Fluke 289 Application Note

2015-09-09

: Fluke Fluke-289-Application-Note-805139 fluke-289-application-note-805139 fluke pdf

Open the PDF directly: View PDF PDF.
Page Count: 3

Application Note
Troubleshooting electrical
problems in high-end TVs
Using logging and TrendCapture in the Fluke 289 DMM
From the Fluke Digital Library @ www.fluke.com/library
These troubleshooting steps feature the Fluke 289 DMM
because of its onscreen trending capabilities. TrendCapture
helps troubleshooters spot anomalies faster by showing results
on-screen as soon as recording sessions are complete. Follow
correct safety procedures for testing live circuits. Use a safety-
rated, fused digital multimeter (DMM) and test a known voltage
source first to make sure your meter is operating properly.
1. Testing over voltage or supply overload shut-
down in televisions:
Most televisions sets that use CRT’s have shut-
down circuits that turn off the TV when the high
voltage goes too high. These same sets also
have circuits that detect excessive supply voltage
or load currents for several low voltage supplies.
It is often a real challenge to find an intermittent
shutdown problem because most TV’s do not
have built in circuits that indicate which supply
is intermittently overloaded or has intermittent
excessive voltage.
You can use the Fluke 289 Digital Multimeter
(DMM) to find the power supply that is caus-
ing the shutdown problem. Connect the meter
across the HV/deflection supply, monitor the
voltage level, and record the voltage level over
a period of time. You can also set the meter in
MIN-MAX mode, wait for the shutdown to occur,
and observe the maximum voltage. If you don’t
see an over voltage problem, test another supply.
For some supplies, you may be able to access
the current sense shunt, and then you can easily
connect the meter across the resistor to monitor
the current in MIN-MAX mode.
2. Testing for cause of repeated failure of con-
vergence amplifiers in projection televisions.
Projection television sets that use three projec-
tion CRT’s have circuits that dynamically modify
the deflection of the electron beams in the three
tubes to converge the three colors displayed on
the screen. The convergence is done by driving
secondary deflection yokes with hybrid power
amplifiers. These amplifiers are usually pack-
aged two or three in a hybrid assembly.
The amplifiers are fairly reliable, but occasion-
ally there may be a set where the amplifiers
fail in less than a year, and the replacement
amplifier fails in six to nine months. Using a 289
DMM and a temperature probe, you can conduct
a simple extra check to determine if a repair
is really complete and the TV will not have a
repeating failure.
After a hybrid amplifier assembly has been
replaced, connect the temperature probe to the
amplifier assembly and heat sink (at the junc-
tion of the body of the amplifier assembly and
2 Fluke Corporation Troubleshooting electrical problems in high-end TVs
the heat sink). Set up the 289 DMM to record
the temperature at ten second intervals and start
recording. Turn on the TV and allow the set to
warm up to its normal operating temperature,
about 20 to 30 minutes. After about 15 minutes,
the temperature should be around 45 °C to
55 °C. Figure 1 shows the screen shots of the
temperature recordings for a normal amplifier
and an amplifier that is running too hot.
3. Troubleshooting intermittent turn-on
problem in projection television.
An older projection TV had a problem where it
intermittently would turn on and have sound,
but no picture. If you turned off the set, waited
about two minutes, then turned the set back on,
you could get a picture about one time in three
tries. The problem was that one of the switching
power supplies did not start up every time the
set was turned on.
Using the 289 DMM, the problem was traced to
the startup circuit where there was a defective
electrolytic capacitor. The DMM was connected
across the supply for the controller IC. When
the controller IC was enabled, the supply would
drop below the minimum supply voltage needed
for the controller to operate. Figure 2 is a screen
shot of the supply voltage for the controller IC
with the TV turned on. The screen shot is a
recording made after the capacitor was replaced.
Before the capacitor was replaced, the volt-
age dropped below 12 V, below the low supply
threshold, shutting the IC down. The Zoom fea-
ture enables the technician to examine the volt-
age levels as the supply is being enabled and
turned on.
4. Verification of power supply input over
voltage shutdown:
On some projection television sets, the power
supply is designed to shut down if the line volt-
age in above a certain level. It is important to
verify that the power supply does shutdown if
the line voltage goes above this set level. Using
the recording function, of the 289 DMM, the test
can easily be done.
The TV connects to the power line through a
Variac (variable auto-transformer). Set the 289
DMM up to record in ac volts at one second
intervals. Turn the TV on with the line voltage
set to approximately 120 V. Connect the 289
DMM across the line and start the recording.
Slowly increase the line voltage until the TV
turns itself off or shuts down. Do not exceed
140 V if the TV does not shut down; there is
something wrong with the shutdown circuit in
the TV.
Figure 1. Left Screen Shot shows an amplifier that is running too hot. Right Screen Shot shows an amplifier running at normal temperature.
Figure 2. Screen shot showing the controller IC supply voltage as the
supply is being turned on. Before the supply is enabled, the voltage is
about 24.75 V and drops to approximately 17 V.
3 Fluke Corporation Troubleshooting electrical problems in high-end TVs
After the TV shuts down or turns off, stop the
recording and save the session. Select Trend
and Zoom to observe when the line voltage
jumps two to three volts within one or two sec-
onds. The voltage at this point is the voltage
level where the TV shuts down. Figure 3 shows
the transition point just as the TV shuts down at
a line voltage of 132 V.
5. Tracing cause of failure of 3-terminal
regulator used as post regulator off of
an off-line switching supply.
In one instance, a three terminal, five volt regu-
lator (TO-220) mounted on a small heat sink
ran much hotter than normal. The solder on the
pins and circuit board had changed color, and
the paper-phonolic board turned a dark brown
around the part pins. The case of the regulator
had cracked, and it was replaced. After the new
regulator was installed, the temperature of the
regulator was checked using the 289 DMM and
a temperature probe. The thermocouple was
connected to the metal tab of the regulator and
289 meter, set to record at 5 second intervals,
recording started.
After about five minutes, the temperature of the
metal tab was about 55 °C. The input voltage
to the regulator was 14.5 V when it should be
around 12 V. The output level of the switch-
ing supply is regulated by sensing the voltage
level at the output of the control/sense circuit,
which generates a voltage proportional to the
Figure 4. Screen shots showing temperature of 3-terminal regulator IC before capacitor replacement, left, and after capacitor replacement, right.
Figure 3. Screen shot showing the voltage level of 136 V where the
TV shuts down.
voltage across a sense winding of the power
transformer. The problem was traced to a faulty
capacitor. Once the capacitor had been replaced,
the input voltage dropped to 12.03 V, and the
temperature of the regulator dropped to 45.2 °C
at five minutes and 46.9 °C at eight minutes.
Screen shots of the temperature recordings are
shown in Figure 4. The left screen is before the
capacitor was replaced, and the right screen is
after the capacitor was replaced.
Fluke Corporation
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
Fluke. Keeping your world
up and running.®
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2008 Fluke Corporation.
Specifications subject to change without notice.
Printed in U.S.A. 12/2008 3392461 A-EN-N Rev A
Modification of this document is not
permitted without written permission
from Fluke Corporation.

Navigation menu