Fluke Electronics FT400 Gemini User Manual

Fluke Electronics Gemini

User Manual

PN 4281773
September 2013
© 2013 Fluke Corporation. All rights reserved.
Specifications are subject to change without notice.
All product names are trademarks of their respective companies.
Ti200, Ti300, Ti400
Thermal Imagers
Users Manual
LIMITED WARRANTY AND LIMITATION OF LIABILITY
Each Fluke product is warranted to be free from defects in material and workmanship under normal
use and service. The warranty period is two years and begins on the date of shipment. Parts, product
repairs, and services are warranted for 90 days. This warranty extends only to the original buyer or
end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries, or
to any product which, in Fluke's opinion, has been misused, altered, neglected, contaminated, or
damaged by accident or abnormal conditions of operation or handling. Fluke warrants that software
will operate substantially in accordance with its functional specifications for 90 days and that it has
been properly recorded on non-defective media. Fluke does not warrant that software will be error free
or operate without interruption.
Fluke authorized resellers shall extend this warranty on new and unused products to end-user
customers only but have no authority to extend a greater or different warranty on behalf of Fluke.
Warranty support is available only if product is purchased through a Fluke authorized sales outlet or
Buyer has paid the applicable international price. Fluke reserves the right to invoice Buyer for
importation costs of repair/replacement parts when product purchased in one country is submitted for
repair in another country.
Fluke's warranty obligation is limited, at Fluke's option, to refund of the purchase price, free of charge
repair, or replacement of a defective product which is returned to a Fluke authorized service center
within the warranty period.
To obtain warranty service, contact your nearest Fluke authorized service center to obtain return
authorization information, then send the product to that service center, with a description of the
difficulty, postage and insurance prepaid (FOB Destination). Fluke assumes no risk for damage in
transit. Following warranty repair, the product will be returned to Buyer, transportation prepaid (FOB
Destination). If Fluke determines that failure was caused by neglect, misuse, contamination, alteration,
accident, or abnormal condition of operation or handling, including overvoltage failures caused by use
outside the product’s specified rating, or normal wear and tear of mechanical components, Fluke will
provide an estimate of repair costs and obtain authorization before commencing the work. Following
repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for
the repair and return transportation charges (FOB Shipping Point).
THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER
WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED
WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE
SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR CONSEQUENTIAL
DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion
or limitation of incidental or consequential damages, the limitations and exclusions of this warranty
may not apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a
court or other decision-maker of competent jurisdiction, such holding will not affect the validity or
enforceability of any other provision.
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
11/99
To register your product online, visit http://register.fluke.com.
i
Table of Contents
Title Page
Introduction ................................................................................ 1
How to Contact Fluke ................................................................. 2
Safety Information ...................................................................... 2
Radio Frequency Data ............................................................... 5
Accessories ................................................................................ 5
Before You Start ......................................................................... 7
Battery .................................................................................... 8
Two-Bay Battery Charger Base........................................... 8
On-Imager AC Power Socket .............................................. 9
Optional 12 V Vehicle Charger ............................................ 10
Power On and Off ................................................................... 10
Operation in Extreme Conditions ............................................... 10
Features and Controls ................................................................ 11
Primary and Secondary Triggers ................................................ 13
Focus ......................................................................................... 13
Control Buttons .......................................................................... 14
Touch Screen ............................................................................. 14
How to Use the Menus ............................................................... 15
Image Capture ........................................................................... 15
IR-PhotoNotes™..................................................................... 16
Voice Annotation (Recording) ................................................. 17
Listen to a Voice Annotation (Recording) ............................... 17
Edit Captured Infrared Image ................................................. 18
Save Captured Infrared Image ............................................... 18
Micro SD Memory Card .............................................................. 19
Temperature Measurement ........................................................ 20
SmartView
Software ................................................................. 20
Ti200, Ti300, Ti400
Users Manual
ii
Menus ......................................................................................... 21
Measurement Menu ................................................................ 21
Range .................................................................................. 21
Level/Span .......................................................................... 21
Emissivity Adjustment .......................................................... 23
Background (Reflected Background
Temperature Compensation) ............................................... 25
Transmission/Transmittance Adjustment ............................. 25
Spot Temperatures .............................................................. 26
Spot Markers ....................................................................... 26
Center Box........................................................................... 27
Image Menu ............................................................................ 28
Palette ................................................................................. 28
IR-Fusion® Technology ........................................................ 29
Color Alarms ........................................................................ 30
Display Graphics Presentation ............................................ 31
Logo .................................................................................... 31
Camera Menu ......................................................................... 32
LaserSharp™ Auto Focus System ...................................... 32
Backlight .............................................................................. 33
Torch ................................................................................... 33
Memory Menu ......................................................................... 34
Review Image Files ............................................................. 34
Delete Image Files ............................................................... 34
Settings Menu ......................................................................... 35
Units .................................................................................... 35
File Format .......................................................................... 35
Auto Off ............................................................................... 36
Date ..................................................................................... 37
Language ............................................................................ 38
Wireless Connectivity .......................................................... 38
Image Storage ..................................................................... 38
Advanced Settings ............................................................... 39
Streaming Video ......................................................................... 40
HDMI
......................................................................................... 40
Remote Control (Ti400) .............................................................. 41
Maintenance ............................................................................... 41
How to Clean the Case ........................................................... 41
Battery Care ............................................................................ 42
General Specifications ................................................................ 43
Detailed Specifications ............................................................... 45
iii
List of Tables
Table Title Page
1. Symbols ..................................................................................... 4
2. Accessories ................................................................................ 5
3. Features and Controls ................................................................ 11
4. Overview of Controls .................................................................. 14
5. Palettes ...................................................................................... 28
List of Figures
Figure Title Page
1. Lens Cover Laser Warning ......................................................... 3
2. Add-On Lens Installation ............................................................ 6
3. SD Memory Card Insertion and Removal ................................... 19
4. Level and Span Settings ............................................................ 23
Ti200, Ti300, Ti400
Users Manual
iv
1
Introduction
The Fluke Ti200, Ti300, and Ti400 Thermal Imagers (the Product or Imager)
are handheld, infrared imaging cameras for use in many applications. These
applications include equipment troubleshooting, preventive and predictive
maintenance, building diagnostics, and research and development.
All Imagers display thermal images on a high-visibility, industrial-quality
640 x 480 LCD touch screen and can save images to a removable memory
card. Saved images and data can be transferred through the memory card to a
PC, a direct USB connection to the PC, or by wireless transfer to a PC or
mobile device.
The Imager includes SmartView® software. This software is a high-
performance, professional software suite that allows for quality analysis and
reporting. SmartView® Mobile software is also available for use on mobile
devices.
A rugged, rechargeable lithium-ion smart battery provides power to the Imager.
Direct AC power is accessible with the included AC power adapter.
All Imagers have the Fluke LaserSharp Auto Focus System, as well as
advanced manual focus capabilities. They also include many useful features
and functions associated with IR-Fusion® technology with Autoblend™ mode,
IR-PhotoNotes™ photo annotation system, WiFi™ connectivity, Bluetooth
connectivity, and HDMI video connectivity.
Ti200, Ti300, Ti400
Users Manual
2
How to Contact Fluke
To contact Fluke, call one of the following telephone numbers:
USA: 1-800-760-4523
Canada: 1-800-36-FLUKE (1-800-363-5853)
Europe: +31 402-675-200
Japan: +81-3-6714-3114
Singapore: +65-6799-5566
Anywhere in the world: +1-425-446-5500
Or, visit Fluke's website at www.fluke.com.
To register your Product, visit http://register.fluke.com.
To view, print, or download the latest manual supplement, visit
http://us.fluke.com/usen/support/manuals.
Safety Information
A Warning identifies conditions and procedures that are dangerous to the user.
A Caution identifies conditions and procedures that can cause damage to the
Product or the equipment under test.
 Warning
To prevent eye damage and personal injury:
Do not look into the laser. Do not point laser directly at
persons or animals or indirectly off reflective surfaces.
Do not open the Product. The laser beam is dangerous to
eyes. Have the Product repaired only through an approved
technical site.
Thermal Imagers
Safety Information
3
Additional laser warning information is on the inside of the Product lens cover,
see Figure 1.
www.patentlabel.com/uke
FLUKE CORPORATION
EVERETT, WA USA
hie05.eps
Figure 1. Lens Cover Laser Warning
Warning
To prevent personal injury:
Read all safety information before you use the Product.
Carefully read all instructions.
Use the Product only as specified, or the protection supplied
by the Product can be compromised.
Replace the batteries when the low battery indicator shows
to prevent incorrect measurements.
Do not use the Product if it operates incorrectly.
Do not use the Product if it is damaged.
Ti200, Ti300, Ti400
Users Manual
4
See emissivity information for actual temperatures.
Reflective objects result in lower than actual temperature
measurements. These objects pose a burn hazard.
Do not use the Product around explosive gas, vapor, or in
damp or wet environments.
Do not put battery cells and battery packs near heat or fire.
Do not put in sunlight.
Table 1 is a list of symbols used on the Imager and in this manual.
Table 1. Symbols
Symbol Description Symbol Description
Important information.
See manual. Warning. Laser.
Connected to ac power.
Battery removed. Battery status. Battery
charging when animated.
On/Off Symbol. Conforms to requirements of
European Union and European
Free Trade Association.
Conforms to relevant
Australian standards. Conforms to relevant North
American Safety Standards.
Conforms to relevant
South Korean EMC
standards.
Japan Quality Association
This Product contains a lithium-ion battery. Do not mix with the
solid waste stream. Spent batteries should be disposed of by a
qualified recycler or hazardous materials handler per local
regulations. Contact your authorized Fluke Service Center for
recycling information.
This product complies with the WEEE Directive (2002/96/EC)
marking requirements. The affixed label indicates that you must not
discard this electrical/electronic product in domestic household
waste. Product Category: With reference to the equipment types in
the WEEE Directive Annex I, this product is classed as category 9
“Monitoring and Control Instrumentation” product. Do not dispose
of this product as unsorted municipal waste. Go to Fluke's website
for recycling information.
Thermal Imagers
Radio Frequency Data
5
Radio Frequency Data
See page 39 for instructions on how to access digital copies of the radio
licenses on the Imager.
Accessories
Table 2 is a list of the accessories available for the Imager.
Table 2. Accessories
Model Description PN
FLK-TI-SBP3 Smart Battery Pack 3440365
FLK-TI-SBC3B Charging Base/Power Supply with Adapters 4354922
TI-CAR CHARGER 12 V Vehicle Charger Adapter 3039779
FLK-TI-VISOR3 Sun Visor 4335377
FLK-TI-TRIPOD3 Tripod Mounting Accessory 4335389
BOOK-ITP Introduction to Thermography Principles 3413459
FLK-LENS/TELE2 Infrared Telephoto Lens 4335377
FLK-LENS/WIDE2 Infrared Wide-Angle Lens 4335361
Optional telephoto and wide-angle lenses expand the flexibility and allow more
applications for your infrared inspection work. See Figure 2 for information
about how to install a lens.
Ti200, Ti300, Ti400
Users Manual
6
hie08.eps
Figure 2. Add-On Lens Installation
Thermal Imagers
Before You Start
7
Before You Start
Carefully unpack the items in the shipment box:
Thermal Imager
AC Power Supply with Mains Adapters
Two-Bay Battery Charge Base
2 Lithium-ion Smart Batteries
Hard Carrying Case
USB Cable
HDMI Cable
Removable Memory Card[1]
Soft Transport Bag
Adjustable Hand Strap (Left-hand or Right-hand use)
Users Manuals in print (English, Spanish, French, German, Chinese)[2]
Users Manuals on CD[2]
SmartView® Software
Warranty Registration Card
Note
[1] Fluke recommends the memory card that is supplied with the
Imager or available from Fluke. Fluke does not warrant the use or
reliability of aftermarket memory cards of different brands or
capacities.
[2] To request a printed manual of a language not supplied with your
product, email Fluke at TPubs@fluke.com. Specify the product name
and language preference in the subject line.
Ti200, Ti300, Ti400
Users Manual
8
Battery
Before you use the Imager for the first time, charge the battery for a minimum
of two and one-half hours. The battery status shows on the five-segment
charge indicator.
Warning
To prevent personal injury, do not put battery cells and battery
packs near heat or fire. Do not put in sunlight.
Note
New batteries are not fully charged. Two to ten charge/discharge
cycles are necessary before the battery charges to its maximum
capacity.
To charge the battery, use one of the options that follow:
Two-Bay Battery Charger Base
1. Connect the ac power supply to the ac wall outlet and connect the dc
output to the charger base.
2. Put one or two smart batteries into bays of charger base.
3. Charge batteries until charge indicators show “full.”
4. Remove smart batteries and disconnect the power supply when batteries
are fully charged.
Thermal Imagers
Before You Start
9
On-Imager AC Power Socket
1. Connect the ac power adapter into an AC wall outlet and connect the dc
output to the Imager’s ac power socket. flashes on the display while
the battery charges with the ac power adapter.
2. Charge until the charge indicator on the display does not flash.
3. Disconnect ac power adapter when the smart battery is fully charged.
Note
Make sure that the Imager is near room temperature before you
connect it to the charger. See the charging temperature specification.
Do not charge in hot or cold areas. When you charge in extreme
temperatures, battery capacity may be decreased.
shows in the lower left-hand corner of the display when the Imager is
connected to ac power and the battery is removed. When the Imager’s power is
off and the ac power adapter is connected, flashes in the center of the
display to show that the battery charge is in process.
Keep the Imager attached to the charger until the battery condition icon shows
a full charge. If you remove the Imager from the charger before a full charge
shows, it may have a reduced run-time.
Note
When the battery is connected to ac power or the unit is in video
mode, the Sleep Mode/Auto Off feature is disabled automatically.
Ti200, Ti300, Ti400
Users Manual
10
Optional 12 V Vehicle Charger
1. Connect the 12 V adapter into the 12 V accessory socket of the vehicle.
2. Connect the output to the ac power socket of the Imager.
3. Charge until the indicator shows full on the screen.
4. Disconnect the 12 V adapter and Imager when battery is fully charged.
Caution
To prevent damage to the Imager, remove it from the DC car
charger before you start or jump start the vehicle.
Power On and Off
To turn on or turn off the Imager, push and hold for two seconds, see
Table . The Imager has Power Save and Auto Off features. See page 36 for
more information about how to set these features.
Note
All thermal imagers need sufficient warm-up time for the most
accurate temperature measurements and best image quality. This
time can often vary by model and by environmental conditions.
Although most imagers are fully warmed up in 3-5 minutes, it is always
best to wait a minimum of 10 minutes if the most accurate temperature
measurement is very important to your application. When you move
an Imager between environments with large differences in ambient
temperature, more adjustment time can be required.
Operation in Extreme Conditions
Storage and/or continual operation of the Imager in extreme ambient
temperature conditions can result in temporary interruption of operation. If this
occurs, let the Imager stabilize (cool down or warm up) before you resume
operation.
Thermal Imagers
Features and Controls
11
Features and Controls
Table 3 shows the Imager features and controls.
Table 3. Features and Controls
Ti200
IR FUSION
TECHNOLOGY
11
10
9
6
7
8
4
5
12
13
3
1
2
hie01.eps
Item Description
HDMI Connection
USB Cable Connection
Microphone
Speaker
LCD Display
Power On/Off
Function Buttons (F1, F2, and F3)
Arrow Buttons
Lithium-ion Smart Battery
Ti200, Ti300, Ti400
Users Manual
12
Table 3. Features and Controls (cont.)
Item Description
Hand Strap
Manual Focus Control
2-Bay Battery Charging Base
AC Power Supply with Mains Adapters
Ti2 00
IR FUSION
TECHN OLOGY
19
20
21
22
23
24
18
16
15
17
14
hie06.eps
Connector Cover
AC Adapter/Charger Input Terminal
Removable Micro SD Memory Card Slot
Micro USB Cable Connection
LED Torch/Flashlight
Retractable Lens Cover
Visual Light Camera Lens
Laser Pointer/Range Finder
Infrared Camera Lens
Secondary Trigger
Primary Trigger
Thermal Imagers
Primary and Secondary Triggers
13
Primary and Secondary Triggers
The two-part trigger is located in the standard trigger position for a pistol-grip
device. The larger, green trigger is the primary trigger. The smaller, black
trigger is the secondary trigger.
In normal operation, the function of the primary trigger is to capture a thermal
image for possible storage to memory by the user.
The secondary trigger operates the LaserSharp™ Auto Focus System and
laser pointer. See page 32 for information about how to enable and disable the
LaserSharp™ Auto Focus System.
Focus
The Imager uses the LaserSharp Auto Focus System and an advanced manual
focus system. Correct focus is important in all imaging applications. Correct
focus makes sure that the infrared energy is correctly directed onto the pixels of
the detector. Without correct focus, the thermal image can be blurry and the
radiometric data may be inaccurate. Out-of-focus infrared images are frequently
unusable or of little value.
The LaserSharp Auto Focus System, only from Fluke, uses a precision-
adjusted and aligned laser pointer to indicate the area of focus. To focus with
the LaserSharp Auto Focus System, depress the secondary (black) trigger on
the Imager handle. Aim the laser pointer at the target. When satisfied with the
location of the laser pointer, release the trigger. The system automatically,
rapidly, and reliably focuses on the precise point where the laser was aimed.
Note
The user can enable and disable the LaserSharp Auto Focus
System.
The laser pointer is aligned parallel to the infrared lens. If using
AutoBlend Mode, the laser pointer dot will always be positioned
just above the center point marker on the LCD display. However,
it is easier to use your eyes to locate the visible laser pointer on
the object under inspection.
To focus with the advanced manual focus system, place finger on the
electronically-driven focus wheel control and rotate the control until the
inspection object is in proper focus.
Ti200, Ti300, Ti400
Users Manual
14
Note
The user is able to use the advanced manual focus as a manual
override of the LaserSharp™ Auto Focus System.
Control Buttons
Three function buttons ( ) and four arrow buttons
( and ) are the primary controls. These buttons move the
cursor through the menu structure to set the features.
Table 4 is an overview of the buttons and their actions. In live Manual Mode,
the arrow buttons are always active to adjust Level and Span.
In general, push:
to set the change and go back to the live view.
to set the change and go back to the previous menu.
to cancel the change and go back to the live view.
Table 4. Overview of Controls
Button Button Label / Action
or Primary Trigger Cancel
or Primary Trigger Done (exit from Menu structure)
Select or OK
Back
Move cursor to highlight an option
Touch Screen
The touch screen enables you to interact directly with what is displayed. To
change parameters or select function and options, touch a target on the display
with a finger. Touch targets are easy to recognize, such as arrows to indicate a
drop-down list for a selection of options.
The touch screen has a backlight for work in dimly lit spaces. To change the
backlight intensity, see page 33.
Thermal Imagers
How to Use the Menus
15
How to Use the Menus
The menus, coupled with the three function buttons (, , ) and
arrow buttons, are the access for thermal image display, camera features,
measurement, advanced functions, memory review, and settings for date, time,
language, units, file format, and Imager information.
To open the primary menu, push . The primary menu shows five secondary
menus for Measurement, Image, Camera, Memory, and Settings. The text
above each function button (, , ) applies to that button throughout
all menu screens.
Push to open the primary menu and push to cycle
through the secondary menus. Each secondary menu lists an options menu.
Push to cycle through the options.
The primary, secondary, and option menus close 10 seconds after the last push
of a function button. The option selection menu stays open until you make the
selection, go up a menu level, or cancel the action.
Image Capture
Point the imager at the object or area of interest. Focus on the object using the
LaserSharp™ Auto Focus System or advanced manual focus (See page 13 on
how to focus). Pull and release the primary trigger. This will capture and freeze
the image. The image is now in the memory buffer and you can choose to save
or edit the image. Push to save the image or pull the primary trigger again
or to cancel and return to the live view.
Depending on the selected file format settings, the Imager shows the captured
image and a menu bar. The menu bar lets you save the image, edit some
image settings, and add voice annotation or IR-PhotoNotes™ digital photos. To
change the file format, see File Format on page 35.
Ti200, Ti300, Ti400
Users Manual
16
IR-PhotoNotes™
The IR-PhotoNotes™ photo annotation system allows the user to capture and
add up to five visible (digital) images of various objects, text, or other
information that is related to the analysis and reporting of an infrared image.
Examples of possible annotations include motor name plates, printed
information or warning signs, larger views of the environment or room, and
related equipment or objects. Up to five images can be captured with the visible
image that is stored in addition to the aligned infrared and visible images used
in IR-Fusion® technology. These visible images are only available in the .is2 file
format and are stored in the file so you do not need to collate multiple files at a
later time.
To add photos using the IR-PhotoNotes annotation system:
1. With an infrared image in the buffer, push to open the EDIT IMAGE
menu.
2. Push / to highlight IR-PhotoNotes.
3. Push to enter the Picture mode.
4. Focus the Imager on the object and pull the primary trigger.
5. Push when done.
6. Pull the primary trigger to capture additional pictures.
7. Push save the pictures with the image.
Thermal Imagers
Image Capture
17
i
Voice Annotation (Recording)
To add a voice (audio) recording:
1. With an infrared image in the buffer, push to open the EDIT IMAGE
menu.
2. Push / to highlight Add Audio.
3. Push to record up to 60 seconds of audio. The display updates to
show the recorded time.
4. Push to pause the recorder.
5. Push when done.
6. Push to review the audio file or to save the audio with the image.
Voice annotation is only available in the .is2 file format and is stored in the file
so you do not need to collate multiple files at a later time.
Listen to a Voice Annotation (Recording)
The voice (audio) recording replays through the speaker.
To play back:
1. Do the steps in the Review Image Files section on page 34 to see the
image on the display.
2. Push .
3. Push to set Audio.
4. Push to listen to the audio.
5. Push again to pause the audio.
Ti200, Ti300, Ti400
Users Manual
18
Edit Captured Infrared Image
Before saving a file, you can edit or modify the image.
To edit:
1. With an image in the buffer, go to EDIT IMAGE > Edit Image.
2. Push / to highlight an option.
3. Push to save the changes with the file.
Save Captured Infrared Image
To save an image as a data file:
1. Focus on the object of interest or inspection area.
2. Pull the trigger to capture the image. The image is now in the buffer and
you can save or edit.
3. Push to save the image as a file and go back to the live view.
Thermal Imagers
Micro SD Memory Card
19
Micro SD Memory Card
To eject a Micro SD memory card, push in on the exposed edge of the card and
then release. The card should pop partially out after you release it. Carefully
pull the card out of the slot.
To use the Micro SD memory card, make sure that the write-protect lock is
open. See Figure 3. Push the card in until it catches.
The Micro SD memory card comes with an SD adapter for insertion into a PC
or multi-function card reader if desired.
hie03.eps
Figure 3. Micro SD Memory Card Insertion and Removal
For information about how to save data, see page 18. For information about
how to view or erase a stored image, see page 34.
Ti200, Ti300, Ti400
Users Manual
20
Temperature Measurement
All objects radiate infrared energy. The quantity of energy radiated is based on
the actual surface temperature and the surface emissivity of the object. The
Imager senses the infrared energy from the surface of the object and uses this
data to calculate an estimated temperature value. Many common objects and
materials such as painted metal, wood, water, skin, and cloth are very good at
radiating energy and it is easy to get relatively accurate measurements. For
surfaces that are good at radiating energy (high emissivity), the emissivity
factor is 90 % (or 0.90). This simplification does not work well on shiny
surfaces or unpainted metals as they have an emissivity of <0.60. These
materials are not good at radiating energy and are classified as low emissivity.
To more accurately measure materials with a low emissivity, an emissivity
correction is necessary. Adjustment to the emissivity setting will usually allow
the Imager to calculate a more accurate estimate of the actual temperature.
Warning
To prevent personal injury, see emissivity information for actual
temperatures. Reflective objects result in lower than actual
temperature measurements. These objects pose a burn hazard.
SmartView Software
SmartView® software is supplied with the Imager. This software is intended for
Fluke Imagers and contains features to analyze images, organize data and
information, and make professional reports. SmartView® allows audio
annotations and photos from the IR-PhotoNotes™ annotation system to be
reviewed on a PC. SmartView is used to export IR and visible images as .jpeg,
.jpg, .jpe, .jfif, .bmp, .gif, .dip, .png, .tif, or .tiff formatted files.
SmartView Mobile Software is also available for flexibility away from your PC or
in the field.
Thermal Imagers
Menus
21
Menus
The menus, together with the three function buttons (, , ) and
arrow buttons, are access for thermal image display, camera features, memory
setup, and settings for date, time, language, units, file format, and Imager
information.
Measurement Menu
The Measurement Menu has settings for the calculation and display of
radiometric temperature measurement data related to the thermal images.
These settings include the Temperature Range selection, Level/Span
adjustment, Emissivity, Background, Transmission, Spot Temperatures,
Markers, and Center Box.
Range
Preset measurement ranges and a Full Automatic Range are available on the
Imager. To select the range:
1. Go to Measurement > Range.
2. Push / to select between the preset ranges or the full automatic
range.
3. Push to set the range.
Level/Span
Level and span are set for automatic or manual adjustment. To choose
between automatic or manual level and span:
1. Go to Measurement > Set Level/Span.
2. Push / to select between the Auto and Manual ranging or
Set Level/Span.
3. Push to set the new selection.
In manual mode the Set Level/Span control is active. Use / to adjust
the level setting. Use / to adjust the span setting. See the section,
Detailed Specifications, for more information about the minimum span.
Ti200, Ti300, Ti400
Users Manual
22
Fast Auto/Manual Range Toggle
When NOT in a menu mode, push and release to toggle between Auto
Range and Manual Range. The icon in the upper right-hand corner of the LCD
screen will indicate the change by saying Auto or Manual.
Fast Auto Rescale (in Manual Mode)
When in Manual Range and NOT in a menu mode, push and release to
automatically rescale the level and span range for objects in the thermal field of
view. This feature operates the Imager in a semi-automatic mode if manual fine
re-adjustment of level and span with the arrow buttons is not necessary.
Rescaling can be done as often, or as little, as needed.
Note
The Imager always powers up in the same Level and Span mode,
Auto or Manual, as when it was powered down.
Temperature Level for Manual Operation Mode
When put into manual ranging, the level setting moves the thermal span up or
down within the total temperature range. See Figure 4. In the live manual
mode, the arrow buttons are always available to adjust the level and span.
To set the level:
1. Push to move the range to a higher temperature level.
2. Push to move the range to a lower temperature level.
While you adjust the manual level, the scale along the right side of the display
shows the thermal span as it moves to different levels within the total range.
Thermal Imagers
Menus
23
Total Imager Range
SpanLevel
hie07.eps
Figure 4. Level and Span Settings
Temperature Span for Manual Operation Mode
When in manual mode, the span setting contracts or expands in a selected
palette in a temperature range within the total range. See Figure 4. In the live
manual mode, the arrow buttons are always available to adjust the level and
span.
To adjust the temperature span:
1. Push to increase or widen the temperature span.
2. Push to decrease or narrow the temperature span.
While you adjust the manual span, the scale along the right side of the display
shows the thermal span increasing or decreasing in size.
Emissivity Adjustment
The correct emissivity values are important for the Imager to make the most
accurate temperature measurement calculations. Emissivity of a surface can
have a large effect on the apparent temperatures that the Imager observes.
Understanding the emissivity of the surface being inspected can, but not
always, allow you to obtain more accurate temperature measurements.
Ti200, Ti300, Ti400
Users Manual
24
If you set a value that is <0.60, shows on the Imager display with this
caution:
Caution:
Emissivity <0.6
(see Manual)
Note
Surfaces with an emissivity of <0.60 make reliable and consistent
determination of actual temperatures problematic. The lower the
emissivity, the more potential error is associated with the Imager's
temperature measurement calculations. This is also true even when
adjustments to the emissivity and reflected background adjustments
are performed properly.
Emissivity is set directly as a value, or from a list of emissivity values for some
common materials.
Note
If the Display is set to Display All, you see the information about
current emissivity as ε = x.xx.
Adjust by Number
To set the emissivity value:
1. Go to Measurement > Emissivity > Adjust Number.
2. Push / to change the value.
A custom emissivity value is indicated when any value not in the standard
emissivity table is selected.
Select by Table
To select from a list of common materials:
1. Go to Measurement > Emissivity > Select Table.
2. Push / to highlight the material.
3. Push to select the material.
Thermal Imagers
Menus
25
Background (Reflected Background Temperature
Compensation)
Compensation for reflected background temperature is set in the Background
tab. Very hot objects or very cold objects can affect the apparent temperature
and measurement accuracy of the target or object of interest, especially when
surface emissivity is low. Adjustment of the reflected background temperature
can make the temperature measurement better in many situations. For more
information, see Emissivity Adjustment.
To adjust the background temperature:
1. Go to Measurement > Background.
2. Push / to change the value.
3. Push or when done.
Note
If the Display is set to Display All, you see the information about
current reflected background temperature as BG = xx.x.
Transmission/Transmittance Adjustment
When you do infrared inspections through infrared-transparent windows
(IR windows), not all of the infrared energy emitted from the objects of interest
is transmitted through the optical material in the window. If the transmission
percentage of the window is known, you can adjust this percentage in the
Imager or in the SmartView® software. Adjustment of the transmission
correction can make the accuracy of the temperature measurement better in
many situations.
To adjust the transmission percentage:
1. Go to Measurement > Transmission.
2. Push / to adjust the percentage between 10 % and 100 %.
3. Push or when done.
Note
If Display Information is set to Display All, you see the information
about current transmission correction as τ = xx.
Ti200, Ti300, Ti400
Users Manual
26
Spot Temperatures
The Spot Temperatures are floating HI and LO temperature indicators that
move on the display as the temperature measurements of the image fluctuate.
To turn on/off the hot and cold spot indicators:
1. Go to Measurement > Spot Temp.
2. Push / to highlight ON or OFF.
3. Push or to set the new value.
Note
If the Display is set to Display All, you see the information about
current transmission correction as τ = xxx%.
Spot Markers
Up to three adjustable, fixed-temperature spot markers are available on the
display. You can use these markers to highlight a region before you save the
image. The marker selection is set as All Off, One Marker, Two Markers, or
Three Markers.
To set a Marker:
1. Go to Measurement > Markers.
2. Push the / to highlight the function between All OFF, One
Marker, Two Markers, and Three Markers.
3. Push to set the marker option and go to the “Move Marker” display.
You will see the Move Marker icon and the labels on the function buttons
change to Done, Next, and Cancel.
To change the Marker position on the display:
1. Push to move the Marker location on the image.
2. Push to highlight the next marker. Do Step 1 again.
3. Do Step 2 for a third marker.
4. Push when done.
Thermal Imagers
Menus
27
Center Box
The Center Box feature is an adjustable temperature measurement zone (box)
that you can center on the infrared image. This zone (box) expands and
contracts to different levels within the infrared image. The zone lets the user
see an approximate maximum (MAX), average (AVG), and minimum (MIN)
temperature measurement in that area. When in AUTO Level and Span mode,
the Imager automatically sets the level and span according the infrared scene
within the parameters of the Center Box.
Note
When using the Center Box, the level and span of the Imager is
adjusted to the thermal scene within the Center Box.
To enable or disable the Center Box feature:
1. Go to Measurement > Center Box.
2. Push / to highlight ON or OFF.
3. Push to set the new value.
To set the size of the Center Box when enabled:
1. Push / to highlight Set Size.
2. Push to go to the adjustment screen.
3. Push to increase the size of the Center Box.
4. Push to reduce the size of the Center Box.
5. When satisfied with the size of the Center Box, push:
to set the change and exit the menus.
to set the change and go back to the previous menu.
to cancel the change and go back to the live view.
Ti200, Ti300, Ti400
Users Manual
28
Image Menu
The Image menu has controls for different features used in the presentation of
the infrared image on the Imager's LCD and some saved image files.
Note
Data saved as .is2 or .is3 formats can easily be modified within
SmartView software. Still images saved in .bmp or .jpg format, as well
as video saved in .avi format will retain image settings at the time of
capture and save.
Palette
The Palette menu lets you change the false-color presentation of the infrared
images on display or captured. Some palettes are more suitable for specific
applications and can be set as required. Two different palette presentation
modes are available, see Table 5. The Standard Palettes offer an equal, linear
presentation of colors that allow for best presentation of detail. The Ultra
Contrast™ Palettes offer a weighted presentation of colors. These palettes
work best in situations with high thermal contrast for extra color contrast
between the high temperatures and low temperatures.
Table 5. Palettes
Standard Palettes Ultra Contrast™ Palettes
Grayscale Grayscale Ultra
Grayscale Inverted Grayscale Inverted Ultra
Blue-Red Blue-Red Ultra
High Contrast High Contrast Ultra
Hot Metal Hot Metal Ultra
Ironbow Ironbow Ultra
Amber Amber Ultra
Amber Inverted Amber Inverted Ultra
Thermal Imagers
Menus
29
To switch between palettes:
1. Go to Image > Palette > Standard or Ultra Contrast.
2. Push / to highlight a palette.
3. Push to set the new palette.
To change the palette color:
1. Go to Image > Palette > Set Palette.
2. Push / to highlight the palette color.
3. Push to set the new palette color.
Saturation Colors is an option that you can choose as off, standard, red/blue, or
white/black.
IR-Fusion® Technology
IR-Fusion® technology makes it easier to understand, analyze, and
communicate infrared images through the use of an aligned visible image and
infrared image. The Imager automatically captures a visible image with every
infrared image to show you precisely where a potential problem might be, and
then allows you to more effectively communicate it to others.
IR-Fusion® technology has different modes:
Picture-in-Picture (PIP) AutoBlend (min, mid, max)
Full AutoBlend (min, mid, max)
Visible
Note
The visible image and infrared image can be customized or separated
in SmartView and SmartView Mobile Software when you use the .is2
file format.
To set the IR-Fusion® technology mode:
1. Go to Measurement > Image > IR-Fusion.
2. Push / to highlight an option.
3. Push to set the new option.
Ti200, Ti300, Ti400
Users Manual
30
Color Alarms
The Imager has various apparent temperature color alarms. The high-
temperature color alarm shows a full visible image and only shows infrared
information on objects or areas that are above the set apparent temperature
alarm level. The low-temperature (or dew point) color alarm shows a full visible
image and only shows infrared information on objects or areas that are below
the set apparent temperature (or set dew point) color alarm level. The user
must manually determine and set these parameters. The imager also displays
color isotherms, or infrared information, inside or outside of a set of both high
and low limits.
Note
The Imager does not sense ambient or surface dew point level
automatically. To use the low-temperature color alarm function as a
dew point color alarm, manual determination and input of surface dew
point temperature will yield the best results. Depending on the
situation, the colors presented may help identify areas of concern with
possible dew point condensation.
To view the Color Alarm menu:
1. Go to Image > Color Alarm.
2. Push to view the menu.
Set High-Temperature Color Alarm
To set a high-temperature color alarm:
1. Go to Image > Color Alarm > Set High Alarm.
2. Push / to adjust the temperature setting.
3. Push or set the new value.
Set Low-Temperature (or Dew Point) Color Alarm
To set a low-temperature/dew point color alarm:
1. Go to Image > Color Alarm > Set Low Alarm.
2. Push / to adjust the temperature setting.
3. Push or set a new value.
Thermal Imagers
Menus
31
Outside/Inside Alarm
If you set values for the high-temperature color alarm and a low-temperature
color alarm, the Imager will have the options for inside or outside isotherm color
alarms.
To set an outside/inside isotherm color alarm:
1. Go to Image > Color Alarm > Outside or Inside.
2. Push:
to set the change and go back to the live view.
to set the change and go back to the previous menu.
to cancel the change and go back to the live view.
Display Graphics Presentation
The options for how you view the on-screen graphics are in the Display menu.
These options are Display All, Details and Scale, Scale Only, and Image Only.
To set the display:
1. Go to Image > Display.
2. Push / to highlight an option.
3. Push or set a new option.
Note
Features that have ON/OFF controls must be turned on and turned off
with those controls.
Logo
A Fluke logo shows on the display and captured images. You can choose to
turn on or turn off the logo:
1. Go to Image > Logo.
2. Push / to highlight on or off.
3. Push to set.
With SmartView software you can upload a custom logo to the Imager from
your PC through the USB connection.
Ti200, Ti300, Ti400
Users Manual
32
Camera Menu
The Camera menu has controls and options for secondary camera features
such as auto focus, backlight level, and torch.
LaserSharp Auto Focus System
The laser pointer on the Imager is both a sighting aid and a part of the
LaserSharp™ Auto Focus System. In addition, the advanced manual focus of
the Imager works both when the auto focus system is enabled or disabled.
 Warning
To prevent eye damage and personal injury, do not look into the
laser. Do not point laser directly at persons or animals or
indirectly off reflective surfaces.
The laser warning symbol () shows in the Header zone of the display when
you pull the secondary trigger.
To turn on or turn off the LaserSharp Auto Focus System and the laser pointer:
1. Go to Camera > Auto Focus.
2. Push / to select on or off.
3. to set the change and go back to the live view.
When the laser pointer is turned on, the secondary trigger is the control for the
LaserSharp Auto Focus System. Point the Imager in the general direction of the
object of interest. Pull and hold the secondary trigger to locate the laser dot on
the specific area of interest. Release the trigger. The auto focus system will
quickly and accurately focus on the object.
Thermal Imagers
Menus
33
Backlight
The backlight level control is set to low, medium, and high. To set the backlight:
1. Go to Camera > Backlight.
2. Push / to highlight high, medium, or low.
3. Push to set a new value.
Torch
The torch illuminates darker work areas.
To set:
1. Go to Camera > Torch.
2. Push to toggle the torch between on and off.
Ti200, Ti300, Ti400
Users Manual
34
Memory Menu
The Memory Menu allows the user to review captured images, as well as audio
annotations and photos from the IR-PhotoNotes™ photo annotation system in a
thumbnail view format. An icon shows to indicate any additional items saved
with the infrared image or IR-Fusion technology image:
IR-PhotoNotes photos
Voice annotation
Review Image Files
To view stored images on the memory card:
1. Go to Memory.
2. Push to highlight the thumbnail of the file for review.
3. Push to review the file.
Delete Image Files
To erase one image from the memory card:
1. Go to Memory.
2. Push to highlight the thumbnail of the file to delete.
3. Push to open the Delete menu.
4. Highlight Selected Image and push . The Imager prompts you to
continue or cancel.
5. Push again to delete the file.
To erase all the images from memory:
1. Go to Memory.
2. Push .
3. Highlight All Images and push . The Imager prompts you to continue
or cancel.
4. Push to delete all files in memory.
Thermal Imagers
Menus
35
Settings Menu
The Settings menu has adjustments for user preferences such as units of
temperature measurement, file format of stored data, “save to” location choice,
auto off settings, WiFi and Bluetooth settings, date, time, and language. This
menu also has a section that displays information about the Imager such as
model number, serial number, and firmware versions. Certificates and licenses
are available from this menu.
Units
To change the temperature units:
1. Go to Settings > Units.
2. Push / to highlight an option.
3. Push to set an option.
File Format
Data can be saved to the internal memory, micro SD memory card, or a USB
flash memory device in different file formats. Image format selections are .bmp,
.jpg, and .is2. These selections remain valid when you turn the Imager off or on.
To change the file format:
1. Go to Settings > File Format.
2. Push / to highlight an option.
3. Push to set the option.
Images saved in the .is2 file format have the consolidation of all data into a
single file and are more flexible for analysis and modification in the included
SmartView software. This file format consolidates the infrared image,
radiometric temperature data, visible image, voice annotation, and photos from
the IR-PhotoNotes™ photo annotation system into one location.
Ti200, Ti300, Ti400
Users Manual
36
For situations where a smaller file size with maximum resolution is needed and
modification is not, choose the .bmp file format. For the smallest file size where
modification is not needed and image quality and resolution are not as
important, choose the .jpg file format.
The .bmp and .jpg files can be emailed and then opened on most PC and MAC
systems without special software. These formats do not allow full analysis
capabilities or modification.
The .is2 file format can be emailed and then opened with SmartView Software.
This format has the maximum versatility. Visit the Fluke website or contact
Fluke to find out how to download SmartView analysis and reporting software at
no charge.
Auto Off
The Auto Off timer is user-defined separately for the LCD and power.
Note
Auto Off is automatically disabled when the battery is connected to AC
Power.
To set the Auto Off feature:
1. Go to Settings > Auto Off.
2. Push  to highlight LCD Time Out or Power Off.
3. Push / to set the timer between 1 minute and 120 minutes.
4. Push to set.
Thermal Imagers
Menus
37
Date
The date can be displayed in one of two formats: MM/DD/YY or DD/MM/YY.
To set the date:
1. Go to Settings > Date.
2. Push / to highlight the date format.
3. Push to set a new format.
4. Push / to highlight Set Date.
5. Push to open the Set Date menu.
6. Push  to select highlight day, month, or year.
7. Push to change the settings.
8. Push to set the date and exit the menu.
To set the time:
1. Go to Settings > Time.
Time displays in two different formats: 24 hour or 12 hour. To set the time
format:
2. Push / to highlight time format.
3. Push to select.
4. Highlight Set Time.
5. Push to open the Set Time menu.
6. Push  to highlight hours or minutes.
The 12 hour format has a selection to set the time as AM or PM.
7. Push or to change the setting.
8. Push to set the change.
Ti200, Ti300, Ti400
Users Manual
38
Language
To change the display to a different language:
1. Go to Settings > Language.
2. Push or to highlight the setting.
3. Push to set a new language.
Wireless Connectivity
Your Imager is equipped with several wireless connectivity options to enhance
your ability to work more efficiently and better communicate results.
WiFi™ Hotspot
You can wirelessly send a picture from the Imager to a PC, an iPhone, and an
iPad through the WiFi connection. A transferred image can be viewed with
SmartView Mobile or SmartView Analysis and Reporting software if installed in
the device.
Bluetooth®
Bluetooth is available to connect a device such as a wireless headset. When
on, shows on the display (upper left corner).
Image Storage
The storage setting allows you to choose to save images to the internal
memory, micro SD memory card, or USB flash memory.
1. Go to Settings > Image Storage.
2. Push or to change the setting.
3. Push to select the new storage setting.
Thermal Imagers
Menus
39
Advanced Settings
Filename Prefix
The default filename starts with IR_. You have the option to change this prefix
to a different 3-character name with the touch screen keyboard.
Reset Filename
You can reset the file number to 00001.
Factory Defaults
Erases all user-set preferences and restores all of the factory default settings.
Imager Information
You can access information about the version, certifications, and licenses for
the Imager from the Settings Menu.
To display the Imager Info:
1. Go to Settings > Advanced > Imager Info.
2. Push / to highlight Version.
3. Push to view the information screen with model number, serial
numbers, and firmware versions.
4. Push to close the information screen.
To display the electronic certifications:
1. Go to Settings > Advanced > Imager Info.
2. Push / to highlight Certificates.
3. Push to view the information screen with the Image certifications.
4. Push to close the information screen.
Ti200, Ti300, Ti400
Users Manual
40
To display the license information:
1. Go to Settings > Advanced > Imager Info.
2. Push / to highlight Licenses.
3. Push to view the information screen with a list of Open Source
Software Licenses.
4. Push / to scroll to a specific license.
5. Touch the license name on the screen to view the information screen with
the specific license agreement.
6. Push to close the information screen.
Streaming Video
The Imager can stream live infrared and IR-Fusion™ technology video to a PC
that has SmartView® software installed.
To set up:
1. Attach the included USB cable to the USB port on the imager.
2. Connect the other end to a USB port on a PC.
3. In SmartView software, choose Remote Display from the drop down menu
to begin live video streaming on the PC.
HDMI
HDMI (High-Definition Multimedia Interface) is a compact audio/video interface
for transferring uncompressed data and compressed/uncompressed digital
audio data from the Imager to a compatible HDMI device.
To set up:
1. Attach the included HDMI cable to the HDMI port on the Imager.
2. Connect the other end to an HDMI video device.
Thermal Imagers
Remote Control (Ti400)
41
Remote Control (Ti400)
The Ti400 Imager can be remotely controlled with a PC that has SmartView
software installed.
To remotely control the Imager:
1. If not installed, install the latest version of SmartView software on your PC.
2. Attach the included USB cable to the USB port on the Imager.
3. Connect the other end of the cable to a USB port on the PC. appears
in the SmartView workspace.
4. Select Remote Display from the drop down menu.
5. Use the buttons and touch screen on the UI the same as the physical
Imager.
6. Follow the instructions in the software help files for additional information
about the controls on the software UI.
Maintenance
The Imager does not require maintenance.
 Warning
To prevent eye damage and personal injury, do not open the Product.
The laser beam is dangerous to eyes. Have the Product repaired only
through an approved technical site.
How to Clean the Case
Clean the case with a damp cloth and a weak soap solution. Do not use
abrasives, isopropyl alcohol, or solvents to clean the case or lens/window.
Ti200, Ti300, Ti400
Users Manual
42
Battery Care
Warning
To prevent personal injury and for safe operation of the Product:
Do not put battery cells and battery packs near heat or fire. Do not
put in sunlight.
Do not disassemble or crush battery cells and battery packs.
Remove batteries to prevent battery leakage and damage to the
Product if it is not used for an extended period.
Connect the battery charger to the mains power outlet before the
charger.
Use only Fluke approved power adapters to charge the battery.
Keep cells and battery packs clean and dry. Clean dirty
connectors with a dry, clean cloth.
Caution
To prevent damage, do not expose Product to heat sources or
high-temperature environments such as an unattended vehicle in
the sun.
To get the best performance from the lithium-ion battery, use the guidelines that
follow:
Do not store the Imager on the charger for more than 24 hours as reduced
battery life may result.
Charge the Imager for a two-hour minimum at six-month intervals for
maximum battery life. Without use, the battery will self-discharge in
approximately six months. Batteries stored for long periods will need two to
ten charging cycles for full capacity.
Always operate in the specified temperature range.
Do not store the batteries in extreme cold environments.
Do not attempt to charge the batteries in extreme cold environments.
Caution
Do not incinerate the Product and/or battery. Go to Fluke’s
website for recycling information.
Thermal Imagers
General Specifications
43
General Specifications
Temperature
Operating ............................................................ -10 °C to 50 °C (14 °F to 122 °F)
Storage ............................................................... -20 °C to 50 °C (-4 °F to 122 °F)
without batteries
Relative Humidity .................................................. 10 % to 95 % non-condensing
Altitude
Operating ............................................................ 2,000 m
Storage ............................................................... 12,000 m
Display ................................................................... 8.9 cm (3.5 in) diagonal landscape
color VGA (640 x 480) LCD with
backlight
Controls and Adjustments
User-selectable Temperature Scale (°C/°F)
Language Selection
Time and Date Settings
Emissivity Selection
Reflected background temperature compensation
Transmission correction
User selectable hot spot and cold spot, and center point on the image
Expand/Contract Measurement Box with MIN-AVG-MAX temp High temperature alarm
Color alarms
User selectable backlight setting
Graphic information display (selectable)
Software ................................................................. SmartView® full analysis and reporting
software included
Power
Batteries .............................................................. 2 Lithium-ion rechargeable smart
battery packs with 5-segment LED
display to show charge level.
Battery Life .......................................................... 4+ hours continuous use for each
battery pack (assumes 50 %
brightness of LCD)
Battery Charge Time ........................................... 2.5 hours to full charge
AC Battery Charge .............................................. Ti SBC3B Two Bay Battery Charger
(110 V ac to 220 V ac, 50/60 Hz,
included), or in-Imager charging. Ac
mains adapters included. Optional
12 V automotive charging adapter.
Ti200, Ti300, Ti400
Users Manual
44
AC Operation ...................................................... AC operation with included power
supply: 110 – 220 Vac, 50/60 Hz, ac
mains adapters included
Power Save ......................................................... User-selectable Sleep and Power Off
modes
Safety Standards
CE Directive ........................................................ IEC/EN 61010-1 2nd Edition Pollution
Degree 2
CAN/CSA ............................................................ C22.2 No. 61010-1-04, UL STD
61010-1 (2nd Edition)
ISA ....................................................................... 82.02.01
Electromagnetic Compatibility
(EMI, RFI, EMC, RF) ............................................... Meets all applicable requirements in
EN61326-1:Industrial
FCC Part 15 Subpart C Sections
15.207, 15.209, 15.249
Applies to use in Korea only ................................ Class B Equipment (For Home Use
Broadcasting & Communication
Equipment) [1]
[1] This equipment is home use (Class B)
electromagnetic wave suitability equipment
that is to be used mainly at home and can
be used in all areas.
Vibration ................................................................. 0.03 g2/Hz (3.8 gm), 2.5 gm,
IEC 68-2-6
Shock ...................................................................... 25 G, IEC 68-2-29
Drop ........................................................................ 2 meter (with standard lens)
Size (H x W x L) ...................................................... 27.7 cm x 12.2 cm x 16.7 cm
(10.9 in x 4.8 in x 6.5 in)
Weight (includes battery)...................................... 1.04 kg (2.3 lb)
Enclosure Rating ................................................... IP54
Warranty ................................................................. 2 years
Calibration Cycle ................................................... 2 years (assumes normal operation
and normal aging)
Supported Languages........................................... Czech, Dutch, English, Finnish, French,
German, Hungarian, Italian, Japanese,
Korean, Polish, Portuguese, Russian,
Simplified Chinese, Spanish, Swedish,
Traditional Chinese, and Turkish
Thermal Imagers
Detailed Specifications
45
Detailed Specifications
Temperature Measurements
Temperature Range (not calibrated below -10 °C)
Ti200, Ti300 .................................................... -20 °C to +650 °C
Ti400 ............................................................... -20 °C to +1200 °C
Accuracy ............................................................. ±2 °C or 2 % (whichever is greater) at
25 °C ambient
On-screen emissivity correction.......................... all models
On-screen reflected background
temperature compensation ................................. all models
On-screen transmission correction ..................... all models
Imaging Performance
Image Capture Frequency .................................. 9 Hz or 60 Hz refresh rate depending
on model variation
Detector Type: Focal Plane Array, uncooled microbolometer
Ti400 ............................................................... 320 X 240
Ti300 ............................................................... 240 X 180
Ti200 ............................................................... 200 X 150
Thermal Sensitivity (NETD)
Ti400, Ti300 .................................................... 0.05 °C at 30 °C target temp (50 mK)
Ti200 ............................................................... 0.075 °C at 30 °C target temp (75 mK)
Total pixels
Ti400 ............................................................... 76,800
Ti300 ............................................................... 43,200
Ti200 ............................................................... 30,000
Infrared spectral band ......................................... 7.5 μm to 14 μm (long wave)
Visual (Visible Light) Camera
Type .................................................................... Industrial performance 5.0 megapixel
Minimum parallax alignment
with standard IR lens .......................................... ~60 cm (~24 in)
Standard Infrared Lens
Field of View ....................................................... 24 ° x 17 °
Spatial Resolution (IFOV)
Ti400 ............................................................... 1.31 mRad
Ti300 ............................................................... 1.75 mRad
Ti200 ............................................................... 2.09 mRad
Minimum Focus Distance ................................... 15 cm (6 in)
Ti200, Ti300, Ti400
Users Manual
46
Optional telephoto infrared lens
Field of View ........................................................ 12 ° x 9 °
Spatial Resolution (IFOV)
Ti400 ............................................................... 0.65 mRad
Ti300 ............................................................... 0.87 mRad
Ti200 ............................................................... 1.05 mRad
Minimum Focus Distance .................................... 45 cm (18 in)
Optional wide-angle infrared lens
Field of View ........................................................ 46 ° x 34 °
Spatial Resolution (IFOV)
Ti400 ............................................................... 2.62 mRad
Ti300 ............................................................... 3.49 mRad
Ti200 ............................................................... 4.19 mRad
Minimum Focus Distance .................................... 15 cm (6 in)
Focus Mechanism
LaserSharp™ Auto Focus System ...................... all models
Advanced Manual Focus ..................................... all models
Image Presentation
Standard Palettes ................................................ Blue-Red, Grayscale, Inverted
Grayscale, High Contrast, Amber,
Inverted Amber, Hot Metal, Ironbow
Ultra Contrast™ Palettes .................................... Blue-Red Ultra, Grayscale Ultra,
Inverted Grayscale Ultra, High
Contrast Ultra, Amber Ultra, Inverted
Amber Ultra, Hot Metal Ultra, Ironbow
Ultra
Level and Span
Smooth Auto-Scaling and Manual scaling of level and span
Fast auto toggle between manual and auto modes
Fast auto rescale in manual mode
Minimum Span (in manual mode) ................... 2.0 °C (3.6 °F)
Minimum span (in auto mode) ......................... 3.0 °C (5.4 °F)
IR-Fusion® Technology
Automatically aligned (parallax corrected) visual and infrared
Picture-In-Picture (PIP) infrared
Full screen infrared
AutoBlend mode
Full screen visible
Color alarms (temperature alarms) High temperature, low temperature, and isotherm
(user selectable)
Thermal Imagers
Detailed Specifications
47
Image and Data Storage
Image Capture, Review, Save Mechanism ........ One-handed image capture, review,
and save capability
Storage Medium
Micro SD Memory Card .................................. includes 4 GB memory card to store at
least 2000 fully radiometric (.is2) IR
and linked visual images each with 60
seconds voice annotations or 5000
basic (.bmp or .jpg) images
Internal on-board flash memory ...................... 4 GB available for storage
USB ................................................................ direct download with USB-to-PC
connection
Note
The addition of IR-PhotoNotes or other saved items can vary the total number
of images that can be stored on the SD memory card.
File Formats ........................................................ Non-Radiometric (.bmp, .jpg) or Fully-
Radiometric (.is2)
No analysis software required for Non-
Radiometric (.bmp, .jpg) files
Export File Formats with
SmartView® Software ......................................... JPEG, JPG, JPE, JFIF, BMP, GIF,
DIP, PNG, TIF, TIFF
Memory Review .................................................. Thumbnail view navigation and review
selection
Productivity Features
Voice Annotation (60 seconds maximum recording time for each image.
Reviewable playback on Imager.)
IR-PhotoNotes™
WIFI connectivity
Streaming video
Remote control and operation
Ti200, Ti300, Ti400
Users Manual
48

Navigation menu