EDAN INSTRUMENTS DX12TREDAN Portable ECG with Bluetooth User Manual PADECG iOS

EDAN INSTRUMENTS, INC. Portable ECG with Bluetooth PADECG iOS

User manual

 I About this Manual P/N: 01.54.000805 MPN: 01.54.000805015 Release Date: February 2017 © Copyright EDAN INSTRUMENTS, INC. 2014-2017. All rights reserved. Statement This manual will help you understand the operation and maintenance of the product better. It is reminded that the  product  shall  be used strictly complying  with  this manual. User’s  operation failing  to  comply  with  this  manual  may  result  in  malfunction  or  accident  for  which  EDAN INSTRUMENTS, INC. (hereinafter called EDAN) cannot be held liable. EDAN  owns  the  copyrights  of  this  manual.  Without  prior  written  consent  of  EDAN,  any materials contained in this manual shall not be photocopied, reproduced or translated into other languages. Materials protected by the copyright law, including but not limited to confidential information such as technical information and patent information are contained in this manual, the user shall not disclose such information to any irrelevant third party. The user shall  understand that  nothing  in  this  manual  grants him,  expressly or implicitly, any right or license to use any of the intellectual properties of EDAN. EDAN holds the rights to modify, update, and ultimately explain this manual. Responsibility of the Manufacturer EDAN only considers itself responsible for any effect on safety, reliability and performance of the equipment if: Assembly  operations,  extensions,  re-adjustments,  modifications  or  repairs  are  carried  out  by persons authorized by EDAN, and   The electrical installation of the relevant room complies with national standards, and   The instrument is used in accordance with the instructions for use.
 II Terms Used in this Manual This guide is designed to give key concepts on safety precautions. WARNING A  WARNING  label  advises  against  certain  actions  or  situations  that  could  result  in  personal injury or death. CAUTION A CAUTION label advises against actions or situations that could damage equipment, produce inaccurate data, or invalidate a procedure. NOTE A NOTE provides useful information regarding a function or a procedure.
 III Table of Contents Chapter 1 Safety Guidance ........................................................................................................... 1 1.1 Intended Use/Indications for Use ........................................................................................ 1 1.2 Warnings and Cautions ........................................................................................................ 1 1.2.1 General Warnings ...................................................................................................... 1 1.2.2 Battery Care Warnings .............................................................................................. 5 1.2.3 General Cautions ....................................................................................................... 5 1.3 List of Symbols ................................................................................................................... 6 Chapter 2 Introduction .................................................................................................................. 8 2.1 Assembling the System ....................................................................................................... 8 2.1.1 PADECG System ...................................................................................................... 8 2.1.2 DX12(iOS) Transmitter ........................................................................................... 10 2.1.2.1 Keys and Icons .............................................................................................. 11 2.1.2.2 Setting the Menu ........................................................................................... 12 2.2 Installing the Software ...................................................................................................... 12 2.2.1 System Running Environment ................................................................................ 12 2.2.2 About Installation .................................................................................................... 13 2.2.3 Matching DX12(iOS) Transmitter with iOS Tablet ................................................ 13 2.3 Features ............................................................................................................................. 14 Chapter 3 Preparations Before Operation ................................................................................. 15 3.1 Preparing the Patient ......................................................................................................... 15 3.1.1 Instructing the Patient ............................................................................................. 15 3.1.2 Cleaning the Skin .................................................................................................... 15 3.2 Connecting the Patient Cable ............................................................................................ 15 3.3 Attaching Electrodes ......................................................................................................... 16 3.3.1 Electrode Placement ................................................................................................ 17 3.3.2 Attaching the Reusable Electrodes .......................................................................... 18 3.3.2.1 Attaching the Limb Electrodes ...................................................................... 18 3.3.2.2 Attaching the Chest Electrodes ..................................................................... 18 3.3.3 Attaching the Disposable Electrodes ....................................................................... 19 Chapter 4 Operation Instructions .............................................................................................. 20 4.1 Entering Patient Information ............................................................................................. 20 4.2 Sampling ECG Data .......................................................................................................... 21 4.3 Analyzing ECG Data ......................................................................................................... 22 4.4 Printing Reports ................................................................................................................ 24 4.5 Processing Patient Records ............................................................................................... 25 4.6 Configuring the System..................................................................................................... 26 4.6.1 Patient Information Setting ..................................................................................... 26 4.6.2 Sampling Storage Setting ........................................................................................ 27
 IV 4.6.3 Filter Setting ............................................................................................................ 28 4.6.4 Transmission Setting ............................................................................................... 28 4.6.5 Other Setting ........................................................................................................... 29 Chapter 5 Hint Information ........................................................................................................ 30 Chapter 6 Cleaning, Care and Maintenance ............................................................................. 31 6.1 General Points ................................................................................................................... 31 6.2 Cleaning ............................................................................................................................ 31 6.2.1 Cleaning the DX12(iOS) Transmitter ..................................................................... 32 6.2.2 Cleaning the Patient Cable ...................................................................................... 32 6.2.3 Cleaning the Reusable Electrodes ........................................................................... 32 6.3 Disinfection ....................................................................................................................... 33 6.3.1 Disinfecting the DX12(iOS) Transmitter ................................................................ 33 6.3.2 Disinfecting the Patient Cable ................................................................................. 33 6.3.3 Disinfecting the Reusable Electrodes ..................................................................... 34 6.4 Care and Maintenance ....................................................................................................... 34 Chapter 7 Accessories .................................................................................................................. 36 Chapter 8 Warranty and Service ................................................................................................ 37 8.1 Warranty ............................................................................................................................ 37 8.2 Contact information .......................................................................................................... 37 Appendix 1 Technical Specifications .......................................................................................... 38 A1.1 Safety Specifications ...................................................................................................... 38 A1.2 Environment Specifications ........................................................................................... 38 A1.3 Physical Specifications ................................................................................................... 39 A1.4 Power Supply Specifications .......................................................................................... 39 A1.5 Performance Specifications ............................................................................................ 39 Appendix 2 EMC Information .................................................................................................... 41 Appendix 3 Abbreviation ............................................................................................................. 47
PADECG User Manual                                                                                                            Safety Guidance - 1 - Chapter 1 Safety Guidance This chapter provides important safety information related to the use of PADECG. 1.1 Intended Use/Indications for Use The intended use of PADECG is to acquire resting ECG signals from adult and pediatric patients through body surface ECG electrodes. It is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by PADECG can  help  users  to  analyze  and  diagnose  heart  disease.  However,  the  interpreted  ECG  with measurements and interpretive statements is offered to clinicians on an advisory basis only. It is mainly used in ECG inpatient department of hospitals or healthcare facilities. WARNING 1.  This system is not designed for intracardiac use or direct cardiac application. 2.  This system is not intended for home use. 3.  This system is not intended for treatment or monitoring. 4.  This system is intended for use on adult and pediatric patients only. 5.  The results given by the system should be examined based on the overall clinical condition of the patient, and they cannot substitute for regular checking. 1.2 Warnings and Cautions To  use  the  system  safely  and  effectively,  firstly  be  familiar  with  the  operation  method  of Windows and read the user manual in detail to be familiar with the proper operation method for the purpose of avoiding the possibility of system failure. The following warnings and cautions must be paid more attention to during the operation of the system. 1.2.1 General Warnings WARNING 1.  The system is intended to be used by qualified physicians or personnel professionally trained.  They  should  be  familiar  with  the  contents  of  this  user  manual  before operation.
PADECG User Manual                                                                                                            Safety Guidance - 2 - WARNING 2.  Only  qualified  service  engineers  can  install  this  equipment,  and  only  service engineers  authorized  by  the  manufacturer  can  open  the  shell.  Otherwise,  safety hazards may happen. 3. EXPLOSION  HAZARD - Do  not  use  the  system  in  the  presence  of  flammable anesthetic mixtures with oxygen or other flammable agents. 4.  Only the patient cable and other accessories supplied by the manufacturer can be used. Or else, the performance and electric shock protection cannot be guaranteed. The system has been safety tested with the recommended accessories, peripherals, and  leads,  and  no  hazard  is  found  when  the  system  is  operated  with  cardiac pacemakers or other stimulators. 5.  Make sure that all electrodes are connected to the patient correctly before operation. 6.  Ensure that the conductive parts of electrodes and associated connectors, including neutral electrodes, do not come in contact with earth or any other conducting objects. 7.  Disposable electrodes must be used during defibrillation. 8.  Electrodes of dissimilar metals should not be used; otherwise it may cause a high polarization voltage. 9.  The disposable electrodes can only be used for one time. 10. Do not touch the patient, bed, table or the equipment while using the ECG together with a defibrillator. 11. Do not touch accessible parts of electrical equipment and the patient simultaneously. 12. The use of equipment that applies high frequency voltages to the patient (including electrosurgical  equipment  and  some  respiration  transducers)  is  not  supported  and may  produce  undesired  results.  Disconnect  the  patient  data  cable  from  the  ECG workstation, or detach the leads from the patient prior to performing any procedure that uses high frequency surgical equipment. 13. Fix attention on the examination to avoid missing important ECG waves. 14. Do  not  connect  any  equipment  or  accessories  that  are  not  approved  by  the manufacturer or that are not IEC/EN 60601-1 approved to the system. The operation or use of non-approved equipment or accessories with the system is not tested or supported, and system operation and safety are not guaranteed. 15. The use of patient cable and other accessories not supplied by the manufacturer may result in increased emissions or decreased immunity of the equipment.
PADECG User Manual                                                                                                            Safety Guidance - 3 - WARNING 16. Any non-medical equipment (such as the external printer) is not allowed to be used within the patient vicinity (1.5m/6ft.). 17. Accessory equipment connected to the analog and digital interfaces must be certified according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data processing equipment and IEC/EN 60601-1 for medical equipment). Furthermore all configuration shall  comply  with  the  valid  version  of  the  standard  IEC/EN  60601-1.  Therefore anybody, who connects additional equipment to the signal input or output connector to configure a medical system, must make sure that it complies with the requirements of the  valid  version  of  the  system  standard  IEC/EN  60601-1. If in  doubt,  consult  our technical service department or your local distributor. 18. Connecting  any  accessory  (such  as  external  printer)  or  other  device  (such  as  the computer)  to  this  system  makes  a  medical  system.  In  that  case,  additional  safety measures  should  be  taken  during  installation  of  the  system,  and  the  system  shall provide: a)  Within the patient environment, a level of safety comparable to that provided by medical electrical equipment complying with IEC/EN 60601-1, and   b)  Outside the patient environment, the level of safety appropriate for non-medical electrical equipment complying with other IEC or ISO safety standards. 19. All the accessories connected to system must be installed outside the patient vicinity, if they do not meet the requirement of IEC/EN 60601-1. 20. You are recommended to purchase the iOS tablet from the manufacturer. Otherwise, the manufacturer will not be held responsible for the maintenance of the hardware, operating system and other accessories. 21. If multiple instruments are connected to a patient, the sum of the leakage currents may exceed the limits given in the IEC/EN 60601-1 and may pose a safety hazard. Consult your service personnel. 22. Connecting  to  other  devices  may  decrease  the  antistatic  gradation  of the  system during operation. 23. Make sure that there is no intense electromagnetic interference source around when using the wireless system of PADECG. Furthermore, Keep a unobstructed distance of at most 3 meters between DX12(iOS) Transmitter and the iOS tablet.
PADECG User Manual                                                                                                            Safety Guidance - 4 - WARNING 24. Do  not  open  the  battery  cover  of  DX12(iOS)  Transmitter  when  using  the  wireless system of PADECG. 25. The iOS tablet shall comply with the valid version of the standard IEC 60950 and be used outside the patient environment (at least 2 meters away from the patient). The iOS tablet shall be charged outside the patient environment, and no operations are permitted during the charging. 26. The device shall not be serviced or maintained while in use with a patient. 27. The medical electrical equipment needs to be installed and put into service according to Appendix 2 EMC information. 28. Portable  and  mobile  RF  communications  equipment  can  affect  medical  electrical equipment, refer to the recommended separation distances provided in Appendix 2 EMC Information. 29. The equipment should not be used adjacent to or stacked with other equipment, refer to the recommended separation distances provided in Appendix 2 EMC Information. 30. Assembly of the ECG workstation and modifications during actual service life shall be evaluated based on the requirements of IEC60601-1. 31. Check the gain and speed of the report thoroughly when confirming diagnosis.   32. The device is MR unsafe. It is not intended for use in an MRI environment. 33. Magnetic and electrical fields are capable of interfering with the proper performance of  the device.  For this  reason  make  sure that  all external  devices operated  in  the vicinity of the device comply with the relevant EMC requirements. X-ray equipment or MRI devices are a possible source of interference as they may emit higher levels of electromagnetic radiation.
PADECG User Manual                                                                                                            Safety Guidance - 5 - 1.2.2 Battery Care Warnings   WARNING 1.  Improper operation may cause the internal battery to be hot, ignited or exploded, and it may lead to the decrease of the battery capacity. It is necessary to read the user manual carefully and pay more attention to warning messages. 2.  Batteries of the same model and specification as manufacture configuration should be used. 3. DANGER  OF  EXPLOSION -- Do not  reverse  the  anode  and  the  cathode  when installing the battery. 4. Do not heat or splash the battery or throw it into fire or water. 5.  Do not destroy the battery; do not pierce battery with a sharp object such as a needle; do not hit with a hammer, step on or throw or drop to cause strong shock; do not disassemble or modify the battery. 6.  When leakage or foul smell is found, stop using the battery immediately. If your skin or cloth comes into contact with the leakage liquid, cleanse it with clean water at once. If the leakage liquid splashes into your eyes, do not wipe them. Irrigate them with clean water first and go to see a doctor immediately. 7.  Properly dispose of or recycle the depleted battery according to local regulations. 8.  Remove the battery from the transmitter if the system won’t be used for a long time. 9.  Replace the depleted battery before use. 1.2.3 General Cautions CAUTION 1.  Avoid  liquid  splash  and  excessive  temperature.  The  temperature  must  be  kept between 5 ºC  and 40 ºC  during operation, and it should be kept between -20 ºC  and 55 ºC  during transportation and storage. 2. Do not  use  the  equipment  in  a  dusty  environment  with  bad  ventilation  or in  the presence of corrosive.
PADECG User Manual                                                                                                            Safety Guidance - 6 - CAUTION 3.  Make sure that there is no intense electromagnetic interference source around the equipment, such as radio transmitters or mobile phones etc. Attention: large medical electrical equipment such as electrosurgical equipment, radiological equipment and magnetic  resonance  imaging  equipment  etc.  is  likely  to  bring  electromagnetic interference. 4.  The device and accessories are to be disposed of according to local regulations after their useful lives. Alternatively, they can be returned to the dealer or the manufacturer for recycling or proper disposal. Batteries are hazardous waste. Do NOT dispose of them together with house-hold garbage. At the end of their lives hand the batteries over to the applicable collection points for the recycling of waste batteries. For more detailed  information  about  recycling of  this  product  or battery,  please  contact your local Civic Office, or the shop where you purchased the product. 5.  Federal (U.S.) law restricts this device to sale by or on the order of a physician. 1.3 List of Symbols No. Symbol Description 1  DEFIBRILLATION-PROOF TYPE CF APPLIED PART 2  Caution 3  Operating instructions 4  Warning (Background: Yellow; Symbol&Outline: Black) 5  Refer to User Manual (Background: Blue; Symbol: White) 6  Non- ionizing electromagnetic radiation
PADECG User Manual                                                                                                            Safety Guidance - 7 - 7  General symbol for recovery/recyclable 8 Part Number 9 SERIAL NUMBER 10  Date of manufacture 11  MANUFACTURER 12  AUTHORISED  REPRESENTATIVE  IN  THE EUROPEAN COMMUNITY 13  CE marking 14  Disposal method 15  Caution: Federal (U.S.) law restricts this device to sale by or on the order of a physician. 16 FCC ID: SMQDX12TREDAN Federal Communications Commission:   FCC ID: SMQDX12TREDAN 17  MR Unsafe–Keep away from magnetic resonance imaging (MRI) equipment NOTE: The user manual is printed in black and white.
PADECG User Manual                                                                                                                      Introduction - 8 - Chapter 2 Introduction PADECG as mobile ECG Workstation has similar functions with an ordinary ECG Workstation. ECG  data  can  be  sampled,  analyzed  and  stored  in  a  Pad,  ECG  waves  can  be  reviewed.  Auto measurement and diagnosis are available, and the diagnosis template can be edited. PADECG includes the following equipment, you can also purchase the iOS tablet.  DX12(iOS) Transmitter  Patient Cable  Electrodes NOTE: The pictures and windows in this manual are for reference only. 2.1 Assembling the System 2.1.1 PADECG System iOS Tablet DX12(iOS) TransmitterPatientPatient CableWireless WARNING DX12(iOS)  Transmitter  of  the  wireless  system  uses  the  Bluetooth  technology,  which could make the patient with the pacemaker uncomfortable. Keep DX12(iOS) Transmitter far away from the pacemaker when using the wireless system of PADECG.
PADECG User Manual                                                                                                                      Introduction - 9 -  WARNING 1.  Accessory equipment connected to the analog and digital interfaces must be certified according  to  the  respective  IEC/EN  standards  (e.g.  IEC/EN  60950  for  data processing equipment and IEC/EN 60601-1 for medical equipment). Furthermore all configuration shall  comply with the  valid version of the  standard IEC/EN 60601-1. Therefore anybody, who connects additional equipment to the signal input or output connector to configure a medical system, must make sure that it complies with the requirements of the valid version of the system standard IEC/EN 60601-1. If in doubt, consult our technical service department or your local distributor. 2.  If multiple instruments are connected to a patient, the sum of the leakage currents may exceed the limits given in the IEC/EN 60601-1 and may pose a safety hazard. Consult your service personnel.
PADECG User Manual                                                                                                                      Introduction - 10 - 2.1.2 DX12(iOS) Transmitter DX12(iOS) Transmitter Appearance                Front Panel  Patient Cable Socket
PADECG User Manual                                                                                                                      Introduction - 11 - 2.1.2.1 Keys and Icons            Main Screen                                                            Menu Screen Keys/Icons Description  After  batteries  are  installed,  press  this  key  to  turn  on/off  DX12(iOS) Transmitter. When the Menu screen is displayed, press this key to return to the previous screen.    When the Main or Menu screen is displayed, press this key to enter the next menu. Press  this  key,  and  then  press              in  1.2  seconds  to  lock  /  unlock  the keypad.  When the Main screen is displayed, press this key to switch the lead. When the Menu screen is displayed, press this key to display an item in black.  Icon for Bluetooth Connection If this icon is not displayed on the main screen, you need to match the device manually.  Icon for Keypad Locked If no operation is taken, the Main screen will be displayed and the keyboard will be locked automatically in 8 seconds.  Icon for Battery Capacity When the battery is weak, a hint will be displayed in PADECG software.
PADECG User Manual                                                                                                                      Introduction - 12 - NOTE: ECG waveforms displayed on the DX12(iOS) Transmitter only indicates whether the  electrodes  are  attached  properly.  Diagnosis  and  evaluation  of  the  ECG should be done based on the display on the iPad. 2.1.2.2 Setting the Menu Menu Description Back Light Select On to turn on the backlight of LCD screen. Select Off to turn off the backlight of LCD screen. Auto Sleep Select  On  to  display  Sleeping  on  the  screen  and  make  DX12(iOS) Transmitter  be  in  low  power  consumption  mode  after  lead  off  for  5 minutes. Select Off to turn off auto sleep function. Language You can set the system language. Lead Electrode You can select IEC or AHA. Bluetooth Device You  can  see  the  Bluetooth  name  and  address  of  the  DX12(iOS) Transmitter. Device Information You can see the related information, such as software version, ID, address of the device, manufacture and release time about the device.   NOTE: The device information is for reference only. 2.2 Installing the Software NOTE: This section is only for reference when the operating system of the  iOS tablet needs to be reinstalled or DX12(iOS) Transmitter is broken. 2.2.1 System Running Environment Applicable iOS Tablets: iPad4, iPad air, iPad mini, iPad mini with Retina display Operating System: iOS 7.0 or above CPU: Apple A5 or above Capacity: 16 G or above   Transmission Built-in Bluetooth and WIFI
PADECG User Manual                                                                                                                      Introduction - 13 - WARNING 1.  Use the PADECG system only on the tablet that is installed with the official operating system (OS) versions released by the tablet's manufacturer. 2.  When upgrading the tablet OS, consult EDAN service engineer if necessary. 2.2.2 About Installation Enter the App Store of the iOS tablet, input PADECG to search for the software. The installation is the same as that of other software available in the iOS tablet. NOTE:   1.  To  uninstall  the  software,  exit  the  software  first.  Local  data  will  be  lost  after uninstalling the software. 2.  Please upgrade the software in time when a new version is available. 2.2.3 Matching DX12(iOS) Transmitter with iOS Tablet   For the first time to match the transmitter with the iOS tablet, perform the following operations: 1. Start the DX12(iOS) Transmitter and the iOS tablet. 2. Enable the Bluetooth function in iPad>Settings>Bluetooth. 3. Click the required DX12(iOS) Transmitter for connection. In later use, when turned on, the transmitter will automatically search for and connect with the iOS tablet of the previous match in 10s. Operation  for  viewing  device  information  of  the  required  transmitter:  turn  on  DX12(iOS) Transmitter  press     press    to display Device Information in black  press  . NOTE:   1.  Before  matching  DX12(iOS)  Transmitter  and  the  iOS  tablet,  ensure  batteries  of DX12(iOS) Transmitter and iOS tablet are full. 2.  You  need to reconnect the DX12(iOS) Transmitter to the iOS tablet after changing system language. 3.  Before  matching  the  transmitter,  please  ensure  that  the  iOS  tablet  Bluetooth  is  in unconnected state.
PADECG User Manual                                                                                                                      Introduction - 14 - WARNING This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:     1)  this device may not cause harmful interference, and   2)  this device must accept any interference received, including interference that may cause undesired operation. NOTE: 1.  This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed  and  used  in  accordance  with  the  instructions,  may  cause  harmful interference  to  radio  communications.  However,  there  is  no  guarantee  that interference will not occur in a particular installation. If this equipment does cause harmful  interference  to  radio  or  television  reception,  which  can  be  determined  by turning  the  equipment  off  and  on,  the  user  is  encouraged  to  try  to  correct  the interference by one or more of the following measures: -  Reorient or relocate the receiving antenna. -  Increase the separation between the equipment and receiver. -  Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. -  Consult the dealer or an experienced radio/TV technician for help. 2. Any  changes  or  modifications  to  this  unit  not  expressly  approved  by  the  party responsible for compliance could void the user's authority to operate the equipment. 2.3 Features  Reliable and handy data recording, suitable for doctors’ inspections and visits  iOS operating system, user friendly interfaces and easy operation  Supporting order downloading function  Perfect data management, sampled ECG data can be transmitted to SE-1515 over LAN  3-/6-/12-channel ECG waves are displayed simultaneously  Supporting amplifying ECG waves, providing manual measurement with an electronic ruler of high precision  0.32  Hz/0.67Hz  DFT  filter  greatly  reduces  the  baseline  fluctuations  without  affecting ECG signals  Supporting auto measurement and diagnosis  Supporting editing the Diagnosis Template  12-lead normal ECG analysis
PADECG User Manual                                                                         Preparations Before Operation - 15 - Chapter 3 Preparations Before Operation 3.1 Preparing the Patient 3.1.1 Instructing the Patient Before  attaching  the  electrodes,  greet  the  patient  and  explain  the  procedure.  Explaining  the procedure  decreases  the  patient’s  anxiety.  Reassure  the  patient  that  the  procedure  is  painless. Privacy is important for relaxation. When possible, prepare the patient in a quiet room or area where others can’t see the patient. Make sure that the  patient is comfortable. The more relaxed the patient is, the less the ECG will be affected by noise. 3.1.2 Cleaning the Skin Thorough  skin  preparation  is  very  important.  The  skin  is  a  poor  conductor  of  electricity  and frequently  creates  artifacts  that  distort  the  ECG  signals.  By  performing  methodical  skin preparation, you can greatly reduce the possibility of noise caused by muscle tremor and baseline drift, ensuring high-quality ECG waves. There is natural resistance on the skin surface due to dry, dead epidermal cells, oils and dirt. To clean the skin 1. Shave hair from electrode sites, if necessary. Excessive hair prevents a good connection. 2. Wash the area thoroughly with soap and water. 3. Dry the skin to increase capillary blood flow and to remove the dead, dry skin cells and oils. 4. Use the disposable frosting film in the standard accessory list to get good ECG waveform.   NOTE: Rub the skin with a gauze pad to increase capillary blood flow if you don’t operate the steps above. 3.2 Connecting the Patient Cable WARNING The  performance and  electric  shock  protection  can be  guaranteed  only  if  the  original patient cable and electrodes of the manufacturer are used.
PADECG User Manual                                                                         Preparations Before Operation - 16 -  The patient cable includes the main cable and lead wires which can be connected to electrodes. 1. Connect the patient cable to the socket of DX12(iOS) Transmitter. 2. Align all lead wires of the patient cable to avoid twisting, and connect the lead wires to the electrodes. Firmly attach them. 3.3 Attaching Electrodes WARNING Make sure that the conductive parts of electrodes and associated connectors, including neutral electrodes, do not come in contact with earth or any other conducting objects. The identifiers and color codes of electrodes used comply with IEC/EN requirements. In order to avoid  incorrect  connections,  the  electrode  identifiers  and  color  codes  are  specified  in  the following table. Moreover the equivalent codes according to American requirements are given in the following table too. Table 3-1 Electrodes and Their Identifiers and Color Codes   IEC AHA WILSON FRANK Identifier Color Code Identifier Color Code Right arm Right arm R Red RA White Left arm Left arm L Yellow LA Black Right leg Right leg N or RF Black RL Green Left leg Left leg F Green LL Red Chest 1 I C1 White/Red V1 Brown/Red Electrode  Connector  Lead  Wires  Connecting  to DX12(iOS) Transmitter Main Cable
PADECG User Manual                                                                         Preparations Before Operation - 17 - Chest 2 E C2 White/Yellow V2 Brown/Yellow Chest 3 C C3 White/Green V3 Brown/Green Chest 4 A C4 White/Brown V4 Brown/Blue Chest 5 M C5 White/Black V5 Brown/Orange Chest 6 H C6 White/Violet V6 Brown/Violet 3.3.1 Electrode Placement  Only for the Reusable Electrodes         Only for the Disposable Electrodes  IEC AHA Electrode Placement C1 V1 Fourth intercostal space at the right border of the sternum C2 V2 Fourth intercostal space at the left border of the sternum C3 V3 Fifth rib between C2 and C4 C4 V4 Fifth intercostal space on the left midclavicular line C5 V5 Left anterior axillary line at the horizontal level of C4 C6 V6 Left midaxillary line at the horizontal level of C4 L LA Left arm/Left deltoid R RA Right arm/Right deltoid
PADECG User Manual                                                                         Preparations Before Operation - 18 - F LL Left leg/Upper leg as close to torso as possible N RL Right leg/Upper leg as close to torso as possible 3.3.2 Attaching the Reusable Electrodes 3.3.2.1 Attaching the Limb Electrodes  Limb Electrode        Limb Electrode Connection: 1) Ensure that the electrodes are clean; 2) Clean  the  electrode  area  which  is  a  short  distance above the ankle or the wrist with 75% alcohol; 3) Daub the electrode area on the limb with gel evenly; 4) Place a small amount of gel on the metal part of the limb electrode clamp; 5) Connect the electrode to the limb, and make sure that the metal part is placed on the electrode area above the ankle or the wrist; 6) Attach all limb electrodes in the same way. 3.3.2.2 Attaching the Chest Electrodes  Chest Electrode Reed Connecting  to  a  Lead  Wire Clamp Suction Bulb Connecting to a Lead Wire Metal  Cup
PADECG User Manual                                                                         Preparations Before Operation - 19 - Chest Electrode Connection: 1) Ensure that the electrodes are clean; 2) Clean the electrode area on the chest surface with 75% alcohol; 3) Daub the round area of 25mm in diameter on each electrode site with gel evenly; 4) Place a small amount of gel on the brim of the chest electrode’s metal cup; 5) Place the electrode on the chest electrode site and squeeze the suction bulb. Unclench it and the electrode is adsorbed on the chest; 6) Attach all chest electrodes in the same way. NOTE: Long-time measurement with a strong negative pressure on the suction bulb may  cause  reddening  of  the  skin.  When  using  the  electrode  on  kids  or patients with delicate skin, squeeze the suction bulb lightly. 3.3.3 Attaching the Disposable Electrodes CAUTION The disposable electrodes can only be used for one time.  Connect the snap socket adapter to the disposable electrode. The quality of ECG waveform will be affected by the contact resistance between the patient and the electrode. In order to get a high-quality ECG, the skin-electrode resistance must be minimized while connecting electrodes.
PADECG User Manual                                                  Operation instructions - 20 - Chapter 4 Operation Instructions 1. Turn on the iOS tablet and enable the Bluetooth function in iPad>Settings>Bluetooth. 2. Connect to the required DX12(iOS) Transmitter. 3. Start the PADECG Analysis Software. 4. Input the user name and password, and then click Login. The default password is ecg (case sensitive). Click Setting, you can change the password, or enable or disable the Remember Password function. If no DX12(iOS) Transmitter is detected and no data exists on the file screen, a dialog box requiring password will be displayed. Enter the correct password and you will enter the login screen.   NOTE: Do not run other applications when running PADECG, or the system response speed will be affected. 4.1 Entering Patient Information   On the Patient screen, you can view or create patient orders. 1. Entering patient information Press    on  the  patient  screen,  and  then  the  system  will  automatically  enter  the  New Patient screen. Input  the  related  patient  information  in  the  inputting  area,  click  OK and  the  new  patient record will be listed on top of the information list. If Sample now is enabled, the system will automatically enter the pre-sampling screen after you click on the OK button. NOTE:  Patient  ID  is  a  must  when  entering  patient  information.  You  can  use  the number generated by the system or input a number manually. Patient ID can be a random character string excluding ‘/’,  ‘\’,  ‘:’,  ‘*’, ‘?’, ‘<’, ‘>’  ,‘|’, '%' and Chinese characters. 2. Searching, modifying and deleting patient information Input  the  patient  name  in  the  search  area,  click  on  ,  and  all  the  patient  information which  meet  the  conditions  will  be  displayed  in  the  information  list.  Patient  records  in emergent state will always be listed on the top. NOTE: Fuzzy search by patient name is supported. Long press the patient information in the information list, you can modify or delete patient information.
PADECG User Manual                                                  Operation instructions - 21 - 3. Description for buttons Key Description STAT ECG Press to enter the ECG sampling screen directly.   Options: Delete, Scan, Download, Download Setting  Delete: Press to delete one or multiple orders.  Scan: Press to input patient information by scanning a bar code.  Download: Press to download patient orders.  Download  Setting:  Press  to  set  the  filter  conditions  for downloading patient information from the server.  Press to load orders from the server (SE-1515). For details about the order settings,  please  refer  to  "Order  Server"  in  Section  4.6.4  "Transmission Setting". NOTE:   1.  1000 patient records can be displayed in the information list, and 200  latest  created  orders  in  the  server  can  be  loaded  at  one time. 2.  This item only applies to the local mode. 3.  Not all options related to SE-1515 are available.  If the number of orders exceeds the maximum of a page, press these keys to turn pages. NOTE: Each page contains 50 orders. File Press to enter the file manager screen. For details about the file manager screen, refer to section 4.5 "Processing Patient Records". System Setting Press to enter the system setting screen. For  details  about  the  system  setting  screen,  refer  to  section  4.6 "Configuring the System". 4.2 Sampling ECG Data Enable Sample now and then click OK on the New Patient screen, the system will automatically enter the pre-sampling screen.   If Sample now is not enabled, click on OK and click on the patient record in the patient list to enter the pre-sampling screen.
PADECG User Manual                                                  Operation instructions - 22 - Key Description Start/Stop Press this button to start/stop sampling ECG data.   NOTE: At most 180s ECG data can be sampled.   100Hz EMG Filter: Off, 25Hz, 35Hz or 45Hz Lowpass Filter: Off, 75Hz, 100Hz or 150Hz NOTE: This setup modified on the pre-sampling screen is only effective for the current patient.   10mm/mV Gain: 2.5 mm/mV, 5 mm/mV, 10 mm/mV or 20 mm/mV 25mm/s Speed: 5mm/s, 12.5mm/s, 25mm/s or 50mm/s Patient Press to enter the patient screen. File Press to enter the file manager screen. NOTE: Long press the tablet screen during ECG data pre-sampling, the display mode of 12-channel ECG waves will be switched among 12×1, 6×2+1 and 3×4+1. 4.3 Analyzing ECG Data The system will automatically enter the ECG Analysis screen after sampling ECG data. ECG data can be displayed with the following style: 12×1, 6×2+1, and 3×4+1. 1. Viewing ECG waveform Long press the rhythm waveform to  display the  Rhythm Lead window on the 6×2+1 or 3×4+1 ECG Analysis screen. You can view other rhythm waveform by selecting one lead in the pop-up window. You  can  view  other  waveforms  by  dragging  or  clicking  on  the  rhythm  waveform  on  the 6×2+1 or 3×4+1 ECG Analysis screen.   2. Amplifying and measuring ECG waveform Multi-touch on the ECG waveforms of the ECG Analysis screen can be used to amplify or minify the ECG waveforms. If  more  than  10s  ECG  data  has  been  sampled,  a  scrollbar  will  appear  above  the  ECG waveforms. You  can  slide  the scrollbar to adjust  the time  period  of  the  waveforms  to  be displayed. During sliding, the start time and end time of the displayed 10s ECG waveforms will be shown. You can click "+" or "-" to view the 10s ECG waveforms 5s later or earlier.
PADECG User Manual                                                  Operation instructions - 23 - Long pressing the waveform on the ECG Analysis screen can amplify the ECG waveform. Press    on the Amplified Waveform screen to measure the ECG waveform. Press this button again to cancel measuring. 3. Modifying measure information Long press one parameter in the top left corner of the ECG Analysis screen, and the Modify measure information window pops up. Make the related settings, and then click on the OK button to save the modifications. The common parameters are displayed as follows. Designation Description HR (bpm) Heart Rate P (ms) P-wave duration of the current lead PR (ms) P-R interval of the current lead QRS (ms) QRS complex duration of the current lead QT/QTc (ms) Q-T interval of the current lead/Normalized QT interval P/QRS/T (° ) Dominant direction of the average integrated ECG vectors RV5/SV1 (mV) The amplitude of R wave of V5 lead/the amplitude of S wave of V1 lead RV5+SV1 (mV) The amplitude of R wave of V5 lead plus the amplitude of S wave of V1 lead RV6/SV2 (mV) The amplitude of R wave of V6 lead/the amplitude of S wave of V2 lead 4. Modifying diagnosis results Long press one diagnosis result in the top right corner of the ECG Analysis screen, and the Auto Diagnosis window pops up. Select one diagnosis result from the Diagnosis List or input diagnosis information directly in the textbox, and click on the OK button to save the modifications. 5. Modifying Diagnosis List Long press one diagnosis result in the Diagnosis List, and you can add, delete, or modify the diagnosis result. NOTE: If Diagnosis List  is modified, you should click on the  Save  button in the Auto Diagnosis window to save the related modifications.
PADECG User Manual                                                  Operation instructions - 24 - 6. Hiding or displaying measure information and diagnosis results On the ECG Analysis screen, you can slide the ECG waveforms upward to hide measure information and diagnosis results or downward to display them again. 7. Email Click on    on the analysis screen, and then you can Email files by clicking on the Send Email button. The procedure is as follows: 1) Add an Email account. If no Email  account is  set,  add  one in  the  following directory:  iPad>Mail,  Conatact, Calendars. 2) Set  default  Email  information  in  the  following  directory:  PADECG>System Settings>Transmission Setting>Email Settings. 3) Click    on the ECG analysis screen and select Sending Email. If no default Email information is set, a dialog box requiring report format selection will be displayed. After the file format is selected, the current file will be added to the mail editing window as an accessory. Input the receiver address before sending it. 8. Edit Click    on  the  analysis  screen,  and  then  you  can  click  Edit  to  modify  patient information. 4.4 Printing Reports We suggest that the report be printed out on a printer to diagnose and sign. To  print a report, the tablet should be connected to a PC which is installed with  the  SE-1515 analysis software or an FTP server by setting the IP address and FTP parameters of the PADECG analysis  software.  For  details,  please  refer  to  "FTP  Setting"  in  section  4.6.4  "Transmission Setting". To use the FTP server, configure it before connecting to the PC. For details please contact the local distributor or the manufacturer. NOTE:   1.  This function is available only under network connection. 2.  Not all options related to SE-1515 are available. WARNING The printouts need to be done with no scaling or sizing done to fit the page.
PADECG User Manual                                                  Operation instructions - 25 - 4.5 Processing Patient Records 1. Search Input the patient name in the search  area,  click  on  , and all  the  examination records which meet the conditions will be displayed in the examination record list. NOTE: Fuzzy search by patient name is supported in the search area. 2. Upload and Delete Long press one examination record in the examination record list, and you can perform the following operations on the selected record: view, upload, modify, or delete. NOTE: 1.  The patient ID cannot be modified. 2. The modification on the patient name and gender will be synchronized to other data of the same patient ID. If you need to delete or upload records in bulk, press    and choose Upload or Delete. Key Description  Press this button to enter submenu.  Press this button to upload the selected examination records to the server.  If the checkbox before Select All is ticked, pressing this button can clear the examination record list.    Set the search criteria for the records to be displayed.
PADECG User Manual                                                  Operation instructions - 26 - 4.6 Configuring the System The system will automatically save the modifications after you make the related settings.    4.6.1 Patient Information Setting Item Description ID Generating Method Choose from: Auto or Manual Input Select Auto, the patient ID can be automatically generated according to the examination date. Select Manual Input,  you should enter the patient  ID manually on the New Patient screen. Patient  Info  Auto Clear  Enabled: The  system  automatically  clears  the  current  patient  information except Gender, Exam.Room, Technician, and Physician. If ID Input Method is set to Manual Input, the patient ID will also be cleared.  Disabled: A Clear key will appear in the New Patient window. If Sample now is on, the current patient information remains when creating a new patient. If  Sample  now  is  off,  the  current  patient  information  will  be automatically cleared when creating a new patient.
PADECG User Manual                                                  Operation instructions - 27 - First/Last Name If enabled, the Name textbox on the New Patient screen will change into the First Name and Last Name textboxes. Age   Choose from: Manual or Date of Birth Other  Display Settings You can configure the items to be displayed on the New Patient window, including:  height,  weight,  blood  pressure,  race,  medication,  department, room  number,  request  number,  exam.  room,  technician,  physician,  and priority.   You can also add other items to be displayed by editing the customization options. Barcode Setting NOTE: If unset, the scanned results may be incorrect. Enter the start and end addresses, the male and female codes and encoding mode, and then click on the OK button confirm. 4.6.2 Sampling Storage Setting Item Description Sampling Mode Choose from: Real-time Sample, Pre-sample  Select Pre-Sample, 10s ECG data before pressing the Start key will be saved.    Select Real-time Sample, 10s ECG data sampled after pressing the Start key will be saved. Sampling  Time Setting It can be set to a value between 10s and 180s. Auto Diagnosis If  selected,  the  system  automatically  generates  diagnosis  results  after sampling finishes. Enter  the  Analysis screen  when sampling finishes If selected, the system automatically enters the wave analysis screen after sampling finishes. File Format Choose from: SCP, FDA-XML, DICOM, PDF, JPG, BMP. If selected, files in the selected format will be generated when sampling finishes. NOTE: This item only applies to the local mode. Report Format Choose from: 12*1, 6*2+1, 3*4+1.
PADECG User Manual                                                  Operation instructions - 28 - Rhythm  Lead Definition It can be set to any one of the leads.   The default value is II. Edit  Analysis Result Locally If selected, you can modify the measurement information and diagnosis results locally. 4.6.3 Filter Setting Item Description DFT Filter DFT filter greatly reduces the baseline fluctuations without affecting ECG signals.  The  purpose  of  this  filter  is  to  keep  the  ECG  signals  on  the baseline of the printout. Choose from: 0.05Hz, 0.32Hz or 0.67Hz The set value is the low limit of the frequency range. EMG Filter EMG filter suppresses the disturbance caused by strong muscle tremor.   The cutoff frequency can be set to Off, 25Hz, 35Hz or 45Hz. Lowpass Filter Lowpass Filter restricts the bandwidth of input signals.   The cutoff frequency can be set to Off, 75Hz, 100Hz or 150Hz.   All  the  input  signals  whose  frequency  is  higher  than  the  set  cutoff frequency will be attenuated.   NOTE: The Lowpass Filter is effective only when the EMG Filter is set to Off. AC Filter AC  filter  suppresses  AC  interference  without  attenuating  or  distorting ECG signals.   Choose from: Off, 50Hz and 60Hz. NOTE: To pass the distortion test, the  ECG workstation has to be configured with the highest bandwidth in filter settings. Otherwise, ECG signal may be distorted. 4.6.4 Transmission Setting Item Description Device No. Type the Device No., within 30 ASCII characters.
PADECG User Manual                                                  Operation instructions - 29 - Auto Upload Select this item, the system will automatically upload files to the server when sampling finishes. NOTE: This item only applies to the local mode. Under the network mode,  the  system  automatically  uploads  files  when sampling finishes. Delete file   after uploading Select  this  item, the  system  will  automatically  delete  files  from  the examination record list after they are uploaded. Order Server Address Set it to IP address of the order server. NOTE:  For  more  information  on  configuring  network  settings, consult your Network Administrator. FTP Setting Set the FTP address, port, user name, and password of the FTP server. Email Settings Set the default recipient, subject, and file format. 4.6.5 Other Setting Item Description DEMO mode Choose from: Off, Normal or Abnormal Lead Electrode Choose from: IEC, AHA If  IEC  is  selected,  the  hint  information  on  the  sampling  screen  will show the IEC electrode name upon lead off. If AHA  is  selected,  the  hint information  on  the  sampling  screen will show the AHA electrode name upon lead off. Hospital Name Enter  the  hospital  name,  within  20  Chinese  characters  or  within  60 English characters. The configured hospital name will appear in the PDF/JPG/BMP report. System Password Type a password that allows you to access the System Setting screen. Restore to factory defaults Press to restore the factory settings. Device Info View or edit the information of sampling devices. NOTE: If  a  DX12(iOS) Transmitter is  connected,  its information (name  and  address)  will  be  displayed  on  the  screen, otherwise, the hint "No sampling device is connected" will be displayed.
PADECG User Manual                                                                                                              Hint Information - 30 - Chapter 5 Hint Information Hint information and the corresponding causes provided by the system are listed as follows. Table 5-1 Hint Information and Causes Hint Information Causes Lead off: X Electrodes  fall  off  the  patient  or  the  patient cable falls off the DX12(iOS) Transmitter. Lead wire reversal The limb leads are reversed. Order server connection fails! The order server is not enabled or the network connection is abnormal. FTP connection fails! The  FTP  server  is  not  enabled  or  the  network connection is abnormal. Battery  of  sampling  device  is  weak, please change the battery after the test! Battery of DX12(iOS) Transmitter is low. Battery is weak, and the sampling device will be powered off! Battery of DX12(iOS) Transmitter runs out.
PADECG User Manual                                                                                Cleaning, Care and Maintenance - 31 - Chapter 6 Cleaning, Care and Maintenance Use only the EDAN-approved substances and methods listed in this chapter to clean or disinfect your  equipment.  Warranty  does  not  cover  damage  caused  by  using  unapproved  substances  or methods. Edan Instruments has validated the cleaning and disinfection instructions provided in this User Manual. It is the responsibility of the healthcare professional to ensure that the instructions are followed so as to ensure adequate cleaning and disinfection. 6.1 General Points Keep your DX12(iOS) Transmitter and accessories free of dust and dirt. To prevent the device from damage, please follow the instructions:  Use only the recommended cleaning agents and disinfectants listed in this manual. Others may  cause  damage  (not  covered  by  warranty),  reduce  product  lifetime  or  cause  safety hazards.  Always dilute according to the manufacturer's instructions.  Unless otherwise specified, do not immerse any part of the equipment or any accessories in liquid.  Do not pour liquid onto the equipment.  Do not allow liquid to enter the case.  Never use abrasive material (such as steel wool or silver polish).  Inspect  the  DX12(iOS)  Transmitter  and  reusable  accessories  after  they  are  cleaned  and disinfected. CAUTION If you spill liquid on the equipment or accessories, or they are accidentally immersed in liquid, contact your service personnel or EDAN service engineer. 6.2 Cleaning If the equipment or accessory has been in contact with the patient, then cleaning is required after each use. The validated cleaning agents for cleaning the DX12(iOS) Transmitter and patient cable are:  Mild near neutral detergent  Ethanol (75%)  Isopropanol (70%) The validated cleaning agent for cleaning the reusable electrodes is:  Mild near neutral detergent
PADECG User Manual                                                                                Cleaning, Care and Maintenance - 32 - Cleaning agents should be applied or removed using a clean, soft, non-abrasive cloth or paper towel. 6.2.1 Cleaning the DX12(iOS) Transmitter WARNING Turn off the power and take out the battery before cleaning. 1. Switch off the DX12(iOS) Transmitter and take out the battery. 2. Wipe the exterior surface of the equipment using a soft cloth dampened with the cleaning solution until no visible contaminants remain. 3. Wipe off the  cleaning solution  with  a  fresh cloth  or  towel  dampened  with  tap water after cleaning until no visible cleaning agent remains. 4. Dry the DX12(iOS) Transmitter in a ventilated and cool place. 6.2.2 Cleaning the Patient Cable 1. Wipe the patient cable with a soft cloth dampened with the cleaning solution until no visible contaminants remain. 2. Wipe off the  cleaning solution  with  a  fresh cloth  or  towel  dampened  with  tap water after cleaning until no visible cleaning agent remains. 3. Wipe off with a dry cloth to remove residual moisture. 4. Leave the patient cable to air dry. CAUTION Any remainder of cleaning solution should be removed from the DX12(iOS) Transmitter and the patient cable after cleaning. 6.2.3 Cleaning the Reusable Electrodes 1. Wipe off with a soft cloth to remove residual gel. 2. Wipe the suction bulbs of chest electrodes and the clamps of limb electrodes with a soft cloth dampened with the cleaning solution until no visible contaminants remain. 3. Wipe off the  cleaning solution  with  a  fresh cloth  or towel  dampened with  tap  water  after cleaning until no visible cleaning agent remains. 4. Wipe off with a dry cloth to remove residual moisture. 5. Leave the suction bulbs and clamps to air dry.
PADECG User Manual                                                                                Cleaning, Care and Maintenance - 33 - 6.3 Disinfection If the equipment or accessory has been in contact with the patient, then disinfection is required. To avoid permanent damage to the equipment, it is recommended that disinfection is performed only when it is considered as necessary according to your hospital' regulations. Clean  the  equipment  and  reusable  accessories  before  they  are  disinfected.  The  validated disinfectants for disinfecting the DX12(iOS) Transmitter and patient cable are:  Ethanol (75%)  Isopropanol (70%) The validated disinfectant for disinfecting the reusable electrodes is:  Isopropanol (70%) If Ethanol or Isopropanol is used for both cleaning and disinfecting, then a new cloth is required to be used for the disinfection step. CAUTION 1. Do  not  use  high-temperature,  high-pressure  vapour  or  ionizing  radiation  as disinfection methods. 2.  Do not use chloric disinfectant such as chloride, sodium hypochlorite etc. 3.  Clean and disinfect reusable electrodes after each use. 6.3.1 Disinfecting the DX12(iOS) Transmitter WARNING Turn off the power and take out the battery before disinfection. 1. Switch off the DX12(iOS) Transmitter and take out the battery. 2. Wipe the exterior surface of the equipment using a soft cloth dampened with the disinfectant solution. 3. Wipe off the disinfectant solution with a dry cloth after disinfection if necessary. 4. Dry the DX12(iOS) Transmitter for at least 30 minutes in a ventilated and cool place. 6.3.2 Disinfecting the Patient Cable 1. Wipe the patient cable with a soft cloth dampened with the disinfectant solution. 2. Wipe off the disinfectant solution with a dry cloth after disinfection. 3. Leave the patient cable to air dry for at least 30 minutes.
PADECG User Manual                                                                                Cleaning, Care and Maintenance - 34 - 6.3.3 Disinfecting the Reusable Electrodes 1. Wipe the suction bulbs of chest electrodes and the clamps of limb electrodes with a soft cloth dampened with the disinfectant solution. 2. Wipe off the disinfectant solution with a dry cloth after disinfection. 3. Leave the suction bulbs and clamps to air dry for at least 30 minutes. 6.4 Care and Maintenance CAUTION Besides the maintenance requirements recommended in this manual, comply with local regulations on maintenance and measurement. The following safety checks should be performed at least every 12 months by a qualified person who has adequate training, knowledge, and practical experience to perform these tests. a)  Inspect the equipment and accessories for mechanical and functional damage. b)  Inspect the safety related labels for legibility. c)  Verify that the device functions properly as described in the instructions for use. d)  Test the enclosure leakage current according to IEC/EN 60601-1: Limit: NC 100 μA, SFC 500 μA. e)  Test the patient leakage current according to IEC/EN 60601-1: Limit: NC a.c. 10 μA, d.c. 10 μA; SFC a.c. 50 μA, d.c. 50 μA. f)  Test the patient auxiliary current according to IEC/EN 60601-1: Limit: NC a.c. 10 μA, d.c. 10 μA; SFC a.c. 50 μA, d.c. 50 μA. g)  Test the patient leakage current under single fault condition with mains voltage on the applied part according to IEC/EN 60601-1: Limit: 50 μA (CF). h)  Test the essential performance according to IEC/EN 60601-2-25, or methods recommended by the hospital or local distributor. The data should be recorded in an  equipment log. If the equipment  is not functioning properly or fails any of the above tests, the equipment has to be repaired. WARNING Failure  on  the  part  of  the  responsible  individual  hospital  or  institution  employing  this equipment  to  implement  a  satisfactory  maintenance  schedule  may  cause  undue equipment failures and possible health hazards.
PADECG User Manual                                                                                Cleaning, Care and Maintenance - 35 - 1) iOS Tablet and DX12(iOS) Transmitter ♦ Avoid excessive temperature, sunshine, humidity and dirt. ♦ Put the dustproof coat on the DX12(iOS) Transmitter after use and prevent shaking it violently when moving it to another place. ♦ Prevent  any  liquid  from  seeping  into  the  equipment;  otherwise  the  safety  and  the performance of the ECG workstation cannot be guaranteed. 2) Patient Cable ♦ Integrity  of  the  patient  cable,  including  the  main  cable  and  lead  wires,  should  be checked regularly. Make sure that it is conductible. ♦ Do not  drag or twist  the  patient  cable with excessive stress  while  using  it.  Hold  the connector plug instead of the cable when connecting or disconnecting the patient cable. ♦ Align the patient cable to avoid twisting, knotting or crooking in a closed angle while using it. ♦ Store the lead wires in a big wheel to prevent any people from stumbling. ♦ Once  damage  or  aging  of  the  patient  cable  is  found,  replace  it  with  a  new  one immediately. 3) Reusable Electrodes ♦ Electrodes must be cleansed after use and make sure there is no remainder gel on them. ♦ Keep suction bulbs of chest electrodes away from sunshine and excessive temperature. ♦ After long-term use, the surfaces of electrodes will be oxidized because of erosion and other causes. By this time, electrodes should be replaced to achieve high-quality ECG records. CAUTION The device and accessories are to be disposed of according to local regulations after their useful lives. Alternatively, they can be returned to the dealer or the manufacturer for recycling or proper disposal.
PADECG User Manual                                                           Accessories - 36 - Chapter 7 Accessories WARNING Only the patient cable and other accessories supplied by the manufacturer can be used. Or else, the performance and electric shock protection cannot be guaranteed. Table 7-1 Accessory List Accessory Part Number DX12 Patient Cable (IEC) 01.57.471030 DX12 Patient Cable (AHA) 01.57.471055 Patient Cable (IEC) 01.57.471278 Patient Cable (AHA) 01.57.471279 Chest electrodes 01.57.040163 Limb electrodes 01.57.040162 Excell Alkaline AA LR6 1.5V 01.21.064086 Portable Bag 01.56.465623 Adult Disposable Adhesive Electrodes 01.57.471056 Pediatric Disposable Adhesive Electrodes 01.57.471057 Clip/Snap/Banana Socket Adapter 01.57.040172 Alligator Clip/Banana Socket Adapters 01.57.040173 Snap/Banana Socket Adapters 01.13.107449 Disposable Resting Tab electrodes 01.57.471031 NOTE:   1.  The chest electrodes and limb electrodes are not available in the U.S. 2.  The part name may vary depending on context, but the part number is constant.
PADECG User Manual                                                                                                    Warranty and Service - 37 - Chapter 8 Warranty and Service 8.1 Warranty EDAN warrants that EDAN’s products meet the labeled specifications of the products and will be free from defects in materials and workmanship that occur within warranty period. The warranty is void in cases of: a) Damage caused by mishandling during shipping. b) Subsequent damage caused by improper use or maintenance. c) Damage caused by alteration or repair by anyone not authorized by EDAN. d) Damage caused by accidents. e) Replacement or removal of serial number label and manufacture label. If a product covered by this warranty is determined to be defective because of defective materials, components, or workmanship, and the warranty claim is made within the warranty period, EDAN will, at its discretion, repair or replace the defective part(s) free of charge. EDAN will not provide a substitute product for use when the defective product is being repaired. 8.2 Contact information If you have any question about maintenance, technical specifications or malfunctions of devices, contact your local distributor. Alternatively, you can send an email to EDAN service department at: support@edan.com.cn.
PADECG User Manual                                                                                                Technical Specifications - 38 - Appendix 1 Technical Specifications A1.1 Safety Specifications Comply with: IEC 60601-1:2005/A1:2012 EN 60601-1:2006/A1:2013 IEC 60601-1-2:2007 EN 60601-1-2:2007/AC:2010 IEC 60601-2-25:2011 Anti-electric-shock type: Internally powered equipment Anti-electric-shock degree: Type CF with defibrillation-proof Degree  of  protection  against harmful ingress of water: Ordinary  equipment  (Sealed  equipment  without  liquid proof) Degree  of  protection  against falling: Handheld device Disinfection/sterilization method: Refer to the user manual for details Degree of safety of application  in the presence of flammable gas: Equipment  not  suitable  for  use  in  the  presence  of flammable gas Working mode: Continuous operation EMC: CISPR 11, Group 1, Class A Patient  Leakage Current: NC <10μA (AC) / <10μA (DC) SFC <50μA (AC) / <50μA (DC) Patient  Auxiliary Current: NC <10μA (AC) / <10μA (DC) SFC <50μA (AC) / <50μA (DC) A1.2 Environment Specifications  Transport & Storage Working Temperature: -20ºC (-4ºF) ~ +55ºC (+131ºF) +5ºC (+41ºF) ~ +40ºC (+104ºF)
PADECG User Manual                                                                                                Technical Specifications - 39 - Relative Humidity: 15%RH ~ 95%RH   Non-Condensing 15%RH ~ 95%RH   Non-Condensing Atmospheric Pressure: 70 kPa ~ 106 kPa 70 kPa ~ 106 kPa A1.3 Physical Specifications Dimensions DX12(iOS) Transmitter:   63mm(L)×107mm(W) ×23mm(H) (2.5in×4.2in×0.9in) Weight DX12(iOS) Transmitter:   Approx. 113g (not including battery) A1.4 Power Supply Specifications Power Supply: DX12(iOS) Transmitter:   Input Power: 2×1.5V Excell Alkaline AA IEC LR6; Operation life of battery≥12 hours A1.5 Performance Specifications HR Recognition HR Range: 30 bpm ~300 bpm Accuracy: 1 bpm Filter (PADECG Analysis Software) AC Filter: 50Hz/60Hz/Off   DFT Filter: 0.32Hz (weak) /0.67Hz (strong) /0.05Hz EMG Filter: 25Hz/35Hz/45Hz/Off Lowpass Filter: 150Hz/100Hz/75Hz/Off DX12(iOS) Transmitter Performance Leads Mode: 12 standard leads Acquisition Mode: Simultaneously 12 leads A/D: 18 bits Resolution: 2.52uV/LSB Sample Frequency: 10,000 /sec/channel (sampling) 500 /sec/channel (analysis)
PADECG User Manual                                                                                                Technical Specifications - 40 - Time Constant: ≥3.2 s Frequency Response: 0.05Hz ~ 150Hz Gain: 2.5, 5, 10, 20 (mm/mV) (±5%) Speed: 5mm/s, 12.5mm/s, 25mm/s or 50mm/s Input Impedance: ≥20MΩ (10Hz) Input Circuit Current: ≤0.05μA Input Voltage Range <±5mVp-p Calibration Voltage: 1mV±2% DC Offset Voltage: ±5 00mV Minimum Amplitude: 20 μVp-p Noise: ≤15μVp-p Recovery Time After Defibrillation <5 s CMRR ≥100 dB Pacemaker Detection (Single Channel Detection) Amplitude ±2  mV to ± 500 mV Width 0.1 ms to 2.0 ms Sampling Frequency 10,000Hz, rhythm lead DX12(iOS) Transmitter Bluetooth Transmitting Frequency 2402 Hz ~ 2480Hz Frequency Band 2402 Hz ~ 2480Hz Modulation Type FHSS, GFSK, DPSK, DQPSK Transmitting Power ≥ 0 dBm NOTE: Operation of the equipment below the minimum amplitude may cause inaccurate results.
PADECG User Manual                                                                                                            EMC Information - 41 - Appendix 2 EMC Information Electromagnetic Emissions Guidance and manufacture’s declaration - electromagnetic emission The  PADECG  is  intended  for  use  in  the  electromagnetic  environment  specified  below.  The customer or the user of the PADECG should assure that it is used in such an environment. Emission test Compliance Electromagnetic environment - guidance RF emissions CISPR 11 Group 1 The  PADECG  uses  RF  energy  only  for  its internal function. Therefore, its RF emissions are very low  and  are  not  likely to cause  any interference in nearby electronic equipment. RF emission CISPR 11 Class A The  PADECG  is  suitable  for  use  in  all establishments, other than domestic and those directly  connected  to  the  public  low-voltage power supply network that supplies buildings used for domestic purposes. Harmonic emissions IEC 61000-3-2 Not applicable Voltage  fluctuations/ flicker emissions IEC 61000-3-3 Not applicable Electromagnetic Immunity Guidance and manufacture’s declaration - electromagnetic immunity The  PADECG  is  intended  for  use  in  the  electromagnetic  environment  specified  below.  The customer or the user of the PADECG should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance Electrostatic discharge (ESD) IEC 61000-4-2 8 kV contact 15 kV air 8 kV contact (only for PADECG) 15 kV air (only for PADECG) Floors  should  be  wood, concrete or ceramic tile. If floor  are  covered  with synthetic  material,  the relative humidity should be at least 30%.
PADECG User Manual                                                                                                            EMC Information - 42 - Electrical  fast transient/burst IEC 61000-4-4 2  kV  for  power supply lines 1kV for input/output lines Not applicable  Not applicable  Surge IEC 61000-4-5 1 kV line to line 2 kV line to ground Not applicable Not applicable Power frequency (50Hz/60Hz) magnetic field IEC 61000-4-8 30A/m 30A/m Power frequency magnetic fields  should  be  at  levels characteristic  of  a  typical location  in  a  typical commercial  or  hospital environment. Voltage  dips,  short interruptions  and voltage  variations on  power  supply input lines IEC 61000-4-11 0 % UT; 0,5 cycle   At 0°, 45°, 90°, 135°, 180°,  225°,  270° and 315°  0 % UT; 1 cycle and 70  %  UT;  25/30 cycles ) Single phase: at 0°  0  %  UT;  250/300 cycle Not applicable Not applicable NOTE      UT is the a.c. mains voltage prior to application of the test level.  Electromagnetic Immunity  Guidance and manufacture’s declaration - electromagnetic immunity The  PADECG  is  intended  for  use  in  the  electromagnetic  environment  specified  below.  The customer or the user of the PADECG should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic  environment  - guidance    Portable  and  mobile  RF communications  equipment  should  be used  no  closer  to  any  part  of  the
PADECG User Manual                                                                                                            EMC Information - 43 - PADECG,  including  cables,  than  the recommended  separation  distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance Conducted RF IEC61000-4-6 3 Vrms 150 kHz to 80 MHz 6Vrmsc)  in  ISM bands between 0,15  MHz  and  80 MHz 3 Vrms 150  kHz  to 80 MHz 6Vrmsc)  in ISM  bands between 0,15  MHz and 80 MHz  Pd 2.1 Radiated RF IEC61000-4-3  3 V/m 80 MHz to 2.7 GHz 3 V/m   80  MHz  to 2.7 GHz  Pd 2.1  80 MHz to 800 MHz Pd 3.2  800 MHz to 2.5 GHz  See Table 1   See Table 1 E/6 Pd at  RF  wireless communications  equipment  bands (Portable  RF  communications equipment  (including  peripherals  such as  antenna  cables  and  external antennas) should be used no closer than 30  cm  (12  inches)  to  any  part  of  the PADECG,  including  cables  specified by the manufacturer). Where P is the maximum output power rating  of  the  transmitter  in  watts  (W) according  to  the  transmitter manufacturer  and  d  is  the recommended  separation  distance  in meters (m). Field  strengths  from  fixed  RF transmitters,  as  determined  by  an electromagnetic site survey,a should be less than the compliance level in each
PADECG User Manual                                                                                                            EMC Information - 44 - frequency range.b Interference  may  occur  in  the  vicinity of  equipment  marked  with  the following symbol:  NOTE 1  At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2  These  guidelines  may  not  apply  in  all  situations.  Electromagnetic  propagation  is affected by absorption and reflection from structures, objects and people. a  Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment  due  to  fixed  RF  transmitters,  an  electromagnetic  site  survey  should  be considered. If the measured field strength in the location in which the PADECG is used exceeds the applicable RF compliance level above, the PADECG should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the PADECG. b  Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. c    The ISM (industrial, scientific and medical) bands between 0,15 MHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz. Table 1 Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment Test frequency (MHz) Band a) (MHz) Service a) Modulation b) Maximum power (W) Distance (m) Immunity test level 385 380-390 TETRA 400 Pulse modulation b) 18 Hz 1,8 0,3 27 450 430-470 GMRS 460, FRS 460 FM c) ± 5 kHz deviation 1 kHz sine 2   0,3   28
PADECG User Manual                                                                                                            EMC Information - 45 -  710 704-787 LTE Band 13, 17 Pulse modulation b) 217 Hz 0,2 0,3 9 745 780 810 800-960 GSM 800/900, TETRA 800, iDEN 820, CDMA 850, LTE Band 5 Pulse modulation b) 18 Hz 2   0,3   28 870 930 1720 1700-1990 GSM 1800; CDMA 1900; GSM 1900; DECT; LTE Band 1, 3, 4, 25; UMTS Pulse modulation b) 217 Hz 2   0,3   28 1845 1970 2450 2400-2570 Bluetooth, WLAN, 802.11 b/g/n, RFID 2450, LTE Band 7 Pulse modulation b) 217 Hz 2   0,3   28 5240 5100-5800 WLAN 802.11 a/n Pulse modulation b) 217 Hz 0,2 0,3 9 5500 5785 NOTE    If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3. a) For some services, only the uplink frequencies are included. b) The carrier shall be modulated using a 50 % duty cycle square wave signal. c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent actual modulation, it would be worst case.
PADECG User Manual                                                                                                            EMC Information - 46 - Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM Recommended separation distances between   portable and mobile RF communications equipment and the PADECG   The  PADECG  is  intended  for  use  in an  electromagnetic  environment  in  which  radiated  RF disturbances  are  controlled.  The  customer  or  the  user  of  the  PADECG  can  help  prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF  communications  equipment  (transmitters)  and  the  PADECG  as  recommended  below, according to the maximum output power of the communications equipment. Rated maximum output power of transmitter (W) Separation distance according to frequency of transmitter (m) 150 kHz to 80 MHz Pd 2.1 80 MHz to 800 MHz Pd 2.1 800 MHz to 2.7 GHz Pd 3.2 0.01 0.12 0.12 0.23 0.1 0.38 0.38 0.73 1 1.2 1.2 2.3 10 3.8 3.8 7.3 100 12 12 23 For  transmitters  rated  at  a  maximum  output  power  not  listed  above,  the  recommended separation  distance  d  in  metres  (m)  can  be  estimated  using  the  equation  applicable  to  the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.  NOTE 1 At  80  MHz  and  800  MHz,  the  separation  distance  for  the  higher  frequency  range applies. NOTE 2 These  guidelines  may  not  apply  in  all  situations.  Electromagnetic  propagation  is affected by absorption and reflection from structures, objects and people.
PADECG User Manual                                                          Abbreviation - 47 - Appendix 3 Abbreviation Abbreviation Statement ECG Electrocardiograph/Electrocardiogram BP Blood Pressure HR Heart Rate P P-wave Duration PR P-R Interval QRS QRS Complexes Duration QT/QTc Q-T Interval of the Current Lead / Normalized QT Interval P/QRS/T Dominant Direction of the Average Integrated ECG Vectors aVF Left Foot Augmented Lead aVL Left Arm Augmented Lead aVR Right Arm Augmented Lead LA Left Arm LL Left Leg RA Right Arm RL Right Leg ID Identification AC Alternating Current USB Universal Serial Bus NC Normal Condition SFC Single Fault Condition