EDAN INSTRUMENTS AX8EDAN Diagnostic Ultrasound System User Manual

EDAN INSTRUMENTS, INC. Diagnostic Ultrasound System

Acclarix AX8 User Manual

 1
 I Preface This manual contains necessary and sufficient information to use the Acclarix AX8 system safely for the purposes for which it is intended. It applies to all 1.3X releases. Read and understand all instructions in this manual prior to using the Acclarix AX8 ultrasound system.   Disregarding instructions, particularly warnings and cautions, is considered abnormal use. Caution:    Federal law restricts this device to use by, or on the orders of, a physician.  Contact Information: For  sales  or  service  information  please  contact  your  local  distributor  or  the  EDAN  service department at: support@edan.com.cn  P/N: 01.54.002481 MPN: 01.54.002481014 Release Date: Dec., 2016 © Copyright EDAN INSTRUMENTS, INC. 2015-2016
 II  Contents 1  Introduction ................................................................................................................................ 1 1.1 Intended Use/ Indications for Use .......................................................................................... 1 1.2 Contra-indications .................................................................................................................. 1 1.3 Device Description ................................................................................................................. 1 1.4 Labeling Symbols .................................................................................................................. 2 2  Safety ......................................................................................................................................... 5 2.1 Warnings ................................................................................................................................. 5 2.2 Cautions ................................................................................................................................. 7 3  Getting Started ......................................................................................................................... 10 3.1 System Configuration .......................................................................................................... 10 3.2 System Overview ................................................................................................................. 11 3.3 Battery Use ........................................................................................................................... 13 3.4 Connecting and Disconnecting a Transducer ....................................................................... 15 3.5 Powering on/ off ................................................................................................................... 17 3.6 Screen Layout....................................................................................................................... 18 3.7 Control Panel ........................................................................................................................ 21 3.8 Touch Screen ........................................................................................................................ 24 3.9 Trackpad ............................................................................................................................... 25 4  Exam Operation ....................................................................................................................... 27 4.1 How to Start an Exam .......................................................................................................... 27 4.2 How to End an Exam ........................................................................................................... 27 4.3 How to Restart an Exam ...................................................................................................... 27 4.4 The Patient Information Page ............................................................................................... 28 4.5 Modality Worklist ................................................................................................................ 30 5  Imaging .................................................................................................................................... 32 5.1 B-mode ................................................................................................................................. 32 5.1.1.  B-mode Touch Screen Controls ................................................................................. 32 5.1.2.  Dual Imaging ............................................................................................................. 35 5.1.3.  Quad Imaging ............................................................................................................ 35 5.1.4.  Panorama ................................................................................................................... 35 5.1.5.  Zoom .......................................................................................................................... 36 5.2 PW-mode .............................................................................................................................. 37 5.2.1.  PW Touch Screen Controls ........................................................................................ 37 5.2.2.  HPRF ......................................................................................................................... 39 5.2.3.  Update ........................................................................................................................ 39 5.3 CW-mode ............................................................................................................................. 39 5.3.1.  CW Touch Screen Controls........................................................................................ 39 5.4 Color-mode .......................................................................................................................... 41 5.4.1.  Color Touch Screen Controls ..................................................................................... 41 5.4.2.  Color Mode Variants .................................................................................................. 43 5.5 M-mode ................................................................................................................................ 43 5.5.1.  M-mode Touch Screen Controls ................................................................................ 43 5.6 3D/4D Mode* ...................................................................................................................... 44
 III 5.6.1.  Pre-3D ........................................................................................................................ 45 5.6.2.  3D Volume Sweeping ................................................................................................ 46 5.6.3.  3D Image Review ...................................................................................................... 46 5.6.4.  4D Volume Acquisition .............................................................................................. 51 5.6.5.  4D Live Volume ......................................................................................................... 52 5.6.6.  4D Cine ...................................................................................................................... 52 6  Transducers .............................................................................................................................. 53 6.1 Transducer Model ................................................................................................................ 53 6.2 Name and Function of Each Part of the Transducer ............................................................ 54 6.3 Transducer Cleaning and Disinfecting ................................................................................. 54 7  Features .................................................................................................................................... 58 7.1 Comments ............................................................................................................................ 58 7.2 Body Mark ........................................................................................................................... 59 7.3 Measurements ...................................................................................................................... 60 7.3.1.  Generic Measurements .............................................................................................. 63 7.3.2.  Application Measurements and Calculations ............................................................ 69 7.4 Patient Information Management ......................................................................................... 73 7.4.1.  Storing Images ........................................................................................................... 73 7.4.2.  Reviewing Images ...................................................................................................... 73 7.4.3.  Patient Database ......................................................................................................... 75 7.4.4.  Archiving Studies....................................................................................................... 77 7.5 Needle Biopsy Guide ........................................................................................................... 78 7.5.1.  Installing Needle Guide Bracket ................................................................................ 78 7.5.2.  Activating Needle Guide Function ............................................................................ 82 7.5.3.  To Adjust the Needle Guide Line .............................................................................. 83 7.6 Needle Visualization ............................................................................................................ 83 8  Presets ...................................................................................................................................... 85 8.1 Preset Organization .............................................................................................................. 85 8.2 Selecting a Preset ................................................................................................................. 86 8.3 Storing a Preset .................................................................................................................... 86 8.3.1.  Presets ........................................................................................................................ 87 8.3.2.  Comments Presets ...................................................................................................... 90 8.3.3.  Bodymarker Presets ................................................................................................... 91 8.3.4.  Measure Presets ......................................................................................................... 93 9  Utilities .................................................................................................................................... 97 9.1 Set-up ................................................................................................................................... 97 9.1.1.  General Set-up ........................................................................................................... 97 9.1.2.  Patient Set-up ............................................................................................................. 99 9.1.3.  Store/Print Set-up ..................................................................................................... 100 9.1.4.  Miscellaneous Set-up ............................................................................................... 101 9.1.5.  User Set-up .............................................................................................................. 102 9.2 Connectivity ....................................................................................................................... 103 9.2.1.  TCP/IP ..................................................................................................................... 103 9.2.2.  DICOM .................................................................................................................... 104 9.2.3.  Network Store .......................................................................................................... 107 9.3 Maintenance ....................................................................................................................... 108 9.3.1.  License ..................................................................................................................... 108 9.3.2.  Version ..................................................................................................................... 108
 IV 9.3.3.  Demo ........................................................................................................................ 109 9.3.4.  Export/Import ........................................................................................................... 109 9.4 Screen Adjust ..................................................................................................................... 111 10  In Between Exams ................................................................................................................. 113 10.1 Unpacking ........................................................................................................................ 113 10.2 Transport .......................................................................................................................... 113 10.3 Storage .............................................................................................................................. 113 11  Order List ............................................................................................................................... 114 12  Troubleshooting and Maintenance ......................................................................................... 116 12.1 Daily Checklist ................................................................................................................. 116 12.2 Troubleshooting ............................................................................................................... 116 12.3 Maintenance ..................................................................................................................... 117 12.4 Cleaning the System ......................................................................................................... 117 12.4.1. Cleaning the System Surface ................................................................................... 118 12.4.2. Needle Guide Bracket Cleaning and Sterilization ................................................... 118 13  Specifications ......................................................................................................................... 119 13.1 Electrical Safety Classifications ....................................................................................... 119 13.2 Power Supply ................................................................................................................... 119 13.3 Machine Specifications .................................................................................................... 120 13.4 Display Specifications ...................................................................................................... 120 13.5 General Technical Specifications ..................................................................................... 120 13.6 Operating, Storage and Transportation Environment ....................................................... 121 13.6.1. Operating Environment ............................................................................................ 121 13.6.2. Storage and Transportation Environment ................................................................. 121 14  EMC Information ................................................................................................................... 122
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                          Introduction - 1 - 1 Introduction 1.1 Intended Use/ Indications for Use For the United States of America The Edan Acclarix AX8 Ultrasound system is intended for use by a qualified physician or allied health professional for ultrasound evaluations. Specific clinical applications include:  Abdominal  Gynecology (including endovaginal)  Obstetric  Cardiac  Small parts (Breast, Testes, Thyroid, etc.)  Urology  Musculoskeletal    Peripheral vascular  Intra-operative  For the EU and other non-USA regions The Edan Acclarix AX8 Ultrasound system is intended for use by a qualified physician or allied health professional for ultrasound evaluations. Specific clinical applications include:  Abdominal  Gynecology (including endovaginal)  Obstetric  Cardiac  Small parts (Breast, Testes, Thyroid, etc.)  Urology  Musculoskeletal    Peripheral vascular  Intra-operative  Pediatric  Neonatal (including abdominal and cephalic)  Adult Cephalic* 1.2 Contra-indications The Edan Acclarix AX8 Ultrasound system is not intended for ophthalmic use. 1.3 Device Description The Edan Acclarix AX8 Ultrasound system consists of a main system along with associated transducers.   The  system  circuitry  generates  an  electronic  voltage  pulse,  which  is  transmitted  to  the transducer.  In  the  transducer,  a  piezo  electric  array  converts  the  electronic  pulse  into  an ultrasonic pressure wave. When coupled to the body, the pressure wave transmits through body tissues. The waves are then reflected within the body and detected by the transducer, which
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                          Introduction - 2 - then converts back to an electrical signal. The Acclarix AX8 system then analyzes the returned signal to generate an image or conduct Doppler processing.   The Acclarix AX8 system gives the operator the ability to measure anatomical structures, and offers analysis packages that provide information used by competent health care professionals to make a diagnosis. The system provides both touch screen and hard buttons for the User Interface. *Descriptions of transducer C5-2MQ, adult cephalic application, 3D/4D function, PW one-key optimization,  acoustic  power  control,  panorama  measurement  and  wireless  function  are  not approved for use until finishing the related registration.  1.4 Labeling Symbols The following labels are used on the Acclarix AX8 system: No. Symbol Definition 1  Serial Number 2 P/N Part Number 3  Date of Manufacture 4  Manufacturer 5  Operating instructions 6  Warning (Background: Yellow; Symbol & outline: Black ) 7  Refer to User Manual (Background: Blue; Symbol: White) 8  Caution 9  Biological Risks 10  The symbol indicates that the device complies with the European Council Directive 93/42/EEC concerning medical devices.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                          Introduction - 3 - 11  Authorized Representative in the European Community 12  Disposal method. Indicates that the equipment should be sent to special agencies according to  local  regulations  for  separate  collection  after  its useful life. 13  General Symbol for Recovery / Recyclable 14  Caution: Federal (U.S.) law restricts this device to sale by or on the order of a physician. 15  Conforms  to  AAMI  Std.  ES  60601-1,  IEC  Std.  60601-2-37 Certified  to  CSA  Std.  No.  60601-1,  No  60601-2-37 16 IPX7 No harm for short time immersion 17  Type BF Applied Part 18  Transducer connector   19  Pencil Transducer connector (reserved) 20  ECG signal input connector (reserved) 21  Transducer lock 22    Transducer unlock 23  Network port 24  Trademark 25  Direct current 26  Video Output port
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                          Introduction - 4 - 27  USB 2.0 port   28  USB 3.0 port   29  Display port 30  Microphone input   31  Headphones   32 FCC ID: SMQAX8EDAN Federal Communication Commission: FCC ID: SMQAX8EDAN NOTE:   The user manual is printed in black and white.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                  Safety - 5 - 2 Safety Throughout this document the following terms are used:  Warning: Advises against certain actions or situations that could result in personal injury or death.  Caution:  Advises  against  actions  or  situations  that  could  damage  equipment,  produce inaccurate data, or invalidate a procedure.  Note: Provides useful information regarding a function or a procedure. Please read  all warnings and cautions prior to using the  Acclarix AX8 ultrasound system. For your convenience all warnings and cautions are provided in this section. These may be duplicated elsewhere in this document in the context of the instructions for use. 2.1 Warnings   Only use an Edan supplied power adapter and power cord.   Only use an Edan supplied battery. Read and understand the battery installation instructions prior to changing the battery.   Only use Edan supplied transducer. Use of other transducers may result in electric shock or system malfunction.   Only use a hospital grade, grounded, power outlet and plug. Do not use with an ungrounded outlet.   The system is ordinary equipment (Sealed equipment without liquid proof). But the console  panel  including  touch  screen,  track  pad  and  hard  keys  is  IPX1. Transducer (not including the transducer connector) is IPX7. Footswitch is IP68. Do  not  immerse  or  expose  any  of  the  parts  to  extended  moisture.  Splash resistance does not extend to transducer connectors. Keep connectors dry.   Do not use in a wet environment or when the relative humidity exceeds 95%.   Do not reverse the positive and negative poles when installing a battery.   Do not use the battery near heat sources or when the ambient temperature is over 40oC. Do not heat or dispose of in fire.   Do not destroy the battery; do not pierce or cause a strong impact to the battery.   Do not touch the connector pins on the transducer port.   Parts  and  accessories  used  must  meet  the  requirements  of  the  applicable IEC/EN60601 series safety standards, and/or the system configuration must meet the requirements of the IEC/EN60601-1.   Use  protective  barriers  (gloves  and  transducer  sheaths)  whenever  possible. Follow sterile procedures when appropriate. Thoroughly clean  Transducers and reusable accessories after each patient examination and disinfect or sterilize as needed. Refer to transducer use and care instructions. Follow all infection control policies  established  by  your  office,  department  or  institution  as  they  apply  to personnel and equipment.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                  Safety - 6 -   Not intended for Ophthalmic use.   If  a  sterile  transducer  cover  becomes  compromised  during  an  intra-operative application involving a patient with transmissible spongiform encephalopathy, such as Creutzfeldt-Jakob disease, follow the guidelines of the U.S. Disease Control Center  and  this  document  from  the  World  Health  Organization: WHO/CDS/APH/2000/3,  WHO  Infection  Control  Guidelines  for  Transmissible Spongiform  Encephalopathies.  The  transducers  for  your  system  cannot  be decontaminated using a heat process.   Contact with natural rubber latex may lead to a severe anaphylactic reaction in persons sensitive to the natural latex protein, Sensitive users and patients must avoid  contact  with  these  items.  EDAN  strongly  recommends  that  health-care professionals identify their latex-sensitive patients, and refer to the March 29, 1991 Medical  Alert  on  Latex  products.  Be  prepared  to  treat  allergic  reactions immediately.   Improper  operation  may  cause  the  internal  lithium  battery  (hereinafter  called battery) to  be  hot,  ignited or  exploded,  and it  may lead  to  the decrease of the battery capacity. It is necessary to read the user manual carefully and pay more attention to warning messages.   Do  not  touch  accessible  contacts  of  electrical  equipment  and  the  patient simultaneously.     This device is not suitable for intra-cardiac use or direct cardiac contact.   The system shall not be serviced or maintained while in use with a patient.   Install the system according the EMC guidance provided in Chapter 14   Do not stack the system on other electronic equipment.     The use of transducer and connecting cable not supplied by the manufacturer may result in increased emissions or decreased immunity of the equipment.   Refer to Chapter 14 for recommended separation distances from other equipment, including portable and RF communication devices.   The power adapter is used to isolate the system from main power. Position the system so that it is easy to disconnect the device.   No modification of this equipment is allowed.   The  system  should  be  maintained  regularly,  at  least  annually,  by  a  qualified technician  who has  adequate training, knowledge and  experience. That person should be familiar with the AX8 Service Manual (P/N: 01.54.002483),  available from your Edan representative.   Keep non-medical equipment (such as the external printer) out of the vicinity of the patient.  (1.5m/6ft.)   Use  of  an  extension  cord  or  multi-socket  outlet  setup  to  provide  power  to  the ultrasound  system  or  to  the  system‟s  peripheral  devices,  may  compromise  the
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                  Safety - 7 - system grounding and cause the system to exceed leakage current limits.   SHOCK HAZARD – Do not connect non-isolated electrical equipment to the same circuit being used to power the system.   Edan  recommends  the  use  of  isolated  connectors  on  any  electrical  equipment attached  to  the  system,  and/or  using  isolation  transformers  that  complies  with IEC60601-1 to power that electrical equipment.     Always  use  sterile  technique  during  a  biopsy  procedure.   Sterilize  the  needle guide assembly between uses.    Use a sterile needle with each use.   Always confirm the clinical accuracy of the IMT result before entering it into the report. Transducer Warnings   To avoid infection, always use protective gloves when cleaning or disinfecting   Read and follow all manufacturer instructions for disinfection agents.   To avoid infection, ensure that expiration date of the disinfecting solution has not passed.   Disinfect the transducer after each intra-cavity or intra-operative procedure. Use a new sterile sheath for each such procedure.   Use a pyrogen-free transducer sheath for neurological intra-operative procedures.     Unplug the transducer from the system prior to cleaning or disinfecting.   Do not immerse the transducer beyond the point indicated in Figure 6-2.   Do not allow the transducer connector to get wet. 2.2 Cautions   The system contains no user serviceable components other than the battery. Do not remove any covers.   Excessive dust and dirt could clog internal airflow and cause overheating. Do not use in a dusty environment.   Do not use a battery that leaks, emits an odor, appears deformed, or discolored. Immediately replace it with a new Edan-supplied battery and dispose of the old battery according to local regulations. Replace a battery that has reached the end of its service life.   Use care when storing or disposing of batteries. Do not allow the leakage to come in  contact  with  each  other.  Do  not  dispose  of  them  together  with  household garbage. At the end of their life hand the batteries over to the applicable collection points for the recycling of waste batteries.   Inspect the system regularly, at least weekly. Before use ensure there is no visible
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                  Safety - 8 - evidence of damage to the equipment, cables, and transducers. If a component is damaged replace it before use.   Do not use in locations subject to vibration.   Read and understand the section A1.2 Ultrasound Safety and the ALARA Principle of Acclarix AX8 Advanced User Manual before using the system. Do not expose a patient to ultrasound energy longer than clinically reasonable.   Practice  ALARA  principle  when  operating  ultrasound  system.  Minimize  the acoustic power without compromising the image quality.   Do not use in the presence of a flammable anesthetic.  The system generates radio frequency energy, which may cause interference with other  devices  in  the  vicinity.  If  interference  is  suspected,  try  re-orienting  or relocating the equipment.   The  use of  electrosurgical units or  other devices that  generate radio  frequency interference may cause image distortion or other malfunction.   During  long  term  storage the  battery  should  be  charged  at  least  once  every  3 months to ensure battery capacity.   The  system  should  only  be  used  by  a  qualified  physician  or  allied  health professional for ultrasound evaluations.   Use only Edan supplied or recommended parts and accessories.   Verify measurement results prior to entering them into a report.   Contact your local distributor or Edan service if there is excessive noise from the system speaker or fans.   Please read and understand cleaning instructions prior to use.   Please read and understand maintenance instructions prior to use.   Please read and understand instructions for system operation prior to use.   Studies stored on the system hard drive should be archived regularly. The system is not intended for  long term storage  of patient information.  Confirm  successful archiving before deleting a study from the hard drive.   Ensure that the system vents are clear and unobstructed.   Confirm  patient  identification  information  prior  to  storing  or  printing  any  exam information.   If you have any questions about maintenance, technical specifications, or system functionality,  please  contact  your  local  distributor  or  Edan  service  at: support@edan.com.cn   Ultrasound images occasionally have artifacts, and should only be used as one part of an overall clinical assessment.   To avoid electrical shock, turn off and disconnect the device from the AC power
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                  Safety - 9 - source before cleaning and disinfecting.   The device and accessories are to be disposed of according to local regulations after  their  useful  lives.  Alternatively,  they  can  be  returned  to  the  dealer  or  the manufacturer for recycling or proper disposal. Transducer Cautions   Do not use disinfection agents beyond their expiration date.   Do not use sterile sheaths beyond their expiration date.   Inspect the transducer connector, cable, and head periodically. Do not use if there is evidence of excessive wear or damage.   Do not operate the transducer to temperatures in excess of 40°C  or storage the transducer to temperatures in excess of 55°C .   Do not kink or pull on the transducer cable.   Broken or bent connector pins can cause image artifacts. Do not use a transducer with broken or bent pins.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                Getting Started - 10 - 3 Getting Started 3.1 System Configuration The Acclarix AX8 system is shipped with the following components:  Acclarix AX8 main unit  1 rechargeable lithium battery  1 UPS battery  1 AC adapter  1 bottle of coupling gel  1 basic user manual and 1 advanced user manual The following options are also available (please see the Order List Appendix for details)  Transducers: C5-2XQ, C5-2Q, L10-4Q, L12-5Q, L17-7HQ, E8-4Q, MC8-4Q, P5-1XQ, L17-7SQ, C5-2MQ*  1 Trolley MT-807  Footswitch  Suitcase  The Acclarix AX8 supports the following external DVD drives: SAMSUNG SE-218GN, SAMSUNG  SE-208GN,  LENOVO  DB75,  ASUS  SDRW-08D2S-U  and  PIONEER DVR-XU01C. The Acclarix AX8 supports the following DVD disks: Philips CD-R  700MB 80min  52x, Sony CD-RW 700MB 1x-4x,Sony DVD-RW 4.7GB 120min 2x/1x, Verbatim DVD+RW 4.7GB 120min 4x. Note: Only support DVD disk single burning.  The Acclarix AX8 supports the following printers. Please contact the manufacturer for details. Printer Model Printer Type Interface SONY UP-25MD Color Image S-Video SONY UP-D25MD Color Image USB SONY UP-X898MD B/W Image USB HP Officejet Pro 251dw Color Report USB HP Laserjet Pro 200 M251n Color Report USB HP Laserjet CP1525n Color Color Report USB HP Deskjet Ink Advantage 2010 Color Report USB HP Deskjet 1010 Color Color Report USB HP Deskjet 1510 Color Color Report USB HP LaserJet 400 M401d B/W Report USB Table 3-1 Printer List
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                Getting Started - 11 - Note:  For  more  printer  models,  please  see: http://hplipopensource.com/hplip-web/supported_devices/combined.html  .  Generally, printer models with driver version 3.14.3 or lower can be supported by Acclarix AX8, but some models cannot be perfectly supported.  Needle Guide Bracket Kit Model Name   Description   BGK-C5-2 BGK-C5-2 Needle Guide Bracket Kit For use with the C5-2XQ/C5-2Q,   Supports: 16G, 18G, 20G, 22G BGK-L40UB L40UB Needle Guide Bracket Kit For use with the L10-4Q,   Supports: 16G, 18G, 20G, 22G BGK-CR10UA CR10UA Needle Guide Bracket Kit For use with the E8-4Q,   Supports: 16G BGK-R15UB R15UB Needle Guide Bracket Kit For use with the MC8-4Q,   Supports: 16G, 18G, 20G, 22G BGK-P5-1X BGK-P5-1X Needle Guide Bracket Kit For use with the P5-1XQ Supports: 14G-22G Table 3-2 Needle Guide Bracket Kits   3.2 System Overview  1    2  3   4  3   5    6 7  8 9  3
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                Getting Started - 12 -  Figure 3-1 Appearance of the System  No. Name   Description   1 Monitor Display  the  images  and  parameters  during scanning. 2 Touch Screen Use to control operation and activate functions 3 Control panel Use to control the operation   4 Handle Used for carrying the system 5 Hinge Used to rotate the monitor 6 Transducer locking lever Use to lock or unlock the transducer 7 Transducer Port Used  for  connecting  a  transducer  to  the  Acclarix AX8 main unit 8 TrackPad Use to change the cursor position 9 Battery Indicator LEDs  indicate  battery  charge.  Each  LED corresponds  to  20%  of  battery  power.  The  LEDs will flicker while the battery is charging. Touch the battery indicator to view the battery power.   10 I/O Ports Use to connect the I/O extend modules Table 3-3 Appearance of the System            10
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                Getting Started - 13 - I/O Ports on the left panel:  Figure 3-2 I/O ports of the system 1. Power supply 5. USB 2.0 port   2. S-Video output port   6. Network port 3. Display port   7. Microphone input port 4. USB 3.0 port   8. Audio output port Table 3-4 I/O Ports of the System NOTE:If an external monitor is connected to the system via display port, configure the resolution  to  1920x1080,  other  resolution  may  cause  abnormal  ultrasound  machine display. 3.3 Battery Use The Acclarix AX8 system comes with one lithium-ion battery. When it is fully charged it can run the system for approximately 60 minutes, depending on use. It is automatically charged when the system is plugged in. CAUTION 1.  If the system will remain unused for more than one week, charge the battery to at least 75%capacity, take the battery out and store the system and battery separately. 2.  During long term storage the battery should be charged at least once every 3 months to ensure battery capacity is more than 75%. 1 3 4 5 6 7 2 8
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                Getting Started - 14 - To install the battery: 1. Turn off and unplug the Acclarix AX8 system. 2. Close the monitor, turn the system upside down and rest it on a flat stable surface. 3. Put  the  battery  gently  to  battery  compartment,  and  the  button  will  lock  the  battery automatically.    Figure 3-3 Installation of the Battery To remove a battery: 1. Turn off and unplug the Acclarix AX8 system. 2. Close the monitor, turn the system upside down and rest it on a flat stable surface. 3. Slide the battery button to unlock the battery. 4. Take the battery out.  Figure 3-4 Battery Removal   To install the UPS battery: 1. Close the monitor, turn the system upside down and rest it on a flat surface.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                Getting Started - 15 - 2. Take out the UPS battery from the package. 3. Put the UPS battery gently to the UPS battery compartment.  Figure 3-5 Installation of the UPS Battery 3.4 Connecting and Disconnecting a Transducer To connect a transducer:    Figure 3-6 Tansducer Locking Handle 1. Align the connector with the transducer port and carefully push into place. 2. Toggle the locking handle to the bottom position. 3. Do not allow the transducer head to hang free. Impact to the transducer head could result in irreparable damage.  Figure 3-7 Lock the tansducer locking handle
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                Getting Started - 16 - To disconnect a transducer:   1. Toggle the locking handle to the top position to unlock the connection of the Transducer. 2. Firmly grasp the transducer connector and carefully remove it from the system port. 3. Store transducer in its protective carrying case prior to transport.  Figure 3-8 Unlock the Tansducer Locking Handle CAUTION 1. Do not touch the pins of the transducer connector. 2. Broken or bent connector pins can cause image artifacts. Do not use a transducer with broken or bent pins. Multi-Transducer Connector (MTC)  The Multi-Transducer Connector  (MTC) is  an  optional  addition  to  the  AX8 that  allows  up  to three transducers to be connected to the system.    The MTC can be mounted to the optional AX8 cart, or lie flat on an adjacent table.  Figure 3-9 Connect to the MTC Note: Pluging and unpluging the MTC should be done while the system is frozen or turned off.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                Getting Started - 17 - 3.5 Powering on/ off Please  review  and  follow  the  steps  described  in  the  Section  12.1.Daily  Checklist  prior  to powering on the device.  To power on: 1. Connect the system to a hospital grade power supply, or use the battery as the power supply. 2. Press the power on/off key on the top left of control panel.    To Login If  Password  Protection  is  enabled  (see  section  9.1.5),  the  system  displays  a  Login  dialog when booting up the system, you can type in or select user name in the drop-down list of User name, then enter password and click Login. For emergency use, you can click on Emergency to log on directly without entering user name and password.  Note: When  both  Admin1  and  Admin2  password  are  forgotten,  please  contact  the serviceman for the system password reset.  To power off: 1. Press  the  power  on/off  key  on  the  top  left  of  control  panel  and  the  system  displays  a confirmation dialog box. 2. Select “Shut Down” from the confirmation dialog box. If the system is unresponsive, a long press of the power on/off key will shut down the system directly. NOTE:   Please unplug the AC adapter from the power socket and disconnect the battery prior to storage.    Sleep mode The system will enter a sleep mode that maintains exam information while using minimal battery. There are three events that can invoke sleep mode:
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                Getting Started - 18 -  Close the cover of the system without powering-off the system.  No user input for a configurable amount of time. Please see System Set-up to configure this time.    Pressing the Sleep button on the confirmation dialog box when powering off.   Figure 3-10 Confirmation Dialog Box When Power off 3.6 Screen Layout Figure 3-11 Main Screen Display ①Information Field: The  top  line  of  this  field  contains your hospital/institution  name.  Please  see  Section 9.1. General Set-up for information on customizing this area. The  second  line  of  this  field  contains the  patient name,  gender,  age  and  ID,  as  entered through the Patient Information screen. This field also contains data fields for:   The currently active transducer ① ③ ④ ⑤
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                Getting Started - 19 -   The currently active preset   MI, TI, current date and time. ②Image Field: The ultrasound image appears in the Image Field, under the Information field. The image field also contains information typically associated with the image, such as scales, TGC, and maps. The system supports two different image sizes within the image field. See 9.1.1 for details ③Mini Report: The left side of the screen displays a “mini-report” which provides a synopsis of applications measurements performed in the current exam. ④Thumbnail Field: The right side of the screen displays thumbnail images of all statics and clips captured in the current exam. ⑤Status Bar: The bottom of the screen is used to display icons that provide system status. These include:   Image Store icon : Number of static images and clips in the current exam.   Battery icon : Battery fully charged, symbol in green. : Battery higher than 80%. : Battery 60%-80%. : Battery 40%-60%. : Battery low, symbol in red. : Battery removed, outline in red. : Battery charging.   Network status icon The network icons show the connection status of the DICOM or  FTP Server. If no Network    Server is defined then the icon does not show at all.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                Getting Started - 20 - :   Outline in grey color: Successful connection with a Server. Outline in green color: Data exchange with a Server. Outline in red color: Failure to connect with a configured Server.   Hard drive icon. : Hard drive available : Hard drive data exchange, symbol in green. : Hard drive 95% full, symbol in red. : Hard drive 95% full with data exchange, symbol in red.     USB icon :   Symbol in grey color: USB available   DVD icon : Symbol in grey: DVD device is connected :  Symbol  in  green:  DVD  device  is  connected,  disc  is  inserted  and  data transmission is available.   Printer icon : Printer available.   Wifi icon * :Wi-Fi available, Subject to regulatory approval. See Connectivity (Section 9.2.1) for enabling wireless if it is licensed.    Clicking on this icon shows a list of all Wi-Fi networks that are currently available. WARNING The system complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:   1)  this device may not cause harmful interference, and   2)  this device must accept any interference  received, including interference that may
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                Getting Started - 21 - cause undesired operation. NOTE: 1.  This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed  and  used  in  accordance  with  the  instructions,  may  cause  harmful interference  to  radio  communications.  However,  there  is  no  guarantee  that interference will not occur in a particular installation. If this equipment does cause harmful  interference  to  radio  or  television  reception,  which  can  be  determined  by turning  the  equipment  off  and  on,  the  user  is  encouraged  to  try  to  correct  the interference by one or more of the following measures: -  Reorient or relocate the receiving antenna. -  Increase the separation between the equipment and receiver. -  Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. -  Consult the dealer or an experienced radio/TV technician for help. 2. Any  changes  or  modifications  to  this  unit  not  expressly  approved  by  the  party responsible for compliance could void the user's authority to operate the equipment.  3.7 Control Panel Figure 3-12 Control Panel of the System No. Key Name Description 1.   Patient Invokes the Patient Information Screen. Typically used to start/end exams or to modify patient information during an exam. See section 4.1
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                Getting Started - 22 - 2.    Preset Allows you to change the Preset being used. See section 8 for details. 3.   End Exam Completes  an  exam  without  invoking  the  Patient Information Screen for the next exam. See section 4.2 4.   Utilities Press  this  button  to  get  the  Set  Up,  Maintenance, Connectivity and Screen Adjust keys on the touch screen. 5.   Alt Store One of three keys that can be configured to store images or  clips,  or  to  set  video  printers.  See  section  7.4  for details. 6.   Report Press to display the report page 7.   Review Press to enter exam database or image review mode. See section 7.4 for details. 8.   Zoom Press to zoom the image in or out. See section 5.1.5 for details. 9.   Depth Press to adjust the depth of the image displayed. 10.    Body Mark Enters or exits the Body Mark function. See section 7.2 for details. 11.    Clear Press  to  clear  all  the  measurements,  calculations, comments,  and  body  marks  displayed  in  the  current image. 12.    Comment Enters or exits the Comment function. See section 7.1 for details.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                Getting Started - 23 - 13.    Update In  Strip  Doppler  this  switches  between  acquiring  the Doppler strip and the reference image. See section 5.2.3 for details.   14.    Cursor Press to hide or display the cursor 15.    Measure Invokes the Measure function for Generic and Application Measurements. See section 7.3 for details. 16.     Freeze/Cine Press this key to switch between the frozen and real-time states. 17.    Store 1 One of three keys that can be configured to store or print images or clips, or to set video printers. See section 7.4 for details. 18.    Store 2 One of three keys that can be configured to store or print images or clips, or to set video printers. See section 7.4 for details. 19.    Gain Controls  the  overall  gain  for  the  imaging  mode  that currently has the UI focus. In normal imaging the gain can also be adjusted by sliding your finger along the top of the trackpad area 20.    F1 User-defined button 21.    F2 User-defined button 22.    Color Press  to  enter  or  exit  Color  Mode.  See  section  5.4  for details 23.     M Press to enter or exit M Mode. See section 5.5 for details. Use the trackpad to adjust the M sample line. 24.    D Press to get the sample line, use the trackpad to adjust the position and press the hardkey again to confirm.   For other information, see section 5.2 and 5.3 for details.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                Getting Started - 24 - 25.    B Press  to  return  to  B-mode  imaging  from  any  display  or imaging mode. See section 5.1 for details. Table 3-4 Buttons on Control Panel 3.8 Touch Screen The Touch Screen contains controls that vary depending on the active function. There are several types of controls used by the touch screen, as illustrated below:  Figure 3-13 Touch screen of the System A. Tabs: Each function that is active has a tab on the touch screen. Usually the function that was most recently activated is the top tab. Pressing on any other tab will bring it to the top and provide access to that function‟s controls. B. Slider: Pressing on the top or bottom of a slider changes the control setting by one value. Pressing anywhere on the control and dragging will continuously change the value. C. Push button: This can either be an on/off control (like “Colorize”) or a one-shot control that immediately performs an action (like “Auto”)   D. Radio Buttons: A collection of buttons where only one is active at any time. Activating one will de-activate all others. E. Folder: Controls can be grouped together into a folder. Pressing on the folder will open it to access any of the controls in it. F. TGC:  The B-mode  tab has  a specialized  control  for  TGC.  Each  slider can be  dragged horizontally  and  individually.  Dragging  vertically  down  across  the  sliders  will  set  all sliders.   G. Pages: When a tab has multiple pages of controls each page is represented by a dot at the top of the page. The current page is indicated by a filled-in dot. You can move between pages by dragging your finger horizontally across the dots. These dots do not appear when there is only one page in the current tab.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                Getting Started - 25 - Customizing the touch screen The touch screen can be customized to meet your needs. Press and hold any control for about a second to put the touch screen in customization mode. Continue pressing and drag the control to a new location.  Creating Folders: Dragging one control on top of another control will create a folder that contains  both  controls.  Dragging  controls  out  of  a  folder  until  only  one  exists  will automatically delete that folder. Folders cannot contain other folders.  Multiple pages: Dragging a control to the side of the screen will move that control to the next page.    Radio Button Cluster: There is no restriction on moving a single radio button. However, we suggest that they are grouped adjacent to each other. When they are grouped in this way  the  system  will  automatically  draw  a  border  around  them  to  indicate  they  are  a related group of radio buttons.  TGC: The TGC is a special control that cannot be moved. It is always on the right side of the B-mode tab. For most tabs there is one arrangement of controls. Whatever way they are customized will be remembered across exams and across power-ups. However, the following tabs can have multiple different arrangements:  Comments  Bodymarkers  Measurements These  functions  are  often  configured  differently  for  different  exams.  For  example,  the measurements for an OB exam are different from those of an abdomen exam. Each configuration of these tabs is stored in an application preset (see Chapter 8 Presets, for details) 3.9 Trackpad In general the Acclarix AX8 trackpad is used like a trackball or similar pointing device used on most ultrasound systems. Using a trackpad instead of a trackball helps make the console sealed, and makes the system easier to clean. In addition, there are several capabilities that the trackpad supports that would not be possible on a traditional trackball:  The trackpad supports multi-finger gestures. For example: o While in B-mode a pinch-out gesture can be used to zoom the image.   o While in Color a pinch gesture can re-size the ROI.  Dragging along the Swipe button on top edge of the trackpad as a primary sliding control. For example: o In most live imaging modes this is Gain.    The top right of the trackpad is reserved as a „Set‟ button. Pressing this corner is generally equivalent to a single click on a mouse or trackball. For example: o In CD it is used to toggle between controlling the size and the position of the pan box. o In Measure it is used to toggle between controlling one side or the other side of a caliper pair.  The  top  left  of  the  trackpad  is  reserved  as  an  „Enter‟  button.  Pressing  this  corner  is
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                Getting Started - 26 - generally equivalent to a double click on a mouse. For example: o In Measure it is used to enter a measurement into a report. The examples provided above are a subset of how the trackpad is used. Additional examples are provided in specific feature descriptions elsewhere in this manual.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                  Exam Operation - 27 - 4 Exam Operation 4.1 How to Start an Exam To start a new exam, press the Patient key and enter in patient information for the exam, or select a scheduled patient from the modality worklist. If there  is  no  previous  exam  this  will  bring  you directly to  the  Patient  Information Page  (see figure 4-2 below). If a previous exam is still active you will see the following dialog:  Figure 4-1 Exam Confirmation Dialog It has the following options:  Edit Current: This lets you edit the Patient Information for the current exam. It does not start a new exam.  New Exam: Select this to start a new exam with the same patient. It confirms that you want to end the previous exam, and then displays the Patient Information Page with the previous patient information still displayed.  Cancel: Exits the dialog without starting or ending an exam.  New  Patient:  Select  this  to  start  a  new  exam.  It  confirms  that  you  want  to  end  the previous exam, and then displays a blank Patient Information Page for the new exam. 4.2 How to End an Exam There are two ways to end an exam:  Pressing the Patient key, as described above, and then selecting the New Patient. This both ends the exam and presents the Patient Information Page for the next exam. This is the most common way of ending the exam, especially when there are several exams in a row.  Pressing the End Exam key. This brings up a dialog to confirm you want to end the exam, but does not invoke the Patient Information Page for the next exam. You might end the exam this way when there is no exam next. When  an  exam  is  ended  the  associated  files  on  the  system  are  closed.  If  a  DICOM  server  is connected successfully and Auto-transfer when End Exam is configured, any remaining images are transferred. 4.3 How to Restart an Exam 1. Select an exam within the time limit in Exam Database. For the setting of time limit, refer to section 9.1.2 Patient Set-up. 2. Press  Restart  on  the  touch  screen  to  continue/edit  the  exam  that  was  performed  on  the selected patient. You  can also  modify the patient information by pressing  Patient-->Edit Current.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                  Exam Operation - 28 - 4.4 The Patient Information Page The Patient Information Page is used to enter or modify patient demographic data. The following figure is an example:  Figure 4-2 Patient Information Page(OB Exam) The top three lines are for entering the patient last name, first name, ID, exam accession number, and DOB (Date-of-Birth) or age. If a date of birth is entered the age is automatically calculated. Note:   By  default  the  patient  name  is  two  fields:  family  name  and  given  name.  It  can  be configured to be one field in the Patient Setup screen (See section 9.1.2 for detail). The next line shows the current transducer, and the preset that will be used for the current exam. See  section  8.3.1.2  for  information  on  presets.  The  preset  can  be  changed  by  clicking  this drop-down and selecting one of the other presets associated with the current transducer. The lines below the preset-selection vary depending on the current preset (see section 8.3.1.2 for detail). They show clinical information that is preset-specific.  Gender: Select the patient‟s gender: “M” (Male), “F” (Female), “O” (Other), or “<blank>”. “F”  is  default  for  gynecology,  obstetric  and  breast  exams;  “M”  is  default  for  prostate
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                  Exam Operation - 29 - exam and testis exam; <blank> is default for others exam.  LMP:  Last  Menstrual  Period  (yyyy/mm/dd),  If  LMP is  entered  then  GA and  EDD  are calculated.  Entering  EDD  does  not  impact  LMP.  An  LMP  more  than  300  days  ago  is considered invalid.  GA: Gestational Age (xxWyD), it is auto-calculated when LMP or EDD is entered (only in OB Exam). A GA of more than 42W6D is considered invalid and not displayed.  EDD:  Estimated  Date  of  Delivery  (yyyy/mm/dd).  It  is  auto-calculated  when  LMP  is entered..  Fetus: Enter 1 up to 4, for multiple gestations.  G/P/A: G stands for Gravida, P stands for Para and A stands for Aborta. Enter the times of them in the fields separated by slashes.  Study Description: enter the study description.  Height: Enter the patient‟s height. The units can be set in the Patient section of Setup.  Weight: Enter the patient‟s weight. The units can be set in the Patient section of Setup.  BSA:  Body  Surface  Area,  it  is  auto-calculated  and  displayed  when  Height/Weight  is entered.  HR: Enter the Heart Rate.  BP: Enter the Blood Pressure.  PSA: Prostate Specific Antigen.  PPSA Coefficient: Predicted Prostate Specific Antigen.  Ref. Physician: Enter the name of the Ref.Physician.    Dx. Physician: Enter the name of the Dx.Physician.  Operator: Enter the name of the person performing the exam. .  CPT code: Current Procedural Terminology code.    <Custom field 1>: Enter the user-defined data.  <Custom field 2>: Enter the user-defined data.  Comments: Enter any additional comments. While the Patient Information Page is displayed the following touch screen controls are shown:
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                  Exam Operation - 30 -  Figure 4-3 Patient Information Touch Screen Press Start Exam to exit the Patient Page function and return to B-mode imaging with the newly entered demographic data entered for the active exam. Press Cancel to exit the Patient Information Page without storing any of the entered data. Press Clear All to clear all of the demographic fields except for name and ID. Input  LMP,  and  press  Prior  Exam  to  enable  the  entry  of  previous  OB  exam  data  for  fetal trending details, it is only available when an OB preset is selected.   4.5 Modality Worklist Modality  worklist  provides  a  list  of  scheduled  patients  derived  from  a  DICOM  server.  It  is available only when a DICOM server is configured and worklist is enabled.   When the modality worklist function is enabled and configured in DICOM Connectivity screen, the worklist is shown to the left of the Patient Information Page, as shown below.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                  Exam Operation - 31 -  Figure 4-4 Modality Worklist Display The worklist is displayed at the left side of the Patient Information Page in two columns showing patient  name  and  patient  ID.  Clicking  on  the  header  of  each  column  will  sort  the  list  by  the corresponding column.   The  worklist  shows  all  scheduled  ultrasound  exams  within  the  date-range  specified  in  the Connectivity Utility (See 9.2.2).    Typing any text in the Patient Name or ID fields will filter the list to exams that contain the entered text. Update: Press it to query the patient data and update the list manually.   Hide List: Press it to hide the list with only a Show List button displayed. Press the Show List button to display the list and other buttons. Select one  patient  from  the  list  and  the  detailed patient  information  is  filled  in  the  associated fields on the patient information page, allowing you edit or complete. Then press Start Exam on the touch screen to start an exam.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                Imaging - 32 - 5 Imaging The imaging modes are entered and exited via dedicated hard keys along the right side of the console.   Key Description  The B-mode key behaves differently than the other mode keys. B-mode imaging is the baseline state of the system. Therefore, the B key returns to B-mode imaging at any time from any display on the system. See section 5.1 for B-mode controls.  Invokes and exits strip Doppler mode. Press D button and press PW/CW on the touch  screen  to  switch  to  CW  mode  (requires  P5-1XQ  transducer).  See  section 5.2/5.3 for Doppler controls.  Invokes and exits color modes, including Color, Power Doppler Imaging (PDI) and Directional Power Doppler Imaging (DPDI). The default color mode is determined by the preset and can be changed while Color is active. See section 5.4 for Color controls.  Invokes and exits M-mode. See section 5.5 for M-mode controls Table 5-1 Mode Control Hard Buttons 5.1 B-mode 5.1.1. B-mode Touch Screen Controls The  following touch  screen  controls  impact  the  B-mode  image.  Touch  screen  controls  can  be located where you want them to be. See section 3.8 for details. Name Control Description TGC  The  Time  Gain  Control  (TGC)  adjusts  the  gain  of  the  image  at different depths. Each slider can be adjusted separately, or you can drag your finger vertically to set all of the sliders to a new setting.    Auto  The Auto push button controls automatically updates the Gain and TGC.  Each  single  press  of  the  button  renews  the  automatic optimization.   Dynamic Range  The  Dynamic  Range,  or  log  compression,  adjusts  how  echo intensities  are  converted  to  brightness.  A  high  dynamic  range presents a flat image, while a low dynamic range present a large range of intensities.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                Imaging - 33 - eSRI  eSRI is Speckle Reduction Imaging. There are 4 levels: Off, Low, Med.  and  High.  High  level  provides  more  aggressive  speckle reduction. Persistence  Persistence averages frames together in order to reduce random noise.  There  are  4  options:  Low,  Med.,  High  and  Off.  The persistence level corresponds to the number of frames averaged. The frame rate is unchanged; as each new frame is acquired it is averaged with the preceding frames. Frequency  This determines the fundamental or harmonic (with harmonic invokes) frequencies used for imaging. It is available on live images.  Harmonic  The Harmonic control invokes and exits harmonic imaging. While in harmonic imaging the control is highlighted and „H‟ in the B-mode frequency field. Depending on the transducer there may be multiple harmonic frequencies. Spatial Compounding  Spatial Compounding combines images fired from multiple angles in order to reduce speckle, reduce shadow artifacts, and enhance contrast. Spatial compounding is an on/off control.   Foc. Num  Focus  Number  adjusts  the  number  of  focus  positions.  As  the number of foci increases the frame rate will decrease, but there will be more image uniformity across depths. Foc. Pos.  Focus  Position  adjusts  the  depth  of  the  focus  or  foci.  Upward presses  move  the  focus  shallower,  regardless  of  the  U/D  invert status of the image. Gray Map  Gray Map adjusts the post processing map used on the B-mode image.  In  general,  higher  map  numbers  correspond  to  more contrast in the image. Colorize  The Colorize control adds a color tint to the B-mode image. Tint  The  Tint  control  changes the  color  tint being  used.  Gold,  Sepia, Blue,  Ice  and  Clear  are  available.  If  Colorize  had  been  off, changing the Color map control will automatically activate it. Left/Right  The Left/Right invert control is indicated by a backward R. It toggles the  left/right  orientation  of  the  image.  The  Edan  E  orientation marker at the top of the image switches with the left/right invert to match the orientation marker on the transducer.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                Imaging - 34 - Up/Down  The Up/Down invert control is indicated by an  upside-down R. It toggles the up/down orientation of the image. The TGC also swaps so the top of the TGC corresponds to the top of the image on the screen. FOV  The Field of View control adjusts the image width. Full, Large, Med. and Small are available. As the image becomes narrower the frame rate increases. Steer  This  control is  only  available  for  linear  transducers. It steers  the B-mode image left or right, without moving the transducer. This can be particularly useful when visualizing needles or other objects that are enhanced by a perpendicular beam.(in    Spatial Compounding, Trapezoid and    Panorama states, steer function is not valid) Trapezoid  The  Trapezoid  control  activates  the  trapezoid  imaging  on  linear transducer. It is a part of the B-mode function and available in live imaging. Panorama  The Panorama control invokes the Panorama function. See section 5.1.4 for details. Image Type    B-mode supports presets for Detail, General, and Penetration. See section 8.3.1.1 for more information. Frame rate  Adjusts  the  tradeoff  between  frame  rate  and  spatial  resolution. Higher settings provide more temporal resolution. Dual B     Press  to  activate  dual  imaging  mode.  Dual  L  activates  the  live image on the left side of image field, and Dual R activates the live image on the right side of image field. Quad  Press to activate quad imaging mode. Refer to section 5.1.3 Quad Imaging for details. 3D/4D  Press to activate 3D/4D imaging mode. Refer to section 5.6 3D/4D mode for details. Needle  Press to invoke the touch screen menu for Needle Enhancement Visualization and Needle Biopsy Guide functions. See section 7.5 and 7.6 for more information.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                Imaging - 35 - Table 5-2 B-mode Touch Screen Controls 5.1.2. Dual Imaging Dual Imaging displays images side by side on the screen. During real-time imaging when Dual imaging is enabled, the active image is displayed in real-time and the other image is frozen. Press the Dual L or Dual R button on the B-mode touch screen to select the left or right image for activation. With both images frozen, press the Dual L or Dual R button to toggle between the images. Dual Exit Dual is exited by any of the following:  Pressing the Dual button corresponding to the currently active side.  Pressing the Quad button.  Pressing the B, M, D hard key.  Pressing the 3D/4D button.  Transducer change, preset recall, or new exam. Color Doppler is available during dual imaging.  5.1.3. Quad Imaging Quad Imaging displays images in four quadrants on the screen. During real-time imaging when Quad imaging is enabled, the active image is displayed in real-time and other images are frozen. Quad imaging is invoked by pressing Quad button on the B-mode touch screen. Quad button is a one-shot button. Each single press on it during real-time imaging activates one of the four images in an order from the top left one to the top right one, then to the bottom left one and to the bottom right one. Each single press on it during frozen status switches among the four images in an order described above but does not change the frozen status of the system.   Quad Exit Quad is exited by any of the following:  Pressing the B, M, D hard key.  Pressing the Dual button  Pressing the 3D/4D button  Transducer change, preset recall, or new exam. Color Doppler is available during Quad imaging. 5.1.4. Panorama Panorama constructs an extended field of view image as the transducer slides along its long axis. Acoustic power*  Adjusts the acoustic output power of the activated transducer. It is only  available  in  live  imaging.  Higher  acoustic  power  numbers correspond to more brightness and contrast in the image and more force of penetration, but the ALARA principle should be followed in actual situations.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                Imaging - 36 - Panorama is available in live B-mode imaging. To acquire a panoramic image:  Press the Panorama button on the B-mode touch screen.  Press the Set button on the trackpad to start acquisition.  Slide the transducer along the long axis  The B-mode image will extend to incorporate the newly imaged anatomy.  When the full anatomy has been acquired, press the Enter button on the trackpad or press the Freeze hard key. System also will enter the panorama review state when exceeding the default time. After  acquisition  the  completed  Panoramic  image  is  redisplayed  to  fit  on  the  screen. Measurements*, Comments and Bodymarkers are then supported.   Note: 1.  If a measurement function in panorama mode is activated, the caution information of “The  accuracy  of  measurements  in  Panorama  is  limited  and  can  be  lower  than measurements in B-image” in yellow font color appears on the bottom of the image and a caution symbol appears on the head of the measurement results. 2.  IMT measurement is not available on panorama image. Panorama is available only with linear transducers. Speed Indicator During an acquisition, a speed indicator bar on the screen will show the current scan speed. Keep the indicator in the green center of the graphic in order to achieve best results.                Most acceptable        normal speed                          too slow                         too fast   Figure 5-1 Panorama Speed Indicator 5.1.5. Zoom Zoom is available in live or frozen B-mode and Color imaging. Press the Zoom hard key to zoom an image. There are 2 ways to zoom an image: The Zoom paddle key or a pinch-out gesture on the track pad.  Zoom Paddle key o Press the upper part of the Zoom paddle to make the image bigger. o Press the lower part of the Zoom paddle to make the image smaller, down to normal size.  Pinch Gesture: Use two fingers on the main part of the trackpad, and move the two fingers outward from each other (“Pinch-out”). The image will enlarge. When  in  Zoom  mode,  use  the  trackpad  to  pan  the  position  of  zoomed  image.  The  B  image remains zoomed after entering other imaging modes.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                Imaging - 37 - 5.2 PW-mode 5.2.1. PW Touch Screen Controls The following touch screen controls impact the PW image. Touch screen controls can be located where you want them to be. See section 3.8 for details. Name Control Description Scale  Scale adjusts the range of velocities that are displayed. Upward presses  increase  scale  and  downward  presses  decrease  it. Increasing  scale  when  the  PW  cursor  is  relatively  deep  may result in invoking HPRF, if it is configured. See section 5.2.2 for details. Baseline  The  Baseline  control  adjusts  the  Doppler  baseline.  Upward presses  move  the  baseline  up  on  the  screen  and  downward presses move the baseline down. Invert  Normally  signals  above  the  baseline  are  positive  velocities (moving  toward  the  transducers).  However,  when  Invert  is pressed then negative velocities are above the baseline. Invert does not affect the baseline position. Angle Correct  Adjust the Doppler scale to account for the angle between the Doppler cursor  and  the  blood  flow.  Upward  presses  increment the  angle.  Downward  presses  decrement  the  angle.  Angles above 80 are not allowed. Quick-Angle  Adjusts the angle correct quickly to one of 60/0/-60. Filter  The  Filter  control  removes  excessive  noise  from  movement  of vessel walls. Options of Low, Med. and High are available.   Colorize  Switches  between  grayscale  and  colorized  (pseudo-color) postprocessing maps. Gray Map Tint   Adjusts the current postprocessing, either grayscale or tinted Dynamic Range  The  Dynamic  Range,  or  log  compression,  adjusts  how  signal intensities  are  converted  to  brightness.  A  high  dynamic  range presents a flat spectrum, while a low dynamic range presents a large range of spectrum brightness.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                Imaging - 38 - SV Size/Gate Size  Gate  adjusts  the  size  of  the  sample  volume  gate.  Upward presses increase the gate size. Downward presses decrease the gate size. Sweep Speed  Sweep adjusts the sweep speed of the Doppler strip. Options of Slow, Low, Med., High and Fast are available. Upward presses increase  sweep  speed.  Downward  presses  decrease  sweep speed. Strip Size  Format changes the relative size of the Doppler strip compared to the reference image. Full, Large, Med. and Small are available.   Volume  Volume adjusts the audio volume of the Doppler strip. This can be adjusted in pre-Doppler to set the initial volume upon invoking Doppler acquisition. Duplex (Triplex)   This  determines  if  the  strip  mode  and  reference  image  are imaging simultaneously or not. If the reference image is B-mode this is labeled “Duplex”. If the reference image includes color this is labeled “Triplex”. Image Type    Strip  Doppler  supports  image  presets  for  Low  Flow,  Medium Flow, and High Flow.   Steer  This control is only available for linear transducers. It steers the Doppler cursor angle left or right. Frequency  This  determines  the  Doppler  transmit  frequency  used  for imaging.   Auto Trace  Press to activate the Auto Trace function on a real-time or frozen PW Doppler strip. The  Auto  Trace  function  automatically  traces  the  spectral Doppler  waveform  and  records  several  measurements  on selected general imaging. See section 7.3.1 for details.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                Imaging - 39 - Table 5-3 PW Mode Touch Screen Controls 5.2.2. HPRF Conventional  PW  Doppler  scales  are  limited  by  the  Nyquist  limit.  High  Pulse  Repetition Frequency (HPRF) allows  the system to exceed  the  Nyquist limit by having  multiple Doppler pulses in the body at the same time. In HPRF imaging multiple Doppler gates are displayed, since the  multiple  Doppler  pulses  could  be  giving  information  from  different  depths.  HPRF  is automatically invoked when needed to maintain the requested depth and scale. For example, if the  system  is  at  a  high  scale  and  you  move  the  cursor  deeper  the  system  may  automatically invoke HPRF. This can also happen if the cursor is deep and you increase scale.   If HPRF is active you will see multiple Doppler gates on the reference image. If you do not want to be in HPRF, decrease the scale or move the cursor to a more shallow location until only one gate is displayed. 5.2.3. Update The Update hard key toggles between acquiring Doppler strip and acquiring the reference image.   5.3 CW-mode Depending on system configuration, CW may be available on the P5-1XQ transducer. There are two ways to invoke CW:  Press the CW button on the Doppler touchscreen to switch between PW and CW modes.  Configure the F1 or F2 programmable hardkey to invoke CW directly (see section 9.1 for details). 5.3.1. CW Touch Screen Controls The following touch screen controls impact the CW image. Touch screen controls can be located where you want them to be. See section 3.8 for details.   Auto Trace Direction  Press  to  specify  which  side  of  the  Doppler  baseline  to  take measurements from. there are 3 options: Up: traces positive portion of waveform (above baseline). Down: traces negative portion of waveform (below baseline). Both: traces waveform on both sides of baseline. Acoustic power*  Adjusts the acoustic output power of the activated transducer. It is only available in live imaging. Higher acoustic power numbers correspond  to  more  brightness  and  contrast  in  the  image  and more  force  of  penetration,  but  ALARA  principle  should  be followed in actual situations. Auto Optimization*  The Auto push button automatically updates the PW gain, DR, baseline and PRF scale. Each single press of the button renews the automatic optimization. Whether Gain, DR or Scale/Baseline is optimized when pressing Auto button can be configured in Set-up. See section 9.1.4 for detail.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                Imaging - 40 - Name Control Description Scale  Scale adjusts the range of velocities that are displayed. Upward presses increase scale and downward presses decrease it.   Baseline  The  Baseline  control  adjusts  the  Doppler  baseline.  Upward presses  move  the  baseline  up  on  the  screen  and  downward presses move the baseline down. Image Type    Strip  Doppler  supports  image  presets  for  Low  Flow,  Medium Flow, and High Flow.   Invert  Normally  signals  above  the  baseline  are  positive  velocities (moving  toward  the  transducers).  However,  when  Invert  is pressed then negative velocities are above the baseline. Invert does not affect the baseline position. Angle Correct  Adjust the Doppler scale to account for the angle between the Doppler cursor and the blood flow. Upward presses increment the  angle.  Downward  presses  decrement  the  angle.  Angles above 80 are not allowed. Quick-Angle  Adjusts the angle correct quickly to one of 60/0/-60. Filter   The Filter control removes excessive noise from movement of vessel walls. Options of Low, Med. and High are available.   Colorize  Switches  between  grayscale  and  tinted  (pseudo-color) postprocessing maps. Gray  Map Tint   Adjusts the current postprocessing, either grayscale or tinted. Dynamic Range  The Dynamic Range, or log compression, adjusts how  signal intensities are converted to brightness. A  high dynamic range presents a flat spectrum, while a low dynamic range presents a large range of spectrum brightness.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                Imaging - 41 - Sweep Speed  Sweep adjusts the sweep speed of the Doppler strip. Options of Slow, Low, Med., High and Fast are available. Upward presses increase  sweep  speed.  Downward  presses  decrease  sweep speed. Strip Size  Format changes the relative size of the Doppler strip compared to  the  reference  image.  Full,  Large,  Med.  and  Small  are available. Volume  Volume adjusts the audio volume of the Doppler strip. This can be  adjusted  in  pre-Doppler  to  set  the  initial  volume  upon invoking Doppler acquisition. Table 5-4 CW Mode Touch Screen Controls 5.4 Color-mode 5.4.1. Color Touch Screen Controls The following touch screen controls impact the color image. Touch screen controls can be located where you want them to be. See section 3.8 for details. Name Control Description Color Mode Variation    A set of radio-buttons display the color modes that are available on the current transducer, and let you switch between them. See section 5.4.2 for details. Scale  Scale adjusts the range of velocities that are displayed. Upward presses increase  scale  and  downward  presses  decrease  it.  It  is  available  in Velocity, PDI, and DPDI. It is not available in Freeze/Cine. Baseline  The Baseline control adjusts the Color baseline. Upward presses move the baseline up on the scale and downward presses move the baseline down.Baseline is not available in PDI mode. Invert  Normally  signals  above  the  baseline  are  positive  velocities  (moving toward the transducers). However, when Invert is pressed then negative velocities  are  above  the  baseline.  Invert  does  not  affect  the  baseline position. Invert is not available in PDI mode.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                Imaging - 42 - Wall Filter  The  Filter  control  removes  excessive  noise  from  movement  of  vessel walls. Options of Low, Med. and High are available. Map  Adjusts the current map for the active color variation. Persist  Persistence determines the number of frames that are averaged together for display. Levels of Off, Low, Med. and High are available.   Smooth Filter  The smoothness filter determines the spatial filtering that is applied to the Color  image.  Higher  filter  levels  create  a  smoother  image.  Upward presses increase the filter. Downward presses decrease the filter. Threshold  When the system receives both B-mode and CD signal from a region the threshold determines whether to display that as grayscale or color. Higher values mean a stronger CD signal is needed to display it in color. Upward presses increase threshold. Downward presses decrease threshold.   Frequency  Frequency  determines  the  transmit  frequency  used  by  color  Doppler. Upward presses increase the frequency. Downward presses decrease the frequency. Steer  This control is only available for linear transducers. It steers the Color ROI box angle left or right. Image Type    Color Doppler supports image presets for Low Flow, Medium Flow, and High Flow. Dual    Live  To activate split screen with simultaneous live B/Color and live B. the live B  image  without  color  and  the  same  live  B  image  with  color  are simultaneously displayed on each side of the image field. Freezing the image will freeze both sides simultaneously. Cine review will review both sides simultaneously.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                Imaging - 43 - Dynamic Range  The Dynamic Range, or log compression, adjusts how echo intensities are converted to brightness. A high dynamic range presents a flat image, while a low dynamic range present a large range of intensities. Only available in PDI/DPDI mode. Frame rate  Adjusts  the  tradeoff  between  frame  rate  and  spatial  resolution.  Higher settings provide more temporal resolution. Acoustic power*  Adjusts the acoustic output power of the activated transducer. It is only available in live imaging. Higher acoustic power numbers correspond to more brightness and contrast in the image and more force of penetration, but ALARA principle should be followed in actual situations. Table 5-5 Color Mode Touch Screen Controls 5.4.2. Color Mode Variants The Acclarix AX8 system supports 3 types of Color Doppler imaging:  Color (Color Doppler): This is velocity Color Doppler that shows direction and velocity of  flow.  Different  colors  represent  different  velocities,  and  positive  flow  has  different colors than negative flow.  PDI (Power Doppler Imaging): PDI shows the power, or intensity, of the Doppler signal. PDI is typically more sensitive to low levels of flow, but can not distinguish the velocity or direction of the flow.  DPDI (Directional Power Doppler Imaging): This is similar to DPI in that it shows the power of the Doppler signal instead of the velocity. However, it does map positive flow to different colors than negative flow. 5.5 M-mode 5.5.1. M-mode Touch Screen Controls The following touch screen controls  impact the  M-mode image. Touch screen controls can be located where you want them to be. See section 3.8 for details. Name Control Description Colorize  Switches  between  grayscale  and  colorized  (pseudo-color) postprocessing maps. Gray Map Tint   Adjusts the current postprocessing, either grayscale or tinted.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                Imaging - 44 - Dynamic Range  The Dynamic Range, or log compression, adjusts how  signal intensities are converted to brightness. A  high dynamic range presents a flat spectrum, while a low dynamic range presents a large range of spectrum brightness. Focus Pos  The focus position set in B-mode applies to M-mode as well. Upward presses move  the  focus  shallower,  regardless of  the U/D invert status of the M-mode. Downward presses move the focus deeper. Sweep Speed  Sweep adjusts the sweep speed of the M-mode strip. Options of Slow, Low, Med High and Fast are available. Upward presses increase  sweep  speed.  Downward  presses  decrease  sweep speed. Strip Size  Format changes the relative size of the M-mode strip compared to the reference image. Full, large, Med and small are available.   Side-by-side  This  is  an  on/off  control.  When  On,  the  M-Mode  strip  is side-by-side  with  the  B-mode  image.  When  Off,  the  M-Mode strip is below the B-mode image Line Persistence  The line Persistence determines how many M-mode lines are averaged  together  for  the  display,  similar  to  persistence  in B-Mode. Levels of Off, Low, Med and High are available.   Frequency  Frequency determines the transmit frequency used by M-Mode. Upward  presses  increase  the  frequency.  Downward  presses decrease the frequency. Acoustic power*  Adjusts the acoustic output power of the activated transducer. It is only available in live imaging. Higher acoustic power numbers correspond to more brightness and contrast in the image and more  force  of  penetration,  but  ALARA  principle  should  be followed in actual situations. Table 5-6 M-Mode Touch Screen Controls 5.6 3D/4D Mode* Depending on the system configuration and regulatory approval, 3D/4D may be available on the C5-2MQ* transducer. The first step in using 3D or 4D is to activate pre-3D mode. There are two ways to activate pre-3D mode: 1. Press the 3D/4D button on the touch screen in B-mode. 2. Configure  the  F1 or  F2  hard  key  to  invoke  3D/4D.  There  are  two  options  to  do  so,  one labeled „3D‟ and the other „4D‟. These both invoke pre-3D, but with different defaults for 3D
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                Imaging - 45 - and 4D. NOTE: Measurement function is not available in 3D/4D. 5.6.1. Pre-3D   The Pre-3D mode supports defining the 3D/4D acquisition location and settings. Figure 5-2 shows an example of the pre-3D image. The ROI (Region of Interest) indicates the portion of the image that will be used for 3D/4D. The position/size of the ROI can be adjusted using the trackpad and Set key. Also note that there is a horizontal line going across the ROI near the top and it can be adjusted to a curved line upwards and downwards by using the trackpad. This is the clip plane. The initial volume rendering will show everything below this line.    Figure 5-2 Pre-3D Image The touch screen controls in pre 3D interface: Name Control Description Image quality  Adjusts  the  trade-off  between  speed  and  quality  of  the acquisition.  Settings  include  High  (quality),  Medium,  and Fast. Angle  Adjusts the angle of the volume acquisition. Upward presses increment  the  angle.  Downward  presses  decrement  the angle. Angle range is from 5ºto 70º in increments of 5º.   Baby Spine/   Baby Face   These  two  radio  buttons  determine  if  the  acquisition  is optimized for  looking at a  Baby Spine (Maximum Intensity Projection), or a Baby Face (Surface Render). Thresh (only in surface render mode), bright and contrast will be adjusted
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                Imaging - 46 - Table 5-7 Pre-3D Touch Screen 5.6.2. 3D Volume Sweeping During the process of 3D volume sweeping, the images swept in ROI box are displayed in the image area.  Figure 5-3 3D Image Scanning While scanning, the sector    in the right bottom is filled gradually,and no buttons on the control panel and touch screen can be activated. 5.6.3. 3D Image Review If 3D was selected, then the system will enter to the 3D review status after finishing one sweep. automatically. 3D/4D    These two radio buttons determine if the acquisition will be 3D or 4D. Start    Start  an  acquisition.  If  3D  was  highlighted  then  a  single volume is acquired. If 4D was highlighted then continuous acquisitions are started. Exit  Exit 3D/4D
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                Imaging - 47 -  Figure 5-4 3D Image Review  There are two image modes: Volume imaging mode and multi-slice imaging mode. Figure 5-4 shows Volume Imaging Mode in quad screen with a Baby Face volume rendering.    Quadrant A shows a slice through the data that mimics the original ultrasound image.  Quadrant B is orthogonal to A, as if the transducer had been rotated 90 degrees.  Quadrant C is orthogonal to A and B, showing a slice that is parallel to the transducer face. The icon    at the bottom-right shows the position of each slice with respect to the full 3D data set. The following table shows the controls that are available in Volume Imaging Mode     Name Control Description Volume   These  two  radio  buttons  toggle  between  the  Volume display and the Multi-Slice display. A/B/C/3D      These four radio buttons select which quadrant is the focus of  the  navigation/panning  controls.    A/B/C  are  three orthogonal  slices  through  the  volume,  while  „3D‟  is  the rendered image. Single, Dual, Quad  These three radio buttons switch the display to show 1, 2, or  4  images  at  once.  Single  shows  the  3D  image,  dual shows the A slice and 3D image, Quad shows three MPR slices and 3D image. Reset  Reset the operation of pan, rotate and zoom to the initial condition. Pan  Pan  image  along  the  x-axis  or  y-axis  of  the  activated window. Use the trackpad to pan image along x-axis and
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                Imaging - 48 - y-axis. VOI  Press  to  activate  the  function  of  adjusting  VOI  or  clip plane. Use the trackpad to adjust and press Set to switch between VOI and clip plane.   Cut  Function of cut. After activating this button, the 3D image is selected, press Set to locate the start point, use trackpad to trace and second press Set to locate the end point The region within these two points will be cut. The system will draw  a  straight  line  between  the  start  and  end  point  to enclose the cut region. Undo  Undo the previous cut operation Redo  Redo the previous undo operation Undo All  Return to the condition before cutting   Threshold  Adjust  the  threshold  of  surface  render  mode,  Upward presses increase threshold. Downward presses decrease threshold.     Bright.  Adjust the bright of 3D image   Contrast  Adjust the contrast of 3D image. 3D Smooth/MPR Smooth      Adjust smoothness of A\B\C slices and 3D images. Higher filter  levels  create  a  smoother  image.  Upward  presses increase the filter. Downward presses decrease the filter.   Rotation  Quickly rotate the  image. Available angles are:  0°/90°/180  °/270°  .
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                Imaging - 49 - Table 5-8 Volume imaging Mode Touch Screen Image area and touch screen controls in multi-slice imaging mode: Rotation X  Rotate the image by the X-axis of the activated window. Rotation Y  Rotate the image by the Y-axis of the activated window. Rotation Z  Rotate the image by the Z-axis of the activated window. New Volume  End  the  3D  volume  sweeping  and  return  to  the  pre-3D interface Exit  Exit 3D/4D MPR Review  MPR  (multi-planar  reconstruction)  review  function.  After this  button  is  activated,  the  3D  image  is  hidden,  two perpendicular  lines  appear  on  the  selected  quadrant. These two lines are color matched with the other 2 MPR quadrants,  and  show  where  the  three  planes  intersect. Left/right  trackpad  movement  moves  one  of  these  lines while up/down moves the other line.   3D Tint  Select 3D tint, options of    Clear,Candle,Copper,Sepia,Gold.  MPR Tint  Select  MPR  tint,  options  of  Gold,  Sepia,  Blue,  Ice  and Clear are available.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                Imaging - 50 -  Figure 5-5 Multi-slice Imaging Mode In figure 5-5, the first image shows the basis slice, other images are orthogonal to the basis slice and parallel to each other.    Lines on the basis slice indicate the locations of the parallel slices. If there are more slices than images then the un-displayed slices are shown with a dotted line. A number below each parallel slice shows the slice location in the basis slice, with 0 being the middle slice.     Control Name Description A/B/C    These three radio buttons determine which slice isthe required parallel slice.    These three choices are the same as on the Volume display in quad mode. 3*3, 2*2, 3*2    These three radio buttons determine the layout of the display. Pan  Pan the basis slice along the x-axis or y-axis of the activated window. Use the trackpad to pan the basis slice along x-axis y-axis. Parallel slices change with the pan operation. Reset  Reset  the  operation  of  pan,  rotate  and  zoom  to  the  initial condition. Number  Select how many slices to use
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                Imaging - 51 - Table 5-9 Multi-slice Mode Touch Screen 5.6.4. 4D Volume Acquisition Press 3D/4D button on the touch screen in B-mode to enter pre-3D, select 4D, select Baby Spine or Baby  Face fist and use the  ROI box  to set  the region. Only ROI will  be  stored during the volume sweep. Use the trackpad to set the size and location of the ROI box. Press Set to confirm. Use the trackpad to set the edge of VOI (Volume of interest) and press Start button on the touch screen or the Enter button on the trackpad to begin sweep.   Distance  Adjust the distance between slices.   Next  When there are more slices than can be displayed at once this control cycles through the subset of slices displayed.   Tint  Select  multi-slice  tint,  options  of  Gold,  Sepia,  Blue,  Ice  and Clear are available. Smooth    Adjust smoothness of A\B\C slices. Higher filter levels create a smoother  image.  Upward  presses  increase  the  filter. Downward presses decrease the filter. Rotation X  Rotate the basis slice by the X-axis of the activated window. Parallel slices change with the rotation operation. Rotation Y  Rotate the basis slice by the Y-axis of the activated window. Parallel slices change with the rotation operation. Rotation Z  Rotate the basis slice by the Z-axis of the activated window. Parallel slices change with the rotation operation.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                                Imaging - 52 - 5.6.5. 4D Live Volume   4D  Live  volume  continuously  sweeps  and  displays  successive  3D  volumes.    During  live  4D imaging  the  following  controls  are  available:  activate  any  window  of  A/B/C/3D, Single/Dual/Quad,VOI, Threshold, Bright, Contrast, 3D Smooth/ MPR Smooth, 3D Tint/ MPR Tint, Rotation, Rotation X, Rotation Y, Rotation Z , Zoom and Gain in B mode.  Figure 5-6 4D Volume Sweeping Image As shown in figure above, image in the left is A slice and image in the right is the live 4D volume rendering. 5.6.6. 4D Cine During the process of 4D live volume imaging; press the Freeze button on the control panel to enter the 4D cine mode. A cine replay scroll bar appears below the images. Use the trackpad to replay  the  reset  3D  images.  For  the  selected  frame,  user  can  execute  all  operations  that  are available 3D review status. Press Freeze again to return to the live image.  Figure 5-7 4D Cine   Touch screen controls are the same as that of static 3D except for Cine button.    Table 5-10 4D Touch Screen Name Control Description cine  Switch between 3D and 4D review state
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                          Transducer - 53 - 6 Transducers 6.1 Transducer Model No. Model Type Application   1 C5-2XQ/ C5-2Q Convex Abdomen Fetal / Obstetrics Urology Gynecology Musculo-skeletal   Nerve 2 E8-4Q Endocavity Fetal / Obstetrics Endo-vaginal Endo-rectal : Urology 3 L10-4Q Linear Small Parts (Thyroid, Testes, Breast) Peripheral Vascular Musculo-skeletal   Neurology 4 L12-5Q Linear Small parts (Thyroid, Testes, Breast)   Vascular   Musculoskeletal   Neurology 5 L17-7HQ Linear Small Parts (Thyroid, Testes, Breast) Peripheral Vascular Musculo-skeletal   Neurology 6 L17-7SQ Linear Intra-operative Musculo-skeletal   Nerve Vascular 7 MC8-4Q Micro Convex Pediatric Neonatal    (including abdominal and cephalic) Neurology Vascular   8 P5-1XQ Phase array Adult Cardiac Abdomen Pediatric Cardiac Adult cephalic* 9 C5-2MQ* Convex Fetal / Obstetrics   Table 6-1 Transducer Model & Application
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                          Transducer - 54 - 6.2 Name and Function of Each Part of the Transducer Figure 6-1 shows an example of a transducer.  Figure 6-1 Typical transducer Name   Function Transducer head Converts electrical signals to sound waves, and then converts received echoes back to electrical signals. The tip of the transducer is the acoustic lens.   Transducer cable Transmits electrical signals between the transducer head and the transducer connector. Transducer connector Connects the transducer to the ultrasonic diagnostic system  6.3 Transducer Cleaning and Disinfecting Transducers  should  be  cleaned  and/or  disinfected  as  necessary  or  between  use  with  a recommended  cleanser  or  disinfectant.  Disconnect  the  transducer  from  the  system  prior  to cleaning and disinfecting. The validated cleaning agents for cleaning the transducers are:    Mild near neutral detergent  Ethanol (75%)  Isopropanol (70%) Cleaning agents should be applied and removed using a clean, soft, non-abrasive cloth or paper towel.   The validated disinfectants for transducer are:  Ethanol (75%) Transducer head Transducer connector Transducer cable
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                          Transducer - 55 -  Isopropanol (70%)  OPA (0.55%) If Ethanol or Isopropanol is used for both cleaning and disinfecting, then a new cloth is required for the disinfection step. WARNING 1.  Unplug the transducer from the system prior to cleaning or disinfecting. 2.  To  avoid  infection,  always  use  protective  gloves  when  performing  cleaning  and disinfecting procedures. 3.  To  avoid  infection, ensure  that  expiration  date  of  the  disinfecting  solution  has  not passed. To clean the transducers: 1. Disconnect the transducer from the system. 2. Wipe the patient contact area with a soft cloth dampened with the cleaning solution until no visible contaminants remain. 3. Wipe off the cleaning solution with a fresh cloth or towel, dampened with tap water after cleaning until no visible cleaning agent remains.   4. Wipe off with a dry cloth to remove residual moisture. 5. Leave the transducer to air dry. NOTE:A single-use sheath should be used on E8-4Q transducer. Before cleaning the transducer,  remove  the  sheath  gently and  discard  it.  Put on  a new single-use  sheath before using the transducer. CAUTION 1.  Always wear gloves to perform the following steps. 2.  To minimize the disease transmission, use of legally marketed, sterile transducer and sheath  is  strongly  recommended  for  intra-cavitary  and  intra-operative  procedures. Legally  marketed,  sterile,  pyrogen  free  transducer  sheath  is  required  to  use  for neurological intra-operative procedures. 3.  DO NOT use an expired transducer sheath. Check whether the term of validity has expired prior to using transducer sheaths. 4.  The single-use sheath should comply with the local regulations. To install transducer sheath: 1. Place an adequate amount of sterile coupling gel on the acoustic window of the transducer; 2. Insert the transducer into the sheath; 3. Pull the sheath over the transducer and cable until the sheath is fully extended; 4. Secure the sheath using the bands or clips supplied with the sheath; 5. Check  and  eliminate  bubbles  between  the  face  of  the  transducer  and  the  sheath.  Bubbles between the face of transducer and the sheath may affect the ultrasound imaging. 6. Inspect the sheath to ensure that there are no damages (i.e. holes or tears).
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                          Transducer - 56 - There are two methods to disinfect the transducer (use the immersion method for the E8-4Q and L17-7SQ): Non-immersion: 1. Disconnect the transducer from the system. 2. Prepare the disinfectant solution (75% ethanol or 70% isopropanol). 3. Spray the solution to the transducer interface or wipe it with a soft cloth dampened with the disinfectant solution. 4. Rinse the transducer according to the disinfectant instructions. Wipe the transducer with a dry cloth or leave the transducer to air dry. 5. Inspect the transducer to ensure that there is no damage. Immersion: 1. Prepare the disinfectant solution  (Cidex OPA) 2. Place the cleaned and dried transducer in contact with the disinfectant (refers to figure 6-2 for the contacting area) for the time specified by the disinfectant manufacturer. 3. Rinse the transducer according to the disinfectant instructions. Wipe the transducer with a dry cloth or leave the transducer to air dry. 4. Inspect the transducer to ensure that there is no damage. .                Figure 6-2 Depth of the Transducer Immerged into Disinfectant WARNING 1.  Do not immerse the transducer connector. If the cable connector is immersed, do not plug the connector into the system. Rinse the connector under running water and dry it thoroughly. If necessary, contact EDAN for service. C5-2XQ/ C5-2Q P5-1XQ L17-7SQ L17-7HQ L10-4Q/ L12-5Q Disinfectant E8-4Q MC8-4Q C5-2MQ*
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                          Transducer - 57 - 2.  Prohibit infiltration of any type of liquid into the device or the transducer. 3.  Do not sterilize the transducer using techniques such as autoclave, ultraviolet, gamma radiation, gas, steam, or heat. Severe damage May result. 4.  Use only ultrasound coupling gel that complies with local regulations. 5.  Do  not  immerse  the  AC  adapter  and  connector  of  the  transducer  into  solutions. Transducers can be submerged to, but not including, the strain relief of the transducer array. Do not immerse or soak any part of a transducer in any cleaning material not listed in the recommended list of disinfectant. 6.  Use the Cidex OPA immersion method to disinfect the endocavity transducer. 7.  Only non-immersion method can be used with solution of ethanol or isopropanol. The solution concentration should not exceed the value provided above. 8.  The immersion time should not exceed the time that is specified by the disinfectant manufacturer. 9.  Patient contact area should be immersed into the solution while using the immersion method, but should not exceed the depth shown in figure 6-2. Proper Use of Transducers In order to extend the service life and to obtain optimum performance of the transducer, please operate as follows: Inspect AC adapter, socket and acoustical window of the transducer periodically. Shut down the machine before connecting or disconnecting the transducer. Do  not  drop  the  transducer  onto  the  floor  or  collide  with  hard  objects.  Otherwise  it  will  be damaged easily. Do not heat the transducer. Do not pull or bend the transducer cable. Coupling gel can only be used on the head of the transducer, and it should be wiped off after use. Clean and disinfect the transducer after every use. The acoustical window and the shell of the transducer should be examined frequently.   CAUTION 1.  Do not disinfect or clean transducers under high temperature, and the temperature should be below 45°C. 2.  In order to avoid damaging the device, the disinfection method is limited to regular maintenance of devices in hospitals. Disinfecting instruments should be cleaned first.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 58 - 7 Features 7.1 Comments The Comments function lets you add annotation to an image. The Comment function is invoked by pressing the Comment hard key on the console. Figure 7-1 shows an example touch screen for Comments. The top portion of the screen shows comments that are pre-defined in pre-sets. See  section  8  for  details  on  how  to  configure these.  The  bottom  portion  of  the  screen  shows controls available while the Comment function is active. Table 7-1 gives an overview of these controls.  Figure 7-1 Comment Touch Screen Button Description Home Set the cursor in a pre-defined location.   Set Home Set the current cursor location as the new Home location Move Block Select comments for moving to desired locations Delete Delete a whole text or text group or an arrow Delete All Clear all comments on the image area Arrow Create a new arrow for annotating Keyboard Toggle between showing pre-defined comments(comment library) or a touch screen keyboard Table 7-1 Comment Touch Screen Description Adding comments You can add either pre-defined comments or use a keyboard. Adding comments using the keyboard 1. Invoke the Comment function. 2. Press the “Keyboard” touch screen key to display a keyboard on the touch screen. 3. Move the cursor to the desired location and enter the desired text.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 59 - Adding comments using pre-defined comments 1. Invoke the Comment function 2. Move the cursor to the desired location and press the desired pre-defined comment. Some of the pre-defined comments may be grouped together with a border surrounding them (such as „Rt‟ and „Lt‟ in figure 7.1). These are special keys that let you quickly replace one term with another, regardless of where the cursor is within a block of text. For example, if you had entered “Rt. Kidney” as a block of text, pressing the “Lt.” button would change that block to “Lt. Kidney” even if the cursor is not at the specific location. Adding arrows 1. Invoke the Comment function 2. Press Arrow button on the touch screen to display an arrow at the current cursor location. 3. Move  the  arrow  to  the  desired  location.  Note  that  while  the  arrow  is  being  moved  the orientation of the arrow is fixed. 4. Press  the  Pos/Angle  button  to  set  the  tip  of  the  arrow  in  its  current  location  and  toggle adjusting  the  angle  of  the  arrow.  The  trackpad  can  now  be  used  to  rotate  the  arrow.  The Pos/Angle button can be used to toggle between moving the head of the arrow and moving the tail of the arrow. The selected button is yellow colorized in the background.   5. Press the Enter button to complete the arrow. You can now enter text at the end of the arrow. Modifying comments 1. Move the cursor to a desired comment, notice that an insert cursor appears in the text as the cursor moves over an existing block of text. Enter text using either the keyboard or pre-defined comments. The new comments will be added to the old text. Deleting comments There are several ways to delete comments:  Move the cursor to the desired block of comments and press the Delete button on the touch screen to delete that block of comments.    Press the Delete All button on the touch screen to remove all comments.  Press  the  Clear  hard  key  on  the  console  to  remove  all  comments,  bodymarkers,  and measurements. 7.2 Body Mark The Body Mark lets you add a bodymarker graphic to an image and indicate the location of the transducer  using  a  transducer  icon  on  that  graphic.  The  Body  Mark  function  is  invoked  by pressing the Body Mark hard key on the console. Figure 7.2 shows an example touch screen for Body Mark. The main portion of the screen shows a grid of bodymarkers that are pre-defined in pre-sets. See section 8 for details on how to configure these. The bottom portion of the screen shows controls available while the Comment function is active. Table 7-2 gives an overview of these controls.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 60 -  Figure 7-2 Body Mark Touch Screen Button Description Delete Delete the graphics on the image area Angle Rotates the transducer icon on the bodymarker graphic.   Table 7-2 Body Mark Touch Screen Description Adding a body mark: 1. Invoke the Body Mark function The default bodymarker graphic appears on the main screen, with the transducer icon shown in the default position. 2. Select a desired bodymarker graphic on the touch screen, that graphic replaces the default. 3. Move the transducer icon as needed. The orientation of the transducer icon can be changed with the Angle touch screen key. Moving the transducer icon: 1. Press Probe/Graphic to toggle moving the transducer icon. The transducer icon can not be moved outside of the bodymarker graphic area. 2. Press Probe/Graphic to confirm the new position of the transducer icon and toggle moving the bodymarker graphic. Probe/Graphic button can be used to toggle between moving the transducer icon and moving the bodymarker graphic. The selected button is yellow colorized in the background.   Deleting a body mark: There are two ways of deleting a bodymarker graphic:  Press Delete on the touch screen to delete the graphics on the image area.  Press the Clear hard key to delete all bodymarkers, comments, and measurements. 7.3 Measurements NOTE:  This  section  provides  general  introduction  of  generic  and  application measurements. For  details  about  specific  application  measurements and  calculations, please refer to the Acclarix AX8 Advanced Manual.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 61 - The Measure function lets you perform measurements on a live or frozen image. The Measure function  is  invoked  by  pressing  the  Measure  hard  key  on  the  console.  Figure  7-3  shows  an example touch screen for Measure. There are two types of measurements: Generic  Measurements:  These  are  simple  tools  like  Distance  or  Area.  There  is  no  specific anatomy associated with these measurements and they do not appear in a report. Application Measurements: These are measurements for specific anatomy or clinical conditions. The results can be entered into a report that can be printed later. On the Measure touch screen the lower part of the touch screen shows Generic Measurements. These are different for each imaging mode, but are the same for all presets. The upper part of the touch screen shows Application Measurements. These are different for each imaging mode and each preset. See section 8 for details on how to configure these.  Figure 7-3 Measure Touch Screen Tool Options Some measurements can be done with a choice of different tools, and each tool has a choice of multiple output result items. The users can configure the default tool and result items for these measurements by Tool Options. For result options, at least one item should be selected.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 62 -  Figure 7-4 Tool Options Touch Screen Moving the measurement result window The measurement result window is displayed on the top left corner of the image field by default, the system supports changing the display position of the result window by dragging it. To change the display position of result window: 1. Press Cursor hard key to invoke a cursor after a measurement is done. 2. Move the cursor over the result window, and press Set key, the result window will turn green. 3. Move on the track pad to locate a desired position within the image field for the result window. 4. Press Set key again to fix the result window on the new position. The system supports a different result window position for each image mode. Note: The measurement result window cannot be moved when measure in review state.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 63 -  Figure 7-5 Moving the Measurement Result Window 7.3.1. Generic Measurements A default generic measurement is automatically started when Measure is invoked. Each imaging mode supports several types of measurements, as described below. You can switch between these types  using  the  touch  screen  buttons  along  the  bottom  of  the  touch  screen.  Some  generic measurements  have  options,  for  the  method  to  perform  the  measurement.  When  the  current generic measurement has options available you can view and change those options through the Tool Options touch screen buttons. 7.3.1.1. B-mode Generic Measurements The Acclarix AX8 supports the following types of generic B-mode measurements:  Distance  Circ/Area  Angle  Volume  Stenosis Distance: The distance measurement is always done with a caliper pair. To measure Distance with a caliper pair: 1. Invoke the measure function on a B-mode image. 2. If Distance is not the default measurement, select Distance from the touch screen. 3. Move the caliper to the desired location. 4. Press the Set key to place that caliper and add the second half of the caliper pair.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 64 - Further presses of the Set key toggle which side of the caliper can be moved. Pressing the Enter key completes the caliper and allows you to create another tool. Circ/Area: Area and Circumference can be measured with an Ellipse or Trace To measure Area or Circumference with an Ellipse: 1. Invoke the measure function on a B-mode image. 2. If Area is not the default measurement, select Area from the touch screen.   If Ellipse is not the default tool, press Tool Options to switch to Ellipse 3. Move the caliper to the desired location. 4. Press the Set key to place that caliper and add the second half of the caliper pair. 5. Press the Set key to place the second half of the caliper. An ellipse appears and can be adjusted with the trackpad. Further presses of the Set key toggle between controlling each side of the ellipse caliper and the ellipse diameter. Pressing the Enter key completes the caliper and allows you to create another tool. To measure Area or Circumference with a Trace: 1. Invoke the measure function on a B-mode image. 2. If Area is not the default measurement, select Area from the touch screen.   If Trace is not the default tool, press Tool Options to switch to Trace. Trace can be done with a Draw option. Draw will follow your movements. 3. Move the caliper to the desired location. 4. Press the Set key to place the caliper and start drawing the trace. 5. Moving the caliper now leaves behind a trail of dots to mark the trace outline 6. Press the Enter key to complete the trace. Angle: The Angle measurement is always done with an angle tool. To measure Angle: 1. Invoke the measure function on a B-mode image. 2. If Angle is not the default measurement, select Angle from the touch screen.   3. Move the caliper to the vertex of the angle you are measuring. 4. Press the Set key to place that point and create a new caliper. Move that to one end of the angle. 5. Press the Set key to place that end and create a new caliper. Move that to the other end of the  angle.  Further  presses  of  the  Set  key  toggle  between  controlling  each  of  the  three points that make up the angle. Pressing the Enter key completes the angle and allows you to create another tool. Volume: Volume can be measured with 3 Distance. To measure Volume with 3 Distance:   1. Invoke the measure function on a B-mode image. 2. If Volume is not the default measurement, select Volume from the touch screen.   Move the caliper to the desired location.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 65 - 3. Press the Set key to place that caliper and add the second half of the first caliper pair. 4. Further presses of the Set key toggle which side of the caliper can be moved. Pressing the Enter key completes the first caliper pair and creates the second caliper pair. 5. Repeat step 3~5 to complete the second caliper pair and create the third caliper pair.   6. Repeat step 3~5 to complete the third caliper pair and allow you to create another tool. Stenosis: Stenosis can be measured with Caliper Pairs, Ellipses, or Traces To measure Stenosis with Caliper Pairs: 1. Invoke the Measure Function on a B-mode image. 2. If Stenosis is not the default measurement, select Stenosis from the touch screen. If Caliper is not the default tool, press Tool Options to switch to Caliper. 3. Move the caliper to the desired location. 4. Press the Set key to place that caliper and add the second half of the caliper pair. 5. Further presses of the Set key toggle which side of the caliper can be moved. Pressing the Enter key completes the caliper and creates the second caliper pair tool. 6. Repeat step 3~5 to complete the second caliper pair and allow you to create another tool. To measure Stenosis with Ellipses: 1. Invoke the Measure Function on a B-mode image. 2. If Stenosis is not the default measurement, select Stenosis from the touch screen. If Ellipse is not the default tool, press Tool Options… to switch to Ellipse. 3. Move the caliper to the desired location. 4. Press the Set key to place that caliper and add the second half of the caliper pair. 5. Press the Set key to place the second half of the caliper. An ellipse appears and can be adjusted with the trackpad. 6. Further presses of the Set key toggle between controlling each side of the ellipse caliper and the ellipse diameter. Pressing the Enter key completes the first Ellipse and creates the second Ellipse tool. 7. Repeat step 3~6 to complete the second Ellipse and allow you to create another tool. To measure Stenosis with Traces: 1. Invoke the Measure Function on a B-mode image. 2. If Stenosis is not the default measurement, select Stenosis from the touch screen. If Traces is not the default tool, press Tool Options to switch to Traces. 3. Move the caliper to the desired location. 4. Press the Set key to place the caliper and start drawing the trace. 5. Moving the caliper now leaves behind a trail of dots to mark the trace outline 6. Press the Enter key to complete the trace. 7. Repeat step 3~6 to complete the second trace and allow you to create another tool.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 66 - 7.3.1.2. Strip Doppler Generic Measurements The Acclarix AX8 supports the following types of generic strip Doppler measurements:  Caliper  Trace  Auto Trace  HR (Heart Rate)  RI (Resistive Index) Caliper: The Caliper measurement can provide a wide range of results. To use the Caliper measurement: 1. Invoke the Measure Function on a Doppler strip. 2. If Caliper is not the default measurement, select Caliper from the touch screen.   3. Move the caliper to the desired location. At this point the measurement can display the following results (the actual results shown will depend on the preset):  Vel  PG   4. If you want a caliper pair, press the Set key to place the caliper and add the second half of the  caliper  pair.  At  this  point  the  measurement  may  display  the  following  results  (the actual results shown will depend on the preset):  V1  V2  RI  S/D  Time  ΔV  Accel  PG1  PG2  PHT Further  presses  of  the  Set  key  toggle  which  side  of  the  caliper  can  be  moved. Pressing the Enter key completes the caliper and allows you to create another tool. The  S/D,  PI,  and  PHT results  assume  that  the  first  caliper  is  placed  on  the  peak systolic point. Trace: The Trace measurement can provide a wide range of results. To use the Trace measurement: 1. Invoke the Measure Function on a Doppler strip. 2. If Trace is not the default measurement, select Trace from the touch screen. 3. Trace can be done with a Draw. Draw will follow your movements   4. Move the caliper to the desired start point.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 67 - 5. Press the Set key to place the start point and start drawing the trace. 6. Moving the caliper now leaves behind a trail of dots to mark the trace outline. Press the Set key to adjust the positions of the PS or ED points in the trace. 7. Press the Enter key to complete the trace. 8. At this point the trace measurement may display the following results (the actual results shown will depend on the preset):  PS  ED  MD  TAMax  PGmax  PGmean  PI  RI  S/D  VTI  Time  AT  DT NOTE: Measure trace can only be active in image frozen status. Auto  Trace:  Auto  Trace  is  capable  of  determining  the  maximum  and/or  mean  blood  flow  in freeze mode. Each result that it calculates is  associated with  either the  maximum or the mean value. For activate auto trace in real time mode, press Auto Trace button in PW touch screen. To use the Auto Trace measurement: 1. Invoke the Measure Function on a Doppler strip. 2. If Auto Trace is not the default measurement, select Auto Trace from the touch screen. The trace waveform(s) automatically appear. If any maximum-related results are enabled then the maximum value waveform is displayed on the Doppler strip. If any mean-related results are enabled then the mean value waveform is displayed. 3. The Trace measurement may display the following results (the actual results shown will depend on the preset):  PS  ED  MD  TAMax  TAMean  PGmax  PGmean  PI  RI  S/D  VTI  Time  AT  DT
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 68 -  HR NOTE: Measure Auto trace can only be active in image frozen status. Live auto trace is available as a separate feature from the Doppler touch screen. HR:The HR measurement can provide the heart rate calculation by measured the time of the heart beat cycles. To use the HR measurement: 1. Invoke the Measure Function on a Doppler strip. 2. If HR is not the default measurement, select HR from the touch screen. 3. Move the caliper to position the first heart beat. Press the Set key to activate a second caliper. 4. Move the second caliper to position the next beat. 5. Press the Enter key to complete or press the Set key to switch moving the first caliper. 6. The  HR  measurement  can  display  the  following  result(the  actual  results  shown  will depend on the preset):  HR RI: The RI measurement can measure the PS and ED to calculate the result of RI and S/D. To use the RI measurement: 1. Invoke the Measure Function on a Doppler strip. 2. If RI is not the default measurement, select RI from the touch screen. 3. Move the caliper to position the PS point. Press the Set key to activate a second caliper.   4. Move the second caliper to position the ED point. 5. Press the Enter key to complete or press the Set key to switch moving the first caliper. 6. The  RI  measurement  can  display  the  following  results  (the  actual  results  shown  will depend on the preset):  PS  ED  RI  S/D 7.3.1.3. M-mode Generic Measurements The Acclarix AX8 supports the following types of generic M-mode measurements:  Caliper  HR (Heart Rate) Caliper: The Caliper measurement can provide a wide range of results. To use the Caliper measurement: 1. Invoke the Measure Function on an M-mode strip. A caliper measurement is automatically invoked. It displays the Depth at the current location. 2. Move the Caliper to the desired location.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 69 - 3. Press the Set key to place that caliper and add the second half of the caliper pair. Further presses of the Set key toggle which side of the caliper can be moved. Pressing the Enter key completes the caliper and allows you to create another tool 4. The Caliper measurement can display the following results:  Distance  Time  Slope HR:The HR measurement can provide the heart rate calculation by measured the time of the heart beat cycles. To use the HR measurement: 1. Invoke the Measure Function on a Doppler strip. 2. If HR is not the default measurement, select HR from the touch screen. 3. Move the caliper to position the first heart beat. Press the Set key to activate a second caliper. 4. Move the second caliper to position the next beat. 5. Press the Enter key to complete or press the Set key to switch moving the first caliper. 6. The  HR  measurement  can  display  the  following  result(the  actual  results  shown  will depend on the preset): 7.3.2. Application Measurements and Calculations Application Measurements have a pre-defined meaning and can be entered into a report. Acclarix AX8  supports  the  following  application  packages,  each  with  its  own  set  of  measurements, calculations and report:  OB (including multiple fetuses)  Abd  Vascular  Cardiac  Gynecology  Small Parts  Urology  Pediatric(including Neonatal) Please see the Advanced User Manual for the complete content of each package. When you select an application measurement from the touch screen it will automatically invoke the type of measurement it needs. For example, if you select “BPD” from the OB application it will  automatically  invoke  a  distance  measurement.  These  measurements  generally  behave  as described above  for  generic measurements.  However, a  final  press  of  the  Enter  key after  the measurement is complete will enter the result into the report. Some application measurements can have multiple variations. For example, in a twin OB exam the OB measurements can be done on either fetus. As another example, in Vascular exams several measurements  can  be  done  in  a  Proximal,  Mid,  or  Distal  location.  When  a  measurement  has
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 70 - multiple variations you will see buttons on the touch screen that let you pick which variation you are measuring. 7.3.2.1. Auto IMT (Note: Auto-IMT is not available in the USA and some other countries.) Intima-media  thickness  (IMT)  is  a  measurement  of  the  thickness  of  tunica  intima  and  tunica media. The system provides an automatic measurement of the Carotid IMT.  Figure 7-6 IMT Touch Screen To measure IMT: 1. Invoke the B-mode Measure Function in Carotid Preset. 2. Select an  IMT measurement  on  the touch  screen.  A  ROI box  will be  displayed on  the image field. 3. Select  the  buttons  on  right  side  of  the  touch  screen  to  measure  a  different  position  of carotid artery. 4. Press Set key to resize the ROI box, move the ROI box to locate the measure position. 5. Press Enter key to  complete  and  get  the  Auto  IMT measure  results.  Press  Enter Key again to save the results in the worksheet. 6. If you are not satisfied with the measurement result, press Undo key to restart the IMT measurement. 7. The IMT measurement can display the following result:  Max (the maximum value of the IMT results)  Min (the minimum value of the IMT results)  Avg (the average value of the IMT results)  Width (the width of the ROI box)  SD (the Standard Deviation of the IMT results)  Points (the number of valid points in the IMT results) NOTE: IMT measurement can only be active in image frozen status.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 71 -  Figure 7-7 ROI Box of Auto IMT  Figure 7-8 Measure Result of Auto IMT 7.3.2.2. Simpson’s (Method of Disks) Simpson is a tool in Cardiac presets used to determine LV diastolic and systolic volumes, and from that derive an ejection fraction.   To use the Simpson measurement: 1. Invoke the B-mode Measure function. 2. Select LV Simpson measurement on the touch screen and then select one measurement label from A4C Dias., A4C Sys., A2C Dias., and A2C Sys.. A single caliper will be displayed on the image field.   3. Move the caliper and press Set key to position the start point A at one end of left ventricular long axis.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 72 - 4. Draw the trace along the endocardium of left ventricular.   5. Press  Set  key  to  fix  the  end  point  B  and  complete  the  trace.  A  new  caliper  positions automatically at the apex (Point D) of the trace, connecting the Point C (Midpoint of point A and B). This distance between point C and D is the longest detected by the system, as shown below:    6. Move the caliper(Point D) to a new desired position if necessary.   7. Press Enter key to complete the measurement.   See  Chapter  6  Cardiac  in  advanced  manual  for  the  detailed  Simpson  measurement  and calculation results and equations.     7.3.2.3. PISA PISA  (Proximal  Isovelocity  Surface  Area)  is  a  tool  in  Cardiac  presets  used  for  quantitative analysis of regurgitation. It is only available in Color and Doppler mode.   To use PISA measurement in Color mode: 1. Adjust the color image until the aliasing appears. Invoke the Measure function. 2. Select PISA measurement on the touch screen and then select a measurement label from MR Rad, MR Als. Vel, AR Rad, AR Als. Vel, TR Rad, TR Als.Vel, PR Rad and PR Vls. Vel. A single caliper will be displayed on the image field. 3. Move the caliper and press Set key to position the caliper as the center of the semicircular. 4. Move the trackpad to adjust the radius length and orientation of the semicircular. 5. Press Enter key to fix the semicircular and then a dialog box of “Color Aliasing Velocity” pops up.   6. Select  “Top  aliasing  velocity”, “Bottom  aliasing  velocity”,  or  manually  enter  the  aliasing velocity from 0 cm/s to 500 cm/s.   7. Click OK and then measurement results are shown.   To use PISA measurement in Doppler mode: 1. Freeze a strip image. Invoke the Measure function. 2. Select PISA measurement on the touch screen and then select a measurement label from MR Trace, AR Trace, TR Trace and PR Trace.   3. Use the Doppler trace tool to get the measurement results. When  all  the  measurements  described  above  have  been  completed,  the  system  automatically calculates the Flow Rate, Flow and EROA(Effective Regurgitant Orifice Area). See Chapter 6 Cardiac  in  advanced  manual  for  the  detailed  PISA  measurement  and  calculation  results  and related equations.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 73 - 7.4 Patient Information Management 7.4.1. Storing Images The control panel has three store keys: Store 1, Store 2, and Alt Store. Each store key can be separately configurable to support any of the following functions:  Store a static image.  Store a clip.  Video print function For details about the configuration of store keys, please refer to Section 8.  Storing an image: 1. If a Store key is configured to store a static image it will always capture what is on the imaging  area  of  the  screen.  This  includes  live,  frozen,  or  Cine  images.  It  also  includes reports or other GUI screens and review. Storing a clip: 1. If a Store key is configured to store a clip image it will capture the moving image in live or auto cine play. If the image is frozen or a non-image screen is displayed then a static image is stored. 2. The store starts with the press of the button and continues for the configured length of the clip or until the clip store is interrupted. The following events can cause the clip store to stop:    A second press of the store button  Display of a GUI screen or dialog  Mode change  Reaching a pre-defined clip duration (refer to System Setup)  Image parameters change (depth and zoom  in B/C mode, M cursor, sweep speed, strip size and depth in M mode, angle, scale, baseline, invert, strip size and sweep speed in Strip Doppler mode.)  Cine play pause (press Freeze button, Set button or dragging left along the Swipe button) 7.4.2. Reviewing Images If Static or  Clip  images  have been stored  for  the  current  exam  then  they can  be  reviewed  by pressing the Review hard key on the console. NOTE: If nothing has been stored in the current exam then the  Review hard  key will invoke the Patient Database function (see the next section). Figure 7-9 shows an example touch screen for Review. The following table explains the action of each button in Review.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 74 -  Figure 7-9 Review Screen Button Description Page Change the display one page at a time.   Speed Change  the  playing  speed  of  the  highlighted  clip;  the  speed range is 12.5%-100%, in increments of 12.5%.The default speed is 50%. L.Margin Change the left margin of a highlighted clip R.Margin Change the right margin of a highlighted clip Frame Scroll the paused clip frame by frame Play Plays/Stops the highlighted clip. This can also be done with the Set key on the trackpad. Overlay Display Show or hide overlay icons on thumbnail images Remove Deleted Permanently remove all deleted images Delete Delete the highlighted image. A deleted image  is  marked by a deleted  icon.  To  cancel  the  delete,  press  this  button  while  the deleted image is highlighted  Change the displayed layout: 4*4, 2*2, 1*1, 2*3 Show Selected Toggle  between  showing  all  images  and  only  showing  the selected images Select Toggle the highlighted image in/out of the selected images Select All Select all the images in the current exam. Slideshow Play all the images as a slideshow. Export Export selected images to a removable media currently available. Plug in  a USB disk, select a image, press this  button and  the following confirmation dialog pop-up:
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 75 -  Select the export path from destination drop list menu and click OK to confirm. If there is only one removable media then this field is  a  read-only  field.  Static  images  export  as  .bmp  files.  Clip images export as .avi files. While export in progress the dialog is displayed as bellow. Cancel button is used to stop the export. The dialog disappears when export finishes.  If there is no USB plugged in, the message “The images cannot be exported because no USB flash drive is inserted” is displayed and no export work to do. Exit Exit review function Table 7-3 Review Touch Screen Controls Select Set:   By default, Review displays all Static and Clip images stored for the current exam in the order they were stored. Images can be placed into the Select Set by highlighting them and pressing the Select key on the touchscreen. Pressing the Show Selected button will then display the selected images in the order they were selected.   Reviewing Clips:   By default stored clips playback at half speed. You can adjust the playback rate Review will show Static and Clip images intermixed in the order they were stored.   Exiting Review: In review state, press Review or B hard key on the control panel, or press the Exit button on the touch screen to exit the review function. Measurement in Review State:   Measurements in review state are supported for the current exam (including restarted). Review a clip or static image for current active exam (please pause the clip when a desired image is got) and then press Measure hardkey on the console to start measurements.   Note:   1) PISA, Auto IMT and Auto Trace measurements are disabled in review state. 2) Measurements on stored panorama image are not available in review state.  7.4.3. Patient Database The Patient Database provides a list of recently performed studies. It can be accessed by pressing the Review hard key on the console when there is no active exam. Figure 7-10 shows an example
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 76 - screen shot from the Patient Database screen.  Figure 7-10 Patient Database screen The main part of the display shows a list of studies. Clicking on the header of a field will sort the list by that field. Fields can be displayed or hidden. Filter: The database filter field at the top-left of the screen provides a powerful tool to find the study of interest. It filters the list based on whatever text is typed in  this field. The filter  applies to all fields. For example, typing “Ab” into the filter will show exams that either have that text in either the patient name or in the Exam preset. By default the filter is set to blank, so the default list shows all exams.   Storage Size: There is a box displaying current disk usage. It contains text with the current usage and is  filled  with  a  solid color  to  the extent  that the  current capacity is  used.  The text shows “<current usage> of <total capacity>”. The units are “Mb” for values less than 1 Gb, and “Gb” for anything larger. The solid color fill is green when usage is less than 75% of capacity, yellow for usage between 75% and 95%, and red when disk usage is above 95%. Destination:  The  location  where  the  highlighted  exam(s)  shall  be  exported,  including  all configured network locations and any writable USB disk or DVD drive currently available. This includes:   DICOM server location(s) defined in the DICOM section of Connectivity. (See 9.2.2)   FTP server location(s) defined in the Network Store section of Connectivity. (See 9.2.3)   Any inserted USB drive or DVD drive. Remove DICOM Header: If this item is checked, static images will be exported as bmp format, and clips will be exported as avi format. Otherwise all the images will be exported as DICOM format. This interaction is only available when exporting to USB or DVD. Send/Burn: Pressing this will send/burn the highlighted exam(s) to the destination. This button is available when one or more exams are selected.  Figure 7-11 Send database
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 77 - Working with one study: A study is selected by clicking on it. When a study is selected thumbnail images from that study are shown on the right side of the screen.   A study can also be send to a DICOM server, saved to a USB device or DVD drive, or deleted. If the study has been copied to either a server or to a USB device or DVD drive a small disk icon appears next to it, indicating that it has been saved. CAUTION:   Studies stored on the system hard drive should be archived regularly. The system is not intended for  long term storage  of patient information. Confirm successful archiving before deleting a study from the hard drive. Working with multiple studies: Multiple studies can be selected by clicking the small box at the left of each listed study. Multiple studies can be stored to a DICOM server, saved to a USB device or DVD drive, or deleted. Only one study can be reviewed at a time. 7.4.4. Archiving Studies All Clips and Static images stored on the Acclarix AX8 are stored internally in DICOM format.   They may be converted to .bmp and .avi as described below. The AX8 supports many methods to export an exam for long-term storage:   The study can be sent to a DICOM server.    In this case the images are in DICOM (.dcm) format.   The study can be send to an FTP server.    In this case the images are in .bmp and .avi format.   The study can be saved to a USB or DVD.    In this case the images can be stored as either DICOM or .bmp/.avi format. Sending a study to a DICOM server: There are three ways to send a study to a DICOM server. See section 9 for details on configuring a DICOM server and choosing one of the following options 1. From the patient database screen, you can select a study or multi-studies and send it to a configured server. 2. You can configure the system to automatically send the study at the end of the exam. With this option the study is sent in the background while you continue to use the system for the next exam. 3. You can configure the system to automatically send each image and clip as it is stored. This has the advantage that the images and clips are available on the server sooner. Note that if you have it configured this way you can still delete stored images and clips on the system, but that will not delete them on the server if they have already been sent. Sending a study to FTP Server Please see section 9 for details on configuring a FTP server. Once an FTP server is configured the exam can be sent to the server from the patient database screen. Select the exam(s) from the list shown, and select the FTP server from the list of currently configured destinations. Saving a study to USB or DVD   If a USB flash drive or DVD is plugged in then studies can be archived to them in either DICOM or  .avi/bmp  format.    Plug  the  device  into  a  USB  port  prior  to  invoking  the  patient  database
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 78 - screen, select the exam(s) from the list  shown, and select the device from the list of currently configured destinations. 7.5 Needle Biopsy Guide NOTE: Use proper sterile technique at all times when performing a biopsy. Always follow these basic precautions: WARNING 1.  Disinfect the needle guide kit before the first use and after each subsequent use. 2.  Calibrate  the  needle  guide  kit  (see  section  7.5.3)  under  any  of  the  following conditions: a)  The first time that each bracket/transducer combination is used. b)  If the bracket or transducer head is dropped or struck, or has evidence of wear. c)  If  previous  use  has  shown  some  drift  of  the  needle  from  the  center  of  the guidelines. 3.  The  displayed  needle  guide  pathway  on  the  EDAN  video  monitor  is  intended  for reference during biopsy procedures. A variety of factors outside EDAN‟s control, such as changing tissue density, bending of the needle, off-axis pressure by the person holding  the  transducer,  etc.,  may  cause  deflection  of  a  needle  outside  of  the displayed video pathway even when the transducer, needle guide, and the system software are all performing as intended and within manufacturing specification. The specialist performing a biopsy procedure must be aware of potential external factors when performing an invasive procedure. 4.  Do not freeze the system when performing a biopsy. 5.  EDAN needle guides are designed and manufactured to attach firmly to designated transducers and should not require excessive force to assemble or disassemble. Do not use a needle guide that requires excessive force or manipulation to assemble or disassemble. 6.  A single-use sheath should be used on transducer when performing a biopsy. 7.5.1. Installing Needle Guide Bracket The  needle  guide  bracket  kit  consists  of  the  needle  guide  bracket  and  between  1  to  4 interchangeable guides for different gauge needles (see table 3-2 for details) Step  1:  Take  the  needle  guide  bracket  kit  out  of  the  package.    Check  to  make  sure  it  is  the correct kit for the transducer being used (see table 3-2) and inspect it carefully for any damage. See figure 7-12 for a typical needle guide bracket kit. WARNING The needle guide bracket kits are not disinfected or sterilized before delivery. Disinfect the  needle  guide  kit  before  the  first  use  and  after  each  subsequent  use  For  detailed operations, see 12.4.2 Needle Guide Bracket Cleaning and Sterilization.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 79 - 圆锥台矩形卡槽旋钮2螺钉旋钮1 Figure 7-12 Structure of Needle Guide Bracket Kit Step 2: Select the correct size guide for the gauge of needle to be used, and secure it to the needle guide holder by tightening knob 1. If the needle guide holder is loose tighten the screw with an allen wrench. Loosen knob 2 all the way prior to installing the bracket on the transducer. Step 3: The transducer has a small circular dimple on one side of the transducer, and a rectangular ridge on the other  side.    Place the needle  guide bracket over the  transducer such that  knob  2 tightens into the circular dimple and the ridge sits inside the corresponding slot on the bracket.   (see figures 7-13 through 7-16). 一字形凸起      凹槽 Figure 7-13 Convex slot of Probe                      Figure 7-14 Notch of Probe Knob 1 Rectangular slot Knob 2 Cone Screw Convex Notch
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 80 -          Figure 7-15 Attaching Bracket Kit to Probe            Figure 7-16 Turning Knob 2 Clockwise Step 4: Insert the needle into the rectangular hole, as shown in Figure 7-17. 矩形孔    Figure 7-17 Inserting Needle into Rectangular Hole Installing needle guide bracket of endocavity probe: Step  1:  Take  the  needle  guide  bracket  kit  out  of  the  package.    Check  to  make  sure  it  is  the correct kit for the endocavity transducer (see table 3-2) and inspect it carefully for any damage. See figure 7-18 for an endocavity needle guide bracket kit. WARNING The needle guide bracket kits are not disinfected or sterilized before delivery. Disinfect the  needle  guide  kit  before  the first use and after  each  subsequent  use.  For detailed operations, see 12.4.22 Needle Guide Bracket Cleaning and Sterilization. Rectangular Hole
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 81 -  Figure 7-18 Structure of Needle Guide Bracket Kit   Step 2: Attach the needle guide bracket kit to the transducer by placing the bracket over the head of the probe, ensure that the three fixing claws of the bracket catch in the notches at the probe tip, turn  the  lock  clockwise  as  the  arrow  direction  shown  in  Figure  7-21,  and  then  turn  knob  1 clockwise until it locks in place, as shown in Figure 7-22.  Figure 7-19 Notches at the Probe Tip  Figure 7-20 Turning the Lock Clockwise Lock Knob 1 Fixing claws Notches
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 82 -  Figure 7-21 Turning the Knob 1 Clockwise Step 3: Insert the needle into the round hole, as shown in Figure 7-22.  Figure 7-22 Insert the Needle into the Round Hole 7.5.2. Activating Needle Guide Function To enable the needle guide function: 1. In the B mode imaging, press Needle button on touch screen, A needle touch screen UI is displayed, Press Enable button to active the Needle Guide function. 2. Press Double Line button to switch double line and single line as the Needle guide Line graphics. 3. Some  needle  guide  brackets  support  multiple  (up  to  3)  angles.  If  the  current  transducer supports such a guide then the Angle button appears. Pressing it cycles through the available angles.   Figure 7-23 Needle Guide Touch Screen NOTE:   The distance between each dot of the needle guide line indicates 0.5 cm. Round hole
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 83 - 7.5.3. To Adjust the Needle Guide Line The needle guide bracket was calibrated during manufacture. However, certain events may cause the bracket alignment to shift, and new calibration required. WARNING 1.  Calibrate the needle guide under any of the following conditions: a)  The first time a needle guide is used with a given transducer. b)  Any time the needle guide or transducer has been dropped or struck against a hard surface. c)  After repeated use. 2.  Do not use the needle guide bracket if the needle does not track with the guide during calibration.  Figure 7-24 Needle Guide Calibration Touch Screen To calibrate the guide line 1. Assemble the needle guide bracket on the transducer, and use the transducer to image a water bath or needle guide phantom. 2. From the Needle function on the B-mode touch screen, press the Calibration button on the touch screen to display the Angle and Position paddle button.  Use the Position paddle to adjust the line horizontally until the origin aligns with the actual needle.    Use the Angle paddle to adjust the angle of the line until the entire line aligns with the actual needle.  Any changes will be saved as the default value automatically. 7.6 Needle Visualization Needle  Visualization  is  an  image  processing  technology  that  enhances  needle  visibility.  It  is available in B-mode for all linear transducers.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                              Features - 84 -  Figure 7-25 Needle Enhancement Visualization Needle Visualization is invoked by pressing the Needle button on B-mode touch screen, and then the Enable button in the Needle Visualization section of the Needle touch screen. The following parameters can be adjusted when it is active.  Enable: Enables or disables Needle Visualization.  L/R: press to display the angle line on left side or right side of the image field.  Medium/Steep: press to enhance the needle visibility for different needle inserted angles: 24° (Medium) and 40° (Steep). For best results use the angle line that is closest to perpendicular to the needle path.  Figure 7-26 Touch Screen for Needle Enhancement Visualization  Angle Line
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 85 - 8 Presets 8.1 Preset Organization The Acclarix AX8 Presets are divided into two levels: the Exam Presets and Application Presets.   Exam Preset: Each transducer has its own set of Exam Presets. Each Exam Preset contains:  The image parameters that optimize its transducer for an intended use.  The list of data fields displayed on the patient demographic page.  Pointers to Application Presets for Comments, Measurements, and BodyMarkers. Application Preset: The Application Presets are independent from the transducer. Application Presets  are  used  to  select  Comments,  BodyMarkers  and  Measurements.  Since  they  are independent from the transducer, several Exam Presets can use the same Application Preset. Figure 8-1  shows  an  example  of  how  Transducers, Exam Presets,  and Application  Presets are related.  For  simplicity,  this  shows  just  a  few  of  the  items  that  are  on  the  real  system.  In  this example  the  C5-2XQ  and  the  E8-4Q  both  have an  OB  Exam  Preset.  The C5-2XQ  OB  Exam Preset optimizes the C5-2XQ imaging for OB while the E8-4Q OB Exam Preset optimizes the E8-4Q for the same use. The parameters used for each transducer can be different, and changing one Exam preset does not change the other. However, both OB Exam Presets point to the same OB Measurement Preset as the Application Preset. This means  you can, for example, just configure OB measurements once, and both the C5-2XQ and E8-4Q will have updated OB measurements. This  example  just  shows  a  single  kind  of  Application  Preset.  The  Acclarix  AX8  actually  has separate Application presets for Comments ,Bodymarkers and Measurements. The same principle applies to each of these: application presets can be shared across exam presets and transducers.   Figure 8-1 Preset Organization
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 86 - 8.2 Selecting a Preset There are two ways to select a preset The Preset Function: The preset function is invoked by pressing the Preset button in the upper left corner of the console. See figure 8-2 for an example preset touch screen. The currently active transducer is shown at the top of the touch screen. Below it are all the presets associated with that transducer. Pressing any preset button selects that preset, recalling its parameters for system use.  If the system is configured with the optional Multi-Transducer Connector (MTC) this screen will show a top row of all connected transducers, allowing you to switch to any one of these and then select the preset to be used with that transducer.  Figure 8-2 Example Preset Screen The Patient Page: An exam preset can also be selected from the patient page, typically at the beginning of an exam. A drop-down list of presets associated with the current transducer is shown near the top of the screen. Select from that drop down to change the preset. This will also change the demographic fields that are displayed. 8.3 Storing a Preset There are two ways to store or modify an exam preset: Snapshot and Editing. Snapshot:    Select the exam preset you want to modify or use as the basis for a new preset.    Make any changes you want to system settings using the standard imaging UI.    Press the Preset hard key on the console.  Press the Save button on the touch screen to update the current exam preset with the new setting.  Press the Save As button to create a new preset and rename the preset. Editing:   You can directly view and change the settings for any preset using the Set-up function.  Press the Utilities hard key on the console.  Press the Set-up button on the touch screen to invoke setup. The screen shown in Figure 8-3 appears.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 87 -  Press System to edit overall system settings unrelated to presets. See section 9 for details.  Press Presets to do any of the following: o Edit the imaging parameters of an exam preset. o Configure the demographic fields that appear for that preset o Configure  which  Comment,  Measure,  or  Bodymarker  presets  are  associated  with each exam preset.  Press Comments to configure the pre-defined text of each comment preset.  Press  Bodymarker  to  configure  which  bodymarker  graphics  appear  with  each bodymarker preset.  Press Measure to configure measurement options of each measure preset.  Figure 8-3 Top level Touch Screen for the Setup Function 8.3.1. Presets Figure 8-4 shows an example of the presets page.  Figure 8-4 The Preset Definition Page
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 88 - 8.3.1.1. Editing Image Settings Figure 8-5 shows an example imaging set-up screen for imaging parameters. This example shows the screen for editing the B-mode settings for the C5-2XQ Abdomen exam preset.  Figure 8-5: Example Set-up Screen for Imaging Parameters Transducer and Exam Preset: There are drop-downs at the top of the screen for selecting the transducer and Exam Preset you want to edit. Each transducer has its own set of exam presets, and these two drop-downs let you select which exam preset, on which transducer, you want to edit. Image Modes: The next row of controls show tabs for each imaging mode. These let you select which  imaging  mode  you  want  to  edit  for  the  current  exam  preset.  The  example  shows  the B-mode tab selected. The image mode is divided into two sections. Parameters that have one value for each preset are shown on the left. Parameters that are associated with Image Types are shown on the right. Image Type: Each imaging mode has a set of image types. For example in B-mode the types are Detail,  General,  and  Penetration.  While  imaging,  the  image  type  lets  you  quickly  change  the aesthetic look of the image without changing settings like depth or invert status. The right side of the preset set-up screen shows a tab for each image look. The parameters shown below those tabs let you customize that image type for the current preset. 8.3.1.2. Configuring Patient Info Figure 8-6 shows an example of the Patient information configuration page. This page allows you to configure which demographic fields are shown for each exam preset.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 89 -  Figure 8-6 Example Patient Information Configuration Page To change the demographic fields shown with any preset:  Select the preset you want to configure. This can either be done at the Preset Definition level before invoking the Patient Information Configuration page, or it can be done directly on this page.  Click the check box on any field you want to display. 8.3.1.3. Selecting Default Measurement Package Measure:  Select  one  of  the  pre-defined  application  measurement  packages  to  be  the  default measure for the preset.   8.3.1.4. Selecting Default Comments Preset Comments: Select one of the pre-defined comments presets to be the default comments for the preset. 8.3.1.5. Selecting Default Bodymark Preset Bodymark: Select one of the pre-defined bodymark presets to be the default bodymark for the preset. 8.3.1.6. Selecting Default TI TI: Select one of the pre-defined TI presets to be the default preset. TI here includes: TIS (Soft Tissue Thermal Index), TIB( Bone Thermal Index) and TIC (Cranial Bone Thermal Index) 8.3.1.7. Selecting Default Output Power Output Power: Select one of the pre-defined acoustic output power.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 90 - 8.3.2. Comments Presets  Figure 8-7 Comments Preset Adding a new Comment: 1. Move the cursor to a blank field in the comments area and press Set. 2. Enter a new comment through keyboard and press Enter to finish editing. Editing an Exist Comment in Comment Preset: 1. Move the cursor to a comment in the comments area and press Set. 2. Edit the selected comment through keyboard and press Enter to finish editing. Deleting an Exist Comment in Comment Preset: 1. Move the cursor to a comment in the comments area and press Set. 2. Press Delete in touch screen to delete the selected comment. Creating a Group of Comments: 1. Move  the  cursor  over  the  intended  comment  then  press-and-hold  the  Enter  button.  The selected comment will pop-up. 2. Drag the  selected  comment  on  top  of  an  adjacent  comment  or  insert  it  into  an  adjacent group, then Press Set. The selected comments are joined as one group. Moving out a Comment from a Comment Group: 1. Move the cursor over the intended comment then press the Set button.   2. Press Delete button on touch screen. The selected comment is removed from the group. The rest of the comments in the group will be re-ordered. If there is only one comment in the group, then the group is disbanded. General Settings:  Auto-erase  comments  when  the  image  is  unfrozen:  If  this  option  is  enabled,  the comments will be erased when the image is unfrozen, otherwise the comments will only be erased on demand or at the start of a new exam.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 91 -  Suppress Comment on Unfreeze: If this option is enabled the Comment function goes away  when  the  system  is  unfrozen,  and  comes  back  again  with  the  next  freeze.   Manually exiting the function causes it to remain inactive until manually activated.  Figure 8-8 Comment Setting 8.3.3. Bodymarker Presets  Figure 8-9 Bodymarker Preset Touch Screen Enabling/Disabling a Bodymarker to a Preset: Press a bodymarker on touch screen to switch it as checked or not, the bodymarker is enabled or disabled to current preset.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 92 -  Figure 8-10 Bodymarker Preset   Deleting a Bodymarker from a Preset: If the selected bodymarker is from a user-defined which does not point to any exam preset, you can delete it with the Delete button. If the selected bodymarker is from a user-defined which points to an exam preset, or the selected bodymarker is from a default preset, the Delete button will be unavailable. 1. Select a bodymarker from pre-defined graphics area. 2. Press Delete on the touch screen. Moving a Bodymarker in Pre-Defined Graphics Area: 1. Move the cursor over a bodymarker in pre-defined graphics area and click Enter. A green box displays outside of the bodymarker. 2. Move the trackpad to a desired position and click Enter. To move it to another page, move the trackpad over page up or down icons. Set a Default Bodymarker for a Preset: Select a bodymarker from pre-defined graphics area and click Default, the bodymarker is set as a default bodymarker to select preset. The default bodymarker can not be deleted from a preset until another bodymarker is selected as a default one. General Settings:  Auto-erase  bodymarker  when  the  image  is  unfrozen:  If  this  option  is  enabled,  the bodymarkers will be erased when the image is unfrozen, otherwise the bodymarkers will only be erased on demand or at the start of a new exam.  Suppress Body Mark on Unfreeze: If this option is enabled the bodymarkers function goes  away  when  the  system  is  unfrozen,  and  comes  back  again  with  the  next  freeze.   Manually exiting the function causes it to remain inactive until manually activated. Pre-defined graphics area
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 93 -  Figure 8-11 Bodymaker Preset   8.3.4. Measure Presets The Preset UI of the Measure Presets set-up page is divided into the main screen display and the touchscreen display.     The main screen display shows the current measure presets. The list of presets is shown along the left side, and the grid of buttons to the right show the current content of the selected preset.     The touch screen display shows a library of all possible measurements that can be added to or deleted from any measure preset.    The column along the left shows different clinical application types  and  the  grid  of  buttons  to  the  right  show  all  the  supported  content  for  that  application.   You can add a measurement from any application to any preset.        Figure 8-12 Measure Preset  Pre-Defined Measurement Area
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 94 - Adding a new Measure Preset: A new measure preset can be added by copying an existing preset. The new measure preset contains the same measurements as the copied preset.  Select a preset from the list along the left, and then press the Copy As button.     Enabling/Disabling a measurement in a Preset Select the preset  you want to configure from the list along the left side of the main screen.   The  current  contents  show  in  the  grid  along  the  right  of  the  main  screen.    Press  a measurement from the measurement library on the touch screen to toggle it in or out of the selected preset.  Figure 8-13 Measure Preset touch screen Editing an Existing Measurement in a Measure Preset 1. Select the preset you want to customize from the list along the left of the main screen. 2. Select the measurement you want to customize from the grid along the right of the main screen. 3. Press the Option button on the touch screen to customize the selected measurement. 4. If the selected measurement can be performed by different tools (Caliper, Trace, etc.) select the tool you want from the tabs provided, and check the default box.   5. Check the boxes for the results that you want to derive from this measurement.     For a Generic measurement, press Default button on the touch screen to configure it as a default measurement for the current selected image mode in a preset.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 95 -  Figure 8-14 Measurement Option 8.3.4.1. General Settings  Figure 8-15 Measurement General Settings  Report Logo The system supports a user-defined report header by importing a logo and setting the logo‟s position and size.   Report logo settings are as follows:    Import: import a bmp format logo from USB disk.  Delete: delete the imported logo.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 96 -  Logo Position: set the logo position to locating at the left or right side of report header.  Logo Size: set the logo size to Small or Large.    OB Settings The system supports the following OB measurement settings:  Fetal Age: select the author / measurement options for Fetal age calculation.    Fetal Growth: select the author / measurement options for Fetal Growth calculation.    Estimated Fetus Weight: select an equation for the Estimated Fetus Weight calculation.    Growth Curves Display: set single or quad curves as the default display format for the Growth Curves.  Default  GA  display:  set  the  CUA  (Composite  Ultrasound  Age)  or  AUA  (Average Ultrasound Age) as a default result displayed in the worksheet.  Cardiac Settings The system supports following Cardiac measurement settings:  HR  Cycle:  set  the  number  of  heartbeats  that  are  assumed  in  the  HR  or  FHR measurement calculation.    Range: 1-8.  EDV/ESV: set the author of the formula that is used in EDV/ESV calculation. There are three options: Teichholz, Gibson and Cube.  Suppress Measurement on Unfreeze If this option is enabled the Measure function goes away when the system is unfrozen, and comes back again with the next freeze.    Manually exiting the function causes it to remain inactive until manually activated.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 97 - 9 Utilities The  Utilities  hard  key  provides  access  to  setup,  screen  adjust,  connectivity  and  maintenance software.    Each of these is described in separate sections below. 9.1 Set-up The Set-up function is divided into several sections, accessed by separate touch screen buttons  System Set-up: Described in the remainder of this section.  Presets: Supports edits of preset organization and image preset.    See section 8.3.1  Comments: Supports edits of comment presets. See section 8.3.2  Body Mark: Supports edits of Body Mark presets. See section 8.3.3  Measure: Support edits of Measure presets.    See section 8.3.4 System setup is used to configure parameters that are unrelated to presets. Generally, there is one value of each system setup parameter that is shared across all presets. Entering/exiting Setup: Press Utilities on the control panel, the Setup touch screen button, and then the System touch screen button to access system setup. Reloading Factory Default Set-up: 1. Press “Restore Factory Settings” on the Set-up screen, you are prompted Restore System Setting to factory default? This will erase all current system setting.   2. Press Yes to restore to factory default, and press Cancel to cancel. 9.1.1. General Set-up  Figure 9-1General Set-up Screen
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 98 - Item Options Description Hospital Input freely Set hospital name displayed on the top left of the screen and diagnosis report Language Chinese,  English,  German, French, Italian, Spanish, Russian, Turkish Set the system language(new language is effective after reboot the system) Keyboard U.S International Set the keyboard layout Date Format YYYY/MM/DD/   MM/DD/YYYY DD/MM/YYYY Set the date format Note: If the license is invalidated because of  changing  the  system  date,  please contact EDAN serviceman. Date Set freely Set the system date Time Format 12-AM/PM, 24-hh:mm:ss Set the time format Time Set freely Set the system time, format: H/M/S Auto-Freeze& Waiting Time √/× 1-999 min Set whether to turn on auto-freeze function by default or not; and set the waiting time to perform auto-freeze Sleep& Waiting time √/× 1-999 min Set  whether  to  turn  on  system  sleep function  by  default  or  not;  and  set  the system waiting time to enter sleep mode Button F1/F2 Color PDI DPDI CW Needle 3D 4D Define  the  F1/F2  keys,  select  one  of  the pull-down  options.  The  options  that  are available will depend on what functions are enabled on the system. Scale Unit cm/s, kHz Set the velocity units of spectrum scale Video Mode PAL/NTSC Set  the  output  video  mode,  the  selected video mode should be the same as that of the  s-video  printer,  otherwise  the  printer can not work. FootSwitch Freeze/Store1/Store2/Alt Store Set a function for the left key or right key of   Footswitch.  If  the  physical  footswitch  only has  one  button  then  the  Left  button selection applies. Image Size Large/Small Set the image area size. Hide non intra-operative indicator for L17-7SQ  √/×  Set  whether  to  hide  the  hint  of  “non intra-operative”  while  using  transducer L17-7SQ. Table 9-1 General Set-up Description
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 99 - 9.1.2. Patient Set-up  Figure 9-2 Patient Set-up Screen Item Options Description Weight Unit kg, lbs Set the weight unit of patients Height Unit cm, ft/in Set the height unit of patients Title  of  custom field 1/2 / You can define two additional fields for data entry on the Patient Information screen Time  Limit  for Exam Restart 0/12h/24h/48h/72h You can define the time limit for restarting exam. Only exams within the time limit can be restart. If 0 is selected, no exam can be restart. Patient  Name Display Option One field/Two fields You can define the patient name display format. The  patient  information  page  displays  “Patient Name”  for  one  field  and  displays  “Last  Name” and “First Name” for two fields.   Table 9-2 Patient Set-up Description
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 100 - 9.1.3. Store/Print Set-up  Figure 9-3 Store/Print Set-up Screen Item Options Description Store1/Store2/ Alt Store Store Image Store static  image  to  internal  hard disk Store Clip Store clip to internal hard disk SONY-UP-D25MD, SONY-UP-D898MD_X898MD Connect  the  corresponding  printer to print the current image. Compression Lossless/Low/Medium/High Set the image compression type Duration ≤10 seconds Set the clip capture duration Target Frame Rate 5/10/15/20/25 Hz Set the clip capture frame rate Report Printer Printer Model Set  the  default  report  printer  from the supported printers list (System supports generic HP printer models, select  Generic  HP  printer  if  you cannot find the printer models in the drop-down list) Table 9-3 Store/Print Set-up Description
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 101 - 9.1.4. Miscellaneous Set-up  Figure 9-4 Miscellaneous Set-up Screen Item Options Description Default review layout 4*4, 2*2, 1*1, 2*3 Set the default review layout Overlay √/× Show  or  hide  overlay  icons  on  thumbnail images Gain √/× Optimize Gain or not when press auto button in PW mode. DR √/× Optimize  Dynamic  Range  or  not  when  press auto button in PW mode. Scale /Baseline √/× Optimize scale/Baseline or not when press auto button in PW mode. Table 9-4 Miscellaneous Set-up Information
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 102 - 9.1.5. User Set-up  Figure 9-5 User Set-up Screen Item Options Description Enable Password Protection √/× Show or hide the login dialog when booting up the system. User List √/× Show  all  users  displaying  user name  and  user  type.  Click  the dialog before the user to select this user for edit operation. Table 9-5 User Set-up Information User type includes Administrator and Operator.   • Administrator  users  have  authority  to  enable/disable  password  protection,  add/delete/edit users and can view all exams in the patient database. There are two pre-defined administrator users called Admin1 and Admin2.   • Operator users  can only  edit  their  own  user information, change  their  own  password, and view exams that they  created.  There‟s  one  pre-defined operator user called Emergency for Emergency login without entering password described below. There‟s a couple of buttons for different user edit operation. It includes Add User, Delete User, Edit User, and Change Password. Add User: Click it to add user in the user list. Change Password: Click it to change the password of the users. Delete User: Click it to delete the highlighted user in user list. Edit User: Click it to modify the user information.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 103 - 9.2 Connectivity The Connectivity screens  support  configuration of network  access and services.    It is divided into 3 tabs:  TCP/IP: Configures access to the network from the AX8.  DICOM: Configures network DICOM services.  Network Store: Configures non-DICOM storage to a network via FTP. 9.2.1. TCP/IP This screen configures access to a network. Access can be either wired or wireless*.      Figure 9-6 TCP/IP settings Wired Settings: The following settings apply when connecting an AX8 to a wired network via the network port on the side of the system. Get Local IP: Either Manually entered (Static) or DHCP.    The correct setting will depend on  how  your  DICOM  network  is  administrated.    If  you  are  unsure,  try  DHCP  first.   Generally a Static IP is only needed if the target PACS system requires it.    If it is needed, contact your network IT manager. IP  Address,  Subnet  Mask,  Gateway,  DNS:  These  settings  determine  how  the  AX8 communicates with your network.    If you chose DHCP then  your network router will set these  automatically.    If  you  chose  Manually  Entered  IP  then  contact  your  network  IT manager for the appropriate settings. Wireless Settings:* (note: availability of this feature will depend on local regulatory status) Enable  Wireless:  This  will  turn  the  wireless  radio  on/off.    No  wireless  networks  are
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 104 - displayed when this is disabled.     Networks: When wireless is enabled the system will search for wireless networks that are available and display them in this list.    Networks that have been previously accessed will show up with a checkmark (“√”).    If there is currently an active network it will show up with  an  asterisk  („*‟).    If  the  network  requires  a  password  it  shows  up  with  a  lock  icon.     Each displayed network shows the strength of the wireless signal. Refresh: Clicking on this button updates the list of displayed networks.    The following controls require that a wireless network has been selected from the network list.    Click on the name of a network to select it. Join:  This  joins  the  selected  network.    If  a  password  is  needed  a  dialog  box  will  be displayed to enter it. Forget Network: This forgets any password of the selected network that had been entered, but the network is still shown in the list. Advanced Settings:  This  brings up  a  dialog box  with  additional settings  for the  selected wireless network.    The dialog box includes the following:  Password: This is useful when the network password has changed.  Method: Either Manually entered (Static) or DHCP.    Generally wireless networks use DHCP, and a Static address is only needed if required by the target PACS system.  IP Address, Subnet Mask, Gateway, DNS: If you chose DHCP then your network router will set these automatically.    If you chose Manually Entered IP then contact your network IT manager for the appropriate settings. 9.2.2. DICOM This screen configures DICOM services, including Store and Modality Worklist. Local Settings:   These settings determine how the AX8 communicates with other DICOM devices.    These are required for both Store and for Modality Worklist.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 105 -  Figure 9-7 DICOM Settings Local  AE  Title:  Any  16  characters  that  uniquely  identify  this  system  on  your  DICOM network.    The  default  “EDAN_STR_SCU”  will  work  unless  you  have  multiple  Edan systems on your network.     Time Out, Transfer Syntax:    These determine how the DICOM messages are configured.   The AX8 default settings are common, and you will typically not have to change them.    If Ping works, but Verify does not, you may need to review the conformance statement of your server to determine the proper settings for these. DICOM  Store:  Determines  when  DICOM  images  and  clips  are  sent  to  the  server.    The options are:  Transfer on demand: Images and clips are only sent when you manually send them.   When there is no exam in progress the Review hard key will show a list of all exams on the  hard  drive.  Select  an exam and  press the  Send  button  at  the  bottom of  the screen.    Auto-Transfer  when  Store:  Each  image  or  clip  is  sent  automatically  when  it  is captured.    This has the advantage that images and clips are immediately available for review at the server while the exam is in progress.    It has the disadvantage that, due to the nature of DICOM, images that are already sent can not be deleted from the server from the AX8.  Auto Transfer when End Exam:    Images and clips are sent automatically when the exam is ended. This is the most common setting. Server Settings:   Server list: The main interaction in this section is a list of configured servers.    It starts off as an empty list, and grows as servers are added.    Most sites will only use one server, but if the system is moved between locations then multiple servers may be entered. Clicking in any field in this list will select that server.    Depending on the exact level of software a second click may be needed to edit that field.    The fields are:
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 106 -  Default:  A  check  mark  on  the  left  side  indicates  which  server  is  the  default  for Auto-Transfer.  Name:  The  name  of  the  server  that  appears  in  the  drop-down  list  of  the  patient database  AE Title, IP  Address, Port, Max  PDU:  These  are  the  settings  of  the  destination DICOM server;  it‟s  how the AX8 finds the DICOM  server on  your network.    The AE  title  and  IP  Address  are  unique  to  your  network;  contact  your  network  IT manager for these settings.    The most common setting of Remote Port for DICOM servers is 104, although your server may be different.  Testing  the  server:  There  are  two  tests  to  ensure  that  the  server  information  is  entered correctly.    Click on any field for a given server to make that server selected, then:    Ping: A successful Ping means that the AX8 can communicate with the server at a low-level; basically that the two computers „see‟ each other.    As a security measure, some servers on the Internet may be configured to not respond to a Ping even if the connection is successful.  Verify: A successful Verify means that the AX8 can communicate with the server at a DICOM level; basically that the DICOM on both computers understand each other.   A successful Verify will typically mean that your DICOM configuration is correct. Other controls:  Add:    Adds another line in the list of servers.  Delete:    Deletes the selected server.  Default: Makes the selected server the default for Auto-transfer.  Modality Worklist: Modality Worklist will query the configured server for a list of all ultrasound exams scheduled within the configured date range. Enable: Turns Modality Worklist on/off. Update Worklist:    Determines when the worklist is updated.    Choices are:  On  Demand:    Updates  only  when  the  Update  button  is  pressed  from  the  patient page.  Every 15 Minutes: Updates on power-up and then regularly every 15 minutes.  Every hour: Updates on power-up and then regularly every hour.  On Start of Exam: Updates when the patient page is displayed at the start of each exam. Comment Source:    Modality Worklist can fill the ultrasound comments section from the server.    Depending on    Patient Comments: Obtains comments from DICOM tag (0x0010,0x4000);  Scheduled  Procedure  Description:  Obtains  comments  from  DICOM  tag (0x0040,0x0007)  Comments  on  Schedule  Procedure:  Obtains  comments  from  DICOM  tag (0x0040,0x0400) Worklist Server: Selects the server to query for the worklist.    Options include any server configured in Server Settings (see previous section). Date Range: The range of dates for the worklist query.    Note, the option of “Yesterday and Tomorrow” includes today‟s exams.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 107 - 9.2.3. Network Store This screen configures the system to connect to an FTP server on the local network. This allows the system to store .avi and .bmp files to a local server without using DICOM.    To use this, a PC or server needs to be configured as an FTP server on the local network.    The mechanism to do this depends on the operating system used.    Please contact your local IT support for assistance.   Once configured, studies can be sent to the FTP server from the Exam Database shown in Review.   Studies are sent to the configured folder, where each study appears in a separate folder with the name of the patient and the study date.    Figure 9-8 Network Store Settings FTP  Server list:  The  main  interaction  in  this  section  is  a  list  of  configured  servers  and directories.    It starts off as an empty list, and grows as servers are added.    Most sites will only use  one,  but  if  the  system is  moved  between  locations then  multiple entries may be useful. Clicking in any field in this list will select that server.      Name:  The  name  of  the  server  that  appears  in  the  drop-down  list  of  the  patient database  IP  Address, Shared  Directory:  These  are  the  address  of  the  FTP  server  and  the directory used on that server.  Testing  the  server:  There  are  two  tests  to  ensure  that  the  server  information  is  entered correctly.    Click on any field for a given server to make that server selected, then:    Ping: A successful Ping means that the AX8 can communicate with the FTP server at a  low-level;  basically  that  the  two  computers  „see‟  each  other.    As  a  security measure, some servers on the Internet may be configured to not respond to a Ping even if the connection is successful.  Connect: A successful Connect means that the AX8 can communicate with the FTP server, and FTP transfers will work. If this fails but Ping works it could mean: o The destination computer is not correctly configured as an FTP server. o The password on the server has changed. Use Edit to re-enter the password. o The Shared Directory does not exist.    Use Edit to confirm the directory. Add: Adds another line in the list of servers.    This automatically invokes Edit.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 108 - Edit: Available when an FTP Server has been selected.    This invokes a dialog that edits the IP address, name, directory, and password. Delete: Available when an FTP Server has been selected. This deletes the selected server. 9.3 Maintenance The Maintenance screens provide access to controls that are not typically needed during normal operation of the system. 9.3.1. License   The License page displays which features are currently licensed for use on the system.    At the top  of  the  screen  is  displayed  the  current  license  key.    Below  that  is  a  list  of  all  licensable features,  along  with  its  current  status.    The  system  supports  temporary  trial  licenses.    If  a feature is so licensed the expiration date is also shown.  Press Import button to import a license file from a USB external storage device.  Press Revert button to restore the previous license if the imported license is not correct. Please contact your distributor or Edan service engineer to obtain a new license key.  Figure 9-9 System License Screen 9.3.2. Version   The Version screen shows the current version of software, firmware, and selected hardware for the system.    This information is only needed if requested by an Edan service engineer.    This page  also  includes  an  Upgrade,  Import  Config  and  Show  Config  button.    Any  released
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 109 - upgrade will include instructions regarding use of these buttons.   9.3.3. Demo   A Demo exam is a special exam with a patient name of “Demo”.    It allows you to collect any set of images from multiple exams.    To create a Demo exam: 1. Export the images you want as DICOM format from the system to a USB external storage device. (See section 7.4.3) 2. Using a computer, create a folder named “Demo” under the folder “Edan_Image” on the USB external storage device. 3. Move the exported images to the “Demo” folder.   4. Connect the USB external storage device to a system, Press Utilities hard key, select Maintenance Demo. 5. Press Import button, and select the USB device. 6. Press OK to start copying the images from the USB device to the Demo exam. If “Show Demo exam in Exam Database” is enabled the imported Demo exam will show up in the Exam Database list.  Figure 9-10 System Demo Screen 9.3.4. Export/Import The system supports export/import of user-configured presets and settings to an external storage device.    This same UI supports the export of system log files. To export user data:
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 110 - 1. Press Utilities hard key, select Maintenance Export/ImportExport. 2. Select an external USB storage device. 3. Check the items to be exported: Image Presets, Comment Presets, Body Mark Presets, User  Settings,  System  Log,  Findings.  The  Preset  items  can  be  expanded  to  select  a specific preset.  Each  type  of  preset  can  be  expanded  using  the  arrow  graphics  on  the  left  side.  This allows you to export/import any subset of any type of preset. 4. Press Export button to begin the export process. The User Settings includes all system setting data described in section 9.1. It also contains the links  between  layers  of  presets.  For  example,  suppose  a  system  has  a  user-created  comment preset that was named „my_ABD_comments‟ that was linked to the Abdomen exam preset on that system.    Exporting those presets and then importing them to a different system would not link „my_ABD_comments‟ to the Abdomen exam preset on the new system, unless the User Settings were also exported/imported.  Figure 9-11 System Export Screen To import user data: 1. Press Utilities hard key, select Maintenance Export/ImportImport. 2. Select an external USB storage device, all the presets on it will be loaded and displayed. The Presets must have been previously exported from a system.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 111 - 3. Check the items to be imported: Image Presets, Comment Presets, Body Mark Presets, User Settings, Findings. The Preset items can be expanded to select a specific preset. 4. Press Import button to begin the import process. Note: Import of System Logs is not supported.  Figure 9-12 System Import Screen 9.4 Screen Adjust The brightness and contrast of the monitor and touch screen in the system can be customized. Press Utilities hard key, select Screen Adjust to customize the brightness and contrast:  Main Screen Brightness: adjusts the brightness value of the monitor.  Main Screen Contrast: adjusts the contrast value of the monitor.  Touch Screen Brightness: adjusts the brightness value of the touch screen.  Touch Pad Brightness: adjusts the brightness value of the touch pad. Press Restore Default  button to  restore the  factory settings for all the  brightness and  contrast values.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      System Setup - 112 -  Figure 9-13 Screen Adjust Screen
Acclarix AX8 Diagnostic Ultrasound System User Manual                                            In Between Exams - 113 - 10  In Between Exams 10.1 Unpacking Visually  examine  the  package  prior  to  unpacking.  If  any  signs  of  mishandling  or  damage  are detected, contact the carrier to claim for damage. After unpacking the device, you should follow the packing list to  check the product  carefully and to  make sure that no damage has occurred during  transportation.  Then,  install  the  device  according  to  the  installation  requirements  and methods. WARNING 1.  Do not use the device if it is found to be damaged or defective. 2.  Do not drop or collide with the transducer. Otherwise you shall give up using it. 3.  Make  sure  the  AC  power  supply  complies  with  the  following  specifications: 100V-240V~, 50Hz/60Hz. 10.2 Transport The system is designed to be portable and easily transported. Power off the system and secure all accessories before moving it to another location. CAUTION 1  Switch off the ultrasound system. Unplug the AC adapter from the power source and secure the power cable. 2  Remove the transducer and place them in a safe place. 3  Disconnect and secure the connecting cable. 4  Connect optional system accessories. 5  Secure the system and complete the system setup, and then perform all the daily checking before using it. To prepare the system for shipment over long distances or rough terrain, repack the system in the factory packing To prepare the system for transport over distances: load the system into a vehicle using a lift gate. To  prevent  lateral  movement  of  the  system,  secure  the  system  with  cargo  straps.  To  prevent sudden jarring of the system during transport, provide anti-shock cushions beneath the system. It  is  suitable  for  transportation  by  air,  railway,  highway  and  ship.  Protect  the  system  from inversion, collision, and splashing with rain and snow. 10.3 Storage  Do not place the device near the ground, walls or the roof.    Keep good indoor ventilation. Avoid strong and direct sunlight, and erosive gas.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                            Order List - 114 - 11  Order List The following accessories are recommended for use on the Acclarix AX8.   WARNING Only  accessories  supplied  or  recommended  by  EDAN  can  be  used,  the  battery  and transducers of EDAN can be only used on EDAN‟s systems. Otherwise, the performance and electric shock protection cannot be guaranteed. If electrical or mechanical equipment from  other  companies  need  to  be  connected  to  the  device,  please  contact  EDAN  or authorized representatives before connection. Part Name   Part Number Transducer C5-2XQ 02.01.211389 Transducer L10-4Q 02.01.211855 Transducer L17-7HQ 02.01.211856 Transducer E8-4Q 02.01.211392 Transducer P5-1XQ 02.01.211391 Transducer L17-7SQ 02.01.211857 Transducer C5-2Q 02.01.212622 Transducer MC8-4Q 02.01.212625 Transducer L12-5Q 02.01.212623 Transducer C5-2MQ* 02.01.211474 BGK-C5-2 Needle Guide Bracket Kit 02.01.211006 L40UB Needle Guide Bracket Kit 02.01.210407 CR10UA Needle Guide Bracket Kit 02.01.102963 R15UB Needle Guide Bracket Kit 02.01.210618 BGK-P5-1X Needle Guide Bracket Kit 02.01.210985 Main Lithium-Ion battery 21.21.064246 UPS Battery 01.21.064353 Adapter 21.21.064243 Power Cable, European Standard 01.13.036638 Power Cable, American Standard 21.13.036384 Ultrasound Gel 01.57.78008
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                            Order List - 115 - USB Disk 11.18.052245 Trolley MT-807 83.63.560199 Footswitch(single button) 01.10.000910 Fottswitch(doule button) 01.10.027323 Suitcase 01.56.466102 Multi-Transducer Connector(MTC) 83.63.560232  NOTE: The part name may vary depending on context, but the part number is constant.
Acclarix AX8 Diagnostic Ultrasound System User Manual                  Troubleshooting and Maintenance - 116 - 12  Troubleshooting and Maintenance In  order  to  ensure  proper  system  operation  and  function,  a  maintenance  and  inspection  plan should be established to periodically check the safety of the system. If any system malfunction is experienced, contact EDAN or authorized representatives. 12.1 Daily Checklist Check before the system is switched on, if any system malfunction is experienced, eliminate the malfunction before use, or contact EDAN or authorized representatives for service if needed.  Visually inspect all the transducers. Do not use any damaged transducer.  Visually inspect all the transducer assembly cables and associated connectors.  Visually inspect all the cords. Do not turn on the power if a cord is frayed or split, or shows signs of wear.  Verify that the controls are clean and free from gel or contaminants.   Check after the system is switched on:  Visually check the on-screen display and lighting. Verify that the monitor displays the current date and time and there isn‟t any error message.  Verify that the transducer identification and indicated frequency on the screen are correct for the activated transducer.  Ensure that there isn‟t obvious abnormal noise, discontinuous image or dark area.  Ensure that it isn‟t smelly or too hot.    Ensure that the ultrasound window isn‟t too hot, checking with your hand.  Verify that the buttons on the keyboard are good to operate.  Ensure that there isn‟t obvious abnormal noise from the loudspeakers.  Ensure that there isn‟t obvious abnormal noise from the air fan. 12.2 Troubleshooting If  any  persistent  system  malfunction  is  experienced,  e.g.  an  onscreen  error  message,  blank imaging screen, absent menus, please refer to the following table below. If the failure cannot be eliminated, please contact EDAN or authorized representatives. Item   Problem Solution 1   When the power switch is on, there isn‟t any image displayed. 1.  Check power supply. 2.  Check wires and plugs.
Acclarix AX8 Diagnostic Ultrasound System User Manual                  Troubleshooting and Maintenance - 117 - 12.3 Maintenance The system should be maintained regularly, at least annually, by a qualified technician who has adequate  training,  knowledge  and  experience.  That  person  should  be  familiar  with  the  AX8 Service Manual (P/N 01.54.002483), available from your Edan representative. 12.4 Cleaning the System Use only the EDAN-approved substances and methods listed in this chapter to clean the system. The warranty does not cover damage caused by using unapproved substances or methods. Edan Instruments has validated the cleaning and disinfection instructions included in this User Manual. It is the responsibility of the healthcare professional to ensure that the instructions are followed so as to ensure adequate cleaning and disinfection. General Points: Keep  your  monitor,  cables  and  accessories  free  of  dust  and  dirt.  To  prevent  the  device  from damage, please follow the procedure:    Use only recommended cleaning substances and disinfectants listed in this manual. Others may  cause  damage  (not  covered  by  warranty),  reduce  product  lifetime  or  cause  safety hazards.  Always dilute according to the manufacturer‟s instructions.  Unless otherwise specified, do not immerse any part of the equipment or any accessories in liquid.  Do not pour liquid onto the system.  Do not allow liquid to enter the case.  Never use abrasive material (such as steel wool or silver polish).  Inspect the monitor and reusable accessories after they are cleaned and disinfected. 2   Strip-shape or snowflake-shape disturbance occurs on the display screen. 1.  Inspect the power supply. 2.  Check whether it is disturbed by the ignition action of any other device. 3.  Check the disturbance of electric or magnetic field in the surrounding environment. 4.  Check whether the plug and socket of power supply and transducer are properly connected.   3   Image is not displayed clearly on the screen. 1.  Adjust overall gain (Gain). 2.  Adjust eight TGC slide controls. 3.  Adjust focus (the number and the position). 4   Image window is dark. 1.  Adjust the brightness and slide the TGC controls on the touch screen. 2.  Check whether the transducer is connected correctly. 5   The button is no response 1.  Check the control panel to see whether the button is blocked and press it several times to release it. 2.  Clean the button
Acclarix AX8 Diagnostic Ultrasound System User Manual                  Troubleshooting and Maintenance - 118 - WARNING The console (including touch screen, track pad and hard keys) of system is splash resistant (IPX 1) but not  waterproof.  Do  not  immerse  or  expose  to  extended  moisture.  Splash  resistance  does  not extend to transducer connectors. Keep connectors dry. For cleaning agents and disinfectants, refer to section 6.3 Transducer Cleaning and Disinfection 12.4.1.Cleaning the System Surface To clean the system surface: 1. Switch off the system and unplug it. 2. Wipe the entire exterior surface, including the screen, of the equipment using a soft cloth dampened with the cleaning solution thoroughly until no visible contaminants remain. 3. Wipe off the cleaning solution with a fresh cloth or towel, dampened with tap water after cleaning until no visible cleaning agent remains. 4. Dry the system in a ventilated and cool place. NOTE:   1.  Make sure the system is free of gel and any other visible residue.   2.  Use a soft  dry cloth without chemicals for cleaning. The monitor surface is easily scratched. 12.4.2.Needle Guide Bracket Cleaning and Sterilization NOTE:   1.  Use proper sterilization technique at all times when performing a biopsy. 2.  Ensure that protective gloves are worn. WARNING The  needle  guide  bracket  kits  are  not  disinfected  or  sterilized  before  delivery.  The operators must clean and sterilize the needle guide bracket kits before and after each use. Cleaning Clean the needle guide assembly using a mild detergent prior to sterilizing. Sterilization Sterilize the Needle Guide Assembly by autoclaving for at least four minutes at 132oC
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      Specifications - 119 - 13  Specifications 13.1 Electrical Safety Classifications According to the type of protection against electric shock Internally powered equipment,   Class I equipment According to the degree of protection against electric shock Type BF According to the degree of protection against harmful ingress of liquid Whole device: IPX0 Console (including touch screen, track pad and hard keys): IPX1 Transducer (not including the transducer connector): IPX7;   Footswitch: IP68 According to the degree of safety of application in the presence of a flammable gas Equipment not suitable for use in the presence of a flammable gas According to the mode of operation Continuous operation According to the grade of EMC CISPR 11 Group 1, Class A Standards Compliance EN 60601-1:2006   idt IEC 60601-1: 2005 +A1:2012 EN 60601-1-2: 2007 idt IEC 60601-1-2: 2007 IEC/EN 60601-2-37 NEMAUD 2 NEMAUD 3 Imaging Compliance The AX8 is a portable diagnostic ultrasound system.    Compilant with : Essential Performance : EN60601-2-37. 2D Imaging: GB10152-2009. Color Flow Imaging: YY 0767-2009. Spectral Doppler Imaging: YY 0767-2009. 13.2 Power Supply Operating Voltage 100 -240 V~ Operating Frequency 50 Hz/60 Hz AC Input Current 2.5A-1.5A DC Input Current 19V 7.8A MAX Lithium battery   Capacity 6150 mAh Voltage 14.4V Average working time 1 Hours Charging time 3 hours with system power-on; 2.5 hours with system power-off Cycle life 300
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      Specifications - 120 - 13.3 Machine Specifications Main unit dimensions 38.8cm(W)x 40.7cm(D)x 7.7cm(H) Net weight 9.25kg  including  battery  (not  including  power  adaptor  or transducer) 13.4 Display Specifications Display   TFT-LCD Diagonal Size   15.6 inches View Angle   Right 80, Left 80, Up 80, Down 80 (min) Pixel Number 1920x1080 White Luminance 300 cd/m2(typ) Contrast Ratio 700 (typ) 13.5 General Technical Specifications Display Modes B Mode: B, 2B,4B   C Mode: B/C(Single, Dual, Quad); B+B/C(Dual Live); B/C/PW (triplex mode) PDI/DPDI Mode: B/PDI(DPDI) (Single, Dual, Quad); B+B/PDI(DPDI) (Dual Live); B/PDI(DPDI)/PW (triplex mode) PW Mode: B/PW;(Update) B/PW; (duplex, simultaneous) B/C/PW(Update) B/C/PW, B/PDI(DPDI)/PW; (triplex mode) CW Mode: B/CW; B/C/CW, B/PDI(DPDI)/CW; M Mode: B/M (Display layout: Up/down, Left/right,1:1) Image Gray Scale 256 levels Image Magnification In area Real time: x1.0,  x1.3, x1.5, x1.8, x2.0. Frozen: x1.0, x1.3, x1.5, x1.8, x2.0. Hard Drive Storage 500 GB Depth Adjustment Adjustable in real time in all modes
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                      Specifications - 121 - Image Conversion Up/Down flip, Left/Right flip Language Display Chinese, English, German, French, Italian, Spanish, Russian, Turkish Focus Number Max. 3 Report Packages Abdomen,  Obstetrics,  Small  Parts,  Vascular,  Cardiac,  Gynecology, Urology, Pediatric B/Color mode   Generic measurement Distance, Circ/Area, Angle, Volume, Stenosis M mode   generic measurement Distance, Slope, Time, HR D mode   generic measurement Velocity, PG(Pressure Gradient), Heart Rate, Time, Acceleration, PI, RI, Auto trace Annotations Patient  name,  age,  sex,  time,  date,  hospital  name,  doctor  name, comment (full-screen character editing) Body Mark 100 types   USB Port Two USB 3.0 and two USB 2.0 ports.  13.6 Operating, Storage and Transportation Environment 13.6.1. Operating Environment Temperature 0 °C ~ +40 °C(+32 °F~+104 °F) Relative humidity range 15% RH ~ 95% RH(non-condensing) Atmospheric pressure range 860 hPa ~ 1060 hPa  13.6.2. Storage and Transportation Environment Temperature -20 °C ~ +55 °C(-4 °F~+131 °F) Relative humidity range 15% RH ~ 95% RH(non-condensing) Atmospheric pressure range 700 hPa ~ 1060 hPa
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                EMC Information - 122 - 14  EMC Information Electromagnetic emissions  Electromagnetic emission The  Acclarix  AX8  is  intended  for  use  in  the  electromagnetic  environment  specified  below;  The customer or the user of the Acclarix AX8 should assure that it is used in such an environment. Emission test   Compliance Electromagnetic environment-guidance RF emissions CISPR 11 Group 1 The  Acclarix  AX8  uses  RF  energy  only  for  its  internal function. Therefore, its RF emissions are very low and are not  likely  to  cause  any  interference  in  nearby  electronic equipment. RF emissions CISPR 11 Class A The Acclarix AX8 is suitable for use in all establishments other  than  domestic  and  those  directly  connected to  the public  low-voltage  power  supply  network  that  supplies buildings used for domestic purposes. Harmonic emissions IEC 61000-3-2 Class A Voltage fluctuations/flicker emissions IEC61000-3-3 Complies
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                EMC Information - 123 - Electromagnetic immunity    Electromagnetic immunity   The Acclarix AX8 is intended for use in the electromagnetic environment specified below. The customer or the user of Acclarix AX8 should assure that it is used in such an environment. Immunity test   IEC 60601 test level   Compliance level   Electromagnetic environment -guidance   Electrostatic discharge  (ESD) IEC 61000-4-2   ±6 kV contact   ±8 kV air   ±6 kV contact ±8 kV air   Floors  should  be  wood, concrete or ceramic tile. If floor  are  covered  with synthetic  material,  the relative humidity should be at least 30%.   Electrical  fast transient/burst  IEC 61000-4-4   ±2  kV  for  power  supply lines   ±2KV  for  power  supply lines   Mains  power  quality should be that of a typical commercial  or  hospital environment.   Surge  IEC 61000-4-5   ±1 kV line to line ±2 kV line to ground ±1 kV line to line ±2 kV line to ground Mains  power  quality should be that of a typical commercial  or  hospital environment.   Power  frequency (50/60Hz)  magnetic field IEC61000-4-8   3A/m   3A/m   Power frequency magnetic fields  should  be  at  levels characteristic  of  a  typical location  in  a  typical commercial  or  hospital environment.   Voltage  dips,  short interruptions  and voltage  variations on  power  supply input  lines  IEC 61000-4-11   <5% UT     (>95% dip in UT)   for 0.5 cycle  40% UT (60% dip in UT) for 5 cycles  70% UT (30% dip in UT) for 25 cycles  <5%  UT  (>95%  dip  in UT) for 5 sec   <5% UT     (>95%  dip  in  UT)  for 0.5 cycle  40% UT (60% dip in UT) for 5 cycles  70% UT (30% dip in UT) for 25 cycles  <5%  UT  (>95%  dip  in UT) for 5 sec   Mains  power  quality should be that of a typical commercial  or  hospital environment. If the user of the  Acclarix  AX8  requires continued operation during power mains interruptions, we  recommend  that  the Acclarix  AX8  be  powered from  an  uninterruptible power supply or a battery.   NOTE UT is the a.c. mains voltage prior to application of the test level.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                EMC Information - 124 - Electromagnetic immunity    Electromagnetic immunity   The  Acclarix  AX8  is  intended  for  use  in  the  electromagnetic  environment  specified  below.  The customer or the user of the Acclarix AX8 should assure that it is used in such an environment.   Immunity test   IEC 60601 test level   Compliance level   Electromagnetic environment -guidance       Conducted RF  IEC 61000-4-6     Radiated RF IEC 61000-4-3     3 Vrms 150  kHz to  80 MHz    3 V/m   80  MHz to  2.5 GHz       3Vrms    3 V/m   Portable  and  mobile  RF  communications  equipment should be used no closer to any part of the Acclarix AX8, including  cables,  than  the  recommended  separation distance calculated from the  equation applicable to the frequency of the transmitter. Recommended separation distance Pd 1.2 150kHz ~ 80 MHz   Pd 1.2  80 MHz ~ 800 MHz Pd 2.3  800 MHz ~ 2.5 GHz Where  P  is  the  maximum  output  power  rating  of  the transmitter  in  watts  (W)  according  to  the  transmitter manufacturer  and  d  is  the  recommended  separation distance in metres (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,a should be less than the compliance level in each frequency range.b Interference  may  occur  in  the  vicinity  of  equipment marked with the following symbol:  NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.   NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.   a  Field  strengths  from  fixed  transmitters,  such  as  base  stations  for  radio  (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,  an electromagnetic site  survey should  be  considered. If the measured field strength in the location in which the Acclarix AX8 is used exceeds the applicable RF compliance level above, the Acclarix AX8 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Acclarix AX8 b  Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Acclarix AX8 Diagnostic Ultrasound System User Manual                                                EMC Information - 125 - Recommended separation distances between portable and mobile RF communication equipment and the EQUIPMENT or SYSTEM  Recommended  separation  distances  between  portable  and  mobile  RF  communications equipment and the Acclarix AX8 The  Acclarix  AX8  is  intended  for  use  in  an  electromagnetic  environment  in  which  radiated  RF disturbances  are  controlled.  The  customer  or  the  user  of  the  Acclarix  AX8  can  help  prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Acclarix AX8 as recommended below, according to the maximum output power of the communications equipment.  Rated  maximum output  power  of transmitter   (W) Separation distance according to frequency of transmitter (m) 150 kHz ~ 80 MHz Pd 2.1 80 MHz ~ 800 MHz Pd 2.1 800 MHz ~ 2.5 GHz Pd 3.2 0.01 0.12 0.12 0.23 0.1 0.38 0.38 0.73 1 1.2 1.2 2.3 10 3.8 3.8 7.3 100 12 12 23 For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.  NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
 - 126 -