:07
Document Title9033112_02_2009_de.book
Document CreatorFrameMaker 8.0
Document Author: engelkec

Dräger X-zone 5000
de
Gebrauchsanweisung
2
en
Instructions for Use

fr
Notice d’utilisation

es
Instrucciones de uso

it
Istruzioni per l'uso
nl
Gebruiksaanwijzing

da
Brugsanvisning

no
Bruksanvisning

03333112.eps

DRAFT 02 - 08/02/10
Contents
For Your Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
What is what . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Switching on the instrument . . . . . . . . . . . . . . . . . . 5
Dräger X-am 5x00 info mode . . . . . . . . . . . . . . . . . 6
Switching off the device . . . . . . . . . . . . . . . . . . . . 6
Power supply for the Dräger X-am 5x00
with NiMH battery in the Dräger X-zone 5000: . . . 6
Perform a function test with gas . . . . . . . . . . . . . . 7
Perform a connection test . . . . . . . . . . . . . . . . . . . 7
Safety housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Alarm-attenuation ring . . . . . . . . . . . . . . . . . . . . . . 7
Establishing a wireless connection . . . . . . . . . . . . 8
Connecting devices via a cable connection . . . . . 10
Relay output: . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
During operation . . . . . . . . . . . . . . . . . . . . . . . . . 12
Alarms (default settings) . . . . . . . . . . . . . . . . . . . . 12
Concentration pre-alarm A1 . . . . . . . . . . . . . . . . . 13
Concentration main alarm A2 . . . . . . . . . . . . . . . 13
Battery pre-alarm . . . . . . . . . . . . . . . . . . . . . . . . . 13
Battery main alarm . . . . . . . . . . . . . . . . . . . . . . . . 13
Device alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Operation with pump (optional) . . . . . . . . . . . . . . 14
Commissioning and performing the
measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Observe the following during measuring
mode with pump . . . . . . . . . . . . . . . . . . . . . . . . . 15
Device configuration . . . . . . . . . . . . . . . . . . . . . . . 15
Faults, Cause, Remedy . . . . . . . . . . . . . . . . . . . . . 16
Fault messages . . . . . . . . . . . . . . . . . . . . . . . . . 18
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Maintenance intervals . . . . . . . . . . . . . . . . . . . . . 19
Replacing the batteries . . . . . . . . . . . . . . . . . . . . 19
Charging the batteries . . . . . . . . . . . . . . . . . . . . . 20
Care . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Disposing of the device . . . . . . . . . . . . . . . . . . . . . 20
Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Key allocation of Dräger X-zone 5000 . . . . . . . . . . 22
Status LED Overview . . . . . . . . . . . . . . . . . . . . . . . 22
Overview of LED Ring and Horn Signals . . . . . . . 24
Order List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
DRAFT 02 - 08/02/10
For Your Safety
Strictly follow the Instructions for Use
Any use of the device requires full understanding and
strict observation of these Instructions for Use. The
device is only to be used for the purposes specified
here.
Maintenance
The device must be inspected and serviced regularly by
trained service personnel. Repair of the device may only
be carried out by trained service personnel.
We recommend that a service contract be obtained with
Dräger and that all repairs also be carried out by Dräger.
Only genuine Dräger parts should be used for
maintenance.
Strictly follow the instructions in the chapter
"Maintenance intervals" on Page 18.
Accessories
Do not use accessory parts other than those specified in
the order list Page 24.
Safe coupling with electrical devices
Electrical connections to devices which are not listed in
these Instructions for Use should only be made
following consultation with the respective manufacturers
or an expert.
Use in areas subject to explosion hazards
Devices or components for use in explosion-hazard
areas which have been tested and approved according
to national, European or international Explosion
Protection Regulations may only be used under the
conditions specified in the approval and with
consideration of the relevant legal regulations.
The equipment or components may not be modified in
any manner. The use of faulty or incomplete parts is
forbidden. The appropriate regulations must be
observed at all times when carrying out repairs on
these devices or components.
Safety symbols used in these Instructions for Use
These Instructions for Use contain a number of
warnings for risks and hazards which might occur when
using the instrument. These warnings contain signal
words which will alert you to the degree of hazard you
may encounter. These signal words and corresponding
hazards are as follows:
CAUTION
Indicates a potential hazardous situation which, if not
avoided, could result in injury or damage to property.
It may also be used to alert against unsafe practices.
NOTICE
Additional information on how to use the device.
Intended Use
The Dräger X-zone 5000 is an explosion-proof, portable
alarm amplifier for the quasi-stationary monitoring of
hazard areas. The Dräger X-zone 5000 can be used in
combination with the Dräger X-am 5000/5600 gas
detection instruments. Several Dräger X-zone 5000
devices can operate in a self-crosslinking network.
NOTICE
The Dräger X-zone 5000 is intended for detection in
ambient air. Any increased exposure to certain
hydrocarbons can result in restrictions in the detection
quality of the electrochemical sensors.
Description
The Dräger X-zone 5000 alarm amplifier is intended for
use under industrial conditions within a specified
temperature range for an uninterrupted operating period
of up to 5 days.
Via a wireless connection and/or communication cable,
several Dräger X-zone 5000 devices can be grouped to
form an alarm chain. In the event of an alarm, all
connected Dräger X-zone 5000 devices will issue the
alarm.
The alarm is generated using the gas detection
instrument
Dräger X-am 5x00. The Dräger X-am 5x00 is connected
via an infrared interface
to the Dräger X-zone 5000. In addition to alarm
generation, the Dräger X-am 5x00 is used as the user
interface for the Dräger X-zone 5000. When the Dräger
X-am 5x00 generates a gas alarm, this is transmitted to
the Dräger X-zone 5000 and amplified acoustically and
optically.
DANGER
Indicates an immediate hazardous situation which, if
not avoided, could result in death or serious injury.
WARNING
Indicates a potential hazardous situation which, if not
avoided, could result in death or serious injury.
DRAFT 02 - 08/02/10
What is what
Front
Carrying handle
LED indicators
LED alarm ring
Inductive charging station
Lock
Device receptacle
Dräger X-am 5x00 holder diffusion mode
Dräger X-am 5x00 holder pump mode (optional)
x-zone 5000
00133112.eps
Rear side
9 Quick-reference guide
10 Switch relay connection / RS485 connection
11 Charging port connection / RS485 connection
00233112.eps
10
11
12
13
14
15
16
17
18
DRAFT 02 - 08/02/10
21
20
19
00333112.eps
Display
12 IR interface
13 Battery LED
14 Data transmission LED
15 Pump LED
16 Grouping LED
17 /wireless network key
18 OK key
19 Power contacts for Dräger X-am 5x00
20 Pump inlet (optional)
21 Pump outlet (optional)
Operation
ON mode
NOTICE
Only Dräger X-am 5x00 devices of software version
3.5 and later are compatible with Dräger X-zone 5000.
Switching on the instrument
STANDBY mode
00433112.eps
z Release the lock on
the holder.
z Remove the holder.
z Hold down the OK key
on the
Dräger X-zone 5000
for approx. 3 seconds.
– The battery LED will be
lit in green, red/green
or red (refer to “Status
LED Overview” on
page 21) depending on
the battery capacity.
– The Dräger Xzone 5000 is in the
STANDBY mode.
CAUTION
When switching on the Dräger
X-zone 5000, Dräger recommends wearing hearing
protection or using the alarm-attenuation ring (order
no. 83 20 110) as the acoustic alarm is activated for a
brief period.
z Switch on the Dräger X-am 5x00 in accordance with
the instructions for use.
– The visual and the audible alarm will be activated for
a short time.
– The Dräger X-am 5x00 switches to the X-zone mode
(refer to “X-zone mode:” on page 5).
– The visual and acoustic life signal (green LED ring
and single tone) is issued depending on the
configuration (1 - 60 seconds; default setting:
2 seconds).
– The Dräger X-zone 5000 is in the ON mode and
ready for operation.
– In ON mode, the alarm signals of the Dräger X-am
5x00 are evaluated, processed, and, if applicable,
transmitted to other Dräger X-zone 5000 devices.
00533112.eps
X-zone mode:
– The battery symbol
on the Dräger X-am 5x00 is
replaced by the X-zone mode symbol on the
Dräger
X-zone 5000.
– The Dräger X-am 5x00 is supplied via the Dräger
X-zone 5000 battery.
– If the activated Dräger X-am 5x00 is removed from
the Dräger X-zone 5000, the device will exit X-zone
mode again after max. 10 seconds.
z Place the Dräger X-am 5x00 into the device
receptacle.
NOTICE
The Dräger X-am 5x00 must be equipped with a NiMH
battery.
The clip on the Dräger X-am 5x00 must be connected.
z Place the holder onto the housing.
z Fasten the lock at the holder.
DRAFT 02 - 08/02/10
Dräger X-am 5x00 info mode
Switching off the device
z Press and hold the OK key of the Dräger X-am 5x00
for approx. 3 seconds in detection mode.
– Press the OK key successively for the next display.
The peak values and the exposition values TWA and
STEV as well as additional X-zone information will
be displayed.
If there are warnings or faults, the corresponding
notes or error codes (e. g. X01) are displayed (refer
to “Fault messages” on page 17).
CAUTION
When switching off the Dräger
X-zone 5000, Dräger recommends wearing hearing
protection or using the alarm-attenuation ring (order
no. 83 20 110) as the acoustic alarm is activated for a
brief period.
STANDBY mode
z Switch off the Dräger X-am 5x00 in the Dräger Xzone 5000 in accordance with the instructions for
use.
– The visual and the audible alarm will be activated for
a short time before the Dräger X-zone 5000 switches
off.
– The Dräger X-zone 5000 switches to the STANDBY
mode.
Info window for Dräger X-am 5x00:
RF: ON
NET: 005
P\N: FFFF
OFF mode
02533112.eps
--
1 Station number in grouping mode:
The station number is not currently supported by the
Dräger X-zone 5000.
2 Wireless status: ON/OFF:
Displays whether the wireless function of the Dräger
X-zone 5000 is switched on or off.
The wireless function can be switched on and off
using the Dräger CC-Vision PC software (refer to
“Device configuration” on page 14).
3 Network number:
If independent radio networks are required, different
network numbers must be assigned (refer to
“Establishing a wireless connection” on page 8).
The network number can be set using the Dräger
CC-Vision PC software (refer to “Device
configuration” on page 14).
4 Grouping ID:
The grouping ID is not currently supported by the
Dräger X-zone 5000.
5 X-zone mode symbol
Displays whether or not there is a connection
between the Dräger X-zone 5000 and the Dräger Xam 5x00.
z Release the lock at the holder.
z Remove the holder.
z If necessary, remove the Dräger X-am 5x00 from the
device receptacle.
z Press the OK key and the
key on the Dräger Xzone 5000 and hold down for approx. 3 seconds.
– The battery LED goes off.
– The Dräger X-zone 5000 is switched off
(OFF mode).
Power supply for the Dräger X-am 5x00
with NiMH battery in the Dräger X-zone
5000:
Device mode
Dräger X-zone 5000: ON mode
Dräger X-am 5x00: Switched on
Power supply for the
Dräger X-am 5x00
Dräger X-am 5x00 is
permanently supplied with
power.
Dräger X-zone 5000: STANDBY mode Dräger X-am 5x00 is
Dräger X-am 5x00: Switched off
supplied with power via
trickle charging.
Dräger X-zone 5000: OFF mode
Dräger X-am 5x00: Switched off
– If no key is pressed for 10 seconds, the Dräger X-am
5x00 returns automatically to detection mode.
DRAFT 02 - 08/02/10
Dräger X-am 5x00 is not
supplied with power.
Perform a function test with gas
Perform a connection test
The connection test can be used to check the correct
connection to all Dräger X-zone 5000 devices.
z Press the OK key on of the the Dräger X-am 5x00
devices being used three times in succession.
A signal via a horn and LED ring is issued three times
on every device connected wirelessly or via a cable.
CAUTION
A function test must be carried out on every device
before use.
The function test can be carried out in two ways.
Function test for the Dräger X-am 5x00:
z Perform a function test in accordance with the
instructions for use for the gas detection instrument
before using with the Dräger X-zone 5000.
Safety housing
Function test on the Dräger X-am 5x00 in
combination with the Dräger X-zone 5000:
NOTICE
The function test can only be carried out with the
Dräger X-am 5x00 holder diffusion mode (order no.
83 20 636).
z Switching on Dräger X-zone 5000 (refer to
“Switching on the instrument” on page 5).
z Place adapter (1) (order no. 83 20 108) on the holder
(diffusion).
z Connect the test gas cylinder to the adapter (2).
z Open the test gas cylinder valve to let test gas flow
over the sensors.
z Wait until the instrument displays the test gas
concentration with sufficient tolerance:
Ex: ±20 %1)
O2: ±0.8 Vol.-%1)
TOX: ±20 %1).
– Depending on the test gas concentration, alarm A1
or A2 is issued when the alarm thresholds are
exceeded.
z Close the test gas cylinder valve and remove the
adapter from the holder.
If the displays are outside of the above-mentioned
ranges: Have the Dräger X-am 5x00 calibrated by
service personnel.
1) Upon application of the Dräger mixed gas (order no. 68 11 130)
the displays should be within this range. Different concentrations
can be set using the Dräger CC-Vision PC software provided.
In exceptionally rough environments, and in very strong
sunlight and high temperatures (> +40 °C), the use of
the safety housing (order no. 83 21 519) is
recommended.
1 Hose/cable holder
Alarm-attenuation ring
WARNING
The alarm-attenuation ring must not be used in
explosion-hazard areas!
03033112.eps
02733112.eps
02933112.eps
When switching on and off and during the function test
of the Dräger X-zone 5000, Dräger recommends
wearing hearing protection or using the alarmattenuation ring (order no. 83 20 110) as the acoustic
alarm is sounded for a short time.
DRAFT 02 - 08/02/10
Establishing a wireless connection
Sub-networks:
WARNING
A maximum of 25 Dräger X-zone 5000 devices may be
connected on a wireless network as otherwise it is not
possible to ensure a reliable connection and alarm
forwarding.
– this device may not cause harmful interference,
and
– this device must accept any interference received,
including interference that may cause undesired
operation.
Modifications not expressly approved by this company
could void the user's authority to operate the
equipment. The internal / external antennas used for
this mobile transmitter must provide a separation
distance of at least 20 cm from all persons and must not
be co-located or operating in conjunction with any other
antenna or transmitter. This device and its antenna
must not be co-located or operating in conjunction with
any other antenna or transmitter.
NOTICE
If the devices are in ON mode, they automatically
establish the wireless connection.
1 Sub-network 1
2 Sub-network 2
3 e. g. steel wall, wall, HGV, etc.
Sub-networks are created via an unintentional division
of the open network into two or more radio networks.
This can occur if the user is running two activated
Dräger X-zone 5000 (with the same network number).
This means that the devices have permanently
interconnected wireless connection (data transmission
LED is lit in green). This means that the user can no
longer tell whether or not the wireless connection is
already set up with the devices already installed. To
avoid sub-networks, the Dräger X-zone 5000 devices
should always be set up and switched on in succession.
Positioning the device:
Up to 25 devices can be connected on a wireless
network via a wireless connection.
The typical radio range is up to 100 m in industrial
surroundings (environmental factors may affect the
range).
The devices can be operated in a chain, star or ring
topology. It is possible to establish an open radio
network or independent radio networks. All devices
within radio range are automatically connected.
NOTICE
The gas supply must be connected at an angle of 360°.
If required, use base (order no. 83 20 645) to raise the
detection position by approx. 30 cm.
Open wireless network (standard configuration):
With an open wireless network, any number of Dräger
X-zone 5000 devices (up to 25) can be added to or
removed from a wireless network number.
NOTICE
When positioning the devices, ensure that subnetworks are avoided.
Independent radio networks:
If independent radio networks are required, different
network numbers need to be assigned for the relevant
radio networks (refer to “Device configuration” on
page 14).
Example:
Two independent radio networks each with three Dräger Xzone 5000 devices should be created. To do this, the
Dräger CC-Vision PC software should be used to set
NET:001 with three Dräger X-zone 5000 devices and
NET:002 for three further Dräger X-zone 5000 devices.
CAUTION
Before positioning the devices, a function test (refer to
“Perform a function test with gas” on page 7) must be
conducted on every device.
z Switch on the first Dräger X-zone 5000 (refer to
“Switching on the instrument” on page 5) and
position at the relevant location.
z Switch on the second Dräger Xzone 5000 and position at a
distance so that the data
transmission LED is lit in green.
– The wireless connection is
established and the data
transmission LED indicates the
connection to at least one device
within radio range.
DRAFT 02 - 08/02/10
01333112.eps
FCC and IC:
This device complies with Part 15 of the FCC Rules.
Operation is subject to the following two conditions:
03533112.eps
CAUTION
WARNING
If the data transmission LED is lit in red, the distance
from the next device must be reduced.
z Position other Dräger X-zone 5000 devices using the
same process.
Examples of use:
Setting A:
Monitoring industrial tanks
(Wireless alarm chain)
01833112.eps
NOTICE
Dräger recommends performing a function test (refer to
“Perform a connection test” on page 7) after positioning
all the devices.
Setting B:
Area monitoring with pump
01933112.eps
Setting C:
Radio monitoring of pipelines
(Wireless alarm chain)
02033112.eps
DRAFT 02 - 08/02/10
Connecting devices via a cable connection Pin configuration XEXT1 / XEXT2 on the device:
CAUTION
Before positioning the devices, a function test (refer to
“Perform a function test with gas” on page 7) must be
conducted on every device.
XEXT2 (male)
NOTICE
Due to widely differing customer-specific requirements,
Dräger does not provide any communication cables. All
relevant cable parameters are described in this
chapter.
Relay output
4 Normally Closed (NC)
5 Normally Open (NO)
6 Closed Only (CO)
7 GND
If radio connections are not allowed or blocked, the
devices can be connected via communication cable.
The maximum cable length between two devices is
25 m.
XEXT1 (female)
RS485
1 PLUS
2 MINUS
3 GND
NOTICE
Combined wireless and cable operation is possible.
Charger
4 Additional voltage (U-In)
5 GND2
01233112.eps
01333112.eps
z Switching on Dräger X-zone 5000 (refer to
“Switching on the instrument” on page 5).
z Plug the communication cable into the charging port/
RS485 connection (1) on the rear side of the device.
z Connect the end of the communication cable to the
switch relay/RS485 connection (2) of the second
device.
– The data transmission LEDs of the
coupled devices light green.
WARNING
If the data transmission LED is lit in red, check the
cable connection.
z If required, connect further devices via
communication cable as described above.
NOTICE
Dräger recommends performing a function test (refer to
“Perform a connection test” on page 7) after positioning
all the devices.
10
DRAFT 02 - 08/02/10
XEXT2
6 1
XEXT1
1 6
02133112.eps
RS485
1 PLUS
2 MINUS
3 GND
Relay output:
RS485 connection:
XEXT2
XEXT1
CO
NO
NC
02333112.eps
03433112.eps
CAUTION
Ensure that the cable strands are not crossed!
The relay output is intended for connection to an
isolation amplifier with intrinsically safe output.
WARNING
It is essential that the parameters of the relay outputs
are observed to prevent the circuit from jeopardising
the intrinsic safety of the device.
Only intrinsically safe circuits are permitted for
connection.
NOTICE
The person responsible for the use of the Dräger Xzone 5000 must prepare a system document to verify
the intrinsic safety level.
max. 25 m
XEXT2
03133112.eps
02433112.eps
1 Plug (male):
Housing: Souriau UTS6JC147P (male)
Contacts: Souriau RM20M12K (male)
2 Plug (female):
Housing: Souriau UTS6JC147S (female)
Contacts: Souriau RC20M12K (female)
3 Cable type: Belden 3107A, 2 x 2 AWG 22
Cable length: max. 25 m between 2 devices
max. 25 m
03233112.eps
1 Plug (female):
Housing: Souriau UTS6JC147S (female)
Contacts: Souriau RC20M12K (female)
2 Configure the pin assignment separately depending
on the application
3 Cable type: Belden 3107A, 2 x 2 AWG 22
Cable length: max. 25 m between 2 devices
11
DRAFT 02 - 08/02/10
Alarms (default settings)
Connection of the relay output
20 V
Maximum voltage (Ui):
Maximum continuous current 0.15 A
(Ii):
Maximum switching capacity 3 W
(Pi):
For ohmic loads only:
Reactances:
Ci ≤ 100 pF Li ≤ 10 µH
During operation
During operation the values for each gas detected are
displayed on the inserted Dräger X-am 5x00.
NOTICE
The displays on the Dräger X-am 5x00 are described in
the instructions for use for the gas detection instrument
being used.
The visual and acoustic life signal (LED ring green and
single tone) is issued depending on the configuration
(1 - 60 seconds; default setting every 2 seconds). The
life signal can be configured using the Dräger CC-Vision
PC software (refer to “Device configuration” on
page 14).
In the event of an alarm, the visual and the audible
alarm will be activated (refer to “Alarms (default
settings)” on page 11).
The Dräger X-zone 500 will amplify the visual and
audible alarm and will permanently transmit the alarm
information, via a wireless connection or via cable, to
further Dräger X-zone 5000 devices.
NOTICE
The alarm settings (e. g. self-latching/
acknowledgeable) can be configured using the Dräger
CC-Vision PC software. The configuration of the
Dräger X-am 5x00 is critical for the correct response of
the Dräger
X-zone 5000.
The Dräger X-zone 5000 is equipped with two different
alarm generators:
– Visual signal: LED ring (360°);
Colours red, green; pulsing.
– Acoustic signal: Intense horn (108 dB (A) in 1 m
distance/120 dB (A) in 30 cm distance).
Triggering device:
As soon as a device detects an increased gas
concentration, this device will become the triggering
device.
The triggering device forwards the alarms to all
connected devices wirelessly and/or via communication
cable.
Receiving device:
All devices that receive an alarm from the triggering
device become receiving devices. The receiving
devices generate a seconary alarm. If the receiving
device does not receive any information from the
triggering device, the secondary alarm on the receiving
devices is cancelled after 10 seconds.
NOTICE
The alarm-triggering device and the receiving devices
give a different visual alarm.
12
DRAFT 02 - 08/02/10
Concentration pre-alarm A1
Battery pre-alarm
The alarm is indicated by an
intermittent alarm message:
The alarm is indicated by an
intermittent alarm message:
Display » A1 « and measured value alternating: not for
O2!
Acknowledging the pre-alarm:
– The Battery LED flashes red.
z Press OK key on the Dräger X-am 5x00 of the
Dräger X-zone 5000, only the acoustic alarm is
switched off.
– After the first battery pre-alarm the battery will last for
another approx. 15 minutes.
– The pre-alarm A1 is not self-latching and stops when
the concentration has dropped below alarm
threshold A1.
The triggering device at A1:
– A single tone sounds and the LED ring flashes red
(main alarm).
The receiving device at A1:
– A single tone sounds and the LED ring flashes red/
green (seconary alarm).
Acknowledging the pre-alarm:
z Press OK key on the Dräger X-am 5x00 of the
triggering
Dräger X-zone 5000.
– Only the audible alarm will be switched off.
Concentration main alarm A2
DANGER
Danger to life! Leave the area immediately.
A main alarm is self-latching and cannot be
acknowledged (see Dräger X-am 5x00 Instructions for
Use).
The alarm is indicated by an
intermittent alarm message:
The triggering device at A2:
– A double tone sounds and the LED ring flashes red
twice (main alarm).
A receiving device at A2:
– A double tone sounds and the LED ring flashes red/
green twice (seconary alarm).
For O2: A1 = lack of oxygen
A2 = excess oxygen
WARNING
Before the area may be entered again, a clearance
measurement must be performed!
Battery main alarm
The alarm is indicated by an
intermittent alarm message:
The battery main alarm cannot be acknowledged
acoustically:
– The Battery LED flashes red.
– The device automatically switches off after
10 seconds.
The visual and the audible alarm will be activated for a
short time before the device switches off.
Device alarm
NOTICE
Indicates a malfunction of the Dräger X-zone 5000 or
the Dräger X-am 5x00.
The alarm is indicated by an
intermittent alarm message:
– The device or one or more sensor channels are not
ready for operation.
– Remedies, refer to “Faults, Cause, Remedy” on
page 15
z If necessary, commission the Dräger Safety Service
Centre to eliminate the error.
Acknowledge the device alarm
z Press the OK key on the Dräger X-am 5x00 or on the
Dräger
X-zone 5000.
Only after the concentration has dropped below the
alarm threshold A2:
z Press OK key on the Dräger X-am 5x00 of the
triggering
Dräger X-zone 5000.
– The alarm messages will be switched off.
13
DRAFT 02 - 08/02/10
Operation with pump (optional)
Pump test
The Dräger X-zone 5000 is equipped with a pump as an
option (refer to “Order List” on page 24.)
NOTICE
Pump test is performed in the same way as for the
Dräger pump
X-am 1/2/5000. During the pump test, in the default
setting the acoustic alarm is automaticallyreduced to
80 dB (A).
Commissioning and performing the
measurement
CAUTION
Pump operation only with filter (order no. 83 19 359)!
Otherwise there is a risk of damage to the pump.
If no filter is used during pump operation, the warranty
on the pump becomes invalid.
z Connect the sampling
hose with the water
trap for the hose set
(order no. 68 05 473)
and screw the
connecting bush onto
the filter in a clockwise
direction.
02633112.eps
z Seal or kink the intake flange or
sampling hose for approx.
2 seconds.
– The pump LED lights red.
– Unblock the intake flange/
sampling hose.
00833112.eps
z Release the lock on the holder.
z Remove the holder.
z Press the OK key of the Dräger X-zone 5000 and
hold it for approx. 3 seconds.
– The Dräger X-zone 5000 is in STANDBY mode.
z Place the Dräger X-am 5x00 into the device
receptacle.
z Place the holder (pump) onto the housing.
z Fasten the lock at the holder (pump).
z Switch on the Dräger X-am 5x00 in accordance with
the instructions for use.
– If the device is in the ON mode, the pump will be
automatically switched on via a switching contact on
the holder (pump).
– Pump LED flashes red/green.
00733112.eps
– The pump LED flashes red/green,
accompanied by a signal tone.
If the test was successful:
00933112.eps
– The pump LED lights green,
accompanied by an acoustic
acknowledgment signal.
z A pump test will then need to be performed. The
pump test will start automatically.
NOTICE
The pump test must be performed within 60 seconds
otherwise a device alarm is issued.
– The pump LED lights red,
accompanied by a continuous
tone.
– The pump switches off
automatically.
Ending operation
z Switch off the Dräger X-am 5x00 in accordance with
the instructions for use.
– The visual and the audible alarm will be activated for
a short time before the device switches off.
– The Dräger X-zone 5000 switches to STANDBY
mode.
z Unscrew the sampling hose or the Dräger probe
from the filter.
z Remove the holder (pump) by releasing the lock.
– The pump switches off.
z Place the holder onto the housing.
z Fasten the lock at the holder.
14
DRAFT 02 - 08/02/10
01033112.eps
00633112.eps
If the test was not successful:
Device configuration
z Wait for the flushing time to elapse:
Before every measurement, flush the Dräger
sampling hose or the Dräger probes with the air
sample to be measured.
– A flushing phase is necessary to eliminate or
minimise all effects associated with the use of a
sampling hose or a probe, e. g. absorption in the
hose, dead volume.
– The duration of the flushing phase depends on
factors such as type and concentration of the gas or
vapour to be measured, material, length, diameter,
and age of the sampling hose or probe. Generally,
when using a sampling hose (new, dry, clean), a
typical flushing time of approx. 10 seconds is
required for each metre. This flushing time applies in
addition to the sensor response time (see the
Instructions for Use for the gas detection instrument
used).
Example:
– In the case of a 10 m sampling hose, the flushing
time is approx. 100 seconds and the sensor
response time is in addition approx. 60 seconds.
Therefore, the total time before reading the gas
detection instrument is approx. 90 seconds.
– The flow-rate alarm is delayed by 10 to 30 seconds
depending on the length of the hose.
Replacing the filter
z Unscrew the connecting bush from the filter.
z Unfasten the lock at the holder (pump).
z Remove the holder (pump).
IR
02833112.eps
Observe the following during measuring
mode with pump
To individually configure a standard-configuration
device, connect the device to a PC.
Communication is carried out via the USB DIRA dongle
(order no. 83 17 409).
The Dräger CC-Vision PC software is used to perform
the configuration.
The following settings can be configured, among others:
– Horn volume
– Wireless function
– Alarm frequencies
– Alarm pattern
– Alarm forwarding
– Life signal (light pattern, horn volume)
– Bump test volume
(Default setting: 80 dB (A))
– Actions of the switch relay
NOTICE
Observe the documentation and online help of Dräger
CC Vision PC software.
z Unscrew the filter anticlockwise.
z Screw a new filter
(order no. 83 19 359)
onto the device in a
clockwise direction.
01433112.eps
NOTICE
A Dräger CC Vision version, which can be used for
Dräger X-zone 5000 PC software, is supplied with the
device on CD-ROM.
z Place the holder (pump) onto the housing.
z Fasten the lock at the holder (pump).
z Screw the connecting bush onto the new filter.
15
DRAFT 02 - 08/02/10
Faults, Cause, Remedy
Fault
Dräger X-zone 5000 cannot
be loaded.
Cause
Charging unit plug is not in correct
contact with the Dräger X-zone
5000.
Battery is not fully charged.
Battery not regularly charged.
Remedy
Ensure that the charging unit plug is
correctly plugged in.
Check the battery LED.
Battery service life is low.
Load battery for at least 12 h.
Charge battery regularly even when
switched off, at least every
2 months.
Outdoor temperature is too low.
Warm up Dräger X-zone 5000.
Outdoor temperature is very high.
Use safety housing where
appropriate.
Battery is faulty.
Have the battery checked by the
Dräger service department.
Inductive charging is not functioning Check the distance, remove any dirt
correctly as the distance between
if necessary.
the Dräger X-zone 5000 and the
charging station is too great.
No wireless connection between the Wireless function is not activated.
Activate the wireless function using
Dräger X-zone 5000 devices.
the Dräger CC-Vision PC software
(refer to “Device configuration” on
page 14).
Dräger X-zone 5000 devices are
Position the Dräger X-zone 5000
very far apart.
devices more closely together.
Position additional devices in the
chain.
Place Dräger X-zone 5000 at a
higher position, use base (order no.
83 20 645) where appropriate.
The wireless connection is made
Position the Dräger X-zone 5000
more difficult by an industrial
devices more closely together.
environment: e. g. steel walls.
Position additional devices in the
chain.
Connect the Dräger X-zone 5000
with a communication cable (refer to
“Connecting devices via a cable
connection” on page 9).
Dräger X-zone 5000 are covered by Ensure there are no obstructions.
conductive materials (e. g. metal
grids).
Network number of the Dräger
Use Dräger X-zone 5000 devices
X-zone 5000 is different.
with the same network numbers.
The network number can be
configured using the Dräger CCVision PC software (refer to “Device
configuration” on page 14).
Wireless frequency of the Dräger
Use Dräger X-zone 5000 devices
X-zone 5000 devices is different.
with the same radio frequency.
Cable connection is not functioning. Cable plug is not correctly plugged Check cable connection and
in, cable assignment is incorrect or assignment. Ensure that the cable
cable is broken.
plug is correctly plugged in.
Alarm contact is not switching.
Cable plug is not correctly plugged Check cable connection and
in, cable assignment is incorrect or assignment. Ensure that the cable
cable is broken.
plug is correctly plugged in.
Pump operation is not functioning.
Incorrect holder (diffusion) in
Position the holder (pump).
position.
16
DRAFT 02 - 08/02/10
Fault
Pump fault during operation.
Cause
Condensate formation with cold and
damp intake air.
Pump outside the specified range.
Function test has failed.
Incorrect holder (pump) in position.
Adapter is incorrectly positioned on
the holder (diffusion).
Flow test has failed.
Flow test has not been performed.
Horn is too quiet.
Holder (pump) is not correctly
positioned.
Volume is set too low.
Optical alarm signals not visible or
poorly visible.
Configuration or pattern incorrectly
set.
Life signal is not functioning.
Configuration is set incorrectly.
Gas detection instrument not
detected.
IR interface dirty.
Incompatible gas detection
instrument.
Incorrect software version in the gas
detection instrument.
IR interface faulty, clip on gas
detection instrument not correctly
positioned.
Holder is not correctly positioned.
Dräger X-am 5x00 switches off
quickly; no power supply.
Device defect displayed.
Remedy
Use hose set with water trap (order
no. 83 21 527).
Have the pump checked by the
Dräger service department.
Use holder (diffusion).
Position adapter correctly on the
holder (diffusion) (refer to “Perform
a function test with gas” on page 7).
Perform flow test, repeat if
necessary.
Re-position holder (pump) and
check that it is correctly fitted.
Set the volume using the Dräger
CC-Vision PC software (refer to
“Device configuration” on page 14).
Configure the alarm signals using
the Dräger CC-Vision PC software
(refer to “Device configuration” on
page 14).
Configure the life signal using the
Dräger CC-Vision PC software
(refer to “Device configuration” on
page 14).
Clean IR interface.
Use Dräger X-am 5x00.
Have a software update performed
by the Dräger service department.
Close clip on Dräger X-am 5x00.
Power contacts dirty or damp.
Re-position holder and check that it
is correctly fitted.
Clean power contacts.
Dräger X-am 5x00 moved away
from Dräger X-zone 5000 during
operation.
Acknowledge alarm on Dräger Xzone 5000, switch off Dräger X-zone
5000.
17
DRAFT 02 - 08/02/10
Fault messages
Special symbol » « and
displayed numerical
code:
Cause
Remedy
01
Dräger X-am 5x00 with alkali supply unit.
02
04
Communication interrupted with
Dräger X-zone 5000.
Communication error with battery
controller
Dräger X-zone 5000.
Main battery alarm Dräger X-am 5x00.
Insert Dräger X-am 5x00 with battery
power pack.
Check IR interface on Dräger X-zone
5000 and on Dräger X-am 5x00.
Contact Dräger service department.
05
Battery pre-alarm Dräger X-am 5x00.
06
Dräger X-am 5x00 charging current too
low.
Holder (pump) detected, but no pump
fitted.
Flow fault
Holder status change (pump) during
operation.
Check sum error program code
Check sum error operating parameter
Check sum error operating parameter
Check sum error operating parameter
Working memory test error
Faulty ADC conversion.
No contact with switch box in grouping
mode.
Charging electronics faulty.
Battery completely discharged.
Battery main alarm of the Dräger X-zone
5000.
Battery pre-alarm of the Dräger X-zone
5000.
Dräger X-am5x00 device error.
Dräger X-am5x00 alarm pattern faulty.
Battery main alarm
Battery completely discharged.
Charging electronics faulty.
Communication error with battery
controller
Dräger X-zone 5000.
03
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23 - X28
29
30
31
32
Check charging contacts on Dräger Xzone 5000 and on Dräger X-am 5x00.
Check charging contacts on Dräger Xzone 5000 and on Dräger X-am 5x00.
Check charging contacts on Dräger Xzone 5000 and on Dräger X-am 5x00.
Use holder for diffusion mode.
Check intake hose.
Check that the holder (pump) is securely
positioned.
Contact Dräger service department.
Contact Dräger service department.
Contact Dräger service department.
Contact Dräger service department.
Contact Dräger service department.
Contact Dräger service department.
Check wireless connection to switch box.
Contact Dräger service department.
Charge Dräger X-zone 5000.
Charge Dräger X-zone 5000.
Charge Dräger X-zone 5000.
Check Dräger X-am 5x00.
Contact Dräger service department.
Charge Dräger X-zone 5000.
Charge Dräger X-zone 5000.
Contact Dräger service department.
Contact Dräger service department.
18
DRAFT 02 - 08/02/10
Maintenance intervals
The device should be inspected annually by suitably
qualified personnel.
z Charge the lead battery after each application, at the
latest however after the battery alarm has been
triggered.
z Maintenance by suitably qualified personnel – every
year.
– The inspection intervals must be established in each
individual case and shortened if necessary,
depending on technical safety considerations,
engineering conditions, and the technical
requirements of the equipment.
– We recommend that a service agreement be
concluded with Dräger Safety and that repairs also
be carried out by Dräger.
NOTICE
For the gas detection instruments used, the
maintenance intervals specified in the relevant
instructions for use apply.
z Release the four M5
nuts.
z Disconnect the cable
connections from the
bottom plate.
z Replace the old battery
block with a new one.
z Re-establish the cable
connection to the
bottom plate.
z Check the correct
position of the O-ring.
z Tighten the four M5
nuts.
01633112.eps
Maintenance
z Establish plug connection to the bottom plate.
z Place the upper part of the housing on the bottom
part
(Note preferred position).
z Tighten the screws (M5 cylinder-head screw with
internal hexagon) at the housing bottom (120 Ncm
±20 Ncm).
NOTICE
It is recommended to fully charge the device after the
battery block has been replaced.
Replacing the batteries
WARNING
WARNING
Explosion hazard!
Do not replace the batteries in areas subject to
explosion hazard! Batteries are part of the Ex approval.
Only the following types may be used:
– Battery pack, small - order no. 83 20 644
– Battery pack, large - order no. 83 20 646
Explosion hazard!
Do not throw used batteries into a fire and do not open
them with force.
Dispose of the batteries in accordance with national
regulations.
z Unfasten the screws
(M5 cylinder-head
screw with internal
hexagon) on the
bottom side of the
housing.
z Lift the upper part of
the housing and
disconnect the plug
connection from the
bottom plate.
01533112.eps
z Switch off the device (refer to “Switching off the
device” on page 6).
19
DRAFT 02 - 08/02/10
Charging the batteries
Inductive charging
WARNING
Explosion hazard!
Do not charge underground or in explosion-hazard
areas.
The chargers are not designed in accordance with the
guidelines for firedamp and explosion protection.
CAUTION
Mains powered charging station should only be
handled by persons wearing implants if the
pacemakers and active implants conform to the
relevant legal requirements. Dräger only guarantees
conformity with Directive 2004/108/EC.
NOTICE
The transmitter coil of the inductive charging station
generates a weak magnetic alternating field. During
operation, all requirements of the relevant standards
regarding electromagnetic faults are observed. The
legal requirements of Directive 2004/108/EC are met.
NOTICE
The warranty on the battery becomes null and void if
the device is not fully charged at least every 2 months
when not in use.
Dräger recommends storing devices in the charging
station (order no. 83 20 626) even when not in use.
To maintain the lifetime of the batteries, charging is
temperature controlled and only performed in a
temperature range of 5 to 35 °C. When this temperature
range is left, the charging is automatically interrupted
and automatically continued after the temperature
range has been reached again.
During the charging, the battery LED flashes red, red/
green or green, depending on the battery status, at a
frequency of one Hz. As soon as the charging is
completed, the battery LED will permanently light green.
z Connect the charging station to the mains using the
power pack.
z Place the device onto the charging station.
– The charging time is typically < 10 hours.
Cabled charging
z Plug the charging cable into the charging port on the
rear side of the device.
z Connect the power pack to the mains.
– The charging time is typically < 14 hours.
NOTICE
Dräger recommends storing devices that are not in use
in the charging station (order no. 83 20 626).
Care
The device does not need any special care.
z Dirt and deposits can be removed from the device by
washing it with cold water. A sponge can be used for
wiping if necessary.
z Carefully dry the device using a cloth.
Disposing of the device
Disposing of electric and electronic equipment:
EC-wide regulations for the disposal of electric
and electronic appliances which have been
defined in the EC Directive 2002/96/EC and in
national laws have been effective since August
2005 and apply to this instrument.
Special collecting and recycling options have been
established for households. However, as this device has
not been registered for household usage, it must not be
disposed of through these means. The device can be
returned to your national Dräger Sales Organisation for
disposal. Please do not hesitate to contact the above if
you have any further questions on this issue.
20
DRAFT 02 - 08/02/10
Technical Data
Dimensions:
Weight:
with battery, 12 Ah
with battery, 24 Ah
Ambient conditions:
During operation
During storage
Alarms:
Visual, 360° LED
Audible, 360°
Battery:
Operating time, 12 Ah
(≥ 20 °C)
Operating time, 24 Ah
(≥ 20 °C)
Charging time, 12 Ah
Charging time, 24 Ah
Pump:
Networking of
devices:
approx. 490 x 300 x 300 mm
(H x W D)
approx. 7 kg
approx. 10 kg
-20 °C to +50 °C
-20 °C to +70 °C
700 to 1300 hPa
max. 95% relative humidity
green life signal;
red life signal;
green/red subsidiary alarm
108 dB (A) at 1 m distance
120 dB (A) at 30 cm distance
60 Hours at 15 minute alarm
per day and ‚
fully equipped
Dräger X-am 5x00
120 hours
< 6 hours
< 10 hours
Relay output:
Max. voltage (Ui):
Max. switching current
(Ii ):
Max. continuous
current:
Max. switching capacity
(Pi ):
For ohmic loads only:
Reactances
20 V
0.25 A
0.25 A
3W
Ci ≤ 100 pF; Li ≤ 10 µH
Capacitive or inductive
loads
No connection with:
Approvals:
ATEX:
I M1 Ex ia I Ma
II 1G Ex ia IIC T3 Ga
IECEx:
Ex ia I Ma
Ex ia IIC T3 Gb
Detection range:
see the Technical Manual for
Dräger X-am 5x00
Frequency ranges:
up to 30 m hose
0.5 l/min
Only service department
can switch the frequency
429/433/868/915 MHz
with a typical range of 100 m in
industrial surroundings
(environmental factors may
affect the range).
Automatic establishment of
wireless connection.
Country, region
Frequency range (MHz)
EU, Switzerland, Norway,
Turkey
868
South Africa
868
USA/Canada
915
Singapore
868
Australia
915
India
915
Japan
429
Russia
433
Up to 25 devices can be
connected on a wireless
network.
Combined wireless/cable
operation possible.
21
DRAFT 02 - 08/02/10
Key allocation of Dräger X-zone 5000
The following key functions refer to the keys of the Dräger X-zone 5000 located underneath the holder.
Action
Meaning
Pressing
OK
key once
Acknowledges Dräger X-zone 5000 device defect
Pressing
OK
key once
Acknowledges Dräger X-zone 5000 battery pre-alarm
Pressing and holding
Pressing
OK
key for 3 sec
- Switches to grouping mode, possible in ON or STANDBY mode.
- During grouping mode: Exits grouping mode.
key once
Pressing and holding
Switches from OFF to STANDBY mode.
OK
Pressing and holding
and
key for 3 sec
key for 3 sec
Switches to the OFF mode.
Deletes the grouping information.
Status LED Overview
Battery LED
(Indicates the battery status.)
Colour
off
LED status
Device mode
Meaning
off
OFF mode
Device is switched off.
on
ON/STANDBY mode
green
red/green
Battery capacity of more than 66 %.
red
Battery capacity of more than 33 %.
Battery capacity of less than 33 %.
Battery pre-alarm
red
ON/STANDBY mode
red
red
OFF mode
Battery main alarm; Dräger X-zone 5000 switches off after
10 sec.
Attempt to switch on in OFF mode when the battery is
empty (on for 10 s).
Dräger X-Zone 5000 is being charged, battery capacity
less than 33 %.
red
red/green
ON/STANDBY mode
in charging station
Dräger X-Zone 5000 is being charged, battery capacity
between 33 % and 66 %.
green
Dräger X-Zone 5000 is being charged, battery capacity
greater than 66 %.
green
Dräger X-zone 5000 is fully charged.
Data transmission LED
(Indicates the connection status if multiple devices are coupled via wireless connection or communication cable.)
Colour
off
LED status
off
Device mode
Meaning
ON/STANDBY mode
Wireless connection deactivated.
green
red
on
ON mode
At least 1 other Dräger X-zone 5000 coupled via wireless
connection or cable is detected.
No other Dräger X-zone 5000 coupled via wireless or
cable connection is detected.
22
DRAFT 02 - 08/02/10
Pump LED
(Indicates the pump status.)
Colour
off
off
LED status
off
Device mode
Meaning
STANDBY mode
Device is in STANDBY mode.
No pump adapter detected.
Flow test required.
red/green
red
ON mode
green
red
Performing flow test.
Flow test successful/pump is running.
on
Flow error (e. g. due to lack of volume flow or no flow
test).
Grouping LED
Not currently active - for use with future functions!
23
DRAFT 02 - 08/02/10
Overview of LED Ring and Horn Signals
Signal name
LED ring
Horn
OFF mode
Switch-on signal and switch-off signal
Off
All red LEDs on for 1 s,
then all green LEDs on for 1 s
and all status LEDs on for 1 s.
Continuous tone for 1 s with reduced1)
volume
Dräger X-zone 5000 device defect,
Intermittent triple flashing of the red
LEDs
Intermittent triple tone at full 3) volume
Intermittent double flashing of the red
LEDs
Intermittent single flashing of the red
LEDs
Intermittent double tone at full volume3)
Intermittent triple flashing of the
Intermittent triple tone at full 3) volume 5)
2)
triggering device
Concentration main alarm, triggering device2)
Concentration pre-alarm, triggering device2)
Dräger X-zone 5000 device defect,
receiving device
4)
Concentration pre-alarm, receiving device4)
Concentration main alarm, receiving device4)
Off
Intermittent single tone at full 3) volume
5)
red+green LEDs
Intermittent double flashing of the red
Intermittent triple tone at full 2) volume 5)
and green LEDs5)
Intermittent single flashing of the red and Intermittent triple tone at full 1) volume 5)
green LEDs5)
Acknowledgment signal
Prompting signal
Sustained single tone at full 1) volume
Intermittent (1 Hz) single tone at
reduced1) volume
Life signal
(suppressed in pump test)
Intermittent single flashing of the green
LEDs
Intermittent single tone at full 3) volume6)
Switching on at battery capacity < battery main alarm
Intermittent triple tone at reduced1)
volume
Battery pre-alarm of the Dräger X-zone 5000
Intermittent triple tone at full 3) volume
Battery main alarm of the Dräger X-zone 5000
Intermittent triple flashing of the red
LEDs
Intermittent triple flashing of the red
LEDs for 10 s, after that the Dräger Xzone 5000 switches to the OFF mode
Flow test prompt
flow test running
Flow test successful
1 Hz red
LED ring green for 2 s
Intermittent double tone with full3)
volume for 10 s, afterwards the
Dräger X-zone 5000 does to OFF mode
Intermittent (1 Hz) single tone.
1) Reduced volume for hearing protection: 80 dB (A) (default setting), the reduced volume must not exceed the "full" volume configured
by the customer.
2) Device that triggered the alarm.
3) Full volume: Maximum volume configured by the customer (e. g. 108 dB (A)).
4) Device that has received the alarm from the triggering device.
5) If user has activated the forwarding of defects from the receiving device.
6) Frequency according to the user configuration.
24
DRAFT 02 - 08/02/10
Order List
Name and Description
Order No.
Name and Description
Dräger X-zone 5000,
868 MHz, 12 Ah
83 20 740
Chargers:
Dräger X-zone 5000,
868 MHz, 24 Ah
83 20 741
Dräger X-zone 5000,
868 MHz, 12 Ah, pump
83 20 742
Dräger X-zone 5000,
868 MHz, 24 Ah, pump
83 20 743
Dräger X-zone 5000,
915 MHz, 12 Ah
83 20 744
Dräger X-zone 5000,
915 MHz, 24 Ah
83 20 745
Dräger X-zone 5000,
915 MHz, 12 Ah, pump
83 20 746
Dräger X-zone 5000,
915 MHz, 24 Ah, pump
83 20 747
Order No.
Inductive charging station
83 20 626
Plug charger
83 20 749
Accessories:
Battery pack, small
(Dräger X-zone 5000)
83 20 644
Battery pack, large
(Dräger X-zone 5000)
83 20 646
Alarm-attenuation ring
(Dräger X-zone 5000)
83 20 110
Safety housing
(Dräger X-zone 5000)
83 21 519
Base (Dräger X-zone 5000)
83 20 645
Holder Dräger X-am 5x00 Diffusion (Dräger X-zone 5000)
83 20 636
Holder Dräger X-am 5x00 Pump (Dräger X-Zone 5000)
83 20 704
Adapter (Dräger X-zone 5000)
83 20 108
Dräger X-zone 5000,
433 MHz, 12 Ah
83 20 104
Dräger X-zone 5000,
433 MHz, 24 Ah
83 20 105
Cap for charging and
communication socket
18 93 632
Dräger X-zone 5000,
433 MHz, 12 Ah, pump
83 20 106
83 17 409
Dräger X-zone 5000,
433 MHz, 24 Ah, pump
83 20 107
USB DIRA with USB cable
(USB infrared adapter for Dräger Xzone 5000 – PC communication)
Pump accessories
Dräger X-zone 5000,
429 MHz, 12 Ah
83 20 710
Hose set (water trap,
10 cm Viton hose, filter)
83 21 527
Dräger X-zone 5000,
429 MHz, 24 Ah
83 20 711
Filter, pump
83 19 359
Water trap
68 05 473
Dräger X-zone 5000,
429 MHz, 12 Ah, pump
83 20 712
Float probe with accessories
83 18 371
Viton hose
12 03 150
Dräger X-zone 5000,
429 MHz, 24 Ah, pump
83 20 713
Rubber hose
(not suitable for H2S)
11 80 681
Tygon hose
83 20 395
25
DRAFT 02 - 08/02/10
Dräger Safety AG & Co. KGaA
Revalstrasse 1
D-23560 Lübeck
Germany
Phone +49 451 882-0
Fax
+49 451 882-20 80
www.draeger.com
90 33 112 - GA 4638.500
© Dräger Safety AG & Co. KGaA
Edition 02 - January 2010 (Edition 01 - October 2009)
Subject to alteration
DRAFT 02 - 08/02/10

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.4
Linearized                      : No
Page Mode                       : UseOutlines
XMP Toolkit                     : Adobe XMP Core 4.0-c316 44.253921, Sun Oct 01 2006 17:14:39
Creator Tool                    : FrameMaker 8.0
Create Date                     : 2010:02:08 09:02:55Z
Modify Date                     : 2010:02:08 15:17:07+01:00
Metadata Date                   : 2010:02:08 15:17:07+01:00
Format                          : application/pdf
Title                           : 9033112_02_2009_de.book
Creator                         : engelkec
Producer                        : Acrobat Distiller 8.1.0 (Windows)
Document ID                     : uuid:9d076909-e80b-4c08-842a-d13c26ed43a9
Instance ID                     : uuid:b89db3f9-b85c-47a0-b205-880e10eb9a81
Has XFA                         : No
Page Count                      : 26
Author                          : engelkec
EXIF Metadata provided by EXIF.tools
FCC ID Filing: X6O-AAC00XX

Navigation menu