Elcam Medical ACAL 13EAUPAWRLSYST Wireless Blood Pressure Transducer User Manual Wireless Transmitter Receiver

Elcam Medical ACAL Wireless Blood Pressure Transducer Wireless Transmitter Receiver

User Manual

Download: Elcam Medical ACAL 13EAUPAWRLSYST Wireless Blood Pressure Transducer User Manual Wireless Transmitter Receiver
Mirror Download [FCC.gov]Elcam Medical ACAL 13EAUPAWRLSYST Wireless Blood Pressure Transducer User Manual Wireless Transmitter Receiver
Document ID757780
Application IDFqq6Uy6mGxCrZbIWOA1Arg==
Document DescriptionUser Manual
Short Term ConfidentialNo
Permanent ConfidentialNo
SupercedeNo
Document TypeUser Manual
Display FormatAdobe Acrobat PDF - pdf
Filesize58.89kB (736118 bits)
Date Submitted2007-02-12 00:00:00
Date Available2007-02-12 00:00:00
Creation Date2007-01-15 13:17:01
Producing SoftwareAcrobat Distiller 5.0 (Windows)
Document Lastmod2007-01-15 13:17:26
Document TitleWireless Transmitter/Receiver
Document CreatorAcrobat PDFMaker 5.0 for Word
Document Author: Hospira User

Wireless System for
DPT
System Operating Manual
430-11431-001
Change History
Part Number
Reason for Revision
430-11167-001
Initial Release
Pages Affected
Page 1 of 34
How to Reach Us
For customer service within the United States, contact:
1-877-9Hospira or 1-877-946-7747
For technical assistance, product return authorization, and to order parts, accessories, or manuals within the United
States, contact Hospira Technical Support Operations.
1-800-241-4002
To order parts using the online eCatalog, download technical publications, technical training courses, and additional
services, visit the website at:
www.hospiraparts.com
After authorization, ship prepaid product returns to the following address:
(Call Hospira customer service for return address.)
Note: Outside the U.S., contact your local Hospira sales office.
Note: Decontaminate all product returns prior to shipping.
Introduction 5
Indications for Use 5
Contraindications 5
Wireless Transpac System Contents
Font Conventions 6
Front Panel 8
Calibration Buttons 8
Transpac Transducer Pigtail Sockets 9
Battery LED 9
Alarm LED 9
Power On/Off 9
OK 10
Next 10
Wireless Transducer Battery 10
Back Panel Features 11
Receiver Modules 12
Receiver Module 1(Required) 12
Receiver Module 2 (Required) 12
Receiver Modules 3, 4, & 5 (Optional) 13
Serial Cables 13
Receiver - Philips Adapter 14
Receiver - GE Adapter 15
System Overview 5
Calibration Buttons 5
Power On/Off 6
Transpac Transducer Pigtail Sockets 6
Battery LED 6
Alarm LED 6
OK 6
Next 7
Calibration 7
Wireless Transducer Battery 8
Accessories 8
Basic Operations 9
Set Up 9
Transducer connection 9
Power On and Wireless Link 10
Power Off 12
Operation 12
Settings 12
Calibration 13
Zeroing 13
Zeroing a Single Transducer 14
Zeroing All Transducers 14
Battery Operation 15
Check Battery Charge 15
Charge Battery 16
Battery Life/Battery Disposal 17
Battery Storage 17
Charging Batteries in the Transmitter 17
How to Check the Battery 17
How to Charge the Battery 18
Battery Charge Indicator 18
Swapping Batteries 18
Cleaning 18
Zero Single Transducer 20
Zero All Transducers 20
Zero A Single Transducer 21
Page 3 of 34
Transfer Between Monitors 21
Calibrating Multiple Monitors 22
Warnings and Precautions 23
Warnings 20
FCC Interference warning 21
Cautions 21
Alarms and Troubleshooting 27
Alarms 27
Transducer Alarms 27
Battery LEDs 28
receiver alarms 29
Troubleshooting 30
Specifications 33
Standards 37
Applicable Standards 38
System Specifications 42
Warranty 45
Chapter 1
Introduction
This document describes the features, functions, operation of the Wireless System for DPT®
transceiver system for use with Hospira Transpac IV® and Hospira Transpac IT® disposable
pressure transducers.
The base system includes a transmitter module and a minimum of 2 receiver channels to support
the wireless transmission of 2 invasive blood pressure signals. Additional receiver modules can be
purchased to support up to 5 transducers per system.
This manual is intended for use by trained medical clinicians familiar with multi-parameter patient
monitors.
INDICATIONS FOR USE
Wireless System for DPT® is indicated for use when invasive blood pressure monitoring is needed
and a cable-free connection is desired between Transpac IT® or Transpac IV® transducers and a
patient monitor. Wireless System for DPT® supports only pressure transducers provided by
Hospira.
CONTRAINDICATIONS
The use of Wireless System for DPT® is contraindicated:
•
Do not use to synchronize external devices.
•
Do not use during MRI or PET scanning.
•
Do not use to trigger aortic balloon assist devices.
WIRELESS SYSTEM FOR DPT SYSTEM CONTENTS
The system and accessories are packaged as follows:
Base System Box (required)
•
•
Transmitter
•
AC power cord
•
pre-installed battery
Receivers
•
1 IBP channel #1 receiver module
•
1 IBP channel #2 receiver module
•
1 receiver cable
•
user manual
Monitor Specific Plug Adapters Box (required)
•
20 monitor-specific plug adapters
Additional IBP Channels #3+ Receiver Modules Box (optional)
•
10 receiver modules for IBP channel #3, #4, or #5
•
10 receiver cables
Additional IBP Channels #1 & #2 Receiver Modules Box (optional)
•
2 IBP channel #1 receiver modules
•
2 IBP channel #2 receiver modules
•
2 receiver cables
Contact Hospira customer service for additional accessories.
Page 5 of 34
Transmitter Front Panel
The front panel features include the following user interface:
•
Calibration Buttons
•
Pressure Transducer Pigtail Sockets
•
Battery LED
•
Alarm LED
•
Power On/Off Button
•
OK Button
•
Next Button
•
Pressure Transducer Mounting Positions
CALIBRATION BUTTONS
The calibration buttons send a 100 mmHg pressure signal to the selected channel for 10 seconds.
The calibration button zeroes the channel if it is pressed while the calibration signal is displayed.
Calibration Buttons (P1-P5)
Alarm LED
Battery LED
Next Button
Tr anspac
Mounting
Positions
OK Button
Power
On/Of f
Tr anspac T r ansducer Pigtail Sockets
Figure 1.1
PRESSURE TRANSDUCER PIGTAIL SOCKETS
The pressure transducer pigtail sockets are color coded to aid in blood pressure channel
traceability from the patient monitor to the patient.
Inserting a pressure transducer plug into a socket triggers power-on.
BATTERY LEDS
The battery indicator has 3 LEDs to indicate battery charge (full, medium, low). See [insert cross
reference to go to the Battery LED table in chapter on Troubleshooting]
When not connected to AC, the system powers-off after 60 minutes if the battery discharges to a
predefined safe low voltage.
When the transmitter is connected to AC, the LEDs flash to indicate that the battery is charging.
ALARM LED
The Alarm LED flashes to indicate and alarm state.
See [insert cross reference to go to the Alarm LED table in chapter on Troubleshooting]
POWER ON/OFF
The transmitter is switched ON or OFF by pressing the power button.
OK
The OK button is used during the wireless linking process.. A Wireless System for DPT equipped
monitor will display a triangular “linking” waveform prior to the formation of a wireless link to a
Wireless System for DPT transmitter. Press the OK button to immediately confirm the link.
The OK button is also used during ZERO ALL. When a calibration signal is displayed, press the OK
button to zero all channels. See Zeroing on page 13 for more
information.
NEXT
The NEXT button cancels a wireless linking process if pressed while the “linking” signal is displayed
on a monitor. The wireless linking process is then initiated with the next available monitor.
If the NEXT button is pressed while the transmitter is already linked, the link is broken and a
wireless linking process is initiated with the next available monitor.
BATTERY
The Li-ion battery is accessed via the back panel. A built-in battery charger automatically charges
the battery whenever the transmitter is connected to AC power.
Battery replacement should only be done by a trained technician.
Contact Hospira customer support for replacement batteries.
Back Panel
Page 7 of 34
Alarm Port
Pole Clamp
Mounting
Area
Battery
Compartment
AC Socket Door
Features
Receiver Modules
Wireless System for DPT receiver modules are connected to the IBP sockets of a patient monitor
and present the same electrical load and response as industry standard 5 µV/V pressure
transducers. They are powered by the monitor and contain no batteries or other serviceable parts.
The illustration shows the modules with GE SolarTM series adapters attached. Contact Hospira
customer service to order adapters for monitors from other manufacturers.
Module 2
Serial Cables
Module 1
CAUTION: ONCE THE MONITOR SPECIFIC ADAPTERS ARE CONNECTED THEY CAN NOT BE
REMOVED.
RECEIVER MODULE 1(REQUIRED)
Receiver module number 1 is required. It simulates the presence of the transducer connected to
transmitter socket “P1” and is color coded red.
RECEIVER MODULE 2 (REQUIRED)
Receiver module number 2 is required. It simulates the presence of the transducer connected to
transmitter socket “P2” and is color coded blue.
RECEIVER MODULES 3, 4, & 5 (OPTIONAL)
Receiver modules for channels 3, 4, and 5 are optional. The channel number is selected by the
switch shown below.
RECEIVER CABLES (REQUIRED)
The receiver cables distribute power and data between the modules.
MONITOR ADAPTERS (REQUIRED)
The receiver modules require plug adapters that are often unique to each monitor manufacturer.
Contact Hospira customer service for a list of compatible monitors and to order adapters.
Page 9 of 34
Wireless Transpac System Operations Manual 430-11431-100 (09/06)
Chapter 2
System Overview
The Wireless Transducer is designed to eliminate the wires and cables that tether patients to
monitoring equipment. The cable function is replaced by the Wireless Transducer.
The Wireless Transducer can be powered by a battery for portable use, but is intended to be
operated using standard AC power when available. While connected to AC power, the Wireless
Transducer also functions as a battery charger.
CALIBRATION BUTTONS
Five buttons provide calibration information. The Transpac Transducer sends a 100 mmhg
calibration signal on the selected channel for 10 seconds. The receiver and monitor use this
calibration signal for a secondary monitor and channel number confirmation. The calibration
buttons zero the channel if pressed while the calibration signal is
Calibration Buttons
Power
On/Of f
Tr anspac T r ansducer Pigtail Sockets
displayed.
Figure 2.1
POWER ON/OFF
The Transpac Transducer is turned on by activating the power button.
TRANSPAC TRANSDUCER PIGTAIL SOCKETS
The Transpac Transducer Pigtail Sockets are color coded for jumper cable attachment. Inserting
the plug into the socket triggers a power on cycle.
BATTERY LED
The Battery LED indicates that the power is on.
When not connected to AC, the LED flashes and the system powers off after 60 minutes if the
battery discharges to a predefined safe low voltage.
When the Transpac Transducer is connected to AC, the LED flashes to indicate that the Transpac
Transducer is charging.
ALARM LED
The Alarm LED flashes when the alarm sounds. It blinks when wireless interference is
detected.
OK
Press the Ok button to immediately confirm and link. The monitor displays a corresponding “linking”
waveform on the monitor.
If you press the Ok button when a calibration signal is displayed, the output of all channels are set to
zero. See Zeroing on page 13 for more
information.
NEXT
The Next Button selects the next receiver if it is pressed when the “linking” signal is displayed. The
“linking” signal is redirected to the next receiver. The Next Button triggers new link.
If the Next Button is pressed when it is already linked, the link is broken, and “linking” signal is
displayed on the next available
monitor.
CALIBRATION
The calibration buttons are ....
Wireless Transducer Battery
The battery is located on the back panel under a compartment door. The Transpac Transducer
supports the LI-ON Battery.
Page 11 of 34
A built-in battery charger automatically charges the battery, whenever the Transpac Transducer is
plugged into AC
Pole Clamp
Mounting
Area
Battery
AC Socket Door
power.
Figure 2.2 Transpac Battery
ACCESSORIES
•
10 cm length cable
•
25 cm length cable
•
additional receiver adaptors (3) qty
Wireless Transpac System Operations Manual 430-11431-100 (09/06)
Chapter 3
Basic Operations
This chapter contains instructions and information about the basic operations of the
transmitter/receiver device.
Set Up
Below are instructions on how to set up the transmitter, transducers, and receiver modules.
TRANSDUCER CONNECTION
WARNING: ONLY USE TRANSDUCERS PROVIDED BY HOSPIRA. USE OF OTHER
TRANSDUCERS MAY DAMAGE THE TRANSMITTER AND VOIDS THE WARRANTY.
NOTE: Transducer connection will trigger power on.
1. SLIDE transducer into mounting rack.
2. PLUG pigtail cable into socket under transducer.
POWER ON AND WIRELESS LINK
For normal operation, always keep transmitter connected to AC power.
1. CONFIRM AC is connected (use GFCI outlet).
2. If not already on, PRESS and HOLD power button [add graphic here] until system beeps.
is OK
CONFIRM that the power status
3. Once the transmitter autotmatically links to the strongest receiver signal:
Figure 3.1 a. CONFIRM triangle waveform is displayed on desired monitor. See
below.
b. PRESS “OK”
Page 13 of 34
c. If no triangle waveform, PRESS “NEXT”
d. If no triangle waverform appears, PRESS “NEXT”
again.
e. If there is still no triangle waveform, POWER OFF and POWER ON.
Figure 3.1 Sample of Waveform
NOTE: Waveform may appear clipped due to scaling. See Figure 4.2 below.
Figure 3.2 Sample of Clipped Waveform
POWER OFF
PRESS and HOLD “POWER”
until system beeps and shuts down.
Operation
Once the batteries have been checked and the Wireless Transducer has been set up, there are
three basic steps to using the Wireless Transducer.
For normal operations, keep AC connected to a GFCI outlet.
SETTINGS
Power On the transmitter and select the appropriate settings from the receiver module channel
selector. The channel selector switch offers the following selections: p3, p4 , or p5. Selection p5 is in
the middle because it is difficult to select. Duplicate channels on bus will prevent system
operation.
picture of receiver setting 09/18/06
CALIBRATION
begin Content from Tips Card here 09/12/06
1. PRESS and RELEASE desired calibration button.
2. CONFIRM monitor displays 100 mmHg on the corresponding channel (signal appears for 10
seconds).
A. IF 100 MMHG SIGNAL DOES NOT APPEAR, reset THE WIRELESS LINK.
b. If signal appears on the wrong channel, contact Technical Services for sytem stetup.
c. If signal is not within 98 - 102 mmHg, contact Technical Services for monitor calibration.
3. REPEAT Steps 1 and 2 for each connected transducer.
The Transpac Transmitter sends a 100 mmhg calibration signal to the selected receiver channel
and monitor for 10 seconds to provide a secondary confirmation of the monitor and channel
number.
ZEROING
WARNING: BEFORE ZEROING THE TRANSDUCERS, MOUNT AND SECURE THE
TRANSMITTER SO THAT THE PORTS TO THE ATMOSPHERE ARE LEVEL WITH THE
PATIENT’S HEART.
Connect and prime patient tubing prior to zeroing.
ZEROING A SINGLE TRANSDUCER
1. OPEN the transducer to the atmosphere (see manufacturer’s instructions).
2. CONFIRM flatline pressure trace on the monitor.
3. PRESS and RELEASE the transducer’s calibration button
(add graphic of cal. button).
4. PRESS and RELEASE the transducer’s calibration button
calibration signal displays.
a second time while the
5. CONFIRM that the pressure trace goes to zero.
6. CLOSE the transducer’s stopcock to the atmosphere.
ZEROING ALL TRANSDUCERS
1. OPEN all transducers to the atmosphere (see manufacturer’s instructions).
2. CONFIRM flatline pressure traces for each on the monitor.
Page 15 of 34
3. PRESS and RELEASE
transducer
4. PRESS and RELEASE
(add graphic of calibration button here) corresponding to first
while the calibration signal displays.
5. CONFIRM that the pressure traces for all transducers go to zero.
6. CLOSE each transducer’s stopcock to atmosphere.
Battery Operation
CHECK BATTERY CHARGE
1. If Transducer is not already powered on, press and hold
until system beeps.
Confirm that the battery LED indicates a sufficient charge. See “Battery LED Table” on page 28.
CHARGE BATTERY
1. Connect transmitter to AC.
Battery LED will blink to indicate charging.
2. Charge battery for 6 hours.
NOTE:
-- Keep AC connected to GFCI outlet when not in transport.
-- Battery capacity is 72 hours of operation.
-- Low battery alarm triggered when less than 1 hour of battery capacity remains.
The battery used for the Wireless Transducer is a 3.6 Volt rechargeable Lithium-ion battery. A fully
charged battery will power a Wireless Transducer Patient Transceiver for a minimum of 24 hours.
Batteries are shipped in the uncharged state, with a plastic film cover covering the contacts.
CAUTION: REMOVE THE PLASTIC FILM COVER FROM THE BATTERY CONTACTS BEFORE USING. DO
NOT SHORT THE BATTERY TERMINALS.
DO NOT TRY TO CONNECT A BATTERY WITH ANY
DO NOT EXPOSE THE BATTERY TO HIGH
DEVICE OTHER THAN A WIRELESS TRANSDUCER.
TEMPERATURE (ABOVE 60ÛC
/ 140ÛF).
CAUTION: DO NOT INCINERATE, SUBMERGE, CRUSH, DISASSEMBLE OR AUTOCLAVE THE BATTERY.
DO
NOT RECHARGE OR REUSE BATTERY THAT HAS BEEN SUBMERGED; DISCARD OR
RECYCLE IT IMMEDIATELY.
BATTERY LIFE/BATTERY DISPOSAL
It is recommended that batteries be discarded after one year of use, due to degradation of capacity.
After extended use, the battery will deliver decreased operational life per charge. Recycle or
dispose of discarded batteries according to national, state, and local regulations.
BATTERY STORAGE
Batteries should be removed from Transmitter and stored separately when not in use for more than
a few days.
NOTE: THE TRANSMITTER WILL CONTINUE TO DRAW A SMALL AMOUNT OF POWER FROM
THE BATTERY EVEN WHEN NOT CONNECTED AND ALL INDICATORS ARE OFF. IF BATTERIES
ARE STORED FOR MORE THAN A MONTH THEY WILL NEED CHARGING.
NOTE: STORING THE BATTERY FOR A LONG PERIOD OF TIME WITHOUT USE MAY DEGRADE
THE BATTERY CAPACITY.
CAUTION: STORE BATTERIES BETWEEN -40ÛC AND 70ÛC/ -40ÛF AND 160ÛF. KEEP BATTERIES IN
LOW HUMIDITY LOCATION WITH LITTLE TEMPERATURE VARIATION.
KEEP
BATTERIES
AWAY FROM DIRECT SUNLIGHT.
CHARGING BATTERIES IN THE TRANSMITTER
Whenever the Wireless Transducer Transceiver is connected to AC power, it functions as a charger for the
battery.
HOW TO CHECK THE BATTERY
1. If not already on, press and hold power button until system beeps.
2. Confirm battery LED indicates sufficient charge.
Page 17 of 34
HOW TO CHARGE THE BATTERY
1. Connect transmitter to AC.
2. Battery LED will blink to indicate charging.
3. Charge battery for 6 hours.
BATTERY CHARGE INDICATOR
The battery charge indicator on the front of the Wireless Transducer displays a strobe when the
battery charger function is active. When the Wireless Transducer is powered by a battery (not
connected to AC power), and the battery discharges to a predefined safe low voltage, the LEDS will
flash and the system will power-off after 60 .
SWAPPING BATTERIES
The Wireless Transducer is intended to be used primarily with the Wireless Transducer connected
to AC power. When the battery runs low, it can quickly be swapped for the fully charged battery.
CAUTION: USE
ONLY A
WIRELESS TRANSDUCER TRANSMITTER
OR A BATTERY CHARGER TO
RECHARGE THE BATTERY.
NOTE: SWAPPING BATTERIES DOES NOT REQUIRE RE-CALIBRATION OF THE WIRELESS
TRANSDUCER.
Cleaning
The Wireless Transducer, battery, and power supply may be cleaned as needed by wiping with a
nearly dry cloth, moistened with one of the following widely used hospital disinfectants: 1:10 ratio
household bleach and water, Sporicidin, Manu-Klenz, warm water and soap, Fantastik®, Cidex®,
70% isopropyl alcohol or T.B.Q.® (Calgon Corp.). Thoroughly wipe off any excess cleaning solution
with a dry cloth.
CAUTION: DO NOT USE THE FOLLOWING SOLVENTS FOR CLEANING THE WIRELESS TRANSDUCER,
BATTERY, AND POWER SUPPLY : BUTYL ALCOHOL, ETHANOL,
FREON, BLEACH, ACETONE,
USE OF ANY OF THESE SOLVENTS MAY LEAD TO DETERIORATION
OF PLASTIC COMPONENTS. DO NOT AUTOCLAVEANY COMPONENTS.
HYDROGEN PEROXIDE.
CAUTION: TRANSCEIVERS SHOULD HAVE BATTERIES INSTALLED WHENEVER THEY ARE CLEANED.
BATTERIES MAY BE CLEANED SEPARATELY BUT CARE MUST BE TAKEN TOAVOID WETTING
THE METAL BATTERY CONTACTS.
WARNING: DO NOT ALLOW LIQUID INTO BATTERY SLOTS, BATTERY CONTACTS,
CONNECTOR PORTS, TOKEN PORTS OR ANY OTHER OPENINGS OR CREVICES. DO NOT
IMMERSE ANY COMPONENT IN LIQUID. SYSTEMS ARE INTENDED FOR SINGLE PATIENT
USE ONLY. THEY MAY BE WIPED WITH A NEARLY DRY CLOTH, MOISTENED WITH WARM
WATER. IF THE SYSTEMS AND/OR THE ARMBAND BECOMES BADLY SOILED, IT SHOULD
BE REPLACED.
NOTE: These recommendations are provided for guidance only. The actual cleanliness of
devices depends upon techniques and actual cleaning practices employed by the user.
WARNING: BEFORE ZEROING THE TRANSDUCERS, MOUNT AND SECURE THE
TRANSMITTER SO THAT THE PORTS TO THE ATMOSPHERE ARE LEVEL WITH THE
PATIENT’S HEART.
1. ZERO the monitor.
2. Confirm that the receiver modules have power and are not linked (un-paired) to any transmitter.
3. If the monitor displays 0.0 mmHg for each pressure channel, ZEROing is not required.
4. If the monitor does not display 0.0 mmHg for each pressure channel, ZERO each channel per
the monitor manufacturer's instructions (GE example attached).
5. If the monitor does not display 0.0 mmHg for each pressure channel after ZEROing, contact
your technical services department.
ZERO SINGLE TRANSDUCER
1. Open transducer to atmosphere (see manufacturer’s instructions).
2. Confirm flatline
ZERO ALL TRANSDUCERS
BEFORE LINKING TO A RECEIVER, CONFIRM THAT THE MONITOR IS
ZEROED AS DESCRIBED ZEROING
1. .
2. Confirm that the transmitter is linked to the desired receiver modules.
3. Prepare all transducers for ZEROing per your established procedures (GE example attached).
4. If the monitor displays 0.0 mmHg for all pressure channels, ZEROing is not required.
5. To ZERO all channels, press OK. ["LAP" WILL BE USED FOR THE CURRENT PROTOTYPE]
6. The pressure traces will go to 0.0 mmHg.
NOTE: The ZERO request will be ignored if the any measured pressure is beyond +/- 150
mmHg or is not stable (within +/- 1 mmHg).
7. Confirm that the monitor displays 0.0 mmHg for all pressure channels.
8. If the monitor does not display 0.0 mmHg for all pressure channels, contact your technical
services department.
ZERO A SINGLE TRANSDUCER
1. Before linking to a receiver, confirm that the monitor is ZEROed as described above.
2. Confirm that the transmitter is linked to the desired receiver modules.
3. Prepare the transducer for ZEROing per your established procedures (GE example attached).
4. If the monitor displays 0.0 mmHg for the pressure channel, ZEROing is not required.
5. To ZERO a single channel, press that channel's calibration button.
6. While the 100 mmHg calibration signal is displayed on the monitor, press that calibration button
again. The pressure trace will go to 0.0 mmHg. NOTE: The ZERO request will be ignored if the
measured pressure is beyond +/- 150 mmHg or is not stable (within +/- 1 mmHg).
7. Confirm that the monitor displays 0.0 mmHg.
8. If the monitor does not display 0.0 mmHg for the ZEROed channel, contact your technical
services department.
GE Specs For Zero Balance
Range
Accuracy
Drift
+/- 150 Mmhg
+/- 1 mmHg
+/- 1 mmHg / 24 hrs
8.
Page 19 of 34
Transfer Between Monitors
[10/04/06 aa. title means transferring from one monitor to another? suggested title: changing
monitors or switching monitors]
To redirect a wireless link from one patient monitor do the following:
1. Press and release the Next Button.
2. Follow the procedure for linking (power-on cycle)
NOTE: THIS PROCEDURES REQUIRES A SECOND PATIENT MONITOR CONFIGURED WITH
HOSPIRA ERECEIVER MODULES. SEE SETTING UP RECEIVER MODULES
CALIBRATING MULTIPLE MONITORS
1. Calibrate one monitor at a time.
This section goes in the setup portion.
Wireless Transpac System Operations Manual 430-11431-100 (09/06)
Warnings and Precautions
WARNINGS
•
Explosion Hazard – In accordance with UL 60601-1, that applies to all medical equipment,
do not use this device in the presence of flammable anesthetics.
•
In order for the Wireless Transducer to work properly, the P1 and P2 receiver modules
must be connected to their respective transmitter sockets.
•
The Wireless Transducer should only be connected as illustrated in the System Set Up
section of this manual and only to those monitors validated for compatibility with the
Wireless Transducer. Refer to the Cautions section of this manual for a list of validated
monitors for the current list of validated monitors.
•
Only use the approved Wireless Transducer Power Supply cord (for Europe - Yunhuan
Group Y003/ST3, for U.S.A. - Yunhuan Group, YY-3/ST3) in order to ensure that the
Wireless Transducer continues to comply with risk (leakage) current and EMI/EMC
requirements.
•
To ensure patient safety, only use Wireless Transducer interfaces recommended by
Hospira. For a complete list of approved interfaces, see System Components and
Accessories sections of this manual.
•
Normal operation of the Wireless Transducer should be verified in the configuration in
which it will be used before engaged in use with a patient. Inspect and test the batteries,
and power supply frequently. Do not operate the Wireless Transducer if any system
component appears damaged or appears not to function properly. See System Functional
Check section of this manual for the procedure to verify normal operation.
•
If the Battery LED flashes you have 60 minutes to connect to AC before the system will
power-off.
•
To ensure patient safety care should be taken to prevent physical contact between
electrosurgical equipment and any Wireless Transducer component. the conductive parts
of the Wireless Transducer should not contact any other conductive parts, including earth
ground, at any time.
•
This device should not be used in a Magnetic Resonance Imaging (MRI) suite. Strong
magnetic fields may affect the device, causing injury to the patient and/or damage to the
equipment.
•
The Wireless Transducer can operate in the presence of pacemaker pulses. The Wireless
Transducer may miss pacer pulses if high background noise is present. The Wireless
Transducer captures signal spikes regardless of whether spikes are due to noise or pacer
pulses. If the Wireless Transducer triggers frequently from background noise due to motion
artifact, EMI, etc., then pacer pulses could be masked or missed by the Wireless
Transducer.
•
The Wireless Transducer should not be used for external pacing.
•
Impedance pneumography detects respiratory effort via changes in chest volume.
Impedance pneumography can be used to detect central apnea, but apnea episodes with
continued respiratory effort, such as obstructive apnea and mixed apnea, may go
undetected. Artifacts due to patient motion, apnea mattress shaking or use of
electrocautery equipment may also cause apnea episodes to go undetected. Impedance
pneumography is not recommended for patients with pacemakers.formation
Page 21 of 34
FCC INTERFERENCE WARNING
•
This device complies with Part 15 of the FCC Rules. Operation is subject to the following
two conditions:
(1) This device may not cause harmful interference.
(2) This device must accept any interference received, including interference that may
cause undesired operation.
•
This equipment has been approved for mobile applications where the equipment should be
used at distances greater than 20cm from the human body (with the exception of hands,
wrists, feet and ankles). Operation at distances less than 20 cm is strictly prohibited.
•
Changes or modifications not expressly approved by Elcam – Medical for compliance could
void the user's authority to operate the equipment
•
The antenna(s) used for this transmitter must not be co-located or operating in conjunction
with any other antenna or transmitter.
CAUTIONS
•
The batteries should be removed when the system is not in use for extended periods of
time.
•
Remove the plastic film cover from the battery contacts before using.
•
Do not autoclave Wireless Transducer components. Do not submerge Wireless
Transducer components in any liquid.
•
The Wireless Transducer transmitter and receiver have been tested and shown to comply
with FCC Part 15. Any changes or modifications to the Wireless Transducer not expressly
approved by Hospira could cause Wireless Transducer emissions to exceed those
permitted under FCC rules, and would thus void the user’s authority to operate the
equipment.
•
If operating under conditions according to the EMC-standard 60601-1-2 (Radiated
Immunity 3V/m), field strengths above
1.5 V/m may cause erroneous measurements of respiration at various frequencies when
the Wireless Transducer is used together with the Welch Allyn Propaq Series Monitor
Systems. Therefore, if it is of critical and clinical importance to monitor respiration, it is
recommended to avoid the use of electrically radiating equipment (such as electrocautery
surgical pencils, and high powered portable and mobile RF communications equipment)
while the Wireless Transducer is being used with the Propaq Monitor. However, if
respiration must be monitored in the presence of high frequency electrically radiated
equipment, always monitor and set alarms for SpO2 when using the Wireless Transducer
together with the Propaq Monitor to monitor respiration.
•
Only use the Wireless Transducer with the following equipment that has been validated for
compatibility:
Bedside Cardiac Monitors
Hewlett Packard Agilent Viridia Series
Portable/Transportable Cardiac Monitors
Welch Allyn Monitoring Propaq Encore Series
Note: Validation testing for additional equipment is ongoing. Please contact Hospira for the
current list of compatible equipment
•
Battery replacement is recommended after one year of use.
•
The Transmitter can be operated as a Class II device per EN 60601-1 or internally
powered.
•
•
This device has been tested and certified to comply with emissions portion of EN 60601-1-2,
Medical Electrical Equipment – ElectroMagnetic Compatibility – Requirements and Tests.
Although this device is shielded against ElectroMagnetic Interference (EMI), it is
recommended that electrically radiating devices not be used in close proximity to this
device.
Portable and mobile RF communications equipment can affect the operation of the Wireless
Transducer.
•
Operation of the Wireless Transducer below a minimum amplitude of patient physiological
signal may cause inaccurate results.
Wireless Transpac System Operations Manual 430-11431-100 (09/06)
Page 23 of 34
Chapter 5
Alarms and Troubleshooting
ALARMS
TRANSDUCER ALARMS
Table 5-1: Transducer Alarm Table
LED and Tone
Indicators
Cause
Action
Blinks for 1 minute
1 beep ever 10
seconds
-Power up failure
-Battery empty
Restart on AC
Lights up for 5
seconds
1 beep every 5
seconds
Communication errors
exceed 3 in 10 seconds
-- Eliminate interference
-- Move closer to monitor
3 blinks every 5
seconds
Battery charge failure
Restart on AC
Blinks once
Power On
Transducer Connected
Beeps once
-Power Off
-Start up
-Greeting
-Active
-Idle after greeting
-Steady blinking and
beeping for 2
minutes
-Square wave on
monitor for 30
seconds
Wireless link lost
Move closer to monitor
1 blink per second
No transducers
connected ???
Confirm connection
-Beeps once 10
seconds after action
A successful
transducer connection
BATTERY LEDS
Table 5-2: Battery LED Table
LED and Tone
Indicators
Blinks and beeps 3
times sequentially
-- consecutively??
Cause
-Less than 60 minutes
of batter life
Action
Connect to AC
-No power
-Off -- Charging only
6 blinks per second
Skipping one monitor
to another
3 blinks per second
New connection
3 LED lights on
Battery at full charge
2 LED lights on
Battery at halfcharge
Connect to AC
1 LED light on
Battery at low charge
Connect to AC
3 beeps repeat
Less the 60 minutes
of battery life left
No LED lights on
60 minutes passed
Connect to AC
transmitter shut
down
Power on
Powering off
Connect to AC
No LED lights on
2 beeps
Power on
RECEIVER ALARMS
Table 5-3: Receiver Alarm Table
LED and Tone
Indicators
Cause
Steady blinking
Wireless linking in
process
1 blink per second
Active State
Monitor passed
greeting cycle
Starts receiving
sensor data
Page 25 of 34
Action
Steady beeping
Improper installation
1 blink
Sensor connected
Contact site technical
services department
TROUBLESHOOTING
Table 5-4:
Symptom
Transmitter Won’t Turn
On
Cause
Solution
Battery dead
• Charge battery, or
replace with fully
charged battery
Shipping tape over
battery contacts
• Remove tape,
charge battery
Cable connector not
fully inserted in
Transpac Pigtail socket
• Reposition and
insert Connector
correctly
Transceiver
continuously restarts
Unit needs servicing
• Call Clinical
Coordinator
Service Light On
Unit needs servicing
• Call Clinical
Coordinator
Can’t establish link
• Token not fully
inserted
• Re-sync units with
token, BEEP
confirms sync
• Patient Transceiver or
Monitor Transceiver not
turned ON
• Ensure both units
are turned ON
• Out of range
• System range
exceeded –move
patient back within
30 feet of Monitor
Transceiver
• 3/5 cable system not
connected
• Connect 3/5 cable
system as required
Periodic beeping sound
heard
• Low battery
• Re-charge battery,
or replace with fully
charged battery
Battery does not last 24
hours
• Monitor Transceiver
runs 12 hours per
battery charge
• Re-charge Monitor
Transceiver battery,
or replace with fully
charged battery
Wireless Transducer
not charging battery or
Wireless Transducer
battery charge indicator
not lit
Intermittent or dropped
link ()
Excessive noise
• Battery not fully
charged at start
• Ensure battery is
fully charged at
installation
• Battery at end of life,
or faulty battery
• Replace battery –
Recycle or dispose of
old batteries as
directed in Wireless
Transducer User’s
Manual
• Power Supply not
plugged in
• Plug in Power
Supply
• Faulty AC power
(110V) outlet
• Call electrician
• Power Supply failure
• Replace Power
Supply – Only use
Wireless Transducer
Power Supply
• Battery contacts
covered
• Uncover contacts
and charge battery
• Out of range, or at
edge of range
• System range
exceeded –move
patient back within
30 feet of Monitor
Transceiver
• Signal is blocked
• Try another unit
• Receiver module too
close to AC power
source
• Separate receiver
module and AC
power source cable
by 1 to 2 feet
• Poor serial cable
connection
• Reseat cable
connector
• Out of range
• System range
exceeded –
Table 5-4:
Page 27 of 34
Wireless Transpac System Operations Manual 430-11431-100 (09/06)
Chapter 6
Specifications
Parameter
Specification
Number of Channels,
single-ended, wrt RL
Channels Active
Input Dynamic Range
ECG Signal Slew Rate
Input Impedance
Leadoff Sense Current
Frequency Response
Method A per EC11:1991
Triangle Response
Method D per EC13:2002
Line Filter (60 Hz)
Ch-Ch Signal noise
Multi-channel Crosstalk
ECG Signal Gain wrt RL
Ch-Ch Gain Difference
CM Rejection
Leadoff
• Leadoff sensed by Patient
Transceiver
• Leadoff simulated by Monitor
Transceiver
DC Offset – any channel
Pacer Pulse Detection,
Transmission, and
Reconstruction
Pacer Pulse Function Trigger
Slew Rate
Signal Latency due to radio
Operating Battery
Operating Battery Life
Size
Weight (includes battery)
Rated Voltage/Current
Operating Temperature
Storage Temperature
Atmospheric Pressure
Housing Material
Radio Protocol
Class
Range
Operating Frequency
Channels
Power output level
PT
MT
Bluetooth Revision
Water Ingress Rating
Patient Applied Parts type
Defibrillation Proof
Recovery time after defibrillation
exposure
Safety
EMC
Page 29 of 34
Wireless Transpac System Operations Manual 430-11431-100 (09/06)
Chapter 7
Standards
This chapter contains a list of the applicable standards that the Wireless Transmitter/Receiver
device meets.
APPLICABLE STANDARDS
The wireless transmitter/receiver device shall comply with the following standards or documents.
The compliance shall be in accordance with the Product Performance Specification for Hospira
Transpac® IV Disposable Blood Pressure Transducer as applicable.
added new from Nichol 10/02/06 300 or 301?
•
ISO 13485:2003(E), Medical devices - Quality management systems - Requirements for
regulatory purposes
•
ISO/IEC 14971:2000, Medical Devices - Application of Risk Management to Medical
Devices
•
IEC/EN/UL 60601-1 Third Edition Medical Electrical Equipment - General requirements for
basic safety and essential performance
•
IEC/EN60601-1-1 Medical electrical equipment - Part 1-1: General requirements for safety
- Collateral standard: Safety requirements for medical electrical systems
•
EC/EN 60601-1-2 Edition 2.1 General requirements for safety - Collateral standard:
Electromagnetic compatibility - Requirements and tests
•
IEC/EN 60601-1-4 Medical electrical equipment - General requirements for safety Collateral Standard: Programmable electrical medical systems
•
EC/EN 60601-1-49 Part 2-49: Particular requirements for the safety of multifunction patient
monitoring equipment
•
IEC/EN60601-2-34 Medical electrical equipment: Particular requirements for the safety,
including essential performance, of invasive blood pressure monitoring equipment, section
44 Overflow, spillage, leakage, humidity, ingress of liquids, cleaning, sterilization,
disinfection and compatibility
•
IEC 245 or IEC 227 for flexible cables and cords
•
UL 94 flammability testing
•
IEC/EN 60529 Degrees of protection provided by enclosures (IP Code) consolidated
edition
•
EN 60950 Information technology equipment - Safety - Part 1: General requirements
•
EN 55011 Measuring Radiated Emission
•
IEC 61000-3-2 Electromagnetic compatibility (EMC) - Part 3-2: Limits - Limits for harmonic
current emissions (equipment input current <= 16 A per phase)
•
IEC 61000-3-3 Electromagnetic compatibility (EMC) - Part 3-3: Limits - Limitation of voltage
changes, voltage fluctuations and flicker in public low-voltage supply systems, for
equipment with rated current <= 16 A per phase and not subject to conditional connection
•
IEC 61000-4-2 Electromagnetic compatibility (EMC)- Part 4-2: Testing and measurement
techniques - Electrostatic discharge immunity test Consolidated Edition
•
IEC 61000-4-3 Electromagnetic compatibility (EMC) - Part 4-3 : Testing and measurement
techniques - Radiated, radio-frequency, electromagnetic field immunity test
•
IEC 61000-4-4 Electromagnetic compatibility (EMC) - Part 4-4: Testing and measurement
techniques - Electrical fast transient/burst immunity test
•
IEC 61000-4-5 Electromagnetic compatibility (EMC) - Part 4-5: Testing and measurement
techniques - Surge immunity test
•
IEC 61000-4-8 Electromagnetic compatibility (EMC) - Part 4-8: Testing and measurement
techniques - Power frequency magnetic field immunity test
•
IEC 61000-4-11 Electromagnetic compatibility (EMC) - Part 4-11: Testing and
measurement techniques - Voltage dips, short interruptions and voltage variations
immunity tests
•
ETSI EN 300 328-1 Title: ElectroMagnetic Compatibility and Radio Spectrum Matters
(ERM);Wideband Transmission systems; data transmission equipment operating in the 2,4
GHz ISM band and using spread spectrum modulation techniques; Part 1: Technical
characteristics and test conditions
•
ETSI EN 300 328-2 Electromagnetic compatibility and Radio Spectrum Matters
(ERM);Wideband Transmission systems; Data transmission equipment operating in the 2,4
GHz ISM band and using spread spectrum modulation techniques; Part 2: Harmonized EN
covering essential requirements under article 3.2 of the R&TTE Directive.
•
ETSI EN 300 489-1 Terminal Equipment (TE);Conformance testing for file transfer over the
Integrated Services Digital Network (ISDN);Part 1: Profile Test Specification Summary
(PTS-Summary) for the FTAM profile (ETS 300 388)
•
ETSI EN 300 489-17 (The PRD lists this, however I am not able to locate the title) maybe its
1 or 7.
•
ETSI EN 300 489-2 Terminal Equipment (TE);Conformance testing for file transfer over the
Integrated Services Digital Network (ISDN);Part 2: Profile Specific Test Specification
(PSTS) for the FTAM profile (ETS 300 388)
•
FCC Bulletin OET 65C Evaluating Compliance With FCC Guidelines for Human Exposure
to Radiofrequency Electromagnetic Fields
•
802.15.1 Bluetooth class 2 Wireless Communications Standard as applicable
•
EN60601-1 for safety
•
UL60601-1 for safety
•
EN60601-1-1 for safety (collateral)
•
IEC 245 or IEC 227 for power cords
•
UL 94 for flammability
•
IEC529 for enclosure protection
•
EN60950 for instructions for use
•
EN 60601-1-4 for software
•
EN60601-1-2 for EMC
•
EN60601-1-2-34 section 44 for spillage and ingress of liquids
•
EN55011 for Conducted / Radiated emission.
•
IEC 61000-3-2 for Harmonic emissions.
•
IEC 61000-3-3 for Voltage fluctuations/flicker emissions
•
IEC61000-4-2 for ESD
•
IEC61000-4-3 for radiated fields
•
IEC61000-4-4 for fast transient/burst
•
IEC61000-4-5 for surges
•
IEC61000-4-8 for power frequency magnetic fields
•
IEC61000-4-11 for voltage dips
•
EN55011 for emissions
•
ETSI EN 300 328-1 for radio
•
ETSI EN 300 328-2 for radio
•
ETSI EN 301 489-1 for radio
•
ETSI EN 301 489-17 for radio
•
FCC Bulletin OET 65C for RF
Page 31 of 34
•
802.15.1, Bluetooth class 2 Wireless Communications Standard as applicable.
System Specifications
Parameter
Specification
Number of Channels, single-ended,
wrt RL
Channels Active
Input Dynamic Range
ECG Signal Slew Rate
Input Impedance
Leadoff Sense Current
Frequency Response
Method A per EC11:1991
Triangle Response
Method D per EC13:2002
Line Filter (60 Hz)
Ch-Ch Signal noise
Multi-channel Crosstalk
ECG Signal Gain wrt RL
Ch-Ch Gain Difference
CM Rejection
Leadoff
• Leadoff sensed by Patient
Transceiver
• Leadoff simulated by Monitor
Transceiver
DC Offset – any channel
Pacer Pulse Detection,
Transmission, and Reconstruction
Pacer Pulse Function Trigger Slew
Rate
Signal Latency due to radio
Operating Battery
Operating Battery Life
Size
Weight (includes battery)
Rated Voltage/Current
Operating Temperature
Storage Temperature
Atmospheric Pressure
Housing Material
Radio Protocol
Class
Range
Operating Frequency
Channels
Power output level
PT
MT
Bluetooth Revision
Water Ingress Rating
Patient Applied Parts type
Defibrillation Proof
Recovery time after defibrillation
exposure
Safety
EMC
Page 33 of 34
Wireless Transpac System Operations Manual 430-11431-100 (09/06)
Chapter 8
Warranty
Subject to the terms and conditions herein, Hospira, Inc., herein referred to as Hospira, warrants
that (a) the product shall conform to Hospira’s standard specifications and be free from defects in
material and workmanship under normal use and service for a period of one year after purchase,
and (b) the replaceable battery shall be free from defects in material and workmanship under
normal use and service for a period of 90 days after purchase. Hospira makes no other warranties,
express or implied, as to merchantability, fitness for a particular purpose, or any other matter.
Purchaser's exclusive remedy shall be, at Hospira’s option, the repair or replacement of the product.
In no event shall Hospira’s liability arising out of any cause whatsoever (whether such cause be
based in contract, negligence, strict liability, other tort, or otherwise) exceed the price of such
product, and in no event shall Hospira be liable for incidental, consequential, or special damages or
losses or for lost business, revenues, or profits. Warranty product returned to Hospira must be
properly packaged and sent freight prepaid.
The foregoing warranty shall be void in the event the product has been misused, damaged, altered,
or used other than in accordance with product manuals so as, in Hospira’s judgment, to affect its
stability or reliability, or in the event the serial number or lot number has been altered, effaced, or
removed.
The foregoing warranty shall also be void in the event any person, including the Purchaser,
performs or attempts to perform any major repair or other service on the product without having
been trained by an authorized representative of Hospira and using Hospira documentation and
approved spare parts. For purposes of the preceding sentence, "major repair or other service"
means any repair or service other than the replacement of accessory items such as batteries and
detachable mains power cords.
In providing any parts for repair or service of the product, Hospira shall have no responsibility or
liability for the actions or inactions of the person performing such repair or service, regardless of
whether such person has been trained to perform such repair or service. It is understood and
acknowledged that any person other than a Hospira representative performing repair or service is
not an authorized agent of Hospira.

Source Exif Data:
File Type                       : PDF
File Type Extension             : pdf
MIME Type                       : application/pdf
PDF Version                     : 1.3
Linearized                      : No
Modify Date                     : 2007:01:15 13:17:26+02:00
Create Date                     : 2007:01:15 13:17:01+02:00
Page Count                      : 34
Creation Date                   : 2007:01:15 11:17:01Z
Mod Date                        : 2007:01:15 11:17:01Z
Producer                        : Acrobat Distiller 5.0 (Windows)
Author                          : Hospira User
Metadata Date                   : 2007:01:15 11:17:01Z
Creator                         : Hospira User
Title                           : Wireless Transmitter/Receiver
Page Mode                       : UseNone
Tagged PDF                      : Yes
EXIF Metadata provided by EXIF.tools
FCC ID Filing: UPA13EAUPAWRLSYST

Navigation menu